LifeSource UA-853

26 pages·4.5 MB·Blood Pressure Monitor·PDF·ENDEFRITNLJA
LifeSource UA-853
Table of Contents 32 sections Premium Automatic Blood Pressure Monitor · Trilingual Instruction Guide - Model UA-853 · important information ✓ · TABLE OF CONTENTS · … ShowHide
Page 1 of the LifeSource UA-853 user manual — Trilingual Instruction Guide - Model UA-853
Page 1 · click to enlarge
Page 1

Premium Automatic Blood Pressure Monitor

LifeSource UA-853 — Premium Automatic Blood Pressure Monitor diagram (page 1)
Fig. 1 · Premium Automatic Blood Pressure Monitor

Trilingual Instruction Guide - Model UA-853

English l Français l Español

Page 2 of the LifeSource UA-853 user manual — important information ✓
Page 2 · click to enlarge
Page 2

important information ✓

Please read this important information before using your monitor. ☞ Please remember that only a medical practitioner is qualified to interpret your blood pressure measurements. Use of this device should not replace regular medical examinations. ☞ Consult your physician if you have any doubt about your readings. Should a mechanical problem occur, contact LifeSource® at 1-888-726-9966. ☞ Do not attempt to service, calibrate, or repair this monitor. ☞ Because your UA-853 monitor contains delicate, high-precision parts, avoid exposing it to extremes in temperature or humidity or to direct sunlight, shock and dust. LifeSource® guarantees the accuracy of this monitor only when it is stored and used within the temperature and humidity ranges noted on page E-21. ☞ Clean the monitor and cuff with a dry, soft cloth or a cloth dampened with water and a mild detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean monitor or cuff. ☞ Remove and replace batteries if monitor is not used for more than six months. Alkaline batteries recommended.

PRECAUTIONS The UA-853 is designed to be used at home, by those who are eighteen (18) years and older, to monitor blood pressure (systolic and diastolic) and pulse rate.

Page 3 of the LifeSource UA-853 user manual — important information ✓
Page 3 · click to enlarge
Page 3

Fast. Easy. Accurate.

Congratulations on purchasing a state-of-the-art LifeSource® blood pressure monitor—one of the most technologically advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen.

Physicians agree that daily self-monitoring of blood pressure is an important step individuals can take to maintain cardiovascular health and prevent the serious consequences associated with undetected and untreated hypertension.

A&D Medical has been manufacturing high quality healthcare products for ove 20 years. We are committed to providing you and your family with monitoring devices that provide the utmost in accuracy and convenience.

LifeSource - Designed for Life.

Page 5 of the LifeSource UA-853 user manual — ?
Page 5 · click to enlarge
Page 5

WHAT DISPLAY SYMBOLS MEAN ?

Condition/CauseRecommended Action The symbol shown while measure M ment is in progress. It blinks while re detecting the pulse.easurement is in progress, main as still as possible. An irregular heartbeat or body movement may have occurred. T Refer to page E-11 for more infor co mation on irregular heartbeats. Previous measurement stored in MEMORY. The battery power indicator during measurement. R o The batteries are low when itake measurement again and nsult with your physician. eplace all batteries with new nes when the symbol blinks. A blinks. Unstable blood pressure due to Tr excessive body movement. R m The systolic and diastolic values are Fa within 10 mmHg of each other. th The pressure value did not increase C during inflation. tulkaline batteries recommended. y the measurement again. emain very still during the easurement. sten the cuff correctly, and try e measurement again. heck for air leaks along the be and around the air socket. The cuff is not fastened correctly. R suefasten the cuff and retake mea rement. There is an air leak in the cuff or M monitor. n Tr The pulse is not detected correctly. R R make sure tube is properly con ected to cuff and monitor. y the measurement again. emain very still during the easurement. M Cuff Inflation Meter reeasurement is in progress, main as still as possible. R Pressure Rating Indicator™ Pr fuefer to the section “About essure Rating Indicator” for rther explanation. The alarm is turned on. Pr aless AM button once to turn arm off.

?

Show 4 more labels / short text on this page
Display/ · Symbol · Battery Full · Battery Low
Page 6 of the LifeSource UA-853 user manual — MONITOR COMPONENTS
Page 6 · click to enlarge
Page 6

MONITOR COMPONENTS

LifeSource UA-853 — MONITOR COMPONENTS diagram (page 6)
Fig. 2

(Orange illumination during the AM) (Blue illumination during the PM) Air Socket

AM Button · Display MEMORY Button

Page 7 of the LifeSource UA-853 user manual — ?!
Page 7 · click to enlarge
Page 7

HOW UA-853 WORKS ?

LifeSource® blood pressure monitors are easy to use, accurate and digitally display full measurement readouts. Our technology is based on the “oscillometric method” – a noninvasive blood pressure determination. The term “oscillation” refers to any measure of vibrations caused by the arterial pulse. Our monitor examines the pulsatile pressure generated by the arterial wall as it expands and contracts against the cuff with each heartbeat.

The cuff is inflated until the artery is fully blocked. The inflation speed is maximized and the pressure level is optimized by the device. The monitor takes measurements while the cuff is both inflating and deflating. This results in a faster measurement providing greater comfort to the user.

AM/PM TimeWise™ Technology This monitor automatically stores each measurement as an AM or PM reading in memory (note: clock must be properly programmed before initial use). Healthcare providers agree that understanding your AM and PM readings can assist in the proper diagnosis and management of your heart health through lifestyle and/or medication modifications.

before you start

You must install 4 type AA (1.5 volt) batteries (alkaline batteries recommended), or use the AC Adapter (see page E-14 for using the Monitor with AC Adapter) and attach the cuff to the monitor before using it. To install batteries (or replace them if the “Low Battery” symbol appears on display), proceed as follows:

1. Remove battery compartment cover by gently pushing down on arrow and sliding cover forward.
2. Put in bottom row of batteries first. Place the batteries in compartment with positive (+) and negative (–) terminals matching those indicated in the compartment. Be sure batteries make contact with compartment terminals.
3. Replace cover by sliding it into the compartment and gently pressing into place.

?!

Step 2 · Step 1 · Step 3

Page 9 of the LifeSource UA-853 user manual — TAKING YOUR BLOOD PRESSURE
Page 9 · click to enlarge
Page 9

TAKING YOUR BLOOD PRESSURE

Adjusting the Built-in Clock The UA-853 Automatic Blood Pressure Monitor has a built-in clock that gives you the time and date and can tell you when a blood pressure reading stored in memory was taken. We recommend that you adjust the clock prior to use. To adjust the clock, follow these simple steps:

1. Press and hold the AM button until the digits start blinking.
2. Select the year using the PM button. Press the AM button to set the current year and move to month/day selection. The date can be set anywhere between the years 2006 and 2056.
3. Select the month using the PM button. Press the AM button to set the current month and move to day selection.
4. Select the day using the PM button. Press the AM button to set the current day and move to hour/minute selection.
5. Select the hour using the PM button. Press the AM button to set the current hour and move to minute selection.
6. Select the minute using the PM button. Press the AM button to set the current minute and activate the clock.
7. The Clock is now set.
Page 10 of the LifeSource UA-853 user manual — UA-853 DIGITAL BLOOD PRESSURE MONITOR
Page 10 · click to enlarge
Page 10

TAKING YOUR BLOOD PRESSURE

Setting the Alarm

This monitor has 3 reminder alarms. You can set 3 different reminder alarms within a 24-hour period.

1. Press and hold the AM button until the digits start blinking. Then press the MEMORY button. “P-1” is shown at the middle line of the display.
2. Select the hour using the PM button and press the AM button to set the hour and move to minute selection. Press the MEMORY button to complete the reminder setting after selecting minute.

UA-853 DIGITAL BLOOD PRESSURE MONITOR

3. Then the display shows “P-2” and the symbol for the second reminder. Repeat the above process if you need a second or third reminder programmed.
4. If you do not want a second or third reminder set, simply press the MEMORY button until the clock is displayed on the screen.

Press and hold

Page 11 of the LifeSource UA-853 user manual — TAKING YOUR bLOOD PRESSURE
Page 11 · click to enlarge
Page 11

TAKING YOUR bLOOD PRESSURE

Tips for Blood Pressure Monitoring: ☞ Relax for about 5 minutes before measurement. ☞ Do not smoke or ingest caffeine at least 30 minutes prior to measurement. ☞ Remove constricting clothing and place cuff on bare arm. ☞ Unless your physician recommends otherwise, use left arm to measure pressure. ☞ Do not talk during measurement. ☞ Do not cross legs and keep your feet flat on floor during measurement.

Now you are ready. Follow these simple steps: 1. Sit comfortably with your left arm resting on a

flat surface so that the center of your upper arm is at the same height as your heart.

2. Lay left arm on the table, palm up and thread cuff end through metal loop, smooth size against arm. Then position the tube off-center toward the inner side of arm in line with the little finger.
3. Secure cuff around arm. The cuff should be snug but not too tight. You should be able to insert two fingers between the cuff and your arm.
4. Confirm that the Index Mark points within the proper fit range.
Page 12 of the LifeSource UA-853 user manual — SYSTOLIC
Page 12 · click to enlarge
Page 12

TAKING YOUR BLOOD PRESSURE

5. Press the START button. As the cuff pressurizes, measurement will begin and the Cuff Inflation Meter will show on the display screen. It is normal for the cuff to feel very tight.

Note: If an appropriate pressure is not obtained, the device automatically starts to inflate again. Note: If you wish to stop inflation at any time, press the START button again.

6. When the measurement is complete, the systolic and diastolic pressure readings and pulse rate are displayed. The cuff deflates and the monitor automatically shuts off after 60 seconds, or you can turn it off by pressing the START button. Your measurement is automatically stored in memory (the monitor can store up to 90 readings in memory).

SYSTOLIC

DIASTOLIC

PULSE

7. Remove cuff and make note of your blood pressure and pulse rate in the Blood Pressure Logbook, indicating the date and time of measurement.
Page 13 of the LifeSource UA-853 user manual — SYSTOLIC
Page 13 · click to enlarge
Page 13
LifeSource UA-853 — SYSTOLIC diagram (page 13)
Fig. 3

Measurement with the desired systolic pressure: If your systolic pressure is expected to exceed 230 mmHg follow these steps:

1. Place the cuff on the left arm, unless your physician tells you otherwise.
2. Press and hold the START button until a number 30 to 40 mmHg higher than your expected systolic pressure appears.
3. Release the START button when the desired number is reached. 280 mmHg is the highest pressure level the monitor can reach. Measurement will then begin. Continue to measure your blood pressure as described on Page E-8.
Page 14 of the LifeSource UA-853 user manual — About cuff inflation METER
Page 14 · click to enlarge
Page 14

WHAT IS AN Irregular HeartBeat

The UA-853 Blood Pressure Monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. The Irregular Heartbeat symbol will appear in the display window in the event an Irregular Heartbeat has occured during measurement. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeat intervals during the blood pressure measurement. It is important that you relax, remain still and refrain from talking during measurements.

About cuff inflation METER

The Cuff Inflation Meter is located on the left side of the display screen to tell you when the blood pressure monitor is inflating and deflating the cuff. The Cuff Inflation Meter moves up during inflation and moves down during deflation.

Inflation in Progress Inflation Complete Deflation/Measurement in

Progress ·

Page 15 of the LifeSource UA-853 user manual — About Pressure Rating Indicator™
Page 15 · click to enlarge
Page 15

About Average Reading

The UA-853 provides an average blood pressure reading based on the total measurements stored in memory. The average blood pressure reading appears briefly when you recall the measurements stored in memory. You can also access your AM/PM Average Readings by pressing the MEMORY button, then the AM or PM button.

number of measurements stored in

About Pressure Rating Indicator™

The Pressure Rating Indicator™ is a feature which provides a snapshot of your blood pressure classification based on your measurements. This will let you quickly know what your blood pressure readings mean. Each segment of the bar indicator corresponds to the Seventh Report of the Joint National Committee (JNC7) on Prevention, Evaluation and Treatment of High Blood Pressure from the National Heart, Lung and Blood Institute - May 2003. For a more detailed look at this blood pressure classification, please refer to “Assessing High Blood Pressure” on page E-16.

Stage 2 Hypertension Stage 1 Hypertension Prehypertension Normal

:The indicator displays a segment, based on the current measurement, corresponding to the JNC7 Classification.

Show 6 more labels / short text on this page
average · memory · JNC7 Classification Indicator · Example · Hypertension · Stage 2
Page 16 of the LifeSource UA-853 user manual — About Memory
Page 16 · click to enlarge
Page 16

About Memory

1. Press the MEMORY button.

Press the MEMORY button

2. The average data of all the measurements and the number of measurements stored are displayed. Then, each time the MEMORY button is pressed, the memory data is displayed starting from the most recent. After the last reading is displayed, the display returns to the clock display.
3. To recall AM/PM readings, press:

Average of all readings

MEMORY + AM button

Average of total AM measurements stored is displayed. Each time the AM button is pressed, the memory data taken during the AM is displayed starting from the most recent. After the last measurement is displayed, the display returns to the clock display.

MEMORY + PM button

Average of total PM measurements stored is displayed. Each time the PM button is pressed, the memory data taken during the PM is displayed starting from the most recent. After the last measurement is displayed, the display returns to the clock display.

If the MEMORY button is pressed again, the display returns to the average data of all the measurements.

Clearing the memory data 4. Press and hold the MEMORY button until the “ ” (memory) mark blinks to clear the stored data.

Averaged Systolic · Averaged Diastolic · Number of Readings

Page 17 of the LifeSource UA-853 user manual — USING THE MONITOR WITH AC ADAPTER
Page 17 · click to enlarge
Page 17

USING THE MONITOR WITH AC ADAPTER

The UA-853 has an AC adapter jack to allow you to supply power from an outlet in your home. We recommend you use only the supplied AC adapter to avoid potential damage to the monitor.

Connecting the AC adapter to the monitor: 1. Gently insert the AC adapter plug into a 120V AC outlet.

2. Connect the AC adapter plug into the jack on the back of the monitor.

Disconnecting the AC adapter from the monitor: 1. Turn the unit off by pressing the START button.

2. Disconnect the plug from the jack of the monitor quickly.
3. Gently remove the AC adapter from the outlet.

Jack · AC Adapter Plug

Page 18 of the LifeSource UA-853 user manual — ABOUT BLOOD PRESSURE
Page 18 · click to enlarge
Page 18

ABOUT BLOOD PRESSURE

What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).

What Affects Blood Pressure? Blood pressure is affected by many factors: age, weight, time of day, activity level, climate, altitude and season. Certain activities can significantly alter one’s blood pressure. Walking can raise systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg. Sleeping can decrease systolic blood pressure by as much as 10 mmHg. Taking your blood pressure repeatedly without waiting an interval of 5 minutes between readings, or without raising your arm to allow blood to flow back to the heart, can also affect it.

In addition to these factors, beverages containing caffeine or alcohol, certain medications, emotional stress and even tight-fitting clothes can make a difference in the readings.

What Causes Variations In Blood Pressure? An individual’s blood pressure varies greatly from day to day and season to season. For hypersensitive individuals, these variations are even more pronounced. Normally, blood pressure rises during work or play and falls to its lowest levels during sleep.

Page 19 of the LifeSource UA-853 user manual — ABOUT BLOOD PRESSURE
Page 19 · click to enlarge
Page 19

ABOUT BLOOD PRESSURE

Assessing High Blood Pressure for Adults The following standards for assessing high blood pressure (without regard to age or gender) have been established as a guideline. Please note that other risk factors (e.g. diabetes, obesity, smoking, etc.) need to be taken into consideration and may affect these figures. Consult with your physician for an accurate assessment.

JNC7 Classification Table – for adults within the U.S.

BP Classification Systolic (mmHg) Diastolic (mmHg)

Normal < 120 and < 80

Prehypertension 120-139 or 80-89

Stage 1 Hypertension 140-159 or 90-99

Stage 2 Hypertension ≥ 160 or ≥ 100

SOURCE: The Seventh Report of the Joint National Committee on Prevention, Evaluation and Treatment of High Blood Pressure for Adults. National Heart, Lung and Blood Institute - May 2003.

WHO Classification Table - for adults outside of the U.S. (e.g. Canada, Mexico)

BP Classification Systolic (mmHg) Diastolic (mmHg)

Optimal < 120 and < 80

Normal < 130 or < 85

High-Normal 130-139 or 85-89

Stage 1 Hypertension 140-159 or 90-99

Stage 2 Hypertension 160-179 or 100-109

Stage 3 Hypertension ≥ 180 or ≥ 110

SOURCE: Standards to assess high blood pressure, without regard to age or gender, have been established by the World Health Organization (WHO).

Page 20 of the LifeSource UA-853 user manual — ABOUT BLOOD PRESSURE
Page 20 · click to enlarge
Page 20

ABOUT BLOOD PRESSURE

What Is Hypertension? Hypertension (high blood pressure) is the diagnosis given when readings consistently rise above normal. It is well known that hypertension can lead to stroke, heart attack or other illness if left untreated. Referred to as a “silent killer” because it does not always produce symptoms that alert you to the problem, hypertension is treatable when diagnosed early.

Can Hypertension Be Controlled? In many individuals, hypertension can be controlled by altering lifestyle and minimizing stress, and by appropriate medication prescribed and monitored by your doctor. The American Heart Association recommends the following lifestyle suggestions to prevent or control hypertension:

Don’t smoke. Have regular physical checkups

Reduce salt and fat intake. Maintain proper weight.

Monitor your blood Exercise routinely. pressure at periodic intervals.

Why Measure Blood Pressure at Home? It is now well known that, for many individuals, blood pressure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. This response is commonly called “white coat hypertension.” In any case, self-measurement at home supplements your doctor’s readings and provides a more accurate, complete blood pressure history. In addition, clinical studies have shown that the detection and treatment of hypertension is improved when patients both consult their physicians and monitor their own blood pressure at home.

Page 21 of the LifeSource UA-853 user manual — ABOUT BLOOD PRESSURE
Page 21 · click to enlarge
Page 21

ABOUT BLOOD PRESSURE

Answers to Why Your Readings Are Different Between Home and at the Doctor’s Office

Why are my readings different between home and at a doctor’s office? Your blood pressure readings taken in a doctor’s office or hospital setting may be elevated as a result of apprehension and anxiety. This response is known as white coat hypertension.

When I bring my monitor to a doctor’s office, why do I get a different measurement from my monitor to that taken by a doctor or nurse? The healthcare professional may be using a different sized cuff. The size of the bladder inside the cuff is critical for the accuracy of the measurement. This may give you a different reading. A cuff too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small will produce a reading that is higher than the correct blood pressure. There may also be other factors that can cause the difference in measurements.

Keys to Successful Monitoring: Blood pressure fluctuates throughout the day. We recommend that you are consistent in your daily measurement routine:

  • Measure at the same time every day.
  • Sit in the same chair/position.
  • Do not cross legs and keep your feet flat on the floor.
  • Relax for 5 minutes before measurement.
  • Use the correct cuff size to get an accurate reading.
  • Sit still during measurement – no talking, eating or sudden movements.
  • Record your measurement in the logbook.
Page 22 of the LifeSource UA-853 user manual — ABOUT BLOOD PRESSURE
Page 22 · click to enlarge
Page 22

ABOUT BLOOD PRESSURE

Establishing Baseline Measurements The most important method to get an accurate blood pressure measurement is consistency. To get the most benefit out of your monitor, it is important to establish a “baseline measurement.” This helps build a foundation of measurements that you can use to compare against future readings. To build this baseline measurement, devote two weeks for consistent blood pressure monitoring. This involves doing everything the same way when you measure (e.g. measuring during the same time of day, in the same location, sitting in the same chair, using the same cuff, etc.). Once you establish your baseline measurement, you can start evaluating if your measurement has been affected based on things lifestyle changes or medication treatment.

How Do I Record My Blood Pressure? Blood pressure readings are typically recorded with the systolic pressure written first, followed by a slash mark and the diastolic pressure. For example, 120 mmHg systolic and 80 mmHg diastolic measurements are written as 120/80. Pulse is simply written with the letter “P” followed by the pulse rate—P 72, for example. Please see the back of the manual for the blood pressure tracking log.

Customer Support Tools Online Additional tools are available on www.LifeSourceOnline.com to help you get the most out of your blood pressure monitoring. These include:

  • Large Print Instruction Manuals
  • Animated Operating Instructions
  • Additional Logbook Sheets
Page 23 of the LifeSource UA-853 user manual — TROUBLESHOOTING
Page 23 · click to enlarge
Page 23

TROUBLESHOOTING

Problem Probable Cause Corrective Action

Nothing appears in the display, after I press the START button.

Batteries are drained.

Replace all batteries with new ones. Alkaline batteries recom mended.

Battery terminals are not in the correct position.

Reinstall the batteries with nega tive and positive terminals match ing those indicated in the battery compartment.

Battery voltage is too low. Low battery symbol blinks. [If the batteries are drained completely, the mark does not appear.]

Replace all batteries with new ones. Alkaline batteries recom mended.

The cuff does not inflate.

The cuff is not fastened prop erly.

Fasten the cuff correctly.

The unit does not measure. Readings are too high or too low.

You moved your arm or body during the measurement.

Make sure you remain very still and quiet during the measurement.

The cuff position is not correct.

Sit comfortably and still. Make sure the cuff is at the same level as your heart.

You are using the wrong size cuff.

See Pg. E-5 "Selecting the Correct Cuff."

Taking too many readings on the same arm in a short period of time.

Relax five minutes before each measurement.

The value is different from that measured at a clinic or doctor’s office.

See Pg. E-18 "Answers to Why Your Readings Are Different Between Home and the Doctor’s Office."

The healthcare professional may be using a different sized cuff.

Your measurements may be elevated by white coat hyperten sion

See Pg. E-17 "Why Measure Blood Pressure at Home."

Page 24 of the LifeSource UA-853 user manual — SPECIFICATIONS
Page 24 · click to enlarge
Page 24

SPECIFICATIONS

Model UA-853 Type Oscillometric Display Digital, 20-mm character height

Pressure/pulse displayed simultaneously Time/Date 12 hour format (AM/PM),

year (2006-2056), month, and day with automatic adjustment for leap years Memory 90 readings Measurement range Pressure: 20 mmHg to 280 mmHg

Pulse: 40 pulses to 180 pulses Accuracy Pressure: ±3 mmHg or ±2%,

whichever is greater Pulse: ±5% Pressurization Automatic, using micropump Depressurization Active exhaust valve Power source 4 type “AA” (1.5 volt) alkaline

batteries (not included) or 120 V AC Adapter Battery life Approximately 4 months with

one daily measurement Operating environment 50ºF to 104ºF (10ºC to 40ºC)

Less than 85% relative humidity Storage environment -14ºF to 140ºF (-10ºC to 60ºC)

Less than 95% relative humidity Dimensions Length: 4.9”(126 mm)

Width: 5.9”(150 mm) Height: 6.1” (156 mm) Weight 12.3 oz. (350 g) without batteries

Blood pressure measurements determined by the UA-853 are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute for electronic or automated sphygmomanometers.

Page 25 of the LifeSource UA-853 user manual — This LifeSource® blood pressure monitor is covered by a Lifetime
Page 25 · click to enlarge
Page 25

CONTACT INFORMATION

Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. Our full product line includes:

Digital Blood Pressure Monitors Manual Blood Pressure Monitors Blood Pressure Cuffs Stethoscopes Personal Health Scales Digital Thermometers

This LifeSource® blood pressure monitor is covered by a Lifetime

Warranty. See warranty card for details.

Visit our web site at www.LifeSourceOnline.com

for warranty registration.

For more information regarding use, care or servicing of your blood pressure monitor, contact:

A&D Medical A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 LifeSource Health Line (Toll-Free): 1-888-726-9966 www.LifeSourceOnline.com

For Canadian Residents, please contact: Auto Control Medical 6695 Millcreek Drive, Unit 5 Missisauga, Ontario L5N 5R8 Canada Auto Control (Toll-Free): 1-800-461-0991 www.autocontrol.com

Page 26 of the LifeSource UA-853 user manual — LifeSource ® Customer Service Line
Page 26 · click to enlarge
Page 26

IMPORTANT!

If You Need Assistance with the Set-Up or Operation

We Can Help!

Please call us FIRST before contacting your retailer at

LifeSource ® Customer Service Line

1-888-726-9966 - Toll Free

A specially trained representative will assist you.

A division of A&D Engineering, Inc. 1555 McCandless Drive Milpitas, CA 95035 www.LifeSourceOnline.com

Pressure Rating Indicator is a trademark of A&D Medical and is patent pending.

TimeWise™ is a trademark of A&D Medical.

LifeSource® is a registered trademark OF A&D Medical.

©2006 A&D Medical. All rights reserved.

Specifications subject to change without notice.

I-MAN:853AC Printed in China 5/06

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Nothing appears in the display, after I press the START button.Batteries are drained.Replace all batteries with new ones. Alkaline batteries recom­ mended. p.23
Nothing appears in the display, after I press the START button.Battery terminals are not in the correct position.Reinstall the batteries with nega­ tive and positive terminals match­ ing those indicated in the battery compartment. p.23
The cuff does not inflate.Battery voltage is too low. Low battery symbol blinks. [If the batteries are drained completely, the mark does not appear.]Replace all batteries with new ones. Alkaline batteries recom­ mended. p.23
The unit does not measure. Readings are too high or too low.The cuff is not fastened prop­ erly.Fasten the cuff correctly. p.23
The unit does not measure. Readings are too high or too low.You moved your arm or body during the measurement.Make sure you remain very still and quiet during the measurement. p.23
The unit does not measure. Readings are too high or too low.The cuff position is not correct.Sit comfortably and still. Make sure the cuff is at the same level as your heart. p.23
The unit does not measure. Readings are too high or too low.You are using the wrong size cuff.See Pg. E-5 "Selecting the Correct Cuff." p.23
The unit does not measure. Readings are too high or too low.Taking too many readings on the same arm in a short period of time.Relax five minutes before each measurement. p.23
The value is different from that measured at a clinic or doctor’s office.The healthcare professional may be using a different sized cuff.See Pg. E-18 "Answers to Why Your Readings Are Different Between Home and the Doctor’s Office." p.23
The value is different from that measured at a clinic or doctor’s office.Your measurements may be elevated by white coat hyperten­ sionSee Pg. E-17 "Why Measure Blood Pressure at Home." p.23

Ask about the LifeSource UA-853

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the LifeSource UA-853

Be the first to ask — other owners or our team will answer.