Product Manual
70-82-25-119
January 2009

Copyright 2008 by Honeywell Revision 5 January 2009
Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace without charge those items it finds defective. The foregoing is Buyer's sole remedy and is in lieu of all other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.
While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.
Honeywell 2500 W. Union Hill Drive Phoenix, Arizona 85027
UDA2182 is a U.S. registered trademark of Honeywell
Other brand or product names are trademarks of their respective owners.
ii UDA2182 Universal Dual Analyzer Product Manual January 2009
Honeywell Process Solutions

This document provides descriptions and procedures for the Installation, Configuration, Operation, and Troubleshooting of your UDA2182 Universal Dual Analyzer.
The following lists Honeywell’s World Wide Web sites that will be of interest to our customers.
Honeywell Organization WWW Address (URL)
Corporate http://www.honeywell.com
Honeywell Field Solutions http://www.honeywell.com/ps
Technical tips http://content.honeywell.com/ipc/faq
Contact us by telephone at the numbers listed below.
United States and Canada Honeywell 1-800-423-9883 Tech. Support 1-800-525-7439 Service
January 2009 UDA2182 Universal Dual Analyzer Product Manual iii
Organization Phone Number

The following table lists those symbols used in this document to denote certain conditions.
This CAUTION symbol on the equipment refers you to the Product Manual for additional information. This symbol appears next to required information in the manual.
Protective Earth (PE) terminal. Provided for connection of the protective earth (green or green/yellow) supply system conductor.
Functional earth terminal. Used for non-safety purposes such as noise immunity improvement. NOTE: This connection shall be bonded to protective earth at the source of supply in accordance with national local electrical code requirements.
Earth Ground. Functional earth connection. NOTE: This connection shall be bonded to Protective earth at the source of supply in accordance with national and local electrical code requirements.
Chassis Ground. Identifies a connection to the chassis or frame of the equipment shall be bonded to Protective Earth at the source of supply in accordance with national and local electrical code requirements.
iv UDA2182 Universal Dual Analyzer Product Manual January 2009
Symbol Definition


The UDA2182 Universal Dual Analyzer is the next level of dual channel analyzers providing unprecedented versatility and flexibility. The UDA2182 can accept single or dual inputs from Honeywell Direct pH, pH from preamp, ORP (Oxidation Reduction Potential), Contacting Conductivity and Dissolved Oxygen sensors. Measurements for Dual channel units can be arranged in any combination of measurement.
“Process Information at a Glance” is a unique feature of the UDA2182 graphical backlit LCD.
Two PV values with corresponding UOM (unit of measure), temperature, alarm state, scales, and limits, tagging, and status messages can be displayed simultaneously.
Ten dedicated keys provide direct access to Setup configuration menus and sub-menus and Calibration.
Menu-driven configuration of the UDA2182 is intuitive, fast and easy. A Setup menu is provided for every configuration task. You will be permitted to configure only those parameters relevant to your application and supported by the Analyzer model you purchased.
In fact, Setup configuration screens will contain only prompts and menu choices that apply to your application.
Multi-language prompts guide the operator step-by-step through the configuration process assuring quick and accurate entry of all configurable parameters. Nine languages are available via configuration: English, French, German, Spanish, Italian, Russian, Turkish, Polish and Czech.
Analytical measurements of Direct pH, pH from preamp, ORP, Conductivity and Dissolved Oxygen (ppm or ppb) can all be done in one analyzer. The unit can be used as a single input or dual input instrument – you decide what measurements are included. The input boards are factory calibrated and easily replaced. Addition of additional relays or an analog output is done with a single board. The “Mix –n- Match” design reduces inventory and increases flexibility. You can purchase a basic unit and then add input and output boards as needed.
Introduction

Two standard Analog outputs 0 –20 or 4–20 mAdc, 750 ohms maximum, isolated from inputs, ground, and each other, and independently assignable to any parameters and ranges Proportional to user-set output range(s) of selected parameter(s).
One optional Analog output 0 –20 or 4–20 mAdc, 750 ohms maximum, isolated from inputs, ground, and each other, and independently assignable to any parameters and ranges.
Two 4A SPDT alarm/control relays are standard; with an additional two 4A relays available as an option.
The infrared connection provides a non-intrusive wireless connection with the instrument and maintains its weather tight integrity when combined with the optional PIE (Process Instrument Explorer).
No need to get access to the back of the analyzer to communicate with the instrument, no need to take your screw driver to wire the communication cable, no wiring mistake possible. You can now duplicate an instrument’s configuration, upload or download a new configuration in a matter of seconds, just by pointing your Pocket PC in the direction of the instrument.
The Communication card provides one Serial Port and one Ethernet Port.
Modbus RTU protocol to read signals and read/write variables
Ethernet port provides:
Multi-language web pages to monitor readings, alarms, statuses, events
Multi-language web pages to setup Ethernet port settings
Multi-language email to send alarm status changes
Modbus TCP protocol to read signals and read/write variables
Both ports can communicate to a PIE tool
Introduction · Serial port provides · RS422/RS485 multi-drop

Measured pH temperature is compensated in one of two ways. Electrode temperature sensitivity is automatically compensated to display the correct pH value at temperature. In addition, displayed pH can be optionally normalized to a solution temperature of 25°C as determined by the current Solution Temperature Coefficient, which is expressed in units of pH/°C with precision to the hundredths decimal place. The parameter “Solu Temp Coeff” allows the selection of Pure Water, Ammonia, Phosphate, Morpholine, and Custom or None (User Entry).
Measured Conductivity and Resistivity can optionally be temperature compensated to 25°C for a specific solution type. TDS and concentration are always measured based on a specific solution type. The cell constant and measurement type determines which solution types are available for selection.
Dissolved Oxygen accurately measures the concentration of dissolved oxygen in water. The Analyzer energizes the probe and receives dissolved oxygen and temperature signals. Optional salinity compensation is provided. The Analyzer provides for Air or Sample calibration with ambient temperature and atmospheric pressure compensation.
High purity water pH can be calculated from Specific and Cation conductivities to be used as a check on in-line high purity water pH measurements.
“Buffer Group” types NIST/USP, USA, or Europe determines the set of standard pH buffer values to be used for Zero and Slope calibration by automatic buffer recognition. Each of the available Buffer Groups is a set of 5 or 6 pH buffer standards.
For high purity water measurement you can select pre-set compensations or configure custom values.
Relays can be configured to alarm on conductivity values as determined by the USP26 standards.
The availability of calculated variables in the list of available sources for alarms, math and control and for status display is determined by similarity of units of measure between the two input boards. For example with Dual Conductivity, %Rejection/Passage, Difference, or Ratio can be displayed and assigned to the outputs or alarms. CO2 concentration in ppm can be calculated from de-gassed conductivity measurement.
Introduction

Keyboard security protects configuration and calibration data. A password (up to four digits) can be configured. If the security feature is enabled, the password will be required to access configuration and calibration software functions.
Built-in real time clock is used to set-up versatile cycles that can be used to initiate automatic sensor cleaning and then calibration.
Continuous diagnostic routines detect failure modes, trigger a failsafe output value and identify the failure to minimize troubleshooting time. The UDA2182 Analyzer performs extensive self-diagnostics as a background task during normal operation. If a problem is detected, a message is displayed on the Message stripe to alert the operator. In addition, the operator can initiate keypad and display tests using Maintenance Menu functions.
The analyzer is designed to provide reliable, error-free performance in industrial environments that often affect highly noise-sensitive digital equipment.
CSA Type 4X (NEMA 4X) rated enclosure permits use in applications where it may be subjected to moisture, dust, or hose-down conditions. The UDA2182 is designed for panel, pipe or wall mounting.
Introduction

Display Graphical LCD with white LED Backlight Viewing Area: 66.8 mm (W) X 35.5 mm (H) Dot Pixels: 128 (W) X 64 (H) Display Ranges pH: 0-14 pH Temperature: -10 to 110°C (14 to 230°F) ORP: -1600 to +1600 mV Conductivity: 0.01 Cell: 0-2 uS/cm displayable to 200 uS/cm; 0-0.2 mS/cm; 0-2,000 ppb TDS; 0-200 ppm TDS 0.1 Cell: 0-20 uS/cm displayable to 2000 uS/cm; 0-2 mS/cm, 0-2,000 ppb TDS; 0-2,000 ppm TDS, 1.0 Cell: 0-200 uS/cm displayable to 20,000 uS/cm; 0-20 mS/cm; 0-200 ppm TDS; 0-20 ppt TDS 10 Cell: 0-2,000 uS/cm displayable to 99999 uS/cm; 0-200 mS/cm; 0-2,000 ppm TDS; 0-200 ppt TDS 25 Cell: 0-20,000 uS/cm displayable to 99999 uS/cm; 0-500 mS/cm; 0-10% Concentration displayable to 20% 50 Cell: 0-20,000 uS/cm displayable to 99999 uS/cm; 0-1,000 mS/cm; 0-20% Concentration Temperature: 0 to + 140°C (32 to 284°F) Dissolved Oxygen: 0 - 20 ppm 0 –200 ppb, displayable to 20000 ppb 0 – 100% saturation, displayable to 200% saturation Temperature: 2 – 60°C (35.6 – 104°F), must not freeze Keypad 10 Button Membrane Switch w/Directional Functionality UV/Solvent/Abrasion Resistant Case Material GE Valox
® 357 (un-reinforced thermoplastic polyester) Performances (Under reference operating conditions)
Accuracy: 0.5% of reading Output Accuracy: +/- 0.01 mA Drift: Negligible Repeatability: 0.05% Temperature Accuracy:
pH and Conductivity Thermistor: +/- 0.1°C from –10 to 100° C, +/- 1.0° C from 101° to 140° C pH 1000 ohm RTD: +/- 0.4° C D.O. Thermistor: +/- 0.1° C from 0 to 60° C Reference Operating Conditions: 25 +/- 1° C; 10-40% RH; 120 or 240 Vac Operating Conditions Ambient Temperature Operating: 0 to 60°C (32 to 140°F) Storage: -30 to 70°C (-22 to 158°F) RH: 5 to 90% max. Non-condensing up to 40°C (104°F). For higher temperatures the RH specification is derated to maintain constant moisture content Vibration: 5-15 Hz disp 8 mm pk to pk 15-200 Hz accel 2 G Standard Analog Output Two 0-20 mAdc or 4-20 mAdc, 750 ohms max., isolated from inputs, ground, and each other, Independently field-assignable to any parameters and ranges. Proportional to user-set output range(s) of selected parameter(s), Optional Analog Output One 0-20 mAdc or 4-20 mAdc, 750 ohms max., isolated from inputs, ground, and each other. Independently field-assignable to any parameters and ranges
Specifications

Control Loop/Outputs Control Loops: 2 standard (one for each PV); current, pulse frequency, or time proportional Control Loop Types: PID (optional), Duplex (optional), On/Off (standard) Auto-tuning: Accutune II, fuzzy logic overshoot suppression, applicable to both PID loops Standard Alarm/ Control Relays
Two SPDT (Form “C”) Relays Resistive Load Rating: 4A, 120/240 Vac Optional Additional Alarm/Control Relays
Two SPDT (Form “C”) Relays Resistive Load Rating: 4A, 120/240 Vac Alarm/Control Settings Alarm/on-off control delay: 0-100 seconds. Alarm/on-off control deadbands: individually set, from 1 count to full scale for pH, ORP, and temperature. On/off cycle period: 0 to 1000 seconds. On/off percent “on” time: 0 to 100%, 1% resolution. Set point and proportional band limit ranges: ±19.99 pH, ±1999 mV, -10 to 130°C, 1 count resolution. DAT cycle period: 1 to 1999 seconds. PFT maximum frequency: 1 to 200 pulses/minute. PFT pulse width: 50 ms, compatible with electronic pulse-type metering pumps. Remote Preamplifier Input Option
Optional input card to accept input signal from Honeywell digital preamplifiers: Meridian II – 31075707 and 31022283 Durafet – 31079288 and Cap Adapter cables pH Temperature Compensation
Conventional compensation for changing electrode output (Nernst response), plus selectable solution temperature compensation for high-purity water. Calculated pH from Differential Conductivity
User selectable when unit has two Conductivity inputs. Used when ammonia or amine is the water treatment chemical. Auto Buffer Recognition (pH)
NaCl, HCl, H2SO4, PO4, NaOH, NH3, C4H9C, Pure Water, Custom (User Selectable)
Max flowrate (probe): 950 ml/min with flow chamber; no dependence on stirring or flowrate Atmospheric pressure: 500-800 mm Hg with internal sensor, for calibration Calibration with either Air or Sample Auto Clean/ Auto Cal Function
Real time clock is used to set-up cycles to initiate a cleaning and calibration sequence. Cycle Set-up is user configurable. Event History Screen Event history screen stores 256 events with a description of the event and a Date/time stamp. Calibration History Screen Calibration history screen stores information on 128 calibration events with a date/time stamp. Power Requirements 90 -264 Vac, 47-63 Hz, 15 VA. Memory retained by E 2PROM when power is off. Wireless Interface
Type: Infrared (IR) Length of Link: 0 –1 M, 0 –15° Offset Baud Rate: 9600 Data Format: Modbus Protocol RS422/RS485 Modbus RTU Slave Communications Interface (Optional)
Baud Rate: 2400, 4800, 9600, 19200, 38400, 57600, or 115200 selectable Data Format:: IEEE floating point and 32-bit integer. Word swap configurable. Length of Link:
2000 ft (600 m) max. with Belden 9271 Twinax Cable and 120 ohm termination resistors 4000 ft (1200 m) max. with Belden 8227 Twinax Cable and 100 ohm termination resistors
Link Characteristics: Two-wire (half-duplex), multi-drop Modbus RTU protocol, 15 drops maximum or up to 31 drops for shorter link length. Modbus RTU slave: Provides monitoring of inputs outputs, statuses, alarms, and variables. Provides writing of variables for remotely modifying parameter settings.
Specifications · Dissolved Oxygen Measurement

Ethernet TCP/IP Communications Interface (Optional)
Type: 10 or 100 BaseT; auto-speed and auto-polarity sensing Length of Link: 330 ft. (100 m) maximum. Use Shielded twisted-pair, Category 5 (STP CAT5) Ethernet cable. Link Characteristics: Four-wire plus shield, single drop, five hops maximum IP Address: IP Address is 192.168.1.254 as shipped from the factory Recommended network configuration: Use Switch rather than Hub in order to maximize UDA Ethernet performance Configuration: Ethernet parameters are configured via the front-panel or web pages. Modbus TCP/IP: Five simultaneous socket connections provide monitoring of inputs outputs, statuses, alarms, and variables. Provides writing of variables for remotely modifying parameter settings. Modbus TCP/IP Data Format: IEEE floating point and 32-bit integer. Word swap configurable. Web server: multiple client support Multi-language Web pages: monitoring inputs, outputs, statuses, alarms, and events Multi-language Email: Alarm notification to eight email addresses. These must be configured using web pages signed in as the administrator. DHCP: ( Dynamic Host Configuration Protocol) selectable via web page or front-panel Safety Compliance UL/CSA General Purpose FM/CSA Approval for Class I, Div 2; Groups A, B, C and D. T4, Ta =60°C CE Compliance CE Conformity (Europe): CE Mark on all models signifies compliance to EMC Directive 84/336/EEC and LVD Directive 73/23/EEC. EMC Classification: Group 1, Class A, ISM Equipment Method of Assessment: Technical File (EN61010-1; EN 61326) Declaration of Conformity: 51453667 Case Dimensions 156 mm X 156 mm X 150 mm (6.14” X 6.14” X 5.91”) Panel cutout: 138.5 mm X 138.5 mm (5.45” X 5.45”) Panel thickness: 1.52 mm (0.06”) min, 9.5 mm (0.38”) max Enclosure Rating CSA Type 4X (NEMA 4X) rated enclosure FM Class 1, Div 2 Installation Ratings Installation Category (Overvoltage Category): Category II Pollution Degree: 2 Altitude: 2000 m Weight Approx 3 lbs (6.6kg) Mounting Panel mounting-hardware supplied. Optional Wall and 1” to 2” pipe mounting. Select option appropriate in Model Number. 2.2 CE Conformity (Europe)
This product is in conformity with the protection requirements of the following European Council Directives: 73/23/EEC, the Low Voltage Directive, and 89/336/EEC, the EMC Directive. Conformity of this product with any other “CE Mark” Directive(s) shall not be assumed.
Product Classification: Class I: Permanently connected, panel-mounted Industrial Control Equipment with protective earthing (grounding) (EN61010-1).
Enclosure Rating: The front panel of the analyzer is rated at NEMA4X when properly installed.
Installation Category (Overvoltage Category): Category II (EN61010-1)
Pollution Degree: Pollution Degree 2: Normally non-conductive pollution with occasional conductivity caused by condensation. (Ref. IEC 664-1)
Specifications

The emission limits of EN61326 are designed to provide reasonable protection against harmful interference when this equipment is operated in an industrial environment. Operation of this equipment in a residential area may cause harmful interference. This equipment generates, uses, and can radiate radio frequency energy and may cause interference to radio and television reception when the equipment is used closer than 30 meters (98 feet) to the antenna (e). In special cases, when highly susceptible apparatus is used in close proximity, you may have to employ additional mitigating measures to further reduce the electromagnetic emissions of this equipment.
If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Specifications

This section contains instructions for unpacking, preparing, and mounting the Analyzer. Instructions for wiring are provided in Section 4 (power wiring) and Section 7 (input wiring). Software configuration is described in Section 6.
The UDA2182 Analyzer can be panel, wall, or pipe mounted.
Each unit has (4) 22.22mm [.87"] dia. holes on the bottom of the unit for lead wires and conduit fittings. The user supplies the conduit fittings.
To avoid damage to the case when connecting to a rigid metallic conduit system, the conduit hub must be connected to the conduit before the hub is connected to the enclosure
When installing the unit, you must select appropriate watertight fittings to insure watertight integrity.
The topics in this section are listed below.
3.1 Overview 9
3.2 Unpacking and Preparing 10
3.3 Mounting 10
Unpacking · Topic See Page

Table 3-1 Procedure for Unpacking and Preparing the UDA2182
ATTENTION
For prolonged storage or for shipment, the instrument should be kept in its shipping container. Do not remove shipping clamps or covers. Store in a suitable environment only (see specifications in Section 2).
1 Carefully remove the instrument from the shipping container.
2 Compare the contents of the shipping container with the packing list.
3 Remove any shipping ties or packing material. Follow the instructions on any attached tags, and then remove such tags.
4 All UDA2182 Analyzers are calibrated and tested at the factory prior to shipment. Examine the model number on the nameplate to verify that the instrument has the correct optional features.
5 Select an installation location that meets the specifications in Section 2. The UDA2182 can be panel- , wall-, or pipe-mounted (see Section 3.3).
Pipe mounting is not recommended if the pipe is subject to severe vibration. Excessive vibration may affect system performance.
6 If extremely hot or cold objects are near the installation location, provide radiant heat shielding for the instrument.
The Analyzer can be mounted on either a vertical or tilted panel or can be pipe or wall mounted (option) using the mounting kit supplied. Overall dimensions and panel cutout requirements for mounting the analyzer are shown in Figure 3-1. Pipe mounting is shown in Figure 3-3. Wall Mounting is shown in Figure 3-4.
For Sample Tap Electrode Mounting recommendations, See Section 15.15 – page 229.
The analyzer’s mounting enclosure must be grounded according to CSA standard C22.2 No. 0.4 or Factory Mutual Class No. 3820 paragraph 6.1.5.
Before mounting the analyzer, refer to the nameplate on the outside of the case and make a note of the model number. It will help later when selecting the proper wiring configuration.
Unpacking · Step Action


Customer will need to provide a rear panel support plate to maintain NEMA4 protection if primary panel thickness is less that 2.3mm [0.09”] thick
Customer will need to provide a rear panel support plate to maintain NEMA4 protection if primary panel thickness is less that 2.3mm [0.09”] thick

156 [6.14]
156 [6.14]
(4) 22.22[.87] holes for lead wires and conduit fittings (conduit fittings supplied by user)
(4) 22.22[.87] holes for lead wires and conduit fittings (conduit fittings supplied by user)
Panel Mounting Procedure
Table 3-2 Panel Mounting Procedure
1 Mark and cut out the analyzer hole in the panel according to the dimension information in Figure 3-1.
2 Orient the case properly and slide it through the panel hole from the front. Customer will need to provide a rear panel support plate to maintain NEMA4 protection if primary panel thickness is less that 2.3mm [0.09”] thick – See Figure 3-2.
3 Remove the mounting kit from the shipping container and clamp the edges of the cutout between the case flange and the supplied U-bracket that is fastened to the rear of the case using (2) M5 X 16mm long screws and (2) M5 lock washers supplied.
Unpacking · Step Action


Unpacking

The analyzer can be mounted vertically or horizontally on a pipe. Use the bracket and hardware supplied in the mounting kit. Select 1 inch or 2 inch U-Bolts.
Pipe mounting is not recommended if the pipe is subject to severe vibration. Excessive vibration may affect system performance.
M5 X 10mm long screw with M5 lock washer (2 places) Note orientation of hole and slot in mounting bracket. Hole is to be in the upper position.
M5 X 10mm long screw with M5 lock washer (2 places) Note orientation of hole and slot in mounting bracket. Hole is to be in the upper position.
M8 Nut M8 Lock Washer M8 Flat Washer
M8 Nut M8 Lock Washer M8 Flat Washer
195.1 [7.68]
195.1 [7.68]
97.5 [3.84]
97.5 [3.84]
188.1 [7.40]
188.1 [7.40]
77.4 [3.05]
77.4 [3.05]
156 [6.14]
156 [6.14]
Do not over tighten fasteners 4.5Nm (40 Lb-in) of torque max.
Do not over tighten fasteners 4.5Nm (40 Lb-in) of torque max.
1 or 2 inch Vertical Rear Pipe Mounting
1 or 2 inch Vertical Rear Pipe Mounting

97.5 [3.84]
97.5 [3.84]
97.5 [3.84]
97.5 [3.84]
188.1 [7.40]
188.1 [7.40]
78 [3.07]
78 [3.07]
78 [3.07]
78 [3.07]
1 or 2 inch Horizontal Rear Pipe Mounting
1 or 2 inch Horizontal Rear Pipe Mounting
Unpacking

The analyzer can be mounted on a wall. Use the bracket and hardware supplied in the mounting kit.

188.1 [7.40]
188.1 [7.40]
195.06 [7.680]
195.06 [7.680]
Left hand Side View
Left hand Side View
97.53 [3.840]
97.53 [3.840]

195.1 [7.68]
195.1 [7.68]
195.1 [7.68]
195.1 [7.68]
97.5 [3.84]
97.5 [3.84]
83.9 [3.30]
83.9 [3.30]
38.5 [1.51]
38.5 [1.51]
77 [3.03]
77 [3.03]
Front View Mounting Bracket
Front View Mounting Bracket
83.9 [3.30]
83.9 [3.30]
Four slots in bracket for 6.0mm [1/4 “] dia mounting bolts supplied by customer
Four slots in bracket for 6.0mm [1/4 “] dia mounting bolts supplied by customer
167.6 [6.60]
167.6 [6.60]

83.9 [3.30]
83.9 [3.30]
167.8 [6.61]
167.8 [6.61]
Front View Mounting Bracket
Front View Mounting Bracket
38.5 [1.52]
38.5 [1.52]
77 [3.03]
77 [3.03]

This section contains instructions for installing ac power wiring for the Analyzer, in preparation for performing configuration setup as described in Section 6.
We recommend that you wait to install input and output wiring (See Section 7) until after Configuration Setup. During configuration the software will determine for you, which relay to use for each feature.
The topics in this section are listed below.
4.1 Overview 15
4.2 General Wiring Practices 16
4.3 Power Wiring Considerations 17
4.4 Installing Power Wiring 17
Power Wiring · Topic See Page

Qualified personnel should perform wiring only.
WARNING
A disconnect switch must be installed to break all current carrying conductors. Turn off power before working on conductors. Failure to observe this precaution may result in serious personal injury.
An external disconnect switch is required for any hazardous voltage connections to the relay outputs.
ATTENTION
This equipment contains devices that can be damaged by electrostatic discharge (ESD). As solid-state technology advances and as solid-state devices get smaller and smaller, they become more and more sensitive to ESD. The damage incurred may not cause the device to fail completely, but may cause early failure. Therefore, it is imperative that assemblies containing static sensitive devices be carried in conductive plastic bags. When adjusting or performing any work on such assemblies, grounded workstations and wrist straps must be used. If soldering irons are used, they must also be grounded. A grounded workstation is any conductive or metallic surface connected to an earth ground, such as a water pipe, with a 1/2 to 1 megohm resistor in series with the ground connection. The purpose of the resistor is to current limit an electrostatic discharge and to prevent any shock hazard to the operator. The steps indicated above must be followed to prevent damage and/or degradation, which may be induced by ESD, to static sensitive devices.
In applications where either the power, input or output wiring are subject to electromagnetic disturbances, shielding techniques will be required. Grounded metal conduit with conductive conduit fittings is recommended.
Connect the AC mains through a fused disconnect switch.
Instrument wiring should conform to regulations of the National Electrical Code.
Power Wiring

Observe all applicable electrical codes when making power connections. Unless locally applicable codes dictate otherwise, use 14-gauge (2.081 mm2) wire for ac power, including protective earth.
The power supply voltage and frequency must be within the limits stated in the specifications in Section 2.
WARNING
Turn power off at mains before installing AC Power Wiring. Do not remove boards with power ON.
The ground terminal must be connected to a reliable earth ground for proper operation and to comply with OSHA and other safety codes. If metal conduit is used, connect a bonding wire between conduits. Do not rely upon the conductive coating of the instrument case to provide this connection. Failure to observe this precaution may result in serious personal injury.
To avoid damage to the case when connecting to a rigid metallic conduit system, the conduit hub must be connected to the conduit before the hub is connected to the enclosure
1 Check the tag on the outside of the case to be sure that the voltage rating of the unit matches the input voltage at your site.
The Unit may be damaged if you apply power with the wrong voltage.
3 Refer to Figure 7-1 for the location of the printed wiring board retainer. Loosen the two
Power Wiring · Step Action

screws that hold the retainer and slide the retainer to the left until the retainer tabs disengage from the terminal boards.
4 Refer to Figure 7-1 for the location of the Power Supply/Analog Output/Relay Output board. Insert a screwdriver into the hole in the middle of the terminal board and pull out gently. Slide the board half way out. There is a notch in the terminal board into which you can slide the retainer tabs and hold the board in place while wiring.
5 Install a fused disconnect switch in the power line that will be connected to the Analyzer.
Fuse must be a Time-Delay or Slo-Blo type.
6 Each unit has (4) 22.22mm [.87"] dia. holes on the bottom of the unit for lead wires and conduit fittings. Conduit fittings to be supplied by the user.
Feed the power wiring through the wiring port on the bottom of the case. Connect the power wiring to terminals L1 and L2/N as shown in Figure 4-1. Connect the Green safety ground wire to the grounding stud on the case.
7 Slide the retainer to the left then slide the terminal board back into place. Slide retainer to engage the tabs and tighten the screws.
8 Close the Bezel and secure four captive screws to a torque value of .20 Nm (1.5 Lb-in). Power up the unit. Do not apply power until the bezel is closed.
Power Wiring · Step Action


Analog Output 1 (+)
Analog Output 1 (+)
Analog Output 1 (+)
Analog Output 1 (+)
Analog Output 1 (–) Analog Output 2 (+)
Analog Output 1 (–) Analog Output 2 (+)
Analog Output 1 (–) Analog Output 2 (+)
Analog Output 1 (–) Analog Output 2 (+)
Analog Output 2 (–) Relay Output 1 (N.O.)
Analog Output 2 (–) Relay Output 1 (N.O.)
Analog Output 2 (–) Relay Output 1 (N.O.)
Analog Output 2 (–) Relay Output 1 (N.O.)
Relay Output 1 (COM) Relay Output 1 (N.C.) Relay Output 2 (N.O.) Relay Output 2 (COM) Relay Output 2 (N.C.)
Relay Output 1 (COM) Relay Output 1 (N.C.) Relay Output 2 (N.O.) Relay Output 2 (COM) Relay Output 2 (N.C.)
Relay Output 1 (COM) Relay Output 1 (N.C.) Relay Output 2 (N.O.) Relay Output 2 (COM) Relay Output 2 (N.C.)
Relay Output 1 (COM) Relay Output 1 (N.C.) Relay Output 2 (N.O.) Relay Output 2 (COM) Relay Output 2 (N.C.)
AC N L2
AC N L2
AC N L2
Figure 4-1 Power Wiring

This section contains instructions for operating the Analyzer.
The topics in this section are listed below.
5.1 Overview 20
5.2 Analyzer Overview 21
5.3 Key Navigation 22
5.4 Displays Overview 23
5.5 Input Displays 25
5.6 PID Displays 26
5.7 Auto Cycle Displays 28
5.8 Pharma Display 33
5.10 Status Display 41
5.11 Event History 46
5.12 Process Instrument Explorer Software 48
Operating the Analyzer · Topic See Page

The UDA2182 Universal Dual Analyzer is the next level of dual channel analyzers providing unprecedented versatility and flexibility.
The analyzer can accept single or dual inputs from Honeywell Direct pH, pH from preamp, ORP (Oxidation Reduction Potential), Contacting Conductivity and Dissolved Oxygen sensors.
Measurement for Dual channel units can be arranged in any combination of measurement.
A Communications card provides one Serial Port (RS485) and one Ethernet Port.
Operating the Analyzer

Table 5-1 shows each key on the operator interface and defines its function.

To view display screens, push the
Display key. Pushing the Display key repeatedly scrolls through screens which show the current status of pH/ORP, Conductivity, or Dissolved Oxygen Concentration. There are displays for PID, Auto Cycle, and Pharma. It also lets you view a Status Display and an Event History Display.
One Input - When only one input board is installed, the online screen displays one PV and its units in a larger font size (Section 5.5). Two Inputs - When two input boards are installed, the online screen displays two PVs and its units in a smaller font size. Press to see single PV screens (Section 5.5). PID - When PID 1 or 2 is active (Section 5.6 ), there is a display screen for each. There is a sub- screen that allows editing of the Setpoint value, Setpoint Source, Control Mode, and Output value. You can also enable or disable Accutune and Tune set. Auto Cycle – When Auto Cycle 1 or 2 is active (Section 5.7), there is a display screen for each. There is a sub-screen that allows you to start or stop the Cycle. Pharma – Enabled in Conductivity inputs. Each Pharma screen monitors standard procedure stages for determining Purified Water. There is a sub-screen that allows you to change the Pharma Test Stage and adjust the Pct Warning value (Stage 1), Test µS/cm value (Stage 2) and the Test pH (Stage 3). (See Section 5.8 for details)
Cation Calc – When cation Calc 1 or 2 is active there is a display screen for either cation or degassed CO2 measurement. (Section 5.9) Status Display - of Alarms Status, PID Alarms Status, Logic Status, Input Status, Output Levels, Relay States, Monitor Status, Math Values, Aux Values, Variables, Comm Status, System Status, and Calculated Values. (Section 5.10) Event History - Event History records events with timestamp. (Section 5.11) Events recorded include setup change, power on, calibrations (no values) and alarms with detail available on alarm type and source by scrolling and selecting event name. Status warns of event history at 50% and 90% and when erasing old records. Contrast Adjustment
When viewing a PV or Control display, you can adjust the contrast by pressing the or key.
Output Bargraphs will represent up to three current output values. On the display, the Bargraphs are the output in Engineering Units. The corner annunciators are the physical relay states (light – de-energized, dark – energized). The third output and the 3 and 4 relays are shown only when the source other than NONE is selected.
An upward-pointing arrow indicator above the menu at the left end of the header appears when there are currently menu items above the screen accessible by moving the cursor up. A downward-pointing arrow indicator below the menu at the left end of the status footer appears when there are currently menu items below the screen accessible by moving the cursor down.
Use the keys.
Operating the Analyzer · Bargraphs Overview

Table 5-2 Display Details Functions
When two input boards are installed, the default online screen displays both PVs and their units of measure, as determined by the input boards, the probe (if memory- embedded) or any measurement configuration options that may be available. When only one input board is installed, the default online screen displays one PV and its units in a larger font size.
The currently selected PV type determines the numerical format and the units of measure on the online PV display. Measured PV is generally displayed in the highest decimal precision possible with five digits and has a potentially displayable range of 0.0000 to 99999. The exceptions are dissolved oxygen, pH, ORP and temperature, which are displayed with fixed decimal precision.
PV Type determines specific ranges and in the case of Conductivity, cell constant determines available PV Types. Each PV measurement and display is updated every 500ms maximum. Each temperature measurement and display is updated every 10 seconds maximum.
See the Specific Input configuration for available ranges. (Section 6.6)
Tag Name The real-time displays of process values show the instrument’s tag name (or other configurable fixed sixteen-character string) at the top of the screen.
PV Temperature Each PV value is accompanied by a temperature value for all measurements except ORP, as ORP probes do not contain temperature sensors and no measurement compensation for temperature is required. Temperature values are displayed in units of degrees Fahrenheit or degrees Celsius as determined by configuration.
Measured temperature is always expressed in fixed tenths decimal precision and has a displayed range according to input type:
PH/ORP -10.0 to 110.0°C or 14.0 to 230.0°F
Conductivity 0 to 140.0°C or 32.0 to 284°F
Dissolved Oxygen 0 to 60.0°C or 32 to 140°F Status Messages A text string appears on the bottom of all displays. Online displays provide messages relaying online diagnostics, alarms and other events. Offline screens display messages relevant to data entry and calibration. See Section 12.
Bargraphs The Bargraphs will represent up to three output values. The corner indicators represent the physical state of the Relay Outputs [1, 2, 3, and 4].
*Note that all values and indicators on the main (input) display screen are maintained in the input setup group.

Display. You will see:

*On the display, the bargraphs are the outputs in Engineering Units, the corner annunciators are the physical relay states.
*On the display, the bargraphs are the outputs in Engineering Units, the corner annunciators are the physical relay states.
Single Displays
For single displays on a two input unit;
Press to display a single display for Input 1.
Press again to display a single display for Input 2.
Press again to return to a Dual Display.
Operating the Analyzer · Press

When PID 1 or 2 is active - there is a display screen for each. There is a sub-screen that allows editing of the Setpoint value, Setpoint Source, Control Mode, and Output value. You can also enable or disable Accutune and Tune Set.
Display until you see the PID Display screen. If PID 1 and 2 have been configured, press DISPLAY again. In each instance, you can edit some control parameters. See Table 5-3.
PV Units PID Loop
PV Units PID Loop
1 or 2 PV1 or PV 2 Value Control Mode Auto or Manual*
1 or 2 PV1 or PV 2 Value Control Mode Auto or Manual*
Relay 1 Physical State
Relay 1 Physical State
PID LOOP 1
Relay 3 Physical State
Relay 3 Physical State
1 2 0.00 0.00
Relay 2 Physical State
Relay 2 Physical State
Relay 4 Physical State
Relay 4 Physical State
Working Setpoint Value* *These Control parameters can be edited. See Table 5-3.

When either PID Display is on the Display screen, you can edit the Setpoint value, Setpoint Source, Control Mode, and the Output value. You can also enable or disable Accutune and Tune Set.
Enter to access the PID Parameters. You will see:
PID LOOP 1
LSP
LSP
SP Source Local SP
SP Source Local SP
Manual 0.00
Manual 0.00
Tune Set 2 Disable
Tune Set 2 Disable
Example – PID Loop 1 Edit Display
to highlight the parameter you want to change.
Enter to access the value or selection of each.
to change the value or selection. Note: Output can only be changed in Manual mode. Refer to “Section 6.4.1 – ”General Rules for Editing”.
Enter to make the edit permanent.
Display to return to the selected PID Display.

Auto Cycling allows each input probe to be automatically rinsed and calibrated on a recurring schedule, in response to an event, or on demand.
Auto cycling is supported with Setup Menus ( Section 6.15- Auto Cycle Configuration), Status Displays (Section 5.10 – Status Display) and Operational Displays (Section 5.7 as well as Event History (Section 5.11) and Calibration History logging (Section 11).
Display until you see:
Current Clock Time or remaining Elapsed Time Auto Cycle 1 or 2 Sequence
Relay 1 Physical State
Relay 3 Physical State
12:48:27
µS/cm 27.2 ºC
Next 06-05-01 12:00:00PM Date and Time of next cycle;
Relay 2 Physical State
Relay 4 Physical State
* See Table 5-5.

Each occurrence of a sequence of actions for the cleaning and calibration of a probe is a cycle. The status message “Auto Cycle n Active” appears for the duration of a non- failing cycle (where n refers to 1 or 2). Each cycle consists of the following sequence that will vary depending on the input type and parameters selected:
Probe Extract (if enabled)
Cal pt 2 (If the input is pH)
Probe Insert (If enabled)
Resume Delay (If a time is selected)
For a more detailed explanation please refer to Table 5-5.
The current sequence step is shown in the upper half of the Auto Cycle display.
When Cycle Stop is displayed, the field to the right displays the current clock time in the format configured in Setup/Maintenance/Clock.
When the cycle is active, the same field provides the remaining elapsed time for the current sequence step.
The lower half of the Auto Cycle display provides either: the date and time of the next cycle, an indication that the cycle timer is off; or an auto cycle fail detail message.

If Hold Active is enabled in Auto cycling Setup, then the values remain in the hold state during auto cycle. When Hold is active on either input, the status message “Hold Active” is displayed and the specific PV value flashes at a very slow rate.
When Hold is activated manually from the front panel Hold button, the values remain in the hold state until its state is changed via the front panel again.
This parameter is available to allow you to automate functions that relate to probe removal and insertion. Once the probe transit parameter is enabled, the extract wait source, insert wait source and max transit mins can be selected. The extract wait source and insert wait source can be set to any digital input so that the extract or insert operations continue until the selected digital input is low. Once the wait source signal is low then the probe extraction or insertion sequence step can end or otherwise time out if the duration of the “max transit mins” is exceeded. If probe transit is enabled and probe extract src/insert src is set to none, then the probe extract/insert step will occur for the duration of the max transit mins.
The Auto cycle can be started in one of three ways. It can start upon the occurrence of a specific digital input changing state from low to high. The cycle can also start when the cycle timer engages. The cycle can also be manually started from the Auto Cycle display screen op panel.
This parameter enables the cycle timer and allows you to set the Auto cycle to recur at a period defined by you. If the cycle interval is set to Custom, Monthly, Weekly or Daily then specific menu items are activated to set-up cycle start and period times. You can select a cycle interval appropriate to the application.
This parameter allows you to select when or if a rinse sequence occurs during a cycle. A selection of 0 indicates that a rinse sequence will not occur. A selection of 1 indicates that a rinse sequence occurs during every cycle. There is an option to set the rinse sequence for less frequent times by selecting values from 2 to 100. For example; a selection of 3 means that the rinse sequence will occur every 3rd cycle.
This parameter allows you to select the duration of a rinse.
This parameter allows you to specify a delay time before the cycle is completed.
Operating the Analyzer

Pressing Enter on the Auto Cycle Operational Display brings up an operator panel menu that enables you to manually start or stop an auto cycle sequence or place the cycle in Hold, regardless of whether or not the cycle timer is configured. Start cycle is visible when the Auto Cycle is not active and Stop Cycle is visible during an auto cycle.
The objective of a Cycle Stop is to terminate the Auto Cycle by putting the probe back into the process. The selection of Cycle Stop causes the sequencer to proceed directly to the Probe Insert stage, if Probe Transit is enabled, and then to the Resume Delay stage. Selecting Stop Cycle during Probe Insert advances the sequencer to resume delay. Finally, a Stop Cycle selection during Resume Delay will stop the cycle completely.
If enabled, the cycle timer will still trigger a cycle at the configured time provided the sequencer returns to Cycle Stop beforehand. Otherwise, the cycle will execute at the next available time.
The Hold Cycle selection is available in both auto cycle active and inactive states. When inactive, enabling Hold Cycle will hold the sequencer at the beginning of the sequence until released to continue by disabling the Hold Cycle.
Selecting Hold Cycle during an active cycle will suspend the sequencer at the current step as well as the step timer. Disabling Hold Cycle will resume the sequencer and step timer. During cycle Hold, the status bar will show the message “Auto Cycle in hold”.
Enter to access the Auto Cycle Operator Panel. You will see:
AUTO CYCLE 1
AUTO CYCLE
AUTO CYCLE 1
12:48:27
12:48:27
1 2 0.00
No Hold Cycle 1 2 0.00
1 2 0.00
µS/cm 27.2 ºC
µS/cm 27.2 ºC
µS/cm 27.2 ºC
µS/cm 27.2 ºC
Next 06.03.29 13:50:24
Next 06.03.29 13:50:24
Example – AUTO CYCLE 1 Operator Panel
To access Start Cycle or Hold Cycle
Enter to access the Start Cycle/Hold Cycle selection.
to change the selection from NO to YES.
Enter to Start or Hold the Auto Cycle.
See Table 5-5 for sequencer steps.

The status message “Auto Cycle n Fail” is displayed during a fail state. Once detected, the current cycle proceeds immediately to the Probe Insert step (if enabled) or to the Resume Delay step. The fail state remains for the duration of the Resume Delay, whereupon the fail state returns to zero and the fail message is cancelled.
A fail state also provides a detail message in the lower half of the Auto Cycle display regarding the specific reason for the error. These messages are listed in Section 12.4.
The digital output “AC n Fail” is also available and is active whenever an auto cycle failure has occurred. The auto cycle digital outputs: AC n Extract, AC n Rinse, AC n Cal Pt 1 and AC n Cal and AC n Cal 2 are available as relay digital input sources (See Table 6-4) to control the operation of valves and solenoids for exposure to rinse water and buffer solutions or air to accomplish the automatic probe rinse and calibration functions.
Table 5-5 Conditional Sequencer Steps for Auto Cycle
Step AC n Extract
AC n Cal 2
Cycle Stop 0 Inactive Cycle inactive Cycle Start (transitional)
0 Active Operator panel Cycle Start is Yes or Cycle Start Src state is 1 or cycle timer engages Probe Extract 1 0 0 0 Active Probe Transit is enabled Probe Rinse 1 1 0 0 Active Rinse Cycle Cnt > 0 and enables current cycle and if Probe Transit enabled, Extract Wait Src is None or state is 1 or Max Transit Timer expires Cal Point 1 0 Active Rinse timer expires if enabled and Cal 1 (Cal) Cycle Cnt > 0 and enables current cycle Cal Point 2 1 1 0 1 Active Rinse timer expires or Cal 1 complete and Cal 2 Cycle Cnt > 0 and enables current cycle Probe Insert 0 1 0 0 Active Auto Cycle Fail or Probe Transit is enabled Resume Delay 0 Active Auto Cycle Fail or Rinse timer expires or Cal 1 or Cal 2 complete and Insert Wait Src is None or state is 1 or Max Transit Timer expires

The Pharma Parameter is available when a Conductivity Input is enabled. Pharma supports USP (United States Pharmacopoeia) and PhEur (Pharmacopoeia Europa) standard procedure stages for determining Purified Water.
Selecting Pharma Type USP or PhEur (in Section 6.6 – Conductivity Input Configuration) enables the Pharma monitor screen and adds it to the sequence of displays accessed by each press of the Display button.
Also, configure Pharma PV High, Pharma PV Low, and Pharma Timer Minutes in this section.
Pharmacopoeia Test Procedure
For Procedure steps in each stage, refer to UPS section Test Procedure for Purified Water and Water for Injection. The procedure for this determination involves a series of three stages or tests. If the sample does not pass the Stage 1 conductivity requirement, the State 2 test can be initiated. If the Stage 2 requirements are not met, then the Stage 3 test can be initiated. If Stage 3 requirements are not met, the sample is not Purified Water.
In Stage 1 the non-temperature-compensated conductivity reading is compared to the value specified in the USP standard for a particular temperature. If the measured conductivity is not greater than the table value, the water meets the requirements of the test for conductivity and the Pharma test is complete. If the conductivity is higher than the table value, then the user can manually proceed with Stage 2.
To complete stage 2, transfer a sufficient amount of water (100 mL or more) to a suitable container, and stir the test specimen. While maintaining the temperature at 25° ± 1°, begin vigorously agitating the test specimen and note the conductivity reading when the change is less than 0.1 m S/cm per 5 minutes. If the conductivity is not greater than 2.1 m S/cm, the water meets the requirements of the test for conductivity. If the conductivity is greater than 2.1 m S/cm, proceed with Stage 3.
Stage 3 must be completed within approximately 5 minutes of the conductivity determination in stage 2. While maintaining the sample temperature at 25° ± 1°, add a saturated potassium chloride solution to the same water sample (0.3 mL per 100 mL of the test specimen), and measure the pH to the nearest 0.1 pH unit, as directed under pH (791). From USP section , determine the desired conductivity value for a pH value between 5pH and 7pH. If the measured conductivity is not greater than the conductivity from USP, section , the water meets the requirements for stage 3 - purified water. If either the measured conductivity is greater than this value or the pH is outside of the range of 5.0 to 7.0, the water does not meet the requirements of stage 3 - purified water.
For Procedure steps in each stage, refer to UPS section Test Procedure for Purified Water and Water for Injection.
Operating the Analyzer

The UDA also supports Pharma Europa (PhEur) section 2.2.38, which specifies tests for determining Highly Purified Water which are identical to USP Stages 1, 2 and 3. PhEur adds a less demanding test for determining Purified Water at the end of the sequence.
Display until you see:
The measured temperature- uncompensated conductivity
The measured temperature- uncompensated conductivity
of the solution in µS/cm
of the solution in µS/cm
Relay 3 Physical State
Relay 3 Physical State
Relay 1 Physical State
Relay 1 Physical State
PHARMA 1
Left Bar graph -Measured solution conductivity scaled by the parameter values Pharma High and Pharma Low
Left Bar graph -Measured solution conductivity scaled by the parameter values Pharma High and Pharma Low
of solution in system
of solution in system
25.7 ºC
Right Bar graph – Graphical percent of measured conductivity relative to test limit 1.30
Right Bar graph – Graphical percent of measured conductivity relative to test limit
Stage 1 74.3%
Left Pointer - Test Limit scaled by parameters Pharma High and Pharma Low
Left Pointer - Test Limit scaled by parameters Pharma High and Pharma Low
4 2 The percent of measured conductivity relative to test limit
4 2 The percent of measured conductivity relative to test limit
Relay 2 Physical State
Relay 2 Physical State
Relay 4 Physical State
Relay 4 Physical State
limit in units of measured value
limit in units of measured value
5.8.4 Displays
The upper left portion of the screen shows the measured temperature uncompensated conductivity of the solution in uS/cm for Stage 1. For stage 2 and 3 it displays the test conductivity value entered during the measurement in stage 2
The process temperature is shown in the upper right of the screen
The lower right portion of the screen shows the USP stage 1 purified water limit, the stage 2 limit of 2.1 uS/cm and the stage 3 purified water test limit at the measured pH.
When the Pharma n display window is active the various stages can be accessed through the pharma op panel. (Table 5-6)


Enter to access the Pharma Op Panel. A pop-up dialog box will appear:
80.00 PCT Warn Various parameters appear for each stage:
Pharma Test Selection Stage 1 Stage 2 Stage 3 Pure H2O (PhEur only)
Pct Warning (% at which the test warn occurs)
Test µS/cm 0 to 10 (default=10) Stage 2 only
Test pH 0 to 14 (default=0) Stage 3 only
to change the selection to Stage 1, Stage 2, Stage 3 or Pure H2O (PhEur only).
Enter to access the Pct Warning Value and allow editing
to select Test µS/cm (After Stage 2 is selected)
to select Test pH (After Stage 3 is selected)
Operating the Analyzer

The Pharma 1 warning limit is entered from the op-panel for stage 1 and is user selectable. The digital output “Pharma n Warn” is available (See Digital Source Selection - Table 6-4).
The Pharma Fail signal is generated whenever any of the following conditions are met:
Stage 1 – Measured Conductivity exceeds 100% Stage 1 – Temperature not within range of 0-100 degrees C Stage 2 – Conductivity is 0.1 µS/cm or greater for 5 minutes Stage 3 – pH not within range of 5 – 7pH Stage 2 and 3 – Temperature not within range of 24 – 26 degrees C.
The digital output “Pharm n Fail” is available (See Digital Source Selection - Table 6-4).
When the Stage 2 or Stage 3 test is successful, the fail signal is cancelled and the Pharma Timer begins to count down from the timer minutes value that was configured. When the Timer countdown is completed, the Pharma function block returns to Stage 1. A fail signal will return if measured conductivity exceeds 100% or warn signal if measured conductivity exceeds Pct Warning value.
See Section 12.5 for Pharma Fail Messages.
Operating the Analyzer

This group allows you to configure dual conductivity inputs for cation or degassed CO2 measurement. The cation selection of Ammonia or Amines will display a calculated pH value from differential conductivity and provide continuous pH monitoring using reliable, maintenance free conductivity cells. An outline of the conductivity cells’ installation is illustrated in Figure 5-6.

Specific Conductivity From Cell #1
Specific Conductivity From Cell #1
Specific Conductivity From Cell #1
Cation Conductivity From Cell #2
Cation Conductivity From Cell #2
Cation Conductivity From Cell #2
Degassed Conductivity From Cell #3
Degassed Conductivity From Cell #3
Degassed Conductivity From Cell #3
UDA for Cation Conductivity UDA for Degassed CO2
UDA for Cation Conductivity UDA for Degassed CO2
UDA for Cation Conductivity UDA for Degassed CO2
5.9.2 How it works
UDA2182 will monitor on-line treated water for Specific Conductivity and Cation bed discharge conductivity, using reliable and maintenance free Honeywell conductivity cells, and calculate the pH, using the assumption that the water is pure water with Amine type treatment chemical and residual trace un-removed salts.
The equipment consists of Cell #1 which is used for specific conductivity determination of the influent water sample. The water sample is then passed through a strong acid cation exchange resin which replaces all cations in the influent stream with hydrogen ion. On passing though the resin, a second Cell #2 is used to measure the effluent or cation conductivity. These two measurements in combination are useful for measuring the amount of base contained in the influent sample and the extent of contamination by unwanted salts such as sodium chloride, sodium sulfate, etc.

The dual input UDA can also be configured for degassed CO2 measurement by employing cells #2 and #3. Here the cation effluent stream is degassed of CO2, typically by heating the cation effluent stream to a near boiling temperature. This heating step results in CO2 out-gassing. The resulting 25°C compensated conductivity measurement of Cell #3 is lower in value in proportion to the amount of dissolved CO2.
5.9.3 pH Calculation from Specific and Cation Conductivity Setup Connect cell 1 to input 1 and cell 2 to input 2. Follow the appropriate instructions to configure the UDA for Cation Calc (See Section 6.18 – Maintenance Configuration.) Under the sub menu selection of Inputs, the Cation Calc parameter offers two possible selections.
Hydrazine Morpholine Ethanolamine Aminomethylpropanol Methoxypropylamine 4-aminobutanol 5-aminopentanol Diaminopropane Cyclohexylamine Methylamine Dimethylamine 1,5-diaminopentane Piperidine Pyrrolidine This generic selection employs ammonia temperature compensation and optimizes pH calculation for these base reagents.
Operating the Analyzer

Note: The relationship between the electrolytic conductivity and the pH of ammonia and amines is well established in the technical literature. It must be understood that the UDA was designed for accurate results over the pH range of 8 to 10.5 based on ammonia or amine chemistries. Other chemistries such as phosphate or systems that employ alternative anions, such as borate, cannot be expected to realize results with similar accuracy.
The UDA allows for a sample calibration of the cation pH value. Here an independent sample is withdrawn from the sampling equipment and pH is determined with equipment of known accuracy. This independent pH value is then entered into the UDA as a pH calibration constant. To avoid process pH changes during standardization, it is very desirable to complete the sample extraction, independent measurement and UDA update as soon as possible.
For Calibration procedure, refer to Section 8.7.7 Cation pH Calibration.
The UDA can be configured for CO2 determination by degassed conductivity. The cation conductivity cell is connected to Input 1 and the degassed sample conductivity cell is connected to channel 2. The UDA performs HCl temperature compensation of both measurements to 25°C. The difference between the cation and degassed 25°C values is taken and ppb CO2 is determined by ASTM D 4519.
Display until you see:
PV 1 Temperature PV 1 Temperature
Relay 3 Physical State
Relay 3 Physical State
Relay 1 Physical State
Relay 1 Physical State
CATION CALC
3.772 µS/cm
3.772 µS/cm
Calculated Water pH 28.4 ºC
Output 1 Bar graph Output 3 Bar graph
Output 1 Bar graph Output 1 Bar graph Output 3 Bar graph
Output 2 Bar graph
Output 2 Bar graph
0.099 µS/cm
0.099 µS/cm
24.1ºC
PV 2 Temperature PV 2 Temperature
PV 2 Value PV 2 Value
Relay 2 Physical State
Relay 2 Physical State
Relay 4 Physical State
Relay 4 Physical State

In normal operation, both the direct electrode pH measurement and the pH from differential conductivity should very closely match each other.
If they DO NOT match each other, the possible causes are listed below:
Operating the Analyzer

The Status Displays let you see the status of the Alarm Status, PID Alarm Status, Logic Status, Input Status, Output levels, Relay states, Monitor Status, Math Values, Aux Values, Auto Cycling, Variables, Comm Status, System Status, and the Calculated values (Calc Values available only if both units of measurement are identical).
STATUS DISPLAY
STATUS DISPLAY
STATUS DISPLAY
STATUS DISPLAY
STATUS DISPLAY
STATUS DISPLAY
STATUS DISPLAY
Logic Status Input Status
Logic Status Input Status
Logic Status Input Status
Logic Status Input Status
Logic Status Input Status
Logic Status Input Status
Logic Status Input Status
Logic Status Input Status
Relay States Output Levels
Relay States Output Levels
Relay States Output Levels
Relay States Output Levels
Relay States Output Levels
Relay States Output Levels
Relay States Output Levels
Relay States Output Levels
Monitor Status Math Values Aux Values
Monitor Status Math Values Aux Values
Monitor Status Math Values Aux Values
Monitor Status Math Values Aux Values
Monitor Status Math Values Aux Values
Monitor Status Math Values Aux Values
Monitor Status Math Values Aux Values
Monitor Status Math Values Aux Values
Auto Cycling (if configured)
Auto Cycling (if configured)
Auto Cycling (if configured)
Auto Cycling (if configured)
Calc Values (if configured)
Calc Values (if configured)
Calc Values (if configured)
Calc Values (if configured)
Table 5-7 Status Display Details
Parameter Status (Read Only)
Alarm Status Alarm 1 Alarm 2 Alarm 3 Alarm 4
ON = Latching Alarm in alarm.
Acknowledge alarm by changing status to OFF.
If status changes back to ON, alarm condition still exists.
PID 1 Alm 1 PID 1 Alm 2 PID 2 Alm 1 PID 2 Alm 2
ON = PID Alarm Active
PID alarms are not latching.

Parameter Status (Read Only)
Logic Status Logic 1 Logic 2 Logic 3 Logic 4
Input Status In 1 Fault In 2 Fault
OK or Fail OK or Fail
Read Only – depends on the Input selected
Digital In 1 Digital In 2
On or Off On or Off
Output 1 mA Output 2 mA Output 3 mA
Output Level in Milliamps
Read Only – depends on the Output type selected at setup “Outputs”, “Output n”, “Source”: None Input 1 PV Input 2 PV Input 1 Tmp Input 2 Tmp Pharma Out 1 Pharma Out 2 Math 1 Math 2 Math 3 Math 4 Func Gen 1 Func Gen 2 Switch 1 Switch 2 Sum Difference Ratio %Passage %Rejection PID 1 PID 2
See Table 6-6 for configuration.
Relay States Relay 1 Relay 2 Relay 3 Relay 4
State of the relay
Read Only – state depends on the Output source selected at Relay Setup Group, parameter “Relay Types”: Digital Output (On or Off) Time Proportional (Value) Frequency Proportional Output (Value) On/Off (On or Off) Pulse Out (On or Off)
See Table 6-7 for configuration.
Monitor 1 Monitor 2 Monitor 3 Monitor 4
Read Only – State depends on the output of the analog monitor blocks.

Parameter Status (Read Only)
Math Values Math 1 Math 2 Math 3 Math 4
Analog Values Read Only – Shows the calculated values of the Math blocks.
Aux Values Switch 1 Switch 2 Func Gen 1 Func Gen 2
Analog Values Read Only – Shows the calculated values of the blocks in the Aux Group. This includes the current output of the Switch and Function Generator blocks.
Variables AnlgVar 1 AnlgVar 2 AnlgVar 3 AnlgVar 4
Read Only – shows values of Analog and Digital variables written from Modbus client.
Dgtl Var 1 Dgtl Var 2 Dgtl Var 3 Dgtl Var 14
Digital Values (On or Off)
Read Only – This displays the status of the Communication card. Information present only if the Communication card is present.
Ok/ Not Present/ HW Failure/ Fail Init
Comm Card Stat gives the status of the communication card
Status shown OK if the communication card is working fine.
Status shown as Not present if the communication card is not present.
Status shown as HW Failure if the communication card is installed but unable to communicate to the Main CPU board.
Status shown as Fail if the communication card is not functioning properly. It could be the result of a software failure, a bad flash chip on the board or DHCP is selected but the DHCP server was not found. Check cable connections and potential network issues for DHCP related problems.
Status shown as INIT if the communication card is getting initialized.
SW Version Value SW Version gives the software version of the Communication Card.
Web Page Value Web Page gives the web page version number. The web pages are separate from the Communication card firmware, and can be upgraded independently. Both the SW version and Web Page version should be the same value to guarantee compatibility.

Parameter Status (Read Only)
WebPgLngSet EE/RT/PC Identifies the web page language set programmed into the Communications card. EE web pages support English, French, German, Italian and Spanish. RT web pages support English, Russian and Turkish PC web pages support English, Polish and Czech
Address Value Address states the Modbus RTU slave ID
Baud Rate Value Baud Rate set for RS485
Word Swap Yes/No Word Swap indicates whether the word order set for Modbus Communications is Little Endian (NO) or Big Endian (YES)
MACaddr Hi MACaddr Low
Value MACaddr Hi and MACaddr Low is the MAC address of the Communication card
DHCP Yes/No DHCP indicates whether Dynamic Host Configuration Protocol is used. DHCP is a protocol used by network devices (clients) to obtain various parameters necessary for the clients to operate in an Internet Protocol (IP) network.
IPaddr Value IPaddr gives the IP address of the Communication card.
SubnetMsk Value SubnetMsk indicates the Subnet mask used by the Communication card.
Gateway Value Gateway Indicates the default Gateway IP address used by the Communication card.
DnsSrvr Value DnsSrvr displays the DNS (Domain Name Service) server IP address used by the Communication card
SMTPsrvr Value SMTPsrvr displays the SMTP (Simple Mail Transfer Protocol) server IP address used by the Communication card.
Next Rinse 1 Next Rinse 2 Next Cal 1 Next Cal 2
Date and Time Read Only – The Status Displays menu includes an Auto Cycling selection when any auto cycle is enabled and its cycle timer is also enabled. This display provides information on the next occurrences of rinses and calibrations for any auto cycle with timer according to the configured cycle count of each operation.

Parameter Status (Read Only)
Value Available only if both units of measure between the two input boards are identical.
See Table 6-5 for configuration.
Sum = Input 1 + Input 2
Difference = Input 1 – Input 2
Ratio = Input 1 / Input 2
%Passage = Min(Input 1 or 2)/Max(Input 1 or 2) *100
%Rejection = (1-Min(Input 1 or 2)/Max(Input1 or 2))*100

Event History records events with timestamp. Events recorded include setup change, power on, calibrations (no values) and alarms with detail available on alarm type and source by scrolling and selecting event name. Status warns of event history at 50% and 90% and when erasing old records.
EVENT HISTORY
Setup Chg Alarm 1 On
Setup Chg Alarm 1 On
04.19 08:58
04.19 08:58
03.15 13:02
03.15 13:02
03.15 12:38 Setup Chg 03.15 13:01
03.15 12:38 Setup Chg 03.15 13:01
Setup Chg 03.15 11:21 Power On 03.09 02:31
Setup Chg 03.15 11:21 Power On 03.09 02:31
HOLD ACTIVE
Event History Display Example (Alarm)
ALARM 1 ON
ALARM 1 ON
Time 13:02:13
Time 13:02:13
ALARM 1 INPUT 1 PV HIGH
Figure 5-10 Alarm Event Display screen example (Read Only)

Enter to enter the sub- menu.
Enter to enter the sub-menu.
Enter to allow change.
Enter to clear the Event History screen.
Operating the Analyzer

Process Instrument Explorer lets you configure your analyzer on a desktop/laptop or Pocket PC. For details see Process Instrument Explorer manual #51-52-25-131.

Pocket PC
Operating the Analyzer

The infrared connection provides a non-intrusive wireless connection with the instrument and maintains its waterproof integrity when used in conjunction with the optional PIE (Process Instrument Explorer Software).
No need to get access to the back of the analyzer to communicate with the instrument, no need to take your screw driver to wire the communication cable, no wiring mistake possible. You can now duplicate an instrument’s configuration, upload or download a new configuration in a matter of seconds, just by pointing your Pocket PC in the direction of the instrument.
It takes just a few seconds to upload a configuration from an instrument. You can then save the configuration file onto your PC or pocket PC for review, modification or archiving. Furthermore, this software also gives you important maintenance information on the analyzer: instantly, get information on the current operating parameters, digital inputs and alarm status, identify internal or analog input problems.
Question: What if I have several analyzers on the same panel? How can I be sure I am communicating with the correct one?
Answer: The infrared port of the analyzer is normally “off”. You activate the infrared port on a particular analyzer by pressing any key. You can now communicate with the analyzer. If no communications are received for 2 minutes, the port will be shut down again.
Operating the Analyzer

The UDA2182 provides Modbus communication support on two communication interfaces using the optional Communication Card. A general summary is listed below. For details see UDA2182 Communications User Guide #70-82-25-126.
The Communications card provides one Serial Port (RS485) and one Ethernet Port.
There are four analog and four digital variables. These variables can be read and written remotely using Modbus function codes. Variables will appear as a selection for various parameters:
Web pages provide the following:
Operating the Analyzer · Communications Card (Optional) · Serial port provides

Configuration is a dedicated operation where you use straightforward keystroke sequences to select and establish (configure) pertinent setup data best suited for your application.
To assist you in the configuration process, there are prompts that appear in the Main Setup menu and associated sub menus. These prompts let you know what group of configuration data (Set Up prompts) you are working with and also, the specific parameters associated with each group.
The topics in this section are listed below.
6.1 Overview 51
6.2 UDA2182 Block Diagram 52
6.3 Main Setup Menu 53
6.4 Basic Configuration Procedure 55
6.6 Inputs Configuration 63
6.7 Outputs Configuration 74
6.8 Relays Configuration 75
6.9 Alarms Configuration 81
6.10 Monitors Configuration 83
6.11 Math Configuration 85
6.12 Logic Configuration 86
6.13 Auxiliary Configuration 89
6.14 PID Control Configuration 92
6.15 Auto Cycling Configuration 100
6.16 Variables Configuration 105
6.17 Communication Configuration 106
6.18 Maintenance Configuration 108
Configuration · Topic See Page

Calc Values (Dual Input
B
% Rejection % Passage
PV
PV
PID (1-2) PID Opt Installed
Alm 1 Alm 2
Figure 6-1 UDA2182 Block Diagram

Setup. The main Menu will appear.
Setup Inputs
Setup Inputs
Setup Inputs
Setup Inputs
Setup Inputs
Setup Inputs
Setup Inputs
*Some item are dependent on the Option selection
An upward-pointing arrow indicator above the menu at the left end of the header appears when there are currently menu items above the screen accessible by moving the cursor up.
A downward-pointing arrow indicator below the menu at the left end of the status footer appears when there are currently menu items below the screen accessible by moving the cursor down.
Use the keys.
Refer to “General Rules for Editing” and Table 6-1 Basic Configuration Procedure to configure the following Setup Groups.
Inputs Configuration (Table 6-5) – configure: Input 1 and Input 2 for pH/ORP, pH Preamp, Conductivity, or Dissolved Oxygen and associated parameters Calc Value 1 and 2 (both units of measurement must be the same) select the Calculation type [Ratio, sum, etc.], High range and Low range. Outputs Configuration (Table 6-6) – configure Output 1, 2, or 3 source, % Range High and Low and associated parameters
Relays Configuration (Table 6-7) – configure Relay 1, Relay 2, Relay 3, and Relay 4 for Relay Types: Digital Out (Digital Output Relay), Time Prop (Time

Proportional Output), Pulse Frequency (Pulse Frequency Type), Frequency Prop (Frequency Proportional), or On/Off type and associated parameters.
Alarms Configuration ( Table 6-8) - configure Alarm 1 through 4 for Alarm’s Source and associated parameters.
Monitors Configuration (Table 6-9) – configure Monitor 1 through 4 for Monitor Type, Source and associated parameters.
Math Configuration (Table 6-10) – configure Math 1, 2, 3, and 4 for Input Source, Math Type, and associated parameters.
Logic Configuration ( Table 6-11) – configure Logic 1, 2, 3, and 4 for Input Sources, Type, and associated parameters.
Auxiliary Configuration (Table 6-12) – configure Switch 1, Switch 2, Function Generator 1 and Function Generator 2(for pre-control linearizing of inputs) for Sources and associated parameters.
PID Control (Option) Configuration (Table 6-13) – configure:
PID 1 and PID 2 Configuration parameters, Tune (Enable Accutune, Fuzzy Logic, Use Prop Band, Use RPM, configure Tuning parameters) and Alarms Parameters (Setpoint types and Values, alarm hysteresis)
Auto Cycling (Table 6-16) – enable Auto Cycle 1 and 2 and set rinse schedule and associated parameters. Auto cycling provides automated timing, control and functionality for the cleaning and calibration of input probes.
Variables (Analog Table 6-3 and Digital Table 6-4) selections can be read and written remotely using Modbus function codes. You are setting up the initial values for the variables when power is applied to the UDA (Refer to Table 6-18 for an Example)
Communication Configuration (Table 6-19) – configure IR Front Panel, Modbus, RS485 and Ethernet. Maintenance Configuration (Table 6-20) – Configure:
System - read the Software version, configured Language, selected Mains Frequency, and PID Control Selections, enter a Password, and reset the Unit.
Input 1 and Input 2 - configure Input types, Conductivity units’ type, wire size, and wire length, and temperature Units.
Display – setup the Main Display Header; and Clear Event and Cal Histories.
Tag Names – configure tag name strings for input names on single channel main display, Auto Cycle display header, Pharma display header, PID display header. Alarm names to appear in status and event history.
Clock - set real time clock date; time; date format; and time format.
Tests - run Display and Keyboard tests, read Output levels, and read Relay States.
Configuration

Each of the Set Up groups and their functions are pre-configured at the factory. If you want to change any of these selections or values, read the “General Rules for Editing” and follow the procedure in Table 6-1. This procedure tells you the keys to press to get to any Setup group and any associated parameter prompt. 6.4.1 General Rules for Editing
If you hold down the key, the cursor will move to the left and increment to the next highest digit available for the particular parameter. If you hold
down the key, the cursor will move to the right and increment the next lowest digit available for the particular parameter.
Enter to store.
Configuration

Table 6-1 Basic Configuration Procedure

The Main Menu is displayed. Use to scroll and select a setup group (Example – Inputs). The selection will be highlighted.
The Setup group selected is shown at the top of the screen and will display all the selections within that group.
Press to highlight the desired selection. (Example – Input 1 PRE PH)
The list of parameters for that selection will be displayed.
The displayed current value for the parameter is displayed.
Depending on whether you are changing a text string or a numerical value, follow the “General Rules for Editing” in section 6.4.1 to make the changes.
Configuration

Enters value or selection made into memory after another key is pressed.
Repeat the procedure for changing any parameter for any group.
Any changes made to a parameter value will revert to the original value before editing.
Until you see the main Setup screen.
Configuration

This section contains a list of signals that are available for connection as digital and analog sources.
Analog Source Selections (Table 6-3)
Alarms (PV) Outputs Math Blocks Auxiliary - Function Generators Monitor 1 through 4 Auxiliary - Switch (A,B) Relays (Time Prop, Pulse Freq) PID (RSP Source, Feedforward Source)
None Input 1-2 PV Input 1-2 Temp Pharma Out 1-2 Math 1-4 Func Gen 1-2 Switch 1-2 Sum Difference Ratio %Passage %Rejection Cation Value PIDout 1-2 AnlgVar 1-4
PV Source Selections PID 1 and 2 PV Source Function Generator 1 and 2
None Input 1 – 2 PV
Digital Source Selections (Table 6-4)
Logic (InA, InB) Relays (Digital Out, Pulse out) Auxiliary - Switch (Select B Source) Alarms Disable PID (Remote setpoint select, Manual select) Auto Cycle – Start Source
None Alarm1-4 Alm Grp 1-2 Monitor 1-4 Logic 1-4 Digital In 1-2 In 1-2 Fault In 1-2 Hold Out 1-3 Fault Hold Pharma 1-2 Fail Pharma 1-2 Warn PID 1 Alarm 1-2 PID 2 Alarm 1-2 Auto Cycle 1 – Extract, Rinse, Cal PT1, Cal PT2, Fail Auto Cycle 2 – Extract, Rinse, Cal PT1, Cal PT2, Fail Input 1-2 Cal Output 1-3 Cal DgtlVar 1-4
Configuration

Analog Signal Description Definition
Input 1 PV Input 1 Process Variable
Input 2 PV Input 2 Process Variable
Input 1 Temp Input 1 Temperature Input 1 Temperature Selection
Input 2 Temp Input 2 Temperature Input 2 Temperature Selection
Pharma Out 1 Pharmacopoeia Output 1
Input 1 Pharmacopia 1 Output (for Conductivity) = percent of USP stage limit
Output = 100 * pv in uScm / USP stage limit
Valid for Conductivity Input
Pharma Out 2 Pharmacopoeia Output 2
Input 2 Pharmacopia 2 Output = percent of USP stage limit
Output = 100 * pv in uScm / USP stage limit
Valid for Conductivity Input
Math 1 Math 1
Math 2 Math 2
Math selections can be connected to any Input PV, secondary variable (Temperature), or Calculated Value. Math blocks include scaling for the linear selection only. See Table 6-10 for Math Configuration
Math 3 Math 3
Math 4 Math 4
Func Gen 1 Function Generator 1
Generates an output characteristic curve based on up to 11 configurable data points for both input (X) and output values (Y). Part of the Auxiliary Configuration group. See Table 6-12 for Function Generator Configuration
Func Gen 2 Function Generator 2
Switch 1 Switch 1
Switch selections have 2 input sources (A and B). A switch block is used to select between two analog signals. The switch block can be used for many monitor and control strategies. A Digital Signal Source when active will select the B input source of the switch as the output. Part of the Auxiliary Configuration group. See Table 6-12 for Switch Configuration
Switch 2 Switch 2
Sum* Input 1 + Input 2
Difference* Input 1 – Input 2
Ratio* Input 1 / Input 2
The availability of calculated variables in the list of available sources for alarms, math and control and for status display is determined by similarity of units of measure between the two input boards.
%Passage* Min(Input 1 or 2) Max(Input 1 or 2) *100
% Rejection* (1-Min(Input 1 or 2)/ Max(Input1 or 2)) *100
Cation Value pH Value Calculated pH value from differential conductivity
PID Out 1 PID Output 1 PID 1 Output in percent (0 to 100). Normally connected to a proportional current (Current Type) or time proportional or frequency proportional relay.
Configuration · PV Source selection · PV Source selection


PID Out 2 PID Output 2 PID 2 Output in percent (0 to 100). Normally connected to a proportional current (Current Type) or time proportional or frequency proportional relay.
Anlg Var 1 Analog Variable 1 Initial values of Analog Variable 1 applied at power on.
Anlg Var 2 Analog Variable 2 Initial values of Analog Variable 2 applied at power on.
Anlg Var 3 Analog Variable 3 Initial values of Analog Variable 3 applied at power on.
Anlg Var 4 Analog Variable 4 Initial values of Analog Variable 4 applied at power on.
Digital Signal Description Definition
Alarm 1 Alarm 1 Any Alarm 1 configuration. See Table 6-8 for Alarm configuration
Alarm 2 Alarm 2 Any Alarm 2 configuration. See Table 6-8 for Alarm configuration
Alarm 3 Alarm 3 Any Alarm 2 configuration. See Table 6-8 for Alarm configuration
Alarm 4 Alarm 4 Any Alarm 4 configuration. See Table 6-8 for Alarm configuration
Monitor 1 Monitor 1 Any Monitor 1 configuration. See Table 6-9 for Monitor configuration
Alarm Group 1 Alarm Group 1 Is the OR of the Alarm 1 - 4 signals. Will be TRUE when any Alarm 1 - 4 is TRUE.
If a single digital signal is needed to go TRUE for any alarm, OR alarm group 1 and alarm group 2 together to create a logic signal.
Alarm Group 2 Alarm Group 2 Is the OR of the PID Control alarm signals. Will be TRUE when any PID Control Alarm is TRUE.
If a single digital signal is needed to go TRUE for any alarm, OR alarm group 1 and alarm group 2 together to create a logic signal.
Monitor 2 Monitor 2 Any Monitor 2 configuration. See Table 6-9 for Monitor configuration
Monitor 3 Monitor 3 Any Monitor 3 configuration. See Table 6-9 for Monitor configuration
Monitor 4 Monitor 4 Any Monitor 4 configuration. See Table 6-9 for Monitor configuration
Logic 1 Logic 1 Any Logic 1 configuration. See Table 6-11 for Logic configuration
Logic 2 Logic 2 Any Logic 2 configuration. See Table 6-11 for Logic configuration
Logic 3 Logic 3 Any Logic 3 configuration. See Table 6-11 for Logic configuration
Logic 4 Logic 4 Any Logic 4 configuration. See Table 6-11 for Logic configuration
Digital In 1 Digital Input 1 Digital Input 1 signal from Option Board (must be installed)
Digital In 2 Digital Input 2 Digital Input 2 signal from Option Board (must be installed)
In 1 Hold In 1 Hold
Input is in Hold. This condition occurs either by pushing the HOLD button on the front panel or when an Auto Cycle is being run.
In 2 Hold In 2 Hold
In 1 Fault Input 1 Fault
Input open conditions. An input board disconnect while powered results in an input fault condition and allows an alarm to be triggered.
In 2 Fault Input 2 Fault
Configuration

Out 1 Fault Output 1 Fault
Output open conditions. This allows an alarm to be triggered if the respective 4-20 mA output opens.
Out 2 Fault Output 2 Fault
Out 3 Fault Output 3 Fault
Hold Hold Engages Hold of Analog Inputs
Pharm 1 Warn Pharmacopoeia 1 Warning
The Pharma 1 Display ( Section 5.8) outputs digital Warning signal whenever the measured conductivity exceeds the Percent Warning Value selected in the “Pharma Op Panel” on the Pharma Display (Stage 1only)
Pharm 1 Fail Pharmacopoeia 1 Failure
The Pharma 1 Display ( Section 5.8) outputs digital Failure signal whenever one of the following conditions occur: Stage 1 – Measured Conductivity exceeds 100% Stage 1 – Temperature not within range of 0-100 degrees C Stage 2 – Conductivity is 0.1 µS/cm or greater for 5 minutes Stage 3 – pH not within range of 5 – 7pH Stage 2 and 3 – Temperature not within range of 24 – 26 degrees C.
Pharm 2 Warn Pharmacopoeia 2 Warning
The Pharma 2 Display ( Section 5.8) outputs digital Warning signal whenever the measured conductivity exceeds the Percent Warning Value selected in the “Pharma Op Panel” on the Pharma Display (Stage 1only)
Pharm 2 Fail Pharmacopoeia 2 Failure
The Pharma 2 Display ( Section 5.8) outputs digital Failure signal whenever one of the following conditions occur: Stage 1 – Measured Conductivity exceeds 100% Stage 1 – Temperature not within range of 0-100 degrees C Stage 2 – Conductivity is 0.1 µS/cm or greater for 5 minutes Stage 3 – pH not within range of 5 – 7pH Stage 2 and 3 – Temperature not within range of 24 – 26 degrees C.
PID 1 Alm 1 PID Control 1 Alarm 1
Control Alarms – See Table 6-15 PID Alarms
PID 1 Alm 2 PID Control 1 Alarm 2
PID 2 Alm 1 PID Control 2 Alarm 1
PID 2 Alm 2 PID Control 2 Alarm 2
AC 1 Extract Auto Cycle 1 Probe Extraction
Auto Cycle 1 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
AC 1 Rinse Auto Cycle 1 Probe Rinse
Auto Cycle 1 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
AC1 Cal Auto Cycle 1 Calibration Point 1
Auto Cycle 1 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
AC 1 Cal 2 Auto Cycle 1 Calibration Point 2
Auto Cycle 1 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
AC 1 Fail Auto Cycle 1 Failure Auto Cycle 1 Failure is active whenever an Auto Cycle 1 failure occurs Auto Cycle 1digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
AC 2 Extract Auto Cycle 2 Probe Extraction
Auto Cycle 2 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
AC 2 Rinse Auto Cycle 2 Probe Rinse
Auto Cycle 2 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
AC 2 Cal Auto Cycle 2 Calibration Point 1
Auto Cycle 2 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
Configuration

AC 2 Cal 2 Auto Cycle 2 Calibration Point 2
Auto Cycle 2 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
AC 2 Fail Auto Cycle 2 Failure Auto Cycle 2 Failure is active whenever an Auto Cycle 2 failure occurs Auto Cycle 2 digital output (Cycle Start Source) configuration selection See Table 6-16 Auto Cycling Configuration.
Input 1 Cal Input 1 Calibration This signal goes TRUE when the calibration factor for input 1 is being calculated. The TRUE state is active for less than one second.
Input 2 Cal Input 2 Calibration This signal goes TRUE when the calibration factor for input 2 is being calculated. The TRUE state is active for less than one second.
Output 1 Cal Output 1 Calibration The signal indicates when the Output 1 calibration values are being changed. The signal goes TRUE when the “4ma Offset” or “20ma Offset” is being modified. The signal goes FALSE when the value is entered.
Output 2 Cal Output 2 Calibration The signal indicates when the Output 2 calibration values are being changed. The signal goes TRUE when the “4ma Offset” or “20ma Offset” is being modified. The signal goes FALSE when the value is entered.
Output 3 Cal Output 3 Calibration The signal indicates when the Output 3 calibration values are being changed. The signal goes TRUE when the “4ma Offset” or “20ma Offset” is being modified. The signal goes FALSE when the value is entered.
DgtlVar 1 Digital Variable 1 Initial values of Digital Variable 1 applied at power on.
DgtlVar 2 Digital Variable 2 Initial values of Digital Variable 2 applied at power on.
DgtlVar 3 Digital Variable 3 Initial values of Digital Variable 3 applied at power on.
DgtlVar 4 Digital Variable 4 Initial values of Digital Variable 4 applied at power on.
Configuration

This group lets you select pH/ORP, Preamp pH, Conductivity, or Dissolved Oxygen Input type and the associated output parameters.
Setup to display the Main menu.
Use the keys to select “Inputs” then press
Enter to enter the sub-menus.
Input 1 and Input 2 – Direct pH/ORP, Preamp pH, Conductivity, or Dissolved Oxygen are available for selection. Select PV type, read the range, select Temp Type, Solution Temp Compensation, Bias, Failsafe and Filter Time. For Dissolved Oxygen, also select the Salinity type and Pressure type.
Enter to display the group of parameters.
Press to highlight the desired menu selection then press
Refer to “Section 6.4.1 – ”General Rules for Editing”.
PV Type pH Glass pH HPW pH Durafet (default) ORP
The PV type determines the numerical format and the units of measure on the online PV display. Measured PV is generally displayed in the highest decimal precision possible to .001 and has a potentially displayable range of 0.000 to 99999. The exceptions are dissolved oxygen, pH, ORP and temperature, which are displayed with fixed decimal precision. PV Type determines specific ranges.
Input 1 or 2
PV Range 0.0 to 14.0 pH
-1600 to 1600 ORP
Temp Input (ORP only)
Enable to allow “Temp Type” selection – see below.
Temp Type 8550Ω Therm (default) 1000Ω RTD Manual
8550Ω Thermistor 1000Ω Resistance Temperature Detector Manual

Temp Deg F or C (Temp Type = Manual)
14.0 to 230.0ºF default = 77ºF
Temp Deg F or C will appear depending on what Temperature Unit was selected in “Maintenance” setup group, parameter “Temp Units”.
-10 to 110ºC default = 25ºC
None (default) Custom H20 NH3 Phosphate Morpholine
Enter “Solution pH/ºC” value Pure Water Ammonia Phosphate Morpholine
0.000 (default) to -0.050
Measured pH is displayed and transmitted normalized to a solution temperature of 25°C as determined by the current Solution Temperature Coefficient. This is expressed in units of pH/°C with precision to the hundredths decimal place. The parameter “Solu Temp Coeff” allows the selection of the following entries. Follow the “General Rules for Editing” in section 6.4.1 to make the changes. (-) Will appear when first digit to the right of decimal point is changed.
(Solu Temp Comp = Custom)
Solution Type Temp Coefficient
None (Default) 0.000
H2O (Pure Water) -0.016
NH3 (Ammonia) -0.032
PO4 (Phosphate) -0.032
C4H9NO (Morpholine) -0.032
Custom User Entry PV Bias -99999 to 99999 default = 0.00
PV Bias Constant - is used to compensate the input for drift of an input value.
Failsafe -99999 to 99999 default = 14.00
The output value to which the output will go to protect against the effects of failure of the equipment.
Filter Time 0 to 120 default = 0
A software digital filter is provided for dampening the process noise. This filter is applied before the limit functions.

Input 1 or 2
The pH Preamp input card measures pH and accepts inputs from a Durafet series Preamp, a glass Meredian II Preamp or a Durafet series Cap Adapter. The pH Preamp input is similar to the pH/ORP input shown previously and has an identical Setup/Inputs parameter menu with the following important differences:
No ORP measurement. ORP is not selectable as a PV Type in Setup/Inputs.
No HPW measurement. HPW is not selectable as a PV Type in Setup/Inputs.
The parameter “Temperature Input” is available for either Durafet or Glass PV type to enable or disable. A temperature input disable accommodates preamps that do not transmit measured temperature from the probe. This will disable all monitored temperature values, temperature input diagnostics and faults and the parameter “Solution Temp Comp” under Setup/Inputs/pH Preamp n.
You need to disable “Temperature Input” for Durafet from External Preamp.
PV Type pH Glass pH HPW pH Durafet (default)
The PV type determines the numerical format and the units of measure on the online PV display. Measured PV is generally displayed with fixed decimal precision.
PV Range 0.0 to 14.0 pH Read Only
Enable to allow “Temp Type” selection – see below.
8550Ω Therm (default) 1000Ω RTD Manual
8550Ω Thermistor 1000Ω Resistance Temperature Detector Manual
(Temp Input = Enable)
Temp Deg F or C (Temp Input = Enable, Temp Type = Manual)
14.0 to 230.0ºF default = 77ºF
Temp Deg F or C will appear depending on what Temperature Unit was selected in “Maintenance” setup group, parameter “Temp Units”.
-10 to 110ºC default = 25ºC
None (default) Custom H20 NH3 Phosphate Morpholine
Enter “Solution pH/ºC” value Pure Water Ammonia Phosphate Morpholine
(Temp Input = Enable)

0.0 (default) to -0.050 pH/ºC
Measured pH is displayed and transmitted normalized to a solution temperature of 25°C as determined by the current Solution Temperature Coefficient. This is expressed in units of pH/°C with precision to the hundredths decimal place. The parameter “Solu Temp Coeff” allows the selection of the following entries. Follow the “General Rules for Editing” in section 6.4.1 to make the changes. (-) Will appear when first digit to the right of decimal point is changed.
(Temp Input = Enable, Solu Temp Comp = Custom)
Solution Type Temp Coefficient
None (Default) 0.000
H2O (Pure Water) -0.016
NH3 (Ammonia) -0.032
PO4 (Phosphate) -0.032
C4H9NO (Morpholine) -0.032
Custom User Entry PV Bias -99999.00 to 99999.00 default = 0.00
PV Bias Constant - is used to compensate the input for drift of an input value.
Failsafe -99999.00 to 99999.00 default = 14.00
The output value to which the output will go to protect against the effects of failure of the equipment.
Filter Time 0 to 120 default = 0.0
A software digital filter is provided for dampening the process noise. This filter is applied before the limit functions.

For every cell constant the PV type includes selections for both conductivity µS/cm and conductivity mS/cm. Conductivity µS/cm displays µS/cm and provides standard range solution type selections: None, NaCl, Morpholine, HCL, Acid, and NH3. Conductivity mS/cm displays mS/cm and provides wide range solution type selections: None, HCl, NaCl, H2SO4, and NaOH.
Upper range limit defaults according to the table below:
For every cell constant the PV type also includes selections for either TDS ppb/TDS ppm or TDS ppm/TDS ppt.: TDS ppb/ppm provides standard or wide solution type selections and TDS ppm/ppt provides standard or wide solution type selections. Solution selections are the same as above with the exception of None. Upper range limit defaults according to the table below:
Cell Const 0.01 Cell Const 0. 1 Cell Const 1 Cell Const 10 Cell Const 25 Cell Const 50

Cond μS/cm (NIST-default)+
These selections are only available with regard to the Cell Constant selected (See “Cell Constant”).
Select Cell Constant First
Cond mS/cm – (NIST) Concentrtn TDS ppb TDS ppm TDS ppt Resistivity
Available Selectable PV Types
Use the keys to select
0.01 Conductance μS/cm (default- NIST), Conductance mS/cm (NIST), Conductance mS/m(default - ISO), Conductance S/m (ISO), TDS ppb, TDS ppm, Resistivity, Conductance μS/m (ISO)
0.1 (Default)
Conductance μS/cm (default- NIST), Conductance mS/cm (NIST), Conductance mS/m(default - ISO), Conductance S/m (ISO), TDS ppb, TDS ppm, Resistivity, Conductance μS/m (ISO)
Cond mS/m (ISO-Default)+ Cond S/m – (ISO) Concentrtn TDS ppb TDS ppm TDS ppt Resistivity Cond μS/m
1 Conductance μS/cm (default- NIST), Conductance mS/cm (NIST), Conductance mS/m(default - ISO), Conductance S/m (ISO), TDS ppm, TDS ppt,
10 Conductance μS/cm (default- NIST), Conductance mS/cm (NIST), Conductance mS/m(default - ISO), Conductance S/m (ISO), TDS ppm, TDS ppt.
25 Concentration (default), Conductance μS/cm (default- NIST), Conductance mS/cm (NIST), Conductance mS/m(default - ISO), Conductance S/m (ISO),
+ parameter selected in MAINTENANCE Æ INPUTS menu
50 Concentration (default), Conductance μS/cm (default- NIST), Conductance mS/cm (NIST), Conductance mS/m(default - ISO), Conductance S/m (ISO),
PV Range Read Only
Cell Constant * 0.01 0.1 (default) 1 10 25 50
The Cell Constant is a value specific to a category of cells for the measurement range required.
Cal Factor * 0.850 to 1.150 default = 1.000
The Cal Factor is a correction value applied to the cell’s Cell Constant, which is unique to each cell to take into account tolerances in manufacture. If a standard cell is attached to the sensor, the Cell Constant defaults to “0.1” and the Cal Factor defaults to “1.000”. These standard cell parameter values are editable and are retained through a power cycle.
*Cell Constant and Cal Factor are automatically uploaded from Honeywell conductivity cells with EEPROM (blue & brown leads) and these values cannot be edited.

TDS Factor (only PV Type TDS)
0.010 1.000(default) 2.000
The TDS Factor is a conversion value applied to conductivity to derive total dissolved solids, in units of ppm per μS/cm.
Temp Type 8550Ω Therm (default) 1000Ω RTD Manual
8550Ω Thermistor 1000Ω Resistance Temperature Detector Manual
Temp Deg C or F
-10.0 to 140.0ºC
If “Manual” is selected at “Temp Type” -Temp Deg F or C will appear depending on what Temperature Unit was selected in “Maintenance” setup group, parameter “Temp Units”.

Solu Temp Comp None Custom H20 NH3 PO4 C4H9NO HCl NaCl (default) H2SO4 NaOH
Measured Conductivity and Resistivity can optionally be temperature compensated to 25°C for a specific solution type. TDS and concentration are always measured based on a specific solution type. The cell constant and measurement type determines which solution types are available for selection, according to the table below:
Available Selectable Solution Types
Use the keys to select
0.01 None (Conductivity/Resistivity only), NaCl (μS/cm, mS/cm, TDS ppb, TDS ppm ), NH3 (μS/cm, TDS ppb, TDS ppm ), C4H9NO (μS/cm, TDS ppb, TDS ppm ), H2SO4;HCL;NaOH (mS/cm)
0.1 (Default)
None (Conductivity/Resistivity only), NaCl(μS/cm, mS/cm, TDS ppb, TDS ppm ), NH3(μS/cm, TDS ppb, TDS ppm ), C4H9NO (μS/cm, TDS ppb, TDS ppm ), H2SO4;HCL;NaOH (mS/cm)
1 None (Conductivity only), NaCl (μS/cm, mS/cm, TDS ppm, TDS ppt ), NH3 (μS/cm, TDS ppm ), C4H9NO (μS/cm, TDS ppm ), H2SO4;HCL;NaOH (mS/cm, TDS ppt)
10 None (Conductivity only), NaCl (μS/cm, mS/cm, TDS ppm, TDS ppt ), NH3 (μS/cm, TDS ppm ), C4H9NO (μS/cm, TDS ppm ), H2SO4;HCL;NaOH (mS/cm, TDS ppt)
25 None (Conductivity only), HCl (mS/cm, Concentration), NaCl (μS/cm, mS/cm, Concentration), H2SO4 (mS/cm, Concentration), NaOH (mS/cm, Concentration)
50 None (Conductivity only), HCl (mS/cm, Concentration), NaCl (μS/cm, mS/cm, Concentration), H2SO4 (mS/cm, Concentration), NaOH (mS/cm, Concentration)
Wire Len Feet + 0 to 1000 ft default = 0
Refer to appendix 15.2 to enter values for lead wire resistance compensation
Wire Len Meters + 0 to 304.80 default = 0

Wire Size AWG + 16 AWG 18 AWG(default) 20 AWG 22 AWG
Wire Size Sq mm + 0.33 to 2.08
default = 0.82
Pharma Type None PhEur USP
PhEur - Pharmacopoeia Europa
USP - United States Pharmacopoeia
standard procedure stages for determining Purified Water
Pharma PV High -99999.00 to 99999.00 (default 10.000)
Pharma PV High Value – Measured solution conductivity value scaled for 100%
Pharma PV Low -99999.00 to 99999.00 (default 0.000)
Pharma PV Low Value - Measured solution conductivity value scaled for 0%
Pharm Tmr Mins 000.0 to 120.0 (default 10.000)
Pharma Timer Minutes - If the Pharma sample does not pass the Stage 1 conductivity requirement a Fail signal is generated, then the State 2 and Stage 3 tests are conducted. When the Stage 2 or Stage 3 test is successful, the fail signal is cancelled and the Pharma Timer begins to count down from the configured minutes value set here. When the Timer countdown is completed, the Pharma function block returns to Stage 1.
PV Bias -9999.00 to 9999.00 default = 0.000
PV Bias Constant - is used to compensate the input for drift of an input value.
Failsafe 0.0 to 2000 default = 2000.000
The output value to which the output will go to protect against the effects of failure of the equipment.
Filter Time 0 to 120.0 default = 0.000
A software digital filter is provided for dampening the process noise and is applied before the limit functions.

PV Type DO% Sat
The concentration of oxygen dissolved in water (or other liquid) may be described by either “dissolved oxygen (DO) concentration” or percent saturation. The units for DO are either parts per million - PPM (equivalent to milligrams per liter) or parts per billion - PPB (equivalent to micrograms per liter). The units of saturation are percent where 100% saturation is equivalent to the concentration of oxygen dissolved in air-saturated water. For instance, at 25°C and one atmosphere pressure, 8.24 ppm = 100% saturation. Although the ppm and ppb concentration units are the most frequently used units by far, % saturation may be appropriate for non-aqueous liquids like vegetable oil.
PV Range 0 – 200 ppb, displayable to 20000ppb
Temp Type 5000Ω Therm Default 1000Ω RTD Manual
5000Ω Thermistor 1000Ω Resistance Temperature Detector Manual
Temp Deg C or F (Temp Type = Manual)
Temp Deg F or C will appear depending on what Temperature Unit was selected in “Maintenance” setup group, parameter “Temp Units”.
Salinity is used to correct for salt in the process water.
Valid only if conductivity board is present.
0.00 to 40.00ppt
(parts per thousand) as sodium chloride
“Manual” Salinity type only
0.0 = No selection
default = 0.00
Allows manual entry of atmospheric pressure compensation
Internal sensor for atmospheric pressure compensation during air calibration
500.0 to 800.0
Atmospheric pressure compensation. Enter a value in mmHg.
(Manual Pressure type only)
default = 760 mmHg

PV Bias Constant - is used to compensate the input for drift of an input value.
-20.00 to 20.00 PPM
If PPM Board is installed.
-20000 to 20000 PPB
If PPB Board is installed.
default = 0.000
The output value to which the output will go to protect against the effects of failure of the equipment.
0.000 to 20.00 PPM
If PPM Board is installed.
0.000 to 20000 PPB
If PPB Board is installed.
default = 20.000
Filter Time 0 to 120.0
A software digital filter is provided for dampening the process. The units are in time constant seconds.
default = 0.0

This group lets you select the signal that will be transmitted.
Enter to enter the sub-menu. • Output 1, Output 2, or Output 3 and their associated parameters are available for selection.
Enter to display the group of parameters. Refer to “Section 6.4.1 - General Rules for Editing”.
Parameter Selection or Range
Output 1 Output 2 Output 3
Source Any Analog Signal See Table 6-3
Process Variable Source - Selects the signal that will be transmitted.
See Note 1 for units.
High Range -99999.00 to 99999.00
High Range Value - value of input that corresponds to 100 % output value.
See Note 1 for units.
Low Range -99999.00 to 99999.00
Low Range Value - value of input that corresponds to 0 % output value.
See Note 1 for units.
Slew Time 0.000 to 999.00 in seconds default = 0.000
Slew Time is the maximum rate of change required to drive the output from full OFF (0% - typically 4 mA) to full ON (100% - typically 20mA).
mA Range High 0 to 20 default = 20
Value of mA output that corresponds to 100 % output signal (for example: 20 mA).
mA Range Low 0 to 20 default = 4
Value of mA output that corresponds to 0 % output signal (for example: 4 mA).
mA Limit High 0 to 21 default = 21
Value of mA that you want to set the High Range Limit.
mA Limit Low 0 to 21 default = 3
Value of mA that you want to set the Low Range Limit.

The entries for any parameter are in the units of that parameter.
For example:
Parameters in engineering units.
So in the SETUP OUTPUT menu, for the SOURCE and Hi Range and Low Range values, these look at the units of that source.
If retransmitting a pH input, the Hi Range and Low Range values would normally be set to 14 pH and 0 pH. (14 pH= 100% output, and 0 pH = 0% output.)
But if the output is to go to a valve, to open the valve or operate a pump through a range of 0-100% open, and has a SOURCE of CONTROL 1, then the units of the CONTROL 1 output is in units of %, so in the SETUP OUTPUT menu, the High Range and Low Range would be in % units.

Overview
Programming the relays consists of selecting the relay type, identifying the input parameter, which activates the relay and selecting whether the relay is energized when the input parameter is on or off. The Relay group lets you select a relay type for up to four relays. When planning relay operation, it is wise to consider the state of the relay when power is not applied to the UDA. The invert parameter of the relay configuration is helpful in assuring that the Off device state is consistent with the relay normal operation.
Each relay output can be independently configured to be one of four basic types:
A Digital Output Relay allows connection to any Alarm, Alarm Group, Monitor, Logic, Digital Input, Input Fault, Output Fault, Hold Key, Pharm Warn, Pharm Fail, Control Alarms, Input 1 and 2 Rinse and Cal Pts, Cycle on or fail.
Time proportional output is a form of a process variable transmitter or control output that pulses the relay as a pulse width modulated signal that is proportional to the input signal over a configured input range. The Time Proportional cycle time is configurable between 0.1 and 999 seconds while the duty cycle is directly proportional to the selected input signal.
Frequency proportional output is a form of a process variable transmitter or control output that pulses the relay as a pulse rate that is proportional to the input signal over a configured input range. The maximum frequency is set by the cycle time that is configurable between 0.1 and 999 seconds. The pulse duration is fixed and configured in seconds by an on time parameter.
On / Off output relay turns On when the input is greater than the high and low ranges and turns off when the input is less than the high and low ranges. This allows an on / off control action with an adjustable dead band. The On state is controlled by a cycle time and on duration parameters such to achieve a selectable output proportion. An invert parameter is available to allow inverse action such that the relay will cycle ON when below the low range limit.
Pulse Output relay will provide a fixed duty cycle when the applied input signal is ON. The cycle time and pulse duration are configurable parameters. The relay will be OFF when the applied input is OFF. An inverse parameter allows the input to be inverted for reverse behavior.
Enter to enter the sub-menu.
Configuration

Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Parameter Selection or Range
Digital Output Relay (default)
Set relay types first
Time Proportional Output Relay
On / Off control relay
Relay 1 Relay 2 Relay 3 Relay 4 Digital Output Relay
A Digital Output Relay allows connection to any Alarm, Control Alarm, Logic, Alarm Event, Hold, Input or Output Fault, or Digital Input.
Source Any Digital Signal See Table 6-4
Invert Enable Disable (default)
Inverts the input state of the applied digital input such that inverse relay operation is achieved
Relay 1 Relay 2 Relay 3 Relay 4 Time Proportional Output Relay
Time proportional output is a form of a process variable transmitter or control output that pulses the relay as a pulse width modulated signal that is proportional to the input signal over a configured input range. The Time Proportional cycle time is configurable between 0.1 and 999 seconds while the duty cycle is directly proportional to the selected input signal.
Source Any Analog Signal See Table 6-3
High Range -99999 to 99999 default = 100.00
The high range is the PV based engineering unit value configured as the value that will produce a 100 percent (always active) duty cycle.
Low Range -99999 to 99999 default = 0.00
The low range is the PV based engineering unit value configured as the value that will produce a 0 percent (always inactive) duty cycle.
Invert Enable Disable (default)
Inverts the proportional range of the applied analog input such that inverse relay operation is achieved.
Cycle Time 0 to 999 seconds default = 10
Cycle time is that time period, in seconds, the relay will be activated.
Min Off Time 0 to 15 default = 0
Minimum off is that time period, in seconds, the relay will be activated.
Min On Time 0 to 15 default = 0
Minimum On is that time period, in seconds, the relay will be activated.

Parameter Selection or Range
Relay 1 Relay 2 Relay 3 Relay 4 Frequency Proportional Output Relay
Frequency proportional output is a form of a process variable transmitter or control output that pulses the relay as a pulse rate that is proportional to the input signal over a configured relay range. The maximum frequency is set by the cycle time that is configurable between 0.1 and 999 seconds. The pulse duration is fixed and configured in seconds by an on time parameter.
Source Any Analog Signal See Table 6-3
High Range -99999 to 99999 default = 100.00
The high range is the PV based engineering unit value configured as the value that will produce a 100 percent (Maximum Frequency) duty cycle.
Low Range -99999 to 99999 default = 0.00
The low range is the PV based engineering unit value configured as the value that will produce a 0 percent (always inactive) duty cycle.
Invert Enable Disable (default)
Inverts the proportional range of the applied analog input such that inverse relay operation is achieved.
Cycle Time 0 to 999 default = 10
Sets the Cycle Time of the maximum output frequency.
Max Freq=1/Cycle Time Freq Output = Max Freq * (% Input/100)
Freq Ouput (50%)= Max Freq * .5
Freq Ouput (25%)=Max Freq * .25
0.0 to 999 default = 5
Sets the pulse duration. This value should be less than the cycle time for proper operation. Typically this value is used to control the pulse duration for the finial output control element.

Parameter Selection or Range
Relay 1 Relay 2 Relay 3 Relay 4 ON/OFF Control Relay
On / Off output relay turns On when the input is greater than the relay high and low ranges and turns off when the input is less than the relay high and low ranges. This allows an on / off control action with an adjustable dead band. The On state is controlled by a cycle time and on duration parameters such to achieve a selectable output proportion. An invert parameter is available to allow inverse action such that the relay will cycle ON when below the low range limit.
Source Any Analog Signal See Table 6-3
High Range -99999 to 99999 default = 100.00
The high range is the PV based engineering unit value configured as the value that will produce a 100 percent (Maximum Frequency) duty cycle.
Low Range -99999 to 99999 default = 0.00
The low range is the PV based engineering unit value configured as the value that will produce a 0 percent (always inactive) duty cycle.
Invert Enable Disable (default)
Inverts the proportional range or input state of the applied digital or analog input such that inverse relay operation is achieved.
Cycle Time 0 to 999 default = 10
Cycle time is that time period, in seconds, between relay activations.
0,0 to 999 default = 5
Sets the pulse duration. This value should be less than the cycle time for proper operation. Typically this value is used to control the pulse duration for the finial output control element.

Parameter Selection or Range
Relay 1 Relay 2 Relay 3 Relay 4 Pulse Output Control Relay
Pulse Output relay will provide a fixed duty cycle when the applied input signal is ON. The cycle time and pulse duration are configurable parameters. The relay will be OFF when the applied input is OFF. An inverse parameter allows the input to be inverted for reverse behavior.
Source Any Digital Signal See Table 6-4
Invert Enable Disable (default)
The digital output relays "invert" parameter can be used to allow direct (invert disabled) or reverse (invert enabled) control actuation.
Cycle Time 0 to 999 default = 10.0
Cycle time is that time period, in seconds, between relay activations.
0 to 999 default = 5.0
The time in seconds that the relay is On during each cycle when the input to the Pulse Output is ON. When the Input to the Pulse output is OFF the relay is not activated.

Alarm 1 through 4
Alarm selections can be connected to any Analog Signal (Table 6-3 Analog Signal Sources). Each alarm supports a setpoint type and value. Alarm selections generate front panel alerts, support latching/acknowledge, with on delay timers. Select any Digital signal (Table 6-4 Digital Signal Sources) to disable the Alarm
This example shows how to use the math blocks to scale the output in multiple ranges and uses a monitor and switch to select the desired amplification for the input. A relay is connected in parallel to the switch to provide an indication as to which range is currently being transmitted.
Range Switch using Math, Monitor, and Switch Blocks

Math 1 (x – low range) \ (high range – low range)
Math 2 (x – low range) \ (high range – low range)
Configuration

Enter to enter the sub-menu.
Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Parameter Selection or Range of Setting
Alarm 1 Alarm 2 Alarm 3 Alarm 4
Source Any Analog Signal See Table 6-3
Process Variable Source – Process Variable to be monitored by the alarm.
Disable Any Digital Signal See Table 6-4
Select any Digital signal to disable the Alarm
Type High (default) Low
Alarm actions may be High or Low.
Setpoint Value -99999 to 99999.9 in Engineering Units default = 0.000
Setpoint value in engineering units
Latch Disable (default) Enable
When enabled, the alarm is latch ON until acknowledged from the Alarm Status display.
Hysteresis 0.0 to 99999.9 in engineering units default = 0.000
Hysteresis - A user-specified hysteresis value in the engineering units of the process variable source is provided. Hysteresis in engineering units can be set from 0 to the input span of the monitored variable.
On Delay 0 to 120 seconds default = 0.000
An on-delay time value up to 120 seconds is available to prevent momentary alarm actions. Number of seconds the alarm is active before activating the Output.

Monitor 1, 2, 3, and 4
A Monitor Block is used to determine when a process value is greater or less than a specified setpoint. Monitor blocks can be used for ON/OFF type control or, in conjunction with switch and math blocks to change process gain based upon control regions. The Monitor block provides a hysteresis value limit output transitions near the set point value. The Monitor block can be configured as either a High or Low Monitor type. There are four monitor blocks provided for general use. Unlike Alarms, Monitor blocks do not create an event in the event history, nor do they cause a status message to appear on the display.
Enter to enter the sub-menu. • Press to highlight the desired Monitor selection then press
Enter to display the group of parameters. Refer to “Section 6.4.1 – “General Rules for Editing”.
Parameter Selection or Range of Setting
Monitor 1 Monitor 2 Monitor 3 Monitor 4
Monitor Type High (default) Low
Alarm actions may be High or Low. (See NOTE 2 on next page)
Source Any Analog Signal See Table 6-3
Analog Signal Source – Process signal to be monitored by the Alarm. Any analog source such as PV, Temperature, Pharma, Math, Function Generator, Switch, PID, or Calculated Values* * units of measure between the two input boards must be similar
Setpoint Value 0 to 99999.9 in Engineering Units default = 0.000
Setpoint Value in Engineering Units used for activation of the output based upon the monitor type
Hysteresis 0.0 to 99999.9 in engineering units default = 0.000
Hysteresis - A user-specified hysteresis value in the engineering units of the process variable source is provided. Hysteresis in engineering units can be set from 0 to the input span of the monitored variable.
On Delay 0 to 999 seconds default = 0.0
An on-delay time value up to 999 seconds is available to prevent momentary alarm actions. Number of seconds the alarm is active before activating the Output.

For Low Monitor If Input less than setpoint Output = ON else if Input greater than set point + hysteresis, Output = OFF else Output is unchanged
Configuration

The Math group has four Math selections (Math 1, Math 2, Math 3, and Math 4). Math selections can be connected to any Analog Signal source (Table 6-3). Math blocks include scaling for the linear selection only.
The Math Block can also be used for proportional control over the math blocks configured range for control of any Input PV, Temperature, or calculated values by connecting it to a current output, TPO relay, or FPO relay. Since multiple outputs can share a common math block, the output range of a math block can be split over multiple outputs or relays with each output or proportional relay using a specific portion of the % output range of the math block.
Enter to enter the sub-menu.
Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Configuration

Sub-menu Parameter Selection or
Math 1 Math 2 Math 3 Math 4
Linear (default) Provide a linear output with Gain and Offset with digital filtering.
Output = Filter (Gain * (Input) + Offset) Linear is simple linear scale used to retransmit the PV using the High Range as scaled 100% output and the Low Range is the scaled to 0% output. There is no restriction on the High and Low ranges. Setting the high range to a value less than the low range will invert the action of the math output. Limit out has no effect on the output.
Log Log (base 10): Output = Log(Input): Input > 10^-10 Output = -10 Input
Abs Value Absolute Value If Input >= 0 then Output = Input If Input < 0 then Output = -Input
Any Analog Signal See Table 6-3
Analog Signal Source – Process signal to be monitored by the Alarm. Any analog source such as PV, Temperature, Pharma, Math, Function Generator, Switch, PID, or Calculated Values* * units of measure between the two input boards must be similar
Source -99999 to 99999.9 in Engineering Units default = 1.000
For Linear Math Types. Gain multiplier for Calculation Output = Gain * Input + Offset
-99999 to 99999.9 in Engineering Units default = 0.000
For Linear Math Types. Offset for Calculation Output = Gain * Input + Offset
-99999 to 99999.9 in Engineering Units default = 0.000
Filter Time 0 to 120 default = 0.0
A software digital filter is provided for dampening the process noise and is applied before the limit functions. The units are in time constant seconds.

The Logic group has four selections (Logic1, Logic 2, Logic 3, and Logic 4). Logic selections have 2 input sources (A and B) and a selection for the Logic Type – “AND”, OR”, or LATCH.
The sources can be any Digital Signal Source (Table 6-4).
Enter to enter the sub-menu.
Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Configuration

Parameter Selection or Range
Logic 1 Logic 2 Logic 3 Logic 4
AND -Turns digital output ON when input IN A Source and IN B Source are ON. Thus,
If all inputs are ON, then: OUT = ON.
If any input is OFF, then: OUT = OFF.
OR - Monitors Input A Source and Input B Source to set state of digital output signal.
If A = OFF and B = OFF, then OUT = OFF.
If A = ON and/or B = ON, then: OUT = ON.
LATCH – Sets and Resets Latch state of the Output. If A=ON, B=OFF The Output is Latched ON. If A=OFF, B=ON, The Output is Latched OFF. If A and B are ON the Output = ON If A and B are OFF the Output = Latch State.
Power On considerations. The output state of the latch is cleared on power on.
Any Digital Signal See Table 6-4
Input A logic source selections, and Input B logic source selections
Invert None (default) IN A IN B In A and B
You can invert Input A or Input B or both. If the input is inverted, an input line that is ON is seen as OFF
On Delay 0 to 120 seconds default = 0.0
An on-delay time value up to 120 seconds is available to prevent momentary logic gate output actions. Number of seconds the logic gate is true before activating the Output.

The Auxiliary group has four selections (Switch 1 and Switch 2) and (Func Gen 1 and Func Gen 2).
Switch selections have 2 input sources (A and B). A switch block is used to select between two analog signals. The switch block can be used for many monitor and control strategies. A Digital Signal Source (Table 6-4) when active will select the B input source of the switch as the output. The Switch Input sources can be any Analog Signal Source (Table 6-3). There are two switch blocks provided for general use.
Function Generators are used for pre-control linearizing of inputs (such as during pH titration). Function Generator selections have 2 input sources (Input 1 PV and Input 2 PV).
It generates an output characteristic curve based on up to 11 configurable “Breakpoints” for both Input (X) and Output (Y) values.
The figure below shows an example of using the Func Gen to characterize the PID control loop output for control valve operation using 9 breakpoints.
Compensating for control valve characteristic
Compensating for control valve characteristic

OUT9
OUT8
OUT8
OUT7
OUT7
FGEN OUTPUT
FGEN OUTPUT
OUT6
OUT6
OUT5
OUT5
OUT4
OUT4
OUT3
OUT3
OUT2
OUT2
OUT1
OUT1
X1 X2 X3 X4 X5 X6 X7 X8
X1 X2 X3 X4 X5 X6 X7 X8
PID OUTPUT
PID OUTPUT

Enter to enter the sub-menu.
Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Parameter Selection or Range of Setting Parameter Definition
Switch 1 Switch 2
Any Analog Signal See Table 6-3
Analog Signal Source – Process signal to be monitored by the Alarm. Any analog source such as PV, Temperature, Pharma, Math, Function Generator, Switch, PID, or Calculated Values* * units of measure between the two input boards must be similar
Any Digital Signal See Table 6-4
Digital Signal Source when active will select the B input source of the switch as the output
Switch InA InB Select B
If Select B is OFF then Switch Output = In A
If Select B is ON then Switch Output = In B
Func Gen 1 Func Gen 2
Source None Input 1 PV Input 2 PV PID 1 Out PID 2 Out
Function Generator selections have 2 input sources (Input 1 PV and Input 2 PV).
The X (n) value must be < X(n+1) value. Thus, if fewer than 11 breakpoints are needed, be sure to configure any unneeded breakpoints with the same X and Y values used for the previous breakpoint.

Parameter Selection or Range of Setting Parameter Definition
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 0.000
X-value at Input Breakpoint 1 Y-value at Input Breakpoint 1
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 10.000
X-value at Input Breakpoint 2 Y-value at Input Breakpoint 2
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 20.000
X-value at Input Breakpoint 3 Y-value at Input Breakpoint 3
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 30.000
X-value at Input Breakpoint 4 Y-value at Input Breakpoint 4
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 40.000
X-value at Input Breakpoint 5 Y-value at Input Breakpoint 5
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 50.000
X-value at Input Breakpoint 6 Y-value at Input Breakpoint 6
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 60.000
X-value at Input Breakpoint 7 Y-value at Input Breakpoint 7
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 70.000
X-value at Input Breakpoint 8 Y-value at Input Breakpoint 8
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 80.000
X-value at Input Breakpoint 9 Y-value at Input Breakpoint 9
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 90.000
X-value at Input Breakpoint 10 Y-value at Input Breakpoint 10
–99999 to 999999 Default= 0.000 –99999 to 999999 Default= 100.000
X-value at Input Breakpoint 11 Y-value at Input Breakpoint 11
Configuration · Sub-menu · selection

PID (Option) - Proportional (P), Integral (I) and Derivative (D), (3-mode) control action based on the deviation or error signal created by the difference between the setpoint (SP) and the Process variable analog input value (PV). PID Tuning parameters are available.
Automatic tuning with Fuzzy Logic Overshoot Suppression can be configured. Other parameters listed in this group deal with how the analyzer will control the process including: PV High and Low, Setpoint High and Low limits, the Control Algorithm and Action, PID Tracking (TRV and TRC), Number of Tuning Parameter Sets and associated parameters, Setpoint Rate, Power-up Recall, Output Limits, Failsafe Output Value, Alarm setpoint type and value, and Alarm Hysteresis.
PID tracking is a means to control a PID’s output without the PID loop winding up. It is accomplished by the use of two inputs. TRC (tracking control) – selects the tracking mode (See Table 6-13) TRV (tracking value) – is the commanded output value in percentage (PID Output = TRV Input when TRC = ON) (See Table 6-13) When TRC is active, the front-panel display will indicate TRV for the PID loop.
Variables can be connected to TRC and TRV to allow remote control of the PID output. TRC can be connected to a digital variable TRV can be connected to an analog variable
Tracking is not the same as manual mode.
It may be desirable to use a discrete input to place a PID into manual momentarily to freeze the output. With tracking, this can be done.
When the discrete input is active, the output is frozen – MANUAL select.
When the discrete input is inactive, the PID runs in the auto mode – AUTO select.
Configuration

While TRV is active, the output can be adjusted using manual mode from the front-panel. After manual mode is terminated, the output will remain at the level because the output is tied to TRV.
Upgrading software on the UDA2182 to a new version will remove PID control (on units where PID has been ordered or been added in the Field).
Therefore, the following steps need to be followed in order to retrieve that option:
If PID was ordered when the unit was originally ordered:
If PID was added after the unit was originally shipped:
The PID will have to be reconfigured to settings prior to upgrade.
Configuration

(See “Maintenance” Menu item (Section 6.18), “System” selection to Enable PID Control)
Enter to enter the sub- menu. PID Control 1 and 2 are divided into 3 sections:
PID(n) Config (Table 6-13), PID(n)Tune (Table 6-14), PID(n)Alarms (Table 6-15)
Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Parameter Selection or Range of Setting
PID 1 Config PID 2 Config
PV Source None Input 1 PV (default) Input 2 PV
PV High -99999 to 99999 default = High Range of PV Input
Input Range of the PV - High Range Value These values are in units of that Input PV, such as 0-14pH.
PV Low -99999 to 99999 default = Low Range of PV Input
Input Range of the PV - Low Range Value These values are in units of that Input PV, such as 0-14pH.
SP High Limit -99999 to 99999 default = High Range of PV Input
Setpoint High Limit Value - prevents the setpoint from going above the value set here.
SP Low Limit -99999 to 99999 default = Low Range of PV Input
Setpoint Low Limit Value - prevents the setpoint from going below the value set here.
Output High Limit -99999 to 99999 default = 100.00
Output High Limit Value - is the highest value of output beyond which you do not want the automatic output to exceed.
Output Low Limit -99999 to 99999 default = 0.00
Output Low Limit Value - is the lowest value of output beyond which you do not want the automatic output to go below.

Parameter Selection or Range of Setting
PIDA (default) PIDB Duplex A Duplex B
PID A - is normally used for 3-mode control. The output can be adjusted somewhere between 100 % and 0 %. It applies all three control actions -Proportional (P), Integral (I), and Derivative (D) - to the error signal.
PID B - Unlike the PID-A equation, the analyzer gives only an integral response to a setpoint change, with no effect on the output due to the Gain or Rate action, and gives full response to PV changes.
In PID B, step changes in setpoint will not bump the output; the output will slew smoothly to the new value.
DUPA - like PID A but provides an automatic method to switch tuning constant sets.
DUPB - like PID B but provides an automatic method to switch tuning constant sets.
Control Action Direct Reverse (default)
DIRECT - PID action causes output to increase as process variable increases.
REVERSE - PID action causes output to decrease as process variable increases.
Power Mode Manual(default) Last
Mode permitted at power up.
Power Out Failsafe (default) Last
Output at Power up
FAILSAFE - Failsafe output value. LAST - Same as at power down.
Failsafe Out -5.00 to 105.00% default = 0.00
Failsafe Output Value – The Output value to which the analyzer will go if there is a power down or Failsafe condition (Input Faults).
Manual Select Any Digital Signal See Table 6-4
SP Power On Last
Setpoint at Power up LAST - Same as at power down.
Local SP – Local Setpoint value.
RSP Source Any Analog Signal See Table 6-3
Selects the analog signal that will be used as the remote setpoint. The remote setpoint should be supplied in PV engineering units.
Ratio 1.0 (Default) -20 to 20
Ratio that is applied to the Remote Setpoint.

Parameter Selection or Range of Setting
Bias 0.0 Default -9999 to 99999
Bias that is applied to the Remote Setpoint.
RSP Select Monitor (1 – 4) Logic (1 – 4) Digital In (1 – 2)
When this input is ON, the Remote Setpoint is used. If set to None, the operator can select the remote setpoint from the PID operator display.
FF Source Any Analog Signal See Table 6-3
Feed Forward value that is applied to the output. A change in the feed forward signal input will result in a proportional change in the output per the feed forward gain parameter.
FF Gain 1.000 (default) 0.1 to 1000.0 Feed Forward Gain used to calculate the change in the PID output based upon a change of the feed forward input signal.
TRC Select Any Digital Signal See Table 6-4
TRC (tracking control) – selects the tracking mode
TRV Select Any Analog Signal See Table 6-3
TRV (tracking value) – is the commanded output value in percentage (PID Output = TRV Input when TRC = ON)
When TRC is active, the front-panel display will indicate TRC for the PID loop.
Variables can be connected to TRC and TRV to allow remote control of the PID output. TRC can be connected to a digital variable TRV can be connected to an analog variable
Manual Permit Enable (default) Disable
Allows the operator to select Manual Operation of the PID loop from the PID Operator Display
Auto Permit Enable (default) Disable
Allows the operator to select Auto Operation of the PID loop from the PID Operator Display
LSP Permit Enable (default) Disable
Allows the operator to select the Local Setpoint from the PID Operator Display
RSP Permit Enable (default) Disable
Allows the operator to select the Remote Setpoint from the PID Operator Display

Parameter Selection or Range of Setting
PID 1 Tune PID 2 Tune
When enabled, the analyzer will start controlling to the setpoint while it identifies the process and adjusts the Gain or Proportional Band (P), Rate (I), and Reset Time (D) tuning constants in response to setpoint changes and/or Process Variable disturbances.
Fuzzy Overshoot Suppression minimizes overshoot after a setpoint change or a process disturbance.
The fuzzy logic observes the speed and direction of the PV signal as it approaches the setpoint and temporarily modifies the internal control response action as necessary to avoid an overshoot.
There is no change to the PID algorithm, and the fuzzy logic does not alter the PID tuning parameters.
This feature can be independently Enabled or Disabled as required by the application to work with Accutune.
When enabled, Proportional band is used instead of Gain (default). See “Gain or Prop Band”.
When enabled, Repeat per minute is used instead of Minutes per Repeat (default). See “Reset”.
Gain or Prop Band
Gain – 0.1% to 1000.0%
Gain (default) – is the ratio of output change (%) over the measured variable change (%) that caused it.
PB – 0.1 to 1000.0%
PB % where Prop Band is the proportional Band (in %of Input Range)
default = 1.000
Proportional Band (Prop Band) - is the percentage of the range of the measured variable for which a proportional controller will produce a 100 % change in its output.
-0.035 to 10.000
RATE action, in minutes affects the control output whenever the deviation is changing; and affects it more when the deviation is changing faster. The amount of corrective action depends on the value of Gain.
default = 0.000

Parameter Selection or Range of Setting
-0.02 to 50
RESET (Integral Time) - adjusts the control output according to both the size of the deviation (SP-PV) and the time it lasts. The amount of corrective action depends on the value of Gain.
default = 1.000
The reset adjustment is measured as how many times proportional action is repeated per minute (Repeats/minute) or how many minutes before one repeat of the proportional action occurs (Minutes/repeat – default).
None (default) Monitor 1 Monitor 2 Monitor 3 Monitor 4 Logic 1 Logic 2 Logic 3 Logic 4 Digital In 1 Digital In 2
Digital Source for selection of Tuning set 2.
When active, this input will override the current tuning set selection and force the PID to use tuning set 2. This applies for non-duplex type control.
Gain or Prop Band 2
0.1 to 1000.0 default = 1.000
Gain or Prop Band2 for Tuning Set 2. Same as Gain or Prop Band.
-0.035 to 10.000 default = 0.000
Rate 2 for Tuning Set 2. Same as Rate.
-0.02 to 50 default = 1.000
Reset 2 for Tuning Set 2 Same as Reset.
Parameter Selection or Range of Setting
PID 1 Alarms PID 2 Alarms
Alm 1 SP1 Type No Alarm (default) PV High PV Low Dev High Dev Low SP High SP Low Output High Output Low
Alarm 1 Setpoint 1 Type High PV Alarm: PV > Alm SP Low PV Alarm: PV < Arm SP High Deviation Alarm: |PV – SP| > Alm SP Low Deviation Alarm: |PV – SP| < Alm SP High Setpoint Alarm: SP > Alm SP Low Setpoint Alarm: SP < Alm SP High Output Alarm: Out > Alm SP Low Output Alarm: Out < Alm SP
Alm 1 SP1 Value -99999 to 99999 default = 0.000
Alarm 1 Setpoint 1 Value

Parameter Selection or Range of Setting
Alm 1 SP2 Type Same as Alarm 1 Setpoint 1
Same as Alarm 1 Setpoint 1 Type
Alm 1 SP2 Value -99999 to 99999 default = 0.000
Alarm 1 Setpoint 2 Value
Alm 2 SP1 Type Same as Alarm 1 Setpoint 1
Same as Alarm 1 Setpoint 1 Type
Alm 2 SP1 Value -99999 to 99999 default = 0.000
Alarm 2 Setpoint 1 Value
Alm 2 SP2 Type Same as Alarm 1 Setpoint 1
Same as Alarm 1 Setpoint 1 Type
Alm 2 SP2 Value -99999 to 99999 default = 0.000
Alarm 2 Setpoint 2 Value
Alm Hysteresis 0 to 100% default = 0.00
Alarm Hysteresis – an adjustable overlap of the ON/OFF states of each alarm.

Auto cycling provides automated timing, control and functionality for the cleaning and calibration of input probes. Each input PV has a dedicated auto cycle function block. The input board type and in the case of pH, the PV type, determines the level of auto cycling capability, as indicated below:
Input Board Type Auto Cycle Operation
Preamp pH Rinse, Auto Buffer Cal 1 (zero offset), Auto Buffer Cal 2 (slope)
pH/ORP, PV Type not ORP
Rinse, Auto Buffer Cal 1, Auto Buffer Cal 2
pH/ORP, PV type is ORP Rinse Only
DO ppm Rinse, Auto Air Cal
DO ppb Rinse, Auto Air Cal
Auto cycling is supported with setup menus, status displays and operational displays (Section 5.7) as well as event (Section 5.11) and calibration history logging (Section 11).
Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Configuration · Conductivity Rinse Only

Table 6-16 Auto Cycling Configuration
Parameter Selection or Range
Parameter Definition and Restrictions
Auto Cycle 1 Auto Cycle 2
Auto Cycling Disable (default) Enable
Allows auto cycling to be selected. This should be enabled after configuration is complete.
Hold Active Enable (default) Disable
When enabled, the output(s) sourced by input n for Auto Clean n is in hold during auto cycling.
Probe Transit Disable (default) Enable
When enabled, allows probe extract and probe insert sequence steps to occur and automated probe extract and insert parameters are made available.
Cycle Start Source None or Any Digital Signal See Table 6-4
Starts Auto Cycle on specific Digital Signal selected changing from 0 to 1.
Extract Wait Src (Probe Transit = Enabled)
None or Any Digital Signal See Table 6-4
Allows selection of a specific Digital Signal that causes a delay in the probe extraction sequence. While the selected digital input is active the probe extraction will not end unless a timeout occurs as determined by the duration configured in Probe Transit Mins. If a source is configured, the timeout results in an Auto Cycle Fail. If a digital signal is not available, the source may be left at “None” and the extract step will occur for the duration of Probe Transit Mins.
Insert Wait Src (Probe Transit = Enabled)
None or Any Digital Signal See Table 6-4
Allows selection of a specific Digital Signal that causes a delay in the probe insertion sequence. While the selective digital input is active the probe insertion will not end unless a timeout occurs as determined by the duration configured in Probe Transit Mins. If a source is configured, the timeout results in an Auto Cycle Fail. If a digital signal is not available, the source may be left at “None” and the extract step will occur for the duration of Probe Transit Mins.
Cycle Interval Off(default) Monthly Weekly Daily Custom
Frequency of Auto Cycle occurrence
Set specific time for Auto-Cycle to start.

Parameter Selection or Range
Parameter Definition and Restrictions
1 to 28 (default = 1)
Cycle Interval is Monthly
Sunday – Saturday (default = Sunday)
Cycle Interval is Weekly
1 to 31 (default = 1)
Cycle Interval is Custom, Start Time enabled
0 to 23 (default = 12)
Cycle Interval is Monthly, Weekly, Daily
0 to 59 (default = 0)
Cycle Interval is Monthly, Weekly or Daily Cycle Interval is Custom, Start Time enabled
0 to 100(default = 0) The period day parameter allows the selection of how often the Auto cycle will occur. For Example: 20 means that the Auto Cycle will occur every 20 days.
0 to 23 (default = 1) For Example: 4 means that the Auto-cycle will occur every 4 hours when the days and minutes are set to 0.
Period Mins 0 to 59 (default = 0) For Example: 30 means that the Auto-cycle will occur every 30 minutes when the days and hours are set to 0.
Rinse Cycle Cnt 0 to 100(default = 1) Allows selection for frequency of rinse occurrence.
0 to 100(default = 1) Allows selection for frequency of calibration occurrence. For Example: 1 indicates that a calibration will occur every cycle while a 10 indicates that a calibration will occur every 10th cycle.
Cal 1 Cycle Cnt
0 to 100(default = 1) Allows selection for frequency of calibration occurrences.
Cal 2 Cycle Cnt
0 to 100(default = 1) Allows selection for frequency of calibration occurrences.
0 to 30.00 (default = 0.50)
Maximum Probe Transit time in minutes.
(Probe Transit = Enabled)
Rinse Mins 0 to 30.00 (default = 0.50)
Duration of Rinse sequence in minutes
0 to 30.00 (default = 0.50)
Maximum calibration time in minutes. Enough time should be entered to allow process stabilization and a 20 second measurement time.


Resume Dly Mins 0 to 30.00 (default = 0.50)
The example in Table 6-17 configures the UDA to perform a rinse function once per day, at 8:00 AM, and once per week perform a 1 point Standardization, using 7 buffer. Then once every 4 weeks, perform a complete 2 point Standardize & Slope, using 7 buffer and 4 buffer. Also assume that the sensor is retracted from the sample line during the rinse and cal (not required).
Parameter Selection or Range
Parameter Definition and Restrictions
Auto Cycle 1 Auto Cycle 2
Auto Cycling Enable Allows auto cycling to be selected. This should be enabled after configuration is complete.
When enabled, the output(s) sourced by input n for Auto Clean n is in hold during auto cycling.
Probe Transit Enable When enabled, allows probe extract and probe insert sequence steps to occur and automated probe extract and insert parameters are made available.
Cycle Start Source None
Digital Signal 1 See Table 6-4
This is the end of travel “out” switch on the extraction device
Digital Signal 2 See Table 6-4
This is the end of travel “in” switch on the extraction device
Cycle Interval Daily Frequency of Auto Cycle to occur daily
Start Hour 8 Cycle to start at 8:00 AM
Rinse Cycle Cnt 1 Rinse to occur every cycle.
Cal 1 Cycle Cnt 7 Standardize occurs once every 7 cycles, or once per week.
Cal 2 Cycle Cnt
28 Slope Cal occurs once per 28 days, or every 4 weeks
0.5 If extraction takes longer than 30 seconds, then get a “AUTOCYCLE FAIL ALARM”
(Probe Transit = Enabled)
Rinse Mins 2 Each rinse duration is 2 minutes

Parameter Selection or Range
Parameter Definition and Restrictions
Max Cal Mins 2 If the reading is unstable after 2 minutes then get a “AUTOCYCLE FAIL ALARM”.
Resume Dly Mins 5 Wait 5 minutes after cycle completes and sensor is reinserted before removing HOLD and returning to On-Line mode.

The Variables menu allows you to configure the values that variables are set to when the UDA is first powered on. This group has two selections:
Analog This selection lets you configure the initial values of the Analog Variables.
Digital This selection lets you configure the initial values of the Digital Variables.
Enter to enter the sub-menu.
Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Parameter Selection or Range
Anlg Var 1 Init
-99999.99 to 99999.99 Initial Values of the Analog Variable applied at power on.
Anlg Var 2 Init
Anlg Var 3 Init
Anlg Var 4 Init
Dgtl Var 1 Init
Initial Values of the Digital Variable applied at power on.
Dgtl Var 2 Init
Dgtl Var 3 Init
Dgtl Var 4 Init

The communication menu allows you to configure the Communications Card. There are four selections:
IR Front Panel – configure the IR Front Panel interface
Modbus – configure the byte order
RS485 – configure the RS485 interface of the Communications Card.
Ethernet – configure the Ethernet interface of the Communication card.
Enter to enter the sub- menu.
Enter to display the group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Parameter Selection or Range
When enabled, Port Reset initializes the IR Interface. IR Front
Port Reset Off (default) Enable
Mode Enable(default) Setup Address Disable
Enable - allows IR to work anytime. No IR address required and on any front-panel screen. Setup – IR only works when the front-panel is in a setup screen. This will allow the IR interface to be password protected if a password is configured. No IR address required.
Address -- The UDA’s IR address must be used to communicate to the UDA.
Disable – The UDA will not respond to any request on the IR interface.
Modbus Word Swap Yes (default) No
Word Swap lets you set the word order for Modbus communications.
YES – sets the order to “Big Endian” format NO – sets the order to “Little Endian” format
RS485 Port Reset Off (default) Enable
Enable selection resets the Communication card. It should be enabled when the Address or Baud Rate or both are changed.

Parameter Selection or Range
Address 0 to 999 (default = 0) Modbus RTU Slave ID – 0 is offline
Baud Rate 2400 (default) 4800 9600 19200 38400 57600 115200
Modbus RTU Baud Rate
Port Reset Off (default) Enable
Enable selection resets the Communication card. It should be enabled when configurations for Ethernet are modified.
DHCP No (default) Yes
When YES, Dynamic Host Configuration Protocol server automatically assigns a dynamic IP address to UDA.
The set dynamic IP can be seen from “Comm Status” display.
IpAddr Octet 1 IpAddr Octet 2 IpAddr Octet 3 IpAddr Octet 4
0 to 255 These parameters are visible only when DHCP option is NO.
Allows you to assign Static IP address to the UDA.
SbntMsk Octet 1 SbntMsk Octet 2 SbntMsk Octet 3 SbntMsk Octet 4
0 to 255 These parameters are visible only when DHCP option is NO.
Allows you to assign Subnet Mask as per the local network settings
Dflt Gtwy Octet 1 Dflt Gtwy Octet 2 Dflt Gtwy Octet 3 Dflt Gtwy Octet 4
0 to 255 These parameters are visible only when DHCP option is NO.
Allows you to assign the Default Gateway as per the local network settings
DNS Srvr Octet 1 DNS Srvr Octet 2 DNS Srvr Octet 3 DNS Srvr Octet 4
0 to 255 These parameters are visible only when DHCP option is NO.
Allows you to assign the DNS server IP address as per the local network settings

Enter to enter the sub- menu.
Enter to display the group of parameters.
Enter to allow changes.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Parameter Selection or Range of
SW Version* Software version number
*See note at end of table
Language Language Set EE
Multi-language prompts guide the operator step- by-step through the configuration process assuring quick and accurate entry of all configurable parameters. Select from: English, French, German, Spanish and Italian (Language Set EE).
English (default) Italiano Deutsch Francais Espanõl
English, Russian and Turkish (Language Set RT)
English (default) Pусский Türkçe
English, Polish and Czech (Language Set PC)
English (default) Polski Česká
Read Only - Language Set
Read only language set of the software EE - English, French, German, Spanish and Italian RT - English, Russian and Turkish PC - English, Polish and Czech
Mains Freq 60 Hz (default) 50 Hz This function determines the frequency of AC line noise suppression for the input ADC circuitry.

Parameter Selection or Range of
Password 0000 (default) to 9999
Setup configuration, calibration and maintenance functions can be password-protected. The password can be any number between 1 and 9999 or letters. (When the password is zero, the operator will not be prompted to enter a password.) Follow the “ General Rules for Editing” to change the digits.
Unit ID Unit Identification Read Only
Option ID* Option ID Number = 0 if PID is not available = ID Number if PID available
*See note at end of table
PID Control* 1 Loop 2 Loops (default) Enables the PID Control configuration Parameters
*See note at end of table
Unit Reset No (default) Yes
Unit Reset initializes all calibration and configuration data to factory default values, with the exception of the Factory Temperature Calibration correction values
Inputs Input 1 Type Input 2 Type
Read Only pH/ORP - pH or Oxidation Reduction Potential pH Preamp – pH with preamplifier Conductivity (Dual if both inputs same) DO ppm - Dissolved Oxygen DO ppb - Dissolved Oxygen

Parameter Selection or Range of
The NIST system of conductivity measurement uses units of centimeters, and in the UDA are specifically µS/cm and mS/cm for conductivity and KΩ-m for resistivity.
The ISO system of conductivity measurement uses units of meters, and in the UDA are specifically μS/m, mS/m and S/m for conductivity and KΩ-m for resistivity.
The conductivity “Units Type” selected here affects the “PV Type” selections available under “Setup/Inputs/pH Preamp n. Selection of units type will scale all live conductivity and Resistivity readings on monitor and input calibration screens according to the factors listed:
NIST ISO ISO/NIST factor
μS/cm μS/m 0.01
μS/cm mS/m 10
MΩ-cm KΩ-m 0.1 Cond Wire Size
AWG (default) Sq mm
“Wire Size Units” allows selection of either AWG or Square millimeters (Sqmm). When changing units, the wire size parameter value is not converted. A pop-up message warns you of this.
Cond Wire Len Feet (default) Meters
“Wire Length Units” allows selection of either meters or feet for the Wire Length parameter in the Inputs group. When changing units, the wire length parameter value is not converted. A pop- up message warns you of this. If the value is no longer within range, it will change to closest range limit.
Ammonia pH – Specific conductivity temperature compensation assumes Ammonia NH3 is the base reagent
Amine pH – Specific conductivity temperature compensation assuming a generic amine base.
CO2 determination by degassed conductivity.

Parameter Selection or Range of
Temp Units º F
“Temperature Units” allows selection of either degrees C or degrees F for the display of measured temperature on monitor, pharmacopoeia, control and input calibration screens and for the entry of manual temperature input values in Setup/Inputs. When changing the temperature units, the manual temperature input value is not converted. A pop-up message warns you of this. If the value is no longer within range, it will change to the closest range limit.
Header 1 Label (default) 2 Labels Label/Time Date/Time
Determines the time and date displayed within the Monitor Display header.
Label Alphanumeric text (max 16 characters)
Header Format is 1 Label
Label Alphanumeric text (max 10 characters)
Header Format is Label/Time
Label 1 Alphanumeric text (max 10 characters)
Header Format is Label 1 / Label 2
Label 2 Alphanumeric text (max 10 characters)
Header Format is Label 1 / Label 2
Clr Evt Hist No (default) Yes
Clear Event History – Yes, clears the Event History Screen (see Section 5.11)
Clr Cal Hist No (default) Yes
Clear Cal History – Yes, clears the Calibration History Screen (see Section 11)

Select Tag and Press “Enter”. Follow the “General Rules for Editing” to edit the character string. Tag Names
Input 1 Input 2 PID Loop 1 PID Loop 2 Auto Cycle 1 Auto Cycle 2 Pharma 1 Alarm 1 Alarm 2 Alarm 3 Alarm 4
0 to 16 Characters
The real-time displays of process values show the instrument’s tag name (or other configurable fixed sixteen-character string) at the top of the screen.
Date Format YYYY/MM/DD (default) MM/DD/YYYY DD/MM/YYYY Time Format 24 Hour (default) 12 Hour Year 2005-2037
The parameters Date Format and Time Format determine how time and date are displayed in both the Monitor Display header and the Event History.
Month 1 to 12
Day 1 – 28, 29, 30 or 31 (default: 1) (determined by year & month) Hour 0 – 23 (24 Hour format) 1 – 12 (12 Hour format) AM/PM AM (default) PM Time Format is 12 Hour
Minutes 0 – 59
Display Test action occurs when the “Enter” key is pressed to accept the selection. Tests
Display Test Off (default) Enable
Keypad Test Off (default) Enable
When the keyboard test is enabled, the Status Message area displays the name of the key currently pressed.
Configuration · Clock

Output Level 1 Output Level 2 Output Level 3
Off (default) 0% 25% 50% 75% 100% Low Limit High Limit
Output action occurs when the “Enter” key is pressed to accept selection.
Actual output current is consistent with selected current range of 0 to 20 mA or 4 to 20 mA.
Relay 1 State Relay 2 State Relay 3 State Relay 4 State
Off (default) Energized De-energized
Relay state action occurs when the “Enter” key is pressed to accept selection.
Upgrading software on the UDA2182 to a new version will remove PID control (on units where PID has been ordered or been added in the Field).
Therefore, the following steps need to be followed in order to retrieve that option:
If PID was ordered when the unit was originally ordered:
If PID was added after the unit was originally shipped:
Configuration

Inputs and Outputs Wiring
This section contains instructions for wiring the inputs and outputs of the Analyzer.
The topics in this section are listed below.
7.1 Overview 114
7.2 General Wiring Practices 115
7.3 Inputs and Outputs 117
7.4 Direct pH/ORP Input Wiring Diagrams 120
7.5 pH Input from External Preamplifier/Cap Adapter Wiring Diagrams 126
7.6 Conductivity 130
7.7 Dissolved Oxygen 131
7.8 Communications Card 133
7.9 Outputs 134
7.10 Option Card 135
Topic See Page

Inputs and Outputs Wiring
Qualified personnel should perform wiring only.
WARNING
A disconnect switch must be installed to break all current carrying conductors. Turn off power before working on conductors. Failure to observe this precaution may result in serious personal injury.
An external disconnect switch is required for any hazardous voltage connections to the relay outputs.
To avoid damage to the case when connecting to a rigid metallic conduit system, the conduit hub must be connected to the conduit before the hub is connected to the enclosure
ATTENTION
This equipment contains devices that can be damaged by electrostatic discharge (ESD). As solid-state technology advances and as solid-state devices get smaller and smaller, they become more and more sensitive to ESD. The damage incurred may not cause the device to fail completely, but may cause early failure. Therefore, it is imperative that assemblies containing static sensitive devices be carried in conductive plastic bags. When adjusting or performing any work on such assemblies, grounded workstations and wrist straps must be used. If soldering irons are used, they must also be grounded. A grounded workstation is any conductive or metallic surface connected to an earth ground, such as a water pipe, with a 1/2 to 1 megohm resistor in series with the ground connection. The purpose of the resistor is to current limit an electrostatic discharge and to prevent any shock hazard to the operator. The steps indicated above must be followed to prevent damage and/or degradation, which may be induced by ESD, to static sensitive devices.
In applications where either the power, input or output wiring are subject to electromagnetic disturbances, shielding techniques will be required. Grounded metal conduit with conductive conduit fittings is recommended.

Inputs and Outputs Wiring
Instrument wiring should conform to regulations of the National Electrical Code.
Table 7-1 Recommended Maximum Wire Size
Gage Number mm2 Description
14 2.081 power, relays, and PE (protective earth)
18 0.823 inputs
18 0.823 isolated outputs
In applications where plastic conduit or open wire trays are used, shielded milticonductor 22 gage (0.326 mm2) or heavier signal input wiring is required.
Instrument wiring is considered Level 1, per section 6.3 of IEEE STD. 518 for plant facilities layout and instrumentation application. Level 1 wiring must not be run close to higher level signals such as power lines or drive signals for phase fired SCR systems, etc. Unprotected input wiring in high electrical noise environments is subject to electromagnetic, electrostatic, and radio frequency interference pickup of sufficient magnitude to overload input filters. The best instrument performance is obtained by keeping the interfering signals out of the instruments altogether by using proper wiring practices.
Refer to the following when wiring the unit.

Inputs and Outputs Wiring
The analyzer can accept single or dual inputs from Honeywell Direct pH, pH Input from External Preamplifier, ORP, Contacting Conductivity and Dissolved Oxygen sensors.
Two analog outputs standard
One additional output optional
Two electromechanical relays standard
Two additional relays optional
Wiring these inputs and outputs is described here.
The wiring is easily accessible through the front and the boards can be pulled out to facilitate the wiring of sensor input.
Open the case.
The display cable can become loose from the connector on the display board. Follow these instructions for re-inserting the cable into the connector:
Open connector by carefully lifting connector as shown:

Once the cable has been inserted carefully close the connector.
Loosen the four captive screws on the front of the bezel. Grasp the bezel on the right side. Lift the bezel gently and swing the bezel open to the left.
Two Digital Inputs

Inputs and Outputs Wiring

Option Board Location Option Board Location Communications Board Location Communications Board Location
Inside case with door open
Inside case with door open
Procedure
WARNING
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open the case while the unit is powered.

Inputs and Outputs Wiring
1 Go to Configuration setup to view the displays showing analog input, relay, and analog output use. Note the assignments shown. You must wire the unit to match these assignments in order for the analyzer to work as expected (See Section 6).
Turn off the power to the analyzer. More than one switch may be required to remove power.
3 Refer to Figure 7-1 for the location of the terminal board retainer. Loosen the screws that hold the retainer and slide the retainer left until the retainer tabs disengage from the terminal boards.
4 Insert a screwdriver into the tab in the terminal board to be wired and pull out gently. Slide the board half way out. There is a notch in the terminal board into which you can slide the retainer tabs and hold the boards in place while wiring.
5 Connect the inputs from the electrode or cells to the terminals in accordance with the configuration setup assignments. Refer to the wiring diagram provided with the electrode or cell, and to Figure 7-2 through Figure 7-20
6 Analog outputs (In addition to the standard outputs, one more is available as an option). See Option Board Wiring - Figure 7-20). Connect the outputs from the Analyzer terminals in accordance with the configuration setup assignments. Refer to the wiring diagrams provided with the field devices receiving the signals, and to Figure 7-2 through Figure 7-20.
7 If the relay outputs are to be used, leave the unit open and powered down. The relays can be used for Time Proportioning Output, Pulse Frequency Output, and Digital Output control as well as alarm annunciation. (In addition to the standard relays, two more are available as an option. See Option Board Wiring - Figure 7-20). Connect the outputs from the Analyzer terminals in accordance with the configuration setup assignments. Refer to the wiring diagrams provided with the external device and to Figure 7-2 through Figure 7-20.
These relays can be programmed to de-energize or energize on alarm. Use the Maintenance configuration setup to specify relay state. (NOTE 1)
8 Slide the retainer to the left then slide the terminal board back into place. Slide retainer to engage the tabs and tighten the screws.
9 Close the Bezel and secure four captive screws to a torque value of .20Nm (1.5 Lb-in). Power up the unit. Do not apply power until the bezel is closed.
Note 1: If set to de-energize on alarm, this means that when an alarm occurs (or the discrete control point becomes active), the relay coil will be de-energized. The NC contacts will then be closed and the NO contacts will be open. Conversely, during normal non-alarm operation (or when the control point is not active) the NC contacts will be open, and the NO contacts will be closed. If de-energize on alarm is selected, a power loss will force all relays to the same position as an alarm condition.
Step Action

Inputs and Outputs Wiring
Cable shield (yellow) to chassis ground screw
Cable shield (yellow) to chassis ground screw

Green with Black stripe
Green with Black stripe
White with Black stripe
White with Black stripe
Orange with Black stripe
Orange with Black stripe
Red with Black stripe
Red with Black stripe
Black with White stripe
Black with White stripe
Blue with Black stripe
Blue with Black stripe

Inputs and Outputs Wiring
Cable shield (yellow) to chassis ground screw
Cable shield (yellow) to chassis ground screw

Green Green with Black stripe
Green Green with Black stripe
White with Black stripe
White with Black stripe
Orange with Black stripe
Orange with Black stripe
Red with Black stripe
Red with Black stripe

Inputs and Outputs Wiring
White with Black stripe
White with Black stripe
Clear (center conductor of coax)
Clear (center conductor of coax)
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards

Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards

Inputs and Outputs Wiring
Cable shield (Yellow) to chassis ground screw
Cable shield (Yellow) to chassis ground screw
Shield Red or Clear (center conductor of coax)
Shield Red or Clear (center conductor of coax)
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Cable shield (Violet) to chassis ground screw
Cable shield (Violet) to chassis ground screw
Cable shield (Violet) to chassis ground screw
Orange Black pigtail of Coax
Orange Black pigtail of Coax
Center conductor of Coax
Center conductor of Coax
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Quick Disconnect Option

Inputs and Outputs Wiring

RTH 3rd Wire Red
RTH 3rd Wire Red
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards

Inputs and Outputs Wiring
White pigtail of Coax
White pigtail of Coax
Center conductor of Coax
Center conductor of Coax
Center conductor of Coax
Center conductor of Coax
RTH Sense RTH Low
RTH Sense RTH Low

Inputs and Outputs Wiring

RTH 3rd Wire
RTH 3rd Wire
Preamplifier
1 When using 022283 preamplifier module, jumper between “SC” and “ET” at the preamp
2 Color of wires corresponds to Honeywell cables:
Æ 31075723
Æ 51309677-001

Inputs and Outputs Wiring
Durafet II External Preamp
(+) 10 Volt Supply
(+) 10 Volt Supply
(-) 10 Volt Supply
(-) 10 Volt Supply
White Jumper terminals 5 and 6
White Jumper terminals 5 and 6
RTH 3rd Wire
RTH 3rd Wire
Preamplifier

Inputs and Outputs Wiring

Red Red with Black stripe
Red Red with Black stripe
Red Red with Black stripe
Cable shield (yellow) to chassis ground screw
Cable shield (yellow) to chassis ground screw
Cable shield (yellow) to chassis ground screw

Inputs and Outputs Wiring

Red with Black stripe
Red with Black stripe
Red with Black stripe
Cable shield (yellow) to chassis ground screw
Cable shield (yellow) to chassis ground screw
Cable shield (yellow) to chassis ground screw

Inputs and Outputs Wiring
4 Wire Cond. 18AWG (Has no shield)
4 Wire Cond. 18AWG (Has no shield)
4 Wire Cond. 18AWG (Has no shield)
4 Wire Cond. 18AWG (Has no shield)
Cable Shield (Violet) to chassis ground screw
Cable Shield (Violet) to chassis ground screw
Center conductor of Coax
Center conductor of Coax
RTH 3rd Wire
RTH 3rd Wire

Inputs and Outputs Wiring
Cable shield (Blue) to chassis ground screw Cathode
Cable shield (Blue) to chassis ground screw Cathode
* Older Dissolved Oxygen probes may have a White/Black Guard wire instead of a Black Guard wire.
* Older Dissolved Oxygen probes may have a White/Black Guard wire instead of a Black Guard wire.
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
CAUTION
When installing the probe, the wiring must be done in the order shown below even if the analyzer is not powered. This is because the DO Input card is continuously supplying a voltage (bias) to the terminals.
Red Green Coax (clear) Guard (Black) Yellow Orange
Orange – first Yellow Guard (Black) Coax (clear) Green Red Blue Shield Wire

Inputs and Outputs Wiring
Cable shield (Violet) to chassis ground screw Cathode
Cable shield (Violet) to chassis ground screw Cathode
Black Pigtail of Coax
Black Pigtail of Coax
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Some cables have connectors on the leads. Cut off the connectors, skin and tin the leads and then wire to the screw terminals on the boards
Quick Disconnect Option

Inputs and Outputs Wiring
RJ45 Ethernet Connection
TX + 4
TX + 4
TX-
TX-
RS 485 Connection
RS 485 Connection
SHIELD
SHIELD
Wire to Chassis Ground Screw
Wire to Chassis Ground Screw

Inputs and Outputs Wiring
Analog Output 1 (+)
Analog Output 1 (+)

Analog Output 1 (–)
Analog Output 1 (–)
Analog Output 2 (+)
Analog Output 2 (+)
Analog Output 2 (–)
Analog Output 2 (–)
Relay Output 1 (N.O.)
Relay Output 1 (N.O.)
Relay Output 1 (COM)
Relay Output 1 (COM)
Relay Output 1 (N.C.)
Relay Output 1 (N.C.)
Relay Output 2 (N.O.)
Relay Output 2 (N.O.)
Relay Output 2 (COM) Relay Output 2 (N.C.)
Relay Output 2 (COM) Relay Output 2 (N.C.)
AC N L2
Figure 7-19 Terminal Designations for Power, Analog Output, and Relay Output
AC Hot L1 · AC Hot L1

Inputs and Outputs Wiring
Analog Output 3 (+)
Analog Output 3 (+)
Analog Output 3 (–)
Analog Output 3 (–)
Digital Input 1 (+)*
Digital Input 1 (+)*
Digital Input 1 (–)* Digital Input 2 (+)*
Digital Input 1 (–)* Digital Input 2 (+)*
Digital Input 2 (–)* Case (earth) Ground
Digital Input 2 (–)* Case (earth) Ground
Relay Output 3 (N.O.)
Relay Output 3 (N.O.)
Relay Output 3 (COM) Relay Output 3 (N.C.)
Relay Output 3 (COM) Relay Output 3 (N.C.)
Relay Output 4 (N.O.)
Relay Output 4 (N.O.)
Relay Output 4 (COM) Relay Output 4 (N.C.)
Relay Output 4 (COM) Relay Output 4 (N.C.)
* Contact Closure only
* Contact Closure only

The section describes the calibration procedures for the following:
Input Cal – calibrate Input 1 and Input 2 for pH/ORP, Conductivity, or Dissolved Oxygen.
For other Calibration Procedures refer to the sections listed below.
Output Cal – calibrate Analog Output 1, Analog Output 2, and Analog Output 3 (See Section 1).
Temp Input Cal – calibrate Temperature 1 and Temperature 2 for pH/ORP or Conductivity (See Section 1).
For Calibration History, refer to Section 11.
The topics in this section are listed below.
8.1 Overview 136
8.2 Calibration Menu 137
8.3 pH/ORP and Conductivity Overview 138
8.4 Recommendations for Successful Measurement and Calibration 139
8.5 pH Calibration 140
8.6 ORP Calibration 151
8.7 Conductivity Calibration 157
8.8 Dissolved Oxygen Calibration 166
Input Calibration · Topic See Page

Calibrate. The Main Calibration Menu will appear.
CALIBRATION
Input PV Cal
Input PV Cal
Input Temp Cal
Input Temp Cal
Output Cal Cal History
Output Cal Cal History
Use the keys to highlight the “Input PV Cal” selection.
Enter to display the sub-menu for that selection. Depending on the Input board installed, you can select from: IN 1 or 2 pH/ORP Cal IN 1 or 2 Pre pH Cal IN 1 or 2 Conduc Cal IN 1 or 2 DO Cal
Use the keys to highlight the Input selection for calibration. Refer to the following sections for “calibration instructions”:
8.5 pH Calibration Page 140
8.6 ORP Calibration Page 151
8.7 Conductivity Calibration Page 157
8.8 Dissolved Oxygen Calibration Page 166
Input Calibration · Press · Press

Calibration of pH or ORP measuring instruments is necessary because similar electrodes may produce slightly different potentials in the same solution, requiring a corrective adjustment at the measuring instrument. Also, electrode outputs change over a period of time, making periodic recalibration necessary for best performance. Determine recalibration intervals based on operating experience.
Each type of cell has an associated cell constant entered during Configuration Setup. (See Section 6.6) This number is part of the cell model number. However, for greater precision, every Honeywell cell is individually tested at the factory, and a calibration factor unique to that cell is determined. The cal factor for a cell can be found on the plastic tag hanging from the cell lead wires. Instructions for entering this cell cal factor are in Section 6.6. The UDA automatically uploads the Cal Factor from Honeywell cells with EEPROM. This value is displayed in the “Setup” menu.
For some conductivity applications even greater accuracy is required. For those applications it is possible to perform a calibration trim procedure. The Analyzer’s reading can be adjusted while the associated cell is measuring a reference solution of known conductivity, as described in Section 8.7. The same procedure can be used to adjust the Analyzer’s reading while the cell is in the process, if a reference instrument is used to determine the conductivity of the process. In this case the process fluid becomes the “reference solution”.
Calibration trim is recommended for acid concentration applications above 5%.
Calibration trim can be reset as described in Section 8.7.
For accurate measurement of total dissolved solids (TDS) a conversion factor is entered for each cell as described in Section 6.6.
Any time a unit reset is performed; the TDS Value will be reset to 1.0. Calibration trim and cal factor will be reset for cells.
Input Calibration

Successful measurements and calibration depend upon selection and care of the electrode system or cells. Always prepare electrodes or cells and their mountings in accordance with the instructions supplied with them, observing temperature, pressure and flow limitations. Note the following recommendations:
For successful measurement in pure water applications where plastic piping is used, you may have to provide an earth ground for the cell. Run a wire from the black electrode terminal of the cell to one of the earth ground screws.
Input Calibration

pH instrument calibration consists of standardization and slope adjustments. Standardization is a pH Offset adjustment to compensate for electrode drift. Slope adjustment is a span adjustment to match the gain of the instrument to the electrode output response. For Durafet III pH electrodes, initial factory default value of offset and slope are automatically uploaded by the UDA. These values will appear in the “pH/ORP Cal” screens Table 8-2, step 4.
The analyzer supports two methods of calibration:
The two reference solutions must have a pH difference of at least 2.
When a Durafet III pH electrode is replaced, its electrode calibration data needs to be updated by the UDA2182 Analyzer. This is done either by power cycling the analyzer or using the restart screen.
Repeatedly until the One Input or Two Inputs display screen appears.
Enter You will see:
Enter to restart the UDA.

Analyzer stores information on multiple buffers
The UDA2182 Universal Dual Analyzer contains (in its permanent memory) information on several commonly used buffer solution standards in three groups, including the pH versus temperature characteristics of each.
By command, the instrument will automatically select one of these buffers in the selected group and use its values in the calibration process. Automatic checks are included to ensure that reasonable and correct values are entered.
The procedure for using the automatic buffer recognition feature in an actual calibration is provided in Table 8-2.
The standard pH Buffer values are listed in Table 8-1.
Input Calibration

Calibrating the pH Offset (Standardization) –. In auto buffer recognition calibration, you can select one of the other buffer pH values directly above or below the recognized buffer value in the current buffer group. (See Table 8-1.)
Calibrating the Slope - In auto buffer recognition calibration, you can select one of the other buffer pH values directly above or below the recognized buffer value in the current buffer group. (See Table 8-1.)
Temperature °C 0 5 10 15 20 25 30 35 40 45 50
NIST/USP 1.68 1.67 1.68 1.68 1.68 1.69 1.69 1.70 1.71
4.01 4.01 4.00 4.01 4.01 4.02 4.03 4.04 4.06
6.86 6.98 6.95 6.92 6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83
9.18 9.46 9.40 9.33 9.28 9.23 9.18 9.14 9.10 9.07 9.04 9.01
12.45 13.42 13.21 13.01 12.80 12.64 12.45 12.30 12.13 11.99 11.84 11.71
USA 2.00 2.01 2.01 2.01 2.00
4.00 4.01 3.99 4.00 3.99 4.00 4.00 4.01 4.02 4.03 4.04 4.06
7.00 7.13 7.10 7.07 7.05 7.02 7.00 6.99 6.98 6.97 6.97 6.97
10.00 10.34 10.26 10.19 10.12 10.06 10.00 9.94 9.90 9.85 9.82 9.78
12.00 12.60 12.44 12.28 12.14 12.00 11.88 11.79 11.66 11.53 11.43 11.32
Europe 1.00 0.98 0.98 0.99 0.99 1.00 1.00 1.01 1.02
3.00 3.02 3.00 3.00 2.99 2.99 2.98 2.98 2.97
6.00 6.03 6.02 6.01 6.00 6.01 6.02 6.04 6.05
8.00 8.15 8.11 8.07 8.03 8.00 7.97 7.94 7.91 7.88 7.87 7.86
10.00 10.22 10.17 10.12 10.05 10.00 9.95 9.90 9.86 9.82 9.78 9.74
13.00 13.81 13.60 13.39 13.19 13.00 12.83 12.68 12.53 12.38 12.25 12.11
Input Calibration · Group Buffer

Make sure you have selected “PV Type –pH Glass, pH Durafet, or pH HPW” in the Inputs configuration -Table 6-5.
Refer to Section 6.4.1 – General Rules for Editing.
1 Prepare containers of two standard reference solutions.
CALIBRATION
Output Cal Cal History
Output Cal Cal History
Use to select Input PV Cal
PV INPUT CAL
In 1 pH/ORP Cal In 2 Conduc Cal
In 1 pH/ORP Cal In 2 Conduc Cal
Use to select Input 1 or 2 pH/ORP Cal
IN 1 pH/ORP Cal
IN 1 pH/ORP Cal
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Use to select ”Buffer Group”
5 Press Enter Use to select NIST/USP (default)

IN 1 pH/ORP Cal
IN 1 pH/ORP Cal
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Use to select ”Auto Buffer Cal”
7 • Put the unit in “Hold” mode
8 Calibrating the pH Offset
“Place probe in Buffer 1” The display will show the pH of the buffer 1 solution as measured by the electrode system. The reading will be automatically adjusted to match the known pH value stored in the UDA2182 memory. “Press Enter when stable”
Press Enter Follow the prompts at the top and bottom of the screen.
9 Once the reading is stable
“Buffer 1 stability check” Use to change the value of the Buffer.
10 Rinse the electrode thoroughly with distilled or de-ionized water
11 Calibrating the Slope
“Place probe in Buffer 2” The display will show the pH of the buffer 2 solution as measured by the electrode system. The reading will be automatically adjusted to match the known pH value stored in the UDA2182 memory. “Press Enter when stable”
Press Enter Follow the prompts at the top and bottom of the screen.

12 Once the reading is stable
“Buffer 2 stability check” Use to change the value of the Buffer.
13 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Error Messages:
Buffer span too low
Make necessary adjustments and re-calibrate.
See Table 12-2 for definitions
Recommended for most applications
This technique is recommended for best accuracy in most applications.
Make sure you have selected “PV Type –pH Glass, pH Durafet, or pH HPW ” in the Inputs configuration - Table 6-5.
Refer to Section 6.4.1 – General Rules for Editing.

CALIBRATION
CALIBRATION
Output Cal Cal History
Output Cal Cal History
Use to select Input PV Cal
PV INPUT CAL
In 1 pH/ORP Cal In 2 Conduc Cal
In 1 pH/ORP Cal In 2 Conduc Cal
Use to select Input 1 or 2 pH/ORP Cal
IN 1 pH/ORP Cal
IN 1 pH/ORP Cal
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Use to select Buffer Cal
4 • Put the unit in “Hold” mode
5 Standardization (adjust instrument zero)
“Place probe in Buffer 1” The display will show the pH of the buffer 1 solution as measured by the electrode system. “Press Enter when stable”
Press Enter Follow the prompts at the top and bottom of the screen.

6 Once the reading is stable
“Change to Buffer 1 value” Use to change the value to match the actual pH of the Buffer 1 solution at its current temperature.
“Enter to save, Exit to cancel”
7 Rinse the electrode thoroughly with distilled or de-ionized water.
8 Percent Theoretical Slope Adjustment
“Place probe in Buffer 2” The display will show the pH of the buffer 2 solution as measured by the electrode system. “Press Enter when stable”
Press Enter Follow the prompts at the top and bottom of the screen.
9 Once the reading is stable
“Change to Buffer 2 value” Use to change the value to match the actual pH of the Buffer 2 solution at its current temperature.
“Enter to save, Exit to cancel”
10 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Error Messages:
Buffer span too low
Slope (Percent Theoretical Slope) underrange
Make necessary adjustments and re-calibrate
Slope(Percent Theoretical Slope) overrange
See Table 12-2 for definitions

Recommended where pH is stable, or for high-purity water applications
This method is recommended only where the pH is stable and changes very slowly. It is also recommended for high-purity water measurement applications. Special instructions for high-purity water applications are provided below.
To use the sample method, follow the instructions in Table 8-4.
Make sure you have selected “PV Type –pH Glass, pH Durafet, or pH HPW” in the Inputs configuration - Table 6-5.
Refer to Section 6.4.1 – General Rules for Editing.
1 Prepare the Calibration meter.
CALIBRATION
Output Cal Cal History
Output Cal Cal History
Use to select Input PV Cal
PV INPUT CAL
In 1 pH/ORP Cal In 2 Conduc Cal
In 1 pH/ORP Cal In 2 Conduc Cal
Use to select Input 1 or 2 pH/ORP Cal

IN 1 pH/ORP Cal
IN 1 pH/ORP Cal
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Use to select Sample Cal
5 • Put the unit in “Hold” mode
6 Press Enter Follow the prompts at the top and bottom of the screen.
“Place probe in Sample” The display will show the pH of the process as measured by the electrode system. “Press Enter when stable”
7 Collect a beaker of the process sample from a point near the electrode mounting and measure its pH value with a calibrated portable instrument.
Special instructions for high-purity water applications
For a high purity water application, do not remove the sample from the process for measurement. Bring the portable instrument to the sampling site and measure a continuously flowing sample that has not been exposed to air. This prevents lowering the sample pH by absorption of carbon dioxide from the air.
8 Once the reading is stable,
Change to Sample Value” Use keys to change the displayed value to match the value on the portable meter.
“Enter to save, Exit to cancel”
9 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Error Messages:
See Table 12-2.
Make necessary adjustments and re-calibrate

If the calibration is suspect, you can reset the pH Offset and pH Slope and calibrate again.
In the same screen as “Sample Cal”, use the keys to highlight “Reset pH Offset” or “Reset pH Slope”.
IN 1 pH/ORP Cal
IN 1 pH/ORP Cal
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
Buffer Group pH Offset pH Slope Reset pH Offset Reset pH Slope
(Read only) (Read only)
(Read only) (Read only)
Press ENTER. The pH Offset or pH Slope will be reset to (Factory calibration default for Durafet III).

ORP calibration consists of adjusting the reading of the analyzer to match a known value. There are two types of ORP calibration supported:
Recommended to adjust for changes in electrode potential over time
An ORP measuring system can be checked by measuring a solution having a known oxidation-reduction potential, then adjusting the UDA2182 to match. Although a reference solution provides only an approximation of ORP potential, the system can be adjusted periodically to compensate for changes in electrode potential over time.
The materials required to use the ORP standardization method are:
To prepare an ORP standardization solution, dissolve 0.1 g of quinhydrone powder in 5 cc of acetone or methyl alcohol (methanol). Add this to not more than 500 cc of a standard pH reference solution (buffer), about 1 part saturated quinhydrone to 100 parts buffer solution. The oxidation potential of this solution is listed below for several temperatures. The polarity sign shown is that of the measuring element with respect to the reference element.
These solutions are unstable and should be used within eight hours of preparation.
All mV values in Table 8-5 have a ± 30 mV tolerance.
Input Calibration

at Specified Temperature

Temperature 20 °C 25 °C 30 °C
4.01 @ 25 °C (31103001) 267 mV 263 mV 259 mV
6.86 @ 25 °C (31103002) 100 mV 94 mV 88 mV
7.00 @ 25 °C (not available from Honeywell) 92 mV 86 mV 80 mV
9.00 @ 25 °C **(not available from Honeywell) –26 mV –32 mV –39 mV
9.18 @ 25 °C (31103003) –36 mV –43 mV –49 mV
Make sure you have selected “PV Type –ORP” in the Inputs configuration – Table 6-5 Refer to Section 6.4.1 – General Rules for Editing.

Use to select Input 1 or 2 pH/ORP Cal
Input Calibration

4 Put the unit in “Hold” mode Remove the electrode from the process. Rinse the electrode thoroughly with distilled or de- ionized water
6 Follow the prompts at the top and bottom of the screen.
“Place probe in Sample” The display will show the Oxidation Reduction Potential of the reference solution as measured by the electrode system. “Press Enter when stable”
7 Once the reading is stable
“Change to Sample value” Use to change the value to match the actual oxidation- reduction potential of the reference solution at its current temperature.
“Enter to save, Exit to cancel”
This will standardize the unit.
9 Take the unit out of “Hold” and return to the calibration menu.
10 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Refer to Table 12-2.

Calibrates Analyzer only
The procedure described in this sub-section calibrates the Analyzer only. It does not involve compensating for electrode drift. Instead, a known millivolt signal is applied to the analyzer input terminals in place of the signal from the electrode, and the UDA2182 is adjusted so that its reading matches the known input.
This procedure can only be used when measuring ORP only
The materials required to calibrate the Analyzer using a voltage input are:
Make sure you have selected “PV Type –ORP” in the Inputs configuration – Table 6-5. Refer to Section 6.4.1 – General Rules for Editing.
To calibrate the ORP Analyzer using Voltage Input, follow the instructions in Table 8-7.
This procedure should be performed by qualified personnel only. Disconnect the power before opening the instrument case. A potentially lethal shock hazard exists inside the case if the unit is opened while powered. More than one switch may be required to disconnect power.
1 Turn off the power to the Analyzer. More than one switch may be required to disconnect power.
2 With the power off open the case:
Loosen the four captive screws on the front of the bezel.
Grasp the bezel on the right side. Lift the bezel gently and swing the bezel open to the left. (The bezel and display assembly is mounted on pivot arms.)
3 Refer to Figure 7-1 for the location of the terminal board retainer. Loose the screws that hold the retainer and slide the retainer right or left until the retainer tabs disengage from the terminal boards.
Input Calibration · Step Action Screen

4 Insert a screwdriver into the tab in the terminal board to be wired and pull. Slide the board half way out. There is a notch in the terminal board into which you can slide the retainer tabs and hold the boards in place while wiring.
5 Label and remove the input wiring from the input terminals. Terminals 8 and 10. (See Figure 7-6 Terminal Designations for ORP).
Slide the Input board back and close the case and power up the unit. Do not apply power until the case is closed.
PV INPUT CAL
In 1 pH/ORP Cal In 2 Conduc Cal
In 1 pH/ORP Cal In 2 Conduc Cal
Use to select Input 1 or 2 pH/ORP Cal
IN 1 pH/ORP Cal
IN 1 pH/ORP Cal
Sample Cal ORP Offset Reset ORP Offset
Sample Cal ORP Offset Reset ORP Offset
9 • Put the unit in “Hold” mode
The display will show the Oxidation Reduction Potential in Millivolts.
The value should match the Input signal.
11 Ignore the instructions to put the electrode in the reference solution. Instead, apply an appropriate millivolt signal (between –2000 and 2000 mV) to the input terminals.
To obtain a negative value, you must reverse the input to the unit as described in Step 5.

12 Once the reading is stable, if it does not match the input
“Change to Sample value” Use to change the value to match the Voltage being applied to the input terminals.
“Enter to save, Exit to cancel”
This will standardize the unit.
14 Take the unit out of “Hold” and return to the calibration menu.
15 Turn off the voltage source and turn off power to the Analyzer. Do not open the case until power is disconnected.
16 Reconnect field wiring removed in Step 5.
17 Re-insert the terminal board into the case.
18 Close the case and power up the unit. Do not apply power until case is closed.
If the calibration is suspect, you can reset the ORP Offset and calibrate again.
In the same screen as “Sample Cal”, use the keys to highlight “Reset ORP Offset”.
IN 1 pH/ORP Cal
IN 1 pH/ORP Cal
Sample Cal ORP Offset Reset ORP Offset
Sample Cal ORP Offset Reset ORP Offset
Press ENTER. The ORP Offset will be reset to 0.000 (default).

Each type of cell has an associated cell constant entered during Configuration setup (see Section 6.6). This number is part of the cell model number. However, for greater precision, every Honeywell cell is individually tested at the factory, and a calibration factor unique to that cell is determined. The cal factor for a cell can be found on the plastic tag hanging from the cell lead wires. Instructions for entering this cell cal factor are in Section 6.6. The UDA automatically uploads the Cal Factor from Honeywell DL4XXX type cells. This value is displayed in the “Setup” menu.
For some conductivity applications even greater accuracy is required. For those applications it is possible to perform a calibration trim procedure. The Analyzer’s reading can be adjusted while the associated cell is measuring a reference solution of known conductivity, as described in Table 8-9. The same procedure can be used to adjust the Analyzer’s reading while the cell is in the process, if a reference instrument is used to determine the conductivity of the process. In this case the process fluid becomes the “reference solution”.
Calibration trim is recommended for acid concentration applications above 5%.
Calibration trim can be removed as described in this section.
For accurate measurement of total dissolved solids (TDS) a conversion factor is entered for each cell as described in Table 6-5 (Input1, Input 2, Conductivity).
Introduction
Each type of cell has an associated cell constant; this number is part of the cell model number. The constant for each cell is entered during Input setup. However, for greater precision, every Honeywell cell is individually tested at the factory, and a calibration factor unique to that cell is determined. The cal factor for a cell can be found on the plastic tag hanging from the cell lead wires.
If you have not done so already, refer to Table 6-5 (Input1/Input2/Conductivity) to enter the cal factor for each cell
Introduction
The UDA2182 measures conductivity. However, the process value can be displayed in terms of total dissolved solids (TDS). If a TDS PV type was specified during Input setup (Section 6.6), then the same menu in will contain an entry for the TDS conversion factor for each cell.
Input Calibration

To determine the TDS conversion factor, it is first necessary to establish the total dissolved solids in a representative sample of the process. The formal determination of TDS is a laboratory standard method performed on a weighed grab sample of the process fluid. To summarize how to obtain a TDS value:
For detailed guidance in determining the official TDS, see “Standard Methods for the Examination of Water and Wastewater,” jointly published by the American Public Health Association, American Water Works Association and Water Pollution Control Federation, Washington, DC.
To determine the conversion factor needed by the Analyzer, first use the laboratory procedure summarized above to give an official TDS value. Next divide the TDS value by the conductivity of the sample to yield the conversion factor for that particular process fluid. The conversion factor is then entered into the analyzer to normalize the TDS readout.
With power plant cation conductivity measurements, ion chromatography results may be used to establish the conversion factor for readout in ppb chloride or sulfate ion. Nominal values are 83 ppb per μS/cm for chloride ion and 111 ppb per μS/cm for sulfate ion. The analyzer does not provide temperature compensation in TDS for chloride or sulfate ions.
As long as the entered TDS value is within the acceptable limits for a given cell constant, the Analyzer accepts the value. If a value is outside the accepted range, the unit will not display an error message; instead it will force the value to either the high or low limit of the range of the cell constant. Refer to Table 6-5 (Input1/Input2/Conductivity) for TDS conversion factor defaults.
If you intend to enter a cal factor or use calibration trim, do so before entering the TDS conversion factor as described here.
If you use calibration trim, first set the solution temperature compensation in Table 6-5 to the non-TDS choice for your process. For example, if you plan to use “NaCl” set the solution temperature compensation type to “NaCl” temporarily for calibration purposes. (Solution temperature compensation type is the one Input setup parameter that can be changed without triggering a cold reset.)
Next, perform the calibration. Once calibration has been completed, go back to Input setup and set the solution temperature compensation type to the TDS choice, for example “NaCl”. At this point you are ready to enter the TDS conversion factor as described in Table 6-5.
Input Calibration

Introduction
For most applications entering the cal factor for each cell will achieve satisfactory system performance. However, it is possible to perform a calibration trim procedure in which the Analyzer and cell combination are used to measure a reference solution of known conductivity; the reading of the Analyzer is adjusted to match.
The same procedure can be used to adjust the Analyzer’s reading while the cell is in the process, if a reference instrument is used to determine the process conductivity. In this case, the process fluid becomes the “reference solution”.
Calibration trim is recommended for acid concentration measurements above 5%.
To perform calibration trim using a standard reference solution, follow the instructions in Table 8-9.
Materials required are:
Input Calibration

Table 8-8 Conductivity of Potassium Chloride Solutions at 25 °C

* M = Molarity; 1M = 74.555g potassium chloride per liter of solution
Table 8-9 Procedure for Performing Calibration Trim Using a Reference Solution

Use to select Input 1 or 2 Conduc Cal
Input Calibration

6 Follow the prompts at the top and bottom of the screen.
“Place probe in Sample” The display will show the conductivity of the reference solution as measured by the cell and Analyzer system.
“Press Enter when stable”
7 Once the reading is stable,
“Change to Sample value” Use to change the value to match the actual conductivity of the reference solution at its current temperature.
“Enter to save, Exit to cancel”
This will save the Calibration Trim Value. If the calibration trim adjustment is successful, the calibration menu will again be displayed.
Return the cell to the process.
Repeat the operation for the other cell.
9 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Refer to Table 12-2.

If the calibration is suspect, you can reset the Calibration Trim and calibrate again.
In the same screen as “Sample Cal”, use the keys to highlight “Reset Trim”.
IN2 Conduc Cal
Sample Cal Cal Trim Reset Cal Trim
Sample Cal Cal Trim Reset Cal Trim
Figure 8-3 Resetting Calibration Trim
Press ENTER. The Calibration Trim will be reset to 1.00 (default).
Input Calibration

The UDA allows for a sample calibration of the specific or influent pH value. Here an independent sample is withdrawn from the sampling equipment and pH is determined with equipment of known accuracy. This independent pH value is then entered into the UDA as a pH calibration constant.
This method is recommended only where the pH is stable and changes very slowly. It is also recommended for high-purity water measurement applications. Special instructions for high-purity water applications are provided below.
To use the sample method, follow the instructions in Table 8-4.
Materials required are:
Make sure both inputs are “Conductivity”.
Refer to Section 6.4.1 – General Rules for Editing.
1 Prepare the Calibration meter.
CALIBRATION
Output Cal Cal History
Output Cal Cal History
Use to select Input PV Cal
Input 1 Conduc Input 2 Conduc Cation pH
Input 1 Conduc Input 2 Conduc Cation pH
Use to select Cation pH

CATION PH
Sample Cal pH Offset 0.00 Rst pH Offset
Sample Cal pH Offset 0.00 Rst pH Offset
Use to select Sample Cal
5 • DO NOT Remove the electrode from the process.
6 Once the reading is stable,
Change to Sample Value” Use keys to change the displayed value to match the value on the portable meter.
“Enter to save, Exit to cancel”
7 Follow the prompts at the top and bottom of the screen.
To recalibrate, press “Enter”.
“Enter = recal, Exit = exit”
8 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Error Messages:
See Table 12-2.
Make necessary adjustments and re-calibrate.

If the calibration is suspect, you can reset the Ph Offset and calibrate again.
In the same screen as “Sample Cal”, use the keys to highlight “Rst pH Offset”.
CATION PH
Sample Cal pH Offset Rst pH Offset
Sample Cal pH Offset Rst pH Offset
Figure 8-4 Resetting pH Offset
Press ENTER. The pH Offset will be reset to 0.00 (default).
Input Calibration

The analyzer supports three methods of Dissolved Oxygen calibration:
Air Calibration - is done with the probe removed from the process. This is the recommended method of calibration and should be completed unless the process set-up prohibits removing the probe. This is recommended prior to installation as it saves system parameters that are used in optimizing error diagnostics.
If the probe has just been removed from a sample low in dissolved oxygen, it takes longer to complete a calibration than that of a probe that is already near ambient conditions (sample high in dissolved oxygen).
Sample Calibration - Sample calibration allows a calibration based on a known dissolved oxygen concentration where a DO value may be entered that is based on a reference measurement. Sample calibration is usually executed by leaving the probe in the measured sample and adjusting the Analyzer to agree with the sample dissolved oxygen measured with a properly calibrated portable dissolved oxygen meter whose probe is held very close to the process probe.
For those situations where sample calibration is preferred, it is recommended that an Air Calibration be performed before the probe is put into service. It is also good practice to Air Calibrate the probe once every 2 - 4 months of service.
Pressure Compensation - The concentration of oxygen dissolved in air-saturated water depends on the air pressure. This dependence is automatically compensated for during air calibration using a pressure sensor built into the Analyzer. The purpose of the pressure calibration is to insure that the atmospheric oxygen level is known at the time of air calibration. Pressure compensation is only employed at the time of Air Calibration.
In this section there is also a procedure for running a Probe Bias Scan.
Do check the key parameters on the Display screen before performing an air calibration for the first time. The parameters should be within the following ranges:
Pressure: 500 to 800 mmHg Salinity: 0.0 if not being used Temperature should be a stable reading Don’t perform a probe bias test while the probe is in normal measurement service. Don’t perform an air calibration while the probe is in either the ppm or ppb process water. Don’t perform a sample calibration when the Dissolved Oxygen reading is in the 0.0 - 2.0 ppb range. Don’t measure the dissolved oxygen in gas streams or air streams. This product measures dissolved oxygen in water.
Input Calibration

Introduction
This is the simplest and most commonly used method of calibration.
If “Initial Installation”, power probe and analyzer for 24 hours before first air calibration.
Table 8-11 Calibrating a Dissolved Oxygen Probe Using Air Calibration Method
CALIBRATION
CALIBRATION
Output Cal Cal History
Output Cal Cal History
PV INPUT CAL
In 1 DO Cal
In 1 DO Cal
Use to select Input 1 or 2 DO Cal
IN1 DO CAL
Sample Cal Air Cal
Sample Cal Air Cal
Pressure Offset Reset Prs Offset
Pressure Offset Reset Prs Offset
Use to select Air Cal

4 Press Enter Follow the prompts at the top and bottom of the screen.
“Place probe in air” The display will show the live Dissolved Oxygen value.
Press Enter when ready”
This screen remains until the Air Calibration is complete. At that time the previous screen is displayed indicating that the air calibration is complete.
“Wait for cal complete”
6 “ Cal Complete”
This screen gives you an option to exit or recalibrate.
Press ENTER to recalibrate.
Press EXIT to return to Input Cal Screen.
7 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Refer to Table 12-2.
Air Calibration is not completed until both the probe temperature and the probe signal are stable. If the probe has just been removed from a sample low in dissolved oxygen or with temperature significantly different from the air temperature, it takes longer to reach stability than if the probe were already near ambient conditions when calibration was initiated.

Introduction
Sample calibration allows a calibration based on a known dissolved oxygen concentration. It is similar to air calibration except that the known DO value may be entered. Assuming an accurate reference is available, use the sample calibration method rather than air calibration if any of the following conditions apply:
Sample calibration is usually executed by leaving the probe in the measured sample and adjusting the analyzer to agree with the sample dissolved oxygen measured with a properly calibrated portable dissolved oxygen meter whose probe is held very close to the probe of the analyzer. Alternatively, the probe may be removed from the measured sample and placed in a sample of known dissolved oxygen concentration.
Table 8-12 Calibrating a Dissolved Oxygen Probe Using Sample Calibration
Method
1 Power the probe for at least one hour. (power the probe for 24 hours if initial installation)
CALIBRATION
Output Cal Cal History
Output Cal Cal History
PV INPUT CAL
In 1 DO Cal
In 1 DO Cal
Use to select Input 1 or 2 DO Cal

IN1 DO CAL
Sample Cal Air Cal
Sample Cal Air Cal
Pressure Offset Reset Prs Offset
Pressure Offset Reset Prs Offset
Use to select Sample Cal
5 • Put the unit in “Hold” mode, if required.
6 Press Enter Follow the prompts at the top and bottom of the screen.
“Place probe in sample” Immerse the probe in the sample of known DO concentration and wait until the DO reading is stable. “Press Enter when stable”
7 Once the reading is stable,
Change to sample value” Use the arrow keys to change the displayed value to match the value of the known sample DO concentration.
“Enter to save” when the value displayed equals the known sample DO concentration.
This screen gives you an option to exit or recalibrate.
Press ENTER to recalibrate.
Press EXIT to return to Input Cal Screen.
10 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Error Messages:
Make necessary adjustments and re-calibrate.
See Table 12-2.

Introduction
The concentration of oxygen dissolved in air-saturated water depends on the barometric pressure. This dependence is automatically compensated for during air calibration using a pressure sensor built into the Analyzer. The purpose of the pressure calibration is to calibrate that pressure sensor. However, this sensor has been factory calibrated and should not require re-calibration.
Determine the true ambient barometric pressure, such as from a calibrated pressure transmitter or a mercury barometer. Absolute barometric pressure is required - not the “relative” sea-level pressure normally reported by the weather bureau.
CALIBRATION
CALIBRATION
Output Cal Cal History
Output Cal Cal History
PV INPUT CAL
In 1 DO Cal
In 1 DO Cal
Use to select Input 1 or 2 DO Cal
IN1 DO CAL
Sample Cal Air Cal
Sample Cal Air Cal
Pressure Offset Reset Prs Offset
Pressure Offset Reset Prs Offset
Use to select Pressure Cal

4 Press Enter Follow the prompts at the top and bottom of the screen.
“Pressure Sensor Cal” Display shows the barometric pressure value in mm Hg.
“Press Enter when stable”
5 Once the reading is stable,
Change to sample value” Use the arrow keys to change the displayed value until the displayed pressure in mmHg agrees with the known pressure.
“Enter to save, Exit to cancel”
6 “ Cal Complete”
This screen gives you an option to exit or recalibrate.
Press ENTER to recalibrate.
Press EXIT to return to Input Cal Screen.
7 If the calibration fails, an error message will be displayed across the bottom stripe of the screen.
Error Messages:
See Table 12-2.
Make necessary adjustments and re-calibrate.
Introduction
The dissolved oxygen probe is an electrochemical cell, which produces an electric current that is directly proportional to the concentration of oxygen dissolved in the sample in which the probe tip is immersed. (When the probe is in air, the current is identical to that produced when the probe is in air-saturated water.) This current is a direct measurement of oxygen level. Usually, the probe is operated at -0.55V with respect to a reference electrode within the probe. (The minus sign is omitted from the screen as well as from the following discussion.) However, in some applications, the performance of the DO probe can be enhanced by using other bias voltages. The purpose of this test is to evaluate whether the probe bias voltage should be adjusted. Possible interference with probe performance may also be inferred from the Probe Bias Test (PBT).
When the test is initiated, the bias voltage is adjusted down from its original value (usually 0.55V) at 25mV/sec until 0V is reached. Then the bias voltage is driven up to 1.0 V at 25mV/sec and finally, it is driven down again until it has returned to the value it had just
Input Calibration · Step Action Screen · press Enter

before the test was initiated. During this voltage sweep, the probe current is monitored and the graph of current as a function of voltage is displayed.
If during the test the probe current rises above a factory-set upper limit, the bias voltage is returned to its pre-test value at 25mV/sec and the test is terminated without completing the full 1.0 Volt sweep. (The bias voltage test may also be terminated at any time by pressing the “EXIT” button.)
Under normal conditions, the completed display shows a graph of current as a function of voltage with the following features: from approximately 0 to 0.2 volts a fairly rapid increase in current is observed; from approximately 0.2 to 0.8 volts, the current exhibits a “flat” region where it is nearly independent of voltage and at some voltage above about 0.8 volts, the current rises quickly.
A typical current-voltage curve is shown below. The Sweep Bias millivoltage (along the bottom of the graph) is a voltage from 0 -1V that is applied to perform the test. The Operating Bias millivoltage is the current position of the cursor on the graph and represents the current bias voltage. The horizontal axis numerals are in hundreds of millivolts.
0.55V 80μA
0 0.2 0.4 0.6 0.8 μA 0
0 0.2 0.4 0.6 0.8 μA 0
Figure 8-5 Display of Probe Bias Test Done in Air
Note that the curve is quite flat at 0.55V. This means that even rather large changes in the probe current-voltage characteristic do not affect the current (and, thus, probe sensitivity) at 0.55V. In general, the curve formed by decreasing voltage is not identical to that formed by increasing voltage. This hysteresis is a function of the voltage scan rate and may be ignored.
As the bias voltage of the oxygen-consuming electrode (relative to an internal reference electrode) is increased, there is an initial increase in current as more and more of the oxygen that approaches the electrode is reacted. However, at about 0.2V, the current stops rising and a flat region, independent of voltage, is observed. It is in this region that probe current is determined by oxygen mass transport limitation. Increasing the voltage cannot increase the current because oxygen movement is diffusion limited. Finally, at a voltage
Input Calibration

exceeding 0.8 volts, a second process (water reduction) begins to occur and the current again rises. To achieve stable results, the probe should be operated within the flat region so that small changes in the probe characteristics result in negligible changes in probe current.
In some industrial wastewater applications, particularly those in petroleum refineries, active gases dissolved in the wastewater can cause this current-voltage characteristic to shift, moving the flat region to other, usually lower, voltages. Also, in some very rare instances, the chemical treatment of boiler water can cause this current-voltage characteristic to shift, moving the flat region to other, usually lower, voltages.
To summarize, the Probe Bias Test automatically varies the probe voltage while displaying the probe current as shown in the figure. At the completion of the test an opportunity to change the bias voltage is provided. Thus, even where significant gaseous contamination might otherwise interfere with the response of the probe to dissolved oxygen, this advanced feature allows the probe to operate.
(If the results of the probe bias test should ever be significantly different from those shown in the figure, Honeywell Service should be consulted.)
Table 8-14 Running a Probe Bias Scan
CALIBRATION
CALIBRATION
Output Cal Cal History
Output Cal Cal History
PV INPUT CAL
In 1 DO Cal
In 1 DO Cal
Use to select Input 1 or 2 DO Cal
IN1 DO CAL
Sample Cal Air Cal
Sample Cal Air Cal
Pressure Offset Reset Prs Offset
Pressure Offset Reset Prs Offset


4 Press Enter to initiate the Bias Scan screen
The bias voltage is adjusted down from its original value (usually 0.55V) at 25mV/sec until 0V is reached.
Then the bias voltage is driven up to 1.0 V at 25 mV/sec until “Scan complete” appears.
and finally, it is driven down again until it has returned to the value it had just before the test was initiated. During this voltage sweep, the probe current is monitored and the graph of current as a function of voltage is displayed.
At the completion of the test an opportunity to change the bias voltage is provided.
Input Calibration

6 Press Enter to save
Screen returns to “IN1 DO CAL” screen. Bias Volts will be indicated on the screen.
Input Calibration · Step Action Screen

If the calibration is suspect, you can reset any of these values and calibrate again.
In the same screen as “IN 1 DO Cal”, use the keys to highlight “Reset Prs Offset” or “Reset Bias Volts”.
IN1 DO CAL
Sample Cal Air Cal
Sample Cal Air Cal
Pressure Offset Reset Prs Offset
Pressure Offset Reset Prs Offset
Press ENTER. The selected value will be reset to (default).

The section describes the calibration procedures for the following:
Output Cal – calibrate Analog Output 1, Analog Output 2, and Analog Output 3
For other Calibration Procedures refer to the sections listed below.
PV Input Cal – calibrate Input 1 and Input 2 for pH/ORP, Conductivity or Dissolved Oxygen (See Section 8)
Temperature Cal – calibrate Temperature 1 and Temperature 2 for pH/ORP or Conductivity (See Section 1)
For Calibration History, refer to Section 11.
The topics in this section are listed below.
9.1 Overview 178
9.2 Output Calibration 179
Outputs Calibration · Topic See Page

The UDA2182 is available with two standard and one optional analog outputs. The output signals can be adjusted to trim the high and low output current or voltage values over a range of ± 0.4 % of span to compensate for component tolerance variations.
Calibrate. The Main Calibration Menu will appear.
CALIBRATION
Input PV Cal
Input PV Cal
Input Temp Cal
Input Temp Cal
Output Cal Cal History
Output Cal Cal History
Use the keys to highlight the “Output Cal” selection.
Enter to display the sub-menu for that selection.
Output calibration involves connecting a meter to the Analyzer’s output terminals. The meter required for output calibration depends on the type of outputs.
A screwdriver to fit the terminal block screws and the screw securing the terminal board retainer is also required.
To calibrate outputs, follow the procedure described in Table 9-1 Procedure for Calibrating Analyzer Outputs. The output terminals are inside the case as shown in Figures 6-1 through 6-6.
Outputs Calibration · Press · Press

While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open the case while the unit is powered. Do not access the output terminal as described below while the unit is powered.
A disconnect switch must be installed to break all current carrying conductors. Turn off power before working on conductors. Failure to observe this precaution may result in serious personal injury.
Outputs Calibration

Table 9-1 Procedure for Calibrating Analyzer Outputs
1 Turn off the power to the Analyzer. More than one switch may be required to disconnect power.
2 With the power off, open the case:
Loosen the four captive screws on the front of the bezel.
Grasp the bezel on the right side. Lift the bezel gently and swing the bezel open to the left.
3 Refer to Figure 7-1 for the location of the terminal board retainer. Loose the screws that hold the retainer and slide the retainer right or left until the retainer tabs disengage from the terminal boards.
4 Insert a screwdriver into the tab in the terminal board to be wired and pull. Slide the board half way out. - Refer to Figure 7-1 for the location. (Output 1 and 2 – Power Supply/Analog Output/Relay Output card) (Output 3 – Option card) There is a notch in the terminal board into which you can slide the retainer tabs and hold the boards in place while wiring.
5 Label and remove the field wiring from the output terminals. Output 1 – Terminals 12– and 13+ Output 2 – Terminals 10– and 11+ Output 3 – Terminals 14– and 15+
6 Feeding the test wiring through the conduit hole in the case and connect the appropriate type meter to the specific output terminals Be sure to observe the correct polarity.
Slide the Input board back and close the case and power up the unit. Do not apply power until the case is closed.
CALIBRATION
Output Cal Cal History
Output Cal Cal History
Use to select ”Output Cal”

OUTPUT CAL
Output 1 Output 2 Output 3
Output 1 Output 2 Output 3
Use to select an Analog Output to be
OUTPUT 1
20mA Offset 4mA Offset 0 0 Reset 20mA Offs Reset 4mA Offs
20mA Offset 4mA Offset 0 0 Reset 20mA Offs Reset 4mA Offs
Use to select ” 20 mA Offset”
OUTPUT 1
20mA Offset 4mA Offset -147 3 Reset 20mA Offs Reset 4mA Offs
20mA Offset 4mA Offset -147 3 Reset 20mA Offs Reset 4mA Offs
The right most digit will be “blinking”.
11 To correct the value on the meter:

12 Use to select ” 4 mA Offset” and repeat the process.
OUTPUT 1
20mA Offset 4mA Offset -147 3 Reset 20mA Offs Reset 4mA Offs
20mA Offset 4mA Offset -147 3 Reset 20mA Offs Reset 4mA Offs
13 Press “Enter” to store the 4mA Offset value.
Press “Exit” to cancel. The previous value is retained.
14 If the calibration is suspect, you can reset the 20mA and 4mA Offset and calibrate again.
To calibrate additional Outputs, repeat the above steps Including powering down the unit before changing the connections to the output terminals.
When output calibration has been completed, re-install the field wiring removed in step 5. Disconnect power before opening the case.
Close the case and power up the unit. Do not apply power until the case is closed.
Outputs Calibration · Step Action Screen

If the calibration is suspect, you can reset the 20mAand 4mA Offset and calibrate again.
In the same screen as “20mA and 4mA Offset”, use the keys to highlight “Reset 20mA Offset” or “Reset 4mA Offset”.
OUTPUT 1
20mA Offset 4mA Offset -147
20mA Offset 4mA Offset -147
3 Reset 20mA Offs Reset 4mA Offs
3 Reset 20mA Offs Reset 4mA Offs
Press ENTER. The 20mA Offset or 4mA Offset will be reset to 0(default).
Outputs Calibration

The section describes the calibration procedures for the following:
Temp Input Cal – calibrate (T1) Temperature 1 or (T2) Temperature 2 for pH/ORP or Conductivity
For other Calibration Procedures refer to the sections listed below.
PV Input Cal – calibrate Input 1 and Input 2 for pH/ORP, Conductivity or Dissolved Oxygen (See Section 8)
Output Cal – calibrate Analog Output 1, Analog Output 2, and Analog Output 3 (See Section 1)
For Calibration History, refer to Section 11.
The topics in this section are listed below.
10.1 Overview 185
10.2 Temperature Input Calibration 186
Temperature Input Calibration · Topic See Page

Temperature Input Calibration lets you monitor a live temperature reading while continuing to monitor the sample. The currently displayed temperature value can be edited through a series of prompts on the screen. The temperature offset value is always displayed in the temperature units selected in the Maintenance setup menu.
Calibrate. The Main Calibration Menu will appear.
CALIBRATION
Input PV Cal
Input PV Cal
Input Temp Cal
Input Temp Cal
Output Cal Cal History
Output Cal Cal History
Use the keys to highlight the “Input Temp Cal” selection.
Enter to display the sub-menu for that selection.
Table 10-1 Procedure for Calibrating the Temperature Inputs
CALIBRATION
CALIBRATION
Output Cal Cal History
Output Cal Cal History
Use to select ” Input Temp Cal”
TEMP INPUT CAL
T1 pH/ORP Cal T2 Conduc Cal
T1 pH/ORP Cal T2 Conduc Cal

Use the keys to highlight the desired “Temperature Input”
selection.
Temp Cal Temp Offset (Read only) Reset Tmp Offs
Temp Cal Temp Offset (Read only) Reset Tmp Offs
5 Follow the prompts at the top and bottom of the screen.
“Place probe in sample” The display will show the temperature of the reference solution as measured by the probe and Analyzer system.
“Press Enter when stable”
6 Once the reading is stable,
“Change to sample value” Use to change the value to match the actual temperature of the reference solution at its current temperature.
“Enter to save, Exit to cancel”
Limit is ± 5ºC (± 9ºF)
This will save the Temperature Offset value. If the calibration is not successful, an error message will be displayed.
8 If the calibration is suspect, you can reset the Temperature Offset and calibrate again.

If the calibration is suspect, you can reset the Temperature Offset and calibrate again.
In the same screen as “Temp Cal”, use the keys to highlight “Reset Tmp Offset”.
Temp Cal Temp Offset (Read only) Reset Tmp Offset
Temp Cal Temp Offset (Read only) Reset Tmp Offset
Press ENTER. The Temperature Offset will be reset to (default).
Temperature Input Calibration · T1 pH/ORP CAL · T1 pH/ORP CAL

11.1 Overview
Calibration History records every successful input or output calibration with timestamp, with detail available on cal type and before and after cal values by scrolling and selecting cal event name. Calibration records are listed top down from most recent to least recent. Each line in the list consists of a calibration event name and the date and time of occurrence.
Successful automatic cals from auto cycling also recorded and identified. Status warns of cal history at 50% and 90% and when erasing old records.
Calibrate. The Main Calibration Menu will appear.
CALIBRATION
Input PV Cal
Input PV Cal
Input Temp Cal
Input Temp Cal
Output Cal Cal History
Output Cal Cal History
Use the keys to highlight the “Cal History” selection. Press “Enter”.
Table 11-1 Cal History items
Item Values Item Values
Calibration type Sample Conduc Sample pH/ORP Buffer pH 1 Buffer pH 2 Auto Buffer pH 1 Auto Buffer pH 2 Auto Cycle pH 1 Auto Cycle pH 2 Sample DO Auto Air DO Auto Cycle DO The list is fully scrollable and individual records are selectable for further detail by highlighting specific event names and pressing Enter. The calibration history has a capacity of 128 records.
In 1 PV Cal In 2 PV Cal In 1 Temp Cal In 2 Temp Cal Out 1 4mA Cal Out 1 20mA Cal Out 2 4mA Cal Out 2 20mA Cal Out 3 4mA Cal Out 3 20mA Cal
Calibration History · Press · Calibration event name

Enter to enter the sub- menu.
Enter to enter the sub-menu.
Enter to allow change.
Enter to clear the Calibration History screen.
Calibration History

This section contains information on status and alarm messages, as well as on diagnostics and system error messages and Fail messages. All these messages are displayed on the “Status Message” stripe. If more than one message is active, the display will cycle through all the messages, and then repeat the cycle.
The topics in this section are listed below.
12.1 Overview 191
12.2 System Status Messages 192
12.3 Calibration Diagnostics 193
12.4 Auto Cycle Fail Messages 194
12.5 Pharma Fail Messages 195
Diagnostics and Messages · Topic See Page

The following table lists all the error messages that can appear for Measurement errors, Input errors, Output errors, and Alarm Conditions.
Table 12-1 Status Messages
HOLD ACTIVE Analog Inputs (PVs) are held at their last active levels by pressing the “HOLD” button, until cancelled by pressing the “HOLD” button again.
n = 1 or 2 Measurement Errors
TEMP n UNDERRANGE Measured temperature is less than the minimum range value according to measurement type, where: n is 1 (Input 1) or 2 (Input 2).
TEMP n OVERRANGE Measured temperature is greater than the maximum range value according to measurement type, where: n is 1 (Input 1) or 2 (Input 2).
PV n UNDERRANGE Measured PV is less than the minimum range value according to measurement type, where n is 1 (Input 1) or 2 (Input 2).
PV n OVERRANGE Measured PV is greater than the maximum range value according to measurement type, where n is 1 (Input 1) or 2 (Input 2).
n = 1 or 2 Input Errors – Output(s), for which Input is source, will go to failsafe level
PROBE TEMP n INPUT FAULT
Probe temperature sensor at Input n is defective.
PROBE PV n INPUT FAULT Probe PV sensor at Input n is defective.
PROBE n INPUT OUT OF SOLUTION
Probe at Input n is out of solution.
TEMP n INPUT OPEN Probe temperature sensor at Input n is not connected. Check Wiring.
PV n INPUT OPEN Probe PV sensor at Input n is not connected. Check Wiring.
INPUT BOARD n FAULT An input board disconnect while powered results in an input fault condition and this status message. The PV value is the failsafe parameter value and the temperature is at the lower limit of the input board’s temperature range. Re-insert the input board, or cycle the unit’s power if the input board is no longer needed.
n = 1, 2, or 3 Output Errors – Output in error goes to failsafe level
OUTPUT n OPEN Analog output n current is less than 3 mA and is less than output minimum mA value. Check wiring.
n = 1, 2, 3, or 4 Alarm Conditions
ALARM n ACTIVE Alarm number n is currently active.
Diagnostics and Messages · Status Message Definition

All of the possible errors are detected during a probe calibration and will abort the calibration process with the message “FAIL” appearing briefly, followed by a return to the online pH/ORP/DO display. At that point, the specific error will be displayed as described. In addition, any of following errors may occur during probe calibration and abort the calibration process.
BUFFER SPAN TOO LOW The span between pH buffer 1 and pH buffer 2 is less than 2 pH. Use a set of buffers that are at least 2 pH apart. As a warning status, will clear when an appropriate buffer 2 value is selected. As an error message, will abort calibration and preserve original slope value.
OFFSET UNDERRANGE Resulting pH offset (standardization) value is less than –2 pH after pH slope calibration. Calibration is aborted and original pH offset and slope values are preserved.
OFFSET OVERRANGE Resulting pH offset (standardization) value is greater than 2 pH after pH slope calibration. Calibration is aborted and original pH offset and slope values are preserved.
SLOPE UNDERRANGE Resulting pH slope is less than 80%. Calibration is aborted and original slope value is preserved.
SLOPE OVERRANGE Resulting pH slope is greater than 105%. Calibration is aborted and original slope value is preserved.
Resulting DO calibration factor is less than 0.001268. DO calibration is aborted and original calibration factor is preserved.
Resulting DO calibration factor is greater than 0.040580. DO calibration is aborted and original calibration factor is preserved.
DO probe current is less than 5 µA. DO bias scan is aborted and original bias voltage is preserved.
DO probe current exceeds the greater of 133% of the probe current at last successful calibration or 160 µA. During DO bias scan, scan is aborted and original bias voltage is preserved.
DO air PV or temperature readings too unstable for successful air calibration. Calibration is aborted and original calibration factor is preserved.
pH solution PV or temperature readings too unstable for successful auto buffer calibration. Calibration is aborted and original pH offset (for buffer 1) or slope value (for buffer 2) is preserved.
SOLUTION TEMP TOO LOW pH solution temperature readings less than minimum of 0 degrees C. Auto buffer calibration is aborted and original pH offset (for buffer 1) or slope value (for buffer 2) is preserved
SOLUTION TEMP TOO HIGH pH solution temperature readings greater than maximum of 100 degrees C. Auto buffer calibration is aborted and original pH offset (for buffer 1) or slope value (for buffer 2) is preserved
Diagnostics and Messages · Status Message Definition

Auto Cycle Fail is active whenever an auto cycle failure has occurred. The status message “Auto Cycle n Fail” is also displayed during a fail state. Once detected, the current cycle proceeds immediately to the Probe Insert step (if enabled) or to the Resume Delay step. The fail state remains for the duration of the Resume Delay, whereupon the fail state returns to 0 and the fail message is cancelled. A fail state also provides a detail message in the lower half of the Auto Cycle display regarding the specific reason for the error. These messages are listed below:
Probe Extract Timeout Probe Transit enabled, Extract Wait Src not None and state not 0 within Max Transit Mins of start of Probe Extract.
Probe Insert Timeout Probe Transit enabled, Insert Wait Src not None and state not 0 within Max Transit Mins of start of Probe Insert.
Input Fault Input board, PV or temperature fault has occurred during calibration (fault type in status message).
Solution Unstable PH PV or temperature not stable for calibration within elapsed time limit Max Cal Mins.
Buffer Span Too Low Difference of PV reading for pH Cal 2 (Slope) and that of last pH Cal 1 (zero offset) < 1.8 pH.
Offset Underrange PH calibration has calculated and rejected a zero offset < -2 pH.
Offset Overrange PH calibration has calculated and rejected a zero offset > 2 pH.
Slope Underrange PH calibration has calculated and rejected a slope < 80 %.
Slope Overrange PH calibration has calculated and rejected a slope > 105 %.
Readings Unstable DO PV or temperature not stable for calibration within elapsed time limit Max Cal Mins.
Probe Current Too Low Probe current is < 5 μA during DO calibration.
Diagnostics and Messages · Fail Message Reason

Pharma Fail is active whenever a Pharma failure has occurred. Status messages are also displayed during a fail state. These messages are listed below:
Warn Condition Diagnostic Message
Stage 1: Measured conductivity exceeds Pct Warning value.
PHARMA n PV LIMIT WARN
Fail Condition Diagnostic Message
PHARMA n PV OVERLIMIT
Stage 1: Temperature not within range of 0 – 100 degrees C.
PHARMA n TEMP OVERRANGE PHARMA n TEMP UNDERRANGE
Stage 2: Conductivity (due to uptake of atmospheric carbon dioxide) is 0.1 µS/cm or greater per 5 minutes
PHARMA n PV OVERLIMIT
PHARMA n PH UNDERRANGE
Stages 2 and 3: Temperature not within range of 24 – 26 degrees C.
PHARMA n TEMP OVERRANGE
PHARMA n TEMP UNDERRANGE
Status Condition Diagnostic Message
Diagnostics and Messages

For all information relating to the UDA2182 and Communications please refer to the UDA2182 Communications User Guide #70-82-25-126.
Ethernet and Communications

Accessories and Replacement Parts List
This section provides part numbers for field-replaceable parts and for accessories.
The topics in this section are listed below.
14.1 Overview 197
14.2 Part Numbers 198
Topic See Page

Accessories and Replacement Parts
Part numbers for field-replaceable parts and accessories are provided in Table 14-1.




(See Appendix B for titanium cells mounted into stainless steel flow chamber 31079198)
If you use standard Honeywell cell lead lengths of 7 or 20 feet connected directly to the Analyzer, no compensation for lead resistance is necessary. Similarly, if a junction box is used to extend the leads up to 150 feet, no compensation is required. However, if longer leads are used (greater than 150 feet), signal accuracy can be adversely affected unless you enter information that will permit the UDA2182 to compensate for lead resistance in the black and white cell leads only. Lead resistance compensation is not necessary, nor applied to the other cell leads.
For lengths up to 1000 feet*, simply specify the gauge and length as described in Table 6-5. Note that the maximum wire size for sensor inputs at the input terminal board is 16AWG. * DirectLine DL4000 series cells have a total lead length limit of 250 feet.
If mixed wired gauges are used, or lead length or wire gauge are not within the stated ranges, the UDA2182 can still perform the compensation. However, you must first calculate the lead resistance, and then put it in terms of the available settings for AWG gauge and length.
The resistance of each available gauge choice (in copper wire) is:
16 AWG = 4.0 ohms per 1000 feet
18 AWG = 6.4 ohms per 1000 feet
20 AWG = 10.2 ohms per 1000 feet 22 AWG = 16.1 ohms per 1000 feet
For example, suppose extension cables between the cell and Analyzer consist of 500 feet of 18-gauge wire and 200 feet of 16-gauge WIRE. The cell has the TC head option.
500 ft of 18 AWG wire 200 ft. of 16 AWG wire
Analyzer Honeywell Junction Conductivity Cell Box
Because there are two different types of wire used in each lead to the cell in this example, the total lead resistance is calculated as follows: (Note: the analyzer accounts for the fact that there is always a pair of conductor wires in the system loop.)
(0.5 x 6.4) + (0.2 x 4.0) = 4.0 ohms
Appendices

Since the analyzer only allows entry of one wire gauge type, we allow for the worst-case condition by dividing the total resistance by the resistance per thousand feet of the higher resistance gauge wire. In our example this would be:
4.0 ohms ÷ 6.4 ohms per thousand feet of 18 AWG wire = 625 feet
Therefore, in our example we would use the procedure in Table 6-5, and specify the wire gauge as 18 AWG and the length as 625 feet.
Appendices

(4973 or DL4311 Titanium cells mounted in stainless steel flow chamber 31079198)
If you use standard Honeywell cell lead lengths of 7 or 20 feet connected directly to the Analyzer, no compensation for lead resistance is necessary. Similarly, if a junction box is used to extend the leads up to 150 feet, no compensation is required. However, if longer leads are used (greater than 150 feet), signal accuracy can be adversely affected unless you enter information that will permit the UDA2182 to compensate for lead resistance in the black and white cell leads only. Lead resistance compensation is not necessary, nor applied to the other cell leads.
For lengths up to 1000 feet*, simply specify the gauge and length as described in Table 6-5. Note that the maximum wire size for sensor inputs at the input terminal board is 16AWG. Coax cable is recommended for extension of the black and white cell leads. * DirectLine DL4000 series cells have a total lead length limit of 250 feet.
If mixed wired gauges are used, or lead length or wire gauge are not within the stated ranges, the UDA2182 can still perform the compensation. However, you must first calculate the lead resistance, and then put it in terms of the available settings for AWG gauge and length. Because the smaller gauge coax cables consist of a low resistance shield and a higher resistance conductor, an average equivalent resistance is used for calculations, i.e. 20 AWG wire is used to simulate 22AWG coax.
The resistance of each available gauge choice (in copper wire) is:
16 AWG = 4.0 ohms per 1000 feet
18 AWG = 6.4 ohms per 1000 feet
20 AWG = 10.2 ohms per 1000 feet (Use 18 AWG values for Input Configuration and calculations)
22 AWG = 16.1 ohms per 1000 feet (Use 20 AWG values for Input Configuration and calculations)
For example, suppose extension cables between the cell and Analyzer consist of 200 feet of 22-gauge coax and 500 feet of 18-gauge coax. The cell has the TC head option.
500 ft of 18 AWG coax 200 ft. of 22 AWG coax
Analyzer Honeywell Junction Conductivity Cell Box
Appendices

Because there are two different types of wire used in each lead to the cell in this example, the total lead resistance is calculated as follows: (Note: the analyzer accounts for the fact that there is always a pair of conductor wires in the system loop.)
(0.5 x 6.4) + (0.2 x 10.2) = 5.24 ohms
Since the analyzer only allows entry of one wire gauge type, we allow for the worst-case condition by dividing the total resistance by the resistance per thousand feet of the higher resistance gauge wire. In our example this would be:
5.24 ohms ÷ 10.2 ohms per thousand feet of 22 AWG wire = 514 feet
Therefore, in our example we would use the procedure in Table 6-5, and specify the wire gauge as 20 AWG and the length as 514 feet. (20 AWG wire simulates 22 AWG coax)
Appendices

Uses of cyanide solutions
Cyanide solutions are used in plating baths for zinc, cadmium, copper, brass, silver and gold. The toxic rinse waters and dumps from these operations require destruction of the cyanide (typically to a level below 0.1 ppm) before its discharge.
The technique most often used for cyanide destruction is a one or two-stage chemical treatment process. The first stage raises the pH and oxidizes the cyanide to less toxic cyanate. When required, the second stage neutralizes and further oxidizes the cyanide to harmless carbonate and nitrogen. The neutralization also allows the metals to be precipitated and separated from the effluent.
Consistent treatment and stable control in this type of process requires well-mixed reaction tanks with enough volume for adequate retention time. See Figure 15-3. Retention time is calculated by dividing the filled or usable tank volume by the waste flowrate. Typically it is 10 minutes or more.

CYANIDE TO CYANATE CYANATE TO CARBONATE AND NEUTRALIZATION
SETTLING SLUDGE
SETTLING SLUDGE
Figure 15-3 Cyanide Treatment System
Appendices

Raise pH and oxidize cyanide
Sodium hydroxide (caustic) is used to raise the effluent to about 11 pH, which will promote the oxidation reaction and ensure complete treatment. The oxidizing agent is usually sodium hypochlorite, NaOCl. The reaction for the first stage is given below using the NaOCl and with cyanide expressed in ionic form (CN- ). The result is sodium cyanate (NaCNO) and chloride ion (Cl- ).
NaOCl CN NaCNO Cl + → + − − This first-stage reaction is analyzed and controlled by independent control loops: caustic addition by pH control and oxidizing-agent addition by ORP control (redox potential or ORP, oxidation-reduction potential). Often an ON-OFF type of control using solenoid valves or metering pumps can be used. The pH controller simply calls for more caustic whenever pH falls below 11. The ORP controller calls for additional hypochlorite whenever ORP potential falls below about +450 mV. (The metal ORP electrode is positive with respect to the reference electrode.)
The ORP titration curve in Figure 15-4 shows the entire millivolt range if cyanide is treated as a batch. For continuous treatment, operation is maintained in the oxidized, positive region of the curve near the +450 mV setpoint. The ORP setpoint can vary between installations, depending upon pH, the oxidizing agent, the presence of various metals in solution, and the type of reference electrode used. Determine the exact setpoint empirically at that potential where all the cyanide has been oxidized without excess hypochlorite feed. This point can be verified with a sensitive colorimetric test kit or similar check for cyanide.
+600 pH= 10.5
pH= 11.0
0 1 2 3 4 5 6 7 -400
Figure 15-4 First Stage Cyanide Oxidation - Typical Titration Curve

As shown in Figure 15-4, pH has a direct effect on the ORP potential and must be closely controlled to achieve consistent ORP control, especially if hypochlorite is used as the oxidizing agent. Hypochlorite raises pH, which lowers the ORP potential, which in turn calls for additional hypochlorite -- a runaway situation. To avoid this situation, use close pH control and locate the ORP electrode at a distance from the hypochlorite addition point.
For this application, a gold ORP electrode gives a more reliable measurement than does a platinum electrode, because platinum may catalyze additional reactions at its surface and is more subject to coating than gold. Note that the solubility of gold in cyanide solutions does not present a problem as it is in contact, primarily, with cyanide. In fact, a slight loss of gold serves to keep the electrode clean.
Neutralize and further oxidize cyanate
The wastewater is neutralized in order to promote additional oxidation and to meet the discharge pH limits. Typically, sulfuric acid is added to lower the pH to about 8.5. At this pH the second oxidation occurs more rapidly.
Failure to comply with these instructions could result in death or serious injury.
An interlock must be provided to prevent the addition of acid before the positive oxidation of ALL cyanide. Failure to observe this precaution can result in the generation of highly toxic hydrogen cyanide.
Additional chlorine or sodium hypochlorite (NaOCl) can be added in proportion to that added in the first stage, or by separate ORP control to complete the oxidation to sodium bicarbonate (NaHCO3) in the following reaction:
2 3 2 3 2 3 2 NaCNO NaOCl H O NaHCO N NaCl + + → + + ORP control in the second stage is very similar to that in the first stage, except that the control point is near +600 mV. Control of pH in the second stage is more difficult than in the first stage, because the control point is closer to the sensitive neutral area. Proportional type pH control is often used.
Following the second stage, a settling tank and/or a filter can be used to remove suspended metal hydroxides. However, further treatment may be required to lower concentrations of some metals below their hydroxide solubilities.
Appendices

Sequence of steps
Continuous treatment is shown in Figure 15-3. However, all of the reactions can be achieved with semi-automatic batch control. Only a single tank with a pH controller and an ORP controller are required. The steps are sequenced, and the pH and ORP setpoints are changed to give the same results as for the continuous treatment. Caustic is added to raise pH to 11; then hypochlorite is added to raise the ORP potential to about +450 mV while more caustic is added as required to maintain 11 pH.
Failure to comply with these instructions could result in death or serious injury.
An interlock must be provided to prevent the addition of acid before the positive oxidation of ALL cyanide. Failure to observe this precaution can result in the generation of highly toxic hydrogen cyanide.
Then the acid can be added to neutralize the batch and further oxidation will complete the cyanate-to-carbonate conversion. A settling period can then be used to remove solids, or the batch can be pumped directly to another settling tank or pond.
Cyanide is reducing ion
An oxidation-reduction reaction involves the transfer of electrons from the ion being oxidized to the oxidizing agent. In cyanide destruction, chlorine or hypochlorite accepts electrons from the cyanide, oxidizing it, while simultaneously the hypochlorite is reduced to chloride. ORP potential is a measure of the status of the oxidation-reduction reaction; i.e., the gold electrode detects the solution’s ability to accept or donate electrons. The hypochlorite, an oxidizing ion, accepts electrons, which makes the electrode more positive. The cyanide, a reducing ion, provides electrons and makes the electrode more negative. The net electrode potential is related to the ratio of concentrations of reducing and oxidizing ions in the solution.
This electrode potential is extremely sensitive in measuring the degree of treatment in the reaction tank. However, it cannot be related to a definite concentration of a cyanide or cyanate; therefore it cannot be used as a monitor of final effluent concentration.
Reliable ORP measurement requires a very clean metal electrode surface. Routinely clean the electrodes with a soft cloth, dilute acids, and/or cleaning agents to promote fast response.
Appendices

Corrosion inhibition
Chromates are used as corrosion inhibitors in cooling towers and in metal-finishing operations including bright dip, conversion coating, and chrome plating.
The wastewater form rinse tanks, dumps, and cooling tower blowdown contains toxic soluble chromium ion, Cr+6, which must be removed, typically to a level less than 0.5 ppm before discharge.
The technique most often used for this chrome removal is a two-stage chemical treatment process. The first stage lowers the pH and adds the reducing agent to convert the chrome from soluble Cr+6 to Cr+3. The second stage neutralizes the wastewater, forming insoluble chromium hydroxide, which can then be removed.
Consistent treatment and stable control in this type of process requires well-mixed reaction tanks with enough volume for adequate retention time (see Figure 15-5). Retention time is calculated by dividing the filled or usable tank volume by the waste flowrate. Typically, it is ten minutes or more.

REDUCTION NEUTRALIZATION SETTLING SLUDGE
Figure 15-5 Chrome Treatment System
Appendices

Lower pH and add reducing agent
Sulfuric acid is used to lower the pH to about 2.5, which promotes the reduction reaction and ensures complete treatment. The reducing agent may be sulfur dioxide, sodium sulfite, sodium bisulfite, sodium metabisulfite, sodium hydrosulfite, or ferrous sulfate. The reaction is given below. The chromate is expressed as chromic acid, CrO3, with a +6 charge on the chromium and the reducing agent is expressed as sulfurous acid, H2SO3, which is generated by sulfites at low pH. The result is chromium sulfate, Cr2(SO4)3, with a +3 charge on the chromium. The reaction is expressed as:
2 3 3 3 2 3 2 4 3 2 CrO H SO Cr SO H O + → + ( ) This first stage reaction is analyzed and controlled by independent control loops: acid addition by pH control; reducing-agent addition by redox potential or ORP (oxidation- reduction potential) control. Often an ON-OFF type of control using solenoid valves or metering pumps can be used. The pH controller simply calls for additional acid whenever the pH rises above 2.5. The ORP controller calls for additional reducing agent whenever the ORP potential rises above about +250 mV. (The metal ORP electrode is positive with respect to the reference electrode.)
The ORP titration curve in Figure 15-6 shows the entire millivolt range if Cr+6 chrome is treated as a batch. With continuous treatment, operation is maintained in the fully reduced portion of the curve near the +250 mV setpoint. The ORP setpoint can vary between installations, depending on pH, reducing agent, presence of additional contaminants and dissolved oxygen, and the type of reference electrode used. Determine the exact setpoint empirically. This ORP setpoint should be at a potential where all of the Cr+6 has been reduced without excess sulfite consumption, which can release sulfur dioxide gas. This point can be verified with a sensitive colorimetric test kit or similar check.
0 1 2 3 4 5 6 7 100
Figure 15-6 Chrome Reduction - Typical Titration Curve

Chrome reduction is slow enough that 10 to 15 minutes may be required for a complete reaction and this time increases if pH is controlled at higher levels. The pH also has a direct effect on the ORP potential as shown in Figure 15-6. Therefore, pH must be controlled to achieve consistent ORP control.
Neutralize the wastewater
In this stage the wastewater is neutralized to precipitate the Cr+3 as insoluble chromium hydroxide, Cr(OH)3. Another reason is to meet the discharge pH limits. Sodium hydroxide or lime is used to raise the pH to 7.5 to 8.5 in the following reaction.
Cr SO NaOH Na SO Cr OH 2 6 3 2 4 3 2 4 3 ( ) ( ) + → + pH control point close to neutral point
Control of pH in the second stage is more difficult than in the first because the control point is in the sensitive area closer to the neutral point. Although this reaction is fast, for stability, a retention time of at least 10 minutes is usually needed for continuous treatment. Proportional pH control is often used in this stage.
Subsequently, a settling tank and/or filter will remove the suspended chromium hydroxide. Flocculating agents are helpful in this separation.
Sequence of steps
Continuous treatment for chrome removal is shown in Figure 15-5. However, all of the reactions can be achieved with semi-automatic batch control. Only a single tank with a pH controller and an ORP controller are required. The steps of the treatment are sequenced, and the pH setpoint is changed to give the same results as for the continuous treatment. Acid is added to lower pH to 2.5; then reducing agent is added to lower ORP potential to +250 mV. After waiting a few minutes to ensure a complete reaction (and possible test for Cr+6), the sodium hydroxide is added to raise pH to 8 as in the second stage of the continuous treatment. The settling period then begins, or the batch is pumped to a separate settling tank or pond.
Sulfite is reducing ion
An oxidation-reduction reaction involves the transfer of the electrons from the reducing agent to the ion being reduced. In the chrome removal application, sulfur in the sulfite ion donates electrons to reduce the chromium; simultaneously the chromium oxidizes the sulfur. The ORP potential is a measure of the status of the oxidation-reduction reaction; the platinum or gold electrode detects the solution’s ability to accept or donate electrons. Sulfite (SO3
+2), a reducing ion, donates electrons which makes the electrode more
Appendices

negative. The chromium, an oxidizing ion, Cr+6, accepts electrons and makes the electrode more positive. The net electrode potential is related to the ratio of concentrations of reducing and oxidizing ions in the solution.
This electrode potential is extremely sensitive in measuring the degree of chrome treatment in the reaction tank. However, it cannot be related to a definite concentration of chrome and, therefore, cannot be used as a final effluent monitor of chrome concentration.
Reliable ORP measurements require a very clean metal electrode surface. Clean the electrodes routinely with a soft cloth; dilute acids, and/or cleaning agents to promote fast response. Control at low pH levels in the first stage of treatment has also been found to help maintain clean ORP electrodes.
Appendices

Ion exchange operations can achieve especially precise control using the conductivity ratio of two points with each bed. Ratio measurement accounts for feedwater variations when the upstream point is measured at the cation bed inlet. With the upstream point in the bed as shown for following stages, it can identify exhaustion before breakthrough.

CELL 1
CATION BED ANION BED MIXED BED
CATION BED ANION BED MIXED BED
Reverse Osmosis
Reverse Osmosis efficiency is monitored by comparing inlet and outlet conductivity (or TDS). Automatic calculations of Percent Rejection or Percent Passage are provided. If readout is in resistivity, cell locations are interchanged. Temperature readout assists with normalized performance comparisons.
CELL 2 CELL 1
RO UNIT
RO UNIT
FEED PERMEATE
FEED PERMEATE
CONCENTRATE
CONCENTRATE
Cell1 Cell2 x100 Cell1 Cell2 x100
Cell1 Cell2 x100 Cell1 Cell2 x100
Typical Range is 0 to 20%
Typical Range is 0 to 20%
Typical Range is 80 to 100%
Typical Range is 80 to 100%
Conductivity/Resistivity/TDS difference using redundant cells on critical processes can provide a valuable diagnostic capability. If the difference in measurements exceeds the alarm points, an operator is summoned for corrective action. Monitoring may be switched to the alternate cell during maintenance. For deviation in either direction, two different

alarms (+ and -) are used. A difference kind of diagnostic can be provided by a precision check resistor in place of one cell to give continuous Analyzer/Controller checking at one value. Also see 15.11 Appendix J – Discussion on Chemical Interferences on Measured DO Currents.
UDA2182 ANALYZER
OUTPUT SIGNAL
OUTPUT SIGNAL
DIFFERENCE ALARM
DIFFERENCE ALARM
CELL 2 CELL 1 PROCESS
CELL 2 CELL 1 PROCESS
Parts Rinsing
Parts rinsing is usually controlled by conductivity to obtain adequate rinsing without wasting excess water, whether a single stage or a counter-current series of tanks. The two-cell ratio approach can determine whether inadequate rinsing is due to low flowrate or due to poor supply water quality.
CELL 1 CELL 2
PARTS FLOW
PARTS FLOW
WASTE WATER
WASTE WATER
RINSE WATER
RINSE WATER
Conductivity Ratio = Typical Ratio Range is 0.1 to 1.1 Cell1 Cell2 Cell1 Cell2
Conductivity Ratio = Typical Ratio Range is 0.1 to 1.1 Cell1 Cell2 Cell1 Cell2 Conductivity Ratio = Typical Ratio Range is 0.1 to 1.1 Cell1 Cell2 Cell1 Cell2
Appendices

Softener monitoring by conductivity ratio gives a continuous indication of performance. Sodium is typically more conductive than the hardness minerals it displaces, yielding a higher conductivity at the outlet. A ratio approaching 1 indicates that hardness ions are breaking through and that regeneration is needed.
(HARD) WATER SUPPLY CELL 2
CELL 1 (SOFT) TREATED WATER
CELL 1 (SOFT) TREATED WATER
Softening Ratio = Typical Ratio Range is 1 to 1.25 Cell1 Cell2 Cell1 Cell2
Softening Ratio = Typical Ratio Range is 1 to 1.25 Cell1 Cell2 Cell1 Cell2
The three conductivity measurements in power plants relate to water chemistry parameters as follows:
Appendices

SPECIFIC CONDUCTIVITY
CATION CONDUCTIVITY
CATION CONDUCTIVITY
UDA2182 ANALYZER
UDA2182 ANALYZER
UDA2182 ANALYZER
SAMPLE
SAMPLE
CELL 1 CELL 2 CATION EXCHANGER
CELL 1 CELL 2 CATION EXCHANGER
SPECIFIC CONDUCTIVITY CATION CONDUCTIVITY
SPECIFIC CONDUCTIVITY CATION CONDUCTIVITY
DEGASSED CONDUCTIVITY
DEGASSED CONDUCTIVITY
UDA2182 ANALYZER
UDA2182 ANALYZER
UDA2182 ANALYZER
(ANIONS)
(ANIONS)
UDA2182 ANALYZER
UDA2182 ANALYZER
UDA2182 ANALYZER
CARBON DIOXIDE BY CALCULATION
CARBON DIOXIDE BY CALCULATION
SAMPLE
SAMPLE
CELL 1 CELL 2 CATION EXCHANGER
CELL 1 CELL 2 CATION EXCHANGER
REBOILER
REBOILER
Sodium Hydroxide and Hydrochloric Acid Concentration Measurements
The measurement range of sodium hydroxide by conductivity is limited by temperature. The conductivity is limited by temperature. The conductivity of sodium hydroxide reaches a maximum value near 14% at 0º C and 29% at 100º C. Near the maximum there is poor resolution and no reliable way to know which side of the peak is being measured. Therefore, the UDA2182 measurement range is restricted by temperature to assure reliable values. Maximum concentrations are 10% at 0º C, 13% at 25º C and 20% at 75º C and above, with linear interpolation between these points. Operation above these limits gives a flashing display.
The measurement range of hydrochloric acid is restricted to less than 15.5% above 40º C and less than 18% below 40º C.
Appendices

The operation of the Analyzer/Controller can be verified by replacing the input from a cell with a precision check resistor across the Analyzer/Controller input terminals. In addition, an 8550 ohm resistor (Honeywell Part No. 31233300) can be wired in place of the inputs from the temperature compensator to simulate 25º C, the reference temperature. The unit will display a simulated “process value” appropriate for the check resistor installed. (Equations showing the relationship between resistor rating and displayed value are provided below.) If the displayed value is incorrect, the Analyzer/Controller should be serviced.
This technique can be used two ways:
When a check resistor is used instead of cell input, the Analyzer/Controller must be set for theoretically ideal conditions to achieve display of the appropriate value for the installed resistor. This means that you set the cell calibration factor to 1.00 and remove the calibration trim for the cell input being replaced by the check resistor.
To verify instrument operation at any point of measurement, calculate the check resistance needed to simulate that value. (It is assumed that you have selected a display measurement value that is within the range of your cell constant; see 2.1for ranges.) The equation used depends on the measurement type. For concentration check values see the table on the following page.
Conductivity check resistance (ohms) = Cell Constant (cm-1) x 106 Conductivity (microSiemens/cm)
Resistivity check resistance (ohms) = Cell Constant (cm-1) x Resistivity (ohm-cm)
TDS check resistance (ohms) = Cell Constant (cm-1) x 106 TDS (ppm)/TDS factor
(TDS factor has units of ppm/microSiemens-cm-1)
Appendices

Example 1: To determine the check resistor value needed to simulate conductivity measurement of 10 μS, use cell constant 0.1 and perform the following calculation:
10 k ohms = (0.1) x (1,000,000) 10
Example 2: To determine the check resistor value needed to simulate resistivity measurement of 10 M ohms, use cell constant 0.01 and perform the following calculation:
100 K ohms = (0.01) x (10,000,000)
Obtain the appropriate check resistance value from the table below.
Material/Weight % Concentration Simulation Resistance (ohms) @ 25º C
Hydrochloric Acid (HCl) 0 1 4
∞ 242.5 68.9
∞ 485.0 137.7
Sulfuric Acid (H2SO4) 0 1 4
∞ 215.5 56.0
∞ 538.7 140.0
∞ 1077.4 280.0
Sodium Chloride (NaCl) 0 1 4
∞ 574.1 195.2
∞ 1435.1 398.0
∞ 2870.3 796.1
Sodium Hydroxide (NaOH) 0 1 4
∞ 189.2 54.0
∞ 473.0 135.1
∞ 946.1 270.1
Appendices · Cell Constant · 10 25 50

Specifications for proper operation of Honeywell dissolved oxygen (DO) probes demand that the alternating current (AC) voltage signal (noise) between anode and shield connections and cathode and shield connections be less than 1 mV AC.
While it is your responsibility to assure that this specification is met, the following are some hints that have been successful in reducing these signals to the required value in a variety of installations.
Anode - Shield 1.2 to 2.0 VDC depending on oxygen level less than 1 mV AC. In low ppb measurements, this value may be zero.
Cathode - Shield < 1 mV DC less than 1 mV AC
If these steps fail to reduce the Anode-Shield and Cathode-Shield AC signals to the specified 1mV AC or less, obtain an isolated transformer and power the analyzer from that.
Appendices

Before performing air leak detection, it is necessary to determine that both Probe and Analyzer are working properly.
Assumptions:
Check for probe membrane leakage
If the probe has membrane leaks, incorrect readings may occur. Follow this procedure to check for probe membrane leakage:
8. Using a DVM that can measure Mohms, attach one DVM lead to the paper clip (or touch side of SS flow through chamber) and the other DVM lead to the cathode (black lead). Measure the impedance between the Cathode and the wire (probe side). If the probe has no leakage problem, this resistance will be greater than 1 Mohm. Go to Step 10. If the reading is in the k ohms or ohms range, there is a leak in the membrane, which can cause erratic readings in the probe. Stop any further testing until the probe is replaced.
Appendices

Check that the analyzer and probe are working together correctly.
Appendices

problem is that the cursor is positioned too far to the left or right of the flat portion of the curve, move the cursor back to the flat portion of the curve.
Appendices

The actual quantity of oxygen that can be present in solution is governed by the partial pressure of the gas in the atmosphere, the solubility in solution, the temperature and purity of the solution.
UDA2182 Universal Dual Analyzers include an internal pressure sensor and software algorithm that automatically compensates for atmospheric pressure variations during calibration. Pressure variations have a direct effect on the dissolved oxygen concentration during normal measurement so no pressure compensation is applied at that time. The information given below is only for reference to published solubility tables and is not needed for operation of the Analyzer.
The equilibrium concentration of oxygen dissolved in a liquid is directly proportional to the partial pressure of oxygen in the vapor phase with which the solution is in contact. Dry air, which contains 20.9% oxygen, will have an oxygen partial pressure of 159 mmHg if the total pressure is 760 mmHg. Tables of oxygen solubility are normally referenced to this value. An altitude or pressure correction must be made when conditions differ from this level. The correction is made using the following equation:
S = S’ (P - p)/(760 - p) where:
S is the solubility at barometric pressure of interest (P) S’ is the solubility at 760 mmHg at a given temperature P is the barometric pressure p is the partial pressure of water at the given temperature
Honeywell dissolved oxygen probes and analyzers include temperature sensors and an automatic temperature compensation algorithm. The algorithm takes the raw oxygen signal from the probe (which is proportional to the partial pressure of oxygen) and converts it into the actual concentration of oxygen at the measuring temperature. The algorithm is based on the decreasing solubility of oxygen with increasing temperature and on the probe temperature coefficient.
The significant effect of dissolved solids on reducing oxygen solubility is well documented. However, the partial pressure of oxygen (raw oxygen probe signal) is the same whether in pure or saline water. Since the actual solubility is reduced, a correction must be made when measuring brackish, sea or other water containing much more than 1 ppt (1000 ppm) of dissolved solids. The Analyzer includes a salinity correction algorithm, which uses input from a fixed value of salinity in ppt (parts-per-thousand) entered from the front panel. Suspended and settled solids have negligible effect on solubility, but may affect the transfer rate of oxygen when in excess of 2%.
Appendices

There are four contributors to measured current:
Faradaic currents are those resulting from oxidation or reduction of chemical species. The reduction of oxygen to water, the oxidation of water to oxygen, and the oxidation of hydrogen, hydrazine or sulfur dioxide, are examples of Faradaic currents.
Residual currents are unwanted Faradaic currents caused by impurities in the probe electrolyte. These impurities are metals (e.g. lead, zinc) in electrolyte reagents, which are capable of being reduced at the cathode and give rise to zero offset currents at “zero ppb oxygen”.
The platinum cathode and anode materials are actually made up of conducting platinum oxides. These oxides exist at the molecular level. The actual platinum surface state strongly affects the observed Faradaic currents. Before methods of wire conditioning were established, upwards of 96 hours was needed to allow these conditioning currents to stabilize. Once wire-conditioning methods were established, it now takes approximately 24 hours for these conditioning currents to completely stabilize. Electrode conditioning currents occur on first probe power-ups, following power interruptions of more than 1 second (back-up power is provided for the probe to prevent this current during a power outage of 1 hour or less) and following a Probe Bias test.
The Dissolved Oxygen (DO) probe consists of closely spaced bi-filar platinum windings separated by a high dielectric constant material. This is a description of a capacitor; the capacitance of a DO probe is in the hundreds if microFarads. When the probe is scanned during a Probe Bias Test(PBT) at 25mV/sec, an appreciable charging current is observed. This is equivalent to several hundred ppb dissolved oxygen.
The purpose of the PBT is to verify the optimum operating range of the current/voltage curve. It further allows one to determine if a reference shift has occurred. Most importantly, it allows one to select to identify a new bias point, if one is needed. To employ this diagnostic, you should be in air or air saturated water (ppm current is in uA range). A PBT should not be performed in a ppb application (ppb current is in nA range), due to charging and electrode currents being at a maximum value (µA range) during one of these scans. Furthermore, the final current rise during the PBT produces both hydrogen and oxygen gases within the probe. Time is needed before these gases can re- establish equilibrium with the outside sample. Therefore, the PBT should be limited to air level conditions and adequate time should be allowed for probe recovery following a PBT.
Appendices

The DO probe responds to oxygen partial pressure as follows:
O2 + 4H+ + 4e- → 2H2O (1)
Reaction (1) is a chemical reduction in which dissolved oxygen is reduced to water. This reduction occurs at the working electrode, commonly referred to as the cathode. The equal and opposite (oxidation) reaction occurs at the counter electrode (anode). Any gaseous substance, which is permeable through the membrane and is capable of being oxidized or reduced (electroactive) at the working electrode will interfere. Cl2, O3, H2, N2H4 and SO2 are examples of interfering dissolved gases.
Cl2+ 2e- → 2Cl- (2)
H2 → 2H+ + 2e- (3)
Reaction (2) is a reduction and hence a positive interference will be observed; reaction (3) is an oxidation, which will result in a negative interference. All amperometric probes are subject to reduction or oxidation interference as shown above. In addition to the direct interference shown in these two equations, the equilibrium probe provides an additional indirect interference. In normal probe operation oxygen is consumed at the working electrode and an equal amount of oxygen is produced at the counter electrode. In a positive interference condition, such as (2) above, chlorine is reduced at the working electrode and an equivalent amount of oxygen is produced at the anode. This oxygen is electroactive, along with the dissolved chlorine and is a contributor to the measured current.
In the absence of dissolved oxygen and in a negative interference situation as in (3), hydrogen gas is consumed at the working electrode and the opposite reaction, the reduction of water to hydrogen gas occurs at the counter electrode. In this hydrogen interference mode, the probe is both consuming and producing equal amounts of hydrogen, and is operating in a hydrogen detection equilibrium mode.
In cases of electrochemical interference, if the interference is positive, dissolved oxygen will be produced at the counter electrode giving a perceived higher oxygen reading. If the interference is negative, dissolved hydrogen gas will be produced at the counter electrode giving a perceived zero oxygen reading.
Often as a quick check to determine if a DO probe can reach 0.0 ppb, you can immerse the probe in a solution of sodium sulfite (Na2SO3) or sodium meta bisulfite (Na2S2O5). A 2 to 5% by weight solution in water is sufficient. If available, a small level of coboltous ion CO2+ will act as a catalyst and speed up the reaction of oxygen with the scavenger. Note: The lifetime of this solution is related to its exposure to air. Namely, the greater the exposure, the shorter the lifetime.
However, a Honeywell proven low ppb DO test using Nitrogen, an oxygen displacer, is recommended in Appendix M – Procedure for Low Level ppb Dissolved Oxygen Testing of this manual.
Appendices

In some special applications, it is desirable to read out in percent saturation rather than concentration. These are usually in non- aqueous solutions where the normal temperature compensation of the Series UDA2182 Analyzer for the solubility of air/oxygen in water does not apply. The percent saturation readout disables this solubility part of the temperature compensation. The readout is 100% when measuring in air or in a solution saturated with air, regardless of the temperature. Thus an air calibration will always produce approximately a 100% saturation readout. With this readout, salinity should be left at zero since the normal salinity correction also does not apply to non-aqueous media. When percent saturation readout is selected, the on-line displays read in percent saturation, however, all the dissolved oxygen settings in the Analyzer remain in concentration units (ppm or ppb). Therefore, percent saturation alarms, output, etc. Should be used only if the process temperature is nearly constant. For example, assume it is desired to have an alarm setpoint at 75% saturation while operating at 20°C. The corresponding setpoint is the 0.75 x 9.07 = 6.80 ppm.
(From Standard Methods for the Examination of Water and Wastewater)
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50
14.60 14.19 13.81 13.44 13.09 12.75 12.43 12.12 11.83 11.55 11.27 11.01 10.76 10.52 10.29 10.07 9.85 9.65 9.45 9.26 9.07 8.90 8.72 8.56 8.40 8.24 8.09 7.95 7.81 7.67 7.54 7.41 7.28 7.16 7.05 6.93 6.82 6.71 6.61 6.51 6.41 6.31 6.22 6.13 6.04 5.95 5.86 5.78 5.70 5.62 5.54
Appendices · Sample Temperature · Solubility (ppm, mg/L)

Before performing air leak detection, it is necessary to determine that both the probe and analyzer are working properly. Refer to Probe and Analyzers tests in Section 15.9
Appendices

The purpose of this procedure is two-fold. First, using a controlled environment, new probes and/or analyzers can be tested to determine if each is performing correctly before being installed in the field. Second, this procedure can be used to re-test the performance of an existing analyzer and/or probe.
You may choose to use this set-up for a zero calibration test. However, a zero calibration test would require, as a minimum, modifications to two of the test parameters. One modification would require a closed loop water system. The sample water must be tapped directly from the customer’s process water. The other modification would be the gas. For zero calibration, a high purity nitrogen gas (very expensive) must be piped into the process sample. Since Honeywell can neither control the quality of the gas the customer purchases nor the quality of the process water used, the company will not guarantee the accuracy of the results of a zero calibration done by this modified method.
Appendices

*Air Sat. Value at T °C x known gas O2 Value x Barometric Pressure = True Value
20.9% 760 mmHg
At 25°C using 250 ppm O2 in N2 at 770 mm Hg
True Value = 8.24 x 10-6 x 250 x 10-6 x 770 = 9.986 x 10-9 or 10 ppb
20.9 x 10-2 760
* If the temperature of the process water is not at 25°C, use O2 Solubility Tables in Table 15-2 and the process water temperature to determine the Air Saturated O2 value.
Special Requirements: Note 1:Gas Mixture is Oxygen in Nitrogen - use GravametricGases analyzed to +/- 1% of contained gas. Get Certificates of Conformance and Analysis on the purchased gas. Note 2:Dual Stage, Ultra high purity, high flow pressure regulator - non-corrosive surface. Try to get this from the gas supplier to be consistent with their recommended setup. Example supplier - Scott Model-18 Series. Note 3:Glass Gas Washing bottles with fritted disc and beaker. Gas bottle should be filled 3/4 with water. Example supplier - Fisher, Cole Palmer Note 4:Piping around glass wash bottle should be heavy wall flexible plastic. Piping that goes into beaker should be submerged about 1/2” into beakerof water. Note 5:All other piping should be rigid polypropylene tubing. Note 6: All calculations are based on 25 Deg C.
Special Requirements: Note 1:Gas Mixture is Oxygen in Nitrogen - use GravametricGases analyzed to +/- 1% of contained gas. Get Certificates of Conformance and Analysis on the purchased gas. Note 2:Dual Stage, Ultra high purity, high flow pressure regulator - non-corrosive surface. Try to get this from the gas supplier to be consistent with their recommended setup. Example supplier - Scott Model-18 Series. Note 3:Glass Gas Washing bottles with fritted disc and beaker. Gas bottle should be filled 3/4 with water. Example supplier - Fisher, Cole Palmer Note 4:Piping around glass wash bottle should be heavy wall flexible plastic. Piping that goes into beaker should be submerged about 1/2” into beakerof water. Note 5:All other piping should be rigid polypropylene tubing. Note 6: All calculations are based on 25 Deg C.
Special Requirements: Note 1:Gas Mixture is Oxygen in Nitrogen - use GravametricGases analyzed to +/- 1% of contained gas. Get Certificates of Conformance and Analysis on the purchased gas. Note 2:Dual Stage, Ultra high purity, high flow pressure regulator - non-corrosive surface. Try to get this from the gas supplier to be consistent with their recommended setup. Example supplier - Scott Model-18 Series. Note 3:Glass Gas Washing bottles with fritted disc and beaker. Gas bottle should be filled 3/4 with water. Example supplier - Fisher, Cole Palmer Note 4:Piping around glass wash bottle should be heavy wall flexible plastic. Piping that goes into beaker should be submerged about 1/2” into beakerof water. Note 5:All other piping should be rigid polypropylene tubing. Note 6: All calculations are based on 25 Deg C.
Special Requirements: Note 1:Gas Mixture is Oxygen in Nitrogen - use GravametricGases analyzed to +/- 1% of contained gas. Get Certificates of Conformance and Analysis on the purchased gas. Note 2:Dual Stage, Ultra high purity, high flow pressure regulator - non-corrosive surface. Try to get this from the gas supplier to be consistent with their recommended setup. Example supplier - Scott Model-18 Series. Note 3:Glass Gas Washing bottles with fritted disc and beaker. Gas bottle should be filled 3/4 with water. Example supplier - Fisher, Cole Palmer Note 4:Piping around glass wash bottle should be heavy wall flexible plastic. Piping that goes into beaker should be submerged about 1/2” into beakerof water. Note 5:All other piping should be rigid polypropylene tubing. Note 6: All calculations are based on 25 Deg C.
Special Requirements: Note 1:Gas Mixture is Oxygen in Nitrogen - use GravametricGases analyzed to +/- 1% of contained gas. Get Certificates of Conformance and Analysis on the purchased gas. Note 2:Dual Stage, Ultra high purity, high flow pressure regulator - non-corrosive surface. Try to get this from the gas supplier to be consistent with their recommended setup. Example supplier - Scott Model-18 Series. Note 3:Glass Gas Washing bottles with fritted disc and beaker. Gas bottle should be filled 3/4 with water. Example supplier - Fisher, Cole Palmer Note 4:Piping around glass wash bottle should be heavy wall flexible plastic. Piping that goes into beaker should be submerged about 1/2” into beakerof water. Note 5:All other piping should be rigid polypropylene tubing. Note 6: All calculations are based on 25 Deg C.
Honeywell Model 2182 Universal Dual Analyzer
Honeywell Model 2182 Universal Dual Analyzer
Honeywell Model 2182 Universal Dual Analyzer
Honeywell Model 2182 Universal Dual Analyzer
Oxygen Probe Model DL5PPB
Oxygen Probe Model DL5PPB
Chamber #31063336 or #31063337
Chamber #31063336 or #31063337
1/ 2” Note 4
1/ 2” Note 4
1/ 2” Note 4
1/ 2” Note 4
1/ 2” Note 4

Many applications tap a sample from a main process stream and, after the flow has passed through the measurement manifold, it is discharged to a sink or floor drain.
A typical probe installation will find the probe mounted in a flow chamber or tee arrangement similar to what is shown in Figure 15-8. Key installation features are provision for flow rate adjustment, a water trap to assure that the probe remains immersed if sample flow is turned off and means to prevent a below atmospheric pressure within the manifold.

OPTIONAL 3-WAY VALVE AND DRAIN FOR DISSOLVED OXYGEN INSTALLATIONS
DRAIN
Figure 15-8 Typical Probe Installation
It is desirable for water to exit the manifold 3 to 6 inches above the sensor tip. This will insure that the sensor remains immersed if sample flow is turned off.
The air vent extension is sized so normal sample flow does not completely fill this tube. Its purpose is to prevent negative pressure within the manifold. Without this air vent, if for example the exit stream is discharged to a floor drain four feet below the manifold,
Appendices

then the pressure at the sensor is four feet water-column below atmospheric pressure. Any fitting leaks at or beyond the flow adjustment valve will result in air infiltration into the sample. This entrapped air can result in noisy and unstable measurement. In the case of a part per billion dissolved oxygen (DO) measurement, the indicated DO value can be substantially higher than the true value.
When it becomes necessary to discharge the sample stream in a loop higher than the manifold, then the air vent should be located above the highest point in the loop.
Appendices

Overview
Although the Honeywell probe accuracy is unaffected by inert fouling, there are two conditions where probe cleaning may be required. (These conditions affect all conventional dissolved oxygen probes as well.)
The first is where the fouling is so thick that the response time of the probe becomes unacceptably long. The second is where organic fouling is consuming oxygen before it reaches the surface of the probe.
A feature allowing automatic cleaning at preconfigured times is included in the UDA2182 analyzer. Cleaning may be initiated with a frequency of every few minutes to monthly.
Functionally, relays within the analyzer are tripped, allowing withdrawal of the probe from the sample, turning on a cleaning spray, turning off the spray, and reinserting the probe into the sample. Execution of automatic cleaning and calibration requires you to install a drive unit, a solenoid valve, and mounting hardware. See Figure 15-9
Similarly, all probes drift with time. Although the Honeywell probe is very stable, included in the analyzer is a feature that allows withdrawal of the probe into air for automatic air calibration at user-configured times.
The sequence of calibrations and cleanings are user-configurable.
One symptom of the need for cleaning is a low dissolved oxygen reading. The UDA2182 dissolved oxygen analyzer can be configured to execute a cleaning cycle if the measured dissolved oxygen falls below a user-selected value. An alarm can be configured to alert you if the cleaning fails to restore the dissolved oxygen to a higher level. The alarm will indicate either a true decrease in dissolved oxygen concentration or unsuccessful cleaning.
Appendices


Appendices

A typical set – up for automatic calibration in a boiler water sampling system is shown in Figure 15-10. The solenoid valve and connections should be supplied by others and must be positively air tight to prevent leakage and erroneous measurements.
The solenoid valve is wired to assigned relay contacts in the UDA2182 analyzer and will operate at a frequency, and for duration as assigned by the end user.

Appendices

Periodic calibration of pH electrodes is necessary for best system performance because electrode outputs change over time. One-point calibration (standardization) is a zero adjustment to compensate for electrode drift. Two-point calibration (standardization and slope adjustment) includes a span adjustment to match the gain of the Analyzer/Controller to the electrode response. Standardization and slope each depend on the electrode measuring a reference solution (buffer) of known pH.
The AutoClean and AutoCal features periodically rinse and calibrate pH electrodes automatically as described in this section. To take advantage of these features you must:
Rinse sequence
The AutoClean operation occurs at the configured intervals. The sequence is described below.
1 All alarm action is held at existing levels. The output(s) can be held or be active, depending on configuration. Even if the outputs are not held, “HOLD ACTIVE” is displayed on the alarm stripe because alarms are always held.
Also, “AUTOSEQUENCE” is displayed. Pressing the DISPLAY key will call up a special display that shows how much time is left in the operation.
2 Relay 1 activates 3-way solenoid valve S1 (see Figure 15-11) to direct rinsing fluid to the electrodes for the configured rinse duration (1 to 1999 seconds). If the measured sample is normally returned to the process but quantities of rinsing fluid cannot be tolerated there, use an additional 3-way solenoid valve S4. It is activated simultaneously with S1 to divert the discharge to drain.
3 At the end of the configured rinse time Relay 1 de-activates the solenoid valve S1 (and S4, if used). After the configured delay period (1 to 1999 seconds) the Analyzer/Controller resumes sampling the process. (Note that even with S4 for diversion, one system volume of washing fluid will pass to the process at this point.)
4 The “HOLD” and “AUTOSEQUENCE” messages are cleared.
Index

S1
Items outside this area provided by user
Select valves and fittings with appropriate pressure ratings
Make the process connections as shown in Figure 15-11. Be sure that valves and fittings (S1) have sufficient pressure ratings to withstand pressure peaks which will occur when process flow is blocked.
Keep pipe sizes small and couplings close to minimize the liquid volume in the system. Smaller volumes require less time to rinse.
Introduction
AutoCal can include one-point calibration (standardization) to adjust zero to compensate for electrode drift, or two-point calibration (standardization and slope adjustment) to also adjust span to match the gain of the Analyzer/Controller to the electrode response. Standardization and slope each depend on the electrode measuring a reference solution (buffer) of known pH.
The AutoCal operation automatically occurs at the configured intervals. It always includes AutoClean rinsing of the pH electrode, in addition to any other AutoClean sequences that are configured to occur between standardization operations. The sequence is described below.
1 All alarm action is held at existing levels. The output(s) can be held or be active, depending on configuration. Even if the outputs are not held, “HOLD ACTIVE” is displayed on the alarm stripe because alarms are always held.
Also, “AUTOSEQUENCE” is displayed. Pressing the DISPLAY key will call up a special display that shows how much time is left in the operation
2 In preparation for the calibration, Relay 1 activates 3-way solenoid valve S1 (see Figure 15-12) to direct rinsing fluid to the electrodes for the configured rinse duration (1 to 1999 seconds). If the measured sample is normally returned to the process but quantities of rinsing fluid cannot be tolerated there, use an additional 3-way solenoid valve S4. It is activated simultaneously with S1 to

divert the discharge to drain.
3 Relay 2 activates solenoid valve S2 for the preset buffer time (1 to 1999 seconds) to direct buffer solution past the electrodes by gravity.
4 After a stable reading is reached or the set maximum buffer time elapses, the 9782 stores the new calibration value using automatic buffer recognition. Diagnostics detect excessive instability or offset, prevent erroneous calibrating and can activate an alarm, depending on configuration. If the diagnostic fails, an error message is always displayed on the alarm stripe (see Section 12). If an unacceptable value is obtained, it will be rejected and the previous value will be retained for uninterrupted operation.
5 All valves are deactivated to resume measurement of the sample.
6 A delay period (1 to 1999 seconds) can be configured to permit the measurement to stabilize on the process sample. At the end of the delay period normal alarm, control and output operation resumes. The “HOLD” and “AUTOSEQUENCE” messages are cleared.
Items outside this area provided by user
S2
S 1
Figure 15-12 Rinse and One-Point Calibration
Rinse and two-point calibration sequence
With this function, rinse and one-point standardization operations are performed as described previously according to the configured schedule. If two-point calibration is to be performed periodically, then after the configured number of standardization operations, Steps 4a and 4b shown below are also performed (before Step 5 above) to make the slope adjustment.
4a Relay 3 activates solenoid valve S3 for the configured buffer time to direct the second buffer flow to the electrodes.
4b After stability is reached or the set maximum buffer time elapses, the instrument calculates and stores a new slope value using automatic buffer recognition. Diagnostics detect excessive instability or offset, prevent erroneous calibrating and can activate an alarm, depending on configuration. If the diagnostic fails, an error message is always displayed on the alarm stripe (see Section 9).
If an unacceptable value is obtained, it will be rejected and the previous value will be retained for uninterrupted operation.

Make the process connections as shown in Figure 15-12 or Figure 15-13. Materials and components should be carefully selected for chemical resistance to process and buffer solutions at anticipated temperatures. Be sure that valves and fittings have sufficient pressure ratings to withstand pressure peaks which will occur when process flow is blocked.
Keep pipe sizes small and couplings close to minimize the liquid volume in the system. Smaller volumes require less buffer solution and less time to rinse and to calibrate.

Slope Duration, Relay 3 Activated (Same Period as Standardize)
Two-Point AutoCal Operation* * One-point AutoCal and AutoClean operations omit steps using Relay 3 and Relays 2 & 3, respectively.

Index











Index



For application assistance, current specifications, pricing, or name of the nearest Authorized Distributor, contact one of the offices below.
ASIA PACIFIC Control Products Asia Pacific Headquarters Phone: +(65) 6355-2828 Fax: +(65) 6445-3033 Asia Pacific Global Technical Support Field Instruments Phone: +65 6580 3156 Fax: +65 6445-3033 Process Instruments Phone: (603) 76950 4777 Fax: (603) 7958 8922 Australia Honeywell Limited Phone: +(61) 7-3846 1255 FAX: +(61) 7-3840 6481 Toll Free 1300-36-39-36 Toll Free Fax: 1300-36-04-70 China – PRC - Beijing Honeywell China Inc. Phone: +(86-10) 8458- 3280 Fax: +(86-10) 8458-4650 China – PRC - Shanghai Honeywell China Inc. Phone: (86-21) 5257-4568 Fax: (86-21) 6237-2826 China – PRC - Chengdu Honeywell China Inc. Phone: +(86-28) 8678- 6348 Fax: +(86-28) 8678-7061 China – PRC - Xi’an Honeywell China Ltd - Xi’an. Phone: +(86-29) 8833- 7490 Fax: +(86-29) 8833-7489 China – PRC - Shenzhen- Honeywell China Inc. Phone: +(86) 755-2518- 1226 Fax: +(86) 755-2518-1221
Malaysia Honeywell Engineering Sdn Bhd Phone: +(60-3) 7950- 4776 Fax: +(60-3) 7958-8922 New Zealand Honeywell Limited Phone: +(64-9) 623-5052 Fax: +(64-9) 623-5060 Toll Free (0800) 202-088 Philippines Honeywell Systems (Philippines) Inc. Phone: +(63-2) 633-2830- 31/ 636 1661-62 Fax: +(63-2) 638-4013 Singapore Honeywell Pte Ltd. Phone: +(65) 6580 3278 Fax: +(65) 6445-3033 South Korea Honeywell Korea Co Ltd Phone: +(822) 799 6315 Fax: +(822) 792 9015 Thailand Honeywell Systems (Thailand) Ltd. Phone: +(662) 693-3099 FAX: +(662) 693-3089 Taiwan R.O.C. Honeywell Taiwan Ltd. Phone: +(886-2) 2245- 1000 FAX: +(886-2) 2245-3241 SE Asia Countries see Honeywell Pte Ltd (Singapore) for: Pakistan, Cambodia, Guam, Laos, Myanmar, Vietnam, East Timor SE Asia Countries see Honeywell Automation India Ltd for: Bangladesh Nepal Sri Lanka EUROPE Austria Honeywell Austria GmbH Phone: +43 (316)400123 FAX: +43 (316)40017 Belgium Honeywell SA/NV Phone: +32 (0) 2 728 24 07 FAX: +32 (0) 2 728 22 45
Bulgaria Honeywell EOOD Phone: +(359) 2 40 20 900 FAX: +(359) 2 40 20 990 Czech Republic Honeywell spol. s.r.o. Phone: +420 242 442 232 FAX: +420 242 442 131 Denmark Honeywell A/S Phone: +(45) 39 55 55 55 FAX: +(45) 39 55 55 58 Finland Honeywell OY Phone: +358 (0)20752 2753 FAX: +358 (0) 20752 2751 France Honeywell SA Phone: +33 (0)1 60198075 FAX: +33 (0)1 60198201 Germany Honeywell AG Phone: +49 (69)8064-299 FAX: +49 (69)806497336 Hungary Honeywell Kft. Phone: +36-1-451 4300 FAX: +36-1-451 4343 Italy Honeywell S.p.A. Phone:+390292146307 FAX: +39 0292146377 The Netherlands Honeywell B.V. Phone: +31 (0) 20 5656200 FAX: +31 (0) 20 5656210 Norway Honeywell A/S Phone: (45) 39 55 55 55 Poland Honeywell Sp. zo.o Phone: +48-22-6060900 FAX: +48-22-6060901 Portugal Honeywell Portugal Lda Phone: +351 21 424 5000 FAX: +351 21 424 50 99 Romania Honeywell Bucharest Phone: +40 (0) 21 2316437 FAX: +40 (0) 21 2316439 Russian Federation (RF), ZAO "Honeywell" Phone: +7 (095) 796 98 00 FAX: +7 (495) 797 99 64
Slovak Republic Honeywell s.r.o. Phone: +421-2-58247 410 FAX: +421-2-58247 415 Spain Honeywell S.A. Phone: +34 (0)91313 61 00 FAX: +34 (0)91313 61 30 Sweden Honeywell AB Phone: +(46) 8 775 55 00 FAX: +(46) 8 775 56 00 Switzerland Honeywell AG Phone: +41 18552448 FAX: +(41) 1 855 24 45 Turkey Honeywell Turkey A.S. Phone: +90 216 578 71 00 FAX: +90 216 575 66 35 Ukraine Honeywell Tel: +380-44-201 44 74 Fax: +380-44-201-44-75 United Kingdom Honeywell Control Systems Ltd. Phone: +44 (0)1344 655251 FAX: +44 (0) 1344 655554
AFRICA Mediterranean & African Distributors Honeywell SpA Phone: +39 (02) 250 10 604 FAX: +39 (02) 250 10 659 South Africa (Republic of) and sub saharan Honeywell Southern Africa Honeywell S.A. Pty. Ltd. Phone: +27 11 6958000 FAX +27 118051504 NORTH AMERICA Canada Honeywell LTD Phone: 1-800-737-3360 FAX: 1-800-565-4130 USA Honeywell Process Solutions, Phone: 1-800-343-0228 FAX: 1-717-771-8251 Email:sc-cp- appssales@ honeywell.com LATIN AMERICA Argentina Honeywell S.A.I.C. Phone: +(54-11) 4383- 3637 FAX: +(54-11) 4325-6470 Brazil Honeywell do Brasil & Cia Phone: +(55-11) 7266- 1900 FAX: +(55-11) 7266-1905 Chile Honeywell Chile, S.A. Phone: +(56-2) 233-0688 FAX: +(56-2) 231-6679 Mexico Honeywell S.A. de C.V. Phone: +(52) 55 5259- 1966 FAX: +(52) 55 5570-2985 Puerto Rico Honeywell Inc. Phone: +(809) 792-7075 FAX: +(809) 792-0053 Trinidad Honeywell Inc. Phone: +(868) 624-3964 FAX: +(868) 624-3969 Venezuela Honeywell CA Phone: +(58-2) 238-0211 FAX: +(58-2) 238-3391
MIDDLE EAST Abu Dhabi U A E Middle East Headquarters Honeywell Middle East Ltd. Phone: +971 2 4041246 FAX: +971 2 4432536 Sultanate of Oman Honeywell & Co Oman LLC Phone: +968 24 701153/ Ext.33 FAX +968 24 787351 Saudia Arabia Honeywell Turki Arabia Ltd Jubail Office Phone: +966-3-341-0140 Fax: +966-3-341-0216 Honeywell - ATCO Dammam Office Phone: 0096638304584 Fax: 0096638338059 Kuwait Honeywell Kuwait KSC Phone: +965 242 1327 to 30 Fax: +965 242 8315 And Phone: +965 326 2934/1821Fax: +965 326 1714
Indonesia PT Honeywell Indonesia Phone: +(62) 21-535-8833 FAX: +(62) 21-5367 1008 India Automation India Ltd. Honeywell Ltd. Phone:+(91) 5603-9400 Fax: +(91) 5603-9600 Japan Honeywell Inc. Phone: +(81) 3 6730 7150 Fax: +(81) 3 6730 7228
Honeywell Process Solutions Honeywell 70-82-25-119 Rev.5 2500 W. Union Hill Drive January 2009 Phoenix, Arizona 85027 ©2009 Honeywell International Inc. www.honeywell.com/ps