Instruction Sheet PN 51A-140/rev.E December 2010
Conductivity Sensors
at www.emersonprocess.com/raihome/liquid/.
CAUTION
BEFORE REMOVING THE SENSOR, be absolutely certain the process pressure is reduced to 0 psig and the process tempera- ture is at a safe level!
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user.
Viton1 is a registered trademark of DuPont Performance Elastomers. Kel-F2 is a registered trademark of 3M


MODEL 140,141, and 142 sensors INSTALLATION

FITTING
PROCESS PIPING
SENSOR
BALL VALVE
A
DWG. NO. REV. 40014008 E
1-in. NPT HEX NIPPLE
PUT WRENCH “A” HERE AND TURN WRENCH “A”
Hold body with wrench B and turn WRENCH A 1 1/4 turns beyond fin- ger tight.
Tighten finger tight before inserting sensor
SIDE VIEW
WRENCH “B”
TOP VIEW
FIGURE 2. Installing model 140 sensor with ball valve kit (PN 23724-00).

MODEL 140,141, and 142 sensors INSTALLATION
INSTALLATION – 141 SENSOR 1. Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee. 2. Remove the plastic shipping cap from the sensor. 3. Screw the sensor into the fitting. Use pipe tape on the threads. See Figure 3.

FIGURE 3. Installing model 141 sensor
INSTALLATION – 142 SENSOR 1. Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee. See Figure 4. 2. Remove the plastic shipping cap from the sensor. 3. Screw the sensor into the fitting. Use pipe tape on the threads. DO NOT tighten the sensor compression fitting until the sensor is correctly positioned. 4. If necessary, loosen the sensor compression fitting and position the sensor so that the tip of the sensor is at least 1-inch (25 mm) from the far wall of the pipe. 5. Tighten the compression fitting using the procedure shown in Figure 2.

FIGURE 4. Installing model 142 sensor


MODEL 140,141, and 142 sensors WIRING WIRING All 140 series sensors have a junction box mounted on the back of the sensor. Wiring connections in the junction box are shown in Figure 5.
9 8 7 6 5 4 3 2 1
Figure 5. Sensor junction box wiring. Terminals in the junction box are not numbered.
141 and 142 sensors have one gray wire (shown). The 140 sensor has two gray wires attached to the terminal.


MODEL 140,141, and 142 sensors WIRING


FIGURE 8. Model Xmt-C-10 Wiring (Panel)
FIGURE 9. Model Xmt-C-11 Wiring (Pipe or Wall)
RTN
WHITE
RCV B
SENSE
WHITE/RED
RCV A
RTD IN
RED
CLEAR
RSHLD
SHLD CLEAR
GRAY
DRVB
ORANGE
DRVA
CLEAR
DSHLD
FIGURE 10. Model 1066 Wiring


FIGURE 11. Model 5081C Wiring
FIGURE 12. Model 6081-C Wiring

MODEL 140,141, and 142 sensors WIRING
CAUTION
Model 140 sensors are retractable. BEFORE RETRACTING THE SENSOR, be absolutely certain the process pressure is less than 100 psig (791 kPa abs) and the process temperature is at a safe level!
RETRACTING THE SENSOR 1. Push in on the sensor junction box and slowly loosen the sensor compression fitting nut by reversing the sensor tightening procedure illustrated in Figure 2. 2. When the sensor compression nut is completely unscrewed, slowly ease the sensor out until the flared tip of the electrode rests firmly within the body of the compression fitting body. 3. Close the ball valve completely.
Before removing the sensor be sure the ball valve is completely closed.
CAUTION
Make sure process O-ring is clean, lubricated, and in place before installing sensor. Replace if worn.
The system pressure must be less than 100 psig (791 kPa).
If the sensor comes free of the valve, refer to Figures 2 and 13 and verify that the valve and associated fittings are as shown. Do not proceed until the sensor is correctly restrained.

MODEL 140,141, and 142 sensors WIRING
Stand clear of the sensor.
For initial installation of the sensor, tighten the com- pression fitting nut 1-1/4 turns after finger tight. If it is a reinstallation, turn no more than 1/4 to 1/2 addi- tional turns!
WARNING
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
REINSTALLING THE SENSOR 1. Slide the sensor into the process fitting and position the sensor the way it was originally installed.
Be sure the sensor is in the original position. The sensor tube takes a permanent set and could become weakened if the new set is adjacent to the original set.

MODEL 140,141, and 142 sensors CALIBRATION
Model 140 sensors are not calibrated at the factory. The cell constant on the label is a nominal value only. The true cell constant can differ from the nominal value by as much as ±5%. For improved accuracy, calibrate the sensor using either a solution of known conductivity or a referee meter and sensor. If using a standard solution, choose one having conductivity in the recommended operating range for the sensor cell constant. Refer to the analyzer manual or product data sheet and use the range recommended for Model 400 series sensors. For the 0.2/cm cell constant Model 140 sensor, use the range for the 0.1/cm Model 400 sensor and multiply the upper and lower limits by two.
Do not use standard solutions having conductivity less than about 100 uS/cm for calibration. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25°C). Because 0.01/cm sensors must be calibrated in low conductivity solutions, they are best cal- ibrated against a referee meter and sensor in a closed system.
For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website.

MODEL 140,141, and 142 sensors MAINTENANCE
MAINTENANCE CLEANING THE SENSOR Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.
CHECKING MODEL 140 RETRACTION RESTRAINT The integrity of the Model 140 will become compromised is the flared tip of the electrode is allowed to blow out against the compression fitting body. In the even a blowout occurs, replace the sensor.
REPLACING MODEL 140 SENSOR SEAL. If the process seal is leaking owing to a pitted or uneven sensor tube, a replacement sensor is required. If the sensor tube surface is smooth and clean yet the process seal is leaking, the process O-ring is damaged and requires replace- ment according to the following procedure (see Figure 2). Replacement parts can be obtained from the Process Fitting Rebuild Kit (PN 23731-00). 1. The junction box with attached compression fitting body, nut and compression fitting must be recovered from the sensor for reuse. Unscrew the junction box cover and set aside. Mark and disconnect the electrical connections from the terminal block. Remove the junction box compression fitting nut from the compression fitting body and separate the junction box from the sensor tube. 2. Remove the nylon ferrule and snap ring (discard both). Remove and save the junction box compression fitting nut. 3. Slide off the sensor compression fitting nut and set aside for reuse. Slide off the remaining PEEK ferrule and split ring (discard both). 4. Remove the sensor compression fitting body and replace the Viton O-ring. Lubricate the O-ring with the barium based lubricant provided. 5. Wrap the threads of the sensor compression fitting body with pipe tape and slide the body on to the sensor tube. 6. Slide on a new PEEK ferrule, beveled side facing the electrode tip, and a new PEEK split ring, flared end towards electrode tip. Slide on the sensor compression fitting nut and thread it onto the compression fitting body. Finger tighten. 7. Reinstall the J-box on the sensor tube. Finger tighten the J-box compression fitting nut. Use a wrench to turn the nut a 1/4 to 1/2 additional turn.

MODEL 140,141, and 142 sensors ACCESSORIES

Junction Box Compression Fitting (PN 9310120)
Sensor Compression Fitting Nut
(1-in. NPT Hex Nipple)
Peek Split Ring (inside)
Included in Kit (PN 23730-00)
Viton O-ring (inside) (PN 9550200)
Ball Valve Kit (PN 23724-00)
FIGURE 13. Model 140 with Ball Valve Kit (PN 23724-00)

MODEL 140,141, and 142 sensors TROUBLESHOOTING
terminal strip in sensor
FIGURE 15. Checking Continuity and Leakage Disconnect cable leads and measure resistance and con- tinuity as shown. Sensor must be dry when checking resistance between electrode leads.


The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2010