Instruction Manual 90003752 11/2003

Instruction Manual AK Protocol Communication Option for MLT 1, MLT 2, MLT 3, MLT 4, MLT 5 and CAT 200

3 rd Edition 11/2003

Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — 3 rd Edition 11/2003 diagram (page 1)
Fig. 1 · 3 rd Edition 11/2003
Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — 3 rd Edition 11/2003 diagram (page 1)
Fig. 2 · 3 rd Edition 11/2003
Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — 3 rd Edition 11/2003 diagram (page 1)
Fig. 3 · 3 rd Edition 11/2003
Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — 3 rd Edition 11/2003 diagram (page 1)
Fig. 4 · 3 rd Edition 11/2003
Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — 3 rd Edition 11/2003 diagram (page 1)
Fig. 5 · 3 rd Edition 11/2003

www.EmersonProcess.com

Page 2 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING!
Page 2 · click to enlarge
Page 2

AK Protocol for MLT 1, 2, 3, 4, 5 & CAT 200 Instruction Manual

ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING!

Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using and maintaining Emerson Process Management (Rosemount Analytical) products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.

  • Read all instructions prior to installing, operating, and servicing the product.
  • If you do not understand any of the instructions, contact your Emerson Process Management (Rosemount Analytical) representative for clarification.
  • Follow all warnings, cautions, and instructions marked on and supplied with the product.
  • Inform and educate your personnel in the proper installation, operation, and maintenance of the product .
  • Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes . Connect all products to the proper electrical and pressure sources.
  • To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
  • When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson Process Management (Rosemount Analytical). Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY . Look-alike substitutions may result in fire, electrical hazards, or improper operation.
  • Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury .

The information contained in this document is subject to change without notice. Misprints reserved.

1st Edition 10/1998 2nd Edition 11/2001 3rd Edition 11/2003 © 2003 by Emerson Process Management

Emerson Process Management GmbH & Co. OHG Industriestrasse 1 D-63594 Hasselroth Germany T +49 (0) 6055 884-0 F +49 (0) 6055 884-209 Internet: www.EmersonProcess.com

Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! diagram (page 2)
Fig. 6
Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! diagram (page 2)
Fig. 7

90003752 11/2003

Page 5 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — I) V24/RS232/485-Interface - Basics
Page 5 · click to enlarge
Page 5

I) V24/RS232/485-Interface - Basics

Protocol settings of a serial interface between a test bench control computer and peripheral analyzers on exhaust test benches

1. Introduction

The serial interface is made for slow point to point connections (f ≤ 10 Hz). The communication between the test bench control computer (TBCC) and the peripheral analyzers works according to the master slave principle. That means that the peripheral analyzers will only answer with a response telegram to the command telegram of the TBCC. They will not send an own message. You can distinguish two cases: (1) Analyzers in a function unit (system)

Some analyzers are combined to a logical unit. They are connected to the TBCC via an front-end computer. In that case the communication will not take place directly between the TBCC and the analyzers, but between the TBCC and the front-end computer. Each analyzer or the whole system unit will be identified by a defined channel number: K0 is the channel number for the whole defined system. ("Assembling command resp. assembling report") Kn (n=1, nmax) is the channel number for each analyzer. KV is the channel number for the front-end computer. (2) Single analyzers

Each analyzer is connected directly to the TBCC. In that case the identification of each analyzer will be done by the hardware connections and not by a software control. That is why the two channel number bytes (Kn) could be deleted. But in spite of that the channel number is generally 0 (K0) to get a uniform protocol. The data transfer will only be done by ASCII code to get an easy handling of the protocol with a terminal for simulation of the TBCC, the system unit and the analyzers. Therefore, no parity check will be done as data saving.

Page 6 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — NC
Page 6 · click to enlarge
Page 6
2. Hardware
1. Baud rate: 1200, 2400, 4800, 9600, 19200
2. Length of signs: 1 start bit 7 or 8 data bits 1 or 2 stop bits
3. Parity: even/odd/none
4. Operating: full duplex, no echo
5. Handshake: Xon/Xoff
6. Plug: 9 pin sub d, socket
7. Pin assignment: RS 232 module
Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — GND diagram (page 6)
Fig. 8

GND

Relay 1 contact NC/NO Relay 2 contact NC/NO Relay 3 contact NC/NO Relay common node

NC

GND

GND

Relay 1 contact NC/NO Relay 2 contact NC/NO Relay 3 contact NC/NO Relay common node

Show 7 more labels / short text on this page
Rxd · TxD · RS 485 module · RxD- · RxD+ · TxD+ · TxD-
Page 7 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — I) V24/RS232/485-Interface - Basics
Page 7 · click to enlarge
Page 7

I) V24/RS232/485-Interface - Basics

3. Protocol settings

The data and command transfer protocol has the following structure: 3.1. Command telegram

STXHE DON'T CARE FUNCT. CODE 1 FUNCT. CODE 2 FUNCT. CODE 3 FUNCT. CODE 4 BLANKVARIAB (number w digits p (othe can also d dependin functio "K" NUMBER D A T A ETXEN
Page 8 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — GND
Page 8 · click to enlarge
Page 8

3.2. Response telegram

STXHE DON'T CARE FUNCT. CODE 1 FUNCT. CODE 2 FUNCT. CODE 3 FUNCT. CODE 4 BLANKFIX DA ERROR STATUS D A T AVARIAB (can also dependin functio ETXEN
Page 9 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — I) V24/RS232/485-Interface - Basics
Page 9 · click to enlarge
Page 9

I) V24/RS232/485-Interface - Basics

3.3. Command telegram for RS485 BUS operating

STXHE BUS ADDRESS FUNCT. CODE 1 FUNCT. CODE 2 FUNCT. CODE 3 FUNCT. CODE 4 BLANKVARIAB (Number w digits p (othe can also d dependin functio "K" NUMBER D A T A ETXEN
Page 10 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — GND
Page 10 · click to enlarge
Page 10

3.4. Response telegram for RS485 BUS operating

STXHE BUS ADDRESS FUNCT. CODE 1 FUNCT. CODE 2 FUNCT. CODE 3 FUNCT. CODE 4 BLANKFIX DA ERROR STATUS D A T AVARIAB (can also dependin functio ETXEN
Page 11 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — I) V24/RS232/485-Interface - Basics
Page 11 · click to enlarge
Page 11

I) V24/RS232/485-Interface - Basics

4. Specifications of data settings
  • 4.1. Head telegram (Header) The begin of each transfer is a "STX" in the first byte. Each "STX" will start a new transfer. Previous transfers will be deleted, if they are not finished by "ETX". That means, only completed telegrams may be interpreted and answered. You can take any content for the "DON'T CARE" byte, excluding control signs or signs reserved by the AK commands. For the RS485 BUS operating an address byte will be used instead of the "DON'T CARE" byte. The analyzers will only answer to this command if the bus address setup will concur with this byte. In the command telegram a function code will be sent to the system unit or the analyzer with the four function bytes. In the response telegram this function code will be sent back as an echo if the transfer is successful. The echo will be four question marks (????), if
  • the command telegram has not minimum the number of bytes of the head telegram, the channel number in the data part and the end telegram (number of bytes = 10; using a channel number with two digits = 11 bytes) or
  • the function code has errors or is unknown. The function code may not contain blanks. There are three groups of function codes: (1) Control commands (2) Read commands (3) Write commands
Page 12 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — GND
Page 12 · click to enlarge
Page 12

4.2. Data block and error status byte The data presentation is variable. A fixed format will not be used. A blank or a with will be used as separating characters of data. The separation with will only be done, if the following complete date will have more than 60 digits. Each data set will begin normally with a blank. The data block of the command telegram has only variable data. These data depend on the function code. They can disappear for some function codes excluding the channel number. The channel number can have more than two bytes. The data block of the response telegram is divided in fixed and variable data. The first digit of the fixed data is a blank followed by an error status byte. The error status number will be zero for an error free running analyzer or system unit. The error status number will be counted up from 1 to 9 with each change in the error status. The error status number will be zero again after the errors will be removed. Changing the status of the system will not change the error status number. The variable data depend on the function code. They can disappear for some function codes. The long and variable floating point format or the E- Format are allowed to display the digits of numbers. You can find in each analyzer protocol which of these formats may be used. The decimal point can disappear for integers. The "+/-" sign may only be used for negative numbers. Digits without physical meaning have to be vanished. You can distinguish the following cases if a date with an error exists for a reading: (1) The transfer of the date is not possible, e.g. an analyzer in a system is missing or it

cannot send a signal. → The date will be replaced by a "#". (2) The date is only valid with restrictions, e.g. FID temperature too low.

→ The date will begin with a "#". Range overflow and range underflow will be displayed in the same way. "Valid" means that no criterions of plausibility will be considered. Example: You ask for a concentration value and the analyzer is in the "stand-by" mode. The date must not be marked with "#" as "valid with restrictions", if the analyzer would work normally in the operation mode.

Page 13 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — I) V24/RS232/485-Interface - Basics
Page 13 · click to enlarge
Page 13

I) V24/RS232/485-Interface - Basics

If an analyzer or a system is not in the "REMOTE" status, the control and write commands have to report "OF" ("Offline") in the data set to the. In system units the channel number has to be reported, too. If one analyzer is missing, a system unit has to send the channel number and "NA" ("Not Available") to the test bench control computer with control and write commands. A response telegram is not possible, if the test bench control computer has a direct contact to the analyzers and one analyzer is missing or the whole system is missing. So the test bench control computer has to realize the missing of devices by "Time Out". If the system or the analyzer is occupied by executing a function, the new start of a control command will lead to the response "BS" (Busy) in the data block of the response telegram. The running function will not be disturbed. Exception: The order was a software reset. If the data or parameters transfer is not complete (i.e. not expected format) in the command telegram to the system or the analyzer, the test bench control computer will get a "SE" (Syntax Error) in the data block of the following response telegram. If the system or the analyzers cannot work with the data or the parameters of the command telegram (data error, parameter error), the test bench control computer will get a "DF" (data error) in the data block of the following response telegram. 4.3. End of telegram Each transfer will end with "ETX" in the last byte.

Page 14 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — GND
Page 14 · click to enlarge
Page 14

4.4. Command and response telegram timing For each command telegram of the TBCC the peripheral analyzer sends a response telegram. It is not allowed for the TBCC to send new command telegrams before this response telegram is received. For most of the command telegrams this response might be sent after some milliseconds. But for some commands the response telegram might be started to sent 2-3 seconds after receiving the command telegram. The time starts the moment the peripheral analyzer receives the ETX character of the command telegram. It is not assured that the response telegram is sent without any delay between the single characters but it might be that there are up to 2-3 seconds in between. 4.5. Handling of malfunctions For the case of a malfunction it might be that the ETX-character of the command telegram is not received by the peripheral analyzer. Then it will not evaluate the command telegram and so will not send a response telegram. In this case it is in the responsibility of the TBCC not to run into an endless loop. It has to have a timeout of 4-5 seconds to react on this communication failure. The TBCC may repeat the last command telegram or doing some other exception handling.

Page 15 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 15 · click to enlarge
Page 15

I) V24/RS232/485-Interface - Basics

5. Examples for potential responses to control or write commands resp. to command telegrams with data (format) errors:
1. Analyzer and/or system unit with several analyzers "Online" and called analyzers are existing.
1. Byte STX
2. Byte DON'T

CARE

3. Byte C
4. Byte O
5. Byte D
6. Byte E
7. Byte BLANK
8. Byte x Error status byte

evtl. variable

Data

n. Byte ETX

Error status byte: Value is zero: Device without error. Value is not zero: Device with one or more errors.

Page 16 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 16 · click to enlarge
Page 16
2. Analyzer and/or system unit with several analyzers "Offline" and called analyzers are existing.
1. Byte STX
2. Byte DON'T

CARE

3. Byte C
4. Byte O
5. Byte D
6. Byte E
7. Byte BLANK
8. Byte x Error status byte
9. Byte BLANK
10. Byte K
11. Byte n
12. Byte BLANK
13. Byte O
14. Byte F

evtl. variable

Data

n. Byte ETX

Error status byte: Value is zero: Device without error. Value is not zero: Device with one or more errors. 11. Byte: Channel number is zero: "The whole system unit offline". Channel number is one to n: "Single analyzer offline".

Page 17 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 17 · click to enlarge
Page 17

I) V24/RS232/485-Interface - Basics

3. Called system unit "online", called single analyzer not available.

If the test bench control computer will call the devices directly and the system unit or the analyzer are not available, you will not get any response telegram. So, the test bench control computer will have to realize the missing of the system or of the analyzer by "Time Out".

1. Byte STX
2. Byte DON'T

CARE

3. Byte C
4. Byte O
5. Byte D
6. Byte E
7. Byte BLANK
8. Byte x Error status byte
9. Byte BLANK
10. Byte K
11. Byte n
12. Byte BLANK
13. Byte N
14. Byte A
15. Byte ETX

Error status byte: Value is zero: Device without error Value is not zero: Device with one or more errors 11. Byte: Channel number one to n: "Called device not available".

Page 18 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 18 · click to enlarge
Page 18
4. Called system unit "offline", called single analyzer not available.

If the test bench control computer will call the devices directly and the system unit or the analyzer are not available, you will not get any response telegram. So, the test bench control computer will have to realize the missing of the system or of the analyzer by "Time Out".

1. Byte STX
2. Byte DON'T

CARE

3. Byte C
4. Byte O
5. Byte D
6. Byte E
7. Byte BLANK
8. Byte x Error status byte
9. Byte BLANK
10. Byte K
11. Byte 0
12. Byte BLANK
13. Byte O
14. Byte F
15. Byte BLANK
16. Byte K
17. Byte n
18. Byte BLANK
19. Byte N
20. Byte A
21. Byte ETX
Page 19 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — I) V24/RS232/485-Interface - Basics
Page 19 · click to enlarge
Page 19

I) V24/RS232/485-Interface - Basics

Error status byte: Value is zero: Device without error. Value is not zero: Device with one or more errors. 11. Byte: Channel number zero: "System unit offline" 17. Byte: Channel number one to n: "Called device not available.

Page 20 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 20 · click to enlarge
Page 20
5. Called unit or channel is busy with a running function.
1. Byte STX
2. Byte DON'T

CARE

3. Byte C
4. Byte O
5. Byte D
6. Byte E
7. Byte BLANK
8. Byte x Error status byte
9. Byte BLANK
10. Byte K
11. Byte n
12. Byte BLANK
13. Byte B
14. Byte S
15. Byte ETX

Error status byte: Value is zero: Device without error. Value is not zero: Device with one or more errors. 11. Byte: Channel number is zero: "The whole unit is busy". Channel number is one to n: "Single analyzer is busy".

Page 21 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 21 · click to enlarge
Page 21

I) V24/RS232/485-Interface - Basics

6. The data are incomplete or the data do not have the expected format.
1. Byte STX
2. Byte DON'T

CARE

3. Byte C
4. Byte O
5. Byte D
6. Byte E
7. Byte BLANK
8. Byte x Error status byte
9. Byte BLANK
10. Byte K
11. Byte n
12. Byte BLANK
13. Byte S
14. Byte E
15. Byte ETX

Error status byte: Value is zero: Device without error. Value is not zero: Device with one or more errors.

Page 22 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 22 · click to enlarge
Page 22
7. The data or the parameters do not have the expected size.
1. Byte STX
2. Byte DON'T

CARE

3. Byte C
4. Byte O
5. Byte D
6. Byte E
7. Byte BLANK
8. Byte x Error status byte
9. Byte BLANK
10. Byte K
11. Byte n
12. Byte BLANK
13. Byte D
14. Byte F
15. Byte ETX

Error status byte: Value is zero: Device without error. Value is not zero: Device with one or more errors.

Page 23 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — I) V24/RS232/485-Interface - Basics
Page 23 · click to enlarge
Page 23

I) V24/RS232/485-Interface - Basics

6. Function sequence and error status after the receipt of the 'SRES' or 'STBY' commands
1. The test bench control computer is sending the control command SRES (Reset) to the system unit or any single analyzer.

All running functions or procedures will be canceled. An initializing will start, that is analogous to the switching on of the system unit or the analyzer: CPU and memory check, regulating or controlling of required temperatures, igniting of the flame in a FID an so on. The operation mode of the system or analyzer is "stand-by" during the initializing, even if the device is not ready and error free. That means, the status STBY will be reported to the read command ASTZ. The test bench control computer can only realize with the read command ASTF (error status), if the device is ready for measurements. The device will be ready to measure, if the essential functions of the current measuring instruction will be error free.

Example:

The status of the system unit or analyzer is SXYZ. The test bench control computer sends SRES:

Test bench control computer sends SRES Kn → System or analyzer will response SRES x

The system unit or the analyzer will cancel the status SXYZ. It will run a CPU and memory check and it will control the temperatures. If the temperatures are out of the allowed setpoint range, the device will regulate it. The FID will control the flame and will try to ignite it, if necessary, and so on. The test bench control computer will read the operation mode and the error status:

Test bench control computer sends ASTZ Kn → System or analyzer will response ASTZ 0 SMAN STBY

or Test bench control computer sends ASTZ Kn System or analyzer will response ASTZ x SMAN STBY

Test bench control computer sends ASTF Kn System or analyzer will response ASTF x n

The error status byte will be zero and the system or the analyzer will be ready to measure, if all temperatures are in the allowed setpoint range, if the FID flame is burning etc. If these parameters are not correct, the error status byte will be different from zero. The test bench control computer will read the operation mode and the error status as long as the system or the analyzer will be ready to measure. The test bench control computer will control the maximum time for this reading.

Page 24 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 24 · click to enlarge
Page 24
2. The test bench control computer is sending the control command STBY (Standby) to the system unit or any single analyzer.

There are two different cases: ! If the system or the analyzer is resting, this mode will be finished. Then, it will be tried to

get the stand-by mode ready for an error free measurement. The system or the analyzer will regulate all temperatures to the required setpoints, that were down during the resting. The FID will control the flame burning and if necessary it will try to ignite the flame etc. The operation mode of the system or analyzer is "stand-by" during these checkups, even if the device is not ready and error free. That means, the status STBY will be reported to the read command ASTZ. The test bench control computer can only realize with the read command ASTF (error status), if the device is ready for measurements. The device will be ready to measure, if the essential functions of the current measuring instruction will be error free.

Example:

The system or the analyzer is resting. No error is existing. The test bench control computer will ask for the operation mode: Test bench control computer sends ASTZ Kn → System or analyzer will response ASTZ 0 SREM SPAU Test bench control computer sends STBY. System or analyzer shall accept the stand-by mode: Test bench control computer sends STBY Kn → System or analyzer will response STBY 0 The system or analyzer is finishing the resting. Then, it will try to get the stand-by mode for an error free measurement. The system or analyzer will check the conditions: Are all temperatures in the setpoint range ? Is the FID flame burning ? etc. The test bench control computer will read the operation mode: Status is error free: Test bench control computer sends ASTZ Kn → System or analyzer will response ASTZ 0 SREM STBY or status has still some errors: Test bench control computer sends ASTZ Kn → System or analyzer will response ASTZ x SREM STBY Test bench control computer sends ASTF Kn → System or analyzer will response ASTF x n The error status byte will be zero and the system or the analyzer will be ready to measure, if all temperatures are in the allowed setpoint range, if the FID flame is burning etc. If these parameters are not correct, the error status byte will be different from zero. The test bench control computer will read the operation mode and the error status as long as the system or the analyzer will be ready to measure. The test bench control computer will control the maximum time for this reading.

Page 25 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — I) V24/RS232/485-Interface - Basics
Page 25 · click to enlarge
Page 25

I) V24/RS232/485-Interface - Basics

# The system or the analyzer is in the operation mode SXYZ. This mode will be finished.

Then, it will be tried to get the stand-by mode ready for an error free measurement. If there will be an error in the function SXYZ, the system or the analyzer will try to remove this error to get the stand-by mode ready for an error free measurement (i.e. FID flame is not burning, the FID will try to ignite). The operation mode of the system or analyzer is "stand-by" during these check-ups, even if the device is not ready and error free. That means, the status STBY will be reported to the read command ASTZ. The test bench control computer can only realize with the read command ASTF (error status), if the device is ready for measurements. The device will be ready to measure, if the essential functions of the current measuring instruction will be error free.

Example:

The system or the analyzer is in the operation mode SXYZ. An error is existing with the error number n, i.e. FID flame is not burning. The test bench control computer will ask for the error status:

Test bench control computer sends ASTF Kn → System or analyzer will response ASTF x n

Test bench control computer sends STBY. System or analyzer shall accept the stand-by mode and get ready for an error free measurement:

Test bench control computer sends STBY Kn → System or analyzer will response STBY x

The system or analyzer is finishing the operation mode SXYZ. Then, it will try to get the stand-by mode for an error free measurement. The system or analyzer will check the conditions and will try to remove the error, i.e. ignition of the FID flame. The test bench control computer will read the operation mode:

Error is removed (e.g. Flame was ignited):

Test bench control computer sends ASTZ Kn → System or analyzer will response ASTZ 0 SREM STBY

Error is still existing (e.g. Flame has not been ignited):

Test bench control computer sends ASTZ Kn → System or analyzer will response ASTZ x SREM STBY

Test bench control computer sends ASTF Kn → System or analyzer will response ASTF x n

The error status byte will be zero and the system or the analyzer will be ready to measure, if the error is removed i.e. the FID flame is still burning. If these parameters are not correct, the error status byte will be different from zero. The test bench control computer will read the operation mode and the error status as long as the system or the analyzer will be ready to measure. The test bench control computer will control the maximum time for this reading.

or

Page 26 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 26 · click to enlarge
Page 26
Page 26 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 27 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 27 · click to enlarge
Page 27

Specifications of the criterions and codes for the communication between

  • the front-end computer (system computer) of an exhaust analyzer system and the test bench control computer.
  • each analyzer of an exhaust analyzer system and the test bench control computer.
  • the front-end computer (system computer) of an exhaust analyzer system and their single devices. The following measurement systems and equipments can also be such single devices.
  • the front-end computer (system computer) of a fuel consumption analyzer and a test bench control computer.
  • the front-end computer (system computer) of an SHED measurement equipment and the test bench control computer. (SHED: Sealed Housing for Evaporative Determination)
  • the front-end computer (system computer) of a sampling system and the test bench control computer. The computer of an analyzer or of a system will be named as FU (Function Unit) and the test bench control computer will be abbreviated with TBCC.
Page 28 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 28 · click to enlarge
Page 28
1. Basic Informations

You can distinguish three cases: (1) Exhaust analyzer system: Some analyzers are combined logically. That means, these analyzers are connected with the TBCC via an front-end computer (system computer). The communication does not take place directly between the TBCC and the analyzers, but it will take place via the front- end computer. The identification of each device resp. of the whole system will be done by a channel number. K0 means the whole configured analyzer system ("assembling command resp. assembling report"). Kn (n=1, nmax) means each physical available analyzer. KV means the corresponding front-end computer. Some analyzers and the sampling devices or systems (e.g. CVS equipment, particle sampler, sampling system, etc.) are combined logically. That means, these analyzers are connected with the TBCC via an front-end computer (system computer). The communi- cation does not take place directly between the TBCC and the devices or systems, but it will take place via the front-end computer. The identification of each device and system will be done by a channel number. The handling of the analyzers will be like described above. All the other devices or systems can only be called directly by the corresponding channel number. The front-end computer must know the mnemonics of these devices and systems. Furthermore, the front-end computer has to be able to send orders and read commands to the channels resp. to send responses to the TBCC. (2) Single exhaust analyzers (single channel analyzers): All analyzers are connected to the TBCC individually. So, an identification of the analyzers by the software would not be necessary, because the analyzers are identified by their hardware connections. But to get a homogeneous protocol, the channel number will be indicated with K0. (3) Single exhaust analyzers (multi channel analyzers): All analyzers are connected to the TBCC individually, but they measure more than a single component (e.g. CO and C02). The identification by the software is necessary, because it will call the single channels resp. components. That is why such a single analyzer will be treated like a system.

Page 29 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 29 · click to enlarge
Page 29

The codes determined in this manual are valid for the communication between • the TBCC and the front-end computer of an exhaust analyzer system.

  • the front-end computer of an exhaust analyzer system and the corresponding single devices.
  • the TBCC and the single exhaust analyzers connected directly to the TBCC.
  • the TBCC and other exhaust measuring or analyzing devices or equipments connected directly to the TBCC. The floating point format is valid for the signal transfer. The physical units are determined as follows:
  • Exhaust values: ppm
  • Temperatures: K
  • Pressures: Pa
  • Flow: l/min The analyzer system or each analyzer can be set to the operation mode "MANUAL" selecting "REMOTE DISABLE" for the parameter "REMOTE EN-/DISABLE". This setup does not depend on the previous status of the system or analyzer. If you select "REMOTE ENABLE", the mode "MANUAL" will retain for the moment, but the TBCC can call this operation mode with a control command. If the TBCC will setup the system/the analyzer to "REMOTE", the system/the analyzer will execute control commands from the TBCC. Precondition: The system/the analyzer is able to start the function selected. In the mode "REMOTE DISABLE", the TBCC can only send read commands. It is only possible to read signals and status informations. If then the system or the analyzer is in the mode "MANUAL", it will ignore the control command from the TBCC. No change of the error status will be done in the response to the TBCC. Instead of that the response will display "MANUAL" as first date. The same will be valid, if the parameter is "REMOTE ENABLE", but the TBCC did switch the system/the analyzer to the mode "MANUAL". Otherwise, the operation mode can only be recognized by reading the status. This is also valid during a test is running.
Page 30 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 30 · click to enlarge
Page 30

If it is possible in a system to put single channels together to lines, so the following definition will be valid:

A line is the summary of "1-x" analyzers to a logical group "y", that can be switched physically to a gas channel "z". Each device can only be assigned to one line at the same time. If you will try to assign a channel to another line and this channel is already assigned, the front-end computer will send as response "DF" (data error). The organization of each defined line will be done in the front-end computer (CODE KV Ln ...). The order must be sent to the front-end computer "KV". A line will be dissolved by the configuration without assignment of channels (CODE KV Ln) or by the reset order (SRES). All available gas inputs can be assigned to a defined line. So it is possible to assign different lines to different gas sampling points, e.g. in front of a catalyst, behind a catalyst. If the gas running time will change in such cases, you have to regard for it.

Page 31 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 31 · click to enlarge
Page 31
2. List of all Codes

2.1. Control commands CODE Function Page

SALI Kn Mm Linearization check with spangas 2 - 8 SARA Kn Autoranging "OFF" (located range will remain) 2 - 9 SARE Kn Autoranging "ON" 2 - 9 SATK Kn Autom. calibration (zero and span calibration + zerogas + stand-by)

SEGA Kn Spangas (spangas will flow with time limit) 2 - 15 SEMB Kn Mm Set range (1, 2, 3, 4) 2 - 16 SFRZ Kn Decimal point setup for floating point format 2 - 18 SGTS Kn Device test 2 - 19 SHDA K0 Hold mode "OFF" 2 - 20 SHDE K0 Hold mode "ON" 2 - 20 SINT Kn Start integration (integral average) 2 - 21 SLCH Kn Mm Linearization check (with gas distribution) 2 - 22 SLIN Kn Mm Linearization (with determination of corrections and saving)

SLST Kn Set linearization step and get values 2 - 24 SMAN Kn Operation mode "Manual" 2 - 25 SMGA Kn Samplegas (will be sucked or will be on) 2 - 26 SNAB Kn Zerogas calibration 2 - 27 SNGA Kn Zerogas (zerogas will flow with time limit) 2 - 28 SPAB Kn Spangas calibration 2 - 29 SPAU Kn Pause (resting status, e.g. pumps, ozonator, deozonator, high voltage off, gas input closed within the device)

SQEF Kn Mm Cross interference (only for CO analyzers) 2 - 31 SREM Kn Operation mode "Remote" 2 - 32 SRES Kn Reset (analyzer will change via initializing mode to stand-by) 2 - 33 SROF Kn Delay modus "OFF" (operation mode: signal output with/without delay time, involved are SINT, AKON, AIKO, AIKG, analog signal)

SRON Kn Delay modus "ON" (analog to "SROF") 2 - 34 SSPL Kn Purging (purge air will be sucked or will be on) 2 - 35 ST9O Kn Set t90 time step (S = fast, M = medium, L = slowly) 2 - 36 STBY Kn Stand-by (get ready for measurement no matter of the previous history)

Optional: SCAL Kn Start system calibration (only for platform) 2 - 13 SENO Kn NO measurement (operation mode CLD) 2 - 17 SNOX Kn NOx measurement (operation mode CLD) 2 - 17 Kn: Channel n Mm: Range m Response

The response to the control commands will contain the CODE of the control command and the error status byte (0-9).

Page 32 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 32 · click to enlarge
Page 32

2.2. Read commands

CODE Function Page AAEG Kn Spangas deviation 2 - 39 AALI Kn Mm Deviations of the last linearization checks with spangas 2 - 40 AANG Kn Zerogas deviation 2 - 41 AEMB Kn Selected range (R1, R2, R3, R4) 2 - 45 AFDA Kn CODE Zero- /spangas time (function length of calibration) 2 - 46 AFDA Kn CODE Purge time 2 - 46 AGID Kn Device identification 2 - 47 AGRW Kn m Limit 2 - 48 AIKG Kn Concentration integral average; all (in ppm) 2 - 49 AIKO Kn Concentration integral average (in ppm) 2 - 51 AKAK Kn Mm Calibration gas concentration (in ppm) 2 - 53 AKAL Kn Stored calibration corrections 2 - 54 AKFG Kn Configuration of the system 2 - 57 AKON Kn Concentration (current value in ppm) 2 - 58 ALCH Kn Mm Deviations of the last linearization checks 2 - 61 ALIK Kn Calculation of linearization curve 2 - 62 ALIN Kn Mm Linearization values in the device (X/Y = Setpoint-/raw value) 2 - 63 ALKO Kn Mm Polynomial coefficents of the linearization curve 2 - 64 ALST Kn Linearization steps 2 - 65 AM90 Kn Actual response time (t90 time) 2 - 66 AMBA Kn Mm Begin of range (in ppm) 2 - 67 AMBE Kn Mm End of range (in ppm) 2 - 68 AMBU Kn Switch levels for autoranging 2 - 69 AMDR Kn Manual adjusted pressure 2 - 70 AQEF Kn Cross interference check result (in ppm) (only for CO analyzers) 2 - 71 ASOL Kn Setpoint value with limits 2 - 72 ASTA Kn General status of the system 2 - 73 ASTF Kn Internal error status 2 - 74 ASTZ Kn Device status 2 - 76 ASYZ Kn System time (year, month, day, hour, min., sec.) 2 - 78 AT9O Kn T90 time (response time) 2 - 79 ATEM Kn m Temperature 2 - 80 ATOL Kn Mm Stability tolerances 2 - 81 AVEZ Kn Delay and synchronization time 2 - 83 AZEI Kn CODE Times (for procedures) 2 - 84 Optional:

ABST Kn Counter of operating hours 2 - 42 ADRU Kn m Pressure (for service) 2 - 43 ADUF Kn m Flow (for service) 2 - 44 AKEN Kn Device tag 2 - 56 AKOW Kn Mm Correction (zerogas calibration and gradient) 2 - 60 AUKA Kn Uncorrected analog value 2 - 82 Kn: Channel n Mm: Range m Response The response to the read command will contain the CODE of the read command, the error status byte (0-9) and the data.

Page 33 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 33 · click to enlarge
Page 33

2.3. Write commands CODE Function Page

EFDA Kn CODE DATA

Zero- /spangas time (function length of calibration) 2 - 85

EFDA Kn CODE Purge time 2 - 85 EGRW Kn DATA Limits 2 - 87 EKAK Kn Mm DATA

Calibration gas concentration (in ppm) Value = 0: no spangas available

EKFG Kn DATA Configuration of the system 2 - 90 ELIN Kn Mm DATA

Linearization values in the device (X/Y = Setpoint-/raw value) 2 - 91

ELKO Kn DATA Polynomial coefficents of the linearization curve 2 - 92 ELST Kn DATA Linearization steps 2 - 93 EMBA Kn Mm DATA

Begin of range (in ppm) Value = 0: no range defined

EMBE Kn Mm DATA

End of range (in ppm) Value = 0: no range defined

EMBU Kn DATA Switch levels for autoranging 2 - 96 EMDR Kn DATA Manual adjusted pressure 2 - 97 ESOL Kn m DATA Setpoint value with limits 2 - 98 ESYZ Kn DATA System time (year, month, day, hour, min., sec.) 2 - 99 ET9O Kn DATA T90 time (response time) 2 -100 ETOL Kn Mm DATA

Stability tolerances 2 -101

EVEZ Kn DATA Delay and synchronization time 2 -102 EZEI Kn CODE DATA

Times (for procedures) 2 -103

Optional:

EKEN Kn DATA Device tag 2 - 89 Kn: Channel n Mm: Range m Response The response to the write commands will contain the CODE of the write command and the error status byte (0 - 9).

Page 34 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 34 · click to enlarge
Page 34
3. Description of all Control Commands

SALI – Control command "Linearization check with spangas" Starting this command the analyzer in a system or the single analyzer will start the spangas flow with all available spangases one after another. It will check the setpoint values automatically. The device will record the determined values and store the raw/ setpoint deviations. These deviations can be read with the command "AALI". Control command SALI Kn Mx Function associated to channel n, range x Code Response SALI 0 Error status Code

Page 35 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 35 · click to enlarge
Page 35

SARA – Control command "Autoranging OFF" SARE – Control command "Autoranging ON" Starting the command "Autoranging ON" the analyzers in a system or the single analyzer will select the best range for the current concentration automatically. With the control command "SEMB" the autoranging will be stopped and the range sent with the "SEMB" command will be selected. The command "Autoranging OFF" will stop this function, but the found range will remain. Control command SARE K0 SARA K0 Function associated to the whole system unit Code SARE Kn SARA Kn Function associated to a single analyzer Code

Page 36 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 36 · click to enlarge
Page 36

SATK – Control command "Automatic calibration" Starting this command the analyzers in a system or the single analyzer will start a calibration procedure to determine the correction values. The required calibration gases and the pumps will be switched on automatically. Then, the calibration procedure will start automatically. Such a procedure has to run until the end. It may not be canceled or interrupted by other functions. Otherwise it is not sure that the correction values will be valid to calculate the exhaust values of analysis. Exceptions are the commands "Reset" or "Stand-by". After the procedure will be over the system, the analyzers in a system or the single analyzer will have to change to the operation mode "Stand-by". Control command SATK K0 Function associated to the whole system unit Code SATK K1 Kn Function associated to channel n Function associated to channel 1 Code SATK Kn M3 Function associated to channel n and range 3 Code Response SATK 0 Error status Code

Page 37 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — CARE
Page 37 · click to enlarge
Page 37

Parameters:

Wait dead time Tt, if gases were switched

To Time-out [xx s] (max. time for stability control) Tt Dead time [xx s] (Wait after gas switching) Ti Integration time [xx,x s] (for K1, Knew) Ts Stability time [xx s] Tol. Tolerance [x,x % range]

Start timer for time-out To

Calculate concentration mean value K1 over time Ti

Set and start timer

Calculate concentration mean value Knew over time Ti

Deviation = | K1 - Knew|

Dev. ≥ Tol.

Show 16 more labels / short text on this page
Stability control procedure · for stability time · Yes · Signal not · Time out ? · stable · No · Yes · K1 = Knew · No · No · Stability · time over ? · Yes · Signal stable · Signal = Knew
Page 38 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 38 · click to enlarge
Page 38

Begin of gas flow

Begin of stability control

New begin of stability control

K1

2 * Deviation K1

Dead time Tt Time-out

Stability controlled procedure of the zero/span calibration

Average over Ti, Read with AAEG

Begin of the function SATK

Average over Ti, Read with AANG

Show 17 more labels / short text on this page
Stability control · Signal = Knew · Knew · Knew · Ti · Stability time Ts · Calibration · Ti Ti · Calibration · Dead time Stability · Dead time Stability · Stability · Stability · time · time · time · time
Page 39 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 39 · click to enlarge
Page 39

Only for platform: SCAL – Control command "Start system calibration" To control the system calibration procedures the commands "SCAL", "STBY" and "ASTZ" have to be used. With "SCAL" the procedures will be started. For more exact description of procedures see also the "documentation of system calibration". Starting condition: All attached analyzer module are in the stand-by mode ("AK STBY") and the variable "CALSTAT" is "0". Otherwise the response is "BUSY" (BS). Control command SCAL Kx m (n) Optional parameters Type of system calibration Function associated to channel x Code

m Kx n 0 = ZERO-CAL K0 n = 1: switch into test mode 1 = ZERO/SPAN-CAL K0 2 = PROGRAM K0 else: switch into normal mode 3 = TEST ZERO-GAS K1...999 time-out in sec 4 = TEST SPAN-GAS1 K1...999 time-out in sec 5 = TEST SPAN-GAS2 K1...999 time-out in sec 6 = TEST SPAN-GAS3 K1...999 time-out in sec 7 = TEST SPAN-GAS4 K1...999 time-out in sec 8 = TEST CLOSE GASES K1...999 time-out in sec If optional parameter "n" is not in the command string the appropriate variable will not be changed. Stop command

Emerson Process Management Multi-Component Gas Analyzer (MLT 1) — K1 diagram (page 39)
Fig. 9
Page 40 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 40 · click to enlarge
Page 40

STBY K0 Function associated to the whole system unit Code Only using the channel number 0 (K0) will stop running "SYSCAL" procedure. Besides, all procedures of the other analyzer modules will be stopped. Response SCAL 0 Error status Code Read command ASTZ K0 Read of the whole FU Code With the command "ASTZ K0" it will be checked, if a system calibration is running or not. "SCAL" will be sent back for a running system calibration. If no system calibration is running this string will be missed.

Page 41 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 41 · click to enlarge
Page 41

SEGA – Control command "Spangas" Starting this command the analyzers in a system or the single analyzer will switch on the calibration valve to spangas and switch on the required pumps. This function will only check the end point. It will not correct the calibration. If continuous line recorders will be available, the paper transport will also be switched on. Control command SEGA K0 Function associated to the whole system unit Code SEGA K1 Kn Function associated to channel n Function associated to channel 1 Code Response SEGA 0 Error status Code

Page 42 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 42 · click to enlarge
Page 42

SEMB – Control command "Set range" Starting this command the analyzers in a system or the single analyzer will set the range that is named in the data. If the function "Autoranging" is running at that moment, it will be stopped and the named range will be selected. Control command SEMB K1 M4 Kn M2 Function associated to channel n range 2 Function associated to channel 1 range 4 Code Response SEMB 0 Error status Code

Page 43 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 43 · click to enlarge
Page 43

Only for CLD module: SENO/SNOX – Control command "Operation mode CLD" Starting the command "SENO" the CLD analyzers in a system or the single CLD analyzer will start the NO measurement. The command "SNOX" will start the NOx measurement. Control command SENO Kn Function associated to channel n Code for NO measurement SNOX Kn Function associated to channel n Code for NOx measurement Response SENO 0 Error status Code for NO measurement SNOX 0 Error status Code for NOx measurement

Page 44 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 44 · click to enlarge
Page 44

SFRZ – Control command "Decimal point setup for floating point format" With this command the format of real numbers will be setup. This command will have an effect to the output of all real numbers. It is not possible to vary it for different channels. Control command SFRZ K0 n Parameter Function associated to the whole system unit Code Description of parameter “n” n = 1..9: n = number of digits after decimal point in normal floating point format. (f-notation of C-language's printf) n = 11..19: (n-10) = is the maximum number of significant digits of either normal floating point format or E-format, whichever is shorter. (g-notation of C-language's printf). Factory setting: n = 16 (= default).

Default is used if no other value for n is specified by a SFRZ command. n = 10 will set n = default

Example: Number: 1234567.821 n = 2 Æ 1234567.82 n = 13 Æ 1.23E06 n = 15 Æ 1234600

Response SFRZ 0 Error status Code

Page 45 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 45 · click to enlarge
Page 45

SGTS – Control command "Device test" Starting this command the analyzers in a system or the single analyzer will switch off the calibration gas and the samplegas. That means, all gas tubes to the analyzer device will be closed and the pumps will be switched off. Then, the device can be checked via a gas input that is located directly in front of the device. Control command SGTS K0 Function associated to the whole system unit Code SGTS K1 Kn Function associated to channel n Function associated to channel 1 Code Response SGTS 0 Error status Code

Page 46 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 46 · click to enlarge
Page 46

SHDE – Control command "Hold status ON" SHDA – Control command "Hold status OFF" We have the possibility to activate the "Hold"-feature not only per calibration. We can do this also by AK-Command "SHDE". With the command "SHDA" we have the possibility to deactivate an activated "Hold" again. Starting the command "SHDE" will switch on the "hold status". So it is possible to start the "Hold"-feature directly by AK command and not only per calibration. With the "SHDA" command the "Hold status" will be deactivated. Control command SHDE K0 SHDA K0 Function associated to the whole system unit Code Response SHDE 0 SHDA 0 Error status Code

Page 47 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 47 · click to enlarge
Page 47

SINT – Control command "Integrator" Starting this command the FU will activate the internal integrators. The previous calculated and stored integral averages will be set to zero. The integrator will calculate new integral averages as long as the control command "SINT" will be received again. The result of the integrator can be read with the command "AIKG". Control command SINT K0 Function associated to the whole system unit Code SINT K1 Kn Function associated to channel n Function associated to channel 1 Code Response SINT 0 Error status Code

Page 48 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 48 · click to enlarge
Page 48

SLCH – Control command "Linearization check" Starting this command the analyzers in a system or the single analyzer will switch on the gas tubing to a gas distribution and a linearization procedure will run. The device will record the correction values to the receiver specific raw curve. The deviations to the correction values of the last determined linearization will be stored. Look at the command "SLIN" for informations about the logic of the device control and of the gas distribution control. Control command SLCH Kn Mn Range n Function associated to channel n Code Response SLCH 0 Error status Code

Page 49 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 49 · click to enlarge
Page 49

SLIN – Control command "Linearization" Starting this command the analyzers in a system or the single analyzer will switch on the gas tubing to a gas distribution. A linearization procedure for the selected range will run. The device will record the determined correction values to the receiver specific raw curve. The values will be stored in the device to calculate the gas concentration. This procedure will be controlled by several commands of the TBCC or the system. The device or the gas distribution will only accept those commands, it they have already received the "SLIN" command. The "SLIN" command prepares the device or the gas distribution to receive and execute further commands being necessary for the linearization procedure. The spangas concentration has to be set up before by the "EKAK" command. Control command SLIN Kn Mn Range n Function associated to channel n Code Response SLIN 0 Error status Code

Page 50 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 50 · click to enlarge
Page 50

SLST – Control command "Set linearization step" Starting this command the gas distribution will switch on the named distribution step. The device will work like described for the commands "SLIN" or "SLCH" depending on the current procedure. The device will only accept the "SLST" command, if the commands "SLIN" or "SLCH" were received before followed by the command "ELST". Control command SLST K1 n Distribution step Function associated to channel 1 Code Response SLST 0 Error status Code

Page 51 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 51 · click to enlarge
Page 51

SMAN – Control command "Operation mode MANUAL" With this command the FU will change to the operation mode "Manual". Then it will only be possible to start functions from an operating unit integrated in the FU. The same operation mode will be enabled, if the service switch of the FU will be in the position "Remote Disable". In that mode it will only be possible to answer to read commands of the TBCC. Control command SMAN K0 Function associated to the whole system unit Code SMAN K1 Kn Function associated to channel n Function associated to channel 1 Code Response SMAN 0 Error status Code

Page 52 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 52 · click to enlarge
Page 52

SMGA – Control command "Samplegas" Starting this command the analyzers in a system or the single analyzer will switch on the sample gas valve and the pumps necessary for the samplegas transport. If continuous line recorders will be available, the paper transport will also be switched on. Control command SMGA K0 Function associated to the whole system unit Code SMGA K1 Kn Function associated to channel n Function associated to channel 1 Code Response SMGA 0 Error status Code

Page 53 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 53 · click to enlarge
Page 53

SNAB – Control command "Zerogas calibration" Starting this command the analyzers in a system or the single analyzer will start a zerogas calibration. The calibration gas flow will start automatically and the calibration procedure will run. After this procedure will be over the system, the analyzer in a system or the single analyzer will change to the stand-by mode. The running calibration procedure can be canceled with the "STBY" command. Control command SNAB K0 Function associated to the whole system unit Code SNAB K1 Kn Function associated to channel n Function associated to channel 1 Code Response SNAB 0 Error status Code

Page 54 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 54 · click to enlarge
Page 54

SNGA – Control command "Zerogas" Starting this command the analyzers in a system or the single analyzer will switch on the zerogas valve and the pumps necessary for the zerogas transport. This function will only check the zero point. It will not correct the calibration. If continuous line recorders will be available, the paper transport will also be switched on. Control command SNGA K0 Function associated to the whole system unit Code SNGA K1 Kn Function associated to channel n Function associated to channel 1 Code Response SNGA 0 Error status Code

Page 55 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 55 · click to enlarge
Page 55

SPAB – Control command "Spangas calibration" Starting this command the analyzers in a system or the single analyzer will start a spangas calibration. The calibration gas flow will start automatically and the calibration procedure will run. After this procedure will be over the system, the analyzer in a system or the single analyzer will change to the stand-by mode. The running calibration procedure can be canceled with the "STBY" command. Control command SPAB K0 Function associated to the whole system unit Code SPAB K1 Kn Function associated to channel n Function associated to channel 1 Code Response SPAB 0 Error status Code

Page 56 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 56 · click to enlarge
Page 56

SPAU – Control command "Pause" With this command the FU will be set to a defined status of interruption. This command will only be accepted, if the FU is already in the stand-by mode. The "SPAU" command will switch off the operation modes (e.g. FID flame, pump of an NO device) or the corresponding setpoints (e.g. temperature of the hot pipe). With the control command „Reset“ or „Stand-by“ the FU will change to the stand-by mode to get ready for operation. The real functionality of the "SPAU" command will depend on the used FU. It is part of each device or system specification. Control command SPAU K0 Function associated to the whole system unit Code SPAU Kn Function associated to channel n Code Response SPAU 0 Error status Code

Page 57 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 57 · click to enlarge
Page 57

SQEF – Control command "Cross interference" Starting this command the CO analyzer will measure wet C02. It will be produced by streaming three percent C02 through water bottles at 20 degrees Celsius. The CO analyzer will measure this gas mixture. The signal will be stored in the analyzer. It can be read by the TBCC with the "AQEF" command. The measured concentration has to be maximum 3 ppm for ranges smaller than 300 ppm. For bigger ranges it has to be maximum 1 % of the end of range value. These limits will be controlled by the TBCC. Control command SQEF Kn Function associated to channel n Code Response SQEF 0 Error status Code

Page 58 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 58 · click to enlarge
Page 58

SREM – Control command "Remote" With this command the FU will change to the computing operation mode. Then, the function start will only be possible by the TBCC. This operation mode may only be set, if the service switch of the FU is in the position "Remote Enable". Control command SREM K0 Function associated to the whole system unit Code SREM K1 Kn Function associated to channel n Function associated to channel 1 Code Response SREM 0 Error status Code

Page 59 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 59 · click to enlarge
Page 59

SRES – Control command "Reset" With this command the FU will get a software reset. This command has the same effect to the FU like the switching off and on of the power supply. All running procedures will be canceled. An initializing will be started, e.g. check and control of temperature setpoints. After that the operation modes "Manual" and "Stand-by" will be enabled. Control command SRES K0 Function associated to the whole system unit Code SRES Kn Function associated to channel n Code Response SRES 0 Error status Code

Page 60 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 60 · click to enlarge
Page 60

SRON – Control command "Delay modus ON" SROF – Control command "Delay modus OFF" Starting this command the analyzers in a system or the single analyzer will determine measurement and integral values (averages), that will be delayed according to the delay time of the write command "EVEZ". The read commands "AKON", "AIKO" and "AIKG" will get an old value according to the synchronization time of the command "EVEZ". The analog output will get the same delay. With the command "Delay modus OFF" the integrators will start immediately and the measurement and integral values will be determined and sent out without delay. Control command SRON K0 SROF K0 Function associated to the whole system unit Code SRON Kn SROF Kn Function associated to channel n Code Response SRON 0 SROF 0 Error status Code

Page 61 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 61 · click to enlarge
Page 61

SSPL – Control command "Purging" Starting this command the analyzers in a system or the single analyzer will switch on the purge gas valve and the pumps necessary for the purge gas transport. This function can be finished either by a new command or by a defined time interval. After the purging will be over the system, the analyzer in a system or the single analyzer will change to the stand-by mode. Control command SSPL K0 Function associated to the whole system unit Code SSPL Kn Function associated to channel n Code Response SSPL 0 Error status Code

Page 62 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 62 · click to enlarge
Page 62

ST9O – Control command "Set t90 time step" With this command the analyzer will use the t90 time according to the current step. The abbreviation "S" means fast time, "M" means medium time and "L" means slow time. After the switching on of the device or after a "Reset" the fastest time will be set. Control command ST9O K1 S Function associated to channel 1, fast time Code Response ST9O 0 Error status Code

Page 63 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 63 · click to enlarge
Page 63

STBY – Control command "Stand-by" With the command "Stand-by" the FU will be set to a defined status of interruption. Running functions like measuring or purging will be canceled. Then, the stand-by mode will be enabled. The ranges will keep selected. The FU will get ready for measurement and operating, no matter which mode was the previous. Control command STBY Kn Function associated to the whole system unit Code Response STBY 0 Error status Code

Page 64 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 64 · click to enlarge
Page 64
Page 64 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 65 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 65 · click to enlarge
Page 65
4. Description of all Read Commands

AAEG – Read command "Deviation to spangas" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The measured and stored signal of the last spangas measurement. ∗ The deviation from the setpoint value of the linearized curve in ppm and percent,

referred to the end of range value. Spangas measurement: Signal after the end of the functions "Automatic calibration" or "Spangas", stability controlled or time controlled depending on the setup in "EFDA". Read command

AAEG K0 AAEG Kn Read of the whole system unit Code Response AAEG 0 M1 XXX YYY ZZ ... Mn XXX YYY ZZ Deviation [%] Deviation [ppm] Signal [ppm] Range n Deviation [%] Deviation [ppm] Signal [ppm] Range 1 Error status Code The values will get the same format for the read of single channels.

Page 66 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 66 · click to enlarge
Page 66

AALI – Read command "Deviations of the last linearization check with spangas" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device) and subchannel (range): ∗ The determined and stored deviations in ppm of the last linearization check with

spangas. Read command AALI Kn Mx Read of channel n and range x Code Response AALI 0 AAA BBB XXX xth difference 2nd difference 1st difference Error status Code

Page 67 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 67 · click to enlarge
Page 67

AANG – Read command "Deviation to zerogas" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The determined and stored signal of the last zerogas measurement with its range. ∗ The deviation from the setpoint value of the linearized curve in ppm and percent,

referred to the end of range value. Zerogas measurement: Signal after the end of the functions "Automatic calibration" or "Zerogas", stability controlled or time controlled depending on the setup in "EFDA". Read command AANG K0 AANG Kn Read of the whole system unit Code Response AANG 0 M1 XXX YYY ZZ ... Mn XXX YYY ZZ Deviation [%] Deviation [ppm] Signal [ppm] Range n Deviation [%] Deviation [ppm] Signal [ppm] Range 1 Error status Code The values will get the same format for the read of single channels.

Page 68 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 68 · click to enlarge
Page 68

ABST – Read command "Counter of operating hours" To this read command the FU will send to the TBCC the following data: ∗ The operating hours until now for the roots fan, the turbo compressor, the sampling

pumps etc. The operating hours will only be sent as integers. Read command ABST K0 Read of the whole system unit Code Response ABST 0 T1 T2 ... Tn Value of each hour of operation Error status Code

Page 69 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 69 · click to enlarge
Page 69

ADRU – Read command "Pressure" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device) and if need be for the subchannel (pressure measurement): ∗ The signal in Pascal. Note: At the moment no subchannels will be used. Read command ADRU K0 ADRU Kn (m) Read of channel n (and subchannel m) Code Response ADRU 0 XXX Pressure value Error status Code

Page 70 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 70 · click to enlarge
Page 70

ADUF – Read command "Flow" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device) and subchannel (flow measurement): ∗ The signal in liter per time unit. Read command ADUF K0 ADUF Kn (m) Read of channel n (and subchannel m) Code Response ADUF 0 XXX Flow value Error status Code

Page 71 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 71 · click to enlarge
Page 71

AEMB – Read command "Selected range" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The selected and used range at this moment. Read command AEMB K0 AEMB Kn Read of channel n Code Response AEMB 0 Mn Range with setup Error status Code

Page 72 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 72 · click to enlarge
Page 72

AFDA – Read command "Function length" To this read command the FU will send to the TBCC the following data for the called channel (device): ∗ The function or procedure times of the function determined in "CODE". Functions like "SATK", "SLIN", "SLCH", "SALI", "SQEF", "SNGA" or "SEGA" will run time controlled according to the times T1 to T4 or stability controlled. Time control: If only T1 is set or if T2 = 0, time control will run with step time T1 (total function time). Stability control: Times T1 to T4 have to be set. Read command AFDA K0 CODE AFDA Kn CODE Code for the function Read of channel n Code Response AFDA 0 T1 (T2 T3 T4) "Time out"; after this time is over, the procedure will be canceled and you will get an error message; this time will start after the wait. Integration time to get the mean value of one signal. Stability time: All signals have to be in a certain tolerance during this time. Time to wait for: after the switching on resp. changing of gases or the stepping time of time control. Error status Code

Page 73 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 73 · click to enlarge
Page 73

AGID – Read command "Device identification" To this read command the gas analyzer will send to the TBCC a text string consisting of several data. These data will be separated by a slash ( / ). Read command AGID K0 Read of the whole system unit Code Response AGID 0 a/b/c Device identification a = Name and serial number b = Program version c = Date Error status Code

Page 74 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 74 · click to enlarge
Page 74

AGRW – Read command "Limits" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The corresponding limits, e.g. maximum deviations of calibration. Read command AGRW K0 m AGRW Kn m Read of channel n and subchannel m m = 0: Zerogas calibration m = 1: Spangas calibration Code Response AGRW 0 XXX Limit Error status Code

Page 75 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 75 · click to enlarge
Page 75

AIKG – Read command "Concentration integral value; all" To this read command the FU will send to the TBCC the following data: ∗ The corrected average signal valid at that moment (e.g. analyzed value), that has been

calculated since the last "SINT" command. The physical parameter is described in the section about FU. The value will be limited to six relevant digits, because it is useless to send gas concentrations in an accuracy less than pars pro mille. Look at the example of the "AKON" command ! Cf. the "SFRZ" command ! Read command AIKG K0 Read of the whole system unit Code AIKG K1 ... Kn Read of channel n Read of channel 1 Code

Page 76 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 76 · click to enlarge
Page 76

Response AIKG 0 123400 12340 1234 123.4 12.34 -1.23 # Channel 7 no signal, invalid or range overflow/underflow Channel 6 negative value 1 digit before/ 2 digits after decimal point. Channel 5 positive value 2 digit before / 2 digits after decimal point. Channel 4 positive value 3 digits before / 1 digit after decimal point. Channel 3 positive value 4 digits Channel 2 positive value 5 digits Channel 1 positive value 6 digits Error status Code The values will get the same format for the read of single channels.

Page 77 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 77 · click to enlarge
Page 77

AIKO – Read command "Concentration integral value" To this read command the FU will send to the TBCC the following data: ∗ The corrected average signal valid at that moment (e.g. analyzed value), that has been

calculated since the last "SINT" command resp. the last "AIKO" command. The physical parameter is described in the section about FU. The value will be limited to six relevant digits, because it is useless to send gas concentrations in an accuracy less than pars pro mille. Look at the example of the "AKON" command ! Read command AIKO K0 Read of the whole system unit Code AIKO K1 ... Kn Read of channel n Read of channel 1 Code

Page 78 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 78 · click to enlarge
Page 78

Response AIKO 0 123400 12340 1234 123.4 12.34 -1.23 # Channel 7 no signal, invalid or range overflow/underflow Channel 6 negative value 1 digit before/ 2 digits after decimal point. Channel 5 positive value 2 digit before / 2 digits after decimal point. Channel 4 positive value 3 digits before / 1 digit after decimal point. Channel 3 positive value 4 digits Channel 2 positive value 5 digits Channel 1 positive value 6 digits Error status Code The values will get the same format for the read of single channels.

Page 79 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 79 · click to enlarge
Page 79

AKAK – Read command "Calibration gas concentration" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device) and range: ∗ The calibration gas concentration in ppm. Read command AKAK Kn [Mx] Range number (optional); x = 1, 2, 3, 4 Output of all ranges, if no range is named. Read of channel n Code Response AKAK 0 M1 XXX M2 XXX ... Mn ZZZ Concentration of calibration gas Range n Concentration of calibration gas Range 2 Concentration of calibration gas Range 1 Error status Code The values will get the same format for the read of single ranges.

Page 80 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 80 · click to enlarge
Page 80

AKAL – Read command "Stored calibration corrections" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The corrections in ppm determined and stored during the last calibration. These

corrections are also necessary to calculate the analyzer values (deviations from the linearized curve). ∗ The sum of deviations (total correction) to the calibration before the last linearization

check for the last range calibrated with zero and spangas. Read command AKAL K0 [Mx] AKAL Kn [Mx] Range number (optional); x = 1, 2, 3, 4 Output of all ranges, if no range is named. Read of channel n Code

Page 81 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 81 · click to enlarge
Page 81

Response AKAL 0 M1 aaa kk AAA KK Mn xxx ff XXX FF Deviation of the final point value Correction for the final point Deviation of the zero value Correction for the zero point nth range Deviation of the final point value Correction for the final point Deviation of the zero value Correction for the zero point 1st range Error status Code

Page 82 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 82 · click to enlarge
Page 82

AKEN – Read command "Device tag" To this read command the FU will send to the TBCC the tag for the called channel (device). Read command AKEN K0 AKEN Kn Read of channel n Code Response AKEN 0 XXX..X Device tag Error status Code

Page 83 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 83 · click to enlarge
Page 83

AKFG – Read command "Configuration of the system" To this read command the system will send to the TBCC the following data: ∗ The adjustment about the channels (analyzers or virtual channels) that are expected to

send the current signals and the corresponding sequence. ∗ The channels that can be called with the total channel command "K0". That means:

Which devices can be included with "K0" or "KV Ln" to operation modes like measuring, zerogas and spangas. ∗ Which channel will analyze which chemical component. ∗ The sequence in which the signals of analyzers or calculated values from virtual

channels will be sent to the total channel read command "K0" or "KV Ln". This information will be done by the string "XYZ" standing for each component, e.g. C0, N0, NOx, BRETT (Lambda according to Brettschneider) etc. Read command AKFG K0 AKFG KV Ln Read of the whole system unit resp. some defined lines Code Response AKFG 0 XYZ Kn XYZ Km XYZ Kx Selected channel x, component XYZ Selected channel m, component XYZ Selected channel n, component XYZ Error status Code

Page 84 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 84 · click to enlarge
Page 84

AKON – Read command "Signal" (measured concentration value) To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data: ∗ The corrected signal (concentration value) valid at that moment. Normally the physical

unit is ppm. The value will be limited to six relevant digits, because it is useless to send gas concentrations in an accuracy less than pars pro mille. Example for four relevant digits (default: six digits):

measured conc. [ppm]

sent conc. [ppm]

123456 123500 12356 12360 1234.4 1234 123.45 123.5 12.56 12.56 1.23 1.23 Cf. the "SFRZ" command ! Read command AKON K0 Read of the whole system unit Code AKON K1 Kn Read of channel n Read of channel 1 Code

Page 85 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 85 · click to enlarge
Page 85

Response AKON 0 123400 12340 1234 123.4 12.34 -1.23 # Channel 7 no signal, invalid or range overflow/underflow Channel 6 negative value 1 digit before/ 2 digits after decimal point. Channel 5 positive value 2 digit before / 2 digits after decimal point. Channel 4 positive value 3 digits before / 1 digit after decimal point. Channel 3 positive value 4 digits Channel 2 positive value 5 digits Channel 1 positive value 6 digits Error status Code The values will get the same format for the read of single channels.

Page 86 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 86 · click to enlarge
Page 86

AKOW – Read command "Correction for zerogas calibration and gradient" To this read command each analyzer in a system or the single analyzer will send to the TBCC the following data: ∗ The correction of the last zero calibration and the gradient of the calibration curve. Read command AKOW K0 Mx AKOW Kn Mx Read of channel n range x Code Response AKOW 0 XXX YYY Steepness of the calibration curve Correction of the zero calibration Error status Code

Page 87 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 87 · click to enlarge
Page 87

ALCH – Read command "Deviations of the last linearization check" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device) and subchannel (range): ∗ The determined and stored deviations of the last linearization check in ppm. ∗ The information if these deviations will be in the lawful tolerances. That means: Is the

check o.k. or not ? Read command ALCH Kn Mx Read of channel n and range x Code Response ALCH 0 IO AAA BBB ... XXX xth difference 2nd difference 1st difference Checked result is o.k. (otherwise NO) Error status Code

Page 88 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 88 · click to enlarge
Page 88

Only for MLT analyzers: ALIK – Read command "Output of the linearization curve" To this read command the x/y values of the linearization curve will be sent. With this values can be determined the desired segment of the linearization curve and the interval between the function values. Read command ALIK Kn a b c Interval between the function values [ppm] End concentration of the segment [ppm] Beginning concentration of the segment [ppm] Read of channel n Only one channel can be checked (no K0). Code Response ALIK s y1 x1 y2 x2 y3 x3... Raw value No. 3 Setpoint value No. 3 Raw value No. 2 Setpoint value No. 2 Raw value No. 1 Setpoint value No. 1 Error status Code

Page 89 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 89 · click to enlarge
Page 89

ALIN – Read command "Linearization values" (X/Y = Setpoint / Raw value) To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device) and subchannel (range): ∗ The determined and stored setpoint/raw values of the last linearization. Read command ALIN Kn [Mx] This information is optional. It will not be evaluated, because the values are valid for all ranges. Read of channel n and range x Code Response ALIN 0 aaa AAA bbb BBB xxx XXX nth pair of values 2nd pair of values 1st pair of values Error status Code

Page 90 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 90 · click to enlarge
Page 90

ALKO – Read command "Polynomial coefficients of the linearization curve" The coefficients of the linearization polynomial calculated by the analyzer linearization will be transferred. These coefficients will be enabled using the polynomial method to linearize. Read command ALKO Kn Mm Read of channel n and range m Code Response ALKO 0 Mm a0 a1 a2 a3... Coefficients of the polynomial Range m Error status Code

Page 91 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 91 · click to enlarge
Page 91

ALST – Read command "Linearization steps" To this read command the gas distribution will send to the TBCC or to the system the following data: ∗ The numbers and the division in percent of the distribution steps (maximum two digits

after decimal point). This command will only be accepted by the gas distribution, if the commands "SLIN" or "SLCH" have been received before. Read command ALST Kn Read of the gas distribution Code Response ALST 0 1 XY 2 XY ... n XY Last distribution step n, division XY % 2nd distribution step, division XY % 1st distribution step, division XY % Error status Code

Page 92 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 92 · click to enlarge
Page 92

AM90 – Read command "Actual response time (t90)" To this read command the analyzer will send the t90 time in seconds that is enabled to calculate the concentration for the called channel at the moment. Cf. "AT90", "ET90", "ST90" and "EM90" Read command AM90 Kn Read of channel n Code Response AM90 0 a t90 time (response time) [s] Error status Code

Page 93 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 93 · click to enlarge
Page 93

AMBA – Read command "Begin of range" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The begin of range values in ppm. Read command AMBA K0 [Mx] AMBA Kn [Mx] Range number (optional) Read of channel n Code Response AMBA 0 Mx XXX Begin of range x Range No. x Error status Code The values will get the same format for the read of single channels.

Page 94 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 94 · click to enlarge
Page 94

AMBE – Read command "End of range" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The end of range values in ppm. Read command AMBE K0 [Mx] AMBE Kn [Mx] Range number (optional) Read of channel n Code Response AMBE 0 Mx XXX End of range x Range No. x Error status Code The values will get the same format for the read of single channels.

Page 95 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 95 · click to enlarge
Page 95

AMBU – Read command "Switching values for autoranging" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The adjusted switching values in ppm for a changing of ranges with autoranging. Read command AMBU K0 AMBU Kn Read of channel n Code Response AMBU 0 M1 xxx XXX M2 yyy YYY Mn zzz ZZZ Switch on value of range n Switch off value of range n Range No. n Switch on value of range 2 Switch off value of range 2 Range No. 2 Switch on value of range 1 Switch off value of range 1 Range No. 1 Error status Code The values will get the same format for the read of single channels.

Page 96 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 96 · click to enlarge
Page 96

AMDR – Read command "Manual adjusted pressure" To this read command the gas analyzer will send the value adjusted for the parameter. This value will be useful, if no pressure measurement will be installed in the analyzer. Read command AMDR Kn Read of channel n Code Response AMDR 0 a Pressure [Pa] Error status Code The values will get the same format for the read of single channels.

Page 97 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 97 · click to enlarge
Page 97

AQEF – Read command "Cross interference" To this read command the CO analyzer or the system will send to the TBCC: ∗ The concentration value in ppm determined and stored in the analyzer. This value will

be stored in the device until a new cross interference will be determined by the "SQEF" command. The TBCC will control, if limits will be exceeded. The TBCC will also start actions if necessary. Read command AQEF Kn Read of channel n Code Response AQEF 0 XXX Interference [ppm] Error status Code

Page 98 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 98 · click to enlarge
Page 98

ASOL – Read command "Setpoint value with limits" To this read command the FU will send to the TBCC the following data for the called channel (FU) and subchannel (e.g. heating): ∗ The adjusted setpoints with deviation limits for error reports. For the actual used devices subchannel "m" will be: m = 0 Concentration m = 1 Temperature m = 2 Pressure m = 3 Flow m = 4 Pocket calculator No. 1 m = 5 Pocket calculator No. 2 m = 6 Pocket calculator No. 3 m = 7 Pocket calculator No. 4 These assignations may be changed for devices used in the future ! Read command ASOL K0 m ASOL Kn m Read of channel n and subchannel m Code Response ASOL 0 WWW xxx XXX Upper limit Lower limit Setpoint value Error status Code

Page 99 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 99 · click to enlarge
Page 99

ASTA – Read command "General status of the system" To this read command the FU will send to the TBCC: ∗ All channels of the FU with any error in their status at that moment. A detailed

description of the errors will not be sent to this read command. It is only possible to read channel No. 0. Read command ASTA K0 Read of channel zero Code Response ASTA 7 K1 K4 Kn Channel n with error status Channel 4 with error status Channel 1 with error status Error status Code

Page 100 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 100 · click to enlarge
Page 100

ASTF – Read command "Error status" To this read command the FU will send to the TBCC: ∗ All error existing at that moment in the called channel (FU). The description of the error

characterization is specific for each device. It will be symbolized with a number. A reading to "K0" will get the errors of devices that are not assigned to single channels (e.g. samplegas cooler). Read command ASTF Kn Read of channel n Code Response ASTF 7 XXX YYY ... NNN nth error number 2nd error number 1st error number Error status Code

Page 101 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 101 · click to enlarge
Page 101

TTypical meanings of errors for certain devices: NDIR/NDUV Analyzers No. 1 = Flow error No. 2 = Chopper failure No. 3 = Thermostat failure No. 4 = RAM error No. 5 = Calibration error zerogas No. 6 = Calibration error spangas No. 7 = Range overflow No. 8 = External error (digital input) No. 9 = Error of pressure measurement No.10 = Error of temperature measurement TFID Analyzers No. 1 = Flow error No. 2 = Flame out No. 3 = Thermostat failure No. 4 = RAM error No. 5 = Calibration error zerogas No. 6 = Calibration error spangas No. 7 = Range overflow No. 8 = Fuel gas error No. 9 = Fuel air error No.10 = H2 generator failure (optional H2 generators, if available) No.11 = Temperature of the heated wires No.12 = Temperature of the heated filters No.13 = Response of ethane too high (optional for methane free measurement)

Page 102 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 102 · click to enlarge
Page 102

ASTZ – Read command "Status" To this read command the FU will send to the TBCC the following data for the called channel (FU): ∗ The device status at that moment ∗ Running procedures. The status will be described by the code used for the activation of the function. The operation modes "REMOTE" or "MANUAL" will also be sent. These modes will always be the first codes in the data string. To the read of channel 0 the statuses of all channels (FU) defined with "EKFG" will be sent. If a defined FU will be defect or not available and it cannot send its status, the statuses of these channels (FU) will be replaced by "#", analogous to the values like described for "AKON". Read command ASTZ K0 ASTZ Kn Read of channel n Code

Page 103 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 103 · click to enlarge
Page 103

Response ASTZ 0 KV SREM CODEn K1 SREM CODEn K2 # Kn SREM CODEn SMAN SMAN SMAN Status code Channel n and its status Device is not available Channel 2 and its status Status code Channel 1 and its status Status code Channel 0 and its status Error status Code ASTZ 0 Kn SREM CODEn SMAN Status code Channel n and its status Error status Code

Page 104 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 104 · click to enlarge
Page 104

ASYZ – Read command "System time" To this read command the FU will send to the TBCC the following data for the called channel (device): ∗ The current system time (calendar time). Read command ASYZ Kn Read of channel n Code Response ASYZ 0 JJMMTT hhmmss Time of the system Error status Code

Page 105 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 105 · click to enlarge
Page 105

AT9O – Read command "T90 time" (response time) To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The t90 time steps. Read command AT9O K0 AT9O Kn Read of channel n Code Response AT9O 0 XXX YYY ZZZ T90 time (fast) T90 time (medium) T90 time (slowly) Error status Code

Page 106 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 106 · click to enlarge
Page 106

ATEM Read command "Temperature" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device) and subchannel (temperature measurement): ∗ The signal in Kelvin. Read command ATEM K0 (m) ATEM Kn (m) Read of channel n (and subchannel m). Note: Subchannel m is not in use at the moment. Code Response ATEM 0 XXX Temperature Error status Code

Page 107 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 107 · click to enlarge
Page 107

ATOL – Read command "Stability tolerances" To this read command the analyzers in a system, the single analyzer or the front-end computer will send to the TBCC the following data: ∗ The necessary tolerances for functions running stability controlled. Read command ATOL Kn [Mx] Range number (optional) Read of channel n Code Response ATOL 0 Mm Tnm Tolerance of channel n Range m as percent of end of range Error status Code

Page 108 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 108 · click to enlarge
Page 108

AUKA – Read command "Uncorrected analog value" To this read command the called analyzer in a system or the single analyzer will send to the TBCC the following data: ∗ The uncorrected analog output value in Volt and the corresponding range. Read command AUKA K0 AUKA Kn Read of channel n Code Response AUKA 0 Mn XXX Uncorrected analog value Range n Error status Code

Page 109 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 109 · click to enlarge
Page 109

AVEZ – Read command "Delay and synchronization time" To this read command the analyzers in a system or the single analyzer will send to the TBCC the following data for the called channel (device): ∗ The delay time used for the record of valid signals or used to start the integrators by

"SINT". ∗ The synchronization time used for the output of values from an internal buffer that were

read by "AKON", "AIKO" and "AIKG". This synchronization time is also the delay time for the output signal of those values. This procedure will be started with the control command "SRON" and will be finished with the control command "SROF". Read command AVEZ K0 AVEZ Kn Read of channel n Code Response AVEZ 0 XXX YYY Time of synchronization Delay time Error status Code

Page 110 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 110 · click to enlarge
Page 110

AZEI – Read command "Times" To this read command the FU will send to the TBCC the following data for the called channel (FU): ∗ The times used to start a function or procedure, e.g. times for the automatic start of a

calibration. Read command AZEI Kn CODE Code of the function Read of channel n Code The following codes will be supported: SNAB, SPAB, SATK, SNGA, SEGA, SQEF, SSPL, SALI, SMGA Response AZEI 0 JJMMTT hhmmss T-0 Function length Start time Start date Error status Code

Page 111 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 111 · click to enlarge
Page 111
5. Description of all Write Commands

EFDA – Write command "Function length" With this write command the FU will get the function length of the function "SXXX" in seconds for the called channel (device), e.g. "Time Out" for purging or for switching on the calibration gases. If the function "SXXX" will be a procedure with several internal steps, the times will be valid for each step and not for the whole procedure. If the function will be for instance "Automatic calibration", the times will be valid for each spangas resp. each range and not for the whole procedure. During this time of flow the analyzer will test, if the setpoint value will be reached and stable. If this test will not succeed during the function time, it will produce a function error. The same effect will exist for other procedures like linearization, converter test etc. If the function length will be set to zero, the function will run without any time limit. Functions like "SATK", "SLIN", "SLCH", "SALI", "SQEF", "SNGA" or "SEGA" will run time controlled according to the times T1 to T4 or stability controlled. Time control: If only T1 is set or if T2 = 0, time control will run with step time T1 (total function time). Stability control: Times T1 to T4 have to be set. Write command EFDA Kn SXXX T1 (T2 T3 T4) "Time out"; after this time is over, the procedure will be canceled and you will get an error message; this time will start after the wait. Integration time to get the mean value of one signal. Stability time: All signals have to be in a certain tolerance during this time. Time to wait for: after the switching on / changing of gases or the stepping time of time control. Function Addressed channel Code Response

Page 112 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 112 · click to enlarge
Page 112

EFDA 0 Error status Code

Page 113 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 113 · click to enlarge
Page 113

EGRW – Write command "Limit" With this write command the analyzers in a system or the single analyzer will get the required limits for the called channel (device), e.g. maximum deviation for calibration. If these limits will be exceeded during the operation, it will cause a changing of the error status byte. – The unit of limits is percent. The deviation value is referred to the setpoint. For zero calibration the deviation value will be referred to the smallest spangas value. Write command EGRW Kn m XXX Limit Addressed subchannel m m = 0: zero calibration; m = 1: spangas calibration Addressed channel n Code Response EGRW 0 Error status Code

Page 114 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 114 · click to enlarge
Page 114

EKAK – Write command "Calibration gas concentration" With this write command the analyzers in a system or the single analyzer will get the spangas values for each range. Write command EKAK K0 M1 YYYY ... Mx ZZZZ EKAK Kn M1 YYYY ... Mx ZZZZ Spangas value Addressed range x Spangas value Addressed range 1 Addressed channel n Code Response EKAK 0 Error status Code

Page 115 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 115 · click to enlarge
Page 115

EKEN – Write command "Device tag" With this write command the FU will get a tag to store in the FU memory. Then it will have to be unchangeable. That means, if this part of memory will be written, the device processor will have to save it against an overwriting automatically. The setup of a new tag will only be possible by changing the processor. The memory size for the tag is 30 ASCII characters. Write command EKEN Kn TAG Data of the device tag Read of channel n Code Response EKEN 0 Error status Code

Page 116 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 116 · click to enlarge
Page 116

EKFG – Write command "Configuration of the system" With this write command the FU will get the channels (analyzers) sending the current signals and their sequence. Furthermore the FU will get the channels that should be called with the total channel command "K0" or "KV Ln", i.e. the channels that shall be included to operation modes like measurement, zerogas and spangas. With the string XYZ the FU will be told which chemical component or virtual device (e.g. Lambda) shall be combined with which channel. The string XYZ will be for instance C0, N0, Nox or BRETT (Lambda according to Brettschneider) etc. The other physical or virtual channels can only be called directly with their channel number. If no component and no channel number will be specified, the FU will get its default confi- guration of the system. That means, that all physical and virtual available components will be included to the read and control commands with the total channel command "K0" or "KV Ln". Besides, the TBCC can get the real physical composition of the FU with the read command "AKFG". Write command

EKFG K0 XYZ Kn XYZ Km XYZ Kx Adjusted channel x, component XYZ Adjusted channel m, component XYZ Adjusted channel n, component XYZ Addressed channel 0 Code

EKFG KV Ln Kn ... Kx

Adjusted channel n to x Lines getting assigned with adjusted channels Addressed channel; front-end computer Code

EKFG 0 Error status Code

Response

Page 117 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 117 · click to enlarge
Page 117

ELIN – Write command "Linearization values (X/Y)" With this write command the analyzers in a system or the single analyzer will get the linearization values for the calculation of gas concentration for the called channel (device) and subchannel (range). A new linearization curve will be calculated with these x/y-pairs. Write command ELIN Kn Mx aaa AAA bbb BBB ... xxx XXX nth pair of values 2nd pair of values 1st pair of values Addressed channel n and range x Code Response ELIN 0 Error status Code

Page 118 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 118 · click to enlarge
Page 118

ELKO – Write command "Polynomial coefficients of the linearization curve" The coefficients of a linearization polynomial will be transferred to the analyzer. These values will then be enabled to calculate the gas concentration using the polynomial method for linearization. Write command ELKO Kn a0 a1 a2 a3... Polynomial coefficients Read of channel n Code Response ELKO 0 Error status Code

Page 119 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 119 · click to enlarge
Page 119

ELST – Write command "Linearization steps" With this write command the device will get the numbers of the distribution steps and each division in percent of the gas distribution used for linearization. But this command will only be accepted, if the commands "SLIN" or "SLCH" were received before. Write command ELST Kn 1 XY 2 XY ... n XY Last distribution step, division XY % 2nd distribution step, division XY % 1st distribution step, division XY % Addressed channel n Code Response ELST 0 Error status Code

Page 120 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 120 · click to enlarge
Page 120

EMBA – Write command "Begin of range" With this write command the analyzers in a system or the single analyzer will get each begin of range of the total range in ppm. The analog output signal will be referred to these values for instance. Write command EMBA Kn M1 YYYY [... Mx ZZZZ] Begin of range x (optional) Addressed range x (optional) Begin of range 1 Addressed range 1 Addressed channel n Code Response EMBA 0 Error status Code

Page 121 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 121 · click to enlarge
Page 121

EMBE – Write command "End of range" With this write command the analyzers in a system or the single analyzer will get each end of range of the total range in ppm. The analog output signal will be referred to these values for instance. Write command EMBE Kn M1 YYYY [... Mx ZZZZ] End of range x (optional) Addressed range x (optional) End of range 1 Addressed range 1 Addressed channel n Code Response EMBE 0 Error status Code

Page 122 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 122 · click to enlarge
Page 122

EMBU – Write command "Switch levels for autoranging" With this write command the analyzers in a system or the single analyzer will get for the called channel (device): ∗ The required values in ppm to switch from one range to another with autoranging. Write command EMBU Kn Mn XXX YYY [Mm XXX YYY ...] (optional) Switching value to change to the bigger range Switching value to change to the smaller range Range n Addressed channel n Code Response EMBU 0 Error status Code

Page 123 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 123 · click to enlarge
Page 123

EMDR – Write command "Manual adjusted pressure" With this write command the analyzer will get a pressure value for the called channel. This value will be used as pressure correction, if no pressure measurement is installed in the device. Cf. "AMDR", "ADRU" ! Write command EMDR Kn a Pressure in Pa Read of channel n Code Response EMDR 0 Error status Code

Page 124 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 124 · click to enlarge
Page 124

ESOL – Write command "Setpoint value with limits" With this write command the FU will get the required setpoint values with acceptable deviations for the called channel (device) and subchannel (e.g. heating). If these limits will be exceeded during the operation, it will cause a changing of the error status byte. For the actual used devices subchannel "m" will be: m = 0 Concentration m = 1 Temperature m = 2 Pressure m = 3 Flow m = 4 Pocket calculator No. 1 m = 5 Pocket calculator No. 2 m = 6 Pocket calculator No. 3 m = 7 Pocket calculator No. 4 These assignations may be changed for devices used in the future ! Write command ESOL Kn m XXX yyy YYY Upper limit Lower limit Addressed setpoint value Addressed subchannel m Addressed channel n Code Response ESOL 0 Error status Code

Page 125 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 125 · click to enlarge
Page 125

ESYZ – Write command "System time" With this write command the FU will get the system time (calendar time) that has to be adjusted for the called channel (device). Write command ESYZ Kn JJMMTT hhmmss Value for the adjustment of the system time Addressed channel n Code Response ESYZ 0 Error status Code

Page 126 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 126 · click to enlarge
Page 126

ET9O – Write command "T90 time" (response time) With this write command the analyzers in a system or the single analyzer will get the t90 time steps in seconds that have to be adjusted for the called channel (device). Write command ET9O Kn XXX YYY ZZZ Value for the adjustment of the t90 time (fast) Value for the adjustment of the t90 time (medium) Value for the adjustment of the t90 time (slowly) Addressed channel n Code Response ET9O 0 Error status Code

Page 127 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 127 · click to enlarge
Page 127

ETOL – Write command "Stability tolerances" With this write command the analyzers in a system, the single analyzer or the front-end computer will get the required tolerances for functions running stability controlled. The tolerance will be specified in percent of the end of range value. The tolerance can be adjusted for each range separately. No tolerance check will be done, if the tolerance value will be set to T = 100%. Write command ETOL Kn M1 Tn1 [...Mm Tnm] Tolerance channel n, range m in percent of the end of range value; tolerance deviation = 2 * tolerance (optional) Tolerance channel n, range 1 Range 1 Addressed channel n Code Response ETOL 0 Error status Code

Page 128 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 128 · click to enlarge
Page 128

EVEZ – Write command "Delay and synchronization time" With this write command the analyzers in a system or the single analyzer will get the delay time and the synchronization time for the called channel (device). The delay time will be used for integrators started by "SINT" or for the delayed record of valid signals. The synchronization time will be used for the output of values that were read by "AKON", "AIKO" and "AIKG" from an internal buffer. The same synchronization time will also delay the analog signal of these values. This procedure will be started by the control command "SRON" or will be finished by the control command "SROF". Write command EVEZ K0 XXX YYY EVEZ Kn XXX YYY Synchronization time Delay time Read of channel n Code Response EVEZ 0 Error status Code

Page 129 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 129 · click to enlarge
Page 129

EZEI – Write command "Times" With this write command the FU will get the times that have to be adjusted in the called channel (device) for the automatic start of functions or procedures (e.g. "Automatic calibration"). Furthermore, the FU will get the function length in seconds. The following controls will be possible: • If the calendar day will be mentioned, the function will start only once at one date. If the calendar day will be missing, the function will start each day.

  • The clock time as starting time and the function length have always to be mentioned.
  • If the function length will be zero, the function will run without any time limit. Then it can only be finished by another control command. A function length of the write command "EFDA" will not be used here.
  • Unused data will be replaced by "#". If the functions shall start several times automatically at the same day, the times can be set up in blocks one after another (maximum 4 digits). Functions with day information will be deleted automatically after they will be ready. Otherwise, the set-ups of functions will be deleted by resetting all values to zero. Examples:
  • CODE 871113 171200 0: The adjusted function will be started for one time at 13/11/87, 5 o'clock p.m. (17 o'clock), 12 minutes and 0 seconds. The function length will be unlimited.
  • CODE 871113 171200 33: The adjusted function will be started for one time at 13/11/87, 5 o'clock p.m. (17 o'clock), 12 minutes and 0 seconds. The function length will be limited to 33 sec.
  • CODE # 051208 0: The adjusted function will be started each day at 5 o'clock a.m., 12 minutes and 8 seconds. The function length will be unlimited.
  • CODE # 051200 1850: The adjusted function will be started each day at 5 o'clock a.m., 12 minutes and 0 seconds. The function length will be limited to 1850 seconds.
  • CODE # 060000 100 # 120000 0 # 180000 0 # 240000 150: The adjusted function will be started each day at 6, 12, 18 (6 p.m.) and 24 (12 p.m.) o'clock.
Page 130 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — K1
Page 130 · click to enlarge
Page 130

Write command EZEI Kn CODE JJMMTT hhmmss T-0 Function length Start time of the function Start day of the function Function that shall be started Addressed channel Code Response EZEI 0 Error status Code

Page 131 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — MLT/CU
Page 131 · click to enlarge
Page 131

Supplement

1. Overview about working AK commands in NGA devices

Code Function MLT local

MLT/CU

AAEG Deviation to span gas √ √ AALI Linearization check results √ √ AANG Deviation to zero gas √ √ ABST Hours of operation √ √ ADRU Pressure √ √ √ ADUF Flow √ √ √ AEMB Actual range number √ √ √ AFDA Function length √ √ AGID Device identification √ √ √ AGRW Limits √ √ AIKG PV average (Concentration integral value); all √ √ AIKO PV average (Concentration integral value); partial √ √ AKAK Span gases √ √ √ AKAL Calibration results √ √ AKEN Tag √ √ AKFG System configuration √ √ √ AKON Concentration √ √ √ AKOW Calibration values (Correction for zero and gradient) √ √ ALCH Linearization check results √ √ ALIK Calculation of linearization curve √ √ ALIN Linearization x/y-values √ √ ALKO Linearization polynomial coefficients √ √ √ ALST Linearization steps √ √ AM90 Actual t90 time √ √ AMBA Begin of ranges √ √ √ AMBE End of ranges √ √ √ AMBU Switch levels for autorange √ √ AMDR Manual pressure value √ √ APAR Parameters general AQEF Cross interference check result √ √ ASOL Setpoints with limits √ √ ASTA Internal status √ √ ASTF Internal error status √ √ ASTZ Action status (running procedure) √ √ ASYZ Time and date of the system √ √ √ AT90 t90-times √ √ ATEM Temperature √ √ √ ATOL Tolerances for stability controlled procedures √ √ AUKA Uncorrected, analog PV value √ √ √ AVEZ Delay and synchronization time √ √ AZEI Times for procedures √ √ EFDA Function length √ √ EGRW Limits √ √ EKAK Span gases √ √ √ EKEN Tag √ √ EKFG System configuration √ √ √ ELIN Linearization x/y-values √ √

90003752(1) [AK-Commands] 10/98 AK Supplement - 1

Other AM

Page 132 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — MLT/CU
Page 132 · click to enlarge
Page 132

Code Function MLT local

MLT/CU

ELKO Linearization polynomial coefficients √ √ √ ELST Linearization steps √ √ EMBA Begin of ranges √ √ √ EMBE End of ranges √ √ √ EMBU Switch levels for autorange √ √ EMDR Manual pressure value √ √ EPAR Parameters general ESOL Setpoints with limits √ √ ESYZ Time and date of the system √ √ ET90 t90 times √ √ ETOL Tolerances for stability controlled procedures √ √ EVEZ Delay and synchronization time √ √ EZEI Times for procedures √ √ SALI Linearization check with span gases √ √ SARA Autoranging OFF √ √ SARE Autoranging ON √ √ SATK Zero and span calibration √ √ SCAL Start system calibration √ SEGA Open spangas valve √ √ SEMB Set range √ √ √ SENO Switch to NO mode for CLD module √ SFRZ Floating point format of real numbers √ √ √ SGTS Switch to device test status SHDA Hold mode OFF √ √ SHDE Hold mode ON √ √ SINT Start integration √ √ SLCH Start linearization check √ √ SLIN Start linearization √ √ SLST Switch linearization step √ √ SMAN Communication: Manual √ √ SMGA Open sample gas valve √ √ SNAB Zero calibration √ √ √ SNGA Open zero gas valve √ √ SNOX Switch to NOx mode for CLD module √ SPAB Span calibration √ √ √ SPAU Set to pause mode √ √ SQEF Start cross interference measurement √ √ SREM Communication: Remote √ √ SRES µP-Reset √ √ √ SROF Delay mode OFF √ √ SRON Delay mode ON √ √ SRUC Purge backwards SSPL Open purge gas valve √ √ ST90 Set t90 time √ √ STBY Set to stand-by √ √

Implemented additional Fisher-Rosemount specific commands (not official AK): ALIK, AMBA, AMDR EMBA, EMDR SCAL, SFRZ, SHDA, SHDE, SNAB, SPAB ASVC, ESVC, SSVC, (additional service commands)

Supplement - 2 AK 90003752(1) [AK-Commands] 10/98

Other AM

Page 133 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — Supplement
Page 133 · click to enlarge
Page 133

Supplement

90003752(1) [AK-Commands] 10/98 AK Supplement - 3

Page 134 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — Supplement
Page 134 · click to enlarge
Page 134
2. AK Service Commands

ASVC K0 S599 ExactNode Name Reading of LON variables of a network node Description: This command is an enlarged alternative to the S600 command. It may use the variables of all network nodes, even if they will not be analyzer modules. You only have to know the right node address (cf. S632). Syntax: ExactNode = Right network node address Name = Name of the variable Response: ASVC 0 a a: Value of the desired variable Notes: Only texts without an AK separating sign will be valid as variable names. The "ExactNode" is an address composed from a node address and its subnode address. The formula is: ExactNode = Subnode * 256 + Node; Example: Node = 2 Subnode = 1 ExactNode = 1*256 + 2 = 258 Cf.: S600, S630, S632 ESVC K0 S599 ExactNode Name a Writing from a LON variable of a network node Description: This command is an enlarged alternative to the S600 command. It may use the variables of all network nodes, even if they will not be analyzer modules. You only have to know the right node address (cf. S632). Syntax: ExactNode = Right network node address Name = Name of the variables a = Value of the variable Notes: Only texts without an AK separating sign will be valid as variable names. The "ExactNode" is an address composed from a node address and its subnode address. The formula is: ExactNode = Subnode * 256 + Node; Example: Node = 2 Subnode = 1 ExactNode = 1*256 + 2 = 258 Cf.: S600, S630, S632 SSVC K0 S599 ExactNode Name a Writing from a LON variable of a network node without checking the range of values Description: This command corresponds to "ESVC K0 S599...", only without checking the range of values.

Supplement - 4 AK 90003752(1) [AK-Commands] 10/98

Page 135 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — Supplement
Page 135 · click to enlarge
Page 135

Supplement

S600: Access to LON network variables Access to network variables of an analyzer module. Notes: ☞ If a variable will contain several values (array), you can access to these values by adding the corresponding number directly to the name of the variable. Example: LINYA = [31, 44, 54] LINYA1 = 31 LINYA2 = 44 LINYA3 = 54 ☞ The network variable knows its data type. The AK command will try to convert the text of value inputs to the required type. Only if this will not be possible a syntax error will be reported. ☞ If the data will be of the so-called "enum" type, the value will have to be inputted as integer. The text on the display cannot be used, because this text information is only existing in the control unit. To find out the integer value corresponding to a certain "enum" text, you have to look into the EXCEL- File. In the EXCEL-File the first column of the "enum" text will correspond to the value "0", the next column to the value "1" etc. ASVC Kn S600 Name Reading of the variable value Syntax: Name = Name of the variable Response: ASVC 0 a a: Value of the desired variable Note: Only texts without an AK separating sign will be valid as variable names. Cf.: S630 (which variables are existing ?) ESVC Kn S600 Name a Writing of the variable value Syntax: Name = Name of the variable a = Value of the variable Cf.: S630 (which variables are existing ?) SSVC Kn S600 Name a Writing of the variable value without checking the range of values Description: This command corresponds to "ESVC Kn S600...", only without checking the range of values.

90003752(1) [AK-Commands] 10/98 AK Supplement - 5

Page 136 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — Supplement
Page 136 · click to enlarge
Page 136

ASVC Kn S615 b Reading of the current data of a DIO board Description: The current status of the named DIO board will be read. Syntax: b: Board number (1, 2, 3 ,4; not SLOT-ID); for b = 0 the data of all available DIO's will be sent Response: ASVC 0 IIIIIIII OOOOOOOO OOOOOOOO OOOOOOOO RLLL I1...8 Status of the digital input pins 1...8 O1...24 Status of the digital output pins 1...24 R Retrigger error L1...3 Overload group 1...3 SSVC Kx S617 n s [n s] Setup of the external switch variable DIGEXTSWITCH Description: The outputs of a DIO board can be assigned to the single bits of the variable DIGEXTSWITCH. For the platform the lowest 8 bits of this variable are available. With the command "SSVC" each switch can be set. This on the other hand may cause a direct switch of an assigned digital output. Syntax: n: Switch number s = Status (0 = OFF; 1 = ON) x: If the channel number will be K0, the variable DIGEXTSWITCH will be called of the module containing the local SIO (platform SIO → control module, MLT SIO → analyzer module). The parameters "n s" can be repeated up to seven times for the platform. Examples: SSVC K0 S617 1 1 2 0 3 1 Set external switch 1 to HIGH Set external switch 2 to LOW Set external switch 3 to HIGH

Supplement - 6 AK 90003752(1) [AK-Commands] 10/98

Page 137 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — Supplement
Page 137 · click to enlarge
Page 137

Supplement

SSVC Kn S621 a Loading/Saving of device specific parameters Description: With this command you can load/save the device specific parameters. This will be possible via the serial interface and from /to the internal FLASH memory (if on ACU available!). Syntax: a = 1: loading new configuration via the serial interface. 2: sending out the current configuration via the serial interface 3: saving the current configuration in the FLASH memory 4: loading the configuration saved in the FLASH memory Notes: The data format for the loading via the serial interface (a = 1) corresponds to the data format of the serial sending out (a = 2). Some peculiarities will have to be heeded, because the serial interface will be used parallel for AK and for in-/output of the data stream: For a = 1 you will have to wait for the AK response, before you will start the sending of the data stream. For a = 2 the AK response ("SSVC 0") will be sent at first. It will not be necessary to treat these response signs in a certain way, because this line will be deleted as invalid stream! Saving the configuration in the FLASH memory (a = 3) will overwrite the factory settings of the device !! With a = 4 the factory settings will be reloaded to the RAM memory and so to the actual working memory. Cf.: Instruction manual

90003752(1) [AK-Commands] 10/98 AK Supplement - 7

Page 138 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — Supplement
Page 138 · click to enlarge
Page 138

ASVC Kn S630 Output of the LON variable names Description: All LON variables that are available in an analyzer module will be sent out with their names. If a variable will contain several values (array), it will be marked by an appendix in brackets. In these brackets you will find the numbers to access to the array values of the variable. Example: LINYA[1-7] contains 7 values. You can access to these values with the names LINYA1, LINYA2, ... LINYA7. Notes: K0 is not possible. At the moment the command is implemented in a way that it will send out for each channel all variables available in the certain module. That means that variable names may exist multiplex. These multiplex available variables will be separated internal and each will be assigned to another channel (subnode). If variables will not exist multiplex, these variables will only exist once. That means that a changing within a channel will have an effect as changing of the other channels. Cf.: S600 (Access to network variables) ASVC Kn S631 Output of the LON node informations Syntax: ASVC Kn S631 Response: ASVC 0 ASVC Kn S632 Output of the LON node informations of all nodes Syntax: ASVC K0 S632 Response: ASVC 0 ExactNode = Exact node address in the network NodeTag = Tag string of this node Notes: The "ExactNode" is an address composed from the node address and its subnode address. The formula is: ExactNode = Subnode * 256 + Node; Example: ExactNode = 258 Subnode = ExactNode mod 256 = 1 Node = ExactNode - ((ExactNode mod 256)*256) = 2 Cf.: S599, S600, S630

Supplement - 8 AK 90003752(1) [AK-Commands] 10/98

Page 139 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — Supplement
Page 139 · click to enlarge
Page 139

Supplement

ASVC Kn S640 a Output of values of the MLT concentration formula Concentration formula:

Conce = FacP * FacT * FacSpan * Lin{(RawAvg - OffP - OffT - OffX) * RGain * Gain}

Meaning of the variables: RawAvg: Raw value average (including t90 time) OffP: Offset correction of physics OffT: Offset correction of temperature OffX: Offset correction of interferences from other channels RGain: Analog pre amplifying factor (BIS) Gain: Factor standardizing to the desired setpoint value (span) Lin: Linearization procedure FacSpan: Span factor standardizing the product "FacP*FacT*FacSpan" to 1.0 during the span cal. FacT: Temperature correction in the sensitivity FacP: Pressure correction in the sensitivity Conce: Concentration value Syntax: ASVC Kn S640 a Response: ASVC 0 ... a = 0: value1 = RawCount = uncorrected signal value2 = RawAvg value3 = OffT value4 = OffX value5 = OffP value6 = Gain value7 = RGain a = 1: value1 = LinInput = (RawAvg - OffP - OffT - OffX) * RGain * Gain value2 = LinOutput = Lin{(RawAvg - OffP - OffT - OffX) * RGain * Gain} value3 = FacSpan value4 = FacT value5 = FacP value6 = Conce a = 2: value1 = Temperature for OffT value2 = Temperature for FacT value3 = Atmospheric pressure for FacP

90003752(1) [AK-Commands] 10/98 AK Supplement - 9

Page 140 of the Emerson Process Management Multi-Component Gas Analyzer (MLT 1) user manual — ASIA-PACIFIC
Page 140 · click to enlarge
Page 140

NGA 2000 MLT AK Protocol

WORLD HEADQUARTERS ROSEMOUNT ANALYTICAL EUROPE Emerson Process Management GmbH & Co. OHG Industriestrasse 1 63594 Hasselroth Germany T 49 6055 884 0 F 49 6055 884209

Emerson Process Management Rosemount Analytical Inc. 6565 P Davis Industrial Parkway Solon, OH 44139 USA T 440.914.1261 Toll Free in US and Canada 800.433.6076 F 440.914.1271 e-mail: gas.csc@EmersonProcess.com

www.raihome.com

GAS CHROMATOGRAPHY CENTER AND LATIN AMERICA

Rosemount Analytical Inc.

11100 Brittmoore Park Drive

T 713 467 6000

F 713 827 3329

EUROPE, MIDDLE EAST AND AFRICA

Emerson Process Management Shared Services Limited Heath Place Bognor Regis West Sussex PO22 9SH England T 44 1243 863121 F 44 1243 845354

ASIA-PACIFIC

Emerson Process Management Asia Pacific Private Limited 1 Pandan Crescent Singapore 128461 Republic of Singapore T 65 6 777 8211 F 65 6 777 0947 e-mail: analytical@ap.emersonprocess.com

© Emerson Process Management GmbH & Co. OHG 2007

Show 4 more labels / short text on this page
Instruction Manual · 90003752 11/2003 · Emerson Process Management · Houston, TX 77041
Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson Process Management Multi-Component Gas Analyzer (MLT 1)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Process Management Multi-Component Gas Analyzer (MLT 1)

Be the first to ask — other owners or our team will answer.

Original document

Emerson Process Management Multi-Component Gas Analyzer (MLT 1) manual PDF

140 pages·1 MB·TV Converter Box·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…