Instruction/ maintenance manual of the product PROPLUS 2000KL Health O Meter
Go to page of 29
PRO PLUS ® STRETCHER SCALE MODEL 2000KL OPERATION MANUAL AA0339-0.
1 PRO PLUS ® STRETCHER SCALE MODEL 2000KL Thank you for your purchase of this product. Please read this manual carefully and keep it handy for ready reference. TABLE OF CONTENTS CAUTION AND WARNING ....................................................
2 CAUTION AND WARNING To prevent injury and damage to your scale, pleas e follow these instructions very carefully. • To ensure proper operation of the ProPlus® Stretcher Scale, this operation manual should be reviewed carefully before operation.
3 SPECIFICATIONS GENERAL Health o meter’s ProPlus ® Stretcher Scale Model 2000KL uses highly sophisticated microprocessor technology. This scale is designed to provide accurate, reliable and repeatable weight measurements and features that make the weighing process simple, fast and convenient.
4 ASSEMBLY INSTRUCTIONS Before Assembly The Model 2000KL ProPlus® Stretcher Scale is shipped disassembled in two (2) separate cartons. Carefully inspect all cartons for shipping damage before unpacking. If damage is found, contact your shipper or Healthometer Professional immediately.
5 ASSEMBLY INSTRUCTIONS Assembling the Patient Lifter 1. Place the Stretcher Storage Tray on the Lifter Base, as shown below, before attaching the Lifter Mast.
6 ASSEMBLY INSTRUCTIONS Attaching the Display and Scale to the Patient Lifter Figure 2 Figure 3 PHILLIPS PAN HEAD SCREW NC8-32x5/16.
7 ASSEMBLY INSTRUCTIONS 4. Connect the Lifter Boom and the Display Support to the Lifter Mast using the NC ⅜ - 20x2 ½” hex screw and tighten the self lock nut on its other side.
8 ASSEMBLY INSTRUCTIONS Figure 4 10. Position the Display Assembly close to the Display Support. Connect the load cell cable connector to the load cell port on the rear of the Display Assembly and insert and secure the cable into the center channel in the display.
9 ASSEMBLY INSTRUCTIONS Figure 5 15. Hang the Load Cell Assembly on the two hooks at the end of the Lifter Boom, using the loops on top of the Load Cell Assembly. 16. Connect the end of the Load Cell Cable coming out of the center of the Load Cell Assembly to its port in the Lifter Boom.
10 SET UP Preparing the Scale for Use 1. Remove protective plastic film from keypad and display. 2. Place batteries in the battery holder (see “Replacing Batteries”). 3. Plug the scale’s AC adapter into the power source. 4. Press the ON/OFF button to turn the scale on.
11 OPERATION INSTRUCTIONS Preweighing the Stretcher 1. Make sure the scale is free of load (nothing is hanging from it), including the stretcher. 2. Press the ON/OFF button to turn the scale on. 3. Wait until ”000 Lb 00 oz ” and “ZERO” on the left side of the display will appear.
12 OPERATION INSTRUCTIONS 4. Release the hanging loops from the lifter. Note : The scale is programmed to save the last setting chosen as well as the weight of the stretcher as tare weight. This enables weighing a number of patients without having to re- set the scale.
13 OPERATION INSTRUCTIONS Figure 7. Keypad ITEM DESCRIPTION FUNCTION POWER Turns scale ON and OFF. ZERO Zeros the scale prior to weighing. HOLD/RELEASE Holds the value of the weighed object on the display until the button is pressed again to clear the value.
14 OPERATION INSTRUCTIONS MENU In the menu screen the user can set preferences and/or instruct the scale how to handle stored data. The menu can be navigated using the up and down keys ( cd ) or by entering the associated menu position number with the keypad.
15 OPERATION INSTRUCTIONS 02 TARE WEIGHT NOTE : Due to the scale’s sensitivity, we recommend using the REWEIGH function prior to setting the TARE weight, in order to eliminate any operator interference with the item to be tared out.
16 OPERATION INSTRUCTIONS 3. The scale will prompt you to enter the patient’s height. Use the keypad to enter the height in 1/4" (1=1/4, 2=1/2, 3=3/4) or 1 cm increments (use inches for weight in lb, cm for kg). Press ENTER. 4. Display will read the patient’s BMI.
17 OPERATION INSTRUCTIONS 04 Transfer now All the values stored in memory are transferred to the PC and the scale memory is cleared of all values. If the transfer was unsuccessful, the values are kept in memory until successfully transferred or cleared.
18 OPERATION INSTRUCTIONS 03 Auto Off Time The user can determine how long the scale will operate before turning off automatically due to inactivity. Default time is 10 minutes. If the value is set to zero, the auto off function is disabled. 04 Tone Volume There is an option to adjust the beeping tone of the scale.
19 OPERATION INSTRUCTIONS 06 Display Backlight The user can set the brightness of the backlight. 07 Display Contrast The user can set the brightness of the LCD. 08 Live Weight By selecting “Yes” the user can set the Live Weight mode to deactivate the motion-sensing mode.
20 OPERATION INSTRUCTIONS Set the date using the up and down keys cd and enter the values on the keypad. 02 About This screen displays the software version of the scale. 06 SYSTEM TEST 01 Battery Test The scale will display the estimated amount of battery life remaining until the batteries will have to be replaced.
21 OPERATION INSTRUCTIONS 03 UI Test The scale has a diagnostic routine where it tests the User Interface (UI) hardware functionality (LCD, keypad). In order to do this the user has to press all the keys according to the messages displayed on the screen.
22 OPERATION INSTRUCTIONS 02 Disp Height & ID When the user selects to display the height and ID of the user, it will be displayed at the bottom of the screen. We recommend the use of this function to ensure that the patient’s correct ID and height have been entered.
23 MAINTENANCE GENERAL This section provides instructions for maintenance, cleaning, troubleshooting and operator replaceable parts for the ProPlus ® Stretcher Scale Model 2000KL. Maintenance operations other than those described in this section should be performed by qualified service personnel.
24 TROUBLESHOOTING Refer to the following instructions to check and correct any failure before contacting service personnel. SYMPTOM POSSIBLE CAUSE CORRECTIVE ACTION Scale does not turn on 1. Dead Battery 2. Faulty electrical outlet 3. Bad power supply 1.
25 CALIBRATION PATH The calibration is performed in Kg or Lb, according to the units of measure used upon entering into the calibration path. Operator Action Display 1. Press ON/OFF button and immediately press and hold HOLD/RELEASE button for 3-4 seconds Calibration Process 2.
26 EXPLODED VIEW OF SCALE (continue).
27 PARTS LIST Key No. Part No . Description Qty. 1 1109403-0 DISPLAY ASSEMBLY 1 2 2268101-0 TILT MECH. ASSEMBLY 1 3 3848401-0 PLASTIC WASHER 2 4 401369 PLASTIC OVAL BAR COVER 3 5 HEX. HEAD SCREW NC 3/8-16*2 1/2 1 6 HEX. HEAD SCREW NC 3/8-16*1 3/4 1 7 2268001-0 LIFTER BOOM ASSEMBLY 1 8 2034701-0 MAIN BOARD TO CON.
28 WARRANTY LIMITED WARRANTY What does the Warranty Cover? Pelstar LLC scales are warranted from date of purchase against defects of materials or in workmanship for a period of three (3) years. If product fails to function properly, return the product, freight prepaid and properly packed to Pelstar.
An important point after buying a device Health O Meter PROPLUS 2000KL (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Health O Meter PROPLUS 2000KL yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Health O Meter PROPLUS 2000KL - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Health O Meter PROPLUS 2000KL you will learn all the available features of the product, as well as information on its operation. The information that you get Health O Meter PROPLUS 2000KL will certainly help you make a decision on the purchase.
If you already are a holder of Health O Meter PROPLUS 2000KL, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Health O Meter PROPLUS 2000KL.
However, one of the most important roles played by the user manual is to help in solving problems with Health O Meter PROPLUS 2000KL. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Health O Meter PROPLUS 2000KL along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center