For additional information, please visit our website at www.rosemountanalytical.com
The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely the responsibility of the user.
The 402 and 402VP sensors are retractable. BEFORE RETRACTING THE SENSOR, be absolutely certain the process pressure
is less than 64 psig (542 kPa abs) and the process temperature is at a safe level!


Figure 1. Sensor Orientation

Keep ¼ in. (6 mm) clearance between electrodes and piping. The electrolytes must be completely submerged in the process liquid, i.e., to the upper edge of the guard (item 10 in Figure 3).




Installation Retraction assembly kit (PN 23765-00) consists of items 1 through 7 shown in Figure 3 as well as a hex key and pipe tape. The sensor assembly includes items 8 through 11.
Do not remove or alter the guard (10) on the sensor tube (9).
Figure 3. 402 Sensor with Retraction Assembly Kit


1. Make sure the system is shut down and there is no residual pressure. 2. Attach the ball valve (2) to the process piping using the 1–¼-inch FNPT port on the valve or the 1–¼ inch NPT nipple (1). Use pipe tape on male threads. 3. Slide the handle lock up on the ball valve handle and close the ball valve (2). If the process will be restarted before the sensor is installed, make sure the system pressure is at or below 542 kPa abs (64 psig) before proceeding. If the system will not be restarted until after the sensor installation, leave the valve in the open position. 4. If the sensor includes a junction box (8), it must be removed from the sensor to install the retraction assembly kit. Disconnect the sensor wires inside the junction box prior to disassembly. 5. Follow Figure 4 to assemble the sensor and retraction assembly. Keep items 3-7 in the proper sequence. Items 3, 6, and 7 are factory assembled. 6. Position the sensor tube (9) so that the electrodes (11) are completely withdrawn inside the packing adapter (3). If it is difficult to slide the sensor through the packing rings, loosen the packing bushing (6). 7. Position the clamp (4) on the sensor tube (9) so that the electrodes (11) will be completely immersed in the process liquid when the sensor is fully inserted through the ball valve. See Figure 3. Use the hex key to secure the clamp to the sensor tube. 8. Slide the retainer (5) onto the sensor tube (9). 9. Screw the junction box (8) — hand tight — onto the sensor tube (9). DO NOT OVERTIGHTEN. Use 2-3 wraps of pipe tape on the tube threads if a NEMA 4 seal is required at the junction box. 10. Apply pipe tape to the packing adapter (3) threads and screw the packing adapter onto the ball valve. 11. Check to ensure that the packing bushing (6) has been tightened. You should be able to push the sensor tube (9) against the resistance provided by the packing rings (7). 12. Open the ball valve (2). 13. There may be some leakage around the packing bushing (6). Tighten the packing bushing to stop the leak. Use the junction box (8) or sensor rear to push the sensor through the valve until the clamp (4) rests against the packing bushing. 14. To secure the sensor tube (9) in place, tighten — hand tighten only — the retainer (5) against the back of the packing adapter (3). If the sensor tube retracts when the system pressure is increased, reduce the pressure to 542 kPa abs (64 psig) or less, remove the retainer (5), and tighten the screw in the clamp (4).
Figure 4. 402 Sensor Installation


1. Make sure the system pressure is at or below 542 kPa abs (64 psig) at the valve before proceeding. Provide adequate working area for retraction. 2. Unscrew the retainer (5) from the packing adapter (3). If it is difficult to unscrew the retainer, system pressure may be forcing the clamp (4) into the retainer. The packing bushing (6) may not be tight enough to hold the sensor tube (9) in place. To prevent a sudden retraction of the sensor tube as the retainer is removed, hold onto the junction box (8) or sensor rear to restrain the sensor. 3. If the sensor tube (9) does not slide back as the retainer is unscrewed, pull back on the junction box (8) or sensor tube to retract the sensor. 4. If the sensor tube (9) cannot be retracted, loosen the clamp (4) using the hex key. Pull the clamp back and slowly loosen the packing bushing (6) in 1/8-turn increments. System pressure may cause the tube to retract. If not, pull back on the junction box (8) or sensor tube (9) to withdraw the sensor. 5. Withdraw the sensor tube until the guard (10) contacts the stop inside the packing adapter (3). The electrodes (11) are now inside the packing adapter. See Figure 5. Slide the handle lock up on the valve handle and close the ball valve (2).
Failure to withdraw the sensor completely may result in damage to the electrodes when the valve is closed.
Figure 5. 402 Sensor in Retracted Position


For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product.
1056, 1057, 56, 5081, 6081, 54e, and XMT : http://www3.emersonprocess.com/raihome/sp/liquid/wiring/XMT/ 1066 and sensors with SMART preamps: http://www2.emersonprocess.com/en-US/brands/rosemountanalytical/Liquid/Sensors/ Pages/Wiring_Diagram.aspx 1055: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/1055/
Wire Color and Connections in Sensor
Figure 6. Wiring for Sensor-Mounted Junction Box Figure 7. Wiring for 54eC and 54C


MODEL 54C & 54eC
ANALYZERS
SENSOR CABLE
GRAY
SENSOR
TB1
RED
RED
WIRES
GROUND
RTD SHIELD
CLEAR
TB
RTD RETURN
WHITE WHITE/RED
RTD SENSE
RED
RTD IN
N/C RECEIVE COMMON
CLEAR
WHITE ORANGE
RED
GRAY
CLEAR
WHT/RED
CLEAR
CLEAR ORANGE
CABLE
RECEIVE
N/C
DRIVE COMMON
CLEAR GRAY
GROUND

Figure 8. Wiring for 56 and 1056
Figure 9. Wiring (Panel) for Xmt-C-10


TB1
WHITE WHITE/RED
RTD RTN
DRIVE
GRAY
RTD SENSE RTD IN
CLEAR
DRIVE COM
RED
CLEAR
RECEIVE
ORANGE
RTD SHLD
CLEAR
4CTB 4CTA SHLD 2CT
RECEIVE COM
CLEAR
RTD IN
RED
ORANGE
SEN 2CTB
RTD SENSE
RED/WHITE
CLEAR
SHLD 2CT
RTD COM
WHITE
GRAY
SEN 2CTA
RTD SHIELD
CLEAR
Figure 10. Wiring for Xmt-C-11 (Pipe or Wall)
Figure 11. Wiring for 1066

RTD SHIELD
CLEAR
TB1 TB2
TB1
WHITE RED/WHITE
RTD COM
RTN
WHITE
RCV B
WHITE
RTD SENSE
SENSE
WHITE/RED
WHITE/RED
RCV A
RED
RTD IN
RTD IN
CLEAR
RSHLD
RED
RED
CLEAR
SHLD
CLEAR
CLEAR
GRAY
GRAY
DRVB
CLEAR
RECEIVE COM
ORANGE
ORANGE
DRVA
ORANGE
RECEIVE
DSHLD
CLEAR
CLEAR
CLEAR
DRIVE COM
GRAY
DRIVE
Figure 12. Wiring for 5081-C
Figure 13. Wiring for 6081-C

2CT-A
2CT-B
DRV SHLD
RTD SHIELD
SHIELD
RCV COM
DRV COM RCV

4CT-A
RCV SHLD
DRV
RTD IN
SENSE
4CT-B
RTD SENSE
RTD COM
RTD RTN
SMART PWR
RTD SHLD HT/FF (-)
RESERVED
HT/FF (+)
WHITE
RED/WHITE
RED
CLEAR
CLEAR
ORANGE
GRAY
CLEAR
CLEAR
WHITE
WHITE/RED
RED
CLEAR
ORANGE
CLEAR
GRAY

Wiring through a Junction Box If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point. Use interconnecting cable 23747-00 (factory-terminated) or 9200275 (no terminations).
Figure 14. VP pin-out

Cleaning the Sensor Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol (rubbing alcohol) can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors. Calibration ENDURANCE conductivity sensors are calibrated at the factory and do not need calibration when first placed in service. Simply enter the cell constant printed on the label into the analyzer.
After a period of service, the sensor may require calibration. The sensor can be calibrated against a solution having known conductivity or against a referee meter and sensor. If using a standard solution, choose one having conductivity in the recommended operating for the sensor cell constant. Refer to the analyzer manual or product data sheet for recommended ranges. Do not use standard solutions having conductivity less than about 100 uS/cm. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25 °C). Because 0.01/cm sensors must be calibrated in low conductivity solutions, they are best calibrated against a referee meter and sensor in a closed system. For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website. Troubleshooting

Rev. L
Figure 15. Checking Temperature Element Disconnect leads and measure resistances shown. The mea sured resistance should be close to the value in the table.
Figure 16. Checking Continuity and Leakage Disconnect electrode leads and measure resistance and continuity as shown. Sensor must be dry when checking resistance between electrode leads.


facebook.com/EmersonRosemountAnalytical
Credit Cards for U.S. Purchases Only.
youtube.com/user/RosemountAnalytical
Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 rosemountanalytical.com
©2014 Rosemount Analytical, Inc. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
© Rosemount Analytical Inc. 2013
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Off-scale reading | Wiring is wrong. | Verify wiring. | p.7 |
| Off-scale reading | Temperature element is open or shorted. | Check temperature element for open or short circuits. See Figure 15 (next page). | p.7 |
| Off-scale reading | Sensor is not in process stream. | Be sure sensor is completely submerged in process stream. | p.7 |
| Off-scale reading | Variopol cable is not properly seated. | Loosen connector and reseat. | p.7 |
| Off-scale reading | Sensor has failed. | Perform isolation checks. See Figure 16 (next page). | p.7 |
| Noisy reading | Sensor is improperly installed in process stream. | Be sure sensor is completely submerged in process stream. | p.7 |
| Noisy reading | Variopol cable is not properly seated. | Loosen connector and reseat. | p.7 |
| Reading seems wrong (lower or higher than expected | Bubbles trapped in sensor. | Be sure sensor is properly oriented in pipe or flow cell. See Figure 1. Apply back pressure to flow cell. | p.7 |
| Reading seems wrong (lower or higher than expected | Wrong temperature correction algorithm. | Check that temperature correction is appropriate for the sample. See analyzer manual for more information. | p.7 |
| Reading seems wrong (lower or higher than expected | Wrong cell constant. | Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. | p.7 |
| Sluggish response | Electrodes are fouled. | Clean electrodes. | p.7 |
| Sluggish response | Sensor is installed in dead area in piping. | Move sensor to a location more representative of the process liquid. | p.7 |
| Sluggish response | p.7 |