For additional information, please visit our website at www.rosemountanalytical.com
SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user.
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!

Temperature Range 32–221 °F (0–105 °C) Sensors tolerate steam sterilization to 135 °C
Elastomers and fluorocarbon resins are compatible with 21CFR177. Elastomers also meet the requirements of USP Class VI. Stainless steel contains
Depending on the option selected, the sensor can be installed in either a 1 ½-inch or 2-inch Tri-Clamp tee. The gasket,
Figure 2. Recommended Installation


clamp, and tee must be supplied by the user. The electrodes must be completely submerged in the process liquid, i.e., up to the inside surface of the flange.
If the sensor is installed in a side stream with the sample draining to open atmosphere, bubbles may accumulate on the electrodes. Trapped bubbles will cause errors. Normally, as bubbles accumulate the conductivity reading drifts down. To control bubble formation, apply a small amount of back pressure to the sensor.

For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product.
1056, 1057, 56, 5081, 6081, 54e, and XMT : http://www3.emersonprocess.com/raihome/sp/liquid/wiring/XMT/ 1066 and sensors with SMART preamps: http://www2.emersonprocess.com/en-US/brands/rosemountanalytical/ Liquid/Sensors/Pages/Wiring_Diagram.aspx 1055: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/1055/

TB1

RTD IN
RTD SHLD
4CT B
4CT A
SHLD 2CT
SEN 2CT B
SHLD 2CT
SEN 2CT A


Figure 6. Wiring (Panel) for Xmt-C-10 Figure 7. Wiring(Pipe or Wall) for Xmt-C-11
DRIVE
RTD SHIELD
DRIVE COM
RTD COM
RECEIVE
RTD SENSE
RECEIVE COM
RTD IN
RTD IN
RECEIVE COM
RTD SENSE
RECEIVE
DRIVE COM
RTD COM
RTD SHIELD
DRIVE
Figure 8. Wiring for 5081-C Figure 9. Wiring for 6081-C
MODEL 5081C TRANSMITTER
DRV SHLD DRV
SHIELD
2CT-B 2CT-A
RCV RTD COM
RTD SHIELD
SHIELD
COM

5 6 7 8 9 10 11 12 13
RTD IN
RTD IN
RTD SENSE
SENSE
4CT-B
RTD COM
RTD RTN
RTD SHLD
HT/FF (-)
SMART PWR
RESERVED
HT/FF (+)
CLEAR
WHITE
WHITE/RED
RED
CLEAR
ORANGE
CLEAR
GRAY
WHITE
RED/WHITE
RED
CLEAR CLEAR
GRAY ORANGE
CLEAR
SENSOR CABLE

Wiring through a Junction Box If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point. Use cable 23747-00 (factory-terminated) or 9200275 (no terminations).

Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol (rubbing alcohol) can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.
PURSense conductivity sensors are calibrated at the factory and do not need calibration when first placed in service. Simply enter the cell constant printed on the label into the analyzer.
After a period of service, the sensor may require calibration. The sensor can be calibrated against a solution having known conductivity or against a referee meter and sensor. If using a standard solution, choose one having conductivity in the recommended operating for the sensor cell constant. Refer to the analyzer manual or product data sheet for recommended ranges. Do not use standard solutions having conductivity less than about 100 uS/cm. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25°C). Because 0.01/cm sensors must be calibrated in low conductivity solutions, they are best calibrated against a referee meter and sensor in a closed system.
For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website.

Figure 11. Checking RTD Disconnect leads and measure resistances shown. The measured resistance at room temperature should be close to the value in the table.
Figure 12. Checking Continuity and Leakage Disconnect electrode leads and measure resistance and continuity as shown. Sensor must be dry when checking resistance between electrode leads.



red · red/white RTD · white



Rev. P
facebook.com/EmersonRosemountAnalytical
Credit Cards for U.S. Purchases Only.
youtube.com/user/RosemountAnalytical
©2015 Rosemount Analytical, Inc. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners.
Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 www.RosemountAnalytical.com
The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Off-scale reading | Wiring is wrong. | Verify wiring. | p.5 |
| Off-scale reading | RTD is open or shorted. | Check RTD for open or short circuits. See Figure 11. | p.5 |
| Off-scale reading | Sensor is not in process stream. | Be sure sensor is completely submerged in process stream. | p.5 |
| Off-scale reading | Variopol cable is not properly seated. | Loosen connector and reseat. | p.5 |
| Off-scale reading | Sensor has failed. | Perform isolation checks. See Figure 12. | p.5 |
| Noisy reading | Sensor is improperly installed in process stream. | Be sure sensor is completely submerged in process stream. | p.5 |
| Noisy reading | Variopol cable is not properly seated. | Loosen connector and reseat. | p.5 |
| Reading seems wrong (lower or higher than expected) | Bubbles trapped in sensor. | Be sure sensor is properly oriented in pipe or fol w cell. See Figure 1. Apply back pressure to fol w cell. | p.5 |
| Reading seems wrong (lower or higher than expected) | Wrong temperature correction algorithm. | Check that temperature correction is appropriate for the sample. See analyzer manual for more information. | p.5 |
| Reading seems wrong (lower or higher than expected) | Wrong cell constant. | Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. | p.5 |
| Sluggish response | Electrodes are fouled. | Clean electrodes. | p.5 |
| Sluggish response | Sensor is installed in dead area in piping. | Move sensor to a location more representative of the process liquid. | p.5 |