Instruction Sheet PN 51A-140/rev.E December 2010

Models 140, 141, and 142

Conductivity Sensors

For additional information, please visit our website

at www.emersonprocess.com/raihome/liquid/.

CAUTION

CAUTION

SENSOR/PROCESS APPLICATION COMPATIBILITY

BEFORE REMOVING THE SENSOR, be absolutely certain the process pressure is reduced to 0 psig and the process tempera- ture is at a safe level!

The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user.

SENSOR SPECIFICATIONS

SPECIFICATIONSMODEL 140MODEL 141MODEL 142 Wetted Materials316SS, PEEK, Viton1316SS, PEEK, Viton1316SS, Viton1 PEEK (high temp options), Kel-F2(stand temp options) Temperature (standard)302ºF (150ºC) maxsee graphsee graph Temperature (high temp)392ºF (200ºC) maxsee graphsee graph Maximum Pressure100 psig (791 kPa abs)see graphsee graph
INSTALLATION HARDWARE FOR 140 SENSOR SPECIFICATIONBall valve kit (PN 23724-00)Fitting kit (PN 23730-00) Wetted Materials316SS316SS, PEEK

Viton1 is a registered trademark of DuPont Performance Elastomers. Kel-F2 is a registered trademark of 3M

Emerson 141 — SENSOR SPECIFICATIONS diagram (page 1)
Fig. 1
Page 2 of the Emerson 141 user manual — BALL VALVE
Page 2 · click to enlarge
Page 2

MODEL 140,141, and 142 sensors INSTALLATION

INSTALLATION

o a r a c a n p r e e c u ie d p e r e th m ity u s n a g p e ly p p e d e r m ta 1 le a s ip e d e u e e d r /4 a tio s in d r la n in r d n g c .th o . m s in a te o r in a o c th s n T r in ll g . o h e n h n in ( a e e is A d p g 6 c r r s m th e o o in o e to tio m b c p le s u n s u e c e . m o . n ta s ) b t tr p le s S b o c e T c o lle e f d o le n tr d le e e liq s b o a s a c u r F a d in a o a c id e ig c , m tm n n k s a u c . u o c tr r u p s s e s o e t i.e r T l p r m ., e id b 1 h b a u s e b e e e p s to r u fo la p s r u tw c e b r , e te r tr e o th b , e d e b m le c a n b u p th to o le u b m e e le v fo m b b th r le e c b te l w e m le c o m ly s e le tr o n d a ith f s n r o s d d m a d u tio u w th e th a e c b n in s -K yIf dm ea ills -a e ,c .tiv° ° FIGURE 1. Sensor Orientation° ° FIGURE 1. Sensor Orientation INSTALLATION – 140 SENSOR WITH BALL VALVE KIT (PN 23724-00)* 1. Install the sensor in either a 1-inch NPT weldalet or in a 1-inch pipe tee. 2. Remove the plastic shipping cap from the sensor. 3. Screw the 1-inch hex nipple into the weldalet or pipe tee. See Figure 2. Use pipe tape on the threads. 4. Position the sensor for easy access to the ball valve handle, sensor compression fitting nut, and junction box. 5. Make sure the ball valve is in the fully open position. 6. Finger tighten the sensor compression fitting nut. Do not over tighten because the next step is to press the sensor into the process pipe. 7. Insert the sensor tube until the sensor tip is no closer than 1 inch (25 mm) from the far wall of the process pipe. See Figure 2. 8. Tighten the sensor compression fitting nut to hold the sensor tip in position. See Figure 2 for instructions. *If the ball valve assembly is already in place and the process line is pressurized, refer to Inserting the 140 Sensor Section on page 6.
Emerson 141 — INSTALLATION diagram (page 2)
Fig. 2

SENSOR COMPRESSION

FITTING

PROCESS PIPING

SENSOR

BALL VALVE

A

DWG. NO. REV. 40014008 E

1-in. NPT HEX NIPPLE

WELDALET

PUT WRENCH “A” HERE AND TURN WRENCH “A”

PUT WRENCH “B” HERE

Hold body with wrench B and turn WRENCH A 1 1/4 turns beyond fin- ger tight.

Tighten finger tight before inserting sensor

BALL VALVE

SIDE VIEW

WRENCH “B”

TOP VIEW

FIGURE 2. Installing model 140 sensor with ball valve kit (PN 23724-00).

Page 3 of the Emerson 141 user manual — DWG. NO. REV. 40014204 B
Page 3 · click to enlarge
Page 3

MODEL 140,141, and 142 sensors INSTALLATION

INSTALLATION – 141 SENSOR 1. Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee. 2. Remove the plastic shipping cap from the sensor. 3. Screw the sensor into the fitting. Use pipe tape on the threads. See Figure 3.

Emerson 141 — DWG. NO. REV. 40014204 B diagram (page 3)
Fig. 3

DWG. NO. REV. 40014204 B

FIGURE 3. Installing model 141 sensor

INSTALLATION – 142 SENSOR 1. Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee. See Figure 4. 2. Remove the plastic shipping cap from the sensor. 3. Screw the sensor into the fitting. Use pipe tape on the threads. DO NOT tighten the sensor compression fitting until the sensor is correctly positioned. 4. If necessary, loosen the sensor compression fitting and position the sensor so that the tip of the sensor is at least 1-inch (25 mm) from the far wall of the pipe. 5. Tighten the compression fitting using the procedure shown in Figure 2.

Emerson 141 — DWG. NO. REV. 40014204 B diagram (page 3)
Fig. 4

DWG. NO. REV. 40014204 B

FIGURE 4. Installing model 142 sensor

Page 4 of the Emerson 141 user manual — DWG. NO. REV. 40014204 B
Page 4 · click to enlarge
Page 4
Emerson 141 — DWG. NO. REV. 40014204 B diagram (page 4)
Fig. 5

MODEL 140,141, and 142 sensors WIRING WIRING All 140 series sensors have a junction box mounted on the back of the sensor. Wiring connections in the junction box are shown in Figure 5.

9 8 7 6 5 4 3 2 1

Figure 5. Sensor junction box wiring. Terminals in the junction box are not numbered.

141 and 142 sensors have one gray wire (shown). The 140 sensor has two gray wires attached to the terminal.

WIRING DIAGRAMS FIGURE 6. Model 54eC WiringFIGURE 7. Model 1056 and 56 Wiring FIGURE 6. Model 54eC Wiring
Emerson 141 — DWG. NO. REV. 40014204 B diagram (page 4)
Fig. 6
Page 5 of the Emerson 141 user manual — TB1 TB2
Page 5 · click to enlarge
Page 5

MODEL 140,141, and 142 sensors WIRING

Emerson 141 — TB1 TB2 diagram (page 5)
Fig. 7
Emerson 141 — TB1 TB2 diagram (page 5)
Fig. 8

FIGURE 8. Model Xmt-C-10 Wiring (Panel)

FIGURE 9. Model Xmt-C-11 Wiring (Pipe or Wall)

TB1 TB2

RTN

WHITE

RCV B

SENSE

WHITE/RED

RCV A

RTD IN

RED

CLEAR

RSHLD

SHLD CLEAR

GRAY

DRVB

ORANGE

DRVA

CLEAR

DSHLD

FIGURE 10. Model 1066 Wiring

Emerson 141 — TB1 TB2 diagram (page 5)
Fig. 9
Emerson 141 — TB1 TB2 diagram (page 5)
Fig. 10

FIGURE 11. Model 5081C Wiring

FIGURE 12. Model 6081-C Wiring

Page 6 of the Emerson 141 user manual — WARNING
Page 6 · click to enlarge
Page 6

MODEL 140,141, and 142 sensors WIRING

RETRACTING AND INSERTING THE MODEL 140 SENSOR

CAUTION

Model 140 sensors are retractable. BEFORE RETRACTING THE SENSOR, be absolutely certain the process pressure is less than 100 psig (791 kPa abs) and the process temperature is at a safe level!

RETRACTING THE SENSOR 1. Push in on the sensor junction box and slowly loosen the sensor compression fitting nut by reversing the sensor tightening procedure illustrated in Figure 2. 2. When the sensor compression nut is completely unscrewed, slowly ease the sensor out until the flared tip of the electrode rests firmly within the body of the compression fitting body. 3. Close the ball valve completely.

CAUTION

Before removing the sensor be sure the ball valve is completely closed.

4. Unscrew the compression fitting body from the reducing bushing and remove the sensor from the ball valve assembly.

INSERTING THE SENSOR

CAUTION

Make sure process O-ring is clean, lubricated, and in place before installing sensor. Replace if worn.

1. DO NOT open the ball valve.

WARNING

The system pressure must be less than 100 psig (791 kPa).

2. Thread the sensor compression fitting body into the reducing bushing in the rear of the ball valve and tighten. NOTE: Do not push past this point. Damage to the sensor could result.

WARNING

If the sensor comes free of the valve, refer to Figures 2 and 13 and verify that the valve and associated fittings are as shown. Do not proceed until the sensor is correctly restrained.

Page 7 of the Emerson 141 user manual — CAUTION
Page 7 · click to enlarge
Page 7

MODEL 140,141, and 142 sensors WIRING

3. Slowly open the valve.

WARNING

Stand clear of the sensor.

4. Insert the sensor up to the desired insertion depth and turn the sensor compression fitting nut until it is finger tight.
5. Position the entire sensor for easy access to the ball valve handle, sensor compression fitting nut and J-box terminal block.
6. Tighten sensor compression fitting nut.

CAUTION

For initial installation of the sensor, tighten the com- pression fitting nut 1-1/4 turns after finger tight. If it is a reinstallation, turn no more than 1/4 to 1/2 addi- tional turns!

REMOVING AND REINSTALLING THE MODEL 142 SENSOR

REMOVING THE SENSOR

WARNING

WARNING

Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!

1. Reduce process temperature and pressure to a safe level. If necessary drain the process line. 2. Loosen the sensor compression fitting and slowly slide the sensor from the pipe fitting or weldalet.

REINSTALLING THE SENSOR 1. Slide the sensor into the process fitting and position the sensor the way it was originally installed.

CAUTION

Be sure the sensor is in the original position. The sensor tube takes a permanent set and could become weakened if the new set is adjacent to the original set.

2. Tighten the sensor compression fitting 1/4 to 1/2 turn after it is finger tight.
Page 8 of the Emerson 141 user manual — CALIBRATION
Page 8 · click to enlarge
Page 8

MODEL 140,141, and 142 sensors CALIBRATION

CALIBRATION

Model 140 sensors are not calibrated at the factory. The cell constant on the label is a nominal value only. The true cell constant can differ from the nominal value by as much as ±5%. For improved accuracy, calibrate the sensor using either a solution of known conductivity or a referee meter and sensor. If using a standard solution, choose one having conductivity in the recommended operating range for the sensor cell constant. Refer to the analyzer manual or product data sheet and use the range recommended for Model 400 series sensors. For the 0.2/cm cell constant Model 140 sensor, use the range for the 0.1/cm Model 400 sensor and multiply the upper and lower limits by two.

Do not use standard solutions having conductivity less than about 100 uS/cm for calibration. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25°C). Because 0.01/cm sensors must be calibrated in low conductivity solutions, they are best cal- ibrated against a referee meter and sensor in a closed system.

For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website.

Page 9 of the Emerson 141 user manual — CALIBRATION
Page 9 · click to enlarge
Page 9

MODEL 140,141, and 142 sensors MAINTENANCE

MAINTENANCE CLEANING THE SENSOR Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.

CHECKING MODEL 140 RETRACTION RESTRAINT The integrity of the Model 140 will become compromised is the flared tip of the electrode is allowed to blow out against the compression fitting body. In the even a blowout occurs, replace the sensor.

REPLACING MODEL 140 SENSOR SEAL. If the process seal is leaking owing to a pitted or uneven sensor tube, a replacement sensor is required. If the sensor tube surface is smooth and clean yet the process seal is leaking, the process O-ring is damaged and requires replace- ment according to the following procedure (see Figure 2). Replacement parts can be obtained from the Process Fitting Rebuild Kit (PN 23731-00). 1. The junction box with attached compression fitting body, nut and compression fitting must be recovered from the sensor for reuse. Unscrew the junction box cover and set aside. Mark and disconnect the electrical connections from the terminal block. Remove the junction box compression fitting nut from the compression fitting body and separate the junction box from the sensor tube. 2. Remove the nylon ferrule and snap ring (discard both). Remove and save the junction box compression fitting nut. 3. Slide off the sensor compression fitting nut and set aside for reuse. Slide off the remaining PEEK ferrule and split ring (discard both). 4. Remove the sensor compression fitting body and replace the Viton O-ring. Lubricate the O-ring with the barium based lubricant provided. 5. Wrap the threads of the sensor compression fitting body with pipe tape and slide the body on to the sensor tube. 6. Slide on a new PEEK ferrule, beveled side facing the electrode tip, and a new PEEK split ring, flared end towards electrode tip. Slide on the sensor compression fitting nut and thread it onto the compression fitting body. Finger tighten. 7. Reinstall the J-box on the sensor tube. Finger tighten the J-box compression fitting nut. Use a wrench to turn the nut a 1/4 to 1/2 additional turn.

Page 10 of the Emerson 141 user manual — DWG. NO. REV. 40014013 D
Page 10 · click to enlarge
Page 10

MODEL 140,141, and 142 sensors ACCESSORIES

REPLACEMENT PARTS

PNDescription 23724-00Ball valve kit for Model 140 sensor 23730-00Process compression fitting kit (3/4-inch NPT) for Model 140 sensor 23731-00Process fitting rebuild kit for Model 140 sensor (includes ferrule, split ring, O-ring, lubricant) 9310120Junction box compression fitting for Model 140 sensor 9550200O-ring, 2-116, Viton for Model 140 sensor 3001882Process compression fitting (3/4-inch NPT) for Model 142 sensor
Emerson 141 — REPLACEMENT PARTS diagram (page 10)
Fig. 11

Junction Box Compression Fitting (PN 9310120)

Sensor Compression Fitting Nut

(1-in. NPT Hex Nipple)

Peek Split Ring (inside)

Included in Kit (PN 23730-00)

Viton O-ring (inside) (PN 9550200)

Ball Valve Kit (PN 23724-00)

DWG. NO. REV. 40014013 D

FIGURE 13. Model 140 with Ball Valve Kit (PN 23724-00)

Show 8 more labels / short text on this page
Junction Box · Junction Box Cover · Nylon Ferrule (inside) · Sensor Tube · Peek Ferrule (inside) · Compression Fitting Body · (Reducing Bushing) · Flared Mechanical Stop
Page 11 of the Emerson 141 user manual — RTD
Page 11 · click to enlarge
Page 11

MODEL 140,141, and 142 sensors TROUBLESHOOTING

TROUBLESHOOTING PROBLEMPROBABLE CAUSESOLUTION Off-scale readingWiring is wrong.Verify wiring. Temperature element is open or shorted.Check temperature element for open or short circuits. See Figure 14. Sensor is not in the process stream.Be sure the sensor is completely submerged in the process stream. Sensor has failed.Perform isolation checks. See Figure 15. Noisy readingSensor is improperly installed in the process stream.Be sure sensor is completely submerged in process stream. Reading seems wrong (lower or higher than expected)Bubbles trapped in the sensor.Be sure the sensor is properly oriented in the pipe or flow cell. See Figure 1. Apply back pressure to the flow cell. Wrong temperature correction algorithm.Check that the temperature correction is appropriate for the sample. See analyzer manual for more information. Wrong cell constant.Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. Sluggish responseElectrodes are fouled.Clean electrodes. Sensor is installed in a dead area in the piping.Move the sensor to a location more representative of the process liquid.

RTD

terminal strip in sensor

Pt 10010K NTC100K NTC 0100.029.49K351K 10103.918.79k208K 20107.812.26K127K 30111.7819479.4K 40115.5559251.0K 50119.4389333.6K

FIGURE 15. Checking Continuity and Leakage Disconnect cable leads and measure resistance and con- tinuity as shown. Sensor must be dry when checking resistance between electrode leads.

Show 5 more labels / short text on this page
orange · R red · red · junction box · gray
Page 12 of the Emerson 141 user manual — RTD
Page 12 · click to enlarge
Page 12
Emerson 141 — RTD diagram (page 12)
Fig. 12

The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE

http://www.raihome.com

Specifications subject to change without notice.

Credit Cards for U.S. Purchases Only.

2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250

http://www.raihome.com

© Rosemount Analytical Inc. 2010

Emerson Process Management

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson 141

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson 141

Be the first to ask — other owners or our team will answer.

Original document

Emerson 141 manual PDF

12 pages·402 KB·Stud Sensor·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…