DE DeVilbiss ® Vacu-Aide ®
EN
Compact Suction Unit Instruction Guide
Kompaktes Absauggerät Bedienungsanleitung
7310 Series
7310 Serie
Made in U.S.A. of US & Imported Parts
Hergestellt in den U.S.A. unter Verwendung amerikanischer und importierter Teile
Aspiratore compatto
IT
DeVilbiss ® Vacu-Aide ® Guida dell’utente
Serie 7310
Serie 7310
Fabricado en EE. UU. con piezas fabricadas en EE. UU. e importadas
Assemblato negli USA con componenti prodotti negli Stati Uniti e importati.
FR
l’Unité d’aspiration compacte Vacu-Aide ® DeVilbiss ®
NL
Instructiehandleiding
DeVilbiss ® Vacu-Aide ® Compact afzuigeenheid
Série 7310
Model 7310
ATTENTIE– De federale wetgeving in de Verenigde Staten schrijft voor dat dit apparaat uitsluitend mag worden verkocht of voorgeschreven door een arts.
Geproduceerd in de Verenigde Staten met Amerikaanse en geïmporteerde onderdelen.
Assemblé aux États-Unis avec des pièces des États-Unis et des pièces importées.
Manual de instrucciones









EN English EN-3 ES Español ES-11 FR Français FR-20 DE Deutsch DE-29 IT Italiano IT-38 NL Nederlands NL-47



INTERNATIoNAL TRAVEL The suction unit is equipped with an AC to DC Adapter/Charger allowing operation on any AC voltage (100-240 VAC, 47-63 Hz). However the correct power cord must be used to connect to adaptable wall power. Your unit is approved for use on commercial aircraft; specific standards are listed in the Specifications in this guide. Certificate is available upon request; if needed, contact your provider for a copy of approval letter before travel.
INTRoDUCTIoN Your DeVilbiss Suction Unit is a compact medical suctioning device which has been designed for reliable, portable operation. Following the recommended operating and maintenance procedures outlined in this instruction guide will maximize the life of this product.
Intended Use Statement The device is to be used to remove fluids from the airway or respiratory support system and infectious materials from wounds. The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection container. The fluids are trapped in the collection container for proper disposal. It is for use on the order of a physician only.
2 4 3 1



The following items can be purchased separately as accessories or replacement for your suction unit.
1. Your unit was shipped with the battery disconnected. Connect battery and charge before first use. Refer to Battery Replacement section for instructions. 2. Place clear side of filter labeled IN directly into container’s lid. 3. With filter attached to container, directly insert into the front port. Both the filter and the bottom support secure the container to the unit (FIG. 1 & 2). 4. The 6’ (1.8m) patient tubing can now be connected to the container lid at the outlet labeled . 5. Ensure that all connections are secure before use to prevent leaks. 6. Verify that the unit is at the desired suction level before beginning patient suctioning.
how To oPERATE yoUR SUCTIoN UNIT
External power: Supplied from AC or DC power cord. Illuminates in GREEN when power is supplied.
Battery charging: Illuminates in YELLOW. The light will go out when the battery is fully charged.
Low battery: illuminates in RED when battery reaches a discharge state.
mmHg Suction level setting: Scale/strength illuminates in GREEN. This scale shows the level of suction strength in mmHg.
Power Source options AC OPERATION - Plug the 90° power connector of AC to DC adapter/charger into DC power input (FIG 3) and attach the line cord. Plug the other end of AC adapter into a grounded AC supply. 12V DC OPERATION - Plug the small 90° power connector of DC power cord into DC power input (FIG 3). Plug large end of cord into 12V DC power receptacle of vehicle. BATTERY OPERATION - Your unit is equipped with a high capacity rechargeable battery. For initial charge on new unit, fully charge the battery for a minimum of 5 hours (see Battery Charging). To operate unit from the rechargeable battery, ensure that no external power sources are plugged into the suction unit.
how to adjust the vacuum level
ACCESSoRy/REPLACEMENT ITEMS · Control Panel Symbols · On/Off

vacuum regulator knob clockwise to increase and counter-clockwise to decrease the vacuum (FIG 4). Release and occlude once more to confirm setting. The desired level of vacuum can be viewed on the LED display.
NOTE–The LEDs have two brightness levels. As the vacuum level is adjusted, the LEDs will illuminate in progression. When an LED is at half brightness, it indicates that the vacuum level is halfway between the previous fully lit LED and the half brightness LED. EXAMPLE: If the 150 mmHg LED is fully illuminated and the 200 mmHg LED is at half brightness, the suction level is 175 mmHg. When the 200 mmHg LED illuminates at full brightness, the unit has reached 200 mmHg.
BATTERy ChARGING The Vacu-Aide Compact series is equipped with a factory-installed high capacity rechargeable battery. Connect battery and charge before first use (refer to Battery Replacement).
1. Connect the unit to either an AC or DC power source. 2. The green external power light should illuminate; the yellow charge indicator will remain lit while the battery is charging. 3. Ensure that the yellow charging light is illuminated when charging begins. As the battery nears a full charge, the yellow battery charging light may flash for several minutes. This is normal. If your unit does not hold a charge, check that the yellow light turns on when external power is applied with the power button “Off.” If problems persist, contact your medical equipment provider.
If unit is not in use for extended periods, the battery should be recharged a minimum of every 6 months. A fully charged battery will provide approximately 45-60 minutes of continuous operation at zero vacuum level (free flow).
Collection Container: NOTE–Elbow is permanently attached to the lid and cannot be removed.
Suction Unit (single-patient use)
BATTERy REPLACEMENT · CLEANING INSTRUCTIoNS

Carrying Case (single-patient use)
Suction Unit (multi-patient use) Device Cleaning and Disinfection When There is a Patient Change When medical devices have already been used with a patient, contamination with human pathogenic germs should be assumed (unless there is evidence to the contrary), and the next patient, user or third party should be protected by appropriate handling and preparation. Therefore, when there is a patient change, people must be protected during the transport and handling of the device and the device must be fully processed, i.e., cleaned and disinfected, by suitably trained personnel before reuse to protect the next patient. The complete processing may only be done by the manufacturer or by a qualified DeVilbiss provider/service technician.
NOTE–IfthefollowingprocessingoftheunitbyaqualifiedDeVilbissprovider/technicianisnotpossible,theunitmustnot be used by another patient! DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualified third party between uses by different patients. 1. Dispose of all accessory components that are not suitable for reuse, i.e., collection container, filter, tubing and carrying case. 2. With the power switch in the “Off” position, disconnect the DeVilbiss Suction Unit from all external power sources. 3. Visually inspect unit for any damage, missing parts, etc. 4. Wipe the housing with a clean cloth and a commercial (bacterial-germicidal) disinfectant that meets the requirements listed in the NOTE below and is used as per the disinfectant manufacturer’s recommended dilution rates and instructions. CAUTION – Do not submerge in water as this will result in damage to the vacuum pump.
Changing Filter & Replacing Container
Tubing (single-patient use)

SPECIFICATIoNS/CLASSIFICATIoNS Size 7.25”H x 7.25”W x 6.75”D (18.4cm x 18.4cm x 17.1cm) Weight 3.375 lbs. (1.53 kg) Electrical Requirements 100-240 VAC 50/60 Hz 0.75 A max; 12 VDC, 33 W max Vacuum Range 50-500 mmHG +/- 10% Air Flow @ pump inlet: 27 LPM (free flow) typical (may be less when running from internal battery) Collection Container Capacity 725 ml (cc) Environmental Conditions Operating Temperature Range 32°F (0°C) - 104°F (40°C) Operating Relative Humidity 0-95% Operating Atmospheric Pressure 10.2 Psi (70 kPA) - 15.4 Psi (106 kPA) Storage & Transport Temperature Range -40°F (-40°C) - 158°F (70°C) Storage & Transport Relative Humidity 0-95% Storage & Transport Atmospheric Pressure 7.3 Psi (50 kPA) - 15.4 Psi (106 kPA) warranty 7310 Series Two-years limited, excluding internal battery and collection container Internal Battery 6 months Approvals IEC 601-1; IEC 68; CAN/CSA-C22.2 No. 601.1-M90; UL 2601-1, CE EN 60601-1-2, ISO10079-1:1999 Meets RTCA/DO-160E DO-160E - section 21 Category M
(for battery operation only; commercial aircraft, airborne equipment) Equipment Classifications With respect to protection from electric shock Class I and internally powered Degree of protection against electric shock Type BF Applied Parts Degree of protection against ingress of liquids IP12 and ordinary power supply Mode of Operation Intermittent Operation: 30 minutes on, 30 minutes off Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. ISo Classification 7310 Series - Electrically powered medical suction equipment for field and transport use according to ISO 10079-1 : 1999 High Flow/High Vacuum
TRoUBLEShooTING

Two-yEAR LIMITED wARRANTy The compressor portion of the DeVilbiss Vacu-Aide Compact Suction Unit 7310 Series (excluding internal rechargeable battery and collection container) is warranted to be free from defective workmanship and materials for a period of two years from date of purchase. Internal rechargeable batteries are warranted for 6 months. Any defective part(s) will be repaired or replaced at DeVilbiss Healthcare’s option if the unit has not been tampered with or used improperly during that period. Make certain that any malfunction is not due to inadequate cleaning or failure to follow the instructions. If repair is necessary, contact your DeVilbiss Provider or DeVilbiss for instructions.
THERE IS NO OTHER EXPRESS WARRANTY. IMPLIED WARRANTIES, INCLUDING THOSE OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW. ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED. THIS IS THE EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION IS PERMITTED BY LAW. SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, OR THE LIMITATION OR EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
Manufacturer’s Note Thank you for choosing a DeVilbiss Vacu-Aide Compact Suction Unit. We want you to be a satisfied customer. If you have any questions or comments, please send them to our address on the back cover. For Service Call Your Authorized DeVilbiss Provider: Phone Purchase Date Serial #
DEVILBISS GUIDANCE AND MANUFACTURER’S DECLARATIoN
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used.
NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in determining the suitability of the Equipment or System for the Electromagnetic Environment of use, and in managing the Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without disturbing other Equipment and Systems or non-medical electrical equipment.


ÍNDICE Figuras EN - 2 Símbolos IEC ES - 11 Precauciones importantes ES - 11 Viajes internacionales ES - 12 Introducción ES - 12 Piezas importantes ES - 13 Accesorios/Piezas de repuesto ES - 13 Instalación ES - 13 Cómo utilizar la unidad de succión ES - 13 Carga de la batería ES - 14 Sustitución de la batería ES - 14 Instrucciones de limpieza ES - 15 Mantenimiento ES - 16 Resolución de problemas ES - 16 Especificaciones/Clasificaciones ES - 17 Garantía ES - 17 Declaración del fabricante y guía de uso de DeVilbiss ES - 18
Atención, consulte el manual de instrucciones Corriente alterna
Consulte las instrucciones de uso Corriente continua
En espera - Encendido / Apagado Utiliza piezas tipo BF
Indicador de polaridad positiva en el centro
SÍMBoLoS IEC
What the fault codes shown on the display mean and how to clear them.
| Code | Meaning | Fix | |
|---|---|---|---|
| IP12 | Vertically falling water drops shall have no harmful effects when the enclosure is tilted at an angle | p.3 |
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Unit does not power on. (Green external power indicator should be illuminated when power is applied.) | 1. Check power sources and connections. 2. E nsure wall outlet is live by plugging in a lamp. 3. Check that battery is fully charged. | p.8 | |
| Pump runs, but no vacuum. | 1. Check that all tubing is connected properly. 2. C heck tubing connections for breaks, leaks, or occlusions. 3. E nsure that float shut-off is not activated due to full container. 4. Check for leaks or cracks in container assembly. | p.8 | |
| Low vacuum. | 1. U se vacuum adjustment knob to increase vacuum level. 2. Check system for leaks. | p.8 | |
| Battery will not hold a charge. | 1. Confirm battery is connected (refer to Set-Up). 2. Verify that charge light turns on. 3. Check electrical connections during charging. 4. E nsure wall outlet is live by plugging in a lamp. If problem continues, contact your equipment provider. | p.8 |