Emerson Process Management Gas Chromatograph (CFA-3000)

74 pages·2.7 MB·TV Converter Box·PDF·ENDEFRITNLJA
Emerson Process Management Gas Chromatograph (CFA-3000)
Table of Contents 80 sections CFA-3000 Ion Selective Electrode (ISE) Analyzer · ESSENTIAL INSTRUCTIONS · WARNINGS · READ THIS PAGE BEFORE PROCEEDING! · … ShowHide
CFA-3000 Ion Selective Electrode (ISE) Analyzerp.1ESSENTIAL INSTRUCTIONSp.2WARNINGSp.2READ THIS PAGE BEFORE PROCEEDING!p.2RISK OF ELECTRICAL SHOCKp.2CAUTIONp.2WARNINGp.2TABLE OF CONTENTS (CON’T)p.5LIST OF FIGURESp.5LIST OF TABLESp.6CFA3000 MANUAL SECTION 1 INTRODUCTIONp.7SECTION 1 INTRODUCTIONp.71.1 Thoroughly Read This Manualp.81.2 Analyzer Benefitsp.81.3 WARNING – Electrical Shock Hazardp.9SECTION 2 INSTALLING THE ANALYZERp.102.0 INSTALLING THE ANALYZERp.102.1 UNPACK PARTSp.102.6 Prepare Electrical Connectionsp.162.8 Prepare Reagent Containersp.19SECTION 3 OPERATING THE ANALYZERp.203.0 Operating the Analyzerp.203.2 Check Sample Streamp.21Enter Test Mode:p.22DECIMALp.22MANUALp.22CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTIONp.23FIGURE 32. KEYPAD PANELp.233.3 Set Reagent Supply Timep.283.4 Set Scaling Outputsp.283.7 Shutdownp.313.8 Quick Startp.32CFA3000 MANUAL SECTION 4 MAINTAINING THE ANALYZERp.33SECTION 4 MAINTAINING THE ANALYZERp.33CFA3000 MANUAL SECTION 5 TROUBLESHOOTING PROCEDURESp.34SECTION 5 TROUBLESHOOTING PROCEDURESp.34OFFp.345.2 Test Functions Table 52: Test Functionsp.35CFA3000 MANUAL SECTION 6 REPAIR PROCEDURESp.37SECTION 6 REPAIR PROCEDURESp.376.2 Replace Circuit Boardsp.386.3 Replace Fusesp.396.4 ISE Reaction Cell Assembly, Electrodesp.39CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTIONp.40SECTION 7 INSTRUMENT DESCRIPTIONp.407.2 Keypad Panelp.42PLEASE NOTE: VALVEPUMP CONNECTOR IS ON CPU BOARDp.44FIGURE 72. CARD CAGE AND BOARDSp.443 ONp.457 OFFp.45FIGURE 74. REAR OF MAIN CABINETp.467.5 Valve Pump Assemblyp.47Figure 75. Valve Pump Assemblyp.477.9: Overflow Sampling Assembliesp.49CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURESp.51SECTION 8 ADDITIONAL FEATURESp.51EIAp.54DISVALVp.56CFA3000 MANUAL APPENDIXp.58APPENDIX A CHEMISTRIESp.58APPENDIX B SPECIFICATIONSp.59APPENDIX C POWER INTERCONNECT DESCRIPTIONp.61APPENDIX D BYPASS FOR CALIBRATIONp.62APPENDIX E KEYPAD HELP SHEETSp.63Table E1: Singlestream Keypad Help Sheetp.64THAN HIGH ALARM)p.64Figure E3. Keypad Panel for Multistream Optionp.66Table E2. Multistream Keypad Help Sheetp.66DESIRED, GO TO SET LOWp.67ALARMp.67CFA3000 MANUAL ORDERINGp.69ORDERING INFORMATIONp.69WARRANTYp.70RETURN OF MATERIALp.70CFA3000 MANUAL SECTION 9 RETURN OF MATERIALSp.71SECTION 9.0 RETURN OF MATERIALp.71IMPORTANTp.71Immediate, Reliable Analytical Supportp.73A Worldwide Network of Sales and Servicep.73GERMANYp.73
Page 1 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA-3000 Ion Selective Electrode (ISE) Analyzer
Page 1 · click to enlarge
Page 1

Instruction Manual PN 51CFA3000ISE/rev.F December 2013

CFA-3000 Ion Selective Electrode (ISE) Analyzer

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA-3000 Ion Selective Electrode (ISE) Analyzer diagram (page 1)
Fig. 1 · CFA-3000 Ion Selective Electrode (ISE) Analyzer
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA-3000 Ion Selective Electrode (ISE) Analyzer diagram (page 1)
Fig. 2 · CFA-3000 Ion Selective Electrode (ISE) Analyzer
Page 2 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — WARNING
Page 2 · click to enlarge
Page 2

ESSENTIAL INSTRUCTIONS

WARNINGS

READ THIS PAGE BEFORE PROCEEDING!

RISK OF ELECTRICAL SHOCK

Your purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in oper ating and maintaining the instrument. To ensure their continued operation to the design specifications, per sonnel should read this manual thoroughly before proceeding with installation, commissioning, opera tion, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.

Equipment protected throughout by double insulation.

  • Installation of cable connections and servicing of this product require access to shock hazard voltage levels.
  • Main power and relay contacts wired to separate power source must be disconnected before servicing.
  • Do not operate or energize instrument with case open!
  • Signal wiring connected in this box must be rated at least 240 V.
  • Nonmetallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bush ings and jumper wires.
  • Unused cable conduit entries must be securely sealed by nonflammable closures to provide enclosure integrity in compliance with personal safety and environmental protec tion requirements. Unused conduit openings must be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (NEMA 4X).
  • Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
  • Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call 18008548257 or 9497578500 to request correct manual.
  • Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA70) and/or any other applicable national or local codes.
  • Operate only with front and rear panels fastened and in place over terminal area.
  • For clarification of instructions, contact your Rosemount representative.
  • Safety and performance require that this instrument be connected and properly grounded through a threewire power source.
  • Proper relay use and configuration is the responsibility of the user.
  • Follow all warnings, cautions, and instructions marked on and supplied with the product.
  • Use only qualified personnel to install, operate, update, program and maintain the product.

CAUTION

  • Educate your personnel in the proper installation, operation, and maintenance of the product.
  • Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electrical and pressure sources specified in this manual.

This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation, or operation, may increase such interfer ence. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class A comput ing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to provide reasonable protection against such interference. Operation of this equipment in a residential area may cause interference, in which case the user at his own expense, will be required to take whatever measures may be required to correct the interference.

  • Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process.
  • All equipment doors must be closed and protective covers must be in place unless qualified personnel are performing maintenance.

WARNING

  • If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.

This product is not intended for use in the light industrial, residential or commercial environments per the instrument’s certification to EN500812.

2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 7578500 Fax: (949) 4747250

Emerson Process Management Gas Chromatograph (CFA-3000) — WARNING diagram (page 2)
Fig. 3

http://www.rosemountanalytical.com

© Rosemount Analytical Inc. 2012

Emerson Process Management

Page 3 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — ESSENTIAL INSTRUCTIONS
Page 3 · click to enlarge
Page 3
Page 3 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — ESSENTIAL INSTRUCTIONS
Page 4 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — ESSENTIAL INSTRUCTIONS
Page 4 · click to enlarge
Page 4
Page 4 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — ESSENTIAL INSTRUCTIONS
Page 5 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — Table of Contents
Page 5 · Contents
Contents

Contents

CFA-3000 Ion Selective Electrode (ISE) Analyzerp.1ESSENTIAL INSTRUCTIONSp.2WARNINGSp.2READ THIS PAGE BEFORE PROCEEDING!p.2RISK OF ELECTRICAL SHOCKp.2CAUTIONp.2WARNINGp.2TABLE OF CONTENTS (CON’T)p.5LIST OF FIGURESp.5LIST OF TABLESp.6CFA3000 MANUAL SECTION 1 INTRODUCTIONp.7SECTION 1 INTRODUCTIONp.71.1 Thoroughly Read This Manualp.81.2 Analyzer Benefitsp.81.3 WARNING – Electrical Shock Hazardp.9SECTION 2 INSTALLING THE ANALYZERp.102.0 INSTALLING THE ANALYZERp.102.1 UNPACK PARTSp.102.6 Prepare Electrical Connectionsp.162.8 Prepare Reagent Containersp.19SECTION 3 OPERATING THE ANALYZERp.203.0 Operating the Analyzerp.203.2 Check Sample Streamp.21Enter Test Mode:p.22DECIMALp.22MANUALp.22CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTIONp.23FIGURE 32. KEYPAD PANELp.233.3 Set Reagent Supply Timep.283.4 Set Scaling Outputsp.283.7 Shutdownp.313.8 Quick Startp.32CFA3000 MANUAL SECTION 4 MAINTAINING THE ANALYZERp.33SECTION 4 MAINTAINING THE ANALYZERp.33CFA3000 MANUAL SECTION 5 TROUBLESHOOTING PROCEDURESp.34SECTION 5 TROUBLESHOOTING PROCEDURESp.34OFFp.345.2 Test Functions Table 52: Test Functionsp.35CFA3000 MANUAL SECTION 6 REPAIR PROCEDURESp.37SECTION 6 REPAIR PROCEDURESp.376.2 Replace Circuit Boardsp.386.3 Replace Fusesp.396.4 ISE Reaction Cell Assembly, Electrodesp.39CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTIONp.40SECTION 7 INSTRUMENT DESCRIPTIONp.407.2 Keypad Panelp.42PLEASE NOTE: VALVEPUMP CONNECTOR IS ON CPU BOARDp.44FIGURE 72. CARD CAGE AND BOARDSp.443 ONp.457 OFFp.45FIGURE 74. REAR OF MAIN CABINETp.467.5 Valve Pump Assemblyp.47Figure 75. Valve Pump Assemblyp.477.9: Overflow Sampling Assembliesp.49CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURESp.51SECTION 8 ADDITIONAL FEATURESp.51EIAp.54DISVALVp.56CFA3000 MANUAL APPENDIXp.58APPENDIX A CHEMISTRIESp.58APPENDIX B SPECIFICATIONSp.59APPENDIX C POWER INTERCONNECT DESCRIPTIONp.61APPENDIX D BYPASS FOR CALIBRATIONp.62APPENDIX E KEYPAD HELP SHEETSp.63Table E1: Singlestream Keypad Help Sheetp.64THAN HIGH ALARM)p.64Figure E3. Keypad Panel for Multistream Optionp.66Table E2. Multistream Keypad Help Sheetp.66DESIRED, GO TO SET LOWp.67ALARMp.67CFA3000 MANUAL ORDERINGp.69ORDERING INFORMATIONp.69WARRANTYp.70RETURN OF MATERIALp.70CFA3000 MANUAL SECTION 9 RETURN OF MATERIALSp.71SECTION 9.0 RETURN OF MATERIALp.71IMPORTANTp.71Immediate, Reliable Analytical Supportp.73A Worldwide Network of Sales and Servicep.73GERMANYp.73
Page 6 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — TABLE OF CONTENTS (CON’T)
Page 6 · click to enlarge
Page 6

CFA3000 ISE Analyzer TABLE OF CONTENTS Instruction Manual

TABLE OF CONTENTS (CON’T)

LIST OF TABLES

Table 51 Troubleshooting Chart 28

Table 52 Test Functions 29

Table 71 Main Cabinet Parts 34

Table 72 Reagent Cabinet Parts 34

Table 73 Buttons/LED Functions 36

Table 74 Card Cage Parts 37

Table 75 CPU Board Components 37

Table 76 Analog Board Components 37

Table 77 Valve Board Components 37

Table 78 Power Supply Board Components 37

Table 79 Analog Board DIP Switch Positions 39

Table 710 Back Cabinet Parts 40

Table 711 Valve Pump Parts 41

Table 712 Interface Ports 42

Table 713 Relay Output Option Pin Assignments 42

Table 714 Multioutput Pin Assignments on “D” Connector 42

Table 715 Overflow Sampling Assembly Parts 43

Table 81 Stream Identification Blips (SIBs) 45

Table 82 Normal Data Output Format 48

Table 83 Data Configuration 48

Table 84 RS232C Pin Configuration 48

Table 85 RS232C Commands 50

Table 86 Printer Port Pin Configuration 51

Table E1 Singlestream Keypad Help Sheet 5859

Table E2 Multistream Keypad Help Sheet 6062

Table E3 Replacement Parts 63

Page 7 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 1 INTRODUCTION
Page 7 · click to enlarge
Page 7

CFA3000 MANUAL SECTION 1 INTRODUCTION

SECTION 1 INTRODUCTION

The Rosemount Analytical CFA3000 Series analyzers are designed for minimum maintenance and maximum reliability. The touch of a button starts the microprocessorcontrolled program, calibrates the analyzer, keeps it calibrated, and reports results directly unattended for months at a time. Maintenance is simplified by a modular design with special fittings, labeled tubing, and colorcoded components.

The CFA3000 Series includes analyzers for parameters commonly measured using Ion Selective Electrode (ISE) technology including sodium, fluoride and chloride. Sample concentration is clearly indicated on the digital display panel, and output signals are provided to drive recorders, alarms, or other external devices such as printers or computers. The specific chemistry of your individual analyzer is discussed in Appendix A, Chemistries, or in an accompanying method sheet..

Your purchase from Rosemount Analytical provides you with one of the finest instruments available for your particular application. These instruments have been designed and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowl edge of the user in operating and maintaining the instrument.

Page 8 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 1.2 Analyzer Benefits
Page 8 · click to enlarge
Page 8

CFA3000 ISE Analyzer Instruction Manual

1.1 Thoroughly Read This Manual

This instruction manual describes installation, operation, and maintenance of the CFA3000 ISE Analyzer. To ensure continued operation to the design specifications, personnel should read this manual thoroughly before proceeding with installation, operation, or maintenance of the analyzer.

  • Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call: 18008548257 or 9497578500 to request the correct
  • manual.
  • For any inquiries about your analyzer, contact your Rosemount Analytical representative.
  • Failure to follow the documented instructions may cause any one of the following situations to occur: damage
  • to this instrument and warranty invalidation.
  • Follow all warnings, cautions, and instructions supplied with the product.
  • Only qualified personnel should install, operate, update, program, and maintain the product.
  • All personnel must learn proper installation, operation, and maintenance of the product.
  • Install, operate, and maintain equipment exactly as specified in this manual.
  • Follow appropriate local and national codes.
  • Only connect the analyzer to electrical sources specified in this manual.
  • Use only components supplied by Rosemount Analytical or recommended by them for replacement. Tampering or unauthorized substitution of parts and procedures may affect the performance of the analyzer

as well as void the warranty. • All protective covers must be in place and cabinet panels must be closed at all times after installation or

maintenance.

1.2 Analyzer Benefits

Rosemount Analytical listened to you and decided to move away from a peristaltic pump assembly requiring monthly replacement of the pump tubing. The new pump assembly has a vacuum/pressure valve system which eliminates pump tubing and used significantly less reagent. Since the new valve pump swiftly moves fluids through the analyzer, the potential for clogging within any internal component is reduced. Significantly. As a result, mainte nance is required less frequently and procedures such as cleaning flow cells and the mixing system are no longer required.

The upgraded software now identifies a fixed standard that allows one to change the range of the 05 V and the 420 mA output signals to your requirements.

Page 9 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 1.3 WARNING – Electrical Shock Hazard
Page 9 · click to enlarge
Page 9

CFA3000 ISE Analyzer Instruction Manual

1.3 WARNING – Electrical Shock Hazard

Installation of cable connections and servicing of this product require access to shock hazard voltage levels. Main power and relay contacts wired to separate power sources must be disconnected before servicing. Nonmetallic cable strain relief does not provide grounding between conduit connections! Use grounding type bushings and jumper wires. Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA 70) and/or any

other applicable national or local codes. Operate only with: 1. Main console rear cover fastened. 2. Left side power terminal cover closed. 3. Keypad assembly latched closed. Safety and performance require that this instrument be connected and properly grounded through a three wire power source. Proper configuration and use is the responsibility of the user.

MarkIDDescription EN 610101Safety of electrical equipment EN 55011Limits of RF equipment EN 500821:92Generic immunity from electrostatic discharge, RF & Fast Transient/Burst Immunity EtL 9700516ETL LISTED CONFORMS TO UL STD. 31011 EtLETL LISTED CERTIFIED TOCAN/CSA C22.2 No. 1010.192
Page 10 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 2.0 INSTALLING THE ANALYZER
Page 10 · click to enlarge
Page 10

CFA3000 ISE Analyzer Instruction Manual

SECTION 2 INSTALLING THE ANALYZER

2.0 INSTALLING THE ANALYZER

2.1 UNPACK PARTS

When you receive your new CFA3000 analyzer, carefully unpack the items in the following list.

√ Overflow sampler assembly

√ Waste tube assembly

√ Valve pump assembly (integrated housing includes pump, labeled tubes with straws)

If any of the items are missing or damaged, contact Rosemount Analytical at: our:

Tollfree number(800) 8548257
Direct number(949) 7578500
Fax(949) 8639159

Inspect each piece to confirm that there is no damage or loose parts. If anything is missing or appears broken or damaged, contact Rosemount Analytical at the tollfree phone number above.

Show 5 more labels / short text on this page
√ Main cabinet · √ Reagent cabinet · √ Mounting brackets · √ Accessories pack · √ Startup reagents
Page 11 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — SECTION 2 INSTALLING THE ANALYZER
Page 11 · click to enlarge
Page 11

CFA3000 ISE Analyzer Instruction Manual

After analyzer has been unpacked and all parts accounted for, install as described in following sections. Refer to Figure 21 to view a completely installed analyzer.

Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 11)
Fig. 4
Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 11)
Fig. 5

Figure 21. Opened Assembled Cabinet

Page 12 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — SECTION 2 INSTALLING THE ANALYZER
Page 12 · click to enlarge
Page 12

CFA3000 ISE Analyzer Instruction Manual

2.2 Mount Main Cabinet

  • This cabinet is not for outdoor installation. Environmental conditions of enclosed facility:
  • Temperature: min. 50°F to max. 113°F (min 10°C – max. 45°C) ±10° within regulated range.
  • Humidity: noncondensing Mount cabinet on a wall or other suitable vertical surface, at least 42” (107 cm) wide and 431/2” (111 cm) off the floor. Mounting surface must be able to support 250 pounds (113.6 kg). Allow 11/2 feet (.5 m) of clearance at left side of analyzer for access to input/output panel, and allow for swing out for rear access. Allow 12” (.3 m) at right side for access to overflow tube assembly (if single stream). Mount cabinet level, or with a slight backward tilt for proper leak detector operation.

To mount main cabinet, proceed as follows (see Figure 22):

1. Securely attach supplied "L" brackets to wall using suitable hardware. Brackets must be anchored properly to support 250 pounds (113.6 kg). 2. Attach main cabinet to "L" brackets using supplied four bolts, split ring washers, and hex nuts.

2.3 Mount Reagent Cabinet

To mount reagent cabinet, proceed as follows:

1. Loosen the four bolts in bottom of main cabinet. 2. Lift reagent cabinet and attach to main cabinet by placing heads of bolts into keyhole slots in top of reagent cabinet. 3. Slide reagent cabinet back. 4. Feed grounding strap on bottom of main cabinet through notched hole in left rear corner and attach it to stud on back of reagent cabinet (see Figure 23). 5. Tighten screws to attach reagent cabinet to main cabinet. 6. For multistream units, plug valve bank cable into positive fit receptacle at left side of cabinet (refer to Figure 71, “Multistream Valve Panel Connector”).

General Recommendations

Page 13 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — SECTION 2 INSTALLING THE ANALYZER
Page 13 · click to enlarge
Page 13

CFA3000 ISE Analyzer Instruction Manual

Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 13)
Fig. 6
Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 13)
Fig. 7

Figure 22. Mounting Main and

Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 13)
Fig. 8
Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 13)
Fig. 9

Figure 23. Attaching

Reagent Cabinets · Grounding Strap

Page 14 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — SECTION 2 INSTALLING THE ANALYZER
Page 14 · click to enlarge
Page 14

CFA3000 ISE Analyzer Instruction Manual

2.4 Install Waste Tube

To install waste tube, proceed as follows (refer to Figure 21):

1. Guide waste tube through main cabinet into reagent cabinet and out through bottom of reagent cabinet. 2. Secure plastic waste tube in clamp in lower right side of main cabinet. 3. Insert Colorimeter drain tube into waste tube.

2.5 Mount Overflow Sampling Assembly

The Sample intake tube should extend into

The overflow tube approximately 1½” – 2”

3.81 –5.08 cm).

To mount overflow sampling assembly proceed as follows (for sample configurations, see Figures 24a and 24b):

1. Mount overflow sampler assembly on outer, right side of reagent cabinet using the provided four machine screws. For multistream units, mount overflow assembly on left side of cabinet. (If using IF100 InLine Filter, refer to instruction sheet accompanying filter for complete installation instructions.) 2. Place ends of air vent tubing and overflow waste tubing into waste pipe. 3. Connect waste pipe to drain. 4. Supply sample stream to regulating or intake valve with 1/4” tubing (plastic or steel) as follows: Insert tubing into tube fitting. Make sure tubing rests firmly on shoulder of fitting and nut is fingertight. Mark nut at 6 o’clock position. While holding fitting body steady with a wrench, tighten nut 11/4 turns (go to 9 o’clock position). 5. For multistream assembly, connect sample intake line to bulkhead fitting. 6. For singlestream assembly, attach sample intake line to valve pump. 7. Make sure that sample liquid overflows into overflow chamber.

Page 15 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — SECTION 2 INSTALLING THE ANALYZER
Page 15 · click to enlarge
Page 15

CFA3000 ISE Analyzer Instruction Manual

Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 15)
Fig. 10
Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 15)
Fig. 11

Figure 24a. Singlestream Overflow Sampling Assembly

Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 15)
Fig. 12
Emerson Process Management Gas Chromatograph (CFA-3000) — SECTION 2 INSTALLING THE ANALYZER diagram (page 15)
Fig. 13

Figure 24b. Multi

stream Overflow Sampling Assembly, threestream option)

Page 16 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 2.6 Prepare Electrical Connections
Page 16 · click to enlarge
Page 16

CFA3000 ISE Analyzer Instruction Manual

2.6 Prepare Electrical Connections

made in accordance with local or national electrical code as applicable. This equipment must be grounded. A qualified electrician should wire this equipment to an electrical circuit.

Electrical connections are made to and from the cabinet at the input/output panel, located on the left side of the analyzer. See Figure 25.

  • Terminals are provided for ground, power supply, high/low alarms, output signals, calibration and status relays
  • Output signals a voltage output of 05 Vdc and a floating, ungrounded current output of 420 mA are available for remote readout or control of a process.
  • Also available are contact closures for the following: High alarm Calibration Low Alarm Leak Alarm

To make electrical connections, proceed as follows:

Use cable glands to hold cables securely in holes above input/output panel.

1. Remove solid caps from the holes above the input/output panel by puncturing with an awl and prying them out. (Note: remove caps only from those holes that will be used.) 2. Replace with permanent connectors. 3. Connect alarms and external computer interfaces (if any). 4. Feed 3wire power cable into analyzer through hole above the input/output panel. 5. Connect wires to appropriate contacts on Terminal Strip TB3.(See next page)
Page 17 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 2.6 Prepare Electrical Connections
Page 17 · click to enlarge
Page 17

CFA3000 ISE Analyzer Instruction Manual

Emerson Process Management Gas Chromatograph (CFA-3000) — 2.6 Prepare Electrical Connections diagram (page 17)
Fig. 14 · 2.6 Prepare Electrical Connections
Emerson Process Management Gas Chromatograph (CFA-3000) — 2.6 Prepare Electrical Connections diagram (page 17)
Fig. 15 · 2.6 Prepare Electrical Connections

Figure 25. Input/Output Panel

2.7 Install Valve Pump Assembly

The pump assembly has a vacuum pressure system that:

l Eliminates internal pump tubing

l Uses significantly less reagent

l Accurately moves fluids through the analyzer.

Perform the following procedure to install valve pump assembly:

1. Tilt pump assembly towards you. 2. Align the hinge pins and sockets, and slide assembly tray as far as possible to right (see Figure 27). Carefully lay tray down into cabinet 3. Plug in the two connectors to the back of the cabinet (6position connector for heating bath; 9position connector for pump) (see Figure 28). 4. Connect 16pin connector from pump assembly to small connector on CPU board (yellow ejector) above connector for keypad panel. 5. Attach colorimeter inlet tubing on left side of pump to flowcell inlet (see Figure 21). 6. Continue with procedure depending on whether system is singlestream or multistream.

For singlestream, continue:

6. Thread tube labeled Sample" through hole 0n bottom of shelf to the right of pump assembly, and out through hole on left side of cabinet. 7. Connect Sample tube to top of overflow sample on left side of cabinet. 8. Guide all other tubes that are attached to valve pump assembly through reagent cabinet access hole.

For multistream, continue:

6. Thread Sample tube from pump assembly through reagent access hole on the right side of the valve pump assembly, and connect to Common (C) on #1 multistream solenoid valve on left side of reagent cabinet.

Guide all other tubes that are attached to valve pump assembly through reagent cabinet access hole.

Page 18 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 2.6 Prepare Electrical Connections
Page 18 · click to enlarge
Page 18

CFA3000 ISE Analyzer Instruction Manual

Emerson Process Management Gas Chromatograph (CFA-3000) — 2.6 Prepare Electrical Connections diagram (page 18)
Fig. 16 · 2.6 Prepare Electrical Connections
Emerson Process Management Gas Chromatograph (CFA-3000) — 2.6 Prepare Electrical Connections diagram (page 18)
Fig. 17 · 2.6 Prepare Electrical Connections

Figure 26. Valve Pump Assembly

Emerson Process Management Gas Chromatograph (CFA-3000) — 2.6 Prepare Electrical Connections diagram (page 18)
Fig. 18 · 2.6 Prepare Electrical Connections
Emerson Process Management Gas Chromatograph (CFA-3000) — 2.6 Prepare Electrical Connections diagram (page 18)
Fig. 19 · 2.6 Prepare Electrical Connections

Figure 27. Pump Connector

Page 19 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 2.8 Prepare Reagent Containers
Page 19 · click to enlarge
Page 19

CFA3000 ISE Analyzer Instruction Manual

2.8 Prepare Reagent Containers

Prepare reagent containers as follows: 1. Remove screw caps from reagent bottles. 2. Using an awl, make one hole in foil for the reagent straw. Do not allow awl to touch reagent or standards. 3. Place Standard and Reagent bottles into reagent cabinet, as shown in Figure 28. 4. Insert colorcoded reagent straws and labeled Standard straw leading from valve pump assembly into holes punched in corresponding bottles.

Emerson Process Management Gas Chromatograph (CFA-3000) — 2.8 Prepare Reagent Containers diagram (page 19)
Fig. 20 · 2.8 Prepare Reagent Containers

Figure 28. Sample Bottle Placement in Cabinet (for all methods; colors refer to reagent straw colors)

Page 20 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 3.0 Operating the Analyzer
Page 20 · click to enlarge
Page 20

CFA3000 ISE Analyzer Instruction Manual

SECTION 3 OPERATING THE ANALYZER

3.0 Operating the Analyzer

3.1 Operational Overview The analyzer uses a batch flow method for automatically analyzing a sample against a fixed standard (refer to block diagram in Figure 31). That is, the touch of a button starts the microprocessorcontrolled program, calibrates the analyzer, takes measurements of samples, and reports results directly unattendedfor months at a time.

The automated monitoring program determines the sequence for activating each of the valve pumps in the valve pump assembly to inject air and fluids into the reaction chamber. Each valve pump is associated with a specific input via the quick disconnect connector on the side of the assembly (the number of reagents is determined by your particular chemistry and procedure).For example, the inputs for sodium chemistry would be:

  • Buffer or Ionic Strength Adjustor, reagent #1
  • Sample
  • Grab sample
  • Baseline
  • Standard
  • Air
  • Evacuation of chamber
  • Recirculate

For a multistream system, solenoid valves in the analyzer reagent cabinet open and close streams for sampling. When sampling, the keypad panel LEDs identify which sample stream (valve number) is associated with the dis played value, and which valve is now open for sampling. Although the system runs samples on its own, you may also perform a manual grab sampling or a manual calibration at any time.

Figure 31. Simple Block Diagram

Show 6 more labels / short text on this page
ISE (ion selective · Valve Pump · Reaction · Electronics Data output · Cell · electrode)
Page 21 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 3.2 Check Sample Stream
Page 21 · click to enlarge
Page 21

CFA3000 ISE Analyzer Instruction Manual

3.2 Check Sample Stream

  • Check that the sample intake tube is 1½ to 2 inch (3.81 to 5.08 cm) into the liquid.
  • Sample should just overflow the overflow tube. There should be no particulate or bubbles in the tube.

Using a MultiStream Overflow Sampler

Multistream analysis requires an overflow sampler assembly that will accommodate the extra sample streams. One multistream chamber allows for up to three sample streams; for 4 6 streams, use two chambers.

The sample tubes connect to solenoid valves in the reagent cabinet. The analyzer uses the valves to cycle through all sample lines, in sequence. First, stream 1 valve is opened, and Valve 1 LED on the keypad panel is lit. When the measurement is complete, Valve 1 will be closed, and Valve 2 is opened to begin sampling stream 2. The keypad panel will show Valve 1 LED lit as the Displayed Stream, and Valve 2 LED is lit for the Selected Stream.

LED's on the valve pump assembly light up when a particular fluid is taken up by the valve pump assembly. Figure 32 identifies the valve pump LED designations.

3.2.1 Check Fluidics

Check the following:

  • Liquid should overflow in overflow chamber.
  • Waste tubing is in waste pipe.
  • Sample tube is connected to Overflow Sampler Assembly for single stream; and, for multistream, it is connected to Common (C) port of sample solenoid valve #1, on left side of reagent cabinet.
  • System has no leaks.

3.2.2 Check Reagents

Check the following:

  • One hole has been punched in all reagent cap liners.
  • Reagent straws are placed in appropriate reagent containers.
  • High and low standard tubing are in appropriate containers.

3.2.3 Turn On Power

1. Open the keypad panel door to expose the circuit boards. 2. Press each board in to ensure proper seating. 3. Simultaneously press Power button on the back wall of upper cabinet and Reset button on the CPU board (yellow ejector), then release. 4. First release the Power button, then the Reset button. The Keypad should display ”H E L O", then four dashes " ". 5. Pull pump forward and check leak detector by dipping finger in tap water and touching detector behind pump. Leak detector LED on keypad panel will light up. 6. Turn main power switch to OFF. 7. Dry leak detector 8. Wait 20 seconds, then switch main power switch back to ON. Leak detector LED should be off.

Page 22 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — DECIMAL
Page 22 · click to enlarge
Page 22

CFA3000 ISE Analyzer Instruction Manual

Reaction cell takes approx. 30 minutes to heat up.

If leak detector LED does not light up:

Adjust potentiometer on analog board counterclockwise for more sensitivity.

Turn off power to reset system. Wait 20 seconds, and simultaneously press Power and Reset buttons. Dip finger in tap water and touch leak detector contacts on bottom of cabinet behind the valve

pump. LED should light up. If problem persists, contact Rosemount Analytical.

3.2.4 Test Valve Pump Operation

The test procedure allows you to confirm that all physical connections are installed properly, and to ensure that the pump is taking up the proper fluids at the correct time.

Enter Test Mode:

1. Put all straws and Sampling tube into a beaker of DI water. 2. Turn power on by simultaneously pressing Power button on the back of upper cabinet and Reset button on CPUboard (yellow ejector). 3. First release Power button, then Reset button. Keypad should display "H E L O", then four dashes " ". To enter test mode, continue with the step appropriate for your analyzer setup:

For singlestream, continue:

For multistream, continue:

Referring to Figure 32, on keypad panel, press posi tion 15 and then 2

On keypad panel, pressthen

DECIMAL

MANUAL

Page 23 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — FIGURE 32. KEYPAD PANEL
Page 23 · click to enlarge
Page 23

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

The keypad buttons may now be used to test pump operation. Do not rush through the test steps – when a specific step is initiated, it takes approx. 2 seconds to pump air through the system. Actual pumping of a liquid takes approximately 4 seconds. After this operation

is finished, another 4 seconds elapses before the pump is primed for the next test step.

To exit the test at any time, press the ENTER button on the keypad panel.

If a test step does not work properly, try the following:

√ Take a syringe and shoot air through straw to make sure it is not clogged.

√ Check that corresponding LED on valve pumps lights up (refer to Figure 41).

√ Check that fluid is picked up in correct tube. If the problem still persists, contact Rosemount Analytical.

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION diagram (page 23)
Fig. 21
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION diagram (page 23)
Fig. 22

FIGURE 32. KEYPAD PANEL

Begin Test:

Confirm that the corresponding Pump LED lights up for each step (refer to Figure 33 for LED designations during test):

Page 24 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION
Page 24 · click to enlarge
Page 24

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

1. Press ZERO to test air. Press ZERO to turn off.
2. Press to simulate one shot of Reagent #4. Look at blue straw to confirm that water is pumped up through tube. Repeat a few times until you see water is in clear section of tube
3. Press to simulate one shot of Reagent #3. Check green straw and its clear tube
4. Press to simulate one shot of Reagent #2. Check orange straw and its clear tube.
5. Press to simulate one shot of Reagent #1. Check red straw and its clear tube
6. Press to simulate one shot of Sample coming from overflow sampler. Check tube labeled "Sample."
7. Confirm that fluid is going to flowcell.
8. Press to simulate one shot of Standard. Check corresponding clear straw and its clear tube.
9. Press to pump one shot of air into analyzer.
10. Press to simulate one shot of Baseline solution. Check clear straw labeled, BASELINE, and tube.
11. Repeat Step 9 until you see fluid shoot into reaction chamber.
12. Press to simulate evacuation from the reaction chamber to the flowcell. Check the tube going up to flowcell.
13. Press to simulate one shot of Grab solution. Check clear straw and tube.
14. Press to wash flowcell.
Page 25 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION
Page 25 · click to enlarge
Page 25

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION diagram (page 25)
Fig. 23
15. Press to empty reaction chamber. If there is any fluid left, it will be shot through to Waste.

valvepump. When the LCD number disappears, the next position may be pressed. The sequence is

Page 26 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION
Page 26 · click to enlarge
Page 26

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

3.2.5 Set Up Stream Parameters

The analyzer is programmed to measure a singlestream system at 4 sample cycles per hour. The multistream default is 1 cycle per stream, but you may define parameters for individual streams, such as modifying the num ber of cycles or the alarm threshold values.

The alarm values are used to activate alarms or controllers when the sample is outside of the specified concen tration range. A high alarm will activate when the sample concentration exceeds a preset limit. A low alarm will activate when the sample concentration is below the preset limit.

The contacts for the alarm outputs are located at the input/output panel on the cabinet side. The system must be in an alarm condition for at least one minute to activate the alarm contact.

Use the logical flows below to define desired stream parameters.

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION diagram (page 26)
Fig. 24
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION diagram (page 26)
Fig. 25

Figure 34. Logic Flow for Singlestream Parameter Definition

Page 27 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION
Page 27 · click to enlarge
Page 27

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

(NoteUse decimal on the keypad using arrows to place decimal in the correct position)
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION diagram (page 27)
Fig. 26
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION diagram (page 27)
Fig. 27

Figure 35. Logic Flow for Multistream Parameter Definition

Page 28 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 3.4 Set Scaling Outputs
Page 28 · click to enlarge
Page 28

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

3.3 Set Reagent Supply Time

Use the keypad panel to remind you when you need to replace the reagent. The analyzer will count pump cycles and count down the number of days left. To specify the duration, proceed as follows:

1. Press once to set counter for 90 days (3 months).
Sample Display0 0 9 0
2. Press to exit and clear display.

To determine how many days are left until you have to replace reagent,

Press: once.

Look at the LCD and record the number of days;

then press: again to get out of mode.

3.4 Set Scaling Outputs

Set output 420 mA 05 Vdc from the keypad panel, (refer to Figures 72 and 73), then adjust 420 mA and 05 Vdc at your external device.

Set output scale to4 mA/0 V
1. Press: .once then .to display 4 mA and for 0 Vdc.

The sample Display for 4 mA and for 0 Vdc: 0 0 0 4

2. To exit, press
Set output scale to20 mA/5 V
1. Press once then to display 20 mA and 5 Vdc.
Page 29 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION
Page 29 · click to enlarge
Page 29

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

2. The sample display for 20 mA and for 5 Vdc: 0020
2. To exit, press .

To display the fullscale concentration embedded in the EProm:

Press two times. To exit, press .

3.4.1 Record mV Use the mV value to determine that the method’s span and slope are correct. The change in millivolts (mV) is the spread between a baseline and a fullscale standard at the time of calibration. The values are relative to the chemical parameter being measured

Record the millivolt values for both baseline and fullscale standard.

1. Press

The sample Display for value of 150 mV is: 0 1 5 0

2. To exit function, press once.

3.4.2 Initiate Auto Calibration

The analyzer is calibrated automatically usually every two days, depending on the chemistry. Auto calibration performs both baseline and fullscale calibration. Baseline calibration measures a solution with 10% concentra tion of the species of interest. Fullscale calibration measures a standard solution in the upper range of the parameter of interest.

Before starting calibration, make sure that the analyzer is running reaqgents and baseline

solution for at least 20 minutes.

Page 30 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION
Page 30 · click to enlarge
Page 30

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

To initiate auto calibration, press on the keypad panel – the CALIBRATION LED will light. When multi stream systems are being calibrated, only Stream 1 LED is lit.

Baseline calibration will run first for approximately 15 minutes. Fullscale calibration follows, lasting approximately 15 minutes (the actual time for calibration depends on the chemistry). When fullscale calibration is finished, CALIBRATION LED goes out and the value is displayed on the LCD. The “ ” display signifies that a delay is required before the next sample is displayed.

Sample analysis automatically begins after calibration. The time delay from the start of sample analysis to dis play of results is approximately 15 minutes. This allows sufficient time for the previous solution to wash through system, for sample and reagents to react, for analyzer to measure sample and display concentration value.

When the CALIBRATION LED and the ”000” display in a singlestream system go out, the displayed value on the keypad panel is the actual stream value.

For a multistream system, after auto calibration, Sample Stream status LED’s will light to show sample stream valve currently selected and sample stream concentration currently displayed. For example, for the first stream sample, VALVE SELECTED LED is lit for that stream; dashes display on the keypad panel. When analyzer finish es sampling one stream and moves on to the next, the display shows the previous sample stream concentration value, and its VALVE DISPLAYED LED lights up. VALVE SELECTED LED will be lit for next valve in sequence.

then, , then to abort the calibration. Press twice in rapid succession on initial startup for priming of sample & reagents

3.5 Process a Grab Sample The analyzer allows you to rapidly analyze a grab sample.

Press on the keypad panel, to introduce a grab sample into the analyzer from the holder on the overflow sample panel.

If visible, suspended solids are present in the sample, filter it to 8 microns or less.

Page 31 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 3.7 Shutdown
Page 31 · click to enlarge
Page 31

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

Place the clear grab sample inlet tube into a beaker containing the grab sample.

2. Press one time. The display shows flashing dashes, “ “ . The system analyzes the grab sample and after 15 minutes, displays the results when complete (The results are “flashed” on the display).
Grab Sample Display10.0 (flashing for three minutes)

After the display of the grab sample’s value, a multistream analyzer begins processing the next stream in sequence.

3. To display measurement for previous grab sample, press twice.
4. Press button to return to normal operation.

3.6 Manually Sample a Particular Stream.

Use this procedure to manually switch between sample streams in a multistream system. This operation can be performed at any time, except during calibration.

Press one time. Sample Display: 0 0 0 1 .

Press and to change stream number. Sample Display: 0 0 0 3

Press to start sampling desired stream.

Press to return to multistream automatic analysis.

3.7 Shutdown

To shut down the CFA3000 for only a few days, press main Power button in back of upper cabinet to OFF position. Leave reagent straws in reagent containers.

Page 32 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 3.8 Quick Start
Page 32 · click to enlarge
Page 32

CFA3000 MANUAL SECTION 3 INSTRUMENT DESCRIPTION

Some chemistries may require specific wash solutions. Check Appendix A,

"Chemistries," for details.

To shut down the analyzer for more than a day, perform the following procedure

1. Remove straws from reagent containers and place them into a beaker of DI water.
2. Press twice within a twosecond interval to activate priming. This allows wash solution or DI water to be injected rapidly through system.
3. After priming finishes (after 15 minutes or when “auto cal” starts), press the main Power button in back of upper cabinet to OFF position.

3.8 Quick Start

If the analyzer has been shut down during normal operations, turn power on and

press once.

If the CFA3000 has been turned off for an extended shutdown:

Press twice to prime; the system will automatically go into :”Auto Cal” after priming

Page 33 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 4 MAINTAINING THE ANALYZER
Page 33 · click to enlarge
Page 33

CFA3000 MANUAL SECTION 4 MAINTAINING THE ANALYZER

SECTION 4 MAINTAINING THE ANALYZER

Your CFA3000 Series analyzer is a precision instrument which, if maintained properly, should provide years of accurate and reliable service.

4.1 Analyzer Equipment.

Even though the analyzer normally runs automatically, perform the following activities, quarterly:

1. Start Auto Calibration after reagent replenishment (see Section, "Perform Auto Calibration"). 2. Check mV value
Page 34 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — OFF
Page 34 · click to enlarge
Page 34

CFA3000 MANUAL SECTION 5 TROUBLESHOOTING PROCEDURES

SECTION 5 TROUBLESHOOTING PROCEDURES

This chapter is divided into two sections:

  • Troubleshooting Chart: a list of symptoms, probable causes, and remedies.
  • Test Functions:– test functions for diagnostic information.

5.1 Troubleshooting Chart Symptoms observed during operation are generally enough to identify a faulty component.Such symptoms, together with probable causesand remedies are listed in the following troubleshooting chart.

Table 51Troubleshooting Chart

Problem Probable Cause Remedy

Problem Probable Cause Remedy

No power Power switch

Display or associated power circuit

OFF

light up or buttons fail to

switch OFF Press switch ON

Blown fuse Replace fuse

Find and fix leaking; check pump’s electrical

Emerson Process Management Gas Chromatograph (CFA-3000) — OFF diagram (page 34)
Fig. 28
Emerson Process Management Gas Chromatograph (CFA-3000) — OFF diagram (page 34)
Fig. 29

Check pump’s plugs to CPU &

Still inoperative Call Rosemount

Circuit boards are not seated

Remove and reinstall circuit

light up or buttons fail to

boards Display or associated power circuit

Show 32 more labels / short text on this page
Keypad lights or · Press power · display fails to · switch ON · Call Rosemount · Analytical · Power switch · Call Rosemount · faulty · failed · Analytical · operate · Pump is inoperative · Valve pump · Connector faulty · Check connection · or not properly · inserted · Leak detector · activated · connection · Bad electrical · connection · connector · Analytical · Keypad lights or · display fails to · properly · operate · Call Rosemount · Analytical · faulty
Page 35 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 5.2 Test Functions Table 52: Test Functions
Page 35 · click to enlarge
Page 35

CFA3000 MANUAL SECTION 5 TROUBLESHOOTING PROCEDURES

Problem Probable Cause Remedy

Problem Probable Cause Remedy

Replace sample filter on overflow

Tubes pinched shut or clogged

lower than high alarm

causing “pinched” lines; replace sample filter at overflow

must exist for minimum of 60 seconds before

Leak detector LED lights up

Leak in fluidics Repair leak and dry leak detector

OFF for 20 seconds

Clogged tubing Replace tubing

High humidity Adjust screw on

No signal, or signal “offscale” high

boards for proper seating

on analog board; turn clockwise to

Check connection to analog board; replace analog

Reagents, standards old, contaminated or

Sensitivity LED ISE on

board.

5.2 Test Functions Table 52: Test Functions

in wrong reagent or

Problem Probable Cause Remedy

Test LED’s and LCD on keypad: a. Press Power button and reset

Press:

button on the CPU board (yel

low ejector) b. Release power, then Reset button.

Show 55 more labels / short text on this page
Alarms not · Incorrect · Readjust settings, making · Loss of sample · activated · settings · flow · sure that low · panel · alarm is set · Decreased pumping volume · Remove condition · from deposits · Insufficient time · Alarm condition · for alarm to · respond · panel · alarm activates · Signal not · Small leak or · Check all tubing · repetitive (same sample · clogging in · lines and connections · contacts; turn · fluidics · main power · not repetitive) · leak detector potentiometer · Connector or · Check circuit · wires loose · Faulty circuit · board · Replace power · decrease sensitivity · supply circuit · Incorrect mV · board · value · Connector faulty · or not properly · Loss of mV · Change reagents · insertede · improperly · prepared · Reagent straws · Check placement of · straws · standard · Front panel · display · Press power
Page 36 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 5 TROUBLESHOOTING PROCEDURES
Page 36 · click to enlarge
Page 36

CFA3000 MANUAL SECTION 5 TROUBLESHOOTING PROCEDURES

Press:

Press:

button.

c. Press the following keys in sequence: (key #’s 15, 5, 6)

Press:

Show 9 more labels / short text on this page
Sensitivity · (mV) · Check · Exit · Test Procedure · then · then · Front panel · display
Page 37 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 6 REPAIR PROCEDURES
Page 37 · click to enlarge
Page 37

CFA3000 MANUAL SECTION 6 REPAIR PROCEDURES

SECTION 6 REPAIR PROCEDURES

The Ionselective electrode analyzers are designed with modular components for quick and easy replacement. The following sections contain instructions to remove parts and replace them. For more complex repairs, or diag nostic assistance, contact Rosemount Analytical at:

  • Telephone: (800) 8548257, (949) 7578500
  • Fax: (949) 4747250
  • Email: <a href="http://www.raihome.com" rel="nofollow noopener" target="_blank">www.raihome.com</a>

6.1 Replace Valve Pump Assembly Use the following procedure to replace pump assembly:

1. Turn off power to analyzer. 2. Pull back of pump assembly towards you to a 45degree angle (see Figure 61). 3. Unplug the two connectors from the back of the cabinet. (9position connector for pump). 4. Disconnect pump assembly’s “D” connector from right side of pump. 5. Remove all liquid tubes from pump. 6. Disconnect waste tube from bottom of reaction cell (right side of valve pump). 7. Slide assembly to the left to disengage from pins, and remove pump assembly. 8. To install a new pump assembly, refer to Section, " Install Valve Pump Assembly".

Figure 61. Remove Valve Pump Assembly

Page 38 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 6 REPAIR PROCEDURES
Page 38 · click to enlarge
Page 38

CFA3000 MANUAL SECTION 6 REPAIR PROCEDURES

6.2 Replace Circuit Boards

Use the following procedure to replace circuit boards:

1. Turn off power. 2. Open latch from right side of keypad panel; and open door. 3. Disconnect all cables and plugs from board to be replaced. 4. Unlatch colored clip from board to be replaced (refer to Figure 62). 5. Pull circuit board straight out. 6. Reverse steps to replace board and reconnect cables. Press firmly to ensure that boards are properly seated. 7. Turn power on. 8. Perform Automatic Calibration (refer to Section, “Initiate Auto Calibration”). Please NOTE: Valvepump connector is on CPU board

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 6 REPAIR PROCEDURES diagram (page 38)
Fig. 30
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 6 REPAIR PROCEDURES diagram (page 38)
Fig. 31

Figure 62. Card Cage

Page 39 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 6.4 ISE Reaction Cell Assembly, Electrodes
Page 39 · click to enlarge
Page 39

CFA3000 MANUAL SECTION 6 REPAIR PROCEDURES

6.3 Replace Fuses

Use the following procedure to replace fuses (all fuses are 1A):

1. Turn off power. 2. From inside upper left corner near pump switch, press down on fuse cap and turn counterclock wise to remove. 3. Pull fuse and cap straight out. 4. Replace fuse in cap (make sure the new fuse is the same type and rating as the one you are replacing). 5. Press fuse cap in place and turn clockwise. 6. Turn power on.

6.4 ISE Reaction Cell Assembly, Electrodes

Emerson Process Management Gas Chromatograph (CFA-3000) — 6.4 ISE Reaction Cell Assembly, Electrodes diagram (page 39)
Fig. 32 · 6.4 ISE Reaction Cell Assembly, Electrodes
Emerson Process Management Gas Chromatograph (CFA-3000) — 6.4 ISE Reaction Cell Assembly, Electrodes diagram (page 39)
Fig. 33 · 6.4 ISE Reaction Cell Assembly, Electrodes

Figure 63. ISE Reaction Cell Assembly, Electrodes

6.4.1 The sodium electrode used in the CFA3011 sodium analyzer is a proprietary “solid state” electrode which is joined with a reference electrode in a sealed cartridge. The buffered sample stream flows into the reaction cell—through the ground stud to the sodium electrode, then into the reference electrode, out through the ground stud and then to waste. The sample concentration is measured in the sodium electrode, compared to the reference value and the sodium concentration is determined.

Page 40 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION
Page 40 · click to enlarge
Page 40

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

SECTION 7 INSTRUMENT DESCRIPTION

7.1 Analyzer Front View (Refer to Fig. 71).

7.1.1 Main Cabinet

Table 71.1Main Cabinet Parts

Main Cabinet Splashproof enclosure protects analyzer

Chemistry Module Tubing Carries sample and

Reagent Cabinet Splashproof enclosure protects reagents and solenoid valves

reagent to valve pump assembly

Door Hinges Pin hinges mount doors on main and reagent cabinets Door Latch Secures cabinet door

Quickdisconnects Connects reagents, samples, and standards to valve pump assembly

Waste Tube Carries waste solutions to drain

Keypad Panel and Driver Board

External side contains touch buttons for controlling analyzer, digital display, and status panels. Internal side contains driver board

Leak Detector Senses leakage of solutions in main cabinet, shuts pump off, and activates leak LED

HB Lamp Indicates when propor tional controller is heating the heat bath

7.1.2 Reagent Cabinet

HB T 1 A 1 A fuse for heating bath

Table 72. Reagent Cabinet Parts

LAMP PS T 1.0 A 1 A fuse for power supply to ise lamp

MultiStream Valve Panel (MultiStream option only)

Allows sequential analysis of up to six sample streams

PUMP T 1.0 A 1 A fuse for pump

XFMR T 1.0 A 1 A fuse for transformer

MultiStream Connector (MultiStream option only)

Positive fit connector connects Multistream valve panel to main cabi net electronics

Pump On/Off Lighted switch controls power to pump

On/Off Main power switch

Bulkhead Fitting Fitting that directs individual sample stream to sample valve

Waste Drain Pan Drain pan directs waste solutions to waste tube

Pump Connector 9position connector connects pump to main cabinet electronics

Supplies continuous sample streams for analyzer

Heating Bath Connector 6position connector connects to heating bath assembly

Valve Pump Assembly Includes pump, PC board, labeled tubes, and reaction vessel assembly (chemistry module with integrated

Show 6 more labels / short text on this page
(heating bath) · Part Function · Part Function · Part Function · Colorimeter Detector · Overflow Sample Assembly
Page 41 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION
Page 41 · click to enlarge
Page 41

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

Figure 71. Analyzer Front View

Page 42 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION
Page 42 · click to enlarge
Page 42

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

7.2 Keypad Panel

The keypad panel has a fourdigit liquid crystal display with an adjustable decimal point. The panel also has indi cator LED’s and pressuresensitive buttons.

Table 73 Buttons/LED Functions

For Both Single and Multistream Options

FunctionDisplays Systems mV

LEAK LED When lit, indicates leak in main cabinet

Initiates processing from grab sample

CALIBRATION LED When lit, indicates calibration in progress ISE/LAMP LED Will light when millivolt level is below default levels.e

Initiates automatic calibration of analyzer

REAGENTS LED When lit, indicates reagents need to be replaced

Used for setting alarm values.

UP or DOWN Arrow

Increases or decreases displayed value

For Multistream option only

VALVE SELECTED shows which valve (stream) is being analyzed, and VALVE DISPLAYED shows which valve is associated with completed and displayed value

Sets value or exits function

Sets high alarm limit

Initiates automatic multistream analysis

Sets low alarm limit

Initiates single stream analysis

Two functions: Sets 90 day clock for reagent usage Displays number of days remaining until next replacement of reagents

Enters function to identify parameters for each stream (such as number of cycles and high and low alarms)

Sets scale to adjust out put to 4 mA and 0 Vdc on recording device Sets scale to adjust out put to 20 mA and 5 Vdc on recording device

Show 4 more labels / short text on this page
Button/LED Function · Button/LED Function · Sample Streams LED’s · (Not used)
Page 43 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION
Page 43 · click to enlarge
Page 43

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

7.3 Card Cage The card cage is located directly behind the keypad panel. To access the card cage, lift latches from right side of keypad panel and open hinged door. (Refer to Figure 72.)

7.3.3 Analog Board Table 76Analog Board Components

Potentiometer Adjusts sensitivity of leak detector – clockwise for less sensitivity; counterclockwise for greater sensitivity

7.3.1 Card Cage and Door Table 74: Card Cage Parts

Analog Board DIP Switch Contains 8 rocker switches

Keypad Panel Door Opens to access card cage

for setting various functional processes (see Figure 73)

Display Board Circuit board contains electronics for keypad and LCD display

Signal Input Connector For process signals

On 180B0250120 multistream output potentiometers

For each stream 2 – 6, potentiometers for 4 mA, 20 mA, 0 V, 5 V (multistream only)

CPU Board 180B02801 Central processing circuit board processes the signal fed to the digital display and any external devices. (Yellow ejector)

7.3.4 Valve Board Table 77Valve Board Components

Valve Board 180B01601 (only supplied for multistream option or for RS232 output on single stream)

Circuit board controls multistream accessory. Jumper should be installed on back of valve board. (Purple ejector)

Connector for cable connector from display board

Analog Board 180B02502 For multistream configurations: 180B02501*

Circuit board converts signals from analog to digital and processes signals from "fault" detec tors (Orange ejector)

7.3.5 Power Supply Board Table 78 Power Supply Board

Power Supply Board 180B01203 Supplies voltage to electrical circuitry (Blue ejector)

18pin Female Connector For relays

6pin Male Connector Operating power

6pin Female Connector Valve Pump Power

7.3.2 CPU Board Table 75CPU Board Components

0Vdc 5Vdc Connector To recording device.

4mA –20mA Connector To recording device.

Computer Memory Reset Clears microprocessor memory

Computer Memory Reset Clears microprocessor memory Valve Pump Connector Connector for cable from valve pump

Display Board Cable Connector

Connector for display board cable

Show 6 more labels / short text on this page
Component Function · Part Function · Component Function · Multistream Panel Connector · Part Function · Component Function
Page 44 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — FIGURE 72. CARD CAGE AND BOARDS
Page 44 · click to enlarge
Page 44

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

PLEASE NOTE: VALVEPUMP CONNECTOR IS ON CPU BOARD

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION diagram (page 44)
Fig. 34
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION diagram (page 44)
Fig. 35

FIGURE 72. CARD CAGE AND BOARDS

Page 45 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — 7 OFF
Page 45 · click to enlarge
Page 45

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

7.3.6. Analog Board DIP Switches The following list describes the conditions for each switch to be in the ON or OFF position:

Table 79. Analog Board DIP Switch Positions

Switch No. Position Condition Remarks

Analog Board has no adjustment

1 ON When using Analog Board, Part # 180B02602

ports for individual output device and/or 05 Vdc, 420 mA outputs.

Analog Board has individual ports

1 OFF When using Analog Board, Part # 180B02601

for adjusting each output device (up to 6) and/or 05 Vdc, 420 mA outputs.

2 ON Quick calibration on

2 OFF Quick calibration off

3 ON

4 ON Nonfunctioning switch

5 ON Nonfunctioning switch

6 ON Nonfunctioning switch

Allows markers to be recorded on

7 ON For stream ID and “brackets” before and after calibration

strip charts and other recording devices.

No markers will be recorded on strip charts and other recording devices.

7 OFF

8 ON For 80 column printers Formats printer output

8 OFF For 40 column printers Formats printer output

Emerson Process Management Gas Chromatograph (CFA-3000) — 7 OFF diagram (page 45)
Fig. 36
Emerson Process Management Gas Chromatograph (CFA-3000) — 7 OFF diagram (page 45)
Fig. 37

Figure 73. Analog Board DIP Switches

(with #3 set in OFF position for multistream analyzer) Colorimeter Assembly

Page 46 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — FIGURE 74. REAR OF MAIN CABINET
Page 46 · click to enlarge
Page 46

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

7.4 Rear of Main Cabinet

Table 710Back Cabinet Parts

In the rear of the main cabinet are a number of electrical components and connectors. The following diagram is for reference, only.

Backplane Board Circuit boards plug into front side

Power Transformer Supplies ac voltage to electronics

Heating Bath Connector Interconnect to heating bath Pump Connector Interconnect to valve pump

220Vdc Transformer Optional part for specific electrical requirements (208 V – 240 V)

FIGURE 74. REAR OF MAIN CABINET

Emerson Process Management Gas Chromatograph (CFA-3000) — FIGURE 74. REAR OF MAIN CABINET diagram (page 46)
Fig. 38
Emerson Process Management Gas Chromatograph (CFA-3000) — FIGURE 74. REAR OF MAIN CABINET diagram (page 46)
Fig. 39

Component Function

Page 47 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — Figure 75. Valve Pump Assembly
Page 47 · click to enlarge
Page 47

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

7.5 Valve Pump Assembly

The vacuum pump assembly includes the pump, PC board, and reaction cell.

Sample and Reagent Tubing

Colorcoded straws and labeled tubes for transport of solutions

Table 7 11Valve Pump Parts

“D” Connector Connects to small, 16pin connector on CPU boards

Pump Housing Contains and protects valve pump electronics

Vent Line Vents reaction cell to waste Pump Drip Tray Catches leaks and directs them to the leak detector

Pump Valve pump routes sample and reagents through the analyzer

Reagent Barb Fittings Male half of tubing connector. Tubes are connected to reagents and sample

LEDS Identify which injector is active.

Contains reaction cell, heating block

* Vacuum/Pressure LED is located on the upper right corner of the ValvePump as it is installed

6position Connector Electrical power to heating bath 9position Connector Electrical power to pump

LED Green is OK; RED is Low pressure

Figure 75. Valve Pump Assembly

Part Function · Part Function · Reaction Vessel Assembly

Page 48 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION
Page 48 · click to enlarge
Page 48

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

Table 713Relay Output Option

7.6 Interface Ports The interface ports (labeled for recorder, printer, relay output, and RS232) are located on the lower, left outside wall.

Pin No. Description

1 NO 9 C Stream #1 Alarm or ID

2 NO 10 C Stream #2 Alarm or ID

Table 712Interface Ports

3 NO 11 C Stream #3 Alarm or ID

4 NO 12 C Stream #4 Alarm or (#1*) ID

RS232C For connection of computer interface or other accessories

5 NO 13 C Stream #5 Alarm or (#2*) ID

6 NO 14 C Stream #6 Alarm or (#3*) ID

Relay Output Optional relay contact (multistream analyzer only)

If there are 3 streams or less, both stream ID and alarms may be optionally configured.

Printer For connection of printer cable Recorder For connection of external recorder; for multioutput, 420 mA and 05 Vdc outputs for streams 2 through 6.

7.8 MultiOutput Board The output for Stream #1 is obtained from the Input/Output Panel (see Figure 25). The additional five outputs are obtained from the lower lefthand side of the analyzer. Table 714 gives the pin assignments on the “D” connector for a multistream system:

\ 1 2 3 4 5 6 7 8 9 10 11 12 13 14 /

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION diagram (page 48)
Fig. 40
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION diagram (page 48)
Fig. 41

\15 16 17 18 19 20 21 22 23 24 25 /

Figure 77. "D" Connector for Relay Output

Table 714Multioutput Pin Assignments on

2 15 2 14 1

3 17 4 16 3

4 19 6 18 5

5 21 8 20 7

6 23 10 22 9

Figure 76. Interface Ports

7.7 Relay Output Option This port (see Figure 77) is a collection of contact closures that can be used to identify either the stream being displayed or the alarm that is activated (multistream systems only). these are not “powered” contacts.

Relay Contact Ratings: 100 Vdc max voltage 10 Vdc max DC current 1.0 A max carry current 0.5 A max switched current

Pin Assignments · Port Function · “D” Connector

Page 49 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION
Page 49 · click to enlarge
Page 49

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

7.9: Overflow Sampling Assemblies

Table 715Overflow Sampling Assembly Parts

Overflow Waste Vent Allows excess sample

to escape.

Intake/Regulating Valve Regulates flow of sample into overflow sampler assembly

Waste Pipe Carries waste to drain.

Overflow Tubing All overflow tubes feed into common waste pipe. Filters (180134108)

Overflow Tube Assures adequate supply

Replaceable filters are mounted down stream from each intake regulating valve. Sample Inlet Tubing 1/8" x 1/4" tubing (provided by customer).

of sample and reproducible results by eliminating positive pressure from sample stream.

Sample Intake Tubing Tubing which carries sample from overflow tube to analyzer (one to six stream configurations) NOTE: Sample intake tube should be no more than 1 to 11/2" into the sample.

Common Panel Mounts multistream Overflow Sampler Assembly to left side of

analyzer (facing front panel). Panel with dimensions 221/4” x 9" will support up to six streams.

Air Vent Allows air to escape

overflow tube as it fills with sample.

Figure 78. SingleStream Overflow Sampling Assembly

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION diagram (page 49)
Fig. 42
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION diagram (page 49)
Fig. 43

Part Function · Part Function

Page 50 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION
Page 50 · click to enlarge
Page 50

CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION diagram (page 50)
Fig. 44
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 7 INSTRUMENT DESCRIPTION diagram (page 50)
Fig. 45

Figure 79. Multistream Overflow Sampling Assembly

(threestream option)

Page 51 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES
Page 51 · click to enlarge
Page 51

CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES

SECTION 8 ADDITIONAL FEATURES

8.1 Chart Recorders and Data Loggers

A chart recorder or other data acquisition device may be connected to the analyzer for a permanent record of sam ple concentration. Setting the output 420 mA 05 Vdc from the keypad panel allows you to adjust 420 mA and 0 5 Vdc to a recording device.

These input devices must be isolated or have a floating ground (not reference to ground). The contacts for recorder output are located at the input/output panel on the side of the analyzer (see Figure 81).

The analyzer will operate normally with data output to the chart recorder or other device. During normal operation, calibration is performed biweekly. This may vary by Chemistry. To indicate the beginning and end of a calibration sequence, the recorder makes a quick trace down to zero, and up to full scale. This is called a calibration marker. After calibration, the recorder resumes recording the sample values (refer to Figure 82).

When a recorder is connected to a multistream system, a ”stream identification blip” (SIB) is produced prior to the output signal for that particular stream. This allows quick identification of individual stream readings on the recorder. SIB values are shown in Table 81, and tracings are illustrated in Figure 83.

Figure 81. Input/Output Panel

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES diagram (page 51)
Fig. 46
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES diagram (page 51)
Fig. 47

1 0.5 10

2 1.0 20

3 1.5 30

4 2.0 40

5 2.5 50

6 3.0 60

Grab 3.5 70

Table 81Stream Identification Blips
Show 7 more labels / short text on this page
Voltage · % Scale of · Stream # · Output · Recording · Marking · (SIB’s)
Page 52 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES
Page 52 · click to enlarge
Page 52

CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES diagram (page 52)
Fig. 48
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES diagram (page 52)
Fig. 49

Figure 82. Typical Trace Recording with

Calibration of Single Stream System

Page 53 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES
Page 53 · click to enlarge
Page 53

CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES diagram (page 53)
Fig. 50
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES diagram (page 53)
Fig. 51

Figure 83. Typical Stream Trace Recording with

Calibration of Multistream System

Page 54 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — EIA
Page 54 · click to enlarge
Page 54

CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES

8.2 Computer Interface The computer interface is an RS232C serial data interchange port operating at 1200 baud. The serial port receives the same information as the printer, and has the capability to transmit information, enabling remote control of the analyzer.

8.2.2 Data Configuration The computer must be set to match the fixed data configuration of the analyzer, which is:

Parameter. Setting

Data 8 bits, high order bit set to zero

8.2.1 Data Output Format The data output formats contain no control characters, with the exception of carriage return, line feed, and the preamble/post amble described below:

Table 82Normal Data Output Format

8.2.3 RS232C Function Commands The RS232C interface can be used as follows: • By direct hookup to a nonMacIntosh® based computer A program on disc can be obtained from

Preamble NUL, STX, NUL = (00, 02, 00 Hex notation) Date/Time ASCII string of 25

Rosemount Analytical for remote control. Cable length must less than 150 feet. • By direct hookup to a nonintelligent terminal. No

characters, followed by 7 spaces

program disc is required. Cable length must be less than 150 feet. • By communication through a modem. One modem located at the analyzer can communicate over a

Concentration ASCII string of 5

characters, followed by 9 spaces

Stream Number One ASCII character

telephone network to another modem hooked up to an IBM PC or compatible computer or a nonintelli gent terminal. Pin configuration is described in Table 84.

Mode An ASCII string,

consisting of 7 spaces, followed by three possible messages: NOT calibrated, Manual, or Automatic.

Table 84RS232C Pin Configuration

Pin No. Circuit

Direction of Signal Description

EIA

Calibration Status There are two ASCII messages transmitted: CAL when calibration is in progress, and blank when good data is available. Postamble: NUL, EOT, NUL = (00, 04, 00 Hex notation) Postamble NUL, EOT, NUL = (00,04,00 Hex notation)

1 AA 101 Protective Ground

*2 BA 103 From Terminal

*3 BB 104 To Terminal Received

*4 CA 105 From Terminal

*5 CB 106 To Terminal Clear to Send

Show 8 more labels / short text on this page
Baud Rate 1200 · Parity None · Stop Bits 1 · Data Format · Circuit CCITT · Transmitted Data · Data · Request to Send
Page 55 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — EIA
Page 55 · click to enlarge
Page 55
Emerson Process Management Gas Chromatograph (CFA-3000) — EIA diagram (page 55)
Fig. 52

8.2.4 Initiating RS232Communications to the Computer 1. With the computer and CFA3000 power turned off, connect RS232C cable to the computer and to the RS232C port on the CFA3000 analyzer. (If a printer is to be used, connect it to the printer port

on the analyzer). 2. Turn on power to the analyzer. 3. From the analyzer keypad panel, turn on the printer by pressing the following buttons: • From a multistream keypad:

  • From a singlestream keypad, referring to Figure 84, press position 5, then 1, and then 10. 4. Turn on computer
Page 56 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — DISVALV
Page 56 · click to enlarge
Page 56

CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES

5. If you purchased program from Rosemount Analytical, run it from your computer, by typing in

SETCLO Sets clock calendar, Date is entered as follows: day of week, year, month, date, hour, minute, second. Day of week is coded as Sunday = 0, Monday = 1, Tuesday = 2, etc. Year is entered as last two digits of year. Month and date must be two digits, use leading zeros as necessary. A 24 hour format is used for hour entry. For example, to set the clock at Thursday, April 12, 2003 at 4:35 p.m., enter SETCLO 4030412163500.

CFATERM, then press Return. 6. When a blinking cursor appears on the screen, press Return. The screen will display M>. 7. At this point, all of the RS232C functions listed in Table 84 are accessible. 8. To set up a file to collect data in the computer, press <altL> (formatted) or <altC> (capture). 9. Type in file name (up to 8 characters). 10. To start collecting data, press Return. The display will show a blinking cursor. 11. To access the command functions again, press Return. The screen will display M>.

Use the commands in Table 85 to control the analyzer.

Table 85RS232C Commands

SETHAL Changes stream high alarm setting. For

AUTOCAL Runs fullscale automatic calibration AUTOVALV Resumes automatic valve selection

example, to change stream 1 to 25, enter SETHAL 1,25.

DATA Forces printing of analyzer data

SETLAL Changes stream low alarm setting. For example, to change stream 2 to 50, enter SETLAL 2,50.

DISVALV

Displays stream parameters, including sample cycles, low and Function

SETVTIM Changes stream valve cycles. For example, to change stream 2 to 2 cycles, enter SETVTIM 2,20.

MANVALV To operate valves manually (for multi stream, one stream continuously). For example, to monitor stream 3, enter MANVALV 3.

RESET Resets the computer for a full wipeout

Command Function · Command

Page 57 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES
Page 57 · click to enlarge
Page 57

CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES

8.2.5 Printer Interface

8.2.6 Set Printer Internal Clock

The printer interface is a female parallel port designed to be compatible with IBM PCtype cables and printers. This builtin capability permits text printing only, no graphics.

Set printer's internal clock before printing; perform the following procedure:

1. From analyzer keypad panel, turn on printer: • From a multistream keypad, press:
Table 86Printer Port Pin Configuration

and , then . • From a singlestream keypad, referring to Figure 84, press position 15, then 1, and then 10

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES diagram (page 57)
Fig. 53
2. Enter data as follows: a. Day of the week first, one digit:

Sunday = 0, Monday =1 Press enter. b. Year, two digits: 2003 = 03, 2007= 07

Press enter. c. Month, two digits: Jan = 01,

Feb = 02 Press enter. d. Day of the month, two digits: 01, 02

Press enter. e. Hour, using 24hour format:

8:00 AM = 08, 1:00 PM = 13 Press enter. f. Minute, two digits: 01, 02

Press enter. ExampleTo set the clock for Friday, August 16, 2003

at 4:35 PM:

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL SECTION 8 ADDIDIONAL FEATURES diagram (page 57)
Fig. 54
Page 58 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL APPENDIX
Page 58 · click to enlarge
Page 58

CFA3000 MANUAL APPENDIX

APPENDIX A CHEMISTRIES

MethodSodium01A (0.110, 1100, 101000 ppb to 1000 ppm (contact Rosemount Analytical for other ranges)
1. The chemicals used for reagents and distilled water should be free of sodium.

General Description: For this procedure for the determination of sodium, the sodium ion is measured potentiometrically using ion selective electrodes . The addition of an ionic strength adjustor to the sample maintains constant background ionic strength and pH.

2. The sample stream must be filtered to 10 microns or less before being introduced to the system with

an inline filter such as Rosemount Analytical' IF 100 (Part #180140001).

Sodium in Water Range: 0.110, 1100, 101000 ppb up to 1000 ppm or higher.

Reagents Kit(reagent & standards) 185301130
REAGENT #1, Buffering Reagent185301134
Working Standards 10 ppb185301133 100 ppb: 182301132

Operating Notes

Page 59 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL APPENDIX
Page 59 · click to enlarge
Page 59

CFA3000 MANUAL APPENDIX

APPENDIX B SPECIFICATIONS

Accuracy: All twodecade ranges up to 1000 ppb maximum: ±10% of reading or ±2 ppb, whichever is greater

single stream, standard for multistream monitors. RS232C output is optional for single stream monitors, standard for multistream monitors.

0.110 ppm range = ±10% of reading or ±0.5 ppm, whichever is greater

Alarms:

  • General: contact closures for
  • leak detection,
  • calibration in progress, Contact closures are rated:2 A at 250 Vac or 3 A at 30 Vdc; delay on/off is 1 minute
Automatic Calibration40 minutes
Total Calibration40 minutes, twice/week
Reagent Consumption5 L of each reagent per 3 months

Streams 2 through 6 (multistream option):

  • One alarm relay for each stream; contacts have dc ratings
  • 100 Vdc (maximum carry only)
  • 0.5 A (maximum switched)
  • delay on/off time is 1 minute
Standard Consumption1 L of high standard,1 L of low standard per 3 months

Sample Requirements:

Flow4050 mL/min Current standard bottles are:

Display: Single line LCD, 4 1/2 digits, readout in concentration units. For multistream option display remains at last value until new stream reading is stable.

“0”(baseline)1.0 L. Fullscale: 5.0 L

Status LEDs:

Pressure<10 psig (<170 kPa) (overflow cup must drain to atmosphere)
  • Leak,
  • Lamp failure,
  • Quarterly maintenance due
  • Calibration in progress. For multistream analyzers.
Temperature050° C (32122° F)

Filter: 8 micron filter in sample line is Recommended; (#180134108).

StatusLEDs show:

  • the sample stream currently selected
  • the sample stream for which data is being displayed.
Multiple Stream OptionGeneral: up to 6 sample streams; measurement is sequential
Response Time15 minutes (methoddependent)

Environmental: The analyzer is not intended for out door use. Install in a clean, indoor environment.

Outputs420 mA or 05 Vdc (nonisolated). (Isolated output is optional.)

Ambient Temperature: 40120° F (1045° C) must be regulated to ± 10° within range, avoid temperature swings greater than ± 10° F (± 5± C) over a 12hour period

Load requirements for 420 mA output: 450 ohms minimum, 1200 ohms? maximum

Load requirements for 05 Vdc output: 10 k Ohms, minimum

Humidity595% (noncondensing) Pollution degree: Standard Altitude: Not applicable

The range of outputs is the same as the measure ment range. Parallel output for printer is optional for

Page 60 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL APPENDIX
Page 60 · click to enlarge
Page 60

CFA3000 MANUAL APPENDIX

APPENDIX B SPECIFICATIONS Cont.

DimensionsHeight of Analyzer Cabinet:

25.75 in. (654 mm) Height of Reagent Cabinet: 18.25 in. (464 mm) Width: 18 in. (457 mm) Depth: 16 in. (406 mm)

Space Requirements: Wall mounting: Allow 18 in. (457 mm) on one side for swing out access to rear panel. Allow 12 in. (305 mm) on the other side for access to input/output panel and overflow sampler. Panel Mounting: Allow 12 in. (305 mm) on both sides for access to input/output panel and overflow panel.

Weight: 117 lb (53 kg).

Add up to 60 lb (27.3 kg) for full weight of

reagents and standards. Shipping Weight of Analyzer: 138 lb (63 kg) Shipping Weight of Reagents: up to 64 lb

(29 kg) per kit (3 months supply)

Page 61 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL APPENDIX
Page 61 · click to enlarge
Page 61

CFA3000 MANUAL APPENDIX

APPENDIX C POWER INTERCONNECT DESCRIPTION

The block diagram in Figure C1 graphically shows the distribution of power and signals throughout the analyzer. The raised numbers (e.g., (6) ) in the description below relate to the circled numbers on the diagram. For clarity, minor components such as fuses are not shown.

3. From TB1, for both options, line power flows to the Colorimeter Lamp Power Supply, (4) , the Power Transformer (3) , the Pump Power Supply (5) and the Power Supply Board (8e) located in the Card Cage (8).
4. The Colorimeter Lamp Power Supply produces highly regulated DC current which energizes the Colorimeter Lamp (4a).
1. For 110 Vac 50/60 Hz option, line power enters the analyzer at the Power Entry barrier strip TB3 (1), with the following connections (refer to Figure 25 and Figure C1):
5. The Power Transformer converts standard line voltage to lowvoltage AC for use by the CFA3000 electronics. The Power Transformer secondaries feed the Power Supply Board, which converts the lowvolt age AC to regulated DC current, and to the heater element in the Heating Bath (7) . Relay contacts on the Power Supply Board are externally available at the I/O Panel (10) for activation of alarm and status indicators.
  • Line (hot) connection is TB31
  • Neutral is TB32
  • Safety (chassis) ground connection is TB34
  • For 220 Vac 50/60 Hz option, the following con nections are used:
  • L1 connection is TB37
  • L2 connection is TB38
  • Safety (chassis) ground connection is TB34

Multistream Analyzer Option In units equipped with the Multistream option, a Valve Board (8c) will be installed. This board contains relay contacts that are externally available at the Interface Ports (11) as well as LED drivers to activate the sample stream indicators located on the Keypad Panel (9) . The Valve Board also provides interfaces for a printer and serial data communications, exter nally available at the Interface Ports.

2. For 110 VAC 50/60 Hz option, line power flows through the Power Switch and other minor compo nents to barrier strip TB1. TB1 serves as the line power distribution point.

For 220 VAC 50/60 Hz option, line power flows through the Power Switch and other minor compo nents to the primary of the stepdown Transformer (2). The Stepdown transformer converts the input 220 VAC to 110 VAC for compatibility with the standard CFA3000 electronics. From the Stepdown Transformer secondary power flows to barrier strip TB1.

If the monitor includes a multioutput analog board, analog recorder drivers for each stream are installed on the Analog Board (8d) . These outputs are exter nally available at the Interface Ports.

Page 62 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL APPENDIX
Page 62 · click to enlarge
Page 62

CFA3000 MANUAL APPENDIX

APPENDIX D BYPASS FOR CALIBRATION

To avoid extraneous alarms from external devices, rewire output as shown below

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL APPENDIX diagram (page 62)
Fig. 55
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL APPENDIX diagram (page 62)
Fig. 56

Figure D1. Calibration Bypass Wiring

Page 63 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL APPENDIX
Page 63 · click to enlarge
Page 63

CFA3000 MANUAL APPENDIX

APPENDIX E KEYPAD HELP SHEETS

E.1. SingleStream Keypad "Help Sheet"

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL APPENDIX diagram (page 63)
Fig. 57
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL APPENDIX diagram (page 63)
Fig. 58

Figure E1. Keypad Panel for Single Stream Option

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL APPENDIX diagram (page 63)
Fig. 59
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL APPENDIX diagram (page 63)
Fig. 60

Figure E2. Number Positions on a SingleStream Keypad

Page 64 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — THAN HIGH ALARM)
Page 64 · click to enlarge
Page 64

CFA3000 MANUAL APPENDIX

Table E1: Singlestream Keypad Help Sheet

Press … Function To Exit, Press …

Initiates automatic calibration No action required

Initiates priming and auto calibration No action required

Sets scale to adjust output to 4mA

and 0 Vdc on external device

Referring to Figure E2, on keypad panel, press position 15 and then 2. Initiate valve pump test operation

Display high alarm setting

or Set high alarm value

Display low alarm setting

Change low alarm value (MUST BE SET LOWER

THAN HIGH ALARM)

or

Page 65 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL APPENDIX
Page 65 · click to enlarge
Page 65

CFA3000 MANUAL APPENDIX

Press … Function To Exit, Press …

Set reagent supply time

Display number of days remaining

until replacement of reagents

Set output scale for 20mA and

. Display mV

Display measurement of previous

Referring to Figure 84, press position 15, then 1, and then 10

Turn on printer No action required

Referring to Figure 84, press posi

tion 15, then 1, and then 10 Turn on printer No action required

Show 4 more labels / short text on this page
5 Vdc · Display Gain value · Initiate grab sample · grab sample
Page 66 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — Figure E3. Keypad Panel for Multistream Option
Page 66 · click to enlarge
Page 66

CFA3000 MANUAL APPENDIX

Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL APPENDIX diagram (page 66)
Fig. 61
Emerson Process Management Gas Chromatograph (CFA-3000) — CFA3000 MANUAL APPENDIX diagram (page 66)
Fig. 62

Figure E3. Keypad Panel for Multistream Option

Table E2. Multistream Keypad Help Sheet

Press … Function To Exit, Press …

Initiates automatic calibration No action required

Initiates priming and auto calibration No action required

Sets scale to adjust output to 4mA

and 0 Vdc on external device

Initiate valve pump test operation

Page 67 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — DESIRED, GO TO SET LOW
Page 67 · click to enlarge
Page 67

CFA3000 MANUAL APPENDIX

Press … Function To Exit, Press …

or Display a particular stream in multi

stream system, to set parameters

Then, or Set number of cycles for a particular

stream in multistream system

Set high alarm value for stream (IF ONLY A LOW ALARM VALUE IS

DESIRED, GO TO SET LOW

ALARM

Press … Function To Exit, Press …

Change low alarm value (MUST BE SET LOWER THAN HIGH ALARM SETTING)

Display low alarm setting

Display high alarm setting

Initiates multistream sampling No action required

Set reagent supply time (90 days)

or · or

Page 68 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL APPENDIX
Page 68 · click to enlarge
Page 68

CFA3000 MANUAL APPENDIX

Press … Function To Exit, Press …

Display number of days remaining

until replacement of reagents

Set output scale for 20mA and 5Vdc

Press … Function To Exit, Press …

Turn on printer No action required

Turn on printer No action required

Show 5 more labels / short text on this page
Display mV · Display Gain value · Initiate grab sample · Display measurement of · previous grab sample
Page 69 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL ORDERING
Page 69 · click to enlarge
Page 69

CFA3000 MANUAL ORDERING

ORDERING INFORMATION

Analyzer reagents are available from Rosemount Analytical on a contract basis. Refer to APPENDIX A, "Chemistries" for specific ordering information or contact Rosemount Analytical

Replacement Parts and Accessories

Table E3 Replacement Parts,

Part Part No.

Part Part No.

E Prom (NOTE: specify chem istry range with decimal places, part #'s of analog, and line frequency, 50/60 Hz, number of streams)

Valve Pump Assembly 185A00101

Solenoid Valve, multistream 180B30101

Waste Tube Assembly 180134601

Overflow Sampler Assembly:

Sample Line Tubing (.0625" ID) 180285201

1 Stream Analyzer 180130001

Filter, disposable (for overflow panel assembly)

2 Stream Analyzer 180130002

3 Stream Analyzer 180130003

4 Stream Analyzer 180130004

5 Stream Analyzer 180130005

6 Stream Analyzer 180130006

IF100 InLine Filter 180140001

Fuse, T 1 A/250 Vdc 625000302

Clamp, singlestream overflow 180132701

Clamp, multistream overflow 180133701

Power Supply Board 180B01203

Multi Output Board 180B02601

Keypad & Display Board Assembly, single stream 185B00301

Keypad & Display Board Assembly, multistream 185B00401

Show 7 more labels / short text on this page
Reagents · Mounting Brackets 180200401 · 180134108 · Clamp, drain 180202901 · CPU Board 180B02801 · Analog Board 180B02602 · Valve Board 180B01601
Page 70 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — RETURN OF MATERIAL
Page 70 · click to enlarge
Page 70

WARRANTY

Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes, membranes, liquid junctions, electrolyte, orings, catalytic beads, etc., and Services are warranted for a period of 90 days from the date of shipment or provision.

Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.

If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por tion of the Goods/Services.

All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.

Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.

RETURN OF MATERIAL

Material returned for repair, whether in or out of warranty, should be shipped prepaid to:

Liquid Division 2400 Barranca Parkway Irvine, CA 92606

The shipping container should be marked:

The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):

1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or nonwarranty service is requested.
5. Complete shipping instructions for return of the material.

Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.

If the material is returned for outofwarranty repairs, a purchase order for repairs should be enclosed.

Emerson Process Management · Return for Repair · Model

Page 71 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — IMPORTANT
Page 71 · click to enlarge
Page 71

CFA3000 MANUAL SECTION 9 RETURN OF MATERIALS

SECTION 9.0 RETURN OF MATERIAL

9.1 GENERAL.

To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for repair, call 19497578500 for a Return Materials Authorization (RMA) number.

9.2 WARRANTY REPAIR.

The following is the procedure for returning instruments still under warranty:

1. Call Rosemount Analytical for authorization.
2. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or subassemblies, the serial number on the unit must be supplied.
3. Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the materials in the same manner as they were received.
4. Send the package prepaid to:

IMPORTANT

Rosemount Analytical Liquid Division 2400 Barranca Parkway Irvine, CA 92606

Please see second section of “Return of Materials Request” form. Compliance with the OSHA requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required.

AttnFactory Repair

RMA No

Mark the packageReturned for Repair

Model No

9.3 NONWARRANTY REPAIR.

The following is the procedure for returning for repair instruments that are no longer under warranty:

1. Call Rosemount Analytical for authorization.
2. Supply the purchase order number, and make sure to provide the name and telephone number of the individ ual to be contacted should additional information be needed.
3. Do Steps 3 and 4 of Section 17.2.
Page 72 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 9 RETURN OF MATERIALS
Page 72 · click to enlarge
Page 72
Page 72 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — CFA3000 MANUAL SECTION 9 RETURN OF MATERIALS
Page 73 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — VISIT OUR WEBSITE AT www.rosemountanalytical.com
Page 73 · click to enlarge
Page 73
Emerson Process Management Gas Chromatograph (CFA-3000) — Immediate, Reliable Analytical Support diagram (page 73)
Fig. 63 · Immediate, Reliable Analytical Support

The right people, the right answers, right now.

Immediate, Reliable Analytical Support

Now there’s a way to quickly get the right answers for your liquid analytical instrumentation ques tions: the Analytical Customer Support Center.

Our staff of trained professionals is ready to provide the information you need. If you are placing an order, verifying delivery, requesting application information, or just want to contact a Rosemount Analytical representative, a call to the Customer Support Center will provide you with the right peo ple, the right answers, right now.

A Worldwide Network of Sales and Service

Emerson Process Management’s field sales offices are your source for more information on the fill line of Rosemount Analytical products. Field sales personnel will work closely with you to supply technical data and application information.

For more information, please contact your nearest Emerson Process Management sales office.

GERMANY

THE AMERICAS HEADQUARTERS

Emerson Process Management Process Gas Analyzer Center of Excellence GmbH & Co. OHG Industriestrasse 1 63594 Hasselroth Germany T 49.6055.884.0 F 49.6055.884.20

Emerson Process Management Rosemount Analytical Inc. Liquid Center of Excellence 2400 Barranca Parkway Irvine, CA 92606 Phone: +1.949.757.8500 Toll Free: +1.800.854.8257 Fax: +1.949.474.7250

ASIAPACIFIC

LATIN AMERICA

Emerson Process Management Asia Pacific Private Ltd. 1 Pandan Crescent Singapore 0512 Republic of Singapore Phone: 65.777.8211 Fax: 65.777.0947

Emerson Process Management Rosemount Analytical 10241 West Little York, Suite #200 Houston, TX 77040 USA T 713.467.6000 F 713.827.3328

MIDDLE EAST AND AFRICA

EUROPE

Emerson Process Management EPM Building P. O. Box 17033 Jebe Ali Free Zone Dubai, United Arab Emirates T 971.4.8835235 F 971.4.8835312

Emerson Process Management AG Blegistrasse 21 CH6341 BaarWalterswil Switzerland T 41.41.768.6111 T 41.41.761.8740

VISIT OUR WEBSITE AT www.rosemountanalytical.com

Page 74 of the Emerson Process Management Gas Chromatograph (CFA-3000) user manual — ONLINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Page 74 · click to enlarge
Page 74
Emerson Process Management Gas Chromatograph (CFA-3000) — Immediate, Reliable Analytical Support diagram (page 74)
Fig. 64 · Immediate, Reliable Analytical Support

The right people, the right answers, right now.

ONLINE ORDERING NOW AVAILABLE ON OUR WEB SITE

http://www.rosemountanalytical.com

Credit Cards for U.S. Purchases Only.

2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 7578500 Fax: (949) 4747250

http://www.rosemountanalytical.com

© Rosemount Analytical Inc. 2013

Emerson Process Management

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
No powerPower switch OFFPress power switch ON p.34
No powerPower switch failedCall Rosemount Analytical p.34
Pump is inoperativeValve pump switch OFFPress switch ON p.34
Pump is inoperativeBlown fuseReplace fuse p.34
Pump is inoperativeLeak detector activatedFind and fix leaking; check pump’s electrical connection p.34
Pump is inoperativeBad electrical connectionCheck pump’s plugs to CPU & connector p.34
Pump is inoperativeStill inoperativeCall Rosemount Analytical p.34
Keypad lights or display fails to light up or buttons fail to operateCircuit boards are not seated properlyRemove and reinstall circuit boards p.34
Keypad lights or display fails to light up or buttons fail to operateDisplay or associated power circuit faultyCall Rosemount Analytical p.34
Keypad lights or display fails to light up or buttons fail to operateConnector faulty or not properly insertedCheck connection p.34
Alarms not activatedIncorrect settingsReadjust settings, making sure that low alarm is set lower than high alarm p.35
Alarms not activatedInsufficient time for alarm to respondAlarm condition must exist for minimum of 60 seconds before alarm activates p.35
Signal not repetitive (same sample not repetitive)Small leak or clogging in fluidicsCheck all tubing lines and connections p.35
Signal not repetitive (same sample not repetitive)Clogged tubingReplace tubing p.35
No signal, or signal “off­scale” highConnector or wires loose Faulty circuit board Incorrect mV valueCheck circuit boards for proper seating Replace power supply circuit board p.35
Loss of mV Sensitivity LED ISE onReagents, standards old, contaminated or improperly preparedChange reagents p.35
Loss of mV Sensitivity LED ISE onReagent straws in wrong reagent or standardCheck placement of straws p.35
Loss of sample flowReplace sample filter on overflow panel p.35
Decreased pumping volumeTubes pinched shut or clogged from depositsRemove condition causing “pinched” lines; replace sample filter at overflow panel p.35
Leak detector LED lights upLeak in fluidicsRepair leak and dry leak detector contacts; turn main power OFF for 20 seconds p.35
Leak detector LED lights upHigh humidityAdjust screw on leak detector potentiometer on analog board; turn clockwise to decrease sensitivity p.35
Leak detector LED lights upConnector faulty or not properly insertedeCheck connection to analog board; replace analog board. p.35
Front panel displayTest LED’s and LCD on keypad: a. Press Power button and reset button on the CPU board (yel­ low ejector) b. Release power, then Reset button. Press powerPress: p.35

Ask about the Emerson Process Management Gas Chromatograph (CFA-3000)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Process Management Gas Chromatograph (CFA-3000)

Be the first to ask — other owners or our team will answer.