ENDURANCE ™ Conductivity Sensors

For additional information, please visit our website at www.rosemountanalytical.com

CAUTION Sensor/Process Application Compatibility

The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely the responsibility of the user.

CAUTION

The 402 and 402VP sensors are retractable. BEFORE RETRACTING THE SENSOR, be absolutely certain the process pressure

is less than 64 psig (542 kPa abs) and the process temperature is at a safe level!

Sensor Specifications

Specifications402402VP Electrode-11, -12, -13 titanium-11, -12, -13 titanium Insulator-11, -12, -13 PEEK (glass filled)-11, -12, -13 PEEK (glass filled) Other Wetted Materials316 SST, EPDM, graphite, Neoprene316 SST, EPDM, graphite, Neoprene Temperature Range32–212 °F (0–100 °C)32–212 °F (0–100 °C) Maximum Pressure200 psig (1481 kPa abs)200 psig (1481 kPa abs) Maximum Retraction Pressure64 psig (542 kPa abs)64 psig (542 kPa abs)
Emerson Process Management Rosemount 402VP — Sensor Specifications diagram (page 1)
Fig. 1 · Sensor Specifications
Emerson Process Management Rosemount 402VP — Sensor Specifications diagram (page 1)
Fig. 2 · Sensor Specifications

Figure 1. Sensor Orientation

Emerson Process Management Rosemount 402VP — Sensor Specifications diagram (page 1)
Fig. 3 · Sensor Specifications

Keep ¼ in. (6 mm) clearance between electrodes and piping. The electrolytes must be completely submerged in the process liquid, i.e., to the upper edge of the guard (item 10 in Figure 3).

Emerson Process Management Rosemount 402VP — Sensor Specifications diagram (page 1)
Fig. 4 · Sensor Specifications
Page 2 of the Emerson Process Management Rosemount 402VP user manual — IMPORTANT!
Page 2 · click to enlarge
Page 2

Installation and Retraction of Sensor

Emerson Process Management Rosemount 402VP — Installation and Retraction of Sensor diagram (page 2)
Fig. 5 · Installation and Retraction of Sensor
Emerson Process Management Rosemount 402VP — Installation and Retraction of Sensor diagram (page 2)
Fig. 6 · Installation and Retraction of Sensor

Installation Retraction assembly kit (PN 23765-00) consists of items 1 through 7 shown in Figure 3 as well as a hex key and pipe tape. The sensor assembly includes items 8 through 11.

IMPORTANT!

Do not remove or alter the guard (10) on the sensor tube (9).

Figure 3. 402 Sensor with Retraction Assembly Kit

Emerson Process Management Rosemount 402VP — IMPORTANT! diagram (page 2)
Fig. 7
Page 3 of the Emerson Process Management Rosemount 402VP user manual — Installation and Retraction of Sensor
Page 3 · click to enlarge
Page 3

1. Make sure the system is shut down and there is no residual pressure. 2. Attach the ball valve (2) to the process piping using the 1–¼-inch FNPT port on the valve or the 1–¼ inch NPT nipple (1). Use pipe tape on male threads. 3. Slide the handle lock up on the ball valve handle and close the ball valve (2). If the process will be restarted before the sensor is installed, make sure the system pressure is at or below 542 kPa abs (64 psig) before proceeding. If the system will not be restarted until after the sensor installation, leave the valve in the open position. 4. If the sensor includes a junction box (8), it must be removed from the sensor to install the retraction assembly kit. Disconnect the sensor wires inside the junction box prior to disassembly. 5. Follow Figure 4 to assemble the sensor and retraction assembly. Keep items 3-7 in the proper sequence. Items 3, 6, and 7 are factory assembled. 6. Position the sensor tube (9) so that the electrodes (11) are completely withdrawn inside the packing adapter (3). If it is difficult to slide the sensor through the packing rings, loosen the packing bushing (6). 7. Position the clamp (4) on the sensor tube (9) so that the electrodes (11) will be completely immersed in the process liquid when the sensor is fully inserted through the ball valve. See Figure 3. Use the hex key to secure the clamp to the sensor tube. 8. Slide the retainer (5) onto the sensor tube (9). 9. Screw the junction box (8) — hand tight — onto the sensor tube (9). DO NOT OVERTIGHTEN. Use 2-3 wraps of pipe tape on the tube threads if a NEMA 4 seal is required at the junction box. 10. Apply pipe tape to the packing adapter (3) threads and screw the packing adapter onto the ball valve. 11. Check to ensure that the packing bushing (6) has been tightened. You should be able to push the sensor tube (9) against the resistance provided by the packing rings (7). 12. Open the ball valve (2). 13. There may be some leakage around the packing bushing (6). Tighten the packing bushing to stop the leak. Use the junction box (8) or sensor rear to push the sensor through the valve until the clamp (4) rests against the packing bushing. 14. To secure the sensor tube (9) in place, tighten — hand tighten only — the retainer (5) against the back of the packing adapter (3). If the sensor tube retracts when the system pressure is increased, reduce the pressure to 542 kPa abs (64 psig) or less, remove the retainer (5), and tighten the screw in the clamp (4).

Figure 4. 402 Sensor Installation

Emerson Process Management Rosemount 402VP — Installation and Retraction of Sensor diagram (page 3)
Fig. 8 · Installation and Retraction of Sensor
Page 4 of the Emerson Process Management Rosemount 402VP user manual — IMPORTANT!
Page 4 · click to enlarge
Page 4

Retraction

1. Make sure the system pressure is at or below 542 kPa abs (64 psig) at the valve before proceeding. Provide adequate working area for retraction. 2. Unscrew the retainer (5) from the packing adapter (3). If it is difficult to unscrew the retainer, system pressure may be forcing the clamp (4) into the retainer. The packing bushing (6) may not be tight enough to hold the sensor tube (9) in place. To prevent a sudden retraction of the sensor tube as the retainer is removed, hold onto the junction box (8) or sensor rear to restrain the sensor. 3. If the sensor tube (9) does not slide back as the retainer is unscrewed, pull back on the junction box (8) or sensor tube to retract the sensor. 4. If the sensor tube (9) cannot be retracted, loosen the clamp (4) using the hex key. Pull the clamp back and slowly loosen the packing bushing (6) in 1/8-turn increments. System pressure may cause the tube to retract. If not, pull back on the junction box (8) or sensor tube (9) to withdraw the sensor. 5. Withdraw the sensor tube until the guard (10) contacts the stop inside the packing adapter (3). The electrodes (11) are now inside the packing adapter. See Figure 5. Slide the handle lock up on the valve handle and close the ball valve (2).

IMPORTANT!

Failure to withdraw the sensor completely may result in damage to the electrodes when the valve is closed.

6. Unscrew the packing adapter (3) from the ball valve (2) to remove the sensor and retraction assembly.

Figure 5. 402 Sensor in Retracted Position

Emerson Process Management Rosemount 402VP — IMPORTANT! diagram (page 4)
Fig. 9
Page 5 of the Emerson Process Management Rosemount 402VP user manual — DRIVE
Page 5 · click to enlarge
Page 5

Wiring

For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product.

1056, 1057, 56, 5081, 6081, 54e, and XMT : http://www3.emersonprocess.com/raihome/sp/liquid/wiring/XMT/ 1066 and sensors with SMART preamps: http://www2.emersonprocess.com/en-US/brands/rosemountanalytical/Liquid/Sensors/ Pages/Wiring_Diagram.aspx 1055: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/1055/

Wire Color and Connections in Sensor

ColorFunction GrayConnects to outer electrode ClearCoaxial shield for gray wire OrangeConnects to inner electrode ClearCoaxial shield for orange wire RedRTD in RTD RTD sense RTD return White with red stripe White ClearShield for all RTD lead wires

Wiring Diagrams

Figure 6. Wiring for Sensor-Mounted Junction Box Figure 7. Wiring for 54eC and 54C

Emerson Process Management Rosemount 402VP — Wiring Diagrams diagram (page 5)
Fig. 10 · Wiring Diagrams
Emerson Process Management Rosemount 402VP — Wiring Diagrams diagram (page 5)
Fig. 11 · Wiring Diagrams

MODEL 54C & 54eC

ORANGE

ANALYZERS

SENSOR CABLE

GRAY

SENSOR

TB1

RED

RED

WIRES

GROUND

RTD SHIELD

CLEAR

TB

RTD RETURN

WHITE WHITE/RED

RTD SENSE

RED

RTD IN

N/C RECEIVE COMMON

CLEAR

WHITE ORANGE

RED

GRAY

CLEAR

WHT/RED

CLEAR

CLEAR ORANGE

CABLE

RECEIVE

N/C

DRIVE COMMON

CLEAR GRAY

DRIVE

GROUND

Page 6 of the Emerson Process Management Rosemount 402VP user manual — RTD IN
Page 6 · click to enlarge
Page 6

Figure 8. Wiring for 56 and 1056

Figure 9. Wiring (Panel) for Xmt-C-10

Emerson Process Management Rosemount 402VP — RTD RTN diagram (page 6)
Fig. 12
Emerson Process Management Rosemount 402VP — RTD RTN diagram (page 6)
Fig. 13

TB1

TB1

WHITE WHITE/RED

RTD RTN

DRIVE

GRAY

RTD SENSE RTD IN

CLEAR

DRIVE COM

RED

CLEAR

RECEIVE

ORANGE

RTD SHLD

CLEAR

4CTB 4CTA SHLD 2CT

RECEIVE COM

CLEAR

RTD IN

RED

ORANGE

SEN 2CTB

RTD SENSE

RED/WHITE

CLEAR

SHLD 2CT

RTD COM

WHITE

GRAY

SEN 2CTA

RTD SHIELD

CLEAR

Figure 10. Wiring for Xmt-C-11 (Pipe or Wall)

Figure 11. Wiring for 1066

Emerson Process Management Rosemount 402VP — TB1 diagram (page 6)
Fig. 14

TB1

RTD SHIELD

CLEAR

TB1 TB2

TB1

WHITE RED/WHITE

RTD COM

RTN

WHITE

RCV B

WHITE

RTD SENSE

SENSE

WHITE/RED

WHITE/RED

RCV A

RED

RTD IN

RTD IN

CLEAR

RSHLD

RED

RED

CLEAR

SHLD

CLEAR

CLEAR

GRAY

GRAY

DRVB

CLEAR

RECEIVE COM

ORANGE

ORANGE

DRVA

ORANGE

RECEIVE

DSHLD

CLEAR

CLEAR

CLEAR

DRIVE COM

GRAY

DRIVE

Figure 12. Wiring for 5081-C

Figure 13. Wiring for 6081-C

MODEL 5081 TRANSMITTER

Emerson Process Management Rosemount 402VP — MODEL 5081 TRANSMITTER diagram (page 6)
Fig. 15

SHIELD

2CT-A

2CT-B

DRV SHLD

RTD SHIELD

SHIELD

RCV COM

DRV COM RCV

Emerson Process Management Rosemount 402VP — SHIELD diagram (page 6)
Fig. 16

RTD IN

4CT-A

RCV SHLD

DRV

RTD IN

SENSE

4CT-B

RTD SENSE

RTD COM

RTD RTN

SMART PWR

RTD SHLD HT/FF (-)

RESERVED

HT/FF (+)

WHITE

RED/WHITE

RED

CLEAR

CLEAR

ORANGE

GRAY

CLEAR

CLEAR

WHITE

WHITE/RED

RED

CLEAR

ORANGE

CLEAR

GRAY

Page 7 of the Emerson Process Management Rosemount 402VP user manual — Pin Out Diagram for 402VP
Page 7 · click to enlarge
Page 7

Wiring through a Junction Box If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point. Use interconnecting cable 23747-00 (factory-terminated) or 9200275 (no terminations).

Pin Out Diagram for 402VP

Figure 14. VP pin-out

Emerson Process Management Rosemount 402VP — Pin Out Diagram for 402VP diagram (page 7)
Fig. 17 · Pin Out Diagram for 402VP

Cleaning the Sensor Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol (rubbing alcohol) can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors. Calibration ENDURANCE conductivity sensors are calibrated at the factory and do not need calibration when first placed in service. Simply enter the cell constant printed on the label into the analyzer.

After a period of service, the sensor may require calibration. The sensor can be calibrated against a solution having known conductivity or against a referee meter and sensor. If using a standard solution, choose one having conductivity in the recommended operating for the sensor cell constant. Refer to the analyzer manual or product data sheet for recommended ranges. Do not use standard solutions having conductivity less than about 100 uS/cm. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25 °C). Because 0.01/cm sensors must be calibrated in low conductivity solutions, they are best calibrated against a referee meter and sensor in a closed system. For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website. Troubleshooting

ProblemProbable CauseSolution Off-scale readingWiring is wrong.Verify wiring. Temperature element is open or shorted.Check temperature element for open or short circuits. See Figure 15 (next page). Sensor is not in process stream.Be sure sensor is completely submerged in process stream. Variopol cable is not properly seated.Loosen connector and reseat. Sensor has failed.Perform isolation checks. See Figure 16 (next page). Noisy readingSensor is improperly installed in process stream.Be sure sensor is completely submerged in process stream. Variopol cable is not properly seated.Loosen connector and reseat. Reading seems wrong (lower or higher than expectedBubbles trapped in sensor.Be sure sensor is properly oriented in pipe or flow cell. See Figure 1. Apply back pressure to flow cell. Wrong temperature correction algorithm.Check that temperature correction is appropriate for the sample. See analyzer manual for more information. Wrong cell constant.Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. Sluggish responseElectrodes are fouled.Clean electrodes. Sensor is installed in dead area in piping.Move sensor to a location more representative of the process liquid. 7
Page 8 of the Emerson Process Management Rosemount 402VP user manual — Pin Out Diagram for 402VP
Page 8 · click to enlarge
Page 8

Rev. L

Figure 15. Checking Temperature Element Disconnect leads and measure resistances shown. The mea sured resistance should be close to the value in the table.

Figure 16. Checking Continuity and Leakage Disconnect electrode leads and measure resistance and continuity as shown. Sensor must be dry when checking resistance between electrode leads.

Emerson Process Management Rosemount 402VP — Pin Out Diagram for 402VP diagram (page 8)
Fig. 18 · Pin Out Diagram for 402VP
Emerson Process Management Rosemount 402VP — Pin Out Diagram for 402VP diagram (page 8)
Fig. 19 · Pin Out Diagram for 402VP

facebook.com/EmersonRosemountAnalytical

AnalyticExpert.com

Credit Cards for U.S. Purchases Only.

twitter.com/RAIhome

youtube.com/user/RosemountAnalytical

Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 rosemountanalytical.com

©2014 Rosemount Analytical, Inc. All rights reserved.

The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners.

The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.

© Rosemount Analytical Inc. 2013

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Off-scale readingWiring is wrong.Verify wiring. p.7
Off-scale readingTemperature element is open or shorted.Check temperature element for open or short circuits. See Figure 15 (next page). p.7
Off-scale readingSensor is not in process stream.Be sure sensor is completely submerged in process stream. p.7
Off-scale readingVariopol cable is not properly seated.Loosen connector and reseat. p.7
Off-scale readingSensor has failed.Perform isolation checks. See Figure 16 (next page). p.7
Noisy readingSensor is improperly installed in process stream.Be sure sensor is completely submerged in process stream. p.7
Noisy readingVariopol cable is not properly seated.Loosen connector and reseat. p.7
Reading seems wrong (lower or higher than expectedBubbles trapped in sensor.Be sure sensor is properly oriented in pipe or flow cell. See Figure 1. Apply back pressure to flow cell. p.7
Reading seems wrong (lower or higher than expectedWrong temperature correction algorithm.Check that temperature correction is appropriate for the sample. See analyzer manual for more information. p.7
Reading seems wrong (lower or higher than expectedWrong cell constant.Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. p.7
Sluggish responseElectrodes are fouled.Clean electrodes. p.7
Sluggish responseSensor is installed in dead area in piping.Move sensor to a location more representative of the process liquid. p.7
Sluggish response p.7

Ask about the Emerson Process Management Rosemount 402VP

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Process Management Rosemount 402VP

Be the first to ask — other owners or our team will answer.

Original document

Emerson Process Management Rosemount 402VP manual PDF

8 pages·763 KB·Marine Instruments·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…