LT801310/C - ©2003 Gilson SAS All rights reserved November 2003




Safety
Read this section carefully before installing and operating the instrument.
For safe and correct use of the instrument, it is essential that both operating and service personnel follow generally accepted safety procedures as well as the safety instructions given in this document, the 402 Dilutor-Dispenser User’s Guide.
The instrument described in this document is a syringe operated pump for transfer- ring liquids. It should only be used in the laboratory or similar indoor environment for analytical purposes, by qualified personnel. If an instrument is used in a manner not specified by Gilson, the protection provided by the instrument may be impaired.
Voltages present inside the instrument are potentially dangerous. If there is a problem with the instrument, the power cable should be removed until qualified service personnel have repaired it. This is to prevent anyone from inadvertently using the instrument, thus causing possible harm to themselves, or damage to the instrument itself.
The leakage current of this instrument is within the limits allowed by international safety standards for laboratory equipment. An efficient ground connection is impera- tive for the physical protection of the user.
Power supply cord reference 7080316106 is for use in France and Germany. Power supply cord reference 7080316105 is for use in USA and Canada. For other countries contact your local Gilson distributor. You must only use the type of fuse described in this document: 1.0 Amp type “T” slow blow.
Adequate protection including clothing and ventilation must be provided if dangerous liquids are used in the analytical work. In case of incidental spillage, carefully wipe with a dry cloth, taking into account the nature of the spilled liquid and the necessary safety precautions.
Cleaning, dismantling, maintenance, and repair should only be performed by personnel trained in such work, and who are aware of the possible dangers involved.
Safety-1

1 Individual components: parts of the valve, the Tee junction, and the syringes (except for the 25 mL syringe) may be autoclaved after dismounting (121°C, 0.1 MPa, 20 min).
This instrument1 must not be sterilized, using an autoclave, or any other mechanical device. When you need to clean an instrument, use one of the three following methods:
If a cloth dampened with soapy water is used to clean the instrument, only domestic soap may be used. No other form of detergent or chemical may be used.
These electronic and hazard symbols appear on the 402 Dilutor Dispenser:
PROTECTIVE CONDUCTOR TERMINAL
O

The following chapter describes the instrument, the service and warranty conditions.

Notational Conventions
The conventions in this User’s Guide are as follows.
Used in the stand-alone mode, this instrument is referred to as ‘402 Dilutor-Dispenser’ or simply 402.
Specific keys to press on the Gilson Keypad are represented in bold capitals, for example: ‘key in the value and press ENTER’; where ENTER is a key on the keypad.
For a soft-key that must be pressed, the function of that key is represented in bold, for example: ‘Press Prime to prime the pump’.
The Gilson Keypad displays on-screen messages as normal typeface inside a frame.

The 402 Dilutor-Dispenser, is used to transfer liquids between vials according to user defined protocols (Methods). You program the 402 Dilutor-Dispenser via the keypad, which includes an LCD screen, for communication purposes.
You may store a total of 99 linkable Methods using the keypad. Each Method consists of one or more of the following tasks: Dilute, Dispense, Mix, Pipette, and Rinse. In addition three functions for direct manipulation are available: Priming, Titration, Volume Measurement.
Gilson’s 402 Dilutor-Dispenser is available in two configurations: Single-syringe, or Dual-syringe. The 402 Dilutor-Dispenser is very versatile; it may be fitted with seven sizes of syringe. The standard 402 Dilutor-Dispenser is equipped with a hand- held probe for aspirating and dispensing liquids. Optionally, the probe may be fitted with disposable tips (Gilson Diamond® Tips, D200 and D1000).
This versatility means that your 402 Dilutor-Dispenser may be used for a wide variety of applications including the chemical field where the ability to handle large volumes or high dilution ratios may be appreciated, and the biological field where small volumes and protection against cross-contamination are important (see Autoclavability, page 6-5). It can even be used with corrosive liquids (see Chemical Compatibility, Appendix B).
Refer to Appendix A for lists of the standard equipment, additional accessories and spare parts. The 402 DD is supplied with a tubing kit for con- necting to a solvent reservoir. Syringes are not included in the standard equipment; you should order syringe(s) appropriate to your Method (volume, systematic error and flow rate requirements; refer to Appendix C). Also, a Tee junction kit (not includ- ing the syringe) is available for upgrading a Single- syringe Dilutor-Dispenser to the Dual-syringe version.

Good Laboratory Practice (GLP) is facilitated by the following:
codes to protect the program files against unau-
thorized modification, counters for the number of syringe and valve
checking protocol for the hydraulic specifica-
correction of the syringe-piston stroke to com-
pensate for environmental factors (temperature, viscosity, etc.).
operating-cycles, · tions (IQ/OQ procedures),

Unpacking
Your 402 Dilutor-Dispenser is shipped in a box together with its various accessories. The original box and packing material should be kept in case the equipment has to be returned to your Gilson distributor.
You should check off the parts received in the box against the parts list (refer to Appendix A), at the time that you receive the shipment, even if the 402 is not to be used immediately. Your Gilson distribu- tor should be notified immediately of any inconsist- encies between the contents of the box and the parts list, or if any of the parts appear to have been damaged in transit.

Warranty
If the 402 Dilutor-Dispenser does not appear to function correctly, first verify the electrical connec- tions are correct and that the instrument is switched ON. Contact your Gilson distributor for technical advice or possibly a service visit. Any service required will be given within the warranty conditions assured by your Gilson distributor.

Customer Service
Gilson and its worldwide network of authorized distributors provide you with four types of assist- ance: sales, technical, applications and service.
Customer Service personnel are able to serve you more efficiently if you provide the following information:


The following chapter describes the mechanics and hydraulics of the Dilutor-Dispenser.
Two models of 402 Dilutor-Dispenser are available: Single-syringe and Dual-syringe. Each model comes with a keypad, a probe, and accessories. You order and fit the syringe(s), from those described on page 2-6, most suited to your application.
Your 402 is delivered with the appropriate modules fitted. If later, you decide to convert a Single-syringe model into a Dual-syringe model, you must order and fit a Tee-junction module to the right-hand side of your dilutor; the left-hand always contains a Valve module.

Single-syringe Model
The Single-syringe model has a Valve module at the left, and a blanking plate at the right.
The Valve module includes one motor to operate the syringe and another motor to switch the valve. It is a liquid transfer station that can be used to deliver up to 25 mL in one stroke, and down to 1 µL (typical value), with high accuracy.
Bottle Clip for Tubing


The front panel holds the hydraulic components and is fitted with an ON/OFF indicator.
The rear panel holds a socket for connecting the keypad, a connector for the probe, the ON/OFF switch, and voltage selector/fuse holder. There is also a Gilson Serial Input Output Channel (GSIOC) connector, which is a communications channel to link the 402 Dilutor-Dispenser with other equipment, if required. There is also a bank of microswitches for use in setting up the communications channel.

The Dual-syringe model is fitted with a Valve module at the left and a Tee-junction module at the right. The modules are linked by plastic tubing.
Bottle Clip for Tubing Tubing Clip (Self-adhesive)

The front panel holds the hydraulic components and is fitted with an ON/OFF indicator.
The Valve module includes one motor to operate the syringe and another motor to switch the valve. The Tee-junction module has only one motor, to operate the syringe.
The Dual-syringe model may be fitted with syringes of different volumes. The left syringe is the large volume syringe and is used to deliver a large volume of liquid in a minimum time. The right syringe can be used to transfer smaller volumes with high reproducibility and accuracy (refer to Appendix C for Technical Data). The benefits of this configuration are a large dilution ratio with the best reproducibil- ity and fast transfer times.
This is the same for both models.

Keypad and Probe
Keypad
The keypad includes an LCD display (two 24- character lines), membrane touch-keys, and a connecting cord.
decimal point key, and the Enter key.
the screen that is displayed (e.g. Next, Prev, Chge...).
STOP key: Press this key to stop the current ac- tion (Method or direct manipulation).
HELP key: Press this key to access help messages relating to the current screen.
ESC key: Press this to quit a menu or return to the first menu of the software branch.
CLEAR key: Press this key to cancel a value and return to the previous value before validating it.

The assembled probe fitted with the standard probe nozzle is shown opposite. Assembly and installation of the standard probe is described in Chapter 3. A kit is available to enable you to fit Gilson’s Diamond® tips (D200, D1000) to the probe (see Chapter 3).

Valve
Probe (or Tee junction)
The two-position valve, which is fitted to the left- hand side of the 402, has three ports. It switches the syringe port to either the reservoir or to the probe side (see opposite figure).
The valve, which has a 0.5 sec rotation period, is driven by a stepper motor. It is designed to with- stand a million rotations, when used with water. The liquid contact surfaces of the valve, which is autoclavable, are in PEEK and ceramic (121°C, 0.1 MPa, 20 min).
From Valve To Probe
The Tee-junction, which is fitted to the right-hand side of the dual-syringe model, has three ports: the inlet from the valve, the outlet to the probe, and the syringe port (see opposite figure).
The Tee-junction, which has no internal moving parts, includes 2 channels, each 1.5 mm in diam- eter oriented at 120°. The liquid contact surfaces of the Tee-junction, which is autoclavable, are in PEEK.
The choice of syringe allows the 402 Dilutor-Dispenser to handle volumes from 1 µL to 25 mL with low values of systematic and random error. The 402 Dilutor-Dispenser is able to transfer liquid as follows:
- from 1 µL to 10 mL from the probe (transfer
tubing volume of up to 10 mL),
- up to 25 mL from the reservoir, in one stroke,
- up to 99 mL from the reservoir, in several strokes.
When using the Dual-syringe model, a theoretical dilution ratio of 1/25000 is achieved in one stroke of both sample and diluent syringes. However, you are advised not exceed a ratio of 1/500 for any given dilution, in order to stay within the RSD range (after efficient mixing).

0.1 mL
There are seven syringe sizes available for the 402, all except the 25 mL are autoclavable. The lifetime of all pistons (see Chapter 6), when they are used with water at a pressure of up to 0.3 MPa is over 30000 cycles (each cycle contains 2 strokes: one up and one down).

0.5 mL 0.25 mL
The available syringe sizes are:
25025341 0.1 mL 25025342 0.25 mL 25025347 0.5 mL 25025343 1 mL

25025344 5 mL 25025345 10 mL 25025346 25 mL
Part of the syringe Material
Syringe body Borosilicate glass Syringe adapter 316L stainless steel Piston Stainless steel (models up to 1 mL)
Aluminum Piston nose PTFE/Ekonol (PTFE for the 25 mL)

This chapter explains how to install the fuses, syringes and hydraulics according to your system requirements.
This chapter explains how to add the extra module to upgrade your 402 to a dual syringe pump.

Assembling the Probe
For Standard Use
In standard use the probe is used with the 2 mL tapered transfer-tubing (see opposite figure).


When you want to increase the length of tubing protruding from the probe, you must first carefully lift the tubing out of the handle, otherwise you will damage the tubing.

The ability to use Diamond Tips will extend the use of your 402 into the biological field, where protec- tion against cross-contamination is important, and in the chemical field to protect the tubing from corrosive liquids.

You must fit a tip-holder to the probe when you want to use Gilson’s Diamond tips. An adapter kit (reference F4015015) is available to permit you to fit them to the probe. The kit includes two sizes of tip-holder (with built-in tip ejector), and a special non-tapered 2 mL transfer tubing, as follows.

2 Special adapters for the probe 1 D200 tips, pack of 10 1 D1000 tips, pack of 10 1 Transfer-tubing (outlet) non-tapered, FEP (2 mL)
(1) The small washer is important in that it helps to ensure that fluid does not leak from the junction between the tip-holder and the orifice of the probe.
(2) The special ‘S’ shaped metallic clips, which are pre-mounted, are for supporting the tubing on the black control cable. The tubing for use with the tip-holder will not pass through the plastic ‘tunnels’ in the pre-mounted cable supports. However, to make the metallic clip easier to attach, you can push these black tubing-supports to the end of the cable.

Connecting the Tubing
Tubing connectors should be hand-tightened, tools should not be used as their use will damage the connector’s thread.

Valve Inlet Tubing
A length of Fluoride Ethylene Propylene (FEP) tubing, supplied with the 402, is used to connect the inlet port to the solvent reservoir. Screw the connector into the inlet port (left-hand) of the valve and put the other end into the reservoir. A bottle clip is provided to hold the tubing securely in the reservoir (see opposite Figure).
An alternative tubing, fitted with a connector at one end and a 20 µm filter at the other, is available as an accessory (reference 3645357).
Connect a suitable length of tubing from the waste outlet of the valve to a suitable receptacle. The valve releases liquid through the waste outlet if the pressure inside the valve is too high. For the pressure limits of the valve, refer to Appendix C, Technical Data.
This tubing (FEP) allows the transfer of samples aspirated from the probe and prevents the contami- nation of the syringe(s). The size of transfer tubing is chosen as a function of the volumes of samples, reagents, and air gaps that have to be handled via the probe. The total volume transferred must be less than the volume of the transfer tubing.
The transfer tubing connects the outlet port (right- hand) of the valve to the probe. A 2 mL tapered tubing is supplied as standard; 1 mL, 5 mL, and 10 mL are available as additional accessories.
Self-adhesive clips are provided for holding the transfer tubing securely against the side of the 402.
The 1 mL transfer tubing is recommended for trans- ferring volumes smaller than 1 mL. To avoid move- ments of the tubing that could affect the precision, the 5 mL and 10 mL transfer tubings are rolled and held by magnets on the left-side of the 402.

After assembling the probe, the other end of the tubing is installed by screwing the connector into the right-hand port of the valve. When you use Gilson Diamond Tips, the special non-tapered tubing, included in the kit, must be fitted.
Dual-syringe Model
Valve Inlet Tubing
A length of Fluoride Ethylene Propyl- ene (FEP) tubing, supplied with the 402, is used to connect the inlet port to the solvent reservoir. Screw the connector into the inlet port (left- hand) of the valve and put the other end into the reservoir. A bottle clip is provided to hold the tubing securely in the reservoir.
to probe From reservoir
An alternative tubing, fitted with a connector at one end and a 20 µm filter at the other, is available as an accessory (reference 3645357).
Connect the length of tubing (supplied as standard) to the waste outlet of the valve, and put the other end into the waste or any other suitable receptacle. The valve releases liquid through the waste outlet, if the pressure inside the valve is too high (see Appendix C, Technical Data).
This tubing has connectors at both ends and is used to connect from the valve outlet port (right-hand) to the Tee-junction inlet port (left-hand). It is made of FEP and has the following dimensions: 70 x 3 x 2 mm. It is autoclavable (20 min. at 0.1 MPa, 121 °C).
This tubing (FEP) allows the transfer of samples aspirated from the probe and prevents the contami- nation of the syringe(s). The size of transfer tubing is chosen as a function of the volumes of samples, reagents, and air gaps that have to be handled via the probe. The total volume transferred must be less than the volume of the transfer tubing.

The transfer tubing connects the outlet port (right- hand) of the Tee-junction to the probe. A 2 mL tapered transfer-tubing is supplied as standard ; 1 mL, 5 mL, and 10 mL are available as additional accessories.
Self-adhesive clips are provided for holding the transfer tubing securely against the side of the 402 (see previous page).
The 1 mL transfer tubing is recommended for transferring volumes smaller than 1 mL. To avoid movements of the tubing that could affect the precision, the 5 mL and 10 mL transfer tubings are rolled and held by magnets on the left-side of the 402.
After assembling the probe, the other end of the tubing is installed by screwing the connector into the right-hand port of the Tee-junction. When you use Diamond tips, the special non-tapered tubing, included in the kit, must be fitted.

The rear panel of the 402 Dilutor-Dispenser houses:
- sector block containing the power socket, fuse
drawer and ON/OFF switch,
- GSIOC connector.
The sector block contains the power socket, fuse holder and ON/OFF switch. Your 402 can be connected to a 100-120 VAC or a 220-240 VAC supply. In either case, two 1A type “T” slow blow fuses must be installed (reference 6730104006).
The instrument has no physical voltage selector. The 402 Dilutor-Dispenser can be connected to 100 or 240 V. Power control is carried out electronically on the power supply board.
For safety reasons, your 402 is delivered without the fuses installed. You must install two 1A type “T” slow blow fuses (reference 6730104006), as follows:

In the figure, the drawer containing the fuses is shown separated from the sector block. In fact the drawer hinges downwards; it is not advisable, nor necessary, to remove it completely from the block.

Connecting the Power Cord
Make sure that the ON/OFF switch is in the OFF (down) position. Connect the power cord to the power socket at the rear of the instrument and then to the power supply source. When the ON/OFF switch is put to the ON position, the instrument initializes; the valve motor rotates the spindle and stops with the valve set to the outlet position. The piston motor raises the piston to the upper position.

Switch off the 402 Dilutor-Dispenser, before con- necting or disconnecting the keypad.
Grip the connector, at the end of the coiled connec- tion cable, so that you squeeze the clip between your thumb and index finger. Then, plug the connector into the socket on the rear panel labeled ‘KEYPAD’ (see figure). You must grip the connector in the same way when disconnecting the keypad.
The coiled connecting cable enables you to use the keypad beside the 402 (as in Chapter 2), or even at some distance away from it. However, you can also install the keypad on top of the 402; to do this you should insert the tab on the rear support of the keypad into the slot on top of the 402 and then gently push the tab in as far as it will go.
Grip the connector, at the end of the black connec- tion cable, between your thumb and index finger. Then plug the connector into the socket on the rear panel labeled ‘PROBE’.

The 402 Dilutor-Dispenser is supplied without syringes. Order a syringe (or syringes), appropriate to your Method, for your 402 Dilutor-Dispenser. Mount the syringe on the valve or Tee-junction.

The syringe and its piston are supplied assembled. However, if you remove the syringe from the piston, ensure that the piston and the piston nose are free from dust and contamination before reinserting the piston into the syringe.
Wet the piston nose (with water), insert the piston nose into the syringe body, and then push the piston into the syringe.
Before fitting one of the smaller syringes (100 µL, 250 µL, 500 µL and 1 mL), prime it manually by placing its open end in the solvent reservoir, then aspirate using the piston. This manual prime is not necessary for the 5 mL, 10 mL and 25 mL syringes.
You must use the keypad software to lower the piston operating rod. You should therefore refer to Chapter 4, which describes the complete procedure for fitting a syringe. When changing a syringe, you should first undo the piston securing-screw in order to free the piston assembly from the piston operating-rod.

The 100 µL, 250 µL and 500 µL syringes are sup- plied with a cover seal to ensure an airtight fit between the syringe and the valve or Tee. Refer to the opposite figure for the correct orientation of the cover seal.
If you change the syringe size, don’t forget to change the declared syringe volume in your soft- ware’s system configuration.
Inlet Outlet

Upgrading to Dual-syringe Model
The single syringe 402 can be upgraded to improve the versatility of the system.
To upgrade to the Dual-syringe 402, you must install a Tee-junction module (kit reference F4015010) at the right-hand-side of the 402. You must first install the syringe module (includes motor etc.) and then fit the Tee-junction.
Before installing the module: make sure that the ON/ OFF switch is in the OFF position, remove the power cord, and disconnect any hydraulic connections from the 402.


Upgrading to Dual-syringe Model
There is a flexible flat cable inside the 402 (attached to the lower chassis with Velcro) that connects the logic board to the Valve module. Detach the cable from the lower chassis, then use the free connector to connect to the motor drive board on the upgrade module. With reference to the opposite figure, the loose-end of the flat cable (with the second connec- tor) and the connecting cable may be re-attached, using their Velcro strips.


Second connector is not used
To tee module

Inserting the Module
Do not try to force the module into position as this may affect the alignment of the component parts. You should proceed carefully as follows.

This chapter describes how to operate the 402 in standard use. Chapter 5 contains more detailed and advanced information, together with less frequently used options. You are advised to read the first section of Chapter 5, ‘Correct Use of the 402’, before proceeding.
To use your 402 Dilutor-Dispenser, you must first install the hardware (syringe, probe, keypad, etc.) suitable to your Method, and then inform the 402’s software of the installation, this is called ‘Configuration’. Then, you write and run your Method, which is stored as a file in your 402’s memory.

Essential Steps
Here is a summary of the things you must do, and the order in which you should do them.
10)Prime the Dilutor-Dispenser.
11)Run the Method.
If you have a Single-syringe or a Dual-syringe, check that you have received all the parts in Appendix A.
Connect the keypad as described in Chapter 3.
Fit the fuses and connect the power supply cord (see Chapter 3).
Assemble and connect the probe (see Chapter 3).
If you have a Single-syringe 402, install the tubings as described in Chapter 3, page 3-4. You must con- nect 3 pieces of tubing: inlet (reservoir), transfer (probe), and waste.

If you have a Dual-syringe 402, install the tubings as described in Chapter 3, page 3-5. You must connect 4 pieces of tubing: inlet (reservoir), Valve to Tee-junction, transfer (probe), and waste.
Switch-on


The syringe-piston operating-rod(s) move to their uppermost positions, while the Homing Screen is displayed on the keypad.

Four soft-touch keys are accessible from the Start-up Screen: Conf (Configuration), Edit, Man (Direct Manipulations), and File; the fifth ‘Run’ appears after you save a Method file. They are called soft- keys because they vary from screen-to-screen; you must press the white key immediately below the corresponding ‘label’ (e.g. Conf) in order to carry out the desired action.
Pressing a soft-touch key leads you to a series of screens containing menus and sub- menus that enable to perform functions related to the key pressed.
Enter the configuration of the Dilutor-Dispenser Conf
Create or modify a Method Edit
Save the Method as a file in the 402’s memory Edit
To operate your 402, you must use Conf, Man, and Run. Five prefabricated Methods (files 90-94, see Chapter 5) are supplied as part of the software; however, you will often want to use Edit to create your own Method files.
Direct manipulation of the Dilutor-Dispenser Man
Run a Method Run
File transactions (Delete, Copy, Lock, Dir.) File

Configuration (Conf)
You only have to enter the Configuration details one-time; the 402 keeps this information in its memory, even after you switch off. The next time you switch on the 402, the Configuration and all Methods previously saved as files will be available to modify or run.
The software must be given details about syringe sizes, flow rates, etc. You enter this information using the configuration function, which you access by pressing the Conf soft-key (see Configuration Screen).

As a minimum, you must enter the volume of syringe(s) that you are going to fit to your 402, check that the default flow rates are appropriate (or enter suitable values), and check the transfer (probe) tubing volume is correctly specified (standard is 2 mL).
The use of the soft-keys is as follows:
Check that the configured volume for the probe tubing is the same as the size fitted.

Press the Vol soft-key, you see the Tubing Volume Screen.
Next is used to go to the next screen, Chge is used to select the tubing size (see Tubing Volume Screen), Prev is used to go to the previous screen (see Air Gap Volume Screen).
Air Gap Volume Screen

Configuration Screen · Tubing Volume Screen

If you want to change the configured volume, press Chge (until you reach the correct value); when the volumes are the same press ENTER, press the ESC key to return to the Configuration Screen, or press Next, after which you see the Air Gap Volume Screen.
Key in the ‘Air gap volume’, press ENTER to confirm then press the ESC key to return to the Configuration Screen (see Chapter 5, for a definition of Air Gap).
You must check that the configured volumes are the same as the sizes to be fitted. Press the Syr soft-key.
The use of the soft-keys is as follows:
Next is used to go to the next screen, Prev is used to go to the previous screen, Corr is used to supply a correction factor (see Chapter 5), Chge is used to select the size(s) of syringe(s), Exch is used to lower the piston rods.
To change a syringe size, press the Chge soft-key until you reach the corresponding volume, then press ENTER to confirm. Press Exch if you are ready to lower the piston rods, or when you have two syringes, press Next. Check that the configured volume for the right syringe is the same as the size to be fitted. Press the Chge soft-key until you reach the corresponding volume, then press ENTER to confirm. Press Exch if you are ready to lower the piston rods or press the ESC key to return to the Configuration Screen.
Whenever you change the configured volume, the software reminds you to ‘Exchange Syringes’ before allowing you to quit the configuration screen. Go directly to page 4-7, which explains how the ‘Ex- change Syringes’ function works, unless you need to change flow rates, etc.
Flow rates should be specified according to the sizes of syringe and tubing fitted to the 402. Refer to Chapter 5 for more details, the maximum values are given in the Tables, these are used as default values by the software.
Press the Flow soft-key; the Air Flow Screen appears.
Syringe Screen · Air Flow Screen

Next is used to go to the next screen (see Air Flow Screen), Prev is used to go to the previous screen (see Resv flow Screen).
Air Flow is the rate in mL/min. that air is drawn into the transfer tubing when aspirating air.

Key in the required value, then press ENTER. Press the ESC key to return to the Configuration Screen or press Next to go to the Resv Flow Screen.
Resv (Reservoir) Flow is the rate in mL/min. that solvent is aspirated from the reservoir by the left syringe.
Key in the required value, then press ENTER. Press the ESC key to return to the Configuration Screen or press Next to go to the Prime Flow Screen.

Prime Flow is the rate in mL/min. that liquid is transferred by either syringe when priming.
Key in the required value, then press ENTER. Press the ESC key to return to the Configuration Screen.
Press the Etc soft-key; the Beeper Screen appears. This is to set an alarm when an error occurs, and to warn that liquid is going to be dispensed to waste.

To switch between ON and OFF, press the Chge soft-key, then press ENTER to confirm.

Press Next to go to the Autorefill Screen. Press the ESC key to return to the Configuration Screen.
When you set Autorefill to ON, the left-hand syringe re-aspirates diluent automatically for the next cycle, while you are (for example) aspirating air via the probe. This feature enables you to work faster with the 402.
To switch between ON and OFF, press the Chge soft-key, then press ENTER to confirm. Press Prev to go to the Beeper Screen. Press the ESC key to return to the Configuration Screen.

This completes the basic configuration process, press ESC to return to the Start-up Screen.

Essential Steps

correct before continuing), - press the Exch soft-key. ‘Please wait ...’ appears on the screen whilst the piston rod(s) descend; then ‘Exchange the Syringe(s)’.


Introduction
In this section you will find the basic information required to enter and save your Method as a file. The steps are as follows:
Alternatively, you could run an adapted copy of one of the five pre-fabricated files (90-94). See Chapter 5 for details.
- describe the Method on paper (protocols and
parameters), - use the keypad to enter the Method as a file (one
or more tasks used sequentially), - save the Method as a file.
Configuration Screen · Syringe Screen · Counter Screen

In-run Prompts
You can defer some parameters (volumes and flow rates) by entering a question mark (?) in place of the value. You enter the actual value after pressing the RUN key. This feature is particularly useful when optimizing parameters. In the following example, the aspirate flow rate (Asp Flow) is submitted as an unknown. Then, each time the Method is run, you can submit a different value (6, 12, 18). After which you evaluate the results and select the most appro- priate value for use in the Method.
Dispense equal volumes of various samples, rinse the probe after distribution.
Common (Fraction) Volume = 20 µL, Number of fractions = 3.
Aspirate flow rate from 6 to 18 mL/minute, dispense flow rate = 6 mL/minute. In this example ‘Aspirate flow rate’ will be submit- ted as an in-run prompt.
Extra volume = 5 µL. You aspirate an extra volume of sample, to avoid diluting the last fraction with solvent. This extra volume is sent to the Waste, after distributing the last fraction.
Rinse with 500 µL of solvent.
When using the software, remember that the ‘HELP’ key is available with most screens.

Beginning at the Start-up Screen, key in the file number for your Method, then press ENTER.
Then press Edit; the Task Selection Screen appears.

You can now access the screens for the required task and enter the task related parameters. The following screens show how to enter the informa- tion (described above) for each task.
In general you must press the Chge soft-key to select an option, then press ENTER to confirm the option selected. Press the Next soft-key to cycle forward through the screens, press Prev to cycle backwards, or ESC when you reach the extremity of a software branch.
Start-up Screen · Entering the Method · Task Selection Screen

For the Dispense task, press Disp; you see the first parameter screen, Dispense Screen (1).

Press Chge until you see ‘Source: PROBE’, press ENTER to confirm, then press Next.
Press Chge until you see ‘Fraction vol: SAME’, press ENTER to confirm, then press Next.

Key in the number of fractions (up to 99 are possi- ble) to dispense (3 in this example), press ENTER to confirm, then press Next. Key in the ‘Common vol:’ (20), press ENTER to confirm, then press Flow.
At ‘Asp. Flow’, key in a typical flow rate (6), then press the question mark (?) soft-key, press ENTER.

Press Next, key in the flow rate (6), press ENTER; Dispense Screen (5) appears.

Press ESC Dispense Screen (4a) reappears. Press Extra; Dispense Screen (6) appears.
Key in the ‘Extra vol: ’(5), press ENTER to confirm, then press ESC, Dispense Screen (4) reappears.

Press ESC, Dispense Screen (7) appears.

This completes the parameter set for the first task (Dispense). If you need to verify (or change) any of the parameters, you would press the Param key. However, in this example you continue by pressing ADD; Task Selection Screen 2 appears.
Press soft-key for required task: Rinse; you now see the Rinse Screen.
Task Selection Screen 2

Key in the rinse volume (500), press ENTER. Press ESC; Task Edit Screen appears.
Press ESC, the Save Screen appears.

Press Yes; Run Screen appears (the name of the first task to be run appears, in this example ‘Dispense’); you are now ready to prime the 402, after which you can run the Method that you have just created.


Prime the 402 (Man)
Prime is used to thoroughly clean out all fluid pathways with fresh solvent and to clear bubbles from the tubings.

Press the Man soft-key, the Manipulation Screen appears.

Press Prime, Initialization Screen 2 appears. [Measure and Titrate are described in Chapter 5.]
You should prime the 402 before each run, to ensure that all fluid pathways contain fresh solvent and no air bubbles.
Put the probe-end of the transfer-tubing into a suitable the waste container, then press OK.
Priming starts immediately and the message ‘Priming ...’ appears on the screen; during which time solvent is expelled at the probe-end of the transfer tubing.
Press STOP, or the End soft-key, when you are satisfied that all fluid pathways have been primed with fresh solvent and are free of air bubbles.
When you press End, the motion of the syringe(s) ceases and ‘Homing ...’ appears on the screen until the syringe(s) have returned to their uppermost positions.
You can press STOP at any time, however, this key is more for use if a problem arises (such as a leak).

When you press STOP, the motion of the syringe(s) ceases and the Abort Screen appears. Press Proceed, when you want to continue the prime from the point at which it was stopped (for example, to refill the solvent reservoir).
Press Abort, when you want to stop priming; Initialization Screen 2 reappears. Ensure that the probe-end of the transfer-tubing is in a suitable waste container, then press OK.
Manipulation Screen · Initialization Screen 2 · Abort Screen

‘Homing ...’ appears on the screen until the syringe(s) have returned to their uppermost (home) positions, during which time any liquid contained in the transfer-tubing is expelled from the probe-end. After this the Action Screen reappears.
Press ESC to return to the start-up screen for the current file.
Whenever you see ‘Homing’ on the screen, the syringe piston(s) are returned to their uppermost position(s). As a result, some (or all) of the liquid contained in the transfer-tubing will be expelled from the probe-end. You must ensure that the probe end of transfer-tubing is already in a suitable waste container.
Introduction
The Run soft-key appears, after you have entered your Method. Press Run to start the Method cur- rently shown on the screen (at the first task, in this example Dispense).

If you want to run a different program, you must key in its file number and press ENTER, before pressing Run.

After pressing Run, a series of in run screens appears. During the run, you must follow the instructions on the screen, and press the command button on the probe when prompted to do so by the command light on the probe.
You press the command button on the probe to aspirate and to dispense: - a steady light means you are going to aspirate or
dispense liquid, - a flashing light means you are going to aspirate air.
When you have finished, or wish to change pro- grams, press STOP, to instantly halt the program.
When you press STOP, the motion of the syringe(s) ceases and the Abort Screen appears. Press Proceed, when you want to continue with your Method.
Press Abort, when you want to stop. Unless the syringe(s) are already at their uppermost (home) positions, Initialization Screen 2 reappears.
Run Screen (1) · Run Screen (33) · Abort Screen

Ensure that the probe-end of the transfer-tubing is in a suitable waste container, then press OK. ‘Homing ...’ appears on the screen until the syringe(s) have returned to their home positions, during which time any liquid contained in the transfer-tubing is expelled from the probe-end. After this the Action Screen reappears.
Press ESC to return to the start-up screen for the current file.
Whenever you see ‘Homing’ on the screen, the syringe piston(s) are returned to their uppermost position(s). As a result, some (or all) of the liquid contained in the transfer-tubing will be expelled from the probe-end. You must ensure that the probe end of transfer-tubing is already in a suitable waste container.
This section describes how to run Method 33 (created as above).
In Run Screen (1)

(i) Press Run, you see In Run Screen 1.
M33/T 1: DISPENS/1 means Method 33 Task 1, Dispense Task, first cycle. The cursor moves from side-to-side below the parameter value field, indicating that you must enter a value (or validate the current value, in this example 6 mL/min.). Key in the required flow rate (for example, 12 mL/min.) and press ENTER. [Similar screens would appear for any other parameters that you had entered as a question mark (?).] The command light flashes and you see alternately In Run Screens 2a and 2b (about every half a second). In the Dispense task, you aspirate an air gap, then liquid. After aspirating, you dispense the liquid in several fractions, finally the ‘Extra volume’ is sent to waste.
In Run Screen (2a)

In Run Screen (2b)


Extra Volume

In Run Screen (3a)
In Run Screen (3b)
button to aspirate AIR. The command light is steady and you see alternately In Run Screens 3a and 3b.


(iii) Put the tip of the
transfer-tubing into the sample then press the command button to aspirate liquid. The command light is steady and you see alternately In Run Screens 4a and 4b.
In Run Screen (4a) In Run Screen (4b)
(iv) Put the tip of the
transfer-tubing into a recipient vessel then press the command button to dispense the fraction, the command light is steady. Repeat for the other fractions. The command light is steady and you see alternately In Run Screens 5a and 5b.
In Run Screen (5a) In Run Screen (5b)

(v)Put the tip into the
waste then press the command button to dispense the ‘Extra volume’. The com- mand light is steady and you see alter- nately In Run Screens 6a and 6b. M33/T 2: RINSE 1 means Method 33 Task 2, Rinse, first cycle.
In Run Screen (6a) In Run Screen (6b)


(vi) Put the tip of the transfer-tubing into the waste
then press the command button to rinse. The command light goes out while the tubing is being rinsed. Then, you are returned to In-run Screen 1 for the next sample. At his point you may continue to cycle through In-run Screens 1 to 6, or end the run by pressing STOP, in which case you see the Abort Screen.

(ii)Press the command · Abort Screen

Press ‘Abort’ to confirm that you wish to stop and you are returned to the Run Screen. Alternatively, you can bring the 402 to a halt by switching off the power, in which case you must ensure that the probe end of transfer-tubing is in a suitable waste container before switching on again.

This chapter explains in details how liquids are aspirated and dispensed using the 402 Dilutor-Dispenser and gives in depth details of various software functions.

Correct Use of a 402
Single-syringe Model
(a) Aspirating from Reservoir (b) Aspirating from Probe (c) Dispensing
From Probe To Probe
Piston descends to aspirate
Piston descends to aspirate
Piston ascends to aspirate
From Reservoir

(a) Left-hand Syringe Aspirating from Reservoir
Left piston descends to aspirate
Right piston remains at the top
To Tee To Probe
Left piston ascends to dispense
Right piston remains at the top
From Reservoir · (b) Left-hand Syringe · Dispensing

(c) Left-hand Syringe Aspirating from Probe
From Tee From Probe
Left piston descends to aspirate
Right piston remains at the top
(d) Right-hand Syringe Aspirating from Probe
From Probe Valve switched to Reservoir
Left piston remains inactive
Right piston descends to aspirate
(e) Right-hand Syringe Dispensing to Probe
Valve switched to Reservoir To Probe
Left piston remains inactive
Right piston ascends to aspirate

You must first enter the configuration details for your 402 Dilutor-Dispenser. Then you may do one of three things:
(i) Perform a direct manipulation (Prime, Measure, Titrate).
(ii) Run a prefabricated Method (files 90 - 94).
(iii) Create your own Method.
A Method consists of a file saved in the 402’s memory, which can hold up to 99 files. Each file will contain one or more of the following tasks: Dilute (Dil), Dispense (Disp), Pipette (Pip), Rinse, and Mix. You can use Link to combine Methods so that they operate sequentially.
Each task consists of a series of instructions and parameters. Some of the parameters may be defined as in-run prompts (entered as a question mark); in which case you are prompted to enter the variable’s value each time you run the Method. To make and run a Method file, you must follow the procedures 1 to 5.
Each procedure is activated by pressing the appropriate soft- key, each of which gives access to a specific menu. The RUN key only appears after you have saved your Method (i.e. after having entered the tasks and parameters).
1 Enter the configuration of the Dilutor-Dispenser Conf 2 Create or modify a Method Edit 3 Save the Method as a file in the 402’s memory Edit 4 Direct manipulation of the Dilutor-Dispenser Man 5 Run a Method Run 6 File transactions File
In carrying out these procedures, you must observe some general rules.
1 Press ENTER, to confirm your entry, after keying in any value or number1.
1After keying a value (or selecting an option), you must press ENTER to make the soft-keys reappear.
2 Press ENTER, to confirm your choice, after selecting any option1.
3 Press the Prev soft-key to go back to the previous screen.
Function Press Key

4 Press the Next soft-key to go to the following screen.
5 Press the Chge soft-key to choose an option or to select from a list of possible values.
6 Press the ESC key to return to the level of the previous screen or menu.
7 Press CLEAR to remove the value (before pressing ENTER) and return to the previous value.
8 Press Deflt to select the default value or option.
You have to tell the controlling software how your 402 Dilutor-Dispenser is configured (syringe sizes, flow rates, etc.). Press the CONF key; you see the Configuration Screen.

Press Key To Change What Items
Vol Volumes Transfer Tubing, Air Gap. Syr Syringe Sizes, Correction, and to change. Flow Flow-rates Air, Prime, Reservoir. Count Counters Number of cycles for syringe(s) and valve. Etc Switch On/Off Beeper and Autorefill function.

Press the Vol soft-key, you see the Tubing Volume Screen.
The use of the soft-keys is as follows:
Next is used to go to the next screen (from Tubing Volume Screen), Chge is used to select the tubing size, Prev is used to go to the previous screen (from Air Gap Volume Screen),
Check that the configured volume for the probe tubing is the same as the size fitted. For the standard probe, you can select from: 1, 2, 5, 10 mL (tapered tubing).
When using disposable tips, you must select the size of tip (Micro or Large Volume), and application type for Micro Tips only (Air or Liquid). When you use disposable tips, the ‘Extra Volume’ parameter is automatically changed to ‘Air Push Volume’.
Configuration Screen · Tubing Volume Screen

If you want to change the configured volume, press Chge (until you reach the correct value); when the volumes are the same press ENTER, press the ESC key to return to the Configuration Screen, or press Next, after which you see the Air Gap Volume Screen.
Key in the ‘Air gap volume’, press ENTER to confirm then press the ESC key to return to the Configuration Screen.
Air gaps may be used to separate liquids in the transfer tubing; the default value for ‘Air gap volume’ is 10 µl, which gives a column of air 6 mm long in the standard transfer-tubing (2 mL).
Check that the configured volumes for the syringes are the same as the sizes to be fitted. Press the Syr soft-key.

The use of the soft-keys is as follows:
Next is used to go to the next screen, Prev is used to go to the previous screen, Corr is used to supply a correction factor for the syringe travel, Chge is used to select the size(s) of syringe(s), Exch is used to lower the piston rods (see page 4-7).
Syringe Screen

When you specify left and right syringe capacities (left must be > right), you must enter compatible volumes. An error message is displayed if you try to enter an incompatible volume, in which case you may need to change the volume of the other syringe, before entering the volume of the one that provoked the error message.

When you turn the syringe correction factor ON, a plus sign (+) appears after the file number, as a reminder.

Press the Flow soft-key; the Air Flow Screen appears.
Air Flow is the rate in ml/min. that air is drawn into the transfer tubing when aspirating air.
Key in the required value, then press ENTER. Or, press Deflt to use the default value (4 mL/min).

Press the ESC key to return to the Configuration Screen or press Next to go to the Resv Flow Screen.
‘Resv asp:’ is the rate in mL/min. that solvent is aspirated from the reservoir by the left syringe.
Key in the required value, then press ENTER. Or, press Deflt to use the default value. Press the ESC key to return to the Configuration Screen or press Next to go to the Prime Flow Screen.

Prime Flow is the rate in mL/min. that liquid is transferred by the syringe(s) when priming.
Key in the required value, then press ENTER. Or, press Deflt to use the default value. Press the ESC key to return to the Configuration Screen. The flow- rate limits depend on the volume of the syringe(s) and type of tubing used, see pages 5-31 and 5-32.

Software
Under normal use you should expect to fit new (or refurbished) syringes after 30000 cycles with water. This feature is made easy by the 402, which automati- cally counts the number of operating cycles performed by the syringe(s) and valve. (A cycle is completed when the syringe leaves its home position.) You should only need to reset these counters after changing one or more of these items. To do this, press Count, the Syringe Counter Screen appears.

Press Chge, then key in the number from which you wish the new count to start, followed by ENTER. Press Next to go to the second syringe (if fitted) or to the valve.
Press Chge, then key in the number from which the new count is to start, followed by ENTER. Do this for each counter that you want to change; followed by ESC to return to the Configuration Screen.
Press the Etc soft-key; the Beeper Screen appears. This is to set an alarm when an error occurs, and to warn that liquid is going to be dispensed to waste.

To switch between ON and OFF, press the Chge soft-key, then press ENTER to confirm.
Press Next to go to the Autorefill Screen. When you set Autorefill to ON, the left-hand syringe re-aspirates diluent automatically for the next cycle, while you are (for example) aspirating air via the probe. This feature enables you to work faster with the 402.

To switch between ON and OFF, press the Chge soft-key, then press ENTER to confirm. Press Prev to go to the Beeper Screen. Press the ESC key to return to the Configuration Screen.
You use the keypad to enter the Method as a file (one or more tasks used sequentially).
Key in the file number for your Method, then press ENTER; press Edit. Now, one of two situations can arise:
(i) if you have not entered a syringe correction factor

in the configuration process, the Task Selection Screen appears, after which you press a soft-key

to access the required task and its parameters,
(ii) if you have entered a syringe correction factor in
the configuration process, the Syringe Correction Screen appears first, after which you select ON or OFF using the Chge soft-key, press ENTER, and then Next, after which the Task Selection Screen appears; you can now access the screens for the task and enter the task related param- eters (see page 5-15).
When using the Edit function to create or edit a Method, you should press the Chge soft-key to select an option, then press ENTER to confirm the option selected. Pressing ENTER moves you auto- matically to the next screen. Press the Next soft-key to cycle forward through the screens, or press Prev to cycle backwards.
When you have created, or modified your Method, you are prompted by the software as follows:

Normally you press Yes, and you are returned to the Run Screen. Other file manipulations are below.

Press the FILE soft-key, the File Menu Screen appears.
The use of the soft-keys is as follows:
Del is used to delete a Method file, Copy is used to copy a Method file, Dir is used to access the file directory, Lock is used to enter your access code.
A plus sign (+) after the file number, indicates that the syringe correction function is ON. An asterisk (*) means that the file is locked.

This action applies to the file currently selected; so to delete another file you must first key in the number of the file to be deleted and press ENTER. Press Del, the Delete Screen appears.

If you select No you are returned to the File Menu Screen.

When you select Yes, the file is deleted from the directory and you see the New-File Directory Screen.
New File Directory Screen

Press ESC to return to the New File Screen, or press Dir to access the file directory.

This action applies to the file currently selected; so to copy another file you must first key in the number of the file to be copied and press ENTER.

Press Copy, the Copy Screen appears.
Key in the file number to which the file currently selected is to be copied, press ENTER, the Copy Method Screen appears.

Normally, you select Yes, in which case you are returned to the File Menu Screen.

When you select No or Prev you are also returned to the File Menu Screen.
If the destination file already exists (and is not locked), you are asked to confirm that you wish to overwrite; select Yes and the Method previously stored in the destination file is destroyed.
The file directory allows you to visualize the file numbers currently assigned. Press Dir, the file directory for the first set of file numbers appears (1 to 9). Thereafter, a file directory exists for each decade (range of ten numbers), for example Method 10 [-1—5-78-]; in this example, files 11, 15, 17 and 18 exist. You use the Prev and Next keys to move between decades.
Files may be numbered from 1 to 99; so the last decade is Method 90. You press the ESC key to return to the File Menu Screen.

To prevent changes to a Method, you can lock the corresponding file. This action applies to the file currently in memory; so to lock another file you must first key in the number of the file to be locked and press ENTER.
Press Lock, key in your code (any number between 1 and 9999), then press ENTER, followed by ESC. The file is now locked, the Unlock File Screen appears.


To unlock this file you must key in the code that was used to lock the file. You may not edit a locked file, so the Edit soft-key is replaced by the View soft-key, which allows you to browse through the contents of that file, without making any changes.

However, you are permitted to copy a locked file. After copying the original file remains locked but the copied file is unlocked automatically by the software.
You can either perform one of three direct manipu- lations (Prime, Measure, Titrate) by pressing the corresponding soft-key. Press the Man soft-key, the Action Screen appears.

You should prime the 402 before each run, to ensure that all fluid pathways contain fresh solvent.

Press Prime, Initialization Screen 2 appears.
Put the probe-end of the transfer-tubing into a suitable waste container, then press OK.
Priming starts immediately and the message ‘Priming ...’ appears on the screen; during which time solvent is expelled at the probe-end of the transfer-tubing.
Press STOP, or the End soft-key, when you are satisfied that all fluid pathways have been primed with fresh solvent and that there are no bubbles in the tubing.
When you press End, the motion of the syringe(s) ceases and ‘Homing ...’ appears on the screen until the syringe(s) have returned to their uppermost positions.
You can press STOP at any time, however, this key is more for use if a problem arises (such as a leak).

When you press STOP, the motion of the syringe(s) ceases and the Abort Screen appears. Press Proceed, when you want to continue the prime from the point at which it was stopped (for example, to refill the solvent reservoir).

Press Abort, when you want to stop priming; Initialization Screen 2 reappears. Ensure that the probe-end of the transfer-tubing is in a suitable waste container, then press OK. ‘Homing ...’ appears on the screen until the syringe(s) have returned to their uppermost (home) positions, during which time any liquid contained in the transfer-tubing is expelled from the probe-end. After this the Action Screen reappears.
Press ESC to return to the start-up screen for the current file.
Whenever you see ‘Homing’ on the screen, the syringe piston(s) are returned to their uppermost position(s). As a result, some of the liquid con- tained in the transfer-tubing will be expelled from the probe-end. You must ensure that the probe end of transfer-tubing is already in a suitable waste container.
You can measure a volume (e.g. supernatant after centrifugation).


Press Run, then follow the instructions given on the screen; you press the command button on the probe to aspirate, when prompted to do so by the com- mand light on the probe: - a steady light means you are to aspirate (or
dispense) liquid, - a flashing light means you are to aspirate air via
the probe.
Measure Screen · Flow Rate Screen

After completely aspirating the liquid being measured, you see (for example) the Measurement Screen.

1High precision is obtained by performing this action at a low flow rate. And, because the piston accelerates after pressing the command button, you are advised to start-and-stop dispensing by repeatedly pushing the command button with an increasing frequency as the liquid nears the tip.
Conclude the current measurement by pressing STOP or End (see Prime for details).
The use of this direct manipulation is the same as classical titration (using a burette) where the idea is to run a quantity of reagent into a sample that includes a pH indicator, up to the point where the color changes. With 402, you may also use titrate to fill plates, vials, or bottles.
To titrate, you aspirate a known volume of reagent, dispense part of the volume required into the sample, agitate the sample, dispense a small amount, agitate again, dispense a smaller amount, agitate again, etc., as you move in finer and finer steps towards the point where you have added the precise volume required to change the color.
Measurement Screen

This is realized using the 402 by pressing the probe’s command button with increasing rapidity as you approach the neutralization point. Care must be taken in selecting the dispense flow rate.



dispense liquid, - a flashing light means you are to aspirate air
via the probe.
Read the volume on the screen, Titrate (4) Screen (for example), after completing each titration.

Conclude the current titration by pressing STOP or End (see Prime for details).
Your Method may consist of one or more of the following tasks. Each task consists of a pre-defined protocol, from which you select the required mode of operation, and into which you insert the required parameter values.
For optimization purposes, all volumes and flow rates, which can be entered as parameters in a task, can be defined as In-run Prompts. When editing the task, you key in a value, and then a question mark (?), the corresponding flashing value has to be validated or changed before the Method runs. It is not necessary to re-edit the method to modify the chosen parameter.

Keys The use of the soft-keys is as follows: - Prev is used to go to the previous task (or screen
within a task), - Next is used to go to the next task (or screen within
a task), - Add is used to add a task to the Method (after the
task displayed), - Param is used to access the parameters for the
current task, - Ins is used to insert a task (before the task displayed), - Del is used to remove the task currently displayed.
Parameters The following list summarizes the parameters used by the tasks; it gives the maximum and minimum values for each parameter. The maximum values are mostly unambiguous. However, in the case of volumes aspirated via the probe, the total volume of fluids (air and liquids) aspirated by a task can never exceed the internal volume of the transfer tubing. The maximum values listed for volumes aspirated via the probe are guidelines that you should follow when considering parameter values to enter. In some cases you may be able to key in a larger value but this will provoke a self explanatory error message, immediately after entry. Even then, depending on the task, you may key in a series of parameter values, for which the software will then calculate the total volume. If the transfer-tubing volume is exceeded, you will be advised of the required configuration. For example, you have configured you 402 with the standard 2 mL transfer tubing. In a task, where you can aspirate more than one liquid (e.g. sample and reagents), the total volume of all fluids adds up to 2110 µL. The software issues a warning message (for example, ‘Tubing too small’). You press the OK key to acknowledge the message, after which will you see ‘Parameters Incompatible!’ on the screen. At this point you can either press Quit and reconfigure the 402 or press OK and renter one or more of the volumes entered in the task. When you configure your 402 for use with dispos- able tips, the ‘Extra volume’ parameter is replaced automatically by ‘Air-push volume’ in the Dilute, Distribute and Pipette tasks, corresponding to the special requirements of this type of configuration; refer to pages 5-33, 5-34 for additional information.

Air-gap Volume (in Configuration) 0 Transfer-tubing Volume Dilution Ratio (Dual-syringe) 1 100000 Dilution Ratio (Single-syringe) 1 500 Number of Serial Dilutions 1 12 Number of Reagents 0 12 Number of Fractions 1 12 if volumes are different, 99 if the same. Number of Cycles (mixing) 0 12 Sample Volume (Vs) 0.1 Volume of transfer-tubing - air gap(s) Reagent Volume (Vr) 0.1 Volume of transfer-tubing - air gap(s) Diluent Volume (reservoir) 0.1 99.999 mL Diluent Volume (via probe) 0.1 Volume of transfer-tubing - air gap(s) Common Volume (reservoir) 0.1 99.999 mL Common Volume (via probe) 0.1 Volume of transfer-tubing - air gap(s) Total Volume (/) 0.1 99.999 mL Total Volume (via probe) 0.1 Volume of transfer-tubing - air gap(s) Extra Volume 0.1 100 µL Final Volume (adjust via probe) 0.1 Volume of transfer-tubing - air gap(s) Final Volume (adjust from reservoir) 0.1 99.999 mL Flow Rate - Aspirate (mL/min.) 0.01 See page 5-31 Flow rate - Dispense (mL/min.) 0.01 See page 5-32
Chaining Tasks Because of the flexibility of the software many combinations of the tasks are possible in your Method. However, it is better to avoid certain combinations of task such as Dispense followed by Dilution. The reason is that Dispense distributes the fractions into several tubes, whereas Dilute, Mix, and Rinse operate on a tube by tube basis. So, if Dilute follows Dispense, the software will first dispense the number of fractions specified (into a series of tubes), then one tube will be diluted, and then the software will start the second dispense cycle. When editing a Method containing (for example) Dispense followed by Dilution, the software issues a warning message indicating that it would be better if Dispense could be placed at the end of the chain.
Parameter Minimum Maximum

Dilute
Three types of dilution task are available ‘Normal’, ‘Unknown’, and ‘Serial’. Three modes are available for ‘Normal’ and ‘Serial’: Total Volume, Diluent Volume, and Dilution Ratio. The first action is to select the type of dilution task that you are going to use, by pressing the appropriate soft-keys.
If you specify a volume of sample greater than that of the right- hand syringe, the left-hand will be used to aspirate the sample.
You use Dilute Normal to dilute samples, with diluent from the reservoir or via the probe, and add up to 12 reagents. There are 3 modes of entering the parameters, in each mode you supply Sample Volume (Vs) plus one of the following variables: Dilution Ratio (Rd), Diluent Volume (Vd), and Total Volume (Vt). The software calculates the remaining variables, taking into account the volumes of the added reagents (Vr).
Vt = Vs x Rd, Vd = Vt - Vs - Vr For example, after selecting Dilution Ratio Mode, you enter Sample Volume = 10 and Dilution Ratio = 3, Reagent Volume (total) = 9; the software calculates the Total Volume as 30 and the required Diluent Volume as 11. In any mode, pressing the Data soft- key gives the calculated values for the other two parameters. 1. Aspirate the required volume of diluent (in opposite figure, from the reservoir). 2. Aspirate an air gap (specified in configuration). 3. Possibly, Aspirate reagent(s) then the sample (with an air gap between each liquid). 4. Dispense all into recipient vessel.
Typical Task Entry Key to press: Key in the File # (e.g. 48), Press ENTER Press Edit Press Dil Press Normal Press Chge to select source, press ENTER, then Next. Press Chge to select the mode, press ENTER,
Screens:







Press Next, then key in the sample volume, press ENTER.

Depending on the mode, you see one of three screens after pressing Next.



Key in the value, press ENTER.
Optionally, for any mode, you can display the variables calculated by the software by pressing Data. Optionally, you can modify the flow rates; press Flow, For any mode, key in a value for Asp flow, press ENTER, press Next, key in a value for Disp flow, press ENTER. Optionally, for any mode, you can supply up to 12 reagents, press Reag, key in number of reagents, press ENTER. Press Next, then for each reagent, key in volume press ENTER. After the data are entered the Reag key changes to REAG. Press ESC until you see this screen (three times if you have added reagents); Press ESC again to reveal the Save Screen; select Yes, the Method is then saved, after which you see the Run Screen.








Dilute Serially You use Dilute Serially to dilute samples, with diluent from the reservoir or via the probe. You also have the option to mix after dilution, but the addition of reagents is restricted to Dilute Normal and Dilute Unknown. There are 3 modes of entering the parameters, in each mode you supply Sample Volume (Vs) plus one of the following variables: Dilution Ratio (Rd), Diluent Volume (Vd), and Total Volume (Vt). The software calculates the remaining variables. In any mode, pressing the Data soft-key gives the calculated values for the other two parameters.
dilution 1 dilution 3 dilution 2


For each dilution in the series: Vt = Vs x Rd, Vd = Vt - Vs For example, after selecting Dilution Ratio Mode, you enter Sample Volume = 10 and Dilution Ratio = 3; the software calculates the Total Volume as 30 and the required Diluent Volume as 20. 1. When diluent comes from the reservoir (RESV); aspirate the required volume of diluent (for all dilutions, up to the capacity of the left-hand syringe). If diluent comes from the probe, an air gap is aspirated before the diluent, and the required volume of diluent is aspirated (for all dilutions, up to the capacity of the right-hand syringe, after which the left-hand syringe is used). 2. Aspirate an air gap (specified in configuration). 3. Aspirate sample. 4. Dispense all into the recipient vessel. 5. Aspirate an air gap (specified in configuration). 6. Aspirate diluted sample. 7. Dispense all into the recipient vessel. 8. Aspirate an air gap (specified in configuration). 9. Aspirate diluted sample. 10. Dispense all into the recipient vessel. 11. And so on ...
Typical Task Entry Key to press: Key in the File # (e.g. 48), Press ENTER Press Edit
Screens:



Press Chge to select source, press ENTER, then Next.

Press Chge to select mode, press ENTER,
Press Next, key in a number of serial dilutions, press ENTER. Press Next, then key in the sample volume, press ENTER.





Depending on the mode, you see one of three screens after pressing Next. Key in the value, press ENTER.

Press Dil · Press Serial

Optionally, for any mode, you can display the variables calculated by the software by pressing Data. Optionally, you can modify the flow rates; press Flow, For any mode, key in a value for Asp flow, press ENTER, press Next; key in a value for Disp flow, press ENTER. Optionally, for any mode, you can mix the diluted liquid; press Mix. The volume used for mixing is 90% of the diluent volume. Key in number of cycles, press ENTER. After the data are entered the Mix key is shown as MIX. Press ESC Press ESC again to reveal the Save Screen; select Yes, the Method is then saved, after which you see the Run Screen.





Dilute Unknown You use Dilute Unknown to dilute an unknown sample volume, with diluent from the reservoir or via the probe, and add up to 12 reagents. The Dilution Ratio mode is used to enter the param- eters, in this mode you supply the Dilution Ratio, the Sample Volume being measured when you press Run. The software calculates the Total Vol- ume and the required Diluent Volume.
Typical Task Entry Press ENTER Press Edit Press Dil Press Unknown Press Chge to select source, press ENTER, then Next. Optionally, you can modify the flow rates; press Flow, For any mode, key in a value for Asp flow, press ENTER, press Next; key in a value for Disp flow, press ENTER. Key in the ratio, press ENTER. Press Next.



Optionally, you can supply up to 12 reagents, press Reag, key in number of reagents, press ENTER. Press Next, then for each reagent, key in volume press ENTER. After the data are entered the Reag key changes to REAG. Press ESC until you see this screen (three times if you have added reagents) Press ESC again to reveal the Save Screen; select Yes, the Method is then saved, after which you see the Run Screen.
Run Procedure Press Run Select the syringe (for the Dual-syringe 402 only), press the Chge soft-key to select LEFT or RIGHT, press ENTER to confirm. You are advised to select the left-hand syringe if the volume to be measured seems to be significantly larger than the volume of the right-hand syringe. The maximum volume that you can measure is limited to the volume of the aspirating syringe or the transfer tubing size, whichever is the smaller. Press Next, Key in the flow rate at which you want the volume measurement to take place, and then press ENTER. The default flow rate is half the maximum possible for the combination of syringe and transfer in use (refer to pages 5-31, 5-32). Press Run, press the command button, aspirate the air gap: Insert the tip of the tubing into the liquid to be measured press the command button once to start the measurement and a second time to stop. Refer to page 5-12 for details of the measuring process. At the end, the measured volume is shown on the screen. At his point you can either aspirate diluent from the reservoir or via the probe. When you aspirate from the reservoir, the diluent volume and the measured volume are chained together in the transfer tubing and are dispensed together. If you use the probe, or if reagents are added, the liquids are dispensed in three steps: (i) you dispense the measured sample, (ii) you aspirate the diluent, reagents, and mea-




sured sample, (iii) you dispense all of the above into a vial.

If for any reason you want to repeat the measure- ment (e.g. you need to select the LEFT syringe instead of the RIGHT), put tip of the transfer- tubing back into the original container, then press Redo to discharge the liquid. When you are ready to dilute the previ- ously measured sample, press End. The command light is steady and you see alternately the next pair of screens. Press the command button, diluent is drawn from the reservoir; the command light is steady and you see alternately the next pair of screens. Put the tip of the transfer-tubing back into a suitable recipient vial then press the command button to dispense the diluted sample. After which you are returned to the following screen for the next cycle. If you want to quit at this point you must press the STOP key.





This task is used to dispense liquid from the reservoir or from the probe, in equal or un- equal fractions, into one or more vessels. You may adjust the final volumes in the recipient vessels to a predetermined final volume, with liquid via the reservoir or the probe. If the source is the reservoir, you can dispense a volume greater than the capacity of the left- hand (or only) syringe, which will complete this operation with several strokes of the piston.
Adjust 2 Adjust 3
Fraction 1 Fraction 2 Fraction 3

When you use this task to aspirate and dispense via the probe, you may also aspirate an ‘Extra volume’. The ‘Extra volume’ is aspirated to ensure that the accuracy of the first volume of liquid is maintained (when it is dispensed) and that the first fraction is not contaminated by liquid coming from the reservoir. The ‘Extra volume’, which is aspirated with the first
Final Volume · Adjust 1 · Extra Volume

volume of liquid, is dispensed, at the end of the process, into the waste.
When you configure disposable tips in air mode, ‘Extra volume’ is replaced by ‘Air push volume’ in this task.
Typical Task Entry Key in the File # (e.g. 45), ENTER. Press Edit.


Disp, then press Chge to switch source. Now, you may select one of two modes, where: (a) Fractions are equal, (b) Fractions are different. Press Chge to change modes, then ENTER, Next.



Key in number of fractions, then ENTER, Next. For SAME mode, key in ‘Common volume’, for DIFFERENT mode, key in volume for fraction 1.





Press Next, key in volume for fraction 2, Next, etc. Press ENTER, Next. Optionally press Extra (only if source is probe), Key in ‘Extra volume’, press ENTER, ESC. At the last volume screen the Adj key replaces Next. Optionally, press Adj. Press Chge to switch source, ENTER, then Next. Key in the final volume, ENTER, ESC. After the data are entered the Adj key is shown as ADJ. Optionally at the last of the volume screens, press Flow.


Key in a value for Asp flow, press ENTER. Press Next. Key in a value for Disp flow, press ENTER.



Press ESC (twice). Press ESC.


Pipette
You use this task to aspirate a series of fractions via the probe, then dispense the total volume aspirated into a recipient vessel. Each liquid aspirated is sepa- rated by an air gap (if air gap is configured as > 0).
You may also aspirate an ‘Extra volume’ to ensure the accuracy of the first fraction. This ‘Extra volume’, which is aspirated with the first fraction, is dispensed at the end of the process, into the waste.
When you configure disposable tips in air mode, ‘Extra volume’ is replaced by ‘Air push volume’ in this task.

1 air gap 2 fraction 1 3 air gap 4 fraction 2
Key to press Key in the File # (e.g. 45), ENTER Press Edit, Press Pip, then select one of two modes, where: (a) Fractions are equal, (b) Fractions are different. Press Chge to switch modes. Press ENTER, then Next. Key in number of fractions, press ENTER, then Next. For SAME mode, key in common volume, for DIFFERENT mode, key in volume for fraction 1, press Next, key in volume for fraction 2, Next, etc. Press ENTER, then Next.

Optionally at the first of the volume screens, press Flow.
Key in a value for Asp flow, press ENTER. Press Next. Key in a value for Disp flow, press ENTER. Optionally at the first of the volume screens, press Extra, Key in ‘Extra volume’, press ENTER.

Press ESC (twice). Press ESC.
The purpose of this task is to minimize carry-over by flushing the transfer-tubing and syringes with solvent from the reservoir, between tasks. The syringe, most appropriate to the volume specified, is used to pump a volume of liquid from the reser- voir through the tubing.
Typical Task Entry Key in file # (45), ENTER. Press Edit. Press Rinse.

Key in volume, ENTER.

Press ESC. Press ESC.

You use this task to mix either by dispensing a predetermined volume of air through the liquid or by aspirating the liquid from a vial and dispensing it back into the same vial several times.
You are advised against using the air mixing tech- nique when the liquids to be mixed are oxidizable or surface active.
Typical Task Entry Key to press: Key in file # (44) ENTER. Press Edit. Press More.


Press Mix, then select the mixing mode (air or asp/ disp).







The purpose of this task is to join Methods so that they will be executed serially.
What it Does Link creates a file (Link File) in which the file numbers corresponding to the Methods to be linked are placed. When you press Run the Methods are executed in the order in which they were entered in the Link File.
Example You want to create Link File #44, which is to contain 4 Methods:
#7 Dilute, #8 Rinse, #9 Mix, #8 Rinse. Key in Link File number (44), press ENTER, press Edit. The only criterion is that the number of the Method file to be linked must exist. If it does not exist, you will see the error message ‘Method does not exist’. At the Task Menu Screen press More, press Link. Key in the file number of the first Method to link (7), press ENTER, press Next.

Key in the file number of the second Method to link(8), press ENTER, press Next. Key in the file number of the third Method to link(9), press ENTER, press Next.
Key in the file number of the fourth Method to link(8), press ENTER, press Next. After a file is linked in this manner it can not be deleted. Press ESC to access the Save Screen, press Yes.
Typical Task Entry Key to Press: File # (44) ENTER
Edit More Link File # (7) ENTER Next File # (8) ENTER Etc until all file numbers have been entered then: ESC




Five prefabricated Methods, to run or to adapt, are present in the 402’s memory as file numbers 90 to 94. Although the tasks are the same, the parameters vary according to whether you have a Single- syringe or a Dual-syringe 402.
Configuration Standard equipment with a 2 mL transfer tubing. Left-hand (or only) syringe: 5 mL. Right-hand syringe (if fitted): 500 µL. Default values are used for parameters not listed.
Value for use with File Number TASK/Parameter Dual-syringe Model Single-syringe Model
Method 90 Dilute Normal Dilution Ratio 10 10 Sample Volume 200 1000 Rinse 300 1500
Method 91 Dilute Serially Total Volume 3000 3000 Sample Volume 100 1000 Nb of Dilutions 3 3

Value for use with File Number TASK/Parameter Dual-syringe Model Single-syringe Model
Method 92 Dispense Sample Source Reservoir (RESV) Reservoir (RESV) Fractions Same Same Nb of Fractions 3 3 Common Volume 1000 1000
Method 93 Dispense Sample Source Probe Probe Fractions Different Different Nb of Fractions 3 3 Fraction 1 50 500 Fraction 2 100 700 Fraction 3 200 500 Adjust Source Reservoir (RESV) Reservoir (RESV) Final Volume 1000 1000
Method 94 Pipette Fractions Different Different Nb of Fractions 4 4 Fraction 1 50 300 Fraction 2 80 400 Fraction 3 100 500 Fraction 4 200 600
The Run soft-key appears, after you have entered your Method. Press Run to start the Method currently shown on the screen. If you want to run a different program, you must key in its file number and press ENTER, before pressing Run. In either case a series of in run screens appears, starting with the in-run prompts (if any).
Values for any parameters defined as in-run prompts (using a question mark) must be confirmed, or changed, before the Method runs.
After pressing Run, the software prompts you for action: you must press ENTER to confirm the value shown on the screen, or to change it, key in a new value and then press ENTER.
When the Method runs you must follow the instruc- tions on the screen. You press the command button on the probe to aspirate and to dispense, when prompted to do so by the command light on the probe: - a steady light means you are to aspirate or dispense
liquid, - a flashing light means you are to aspirate air via
the probe.

When you have finished, or wish to change programs, press STOP, to instantly halt the program. Alterna- tively, you can bring the 402 to a halt by switching off the power, in which case you must ensure that the probe end of transfer-tubing is in a suitable waste container before switching on again.

When you press STOP, the motion of the syringe(s) ceases and the Abort Screen appears. Press Proceed, when you are ready to continue with your Method.
Press Abort, when you want to stop; Initialization Screen 2 reappears. Ensure that the probe-end of the transfer-tubing is in a suitable waste container, then press OK. ‘Homing …’ appears on the screen until the syringe(s) have returned to their uppermost positions, during which time any liquid contained in the transfer-tubing is expelled from the probe-end.
Whenever you see ‘Homing’ on the screen, the syringe piston(s) are returned to their uppermost position(s). As a result, some (or all) of the liquid contained in the transfer-tubing will be expelled from the probe-end. You must ensure that the probe end of transfer-tubing is already in a suitable waste container.

Flow Rates
Flow rates depend on the nature of the liquid (viscosity), the capacity of the syringe, and the internal diameter of tubing. For liquids transferred via the probe, it is possible to adjust the aspirate and dispense flow rates according to the nature of the liquid (e.g. viscous liquids must be aspirated slowly). The default value for ‘Air flow rate’ is 4 ml/min. Whatever the capacity of the syringe(s).
Syringe size Flow rate
100 µL 4 250 µL 10 500 µL 20 1 mL 40 5 mL 120 10 mL 120 25 mL 120
In the configuration menus, you can modify the air flow rate and the flow rates of liquids coming from the Reservoir (Prime flow rate and Reservoir aspirate flow rate).
In the Flow menu of each task, you can adjust the flow rates of liquid aspirated via the probe. Maxi- mum values (see tables) are given according to syringe capacity and internal diameter of the tubing. These maximum values have been opti- mized without air gap. When using air gap, it is recommended that you decrease these values.
Flow rates may be optimized using in-run prompts: when you enter a question mark in the place of a value (in task edit), that value has to be validated before the run. Optimization of flow rates is essential to assure maximum rapidity and volu- metric precision. The default values used by the software are shown in bold in the tables that follows.
When using Titrate or Measure, you are advised to halve the maximum values shown in the tables.
Maximum Aspirate Flow Rate
Syringe Volume (µL) Tubing Type 100 250 500 1000 5000 10000 25000
1 ml 4 10 20 40 120 120 120 2 ml 4 10 20 40 100 100 100 5 ml 4 10 20 40 80 80 80 10ml 4 10 20 30 30 30 30 Micro Tips (liquid) 4 10 20 40 120 120 120 Micro Tips (air) 4 10 20 40 120 120 120 Large Tips (air) 4 10 20 40 120 120 120
Configuration Flow Rates · (mL/min) · (No Air Gaps)

Maximum Dispense Flow Rate (No Air Gaps)
Syringe Volume (µL) Tubing Type 100 250 500 1000 5000 10000 25000
1 mL 4 10 20 40 120 120 120 2 mL 4 10 20 40 120 120 120 5 mL 4 10 20 40 120 120 120 10 mL 4 10 20 40 120 120 120 Micro Tips (liquid) 4 10 20 40 120 120 120 Micro Tips (air) 4 10 20 40 120 120 120 Large Tips (air) 4 10 20 40 120 120 120
Recommended Flow Rate with an Air Gap (10 µL)
Flow rates are the values entered in Flow menus (Asp Flow or Disp Flow).
Syringe size Flow rate
100 µL 4 250 µL 10 500 µL 20 1 mL 30 5 mL Asp: 30 - Disp: 50 10 mL Asp: 30 - Disp: 50 25 mL Asp: 30 - Disp: 50
Example When using the 2 mL tapered transfer-tubing, a 5 mL syringe at the left-hand side and a 500 µL syringe at the right-hand side, the default aspirate flow rates are respectively 100 mL/min and 20 mL/min. The dispense flow rate, accessible by pressing Deflt soft-key, is 120. The implicit flow rate of the small syringe is 20 mL/min. If the value of the flow rate is between 120 and 20, the flow rate of the left-hand syringe takes this value and the flow rate of the right-hand syringe is 20. To reduce to the flow rate of the right-hand syringe, it is necessary to enter values below 20 mL/min.
(mL/min)

Additional Information
A correction factor may be applied to all volumes of liquid to compensate for the effects of temperature (for example, when the 402 is used in a culture room) or for the effects of viscosity. This factor should be calculated from gravimetric assays and entered in the Configuration menu. With the Dual-syringe model, you may apply different corrections to each syringe.
When the syringe correction is switched on, a plus sign (+) appears after the Method number. All volumes in all the tasks are affected by this correc- tion, when it is switched on. At the start of a task edit, a message appears to inform you of the status (Syringe Correction Screen, see page 5-8).
A correction factor of 1% (for example) under the same condition of temperature and pressure in- creases the volume of liquid aspirated or dispensed by this amount.
The correction allows you to increase the volume transferred, for example: 250 µL syringe, nominal value is 100 µL, the actual value (after gravimetric measurements) is 99 µL. In this case, a 1% correction results in a nominal value of 101 µL, and an actual value of 100 µL. The application of a 1 % correction factor effectively reduces the capacity of the syringe by 1 %.
You should follow the procedure for gravimetric analysis described in the Operational Qualification (OQ) document that accompanies this User’s Guide. Check the accuracy and RSD of the volumes with the liquid to be transferred under the conditions of pressure and temperature of the working environment.
Using the disposable tip kit, Gilson D200 tips or Gilson D1000 tips can be mounted at the end of the probe. Two modes are available: air and liquid.
The use of 402 in ‘Air Mode’ is, for those users familiar with Gilson’s pipettes, the same as that of ‘Pipetman’.

Gilson D200 tips are used for volumes from 1 µL to 78 µL (with a 100 µL syringe, minimum recommended air-push-volume is 20 µL). Gilson D1000 tips are used for volumes from 50 µL to 900 µL (minimum recommended air-push-volume is 100 µL).
The liquid is ejected with an Air push volume of at least 20 µL when using a Gilson D200 tip and 100 µL when using a Gilson D1000 tip. An automatic software correction of 2% of the piston stroke is made to improve accuracy with Gilson D200 and D1000 tips. For example, when using a Gilson D200 tip, if an Air push volume of 20 µL is chosen, the maximum sample volume will be 78 µL; this volume will be handled with the 100 µL syringe. For the dual syringe model, the left syringe operates if the volume exceeds 78 µL; for the single syringe model, an error message is displayed.
A disposable tip should not be fitted, when priming the 402 in ‘Air Mode’. Droplets of liquid adhering the tubing after purging must be wiped from the end, before fitting a tip.
To minimize carry-over, dilution can be performed with one or more tips. If one tip is used, you use the Dilute task where the diluent and the sample are separated by an air gap (volume defined in the Configuration menu). When several tips are used, you use multiple runs of the Dispense task to transfer the liquids, changing the tips where appro- priate: one to aspirate and dispense the diluent, one to aspirate and dispense each sample.
You can use this mode in the same way as the standard tip with tapered tubing. However, it is recommended not to use an air gap, or at least to use it in a way that keeps the air gap in the disposable- tip. The use of Gilson D200 tips does not adversely affect the performance of the 402.
The Gilson D200 tip is installed to protect the end of the probe against corrosive chemicals, and by changing the tips within a task.
A disposable tip should be fitted, when priming the 402 in ‘Liquid Mode’. The following flow rates are suggested when priming: small syringes up to 10 mL/ min, larger syringes at least 80 mL/min.

This chapter explains how to perform the routine maintenance procedures.
The only maintenance operations necessary for the 402 DD are:

Cleaning
Cleaning the Exterior
The instrument cover and front panel may be cleaned. Ensure that the instrument is switched OFF and the power cord is removed when you clean the instrument.
When you need to clean the instrument, use one of the three following methods:
If a cloth dampened with soapy water is used to clean the instrument, only domestic soap may be used. No other form of detergent or chemical may be used.
Parts in the liquid flow path may be cleaned with ethanol or bleach (10 % Sodium Hypochlorite).
The surfaces of the ceramic stator and rotor inside 402’s valve are treated to give optimum airtight operation. If the valve has been used with alcohol based liquids, and then left unused for a while, a grinding noise may be heard during valve switching the next time the pump is used. This is because of friction between the ceramic surfaces. To correct this problem, you should disassemble the valve and clean the ceramic surfaces with a damp cloth.
The valve must be cleaned after using the 402 with corrosive or hazardous liquids, if leakage has occurred, or if liquid flows back into the waste tubing, or if the dilutor has been left unused for a while after pumping alcohol based liquids. The valve must be dis-mounted from the 402 and separated from the syringe, then disassembled. The valve can then be cleaned or autoclaved. It is not necessary to autoclave the valve axle or the spring.
Hold the valve body firmly in one hand, undo the valve axle guide, using a 17 mm open ended spanner. Separate the axle guide from the body of the valve.


Spring Ceramic and PEEK rotor
The Tee junction is a single piece. Apart from dis- mounting it from the 402, there is no disassembly procedure. The Tee junction can be cleaned or autoclaved.

Cleaning the Syringe Assembly
The 25 mL syringe is not autoclavable, all of the other size syringes are autoclavable.
You must dismount the syringe and piston assembly from the instrument and then separate the syringe and piston assembly before cleaning or autoclaving them.
11)Tighten the piston holding screw so that the piston assembly is secured against the piston operating rod.

Certain parts of the 402 can be autoclaved (20 min. at 0.1 MPa, 121° C). They must be removed from the instrument for autoclaving.
The autoclavable parts are:
The tubing can be cleaned or autoclaved.
This procedure prevents the end of the tubing from deforming. It is only necessary for the first time you autoclave the tubing.
Precautions to take when autoclaving the tubing:
Union connector Transfer tubing
Tubing connector · Seal (threaded)

Changing the Syringe Piston Assembly
100 µL piston kit F4015072
250 µL piston kit F4015073
For all syringe sizes, it is necessary to replace the piston assembly after approximately 30 000 cycles (using water).
500 µL piston kit F4015074
1 mL piston kit F4015075
5 mL piston kit F4015076
10 mL piston kit F4015077
The correct reference piston kit must be ordered:
25 mL piston kit F4015078
Item Part Number

The piston nose should be replaced after the pump has performed approximately 30 000 cycles. All piston noses are replaceable, the process is relatively easy, but for most pistons (except the 25 mL pistons), special tools are required to mount the piston nose.
100 µL Piston: F4015062 5 x piston noses, 2 x support collars and mounting tool (for 100 and 250 µL pistons). F4015072 1 piston assembly kit
250 µL Piston: F4015063 5 x piston noses and mounting tool (for 100 and 250 µL pistons) F4015073 1 piston assembly kit
1 mL Piston: F4015065 5 x 1 mL piston noses and mounting tool F4015005 1 nose mounting tool
F4015062 5 x piston noses, 2 x support collars and mounting tool (for 100 and 250 µL pistons)
F4015063 5 x piston noses and mounting tool (for 100 and 250 µL pistons)
To optimize the fitting of the 100 µL and 250 µL syringes on the syringe pump valve, each syringe is supplied with a cover seal. The orientation of this seal must be respected to ensure that the connec- tion with the syringe is airtight. Figure page 6-10 shows the correct orientation of the cover seal on a syringe.
Part Number Description · Part Number Description



Changing the Syringe Piston and Piston Noses
F4015064 5 x 500 µL piston noses and mounting tool kit
Piston nose holding tool

Before Mounting a 500 µL Syringe: To optimize the fitting of the 500 µL syringe on the Dilutor valve, the 500 µL syringe is supplied with a cover seal. The orientation of this seal must be respected to ensure that the connection with the syringe is airtight. (See the opposite figure).
You need the special tool for mounting the piston nose on the 1 mL piston. The nose push-fits onto the piston shaft.
F4015065 5 x 1 mL piston noses and mounting tool.
F4015005 1 nose mounting tool.

Piston nose holding tool

Changing the Syringe Piston and Piston Noses
You don’t need a special tool for mounting the piston nose on the 25 mL piston. The nose screw- fits onto the piston shaft.
F4015068 1 x 25 mL piston nose.
The piston nose mounting kits are composed of a piston nose holder and a piston shaft holder.
F4015003 5 mL piston nose mounting kit.
F4015004 10 mL piston nose mounting kit.
F4015066 1 x 5 mL piston nose and O-ring.
F4015067 1 x 10 mL piston nose and O-ring.
Part Number Description · Part Number Description


Piston shaft holder Piston nose holder O-ring
Piston shaft Piston nose

This chapter explains how to diagnose possible problems and how to overcome them.

Mechanical Problems
Problem Possible Cause Solution
Valve does not fit on to the 402. Valve axle not properly aligned Ensure that the flat on the axle with the motor drive. is aligned with the drive slot on the motor.
Valve moves during aspirate or Valve mounting screw not tight. Check that the mounting screws, dispense. syringe and syringe mounting screw are all tight.
Syringe does not move properly. Tubing blocked. Check that the inlet tubing or transfer tubing is not blocked.
Valve sticks Instrument left unused. Remove the valve (see page 6-2). Use the valve key, reference F123674 to turn the valve axle and turn the key to free the valve. Do not try turning the axle with your fingers, it is not possible to apply the necessary torque.


Problem Possible Cause Solution
Liquid leakage Connections incorrectly made. Tighten all tubing connections.
Liquid leaving by the Tubing blocked. Check all tubing for blockages. waste tube. Ensure that the inlet filter is clear.
Dirt trapped between the ceramic Clean the valve. Refer to page 6-2. surfaces.

Electrical Problems
Problem Possible Cause Solution
Instrument does not Power cord not connected. Check power cord is properly connected. respond.
402 is not switched on. Check ON/OFF switch.
Fuse blown. Check the fuses. See page 3-7.

Parts lists for the 402 Dilutor Dispenser, consisting of Standard Equipment, Additional Accessories, and Spare parts.
Appendix A-1

402 Single Syringe
Part Number Qty Description
F410514 1 402 Dilutor Dispenser (Single-syringe)
F4015016 1 Probe (standard)
F40151403 1 Probe support
4011534125 1 Fixing screw for probe support
F4015032 1 Tubing, tapered transfer (probe), FEP (2 mL, 1300 x 1.5 mm id)
365118101 1 Clip for reservoir bottle
F1410108 2 Plastic seals for autoclaving (PVDF)
F1410153 2 Union connector for autoclaving (PVDF)
F4420577 1 Tubing, waste, FEP (200 x 3 mm id)
49967601 1 Tubing, inlet, FEP (600 x 1.5 mm id)
7080316106 1 Power cord 220-240 V
7080316105 1 Power cord 100-115 V
6730104006 5 1.0 Amp type T slow-blow fuses
54080181 2 Clips for tubing, self-adhesive
LT801310 1 402 Dilutor Dispenser User’s Guide
LT801311 1 402 Dilutor Dispenser Quick Guide cards
LT804021 1 IQ Procedure
LT804035 1 Quality Control Certificate
LT804046 1 OQ Procedure
130221 1 Keypad · Appendix A-2

Part Number Qty Description
F410515 1 402 Dilutor Dispenser (Dual-syringe)
F4015016 1 Probe (standard)
F40151403 1 Probe support
4011534125 1 Fixing screw for probe support
F4015032 1 Tubing, tapered transfer (probe), FEP (2 mL, 1300 x 1.5 mm id)
F4015002 1 Tubing, valve to Tee junction (70 x 2 mm id)
365118101 1 Clip for reservoir bottle
F1410108 2 Plastic seals for autoclaving (PVDF)
F1410153 2 Union connector for autoclaving (PVDF)
F4420577 1 Tubing, waste, FEP (200 x 3 mm id)
49967601 1 Tubing, inlet, FEP (600 x 1.5 mm id)
7080316106 1 Power cord 220-240 V
7080316105 1 Power cord 100-115 v
6730104006 5 1.0 Amp type T slow-blow fuses
54080181 2 Clips for tubing, self-adhesive
LT801310 1 402 Dilutor Dispenser User’s Guide
LT801311 1 402 Dilutor Dispenser Quick Guide cards
LT804021 1 IQ Procedure
LT804035 1 Quality Control Certificate
LT804046 1 OQ Procedure
130221 1 Keypad · Appendix A-3

Additional Accessories
F4015015 Adapter kit for disposable tips 25025341 100 µL syringe
25025342 250 µL syringe
25025347 500 µL syringe
25025343 1 mL syringe
25025344 5 mL syringe
25025345 10 mL syringe
25025346 25 mL syringe
F4015031 Tubing, tapered transfer (probe), FEP (1 mL), (1 x 2 x 1300 mm)
F4015033 Tubing, tapered transfer (probe), FEP (5 mL), (1.5 x 2.1 x 3000 mm)
F4015034 Tubing, tapered transfer (probe), FEP (10 mL), (1.5 x 2 x 6000 mm)
3645357 Tubing, inlet FEP (640 x 2 mm id) with filter (20 µm)
F1410050 5 PVDF union connectors
F4015010 Tee junction upgrade kit
Part Number Description · 3651181 Tubing clip · Appendix A-4

F4015072 100 µL piston kit
F4015073 250 µL piston kit
F4015074 500 µL piston kit
F4015075 1 mL piston kit
F4015076 5 mL piston kit
F4015077 10 mL piston kit
F4015078 25 mL piston kit
Part Number Description · Appendix A-5

Appendix A-6

This chapter contains information about the chemical compatibility of the syringes and the tubing.
Appendix B-1

Chemical Compatibility
Materials in Contact with Liquids
In the 402 Dilutor Dispenser flow-path, several materials are in contact with liquids:
FEP Inlet tubing, Transfer tubing PTFE/Ekonol Piston nose of the syringes (except 25 mL) PTFE Piston nose of the 25 mL syringe
Borosilicate glass Syringe Ceramic Valve PEEK Valve, Tee junction
The syringe assemblies and the tubings are compatible with a wide range of liquids:
Acetic acid (glacial) Acetone Acetonitrile Ammonium hydroxide
Ethyl acetate Ethylene glycol Hexane Hydrochloric acid - 37%
Benzene Benzyl alcohol Boric acid Butanol
Isooctane Methanol Nitric acid - 50% Propanol
Carbon tetrachloride Chlorobenzene Chloroform Cyclohexane
Sodium hydroxide - 30% Sodium hypochlorite -10% Sulfuric acid - 96% Toluene
Cyclohexanone Dichloroethane Dichlorobenzene Diethyl ether
Trichloroethane Trichloroethylene Tetrahydrofuran Toluene
Dimethylformamide Ethanol · Xylene · Appendix B-2

This chapter contains the Technical Data concerning the 402 Dilutor Dispenser.
Appendix C-1

Hydraulics
Dilution Ratio
When using the dual syringe model equipped with a 25 mL syringe on the left and a 100 µL syringe on the right, the dilution ratio is 1/25000, but in order to stay within the RSD range a maximum ratio of 1/500 is recommended.
The minimum programmable transfer volume is 1 µL. However, precise and accurate transfer of such volumes depends on the following parameters:
The maximum volume that can be aspirated from a vial is limited to the capacity of the syringe or on the transfer tubing, whichever is the smaller. The total aspirated volume (air gaps plus liquid volumes) must be inferior to this volume.
Flow rates depend on the nature of the liquid (viscosity), the capacity of the syringe, and the internal diameter of tubing. For liquids transferred via the probe, it is possible to adjust the aspirate and dispense flow rates according to the nature of the liquid (e.g. viscous liquids must be aspirated slowly). The default value for ‘Air flow rate’ is 4 mL/min. Wathever the capacity of the syringe(s).
Appendix C-2

Hydraulics
Syringe capacity (µL) 100 250 500 1000 5000 10000 25000
Maximum Flow (mL/min) 4 10 20 40 120 120 120
Minimum Specified Volume (µL) 10 25 50 100 500 1000 2500
Nominal Volume (µL) 100 250 500 1000 5000 10000 25000
The following values are obtained using a defined protocol. If this protocol is followed, the obtained values must be inferior or equal to the listed figures.
The values are obtained by pipetting de-ionized water (grade 3, ISO3696) at 21 °C ± 0.5 °C, relative humidity > 50 % and athmospheric pressure 1013 MPa.
Syringe Volume Systematic error Random error capacity (µL) pipetted (µL) % µL % µL
10 1.8 0.18 0.8 0.08 100 50 1.5 0.75 0.25 0.125 100 1 1 0.15 0.15
20 1.5 0.38 0.8 0.2 250 125 1.0 1.25 0.25 0.32 250 0.8 2 0.15 0.38
50 1.5 0.75 0.50 0.25 500 250 1.0 2.5 0.25 0.63 500 0.8 4 0.15 0.75
100 1.5 1.5 0.50 0.5 1000 500 0.8 4 0.25 1.3 1000 0.6 6 0.15 1.5
500 1.2 6 0.5 2.5 5000 2500 0.8 20 0.25 6.3 5000 0.6 30 0.15 7.5
1000 1.2 12 0.5 5 10000 5000 0.8 40 0.25 12.5 10000 0.5 50 0.15 15
2000 1.0 25 0.50 5 25000 12500* 0.8 100 0.25 31.3 25000* 0.5 125 0.15 37.5 * Dispensing mode only.
Configuration: 1 mL tubing for volumes up to 1 mL, tapered transfer-tubing of 2 mL (standard accessory) for volumes up to 2 mL, 5 mL or 10 mL tapered transfer tubings are used for volumes up to 10 mL.
Transfer of de-ionized water using the Pipette Task, aspiration of liquid from a vial and delivery at the same flow rate into another vial containing water.
Appendix C-3

The same specifications are obtained in the dispensing mode with liquid from the reservoir.
See OQ Document for procedure and calculations.
For 1 % syringe capacity, typical values are given.
Syringe Volume Accuracy RSD capacity (µL) pipetted (µL) % µL % µL
100 1 5.0 0.05 3.0 0.03
250 2.5 3.0 0.075 2.0 0.05
500 5 3.0 0.15 2.0 0.1
1000 10 3.0 0.3 1.0 0.1
5000 50 3.0 1.5 0.8 0.4
10000 100 3.0 3 0.5 0.5
25000 250 3.0 7.5 0.5 1.25
Typical values (all syringes)
Syringe capacity Max. accuracy Max.error imprecision
100 0.50 0.10 50 0.80 0.20 10 1.00 0.50
(between valve and Tee)
Material: FEP, with two PVDF union connectors.
Tubing dimensions: 2.0 mm i.d., 70 mm length.
Tubing dead volume: 220 µL.
Appendix C-4

Valve
Type: Three port, two position.

Angle between ports: 120°.
Valve drive: Stepper motor.
Rotation time: 0.5 Second.
Liquid contact materials: PEEK and ceramic.
Lifetime: 1 000 000 cycles.
Dead volume: 123 µL.
Maximum Pressure: 0.8 MPa.
Type: Three port, no moving parts.
Diameter of channels: 1.5 mm.
Length of channels: 14 mm.
Angle between channels: 120°.
Liquid contact material: PEEK.
Dead volume: 51 µL.
Cross section of Tee
Piston drive: Stepper motor.
Piston displacement control: Coded disk.
Piston movement resolution: 1.6 µm.
Minimum piston travel time: 1.5 seconds for one complete stroke.
Maximum syringe displacement: 60 mm ± 20 µm.
Minimum syringe speed: 60 mm in 10 hours.
Maximum syringe speed: 60 mm in 1.5s.
Appendix C-5

Environmental
Conformity to Norms
Refer to separate leaflet LT801354 ‘Declaration of Conformity’.
Indoor use only.
Installation: Category II.
Altitude: operating up to 2000 m. storage up to 4600 m.
Temperature range: operating 4°C to 40°C. storage -15°C to +40°C.
Pollution degree 2.
Humidity: Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C.
Frequency: 50 to 60 HZ.
Voltage: 100 to 240 V ac, mains supply voltage fluctuations should not exceed ±10% of the nominal voltage.
Power rating: 110 VA.
Dimensions: (W x D x H) 70 x 215 x 240 mm (all models).
Appendix C-6

This chapter contains the Error Messages concerning the 402 Dilutor Dispenser.
Appendix D-1

Error Messages
Error Message Cause/Possible Action
Adjust volume too small In DISPENSE, you have, you have entered a value for “Final vol.” that is too small. “Final vol.” > volume of biggest fraction dispensed.
Asp flow missing Re-enter Method, supplying the missing parameter, when prompted to do so (see DILUTE, DISPENSE, PIPETTE, and MIX tasks).
Diluent volume missing Re-enter Method, supplying the missing parameter, when prompted to do so (see DILUTE task).
Disp flow missing Re-enter Method, supplying the missing parameter, when prompted to do so (see DILUTE, DISPENSE, PIPETTE, and MIX tasks).
DISP task in wrong place Within a Method, DISPENSE must follow DILUTE or MIX. To DISPENSE before DILUTE you should create a Method file containing only the DISPENSE task.
Incorrect flow rates Use flow rates that correspond to the configured syringe size(s); refer to page 5-31.
Left syr life exhausted! For the Dual-syringe model: piston of left-hand syringe may need replacing.
LINK method not allowed! You can’t link files that are already linked.
Max sample vol. reached In DILUTE UNKNOWN, you need a larger syringe/transfer- tubing to measure volumes greater than the maximum reached with the existing configuration.
Mechanical problem: Help Occurs when the valve or a syringe jams; switch off and correct the problem: the syringe and/or valve need servicing (clean or change).
Method does not exist! First, you must create the Method to be linked, then enter its file number using LINK.
Method used in link file You have attempted to delete a Method number that has been assigned in LINK. You must delete the file containing LINK, before deleting the Method.
Method xxx is locked Unlock Method file, to continue with this action.
Mix volume is missing Re-enter Method, supplying the missing parameter, when prompted to do so (see MIX task).
Nb mixes is missing Re-enter Method, supplying the missing parameter, when prompted to do so (see MIX task).
No room to copy meth xxx You need to free up space by deleting one or more Methods.
Not enough memory! Reduce the size and/or number of stored Method files.
Appendix D-2

Error Message Cause/Possible Action
Parameters incompatible! Refer to page 5-16.
Ratio impossible Dilution ratio calculated by software is not valid; check parameters using formulae given in page 5-18 (DILUTE task) and correct one or both of the other values.
Ratio missing Re-enter Method, supplying the missing parameter, when prompted to do so (see DILUTE task).
Reservoir not allowed Liquids must be aspirated using the probe.
Right syr life exhausted! For the Dual syringe model: piston or right-hand syringe may need servicing (refer to page 6-6).
Sample too small Enter a larger value for Sample Volume. Verify parameters in the task.
Sample volume missing Re-enter Method, supplying the missing parameter, when prompted to do so (see DILUTE task).
Syr capacity too small Check volumes used in the task correspond to the configured syringe size(s).
Syringe Life Exhausted! For the Single syringe model: piston syringe may need servicing (refer to page 6-6).
Total volume is missing! Re-enter Method, supplying the missing parameter, when prompted to do so (see DILUTE task).
Total volume missing Re-enter Method, supplying the missing parameter, when prompted to do so (see DILUTE task).
Total volume too small Enter a larger value for the Total Volume, or check the volumes used in the task.
Valve life exhausted! Valve may need cleaning or servicing (refer to page 6-2).
xxx exists, overwrite? Select another file number, unless you want to overwrite file xxx.
Appendix D-3

Appendix D-4

Probe 2-4, 3-2 Problems
electrical 7-4 hydraulic 7-3 mechanical 7-2 Proceed 4-11
Large Volumes 5-4 Liquid contact materials B-2 Link 5-27 Lock 5-11
Rear panel 2-2, 3-8 Reservoir 3-4, 3-5 Rinse 5-28 RSD C-3 Running a method 4-11
Autoclaving 6-5 Changing the piston assembly 6-7 Cleaning the instrument 6-2 Cleaning the syringe assembly 6-4 Cleaning the tubing 6-5 Cleaning the valve 6-2 Disassembling the valve 6-2 Dismounting the valve 6-2 Measure 5-13 Mechanical
Safety Safety-1 Seal 3-9 Sector block
Fuses - Installing/changing 3-7 Installing the fuses 3-7 Power cord - connecting 3-8 Power cord - types 3-7 Voltage 3-7 Service 1-7 Single-syringe model 2-2, 5-2 Small Volumes 5-4 Soft-keys 1-2 Spare parts list A-5 Standard equipment lists A-2 STOP 4-10, 4-11 Specifications C-3 Syringe
problems 7-2 syringes C-5 tee-junction C-5 valve C-5 Method Example 4-8 Method
entering 4-7, 5-9 linking 5-27 prefabricated 5-29 storing 4-9, 5-10 Microswitches 2-2 Mix 5-26
changing 4-5, 4-7 cleaning 6-6 correction 5-33 installing 3-9, 4-5, 4-7 technical data 2-5
Next (key) 4-4 Nose (piston) 6-7 Nozzle 2-4, 3-2
advanced use 5-1 standard use 4-1
chaining 4-5, 4-7 detailed descriptions 5-15 Technical data
additional accessories list A-4 spare parts list A-5 standard equipment lists A-2, A-3 Performance Specifications C-3, C-4 Pipette 5-25 Piston
dead volumes C-5 dilution ratio C-2 systematic error C-3 random error C-3 liquid contact materials B-2 norms - conformity to C-6 Tee junction
assembly - changing 6-6 kits 6-7 Prefabricated Methods 5-29
cleaning 6-3 outlet connections 3-5

electrical problems 7-4 hydraulic problems 7-3 mechanical problems 7-2 Tubing
connecting 3-4 transfer 3-4 valve inlet 3-4 valve waste 3-5 Typical Values C-4
Union connector 6-5 Unlock 5-11 Unpacking 1-5 Upgrading 3-10
disassembly 6-2 dismounting 6-2 inlet tubing connections 3-4 outlet connections 3-5 Viscous liquids 5-33 Voltage
range C-6 selection 3-7 Volumetric analyses 5-31, C-3

Index-4