Instruction/ maintenance manual of the product PXI-8460 National Instruments
Go to page of 56
CAN Getting Started with Your PXI-8461 or PXI-8460 and the NI-CAN ™ Software for Windows 98/95 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 September 1999 Edition Part Number 3220 05B-01.
Worldwide Technical Support and Product Info rmation www.natinst.com National Instruments Corporate Headquarters 11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 794 0100 Worldwide O.
Important Information Warranty The PXI hardware is warranted against defects in materi als and workmanship for a peri od of one year from the date of shipment, as evidenced by receipts or other documentat ion. National Instruments will, at its opt ion, repair or replace equipment that pro ves to be defective during th e warranty p erio d.
Compliance FCC/DOC Radio Frequency Interference Class A Compliance This equipment generates and uses radio frequency energy and, if not installed and used in strict accorda nce with the instructions in this manual, may cause interference to radio and television reception.
© National Instruments Corporation v PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Contents About This Manual How To Use the Manual Set ............ .......................... ......................... ........... ............ .... ix Conventions .
Contents PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 vi www.natinst.com Appendix B Cabling Requirements for PXI-8461 High-Speed CAN Connector Pinouts .......................... ......................... ......................... ................
Contents © National Instruments Corporation vii PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Appendix F Technical Support Resources Glossary Figures Figure 2-1. Add/Remove Programs Properties Dialog Box............. ........................ 2-1 Figure 2-2.
Contents PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 viii www.natinst.com Table C-1. Power Requirements for the Low-Speed CAN Physical Layer for Bus-Powered Versions ....................... ........... ........... ............... .............. C-3 Table C-2.
© National Instruments Corporation ix P XI-8461 or PXI-8460 and NI-CAN for Windows 98/95 About This Manual This manual contains instructions to help you install and configure your PXI-8461 (one port and two port) or PXI-8460 (one p ort or two port) and the NI-CAN software for Windo ws 98/95.
About This Manual PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 x www.natinst.com Use the NI-CAN Programmer Refere nce Manual for specif ic information about each NI-CAN function and object, including format, parameters, and possible errors.
About This Manu al © National Instruments Corporation xi P XI-8461 or PXI-8460 and NI-CAN for Windows 98/95 • CiA Draft Standard 102, V ersion 2.0, CAN Physical Layer for Industrial Applications • CompactPCI Specification , Revision 2.0, PCI Industrial Computers Manufacturer’ s Group.
© National Instruments Corporation 1-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 1 Introduction This chapter lists what you need to get started, provides an overview of the PXI hardware and the NI-CAN software for Windows 98/95, and describes optional equipment you can order.
Chapter 1 Introduction PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 1-2 www .natinst.com PXI-8460 physical layer conforms to the ISO 1 1519-2 physical lay er specification for CAN and is also optically isolated to 500 V . CAN interfacing is accomplished using the Intel 82527 CAN controller chip.
Chapter 1 Introduction © National Instruments Corporation 1-3 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 • Configuration utility • Language interface libraries for Microsoft V isual C/C++ 2.0 or later , LabW indows/CVI 4.0 or later , and LabVIEW 4.
© National Instruments Corporation 2-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 2 Installation and Configuration This chapter describes how to install and configure the NI-CAN so ftware for Windows 98/95 and the PXI-8461 or PXI-8460 hardware.
Chapter 2 Installation and Configuration PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 2-2 www .natinst.com Y ou can use this same applet to uninstall the NI-CAN software at a later time. Refer to Appendix A, Uninstalling the Hard ware and Software , for more information.
Chapter 2 Installation and Configuration © National Instruments Corporation 2-3 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 4. T ouch a metal part on your chassis to discharge an y static electricity that might be on your clothes or body . 5. Insert the PXI board into the selected 5 V slot.
Chapter 2 Installation and Configuration PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 2-4 www .natinst.com If the New Hardware Found dialog box does not appear when you restart W indows, refer to the section Device Manager Pro blem s in Appendix D, Troubleshooting and Common Questions .
Chapter 2 Installation and Configuration © National Instruments Corporation 2-5 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Figure 2-4. Device Manager Shows PXI-8461 That Is Not Working Properl.
Chapter 2 Installation and Configuration PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 2-6 www .natinst.com Each port of the PXI-8461 or PXI-8460 is configured from the same tab . Use the drop-do wn box nearest the top of the tab to select the physical port number to configure.
© National Instruments Corporation 3-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 3 V erify the Installation This chapter describes how to verify the hardware and software installation. Y ou can use the NI-CAN Diagnostic uti lity , installed with you r NI-CAN software, to test the hardware and software installati on.
Chapter 3 Verify the Installation PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 3-2 www .natinst.com and Common Questions , to troubleshoot the problem. T roubleshooting information is also av ailable in the online help for the NI-CAN Diagnostic utility , which you can access by clicking on the Help button.
© National Instruments Corporation 4-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 4 Begin to Use the NI-CAN Software This chapter helps you get started with the NI-CAN software for Windows 98/95. Using the NI-CAN Software The functions provided by the NI-CAN software are similar to those provided by many other device drivers.
Chapter 4 Begin to Use the NI-CAN Software PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 4-2 www .natinst.com General Programming Considerations As you begin develo ping your Win32 N I-CAN application, rememb er the following p oints: • For your LabVIEW or BridgeVIEW application, you must use the NI-CAN LabVIEW functions in nican.
© National Instruments Corporation A-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 A Uninstalling the Hardware and Software This appendix describes ho w to uninstall the PXI hardware and the NI-CAN software from W indo ws 98/95.
Appendix A Uninstalling the Hardware and Software PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 A-2 www .natinst.com Figure A-1. Selecting an Interface to Remove from Windows 98/95 After you remov.
Appendix A Uninstalling the Hardware and Softwa re © National Instruments Corporation A-3 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Uninstalling the NI-CAN Software from Windows 98/95 Before uninstalling the NI-CAN software, you should remove all PXI-8461 or PXI-8460 boards from your computer , as explained in the pre vious section.
Appendix A Uninstalling the Hardware and Software PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 A-4 www .natinst.com The uninstall program removes o nly items that the installation program installed.
© National Instruments Corporation B-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 B Cabling Requirements for PXI-8461 High-Speed CAN This appendix describes the cabling requirements for the hardware. Cables should be constructed to meet these requirements and the requirements of the other CAN or De viceNet devices in the network.
Appendix B Cabling Requirements for PXI-8461 High-S peed CAN PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 B-2 www .natinst.com The 5-pin Combicon-style pluggable scre w terminal follows the pinout required by the DeviceNet Specification . Figure B-2 shows the pinout for this connector .
Appendix B Cabling Requirements for PXI-8461 High-S peed CAN © National Instruments Corporation B-3 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Figure B-3. PXI-8461 Parts Locator Diagram Connecting pins 1 and 2 of a jumper configures the CAN physical layer to be powered e xternally (from the bus cable power).
Appendix B Cabling Requirements for PXI-8461 High-S peed CAN PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 B-4 www .natinst.com Figure B-4 shows ho w to configure your jumpers for intern al or external power supplies. Figure B-4. Power Sour ce Jumpers The CAN physical layer is still isolated re gardless of the power source chosen.
Appendix B Cabling Requirements for PXI-8461 High-S peed CAN © National Instruments Corporation B-5 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Cable Specifications Cables should meet the physical medium requirements specif ied in ISO 11898, shown in T able B-2.
Appendix B Cabling Requirements for PXI-8461 High-S peed CAN PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 B-6 www .natinst.com Number of Devices The maximum number of de vices that you can connect to a CAN port depends on the electrical characteristics of the de vices on the network.
Appendix B Cabling Requirements for PXI-8461 High-S peed CAN © National Instruments Corporation B-7 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Cabling Example Figure B -6 shows an e x ample of a cable to connect two CAN de vices. For the internal power configuration, no V+ connection is required.
© National Instruments Corporation C-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 C Cabling Requirements for PXI-8460 Low-Speed CAN This appendix describes the cabling requirements for the PXI-8460 hardware. Cables should be constructed to meet these requirements, as well as the requirements of the other CAN devices in the network.
Appendix C Cabling Requirements for PXI-8460 Low-Speed CA N PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 C-2 www .natinst.com CAN_H and CAN_L are signal lines that carry the data on the CAN network. These signals should be connected using twisted-pair cable.
Appendix C Cabling Requirements for PXI-8460 Low-Speed CAN © National Instruments Corporation C-3 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Connecting pins 1 and 2 of a jumper configures the CAN physical layer to be powered e xternally (from the bus cable power).
Appendix C Cabling Requirements for PXI-8460 Low-Speed CA N PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 C-4 www .natinst.com Cable Specifications Cables should meet the physical medium requirements specif ied in ISO 11519-2, sho wn in T able C-2.
Appendix C Cabling Requirements for PXI-8460 Low-Speed CAN © National Instruments Corporation C-5 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Low-Speed T ermination Every de vice on the low-speed CAN netw ork requires a termination resistor for each CAN data line: R RT H for CAN_H and R RT L for CAN_L.
Appendix C Cabling Requirements for PXI-8460 Low-Speed CA N PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 C-6 www .natinst.com v alues o f 510 Ω ±5% per port.
Appendix C Cabling Requirements for PXI-8460 Low-Speed CAN © National Instruments Corporation C-7 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Figure C-5. Location of T ermination Resistors on a PXI-8460 2. Cut and bend the lead wires of the resistors you want to install.
Appendix C Cabling Requirements for PXI-8460 Low-Speed CA N PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 C-8 www .natinst.com Cabling Example Figure C -7 shows an e x ample of a cable to connect two lo w-speed CAN de vices.
© National Instruments Corporation D-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 D T roubleshooting and Common Questions This appendix describes ho w to troubleshoot problems and answers some common questions. Device Manager Problems The W indows De vice Manager contains configuration information for all of the hardware in your system.
Appendix D T roubleshooting and Common Questions PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 D-2 www .natinst. com T o fix the problem, complete the follo wing steps: 1. Select Start»Settings»Con trol Panel . 2. Double-cl i ck on the System icon to display the System Properties dialog box for the computer .
Appendix D T roubleshooting and Common Questions © National Instruments Corporation D-3 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Problem Shown in Device Manager If a PXI-8461or PXI-8640 is not working properly , its icon has a circled X or exclamation mark (!) o verlaid on it, as shown in Figure D-1.
Appendix D T roubleshooting and Common Questions PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 D-4 www .natinst. com Code 12: Y our interface w as not assigned a physical memory range. If your computer does not ha ve 8 KB of av ailable memo r y , W indows might configure your interface without a ph ysical memory assignment.
Appendix D T roubleshooting and Common Questions © National Instruments Corporation D-5 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Diagnostic Utility Failures The following sections explain common error messages generated by the NI-CAN Diagnostic utility .
Appendix D T roubleshooting and Common Questions PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 D-6 www .natinst. com Missing Board in Diagnostic Utility If a board is physically installed in your sy stem, but is not listed in the NI-CAN Diagnostic utility , check the W indows De vice Manager to see if W indo ws has detected the board.
Appendix D T roubleshooting and Common Questions © National Instruments Corporation D-7 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 What do I do if my board is listed in the Windo ws Device Man.
© National Instruments Corporation E-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 E Specifications This appendix describes the physical characteristics of the hardw are and the recommended op erating cond i tions. PXI-8461 One Port and T wo Port Dimensions.
Appendix E Specifications PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 E-2 www .natinst.com Operating environment Component temperature ................. . 0 to 55 °C Relative humidity ............. .............. .10% to 90%, noncondensing Storage environment Temperature.
© National Instruments Corporation F-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 F T echnical Support Resources This appendix describes the comprehensiv e resources av ailab le to yo u in the.
Appendix F T echnical Support Resources PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 F-2 www .natinst.com Software-Related Resources • Instrument Driver Network —A library with hundreds of instrument dri v ers for contro l of standalone instruments via GPIB, VXI, or serial interfaces.
© National Instruments Corporation G-1 PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 Glossar y Pref ix Meaning V alue n- nano - 10 –9 m- milli- 10 –3 k- kilo- 10 3 M- mega- 10 6 ° degrees .
Glossary PXI-8461 or PXI-8460 and NI-CAN for Windows 98/95 G-2 www .natinst. com HMI Human Machine Interface Hz hertz IEEE Institute of Electrical and Electronic Engineers in.
An important point after buying a device National Instruments PXI-8460 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought National Instruments PXI-8460 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data National Instruments PXI-8460 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, National Instruments PXI-8460 you will learn all the available features of the product, as well as information on its operation. The information that you get National Instruments PXI-8460 will certainly help you make a decision on the purchase.
If you already are a holder of National Instruments PXI-8460, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime National Instruments PXI-8460.
However, one of the most important roles played by the user manual is to help in solving problems with National Instruments PXI-8460. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device National Instruments PXI-8460 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center