Abbott Diabetes Care Inc. 1360 South Loop Road Alameda, CA 94502, USA
Precision Xceed Pro, TrueID, TrueMeasure, TrueAccess, PrecisionWeb, Precision, MediSense, Optium and Precision PCx are trademarks of Abbott Group of Companies in different jurisdictions. Other trademarks are owned by their respective owners.
ART14004 Rev. C 03/09
© 2009 Abbott


Intended use
TrueID, TrueMeasure and TrueAccess
Icons and warnings
Conventions
System components

Intended use The Precision Xceed Pro Point of Care blood glucose and blood beta ketone control system is intended for In vitro diagnostic quantitative determination of glucose (D-glucose) in fresh capillary, venous, arterial and neonatal whole blood samples and β ketones (betahydroxybutyrate) in Fresh Capillary and Vein Whole Blood Samples. The system is not intended for use in diagnostics of diabetes mellitus but is intended to help monitor the effectiveness of diabetes regulation.
The Precision Xceed Pro system simplifies point-of-care tests for healthcare professionals using functions that increase the reliability of the test process and ensure best compliance with testing policies.
Easy Choose between the built-in barcode scanner and the alphanumeric keypad for the data input entering data required under the test policy. The meter can be configured in such a way that patient and operator identification numbers, test strip lottery numbers, comment codes and up to two fields of free text, such as doctors' names, may be entered.
Confirmation of the The meter can display the name, date of birth and sex of the patient in support Patient ID of good patient identification procedures. The year of birth may also be required to confirm.
Data integrity There are many options available that help to collect the correct data per test, such as: specifying the minimum and maximum length of an ID; limiting the bar code types accepted; using a bar code verification check digit; or requesting a repetition of keypad input when not using the barcode scanner. The meter also records the method by which patient and operator identification numbers have been entered (with barcode scanner or manually with keyboard) as a help for point-of-care coordinator in identifying and correcting errors.
Quick and easy testing A test is performed simply by inserting a strip into the gauge's test strip port, a
Put a drop of blood in the bag and wait for the countdown to be completed. The sample may be applied to the top and end of the test strip for convenience. The tests will only start when sufficient blood has been applied.
Simple calibration By scanning the bar code of each foil packaging with test strips before use, the meter is automatically calibrated and its expiry date checked, which contributes to reliable and accurate test results. Testing may also be limited to approved strip slots.
Exclusion The meter can be programmed in such a way that periodic testing with control solutions is required quality control is and that access to the system is only possible again after these tests have been carried out. Results of quality checks shall only be displayed as OK/fault, preventing patient tests while the meter is in control test mode.
Operator Management Access to the meter may be restricted to entitled operators, thereby excluding use by untrained users. A warning can also be displayed for operators whose ID is almost expired.
Storage of results The Precision Xceed Pro meter can store up to 2500 patient test results, 1000 control test results and 20 skill test results as well as 2 linearity spanners. All stored data can be automatically uploaded to the data management system using a docking station (optional) or a data upload cable (optional).
Network connectivity Precision Xceed Pro meters can be uploaded and configured anywhere in the institution's computer network via a central data management system. This is a simple and automated way of managing multiple instruments, point-of-care monitoring and transmitting test data to the Laboratory Information System (LIS) or Hospital Information System HIS for inclusion in the Electronic Medical Record (EMR).

Icons and warnings Information of special interest to the reader is given in grey boxes throughout the manual with an appropriate pictogram and warning level.
instruction not followed.
NB: Provides the operator with useful additional information or background information, or refers to it.
Conventions The glucose unit of measurement is set up via the data management system. This unit cannot be changed with the meter keypad. When this manual shows a screenshot of the meter with glucose test result in one unit of measurement (e.g. mmol/L) then the equivalent value is shown in the other unit of measure (e β-ketone test results are always shown in mmol/L.


1-2 Do you have any questions? See pag. 10-10 for instructions on technical assistance. Chapter 1: Overview
105 mg/dL

System components The Precision Xceed Pro meter is part of a total system designed to simplify point-of-care tests. These parts are sold separately and some of them are optional.
Precision Xceed Pro System Parts The Precision Xceed Pro system consists of the following components:
not used in accordance with the manufacturer's documentation This may be at the expense of safety in the product (s).
Meter The Precision Xceed Pro meter features many functions designed to simplify testing. The meter can be operated with both the left and right hand. The view window has large text and backlight to make it easy to read. The keypad looks like a phone and all the keys click when pressed. The use of the meter is described in Chapters 2 to 8 of this manual and in the short guide (available separately).

The test strip port is located at the top of the meter and is covered by the gate protector. The port protector is designed to minimise the risk of moisture entering the meter via the test strip port (Chapter 11 contains instructions for replacing the gate protector). The barcode scanner window is under the test strip port. The battery compartment is located at the rear of the meter (Chapter 11 also contains instructions for replacing batteries). Right underneath the battery compartment is a label with the unique serial number of the meter. On the side of the meter is room for gauge identification. On this non-textured surface (22 mm wide, 11 mm high) an adjustment specific location label can be glued. At the bottom is a data port that allows the meter to be connected to a data upload cable or docking station (Chapter 9 contains information on uploading data to a given management system).
Each Precision Xceed Pro meter comes with this manual (on CD-ROM), one port protector (installed) and two AA batteries.


Test strips Precision Xceed Pro test strips are equipped with the latest in biosensor technology. The test strips work by inserting the contact stripes into the meter first. The sample shall then be applied to the test surface covering both the working electrode and the reference electrode. This area is coated with active substances that react with the glucose present, creating a small electrical current. This stream is led through the stripe to the contact lines and meter, which calculates a glucose result.

Precision Xceed Pro glucose test strips apply a glucose-specific chemical process that is owned by the company and in which the enzyme glucose.dehydrogenase with NAD cofactor and PQ mediator (GDH-NAD) plays a role. This chemical process is used for glucose testing as it is not affected by normal therapeutic concentrations of maltose, icodextrin and other common substances. This process also ensures minimal distortion of measurement by oxygen, haematocrit and other physiological variables.
Precision Xceed Pro blood beta ketone test strips are also available for use with the Precision Xced Pro system. These test strips use a chemical process that specifically measures β-hydroxybutyrate, the main ketone body produced when a patient develops ketoacidosis.

Both glucose and β ketone test strips are further equipped with exclusive blood application functions that ensure reliable sampling. First, the test surface can be completely covered by applying a drop of the sample to the top of the test strip or by baptizing the end of the testing strip into the sample. Second, the reaction start electrode ensures that the test does not commence until sufficient blood has been applied. Together, these functions minimise the risk of false results due to the technique used for applying the sample.

Each test strip is protected by a foil packaging with its own bar code. The foil packaging keeps the test strip intact by protecting it from exposure to air, moisture and contamination. The bar code label shall contain information on the test strip, including its lot number, calibration, expiry date and expected results of the control solution. With one quick scan, all this information is entered into the meter at the time of test.

If you do not, it may produce false results. Use the test strip immediately after use of opening the foil packaging. Do not use the test strip if foil packaging is punctured or ruptured.
There are three types of test strips that can be used for testing with the Precision Xceed Pro system: the light blue Precision Xced Pro blood glucose test strip (in blue foil packaging) and the green Precision PCx Plus blood glucose Test Strip (in gold foil packaging), to test the glucose level in fresh whole blood, as well as the purple Precision XCed Pro Blood β ketone test strip for testing the β ketone level in full blood. Each carton of test strips contains a package leaflet with an extensive instruction manual.

Test strips. Using other test strips may result in incorrect results.
1-4 Do you have any questions? See pag. 10-10 for instructions on technical assistance. Chapter 1: Overview
Test surface · Contact stripes

Control solutions Control solutions are used to perform quality checks with the meter regularly to check that it is still working properly. Control tests are further described in Chapter 4. Use Precision, Optium or MediSense blood glucose and beta-ketone control solutions in the Precision Xceed Pro system. These control solutions are available in configurations with one, two and three levels.
Calibration verification checks Use glucose and β-ketone calibration verification checks of RNA Medical to check the calibration and analytical measurement range of the meter for Precision Xceed Pro blood glucose and beta-keton test strips. Use glucose calibration verification checks of RNA Medical with Precision PCx Plus blood glucose test strips. This is called linearity tests; a set of 5 levels control solution is required. Linearity tests are further described in Chapter 7.
Insulation bag (optional) Abbott's insulation bags are disposable plastic bags for use with the Precision Xceed Pro system. The isolation bags form a partial barrier between the point-of-care meter and patients isolated due to the risk of infection. The meter does not need to be specially prepared before using the bag and the bag has no influence on the functions of the gauge. Please note that Abbott's isolation bags have not been sterilised.
Suitcase (optional) The case provides space for the meter, test strips, control solutions, short guide and other testing equipment. The transparent lid allows you to check the stock without having to open the case.


Docking station (optional) The docking station provides a convenient way to upload the data from the meter to the data management system. When the meter is placed in the docking station, data are automatically uploaded to the data management system and then the meter will be switched off. Uploading data is further described in chapter 9.
Data upload cable (optional) A data upload cable is also available to connect the meter to the computer on which the data management system runs.
Data management system A computer with a data management system is needed to configure the Precision Xceed Pro meter and upload results. The PrecisionWeb Point-of-Care data management system of Abbott is available for this purpose, as well as software from other companies. For uploading data from remote locations, terminal servers or network PCs are required to run Abbott Data Repeater software.

Strip code scanner
Data input keypad
Sound indicator
Data port
Battery compartment
Port protector
Meter identification
Menu Tree

Strip code scanner The barcode scanner allows you to scan the information in the meter instead of entering it manually via the keypad. By scanning the bar code on the foil packaging of the test strips, the test strip information (calibrate, control range, lot number and expiry date) is automatically introduced. The scanner can also be used for entering the operator ID, patient ID, control solution lot number, comment code and free text. Control the meter as follows:



NB: . . If you hold the Scan key for three seconds, the scanner stops. Redirect the scanner and try
Scan again. The optimal distance depends on the type of bar code.
Some precautions should be taken when learning to use the bar code scanner. It is important that you place the object to be scanned on a flat surface or hold it. This prevents other objects from being scanned by accident.
Manually entered data is erased and replaced by scanned data, except when the manual input is confirmed by pressing the Enter key. This input is considered scanned. After a barcode has been successfully scanned, the system automatically moves to the next screen. The scan beam will be switched off after three seconds if no bar code is seen.

Someone's eyes. The beam can cause permanent eye damage. CDRH Class II/IEC Class 2 laser product: avoid long-term looking at Direct laser light.
Data input keypad The keypad for data input allows you to enter identification numbers and letters (only the 26 letters of the Dutch alphabet) or select an option that appears in the viewing window. The keypad contains 10 number keys numbered 0 to 9. On the keys 2 to 9 are also letters below the number. The keyboard also contains six special keys: the Scan key, the On/Off key (on/off), the backlight/alphanumeric key, Clear key (erase), Menu key and Enter key.

Scan Press this key to use the barcode scanner. When the sound indicator is activated, the meter beeps once when a bar code has been successfully scanned. By holding down the Scan key, the scanned barcode appears in the view window until the key is released. The scanner will be disabled if no information is scanned for 3 seconds.

On/Off By pressing this key, turn on the meter. Every time the meter is turned on, the Abbott logo screen appears in black for a few seconds and then in grey in the viewing window. This allows you to check that all pixels work properly.
You disable the meter by holding this key for 2 seconds. To save the batteries, automatically switch off the meter after 4 minutes. The period before automatic switch-off of the meter can be set to 4-10 minutes, but only after patient tests.
Background lighting/alphanumeric You activate the backlight by holding this button for about 2 seconds.

While entering data manually, you can press this key to switch back and forth between numerically, uppercase alpha mode and lowercase alpha Mode. In alpha mode, enter alpha characters on the screen by pressing keys 2 to 9 multiple times, with different characters in turn appearing on the display. In alpha mode, enter a space by pressing the 1 key.
The meter indicates with an alpha indicator with a capital/small letter at the bottom right of the view window which mode you are working in.
Clear If you press this key, you will erase the previous character when entering alphanumeric information with the keyboard.
This key is also used to return to a previous screen if necessary.
Menu This key will switch the meter from Test mode to Menu mode. These two modes are described later. When you view configuration information in Menu mode, return to the top of the menu tree if you press this key once. If you press the button a second time, the meter switches to Test mode.
Enter After entering data with the keypad, proceed to the next screen by pressing Enter key.
Sound indicator (not visible) The meter is equipped with a sound indicator that can be switched on or off. This indicator produces two notes: a high tone to indicate that an action has been successful, a low tone of voice to indicate there is a problem. When the sound indicator is switched on, the meter shall beep in the following cases:
Data port The data port is located at the bottom of the meter. The meter can automatically transfer the data to the data management system via a cable or docking station.
Battery compartment The meter uses two AA batteries as a power source. To extend the battery life, the meter should be switched off if it is not used. The data will remain unaffected if the batteries are removed.
Port protector The gate protector covers the test strip port. The gate protector is designed to minimise the chance of moisture entering the meter via the test strip port. If the gate protector comes into contact with blood or control solution, then the meter should be cleaned and dried off and the port protector replaced. Chapter 11 contains instructions for replacing the gate protector.

Each meter has a unique serial number, mounted on the back of the device right below the battery compartment. At the side of Precision Xceed Pro meter is a non-textured plane. In this area of 22 mm wide and 11 mm high, the setting can label with information about the meter. See Chapter 1 for an image of this.
Menu Tree When using the Precision Xceed Pro meter, navigate between two main menu menu templates: Test mode and Menu mode. The Test mode is used for performing patient and control tests. The Menu mode is used for viewing data, performing linearity and skill tests, and viewing the configuration settings and status (e.g. battery voltage) of the meter. The configuration options in Menu mode must be configured with a data management system. You can visually check existing meter settings, but do not change them.
The menu mode of the software is set up as a tree, with submenus at different levels. The Precision Xceed Pro menu tree (see below) shows the different layers with submenu .
Test mode 1. Patient test
Menu Mode 1. Data overview
Upload interval Operator-ID operator-ID truncate Operator- ID command
Glucosic units Glucosine operating area Glucose Critical range Ketones Operation area Ketons Critical range
Patient ID Truncate Patient ID Comment code As within range Comment Code As out of range Patient ID command Repeat manual ID input
Levels Results Comment code As in range Comment Code As out of range
QC interval Operator ID Unauthorized Striplot Not confirmed in list Patient data Confirm patient data

Use guidelines for all samples
Take blood samples
Take a capillary blood sample
Take a venous blood sample
To take an arterial blood sample
Take a neonatal capillary blood sample
Disposal
Patient testing procedure - glucose
Patient test procedure - β ketones

This chapter describes how fresh whole blood from patients with the Precision Xceed Pro system should be tested for blood glucose or blood beta ketones. Each carton of test strips contains a package leaflet with an extensive instruction manual.
Use guidelines for all samples Follow the recommended guidelines to obtain as accurate results as possible:
Always wear gloves when performing tests with patient blood samples and follow the policies and procedures for biologically hazardous material of your institution.
Make sure the meter and test strips are at room temperature. If the meter is moved to a room • where it is warmer or cooler than in the space where the meter was first, let the meter come at the new room temperature before testing.
Use only control solutions mentioned in the package leaflet accompanying the test strips for verifying the operation of the Precision Xceed Pro meter.
Only use Precision Xceed Pro blood glucose or beta-ketone test strips. Precision PCx Plus • Blood glucose test strips may also be used in the Precision Xceed Pro meter.
For specific instructions on how to store and use the test strips, please refer to the package leaflet. •
Do not use the test strips on foil packaging and outer carton after expiry date. •
Do not use the test strip if the foil packaging is punctured or torn. •
Do not use test strips that are wet, bent, scratched or damaged. Use the test strip immediately after opening the • foil packaging.
Use each test strip only once. •
Always use test strips from the pack you scanned for bar code. If you do not, the • results may be inaccurate.
Apply a drop of blood to the test surface at the end of the test strip. Make sure that the entire test surface is filled with • blood. If the test surface is briefly touched by the patient's finger, a capillary tube, a syringe or pipette, then this does not affect the test result.
The test strip must not be removed or disturbed after the blood has been applied to the test strip and the countdown • of the test has begun.

The meter is coming in. If it does, the meter can be irreparable damaged Hit.




Please observe the following guidelines when disposing of biological hazardous waste:

Patient testing procedure - glucose Use the following procedure to perform a patient test with either Precision Xceed Pro or Precision PCx Plus blood glucose test strip.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The Abbott logo screen appears in black for a few seconds and then grey to check that the view window is working properly. Next, the software version number appears briefly in view.

The meter starts in the Test mode. The Menu key will be used to insert the meter in the Menu Mode The Commission has already taken a decision on this matter.

The Operator ID can consist of 30 digits and therefore does not always fit in the viewing window. The meter can also be set to truncate (ignoring) first, last and/or selected digits of the bar code.


The patient ID can consist of 30 digits and therefore does not always fit fully in the viewing window. The meter can also be set to truncate (ignoring) selected digits of the bar code. .
If you enter the ID manually, it may be that you need to enter the id a second time to make sure the ID is correct.

Enter the ID again with the keypad.

Press 2 before Confirm or on 1 for ID re-enter.

Enter the year of birth (for example .63 .73 .53 or .01 .71 for 2001) and press Enter-test.
Press Clear to re-enter the ID. If numbers are entered, then they will be deleted by pressing the Clear key first. The Clear key may need to be pressed several times to return to the input screen Patient ID in step 4.

Press 2 before Continue with testing or on 1 before Re-enter ID.

By scanning the bar code, the type of strip is determined, the meter calibrated, checked whether the expiry date has not expired, the lot strips used registered and verified if your setup approved it for use.

Use mode Veneus only for venous samples from sealed tubes and not mixed with air. Use mode for all other samples Arterial/Capillar. If the screen is not shown Monster type, continue with step 8.

With clean, dry hands you can gently touch the test strip anywhere when removing the strip from the foil packaging to insert into the meter.



Before inserting the test strip, check that the gate protector is in place and it is clean and dry. Replace the gate protector if necessary. In the view window, Enter Strip alternated with Glucose.

When sufficient blood is applied, the meter beeps, Sample good the test is automatically started. Refer to the Test Strips section in Chapter 1 for a detailed description of the test strip surface.

The meter is counted and then shows the test result. If an error is detected during the test, then the test shall be terminated and no result shall be displayed. From the start of the test, tests shall no longer respond to impressions until the test is complete.

The meter can be set to scan or manually enter a 1- or 2-digit comment code.

If no comment code is requested, proceed to step 13.
Scan if requested the information in free text or enter it manually and press Enter.
The meter can be set up to allow the scanning or input of information as a free text. A maximum of two fields of free text can appear.

If you do not ask for the free text to be entered, proceed with step 13.
105 mg/dL · 105 mg /dL

If you have results that are out of reach, please refer to your institution's policy.
Your institution may require a scope of application The Commission's proposal is based on the following principles: The meter shows results beyond this range. Results outside the scope may require a comment code.

207 mg/dL Higher than the scope of application
If you have results that are out of reach, please refer to your institution's policy.
Your institution may require a critical range The Commission's proposal is based on the following principles: The meter does not display results beyond this range. Results outside the critical range may require a comment code.

>400 mg/ dl Higher than critical range
The view window shows: Patient ID or Name (p: assignment) Comment code (number in box) Test type date and time (Glucose) If 1 select, then the screen appears Scan Patient ID. Go back to step 4 and continue testing the next patient. If 2 If selected, the meter will show this patient's latest test result and you can also view previous results. Refer to Patient Data Summary for further information per patient ID in Chapter 5.
Then there may be a problem with the test strip. Repeat the test with a new test strip. Incorrect results can have serious medical consequences. Please refer to the Prescriptive physician before changing diabetes medication if:
105 mg/dL

Use the following procedure for performing patient tests with the Precision Xceed Pro blood beta ketone test strip. The Precision Xceed Pro meter recognizes the type of test strip by using the scanned bar code of the strip.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
Before inserting the test strip, check that the gate protector is in place and it is clean and dry. Replace the gate protector if necessary. In the view window, Enter Strip alternated with Keton.

When sufficient blood is applied, the meter beeps, Sample good the test is automatically started. Refer to the Test Strips section in Chapter 1 for a detailed description of the test strip surface.
The meter is counted and then shows the test result. If an error is detected during the test, then the test shall be terminated and no result shall be displayed. From the start of the test, tests shall no longer respond to impressions until the test is complete.

The meter can be set to scan or manually enter a 1- or 2-digit comment code.

The meter may display results that are outside the range in the same way as glucose when the meter is activated for beta-ketone testing.
. If you do not request a comment code, proceed to step 5.
The meter can also be set up to allow the scanning or input of information in two free text fields.
The view window shows: Patient ID or Name (p: assignment) Comment code (number in box) Test type date and time (Keton) If 1 select, then the screen appears Scan Patient ID. Go back to step 1 and continue testing the next patient. If 2 If selected, the meter will show this patient's latest test result and you can also see previous results.

Guidance for testing with control solutions
Test procedure - glucose
Test procedure - β ketones

This chapter describes the steps for carrying out a control test with control solutions. Control tests will help to determine the operation of Precision Xceed Pro meter and test strips verified.
The quality control functions of the meter can be adapted to the requirements of each setting. The meter can be set as follows:
Tests with low, normal and high level control solutions or variations depending on the institution's policy.
Tests with control solutions at relative times (e.g. once every 24 hours) or fixed times • (e
Check test results are displayed as numerical value or OK/fault. •
These quality control functions may be set up via the data management system.
Guidance for testing with control solutions Take for optimal quality control results with the Precision Xceed Pro system
Check the operation of Precision Xceed Pro meter and test strips only with control solutions that are • approved for use with Precision Xced Pro and Precision PCx Plus Test Strips.
When opening a new bottle, write the date of disposal on the label. Each bottle with control solution remains stable • for 90 days after opening or until the expiry date printed on the label (the first one).
Do not use the control solutions after the expiry date printed on the bottles and carton. Do not use after the • disposal date, which is noted on the bottle after opening.
Before use, turn the control bottle several times to mix well. •
Turn the bottle with control solution over and tap it to remove air bubbles from the tip of the • bottle.
Wipe the bottle tip before and after each test with a clean gauze or cloth. •
Scan the bar code of foil packaging from the test strip used and not any other film packaging. If you do not • this may result in incorrect results.
Immediately after each use, place the correct cap back on the bottle and fasten the cap firmly. •
Where necessary in the framework of your institution's quality policy or on-site rules. •
To verify the operation of the Precision Xceed Pro system. •
Work schedule The meter expects control tests to be performed in the following order. If a control level is disabled, the meter expects this level to be skipped. If beta-ketone tests are disabled, the operator is not asked to perform this test.

Test procedure - glucose Use the following procedure for glucose control tests.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The Abbott logo screen will then appear in black for a few seconds and then grey to check that the view window is working properly. Next, the software version number appears briefly in view.
The meter starts in the Test mode. The Menu key will be used to insert the meter in the Menu Mode The Commission has already taken a decision on this matter.

The Operator ID can consist of 30 digits and therefore does not always fit in the viewing window. The meter can also be set to truncate (ignoring) first, last and/or selected digits of the bar code.
You can scan or enter the information of the barcode of the control solution manually. Manually: Enter the five numerical characters of the lottery number on the control-solution bottle. You do not have to enter the alpha character at the end of the five digit number. Next the screen appears Lot number strip.

If the screen Wrong opl. appears, you can:
The meter is programmed to expect the control tests in the following order: Low, Normal and High glucose, then Low, Normally and High ketones. A warning will appear if the checks are used in a different order, but you can continue with the test.



By scanning the bar code, the type of strip is determined, the meter calibrated, checked whether the expiry date has not expired, the lot strips used registered and verified if your setup approved it for use.
NB: Place the meter on a flat surface during control tests.
With clean, dry hands you can gently touch the test strip anywhere when removing the strip from the foil packaging to insert into the meter.




Before inserting the test strip, check that the gate protector is in place and it is clean and dry. Replace the gate protector if necessary. In the view window, Enter Strip alternated with Glucose.
When sufficient blood is applied, the meter beeps, Sample good the test is automatically started. Refer to the Test Strips section in Chapter 1 for a detailed description of the test strip surface.


NB: . .Take the meter to check the reading window after removal of the test strip. Do that sooner,
control solution can drop on the gate protector or in the strip port, causing damage to the meter. If the gate protector gets wet, it must be replaced by a new one.
/

The meter is counted and then shows the test result. If an error is detected during the test, then the test shall be terminated and no result shall be displayed. From the start of the test, tests shall no longer respond to impressions until the test is complete.

Check test results can be displayed numerically or as OK/Error depending on how the test is set. A .X.X is shown before numerical test results that are out of range.

If necessary, scan the comment code or enter it manually and press Enter.
The meter can be set to scan or manually enter a 1- or 2-digit comment code.
If you do not request a comment code, proceed to step 11.
The view window shows: Level (Layer, Norm., High) Comment. code (the number in the box) Test type (Glucose) Date and time of test Operator ID (o: command) Please note: the date and time of the test and operator ID are displayed alternately. If 1 select, then the screen appears Scan or Enter Control Lock. Return to step 4 Next Level perform the control test. If 2 select, then the screen appears Lot Number Strip Scan or Enter. Return to step 5 Repeat test to repeat this test at the same level.

NB: If beta-ketone testing is deactivated, then 1-Next level changes after a full set
glucose controls in 1-Quit. If beta ketone tests are activated and glucose quality control is performed first, then 1-Next level changes to 1-QC Keton after a complete set of glucose controls. If beta-ketone tests are activated and the β-keton quality control is performed first, then 1-Next level after a complete set of β- ketone checks changes to 1-QC Glucose.
99 mg/dL

Test procedure - β ketones Use the following procedure to perform β-ketone control tests. If beta-ketone tests are activated, the meter requires a blood β-keton test strip.
NB: The following section explains some representative institutions.
If the meter does not display these screens, please refer to chapter 10, Troubleshooting.
Before inserting the test strip, check that the gate protector is in place and it is clean and dry. Replace the gate protector if necessary. In the view window, Enter Strip alternated with Keton.
When sufficient blood is applied, the meter beeps, Sample good the test is automatically started. Refer to the Test Strips section in Chapter 1 for a detailed description of the test strip surface.
The meter is counted and then shows the test result. If an error is detected during the test, then the test shall be terminated and no result shall be displayed. From the start of the test, tests shall no longer respond to impressions until the test is complete.
Check test results can be displayed numerically or as OK/Error depending on how the test is set. A .X.X is shown before numerical test results that are out of range.

. If necessary, scan the comment code or enter it manually and press Enter. If you do not request a comment code, proceed to step 5.
The meter can be set to scan or manually enter a 1- or 2-digit comment code.
The view window shows: Level (Layer, Norm., High) Comment code (number in box) Test type (Keton) Date and time of test Operator ID (o: command) Please note: the date and time of the test and operator ID are displayed alternately. If 1 select, then the screen appears Scan or Enter Control Lock. Return to step 1 Next Level perform the control test. If 2 select, then the screen appears Lot Number Strip Scan or Enter. Repeat the test.


Patient Data Overview by Operator ID
Patient Data Summary by Patient ID
Data overview for all patient data
Data overview for control data
Data overview for skill data
Data overview for linearity data

The Precision Xceed Pro The results of up to 2500 patient tests, 1000 control tests, 20 skill tests and 2 linearity spans are stored in meters. This information may be called up in the following categories:
Patient by Operator ID (OperID): This will show the results of tests carried out by a specific operator, starting with the most recent test.
Patient by patient ID (PatID): This will show the test results of a specific patient, starting with the most recent test.
All patient data: This will show the test results of all patients, starting with the most recent test.
Control data: This will show the control test results, starting with the latest test.
Skill data: This will show the skill test results, starting with the latest test.
Linearity data: This will show the linearity test results, by level and in order.
NB: The meter can be configured to require a valid operator ID for results
bekijken.
Please perform the following steps to open the Data Overview menu:
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.

The Abbott logo screen will then appear in black for a few seconds and then grey to check that the view window is working properly. Next, the software version number appears briefly in view.


The Operator ID can consist of 30 digits and therefore does not always fit in the viewing window. The meter can also be set to truncate (ignoring) first, last and/or selected digits of the bar code.


Patient Data Overview by Operator ID Please refer to the beginning of this section Data overview for step-by-step instructions before opening the Data Overview menu.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The meter shows the result of your latest patient test.
The number that is on the right of 1 - Previous and 2 - Next shows up on the meter, indicates the number of tests that can be viewed. The view window also shows: Patient ID or Name (p: assignment) Comment code (number in box) Test type (Glucose or Keton) Date and time of test Operator ID (o: command) Please note: the date and time of the test and operator ID are displayed alternately.

105 mg/dL

Patient Data Summary by Patient ID Please refer to the beginning of this section Data overview for step-by-step instructions before opening the Data Overview menu.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The viewing window shows the test results for a specific patient.
If you enter the ID manually, it may be that you need to enter the id a second time to make sure the ID is correct.

The patient ID can consist of 30 digits and therefore does not always fit fully in the viewing window. The meter can also be set to truncate (ignoring) selected digits of the bar code.
The number that is on the right of 1 - Previous and 2 - Next shows up on the meter, indicates the number of tests that can be viewed. The view window also shows: Patient ID or Name (p: assignment) Comment code (number in box) Test type (Glucose or Keton) Date and time of test Operator ID (o: assignment)
Please note: the date and time of the test and operator ID are displayed alternately.
105 mg/dL

Data overview for all patient data Please refer to the beginning of this chapter for step-by-step instructions to open the Data Overview menu.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The view window shows all the patient test results, starting with the most recent result.
The number that is on the right of 1 - Previous and 2 - Next shows up on the meter, indicates the number of tests that can be viewed. The view window also shows: Patient ID or Name (p: assignment) Comment code (number in box) Test type (Glucose or Keton) Date and time of test Operator ID (o: assignment)
Please note: the date and time of the test and operator ID are displayed alternately.
105 mg/dL

Data overview for control data The meter stores the last 1,000 control test results. Please refer to the beginning of this chapter Data overview for step-by-step instructions before opening the Data Overview menu.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The meter shows the result of the most recent check.
The number that is on the right of 1 - Previous and 2 - Next shows up on the meter, indicates the number of tests that can be viewed. The reading window also shows the acceptable range for this test, plus: Comment code (number in box) Level (Layer, Normal, High) Range (if numerical) Test type (Glucose or Keton) Date and time of test Operator ID (o: assignment)

Please note: the date and time of the test and operator ID are displayed alternately.

Data overview for skill data The meter stores the last 20 skill results. Please refer to the beginning of this chapter Data overview for step-by-step instructions before opening the Data Overview menu.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The meter shows the latest skill results.
The number that is on the right of 1 - Previous and 2 - Next shows up on the meter, indicates the number of tests that can be viewed. The view window also shows: Sample ID (m: assignment) Comment code (number in box) Test type (Glucose or Keton) Date and time of test Operator ID (o: assignment)

Please note: the date and time of the test and operator ID are displayed alternately.
105 mg/dL

Data overview for linearity data The meter stores the last linearity panel with up to 4 repeats for 5 levels (20 tests). Please refer to the beginning of this section Data overview for step-by-step instructions before opening the Data Overview menu.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The meter shows linearity tests first per level (1, 2, 3, 4, 5) and then in order within each level. If no tests are stored for a level, that level is skipped. At the top of the viewing window are the level (1 to 5) and test repetition (1 to 4) shown.
The number that is on the right of 1 - Previous and 2 - Next shows up on the meter, indicates the number of tests that can be viewed. The view window also shows: Level (1 t/m 5), Testrepetition (1 to 4) Comment code (number in box) Test type (Glucose or Keton) Date and time of test Operator ID (o: assignment)

Please note: the date and time of the test and operator ID are displayed alternately.
45 mg/ dL

Skill test - glucose
Skill test - β ketones

Skill test - glucose Use the following procedures for performing a skill test. The Skill Test menu allows you to perform test tests with unknown samples if your setting's policy so requires. This functionality may or may not be available depending on how the meter is configured via the data management system. The skill test is based on comment codes in the patient tests.
specific settings of your settings some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The Abbott logo screen will then appear in black for a few seconds and then grey to check that the view window is working properly. Next, the software version number appears briefly in view.

The Operator ID can consist of 30 digits and therefore does not always fit in the viewing window. The meter can also be set to truncate (ignoring) first, last and/or selected digits of the bar code.
The meter indicates that you need to scan or enter the sample ID. VAAR means skill test.



By scanning the bar code, the type of strip is determined, the meter calibrated, checked whether the expiry date has not expired, the lot strips used registered and verified if your setup approved it for use.
With clean, dry hands you can gently touch the test strip anywhere when removing the strip from the foil packaging to insert into the meter.



Before inserting the test strip, check that the gate protector is in place and it is clean and dry. Replace the gate protector if necessary. In the view window, Enter Strip alternated with Glucose.

/

When sufficient blood is applied, the meter beeps, Sample good the test is automatically started. Refer to the Test Strips section in Chapter 1 for a detailed description of the test strip surface.

The meter is counted and then shows the test result. If an error is detected during the test, then the test shall be terminated and no result shall be displayed. From the start of the test, tests shall no longer respond to impressions until the test is complete.

The meter can be set to scan or manually enter a 1- or 2-digit comment code.

. If you do not request a comment code, proceed to step 12.
The view window shows: Sample ID (m: assignment) Comment code (number in box) Test type (Glucose) Date and time of test Operator ID (o: command) Please note: the date and time of the test and operator ID are displayed alternately. If 1 select, then the screen appears Scan or Enter Sample ID. Return to step 5 Next test to be carried out with another skill sample.

105 mg/dL · 105 mg /dL

Skill test - β ketones Use the following procedures for conducting a β-ketone skill test.
Before inserting the test strip, check that the gate protector is in place and it is clean and dry. Replace the gate protector if necessary. In the view window, Enter Strip alternated with Keton.
When sufficient blood is applied, the meter beeps, Sample good the test is automatically started. Refer to the Test Strips section in Chapter 1 for a detailed description of the test strip surface.
The meter is counted and then shows the test result. If an error is detected during the test, then the test shall be terminated and no result shall be displayed. From the start of the test, tests shall no longer respond to impressions until the test is complete.
The meter can be set to scan or manually enter a 1- or 2-digit comment code.

The view window shows: Sample ID (m: assignment) Comment code (number in box) Test type (Keton) Date and time of test Operator ID (o: command) Please note: the date and time of the test and operator ID are displayed alternately. If 1 select, then the screen appears Scan Sample ID or enter. Return to step 1 Next test to be carried out with another skill sample.



The Linearity Test menu allows you to verify the linearity of the method with a glucose and β ketone calibration verification set (KVC) from RNA Medical®. These sets contain tested materials for use in the confirmation of calibration and linearity of glucose and β ketone test at upper and lower limit of reporting range and three (3) points within range. KVC sets are designed for specific types of test strips. Refer to the package leaflet in KVC set for compatible types of test strips and detailed instructions.
Use the following linearity test procedure. This functionality may or may not be available depending on how the meter is configured via the data management system.
common institutions. Depending on the specific settings of your setting, some screens are displayed differently or not at all. If the meter does not display the screens shown, please refer to chapter 10, Troubleshooting.
The Linearity Test mode shall only be used to assess the linearity performance of the system.
The Abbott logo screen will then appear in black for a few seconds and then grey to check that the view window is working properly. Next, the software version number appears briefly in view.

The Operator ID can consist of 30 digits and therefore does not always fit in the viewing window. The meter can also be set to truncate (ignoring) first, last and/or selected digits of the bar code.
NB: . . When performing a linearity test, you must have strips with the same lottery number for the entire panel
gebruiken.

If beta-ketone tests are deactivated, this screen will not appear.

LIN means linearity.

If the input lot number of the KVC set is different from the previous one, then the meter will ask you to perform one of the following two actions:

Press 1 - Lot set again inv.
Press 2 - Replace Panel.
re-enter the existing lottery number or replace the existing batch number with a new one.
There can be up to 4 repeats per level. The number in brackets indicates the number of repeats already performed for this level. When a level is full, the number 4 appears in brackets, indicating that all four repetitions are ready for this level. For example 1 - Level 1(4).



By scanning the bar code, the type of strip is determined, the meter calibrated, checked whether the expiry date has not expired, the lot strips used registered and verified if your setup approved it for use.
With clean, dry hands you can gently touch the test strip anywhere when removing the strip from the foil packaging to insert into the meter.



Before inserting the test strip, check that the gate protector is in place and it is clean and dry. Replace the gate protector if necessary. The meter asks you to apply the sample on the test strip.
/

When sufficient blood is applied, the meter beeps, Sample good the test is automatically started. Refer to the Test Strips section in Chapter 1 for a detailed description of the test strip surface.

The meter is counted and then shows the test result. If an error is detected during the test, then the test shall be terminated and no result shall be displayed. From the start of the test, tests shall no longer respond to impressions until the test is complete.
The meter can be set to scan or manually enter a 1- or 2-digit comment code.

. If no comment code is requested, proceed to step 15.
The view window shows: Level (test levels 1 to 5) Test (repeat 1 to 4 for that level) Comment code (number in box) Date and time of test Operator ID (o: command) Please note: the date and time of the test and operator ID are displayed alternately. If 2 select, then the screen appears Lot Number Strip Scan or Enter. Return to step 8 to continue linearity tests on Equal Level. If 1 select, then the screen appears Select Level. Return to step 7 for linearity tests for a New Level.

45 mg/ dL · 45 mg / dL

Configuration of the meter
Settings Overview
Overview of test settings for all tests
Test settings summary per test type
Overview of patient testing settings
Overview of test settings for control tests
Overview of security test settings
System status

Configuration of the meter The Precision Xceed Pro meter settings are configured using the data management system. These settings can be seen on the portable device, but cannot be changed with it. Institutions that can be configured via the data management system:

Determine whether the operator can continue testing (allow or warn) or not (Block) when a strip lock is scanned or manually entered into the meter for a lot of strips that are not listed as comic strips. The maximum number of strip slots that can be downloaded to the Precision Xceed Pro meter is 36:18 glucose and 18 ketones.

NB: .The section explains some representative institutions.
If the meter does not display these screens, please refer to chapter 10, Troubleshooting.
The Abbott logo screen will then appear in black for a few seconds and then grey to check that the view window is working properly. Next, the software version number appears briefly in view.


NB: .The section explains some representative institutions.
If the meter does not display these screens, please refer to chapter 10, Troubleshooting.







NB: .The section explains some representative institutions.
If the meter does not display these screens, please refer to chapter 10, Troubleshooting.







NB: .The section explains some representative institutions.
If the meter does not display these screens, please refer to chapter 10, Troubleshooting.







NB: .The section explains some representative institutions.
If the meter does not display these screens, please refer to chapter 10, Troubleshooting.









NB: .The section explains some representative institutions.
If the meter does not display these screens, please refer to chapter 10, Troubleshooting.








NB: .The section explains some representative institutions.
If the meter does not display these screens, please refer to chapter 10, Troubleshooting.
The Abbott logo screen appears in black for a few seconds and then grey to check that the view window is working properly. Next, the software version number appears briefly in view.









The number that appears on your meter may be different from the number that is here.

Docking Station
Assembly instructions
Allow upload of data

After the test results have been collected for a prescribed period, they may be uploaded to the data management system. The data management system provides a simple and automated way to collect, report and transmit data.
Uploading data starts when the Precision Xceed Pro meter is placed in the docking station. The meter is switched on first if it was not yet enabled, after which the data will automatically be uploaded to the data management system.

The screen will appear during communication Data upload and turn the arrows to indicate that the system is working. The arrows can stop for a moment. When uploading data, the meter cannot be used for testing.
After uploading the meter gives Upload passed, Disable and then is eliminated.
NB: . .If the meter is deleted before the data transmission is complete, some data are possible
Not uploaded. This data will then be uploaded with the following data transmission.
If there is a problem while uploading data, an error message may appear on the meter. If the meter is removed from the docking station during uploading, then it shows: Last upload incomplete, turn meter into dock, Turn off. The meter will be shut down. Error messages are described in chapter 10, Troubleshooting. Additional information on how to solve data uploading problems is provided in the manual with the data management system.
After an error occurred when uploading data, the meter displays a warning every time it is switched on. The meter can still be used for tests by on 1 press before Continue. After the meter has been successfully docked, the warning disappears.


Docking Station The Precision Xceed Pro system provides an optional docking station for hand-free automatic data transmission (upload/download) between the Precision Xced Pro meter and a PC on which the data management application runs.

Place the docking station away from the patient. The docking station has a hinged base that allows the station to stand on a desk or be attached to a wall with the included mounting plate.
The Precision Xceed Pro meter is compatible with older Abbott docking stations with LED lights, which are lit using an optional power adapter. The following LED information applies only to older docking stations.
.The green LED indicates that the docking station is connected to an energy source.
.The yellow LED (continuously burning) indicates that the meter is well placed in the docking station.
The yellow LED (flashing) indicates that data is being sent (upload/download).
The power adapter is not required. If no power adapter is used, the data will be sent but the LEDs will not burn.
Another adapter can severely damage the docking station and Precision Xceed Pro meter in the dockstation.
Assembly instructions The docking station can be mounted vertically if required. The mounting is similar to the installation of a wall phone.

Instructions: 1. Drill holes in the wall that match the holes in a mounting plate.

Allow upload of data The docking station has two ports at the back: a port for power adapter (only older docking stations) and a data transfer port . The docking station data transfer port can be connected to the included serial cable in different ways for data transmission.
The docking station can be connected directly to the data management system via a network computer or terminal server. These connections are shown in images 1 and 2.

Figure 1 . . Upload data via a direct connection or network computer 1. Connect the serial cable to the docking station data port and computer serial port.

Image 2 . . Uploading data via a terminal server 1. Connect the power adapter of the terminal server to a power source.
NB: . .Devices which must be connected to an outlet and grounded shall be fitted with:
a grounded outlet.

Solving problems with patient test results
Resolving problems with control test results beyond range
•Solving problems with linearity test results that are out of range
Error Messages
Instructions for technical assistance
Return an instrument
Contact Abbott for service

Solving problems with patient test results This chapter describes circumstances in which false patient test results may arise. Refer to the package leaflet in the test strips for specifications and detailed instructions for use.
Resolving problems with control test results beyond range Repeat the test for that control solution and ensure that the operator follows this checklist:
If the above criteria are met but test results are still out of range, repeat the test with a new box of control solutions and/or test strips. If the results are still out of range, please contact your local office or distributor of Abbott Diabetes Care.

Solving problems with linearity test results that are out of range Repeat the test for linearity level and check that the operator meets all conditions in this checklist:
Error Messages Below you will find information about error messages that appear in the viewing window when Precision Xceed Pro meters detect errors.
Each notification shall contain an explanation and describe what needs to be done. In many cases, certain functions remain operational at least temporarily before the problem is solved. (For example, if the battery voltage is too low to perform tests, then viewing data can still be possible for a short time.)
If the problem persists, please note the displayed error message, including any 4-digit error code, and contact your local office or distributor of Abbott Diabetes Care.
Symptom/Black message Explanation What to do
The meter gets little or no power. Check that the batteries are properly fitted.
If the problem persists, then place new batteries.
See Chapter 11, Maintenance for further information.
The view window is empty.
Battery voltage's getting low. There is only a limited time left to test.
Place new batteries. See Chapter 11, Maintenance for further information.

Press 1 to turn off the meter.
This occurs when selecting a test when the temperature falls outside the range of the operating temperature of the meter.
Press 1 before Quit. Turn off the meter. Give the meter time to reach room temperature.


Symptom/Black message Explanation What to do
The meter is sent unconfigured from the factory. This screen will appear when you activate the meter for the first time after receiving it.
Use PrecisionWeb or other software provided by Abbott to configure the device according to your setting's specific requirements before it is put into service.

An error occurred during the last data transfer. This screen will appear when you turn on the meter.
Put the meter in the docking station to complete upload. After the meter has been successfully docked, the warning disappears.

Press 1 before Continue testing.
There may be a problem with the meter which makes it not working properly. A 4-digit error code can be displayed. These codes provide Abbott with further information on the problem.

The meter found a problem with the barcode scanner which prevented it from working. This is a temporary error that can be corrected by restarting the meter.

The test strip was inserted into the meter at the wrong time. The meter shall beep when a function is started before the strip is removed or if the test strip is left in the gauge when the test indicates that the strip must be removed.
Remove the test strip from the test strips.

Or · . .

Symptom/Black message Explanation What to do
The type of strip inserted into the meter does not correspond to the type of strips scanned.

If glucose quality control is to expire soon, the report . .Glucose QC over xx min. on the main screen, where: xx The countdown is to the gradient. If glucose quality control has expired, the report . .Glucose QC verl. now.

If the ketone quality check is due to expire, the notification .Keton QC over xx min. on the main screen, where: xx The countdown is to the gradient. If the ketone quality check has expired, the notification .Keton QC is gonna lose. now.

If the meter requires an upload to the data management system, .Upload Now Need.

NB: Before each test, check that the Precision Xceed Pro readout window is working properly. As the viewer
if you become ill-readable, stop using the meter and contact your local office or distributor of Abbott Diabetes Care.

Symptom/Black message Explanation What to do
The interval for one or more control tests has expired and the option to set what should happen when a glucose quality check interval has passed is on Warning or Blocking.
Perform the remaining glucose control tests indicated in the report.

(The menu item 2 - Continue only appears if this option is set to Warning. If the option is set to Block, each check screen display must be performed before patient tests begin.)
The Upload interval option is set to Warning or Block and the specified interval is exceeded. (The menu item 2 - Continue only appears if this option is set to Warning. If the option is set to Block, data should be uploaded before testing starts.)

Press 2 before Continue testing.
The interval for one or more control tests has expired and the option to set what should be done when a ketone quality check interval has passed is on Warning or Blocking.
Perform the remaining ketone control tests indicated in the notification.

(The menu item 2 - Continue only appears if this option is set to Warning. If the option is set to Block, each check screen display must be performed before patient tests begin.)
The system cannot confirm the patient ID.
Press 1 before ID re-enter.

Press 2 before Continue testing.
The scanned patient ID is the same size as the bar code of the test strip.

Or · Or

Symptom/Black message Explanation What to do
The bar code just scanned is not accepted. Possible causes are:

The bar code just scanned is part of a foil packaging test strips whose expiry date has expired.

The newly scanned or entered ID or lottery number is longer or shorter than the size determined during setting this meter.

The newly scanned or entered ID or lot number is not on the list of acceptable IDs or lot numbers for this meter. (The menu item 2 - Continue will only appear if this option is set to Warning.)

Press 2 before Continue testing.
The strip inserted is a glucose strip while the meter expected a ketone strip (or vice versa).
Press 1 and replace the strip with the expected test strip.

Press 2 before Quit.
Or · Or

Symptom/Black message Explanation What to do
The operator ID just entered is not on the list of accepted operator IDs for this meter.

(The menu item 2 - Continue will only appear if this option is set to Warning.)
Press 2 before Continue testing.
The operator ID entered is not on the list of accepted operator IDs for this meter.

The operator ID entered has expired.

(The menu items 1 - Re-enter ID and 2 - Continue only appear if this option is set to Warning.)
Press 2 before Continue testing.
The operator ID is almost expired. Testing is still allowed.

Or · Or

Symptom/Black message Explanation What to do
The test result is lower than what the system can measure.

The test result is higher than what the system can measure.

The test strip or the test strip port is wet, broken or dirty, or a wrong test strip has been inserted.

Press 2 before Quit and return to the Test Menu.
There may be a problem with the test strip. A 4-digit error code can be displayed. These codes provide Abbott staff with further information on the problem. Errors are:

4327 - The strip was removed during testing.
Repeat the test with a new test strip.
4330 - Blood glucose levels may be too high to be measured by the system or there may be a problem with the test strip.
Repeat the test with a new test strip. If the error occurs again, verify the result by performing a laboratory reference test.
20 mg/ dL · Or

When selecting a control test:
A lottery number of a control solution has been introduced for a different level check test than the meter expected (in the usual order from layer to high). The user may choose to enter another lot number (usually from the expected level) or run a level test corresponding to the input lot number.
Press 1 for Lot opn. enter.
Press 2 to perform the level test corresponding to the lottery number already entered.
When selecting a linearity test:
A new batch number of a linearity set has been introduced that does not correspond to the linearity panel stored in the meter at that time. Only one panel of data is stored. The set's lottery number may have been mis-entered, but the user can also choose to replace the previous data.
Press 1 for Lot Set again inv. to re-enter the lot number of another linearity set (usually the number of the previous set).

Press 2 before Replace Panel to use the new set's lottery number. The existing linearity data is replaced by new fate data.
When selecting Data Summary:
There are no stored test results for the type of data requested.
Press 1 before Quit and return to the Data Overview Menu.


The meter cannot retrieve a previous test result. This test result may not have been uploaded from the meter.
Press 1 to Previous The Commission will also examine the results of this study.
Press 2 to Next The Commission will also examine the results of this study.
Or · Or ·

Instructions for technical assistance The contact information of Abbott's technical service is on the following two pages. When you contact, an Abbott representative will handle the problem and/or instruct you to return the meter, test strips, control solution and/ or linearity set. The Precision Xceed Pro meter or any part of the system may not be returned for repair until an Abbott representative has authorized it.
To resolve the problem as efficiently as possible, please perform the steps below before contacting Abbott.


Return an instrument Requirements in various jurisdictions require that medical instruments be disinfected before being sent to minimise exposure risk for humans involved in handling or transporting the instrument. We ask you to perform the following steps before returning an instrument to Abbott:
If you want to drain a meter, please. contact your local office or distributor of Abbott Diabetes Care.

Contact Abbott for service Abbott will do everything in its power to help you if you have problems with the Precision Xceed Pro system. For technical assistance, please contact your local office or distributor from the list below:
AZIË Australia and New Zealand Abbott Australia Abbott Diabetes Care Victoria, Australia Tel: +1-800-801-478 New Zealand Tel:+0800-106-100 China Abbott Laboratories Ltd. Shanghai, China Tel: +800-820-3959 Philippines Abbott Laboratories Metro Manila, Philippines Tel: +632-6890-495 Hong Kong Abbott Laboratories Ltd. North Point, Hong Kong Tel: +852-2806-4488 Indonesia Abbott Indonesia Jakarta Selatan, Indonesia Tel: +62-21-52961529 Japan Abbott Japan Co. Ltd. Chiba-ken, Japan Tel: +0120378-055 Korea Abbott Korea Ltd. Seoul, Korea Tel: +080-014-5757 Malaysia Abbott Laboratories (M) Sdn. Bhd. Shah Alam, Selangor, Malaysia Tel: +00-603-5569-1919 Pakistan Abbott Laboratories (Pakistan) Ltd Karachi, Pakistan Tel: +92-21-504-5680 Singapore Abbott Laboratories (S) Pte Ltd Singapore Tel: +65-6272-2881
EUROPE Belgium and Luxembourg Abbott Diabetes Care Louvain-la-Neuve, Belgium Tel: +0800 167-72 Luxembourg Tel:+8002 54 87 Denmark Abbott Laboratories A/ S Abbott Diabetes Care Gentofte, Denmark Tel: +45-3977-0190 Germany Abbott GmbH & CO. KG Abbott Diabetes Care Wiesbaden, Germany Tel: +0800-519-9519 Finland Abbott Oy Abbott Diabetes Care Espoo, Finland Tel: +0800-555-500 France Abbott France Abbott Diabetes Care Rungis Cedex, France Tel: +0800-10-11-56 Greece Abbott Diabetes Care Athens, Greece Athens Tel: +30-210-998-5220 Thessaloniki Tel: + 30-301-047-2947 Ireland Abbott Diabetes Care Dublin, Ireland Tel: +1-800-776-633 Italy Abbott Diabetes Care Italia Rome, Italy Tel: +800-334-216 Netherlands Abbott B. V. Abbott Diabetes Care Amersfoort, Nederland Tel: +0800-022-8828 Norway Abbott Norge as Abbott Diabetes Care Fomebu, Norway Tel: +800-87-100
Austria Abbott Ges.m.b.H Abbott Diabetes Care Vienna, Austria Tel: +0800-93093 Poland Abbott Laboratories Poland Sp. z o.o. Abbott Diabetes Care Warsaw, Poland Tel: +0-801-379-799 Portugal Abbott Laboratórios, Lda- Abbott Diabetes Care Amadora, Portugal Tel: +800-200-891 Spain Abbott Cientifica S. A. Madrid, Spain Tel: +900-300-119 Czech Republic Abbott Laboratories s.r.o Abbott Diabetes Care Prague, Czech Republic Tel: +800-189-564 United Kingdom Abbott Diabetes Care Maidenhead, Berkshire, UK Tel: +0500-467-466 Sweden Abbott Scandinavia AB Abbott Diabetes Care Solna, Sweden Tel: +020-190-11-11
Switzerland ABBOTT AG Abbott Diabetes Care Baar, Switzerland Tel: +041-768-44-47
Taiwan Abbott Laboratories LTD Taipei, Taiwan Tel: +0800-521-125 Thailand Abbott Laboratories Ltd Bangkok, Thailand Tel: +66-2-252-2004

LATIN AMERICA Argentina Abbott Laboratorios de Argentina SA Buenos Aires, Argentina Tel: +0800-800-6334 Brazil Abbott Laboratorios do Brasil Ltda São Paulo, Brazil Tel: +0800-703-0128 Chile Abbott Laboratorios de Chile Ltda. Santiago, Chile Tel: +800-80-2226
MIDDLE EAST AND AFRIKA Israel ILEX Medical Ltd. Rosh-Ha'ayin, Israel Tel: +972-3-938-5501 Saudi Arabia Abbott Diabetes Care Mediserv Riyadh, Saudi Arabia Tel: +800-0124-1238 Turkey Abbott Diabetes Care Goztepe-Istanbul, Turkey Tel: +0500-467-466 South Africa ADC South Africa Abbott Laboratories Gauteng, South Africa Tel: +27-11-858-2000
NORTH AMERICA Canada Abbott Diabetes Care Mississauga, Ontario, Canada Tel: +1-800-387-8378 (English) Tel:+1-800-465-2675 (French) Puerto Rico Abbott Laboratories Abbott Diabetes Care San Juan, Puerto Rico, USA Tel: +1-787-750-5454 VS Abbott Diabetes Care Alameda, CA, USA Tel: +1-877-529-7185
Colombia Abbott Laboratorios de Colombiá SA Bogota, Colombia Tel: +01-8000-522268 Dominican Republic Abbott Laboratorios Republica Dominicana Santo Domingo, Dominican Republic Tel: +809-566-7181 (machine 263) Guatemala Abbott Laboratorios Guatemala Cotio, Guatemala Tel: 597-4109 Mexico Abbott Laboratorios Mexico Mexico City, Mexico Tel: +01-800-711-2208 Peru Abbott Laboratorios Peru Lima, Peru Tel: +0-800-1-1215 Uruguay Abbott Laboratorios Uruguay SA Montevideo, Uruguay Tel: +000-4054-342 Venezuela Abbott Laboratories C. A. Caracas, Venezuela Tel: +0-800-Glucosa

Clean the outside
Replace the port protector
Battery life
Replace the batteries

The Precision Xceed Pro meter requires little routine maintenance. During testing, samples remain outside the meter and thus significantly reduces the risk of contamination. Follow your system maintenance and inspection facility procedures.
This chapter describes the maintenance of the meter.
Clean the outside It is recommended to clean the exterior of the Precision Xceed Pro meter daily. Keep to your institution's policies and procedures for infection prevention, which may require more frequent cleansing.
The meter only needs to be cleaned with a damp cloth or sponge and a mild cleaning agent. Acceptable cleaning solutions include alcohol-based and ammonia-based cleaners. Recommended solutions include Sani- Cloth® HB, Sani Cloth Plus and Super Sani Club®.
Bleaching or hydrogen peroxide-based cleaning agents make the meter's keyboard pale. Use of wipes with Hype-Wipe®, Dispatch® and Virox® 5 is not recommended because it pales the meter's keypad. Non-mentioned cleaning agents have not been tested and may damage the meter. Contact your local office or distributor of Abbott Diabetes Care for the latest information about cleaning solutions.
Turn off the meter before cleaning. Do not apply a cleaning solution directly to the meter. When using pre-packaged wipes, squeeze the excess moisture from the swabs before they are used. The meter must not be immersed in or rinsed with any liquid or sterilised into an autoclave. These steps help prevent liquid from entering the meter and it gets damaged.
damaged.

Replace the port protector The gate protector is designed to minimise the chance of moisture entering the meter via the test strip port. If the gate protector comes into contact with blood or control solution, then the meter should be cleaned and dried and the port protector replaced. Ensure that the area around the test strip port is well dry before applying the new gate protector.
Battery life The average life of alkaline batteries is approximately 60 days with an average of 9 tests per day and when the meter is switched off after 2 minutes without activity (depending on bar code usage). The battery life may vary depending on actual use.
Replace the batteries Insert new batteries in the Precision Xceed Pro meter as follows. The Precision Xceed Pro meter requires two (2) standard AA alkaline batteries or rechargeable nickel metal hydride (NiHM) batteries (not supplied by Abbott Diabetes Care).


Precision Xceed Pro meter
Symbols
Type of bar codes
Important safety instructions

Precision Xceed Pro meter
Food source: Two standard AA alkaline batteries or rechargeable nickel metal hydride (NiHM) batteries (not supplied by Abbot Diabetes Care).
Battery performance depends on the meter's operating frequency and time of testing.
Rechargeable NiMH batteries (2) cells 1,2 volts, 1,1 Ah minimum capacity (not provided by Abbott Diabetes Care)
Environment specifications: This product is for indoor use only.
Operating temperature meter: 15 °C to 40 °C Storage temperature meters: -20 °C up to 50 °C Height: up to 2195 meters Humidity: from 10% to 90% non-condensing power fluctuations: ±10% Temporary span: ±1,5 KV nominal Pollution category: 2

Symbols The Precision Xceed Pro meter, test strips and docking station are provided with special symbols. Below you will find the full list of symbols, with explanation. Please note that some of the symbols below may not appear on your Precision Xceed Pro system due to differences in local regulations.
Indicates glucose test (black text on white).
Indicates a β-ketone test (β-hydroxybutyrate) (white text on purple).
Refers to equipment type BF according to EN60601-1.
Refers to type B equipment according to EN60601-1.
The double arrow indicates the connector to be used for data transmission. The double arrow can be on the front panel of the docking station, underneath the LED screen. (See Upload Data, chapter 9.)
Medical equipment only in respect of electrical shocks, fire and mechanical hazards according to UL 60601-1 and CAN/CSA 22.2 NO. 601.1.
Laser hazard. This product meets 21 CFR 1040.10 and IEC 60825-1:2001, CDRH Class II/IEC Class 2 laser product.
Careful: Use of control buttons, modifications or the conduct of procedures other than those mentioned in this document may lead to dangerous exposure to laser light.
This label is permanently at the back of the
Class 2 laser scanners use a low voltage diode for visible light. As with other sources of bright light, such as the sun, users should not look directly into the beam of light. Short-term exposure to a class 2 laser is not harmful as far as known.
Meter confirmed.
This symbol indicates that this product provides the electrical protection as of a Class II device as described in IEC 601-1 Part 1.
This way upstairs.
Indicates temperature limits for storage.
This product should not be disposed of with ordinary household waste. In the European Union, electrical and electronic waste must be collected separately in accordance with Directive 2002/96/EC. Please contact the manufacturer for details.
Lot number.
Deadline. To be used before YYYY-MM-DD or YYYY-mm.
Note: Provides the operator with useful additional information or background information, or refers to it.
Do not re-use.
Order number/catalogue number.

Delegated representative.
Don't drink it.
CE is the marking indicating that an article is authorised for distribution in the EU.
Manufacturer under European Directive 98/79/EC on medical devices for the use of human beings
In vitro diagnostics - Abbott Diabetes Care.
Refer to the instructions for use.
Recyclable.
For use with In vitro diagnostics.
Sterilisation method for lancets only. Sterilized by irradiation (lancets only).
Codabar is a self-checking, numerical bar code only. Codabar allows the numbers 0 to 9, six symbols (- . $/+) and the start/stop characters A, B, C, D, E, *, N or T are encoded. Codabar is used in libraries, blood banks, parcel delivery companies and a range of other information processing applications.
A type of bar code that provides excellent density for numerical data and good density for alphanumeric data. The Code 128 standard is supported by AIM (Automatic Identification Manufacturers).
A type of bar code that is widely used in many industries and the standard is for many barcode specifications from government agencies, including the US Department of Defense. Code 39 is defined in standard MH10.8M-1983 of the American National Standards Institute (ANSI).
A type of bar code that encodes exactly the same characters as Code 39 but uses 9 bar code elements per character instead of 15.
Interleaved 2 or 5 (I 2 or 5). The data must be equal in number.
EAN: EAN is the abbreviation of European Article Numbering. EAN-13 is used worldwide for marking retail goods. EAN-13 encodes 13 characters: the first two characters are a country code, then follow 10 data numbers and one control number. Two-digit and five-digit additional bar codes may be added for a total of 14 or 17 data digits. EAN-8 is a short version of EAN-13. EAN-8 consists of a two digit country code, five data numbers and one control number. A two-digit and five-digit bar code may be added.

RSS: RSS is the abbreviation of Reduced Symbolology Set (set with reduced symbology), which produces very small bar codes suitable for labeling electronic components and healthcare devices. There are seven different types of RSS indent codes. RSS can encode up to 74 characters.
displayed as spaces. Some bar codes may contain control characters and non-alphanumeric characters. In the Type bar code section of the MeterCom screen, you can set all options for entering barcode data for operator ID and patient ID. The selected types of bar codes determine which codes can be scanned in the meter. For detailed information, see the Meter Component manual.
Signal output and input components The following is recommended by UL for the .u. intended signal input and signal output connections that meet this requirement. (UL 60601-1 . . 1st edition)
Use of TOE HORSE that does not meet the safety requirements set for this device can make the overall system less safe. When selecting accessories, consideration should be given to the following:
whether the accessories are used in the IMMEDIATE CIRCUMSTANCE OF THE PATIENT •
there is evidence that the safety certification of the AUTHORIZED person has been carried out in accordance with the applicable IEC 601-1 • and/or IEC 6001-1-1 harmonised national standard
Immediate patient environment (UL 60601-1 . . 1st edition) In areas where patients are usually cared for, the immediate environment of the patient is the area with surfaces that may touch the patient or an presence which may touch it. This is a room which, with the bed (study table, dental chair, treatment room and similar), extends horizontally 1.83 m around the circumference of the bell and vertically to 2.29 m above the floor.
Other information The device is approved for use in an oxygen-rich environment. The device is approved for continuous use.
This device has been evaluated for compliance with IEC 606011-1-2. The operation of this device may disturb or be disturbed by other devices in the environment. If observed, the distance between devices should be increased.

Compliance with standards The Precision Xceed Pro system has been evaluated for compliance with the following standards:
UL 60601-1, 1st ed. Medical electrical equipment . .
CSA C22.2 NO. 601.1-M90, Medical electrical equipment .
EN 60601-1, 2nd ed. Medical electrical equipment . .
CSA 22.2 No. 61010.1-4
EN 60601-1 . . Medical electrical equipment .
IEC 61010-2-101 (2002-01) . Safety requirements for electrical equipment for measuring, checking and laboratory use Part 2-101: specific requirements for in vitro diagnostics (IVD) medical equipment 1st edition
EN 60601-1-2 .2001 class A . . Medical electrical equipment .
IEC 60601-1 . . Medical electrical equipment .
IEC 60601-1-1, 2nd ed. Medical electrical equipment . . Part 1-1: General safety requirements .
IEC 606011-1-2 (1993-04) . . Medical electrical equipment . Additional standard: electromagnetic compatibility . . requirements and tests first edition; (CENELEC EN 60601-1-2: 1993)
Precision Xceed Pro meter . . Class II (fittings supplied by an internal electric power source).
Docking station . . Class II (fittings supplied by an internal electric power source).
What the fault codes shown on the display mean and how to clear them.
| Code | Meaning | Fix | |
|---|---|---|---|
| EN2 | Continue to appear only if these | p.71 |