Use only outside the body (In vitro Diagnostic use) for self-control or professional use in the treatment of diabetes.
Potential infection risk: Diabetes treatment with this system of blood tests in several patients must always wear gloves and comply with the policies and procedures applicable to infection prevention in their institution.
Use to measure blood glucose in samples taken from the forearm, upper arm, thumb or fingers.
Read the instructions in this user manual. If you do not follow the instructions, the results are incorrect. Practice the test procedures before using the meter. Keep the advice of your diabetes doctor when you test blood glucose levels and blood beta ketone levels.

IMPORTANT: Use only FreeStyle Precision- blood glucose test strips and FreeStyle Precision-blood β-ketone test strips. Other test strips may produce inaccurate results. How your FreeStyle Precision blood glucose and ketone measurement system works When you insert a test strip into your meter, the notification Applying Monster appears in the reading window of your meter. When a blood sample or control solution sample is applied to the test strip, glucose or β ketones react with the chemicals on the test strips. This reaction generates a weak electrical current that is measured. The result is shown in the display window of your meter.
Be careful using the meter near children. Small components may present a risk of suffocation



Press and hold this button until plus appear. This indicates that the pager is switched on.
Press or press to change the pager setting. Press to save the pager setting and continue to set time.

Date Display
The hour is blinking. Press or press until the right hour appears. Press to save the hour and continue on to set up minutes. The minutes blink. Press or in until the right minutes appear. Press to save the minutes and continue to set the date.
The date display is blinking. Press or to change the date view. Press to save the date view and continue to set up the unit of measurement.
Bye.Month

The year's blinking. Press or in until the right year appears. Press to save the year and continue on setting up the month. The month's blinking. Press or press until the right month appears. Press to save the month and continue on setting up the day. The day is blinking. Press or press until the right day appears. Press to save the day and continue to set up time view. Month . .
The unit of measurement set up by the factory appears. You cannot change this setting. Please contact Customer Service for more information. Press and hold the button to turn off the meter.
NB: When using the 12-hour display AM and PM appear. Time display is blinking. Press or press to change the time view. Press to save the time view and continue to set the date view.
Month . . day · year · 24-hour view


This screen always appears when the meter is switched on. Each time you turn on the meter, check that your gauge matches exactly the example shown here.
Do not use the meter if the display doesn't look exactly like the example.
If the screen is not exactly the same, there is a risk that the meter will display an incorrect result. Contact Customer Service.

A138
2012/03

Perform a blood glucose test
The collection of a blood sample from the alternative places . . forearm, upper arm and mouse of thumb . Contact your diabetes treatment before starting testing at alternative places.
Mouse of the thumb

Use none blood samples from alternative sites where:
Upper arm · Forearm

Keep the blood drop on the white surface until the meter beeps (if sound is on) and the status bar appears in the display window. This indicates that the test strip has taken enough blood.
Then the symbols for Monster appear and, indicating that the meter is ready and you can apply a sample to the test strip.
Take a blood sample Pick a test site. Use the puncture device to take a blood sample. (See instructions for use with the needle device for further information.)
The countdown starts in the reading window while your meter checks your glucose level. NB: During the countdown, test strip may be not are removed or disturbed.
Apply blood to the test strip Apply the blood drop to the white area at the end of the test strip. The blood is pulling into the test strip.
View the result The meter beeps (if the sound is on) when the result appears in the viewing window. The test is complete (see example). The result is stored in memory or can be recorded by you in a log. Turn off the meter Use the opened foil pack to remove used test strip. This will disable the meter. Dispose of the used test strip correctly. NB: You can also hold down for at least 2 seconds to turn off the meter. The meter is also switched off after 60 seconds of inactivity.


Meter:
The meter shows results in mmol/L. The unit of measurement shall be pre-set. You cannot change this setting.

General:

Meter:
Reading window What it means What to do
Severely high blood sugar or there may be a problem with the test strip
Repeat the test with a new test strip. If the result is HI (High), take immediate contact your diabetes treatment doctor. Check the blood beta ketones if monitoring of ketones is part of your diabetes treatment program.
A17

Use none expired test strips; these may produce inaccurate results. The expiry date is the last day of the month if only the month and year are displayed. Do none urine on the test strip. Use none damp, bent, scratched or damaged test strips. Use the Test strip not if there is a puncture or crack in the foil packaging.

NB: If you see the error messages E-3 or E-4, please refer to the Error Messages chapter in this user manual.
NB: See the operating range test strips instructions and more information on restrictions of use.
Check your ketones:
2012/03


NB: Don't squeeze the puncture.
For testing for blood β ketones, use only blood samples from the finger tip.
Take the foil packaging of the test strip at the notch, tear open the package and remove the test strips.
Abbott Diabetes Care Ltd.
Witney, Oxon, OX29 0YL
Made in UK.
Abbott Diabetes Care Ltd.
Witney, Oxon, OX29 0YL
Made in UK.
Apply the test strip Insert the test strip with the contact lines (3 black lines) upwards into the meter until the strip does not go further. This activates the meter. NB: The meter is switched off after 3 minutes of inactivity. Remove the unused test strip and insert it back into the meter to re-set the gauge.

Apply blood to the test strip Apply the blood drop to the white area at the end of the test strip. The blood is pulling into the test strip.
View the result The meter beeps (as long as the sound is on) when the result appears and in the view window. The test is complete (see example). The result is stored in memory or can be recorded by you in a log. Turn off the meter Use the opened foil pack to remove used test strip. This will disable the meter. Dispose of the used test strip correctly. NB: You can also hold down for at least 2 seconds to turn off the meter. The meter is also switched off after 60 seconds of inactivity.

Keep the blood drop on the white surface until the meter beeps (if sound is on) and the status bar appears in the display window. This indicates that the test strip has taken enough blood.
The countdown starts in the reading window while the meter checks your blood β ketone level. NB: During the countdown, test strip may be not are removed or disturbed.

Meter:
Reading window What it means What to do
Contact your diabetic treatment provider for disease management instructions.
The meter shows results in mmol/L. The unit of measurement shall be pre-set. You cannot change this setting.
There may be a problem where medical attention is needed

Take immediate contact your diabetes treatment doctor.
Blood beta ketones should be less than 0.6 mmol/L.2 High blood β ketones may be caused by disease, sobriety, strenuous exercise or poorly controlled blood glucose.1-3
if you are at risk of diabetic ketoacidosis (DKA) 2-6
Repeat the blood β ketone test with a new blood beta ketone test strip when:
Very high blood β ketones or there may be a problem with the test strip
Repeat the test with a new test strip. If the result is HI (High), take immediate contact your diabetes treatment doctor.
High blood β ketones

MediSense control solutions for glucose and ketones


Use none control solution with expiry date. Add none water or other liquid to control solution. Control solution may be used not Swallow or inject. Use control solution not like eye drops.
Control solutions may be used with a glucose or β ketone test strip. Use control solution:

Contact Customer Service for information on how to obtain control solution.


The symbols for Monster apply and appear as well.
Abbott Diabetes Care Ltd.
Witney, Oxon, OX29 0YL
Take the foil packaging of the test strip at the notch, tear open the package and remove the test strips.
NB: and appear when a test with control solution for β ketones is performed.
Made in UK.
Abbott Diabetes Care Ltd.
Witney, Oxon, OX29 0YL
Made in UK.
Press to mark the test as a control solution test. The show. The meter is now ready and you can apply control solution to the test strip.
Apply the test strip Insert the test strip with the contact lines (3 black lines) upwards into the meter until the strip does not go further. This activates the meter.

NB: The meter is switched off after 3 minutes of inactivity. Remove the unused test strip and insert it back into the meter to re-set the gauge.
Apply control solution to the test strip Shake the bottle with control solution to mix the solution. Apply a drop of control solution at the indicated location in the white box at the end of the test strip. The control solution shall be drawn into the test strip.
This screen always appears when the meter is switched on.
Do not use the meter if the display doesn't look exactly like the example. Contact Customer Service.
See .Your blood glucose tests . for more information.
Keep the drop of control solution on the white area until the meter beeps (as long as sound is on) and the status bar appears in the display window. This indicates that the test strip has absorbed enough control solution.
Then the time and date appear (if set).

The countdown starts in the reading window while the meter checks glucose or β ketone levels.
Compare the control solution result with the range in instructions for use on blood glucose or beta-ketone test strips. The result should be within the range. Causes of results that are out of reach include:
NB: During the countdown, test strip may be not are removed or disturbed.
If the countdown does not start, remove the used test strip, discard it, turn off the meter and try again with a new strip.
Repeat the test with a new test strip. Perform the test instructions carefully. Contact Customer Service if the result is still outside the printout range.
Turn off the meter Use the opened foil pack to remove used test strip. This will disable the meter. Dispose of the used test strip correctly. NB: You can also hold down for at least 2 seconds to turn off the meter. The meter is also switched off after 60 seconds of inactivity.
View the result The meter beeps (if the sound is on) when the result appears in the viewing window. The test is complete (see examples shown). The result is stored in memory as a control solution result or can be recorded by you in a logbook as a check result.
NB: appears with the result when a test is performed using control solution for β ketones.

Your most recent result will then appear, with the time and date (if set) on which it was obtained. The following symbols indicate the type of result:
The meter's memory stores up to 450 tests for enoate control solution, blood glucose and blood beta ketone results, and other meters information. The most recent event appears first.
Start with disabled meter (no test strip inserted).
Press to turn the meter on.
Press to view the previous result. Repeat this for any result you want to view.
This screen always appears when the meter is switched on.
NB: You can exit the meter's memory at any time by pressing and holding it down.
Do not use the meter if the display doesn't look exactly like the example. Contact Customer Service.
See .Your blood glucose tests . for more information.
Appears in the viewing window when there are no previous results to view.
Press to return to a previously viewed result. Repeat this for any result you want to view.
NB: You can exit the meter's memory at any time by pressing and holding it down.

Press in memory. Your average will appear in 7 days.

NB: n = the number of results included in your average over 7, 14 or 30 days.
Press to view the average in 14 days.
Press to view the average in 30 days.
Press to return to the average in 7 days, or
Press to return to the memory screen.
Hold down to disable the meter. The meter is switched off after 30 seconds of inactivity.

Notification What it means What to do
Transfer of test results to a computer requires a compatible data management system. For more information, please visit our website or contact Customer Service.
Blood glucose levels may be so low that the system cannot measure them or there may be a problem with the test strip
Notification What it means What to do
The meter does not work properly because the temperature is too high or too low
Blood glucose levels may be so high that the system cannot measure them or there may be a problem with the test strip
Blood was applied too early to the test strip
Meter error 1. Take out the meter. 2. Repeat the previous test steps. 3. Contact Customer Service if the error appears again.

Notification What it means What to do
Meter error 1. Check that you are using the right strip for this meter. (See the instructions for use on the test strip to check that your strip is compatible with this meter.) 2. Repeat the test with a test strip for use with your meter. 3. Contact Customer Service if the error appears again.
What it means to do
The meter shall not enter the test mode after inserting a test strip.
The test strip is not inserted correctly or in the gauge
Encoding is not necessary or the test strip may be damaged or used, or the meter does not recognise the test strips
No battery installed The battery is misinstalled
Meter error 1. Take out the meter. 2. Repeat the previous test steps. 3. Contact Customer Service if the error appears again.
The battery is dead 1. Replace the battery. Reset the date and time if necessary.
or

What it means to do
What it means to do
The test shall not start after the blood sample has been applied.
The blood sample is too small 1. See instructions for using the test strips for re-application. 2. If the countdown still does not start, remove the used strip and throw it away properly. 3. Repeat the test with a new test strip. 4. If the test is still not started, please contact Customer Service.
The test shall not start after the blood sample has been applied.
The meter or test strip is defective
The sample is applied after the meter has been switched off

If the meter does not have plastic tabs, push up and remove the battery.
The meter is supplied with an installed CR 2032 lithium battery (button cell). This battery supplies power for about 1000 tests.
When it appears, you can still use the meter and its results are accurate. The display window's screen lights are not working. Replace the battery immediately.
NB: Run used batteries properly.
When only appears, the meter is not usable. The meter will automatically be switched off. Replace the battery immediately.

Keep the meter
Avoid dirt, dust, blood, control solution or liquid entering the meter's test strip port. Clean the outside of the meter with a damp cloth and mild soap. Diabetes treatment: Acceptable cleaning solutions include:
Keep the meter in the supplied case.
Width Upper 5,33 cm

Try not to clean the test strip port. Pour none liquid in the test strip port or on buttons. Immerse the meter not under water or other liquids.
Weight 40 to 46 grams
Food source 1 CR 2032 lithium battery (button cell)
Battery life Approximately 1000 tests
Up to 450 tests, including control-solution results, blood glucose results and blood beta ketones results, and other meter information.
Meters -25 °C to 55 °C (-13 °F to 131 °F)


Display results disabled. Stays on while results are transferred to a computer. Data port Yes
Temperature meter 10 °C to 50 °C (50 °F to 122 °F)
Temperature system See the working range of the test strip you use.
Clinical tests have shown that the results are not affected in testing at altitudes up to 2195 metres above sea level.