Congratulations. You have purchased a LifeSource state-of-the-art blood pressure monitor, one of the most technologically advanced yet easy to use products available in the marketplace today. This LifeSource monitor is designed to make your daily regimen useful, and convenient.

Physicians agree that daily self-monitoring of blood pressure is an important way individuals can contribute to maintaining their cardiovascular health and preventing the serious consequences of hypertension left undiagnosed and untreated.

LifeSource has been manufacturing quality healthcare products for more than 20 years. Rest assured that we are committed to providing you and your family with monitoring devices specifically designed for high accuracy and ease of use. LifeSource—your source for a lifetime of health.

Page 2 of the LifeSource UA-702 user manual — TM
Page 2 · click to enlarge
Page 2
Page 2 of the LifeSource UA-702 user manual — TM
Page 3 of the LifeSource UA-702 user manual — CUFF HOSE
Page 3 · click to enlarge
Page 3

WHAT DISPLAY SYMBOLS MEAN ?

❤

Appears during measure- ment; blinks and beeps when pulse is detected.

Remain still and refrain from talking; measurement in progress.

MEASURING

Appears on display prior to and during measurement. Blinks if there is air in cuff at power up.

Press exhaust button to release air.

ERROR

Measurement begun, but pressure is not sufficient for reading.

Increase pressurization by 30 mmHg to 40 mmHg and repeat measurement.

RE-MEASURE

Battery voltage is too low for monitor to work properly.

Replace all batteries with new ones.

LOW BATTERY

Appears when blood pres- sure cannot be measured accurately due to erratic pulse or noise interference.

Exhaust air from cuff, reap- ply cuff properly, power up again and repeat measure- ment in quiet environment.

ERROR

Appears when pulse cannot be measured accurately.

Exhaust air from cuff, reapply cuff properly, power up again and repeat measurement. Adjust exhaust rate to 2 mmHg to 5 mmHg.

PULSE ERROR

Occurs when exhaust velocity rate is more than 8 mmHg.

EXHAUST SPEED FLASHING

Check for correct air con- nector. Make sure air con- nector is inserted correctly into the hoses and monitor.

CUFF HOSE

BULB HOSE

MONITOR

Show 4 more labels / short text on this page
Display Symbol · Condition/ Cause · Recommended Action · No Display Symbol
Page 4 of the LifeSource UA-702 user manual — WHAT DISPLAY SYMBOLS MEAN ?
Page 4 · click to enlarge
Page 4
LifeSource UA-702 — WHAT DISPLAY SYMBOLS MEAN ? diagram (page 4)
Fig. 1
Page 5 of the LifeSource UA-702 user manual — BEFORE YOU START
Page 5 · click to enlarge
Page 5

HOW 702 WORKS ?

LifeSource Blood Pressure Monitors are easy to use, accurate and digitally display full measurement readouts. Our technology is based on the “oscillometric method” — a noninvasive blood pressure determination. The term “oscillation” refers to any measure of vibrations caused by the arterial pulse. The cuff is first inflated until the artery is fully occluded. Then, the monitor takes measurements while the cuff deflates. Our monitors exam- ine the pulsatile pressure generated by the arterial wall as it expands and contracts against the cuff with each heartbeat.

BEFORE YOU START

You must install 4 type AA (1.5 volt) batteries and attach the cuff to the monitor before using it. To install batteries (or replace them if the “Low Battery” symbol appears on display), proceed as follows:

1. Remove battery compartment cover by gently pushing down on arrow and sliding cover forward.
2. Put in top row of batteries first. Place the batteries in compartment with positive (+) and negative (–) terminals matching those indicated in the compartment. Be sure batteries make contact with compartment terminals.
3. Replace cover by sliding it into the compartment and gently pressing into place.
Page 6 of the LifeSource UA-702 user manual — CUFF HOSE
Page 6 · click to enlarge
Page 6

SELECTING THE CORRECT CUFF

Using the correct cuff size is important for an accurate reading. A cuff that is too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small yields a measurement that is higher. With your arm hanging at the side of your body, measure the circumference of your upper arm at the midpoint between shoulder and elbow.

ARM SIZERECOMMENDED CUFF SIZEREPLACEMENT CUFF MODEL# 5.1” - 7.9” (13-20cm)SmallUA-279 7.5” - 12.2” (19-31cm)MediumUA-280 11.8” - 17.7” (30-45cm)LargeUA-281

CUFF HOSE

BULB HOSE

MONITOR

AIR SOCKET

PRESSURE AND PULSE DISPLAY

SYSTOLIC DIASTOLIC (PULSE)

UA-702

DIGITAL BLOOD PRESSURE MONITOR

MEASURING PRESS EXHAUST RE-MEASURE BATTERY LOW RE-MEASURE

POWER

POWER BUTTON

The air connector is attached to the cuff and inflator bulb tubing. Attach this into the back of the monitor. When changing the cuff size, the exhaust velocity may need to be adjusted (see page 11).

Page 7 of the LifeSource UA-702 user manual — TAKING YOUR BLOOD PRESSURE
Page 7 · click to enlarge
Page 7

TAKING YOUR BLOOD PRESSURE

☞Relax for about 5 to 10 minutes before measurement. ☞Remove constricting clothing and place cuff on bare arm. ☞Unless your physician recommends otherwise, use left arm to measure pressure.

Now you are ready. Follow these simple steps:

Now you are ready. Follow these simple steps:

1. Sit comfortably with your left arm resting on a flat surface so that the center of your upper arm is at the same height as your heart.
1. Sit comfortably with your left arm resting on a flat surface so that the center of your upper arm is at the same height as your heart.
LifeSource UA-702 — TAKING YOUR BLOOD PRESSURE diagram (page 7)
Fig. 2
LifeSource UA-702 — TAKING YOUR BLOOD PRESSURE diagram (page 7)
Fig. 3
2. Lay left arm on the table, palm up and thread cuff end through metal loop, smooth side against arm. Then position the tube off-center toward the inner side of arm in line with the little finger.
2. Lay left arm on the table, palm up and thread cuff end through metal loop, smooth side against arm. Then position the tube offcenter toward the inner side of arm in line with the little finger.
LifeSource UA-702 — TAKING YOUR BLOOD PRESSURE diagram (page 7)
Fig. 4
LifeSource UA-702 — TAKING YOUR BLOOD PRESSURE diagram (page 7)
Fig. 5
LifeSource UA-702 — TAKING YOUR BLOOD PRESSURE diagram (page 7)
Fig. 6
LifeSource UA-702 — TAKING YOUR BLOOD PRESSURE diagram (page 7)
Fig. 7
3. Pull the end of the cuff to tighten it, fold back the extra material, and fasten the Velcro. The cuff should be snug but not too tight. You should be able to insert two fingers between the cuff and your arm.
3. Pull the end of the cuff to tighten it, fold back the extra material, and fasten securely. The cuff should be snug but not too tight. You should be able to insert two fingers between the cuff and your arm.
LifeSource UA-702 — TAKING YOUR BLOOD PRESSURE diagram (page 7)
Fig. 8
LifeSource UA-702 — TAKING YOUR BLOOD PRESSURE diagram (page 7)
Fig. 9

☞IMPORTANT: Measure pressure at the same time each day.

Page 8 of the LifeSource UA-702 user manual — TAKING YOUR BLOOD PRESSURE
Page 8 · click to enlarge
Page 8

TAKING YOUR BLOOD PRESSURE

☞IMPORTANT: Measure pressure at the same time each day.

4. Press the POWER button.
5. Watch for all display symbols to appear briefly, followed by a zero, indicating that monitor is ready for measurement.

6. Watch the display and squeeze the bulb quickly until cuff pressure reaches about 30 mmHg to 40 mmHg above expected systolic pressure. To help you determine the correct inflation level, the monitor is programmed to beep quickly three times at each of the four preset pressure levels (160, 200, 240, and 280). We recommend using the chart shown below to determine which preset measurement to use. If the “re-measure“ symbol appears, pressure is not sufficient for reading and measurement must be taken again.

Usual SystolicInflate to Up to 130160 131-170 171-210200 240 211-250280

Watch the display for the correct inflation level. Be sure to completely deflate the cuff before use.

Page 9 of the LifeSource UA-702 user manual — TAKING YOUR BLOOD PRESSURE
Page 9 · click to enlarge
Page 9

TAKING YOUR BLOOD PRESSURE

7. Sit quietly during measurement. Pressure values are displayed at left, then right, along with ❤ symbol blinks and beeps with pulse beat.
8. Wait for the long beep, indicating that the measurement is complete. Your systolic pressure will appear on the left and your diastolic pressure on the right of the display screen, alternating with pulse reading.
9. Press the exhaust valve to release the air from the cuff. The monitor shuts off automatically in about two minutes, or press POWER to shut off.
Page 10 of the LifeSource UA-702 user manual — TAKING YOUR BLOOD PRESSURE
Page 10 · click to enlarge
Page 10

TAKING YOUR BLOOD PRESSURE

10. Remove cuff and make a note of your blood pressure and pulse rate on the chart (see page 16), indicating date and time of measurement.

We advise that you record the date and time after each measure- ment because an accurate blood pressure history relies not on single or sporadic readings but on a pattern over time.

Page 11 of the LifeSource UA-702 user manual — INFLATOR
Page 11 · click to enlarge
Page 11

CHECK/ADJUST EXHAUST VELOCITY

Exhaust velocity is the rate at which the air pressure in the cuff deflates during the course of measurement. It is preset at the factory and generally does not need any further adjustment.

To check the exhaust velocity rate, proceed as follows:

1. Put on the arm cuff.
2. Press the POWER button. The unit will begin taking a reading.
3. When zero appears, inflate the cuff by squeezing the handheld inflator bulb. Pump it until it shows a pressure level 40 mmHg higher than your usual systolic measurement.

INFLATOR

BULB TUBING

EXHAUST

VALVE

INFLATOR

BULB

Page 12 of the LifeSource UA-702 user manual — CHECK/ADJUST EXHAUST VELOCITY
Page 12 · click to enlarge
Page 12
LifeSource UA-702 — CHECK/ADJUST EXHAUST VELOCITY diagram (page 12)
Fig. 10

CHECK/ADJUST EXHAUST VELOCITY

4. Watch the display screen. As pressure automatically decreases, diastolic numbers should decline in increments of 2 to 5 mmHg.

If the exhaust velocity range is not within 2 to 5 mmHg, make the following adjustment with the unit turned off:

1. The adjustment screw is located on the exhaust valve attached to the inflator bulb. It is covered by a thin grey protection cap with a hole in the center. To remove and discard the cap (you do not need to replace it), insert the end of a paper clip and lift off to see the the gold adjustment screw underneath.
2. Using a small screwdriver, rotate the adjustment screw about 45 degrees clockwise (to raise the rate) or counterclockwise (to lower the rate).

Turn screw very gradually.

3. Repeat the test until the exhaust velocity is between 2 to 5 mmHg.
Page 13 of the LifeSource UA-702 user manual — Fluctuation within a day (case: 35 year old male)
Page 13 · click to enlarge
Page 13

ABOUT BLOOD PRESSURE ❤

What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).

What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).

What Affects Blood Pressure? Blood pressure is affected by many factors: age, weight, time of day, activity level, climate, altitude and season. Certain activities can significantly alter blood pressure. Walking can raise systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg. Sleeping can decrease systolic blood pressure by as much as 10 mmHg. Taking your blood pressure repeatedly without waiting an interval of 5 to 10 minutes between readings, or without raising your arm to allow blood to flow back to the heart, can also affect it.

What Affects Blood Pressure? Blood pressure is affected by many factors: age, weight, time of day, activity level, climate, altitude and season. Certain activities can significantly alter blood pressure. Walking can raise systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg. Sleeping can decrease systolic blood pressure by as much as 10 mmHg. Taking your blood pressure repeatedly without waiting an interval of at least 3 minutes between readings can also affect it.

In addition to these factors, beverages containing caffeine or alcohol, certain medications, emotional stress and even tight-fitting clothes can make a difference in the readings.

In addition to these factors, diet beverages containing caffeine or alcohol, certain medications, emotional stress and even tight- fitting clothes can make a difference in the readings.

What Causes Variations In Blood Pressure? An individual’s blood pressure varies greatly from day to day and season to season. In hypersensitive individuals, these variations are even more pronounced. Normally, blood pres- sure rises during work or play and falls to its lowest levels during sleep.

What Causes Variations In Blood Pressure? An individual’s blood pressure varies greatly from day to day and season to season. In hypersensitive individuals, these variations are even more pronounced. Normally, blood pressure rises during work or play and falls to its lowest levels during sleep.

Fluctuation within a day (case: 35 year old male)

200 150 100 50 0

Systolic Blood Pressure Diastolic Blood Pressure

Show 8 more labels / short text on this page
Arising · Telephone · Argument · Sleep · Start · Work · Leave · Work
Page 14 of the LifeSource UA-702 user manual — ABOUT BLOOD PRESSURE ❤
Page 14 · click to enlarge
Page 14

ABOUT BLOOD PRESSURE ❤

Assessing High Blood Pressure The following standards for assessing high blood pressure (without regard to age) have been established by the National Institutes of Health JNCVI.

Assessing High Blood Pressure The following standards for assessing high blood pressure (without regard to age) have been established as a guideline.

BP Classification Systolic (mmHg) Diastolic (mmHg) Normal

Category Systolic (mmHg) Diastolic (mmHg) Optimal

Stage 1 140 – 159 90 – 99 Stage 2 160 – 179 100 – 109 Stage 3 ≥180 ≥110

SOURCE: The Seventh Report of the Joint National Committee on Prevention, Evaluation and Treatment of High Blood Pressure. National Heart, Lung and Blood Institute - May 2003.

What Is Hypertension? Hypertension (high blood pressure) is the diagnosis given when readings consistently rise above normal. It is well known that hypertension can lead to stroke, heart attack or other illness if left untreated. Referred to as a “silent killer” because it does not always produce symptoms that alert you to the problem, hypertension is treatable when diagnosed early.

What Is Hypertension? Hypertension (high blood pressure) is the diagnosis given when readings consistently rise above normal. It is well known that hypertension can lead to stroke, heart attack or other illness if left untreated. Referred to as a “silent killer” because it does not always produce symptoms that alert you to the problem, hypertension is treatable when diagnosed early. Those who fall within the Prehypertensive range are more likely to develop hypertension in the future.

Can Hypertension Be Controlled? In many individuals, hypertension can be controlled by altering lifestyle and minimizing stress, and by appropriate medication prescribed and monitored by your doctor. The American Heart Association recommends the following lifestyle suggestions to prevent or control hypertension:

Can Hypertension Be Controlled? In many individuals, hypertension can be controlled by altering lifestyle and minimizing stress, and by appropriate medication prescribed and monitored by your doctor. The American Heart Association recommends the following lifestyle suggestions to prevent or control hypertension:

Don’t smoke. Have regular physical Reduce salt and fat intake. checkups. Maintain proper weight. Monitor your blood Exercise routinely. pressure at periodic intervals.

Don’t smoke. Have regular physical Reduce salt and fat intake. checkups. Maintain proper weight. Monitor your blood Exercise routinely. pressure at periodic intervals. In Canada (Toll Free): 1 800 463 5414

Page 15 of the LifeSource UA-702 user manual — ABOUT BLOOD PRESSURE ❤
Page 15 · click to enlarge
Page 15

ABOUT BLOOD PRESSURE ❤

Why Measure Blood Pressure at Home? It is now well known that, for many individuals, blood pressure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. This response is commonly called “white coat hypertension.” One way to determine whether this is the case for you is to take your home monitor to the doctor’s office and, before the doctor or nurse takes your pressure, do it yourself on your home monitor and compare it to your record of home readings.

Why Measure Blood Pressure at Home? It is now well known that, for many individuals, blood pres- sure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. This response is commonly called “white coat hyperten- sion.” One way to determine whether this is the case for you is to take your home monitor to the doctor’s office and, before the doctor or nurse takes your pressure, do it your- self on your home monitor and compare it to your record of home readings. In any case, self-measurement at home supplements your doctor’s readings and provides a more accurate, complete blood pressure history. In addition, clinical studies have shown that the detection and treatment of hypertension is improved when patients both consult their physicians and monitor their own blood pressure at home.

In any case, self-measurement at home supplements the doctor’s readings and provides a more accurate, complete blood pressure history. In addition, clinical studies have shown that the detection and treatment of hypertension is improved when patients both consult their physicians and monitor their own blood pressure at home.

Keys to Successful Monitoring: Blood pressure fluctuates throughout the day. We recom- mend that you are consistent in your daily measurement routine:

How Do I Record My Blood Pressure? Blood pressure readings are typically recorded with the systolic pressure written first, followed by a slash mark and the diastolic pressure. For example, 120 mmHg systolic and 80 mmHg diastolic measurements are written as 120/80. Pulse is simply written with the letter “P” followed by the pulse rate—P 72, for example.

  • Measure at the same time every day.
  • Sit in the same chair/position.
  • Relax for 5 minutes before measurement.
  • Sit still during measurement – no talking, eating or sudden movements.
  • Record your measurement in a logbook.

How Do I Record My Blood Pressure? Blood pressure readings are typically recorded with the sys- tolic pressure written first, followed by a slash mark and the diastolic pressure. For example, 120 mmHg systolic and 80 mmHg diastolic measurements are written as 120/80. Pulse is simply written with the letter “P” followed by the pulse rate—P 72, for example. Please see Page 16 for the blood pressure record.

Page 16 of the LifeSource UA-702 user manual — 1/14 9:30 132/98 P 69 6:30 128/87 P 63
Page 16 · click to enlarge
Page 16
LifeSource UA-702 — 1/14 9:30 132/98 P 69 6:30 128/87 P 63 diagram (page 16)
Fig. 11

B L O O D P R E S S U R E R E C O R D

1/14 9:30 132/98 P 69 6:30 128/87 P 63

Page 17 of the LifeSource UA-702 user manual — IMPORTANT INFORMATION ✓
Page 17 · click to enlarge
Page 17

IMPORTANT INFORMATION ✓

Please read this important information before using your monitor.

☞Please remember that only a medical practitioner is qualified to interpret your blood pressure measurements. Use of this device should not replace regular medical examinations. ☞Have your physician review your procedure for using this device. He or she will want to verify blood pressure measurements before prescribing or adjusting medications. ☞Consult your physician if you have any doubt about your readings. Should a mechanical problem occur, contact LifeSource. ☞Do not attempt to service, calibrate, or repair this device. ☞Because your UA-702 monitor contains delicate, high-precision parts, avoid exposing it to extremes in temperature or humidity or to direct sunlight, shock and dust. LifeSource guarantees the accuracy of this device only when it is stored and used within the temperature and humidity ranges noted on page 18. ☞Clean the monitor and cuff with a dry, soft cloth or a cloth dampened with water and a mild detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean monitor or cuff. ☞Remove and replace batteries if monitor is not used for more than six months.

PRECAUTIONS UA-702 is designed to be used at home, by those who are eighteen (18) years and older, to monitor blood pressure (systolic and diastolic) and pulse rate. This monitor is not designed to measure the blood pressure of people with common arrhythmias, such as atrial or ventricular premature beats or atrial fibrillation. It is not designed for ambulatory use.

Page 18 of the LifeSource UA-702 user manual — SPECIFICATIONS SPECIFICATIONS ✓
Page 18 · click to enlarge
Page 18

SPECIFICATIONS SPECIFICATIONS ✓

Model UA-702 Type Oscillometric Display Digital, 10-mm character height

Pressure/pulse displayed alternately Measurement range Pressure: 20 mmHg to 280 mmHg

Pulse: 40 pulses to 200 pulses/minute Accuracy Pressure: +3 mmHg or 2%,

whichever is greater Pulse: +5% Pressurization Manual by inflator bulb Depressurization Constant-air release-valve system Deflation Hand-held push button Power source 4 type AA (1.5 volt) alkaline

batteries (not included) Battery life Approximately 6 months with

3 daily measurements Operating environment 50ºF to 104ºF (10ºC to 40ºC)

Less than 85% relative humidity Storage environment -4ºF to 140ºF (-20ºC to 60ºC)

Less than 95% relative humidity Dimensions Length: 4.1”(105 mm)

Width: 5.3”(135 mm) Height: 1.7” (44 mm) Weight 4.9 oz. (140 g) without batteries

Blood pressure measurements determined by the UA-702 are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute for electronic or automated sphygmomanometers.

Page 19 of the LifeSource UA-702 user manual — CONTACT INFORMATION
Page 19 · click to enlarge
Page 19

CONTACT INFORMATION

Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. The product line includes:

Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. Our full product line includes:

Aneroid Blood Pressure Kits Blood Pressure Cuffs Digital Blood Pressure Monitors Digital Thermometers Heart Rate Monitors Nebulizers Personal Health Scales Stethoscopes

Aneroid Blood Pressure Kits Blood Pressure Cuffs Digital Blood Pressure Monitors Digital Thermometers Stethoscopes

This LifeSource product is covered by a Lifetime Warranty. See warranty card for details.

This LifeSource blood pressure product is covered by a Lifetime Warranty. See warranty card for details.

Visit our web site at www.LifeSourceOnline.com

for warranty registration.

For more information regarding use, care or servicing of your Blood Pressure Monitor, contact:

For more information regarding use, care or servicing of your blood pressure monitor, contact:

LifeSource A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 LifeSource Health Line (Toll-Free): 1-888-726-9966 www.LifeSourceOnline.com

A&D Medical A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 LifeSource Health Line (Toll-Free): 1-888-726-9966 www.LifeSourceOnline.com For Canada Residents, please contact: Auto Control Medical 6695 Millcreek Drive, Unit 5 Mississauga, Ontario L5N 5R8 Canada Auto Control (Toll-Free): 1-800-463-5414 www.autocontrol.com

For Canada Residents, please contactAuto Control Medical 206A boul. Brunswick Pointe-Claire, Quebec H9R 5P9 Canada
Page 20 of the LifeSource UA-702 user manual — LifeSource Health Line
Page 20 · click to enlarge
Page 20

CONTACT INFORMATION

IMPORTANT!

If You Need Assistance with the Set-Up or Operation

We Can Help!

Please call us FIRST before contacting your retailer at

LifeSource Health Line

1-888-726-9966 - Toll Free

1-800-463-5414 - In Canada

A specially trained representative will assist you

A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 www.LifeSourceOnline.com

LifeSource is a trademark of A&D Medical. ©2003 A&D Medical. All rights reserved. Specifications subject to change without notice.

I-MAN:702 Printed in China 12/03

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the LifeSource UA-702

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the LifeSource UA-702

Be the first to ask — other owners or our team will answer.

Original document

LifeSource UA-702 manual PDF

20 pages·317 KB·Blood Pressure Monitor·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…