DE
IT
ES
NL
RU
Thank you for purchasing the OMRON Body Composition Monitor. Before using this unit for the first time, please be sure to read this Instruction Manual carefully and use the unit safely and properly. Please keep this Instruction Manual at hand all the time for future reference.
9054902-7E



Never use this unit in combination with medical electronic devices such as: (1) Medical electronic implants such as pacemakers.
(2) Electronic life support systems such as an artificial heart/lung. (3) Portable electronic medical devices such as an electrocardiograph. • This unit may cause the above mentioned medical electronic devices to malfunction.

Medical electrical equipment is subject to special safety measurement regarding electromagnetic compatibility. The documentation in accordance with EN60601-1-2:2001 is available. Don’t use mobile telephones, microwave and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit.
Save these instructions for future reference.

Features of the Product
More accurate measurement of the entire body As measurement is conducted between both upper and lower limbs, the measurement values are less affected by the variation of physique as compared to traditional body fat monitors.
Visceral fat is displayed in 30 levels Visceral fat is closely related to lifestyle-related diseases, such as hypertension, diabetes, and hyperlipidemia. * OMRON BF500 can help you check your visceral fat level easily.
Easily check skeletal muscle percentage Skeletal muscle is displayed as a percentage. The unit only measures skeletal muscle, muscle connected to bone and used to move the body. It does not include smooth muscle, found in internal organs.
Display of BMI and resting metabolism In addition to calculating body fat percentage, it can also access your health using the BMI method*1, while taking your physique into consideration. It estimates the resting metabolism from personal data input and the measured body weight. Use those calculated values as reference for planning your weight reduction program. *1 BMI: Body Mass Index = Body weight (kg) / (height (m))²
Memory function to store up to four users’ personal data For the measurement of body composition, it is necessary to set personal data such as age, gender, and height. Once you set and store the data, the unit is ready for use by pressing the personal profile number button. This unit can be shared as it can store up to four different users’ data.
Comparison with previous results to monitor progress The unit stores the measurement results, so you can easily monitor any changes. You can display measurements for 1 day, 7, 30, and 90 days before.
The display unit can be operated easily and is easy to read.
For additional information, please see Section “Understanding the Results”.

What You Should Know When Using This Unit
Body fat has low electric conductivity The BF500 estimates the body fat percentage by the Bioelectrical Impedance (BI) method. Muscles, blood vessels and bones are body tissues with a high water content that conducts electricity easily. Body fat is tissue that has little electric conductivity. The BF500 sends an extremely weak electrical current of 50 kHz and less than 500 µA through your body to determine the amount of fat tissue. This weak electrical current is not felt while operating the BF500. In order for the scale to determine body composition, it uses the electrical impedance, along with your height, weight, age and gender information to generate results based on OMRON’s data of body composition.
Measures the whole body to avoid the influence of fluctuations During the course of a day, the amount of water in the body tends to gradually shift to the lower limbs. This is why there is a tendency for the legs and ankles to swell in the evening or at night. The ratio of water in the upper body and lower body is different in the morning and evening, and this means that the electrical impedance of the body also varies. Since the BF500 uses electrodes for both hands and feet to take measurements, it can reduce the influence of these fluctuations on measurement results.
What is the Bioelectrical Impedance method? Muscles, blood vessels and bones are body tissues having a high water content that conduct electricity easily. Body fat is tissue that has little electric conductivity. The BF500 sends an extremely weak electrical current of 50 kHz and less than 500 µA through your body to determine the amount of fat tissue. This weak electrical current is not felt while operating the BF500. The Bioelectrical Impedance Method safely combines the electric resistance with the distance of the electricity conducted. Correct posture and consistent measuring conditions need to be maintained for the best results.
Uses the best methods to measure different aspects of body composition OMRON has conducted research using exhalation and MRI analysis to develop an algorithm by which the bio impedance method of the BF500 works. Body Fat Percentage: This is based on the DXA (Dual Energy X-Ray Absorptiometry) method, which has been the established method for accurate evaluation of body composition. OMRON has used research information from several hundred people using the DXA method to develop the formula by which the BF500 works. The body fat mass and body fat percentage is calculated by a formula that includes five factors: electric resistance, height, weight, age and gender. The DXA method uses two different frequency X-Rays and rates of absorption of the body and determines the value based on the difference between the two. Resting Metabolism: This is based on exhalation analysis, where the composition of exhaled air is analysed. This makes it possible to calculate the resting metabolism by measuring the consumption of oxygen in exhaled air. Visceral Fat Level: This is based on MRI (Magnetic Resonance Imaging) analysis, which uses nuclear magnetic resonance that take non-intrusive images of the body. This makes it possible to calculate the area of visceral fat in the body.

What You Should Know When Using This Unit
Skeletal Muscle Percentage: This is based on MRI (Magnetic Resonance Imaging) analysis, which uses nuclear magnetic resonance that take non-intrusive images of the body. This makes it possible to calculate the amount of skeletal muscle in the body.
The reason calculated results may differ from actual body fat percentage There are certain conditions, such as with dialysis patients, when significant differences may occur between the estimated and actual body fat values. These differences may be related to changing ratios of body fluid and/or body composition.

What You Should Know When Using This Unit
Recommended measurement times Understanding the normal changes in your body fat percentage can help you in preventing or reducing obesity. Being aware of the times when the body fat percentages shift within your own daily schedule will assist you in obtaining an accurate trending of your body fat. It is recommended to use this unit in the same environment and daily circumstances. (See chart)

Before lunch and about 2 hours or more after breakfast
In the afternoon about 2 hours or more after lunch and before taking a bath or eating dinner
Before going to bed and about 2 hours or more after dinner or bath
Avoid taking measurements under the following conditions If a measurement is made under the following physical conditions, the estimated body fat percentage may differ significantly from the actual one because the water content in the body is changing. Avoid taking measurements:
- immediately after vigorous exercise Please wait for 1 to 2 hours:
- after a bath or sauna - after drinking alcohol - after drinking lots of water - after eating a meal
Measurement should be taken on a hard, level surface To ensure accurate measurement results, measurements should be taken on a hard, level surface.

What You Should Know When Using This Unit
Long-term planning for successful weight loss Ignoring a balanced diet and simply reducing food intake to lose weight can lead to a decline in muscle and bone. This may result in weight loss but not in the reduction in fat levels. So even though your weight goes down, your resting metabolism is reduced making the body more prone to putting on fat.
Example comparing the same loss of 10 kg

A starvation-like diet produces rapid weight loss of 10 kg, but disrupts body mass...
A healthy diet with weight loss of 10 kg...
Lean body mass reduction (muscle, bone etc.): 1 kg
Lean body mass reduction (muscle, bone etc.): 7 kg
Lean body mass (muscle, bone etc.) is greatly reduced, while the body fat percentage has not changed.
Repeated dieting and regaining weight leads to increased visceral fat levels It is easy to regain weight after sudden diets. This regained weight tends to appear as visceral fat rather then subcutaneous fat. Increased visceral fat levels are thought to increase the likelihood of common diseases. Repeated dieting and subsequent regaining of weight leads to increased visceral fat levels.


A. Grip Electrodes B. Display C. Personal Profile Number Button D. SET Button
F. Guest/Memory Button
Use this button to set the personal data for each measurement or to view measurement results stored in memory. G. Body Fat Button H. Resting Metabolism Button I. BMI Button J. Skeletal Muscle Button K. Display Unit Holder L. Cord M.Foot Electrodes N. Heel Electrodes O.Power Switch
Use this button to confirm the personal data or move to the next item after confirming the set item. E. Up/Down Buttons
Use these buttons to set the value or the gender for personal data. These are also used to select the measurement units used for the height and weight display.

1.Know Your Unit

P. Weight Indicator Q.Guest Indicator R. Personal Profile Number S. Weight Results T. Display Mode Indicator U. Body Composition Results V. Body Fat Percentage and BMI
Y. Measurement Start Indicator Z. Height Indicator AA.Gender Indicator
Assessment 12 Levels Bar W.Body Fat Percentage and BMI

1.Know Your Unit
Storing the Display Unit Extend the cord and store the display unit as described below.

Push the display unit until it clicks into place.

2.Inserting and Replacing the Batteries


Make sure that their polarity (+/–) is aligned with the polarity (+/–) indicated in the battery compartment.

When the battery low indicator ( ) appears, replace all four batteries with new ones.

3.How to Set the Date and Time
When the power is turned on for the first time, it is necessary to set the date and time.

SET button Use this button to confirm the setting or move to the next item after confirming the set item.
Down ( ) button Press this button to go back one. Press continuously to rapidly decrease.
Up ( ) button Press this button to advance one. Press continuously to rapidly advance.
If this is the first time that the unit has turned on, the year setting flashes.
The year setting is confirmed and the month flashes on the display.

[TO SET THE MONTH] 4. Press the or button to adjust the month setting to the desired month.

3.How to Set the Date and Time
The month setting is confirmed and the day flashes on the display.

The day setting is confirmed and the hour flashes on the display.

[TO SET THE HOUR] 8. Press the or button to adjust the hour setting to the desired hour.
The hour setting is confirmed and the minute flashes on the display.

The minute setting is confirmed.
After the settings for year, month, day, hour and minute are displayed in that sequence, the power is automatically turned off.

4.Setting and Storing Personal Data
For the measurement of body fat percentage and visceral fat level, it is necessary to set your personal data (age, gender, height). This section describes setting the personal profile setting for number 1 as an example.

SET button Use this button to confirm the personal data and move to the next item after confirming the set item.
Down ( ) button Press this button to go back one. Press continuously to rapidly decrease.
Up ( ) button Press this button to advance one. Press continuously to rapidly advance.
“CAL” blinks on the display, then the display changes to 0.0 kg. Wait until 0.0 kg appears on the display.
and weight measurements are displayed using the new measurement units. The change remains in effect until you change the measurement unit setting again.To change the measurement unit:


4.Setting and Storing Personal Data

The display of the selected personal profile number flashes.
The personal profile number is confirmed and the default age setting blinks on the display.
The age setting is confirmed, and the gender icons blink on the display.
To select Gender (male) or (female), press the or button.
The gender setting is confirmed, and the default height setting blinks on the display.

4.Setting and Storing Personal Data
Setting range: 100.0 to 199.5 cm (Display range in inch display mode: 3' 4" to 6' 6 3/4")
![Omron Body Composition Monitor (BF500) — 6. [TO SET HEIGHT] diagram (page 18)](/cdn/reader/df/df02c57c072c3dfeffb97ae760966eeb043dbd092c9ccf5149281935746b4863/images/fig-18-1.webp)
After the settings for age, gender, and height are displayed in that sequence, then 0.0 kg appears on the display. You can now start taking measurements. (Proceed to Step 3 in Section 5 “Taking a Measurement”.)
When using GUEST mode, the unit starts measurement after the height setting has been confirmed. Refer to Section 5.
The personal data has been stored in selected personal profile number.
If the personal data stored for any of the personal profile numbers (1-4) changes, the settings for that personal profile number will need to be modified.
“CAL” blinks on the display, then the display changes to 0.0 kg. Wait until 0.0 kg appears on the display.
The display of the selected personal profile number flashes once.

The personal profile number is confirmed and the selected age setting blinks on the display.
a)Press the or button to adjust the age setting to the desired age.
b)Press the SET button.
The age setting is confirmed, and the gender icon blinks on the display.

4.Setting and Storing Personal Data
a)Press the or button to adjust the gender setting to the desired gender.
b)Press the SET button.
The gender setting is confirmed, and the height setting blinks on the display.
a)Press the or button to modify the height setting.
b)Press the SET button to confirm the change.
If you are going to change other data, repeat this procedure.
“CAL” blinks on the display, then the display changes to 0.0 kg. Wait until 0.0 kg appears on the display.

The display of the selected personal profile number flashes once.
The personal profile number is confirmed and the selected age setting blinks on the display.
Press the Guest/Memory button for two seconds or longer.
“Clr” will appear in the display and the personal data and measurement values are deleted from memory.
Press the power switch to turn off the power.

5.Taking a Measurement
How to Measure Body Composition Please see Step 4 on this section for correct posture.
“CAL” blinks on the display, then the display changes to 0.0 kg.
the display. The weight of the display is then added to your body weight which causes an incorrect result.

Place both middle fingers along the dents of the grip electrodes.
Hold the inner grip electrodes firmly with your thumb and index finger.
Hold the outer grip electrodes with your ring finger and small finger.

5.Taking a Measurement
If you have personal data (Personal profile number) stored on the unit:
If you do not have personal data stored on the unit (GUEST mode):
Press the stored personal profile number button while holding the display unit. The stored personal profile number will appear after blinking once.
Even if you do not have personal data stored on the unit, you can measure your body composition by entering your age, gender, and height before taking a measurement.
Ex.: To select personal profile number “1”, press button .
“G” will be indicated at the top of the display while “AGE” data flashes.
If the following display is indicated:
Your personal data is not stored in the personal profile number you have selected. Please see Section 4 to store personal data.

5.Taking a Measurement


The display will show your weight and then the weight result will blink twice. The unit will then start to measure your body fat percentage and visceral fat level.
Step onto the unit. Distribute weight evenly.


The indicators in the measurement progress bar at the bottom of the display will gradually appear, from left to right.

5.Taking a Measurement
Correct Posture for Measurement When taking a measurement using a personal profile number, you need to store the personal data before hand. (Refer to Section 4.)

After measuring your weight. The arm is horizontally raised, and the elbow is extended straight. Extend your arms straight at 90° angle to your body.
Stand with your knees and back straight and look straight ahead.
Hold the display unit so that you can see the display.
Make sure each of your heels is positioned on a heel electrode. Stand with your weight evenly distributed on the measurement platform.

5.Taking a Measurement

POSTURES TO AVOID DURING MEASUREMENT Incorrect posture may result in inaccurate measurement of body composition.

5.Taking a Measurement
Press the appropriate button to view the desired measurement results.
For an explanation of the body composition results, see “Understanding the Results”.


5.Taking a Measurement

BMI BMI (Designation by the WHO) BMI Classification Bar BMI Rating
7.0 - 10.7 10.8 - 14.5 14.6 - 18.4
18.5 - 20.5 20.6 - 22.7 22.8 - 24.9
25.0 - 26.5 26.6 - 28.2 28.3 - 29.9
30.0 - 34.9 35.0 - 39.9 40.0 - 90.0
The above-mentioned indices refer to the values for obesity judgment proposed by the WHO, the World Health Organization.
– (Low) 0 (Normal) + (High) ++ (Very High)
Visceral Fat Level 9 __
Visceral fat area (0 - approx. 300 cm2, 1 inch=2.54 cm) distribution 30 levels. Example level determined 0: 1 - 9 level +: 10 - 30 level * The data referenced by Omron Healthcare.
Press the power switch to turn the power off.
For additional information, please see Section “Understanding the Results”.

6.Measuring Weight Only
“CAL” blinks on the display, then the display changes to 0.0 kg.
Your weight is displayed on the display and blinks twice to indicate that measurement is complete.
out the display unit.

7.Using the Memory Function
Measurement results are automatically stored in memory when you take a measurement using a personal profile number button. In order to assist your tracking of measurements, the device has stored 97 memories, and allows you to view the results for only 1 day, 7, 30, and 90 days before. When 97 sets of measurement values are stored in the memory, the oldest record is deleted to save the most recent measurement values.

Guest/Memory button Press this button to view previous results for only 1 day, 7 days, 30 days, and 90 days before.

Current weight/Body Fat Percentage

7.Using the Memory Function

Weight and Body Fat Percentage
- 1 day ago -

Weight and Body Fat Percentage
- 7 days ago -

Weight and Body Fat Percentage
- 30 days ago -

Weight and Body Fat Percentage
- 90 days ago -

7.Using the Memory Function
[Example only. The screen for the item selected will be displayed.]

Previous Body Fat results for the currently selected period. Press the Guest/Memory button to cycle through the other results.
Previous Skeletal Muscle result for the currently selected period. Press the Guest/Memory button to cycle through the other results.
Previous Resting Metabolism results for the currently selected period. Press the Guest/Memory button to cycle through the other results.
Previous BMI results for the currently selected period. Press the Guest/Memory button to cycle through the other results.

7.Using the Memory Function
Follow this procedure if you just want to view and compare previous measurement results, without taking a measurement.

“CAL” blinks on the display, then the display changes to 0.0 kg.

Ex.: To select personal profile number “1”, press button .

7.Using the Memory Function
The results for the previous day are displayed.
[Example]

8.Error Displays
Press your palms or soles firmly to the electrodes, then measure. (Refer to Section 5.)
Your palms or soles are not in firm contact with the electrodes.
The posture for measurement is wrong or the palms or soles are not in firm contact with the electrodes.
Measure without moving the hands or soles. (Refer to Section 5.)
Moisten the palms or soles with a wet towel, then repeat measurement. (Refer to Section 5.)
The palms or soles are too dry.
Please ensure that your personal data is within the measurable range of the device. (see Section 11).
Values of body composition are out of measurable range.
Insert the batteries again and repeat measurement. If this error still occurs, consult your OMRON service representative.
Abnormal operation.
You stepped on the unit before the display indicated 0.0 kg.
Step on the main unit after the display indicates 0.0 kg.
You moved the main unit before the display indicated 0.0 kg.
Do not move the main unit until the display indicates 0.0 kg.
You moved your body while measuring body weight.
Do not move while measuring body weight.
Body weight of 135 kg or over is out of the measuring range of this unit. You cannot use this unit.
Your weight is 135 kg or over.
Care and maintenance

9.Troubleshooting
The displayed value of body composition is abnormally high or low.
Refer to “The reason calculated results may differ from actual body fat percentage” and “Recommended measurement times” in “ What You Should Know When Using This Unit”.
The power turns off approximately 10 seconds after the body weight is confirmed and before your body fat percentage and visceral fat level is measured.
You did not press the correct personal profile number or the Guest/Memory button.(The personal profile number or “G” is not displayed on the display unit.)
Press the correct personal profile number or the Guest/ Memory button. (Refer to Section 5.)
Batteries are not inserted. Insert the batteries.
The polarities of batteries are not aligned correctly.
Insert the batteries in correct alignment.
Nothing is displayed when the power is turned on.
The batteries are worn out. Replace all four batteries with new ones. (Refer to Section 2.)
The cord connecting the main unit and the display unit is damaged or worn.
Contact the nearest OMRON service representative.

10.How to Care and Store the Unit


present.
Thinner · Gasoline · Benzene

11.Technical Data
Name BF500 Product Body Composition Monitor Model HBF-500-E Display Display of minimum body weight: 0 to 135 kg with an increment of 100 g Body Fat percentage: 5.0 to 60.0% with an increment of 0.1% Skeletal Muscle percentage: 5.0 to 50.0% with an increment of 0.1% BMI: 7.0 to 90.0 with an increment of 0.1 Resting Metabolism: 385 to 5000 kcal with an increment of 1 kcal Visceral Fat Level: 30 levels with an increment of 1 level BMI classification: – (Underweight) / 0 (Normal) / + (Overweight) / ++ (Obese) with 12 levels of Bar display Body fat percentage classification: – (Low) / 0 (Normal) / + (High) / ++ (Very High) with 12 levels of Bar display Visceral fat classification: 0 (Normal) / + (High) Memory: 1 day ago / 7 days ago / 30 days ago / 90 days ago Set Items Height: 100.0 to 199.5 cm (3' 4" to 6' 6 3/4") Age: 10 to 80 years old Gender: Male/Female * Height and the contents of a memory are displayed in the unit. *The age range for the skeletal muscle percentage, resting metabolism and visceral fat level is 18 to 80 years old. The age range for the body fat percentage classification is 20 to 79 years old. Weight Accuracy 0.0 kg to 40.0 kg: ± 400 g 40.1 kg to 135.0 kg: ± 1% Accuracy (S.E.E.) Body Fat percentage: 3.5% Skeletal Muscle percentage: 3.5% Resting Metabolism: 120 kcal Visceral Fat Level: 3 levels Power Supply 4 AA batteries (R6) (You may also use AA alkaline batteries (LR6).) Battery Life Approximately 1 year (when manganese batteries are used with four measurements a day) Operating Temperature /Humidity
+10°C to +40°C, 30% to 85% RH
Storage Temperature/ Humidity/Air Pressure
-20°C to +60°C, 10% to 95% RH, 700 hPa - 1060 hPa
Weight Approximately 2.2 kg (including batteries) External Dimensions Display unit: Approx. 300 (W) × 35 (H) × 155 mm (D) Main unit: Approx. 300 (W) × 57 (H) × 325 mm (D) Package Contents 4 AA manganese batteries (R6), instruction manual
Please read the instruction manual carefully before using the device.
= Type BF

11.Technical Data
Correct Disposal of This Product (Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.
This product does not contain any hazardous substances. Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries
With the increased number of electronic devices such as PC.s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.
This medical device manufactured by OMRON Healthcare conforms to this EN60601-1-2:2001 standard for both immunity and emissions. Nevertheless, special precautions need to be observed: • Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2001 is available at OMRON Healthcare Europe at the address mentioned in this instruction manual. Documentation is also available at www.omron-healthcare.com

11.Technical Data

OMRON HEALTHCARE CO., LTD. 24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto, 615-0084, Japan
OMRON HEALTHCARE EUROPE B.V. Kruisweg 577, 2132 NA Hoofddorp, The Netherlands www.omron-healthcare.com
OMRON HEALTHCARE UK LTD. Opal Drive Fox Milne, Milton Keynes MK15 0DG, United Kingdom
OMRON Medizintechnik Handelsgesellschaft mbH John-Deere-Str. 81a, 68163 Mannheim, Germany www.omron-medizintechnik.de
OMRON SANTÉ FRANCE SAS 14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, France
Made in China



Body fat is classified as subcutaneous fat and visceral fat etc., depending on where it is located in the body. In particular, visceral fat level is known to have a close connection to susceptibility common diseases.
What is Body Fat Percentage? Body fat percentage refers to the amount of body fat mass in regards to the total body weight expressed as a percentage. The BF500 uses BI(Bioelectrical Impedance) to calculate your body fat percentage using the following simple formula.
Body fat percentage (%) = {Body fat mass (kg) / Body weight (kg)} × 100 The BF500 uses body fat percentage to measure your body fat percentage using the above simple formula.
Body fat percentage classification differs for men and women When most people think of body fat, they have a negative image of it. However, it serves a vital role in storing energy, protecting internal organs, etc. While too much body fat is unhealthy for you, having too little is also unhealthy. The distribution of body fat in men and women is different, so the basis of classifying the body fat percentage for male and female is different.


20-39 < 21.0 21.0 - 32.9 33.0 - 38.9 39.0 40-59 < 23.0 23.0 - 33.9 34.0 - 39.9 40.0 60-79 < 24.0 24.0 - 35.9 36.0 - 41.9 42.0
20-39 < 8.0 8.0 - 19.9 20.0 - 24.9 25.0 40-59 < 11.0 11.0 - 21.9 22.0 - 27.9 28.0 60-79 < 13.0 13.0 - 24.9 25.0 - 29.9 30.0
Based on NIH/WHO guidelines for BMI Based on Gallagher et al., American Journal of Clinical Nutrition, Vol. 72, Sept. 2000

Body Fat Ratio & Visceral Fat Level

Visceral fat area (0 - Approx. 300 cm2, 1 inch=2.54 cm) distribution 30 levels. Example level determined 0: 1 - 9 level +: 10 - 30 level * The data referenced by Omron Healthcare.
Visceral fat area (0 - Approx. 300 cm2, 1 inch=2.54 cm) distribution 30 levels. Example level determined 0: 1 - 9 level +: 10 - 30 level * The data referenced by Omron Healthcare.
Visceral fat = fat surrounding internal organs Too much visceral fat is thought to be closely linked to increased levels of fat in the bloodstream, which can lead to common diseases such as hyperlipidemia and diabetes, which impairs the ability of insulin to transfer energy from the bloodstream and using it in cells. In order to prevent or improve conditions of common diseases, it is important to try and reduce visceral fat levels to an acceptable level. People with high visceral fat levels tend to have large stomachs. However, this is not always the case and high visceral fat levels can lead to hidden fat.
Subcutaneous fat = fat below the skin Subcutaneous fat not only accumulates around the stomach but also around the upper arms, hips and thighs, and can cause a distortion of the body’s proportions. Although not directly linked to increased risk of disease, it is thought to increase pressure on the heart and other complications. Subcutaneous fat is not displayed in this unit, but is included in the body fat percentage.


Skeletal muscle is muscle that is connected to bone and used to move parts of the body. The maintenance and increase of this skeletal muscle is closely linked to resting metabolism. This section describes resting metabolism and skeletal muscle.
What is Resting Metabolism? Regardless of your activity level, a minimum level of energy is required to sustain the body’s everyday functions. Resting metabolism, the amount of calories needed to supply the body with the minimum level of energy, differs between individuals depending on variables such as age, weight, body composition, and energy expenditure.
60-70% of daily energy use is for resting metabolism The total amount of energy burned by the body in a typical day is as follows.
Resting metabolism Energy required to maintain vital functions.
Energy used for daily activities such as commuting to work, household chores, hobbies etc.
Energy emitted after eating a meal.
The ratio of these is 60%-70% for resting metabolism, 20%-30% for daily activity, and 10% for diet induced thermogenesis. This means that resting metabolism accounts for most of our daily energy consumption. If our daily food intake exceed the amount of energy required for these activities, the additional energy is stored as fat.

Resting Metabolism & Skeletal Muscle

What is Skeletal Muscle? Muscle is divided into two types, muscle in internal organs, such as the heart, and muscle attached to bones that is used to move the body. Skeletal muscle can be increased through exercise and other activity.
Resting metabolism reduces as we get older Resting metabolism level peaks in the late teenage years and then gradually reduces over the years. This leads to reductions in the body’s functions as we get older and in particular, it is one of the main causes for muscle reduction as we get older. Even when not being used to move parts of the body, muscles burn energy throughout the day to generate heat for the body which is part of the resting metabolism. As the amount of muscle decreases, so does the amount of energy burned. If people continue to eat the same amount of food as when they are younger, then they develop an increasing abdominal circumference. In order to avoid this, it is important to know your resting metabolism and maintain muscles through exercise.
Average Female Approximately 28%*
Average Male Approximately 37%*
* According to Omron Healthcare figures.

Resting Metabolism & Skeletal Muscle

Resting Metabolism & Skeletal Muscle
Use this as a quick reference guide only. If you are using this device for the first time, please carefully read the full instruction manual.
Make sure that their polarity (+/–) is aligned with the polarity (+/–) indicated in the battery compartment.
3. Set a personal profile number. (Refer to Section 4.)
In order to measure body composition, personal data for age, gender, and height, needs to be set.
IF USING A PERSONAL PROFILE NUMBER... Once you have stored the personal data, you do not need to set the personal data each time you take a measurement. Press the personal profile number button for future measurements.
FOR GUEST USERS For guest users, set the personal data with each measurement. When using the Guest/Memory button, set the personal data for every measurement.
5. Check the measurement results. (Refer to Section 5.)
To view the measurement results, press the Body Fat, Resting Metabolism, Skeletal Muscle or BMI buttons.
Quick Reference Guide · !
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| The displayed value of body composition is abnormally high or low. | Refer to “The reason calculated results may differ from actual body fat percentage” and “Recommended measurement times” in “ What You Should Know When Using This Unit”. | p.34 | |
| The power turns off approximately 10 seconds after the body weight is confirmed and before your body fat percentage and visceral fat level is measured. | You did not press the correct personal profile number or the Guest/Memory button.(The personal profile number or “G” is not displayed on the display unit.) | Press the correct personal profile number or the Guest/ Memory button. (Refer to Section 5.) | p.34 |
| Nothing is displayed when the power is turned on. | Batteries are not inserted. | Insert the batteries. | p.34 |
| Nothing is displayed when the power is turned on. | The polarities of batteries are not aligned correctly. | Insert the batteries in correct alignment. | p.34 |
| Nothing is displayed when the power is turned on. | The batteries are worn out. | Replace all four batteries with new ones. (Refer to Section 2.) | p.34 |
| Nothing is displayed when the power is turned on. | The cord connecting the main unit and the display unit is damaged or worn. | Contact the nearest OMRON service representative. | p.34 |
Owners and our team usually reply within a day — and your question helps the next person too.