’ S MANUAL
FUTREX ® -5500A/WL
BODY FAT AND FITNESS ANALYZER

FUTREX INC. 6 Montgomery Village Avenue Suite 620 Gaithersburg, MD 20879 Version 1.0



Appearance and Part Names




Light Wand AC Adapter





Light Shield RS-232 Cable Fitness

Insert 6 1.5V “AA” size alkaline batteries into the battery compartment located on the rear of the Main Body. Be sure the batteries are oriented as indicated in the battery compartment.
Connect the Light Wand Cable to the connector jack located on the right hand side of the Main Body.
It is recommended that the Light Wand always remain attached to the Main Body. This will avoid possible damage to the connector.
Connect the A/C Adapter cable to the connector jack located on the right side of the Main Body (the jack closest to the top of the unit).
Preparation

Remove the clear plastic cover from the printer.
Start feeding the paper into the slit inside the Paper Roll Compartment by pressing the PAPER FEED key.
Printer Paper Installation

Models of the FUTREX Body Composition Analyzer
There are three models available for estimating percent body fat in humans:
Estimates the percent body fat of Adults only (ages 18-80)
Estimates the percent body fat of:
1.) Adults 2.) Adolescent Children (13-17) 3.) Pre-Adolescent Children (5-12)
When age is entered, the FUTREX-5500A/ZL automatically sets the instrument to the proper calibration (e.g. if you entered the subject’s age as 14 during the Body Fat Measurement Sequence (page 7), the FUTREX-5500A/ZL will use its Adolescent calibration to determine percent body fat).
Includes a special calibration for high school aged athletes (13-23) as well as all features of the FUTREX-5000A/ZL

Percent Body Fat Measurement
1.) Turn Power Switch ON.
The display will show a software version number (e.g. AO.51) and will then start a 15 second countdown. The display will show Sd.
2.) Insert Light Wand into Optical Standard.
Be sure the silver stripe from the Light Wand is aligned with the silver stripe from the Optical Standard.
3.) Press the ZERO ADJUST Key. The display will now show 0.0. The zero adjustment has been made.
Remove Light Wand from the Optical Standard.
Install the Light Shield on the Wand as shown in the sketch (allow the Wand to protrude from the edge of the shield by about 1/8 of an inch). The Light Shield was provided to ensure that no external light interferes with the estimation of percent body fat. Therefore, this shield should always be used when estimating percent body fat.
The FUTREX-5500A/WL is ready to start the body fat estimation sequence.
4.) The printer will ask “Athletic Cal: 1 = Yes, 2 = No” If the desire is to measure an athlete between the ages of 13 and 23, press the “1” key, otherwise press the “2” key.

4a) The printer will ask “ENTER # SITES: 1 or 3”.
There are two alternative measurement methods.

4b) Press ENTER to start the body fat estimation sequence.
4c) Enter Weight in lbs. or kg. (refer to page 29 to switch between English and Metric Units). Press ENTER.
4d) Enter Height in inches or cm.
4e) Printer will show MEASURE BICEPS.
Place Light Wand firmly on biceps of the dominant arm (see page 15 for locating exact measurement point)
Use the Light Shield to block the external light. Always avoid measurements under strong lights (e.g. direct sunlight) because errors may occur.
4f) Press GREEN BUTTON on the Light Wand The display will flash a decimal point following the number “1a”.
Then the number “2a” will be displayed.

4g) Remove the Light Wand and replace it on the same point on the biceps. Press GREEN BUTTON on the Light Wand. The display will flash a decimal point, then the number “1b”.
4h) Place Light Wand firmly on Triceps of same arm (see page 16 for exact measurement point).
Press ENTER. The display will flash a decimal point. Don’t move the Light Wand until “2b” is displayed.
4i) Remove Light Wand and replace it on the same point on the triceps. Press ENTER. The display will flash a decimal point. Don’t move Light Wand until “1c” is displayed.
4j) Place Light Wand firmly on thigh at same side of body that the arm was measured; e.g. if right arm, then measure right thigh. (See page 16 for exact measurement point.)
4k) Remove Light Wand and replace it on the same point on the thigh. Press ENTER. Display will again flash a decimal point. Don’t move the Light Wand until percent fat is displayed and printing starts.

5.) Press ENTER to start the body fat estimation sequence.
(i.e. adults 18 and above, adolescents 13-17, pre-adolescents 5-12)
There are many methods for measuring Frame Size. Futrex does not recommend any specific method but does offer two techniques often used for determining frame size:
Small Frame Medium Frame Large Frame
Men Less than 8 inches 8 to 9.25 inches Greater than 9.25 inches
Women Less than 7.5 inches 7.5 to 8.75 inches Greater than 8.75 inches

The following procedure provides a quantitative method to determine a subject’s Physical Activity Rating.
A person’s F.I.T. value is a function of 3 parameters:
The F.I.T. value is calculated by multiplying all three parameters: F.I.T. = Frequency x Intensity x Time
The F.I.T. values help to account for the long term effects of physical activity.
When entering the F.I.T. values, you should consider what the individual’s average Frequency, Intensity and Time have been over the past six months.
For example, if an individual has not been particularly active for a short period of time (i.e. several weeks), but had been previously training for competitive athletics, you should take that previous training regimen into account for the F.I.T. values. Conversely, if an individual has been training hard for a short time (i.e. several weeks), but was much less active previously, you should take the previous light activity into account for the F.I.T. values.
11A) Enter Frequency Value. Press ENTER.
FREQUENCY 5 Daily or almost daily (6 or 7 times per week) 4 3 to 5 times per week 3 1 or 2 times per week 2 A few times per month 1 Less than once per month

11B) Enter Intensity Value. Press ENTER.
INTENSITY 5 Aerobic activities that result in sustained heavy breathing and perspiration (e.g. high impact aerobics, running, speed swimming, distance cycling). 4 Intermittent aerobic activities that result in sustained heavy breathing and perspiration (e.g. tennis, racquet-ball, squash). 3 Moderate aerobic activity (e.g. normal bike riding, jogging, low impact aerobics). 2 Moderate aerobic activity (e.g. recreational volleyball, moderate speed walking). 1 Light aerobic activity (e.g. normal walking, golfing).
11C) Enter Time Value. Press ENTER.
TIME 4 Over 30 minutes 3 20 - 30 minutes 2 10 - 20 minutes 1 Under 10 minutes

Use the Light Shield to block the external light. Always avoid measurements under strong lights (e.g. direct sunlight) because errors may occur.



Men between 6 - 24% Fat; 69 - 57% Water Women between 9 - 35% Fat; 67 - 50% Water
To Repeat the Measurement Sequence - Press ENTER and follow the steps outlined under 5-16.
OFF Message - An Off Message will appear and the FUTREX-5500A/WL Automatic Power Down feature will turn the unit off if the user has not pressed any key for 6 minutes. To use the unit again, turn the Power Switch OFF and ON again.

Selecting the Measurement Site

Prior to conducting the actual body composition analysis, you should locate the proper measurement site at the biceps of the dominant arm.
Placing the Biceps Locator

Biceps Locator

Selecting the Triceps Measuring Site
(for use with 3-site test using the athletic calibration)
1) Have the test subject bend his/her arm at a 90 degree angle, then using a tape measure or ruler of adequate length, measure from the top of shoulder to the tip of elbow. 2) The measurement point will be half way between these two points. For example, if the distance from the top of the shoulder to the tip of the elbow is 16 inches, then the measurement point will be at the 8 inch mark. 3) It is recommended that you mark this spot by placing a small line on the spot with a highlighter or washable marker.
Selecting the Thigh Measuring Site
(for use with 3-site test using the athletic calibration)
1) Have the test subject sit down, then using a tape measure or ruler of adequate length, measure from the groin line to the top of the patella (knee). 2) The measurement point will be half way between these two points. For example, if the distance from the groin line to the top of the patella (knee) is 20 inches, then the measurement point will be at the 10 inch mark. 3) It is recommended that you mark this spot by placing a small line on the spot with a highlighter or washable marker.

Adjusting The Percent Body Fat Reading
The FUTREX-5500A/WL has been carefully calibrated to provide body fat estimations. The calibration is traceable to the generally accepted hydrostatic procedure.
However, some users may desire to have the FUTREX-5500A/ZL estimations traceable to their own laboratory. For those users the following adjustment procedure is provided:
Press ENTER

Estimating Relative Fat Distribution
In addition to estimating percent body fat, a relative body fat estimation on other sites of the body (e.g. quadriceps, abdominal region) can be obtained using the FUTREX-5500A/WL.

monitor changes in fat distribution within the location estimated.
To Estimate Another Site or Repeat an Estimation - Place the Light Wand firmly on the particular body site and follow steps outlined in 6 and 7. OFF Message - An OFF Message will appear and the FUTREX-5500A/WL Automatic Power Down feature will turn the unit OFF if the user has not pressed any key for 6 minutes. To use the unit again, turn the Power Switch OFF and then ON again.

The FUTREX-5500AWL allows you to display the date on each printed test result. The default setting from the factory is to not display the date.
If you wish to display the date, you must first activate that feature:
If you activate the Date function, the FUTREX-5500A/WL requires that the date be manually changed each day. Use the following procedure to change the date:
Displaying the Date · Changing the Date

Connecting the Futrex to a Computer
The FUTREX-5500A/WL allows you to communicate with a PC to make use of software that is included with your FUTREX-5500A/ WL. In order to connect to the PC, you must use the RS-232 cable that was included with your instrument.

Illustration of a Typical Serial Connection on the Back of a PC FUTREX-5500A/WL
RS-232 Cable

Using its built-in thermal printer, the FUTREX-5500A/WL can provide two analysis programs for Adult User’s only:
The Fitness Analysis is based upon the Metropolitan Life Insurance Tables combined with the National Institutes of Health published data on the interrelationship between body weight/fat and longevity/health risks. The program prints diet and activity suggestions to allow you to reach the recommended fat/weight levels.
This analysis provides the user with information that covers a broad spectrum of the technical literature. This is deemed necessary, because there is no “one best method” of performing a fitness analysis.
To Obtain a Fitness Analysis Printout on the FUTREX-5500A/WL Built-In Thermal Printer:
Printout Programs · Fitness Analysis

To Estimate Percent Body Fat - Turn the FUTREX-5500A/WL OFF and then ON again.
OFF Message - An OFF Message will appear and the FUTREX-5500A/WL Automatic Power Down feature will turn the unit OFF if the user has not pressed any key for 6 minutes. To use the unit again, turn the Power Switch OFF and then ON again.

To Obtain a Fitness Analysis Printout using the optional Epson Printer Kit
The FUTREX-5500A/WL can print the Fitness Analysis Report directly to an external printer without need for any external computer. The following is required if you desire to use this feature:
This printing function is performed as follows: (NOTE: F4 must have been set to “1”; see page 27).
1.) Turn on both the printer and FUTREX-5500A/WL. Wait for countdown to finish or press CLEAR. The RS-232 serial cable must be attached between the units.
2.) Press SELECT PROGRAMS, then COLOR PRINTER. The FUTREX-5500A/ZL internal printer will display:
“Use Previous Data; 1 = YES, 2 = NO”. If you type in “NO” (i.e. “2”), follow the Fitness Analysis Procedure described on page 18 starting with Step 5.
3.) If “YES” (i.e. “1”) was typed in, the FUTREX-5500A/WL will use the data from the last person that was measured and the instrument’s printer will ask:
4.) The Epson Printer will then print the Fitness Analysis Report. CAUTION: Paper must be in printer.

This printout (for Adult Users only) allows you to compare your previous weight and percent body fat with your current values. It is available only from the FUTREX-5500A/WL’s built in thermal printer.
To Obtain an Update Analysis Printout:
Update Fitness Analysis

To Print Another Update Analysis: follow steps 3 through 16.
To Estimate Percent Body Fat - turn the FUTREX-5500A/WL OFF and then ON again.
OFF Message - An Off Message will appear and the FUTREX-5500A/WL Automatic Power Down feature will turn the unit OFF if the user has not pressed any key for 6 minutes. To use the unit again, turn the Power Switch OFF and then ON again.

This printout allows you to print a fitness analysis report for children from the ages of 5 to 17. To obtain a Youth Analysis Printout:
Youth Fitness Analysis

Color Fitness Analysis Reports
The FUTREX-5500A/WL can directly communicate with an external printer without need for any external computer using Futrex’s unique full color fitness forms. The following is required if you desire to use this feature:
This printing function is performed as follows: (NOTE: F4 must have been set to “0”; see page 26).
“Use Previous Data; 1 = YES, 2 = NO”. If you type in “NO” (i.e. “2”), follow the Fitness Analysis Procedure described on page 18 starting with Step 5.

The Function Mode provides the user with 7 program options:
F1 - Selection of English or Metric Units
The subject’s height and weight values can be printed in either English (lbs/in) or Metric (kg/cm) units.
After entering the Function Mode:
a) Press 1 The display will show either “F1 0” or “F1 1”.
If “F1 0” is displayed, English Units are being used (lbs/in). If “F1 1” is displayed, Metric Units are being used (kg/cm).
b) To Change Between Units Press either “0” or “1” and the display will show the change.
c) Press ENTER if this is correct.
d) To Exit the Function Mode, turn the FUTREX-5500A/WL OFF.
Function Mode

F2 - Selection Calorie Limit Mode
The Fitness Analysis Program can be operated in two different calorie modes. The first mode allows the FUTREX-5500A/WL to calculate the subject’s daily caloric intake. The second mode allows the subject to enter his/her daily caloric intake value.
After entering the Function Mode:
a) Press “2” The display will show either “F2 0” or F2 1".
If “F2 0” is displayed, then the unit will automatically calculate the subject’s daily caloric intake. If “F2 1” is displayed, the user will be asked to enter a specific calorie intake value.
b) To Change Between Modes: Press either “0” or “1” and the display will show the change.
c) Press ENTER if this is correct.
d) To Exit the Function Mode, turn the FUTREX-5500A/WL OFF.

F3 - Use of Dotson-Davis Equations
After entering the Function Mode:
a) Press “3.” The display will show either “F3 0” or “F3 1”.
If “F3 0” is displayed, the Dotson-Davis Equation will be used.
If “F3 1” is displayed, the Dotson-Davis Equation will be omitted from the Body Fat Estimation Program.
b) To Change Between Equations: Press either “0” or “1” and the display will show the change. Press ENTER if this is the correct change.
c) To Exit the Function Mode, turn the unit OFF.
F4 - Color Fitness Analysis/Standard Fitness Analysis Selection
After entering the Function Mode:
a) Press “4” If “0” is displayed, the unit is set to print using the Color Forms. If “1” is displayed, unit is set to print Fitness Analysis on plain paper.
b) To change, press “0” or “1”. Press ENTER to save the change.
c) To Exit the Function Mode, turn the unit OFF.

F5 & F6 - F.I.T. Value Settings
You may utilize the F5 and F6 settings to customize how your FUTREX-5500A/WL asks the F.I.T. Value questions. The following table provides the F.I.T. settings for each configuration:
F5 F6 Setting 0 0 No Exercise Questions 1 0 One Exercise Level Question (Light, Moderate, Heavy) 0 1 F.I.T. Values (entered by operator for each reading) 1 1 F.I.T. Values (pre-set values entered by operator)*
* NOTE: If you select this configuration, you will be prompted by the FUTREX-5500A/WL to pre-set the F.I.T. levels once you have set both F5 and F6 to “1.” You will see the following output from the printer:
F.I.T. Mode: Preset Frequency = 2 Intensity = 3 Time = 2 Change? (1 = Yes 2 = No)
If these are the values you want simply press “2.” If you want to change the values, press “1” and the instrument will prompt you for each F.I.T Value
Important Note It is important to remember that once the F5 and F6 values are both set to “1,” every time you enter the Function setting mode and press “ENTER,” the instrument will prompt you to verify the F.I.T. Settings. If, for example, you use the F1 setting to switch from English to Metric measurement units, once you have made that F setting change and pressed ENTER, the printer will begin to verify the F.I.T. values. You may simply turn the instrument off at this time to log the F setting change that you did make, and no change will be made to the pre-set F.I.T. Values.

F7 - Date Display
After entering the Function Mode:
a) Press “7” If “0” is displayed, the unit will display the date on the printout. If “1” is displayed, unit will not display the date on the printout.
b) To change, press “0” or “1”. Press ENTER to save the change.
c) To Exit the Function Mode, turn the unit OFF.
F8 - Data Stream
This function allows the FUTREX-5500A/WL measured data to be sent to an external PC computer that is utilizing Futrex’s Bodyometry™ software or other compatible software.
After entering the Function Mode:
a) Press “8” If “0” is displayed, no data is being sent. If “1” is displayed, all measured data will be sent to the external computer.
b) To change, press “0” or “1”. Press ENTER to save the change.
c) To Exit the Function Mode, turn the unit OFF.

a) Be sure the Light Wand is properly inserted into to Probe Socket
b) Too much/too little light
b) Be sure the Light Wand is properly seated in the Optical Standard when Zero Adjusting.
Troubleshooting

a) Ensure that there us enough pressure applied between the Light Wand and the subject’s arm
b) Incorrect Zero Adjustment Procedure
b) When taking a measurement, ensure that the Silver Stripe on the Light Shield is pointing towards the subject’s shoulder (the Silver Stripe on the Light Wand should be aligned with the Silver Stripe on the Light Shield).
c) Not measuring at the proper location on the arm
c) Always use the Light Shield to block external light.
d) Avoid taking measurements under strong lights (e.g. direct sunlight) as errors may occur
e) When Zero Adjusting, ensure that the silver line on the Light Wand is aligned with the silver line on the Optical Standard.
f) When Zero Adjusting, make sure that the Light Shield has been removed from the Light Wand. This is to ensure that the Light Wand is properly seated in the Optical Standard.
g) Always use the Biceps Locator to find the measurement site. Instructions can be found on page 12.

a) Too much light present during the measurement.
a) Do not lift the Light Wand from the subject’s biceps until a “2” appears on the screen during the first estimation or the printer starts printing after the second estimation.
b) Use the Light Shield to block the external light. Always avoid measurements under strong lights (e.g. direct sunlight) because errors may occur.
a) Replace batteries (see page 4).
b) Bad battery contact.
b) Remove batteries and clean tabs with isopropyl alcohol (rubbing alcohol).
c) Bad AC connection.
c) Unplug the AC Adapter from the wall socket and from the unit. Clean the adapter contact with isopropyl alcohol (rubbing alcohol).

a) Make sure paper is inserted to print on the “shiny” side of the paper.
b) Wrong type of printer paper.
b) Paper must be thermal paper.
c) Print head is dirty.
c) Clean the print head by carefully removing the Paper Cutter. Insert a clean piece of rough paper (a typical business card is ideal) between the print head and rear paper roller.
a) Replace batteries (Refer to page 4).
b) Bad battery connections.
b) Remove batteries and clean tabs with isopropyl alcohol (rubbing alcohol).

Department of Agriculture. • Estimation Range: 2.5% to 50%
(one roll is supplied installed in the instrument’s printer. Paper must be thermal paper type)
General Specifications

Summary:
Key Name ZERO ADJUST : When the Display shows “Sd”, insert the probe firmly into the “Optical Standard”, and press this key to automatically calibrate the unit.
SELECT PROGRAM: This key is pressed prior to the Program Keys in order to select one of the “Computation Programs”.
FITNESS : One of the “Computation Programs”. This key selects the “Fitness Analysis” program. UPDATE : One of the “Computation Programs”. This key selects the “Update Analysis” program.
COLOR PRINTER: Activates RS-232 Serial Port to send data to be printed onto the four page preprinted color form.
SMALL : These keys are used to enter your body frame size in the “Fitness Analysis” or MEDIUM the “Update Analysis” programs. Press the proper key when the “Enter body frame” message LARGE appears on the printer.
MALE : These keys are used to enter your sex into the “Fitness Analysis” or “Update Analysis” FEMALE programs. Press the appropriate key when the “Enter sex” message appears on the printer.
CLEAR : This key cancels incorrectly entered numbers. ENTER : This key is used to store numbers that have been entered. It is also used to start the body fat estimation sequence.
PAPER FEED : For feeding paper into the printer.
RISKS : Allows adjustment of the instrument’s calibration.
Keypad Descriptions

Printed In U.S.A.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| 1) Display shows “LOBAT” 2) Display shows “1999.9” or “-1999.9” | Batteries are low | p.34 | |
| 1) Display shows “LOBAT” 2) Display shows “1999.9” or “-1999.9” | a) Bad connection between Light Wand and Probe Socket b) Too much/too little light | p.34 | |
| 3) Bad Repeatability | a) Inconsistent Measurement Technique b) Incorrect Zero Adjustment Procedure c) Not measuring at the proper location on the arm | p.35 | |
| 4) Unusually high body fat estimates (40-50%) | a) Too much light present during the measurement. | p.36 | |
| 5) Printer Stops | a) Low batteries. b) Bad battery contact. c) Bad AC connection. | p.36 | |
| 6) Print is too light or does not print. | p.37 | ||
| 7) Printer prints black lines. | p.37 |