DeVilbiss Suction Unit (7314 SU)

38 pages·543 KB·Medical Device·PDF·ENDEJA
DeVilbiss Suction Unit (7314 SU)
Table of Contents 14 sections 7305 SUCTION UNIT SERIES 0044 · EN DeVilbiss ® Vacu-Aide ® QSU Suction Unit Instruction Guide 7314 Series · Vacu-Aide ® QSU · FR Manuel utilisateur de l’apirateur de mucosités Vacu-Aide ® QSU Série 7314 · … ShowHide
Page 1 of the DeVilbiss Suction Unit (7314 SU) user manual — Série 7314
Page 1 · click to enlarge
Page 1
DeVilbiss Suction Unit (7314 SU) — 7305 SUCTION UNIT SERIES 0044 diagram (page 1)
Fig. 1

7305 SUCTION UNIT SERIES 0044

EN DeVilbiss ® Vacu-Aide ® QSU Suction Unit Instruction Guide 7314 Series

Made in U.S.A.

ES Guía de Instrucciones para la Unidad de Succión de la Serie 7314 de DeVilbiss®

Vacu-Aide ® QSU

Hecho en EE.UU.

FR Manuel utilisateur de l’apirateur de mucosités Vacu-Aide ® QSU Série 7314

Fabriqué aux U.S.A.

IT Aspiratore serie 7314 DeVilbiss ® Vacu-Aide ® QSU Guida dell’utente

NL Gebruikershandleiding voor DeVilbiss ® Vacu-Aide ® QSU Model 7314

uitzuigapparaat

ATTENTIE–De federale wetgeving in de Verenigde Staten schrijft voor dat dit apparaat uitsluitend mag worden verkocht of voorgeschreven door een arts.

Geproduceerd in de Verenigde Staten

PT DeVilbiss® Vacu-Aide® QSU Manual de Instruções da Unidade de Sucção

Série 7314

Feito nos EUA.

Prodotto in USA

Page 2 of the DeVilbiss Suction Unit (7314 SU) user manual — 7305 SUCTION UNIT SERIES 0044
Page 2 · click to enlarge
Page 2
Page 2 of the DeVilbiss Suction Unit (7314 SU) user manual — 7305 SUCTION UNIT SERIES 0044
Page 3 of the DeVilbiss Suction Unit (7314 SU) user manual — ACCESSORY/REPLACEMENT ITEMS
Page 3 · click to enlarge
Page 3

3. Keep the power cord away from heated surfaces. 4. Never use while drowsy or asleep. 5. Do not cover the unit or the AC to DC adapter while power is applied. 6. Never operate this product if a. It has a damaged power cord or plug. b. It is not working properly. c. It has been dropped or damaged. d. It has been dropped into water. Instead return the product to an authorized DeVilbiss Healthcare service center for examination and repair.

1. All suctioning should be done in strict accordance with appropriate procedures that have been established by a licensed medical authority. 2. Some attachments or accessories may not fit the tubing supplied. All attachments or accessories should be checked prior to use to assure proper fit.

INTERNATIONAL TRAVEL

The 7314 series is equipped with an AC to DC adapter allowing operation on any AC voltage (100-240 VAC, 50/60 Hz). However the correct power cord must be used to connect to adaptable wall power.

INTRODUCTION

Your DeVilbiss Suction Unit is a compact medical suctioning device which has been designed for reliable, portable operation. Following the recommended operating and maintenance procedures outlined in this instruction guide will maximize the life of this product.

Intended Use Statement The device is to be used to remove fluids from the airway or respiratory support system and infectious materials from wounds. The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection container. The fluids are trapped in the collection container for proper disposal. It is for use on the order of a physician only.

1. Vacuum gauge 2. Vacuum regulator knob 3. 6’ patient tubing 4. Tubing connector 5. 800 ml disposable container with lid (float shut off incorporated into lid) and filter cartridge
DeVilbiss Suction Unit (7314 SU) — INTRODUCTION diagram (page 3)
Fig. 2
6. Filter cartridge with 4 3/8 ” tubing 7. DC power input (on side) 8. Power switch 9. LED power lights AC to DC adapter (not shown) DC power cord (not shown) Internal rechargeable battery (not shown) Carrying case (not shown)

ACCESSORY/REPLACEMENT ITEMS

The following items can be purchased separately as accessories or replacement items for your 7314 Series DeVilbiss Suction Unit:

DescriptionPart No.DescriptionPart No. 6’ patient tubing6305D-611Power cord for USDV51D-606 12V DC power cord (1 each)7304D-619Power cord for Continental EuropeDV51D-607 800 ml disposable container with filter cartridge and 43/8” tubing (48 each)7305D-632Power cord for UKDV51D-608 AC to DC adapter/charger (see Specifications for manufacturer information)7314P-613Power cord for AustraliaDV51D-609 Carrying case7314D-606Power cord for JapanDV51D-613 Filter cartridge (12 pack)7305D-635Power cord package (Europe, UK, Australia cords)DV51D-611 Collection Container Kit (Filter cartridge, 800 ml container, 43/8” and 6’ tubing package)7305D-633
Page 4 of the DeVilbiss Suction Unit (7314 SU) user manual — 7 9 10 11
Page 4 · click to enlarge
Page 4

SET-UP & OPERATION

DeVilbiss Suction Unit (7314 SU) — SET-UP & OPERATION diagram (page 4)
Fig. 3
DeVilbiss Suction Unit (7314 SU) — SET-UP & OPERATION diagram (page 4)
Fig. 4

5 2 3 4

DeVilbiss Suction Unit (7314 SU) — 5 2 3 4 diagram (page 4)
Fig. 5
DeVilbiss Suction Unit (7314 SU) — 5 2 3 4 diagram (page 4)
Fig. 6
DeVilbiss Suction Unit (7314 SU) — 5 2 3 4 diagram (page 4)
Fig. 7
DeVilbiss Suction Unit (7314 SU) — 5 2 3 4 diagram (page 4)
Fig. 8

Fully charge battery for 10-17 hOURS.

Insert container into holder.

Attach 43/8” tubing from filter cartridge to tubing connector.

Attach 6’ patient tubing to container lid at outlet labeled .

Ensure power switch is “off”.

Select desired power source. (Skip steps 7-8 if using internal battery power.)

DeVilbiss Suction Unit (7314 SU) — 5 2 3 4 diagram (page 4)
Fig. 9

7 9 10 11

DeVilbiss Suction Unit (7314 SU) — 7 9 10 11 diagram (page 4)
Fig. 10
DeVilbiss Suction Unit (7314 SU) — 7 9 10 11 diagram (page 4)
Fig. 11
DeVilbiss Suction Unit (7314 SU) — 7 9 10 11 diagram (page 4)
Fig. 12

Turn the unit “on”. If using AC or DC power, plug the small connector into the DC power input on the side of unit.

Plug the other end into an AC wall outlet or DC receptacle. NOTE– The AC adapter may become warm to the touch during charging or running of the unit. This is normal.

Adjust the suction level. Verify suction level. NOTE - Always verify suction level before beginning by occluding open end of patient tubing while observing gauge. Adjust knob to desired level.

bATTERY CHARGING & FILTER MAINTENANCE

DeVilbiss Suction Unit (7314 SU) — 7 9 10 11 diagram (page 4)
Fig. 13
DeVilbiss Suction Unit (7314 SU) — 7 9 10 11 diagram (page 4)
Fig. 14
DeVilbiss Suction Unit (7314 SU) — 7 9 10 11 diagram (page 4)
Fig. 15
DeVilbiss Suction Unit (7314 SU) — 7 9 10 11 diagram (page 4)
Fig. 16

Ensure power switch is “off”.

Plug the small connector of the AC or DC adapter into the DC power input.

Plug the other end into an AC wall outlet or DC receptacle.

Battery charging begins; 10-17 hours for full charge.

Battery charging complete.

battery Charging

Page 5 of the DeVilbiss Suction Unit (7314 SU) user manual — 1 2 3 4
Page 5 · click to enlarge
Page 5
DeVilbiss Suction Unit (7314 SU) — CLEANING INSTRUCTIONS diagram (page 5)
Fig. 17
DeVilbiss Suction Unit (7314 SU) — CLEANING INSTRUCTIONS diagram (page 5)
Fig. 18
DeVilbiss Suction Unit (7314 SU) — CLEANING INSTRUCTIONS diagram (page 5)
Fig. 19

Turn unit “off”.

Remove filter cartridge and 43/8” tubing.

Install new cartridge and attach tubing.

CLEANING INSTRUCTIONS

DeVilbiss Suction Unit (7314 SU) — CLEANING INSTRUCTIONS diagram (page 5)
Fig. 20
DeVilbiss Suction Unit (7314 SU) — CLEANING INSTRUCTIONS diagram (page 5)
Fig. 21

1 3 4 5 6

DeVilbiss Suction Unit (7314 SU) — 1 3 4 5 6 diagram (page 5)
Fig. 22
DeVilbiss Suction Unit (7314 SU) — 1 3 4 5 6 diagram (page 5)
Fig. 23

Disconnect from power source. Disconnect tubing and remove

Carefully remove lid and empty contents. NOTE– The 800ml disposable collection container and lid are meant for single-patient use.

Remove filter cartridge and 43/8” tubing and set aside. Wash container and lid in warm water/dishwashing solution. Rinse with clean, warm water. NOTE– If desired, the float ball can be removed for washing by pushing ball out of lid assembly.

Turn unit “off” and allow vacuum to drop.

container from holder.

DeVilbiss Suction Unit (7314 SU) — 1 3 4 5 6 diagram (page 5)
Fig. 24

Wash in 1 part vinegar to 3 parts warm water solution. Rinse with clean, warm water and air dry.

AC to DC Adapter

DeVilbiss Suction Unit (7314 SU) — 1 3 4 5 6 diagram (page 5)
Fig. 25

1 2 3 4

DeVilbiss Suction Unit (7314 SU) — 1 2 3 4 diagram (page 5)
Fig. 26
DeVilbiss Suction Unit (7314 SU) — 1 2 3 4 diagram (page 5)
Fig. 27
DeVilbiss Suction Unit (7314 SU) — 1 2 3 4 diagram (page 5)
Fig. 28
DeVilbiss Suction Unit (7314 SU) — 1 2 3 4 diagram (page 5)
Fig. 29

Disconnect from lid.

Rinse thoroughly by running warm tap water through it.

Follow by running a solution of 1 part vinegar to 3 parts warm water through it. Rinse with warm tap water and air dry.

Keep outer surface clean by wiping with clean, damp cloth .

Disconnect AC to DC adapter from unit and from power source.

Wipe AC to DC adapter housing and cords with a dry cloth.

DeVilbiss Suction Unit (7314 SU) — 1 2 3 4 diagram (page 5)
Fig. 30
DeVilbiss Suction Unit (7314 SU) — 1 2 3 4 diagram (page 5)
Fig. 31
DeVilbiss Suction Unit (7314 SU) — 1 2 3 4 diagram (page 5)
Fig. 32

Turn unit “off” and allow vacuum to drop.

Disconnect from power source.

Wipe the housing with a clean cloth and any commercial (bacterial-germicidal) disinfectant.

Wipe with clean cloth dampened with detergent or disinfectant.

Show 5 more labels / short text on this page
Filter Cartridge Changing · Collection Container · 6’ Patient Tubing · Suction Unit · Carrying Case
Page 6 of the DeVilbiss Suction Unit (7314 SU) user manual — SPECIFICATIONS/CLASSIFICATIONS
Page 6 · click to enlarge
Page 6
PRObLEMACTION Unit does not turn on (no external power is connected).1. Check that battery is fully charged and/or charge battery. Unit does not turn on when external power is connected. Green external power light does not illuminate.1. Check power sources and connections. 2. Ensure wall outlet is live by plugging in a lamp. Pump runs, but there is no suction.1. Check that all tubing is connected properly. 2. Check tubing connections for breaks or leaks. 3. Ensure that float shut-off in collection container is not activated or filter cartridge occluded. 4. Check for leaks or cracks in collection container assembly. Low suction.1. Use vacuum regulator knob to increase suction level. 2. Check system for leaks. Battery will not charge (external power and charge indicator lights should be illuminated during charge mode)1. Verify that both external power and charge indicator lights illuminate. 2. Check power sources and connections. 3. Ensure wall outlet is live by plugging in a lamp.

SPECIFICATIONS/CLASSIFICATIONS

Size (including container)8.3 H x 8.0 W x 8.5 D (21.1 cm x 20.3 cm x 21.6 cm) (not including AC to DC adapter) Weight (including container)6.6 lb. (3 kg) (not including AC to DC adapter) Typical Operating Sound Level55 dBA Electrical Requirements100-240V~, 50/60Hz, 1.2A max ; 12V ; 33 W Max Vacuum Range50-550 mm Hg +/- 10%* Air Flow @ Pump Inlet:27 LPM (free flow) typical (may be less when running from internal battery)* Collection Container Capacity800 ml (cc) Disposable WarrantyTwo-years limited, excluding internal battery and collection container Internal Battery90-day ApprovalsIEC 60101-1-2; CSA-C22.2 # 601.1; UL 60601-1; EN 60601-1-2; ISO 10079-1; IEC 60601-1; IEC 60529 IP12; IEC 60601-1-6; CENELEC EN 60601-1 Adapter Manufacturer InformationEmerson Model # AD5012N2LM or Autec Power Systems Model # DTM36-12 or SL Power/Ault Model # MENB1040A1240N02 Environmental Conditions Operating Temperature Range32°F (0°c) - 104°F (40°c) Operating Relative Humidity0-95% Operating Atmospheric Pressure10.2 psi (70 kPa) - 15.4 psi (106 kPa) Storage & Transport Temperature Range-40°F (-40°c) - 158°F (70°c) Storage & Transport Relative Humidity0-95% Storage & Transport Atmospheric Pressure7.3 psi (50 kPa) - 15.4 psi (106 kPa) Equipment Classifications With respect to protection from electric shockClass II and internally powered Degree of protection against electric shockType BF Applied Parts Degree of protection against ingress of liquidsIP12 and ordinary power supply Mode of OperationIntermittent Operation: 30 minutes on, 30 minutes off Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide. ISO Classification Electrically powered medical suction equipment for field and transport use according to EN ISO 10079-1 : 2009 High Flow/High Vacuum

Manufacturer’s Note Thank you for choosing a DeVilbiss Suction Unit. We want you to be a satisfied customer. If you have any questions or comments, please send them to our address on the back cover. For Service Call Your Authorized DeVilbiss Healthcare Provider:

Phone Purchase Date Serial #

DEVILbISS GUIDANCE AND MANUFACTURER’S DECLARATION

Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.

The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used.

TROUbLESHOOTING

Page 7 of the DeVilbiss Suction Unit (7314 SU) user manual — SPECIFICATIONS/CLASSIFICATIONS
Page 7 · click to enlarge
Page 7

NOTE– The EmC tables and other guidelines provide information to the customer or user that is essential in determining the suitability of the Equipment or System for the Electromagnetic Environment of use, and in managing the Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without disturbing other Equipment and Systems or non-medical electrical equipment.

Guidance and Manufacturer’s Declaration – Emissions All Equipment and Systems This device is intended for use in the electromagnetic environment specified below. The customer or user of this device should assure that it is used in such an environment. Emissions TestComplianceElectromagnetic Enforcement – Guidance RF Emissions CISPR 11Group 1This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF Emissions CISPR 11Class B Radiated and Conducted EmissionsThis device is suitable for use in all establishments including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonics IEC 61000-3-2Class A Flicker IEC 61000-3-3Complies Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment - Guidance Electrostatic Discharge (ESD) IEC 61000-4-2±6kV contact ±8kV air±6kV contact ±8kV airFloors should be wood, concrete or ceramic tile. If floors are synthetic, the relative humidity should be at least 30%. Electrical Fast Transient/ burst IEC 61000-4-4±2kV on AC Mains±2kV on AC MainsMains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5±1kV Differential ±2kV Common±1kV Differential ±2kV CommonMains power quality should be that of a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11>95% Dip for 0.5 Cycle 60% Dip for 5 Cycles 30% Dip for 25 Cycles >95% Dip for 5 Seconds>95% Dip for 0.5 Cycle 60% Dip for 5 Cycles 30% Dip for 25 Cycles >95% Dip for 5 SecondsMains power quality should be that of a typical commercial or hospital environment. If the user of this device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or battery. Power Frequency 50/60Hz Magnetic Field IEC 61000-4-83A/m3A/mPower frequency magnetic fields should be that of a typical location in a typical commercial or hospital environment. Conducted RF IEC 61000-4-63 Vrms from 150 kHz to 80 MHz3 VrmsPortable and mobile RF communications equipment should be separated from the device by no less than the recommended separation distances calculated/listed below: D=(1.2)√P Radiated RF IEC 61000-4-33 V/m 80 MHz to 2.5 GHz3V/mD=(1.2)√P 80 to 800 MHz D=(2.3)√P 800 MHz to 2.5 GHz Where P is the maximum power rating in watts and D is the recommended separation distance in meters. Field strengths from fixed transmitters, as determined by an electromagnetic site survey, should be less than the compliance levels (3 Vrms and 3V/m). Interference may occur in the vicinity of equipment containing a transmitter. For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: T hese guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. Recommended Separation Distances between Portable and Mobile RF Communications Equipment and this device. This device and system are NOT Life-Supporting This device is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the device as recommended below, according to the maximum output power of the communications equipment. Maximum Output Power (Watts)Recommended Separation Distances for the device (meters) 150 kHz to 80 MHz D=(1.2)√P80 to 800MHz D=(1.2)√P800 MHz to 2.5 GHz D=(2.3)√P 0.010.120.120.23 0.10.380.380.73 11.21.22.3 103.83.87.3 100121223 For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: T hese guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Unit does not turn on (no external power is connected).1. Check that battery is fully charged and/or charge battery. p.6
Unit does not turn on when external power is connected. Green external power light does not illuminate.1. Check power sources and connections. 2. Ensure wall outlet is live by plugging in a lamp. p.6
Pump runs, but there is no suction.1. Check that all tubing is connected properly. 2. Check tubing connections for breaks or leaks. 3. Ensure that float shut-off in collection container is not activated or filter cartridge occluded. 4. Check for leaks or cracks in collection container assembly. p.6
Low suction.1. Use vacuum regulator knob to increase suction level. 2. Check system for leaks. p.6
Battery will not charge (external power and charge indicator lights should be illuminated during charge mode)1. Verify that both external power and charge indicator lights illuminate. 2. Check power sources and connections. 3. Ensure wall outlet is live by plugging in a lamp. p.6

Ask about the DeVilbiss Suction Unit (7314 SU)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the DeVilbiss Suction Unit (7314 SU)

Be the first to ask — other owners or our team will answer.