Distributed by/Gedistributed door:
Abbott S.r.l. Abbott Diabetes Care
Abbott B.V. Abbott Diabetes Care
Wegalaan 9 2132 JD Hoofddorp Nederland 0800 022 88 28
00144 Roma, Italia Per assistenza tecnicoaddestrativa
contattare 800 33 42 16
Abbott S.A. / N.V. Abbott Diabetes Care
Blood glucose monitoring system Bloedglucosemeetsysteem
Abbott AG Abbott Diabetes Care

6341 Baar Schweiz / Switzerland / Svizzera
Belgium Belgium 0800 167 72 Luxembourg 8002 54 87


08 00 / 804 404
Abbott GmbH & Co. KG
Abbott Gesellschaft m.b.H.
65205 Wiesbaden Deutschland 08 00 / 519 95 19
1230 Wien Österreich 08 00 / 93 00 93
This product(s) and/or its use are protected by one or more of the following patents: US6,071,391; US6,120,676; US 6,143,164; US6/156,173; US629,757; US6.338,790; US6503,381; USD6,591,125; US6.616,819; US66,6834; US6.0076,816; United States6,679,816; US7,749,740; US6.8893,545; US6.942,518; US7.058,437; US7.615,637; U.S.713,40,406; USD7,740,581; US611,854S; USD615,890S; EP1119637; EP1145000; EP0958495, CA2346415; CA2358993; CA2351796; CA2423,837; and CA121137. Additional patents may be issued and/or hanging.
FreeStyle, FreeStyle Lite and FreeSthle Freedom Lite are trademarks of the Abbott Group of Companies in various jurisdictions.
Abbott Diabetes Care Ltd. Range Road Witney, Oxon OX29 0YL, UK Abbott Diabetes Care Inc. 1360 South Loop Road Alameda, CA 94502 USA
©2007, 2009 Abbott ART20055 Rev. B 11/10
ART20055 Rev-B-Cvr.indd 1-2 11/19/10 12:06 PM

Contents Important information about your blood sugar monitoring system FreeStyle Freedom Lite 2 Health-related important information 4 FreeStyle Freedom Lite and FreeStyl Lite Dose Electrode FreeStyle Freedom Lite player and its 5L operation screen of the FreeStyl Freedom Lte 7 L Serbo Test Electrode FreeStile Lite 9 FreeStyle Control Solution FreeStyle 10 Control Solution Player Configuration Set up your player 12 Set-up of reminder alarms 12 Set the sound, time and format of the hour 14 Set date and format 15 The determination using the FreeStyle control solution A test solution is used to measure the dose 16 Determination of your blood sugar Determination of your blood sugar 18 Measurement of blood glucose 19 LO (low) and HI (high) 22 Reader results Reader's memory, 7, 14 and 30 days average 23 Test results (control solution, blood glucose) 24 Download data 25 Reader maintenance Battery 26 Battery Replacement 26 Player Maintenance 27 System troubleshooting Error Messages 28 Troubleshooting 31 System Specifications FreeStyle Freedom Lite 32
See directions for use Temperature limits Manufacturer CE Marking
Lot code Use before Do not reuse Diagnostic medical device in vitro Reference Number Recycling
Term of office established in the European Community

The FreeStyle Freedom Lite system uses a very small sample of 0.3 μl to test at different locations
Important information about your FreeStyle Freedom Lite blood sugar monitoring system
How to use the FreeStyle Freedom Lite system • FreeStyle Freedom Lite is for use outside the body (in other words, for diagnostic purposes). in vitro Only).
Suggested test areas for hand
The above human figure illustrates the locations where you can test with FreeStyle Freedom Lite. Avoid the seeds of beauty, veins, bones and tendons.

Important information about your FreeStyle Freedom Lite blood sugar monitoring system (continued)
Warning: Please read all instructions provided in this section. user manual and practice testing procedures before using the FreeStyle Freedom Lite system. Blood glucose monitoring should be performed according to the recommendations of a health care professional.

Data transfer port
Download test results. Requires a compatible data management system. For more information, please visit our website or call customer service.
System Check Screen This screen briefly displays at each drive-on, to ensure the correct operation of the display. Do not use the drive if system check screen does not match the example given here. Please call customer service.


Displays your test results and other important information.
Port of insertion of the dosing electrode
Insert side Higher d-a new FreeStyle Lite dosing electrode here. The drive is turned on when you insert the dosing electrode.
Deposition zones (dark half circles on the dosing electrode)
FreeStyle Lite dosing electrode
Apply blood or control solution to only one deposit area. Insert the electrode into the reader, with the printed side pointing upwards (see picture on right).


Appears when you need to replace the stack of your drive.
Symbols of a drop in blood and the dosage electrode
the control solution Indicates you have scored a result
indicate when the reader is ready for application of blood or solution
FreeStyle Player Unit of Measure configuration
Shows the date, time and number of tests.
Indicates the mean blood glucose score of 7, 14 and
12 hour format (e.g. 4:00 p.m.).

FreeStyle Lite dosing electrode Apply blood or control solution to one deposit area of the dosing electrode by test.
Deposit areas of the electrode
FreeStyle Control Solution Each vial of control solution contains a red liquid containing a specified amount of glucose used to:
A dosing should be performed using the control solution in case of doubt as to the results obtained and if you want to confirm that the metering reader and electrodes are working properly. For information on how to get control solutions, call customer service.
dosage

Information on the control solution The FreeStyle control solution is retained for three months after the bottle has been opened or until expiry date printed on the label, whichever is earlier. Important: Count three months after the date of opening a new vial of control solution. Enter this date on the side of the control solution vial. This is your scrap date, unless the expiry date printed on the vial is earlier. Example: You open the control solution on January 15. Count three months until April 15. This is the date of discard to be entered on the vial.
(incl. dosage) Press the button " m "and hold it down for two seconds, until the screen display.
Two seconds, until the drive goes off. • To move through screens Press the button " m » to move through all
The configuration screens of the type alarms for reminder, sound level, time, date and so on. • To change settings inside a menu Press the button " c ».
Setting reminder alarms You can configure any or all of the reminder alarms (1 to 4). Note: No alarms are triggered during a test.
To configure one or more reminder alarms, proceed as follows.
Warning: The results of the FreeStyle test solution do not reflect your blood sugar level.
Select the reminder alarm

Sound setting There are three (3) sound volume options:
There are three sound volume options to let you know when the dosing electrode is full, when the result is reached and when an error occurs: 1. Press the button " c until the desired sound level appears. 2. Press the button " m "The time is set."
Time setting 1. When the time flashes, press " c » until the exact time appears. The " P The following is a list of the events in which the participants are invited to attend. 2. Press the button " m » to switch to the minute setting. When the minutes flash, press " c » until the exact minutes are displayed. 3. Press the button " m "to change to the time format setting."
Select Volume of
Setting the time format 1. When the time mode flashes, press " c To move from 12 hours to 24 hour format and vice versa. The time adjusts automatically according to the chosen format. 2. Press the button " m "to change to the date and format setting."
Note: To save your alarm settings and exit the configuration mode, press " m and hold it down for two seconds, until the drive is turned off. Press the button " m When the "0" flashes to switch to next configuration screen without setting a reminder alarm.

Setting the date and format 1. When the month flashes, press " c Until the exact month appears.
Insert the top of the dosing electrode into the reader until it stops. This turns the drive on.
System Check Screen Whenever the player turns on, this screen appears briefly to allow you to ensure that the display is working properly.
The date and time appear briefly after the system check screen, to make sure your drive is set at correct times and dates.
Patient until the reader displays the symbols of blood drop and dosage electrode.
Ne faire toucher délicatement that only one of the Deposit areas from the dosing electrode to the control solution.
If the sound is activated, the player will emit a beep once the dosing electrode is full. A small line then moves in the direction of needles a watch around the screen during sample processing.

The result appears on the screen and if sound is enabled, the player will emit two beeps. To mark the result as a control solution result, press " c and hold it for two seconds until the control solution symbol ( ) appears in the upper right corner of the screen.
If you mistakenly marked a blood glucose assay as a dose using the control solution, you can cancel it by pressing again the button " c and holding it down for two seconds.
Compare the result of the assay using the control solution to the range printed on the label of the dosing electrode vial. The results of the assays using the control solution shall be within the range printed on the side of the dosing electrode vial.
You can test to determine if hypoglycaemia is involved when:

Possible hematomas It is possible that a haematoma occurs at the sampling site. If so, you can choose to use another sampling site.
Preparation of the reader Insert a FreeStyle Lite dosing electrode into the drive. The reader is getting on.
Obtaining a blood sample Select a sampling site. There are differences in the results obtained from fingers compared to other sites. Use the autopicier device to obtain a blood sample. See the package leaflet for detailed instructions on how to use FreeStyle.
System Check Screen Whenever the player turns on, this screen appears briefly to allow you to ensure that the display is working properly.
FPO
a. Ensure that FreeStyle dosing electrode
Lite is well in the drive, that it is energized and that symbols of blood drop and dosing electrode are displayed. You are now ready to apply the blood sample.
The date and time appear briefly after the system check screen, to make sure your drive is set at correct times and dates.
b. Bring the FreeStyle Lite dosing electrode
slightly biasing on the blood sample.
The symbols of a drop of blood and the dosing electrode are displayed on the screen. The reader is now ready for the application of a blood sample.
Symbols of a drop in blood and the dosage electrode
c. FreeStyle Lite dosing electrode aspirates the
blood in the dosing electrode through the deposit area.
Deposition zone of the dosing electrode
Verification of · system

LO (low) and HI (high) results Important: Low or high blood glucose levels may indicate a potentially serious medical problem. The FreeStyle Freedom Lite player displays results ranging from 20 to 500 mg/dL (1.1 to 27.8 mmol/L).
If your test result is less than 20 mg/dL (1.1 mmol/L), LO (low) will appear on the player screen. This indicates severe hypoglycaemia (low blood glucose levels).
LO result with if you get a low result (LO) and that you have symptoms of weakness, sweating, nervousness, headache or confusion, Then follow your doctor's recommendations to treat hypoglycaemia.
LO result without if you get a low result (LO), but don't have any symptoms of hypoglycaemia, Then Repeat the test with a new dosing electrode on your fingers. If you still get a low result (LO), follow your doctor's recommendations to treat hypoglycaemia.
If your test is greater than 500 mg/dL (27.8 mmol/L), HI (high) appears on the screen. This indicates severe hyperglycaemia (high blood glucose levels).
HI result with if you experience symptoms such as fatigue, thirst, excessive urination or blurred vision, Then follow your doctor's recommendations to treat hyperglycaemia.
HI result without if you get a high result (HI), but do not have any symptoms of hyperglycaemia, Then Repeat the test with a new dosing electrode. If you still get a high result (HI), follow your doctor's recommendations to treat hyperglycaemia.
LO result · HI result

Test results The drive stores the last 400 test results with time and date. You can scroll forward through all the results by pressing the button " c " The button " m » allows you to scroll back. The first result you see is the most recent.
Start with the power off drive (without inserted dosing electrode). Press the button " m The first memory screen that appears is the 7-day average.
Temperature: A result with a thermometer symbol ( ) indicates that it was performed outside the specified range of operating temperatures of the reader and may not be accurate. This value is included in your 7, 14 and 30 day averages.
Press the button " m and hold it down for two seconds, until the drive is turned off.
Average on 7, 14, 30 days The first memory screen you see shows your average of 7 days. This includes all the results of the last 7 days, from the current date and time indicated on the player. (The results of the control solution, LO and HI are not included in the average.)
OL result: A LO result (less than 20 mg/dL or 1.1 mmol/L) is entered in the memory as a OL; it is not included in your 7, 14 and 30 day averages. HI result: A HI result (over 500 mg/dL or 27.8 mmol/L) is entered in memory as an IH; it is not included in your 7, 14 and 30 day averages.
Example:
7davg = your average over 7 days n =
Determination using the control solution A value with ( ) indicates that you marked this result as a dosage using the control solution. The date and time of the result of the control solution appear at the bottom of the screen.
The button " c allows you to scroll forward and move on to the averages over 14 days and 30 days. Continue pressing the button " c » to display the results of saved tests. The most recent result is shown first.
Determination of blood glucose A result displayed without the control solution symbol indicates that it is a blood glucose result. The date and time of the test appear at the bottom of the screen.

Maintenance of the Pile reader The FreeStyle Freedom Lite player is available with a 3 volt lithium battery #2032 pre-installed. The battery provides the drive with enough power to perform about 1000 tests. If the stack is weak, the battery symbol ( ) appears on the screen until you replace the pile.
The "PC" symbol will appear when a data transfer cable is inserted into the drive's data transfer port. You cannot perform a test when a data transfer cable is connected to the drive.
Le lecteur affiche « PC » jusqu’à ce que le câble soit retiré du port de transfert de données.
For more information, please contact:
please visit our website or call customer service.
Turn the drive. Sliding the component of
cell compartment to the right for opening.
To remove the stack, pull on the tab of
plastic up to release the battery.

Troubleshooting the system Error Messages
Install the stack with positive sign (+) directed up and towards you.
To close, slide the compartment flap of the stack inwards until it is set to actuate.
What you need to do 1. If you have symptoms of weakness, sweating, nervousness, headache or confusion, follow your doctor's advice to treat hypoglycaemia.
Error 2

What you need to do 1. If you have symptoms of thirst, fatigue, excessive urination or blurred vision, follow your doctor's recommendations to treat hyperglycaemia.
What you need to do
Possible causes • Incorrect dosing procedure. For example, put blood on the dosing electrode before inserting it into the reader or applying blood prior to displaying the symbols of the drop of blood and the dosage electrode.
What you need to do 1. Ensure that the blood drop and dosage electrode symbols are seen on the screen before applying any blood or control solution.
Possible causes

The reader does not enter test mode after insertion of a dosing electrode.
Method of measurement Electrochemical coulometric sensor
Possible cause What you can do
Automatic shutdown Two minutes after the last user action
The battery's exhausted. Replace the stack (and reconfigure date and time, if necessary).
Battery life 1000 tests
The battery is not installed or there are no batteries in the drive.
Check if the battery is installed correctly with positive (+) sign pointing up and towards you.
Inverted dosing electrode inserted on the wrong side or not fully inserted into the drive.
Insert the FreeStyle Lite dosing electrode with printed side up and top of the dosage electrode into the drive.
Hematocrite 15% to 65%
Units of measurement mg/dL or mmol/L
Test electrode or defective drive.
Call customer service.
Temp. storage of the drive
-20 °C à 60 °C (-4 °F à 140 °F)
Blood or foreign matter introduced into the port of insertion of the dosing electrode.
Call customer service.
Memory 400 blood glucose measurements and performed using the control solution with date and time
The dosage does not start after applying the blood sample.
Possible cause What you can do
The blood sample is too small.
You can add blood to the same FreeStyle Lite dosing electrode deposit area where you started, up to 60 seconds after first application. If necessary, repeat the test using a new FreeStyle Lite dosing electrode and larger blood sample.
Relative humidity of operation
5 to 90 % (without condensation)
4 °C à 40 °C (40 °F à 104 °F)
Defective dosing electrode. Restart the test with a new FreeStyle Lite dosing electrode.
Power source A 3 volt lithium battery No. 2032, replaceable
Results range 20 mg/dL at 500 mg/ dL (1.1 mmol/L to 27.8 mmol/l)
Sample applied after the delay of the reader has elapsed (two minutes) and it is not turned off.
Repeat the test using a new dosing electrode. Wait to see the symbols of blood drop and dosage electrode on screen before applying d-blood sample.
Fresh sample of total capillary blood
Test electrode or defective drive.
Call customer service.
Sample size 0.3 microlitre (300 nanolitres)
Dimensions Width 5.1 cm (2,00") x length 8.4 cm (3.30) x thickness 1.2 cm (0.50) (at the thickest point)
Weight 39.7 grams to 45.4 grams (1.4 ounces to 1.6 ounce), including pile
Troubleshooting · Calibration Plasma equivalent · Operating temperature
Owners and our team usually reply within a day — and your question helps the next person too.