Models 3980200, 3980201, 3980202, 3980203, 3980220, 3980221, 3980222, 3980223, 3980300, 3980301, 3980302, 3980303, 3980320, 3980321, 3980322, 3980323, 3980400, 3980401, 3980402, 3980403, 3980420, 3980421, 3980422, 3980423

Labby the LABster

For more information, please contact us: ExpotechUSA 10700 Rockley Road Houston, Texas 77099 USA 281-496-0900 [voice] 281-496-0400 [fax] E-mail: sales@expotechusa.com Website: www.ExpotechUSA.com

Labconco HEPA Filtered Enclosure Manual Warranty Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of defective material and workmanship, provided such defect results from normal and proper use of the equipment. The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner, except the following:
Part #3905500, Rev. A

TABLE OF CONTENTS CHAPTER 1: INTRODUCTION 1 About This Manual 2 Typographical Conventions 3 CHAPTER 2: PREREQUISITES 5 Support, Vibration & Movement Requirements 6 Location and Air Current Requirements 6 Exhaust & Blower Requirements 6 Electrical Requirements 7 Space Requirements 7 CHAPTER 3: GETTING STARTED 9 Unpacking Your Enclosure 10 Installing the Filtered Enclosure on a Supporting Structure and Work Surface 10 Connecting to the Exhaust System (Optional) 13 Installation of HEPA Filters and Accessory Odor- Control Carbon Filters 15 Connect the Electrical Supply Source to Filtered Enclosure 16 Set the Face Velocity with Speed Control Adjustment 17 Validating the Vented Enclosure 17 Sealing the Filtered Enclosure to the Work Surface 18 CHAPTER 4: HIGH PERFORMANCE FEATURES AND
SAFETY PRECAUTIONS 19 Components 20 Safety Precautions 27 CHAPTER 5: APPROPRIATE APPLICATIONS FOR YOUR
FILTERED ENCLOSURE 31 Routine Daily Work Procedures 31 Suitable Applications 33 HEPA Filter Applications, Suitability & Guidelines 33 Odor Control Carbon Filter Applications 34 Definition of Terms 35 Appropriate Chemicals for Odor Control Carbon Filters 36 Hazardous Misapplications for Odor Control Carbon Filters with Volatile Chemicals 37 Chemical Carcinogen Use with Odor Control Carbon Filters 37

Prohibited Acid Use 38 CHAPTER 6 MAINTAINING YOUR FILTERED ENCLOSURE 39 Routine Maintenance Schedule 40 Decontamination 41 Determination of when to Replace HEPA Filters 41 How to Install a New HEPA Filter 42 HEPA Filter Leak Test 44 Setting the Inflow Face Velocity with the Speed Control Adjustment 46 Calibrate and Operate the Airflow Monitor 46 Determination of When to Replace Odor control Carbon Filters and How to Replace 50 Calculating Odor control Carbon Filter Life 52 Initial Certification 53 Re-Certification 53 Fluorescent Light Replacement 54 UV Light Replacement 54 Motorized Impeller Replacement 54 Speed Control Replacement 56 CHAPTER 7 ACCESSORIZING & MODIFYING YOUR
FILTERED ENCLOSURE 57 CHAPTER 8 TROUBLESHOOTING & SERVICER OPERATING LOG 63 APPENDIX A: FILTERED ENCLOSURE COMPONENTS AND REPLACEMENT PARTS 67 APPENDIX B: DIMENSIONS AND EXHAUST OPTIONS 72 APPENDIX C: FILTERED ENCLOSURE SPECIFICATIONS 75 APPENDIX D: QUICK CHART FOR THE FILTERED

CHAPTER 1 INTRODUCTION Congratulations on your purchase of a Labconco Purifier® HEPA Filtered Enclosure or Purifier® Class I Filtered Enclosure. Your enclosure provides personnel protection through superior containment. It is the result of Labconco’s more than 50 years experience in manufacturing fume hoods and more than 30 years experience in manufacturing filtered enclosures. These enclosures will effectively contain toxic, noxious, or biohazardous particulates when properly installed and operated. Each enclosure uses a single HEPA filter, which is rated at least 99.99% efficient for 0.3-micron particles. Additionally, an accessory Odor Control carbon filter may be added to adsorb nuisance odors from organic vapors, formaldehyde or ammonia and amines. The Purifier enclosures offer many unique features to enhance safety, performance, and energy savings. To take full advantage of them, please acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how high performance HEPA filtered enclosures operate, please review Chapter 4: High Performance Features and Safety Precautions before you begin working in the enclosure. Even if you are an experienced user, please review Chapter 5: Using Your HEPA Filtered Enclosure, which describes the Purifier features so that you can use the filtered enclosure efficiently. For hazardous biological applications requiring user protection but not product protection, Labconco recommends the purchase of a Class I enclosure, which includes a UV light for general surface decontamination. For general hazardous biological applications,

particulates, and nuisance powders, Labconco recommends the use of the Purifier HEPA Filtered Enclosure. See chart below and contact Labconco for additional ordering information.
* It is recommended that a Labconco product specialist review the chemical application to determine if it is suitable. Consult Chapter 5 and Chapter 6.
This manual is designed to help you learn how to install, use, and maintain your filtered enclosure. Instructions for installing optional equipment or accessory carbon filters on your filtered enclosure are also included. Chapter 1: Introduction provides a brief overview of the filtered enclosure, explains the organization of the manual, and defines the typographical conventions used in the manual. Chapter 2: Prerequisites explains what you need to do to prepare your site before you install the filtered enclosure. Electrical and service requirements are discussed. Chapter 3: Getting Started contains the information you need to properly unpack, inspect, install, and certify the filtered enclosure. Chapter 4: High Performance Features and Safety Precautions explains how the Purifier filtered enclosure operates and the appropriate precautions you should take when using it.

Chapter 5: Using Your Filtered Enclosure discusses the basic operation of how to prepare, use and shut down your filtered enclosure. Chapter 6: Maintaining Your Filtered Enclosure explains how to perform routine maintenance on the filtered enclosure. Chapter 7: Accessorizing Your Filtered Enclosure explains acceptable modifications to the filtered enclosure or how to add accessories. Chapter 8: Troubleshooting contains a table of problems you may encounter while using the filtered enclosure including the probable causes of the problems and suggested corrective actions. Appendix A: Components and Replacement Parts contains labeled diagrams of all of the components of the filtered enclosures. Appendix B: Dimensions and Exhaust Options contains comprehensive diagrams showing all of the dimensions for the filtered enclosures. Appendix C: Filtered Enclosure Specifications contains the electrical requirements for filtered enclosures. Wiring diagrams are also included. Appendix D: Quick Chart provides a table of airflow and test data for the filtered enclosures. Appendix E: References lists the various resources available that address laboratory ventilation and biosafety.


CHAPTER 2 PREREQUISITES Before you install the filtered enclosure, you need to prepare your site for installation. You must be certain that the area is level and of solid construction. In addition, a dedicated source of electrical power should be located near the installation site to power the filtered enclosure, and other apparatus. Additionally, the enclosure should be strategically placed in the lab to provide efficient workflow. Carefully read this chapter to learn the requirements for your installation site:

At a minimum, the supporting structure usually consists of a base cabinet and chemically-resistant work surface. • A stand or mobile bench is not recommended if the enclosure will be used for precise weighing of powders. Location and Air Current Requirements
The Purifier Class I Enclosure and Purifier Filtered Enclosure use an integral motorized impeller to draw room air past the operator and through the enclosure. This contaminated air is then pushed through the HEPA filter. The HEPA-filtered exhaust air is then forced out the top of the enclosure. An optional carbon filter may be installed on the downstream side of the HEPA filter to protect against nuisance odors. The HEPA-filtered exhaust air can be recirculated into the laboratory or exhausted outside with the addition of the exhaust connection kit and remote blower listed in Chapter 7. Data for the exhaust volume and enclosure static pressure loss are listed for each filtered enclosure model at face velocities of 60, 75, 80, 90, 100, and 105 fpm.

Standard duplex electrical receptacles should be nearby for connecting the filtered enclosure, or other equipment. The enclosures include iris pass-throughs to allow electrical cords through the back of the enclosure without leaving a large hole for contaminants to escape.
The dimensions for the different models are shown in Appendix B: Dimensions


CHAPTER 3 GETTING STARTED Now that the site for your filtered enclosure is properly prepared, you are ready to unpack, inspect, install, and validate your system. Read this chapter to learn how to:
Each enclosure model weighs between 125 to 195 lbs. each (55 to 85 kg). The shipping container allows for lifting with a mechanical lift truck or floor jack. If you must lift the enclosure manually, follow safe-lifting guidelines. Do not lift by the front air foil.

Carefully remove the shrink-wrap or carton on the enclosure and inspect it for damage that may have occurred in transit. If damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
DO NOT RETURN GOODS WITHOUT THE PRIOR AUTHORIZATION OF LABCONCO. UNAUTHORIZED RETURNS WILL NOT BE ACCEPTED. IF ENCLOSURE WAS DAMAGED IN TRANSIT, YOU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER. LABCONCO CORPORATION AND ITS DEALERS ARE NOT RESPONSIBLE FOR SHIPPING DAMAGES. Do not discard the packing material until you have checked all of the components and tested the enclosure. We recommend that you do not remove the enclosure from its shipping container until it is ready to be placed into its final location. Move the unit by placing a flat, low dolly under the shipping skid, or by using a floor jack.
☞
!
Do not move the enclosure by tilting it onto a hand truck. Installing the Filtered Enclosure on a Supporting Structure and Work Surface
Use caution when lifting or moving the enclosure. When installing the enclosure onto a chemical-resistant work surface or benchtop, ensure that the structure can safely support the combined weight of the enclosure and any related equipment. The work surface should be at least as wide as the enclosure to properly support it. The front of the enclosure should be aligned within

0.36" of the front of the work surface. Mounting holes are provided in the Labconco accessory work surfaces to secure the enclosure. Work Surface Specifications The work surface should be smooth, rigid, and durable, such as a chemical-resistant epoxy resin. The surface should be non-porous and resistant to the biohazards, powders, solvents and chemicals used in conjunction with the Purifier Filtered Enclosure. The work surface should also contain a dished recessed area for containing primary spills. Work Surface and HEPA Filtered Enclosure Installation
1. Level the base cabinets and the work surface. Work surface should be placed flush with the front of the base cabinet as shown in Figure 3-1. 2. Position the work surface in its intended location and with the front of the work surface towards you. (Rear mounting holes are located close to the rear edge.) 3. Secure the work surface to the base cabinet with a structural adhesive or silicone sealant. 4. Insert the supplied mounting screws in the four holes. Allow a minimum of 1/8" clearance under the head of the screw for positioning the enclosure. 5. Place the enclosure on the work surface and slide the rear flange and front air foil flanges under the mounting screw heads. 6. Tighten the four screws to complete the installation.


Figure 3-1 Filtered Enclosure Installation

!
The exhaust system should be installed by a qualified HVAC contractor.
Exhaust transition kits aid in the removal of chemicals or applications where a higher degree of biohazard and particulate removal is required. See Chapter 7 for ordering exhaust transition kits for the Purifier HEPA Filtered Enclosure or Purifier Class I Enclosure. Review Chapter 2 for exhaust prerequisites and review Chapter 7 for ordering blower exhaust equipment. See Figure 3-2 for exhaust kit options on Purifier filtered enclosures. Consult Labconco Customer Service should you require help sizing your blower for the exhaust volume and system static pressure loss. To ensure compatibility, the selected exhaust duct material should match the enclosure, procedures and chemical applications.


Figure 3-2 Exhaust Duct Connection Kit shown installed on
Purifier Models (Optional) Kit #3924400 (2'), 3924401 (3'), and 3924402 (4')

HEPA Filters HEPA filters are shipped installed with the gasket on the downstream side. The HEPA filters are leak checked at Labconco. A second leak check is recommended before using the enclosure and at least annually thereafter. Consult your Safety Officer and Chapter 6 for the HEPA Filter Leak Test. See Figure 4-2 for HEPA filter location, HEPA filter gasket, and filter clamp bolts. The Purifier enclosures require decontamination of biohazardous materials or are used with nonbiohazardous materials or nuisance powders. All seams upstream of the HEPA filter are jacketed by negative pressure. This intrinsically safe design ensures containment. Carbon Filters (Accessory for Odor Control only) For carbon filter installation, first remove the upper perforated cover, secured by Phillips head screws, on the top of the enclosure. See Figures 4-1 and 4-2 for location of the upper perforated cover. See Chapter 7 for a list of available carbon filters. All carbon filters are simply gravity installed with the gasket side down and down stream on the exhaust side of the HEPA filter. Install the correct carbon filter for the application.
CARBON

Filter Type Appropriate Use HEPA 2' 3707900 3' 3707901 4' 3707902
HEPA filters are high-efficiency particulate air filters having a particulate removable efficiency of 99.99% for particles with a diameter of 0.3 micron. Organic Vapor Activated Carbon
Adsorbs organic compounds designated by NIOSH guidelines as acceptable for use with chemical cartridge-type respirators. Concentrations in the enclosure’s work area must not exceed the IDLH for the chemical and the exhaust from the enclosure must not exceed the TWA. Formaldehyde
2' 3937200 3' 3937300 4' 3937400
Formaldehyde requires the use of an impregnated carbon for the treatment of formaldehyde. Concentrations in the enclosure’s work area must not exceed the IDLH for the chemical and the exhaust from the enclosure must not exceed the TWA. Ammonia and
(Formalin) 2' 3937201 3' 3937301 4' 3937401
Treats ammonia, low molecular weight amines and other bases designated by NIOSH as acceptable for use with ammonia cartridge type respirators. Requires the use of an impregnated carbon. Concentrations in the enclosure’s work area must not exceed the IDLH for the chemical and the exhaust from the enclosure must not exceed the TWA. Important: Carbon filters do not provide any particulate protection, but provide odor control.
Amines 2' 3937202 3' 3937302 4' 3937402
Connecting the Electrical Supply Source to the Filtered Enclosure
115V Models Simply connect the 115V power cord supplied to the IEC electrical supply plug on the back of the enclosure. If using at 50 Hz operations, blower performance maximum airflow will be reduced by 17%.
230V Models The above procedure applies for the 230V except it is shipped without a plug. Install the appropriate plug for your electrical specifications per local codes.

Adjustment of the speed control gives the correct face velocity and is located behind the front panel. The face velocity should be from 75 to 105 fpm for biohazardous operations. (Consult your Safety Officer for airflow recommendations for your application). Containment is maximized at a setting within this range. Working at the lowest face velocity appropriate for the application will give the quietest operation. Face velocity measurements are made using an anemometer. An electric anemometer can be obtained from your laboratory supply dealer. Face velocity measurements should be taken in accordance with the Industrial Ventilation Manual. Using a small Phillips screwdriver, adjust the speed control to give the required face velocity. See Figures 6-1 and 6-2 in Chapter 6 to locate the speed control. The face velocity is increased by turning the speed control counter clockwise and clockwise to decrease face velocity.
To determine the actual face velocity at the sash opening, airflow velocity readings are taken. This should be done across the sash opening of the enclosure in accordance with the Industrial Ventilation Manual. (See Appendix E) The “average face velocity” is achieved by taking readings in two rows across the enclosure with the readings 6" from the ends and evenly spaced every 12"; the first row is 3" down from the upper sash foil and the second row is 3" up from the work surface. A total of four readings will be taken for the 2' enclosure, six readings taken for the 3' enclosure and eight readings taken for the 4' enclosure and then averaged. Refer to Chapter 2 for proper airflow volumes for your particular model. The Purifier enclosures have been tested at Labconco’s airflow test facility per ASHRAE 110-1995. All enclosures achieve an “as manufactured rating” of less than 0.05 part per million (ppm) at 4 liters per minute (lpm); AM

The Purifier Enclosures were also subjected to Biological Containment testing, using a modified Personnel Protection Test, as described in NSF/ANSI Standard Number 49. An aerosol challenge of approximately 5 x 108 spores of Bacillus subtilus var Niger were released in the enclosure and appropriate air sampling equipment established the number of spores that escaped through the sash opening. Each size enclosure was tested three times. All three sizes of the enclosure were tested at 75 FPM inflow and typically 1 or less than one spore escaped from the enclosure during each of the tests. The Purifier Enclosures were also subjected to an industrial hygiene air-monitoring study to confirm the performance of the enclosure for powders. The Purifier Enclosures demonstrated excellent containment when used by an operator using excellent technique and good containment when used by an operator using marginal technique. While no enclosure can compensate for improper technique, these tests confirm the Purifier Enclosures provide a safe working environment.
When the filtered enclosure has been set in place, ducted, (if necessary), and wired, it may be sealed at the work surface to prevent spilled materials from collecting under the walls. A bead of silicone sealant is recommended to seal the filtered enclosure to the work surface.

CHAPTER 4 HIGH PERFORMANCE FEATURES AND SAFETY PRECAUTIONS
High Performance Features: The patented* Purifier HEPA Filtered Enclosure and Purifier Class I Enclosure are designed to meet the needs of the laboratory scientist, and provide superior containment while operating at velocities between 75-105 feet per minute. The filtered enclosures have been tested to effectively contain toxic, noxious, and biohazardous materials when properly installed and operated. What makes the filtered enclosures so unique is the revolutionary way they direct air into and through the contaminated air chamber. Labconco engineered the filtered enclosures to minimize the effects of turbulence. The containment-enhancing and aerodynamic designs of the upper sash foil, side air foils, lower air foil, upper dilution air supply, and rear perforated baffle all work in concert to produce horizontal airflow patterns that significantly reduce powder, chemical and biological material concentrations through the work area.
*(U.S. Patent No. 6,461,233)

These concentrations of materials are predominantly removed on the “first pass” of airflow through the chamber resulting in high performance containment.
The plenum and the HEPA filter are jacketed by negative pressure. Should a leak occur in the filter gasket or the plenum, the contaminated air is recaptured and refiltered. The Purifier Class I Enclosure and Purifier HEPA Filtered Enclosure have an access port located behind the front panel. For HEPA filter removal in biohazardous applications on Purifier enclosures, the HEPA filter will require decontamination before removal. Users of all equipment are encouraged to routinely check airflow or purchase a model with an airflow monitor.

Not Shown 7, 9, 10, 11, 12, 13, 20, 21, 22, 23, 24, 25, 28 (See Figure 4-2)

Figure 4-1


Figure 4-2 HEPA Filtered Enclosure Airflow Diagram



Figure 4-3 · Figure 4-4


Monitor is an available option on all Purifier models. See Figure 4-1 and 4-2.



11. The use of safety goggles, protective clothing, gloves and any other personal protective equipment recommended by your safety officer should be used. 12. The sash must remain in the down position while using the enclosure. 13. Proper performance of the enclosure depends largely upon its location and the operator’s work habits. Consult the references in Appendix D. 14. The enclosure should be recertified whenever it is serviced or at least annually thereafter. 15. The HEPA filter provides personnel and environmental protection from particulate matter. Because room air is drawn over the work surface during operation, this enclosure should not be used for operations requiring product protection from environmental contamination. 16. Avoid the use of flammable gases or solvents in the enclosure if possible. Care must be taken to ensure against the concentration of flammable or explosive gases or vapors. Use of an open flame should be avoided in the enclosure. Open flames may disrupt the airflow patterns in the cabinet, burn the HEPA filter and damage the filter’s adhesive. Gases under high pressure should not be used in the enclosure as they may disrupt the airflow patterns of the cabinet. 17. HEPA filters are only effective for entrapment of particulate matter. Manipulations that generate gases or vapors from toxic chemicals or radionuclides, must be evaluated carefully from the standpoint of buildup to dangerous levels, the decontamination of the enclosure and compliance with applicable regulations. 18. The surface of the HEPA filter is fragile and should not be touched. Care must be taken to avoid puncturing the HEPA filter during installation or normal operation. If you suspect that a HEPA filter has been damaged DO NOT use the enclosure; contact a local certification agency or Labconco.

19. The HEPA filter in the enclosure will gradually accumulate airborne particulate matter from the room and from work performed in the enclosure. The rate of accumulation will depend upon the cleanliness of the room air, the amount of time the enclosure is operating and the nature of work being done in the enclosure. With normal usage, the HEPA filters will last two to five years before requiring replacement. 20. Avoid direct exposure to ultraviolet (UV) radiation; NEVER work in the Purifier Class I Enclosure when the UV light is on. 21. Ensure that the filtered enclosure is connected to an electrical service in accordance with local and national electrical codes. Failure to do so may create a fire or electrical hazard. Do not remove or service any electrical components without first disconnecting the filtered enclosure from electrical service. 22. Use of an open flame must be avoided with the enclosure. Open flames may disrupt the airflow patterns in the enclosure and cause a fire hazard with volatile solvents. 23. Increase the face velocity by adjusting the speed control of the motorized impeller or change the HEPA filter when the face velocity falls below acceptable limits established by your Safety Officer. 24. Exhaust carbon filters and HEPA filters are disposed as hazardous waste. The user is responsible for recording the chemicals adsorbed or treated by the filters and disposing properly. HEPA filters should be decontaminated before disposal on Purifier enclosures. 25. Ensure only trained operators use the enclosure. New users should review the User’s Manual and become familiar with the operation of the enclosure. 26. If the filtered enclosure is to be used in a confined space, make sure the space is well ventilated and the concentration of toxic contaminants cannot accumulate greater than the TWA. 27. Proper operation of the enclosure depends largely upon the enclosure location and the operator’s work habits. The enclosure should be located away from traffic patterns, doors, fans, ventilation registers, fume hoods, and any other air-handling device that could disrupt its airflow patterns. Consult Chapter 2: Prerequisites and Chapter 3: Getting Started sections of this manual for further details.


CHAPTER 5 APPROPRIATE APPLICATIONS FOR YOUR FILTERED ENCLOSURE
Now that the installation of your filtered enclosure is completed, you are ready to use your filtered enclosure. Read this chapter to learn about:
Planning


Unloading Materials and Equipment
NIOSH (National Institute for Occupational Safety and Health, USA) has established guidelines for chemical cartridge and HEPA filtered respirators. Suitable applications for the filtered enclosures are based on these guidelines. As with respirators, chemical contaminants are adsorbed or treated by carbon; particulates, powders and biohazardous materials are filtered by the HEPA filter before air is returned to the room. Biohazardous materials or other hazardous solids can be used with the HEPA filtered enclosure as long as appropriate decontamination occurs before filter changing operations. See Chapter 6 for decontamination procedures. Listed below are suitable applications for HEPA filters and accessory Odor Control Carbon Filters.

REFERENCES TO NIOSH OR OSHA GUIDELINES AND REGULATIONS APPLY TO ANY WORK PLACE UNDER THE JURISDICTION OF THE U.S. DEPARTMENT OF LABOR. OTHER COUNTRIES OUTSIDE THE U.S. HAVE ESTABLISHED STANDARDS, WHICH MAY DIFFER SLIGHTLY FROM

THOSE USED AS GUIDELINES FOR THIS PRODUCT. IT IS THE USER’S RESPONSIBILITY TO BECOME AWARE OF LOCAL REGULATIONS GOVERNING THE SAFE USE AND DISPOSAL OF CHEMICALS, CARBON AND HEPA FILTERS. KNOWLEDGE OF ESTABLISHED SAFE EXPOSURE LEVELS IS IMPERATIVE TO THE PROPER USE OF FILTERED ENCLOSURES.
NIOSH – National Institute for Occupational Safety and Health/Mine Safety and Health Administration. (U.S.A.) TWA – Recommended Exposure Limits expressed as a Time Weighted Average. The exposure limit for that chemical for up to a 10-hour workday, 40 hours a week. Expressed in units of parts per million or milligrams per cubic meter. Odor Threshold – The value in parts per million or milligrams per cubic meter for which one might expect to smell a chemical’s presence in the air. This value is very subjective and detection will vary with the sensitivity of one’s nose. The period of time until the odor threshold is reached in the exhaust stream can be estimated from Labconco exclusive computerized filter modeling program. Contact Labconco on carbon filter life for specific applications. See Chapter 6. Saturation Level or Time – There is a limit to the amount of chemical that can be adsorbed by activated carbon, or neutralized by chemically-treated carbon. Once the capacity of the carbon is reached, it is considered to be saturated and will adsorb (or neutralize) no further material; the outlet concentration of the chemical will equal the inlet concentration from that point until the filter is replaced. (Note that the capacity of activated carbon is not a constant, but varies with the inlet concentration). Labconco Technical specialists can determine with the computerized carbon- modeling program the estimated time saturation for a particular chemical. When using a HEPA filter alone in all filtered enclosures or in combination with an Odor Control carbon filter, the speed control will need to be increased to allow for HEPA filter loading. Measure filtered enclosure face velocity with an anemometer.

IDLH (Immediately Dangerous to Life and Health). An atmosphere that poses an immediate hazard to life or produces immediate irreversible health effects. IDLH concentrations should not be approached in the enclosure.
Below is a general set of rules to determine appropriateness of chemical usage.
Selected organic chemicals considered to be occupational carcinogens by NIOSH can be used in the filtered enclosure with carbon filters under rigid restrictions. See separate discussion on carcinogens for special instructions.
Organics must have time weighted exposure limits (TWA) of 1 PPM or greater.
Chemicals must have a detectable odor at concentrations below the TWA for the chemical.
Chemicals must be designated by NIOSH guidelines as acceptable for use with chemical cartridge-type respirators (the exception is formaldehyde and ammonia/amines, which used impregnated carbon). Chemicals not listed by NIOSH in the Pocket Guide must be approved by Labconco Product Specialist (or Engineering).
Inlet concentration must never exceed the IDLH (Immediately Dangerous to Life and Health) concentrations.
Chemicals having a recommendation by NIOSH of at least “Escape GMFOV” (Gas Mask Full-Face Respirator).
When evaporating a mixture of chemicals, the chemical having the lowest TWA will be used to determine if the mixture meets the guidelines.
Call a Labconco Product Specialist at 1-800-821-5525 for assistance in chemical appropriateness.

There is one scenario where the accessory carbon filter misapplication would be a part of a hazardous condition. If the user continues to operate the enclosure with any of the following conditions present a potentially hazardous condition will exist:
Selected carcinogens may be used safely with Odor Control carbon filters under the following restrictions.
The use of a vented fume hood or ventilated enclosure with ducting to the outside is always the preferred method when working with carcinogens. The Odor Control carbon filters should only be used, as a last resort when venting to the outside is not an option. The potential carcinogens are listed in the NIOSH Pocket Guide to Chemical Hazards as “Ca.” Each potential carcinogen must have a TWA of 1 or greater; have minimum respirator recommendation of Escape GMFOV, and an odor threshold significantly lower than the TWA for the chemical.
The inlet concentration or the evaporation rate of the chemical must never exceed the TWA for the chemical.

Consult a Labconco Technical Specialist for estimated saturation life. See Chapter 6 for an example of estimating saturation life. Another source is the Labconco chemical guide for carbon filtered enclosures.
The Purifier HEPA and Purifier Class I filtered enclosures motorized impeller cannot be exposed to acids. Where applications require the use of acids, a separate ventilated enclosure or vented fume hood should be used with a remote blower ducted to the outside. No exceptions are permitted, as the impeller life span will be limited with acid use.

CHAPTER 6 MAINTAINING YOUR FILTERED ENCLOSURE Monitoring airflow and changing the filters is the primary maintenance required. Decontamination may be required and is reviewed in Chapter 6. Certification and recertification is reviewed in Chapter 6. Review this chapter on maintenance for the following: 1. Routine Maintenance. 2. Decontamination. 3. Determination of when to replace the HEPA filters. 4. How to install a new HEPA filter. 5. HEPA filter leak test. 6. Speed control adjustment and setting the inflow face velocity. 7. Operating and calibrating the airflow monitors. 8. Determination of when to replace Odor Control carbon filters and how to replace. 9. Calculating Odor Control carbon filter life. 10. Initial certification. 11. Re-certification. 12. Fluorescent light replacement. 13. UV light replacement on Purifier Class I. 14. Motorized impeller replacement. 15. Speed control replacement.

CARBON
CARBON
Weekly

The HEPA filters in the filtered enclosures gradually accumulate airborne particulate matter and powders from the enclosure and room. The rate of accumulation will depend upon the cleanliness of the room air, the amount of time the enclosure is operating, and the nature of work being done in the enclosure. In typical installations and usage, the HEPA filters will last two to five years before requiring replacement. Replace HEPA filters when face velocity drops below the recommended 75-105 fpm velocity, and the speed control is adjusted to full speed. Replace HEPA filters if it fails the HEPA Filter Leak Test in Chapter 6.
Annually

servicing the HEPA filter. Only a qualified certifier should service the HEPA filter. After the HEPA filter is replaced, the enclosure MUST be certified. See Figure 6- 1.
1. Unplug the enclosure. 2. Remove the front panel by loosening the two screws that secure it, and then remove the filter access cover. 3. Using a 9/16" deep socket, loosen the filter clamp bolts located on top. Refer to Figures 4-1 and 4-2. 4. With the clamp bolts loosened, the HEPA filter-clamping frame should be clear of filter. Carefully pull the filter straight out of the enclosure and discard properly. 5. With the filter removed, inspect the clamping frame and the frame of the cabinet for damage. 6. Cover the surface of the new HEPA filter gasket with a light coating of silicone grease, if desired. 7. Install the new HEPA filter by pushing it straight into the cabinet, ensuring that it is correctly oriented with the gasket facing up on the exhaust side of the enclosure. The filter gasket is surrounded by a negative pressure plenum and the gasket will not leak. Be sure the filter fits properly. 8. Tighten the clamp bolts uniformly until the filter gasket is properly compressed against the frame. Inspect the seal thoroughly before proceeding. 9. Reinstall the filter access cover and the front panel. 10. Plug the enclosure in and have it recertified before use.


Figure 6-1 HEPA Filter Changing Diagram & Filter Leak Test Diagram

Purpose After installing the new HEPA filter, the HEPA filter should be leak checked. This test is performed to determine the integrity of the HEPA filter, the filter housing, and the filter mounting frames. Leak testing is to be done by a qualified technician with calibrated equipment. Remove the top perforated exhaust cover by using a Phillips screwdriver to unfasten the (2) screws used to secure (2) clips. See Figure 6-1. The filter passes the leak test at .01% or better. Reference Leak Testing and Photometer scanning from the Institute of Environmental Services (IES-RP-CC001.3)
Apparatus 1. An aerosol photometer ATI model 2D, 2E, 2G or equivalent. Air Techniques Hamilton Associates inc. 11403 Cron Ridge Dr. Owings Mills, MD 21117 2. One aerosol generator of the Laskin nozzle(s) type. An aerosol of mineral oil or suitable liquid shall be created by flowing air through it. The compressed air supplied to the generator should be adjusted to a pressure of 10± 1 psig. during operation. Air Techniques Inc. Model TDA-4A or equal. One nozzle at 10 psig is (67.5 cfm x 100ug/l)/(Vol. of air), For the 2' at 90 fpm or 130 cfm, one nozzle @10 psig is 13,500/130 cfm = 52 ug/l. For the 3' at 90 fpm or 200 cfm, one nozzle @ 10 psig is 34 ug/l. For the 4' at 90 fpm or 265 cfm, one nozzle at 10 psig is 26 ug/l. 3. Mineral oil (Catalog #1491400). 4. Sampling Nozzle, Rectangular 1/2" x 3-1/4", Air Techniques, Inc. Procedure For the ATI 2G Photometer

11. Allow the generator to operate for a minimum of 15 seconds. For all integral motorized impeller models, scan the downstream exhaust side of the HEPA filter by passing the sampling nozzle of the gun in slightly overlapping strokes over the entire surface the filter, with the sampling port not more than 1 inch from the surface of the filter media. Scan the entire periphery of the filter and the gasket between the filter frame and the enclosure frame. Scanning shall be done at a traverse rate of not more than 2 inches per second.
place the sampling nozzle in the center of the remote blower exhaust. Acceptance
Aerosol penetration shall not exceed 0.01 percent measured by the photometer.

1. Remove the front panel by loosening the (2) Phillips screws on top that secure the front panel. 2. The speed control is located on the electrical subassembly located behind the switched control panel and below the front panel. See Figure 6-1. 3. Adjust the speed control with a small Phillips screwdriver by turning the screw counterclockwise to increase blower speed or clockwise to decrease the blower speed. The speed control is very sensitive, so proceed with caution. 4. Measure the inflow velocity per the averaging technique outlined in Chapter 3 and adjust the speed control slowly for the desired speed. Allow the speed to stabilize and re- measure the inflow velocity to confirm. 5. Replace the front panel and tighten the screws.
Guardian Airflow Monitor (LED Monitor) Refer to Figure 6-2 for operation and calibration. Labconco Airflow Monitor / Airflow Switch Operation The Guardian Airflow Monitor (LED) consists of a circuit board and an airflow switch. This switch indicates airflow as safe or low. It does not provide an actual face velocity, but a small setscrew in the back of the sensor can adjust the airflow level that it classifies as “good/safe” or “low/alert.” The circuit board provides power to the sensor and also contains a “safe (green)” and “alert (red)” airflow LED indicators, as well as a “SILENCE ALARM” button to quiet the audio alarm. When first powered up, the PCB will light both red and green LED indicators and sound the alarm to indicate it is working. After 5 seconds, the air monitor will indicate either good or bad airflow based on what the connected airflow switch detects. For low airflow, the unit will wait for 10 seconds of bad indications before it sounds both the audio alarm and the red “alert” LED indicator. If

1. Ensure the flow switch and alarm circuit board are installed and operational. 2. Allow the enclosure to operate for at least two minutes. 3. If factory installed, the monitor will alarm at 60±10 fpm with the inflow velocity set at 90±10 fpm. 4. To change the factory setting, set the inflow velocity required by your Safety Officer to the desired alarm condition using the speed control adjustment procedure outlined in Chapter 6. Once the alarm condition is set, use a small screwdriver to turn the adjustment screw on the airflow switch counterclockwise (facing the screw) until the “low” airflow red LED lights and the audible flow alarm sounds.

5. Adjust the inflow velocity to the nominal operating point required by your Safety Officer. 6. Over time the HEPA filter will load and eventually slow the inflow velocity. Once the alarm condition is met, simply increase the speed control outlined in Chapter 6 or replace the HEPA filter if the speed control is maximized. 7. The table below lists typical alarm conditions based on normal operating conditions. Typical alarm conditions are set at face velocities of 10 to 30 feet per minute below the normal operating conditions due to supply air and exhaust air fluctuations, as well as room air cross drafts. Consult your Safety Officer for proper operating speeds.
Alarm Condition Set Point
Speed 100 ± 10 fpm 70-90 fpm 90 ± 10 fpm 60-80 fpm 80 ± 10 fpm 50-70 fpm 70 ± 10 fpm 50-60 fpm 60 ± 10 fpm 50 fpm
Enclosure Operating In-Flow · Speed


Figure 6-2 Guardian Airflow Monitor (LED) with Airflow Switch

Determination of When to Replace Odor Control Carbon Filters and How to Replace
The carbon filters MUST be replaced when any one of the following two conditions are met:
There are four means of determining when its time to change the carbon filters (not shown in the order of preference).
Odor - A person’s sensitivity to odor, tolerance of odor and their comfort level under odoriferous conditions vary with the individual. While odor is an indicator that chemicals are passing through the carbon filter, several points need to be understood:
Detection Tubes - Color change indicators can be used to measure the concentration of the chemical at the exit side of the carbon filter or in the outlet exhaust. A kit including syringe pump and flexible tubing can be purchased as an accessory from Labconco (Catalog # 6924900). Labconco Customer Service Representatives are supplied with detector tube catalog numbers, as well as telephone numbers to direct you to where to purchase these items.
For Organic, Formaldehyde and Ammonia, chemical specific detector tubes should be purchased when installing fresh filters. Each kit contains instructions on how many strokes of the syringe are required to obtain the stated sensitivity. The sampling syringe is connected to the filtered enclosure exhaust. Connect the syringe to the detector tube while the system is running and pull the air through the tube with the syringe. Each stroke of the

syringe represents a 100-ml sample and corresponds to the number of strokes necessary to give the indicated color changes. Due to the wide variety of organics and varying TWA’s, it is recommended that specific detector tubes be purchased directly from Sensidyne, Draeger or your laboratory supply dealer. Alternate detector pumps can also be purchased from your laboratory supply dealers. The vast majority of detector tubes available start measuring at the TWA. When a user observes a color change in the tube, they should replace the filter immediately. If no detector tube for your specific chemical is available, other means of detection must be used.
Time - For applications that have very consistent inlet concentrations and operating time, “Time” can be used to anticipate saturation or TWA levels based on prior experience. However, this does not replace the need for sampling. Consult Labconco technical specialist for an estimate of carbon filter life based on chemical usage. Detector tubes, or analytical instrumentation should always be used to determine concentrations in the carbon filter. It is recommended that the carbon filters be checked with detector tubes or other means at intervals of 20% of the total estimated filter life. The exception to the 20% recommendation is formaldehyde and any carcinogen or suspected carcinogen. These more hazardous chemicals must be checked at least every 10% of the total estimated time.
Analytical Instrumentation - This is the most accurate means of measuring concentrations of any chemical. It is the method of choice when no detector tubes are available or the tubes are not sensitive enough to measure at the TWA concentration for the chemical. This method is also to be used to determine saturation when the chemical concentration is below the measurement range of detector tubes.
Odor Control Carbon Filter Replacement Procedure – See Figures 4-1 and 4-2 for Carbon Filter Replacement.

Labconco developed a modeling program to estimate the filter life for typical carbon filters. Since filter life is dependent on the chemical used, the airflow, filter size, and the dwell time, refer to the Chemical Guide for the Paramount® Filtered Enclosure. The estimated life for Odor Control carbon filters for the filtered enclosures is conservatively calculated at 50% or half of the published values for the Paramount in the Chemical Guide. For example, if you use isopropyl alcohol to disinfect and use approximately 100 ml per week during 2 hours of use per day then follow these steps to calculate the concentration in parts per million (ppm).
Steps for Calculating PPM and Filter Life 1. Determine the amount of the proposed chemical lost to evaporation over a given amount of time. For example, if you use isopropyl alcohol and lose approximately 100 ml per week during 2 hours of use per day. 2. Convert the amount lost into ml/min. For this example: 100 ml X 1 week X 10 hours = 100 ml lost week 10 hours use 600 minutes 600 minutes 3. Convert ml/min to PPM by multiplying ml/min by the conversion factor found in the second to the last column on the right. For isopropyl alcohol .17 x 41 = 7.0 PPM. 4. Find the PPM value on the chart that comes closest to the value you just calculated in step #3. In this example, round up to 10 PPM, which is close to the calculated 7.0. We may approximate the filter life to be around 155 hours of actual use, but use 50% of this for the Odor Control filters or 78 hours. 5. Insert the estimated filter life into the estimated usage to determine how long filters will last.
=.17 ml/min
78 hours filter life 10 hours per week use =7.8 weeks before filter saturation

Formaldehyde only For formaldehyde, use 10% of the impregnated carbon weight. Formalin is 37% formaldehyde by weight. The density of formalin is 1.08 g/ml.
Formalin 2' 18 x 18 x 1 7.0 0.7 lbs./318g 796 ml 3' 30 x 18 x 1 12.0 1.2 lbs./545g 1364 ml 4' 42 x 18 x 1 16.8 1.68 lbs./763g 1910 ml Ammonia only For ammonia, use 10% of the impregnated carbon weight. Assume use of a 50% solution of ammonia for these calculations.
Adsorbed Volume of 50%
II Carbon Adsorbed weight of
Solution 2' 18 x 18 x 1 9.2 0.92 lbs./418g 836 ml 3' 30 x 18 x 1 13.5 1.35 lbs./613g 1226 ml 4' 42 x 18 x 1 18.9 1.89 lbs./859g 1718 ml
The filtered enclosure has been certified at the factory for an inflow velocity of 90±10 fpm along with the HEPA Filter Leak Test. The filtered enclosure should be certified for the proper inflow velocity required by your Safety Officer. It is also a conservative recommendation to perform the HEPA Filter Leak Test again should there be any damage caused during transport.
In addition, the following tests should also be performed at the user’s discretion:

The motorized impeller must be replaced as a complete unit. When the motorized impeller is replaced, the capacitor may also be replaced. See Appendix A for Replacement Parts Diagram. See Figure 6-5 for an isometric view of the motorized impeller plenum assembly. The HEPA filter rests on top of the motorized impeller assembly.


Figure 6-3 Motorized Impeller Replacement


CHAPTER 7 ACCESSORIZING AND MODIFYING YOUR FILTERED ENCLOSURE
There are several ways to accessorize and modify the filtered enclosure for your individual requirements. These include the addition of accessory work surfaces, airflow monitors; exhaust transition kits, remote blowers, exhaust dampers, filters, storage cabinets and utility shelf kit.
An optional dished work surface is available to attach to the filtered enclosure. Dished work surfaces are contoured to fit the dimensions of the filtered enclosures to contain spills. Epoxy is chemical resistant. See Figure 3-1 for installation. Catalog # Description Dimensions (W x D x H) 3909900 Black, 2-foot wide 24" x 29" x 1" 3909901 Black, 3-foot wide 36" x 29" x 1" 3909902 Black, 4-foot wide 48" x 29" x 1" 3909903 Gray, 2-foot wide 24" x 29" x 1" 3909904 Gray, 3-foot wide 36" x 29" x 1" 3909905 Gray, 4-foot wide 48" x 29" x 1"

The Guardian LED Airflow Monitor allows you to continuously monitor airflow through the enclosure. The Guardian LED monitor can be placed on any Purifier enclosure. Description Order Part #’s Individual Parts Required Guardian Airflow Monitor 115V or 230V
1 ea. 3811500 Airflow Monitor Printed Circuit Board 1 ea. 3910700 Airflow Sensor w/ Connector 2 ea. 1885405 #6-32 x .31" Phillips machine screw to mount Airflow Monitor Circuit Board 2 ea. 1891808 #6-32 x .50" Phillips thread cutting screw to mount Airflow Sensor
Exhaust transition connects to the enclosures with integral motorized impellers from the top so the duct can be routed up and outside the building. The transition is available for 6" PVC duct. The transition should be sealed to the top of the enclosure with silicone sealant to prevent leaks. See Figure 3-3. Order the following: Part Description 3924400 2' Exhaust Transition Kit 3924401 3' Exhaust Transition Kit 3924402 4' Exhaust Transition Kit 4. Remote Blowers For Purifier Class I Enclosure The remote blower has 1/4 hp direct drive motor and corrosion-resistant epoxy-coated steel housing and wheel with blower inlet of 5.94" OD. sized to accept 6" nominal PVC duct. Includes a blower transition adapter outlet for 6" PVC duct with integral back draft damper.
CFM @ Static Pressure-Inches of H2O S.P. 0.0" 0.125" 0.25" 0.50" 0.75" 0.87" CFM 595 560 515 420 300 167

Used With Catalog # Description Shipping Wt. (lbs./kg.) 3716000 Remote Blower, 115V, 60 Hz. 4.4 amps
3716001 Remote Blower, 115/230V, 50 Hz, 5.6/2.8 amps 35/15

For Purifier HEPA Filtered Enclosure Has a 1/4 hp direct drive motor and corrosion-resistant epoxy- coated steel housing and wheel with blower inlet of 6.00" ID. Outlet dimensions are 4.25" x 7.38" OD.
CFM @ Static Pressure-Inches of H2O S.P. 0.0" 0.125" 0.25" 0.50" 0.75" 0.87"
CFM 595 560 515 420 300 167 Catalog # Description Shipping Wt.
(lbs./kg.) 4863500 Remote Blower, 115 V, 60 Hz. 4.4 amps 35/16 4863501 Remote Blower, 115/230 V, 50 Hz, 5.6/2.8 amps 35/16 Used with Purifier
Enclosure 7053501 Explosion-Proof Remote Blower, 115 V, 60 Hz, 4.4 amps 40/18


Exhaust dampers allow adjustments required to maintain proper airflow for roof-mounted blowers or house exhaust systems. Catalog # Description 3924000 6" Epoxy Coated Steel In-Line adjustable damper 4724200 6" PVC In-line adjustable damper 6. Accessory Filters HEPA Filter HEPA Filter is 99.99% efficient on particles 0.3 micron.
Width Catalog # Size 2' 3707900 18" x 18" x 3.31" 3' 3707901 30" x 18" x 3.31" 4' 3707902 42" x 18" x 3.31" Odor Control Carbon Filter Provides granular activated carbon or impregnated carbon used for Odor Control applications on integral motorized impeller models exhausted back into the room. Filter Classification and Size Catalog # X Pounds 2' Organic 18 x 18 x 1 3937200 5.5 lbs. activated 3' Organic 30 x 18 x 1 3937300 9.2 lbs. activated 4' Organic 42 x 18 x 1 3937400 12.9 lbs. activated 2' Formaldehyde 18 x 18 x 1 3937201 7.0 lbs impregnated 3' Formaldehyde 30 x 18 x 1 3937301 12.0 lbs. impregnated 4' Formaldehyde 42 x 18 x 1 3937401 16.8 lbs. impregnated 2' Ammonia & Amines 18 x 18 x 1 3937202 8.0 lbs. impregnated 3' Ammonia & Amines 30 x 18 x 1 3937302 13.5 lbs. impregnated 4' Ammonia & Amines 42 x 18 x 1 3937402 18.9 lbs. impregnated

SOLVENT ACID Size/Description Dual Doors Right Hinge Left Hinge Dual Doors Right Hinge Left Hinge 48" 9902000 - - 9901000 - - 36" 9902100 - - 9901100 - - 30" 9902200 - - 9901200 - - 24" - 9902300 9902400 - 9901300 9901500 18" - 9901400 9901600 12" - 48" w/Self-Closing Doors 9903000 - 36" w/Self-Closing Doors 9903100 - 30" w/Self-Closing Doors 9903200 - 24" w/Self-Closing Doors - 9903300 9903400 - - - 24" ADA - 9906000 9906100 - 9905000 9905200 24" ADA w/Self-Closing Doors
- 9906200 9906300 - - -
18" ADA - 9905100 9905300 12" ADA -
STANDARD BASE VACUUM PUMP Size/Description Dual Doors Right Hinge Left Hinge Dual Doors Right Hinge Left Hinge 48" 9900000 - 36" 9900100 - 30" 9900200 - 24" - 9900300 9900600 - - - 18" - 9900400 9900700 - 9907000 9907100 12" - 9900500 9900800 - - - 48" w/Self-Closing Doors - 36" w/Self-Closing Doors - 30" w/Self-Closing Doors - 24" w/Self-Closing Doors - 24" ADA - 9904000 9904300 - - - 24" ADA w/Self-Closing Doors
-
18" ADA - 9904100 9904400 - - - 12" ADA - 9904200 9904500 - - -



Refer to the following table if your filtered enclosure fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance. PROBLEM CAUSE CORRECTIVE ACTION Contaminants outside of enclosure.
See “Certifying the Enclosure” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix D). Restriction of the baffle air slots or blockage of the exhaust outlet.
Improper user techniques for the enclosure.
Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed.
External factors are disrupting the enclosure airflow patterns or acting as a source of contamination.
See “Location Requirements” Chapter 2, “Certifying the Enclosure” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix D).
Enclosure has improper face velocity.
Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm for weighing procedures and 75-105 fpm for biohazardous procedures.


Plug the enclosure into appropriate electrical service. Circuit breaker(s) or Ground Fault Interrupter.
Plug enclosure into appropriate electrical service. Circuit breaker(s) tripped.
Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Verify that all airflow monitor interface cables are connected. Check fuses on enclosure. No audible alarm. Alarm has been temporarily silenced using “SILENCE ALARM”.

PROBLEM CAUSE CORRECTIVE ACTION Airflow Monitor Malfunction (Continued)
Airflow monitor was not properly adjusted. Repeat calibration steps outlined in this manual in Chapter 6. Constant audible alarm.
Wrong alarm set point.
Check airflow and calibration of airflow monitor. See Chapter 6. Continuous alarm. Check the face velocity of the enclosure as the airflow of the system may have changed. If incorrect, adjust the speed control to increase face velocity. The HEPA filter may have become loaded. If face velocity is correct, calibrate the airflow monitor outlined in this manual in Chapter 6. Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris.
Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components.
Audible disable will not stay operational.
An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower.

Customer Name Model Number
Date Installed Serial Number
Unit Location · Date Comments

APPENDIX A FILTERED ENCLOSURE COMPONENTS AND REPLACEMENT PARTS The components that are available for your filtered enclosure are listed. The parts shown are the most commonly requested. If other parts are required, please contact Product Service.

Item Qty. Part Number Description 1 1 3911000 Glass, Side 2A 1 3911900 Glass, Sash 2' 2B 1 3911901 Glass, Sash 3' 2C 1 3911902 Glass, Sash 4' 3A 1 3911800 Sash Assembly, 2' 3B 1 3911801 Sash Assembly, 3' 3C 1 3911802 Sash Assembly, 4' 4A 1 3905600 Lower Air Foil, 2' 4B 1 3905601 Lower Air Foil, 3' 4C 1 3905602 Lower Air Foil, 4' 5A 1 3907400 Bracket, Air Foil R.H. 5B 1 3907401 Bracket, Air Foil L.H. 6A 1 3907500 Stop, Sash R.H. 6B 1 3907501 Stop, Sash L.H. 7A 1 6916500 Latch, Sash 7B 1 3906700 Bracket, Latch 7C 1 1927403 Spring, Compression Latch 7D 2 1893206 Screw, 8-32 x .38" Lg. Phillips Pan S.S. Type F 9A 2 1889316 Screw, 10-24 x 1.00 PH Phil S.S. 9B 2 1912108 Washer, Plastic .194 ID. x .380 OD 9C 2 7868402 Bushing, Spacer .31 x .63 10 1.9 Ft 6913700 Wiper, Sash Seal 11A 2 1936800 Bushing, Heyco closed 1.50 Dia. 11B 2 1934601 Bushing, Heyco w/ Flex Shutter 1.50 Dia. 12A 1 3915401 Side Air Foil 12B 2 1889912 Screw, #6-32 x .75" Oval Hd. Type F (S.S.) 13A 1 3907700 Reflector, Light 2' 13B 1 3907701 Reflector, Light 3' 13C 1 3907702 Reflector, Light 4' 13D 2 1889308 Screw, Mach. 10-24 x .50 PH Phillips 14A 1 3708300 Lamp, Fluorescent 2' Enclosure (F15T8-SP35) 14B 1 9721902 Lamp, Fluorescent 3' Enclosure (F17T8) 14C 1 9721901 Lamp, Fluorescent 4' Enclosure (F25T8) 15A 1 1270100 Lamp, UV 2' or 3' Enclosure (G15T8) 15B 1 1271300 Lamp, UV 4' enclosure (G30T8) 16A 1 3911700 Bracket, Right Reflector Support 16B 1 3911701 Bracket, Left Reflector Support 16C 4 1893108 Screw, #8 x 1/2 AB Hex Washer Head

Item Qty. Part Number Description 17A 1 3918500 Wiring Harness Main, 115V 17B 1 3918600 Wiring Harness Main, 230V 18A 1 3811500 Guardian Airflow Monitor Printed Circuit Board 18B 2 1885405 Screw Mach #6-32 x .31 PH Phil. S.S. 19A 1 3926600 Label, Front Class I 19B 1 3926700 Label, Front HEPA 20A 1 3910700 Airflow Switch 20B 2 1891808 Screw, Thread Cutting #6-32 x .50 PH 21A 1 3906200 Cover Plate for HEPA 2' 21B 1 3906201 Cover Plate for HEPA 3' 21C 1 3906202 Cover Plate for HEPA 4' 21D Multiple 1905617 Nut, KEPS 10-24 22A 1 or 2 1307000 Switch, Rocker (2 position) 22B 1 1306300 Switch, Rocker (3 position) – UV Light Only 23A 1 or 2 3904800 Front Panel, 2' 23B 1 3904801 Front Panel, 3' 23C 1 3904802 Front Panel, 4' 23D 1 4472400 Screw, Thread cut 10-24 x .50 Type F PH Phillips 24A 1 3916500 Impeller, Motorized 2' or 3' Enclosure, 115V 24B 1 3916501 Impeller, Motorized 2' or 3' Enclosure, 230V 24C 1 3903300 Impeller, Motorized 4', 115V 24D 1 3903301 Impeller, Motorized 4', 230V 24E 4 1601800 Vibration Isolator Mount 25A 1 1306900 Capacitor, 2', 115V 10 MFD 25B 1 1306900 Capacitor, 3', 115V 10 MFD 25C 1 1306600 Capacitor, 4', 115V 15 MFD 25D 2 1306800 Capacitor, 2', 230V 4 MFD (effective 2MFD) 25E 2 1306800 Capacitor, 3', 230V 4 MFD (effective 2MFD) 25F 1 1306800 Capacitor, 4', 230V 4 MFD 26A 1 1235400 Ballast, Fluorescent 2', 115V 26B 1 1235500 Ballast, Fluorescent 2', 230V 26C 1 1294000 Ballast, Fluorescent 3' or 4', 115V or 230V 27A 1 1235400 Ballast, UV 2' or 3', 115V 27B 1 1235500 Ballast, UV 2' or 3', 230V 27C 1 1233400 Ballast, UV 4', 115V 27D 1 1233600 Ballast, UV 4', 230V 28A 1 1325500 Speed Control, 115V 28B 1 1325501 Speed Control, 230V 29A 1 3910800 Upper Diffuser Screen 2' 29B 1 3910801 Upper Diffuser Screen 3' 29C 1 3910802 Upper Diffuser Screen 4' 30A 2 3916700 Clip, Upper Diffuser

Item Qty. Part Number Description 30B 2 1893708 Screw, #10 x .50 PH Phil. Thd. Forming 31A 2 1881196 Capscrew, 3/8 –16 x 6.0 Lg. Hx. Hd. 31B 2 1924403 Sealing Washer 3/8 32A 1 3707900 HEPA Filter (2') 18 x 18 x 3.32 w/ Gasket 32B 1 3707901 HEPA Filter (3') 30 x 18 x 3.32 w/ Gasket 32C 1 3707902 HEPA Filter (4') 42 x 18 x 3.32 w/ Gasket 33A 1 3937200 2' Odor Control Organic Vapor Carbon Filter 33B 1 3937201 2' Odor Control Formaldehyde Carbon Filter 33C 1 3937202 2' Odor Control Ammonia Carbon Filter 33D 1 3937300 3' Odor Control Organic Vapor Carbon Filter 33E 1 3937301 3' Odor Control Formaldehyde Carbon Filter 33F 1 3937302 3' Odor Control Ammonia Carbon Filter 33G 1 3937400 4' Odor Control Organic Vapor Carbon Filter 33H 1 3937401 4' Odor Control Formaldehyde Carbon Filter 33I 1 3937402 4' Odor Control Ammonia Carbon Filter 34 1 7907802 Digital Monitor Outlet, 115V only, 230V connected at laboratory outlet location. 35A 1 1333800 IEC Power Inlet 35B 1 1305800 Power Cord, 115V, Right Angle 35C 1 1334200 Power Cord, 230V, Right Angle 36A 1 - (115V) 1327200 Circuit Breaker, 3 AMP 36B 2 - (230V) 1327205 Circuit Breaker, 2 AMP 40A 1 1952500 Gauge, Pressure Mini Helic II 40B 1 3667400 Bracket, Gauge 40C 2 1905617 Nut, #10-24 KEPS 40D 2 1551700 Tubing, 3/16" ID x 5/16" OD, PBC 7.00" Lg. 40E 2, 1" Long 3788200 Snubber, Filter Pressure Gauge

Use for Purifier Class I and Purifier HEPA Filtered Enclosures

33 Placed on ledge
above HEPA · filter

See the following dimensions and exhaust options for all the Purifier Filtered Enclosures.


Figure B-1 Purifier Class I and Purifier HEPA Filtered Enclosure


Figure B-2 Exhaust Duct Connection Kit (Optional Installation on Integral Blower Models
can be installed on Purifier Class I and Purifier
HEPA Filtered Enclosure)











APPENDIX D QUICK CHART FOR THE FILTERED ENCLOSURES To determine the actual face velocity at the sash opening, airflow velocity readings will need to be taken. This should be done across the sash opening of the enclosure in accordance with the Industrial Ventilation Manual. (See Appendix E) The “average face velocity” is achieved by taking readings in two rows across the enclosure with the readings 6" from the ends and evenly spaced every 12"; the first row is 3" down from the upper sash foil and the second row is 3" up from the work surface. A total of four readings will be taken for the 2' enclosure, six readings taken for the 3' enclosure and eight readings taken for the 4' enclosure and then averaged.
Model Size 2' 3' 4' Sash Height from Work Surface (inches) 9.44 9.44 9.44 Total Open Area with Bypass (sq. ft.) 1.45 2.20 2.95 Exhaust Airflow Volume at 90fpm (CFM) 130 200 265 Alarm Airflow Volume at 60 fpm (CFM) 85 130 175 Initial Mag. Gauge Pressure Reading (inches H2O) 0.20 to .045 0.15 to 0.40 0.10 to 0.35 Number of Laskin Nozzles needed at 10 psig 1 1 1 Exhaust HEPA Filter w/ Gasket Dimensions (inches) 18x18x3.31 30x18x3.31 42x18x3.31 *Motor Horsepower (HP) 0.13 0.13 0.30 *Capacitor (MFD, Volts) 10MFD, 115V (2) 4MFD, 230V
15MFD, 115V 4MFD, 230V Fluorescent Lights F15T8-SP35 F17T8 F25T8 UV Light on Class I only G15T8 G15T8 G30T8




General References American Conference of Governmental Industrial Hygienists. Industrial Ventilation, A Manual of Recommended Practice, Cincinnati, OH ASHRAE Standard Committee. ASHRAE Standard Atlanta: ASHRAE Publications Sales Department, 1995 British Standards Institution, Laboratory Fume Cupboards. Parts 1, 2 and 3, London: 1990 Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910, Occupational Exposures to Hazardous Chemicals in Laboratories, Final Rule. Vol. 55, No. 21. Washington D.C.:1990 DiBerardinis. L. et al. Guides for Laboratory Design, Health and Safety Considerations. Wiley & Sons, 1987 McDermott, Henry, Handbook of Ventilation for Contaminant Control, 2nd Edition. Butterworth Publishers, 1985. Miller, Brinton M. et al. Laboratory Safety: Principles and Practices. American Society for Microbiology, Washington, D.C.: 1986 NIH Guidelines for the Laboratory Use of Chemical Carcinogens. NIH Publication No. 81-2385. Rayburn, Stephen R. The Foundation of Laboratory Safety, A Guide for the Biomedical Laboratory. Springer-Verlag, New York: 1990 Sax, N. Irving and Lewis, JR., Richard J. Rapid Guide to Hazardous Chemicals in the Workplace. Van Nostrand Reinhold, 1987. Schilt, Alfred A. Perchloric Acid and Perchlorates. The G. Frederick Smith Chemical Company, Columbus, OH: 1979. Steere, Norman. CRC Handbook of Laboratory Safety, 2nd Edition. CRC Press, 1971. Collins, C.H., 1988. Laboratory Acquired Infections, History, Incidence, Causes and Prevention, 2nd ed., Butterworths, London.

Kruse, R.H., WH. Puckett and J. H. Richardson. 1991. Biological Safety Cabinetry, Clin. Microbiol. Rev. 4:207- 241. Miller, C.D., D.H.M. Groschel, J. H. Richardson, D. Vesley, J. R. Songer, R. D. Housewright and W. E. Barkley. 1986. Laboratory Safety, Principles and Practices, American Society for Microbiology, Washington, D.D. National Research Council (U.S.) Committee on Hazardous Biological Substances in the Laboratory. 1989. Biosafety in the Laboratory. Prudent Practices for Handling and Disposal of Infectious Materials. National Academy Press, Washington, D.C. Rayburn, S.R. 1990. The Foundations of Laboratory Safety, Springer-Verlag, New York. U. S. Department of Health, Education and Welfare. Formaldehyde Decontamination of Laminar Flow Biological Safety Cabinets, National Institutes of Health, Division of Safety, Bethesda, MD 20892. U. S. Department of Health, Education and Welfare. Effective use of a Laminar Flow Biological Safe Cabinet. National Audiovisual Center (GSA), Sales Branch, Catalog No., NAC 005133, Washington, D.C. U. S. Department of Health, Education and Welfare. 1975. Selecting a Biological Safety Cabinet. National Audiovisual Center (GSA), Sales Branch, Catalog No. NAC 000709, Washington, D.C. U. S. Department of Health and Human Services. 1999. Biosafety in Microbiological and Biomedical Laboratories, 4th ed. U.S. Department of Health and Human Services, Publication No. (CDC) 93-8395. U. S. Government Printing Office, Washington, D.C.



For more information, please contact us: ExpotechUSA 10700 Rockley Road Houston, Texas 77099 USA 281-496-0900 [voice] 281-496-0400 [fax] E-mail: sales@expotechusa.com Website: www.ExpotechUSA.com