Trilingual Instruction Guide – MODEL UA-767

Page 3 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — MONITOR COMPONENTS
Page 3 · click to enlarge
Page 3
LifeSource One-Step Blood Pressure Monitor (UA-767) — ERROR 3 diagram (page 3)
Fig. 3
LifeSource One-Step Blood Pressure Monitor (UA-767) — ERROR 3 diagram (page 3)
Fig. 4

ERROR 3

Appears when pressure value did not increase during cuff inflation: 1) cuff is not fastened 2) air connector is not connected properly.

Appears if pressure value is unstable due to movement during reading.

Appears when systolic and diastolic measure- ments are within 10 mmHg of each other.

Battery voltage is too low for monitor to work properly.

Appears during measure- ment; blinks and beeps when pulse is detected.

MONITEUR TUYAU DU BRASSARD

Make sure cuff tube is properly connected to monitor, cuff is securely fastened and exhaust rate is between 2 mmHg and 5 mmHg. Make sure air connector is inserted properly into tube and monitor.

Check for air leakage; make sure cuff tube is properly connected to monitor and exhaust rate is in increments of 2 mmHg. to 5 mmHg

Remain still and refrain from talking; measure- ment in progress.

ADJUSTMENT SCREW

LifeSource One-Step Blood Pressure Monitor (UA-767) — ADJUSTMENT SCREW diagram (page 3)
Fig. 5
LifeSource One-Step Blood Pressure Monitor (UA-767) — ADJUSTMENT SCREW diagram (page 3)
Fig. 6

MONITOR COMPONENTS

BLACK PROTECTION CAP ON ADJUSTMENT SCREW

SYSTOLIC DISPLAY

START

240 210 180 150 PRESSURE SWITCH

START BUTTON

DIASTOLIC DISPLAY

TUBE

?

Show 6 more labels / short text on this page
Display Symbol · Condition/ Cause · Recommended Action · PUL / min · DIASTOLIC mmHg · SYSTOLIC mmHg
Page 4 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — ARM SIZE RECOMMENDED CUFF SIZE
Page 4 · click to enlarge
Page 4
3. Replace cover by sliding it into the compartment and gently pressing into place.
2. Put in top row of batteries first. Place the batteries in compartment with positive (+) and negative (–) terminals matching those indicated in the compartment. Be sure batteries make contact with compartment terminals.
1. Remove battery compartment cover by gently pushing down on arrow and sliding cover forward.

You must install 4 type AA (1.5 volt) batteries and attach the cuff to the monitor before using it. To install batteries (or replace them if the “Low Battery” symbol appears on display), proceed as follows:

BEFORE YOU START

ARM SIZERECOMMENDED CUFF SIZEREPLACEMENT CUFF MODEL# 5.1” - 7.9” (13-20cm)SmallUA-279 7.5” - 12.2” (19-31cm)MediumUA-280 11.8” - 17.7” (30-45cm)LargeUA-281

To attach cuff, insert air connector at end of cuff tube into air socket on left side of monitor. Should you need to change the cuff size, the preset exhaust velocity rate may have to be adjusted (see page 11).

Using the correct cuff size is important for an accurate reading. A cuff that is too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small yields a measurement that is higher. With your arm hanging at the side of your body, measure the circumference of your upper arm at the midpoint between shoulder and elbow.

TUBE

11.8” - 17.7” (30-45cm)

7.5” - 12.2” (19-31cm)

5.1” - 7.9” (13-20cm)

ARM SIZE RECOMMENDED CUFF SIZE

SELECTING THE CORRECT CUFF

AIR CONNECTOR

AIR CONNECTOR

MONITOR

REPLACEMENT CUFF MODEL#

UA-281

UA-280

UA-279

Large · Medium · Small

Page 5 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — PRESSURE
Page 5 · click to enlarge
Page 5
3. Pull the end of the cuff to tighten it, fold back the extra material, and fasten the Velcro. The cuff should be snug but not too tight. You should be able to insert two fingers between the cuff and your arm.
2. Lay left arm on the table, palm up and thread cuff end through metal loop, smooth side against arm. Then position the tube off-center toward the inner side of arm in line with the little finger.
1. Sit comfortably with your left arm resting on a flat surface so that the center of your upper arm is at the same height as your heart.

Now you are ready. Follow these simple steps:

Usual SystolicSet at up to 120150 121-150180 151-180210 181-210240 211+see NOTE

4. Before taking your blood pressure, you will need to set the pressure switch to a number that is at least 30 mmHg higher than your expected systolic pressure. Use the chart below as a guide. For example, if you believe your blood pressure to be 140 (systolic) over 90 (diastolic), set the switch to 180. If you know it is 120/80, you should set the switch to 150. Don’t worry if you set the pressure switch too low. The monitor will automatically re-inflate to the next higher setting if it is unable to detect the systolic pressure at the current setting.

If you think you have set the pressure too high, stop the inflation process by pressing START. Then reset the switch to the next lower setting.

PRESSURE

TAKING YOUR BLOOD PRESSURE

Usual Systolic Set at

Show 6 more labels / short text on this page
211+ · 181-210 · 151-180 · 121-150 · up to 120 · see NOTE
Page 6 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — TAKING YOUR BLOOD PRESSURE TAKING YOUR BLOOD PRESSURE
Page 6 · click to enlarge
Page 6
8. Remove cuff and make a note of your blood pressure and pulse rate on the chart (see page 16), indicating date and time of measurement.
7. Wait for the long beep indicating the measurement is complete. The systolic and diastolic pressure readings and pulse rate are then displayed. The cuff deflates and the monitor shuts itself off after a few moments, or you can turn it off by pressing the start button.
6. It is normal for the cuff to feel very tight. You will see all the display symbols appear briefly. As the cuff pressurizes, blinking numerals—starting at zero—are displayed. When inflation is complete, the MEASUREMENT IN PROGRESS symbol blinks and beeps with each pulse beat.
5. Press the START button, remain still and do not talk as the cuff automatically inflates to the correct level.
LifeSource One-Step Blood Pressure Monitor (UA-767) — TAKING YOUR BLOOD PRESSURE TAKING YOUR BLOOD PRESSURE diagram (page 6)
Fig. 7

TAKING YOUR BLOOD PRESSURE TAKING YOUR BLOOD PRESSURE

➠ ➠

We advise that you record the date and time after each measure- ment because an accurate blood pressure history relies not on single or sporadic readings but on a pattern over time.

Page 7 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — CHECK/ADJUST EXHAUST VELOCITY
Page 7 · click to enlarge
Page 7
3. Watch the display screen as the cuff deflates. When the MEASUREMENT IN PROGRESS symbol appears, the numbers should decline in increments of 2 to 5 mmHg.
2. Press the START button. The unit will begin taking a reading.
1. Put on the arm cuff.

To check the exhaust velocity rate, proceed as follows:

NOTE: The exhaust rate of your monitor was preset and tested at the factory to ensure proper measurement. The rate set is dependent on cuff size, not arm size. If using a cuff other than the one shipped with the unit, you may need to adjust the exhaust velocity rate. There is no need to adjust exhaust velocity when changing between a medium cuff and a large cuff with a valve.

Exhaust velocity is the rate at which the air pressure in the cuff deflates during the course of measurement. It is preset at the factory and generally does not need any further adjustment.

LifeSource One-Step Blood Pressure Monitor (UA-767) — CHECK/ADJUST EXHAUST VELOCITY diagram (page 7)
Fig. 8

➠ ➠

3. Repeat the test until the exhaust velocity is between 2 and 5 mmHg.
2. Using a small screwdriver, rotate adjustment screw about 45 degrees clockwise (to raise the rate) or counterclockwise (to lower the rate).
1. With the edge of your fingernail, peel off the thin black plastic disk covering the adjustment screw. If there is no thin plastic disk, proceed to step 2.

If the exhaust velocity range is not within 2 to 5 mmHg, make the following adjustment with the unit turned off:

CHECK/ADJUST EXHAUST VELOCITY

REMOVE THE COVER FROM THE ADJUSTMENT SCREW

Page 8 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — ABOUT BLOOD PRESSURE
Page 8 · click to enlarge
Page 8

What Causes Variations In Blood Pressure? An individual’s blood pressure varies greatly from day to day and season to season. In hypersensitive individuals, these variations are even more pronounced. Normally, blood pressure rises during work or play and falls to its lowest levels during sleep.

In addition to these factors, diet beverages containing caffeine or alcohol, certain medications, emotional stress and even tight- fitting clothes can make a difference in the readings.

What Affects Blood Pressure? Blood pressure is affected by many factors: age, weight, time of day, activity level, climate, altitude and season. Certain activities can significantly alter blood pressure. Walking can raise systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg. Sleeping can decrease systolic blood pressure by as much as 10 mmHg. Taking your blood pressure repeatedly without waiting an interval of 5 to 10 minutes between readings, or without raising your arm to allow blood to flow back to the heart, can also affect it.

What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).

ABOUT BLOOD PRESSURE ❤

Fluctuation within a day (case35 year old male)

200 150 100 50 0

Systolic Blood Pressure Diastolic Blood Pressure

Can Hypertension Be Controlled? In many individuals, hypertension can be controlled by altering lifestyle and minimizing stress, and by appropriate medication prescribed and monitored by your doctor. The American Heart Association recommends the following lifestyle suggestions to prevent or control hypertension:

What Is Hypertension? Hypertension (high blood pressure) is the diagnosis given when readings consistently rise above normal. It is well known that hypertension can lead to stroke, heart attack or other illness if left untreated. Referred to as a “silent killer” because it does not always produce symptoms that alert you to the problem, hypertension is treatable when diagnosed early.

Category Systolic (mmHg) Diastolic (mmHg) Optimal

Assessing High Blood Pressure The following standards for assessing high blood pressure (without regard to age) have been established by the National Institutes of Health JNCVI.

Don’t smoke. Have regular physical Reduce salt and fat intake. checkups. Maintain proper weight. Monitor your blood Exercise routinely. pressure at periodic intervals.

ABOUT BLOOD PRESSURE

❤

Show 6 more labels / short text on this page
Arising · Start Work · Telephone · Argument · Leave Work · Sleep
Page 9 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — E X A M P L E
Page 9 · click to enlarge
Page 9
LifeSource One-Step Blood Pressure Monitor (UA-767) — DATE AM SYS/DIA PULSE PM SYS/DIA PULSE diagram (page 9)
Fig. 9

How Do I Record My Blood Pressure? Blood pressure readings are typically recorded with the systolic pressure written first, followed by a slash mark and the diastolic pressure. For example, 120 mmHg systolic and 80 mmHg diastolic measurements are written as 120/80. Pulse is simply written with the letter “P” followed by the pulse rate—P 72, for example.

In any case, self-measurement at home supplements the doctor’s readings and provides a more accurate, complete blood pressure history. In addition, clinical studies have shown that the detection and treatment of hypertension is improved when patients both consult their physicians and monitor their own blood pressure at home.

Why Measure Blood Pressure at Home? It is now well known that, for many individuals, blood pressure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. This response is commonly called “white coat hypertension.” One way to determine whether this is the case for you is to take your home monitor to the doctor’s office and, before the doctor or nurse takes your pressure, do it yourself on your home monitor and compare it to your record of home readings.

DATE AM SYS/DIA PULSE PM SYS/DIA PULSE

1/14 9:30 132/98 P 69 6:30 128/87 P 63

NameAge: Weight:

B L O O D P R E S S U R E R E C O R D

E X A M P L E

Page 10 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — SPECIFICATIONS
Page 10 · click to enlarge
Page 10

PRECAUTIONS UA-767 is designed to be used at home, by those who are eighteen (18) years and older, to monitor blood pressure (systolic and diastolic) and pulse rate. This monitor is not designed to measure the blood pressure of people with common arrhythmias, such as atrial or ventricular premature beats or atrial fibrillation. It is not designed for ambulatory use.

Please read this important information before using your monitor. Please remember that only a medical practitioner is qualified to interpret your blood pressure measurements. Use of this device should not replace regular medical examinations. Have your physician review your procedure for using this device. He or she will want to verify blood pressure measurements before prescribing or adjusting medications. Consult your physician if you have any doubt about your readings. Should a mechanical problem occur, contact LifeSource. Do not attempt to service, calibrate, or repair this device. Because your UA-767 monitor contains delicate, high-precision parts, avoid exposing it to extremes in temperature or humidity or to direct sunlight, shock and dust. LifeSource guarantees the accuracy of this device only when it is stored and used within the temperature and humidity ranges noted on page 18. Clean the monitor and cuff with a dry, soft cloth or a cloth dampened with water and a mild detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean monitor or cuff. Remove and replace batteries if monitor is not used for more than six months.

SPECIFICATIONS IMPORTANT INFORMATION

Blood pressure measurements determined by the UA-767 are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute for electronic or automated sphygmomanometers.

Model UA-767 Type Oscillometric Display Digital, 16-mm character height Pressure/pulse displayed simultaneously Measurement range Pressure: 20 mmHg to 280 mmHg Pulse: 40 pulses to 200 pulses/minute Accuracy Pressure: +3 mmHg or 2%, whichever is greater Pulse: +5% Pressurization Automatic, using micropump Depressurization Constant-air release-valve system Power source 4 type AA (1.5 volt) alkaline batteries (not included) Battery life Approximately 4 months with one daily measurement Operating environment 50ºF to 104ºF (10ºC to 40ºC) Less than 85% relative humidity Storage environment -4ºF to 158ºF (-20ºC to 70ºC) Less than 95% relative humidity Dimensions Length: 4.2”(106 mm) Width: 6.4”(163 mm) Height: 2.8” (70 mm) Weight 10.9 oz. (308 g) without batteries

SPECIFICATIONS

Page 11 of the LifeSource One-Step Blood Pressure Monitor (UA-767) user manual — Call Toll-Free LifeSource Health Line:
Page 11 · click to enlarge
Page 11
19 LifeSource Health Line (Toll-Free)1-888-726-9966
For Canada Residents, please contactAuto Control Medical 206A boul. Brunswick Pointe-Claire, Quebec H9R 5P9 Canada AutoControl (Toll-Free): 1-800-461-0991

LifeSource A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 LifeSource Health Line (Toll-Free): 1-888-726-9966 www.LifeSourceOnline.com

For more information regarding use, care or servicing of your Blood Pressure Monitor, contact:

This LifeSource product is covered by a Lifetime Warranty. See warranty card for details.

Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. The product line includes:

LifeSource One-Step Blood Pressure Monitor (UA-767) — CONTACT INFORMATION Need Help? diagram (page 11)
Fig. 10 · CONTACT INFORMATION Need Help?

Aneroid Blood Pressure Kits Blood Pressure Cuffs Digital Blood Pressure Monitors Digital Thermometers Stethoscopes

CONTACT INFORMATION Need Help?

UA-767 Printed in China 12/99

A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 www.LifeSourceOnline.com

Call Toll-Free LifeSource Health Line:

1-888-726-9966

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the LifeSource One-Step Blood Pressure Monitor (UA-767)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the LifeSource One-Step Blood Pressure Monitor (UA-767)

Be the first to ask — other owners or our team will answer.

Original document

LifeSource One-Step Blood Pressure Monitor (UA-767) manual PDF

11 pages·264 KB·Blood Pressure Monitor·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…