Model 225
Sanitary and CIP Toroidal Sensor
For additional information, please visit our website
at www.emersonprocess.com/raihome/liquid/.
WARNING
CAUTION
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsi- bility of the user.
1 Sensor meets 3A sanitary requirements for sensors and sensor fittings and connections used on milk and milk products equipment (74-03).
INSTALLATION The sensor may be installed in either a tank or pipe using a customer-supplied Tri- clamp and tee assembly. The sensor requires a 2-inch Tri-clamp, a 2-inch type 1 gasket, and a 2-inch tank ferrule or tee. Keep at least 1 in. (25 cm) between sensor and pipe wall. If clearance is too small, calibrate the sensor in place. Ensure that the sensor is completely submerged in liq- uid to the flange. Mounting the sensor in a vertical pipe run with the flow from bottom to top is best. If the sensor must be mounted in a hori- zontal pipe run, orient the sensor in the 3 o’clock or 9 o’clock position.


WIRING
Keep sensor wiring away from ac conductors and high current demanding equipment. Do not cut cable. Cutting the cable may void the warranty.



FIGURE 3. Wiring 225-54 and 225-56 sensors to Model 1056 and 56 analyzers



FIGURE 5. Wiring 225-54 and 225-56 sensors to Model Xmt-T panel mount transmitter


FIGURE 6. Wiring 225-54 and 225-56 sensors to Model Xmt-T pipe/wall mount transmitter

FIGURE 7. Wiring 225-54 and 225-56 sensors to Model 1066 transmitter.



AND PN 23294-04
FIGURE 9. Wiring sensors through a remote junction box Wire sensors point to point. For wiring at the analyzer end, refer to the appropriate analyzer wiring diagram. For interconnecting cable 23294-00, use the 225-54 wiring diagram. For interconnecting cable 23294-04 and 23294-05, use the 225-56 wiring diagram.


CALIBRATION
the process piping. Model 225 sensors are often installed in sanitary flange tees, so an appropriate cal- ibration vessel is another tee with blank flanges on the ends. If removing the sensor from the process piping for calibration is impractical, calibrate the sensor in place against a referee sensor and analyzer while both sen- sors are simultaneously measuring the process liquid. If this is not practical the sensor can also be calibrated against the results of a measurement made on a grab sample.
The nominal cell constant of the Model 225 sensor is 2.7/cm. The error in the cell constant is about ±10%, so conductivity readings made using the nominal cell con- stant will have an error of at least ±10%. Wall effects, discussed below, will likely make the error greater. For higher accuracy, the sensor must be calibrated.
Wall effects arise from the interaction between the cur- rent induced in the sample by the sensor and nearby pipe or vessel walls. As Figure 11 shows, the measured conductivity can either increase or decrease depend- ing on the wall material. Because wall effects do not disappear until the Model 225 sensor is at least 1 inch (25 mm) away from the nearest wall, wall effects are present in most installations. Correcting for them is an important part of calibration.
For more information about calibrating toroidal conduc- tivity sensors, refer to application sheet ADS 43-025 available on the Rosemount Analytical website.
Conductivity sensors are calibrated against a solution of known conductivity (a standard) or against a previ- ously calibrated referee sensor and analyzer. If wall effects are present, calibrating a toroidal sensor with a standard solution is possible only if the vessel hold- ing the standard has exactly the same dimensions as
Be sure sensor has been cleaned of process liquid before handling.
Generally, the only maintenance required is to keep the opening of the sensor clear of deposits. Cleaning fre- quency is best determined by experience.

FIGURE 11. Measured conductivity as a function of clearance between sensor and walls

TROUBLESHOOTING



FIGURE 12. Resistance check. Disconnect leads from analyzer before measuring resistances.

The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2010
Emerson Process Management
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Off-scale reading | Wiring is wrong. | Verify and correct wiring. | p.7 |
| Off-scale reading | RTD is open or shorted. | Check the RTD for open or short circuits. See Figure 12. | p.7 |
| Off-scale reading | Sensor is not in process stream. | Confirm that the sensor is fully submerged in the process stream. See installation section. | p.7 |
| Off-scale reading | Sensor is damaged. | Perform isolation checks. See Figure 12. | p.7 |
| Noisy reading | Sensor is improperly installed in the process stream. | Confirm that the sensor is fully submerged in the process stream. See installation section. | p.7 |
| Noisy reading | Sensor cable is run near high voltage conductors. | Move cable away from high voltage conductors. | p.7 |
| Noisy reading | Sensor cable is moving. | Keep sensor cable stationary. | p.7 |
| Reading seems wrong (lower or higher than expected) | Bubbles are trapped in the sensor, particularly in the toroid opening. | Install the sensor in a vertical pipe run with the flow against the toroid opening. Increase flow if possible. | p.7 |
| Reading seems wrong (lower or higher than expected) | Sensor is not completely submerged in the process stream. | Confirm that the sensor is fully submerged in the process stream. See installation section. | p.7 |
| Reading seems wrong (lower or higher than expected) | Cell constant is wrong. Wall effects are present. | Calibrate the sensor in place in the process piping. See calibration section. | p.7 |
| Reading seems wrong (lower or higher than expected) | Wrong temperature correction algorithm is being used. | Check that the temperature correction is appropriate for the sample. See analyzer manual for more information. | p.7 |
| Reading seems wrong (lower or higher than expected) | Temperature reading is inaccurate. | Disconnect the RTD leads (Figure 12) and measure the resistance between the in and common leads. Resistance should be close to the value in Table 1. | p.7 |
| Reading seems wrong (lower or higher than expected) | Slow temperature response to sudden changes in temperature. | Use an RTD in a metal thermowell for temperature compensation. | p.7 |
| Sluggish response | Sensor is installed in a dead area in the piping. | Move sensor to a location more representative of the process liquid. | p.7 |
| Sluggish response | Slow temperature response to sudden changes in temperature. | Use an RTD in a metal thermowell for temperature compensation. | p.7 |