EN DeVilbiss ® Suction Unit Instruction Guide 7305 Series

Rx ONly Made in U.S.A.

ES Guía de Instrucciones para la Unidad de Succión de la Serie 7305 de DeVilbiss®

Rx úNIcamENtE Hecho en EE.UU.

FR Guide d’Instructions l’Unite d’aspiration DeVilbiss ® Série 7305

UNIqUEmENt SUR pREScRIptION Fabriqué aux U.S.A.

Page 2 of the DeVilbiss Suction Unit series (7305) user manual — Approved
Page 2 · click to enlarge
Page 2
Page 2 of the DeVilbiss Suction Unit series (7305) user manual — Approved
Page 3 of the DeVilbiss Suction Unit series (7305) user manual — Approved
Page 3 · click to enlarge
Page 3

d. It has been dropped into water. Return the product to an authorized DeVilbiss Healthcare service center for examination and repair. 4. Keep the power cord away from heated surfaces. 5. Never use while drowsy or asleep. NOTE– The 7305D series is not factory equipped with an internal rechargeable battery; it may be purchased separately and installed by your DeVilbiss Healthcare provider. 7305P series is factory equipped with an internal rechargeable battery and all information regarding battery operation in this guide is applicable.

1. All suctioning should be done in strict accordance with appropriate procedures that have been established by a licensed medical authority. 2. Some attachments or accessories may not fit the tubing supplied. All attachments or accessories should be checked prior to use to assure proper fit.

The 7305 series is equipped with a switch mode power supply allowing operation on any AC voltage (100-240 VAC, 50/60 Hz). However the correct power cord must be used to connect to adaptable wall power. NOTE– Check power cord for adaptability before using.

Your DeVilbiss Suction Unit is a compact medical suctioning device which has been designed for reliable, portable operation. Because of the small size, light weight and DC operation, the DeVilbiss Suction Unit is ideal for providing suction in the home, in transport with optional DC cord, or if your model has an internal rechargeable battery, the unit can be operated anywhere. Three container options give the choice between the standard disposable containers or an optional long-term reusable container. Following the recommended operating and maintenance procedures outlined in this instruction guide will maximize the life of this product.

  • contraindications The DeVilbiss Suction Unit should not be used for:
  • thoracic drainage
  • nasogastric suction

Intended Use Statement The device is to be used to remove fluids from the airway or respiratory support system and infectious materials from wounds.The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection bottle. The fluids are trapped in the collection bottle for proper disposal. It is for use on the order of a physician only.

The following items can be purchased separately as accessories or replacement items for your 7305 Series DeVilbiss Suction Unit:

Descriptionpart No.Descriptionpart No. 6’ patient tubing6305D-611Collection Container Kit (1200 ml reusable container, external bacteria filter, elbow, 43/8” tubing)7314D-603 800 Disposable Container (container and lid only) for use with external bacteria filter (48 pack)7305D-6021200 ml reusable container (external bacteria filter, elbow, 43/8” tubing) (6 pack)7314D-604 Collection Container Kit (800 ml disposable container, external bacteria filter, elbow, 3 3/8” and 6’tubing package)7305D-603Carrying case7305D-606 External bacteria filter (non-sterile) (12 pack) For reusable container and disposable container with external filter7305D-608AC to DC adapter/charger7305P-613 Collection Container Kit (Internal filter cartridge, splash guard, 800 ml container, 43/8” and 6’ tubing package)7305D-63312V DC power cord (1 each)7304D-619 800 ml disposable container with internal filter cartridge, splash guard and 43/8” tubing (48 each)7305D-632Hospital grade power cord (120 VAC)099HD-614 Filter cartridge (12 pack) (For Disposable Container with internal filter)7305D-635Power cord for US6710D-609 Splash Guard (12 pack) for Disposable Container with internal filter cartridge7305D-641Power cord for Continental Europe7305P-631

INtERNatIONal tRaVEl · INtRODUctION · accESSORy/REplacEmENt ItEmS

Page 4 of the DeVilbiss Suction Unit series (7305) user manual — Disposable Container with Internal Filter Cartridge and Splash Guard
Page 4 · click to enlarge
Page 4

7305 Series DeVilbiss Suction Unit

With Disposable Container and Internal Filter Cartridge

With Disposable Container and External Bacteria Filter

With Reusable Container and External Bacteria Filter

1. External bacteria filter 2. Vacuum gauge 3. 4 3/8 ” connection tubing 4. Vacuum regulator knob 5. DC power input (on side) 6. Power switch 7. Collection container 8. Lid 9. Patient tubing 10. Connection elbow AC to DC Adapter (not shown)
1. 4 3/8 ” connection tubing 2. Vacuum gauge 3. Vacuum regulator knob 4. DC power input (on side) 5. Power switch 6. Splash guard 7. Collection container 8. Lid 9. Patient tubing 10. Filter cartridge AC to DC Adapter (not shown)
1. 3 3/8 ” connection tubing 2. Vacuum gauge 3. Vacuum regulator knob 4. DC power input (on side) 5. Power switch 6. Collection container 7. Lid 8. Patient tubing 9. Connection elbow 10. External bacteria filter AC to DC Adapter (not shown)

DC power cord (not shown) optional Internal rechargeable battery (not shown) 7305P series only Carrying case (not shown) 7305P series only

DC power cord (not shown) optional Internal rechargeable battery (not shown) 7305P series only Carrying case (not shown) 7305P series only

DC power cord (not shown) optional Internal rechargeable battery (not shown) 7305P series only Carrying case (not shown) 7305P series only

DeVilbiss Suction Unit series (7305) — With Disposable Container and Internal Filter Cartridge diagram (page 4)
Fig. 2 · With Disposable Container and Internal Filter Cartri…
DeVilbiss Suction Unit series (7305) — With Disposable Container and Internal Filter Cartridge diagram (page 4)
Fig. 3 · With Disposable Container and Internal Filter Cartri…
DeVilbiss Suction Unit series (7305) — With Disposable Container and Internal Filter Cartridge diagram (page 4)
Fig. 4 · With Disposable Container and Internal Filter Cartri…

7305 Series DeVilbiss containers

Disposable Container with Internal Filter Cartridge and Splash Guard

Disposable Container with External Bacteria Filter

Reusable Container with External Bacteria Filter

1. 4 3/8 ” connection tubing 2. Filter cartridge (Do not get wet) 3. Lid 4. Jar 5. Splash guard 6. Patient tubing connector
1. 3 3/8 ” connection tubing 2. Patient tubing connector 3. Jar 4. Connection elbow 5. Bacteria filter
1. 4 3/8 ” connection tubing 2. Lid with o-ring 3. Overflow valve 4. Jar 5. Patient tubing connector 6. Connection elbow 7. Bacteria filter
DeVilbiss Suction Unit series (7305) — Disposable Container with Internal Filter Cartridge and Splash Guard diagram (page 4)
Fig. 5 · Disposable Container with Internal Filter Cartridge…
DeVilbiss Suction Unit series (7305) — Disposable Container with Internal Filter Cartridge and Splash Guard diagram (page 4)
Fig. 6 · Disposable Container with Internal Filter Cartridge…
DeVilbiss Suction Unit series (7305) — Disposable Container with Internal Filter Cartridge and Splash Guard diagram (page 4)
Fig. 7 · Disposable Container with Internal Filter Cartridge…
Page 5 of the DeVilbiss Suction Unit series (7305) user manual — With Disposable Container and Internal Filter Cartridge
Page 5 · click to enlarge
Page 5

Internal Filter cartridge Set-Up

Disposable Container w/Internal Filter Cartridge & Splash Guard

1. Fully charge battery for 17 hours (7305P series only). 2. Ensure splash guard is securely attached to inside of lid over filter cartridge. 3. Securely attach lid to container. 4. Insert container into holder. 5. Attach 43/8” tubing from filter cartridge to tubing connector. 6. The patient tubing should be connected to the container lid at the outlet labeled . 7. Please assure that all connections are secure and without leaks before using. 8. Verify that unit is at desired suction level before beginning patient suction.

Disposable Container or Reusable Container with External Bacteria Filter

1. Fully charge battery for 17 hours (7305P series only). 2. Securely attach lid to container. 3. Insert container into holder. 4. Connect either end of the 43/8” reusable tubing or 33/8 ” disposable tubing to the tubing connector then connect the other end to the bacteria filter. Ensure that the clear side of the bacteria filter is toward elbow and container when installing/re-installing. Do not reverse direction of filter. 5. The bacteria filter should then be connected to the 90˚ elbow connection, and the 90˚ connection should be connected to the top of the container lid where it says . 6. The patient tubing should be connected to the container lid at the outlet labeled . 7. Please assure that all connections are secure and without leaks before using. 8. Verify that unit is at desired suction level before beginning patient suction.

HOw tO OpERatE yOUR 7305 SERIES DEVIlbISS SUctION UNIt

Before connecting the unit to the AC adapter or 12V DC cord, make sure that the power switch located on the side of the unit is in the “Off” position. Select power source desired.

lED Explanations: L1- Green– External power supplied to unit from AC power source or DC cord. Illuminated when external power is supplied. L2 - Yellow– Battery is being charged. Light will go out when battery is fully charged. (7305P Series only) L3 - Red– Low battery. Seek another power source and charge battery as soon as possible when light remains on continuous. (7305P Series only)

AC OPERATION– Plug the small connector of the AC adapter into the DC power input located on the side of the unit. Plug the AC end into a grounded wall-outlet power source.

12V DC OPERATION– (such as a car lighter receptacle). Plug the small connector of the DC power cord into the DC power input on the side of the unit. Plug the large connector into the 12V DC power receptacle of the automobile.

BATTERy OPERATION– Verify that your unit has an internal rechargeable battery; factory installed on models 7305P series or provider installed on models 7305D series (installed as an option). To ensure proper operation from internal battery, fully charge the battery for 17 hours as explained in the Battery Charging section. To operate the unit from an internal rechargeable battery, ensure that no external power sources are plugged into the DC power input on the side of the unit.

Turn the unit “On” using the power switch located on the side of the unit. The power indicator light displayed on the top of the unit will indicate which power source is being used by staying continuously lit when external power is connected.

Adjust the vacuum level from 80 to 550 mm Hg by turning the vacuum regulator knob located on the side of the unit (clockwise to increase vacuum and counter-clockwise to decrease vacuum). The gauge located on the top of the unit near the handle will allow you to select a specific level of vacuum. To accurately read the gauge, block the patient end of the hose or cap off the collection bottle and allow the gauge to reach a stable vacuum reading.

Suction at the patient is automatically obstructed when liquid level reaches the float shut-off located on the underside of the container lid.

SEt-Up · External Filter Set-Up

Page 6 of the DeVilbiss Suction Unit series (7305) user manual — b
Page 6 · click to enlarge
Page 6

On 7305P series, the units are equipped with a factory-installed rechargeable battery. The unit will have a light for low battery and charge indication. On 7305D series, the units are not factory equipped with a rechargeable battery; check with your equipment provider to determine if your unit has been upgraded with a rechargeable battery.

When you connect the unit to an AC or DC power source using the AC adapter or DC power cord, the green external power light will illuminate. The yellow charge indicator will be illuminated while the battery is charging. Verify that this illuminates when charging begins. As the charge nears a full charge, the yellow LED may flash on and off for several minutes. This is normal.

STORAGE NOTE - Battery should be charged for a minimum of 17 hours prior to storage, and at least once every 6 months. Important - If battery recharge is delayed beyond 6 months, battery may be able to provide full run time after completion of 3 full charge and discharge cycles.

1. Shut off unit using power switch and allow vacuum to drop. Disconnect power source from the DC input receptacle on the unit. 2. Disconnect tubing and remove container from holder. 3. Carefully remove lid and empty contents.

a

DeVilbiss Suction Unit series (7305) — a diagram (page 6)
Fig. 8 · a

Disposable collection container with Internal Filter cartridge

The disposable collection container and lid are meant for single-patient use.

1. Remove filter cartridge and 4-3/8” tubing and set aside. NOTE – Filter MUST NOT get wet. The filter material cannot be removed from the elbow (figure A).
2. Wash container, lid and splash guard in warm water/dishwashing solution. Rinse with clean, warm water. 3. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131˚F-149ºF/55˚C-65˚C) solution for 60 minutes. Rinse with clean, warm water and air dry.
DeVilbiss Suction Unit series (7305) — a diagram (page 6)
Fig. 9 · a

b

Disposable collection container with External bacteria Filter

1. Remove bacteria filter, 3-3/8” tubing, and connection elbow and set aside. 2. Wash jar and lid in a solution of warm water with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water. 3. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131˚F-149ºF/55˚C-65˚C) solution for 60 minutes. Rinse with clean, warm water and air dry.

Reusable collection container with External bacteria Filter

1. Remove bacteria filter, 4-3/8” tubing, and connection elbow and set aside. Remove o-ring and overflow valve from lid. 2. Wash jar, lid, o-ring, and overflow valve in a solution of warm water with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water. 3. For single patient use: After washing, disinfect using one of the following methods: a. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) solution for 60 minutes. Rinse with clean, warm water and air dry in a clean environment. b. Wash with rubbing alcohol and air dry in a clean environment. c. Wash with a commercial (bacterial-germicidal) disinfectant. Follow disinfectant manufacturer’s recommended dilution rates and instructions carefully.

battERy cHaRGING · clEaNING INStRUctIONS · preparation

Page 7 of the DeVilbiss Suction Unit series (7305) user manual — With Disposable Container and Internal Filter Cartridge
Page 7 · click to enlarge
Page 7
4. For multi-patient use: After washing: a. After parts are completely dry, place jar and lid in autoclave with the open end down. Ensure parts are not touching. Run one sterilization steam cycle at 250°F (121°C) for 15 minutes. NOTE - Jar is guaranteed up to 30 cycles of autoclave sterilization at the indicated conditions. b. Dispose of and replace filter, tubing and elbow between patients.

Suction Unit (single-patient use)

1. With the power switch in the “Off” position, disconnect the DeVilbiss Suction Unit from all external power sources. 2. Wipe the housing with a clean cloth and any commercial (bacterial-germicidal) disinfectant.
1. Disconnect the tubing from the unit. 2. Rinse thoroughly by running warm tap water through it. 3. Follow by soaking in a solution of 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) for 60 minutes. Rinse with clean, warm water and air dry. 4. Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.

carrying case (single-patient use)

1. Wipe the case using a clean cloth dampened with detergent and/or disinfectant.

Suction Unit (multi-patient use)

Device Cleaning and Disinfection When There is a Patient Change When medical devices have already been used with a patient, contamination with human pathogenic germs should be assumed (unless there is evidence to the contrary), and the next patient, user or third party should be protected by appropriate handling and preparation. Therefore, when there is a patient change, people must be protected during the transport and handling of the device and the device must be fully processed, i.e., cleaned and disinfected, by suitably trained personnel before reuse to protect the next patient. The complete processing may only be done by the manufacturer or by a qualified DeVilbiss provider/service technician.

DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualified third party between uses by different patients. 1. Dispose of all accessory components that are not suitable for reuse, i.e., collection container, filter, tubing and carrying case. 2. With the power switch in the “Off” position, disconnect the DeVilbiss Suction Unit from all external power sources. 3. Visually inspect unit for any damage, missing parts, etc. 4. Wipe the housing with a clean cloth and a commercial (bacterial-germicidal) disinfectant that meets the requirements listed in the NOTE below and is used as per the disinfectant manufacturer’s recommended dilution rates and instructions. CAUTION – Do not submerge in water as this will result in damage to the vacuum pump.

Inspect suction tubing and container for leaks, cracks, etc. before each use.

changing Filter cartridge (single-patient use) Disposable container

1. Change filter cartridge if overflow occurs or every two months, whichever comes first. 2. Turn unit “off”. 3. Remove filter cartridge and 4 3/8 ” tubing. 4. Install new cartridge and tubing. NOTE – Do not substitute any other material for this filter cartridge. Substitution may lead to contamination or poor performance; use only DeVilbiss filter cartridges.

changing bacteria Filter (single-patient use) Reusable container and/or Disposable container

1. Change bacteria filter if overflow occurs or every two months, whichever comes first. 2. Remove filter by disconnecting it from suction unit and lid assembly. 3. Replace with a clean DeVilbiss bacteria filter (non-sterile) and remount to suction unit and lid. Ensure that the clear side of the bacteria filter is toward elbow and container when installing/re-installing. Do not reverse direction of filter. Additional filters (7305D-608 12/pack) may be purchased from your authorized DeVilbiss Healthcare provider. NOTE– Do not substitute any other material for this bacteria filter. Substitution may lead to contamination or poor performance; use only DeVilbiss filters. NOTE– Bacteria filter must be changed between patients.

tubing (single-patient use) · maINtENaNcE

Page 8 of the DeVilbiss Suction Unit series (7305) user manual — With Disposable Container and Internal Filter Cartridge
Page 8 · click to enlarge
Page 8
pROblEmactION Unit does not turn on, but green external power light is illuminated.1. Check power sources and connections. 2. Ensure wall outlet is live by plugging in a lamp. 3. If running from an internal battery, ensure that your unit has a battery installed. 4. If battery is installed, check that it is fully charged. Pump runs, but there is no vacuum.1. Check that all tubing is connected properly. 2. Check tubing connections for breaks or leaks. 3. Ensure that float shut-off is not activated. 4. Check for leaks or cracks in container assembly. Low vacuum.1. Use vacuum adjustment knob to increase vacuum level. 2. Check system for leaks. 3. Push vacuum adjustment knob in toward unit and then release. Battery will not charge but charge light is illuminated.1. Ensure that unit is equipped with an internal battery by contacting your DeVilbiss equipment provider. 2. Verify that charge light turns on. 3. Check electrical connections during charging. 4. Ensure wall outlet is live by plugging in a lamp.
SizeH x w x D inches (cm) 7305P & 7305D Series9.0 x 7.0 x 8.0 (22.9 x 17.8 x 20.3) weightlb. (kg) 7305D Series3.8 (1.7) 7305P Series6.3 (2.9) Electrical Requirements100-240 V AC 50/60 Hz .75 A max; 12 V DC, 33 W max Internal Rechargeable battery 7305P SeriesFactory Equipped 7305D SeriesNot Factory Equipped (provider-installed option) Vacuum Range 7305P & 7305D Series80 to 550 mm Hg Air Flow @ pump inlet:27 LPM (free flow) typical (may be less when running from internal battery) container capacity 7305 D & P Series800 ml (cc) Disposable (2 filter options) 7305 D & P Series - Optional1200 ml (cc) Reusable Environmental conditions Operating Temperature Range32°F (0°C) - 104°F (40°C) Operating Relative Humidity0-95% Operating Atmospheric Pressure10.2 psi (70 kPa) - 15.4 psi (106 kPa) Storage & Transport Temperature Range-40°F (-40°C) - 158°F (70°C) Storage & Transport Relative Humidity0-95% Storage & Transport Atmospheric Pressure7.3 psi (50 kPa) - 15.4 psi (106 kPa) warranty 7305P SeriesTwo-years limited, excluding internal battery and container 7305D SeriesTwo-years limited, excluding container Internal Battery90-day approvals 7305D & 7305P SeriesIEC 601-1; CAN/CSA-C22.2 No. 601.1-M90; UL 60601-1, EN 60601-1-2 7305P meets RTCA/D-160DDO-160D - section 21 Category M For battery operation only Airline use Equipment Classifications With respect to protection from electric shockClass I and internally powered Degree of protection against electric shockType BF Applied Parts Degree of protection against ingress of liquidsIPX2 and ordinary power supply Mode of OperationIntermittent Operation: 30 minutes on, 30 minutes off Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. ISO Classification 7305P Series only - Electrically powered medical suction equipment for field and transport use according to ISO 10079-1 : 1999 High Flow/High Vacuum 7305D Series - Electrically powered medical suction equipment for non-transport use according to ISO 10079-1 : 1999

tROUblESHOOtING · SpEcIFIcatIONS/claSSIFIcatIONS

Page 9 of the DeVilbiss Suction Unit series (7305) user manual — With Disposable Container and Internal Filter Cartridge
Page 9 · click to enlarge
Page 9

No routine calibration or service is required provided the device is used in accordance with the manufacturer’s directions. In case of a change of patient, the device must be reconditioned to protect the user. Reconditioning must only be carried out by the manufacturer or service provider. Between patients:

1. Visually inspect unit for any damage, missing parts etc. 2. Ensure that unit and accessories are clean. 3. Use an independent vacuum gauge to verify the unit provides the proper vacuum level as stated in Specifications. 4. Discard and replace collection container, filter, and tubing between patients. 5. Wipe the surface using a clean cloth dampened with disinfectant.

The compressor portion of the DeVilbiss Suction Unit 7305P & 7305D Series (excluding internal rechargeable batteries) is warranted to be free from defective workmanship and materials for a period of two years from date of purchase. Internal rechargeable batteries are warranted for 90 days. Any defective part(s) will be repaired or replaced at DeVilbiss Healthcare’s option if the unit has not been tampered with or used improperly during that period. Make certain that any malfunction is not due to inadequate cleaning or failure to follow the instructions. If repair is necessary, contact your DeVilbiss Healthcare Provider or DeVilbiss Service Department for instructions: U.S.A. 800-338-1988 or 814-443-4881, Europe +49-(0) 621-178-98-0.

THERE IS NO OTHER EXPRESS WARRANTY. IMPLIED WARRANTIES, INCLUDING THOSE OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED. THIS IS THE EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION IS PERMITTED BY LAW. SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, OR THE LIMITATION OR EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU.

This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.

manufacturer’s Note Thank you for choosing a DeVilbiss Suction Unit. We want you to be a satisfied customer. If you have any questions or comments, please send them to our address on the back cover.

For Service Call Your Authorized DeVilbiss Healthcare Provider:

Phone Purchase Date Serial #

DEVIlbISS GUIDaNcE aND maNUFactURER’S DEclaRatION

Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment. The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used. NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in determining the suitability of the Equipment or System for the Electromagnetic Environment of use, and in managing the Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without disturbing other Equipment and Systems or non-medical electrical equipment.

Guidance and manufacturer’s Declaration – Emissions all Equipment and Systems This device is intended for use in the electromagnetic environment specified below. The customer or user of this device should assure that it is used in such an environment. Emissions TestComplianceElectromagnetic Enforcement – Guidance RF Emissions CISPR 11Group 1This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF Emissions CISPR 11Class B Radiated and Conducted EmissionsThis device is suitable for use in all establishments including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonics IEC 61000-3-2Class A Flicker IEC 61000-3-3Complies Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment - Guidance Electrostatic Discharge (ESD) IEC 61000-4-2±6kV contact ±8kV air±6kV contact ±8kV airFloors should be wood, concrete or ceramic tile. If floors are synthetic, the relative humidity should be at least 30%. Electrical Fast Transient/burst IEC 61000-4-4±2kV on AC Mains±2kV on AC MainsMains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5±1kV Differential ±2kV Common±1kV Differential ±2kV CommonMains power quality should be that of a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11>95% Dip for 0.5 Cycle 60% Dip for 5 Cycles 30% Dip for 25 Cycles >95% Dip for 5 Seconds>95% Dip for 0.5 Cycle 60% Dip for 5 Cycles 30% Dip for 25 Cycles >95% Dip for 5 SecondsMains power quality should be that of a typical commercial or hospital environment. If the user of this device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or battery.

pROVIDER’S NOtES · twO-yEaR lImItED waRRaNty

Page 10 of the DeVilbiss Suction Unit series (7305) user manual — With Disposable Container and Internal Filter Cartridge
Page 10 · click to enlarge
Page 10
Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment - Guidance Power Frequency 50/60Hz Magnetic Field IEC 61000-4-83A/m3A/mPower frequency magnetic fields should be that of a typical location in a typical commercial or hospital environment. Conducted RF IEC 61000-4-63 Vrms from 150 kHz to 80 MHzV1 = 3 VrmsPortable and mobile RF communications equipment should be separated from the device by no less than the recommended separation distances calculated/listed below: D=(3.5/V1)√P Radiated RF IEC 61000-4-33 V/m 80 MHz to 2.5 GHzE1 = 3V/mD=(3.5/E1)√P 80 to 800 MHz D=(7/E1)√P 800 MHz to 2.5 GHz Where P is the maximum power rating in watts and D is the recommended separation distance in meters. Field strengths from fixed transmitters, as determined by an electromagnetic site survey, should be less than the compliance levels (V1 and E1). Interference may occur in the vicinity of equipment containing a transmitter. For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. Recommended Separation Distances between portable and mobile RF communications Equipment and this device. this device and system are NOt life-Supporting This device is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the device as recommended below, according to the maximum output power of the communications equipment. Maximum Output Power (Watts)Recommended Separation Distances for the device (meters) 150 kHz to 80 MHz D=(1.1667)√P80 to 800MHz D=(1.1667)√P800 MHz to 2.5 GHz D=(2.3333)√P 0.010.116670.116670.23333 0.10.368940.368940.73785 11.16671.16672.3333 103.68943.68947.3785 10011.66711.66723.333 For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Display codes

Error codes

What the fault codes shown on the display mean and how to clear them.

CodeMeaningFix
DO-160Dsection 21 Category M For battery operation only Airline use p.8
Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Unit does not turn on, but green external power light is illuminated.1. Check power sources and connections. 2. Ensure wall outlet is live by plugging in a lamp. 3. If running from an internal battery, ensure that your unit has a battery installed. 4. If battery is installed, check that it is fully charged. p.8
Pump runs, but there is no vacuum.1. Check that all tubing is connected properly. 2. Check tubing connections for breaks or leaks. 3. Ensure that float shut-off is not activated. 4. Check for leaks or cracks in container assembly. p.8
Low vacuum.1. Use vacuum adjustment knob to increase vacuum level. 2. Check system for leaks. 3. Push vacuum adjustment knob in toward unit and then release. p.8
Battery will not charge but charge light is illuminated.1. Ensure that unit is equipped with an internal battery by contacting your DeVilbiss equipment provider. 2. Verify that charge light turns on. 3. Check electrical connections during charging. 4. Ensure wall outlet is live by plugging in a lamp. p.8

Ask about the DeVilbiss Suction Unit series (7305)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the DeVilbiss Suction Unit series (7305)

Be the first to ask — other owners or our team will answer.

Original document

DeVilbiss Suction Unit series (7305) manual PDF

32 pages·1.1 MB·Medical Device·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…