Appears when pressure value did not increase during cuff inflation: 1) cuff is not fastened 2) air connector is not connected properly.
Appears if pressure value is unstable due to movement during reading.
Appears when systolic and diastolic measure- ments are within 10 mmHg of each other.
Battery voltage is too low for monitor to work properly.
Appears during measure- ment; blinks and beeps when pulse is detected.
Make sure cuff tube is properly connected to monitor, cuff is securely fastened and exhaust rate is between 2 mmHg and 5 mmHg. Make sure air connector is inserted properly into tube and monitor.
Check for air leakage; make sure cuff tube is properly connected to monitor and exhaust rate is in increments of 2 mmHg. to 5 mmHg
Remain still and refrain from talking; measure- ment in progress.


BLACK PROTECTION CAP ON ADJUSTMENT SCREW
SYSTOLIC DISPLAY
START
240 210 180 150 PRESSURE SWITCH
START BUTTON
DIASTOLIC DISPLAY
TUBE
?

You must install 4 type AA (1.5 volt) batteries and attach the cuff to the monitor before using it. To install batteries (or replace them if the “Low Battery” symbol appears on display), proceed as follows:
To attach cuff, insert air connector at end of cuff tube into air socket on left side of monitor. Should you need to change the cuff size, the preset exhaust velocity rate may have to be adjusted (see page 11).
Using the correct cuff size is important for an accurate reading. A cuff that is too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small yields a measurement that is higher. With your arm hanging at the side of your body, measure the circumference of your upper arm at the midpoint between shoulder and elbow.
11.8” - 17.7” (30-45cm)
7.5” - 12.2” (19-31cm)
5.1” - 7.9” (13-20cm)
SELECTING THE CORRECT CUFF
AIR CONNECTOR
AIR CONNECTOR
MONITOR
REPLACEMENT CUFF MODEL#
UA-281
UA-280
UA-279
Large · Medium · Small

Now you are ready. Follow these simple steps:
4. Before taking your blood pressure, you will need to set the pressure switch to a number that is at least 30 mmHg higher than your expected systolic pressure. Use the chart below as a guide. For example, if you believe your blood pressure to be 140 (systolic) over 90 (diastolic), set the switch to 180. If you know it is 120/80, you should set the switch to 150. Don’t worry if you set the pressure switch too low. The monitor will automatically re-inflate to the next higher setting if it is unable to detect the systolic pressure at the current setting.
If you think you have set the pressure too high, stop the inflation process by pressing START. Then reset the switch to the next lower setting.
TAKING YOUR BLOOD PRESSURE
Usual Systolic Set at


➠ ➠
We advise that you record the date and time after each measure- ment because an accurate blood pressure history relies not on single or sporadic readings but on a pattern over time.

To check the exhaust velocity rate, proceed as follows:
NOTE: The exhaust rate of your monitor was preset and tested at the factory to ensure proper measurement. The rate set is dependent on cuff size, not arm size. If using a cuff other than the one shipped with the unit, you may need to adjust the exhaust velocity rate. There is no need to adjust exhaust velocity when changing between a medium cuff and a large cuff with a valve.
Exhaust velocity is the rate at which the air pressure in the cuff deflates during the course of measurement. It is preset at the factory and generally does not need any further adjustment.

If the exhaust velocity range is not within 2 to 5 mmHg, make the following adjustment with the unit turned off:
REMOVE THE COVER FROM THE ADJUSTMENT SCREW

What Causes Variations In Blood Pressure? An individual’s blood pressure varies greatly from day to day and season to season. In hypersensitive individuals, these variations are even more pronounced. Normally, blood pressure rises during work or play and falls to its lowest levels during sleep.
In addition to these factors, diet beverages containing caffeine or alcohol, certain medications, emotional stress and even tight- fitting clothes can make a difference in the readings.
What Affects Blood Pressure? Blood pressure is affected by many factors: age, weight, time of day, activity level, climate, altitude and season. Certain activities can significantly alter blood pressure. Walking can raise systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg. Sleeping can decrease systolic blood pressure by as much as 10 mmHg. Taking your blood pressure repeatedly without waiting an interval of 5 to 10 minutes between readings, or without raising your arm to allow blood to flow back to the heart, can also affect it.
What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).
200 150 100 50 0
Systolic Blood Pressure Diastolic Blood Pressure
Can Hypertension Be Controlled? In many individuals, hypertension can be controlled by altering lifestyle and minimizing stress, and by appropriate medication prescribed and monitored by your doctor. The American Heart Association recommends the following lifestyle suggestions to prevent or control hypertension:
What Is Hypertension? Hypertension (high blood pressure) is the diagnosis given when readings consistently rise above normal. It is well known that hypertension can lead to stroke, heart attack or other illness if left untreated. Referred to as a “silent killer” because it does not always produce symptoms that alert you to the problem, hypertension is treatable when diagnosed early.
Category Systolic (mmHg) Diastolic (mmHg) Optimal
Assessing High Blood Pressure The following standards for assessing high blood pressure (without regard to age) have been established by the National Institutes of Health JNCVI.
Don’t smoke. Have regular physical Reduce salt and fat intake. checkups. Maintain proper weight. Monitor your blood Exercise routinely. pressure at periodic intervals.
❤


How Do I Record My Blood Pressure? Blood pressure readings are typically recorded with the systolic pressure written first, followed by a slash mark and the diastolic pressure. For example, 120 mmHg systolic and 80 mmHg diastolic measurements are written as 120/80. Pulse is simply written with the letter “P” followed by the pulse rate—P 72, for example.
In any case, self-measurement at home supplements the doctor’s readings and provides a more accurate, complete blood pressure history. In addition, clinical studies have shown that the detection and treatment of hypertension is improved when patients both consult their physicians and monitor their own blood pressure at home.
Why Measure Blood Pressure at Home? It is now well known that, for many individuals, blood pressure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. This response is commonly called “white coat hypertension.” One way to determine whether this is the case for you is to take your home monitor to the doctor’s office and, before the doctor or nurse takes your pressure, do it yourself on your home monitor and compare it to your record of home readings.
1/14 9:30 132/98 P 69 6:30 128/87 P 63
B L O O D P R E S S U R E R E C O R D

PRECAUTIONS UA-767 is designed to be used at home, by those who are eighteen (18) years and older, to monitor blood pressure (systolic and diastolic) and pulse rate. This monitor is not designed to measure the blood pressure of people with common arrhythmias, such as atrial or ventricular premature beats or atrial fibrillation. It is not designed for ambulatory use.
Please read this important information before using your monitor. Please remember that only a medical practitioner is qualified to interpret your blood pressure measurements. Use of this device should not replace regular medical examinations. Have your physician review your procedure for using this device. He or she will want to verify blood pressure measurements before prescribing or adjusting medications. Consult your physician if you have any doubt about your readings. Should a mechanical problem occur, contact LifeSource. Do not attempt to service, calibrate, or repair this device. Because your UA-767 monitor contains delicate, high-precision parts, avoid exposing it to extremes in temperature or humidity or to direct sunlight, shock and dust. LifeSource guarantees the accuracy of this device only when it is stored and used within the temperature and humidity ranges noted on page 18. Clean the monitor and cuff with a dry, soft cloth or a cloth dampened with water and a mild detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean monitor or cuff. Remove and replace batteries if monitor is not used for more than six months.
Blood pressure measurements determined by the UA-767 are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute for electronic or automated sphygmomanometers.
Model UA-767 Type Oscillometric Display Digital, 16-mm character height Pressure/pulse displayed simultaneously Measurement range Pressure: 20 mmHg to 280 mmHg Pulse: 40 pulses to 200 pulses/minute Accuracy Pressure: +3 mmHg or 2%, whichever is greater Pulse: +5% Pressurization Automatic, using micropump Depressurization Constant-air release-valve system Power source 4 type AA (1.5 volt) alkaline batteries (not included) Battery life Approximately 4 months with one daily measurement Operating environment 50ºF to 104ºF (10ºC to 40ºC) Less than 85% relative humidity Storage environment -4ºF to 158ºF (-20ºC to 70ºC) Less than 95% relative humidity Dimensions Length: 4.2”(106 mm) Width: 6.4”(163 mm) Height: 2.8” (70 mm) Weight 10.9 oz. (308 g) without batteries

LifeSource A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 LifeSource Health Line (Toll-Free): 1-888-726-9966 www.LifeSourceOnline.com
For more information regarding use, care or servicing of your Blood Pressure Monitor, contact:
This LifeSource product is covered by a Lifetime Warranty. See warranty card for details.
Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. The product line includes:

Aneroid Blood Pressure Kits Blood Pressure Cuffs Digital Blood Pressure Monitors Digital Thermometers Stethoscopes
UA-767 Printed in China 12/99
A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 www.LifeSourceOnline.com
1-888-726-9966
Pages referenced in this manual — manufacturer, support and downloads.
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.