Blood Pressure Monitor

English l Español

Please read this important information before using your monitor. ☞ Please remember that only a medical practitioner is qualified to interpret your blood pressure measurements. Use of this device should not replace regular medical examinations. ☞ Consult your physician if you have any doubt about your readings. Should a mechanical problem occur, contact LifeSource® at 1-888-726-9966. ☞ Do not attempt to service, calibrate, or repair this monitor. ☞ Because your UA-851 monitor contains delicate, high-precision parts, avoid exposing it to extremes in temperature or humidity or to direct sunlight, shock and dust. LifeSource® guarantees the accuracy of this monitor only when it is stored and used within the temperature and humidity ranges noted on page E-21. ☞ Clean the monitor and cuff with a dry, soft cloth or a cloth dampened with water and a mild detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean monitor or cuff. ☞ Remove and replace batteries if monitor is not used for more than six months. Alkaline batteries recommended.
PRECAUTIONS The UA-851 is designed to be used at home, by those who are eighteen (12) years and older, to monitor blood pressure (systolic and diastolic) and pulse rate.
CAREFULLY FOLLOW THESE INSTRUCTIONS. ☞ For connection to a supply not in the U.S.A., use an attachment plug adapter of the proper configuration for the power outlet. ☞ The power unit (AC Adapter) is intended to be correctly oriented in a vertical or floor mount position.

Fast. Easy. Accurate.
Congratulations on purchasing a state-of-the-art LifeSource® blood pressure monitor—one of the most technologically advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen.
Physicians agree that daily self-monitoring of blood pressure is an important step individuals can take to maintain cardiovascular health and prevent the serious consequences associated with undetected and untreated hypertension.
A&D Medical has been manufacturing high quality healthcare products for over 20 years. We are committed to providing you and your family with monitoring devices that provide the utmost in accuracy and convenience.
LifeSource - Designed for Life.





Display

LifeSource® blood pressure monitors are easy to use, accurate and digitally display full measurement readouts. Our technology is based on the “oscillometric method” – a noninvasive blood pressure determination. The term “oscillation” refers to any measure of vibrations caused by the arterial pulse. Our monitor examines the pulsatile pressure generated by the arterial wall as it expands and contracts against the cuff with each heartbeat.
The cuff is inflated until the artery is fully blocked. The inflation speed is maximized and the pressure level is optimized by the device. The monitor takes measurements while the cuff is both inflating and deflating. This results in a faster measurement providing greater comfort to the user.
You must install 4 type AA (1.5 volt) batteries (included), or use the optional AC Adapter (see page E-14 for using the Monitor with AC Adapter) and attach the cuff to the monitor before using it. To install batteries (or replace them if the “Low Battery” symbol appears on display), proceed as follows:
Step 2 · Step 1 · Step 3

Note: If an appropriate pressure is not obtained, the device automatically starts to inflate again. Note: If you wish to stop inflation at any time, press the START button again.
DIASTOLIC
PULSE


Measurement with the desired systolic pressure: If your systolic pressure is expected to exceed 230 mmHg follow these steps:

The UA-851 Blood Pressure Monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. The Irregular Heartbeat symbol will appear in the display window in the event an Irregular Heartbeat has occured during measurement. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeat intervals during the blood pressure measurement. It is important that you relax, remain still and refrain from talking during measurements.
The Cuff Inflation Meter is located on the left side of the display screen to tell you when the blood pressure monitor is inflating and deflating the cuff. The Cuff Inflation Meter moves up during inflation and moves down during deflation.
Inflation in Progress Inflation Complete Deflation/Measurement in
Progress ·

The UA-851 provides an average blood pressure reading based on the total measurements stored in memory. The average blood pressure reading appears briefly when you recall the measurements stored in memory. You can also access your average readings by pressing and holding the START button.
number of measurements stored in
The Pressure Rating Indicator™ is a feature which provides a snapshot of your blood pressure classification based on your measurements. This will let you quickly know what your blood pressure readings mean. Each segment of the bar indicator corresponds to the Seventh Report of the Joint National Committee (JNC7) on Prevention, Evaluation and Treatment of High Blood Pressure from the National Heart, Lung and Blood Institute - May 2003. For a more detailed look at this blood pressure classification, please refer to “Assessing High Blood Pressure” on page E-16.
Stage 2 Hypertension Stage 1 Hypertension Prehypertension Normal
:The indicator displays a segment, based on the current measurement, corresponding to the JNC7 Classification.

Press and hold the
Average of all readings

The UA-851 has an AC adapter jack to allow you to supply power from an outlet in your home. We recommend you use only the optional LifeSource AC adapter to avoid potential damage to the monitor. To purchase the optional AC Adapter, please contact our LifeSource Health Line at 1-888-726-9966.
Connecting the AC adapter to the monitor: 1. Gently insert the AC adapter plug into a 120V AC outlet.
Disconnecting the AC adapter from the monitor: 1. Turn the unit off by pressing the START button.
Jack · AC Adapter Plug

What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).
What Affects Blood Pressure? Blood pressure is affected by many factors: age, weight, time of day, activity level, climate, altitude and season. Certain activities can significantly alter one’s blood pressure. Walking can raise systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg. Sleeping can decrease systolic blood pressure by as much as 10 mmHg. Taking your blood pressure repeatedly without waiting an interval of 5 minutes between readings, or without raising your arm to allow blood to flow back to the heart, can also affect it.
In addition to these factors, beverages containing caffeine or alcohol, certain medications, emotional stress and even tight-fitting clothes can make a difference in the readings.
What Causes Variations In Blood Pressure? An individual’s blood pressure varies greatly from day to day and season to season. For hypersensitive individuals, these variations are even more pronounced. Normally, blood pressure rises during work or play and falls to its lowest levels during sleep.

Assessing High Blood Pressure for Adults The following standards for assessing high blood pressure (without regard to age or gender) have been established as a guideline. Please note that other risk factors (e.g. diabetes, obesity, smoking, etc.) need to be taken into consideration and may affect these figures. Consult with your physician for an accurate assessment.
JNC7 Classification Table – for adults within the U.S.
BP Classification Systolic (mmHg) Diastolic (mmHg)
Normal < 120 and < 80
Prehypertension 120-139 or 80-89
Stage 1 Hypertension 140-159 or 90-99
Stage 2 Hypertension ≥ 160 or ≥ 100
SOURCE: The Seventh Report of the Joint National Committee on Prevention, Evaluation and Treatment of High Blood Pressure for Adults. National Heart, Lung and Blood Institute - May 2003.
WHO Classification Table - for adults outside of the U.S. (e.g. Canada, Mexico)
BP Classification Systolic (mmHg) Diastolic (mmHg)
Optimal < 120 and < 80
Normal < 130 or < 85
High-Normal 130-139 or 85-89
Stage 1 Hypertension 140-159 or 90-99
Stage 2 Hypertension 160-179 or 100-109
Stage 3 Hypertension ≥ 180 or ≥ 110
SOURCE: Standards to assess high blood pressure, without regard to age or gender, have been established by the World Health Organization (WHO).

What Is Hypertension? Hypertension (high blood pressure) is the diagnosis given when readings consistently rise above normal. It is well known that hypertension can lead to stroke, heart attack or other illness if left untreated. Referred to as a “silent killer” because it does not always produce symptoms that alert you to the problem, hypertension is treatable when diagnosed early.
Can Hypertension Be Controlled? In many individuals, hypertension can be controlled by altering lifestyle and minimizing stress, and by appropriate medication prescribed and monitored by your doctor. The American Heart Association recommends the following lifestyle suggestions to prevent or control hypertension:
Don’t smoke. Have regular physical checkups
Reduce salt and fat intake. Maintain proper weight.
Monitor your blood Exercise routinely. pressure at periodic intervals.
Why Measure Blood Pressure at Home? It is now well known that, for many individuals, blood pressure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. This response is commonly called “white coat hypertension.” In any case, self-measurement at home supplements your doctor’s readings and provides a more accurate, complete blood pressure history. In addition, clinical studies have shown that the detection and treatment of hypertension is improved when patients both consult their physicians and monitor their own blood pressure at home.

Answers to Why Your Readings Are Different Between Home and at the Doctor’s Office
Why are my readings different between home and at a doctor’s office? Your blood pressure readings taken in a doctor’s office or hospital setting may be elevated as a result of apprehension and anxiety. This response is known as white coat hypertension.
When I bring my monitor to a doctor’s office, why do I get a different measurement from my monitor to that taken by a doctor or nurse? The healthcare professional may be using a different sized cuff. The size of the bladder inside the cuff is critical for the accuracy of the measurement. This may give you a different reading. A cuff too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small will produce a reading that is higher than the correct blood pressure. There may also be other factors that can cause the difference in measurements.
Keys to Successful Monitoring: Blood pressure fluctuates throughout the day. We recommend that you are consistent in your daily measurement routine:

Establishing Baseline Measurements The most important method to get an accurate blood pressure measurement is consistency. To get the most benefit out of your monitor, it is important to establish a “baseline measurement.” This helps build a foundation of measurements that you can use to compare against future readings. To build this baseline measurement, devote two weeks for consistent blood pressure monitoring. This involves doing everything the same way when you measure (e.g. measuring during the same time of day, in the same location, sitting in the same chair, using the same cuff, etc.). Once you establish your baseline measurement, you can start evaluating if your measurement has been affected based on things lifestyle changes or medication treatment.
How Do I Record My Blood Pressure? Blood pressure readings are typically recorded with the systolic pressure written first, followed by a slash mark and the diastolic pressure. For example, 120 mmHg systolic and 80 mmHg diastolic measurements are written as 120/80. Pulse is simply written with the letter “P” followed by the pulse rate—P 72, for example. Please see the back of the manual for the blood pressure tracking log.
Customer Support Tools Online Additional tools are available on www.LifeSourceOnline.com to help you get the most out of your blood pressure monitoring. These include:

Problem Probable Cause Corrective Action
Nothing appears in the display, after I press the START button.
Batteries are drained.
Replace all batteries with new ones. Alkaline batteries recom- mended.
Battery terminals are not in the correct position.
Reinstall the batteries with nega- tive and positive terminals match- ing those indicated in the battery compartment.
Battery voltage is too low. Low battery symbol blinks. [If the batteries are drained completely, the mark does not appear.]
Replace all batteries with new ones. Alkaline batteries recom- mended.
The cuff does not inflate.
The cuff is not fastened prop- erly.
Fasten the cuff correctly.
The unit does not measure. Readings are too high or too low.
You moved your arm or body during the measurement.
Make sure you remain very still and quiet during the measurement.
The cuff position is not correct.
Sit comfortably and still. Make sure the cuff is at the same level as your heart.
You are using the wrong size cuff.
See Pg. E-5 "Selecting the Correct Cuff."
Taking too many readings on the same arm in a short period of time.
Relax five minutes before each measurement.
The value is different from that measured at a clinic or doctor’s office.
See Pg. E-18 "Answers to Why Your Readings Are Different Between Home and the Doctor’s Office."
The healthcare professional may be using a different sized cuff.
Your measurements may be elevated by white coat hyperten- sion
See Pg. E-17 "Why Measure Blood Pressure at Home."

Model UA-851 Type Oscillometric Display Digital, 20-mm character height
Pressure/pulse displayed simultaneously Memory 90 readings Measurement range Pressure: 20 mmHg to 280 mmHg
Pulse: 40 pulses to 180 pulses Accuracy Pressure: ±3 mmHg or ±2%,
whichever is greater Pulse: ±5% Pressurization Automatic, using micropump Depressurization Active exhaust valve Power source 4 type “AA” (1.5 volt) alkaline
batteries (included) or 120 V AC Adapter (optional) Battery life Approximately 4 months with
one daily measurement Operating environment 50ºF to 104ºF (10ºC to 40ºC)
Less than 85% relative humidity Storage environment -14ºF to 140ºF (-10ºC to 60ºC)
Less than 95% relative humidity Dimensions Length: 4.7”(120 mm)
Width: 5.9”(150 mm) Height: 6.1” (156 mm) Weight 17.6 oz. (500 g) without batteries
Blood pressure measurements determined by the UA-851 are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute for electronic or automated sphygmomanometers.

Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. Our full product line includes:
Digital Blood Pressure Monitors Manual Blood Pressure Monitors Blood Pressure Cuffs Stethoscopes Personal Health Scales Digital Thermometers
Warranty. See warranty card for details.
Visit our web site at www.LifeSourceOnline.com
for warranty registration.
For more information regarding use, care or servicing of your blood pressure monitor, contact:
A&D Medical A division of A&D Engineering, Inc. 1756 Automation Parkway San Jose, CA 95131 LifeSource Health Line (Toll-Free): 1-888-726-9966 www.LifeSourceOnline.com

other LifeSource® products.
acerca de los productos LifeSource®.








If You Need Assistance with the Set-Up or Operation
Please call us FIRST before contacting your retailer at
1-888-726-9966 - Toll Free
A specially trained representative will assist you.
A division of A&D Engineering, Inc. 1756 Automation Parkway San Jose, CA 95131 www.LifeSourceOnline.com
Pressure Rating Indicator is a trademark of A&D Medical and is patent pending.
TimeWise™ is a trademark of A&D Medical.
LifeSource® is a registered trademark OF A&D Medical.
©2006 A&D Medical. All rights reserved.
Specifications subject to change without notice.
I-MAN:853AC Printed in China 5/06
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Nothing appears in the display, after I press the START button. | Batteries are drained. | Replace all batteries with new ones. Alkaline batteries recom- mended. | p.21 |
| Nothing appears in the display, after I press the START button. | Battery terminals are not in the correct position. | Reinstall the batteries with nega- tive and positive terminals match- ing those indicated in the battery compartment. | p.21 |
| The cuff does not inflate. | Battery voltage is too low. Low battery symbol blinks. [If the batteries are drained completely, the mark does not appear.] | Replace all batteries with new ones. Alkaline batteries recom- mended. | p.21 |
| The unit does not measure. Readings are too high or too low. | The cuff is not fastened prop- erly. | Fasten the cuff correctly. | p.21 |
| The unit does not measure. Readings are too high or too low. | You moved your arm or body during the measurement. | Make sure you remain very still and quiet during the measurement. | p.21 |
| The unit does not measure. Readings are too high or too low. | The cuff position is not correct. | Sit comfortably and still. Make sure the cuff is at the same level as your heart. | p.21 |
| The unit does not measure. Readings are too high or too low. | You are using the wrong size cuff. | See Pg. E-5 "Selecting the Correct Cuff." | p.21 |
| The unit does not measure. Readings are too high or too low. | Taking too many readings on the same arm in a short period of time. | Relax five minutes before each measurement. | p.21 |
| The value is different from that measured at a clinic or doctor’s office. | The healthcare professional may be using a different sized cuff. | See Pg. E-18 "Answers to Why Your Readings Are Different Between Home and the Doctor’s Office." | p.21 |
| The value is different from that measured at a clinic or doctor’s office. | Your measurements may be elevated by white coat hyperten- sion | See Pg. E-17 "Why Measure Blood Pressure at Home." | p.21 |