Instruction Manual PN 51CFA3000ISE/rev.F December 2013



WARNINGS
RISK OF ELECTRICAL SHOCK
Your purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in oper ating and maintaining the instrument. To ensure their continued operation to the design specifications, per sonnel should read this manual thoroughly before proceeding with installation, commissioning, opera tion, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.
Equipment protected throughout by double insulation.
This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation, or operation, may increase such interfer ence. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class A comput ing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to provide reasonable protection against such interference. Operation of this equipment in a residential area may cause interference, in which case the user at his own expense, will be required to take whatever measures may be required to correct the interference.
This product is not intended for use in the light industrial, residential or commercial environments per the instrument’s certification to EN500812.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 7578500 Fax: (949) 4747250

http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2012
Emerson Process Management






CFA3000 ISE Analyzer TABLE OF CONTENTS Instruction Manual
LIST OF TABLES
Table 51 Troubleshooting Chart 28
Table 52 Test Functions 29
Table 71 Main Cabinet Parts 34
Table 72 Reagent Cabinet Parts 34
Table 73 Buttons/LED Functions 36
Table 74 Card Cage Parts 37
Table 75 CPU Board Components 37
Table 76 Analog Board Components 37
Table 77 Valve Board Components 37
Table 78 Power Supply Board Components 37
Table 79 Analog Board DIP Switch Positions 39
Table 710 Back Cabinet Parts 40
Table 711 Valve Pump Parts 41
Table 712 Interface Ports 42
Table 713 Relay Output Option Pin Assignments 42
Table 714 Multioutput Pin Assignments on “D” Connector 42
Table 715 Overflow Sampling Assembly Parts 43
Table 81 Stream Identification Blips (SIBs) 45
Table 82 Normal Data Output Format 48
Table 83 Data Configuration 48
Table 84 RS232C Pin Configuration 48
Table 85 RS232C Commands 50
Table 86 Printer Port Pin Configuration 51
Table E1 Singlestream Keypad Help Sheet 5859
Table E2 Multistream Keypad Help Sheet 6062
Table E3 Replacement Parts 63

SECTION 1 INTRODUCTION
The Rosemount Analytical CFA3000 Series analyzers are designed for minimum maintenance and maximum reliability. The touch of a button starts the microprocessorcontrolled program, calibrates the analyzer, keeps it calibrated, and reports results directly unattended for months at a time. Maintenance is simplified by a modular design with special fittings, labeled tubing, and colorcoded components.
The CFA3000 Series includes analyzers for parameters commonly measured using Ion Selective Electrode (ISE) technology including sodium, fluoride and chloride. Sample concentration is clearly indicated on the digital display panel, and output signals are provided to drive recorders, alarms, or other external devices such as printers or computers. The specific chemistry of your individual analyzer is discussed in Appendix A, Chemistries, or in an accompanying method sheet..
Your purchase from Rosemount Analytical provides you with one of the finest instruments available for your particular application. These instruments have been designed and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowl edge of the user in operating and maintaining the instrument.

CFA3000 ISE Analyzer Instruction Manual
This instruction manual describes installation, operation, and maintenance of the CFA3000 ISE Analyzer. To ensure continued operation to the design specifications, personnel should read this manual thoroughly before proceeding with installation, operation, or maintenance of the analyzer.
as well as void the warranty. • All protective covers must be in place and cabinet panels must be closed at all times after installation or
maintenance.
Rosemount Analytical listened to you and decided to move away from a peristaltic pump assembly requiring monthly replacement of the pump tubing. The new pump assembly has a vacuum/pressure valve system which eliminates pump tubing and used significantly less reagent. Since the new valve pump swiftly moves fluids through the analyzer, the potential for clogging within any internal component is reduced. Significantly. As a result, mainte nance is required less frequently and procedures such as cleaning flow cells and the mixing system are no longer required.
The upgraded software now identifies a fixed standard that allows one to change the range of the 05 V and the 420 mA output signals to your requirements.

CFA3000 ISE Analyzer Instruction Manual
Installation of cable connections and servicing of this product require access to shock hazard voltage levels. Main power and relay contacts wired to separate power sources must be disconnected before servicing. Nonmetallic cable strain relief does not provide grounding between conduit connections! Use grounding type bushings and jumper wires. Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA 70) and/or any
other applicable national or local codes. Operate only with: 1. Main console rear cover fastened. 2. Left side power terminal cover closed. 3. Keypad assembly latched closed. Safety and performance require that this instrument be connected and properly grounded through a three wire power source. Proper configuration and use is the responsibility of the user.

CFA3000 ISE Analyzer Instruction Manual
2.1 UNPACK PARTS
When you receive your new CFA3000 analyzer, carefully unpack the items in the following list.
√ Overflow sampler assembly
√ Waste tube assembly
√ Valve pump assembly (integrated housing includes pump, labeled tubes with straws)
If any of the items are missing or damaged, contact Rosemount Analytical at: our:
Inspect each piece to confirm that there is no damage or loose parts. If anything is missing or appears broken or damaged, contact Rosemount Analytical at the tollfree phone number above.

CFA3000 ISE Analyzer Instruction Manual
After analyzer has been unpacked and all parts accounted for, install as described in following sections. Refer to Figure 21 to view a completely installed analyzer.


Figure 21. Opened Assembled Cabinet

CFA3000 ISE Analyzer Instruction Manual
2.2 Mount Main Cabinet
To mount main cabinet, proceed as follows (see Figure 22):
2.3 Mount Reagent Cabinet
To mount reagent cabinet, proceed as follows:
1. Loosen the four bolts in bottom of main cabinet. 2. Lift reagent cabinet and attach to main cabinet by placing heads of bolts into keyhole slots in top of reagent cabinet. 3. Slide reagent cabinet back. 4. Feed grounding strap on bottom of main cabinet through notched hole in left rear corner and attach it to stud on back of reagent cabinet (see Figure 23). 5. Tighten screws to attach reagent cabinet to main cabinet. 6. For multistream units, plug valve bank cable into positive fit receptacle at left side of cabinet (refer to Figure 71, “Multistream Valve Panel Connector”).
General Recommendations

CFA3000 ISE Analyzer Instruction Manual


Figure 22. Mounting Main and


Figure 23. Attaching
Reagent Cabinets · Grounding Strap

CFA3000 ISE Analyzer Instruction Manual
2.4 Install Waste Tube
To install waste tube, proceed as follows (refer to Figure 21):
2.5 Mount Overflow Sampling Assembly
The Sample intake tube should extend into
The overflow tube approximately 1½” – 2”
3.81 –5.08 cm).
To mount overflow sampling assembly proceed as follows (for sample configurations, see Figures 24a and 24b):
1. Mount overflow sampler assembly on outer, right side of reagent cabinet using the provided four machine screws. For multistream units, mount overflow assembly on left side of cabinet. (If using IF100 InLine Filter, refer to instruction sheet accompanying filter for complete installation instructions.) 2. Place ends of air vent tubing and overflow waste tubing into waste pipe. 3. Connect waste pipe to drain. 4. Supply sample stream to regulating or intake valve with 1/4” tubing (plastic or steel) as follows: Insert tubing into tube fitting. Make sure tubing rests firmly on shoulder of fitting and nut is fingertight. Mark nut at 6 o’clock position. While holding fitting body steady with a wrench, tighten nut 11/4 turns (go to 9 o’clock position). 5. For multistream assembly, connect sample intake line to bulkhead fitting. 6. For singlestream assembly, attach sample intake line to valve pump. 7. Make sure that sample liquid overflows into overflow chamber.

CFA3000 ISE Analyzer Instruction Manual


Figure 24a. Singlestream Overflow Sampling Assembly


Figure 24b. Multi
stream Overflow Sampling Assembly, threestream option)

CFA3000 ISE Analyzer Instruction Manual
made in accordance with local or national electrical code as applicable. This equipment must be grounded. A qualified electrician should wire this equipment to an electrical circuit.
Electrical connections are made to and from the cabinet at the input/output panel, located on the left side of the analyzer. See Figure 25.
To make electrical connections, proceed as follows:
Use cable glands to hold cables securely in holes above input/output panel.

CFA3000 ISE Analyzer Instruction Manual


Figure 25. Input/Output Panel
2.7 Install Valve Pump Assembly
The pump assembly has a vacuum pressure system that:
l Eliminates internal pump tubing
l Uses significantly less reagent
l Accurately moves fluids through the analyzer.
Perform the following procedure to install valve pump assembly:
1. Tilt pump assembly towards you. 2. Align the hinge pins and sockets, and slide assembly tray as far as possible to right (see Figure 27). Carefully lay tray down into cabinet 3. Plug in the two connectors to the back of the cabinet (6position connector for heating bath; 9position connector for pump) (see Figure 28). 4. Connect 16pin connector from pump assembly to small connector on CPU board (yellow ejector) above connector for keypad panel. 5. Attach colorimeter inlet tubing on left side of pump to flowcell inlet (see Figure 21). 6. Continue with procedure depending on whether system is singlestream or multistream.
For singlestream, continue:
For multistream, continue:
Guide all other tubes that are attached to valve pump assembly through reagent cabinet access hole.

CFA3000 ISE Analyzer Instruction Manual


Figure 26. Valve Pump Assembly


Figure 27. Pump Connector

CFA3000 ISE Analyzer Instruction Manual
Prepare reagent containers as follows: 1. Remove screw caps from reagent bottles. 2. Using an awl, make one hole in foil for the reagent straw. Do not allow awl to touch reagent or standards. 3. Place Standard and Reagent bottles into reagent cabinet, as shown in Figure 28. 4. Insert colorcoded reagent straws and labeled Standard straw leading from valve pump assembly into holes punched in corresponding bottles.

Figure 28. Sample Bottle Placement in Cabinet (for all methods; colors refer to reagent straw colors)

CFA3000 ISE Analyzer Instruction Manual
3.1 Operational Overview The analyzer uses a batch flow method for automatically analyzing a sample against a fixed standard (refer to block diagram in Figure 31). That is, the touch of a button starts the microprocessorcontrolled program, calibrates the analyzer, takes measurements of samples, and reports results directly unattendedfor months at a time.
The automated monitoring program determines the sequence for activating each of the valve pumps in the valve pump assembly to inject air and fluids into the reaction chamber. Each valve pump is associated with a specific input via the quick disconnect connector on the side of the assembly (the number of reagents is determined by your particular chemistry and procedure).For example, the inputs for sodium chemistry would be:
For a multistream system, solenoid valves in the analyzer reagent cabinet open and close streams for sampling. When sampling, the keypad panel LEDs identify which sample stream (valve number) is associated with the dis played value, and which valve is now open for sampling. Although the system runs samples on its own, you may also perform a manual grab sampling or a manual calibration at any time.
Figure 31. Simple Block Diagram

CFA3000 ISE Analyzer Instruction Manual
Using a MultiStream Overflow Sampler
Multistream analysis requires an overflow sampler assembly that will accommodate the extra sample streams. One multistream chamber allows for up to three sample streams; for 4 6 streams, use two chambers.
The sample tubes connect to solenoid valves in the reagent cabinet. The analyzer uses the valves to cycle through all sample lines, in sequence. First, stream 1 valve is opened, and Valve 1 LED on the keypad panel is lit. When the measurement is complete, Valve 1 will be closed, and Valve 2 is opened to begin sampling stream 2. The keypad panel will show Valve 1 LED lit as the Displayed Stream, and Valve 2 LED is lit for the Selected Stream.
LED's on the valve pump assembly light up when a particular fluid is taken up by the valve pump assembly. Figure 32 identifies the valve pump LED designations.
3.2.1 Check Fluidics
Check the following:
3.2.2 Check Reagents
Check the following:
3.2.3 Turn On Power
1. Open the keypad panel door to expose the circuit boards. 2. Press each board in to ensure proper seating. 3. Simultaneously press Power button on the back wall of upper cabinet and Reset button on the CPU board (yellow ejector), then release. 4. First release the Power button, then the Reset button. The Keypad should display ”H E L O", then four dashes " ". 5. Pull pump forward and check leak detector by dipping finger in tap water and touching detector behind pump. Leak detector LED on keypad panel will light up. 6. Turn main power switch to OFF. 7. Dry leak detector 8. Wait 20 seconds, then switch main power switch back to ON. Leak detector LED should be off.

CFA3000 ISE Analyzer Instruction Manual
Reaction cell takes approx. 30 minutes to heat up.
If leak detector LED does not light up:
Adjust potentiometer on analog board counterclockwise for more sensitivity.
Turn off power to reset system. Wait 20 seconds, and simultaneously press Power and Reset buttons. Dip finger in tap water and touch leak detector contacts on bottom of cabinet behind the valve
pump. LED should light up. If problem persists, contact Rosemount Analytical.
3.2.4 Test Valve Pump Operation
The test procedure allows you to confirm that all physical connections are installed properly, and to ensure that the pump is taking up the proper fluids at the correct time.
For singlestream, continue:
For multistream, continue:
Referring to Figure 32, on keypad panel, press posi tion 15 and then 2
MANUAL

The keypad buttons may now be used to test pump operation. Do not rush through the test steps – when a specific step is initiated, it takes approx. 2 seconds to pump air through the system. Actual pumping of a liquid takes approximately 4 seconds. After this operation
is finished, another 4 seconds elapses before the pump is primed for the next test step.
To exit the test at any time, press the ENTER button on the keypad panel.
If a test step does not work properly, try the following:
√ Take a syringe and shoot air through straw to make sure it is not clogged.
√ Check that corresponding LED on valve pumps lights up (refer to Figure 41).
√ Check that fluid is picked up in correct tube. If the problem still persists, contact Rosemount Analytical.


Begin Test:
Confirm that the corresponding Pump LED lights up for each step (refer to Figure 33 for LED designations during test):



valvepump. When the LCD number disappears, the next position may be pressed. The sequence is

3.2.5 Set Up Stream Parameters
The analyzer is programmed to measure a singlestream system at 4 sample cycles per hour. The multistream default is 1 cycle per stream, but you may define parameters for individual streams, such as modifying the num ber of cycles or the alarm threshold values.
The alarm values are used to activate alarms or controllers when the sample is outside of the specified concen tration range. A high alarm will activate when the sample concentration exceeds a preset limit. A low alarm will activate when the sample concentration is below the preset limit.
The contacts for the alarm outputs are located at the input/output panel on the cabinet side. The system must be in an alarm condition for at least one minute to activate the alarm contact.
Use the logical flows below to define desired stream parameters.


Figure 34. Logic Flow for Singlestream Parameter Definition



Figure 35. Logic Flow for Multistream Parameter Definition

3.3 Set Reagent Supply Time
Use the keypad panel to remind you when you need to replace the reagent. The analyzer will count pump cycles and count down the number of days left. To specify the duration, proceed as follows:
To determine how many days are left until you have to replace reagent,
Press: once.
Look at the LCD and record the number of days;
then press: again to get out of mode.
Set output 420 mA 05 Vdc from the keypad panel, (refer to Figures 72 and 73), then adjust 420 mA and 05 Vdc at your external device.
The sample Display for 4 mA and for 0 Vdc: 0 0 0 4

To display the fullscale concentration embedded in the EProm:
Press two times. To exit, press .
3.4.1 Record mV Use the mV value to determine that the method’s span and slope are correct. The change in millivolts (mV) is the spread between a baseline and a fullscale standard at the time of calibration. The values are relative to the chemical parameter being measured
Record the millivolt values for both baseline and fullscale standard.
The sample Display for value of 150 mV is: 0 1 5 0
3.4.2 Initiate Auto Calibration
The analyzer is calibrated automatically usually every two days, depending on the chemistry. Auto calibration performs both baseline and fullscale calibration. Baseline calibration measures a solution with 10% concentra tion of the species of interest. Fullscale calibration measures a standard solution in the upper range of the parameter of interest.
Before starting calibration, make sure that the analyzer is running reaqgents and baseline
solution for at least 20 minutes.

To initiate auto calibration, press on the keypad panel – the CALIBRATION LED will light. When multi stream systems are being calibrated, only Stream 1 LED is lit.
Baseline calibration will run first for approximately 15 minutes. Fullscale calibration follows, lasting approximately 15 minutes (the actual time for calibration depends on the chemistry). When fullscale calibration is finished, CALIBRATION LED goes out and the value is displayed on the LCD. The “ ” display signifies that a delay is required before the next sample is displayed.
Sample analysis automatically begins after calibration. The time delay from the start of sample analysis to dis play of results is approximately 15 minutes. This allows sufficient time for the previous solution to wash through system, for sample and reagents to react, for analyzer to measure sample and display concentration value.
When the CALIBRATION LED and the ”000” display in a singlestream system go out, the displayed value on the keypad panel is the actual stream value.
For a multistream system, after auto calibration, Sample Stream status LED’s will light to show sample stream valve currently selected and sample stream concentration currently displayed. For example, for the first stream sample, VALVE SELECTED LED is lit for that stream; dashes display on the keypad panel. When analyzer finish es sampling one stream and moves on to the next, the display shows the previous sample stream concentration value, and its VALVE DISPLAYED LED lights up. VALVE SELECTED LED will be lit for next valve in sequence.
then, , then to abort the calibration. Press twice in rapid succession on initial startup for priming of sample & reagents
3.5 Process a Grab Sample The analyzer allows you to rapidly analyze a grab sample.
Press on the keypad panel, to introduce a grab sample into the analyzer from the holder on the overflow sample panel.
If visible, suspended solids are present in the sample, filter it to 8 microns or less.

Place the clear grab sample inlet tube into a beaker containing the grab sample.
After the display of the grab sample’s value, a multistream analyzer begins processing the next stream in sequence.
3.6 Manually Sample a Particular Stream.
Use this procedure to manually switch between sample streams in a multistream system. This operation can be performed at any time, except during calibration.
Press one time. Sample Display: 0 0 0 1 .
Press and to change stream number. Sample Display: 0 0 0 3
Press to start sampling desired stream.
Press to return to multistream automatic analysis.
To shut down the CFA3000 for only a few days, press main Power button in back of upper cabinet to OFF position. Leave reagent straws in reagent containers.

Some chemistries may require specific wash solutions. Check Appendix A,
"Chemistries," for details.
To shut down the analyzer for more than a day, perform the following procedure
If the analyzer has been shut down during normal operations, turn power on and
press once.
If the CFA3000 has been turned off for an extended shutdown:
Press twice to prime; the system will automatically go into :”Auto Cal” after priming

SECTION 4 MAINTAINING THE ANALYZER
Your CFA3000 Series analyzer is a precision instrument which, if maintained properly, should provide years of accurate and reliable service.
4.1 Analyzer Equipment.
Even though the analyzer normally runs automatically, perform the following activities, quarterly:

SECTION 5 TROUBLESHOOTING PROCEDURES
This chapter is divided into two sections:
5.1 Troubleshooting Chart Symptoms observed during operation are generally enough to identify a faulty component.Such symptoms, together with probable causesand remedies are listed in the following troubleshooting chart.
Problem Probable Cause Remedy
Problem Probable Cause Remedy
No power Power switch
Display or associated power circuit
light up or buttons fail to
switch OFF Press switch ON
Blown fuse Replace fuse
Find and fix leaking; check pump’s electrical


Check pump’s plugs to CPU &
Still inoperative Call Rosemount
Circuit boards are not seated
Remove and reinstall circuit
light up or buttons fail to
boards Display or associated power circuit

Problem Probable Cause Remedy
Problem Probable Cause Remedy
Replace sample filter on overflow
Tubes pinched shut or clogged
lower than high alarm
causing “pinched” lines; replace sample filter at overflow
must exist for minimum of 60 seconds before
Leak detector LED lights up
Leak in fluidics Repair leak and dry leak detector
OFF for 20 seconds
Clogged tubing Replace tubing
High humidity Adjust screw on
No signal, or signal “offscale” high
boards for proper seating
on analog board; turn clockwise to
Check connection to analog board; replace analog
Reagents, standards old, contaminated or
Sensitivity LED ISE on
board.
in wrong reagent or
Problem Probable Cause Remedy
Test LED’s and LCD on keypad: a. Press Power button and reset
Press:
button on the CPU board (yel
low ejector) b. Release power, then Reset button.

Press:
Press:
button.
c. Press the following keys in sequence: (key #’s 15, 5, 6)
Press:

SECTION 6 REPAIR PROCEDURES
The Ionselective electrode analyzers are designed with modular components for quick and easy replacement. The following sections contain instructions to remove parts and replace them. For more complex repairs, or diag nostic assistance, contact Rosemount Analytical at:
6.1 Replace Valve Pump Assembly Use the following procedure to replace pump assembly:
1. Turn off power to analyzer. 2. Pull back of pump assembly towards you to a 45degree angle (see Figure 61). 3. Unplug the two connectors from the back of the cabinet. (9position connector for pump). 4. Disconnect pump assembly’s “D” connector from right side of pump. 5. Remove all liquid tubes from pump. 6. Disconnect waste tube from bottom of reaction cell (right side of valve pump). 7. Slide assembly to the left to disengage from pins, and remove pump assembly. 8. To install a new pump assembly, refer to Section, " Install Valve Pump Assembly".
Figure 61. Remove Valve Pump Assembly

6.2 Replace Circuit Boards
Use the following procedure to replace circuit boards:
1. Turn off power. 2. Open latch from right side of keypad panel; and open door. 3. Disconnect all cables and plugs from board to be replaced. 4. Unlatch colored clip from board to be replaced (refer to Figure 62). 5. Pull circuit board straight out. 6. Reverse steps to replace board and reconnect cables. Press firmly to ensure that boards are properly seated. 7. Turn power on. 8. Perform Automatic Calibration (refer to Section, “Initiate Auto Calibration”). Please NOTE: Valvepump connector is on CPU board


Figure 62. Card Cage

6.3 Replace Fuses
Use the following procedure to replace fuses (all fuses are 1A):


Figure 63. ISE Reaction Cell Assembly, Electrodes
6.4.1 The sodium electrode used in the CFA3011 sodium analyzer is a proprietary “solid state” electrode which is joined with a reference electrode in a sealed cartridge. The buffered sample stream flows into the reaction cell—through the ground stud to the sodium electrode, then into the reference electrode, out through the ground stud and then to waste. The sample concentration is measured in the sodium electrode, compared to the reference value and the sodium concentration is determined.

SECTION 7 INSTRUMENT DESCRIPTION
7.1 Analyzer Front View (Refer to Fig. 71).
7.1.1 Main Cabinet
Main Cabinet Splashproof enclosure protects analyzer
Chemistry Module Tubing Carries sample and
Reagent Cabinet Splashproof enclosure protects reagents and solenoid valves
reagent to valve pump assembly
Door Hinges Pin hinges mount doors on main and reagent cabinets Door Latch Secures cabinet door
Quickdisconnects Connects reagents, samples, and standards to valve pump assembly
Waste Tube Carries waste solutions to drain
Keypad Panel and Driver Board
External side contains touch buttons for controlling analyzer, digital display, and status panels. Internal side contains driver board
Leak Detector Senses leakage of solutions in main cabinet, shuts pump off, and activates leak LED
HB Lamp Indicates when propor tional controller is heating the heat bath
7.1.2 Reagent Cabinet
HB T 1 A 1 A fuse for heating bath
Table 72. Reagent Cabinet Parts
LAMP PS T 1.0 A 1 A fuse for power supply to ise lamp
MultiStream Valve Panel (MultiStream option only)
Allows sequential analysis of up to six sample streams
PUMP T 1.0 A 1 A fuse for pump
XFMR T 1.0 A 1 A fuse for transformer
MultiStream Connector (MultiStream option only)
Positive fit connector connects Multistream valve panel to main cabi net electronics
Pump On/Off Lighted switch controls power to pump
On/Off Main power switch
Bulkhead Fitting Fitting that directs individual sample stream to sample valve
Waste Drain Pan Drain pan directs waste solutions to waste tube
Pump Connector 9position connector connects pump to main cabinet electronics
Supplies continuous sample streams for analyzer
Heating Bath Connector 6position connector connects to heating bath assembly
Valve Pump Assembly Includes pump, PC board, labeled tubes, and reaction vessel assembly (chemistry module with integrated

Figure 71. Analyzer Front View

7.2 Keypad Panel
The keypad panel has a fourdigit liquid crystal display with an adjustable decimal point. The panel also has indi cator LED’s and pressuresensitive buttons.
Table 73 Buttons/LED Functions
For Both Single and Multistream Options
LEAK LED When lit, indicates leak in main cabinet
Initiates processing from grab sample
CALIBRATION LED When lit, indicates calibration in progress ISE/LAMP LED Will light when millivolt level is below default levels.e
Initiates automatic calibration of analyzer
REAGENTS LED When lit, indicates reagents need to be replaced
Used for setting alarm values.
UP or DOWN Arrow
Increases or decreases displayed value
For Multistream option only
VALVE SELECTED shows which valve (stream) is being analyzed, and VALVE DISPLAYED shows which valve is associated with completed and displayed value
Sets value or exits function
Sets high alarm limit
Initiates automatic multistream analysis
Sets low alarm limit
Initiates single stream analysis
Two functions: Sets 90 day clock for reagent usage Displays number of days remaining until next replacement of reagents
Enters function to identify parameters for each stream (such as number of cycles and high and low alarms)
Sets scale to adjust out put to 4 mA and 0 Vdc on recording device Sets scale to adjust out put to 20 mA and 5 Vdc on recording device

7.3 Card Cage The card cage is located directly behind the keypad panel. To access the card cage, lift latches from right side of keypad panel and open hinged door. (Refer to Figure 72.)
Potentiometer Adjusts sensitivity of leak detector – clockwise for less sensitivity; counterclockwise for greater sensitivity
7.3.1 Card Cage and Door Table 74: Card Cage Parts
Analog Board DIP Switch Contains 8 rocker switches
Keypad Panel Door Opens to access card cage
for setting various functional processes (see Figure 73)
Display Board Circuit board contains electronics for keypad and LCD display
Signal Input Connector For process signals
For each stream 2 – 6, potentiometers for 4 mA, 20 mA, 0 V, 5 V (multistream only)
CPU Board 180B02801 Central processing circuit board processes the signal fed to the digital display and any external devices. (Yellow ejector)
Valve Board 180B01601 (only supplied for multistream option or for RS232 output on single stream)
Circuit board controls multistream accessory. Jumper should be installed on back of valve board. (Purple ejector)
Connector for cable connector from display board
Analog Board 180B02502 For multistream configurations: 180B02501*
Circuit board converts signals from analog to digital and processes signals from "fault" detec tors (Orange ejector)
7.3.5 Power Supply Board Table 78 Power Supply Board
Power Supply Board 180B01203 Supplies voltage to electrical circuitry (Blue ejector)
18pin Female Connector For relays
6pin Male Connector Operating power
6pin Female Connector Valve Pump Power
0Vdc 5Vdc Connector To recording device.
4mA –20mA Connector To recording device.
Computer Memory Reset Clears microprocessor memory
Computer Memory Reset Clears microprocessor memory Valve Pump Connector Connector for cable from valve pump
Display Board Cable Connector
Connector for display board cable

PLEASE NOTE: VALVEPUMP CONNECTOR IS ON CPU BOARD



7.3.6. Analog Board DIP Switches The following list describes the conditions for each switch to be in the ON or OFF position:
Table 79. Analog Board DIP Switch Positions
Switch No. Position Condition Remarks
Analog Board has no adjustment
1 ON When using Analog Board, Part # 180B02602
ports for individual output device and/or 05 Vdc, 420 mA outputs.
Analog Board has individual ports
1 OFF When using Analog Board, Part # 180B02601
for adjusting each output device (up to 6) and/or 05 Vdc, 420 mA outputs.
2 ON Quick calibration on
2 OFF Quick calibration off
4 ON Nonfunctioning switch
5 ON Nonfunctioning switch
6 ON Nonfunctioning switch
Allows markers to be recorded on
7 ON For stream ID and “brackets” before and after calibration
strip charts and other recording devices.
No markers will be recorded on strip charts and other recording devices.
8 ON For 80 column printers Formats printer output
8 OFF For 40 column printers Formats printer output


Figure 73. Analog Board DIP Switches
(with #3 set in OFF position for multistream analyzer) Colorimeter Assembly

7.4 Rear of Main Cabinet
In the rear of the main cabinet are a number of electrical components and connectors. The following diagram is for reference, only.
Backplane Board Circuit boards plug into front side
Power Transformer Supplies ac voltage to electronics
Heating Bath Connector Interconnect to heating bath Pump Connector Interconnect to valve pump
220Vdc Transformer Optional part for specific electrical requirements (208 V – 240 V)


Component Function

7.5 Valve Pump Assembly
The vacuum pump assembly includes the pump, PC board, and reaction cell.
Sample and Reagent Tubing
Colorcoded straws and labeled tubes for transport of solutions
“D” Connector Connects to small, 16pin connector on CPU boards
Pump Housing Contains and protects valve pump electronics
Vent Line Vents reaction cell to waste Pump Drip Tray Catches leaks and directs them to the leak detector
Pump Valve pump routes sample and reagents through the analyzer
Reagent Barb Fittings Male half of tubing connector. Tubes are connected to reagents and sample
LEDS Identify which injector is active.
Contains reaction cell, heating block
* Vacuum/Pressure LED is located on the upper right corner of the ValvePump as it is installed
6position Connector Electrical power to heating bath 9position Connector Electrical power to pump
LED Green is OK; RED is Low pressure
Part Function · Part Function · Reaction Vessel Assembly

7.6 Interface Ports The interface ports (labeled for recorder, printer, relay output, and RS232) are located on the lower, left outside wall.
Pin No. Description
1 NO 9 C Stream #1 Alarm or ID
2 NO 10 C Stream #2 Alarm or ID
3 NO 11 C Stream #3 Alarm or ID
4 NO 12 C Stream #4 Alarm or (#1*) ID
RS232C For connection of computer interface or other accessories
5 NO 13 C Stream #5 Alarm or (#2*) ID
6 NO 14 C Stream #6 Alarm or (#3*) ID
Relay Output Optional relay contact (multistream analyzer only)
If there are 3 streams or less, both stream ID and alarms may be optionally configured.
Printer For connection of printer cable Recorder For connection of external recorder; for multioutput, 420 mA and 05 Vdc outputs for streams 2 through 6.
7.8 MultiOutput Board The output for Stream #1 is obtained from the Input/Output Panel (see Figure 25). The additional five outputs are obtained from the lower lefthand side of the analyzer. Table 714 gives the pin assignments on the “D” connector for a multistream system:
\ 1 2 3 4 5 6 7 8 9 10 11 12 13 14 /


\15 16 17 18 19 20 21 22 23 24 25 /
Figure 77. "D" Connector for Relay Output
2 15 2 14 1
3 17 4 16 3
4 19 6 18 5
5 21 8 20 7
6 23 10 22 9
Figure 76. Interface Ports
7.7 Relay Output Option This port (see Figure 77) is a collection of contact closures that can be used to identify either the stream being displayed or the alarm that is activated (multistream systems only). these are not “powered” contacts.
Relay Contact Ratings: 100 Vdc max voltage 10 Vdc max DC current 1.0 A max carry current 0.5 A max switched current
Pin Assignments · Port Function · “D” Connector

7.9: Overflow Sampling Assemblies
Overflow Waste Vent Allows excess sample
to escape.
Intake/Regulating Valve Regulates flow of sample into overflow sampler assembly
Waste Pipe Carries waste to drain.
Overflow Tubing All overflow tubes feed into common waste pipe. Filters (180134108)
Overflow Tube Assures adequate supply
Replaceable filters are mounted down stream from each intake regulating valve. Sample Inlet Tubing 1/8" x 1/4" tubing (provided by customer).
of sample and reproducible results by eliminating positive pressure from sample stream.
Sample Intake Tubing Tubing which carries sample from overflow tube to analyzer (one to six stream configurations) NOTE: Sample intake tube should be no more than 1 to 11/2" into the sample.
Common Panel Mounts multistream Overflow Sampler Assembly to left side of
analyzer (facing front panel). Panel with dimensions 221/4” x 9" will support up to six streams.
Air Vent Allows air to escape
overflow tube as it fills with sample.
Figure 78. SingleStream Overflow Sampling Assembly


Part Function · Part Function



Figure 79. Multistream Overflow Sampling Assembly
(threestream option)

SECTION 8 ADDITIONAL FEATURES
8.1 Chart Recorders and Data Loggers
A chart recorder or other data acquisition device may be connected to the analyzer for a permanent record of sam ple concentration. Setting the output 420 mA 05 Vdc from the keypad panel allows you to adjust 420 mA and 0 5 Vdc to a recording device.
These input devices must be isolated or have a floating ground (not reference to ground). The contacts for recorder output are located at the input/output panel on the side of the analyzer (see Figure 81).
The analyzer will operate normally with data output to the chart recorder or other device. During normal operation, calibration is performed biweekly. This may vary by Chemistry. To indicate the beginning and end of a calibration sequence, the recorder makes a quick trace down to zero, and up to full scale. This is called a calibration marker. After calibration, the recorder resumes recording the sample values (refer to Figure 82).
When a recorder is connected to a multistream system, a ”stream identification blip” (SIB) is produced prior to the output signal for that particular stream. This allows quick identification of individual stream readings on the recorder. SIB values are shown in Table 81, and tracings are illustrated in Figure 83.
Figure 81. Input/Output Panel


1 0.5 10
2 1.0 20
3 1.5 30
4 2.0 40
5 2.5 50
6 3.0 60
Grab 3.5 70



Figure 82. Typical Trace Recording with
Calibration of Single Stream System



Figure 83. Typical Stream Trace Recording with
Calibration of Multistream System

8.2 Computer Interface The computer interface is an RS232C serial data interchange port operating at 1200 baud. The serial port receives the same information as the printer, and has the capability to transmit information, enabling remote control of the analyzer.
8.2.2 Data Configuration The computer must be set to match the fixed data configuration of the analyzer, which is:
Parameter. Setting
Data 8 bits, high order bit set to zero
8.2.1 Data Output Format The data output formats contain no control characters, with the exception of carriage return, line feed, and the preamble/post amble described below:
8.2.3 RS232C Function Commands The RS232C interface can be used as follows: • By direct hookup to a nonMacIntosh® based computer A program on disc can be obtained from
Preamble NUL, STX, NUL = (00, 02, 00 Hex notation) Date/Time ASCII string of 25
Rosemount Analytical for remote control. Cable length must less than 150 feet. • By direct hookup to a nonintelligent terminal. No
characters, followed by 7 spaces
program disc is required. Cable length must be less than 150 feet. • By communication through a modem. One modem located at the analyzer can communicate over a
Concentration ASCII string of 5
characters, followed by 9 spaces
Stream Number One ASCII character
telephone network to another modem hooked up to an IBM PC or compatible computer or a nonintelli gent terminal. Pin configuration is described in Table 84.
Mode An ASCII string,
consisting of 7 spaces, followed by three possible messages: NOT calibrated, Manual, or Automatic.
Pin No. Circuit
Direction of Signal Description
Calibration Status There are two ASCII messages transmitted: CAL when calibration is in progress, and blank when good data is available. Postamble: NUL, EOT, NUL = (00, 04, 00 Hex notation) Postamble NUL, EOT, NUL = (00,04,00 Hex notation)
1 AA 101 Protective Ground
*2 BA 103 From Terminal
*3 BB 104 To Terminal Received
*4 CA 105 From Terminal
*5 CB 106 To Terminal Clear to Send


8.2.4 Initiating RS232Communications to the Computer 1. With the computer and CFA3000 power turned off, connect RS232C cable to the computer and to the RS232C port on the CFA3000 analyzer. (If a printer is to be used, connect it to the printer port
on the analyzer). 2. Turn on power to the analyzer. 3. From the analyzer keypad panel, turn on the printer by pressing the following buttons: • From a multistream keypad:

SETCLO Sets clock calendar, Date is entered as follows: day of week, year, month, date, hour, minute, second. Day of week is coded as Sunday = 0, Monday = 1, Tuesday = 2, etc. Year is entered as last two digits of year. Month and date must be two digits, use leading zeros as necessary. A 24 hour format is used for hour entry. For example, to set the clock at Thursday, April 12, 2003 at 4:35 p.m., enter SETCLO 4030412163500.
CFATERM, then press Return. 6. When a blinking cursor appears on the screen, press Return. The screen will display M>. 7. At this point, all of the RS232C functions listed in Table 84 are accessible. 8. To set up a file to collect data in the computer, press <altL> (formatted) or <altC> (capture). 9. Type in file name (up to 8 characters). 10. To start collecting data, press Return. The display will show a blinking cursor. 11. To access the command functions again, press Return. The screen will display M>.
Use the commands in Table 85 to control the analyzer.
SETHAL Changes stream high alarm setting. For
AUTOCAL Runs fullscale automatic calibration AUTOVALV Resumes automatic valve selection
example, to change stream 1 to 25, enter SETHAL 1,25.
DATA Forces printing of analyzer data
SETLAL Changes stream low alarm setting. For example, to change stream 2 to 50, enter SETLAL 2,50.
Displays stream parameters, including sample cycles, low and Function
SETVTIM Changes stream valve cycles. For example, to change stream 2 to 2 cycles, enter SETVTIM 2,20.
MANVALV To operate valves manually (for multi stream, one stream continuously). For example, to monitor stream 3, enter MANVALV 3.
RESET Resets the computer for a full wipeout
Command Function · Command

8.2.5 Printer Interface
8.2.6 Set Printer Internal Clock
The printer interface is a female parallel port designed to be compatible with IBM PCtype cables and printers. This builtin capability permits text printing only, no graphics.
Set printer's internal clock before printing; perform the following procedure:
and , then . • From a singlestream keypad, referring to Figure 84, press position 15, then 1, and then 10

Sunday = 0, Monday =1 Press enter. b. Year, two digits: 2003 = 03, 2007= 07
Press enter. c. Month, two digits: Jan = 01,
Feb = 02 Press enter. d. Day of the month, two digits: 01, 02
Press enter. e. Hour, using 24hour format:
8:00 AM = 08, 1:00 PM = 13 Press enter. f. Minute, two digits: 01, 02
at 4:35 PM:


APPENDIX A CHEMISTRIES
General Description: For this procedure for the determination of sodium, the sodium ion is measured potentiometrically using ion selective electrodes . The addition of an ionic strength adjustor to the sample maintains constant background ionic strength and pH.
an inline filter such as Rosemount Analytical' IF 100 (Part #180140001).
Sodium in Water Range: 0.110, 1100, 101000 ppb up to 1000 ppm or higher.
Operating Notes

APPENDIX B SPECIFICATIONS
Accuracy: All twodecade ranges up to 1000 ppb maximum: ±10% of reading or ±2 ppb, whichever is greater
single stream, standard for multistream monitors. RS232C output is optional for single stream monitors, standard for multistream monitors.
0.110 ppm range = ±10% of reading or ±0.5 ppm, whichever is greater
Alarms:
Streams 2 through 6 (multistream option):
Sample Requirements:
Display: Single line LCD, 4 1/2 digits, readout in concentration units. For multistream option display remains at last value until new stream reading is stable.
Status LEDs:
Filter: 8 micron filter in sample line is Recommended; (#180134108).
StatusLEDs show:
Environmental: The analyzer is not intended for out door use. Install in a clean, indoor environment.
Ambient Temperature: 40120° F (1045° C) must be regulated to ± 10° within range, avoid temperature swings greater than ± 10° F (± 5± C) over a 12hour period
Load requirements for 420 mA output: 450 ohms minimum, 1200 ohms? maximum
Load requirements for 05 Vdc output: 10 k Ohms, minimum
The range of outputs is the same as the measure ment range. Parallel output for printer is optional for

APPENDIX B SPECIFICATIONS Cont.
25.75 in. (654 mm) Height of Reagent Cabinet: 18.25 in. (464 mm) Width: 18 in. (457 mm) Depth: 16 in. (406 mm)
Space Requirements: Wall mounting: Allow 18 in. (457 mm) on one side for swing out access to rear panel. Allow 12 in. (305 mm) on the other side for access to input/output panel and overflow sampler. Panel Mounting: Allow 12 in. (305 mm) on both sides for access to input/output panel and overflow panel.
Weight: 117 lb (53 kg).
Add up to 60 lb (27.3 kg) for full weight of
reagents and standards. Shipping Weight of Analyzer: 138 lb (63 kg) Shipping Weight of Reagents: up to 64 lb
(29 kg) per kit (3 months supply)

APPENDIX C POWER INTERCONNECT DESCRIPTION
The block diagram in Figure C1 graphically shows the distribution of power and signals throughout the analyzer. The raised numbers (e.g., (6) ) in the description below relate to the circled numbers on the diagram. For clarity, minor components such as fuses are not shown.
Multistream Analyzer Option In units equipped with the Multistream option, a Valve Board (8c) will be installed. This board contains relay contacts that are externally available at the Interface Ports (11) as well as LED drivers to activate the sample stream indicators located on the Keypad Panel (9) . The Valve Board also provides interfaces for a printer and serial data communications, exter nally available at the Interface Ports.
For 220 VAC 50/60 Hz option, line power flows through the Power Switch and other minor compo nents to the primary of the stepdown Transformer (2). The Stepdown transformer converts the input 220 VAC to 110 VAC for compatibility with the standard CFA3000 electronics. From the Stepdown Transformer secondary power flows to barrier strip TB1.
If the monitor includes a multioutput analog board, analog recorder drivers for each stream are installed on the Analog Board (8d) . These outputs are exter nally available at the Interface Ports.

APPENDIX D BYPASS FOR CALIBRATION
To avoid extraneous alarms from external devices, rewire output as shown below


Figure D1. Calibration Bypass Wiring

APPENDIX E KEYPAD HELP SHEETS
E.1. SingleStream Keypad "Help Sheet"


Figure E1. Keypad Panel for Single Stream Option


Figure E2. Number Positions on a SingleStream Keypad

Table E1: Singlestream Keypad Help Sheet
Press … Function To Exit, Press …
Initiates automatic calibration No action required
Initiates priming and auto calibration No action required
Sets scale to adjust output to 4mA
and 0 Vdc on external device
Referring to Figure E2, on keypad panel, press position 15 and then 2. Initiate valve pump test operation
Display high alarm setting
or Set high alarm value
Display low alarm setting
Change low alarm value (MUST BE SET LOWER
or

Press … Function To Exit, Press …
Set reagent supply time
Display number of days remaining
until replacement of reagents
Set output scale for 20mA and
. Display mV
Display measurement of previous
Referring to Figure 84, press position 15, then 1, and then 10
Turn on printer No action required
Referring to Figure 84, press posi
tion 15, then 1, and then 10 Turn on printer No action required



Table E2. Multistream Keypad Help Sheet
Press … Function To Exit, Press …
Initiates automatic calibration No action required
Initiates priming and auto calibration No action required
Sets scale to adjust output to 4mA
and 0 Vdc on external device
Initiate valve pump test operation

Press … Function To Exit, Press …
or Display a particular stream in multi
stream system, to set parameters
Then, or Set number of cycles for a particular
stream in multistream system
Set high alarm value for stream (IF ONLY A LOW ALARM VALUE IS
ALARM
Press … Function To Exit, Press …
Change low alarm value (MUST BE SET LOWER THAN HIGH ALARM SETTING)
Display low alarm setting
Display high alarm setting
Initiates multistream sampling No action required
Set reagent supply time (90 days)
or · or

Press … Function To Exit, Press …
Display number of days remaining
until replacement of reagents
Set output scale for 20mA and 5Vdc
Press … Function To Exit, Press …
Turn on printer No action required
Turn on printer No action required

ORDERING INFORMATION
Analyzer reagents are available from Rosemount Analytical on a contract basis. Refer to APPENDIX A, "Chemistries" for specific ordering information or contact Rosemount Analytical
Replacement Parts and Accessories
Table E3 Replacement Parts,
Part Part No.
Part Part No.
E Prom (NOTE: specify chem istry range with decimal places, part #'s of analog, and line frequency, 50/60 Hz, number of streams)
Valve Pump Assembly 185A00101
Solenoid Valve, multistream 180B30101
Waste Tube Assembly 180134601
Overflow Sampler Assembly:
Sample Line Tubing (.0625" ID) 180285201
1 Stream Analyzer 180130001
Filter, disposable (for overflow panel assembly)
2 Stream Analyzer 180130002
3 Stream Analyzer 180130003
4 Stream Analyzer 180130004
5 Stream Analyzer 180130005
6 Stream Analyzer 180130006
IF100 InLine Filter 180140001
Fuse, T 1 A/250 Vdc 625000302
Clamp, singlestream overflow 180132701
Clamp, multistream overflow 180133701
Power Supply Board 180B01203
Multi Output Board 180B02601
Keypad & Display Board Assembly, single stream 185B00301
Keypad & Display Board Assembly, multistream 185B00401

Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes, membranes, liquid junctions, electrolyte, orings, catalytic beads, etc., and Services are warranted for a period of 90 days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Liquid Division 2400 Barranca Parkway Irvine, CA 92606
The shipping container should be marked:
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.
If the material is returned for outofwarranty repairs, a purchase order for repairs should be enclosed.
Emerson Process Management · Return for Repair · Model

SECTION 9.0 RETURN OF MATERIAL
9.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for repair, call 19497578500 for a Return Materials Authorization (RMA) number.
9.2 WARRANTY REPAIR.
The following is the procedure for returning instruments still under warranty:
Rosemount Analytical Liquid Division 2400 Barranca Parkway Irvine, CA 92606
Please see second section of “Return of Materials Request” form. Compliance with the OSHA requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required.
RMA No
Model No
9.3 NONWARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:




The right people, the right answers, right now.
Now there’s a way to quickly get the right answers for your liquid analytical instrumentation ques tions: the Analytical Customer Support Center.
Our staff of trained professionals is ready to provide the information you need. If you are placing an order, verifying delivery, requesting application information, or just want to contact a Rosemount Analytical representative, a call to the Customer Support Center will provide you with the right peo ple, the right answers, right now.
Emerson Process Management’s field sales offices are your source for more information on the fill line of Rosemount Analytical products. Field sales personnel will work closely with you to supply technical data and application information.
For more information, please contact your nearest Emerson Process Management sales office.
THE AMERICAS HEADQUARTERS
Emerson Process Management Process Gas Analyzer Center of Excellence GmbH & Co. OHG Industriestrasse 1 63594 Hasselroth Germany T 49.6055.884.0 F 49.6055.884.20
Emerson Process Management Rosemount Analytical Inc. Liquid Center of Excellence 2400 Barranca Parkway Irvine, CA 92606 Phone: +1.949.757.8500 Toll Free: +1.800.854.8257 Fax: +1.949.474.7250
LATIN AMERICA
Emerson Process Management Asia Pacific Private Ltd. 1 Pandan Crescent Singapore 0512 Republic of Singapore Phone: 65.777.8211 Fax: 65.777.0947
Emerson Process Management Rosemount Analytical 10241 West Little York, Suite #200 Houston, TX 77040 USA T 713.467.6000 F 713.827.3328
EUROPE
Emerson Process Management EPM Building P. O. Box 17033 Jebe Ali Free Zone Dubai, United Arab Emirates T 971.4.8835235 F 971.4.8835312
Emerson Process Management AG Blegistrasse 21 CH6341 BaarWalterswil Switzerland T 41.41.768.6111 T 41.41.761.8740


The right people, the right answers, right now.
http://www.rosemountanalytical.com
Credit Cards for U.S. Purchases Only.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 7578500 Fax: (949) 4747250
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2013
Emerson Process Management
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| No power | Power switch OFF | Press power switch ON | p.34 |
| No power | Power switch failed | Call Rosemount Analytical | p.34 |
| Pump is inoperative | Valve pump switch OFF | Press switch ON | p.34 |
| Pump is inoperative | Blown fuse | Replace fuse | p.34 |
| Pump is inoperative | Leak detector activated | Find and fix leaking; check pump’s electrical connection | p.34 |
| Pump is inoperative | Bad electrical connection | Check pump’s plugs to CPU & connector | p.34 |
| Pump is inoperative | Still inoperative | Call Rosemount Analytical | p.34 |
| Keypad lights or display fails to light up or buttons fail to operate | Circuit boards are not seated properly | Remove and reinstall circuit boards | p.34 |
| Keypad lights or display fails to light up or buttons fail to operate | Display or associated power circuit faulty | Call Rosemount Analytical | p.34 |
| Keypad lights or display fails to light up or buttons fail to operate | Connector faulty or not properly inserted | Check connection | p.34 |
| Alarms not activated | Incorrect settings | Readjust settings, making sure that low alarm is set lower than high alarm | p.35 |
| Alarms not activated | Insufficient time for alarm to respond | Alarm condition must exist for minimum of 60 seconds before alarm activates | p.35 |
| Signal not repetitive (same sample not repetitive) | Small leak or clogging in fluidics | Check all tubing lines and connections | p.35 |
| Signal not repetitive (same sample not repetitive) | Clogged tubing | Replace tubing | p.35 |
| No signal, or signal “offscale” high | Connector or wires loose Faulty circuit board Incorrect mV value | Check circuit boards for proper seating Replace power supply circuit board | p.35 |
| Loss of mV Sensitivity LED ISE on | Reagents, standards old, contaminated or improperly prepared | Change reagents | p.35 |
| Loss of mV Sensitivity LED ISE on | Reagent straws in wrong reagent or standard | Check placement of straws | p.35 |
| Loss of sample flow | Replace sample filter on overflow panel | p.35 | |
| Decreased pumping volume | Tubes pinched shut or clogged from deposits | Remove condition causing “pinched” lines; replace sample filter at overflow panel | p.35 |
| Leak detector LED lights up | Leak in fluidics | Repair leak and dry leak detector contacts; turn main power OFF for 20 seconds | p.35 |
| Leak detector LED lights up | High humidity | Adjust screw on leak detector potentiometer on analog board; turn clockwise to decrease sensitivity | p.35 |
| Leak detector LED lights up | Connector faulty or not properly insertede | Check connection to analog board; replace analog board. | p.35 |
| Front panel display | Test LED’s and LCD on keypad: a. Press Power button and reset button on the CPU board (yel low ejector) b. Release power, then Reset button. Press power | Press: | p.35 |