IMPORTANT INFORMATION

Please read this important information before using your monitor.

Please remember that only a medical practitioner is qualified

only when it is stored and used within the temperature and

alcohol, benzene, thinner or other harsh chemicals to clean

monitor or cuff. Remove and replace batteries if monitor is not used for more

office, by those who are eighteen (18) years and older, to monitor

to interpret your blood pressure measurements. Use of this

measurements before prescribing or adjusting medications. Consult your physician if you have any doubt about your

humidity ranges noted on Page E-15. Clean the monitor and cuff with a dry, soft cloth or a cloth

The UA-787 is designed to be used at home or in doctor’s/nurse’s

temperature or humidity or to direct sunlight, shock and

dust. A&D Medical guarantees the accuracy of this device

device should not replace regular medical examinations. Have your physician review the procedure for using this

A&D Medical. Do not attempt to service, calibrate, or repair this device. Because your UA-787 monitor contains delicate,

dampened with water and a mild detergent. Never use

readings. Should a mechanical problem occur, contact

high-precision parts, avoid exposing it to extremes in

device. He or she will want to verify blood pressure

blood pressure (systolic and diastolic) and pulse rate.

than six months.

PRECAUTIONS

Page 3 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — TM
Page 3 · click to enlarge
Page 3
LifeSource Health Line (Toll-Free)1-888-726-9966 E-1
LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) — TM diagram (page 3)
Fig. 2

WHAT DISPLAY SYMBOLS MEAN E-2

MONITOR COMPONENTS E-3

HOW UA-787 WORKS E-4

BEFORE YOU START E-4

SELECTING THE CORRECT CUFF SIZE E-5

TAKING YOUR BLOOD PRESSURE E-6 — E-8

WHAT IS AN IRREGULAR HEARTBEAT E-8

ABOUT MEMORY E-9

USING THE MONITOR WITH AC ADAPTER E-10

ABOUT BLOOD PRESSURE E-11 — E-13

BLOOD PRESSURE RECORD E-14

SPECIFICATIONS E-15

CONTACT INFORMATION E-16

FRENCH INSTRUCTIONS F-1

SPANISH INSTRUCTIONS S-1

In Canada (Toll-Free)1-800-463-5414

?

❤

healthcare products for over 20 years. We are committed to

providing you and your family with monitoring devices that

maintain cardiovascular health and prevent the serious

consequences associated with undetected and untreated

LifeSource blood pressure monitor—one of the most

Designed for ease of use and accuracy, this monitor

Physicians agree that daily self-monitoring of blood

pressure is an important step individuals can take to

technologically advanced monitors available today.

A&D Medical has been manufacturing high quality

TM

will facilitate your daily blood pressure regimen.

provide the utmost in accuracy and convenience.

Congratulations on purchasing a state-of-the-art

hypertension.

Page 4 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — PUL. DISPLAY ERROR
Page 4 · click to enlarge
Page 4
LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) — PUL. DISPLAY ERROR diagram (page 4)
Fig. 3
LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) — PUL. DISPLAY ERROR diagram (page 4)
Fig. 4
Display SymbolCondition/CauseRecommended Action The indicator while measurement is in progress. It blinks while detecting the pulse.Measurement is in progress, remain as still as possible. An irregular heartbeat or body movement may have occurred. Refer to page E-8 for more information on irregular heartbeats.Take measurement again and consult with your physician. Previous measurements stored in MEMORY. Battery FullThe battery power indicator during measurement.Replace all batteries with new ones when the indicator blinks. Battery LowThe battery is low when it blinks. Unstable blood pressure due to excessive body movement.Try the measurement again. Remain very still during the measurement. The systolic and diastolic values are within 10 mmHg of each other.Fasten the cuff correctly, and try the measurement again. The pressure value did not increase during inflation.Check for air leaks along the tube and around the air socket. The cuff is not fastened correctly.Refasten the cuff and retake measurement PUL. DISPLAY ERRORThe pulse is not detected correctly.Try the measurement again. Remain very still during the measurement.

Try the measurement again.

and try the measurement

Check for air leaks along

the tube and around the

Remain very still during

Fasten the cuff correctly,

Refasten the cuff and

the measurement.

measurement.

air socket.

again.

measurement is in progress.

It blinks while detecting the

The battery power indicator

Unstable blood pressure due

values are within 10 mmHg

occurred. Refer to page E-8

The battery is low when it

The pressure value did not

body movement may have

The systolic and diastolic

increase during inflation.

The pulse is not detected

An irregular heartbeat or

for more information on

The cuff is not fastened

during measurement.

irregular heartbeats.

stored in MEMORY.

of each other.

movement.

correctly.

correctly.

blinks.

pulse.

PUL. DISPLAY ERROR

Show 8 more labels / short text on this page
retake measurement · Condition/Cause · Previous measurements · The indicator while · to excessive body · Battery Full · Battery Low · Display
Page 5 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — BEFORE YOU START
Page 5 · click to enlarge
Page 5
LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) — BEFORE YOU START diagram (page 5)
Fig. 5

**The EasyCuff feature is a semi-hard shell cuff which provides easy placement on your upper arm. The UA-787 comes with the EasyCuff feature. If arm circum-

to purchase the proper size cuff to ensure correct measurement. If this is the case,

correct cuff or not. Place the cuff on your arm (see Pg. E-6 to learn

LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) — BEFORE YOU START diagram (page 5)
Fig. 6

Using the correct cuff size is important for an accurate reading. A

cuff that is too large will produce a reading that is lower than the

correct blood pressure; a cuff that is too small will produce a reading

your upper arm at the midpoint between the shoulder and elbow.

ference is greater than 14.2”(36 cm) or smaller than 7.5”(19 cm), you will need

Our cuff has an indicator which tells you whether you are using the

how to put the cuff on correctly) and if the Index Mark L points

within the Proper Fit Range, you are using the correct cuff. If the

Index points outside of the Proper Fit Range, contact A&D Medical

hanging at the side of your body, measure the circumference of

that is higher than the correct blood pressure. With your arm

please call A&D Medical at 1-888-726-9966 (in Canada - 1-800-463-5414).

at 1-888-726-9966 (in Canada - 1-800-463-5414) for a cuff

LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) — BEFORE YOU START diagram (page 5)
Fig. 7

Cuff Size Indicator On The Cuff

ARM SIZERECOMMENDED CUFF SIZEREPLACEMENT CUFF MODEL# 7.5” - 14.2” (19-36cm)EasyCuff ™Feature**UA-380 5.1” - 7.9” (13-20cm)Small CuffUA-279 7.5” - 12.2” (19-31cm)Medium CuffUA-280 11.8” - 17.7” (30-45cm)Large CuffUA-281

and attach the cuff to the monitor before using. To install batteries

(or replace them if the “Low Battery” symbol appears on display),

LifeSource TM blood pressure monitors are easy to use, accurate and

examines the pulsatile pressure generated by the arterial wall as it

The cuff is inflated until the artery is fully blocked. The inflation

You must install 4 type “AA” (1.5 volt) batteries, or use the AC

1. Remove battery compartment cover by gently pushing down on arrow and sliding cover down.

measure of vibrations caused by the arterial pulse. Our monitor

device. The monitor takes measurements while the cuff is both

adapter (see Pg. E-10 for Using the Monitor with AC Adapter),

digitally display full measurement readouts. Our technology is

speed is maximized and the pressure level is optimized by the

3. Replace cover by sliding it into the compartment and gently pressing down.

based on the “oscillometric method”—a non-invasive blood

pressure determination. The term “oscillation” refers to the

expands and contracts against the cuff with each heartbeat.

inflating and deflating. This results in a faster measurement

BEFORE YOU START

HOW UA-787 WORKS ?

positive (+) and negative (-) terminals

compartment. Make sure the batteries

2. Put in top row of batteries first. Place batteries in the compartment with

make contact with the compartment

providing greater comfort to the user.

matching those indicated in the

proceed as follows:

terminals.

Step 2 · Step 1 · Step 3

Page 6 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — TAKING YOUR BLOOD PRESSURE
Page 6 · click to enlarge
Page 6
LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) — DIASTOLIC diagram (page 6)
Fig. 8
5. All display symbols appear briefly, then the display changes as the measurement begins. The cuff starts to inflate to the
6. When inflation is complete, deflation starts automatically and the ❤ blinks, indicating that the measurement is in progress. Once

The cuff deflates and the monitor automatically shuts off after 45

7. When the measurement is complete, the systolic and diastolic pressure readings and pulse rate are displayed.

seconds, or you can turn it off by pressing the START button.

8. Remove cuff and make note of your blood pressure and pulse rate on the Blood Pressure Record chart (see Pg. E-14),

the pulse is detected, the mark flashes with each pulse beat.

correct level. It is normal for the cuff to feel very tight.

indicating the date and time of measurement.

DIASTOLIC

SYSTOLIC

PULSE

automatically starts to inflate again.

button again.

Unless your physician recommends otherwise, use left arm

Remove constricting clothing and place cuff on bare arm.

Relax for about 5 to 10 minutes before measurement.

TAKING YOUR BLOOD PRESSURE

Now you are ready. Follow these simple steps:

center of your upper arm is at the same

2. Lay left arm on the table with palm up. Place the semi-hard shell cuff on upper

Fasten cuff securely with the hook and

fingers between the cuff and your arm.

3. The cuff should be snug but not too tight. You should be able to insert two
4. Press the START button. Remain still and do not talk.

arm with the tube facing downward

and toward the inside of your arm.

1. Sit comfortably with your left arm resting on a flat surface so that the

to measure pressure.

height as your heart.

loop fastener strap.

Hose · Cuff

Page 7 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — TAKING YOUR BLOOD PRESSURE
Page 7 · click to enlarge
Page 7

memory are assigned an index number in the order of the newest

(e.g. n20) of the newest reading followed by the measurement.

2. If you want to retrieve other measurements in memory, press and quickly release the Memory (“M”) button repeatedly until

This device automatically stores up to thirty (30) blood pressure

the Memory (“M”) button for at least 3 seconds. You will see the

number “n01” followed by a measurement of three 0's displayed

To clear the measurement history from memory, press and hold

If there are no measurements in memory, you will see the index

1. When the display screen is blank, press and quickly release the Memory (“M”) button. You will see the index number

M symbol appear in the lower left corner of the display. At the

and pulse measurements in memory. Measurements stored in

The UA-787 automatically turns off after several seconds.

end of the process, the M symbol will flash and disappear.

ABOUT MEMORY

the desired index number is reached (e.g. n14). The

to the oldest. The oldest reading displays as “n01” .

To retrieve readings, follow these simple steps:

measurement will then be displayed.

are removed.

vertically.

occurs. An irregular heartbeat is defined as a heartbeat that varies

pressure measurement. It is important that you be relaxed, remain

If your systolic pressure is expected to exceed 230 mmHg follow

1. Place the cuff on the left arm, unless your physician tells you otherwise.
2. Press and hold the START button until a number 30 to 40 mmHg higher than your expected systolic pressure appears.
3. Release the START button when the desired number is reached. Measurement will then begin. Continue to measure

The UA-787 blood pressure monitor provides a blood pressure

and pulse rate measurement even when an irregular heartbeat

by 25% from the average of all heartbeats during the blood

TAKING YOUR BLOOD PRESSURE

WHAT IS AN IRREGULAR HEARTBEAT

Measurement with the desired systolic pressure:

your blood pressure as described on Page E-6.

still and do not talk during measurements.

indicator frequently.

these steps:

Page 8 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — USING THE MONITOR WITH AC ADAPTER
Page 8 · click to enlarge
Page 8

systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg.

❤

diastolic pressure occurs when the heart expands. Blood pressure

your arm to allow blood to flow back to the heart, can also affect it.

An individual’s blood pressure varies greatly from day to day and

are even more pronounced. Normally, blood pressure rises during

Fluctuation within a day (case: 35 year old male)

Blood pressure is the force exerted by blood against the walls of

mmHg. Taking your blood pressure repeatedly without waiting an

In addition to these factors, diet beverages containing caffeine or

season to season. For hypersensitive individuals, these variations

the arteries. Systolic pressure occurs when the heart contracts;

day, activity level, climate, altitude and season. Certain activities

interval of 5 to 10 minutes between readings, or without raising

Blood pressure is affected by many factors: age, weight, time of

Sleeping can decrease systolic blood pressure by as much as 10

alcohol, certain medications, emotional stress and even tight-

can significantly alter one’s blood pressure. Walking can raise

ABOUT BLOOD PRESSURE

What Causes Variations In Blood Pressure?

work or play and falls to its lowest levels during sleep.

fitting clothes can make a difference in the readings.

is measured in millimeters of mercury (mmHg).

What Affects Blood Pressure?

What Is Blood Pressure?

200 150 100 50 0

The UA-787 has an AC adapter jack to allow you to supply power

1. Connect the AC adapter plug into the jack on the right side of the monitor.

exclusive AC adapter (Model TB:181) to avoid potential damage

(in Canada - 1-800-463-5414) if you would like to purchase the

USING THE MONITOR WITH AC ADAPTER

from an outlet in your home. We recommend you use only the

to the monitor. Please contact A&D Medical at 1-888-726-9966

2. Gently insert the AC adapter plug into a 120V AC outlet.
2. Then disconnect the plug from the jack of the monitor.

Disconnecting the AC adapter from the monitor:

1. Gently remove the AC adapter from the outlet.

Connecting the AC adapter to the monitor:

TB:181 AC Adapter Specification:

OutputDC 5V 500mA
Polaritycenter positive
InputAC 120V 60Hz

AC adapter.

Show 15 more labels / short text on this page
Systolic Blood · Diastolic Blood · Pressure · Pressure · Sleep · Work · Leave · Argument · Telephone · Work · Start · Arising · AC Adapter · Jack · Plug
Page 9 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — ABOUT BLOOD PRESSURE ❤
Page 9 · click to enlarge
Page 9

ABOUT BLOOD PRESSURE ❤

history. In addition, clinical studies have shown that the detection

One way to determine whether this is the case for you is to take

In any case, self-measurement at home supplements the doctor’s

readings and provides a more accurate, complete blood pressure

pressure written first, followed by a slash mark and the diastolic

for example. Please see Page E-14 for the blood pressure record.

your home monitor to the doctor’s office and, before the doctor

and treatment of hypertension is improved when patients both

Blood pressure readings are typically recorded with the systolic

pressure readings taken in a doctor’s office or hospital setting

This response is commonly called “white coat hypertension.”

diastolic measurements are written as 120/80. Pulse is simply

written with the letter “P” followed by the pulse rate—P 72,

might be elevated as a result of apprehension and anxiety.

or nurse takes your pressure, do it yourself on your home

pressure. For example, 120 mmHg systolic and 80 mmHg

It is now well known that, for many individuals, blood

consult their physicians and monitor their own blood

Why Measure Blood Pressure at Home?

How Do I Record My Blood Pressure?

monitor and compare it to your home readings.

pressure at home.

when readings consistently rise above normal. It is well known that

Don’t smoke Have regular physical checkups

hypertension can lead to stroke, heart attack or other illness if left

In many individuals, hypertension can be controlled by altering

prescribed and monitored by your doctor. The American Heart

(without regard to age) have been established by the National

not always produce symptoms that alert you to the problem,

lifestyle and minimizing stress, and by appropriate medication

Association recommends the following lifestyle suggestions to

Hypertension (high blood pressure) is the diagnosis given

The following standards for assessing high blood pressure

Maintain proper weight Monitor your blood

Category Systolic (mmHg) Diastolic (mmHg)

untreated. Referred to as a “silent killer” because it does

Reduce salt and fat intake Exercise routinely

Stage 1 140 – 159 90 – 99 Stage 2 160 – 179 100 – 109 Stage 3 ≥180 ≥110

High Normal 130 - 139 85 – 89 Hypertension

hypertension is treatable when diagnosed early.

Can Hypertension Be Controlled?

ABOUT BLOOD PRESSURE ❤

Assessing High Blood Pressure

prevent or control hypertension:

Institutes of Health JNCVI.

Show 4 more labels / short text on this page
pressure regularly · Optimal <120 <80 · Normal <130 <85 · What Is Hypertension?
Page 10 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — TM
Page 10 · click to enlarge
Page 10
LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) — DATE AM SYS/DIA PULSE PM SYS/DIA PULSE diagram (page 10)
Fig. 9

Measurement range Pressure: 20 mmHg to 280 mmHg

batteries (not included) or 120 V

Less than 85% relative humidity

Less than 95% relative humidity

Weight 12.3 oz. (350 g) without batteries

Blood pressure measurements determined by the UA-787 are equivalent to those

obtained by a trained observer using the cuff/stethoscope auscultation method

Display Digital, 20-mm character height

Power source 4 type “AA” (1.5 volt) alkaline

Pulse40 pulses to 200 pulses

Depressurization Active exhaust valve (ECEVII)

Battery life Approximately 4 months with

Operating environment 50ºF to 104ºF (10ºC to 40ºC)

Storage environment -4ºF to 140ºF (-20ºC to 60ºC)

within the limits prescribed by the American National Standards Institute for

Pressurization Automatic, using micropump

AC Adapter (model TB:181)

Accuracy Pressure: +3 mmHg or 2%,

Dimensions Length: 4.4”(112 mm)

Width6.4 ” (163 mm)
Height2.4 ” (62 mm)

Type Oscillometric

Memory 30 readings

Pulse+5%

Model UA-787

electronic or automated sphygmomanometers.

In Canada (Toll-Free)1-800-463-5414 LifeSource Health Line (Toll-Free): 1-888-726-9966

DATE AM SYS/DIA PULSE PM SYS/DIA PULSE

1/14 9:30 132/98 P 69 6:30 128/87 P 63

B L O O D P R E S S U R E R E C O R D

E X A M P L E

NameAge: Weight:

TM

Show 4 more labels / short text on this page
Pressure/pulse displayed · one daily measurement · whichever is greater · simultaneously
Page 11 of the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) user manual — SPECIFICATIONS CONTACT INFORMATION
Page 11 · click to enlarge
Page 11

scientific and technological methods, and offer accurate, easy to

This LifeSource blood pressure product is covered by a Lifetime

use, home monitoring and treatment options. Our full product

LifeSource Health Line (Toll-Free)1-888-726-9966

Our products are designed and manufactured using the latest

For more information regarding use, care or servicing of your

Visit our web site at www.LifeSourceOnline.com

LifeSource Health Line (Toll-Free)1-888-726-9966

for warranty registration.

SPECIFICATIONS CONTACT INFORMATION

Warranty. See warranty card for details.

AutoControl (Toll-Free)1-800-463-5414

Digital Blood Pressure Monitors

For Canada Residents, please contact:

A division of A&D Engineering, Inc.

Aneroid Blood Pressure Kits

blood pressure monitor, contact:

6695 Millcreek Drive, Unit 5 Mississauga, Ontario

www.LifeSourceOnline.com

1555 McCandless Drive Milpitas, CA 95035

www.autocontrol.com

line includes:

Show 8 more labels / short text on this page
Personal Health Scales · Digital Thermometers · Blood Pressure Cuffs · Heart Rate Monitors · Auto Control Medical · Stethoscopes · L5N 5R8 Canada · A&D Medical
Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787)

Be the first to ask — other owners or our team will answer.

Original document

LifeSource One-Step Auto-Inflation Blood Pressure Monitor (UA-787) manual PDF

11 pages·352 KB·Blood Pressure Monitor·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…