Page 1 of the Emerson 403VP user manual — Figure 1. Sensor Orientation
Page 1 · click to enlarge
Page 1

PURSense ™ Conductivity Sensors

For additional information, please visit our website at www.rosemountanalytical.com

CAUTION

SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user.

WARNING

Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!

Specifications - Sensor

Emerson 403VP — Specifications - Sensor diagram (page 1)
Fig. 1 · Specifications - Sensor

Temperature Range 32–221 °F (0–105 °C) Sensors tolerate steam sterilization to 135 °C

Elastomers and fluorocarbon resins are compatible with 21CFR177. Elastomers also meet the requirements of USP Class VI. Stainless steel contains

Depending on the option selected, the sensor can be installed in either a 1 ½-inch or 2-inch Tri-Clamp tee. The gasket,

Figure 1. Sensor Orientation

Figure 2. Recommended Installation

Emerson 403VP — Figure 1. Sensor Orientation diagram (page 1)
Fig. 2 · Figure 1. Sensor Orientation
Emerson 403VP — Figure 1. Sensor Orientation diagram (page 1)
Fig. 3 · Figure 1. Sensor Orientation

clamp, and tee must be supplied by the user. The electrodes must be completely submerged in the process liquid, i.e., up to the inside surface of the flange.

If the sensor is installed in a side stream with the sample draining to open atmosphere, bubbles may accumulate on the electrodes. Trapped bubbles will cause errors. Normally, as bubbles accumulate the conductivity reading drifts down. To control bubble formation, apply a small amount of back pressure to the sensor.

Page 2 of the Emerson 403VP user manual — RTD SENSE
Page 2 · click to enlarge
Page 2

Wiring

NOTE

For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product.

1056, 1057, 56, 5081, 6081, 54e, and XMT : http://www3.emersonprocess.com/raihome/sp/liquid/wiring/XMT/ 1066 and sensors with SMART preamps: http://www2.emersonprocess.com/en-US/brands/rosemountanalytical/ Liquid/Sensors/Pages/Wiring_Diagram.aspx 1055: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/1055/

Wire Color and Connections in Sensor

COLORFUNCTION GrayConnects to outer electrode ClearCoaxial shield for gray wire OrangeConnects to inner electrode ClearCoaxial shield for orange wire RedRTD in RTD RTD sense RTD return White with red stripe White ClearShield for all RTD lead wires
Emerson 403VP — Wire Color and Connections in Sensor diagram (page 2)
Fig. 4 · Wire Color and Connections in Sensor

Wiring Diagrams

Figure 4. Wiring for 56 and 1056 Figure 3. Wiring for 54eC

TB1

Emerson 403VP — Figure 4. Wiring for 56 and 1056 Figure 3. Wiring for 54eC diagram (page 2)
Fig. 5 · Figure 4. Wiring for 56 and 1056 Figure 3. Wiring fo…

RTD SENSE

RTD IN

RTD SHLD

4CT B

4CT A

SHLD 2CT

SEN 2CT B

SHLD 2CT

SEN 2CT A

Show 8 more labels / short text on this page
White RTD RTN · White/Red · Red · Clear · Clear · Orange · Clear · Gray
Page 3 of the Emerson 403VP user manual — 4CT-A
Page 3 · click to enlarge
Page 3

Figure 5. Wiring for 1066

Emerson 403VP — Figure 5. Wiring for 1066 diagram (page 3)
Fig. 6 · Figure 5. Wiring for 1066

Figure 6. Wiring (Panel) for Xmt-C-10 Figure 7. Wiring(Pipe or Wall) for Xmt-C-11

TB1 TB1

DRIVE

RTD SHIELD

DRIVE COM

RTD COM

RECEIVE

RTD SENSE

RECEIVE COM

RTD IN

RTD IN

RECEIVE COM

RTD SENSE

RECEIVE

DRIVE COM

RTD COM

RTD SHIELD

DRIVE

Figure 8. Wiring for 5081-C Figure 9. Wiring for 6081-C

MODEL 5081C TRANSMITTER

DRV SHLD DRV

SHIELD

2CT-B 2CT-A

RCV RTD COM

RTD SHIELD

SHIELD

COM

Emerson 403VP — TB1 TB1 diagram (page 3)
Fig. 7

5 6 7 8 9 10 11 12 13

4CT-A

RTD IN

RTD IN

RTD SENSE

SENSE

4CT-B

RTD COM

RTD RTN

RTD SHLD

HT/FF (-)

SMART PWR

RESERVED

HT/FF (+)

CLEAR

WHITE

WHITE/RED

RED

CLEAR

ORANGE

CLEAR

GRAY

WHITE

RED/WHITE

RED

CLEAR CLEAR

GRAY ORANGE

CLEAR

SENSOR CABLE

Show 16 more labels / short text on this page
Gray 8 · Clear · 1 White · Clear · Orange · Red/White · Clear · Red · Red · Clear · Red/White · Orange · Clear · White · Clear · Gray
Page 4 of the Emerson 403VP user manual — Calibration
Page 4 · click to enlarge
Page 4

Wiring through a Junction Box If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point. Use cable 23747-00 (factory-terminated) or 9200275 (no terminations).

Pin Out Diagram for 400VP

Figure 10. VP pin-out (viewed from connector end of sensor, looking down)

Emerson 403VP — Figure 10. VP pin-out (viewed from connector end of sensor, looking down) diagram (page 4)
Fig. 8 · Figure 10. VP pin-out (viewed from connector end of…

Cleaning the Sensor

Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol (rubbing alcohol) can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.

Calibration

PURSense conductivity sensors are calibrated at the factory and do not need calibration when first placed in service. Simply enter the cell constant printed on the label into the analyzer.

After a period of service, the sensor may require calibration. The sensor can be calibrated against a solution having known conductivity or against a referee meter and sensor. If using a standard solution, choose one having conductivity in the recommended operating for the sensor cell constant. Refer to the analyzer manual or product data sheet for recommended ranges. Do not use standard solutions having conductivity less than about 100 uS/cm. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25°C). Because 0.01/cm sensors must be calibrated in low conductivity solutions, they are best calibrated against a referee meter and sensor in a closed system.

For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website.

Page 5 of the Emerson 403VP user manual — Troubleshooting
Page 5 · click to enlarge
Page 5

Troubleshooting

PROBLEMPROBABLE CAUSESOLUTION Off-scale readingWiring is wrong.Verify wiring. RTD is open or shorted.Check RTD for open or short circuits. See Figure 11. Sensor is not in process stream.Be sure sensor is completely submerged in process stream. Variopol cable is not properly seated.Loosen connector and reseat. Sensor has failed.Perform isolation checks. See Figure 12. Noisy readingSensor is improperly installed in process stream.Be sure sensor is completely submerged in process stream. Variopol cable is not properly seated.Loosen connector and reseat. Reading seems wrong (lower or higher than expected)Bubbles trapped in sensor.Be sure sensor is properly oriented in pipe or fol w cell. See Figure 1. Apply back pressure to fol w cell. Wrong temperature correction algorithm.Check that temperature correction is appropriate for the sample. See analyzer manual for more information. Wrong cell constant.Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. Sluggish responseElectrodes are fouled.Clean electrodes. Sensor is installed in dead area in piping.Move sensor to a location more representative of the process liquid.

Figure 11. Checking RTD Disconnect leads and measure resistances shown. The measured resistance at room temperature should be close to the value in the table.

Figure 12. Checking Continuity and Leakage Disconnect electrode leads and measure resistance and continuity as shown. Sensor must be dry when checking resistance between electrode leads.

Emerson 403VP — Troubleshooting diagram (page 5)
Fig. 9 · Troubleshooting
Emerson 403VP — Troubleshooting diagram (page 5)
Fig. 10 · Troubleshooting
Emerson 403VP — Troubleshooting diagram (page 5)
Fig. 11 · Troubleshooting

red · red/white RTD · white

Page 6 of the Emerson 403VP user manual — Notes:
Page 6 · click to enlarge
Page 6

Notes:

Page 7 of the Emerson 403VP user manual — Notes:
Page 7 · click to enlarge
Page 7

Notes:

Page 8 of the Emerson 403VP user manual — Notes:
Page 8 · click to enlarge
Page 8

Rev. P

facebook.com/EmersonRosemountAnalytical

AnalyticExpert.com

Credit Cards for U.S. Purchases Only.

twitter.com/RAIhome

youtube.com/user/RosemountAnalytical

©2015 Rosemount Analytical, Inc. All rights reserved.

The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners.

Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 www.RosemountAnalytical.com

The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Off-scale readingWiring is wrong.Verify wiring. p.5
Off-scale readingRTD is open or shorted.Check RTD for open or short circuits. See Figure 11. p.5
Off-scale readingSensor is not in process stream.Be sure sensor is completely submerged in process stream. p.5
Off-scale readingVariopol cable is not properly seated.Loosen connector and reseat. p.5
Off-scale readingSensor has failed.Perform isolation checks. See Figure 12. p.5
Noisy readingSensor is improperly installed in process stream.Be sure sensor is completely submerged in process stream. p.5
Noisy readingVariopol cable is not properly seated.Loosen connector and reseat. p.5
Reading seems wrong (lower or higher than expected)Bubbles trapped in sensor.Be sure sensor is properly oriented in pipe or fol w cell. See Figure 1. Apply back pressure to fol w cell. p.5
Reading seems wrong (lower or higher than expected)Wrong temperature correction algorithm.Check that temperature correction is appropriate for the sample. See analyzer manual for more information. p.5
Reading seems wrong (lower or higher than expected)Wrong cell constant.Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. p.5
Sluggish responseElectrodes are fouled.Clean electrodes. p.5
Sluggish responseSensor is installed in dead area in piping.Move sensor to a location more representative of the process liquid. p.5

Ask about the Emerson 403VP

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson 403VP

Be the first to ask — other owners or our team will answer.

Original document

Emerson 403VP manual PDF

8 pages·771 KB·Stud Sensor·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…