Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 1 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Part Number 3-9000-522
Page 1 · click to enlarge
Page 1

MON2000 Software for Gas Chromatographs

User Manual

Applies to BothDaniel Danalyzer On-Line Gas Chromatographs Rosemount Analytical Process Gas Chromatographs

Part Number 3-9000-522

Revision R JULY 2010

Emerson Process Management Chromatograph Software (MON2000) — Part Number 3-9000-522 diagram (page 1)
Fig. 1 · Part Number 3-9000-522
Page 2 of the Emerson Process Management Chromatograph Software (MON2000) user manual
Page 2 · click to enlarge
Page 2
Page 2 of the Emerson Process Management Chromatograph Software (MON2000) user manual
Page 3 of the Emerson Process Management Chromatograph Software (MON2000) user manual — NOTICE
Page 3 · click to enlarge
Page 3

MON2000 Software for Gas Chromatographs

Manual

NOTICE

DANIEL MEASUREMENT AND CONTROL, INC. AND ROSEMOUNT ANALYTICAL (COLLECTIVELY, “SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL. SELLER MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL SELLER BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, ETC.

PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES.

THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.

SELLER DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY SELLER PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.

DANIEL AND THE DANIEL LOGO ARE REGISTERED TRADEMARKS OF DANIEL MEASUREMENT AND CONTROL, INC. ROSEMOUNT AND THE ROSEMOUNT ANALYTICAL LOGO ARE REGISTERED TRADEMARKS OF ROSEMOUNT ANALYTICAL. THE EMERSON LOGO IS A TRADEMARK AND SERVICE MARK OF EMERSON ELECTRIC CO.

COPYRIGHT © 2010 BY DANIEL MEASUREMENT AND CONTROL, INC., HOUSTON, TEXAS, U.S.A.

All rights reserved. No part of this work may be reproduced or copied in any form or by any means - graphic, electronic, or mechanical — without first receiving the written permission of Daniel Measurement and Control, Inc. Houston, Texas, U.S.A.

Page 4 of the Emerson Process Management Chromatograph Software (MON2000) user manual — WARRANTY
Page 4 · click to enlarge
Page 4

WARRANTY

1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly provided herein, Daniel Measurement and Control, Inc. and Rosemount Analytical, (collectively“Seller”) warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services. Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller rep- resentative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller. Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESSED OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. It is understood that corrosion or erosion of materials is not covered by our guarantee.

2. LIMITATION OF REMEDY AND LIABILITY: SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY HEREUNDER SHALL BE LIMITED TO REPAIR, CORRECTION, REPLACEMENT OR REFUND OF PURCHASE PRICE UNDER THE LIMITED WARRANTY CLAUSE IN SECTION 1 HEREIN. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES PROVIDED BY SELLER GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES. THE TERM "CONSEQUENTIAL DAMAGES" SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE AND COST OF CAPITAL.

Page 5 of the Emerson Process Management Chromatograph Software (MON2000) user manual — IMPORTANT INSTRUCTIONS
Page 5 · click to enlarge
Page 5

IMPORTANT INSTRUCTIONS

  • Read all instructions prior to installing, operating, and servicing this product.
  • Follow all warnings, cautions, and instructions marked on and supplied with this product.
  • Inspect the equipment packing case and if damage exists, notify your local carrier for liability.
  • Open the packing list and carefully remove equipment and spare or replacement parts from the case. Inspect all equipment for damage and missing parts.
  • If items are damaged or missing, contact the manufacturer at 1 (713) 827-6314 for instructions about receiving replacement parts.
  • Install equipment as specified per the installation instructions and per applicable local and national codes. All connections shall be made to proper electrical and pressure sources.
  • Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent personal injury.
  • Use of this product for any purpose other than its intended purpose may result in property damage and/or serious injury or death.
  • Before opening the flameproof enclosure in a flammable atmosphere, the electrical circuits must be interrupted.
  • Repairs must be performed using only authorized replacement parts as specified by the manufacturer. Use of unauthorized parts can affect the product's performance and place the safe operation of the product at risk.
  • When installing or servicing ATEX-certified units, the ATEX approval applies only to equipment without cable glands. When mounting the flameproof enclosures in a hazardous area, only flameproof cable glands certified to IEC 60079-1 must be used.
  • Technical assistance is available 24 hours a day, 7 days a week by calling 1 (713) 8276314.
Page 6 of the Emerson Process Management Chromatograph Software (MON2000) user manual
Page 6 · click to enlarge
Page 6

This page is intentionally left blank.

Page 7 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of Contents
Page 7 · Contents
Contents

Contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 8 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 8 · click to enlarge
Page 8

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 9 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 9 · click to enlarge
Page 9

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 10 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 10 · click to enlarge
Page 10

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 11 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 11 · click to enlarge
Page 11

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 12 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 12 · click to enlarge
Page 12

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 13 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 13 · click to enlarge
Page 13

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 14 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 14 · click to enlarge
Page 14

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 15 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 15 · click to enlarge
Page 15

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 16 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.5 Disconnect Power and Disassemble............... E-10
Page 16 · click to enlarge
Page 16

x TABLE OF CONTENTS

C.2 Peak Detection C-2

C.3 Analysis Computations ..................................C-5

C.3.1 Conc. Analysis with Response Factor C-5

Calibration C-5 Calc. in Mole Percent w/o Normalization C-7

C.4 Post Analysis Computations C-9

C.4.1 Liquid Equivalent Computations C-9

C.4.2 Heating Value Calculations C-9

C.4.3 Multi-Level Calibration C-17

C.4.4 Indirect Calibration C-18

APPENDIX D, ANALOG OUTPUT CAL. FOR 2350A

D.1 Calibrating by Volts.......................................D-1

D.2 Calibrating by Percentages .............................D-5

APPENDIX E, UPGRADE

E.1 Connect to GC and Halt Analysis .................... E-1

2350A GC S/W AND 2350 EPROMS

E.2 Offline Edit to Upload App. & Rename ............. E-1

E.3 Upgrade User-Defined Applications.................. E-3

E.4 Upgrade from Disk ........................................ E-3

E.4.1 Upgrade from GC Controller E-6

E.4.2 Configure GC Connection E-9

E.5 Disconnect Power and Disassemble E-10

E.6 Replace EPROMS/Reset CPU ........................ E-12

E.7 Reassembly Procedures ............................... E-13

E.8 Set-Up and Programming ............................. E-14

E.8.1 Procedure E-14

E.8.2 Reprogramming the DiskOnChip E-16

E.8.3 GC Reassembly E-17

Appendix D, Analog Output Cal. For 2350A JULY 2010

Page 17 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F.2 User_Modbus Register List ............................. F-3
Page 17 · click to enlarge
Page 17

TABLE OF CONTENTS xi MON2000

E.9 Connect to GC for Upgraded App E-18

E.10 Guide to Standard Application Files ............... E-19

E.10.1 Importance of Standard Application Files E-19

E.10.2 Standard Applications v1.50 and Later E-20

E.10.3 Standard Applications Prior to v1.50 E-22

APPENDIX F, MODBUS REG. LIST FOR 2350A GC

F.1 Introduction – SIM_2251 & User_Modbus ........ F-1

F.1.1 Notes on User_Modbus F-1

F.1.2 Notes on SIM_2251 Modbus F-2

F.2 User_Modbus Register List F-3

F.2.1 SIM_2251 Modbus Register List F-6

JULY 2010 Appendix F, Modbus Reg. List for 2350A GC

Page 18 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of Contents
Page 18 · click to enlarge
Page 18

xii TABLE OF CONTENTS

This page is intentionally left blank.

Appendix F, Modbus Reg. List for 2350A GC JULY 2010

Page 19 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Section 1 – Introduction
Page 19 · click to enlarge
Page 19

INTRODUCTION 1-1 MON2000

INTRODUCTION

Welcome to the MON2000 User Manual (P/N 3-9000-522), a user guide that accompanies the MON2000 software produced by Daniel Measurement and Control, Inc., a Division of Emerson Process Management.

Use this manual for installing the MON2000 and Modbus Test (WinMB) software programs, getting started, checking various gas chromatograph (GC) application settings, and configuring and monitoring your GC system.

1.1 DESCRIPTION OF MANUAL

See the following section summaries or the Table of Contents for more information.

Section 1 – Introduction

This section includes:

  • summary listing of the manual sections
  • description of the MON2000 User Guide online help file
  • description of the MON2000 software program

JULY 2010 Description of Manual

Page 20 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Section 4 – Control Functions
Page 20 · click to enlarge
Page 20

1-2 INTRODUCTION

Section 2 – Installation and Startup

This section includes:

  • short description of the MON2000 software
  • minimum system requirements for installing MON2000 on a 32-bit Microsoft ® Windows ® platform
  • installation and Startup instructions
  • establishing communications and Logon procedures
  • instructions for customizing MON2000 and setting up security
  • instructions for downloading an application and using the Offline Edit function to change an application’s configuration

Section 3 – Getting Started

This section includes:

  • instructions on how to log on and log off
  • navigation instructions for MON2000
  • listing of available procedures and keyboard shortcuts
  • access and navigation instructions for the MON2000 User Guide online help file

Section 4 – Control Functions

This section includes:

  • description of available Control functions and detailed step procedures

Description of Manual JULY 2010

Page 21 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Section 8 – Logs
Page 21 · click to enlarge
Page 21

INTRODUCTION 1-3 MON2000

Section 5 – Application Functions

This section includes:

  • description of available GC Application functions and detailed step procedures

Section 6 – Chromatogram Viewer

This section includes:

  • description of available Chromatogram Viewer functions and detailed step procedures to view, save, and print chromatograms

Section 7 – Reports

This section includes:

  • descriptions of available reports and sample outputs
  • instructions on how to generate and print a given report

Section 8 – Logs

This section includes:

  • description of Maintenance, Alarm, and Event logs
  • description of the Parameter List
  • step procedures for viewing, editing, and clearing logs

JULY 2010 Description of Manual

Page 22 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Section 10 – Modbus Test
Page 22 · click to enlarge
Page 22

1-4 INTRODUCTION

Section 9 – Data Collection

This section includes:

  • descriptions of Data Collection and AutoSequencing
  • instructions on how to configure, generate and run the Data Collection and AutoSequencing function using MON2000 PLUS

Section 10 – Modbus Test

This section includes:

  • short description of the WinMB software
  • minimum system requirements for installing WinMB on a 32-bit Microsoft ® Windows ® platform
  • installation and start up instructions
  • description of available Modbus Test functions and detailed step procedures
  • description of the WinMB online help file and navigation instructions

Description of Manual JULY 2010

Page 23 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Appendix F, Modbus Reg. List for 2350A GC
Page 23 · click to enlarge
Page 23

INTRODUCTION 1-5 MON2000

Appendix A, PC Config Report

This appendix provides a sample PC Config Report for reference only.

Appendix B, Component Data Table

This appendix provides two sample Component Data Tables for reference only.

Appendix C, Data Computations

This appendix discusses the various equations and computations involved with acquisition and analysis tasks.

Appendix D, Analog Output Cal. 2350A GC

This appendix demonstrates how to calibrate an analog output for a 2350A GC Controller, using the MON2000 software.

Appendix E, Upgrade 2350A GC S/W and 2350 EPROMS

This appendix describes how to upgrade the GC software and the 2350 GC Controller EPROMS, the 2350A WinSystems CPU, P/C104 Bus, and DiskOnChip.

Appendix F, Modbus Reg. List for 2350A GC

This appendix explains the differences between SIM_2251 and User_Modbus protocols, and lists the corresponding Modbus registers used by the 2350A GC Controller.

JULY 2010 Description of Manual

Page 24 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 1.3 DESCRIPTION OF MON2000
Page 24 · click to enlarge
Page 24

1-6 INTRODUCTION

1.2 DESCRIPTION OF ONLINE HELP

Use the User Guide online help file to quickly access information regarding any MON2000 function. See Section 3.8 for more detailed information.

1.3 DESCRIPTION OF MON2000

MON2000 is a menu-driven, Windows-based software program designed to operate the gas chromatograph (GCs). MON2000 runs on an IBM-compatible personal computer (PC) and serves as an interface between you and the GC unit. MON2000 can run on a Windows® 95, Windows® 98, Windows® 2000, Windows® XP or Windows® NT operating system.

MON2000, combined with the GC applica- tion(s), offers a complete software package for operating and monitoring one or more GC systems from a single PC. MON2000 includes security features to help prevent unauthorized access to GC data or control. MON2000 also includes, for trouble-shooting purposes, an auxiliary program (WinMB) to selectively poll the GC Modbus registers (see Section 10 for more information).

When configured for RS-485 multi-drop networking, MON2000 can interface with up to 32 GC units, either in a stand-alone configuration or via a network.

When configured for Ethernet networking MON2000 can interface with a number of GC Units limited only by the number of available TCP/IP addresses.

GC functions that can be initiated or controlled by MON2000 include:

  • alarm parameters
  • alarm and event processing
  • analog scale adjustments
  • analyses
  • baseline runs
  • calculation assignments and configurations
  • calibrations
  • component assignments and configurations

Description of Online Help JULY 2010

Page 25 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INTRODUCTION 1-7 MON2000
Page 25 · click to enlarge
Page 25

INTRODUCTION 1-7 MON2000

  • diagnostics
  • event sequences
  • halt operations
  • stream assignments and sequences
  • valve activations
  • timing adjustments

Reports that can be generated by MON2000 per the GC application:

  • 24-Hour Averages
  • Analysis
  • Calibration
  • Final Calibration
  • Hourly Averages
  • Monthly Averages
  • PC Configuration
  • Raw Data
  • Variable Averages
  • Weekly Averages

Logs that are maintained by MON2000:

  • Alarm Log
  • Event Log
  • Parameter List
  • Maintenance Log

JULY 2010 Description of MON2000

Page 26 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 1-8 INTRODUCTION
Page 26 · click to enlarge
Page 26

1-8 INTRODUCTION

This page is intentionally left blank.

Description of MON2000 JULY 2010

Page 27 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.1 SYSTEM REQUIREMENTS
Page 27 · click to enlarge
Page 27

INSTALLATION AND SETUP 2-1 MON2000

INSTALLATION AND SETUP

This section lists the system requirements to run MON2000 and provides installation procedures as well as initial logon instructions, communications setup, and software configuration.

2.1 SYSTEM REQUIREMENTS

To achieve maximum performance when running the MON2000 software, ensure your PC system contains the following hardware.

  • PC with a 486/90MHz or higher processor (Pentium/100MHz or higher recommended) running either

Windows® 95 (service pack 1 or better) or later equipment and Microsoft® Windows® operating system.

Microsoft Internet Explorer 5.0 (or later) is required to view spreadsheets or reports saved in HTML format.

If running Windows® 95 with the 2350A optional Ethernet card, the user must download Socket 2 from www.microsoft.com/windows95/downloads to utilize MON2000’s ethernet feature.

JULY 2010 System Requirements

Page 28 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-2 INSTALLATION AND SETUP
Page 28 · click to enlarge
Page 28

2-2 INSTALLATION AND SETUP

- Windows® 98 version 1 or later

- Windows® 2000 version 1 or later

- Windows® XP version 1 or later (see note for system requirements) - Windows® NT Server version 4 (service pack 3 or later)

Use the Settings>Control Panel>System/>General Page menu to check the system version number.

For Windows® 95, the version number should be 4.00.950A/B or later.

For Windows® 98 or Windows® 2000, the version number should be 1 or later.

To use Windows® XP you need:

  • Computer/Processor PC with 300 MHz or higher processor clock speed recommended;
  • 233 MHz minimum required (single or dual processor system)
  • *Intel ® Pentium ® /Celeron TM family, or AMD-K6 ® /AMD Athlon TM /AMD Duron TM family, or compatible processor recommended.
  • Memory: 128 MB of RAM or higher recommended (64 MB minimum supported; may limit performance and some features).
  • Hard Disk Minimum: 1.5 GB of available hard disk space.

For Windows® NT4, the version number should be 4.00.1381 or later.

  • 16 megabytes (MB) of RAM (32 MB or higher recommended)
  • 5 MB of free hard disk space
  • Super VGA monitor with 800x600 resolution
  • free serial port for remote/local connection to gas chromatograph (for online operations)
  • free parallel port for connection to printer
  • Windows ® -compatible modem (for remote connection only)
  • Ethernet connection (when using Ethernet networking)

System Requirements JULY 2010

Page 29 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-3 MON2000
Page 29 · click to enlarge
Page 29

INSTALLATION AND SETUP 2-3 MON2000

2.2 INSTALLING MON2000

To install MON2000, do the following:

If you are upgrading MON2000, you must install the new software with same directory as the current version.

1. Either place the MON2000 CD in your CD-ROM drive or insert Installation Disk 1 into your floppy drive.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-3 MON2000 diagram (page 29)
Fig. 2
2. Launch Windows ® Explorer ® by either:
  • Clicking the button (see the taskbar) to access the Programs menu option.
  • Double-clicking the Windows ® Explorer ® icon on your desktop.
3. Access either your CD-ROM drive or floppy drive.
4. Double-click the file “setup.exe”.
5. The Installation Wizard begins. Follow the instructions provided on each screen.

Upon successful installation, Windows® automatically creates a MON2000 icon on your desktop.

JULY 2010 Installing MON2000

Page 30 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.4 COPYING MON2000 TO DISKS
Page 30 · click to enlarge
Page 30

2-4 INSTALLATION AND SETUP

2.3 INSTALLING THE CRYPKEY LICENSE SERVICE FOR MON2000 PLUS

The CrypKey License Service must be installed on all NT-based systems. If you try to run MON2000 PLUS before installing the CrypKey License Service, an error will result.

Emerson Process Management Chromatograph Software (MON2000) — 2-4 INSTALLATION AND SETUP diagram (page 30)
Fig. 3

To install the CrypKey License Service, do the following:

1. After installing MON2000 PLUS, the program folder should display automatically; if it doesn’t, open Windows ® Explorer ® and access the MON2000 PLUS program folder.
2. Double-click the SETUPEX.EXE file.

MON2000 PLUS can now be started. You have 30 days to evaluate the application. Once the trial period expires, you must obtain a license to continue to use MON2000 PLUS.

for more information, refer to the MON2000 PLUS Quick Start Guide (3-9000-743).

2.4 COPYING MON2000 TO DISKS

Use this procedure to copy the MON2000 installation files to 3 floppy disks for future installation.

Installing the CrypKey License Service for MON2000 PLUS JULY 2010

Page 31 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-5 MON2000
Page 31 · click to enlarge
Page 31

INSTALLATION AND SETUP 2-5 MON2000

1. Label 3 blank formatted 3.5-inch floppy disks “MON2000 Install Disk 1”, “MON2000 Install Disk 2”, and “MON2000 Install Disk 3”.
2. Place the MON2000 CD in your CD-ROM drive.
3. Launch Windows ® Explorer ® .
4. Ensure that the Show all files option in Windows ® Explorer ® is selected.

(a) Use the View > Folder Options menu to

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-5 MON2000 diagram (page 31)
Fig. 4

access the Folder Options dialog. (b) Click the View tab.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-5 MON2000 diagram (page 31)
Fig. 5

(c) In the Advanced Settings pane, find the

gray folder icon labeled Hidden Files. (d) Click the Show all files radio button.

(e) Click the button.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-5 MON2000 diagram (page 31)
Fig. 6
5. Use the Windows ® Explorer ® directory tree to open the MON2000 Installation Files folder on your CD.
6. Place the “MON2000 Install Disk 1” in your floppy drive.

JULY 2010 Copying MON2000 to Disks

Page 32 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.5 TRANSFERRING A MON2000 PLUS LICENSE
Page 32 · click to enlarge
Page 32

2-6 INSTALLATION AND SETUP

7. Open the Disk1 folder on your CD.
8. Copy all files from the Disk1 folder to the “MON2000 Install Disk 1” floppy.
9. Remove the “MON2000 Install Disk 1” from your floppy drive.

10.Repeat Steps 5 through 8 to copy the Disk2 and Disk3 folders to your “MON2000 Install Disk 2” and “MON2000 Install Disk 3” floppy disks.

11.Remove the MON2000 CD from your CD-ROM drive.

2.5 TRANSFERRING A MON2000 PLUS LICENSE

To physically transfer the license file from one computer to the other you will need a removeable storage medium such as a 3.5-inch floppy disk or a USB flash drive. Also, there must be a licensed version on MON2000 PLUS installed on the source computer and an unlicensed version of MON2000 PLUS on the targeted computer.

1. On the target computer, if using a floppy disk, insert it into the disk drive; if using a flash drive, insert it in to a USB slot.

Transferring a MON2000 PLUS License JULY 2010

Page 33 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-7 MON2000
Page 33 · click to enlarge
Page 33

INSTALLATION AND SETUP 2-7 MON2000

2. Start MON2000 on the target computer. The License Configuration screen displays.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-7 MON2000 diagram (page 33)
Fig. 7
3. Select Transfer in from another computer... from the License menu. The

JULY 2010 Transferring a MON2000 PLUS License

Page 34 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-8 INSTALLATION AND SETUP
Page 34 · click to enlarge
Page 34

2-8 INSTALLATION AND SETUP

Transfer License In (Step 1 of 3) screen displays.

Emerson Process Management Chromatograph Software (MON2000) — 2-8 INSTALLATION AND SETUP diagram (page 34)
Fig. 8
4. Click Browse to select a disk path to the removeable storage medium that holds the license file. Click OK to accept your selection.

Transferring a MON2000 PLUS License JULY 2010

Page 35 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-9 MON2000
Page 35 · click to enlarge
Page 35

INSTALLATION AND SETUP 2-9 MON2000

5. Click Next . The Transfer License In (Step 2 of 3) screen displays.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-9 MON2000 diagram (page 35)
Fig. 9
6. Remove the removeable storage medium from the targeted computer and insert it into the source computer.
7. Launch MON2000 PLUS on the source computer. When the startup screen displays, press Enter . The License

JULY 2010 Transferring a MON2000 PLUS License

Page 36 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-10 INSTALLATION AND SETUP
Page 36 · click to enlarge
Page 36

2-10 INSTALLATION AND SETUP

Configuration screen displays with the message “Unlimited license”.

Emerson Process Management Chromatograph Software (MON2000) — 2-10 INSTALLATION AND SETUP diagram (page 36)
Fig. 10
8. Select Transfer out to another computer... from the License menu of the

Transferring a MON2000 PLUS License JULY 2010

Page 37 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-11 MON2000
Page 37 · click to enlarge
Page 37

INSTALLATION AND SETUP 2-11 MON2000

License Configuration screen. The Transfer License Out (Step 1 of 2) screen displays.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-11 MON2000 diagram (page 37)
Fig. 11
9. Click Browse to select a disk path to the removeable storage medium that holds the license file. Click OK to accept your selection.

JULY 2010 Transferring a MON2000 PLUS License

Page 38 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-12 INSTALLATION AND SETUP
Page 38 · click to enlarge
Page 38

2-12 INSTALLATION AND SETUP

10.Click Next. The Transfer License Out (Step 2 of 2) screen displays.

Emerson Process Management Chromatograph Software (MON2000) — 2-12 INSTALLATION AND SETUP diagram (page 38)
Fig. 12

Transferring a MON2000 PLUS License JULY 2010

Page 39 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-13 MON2000
Page 39 · click to enlarge
Page 39

INSTALLATION AND SETUP 2-13 MON2000

11.Click Finish. MON2000 PLUS on the source computer is no longer licensed.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-13 MON2000 diagram (page 39)
Fig. 13

12.Remove the removeable storage medium from the source computer and reinsert it into targeted computer.

JULY 2010 Transferring a MON2000 PLUS License

Page 40 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-14 INSTALLATION AND SETUP
Page 40 · click to enlarge
Page 40

2-14 INSTALLATION AND SETUP

13.On the targeted computer, click Next on the Transfer License In (Step 2 of 3) screen. The Transfer License In (Step 3 of 3) screen displays.

Emerson Process Management Chromatograph Software (MON2000) — 2-14 INSTALLATION AND SETUP diagram (page 40)
Fig. 14

Transferring a MON2000 PLUS License JULY 2010

Page 41 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.6 UNINSTALLING MON2000
Page 41 · click to enlarge
Page 41

INSTALLATION AND SETUP 2-15 MON2000

14.Click Finish. The License Configuration screen displays.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-15 MON2000 diagram (page 41)
Fig. 15

MON2000 PLUS is now licensed for unlimited use on the targeted computer.

2.6 UNINSTALLING MON2000

To uninstall MON2000 using Windows® 95, Windows® 98, Windows® 2000 or Windows® NT,

1. Click the button (see the taskbar).

JULY 2010 Uninstalling MON2000

Page 42 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-16 INSTALLATION AND SETUP
Page 42 · click to enlarge
Page 42

2-16 INSTALLATION AND SETUP

2. Click Settings and then Control Panel . The Control Panel dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 2-16 INSTALLATION AND SETUP diagram (page 42)
Fig. 16
Emerson Process Management Chromatograph Software (MON2000) — 2-16 INSTALLATION AND SETUP diagram (page 42)
Fig. 17
3. Double-click the Add/Remove Programs icon.

If using Windows XP, click the Start button, then Add/Remove Programs.

4. The Add/Remove Program Properties dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 2-16 INSTALLATION AND SETUP diagram (page 42)
Fig. 18
5. Select MON2000 and click the button.

Uninstalling MON2000 JULY 2010

Page 43 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.7 STARTING MON2000
Page 43 · click to enlarge
Page 43

INSTALLATION AND SETUP 2-17 MON2000

6. Windows ® displays the following confirmation dialog.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-17 MON2000 diagram (page 43)
Fig. 19
7. Click the button to continue. Windows ® deletes the program files only. Any files shared by other programs or created by the user are not deleted.

Click the button to abort and return to the Add/Remove Program Properties dialog.

2.7 STARTING MON2000

To start MON2000 after a successful installation,

Use the Windows® Start menu (Start > Programs >MON2000 menu) by clicking the

button.

Double-click the MON2000 icon automatically created on your desktop during installation.

You cannot access the MON2000 functions until you are successfully logged on.

To start MON2000 directly from the executable file, use the directory you specified when installing MON2000. Note that c:\Program

JULY 2010 Starting MON2000

Page 44 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.8 PERFORMING YOUR FIRST LOGON
Page 44 · click to enlarge
Page 44

2-18 INSTALLATION AND SETUP

Files\MON2000\MON2000 is the default setting.

Emerson Process Management Chromatograph Software (MON2000) — 2-18 INSTALLATION AND SETUP diagram (page 44)
Fig. 20

2.8 PERFORMING YOUR FIRST LOGON

2.8.1 The Initial Logon

Each new GC unit is shipped with one super user named “Emerson”. After logon, ensure that this user name appears in the Users list as a super user (see Section 2.11.1).

To log on for the first time,

1. Start the MON2000 software program by clicking the desktop icon or by using the Windows ® Start menu ( see Section 2.7 ).
Emerson Process Management Chromatograph Software (MON2000) — 2.8 PERFORMING YOUR FIRST LOGON diagram (page 44)
Fig. 21
2. Click in the User Name data field.

A PIN is not required for the initial logon.

3. Type emerson . Note that the user name is not case-sensitive.

Performing Your First Logon JULY 2010

Page 45 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.8.2 Registering MON2000
Page 45 · click to enlarge
Page 45

INSTALLATION AND SETUP 2-19 MON2000

4. When you have successfully logged on, the MON2000 main window appears.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-19 MON2000 diagram (page 45)
Fig. 22

2.8.2 Registering MON2000

After your have successfully performed your initial logon (see Section 2.8.1), MON2000 automatically prompts you to register your copy of MON2000 software. An active Internet connection is required for registration.

Emerson Process Management Chromatograph Software (MON2000) — 2.8.2 Registering MON2000 diagram (page 45)
Fig. 23 · 2.8.2 Registering MON2000

Registering your MON2000 software allows you to receive information about free updates and related products. A direct link to the Emerson Internet website is provided via the File > Update MON2000 menu.

You may choose to postpone registration. You can register at any time via the File > Register MON2000 menu.

JULY 2010 Registering MON2000

Page 46 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-20 INSTALLATION AND SETUP
Page 46 · click to enlarge
Page 46

2-20 INSTALLATION AND SETUP

1. Use the File>Register MON2000 menu. Follow the prompts in the Register MON2000 dialog to input your name, MON2000 serial number (located inside the CD jewel case), and other relevant information.
Emerson Process Management Chromatograph Software (MON2000) — 2-20 INSTALLATION AND SETUP diagram (page 46)
Fig. 24
2. Click the button to continue.
3. Choose the desired registration method by clicking the corresponding radio button.
Emerson Process Management Chromatograph Software (MON2000) — 2-20 INSTALLATION AND SETUP diagram (page 46)
Fig. 25

You must have an active internet connection to register via the website.

4. Click the button.

Registering MON2000 JULY 2010

Page 47 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-21 MON2000
Page 47 · click to enlarge
Page 47

INSTALLATION AND SETUP 2-21 MON2000

2.8.3 Update MON2000

Provided there is an active internet connection to the computer on which MON2000 is installed, clicking the File>Update MON2000 submenu sends the user to the Emerson® Chromatograph Controllers website (www.emersonprocess.com/daniel/products/GC/ Controllers/Productdetail.htm).

Updating the software requires a previously installed, registered copy of MON2000.

Use the File>Update MON2000 menu to download the latest versions of MON2000 and the BOS software programs.

To download the latest versions of the software,

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-21 MON2000 diagram (page 47)
Fig. 26
1. Use the computer on which MON2000 is installed and connect to the Internet. Minimize the Internet window.
2. From MON2000, select the File>Update MON2000 menu. This opens the Emerson Chromatograph Controllers website ( www.emersonprocess.com/daniel/products/ GC/Controllers/Productdetail.htm ).
3. From the Emerson Chromatograph Controller website, you can download the latest MON2000 and BOS updates, view and/or download catalogs, brochures, equipment specification sheets, product manuals, drawings, application notes, and white papers. Other links to products and services are also available.

JULY 2010 Update MON2000

Page 48 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.10.1 Connect to the GC Unit
Page 48 · click to enlarge
Page 48

2-22 INSTALLATION AND SETUP

2.9 CHECKLIST FOR SETTING UP MON2000

To ensure optimum performance,

  • Configure your system security ( see Section 2.11.1 ).

Verify that you have assigned a password to the super user, and/or created a secure system Admin account.

  • Configure how MON2000 will save your analytic and diagnostic information ( see Section 2.16.2 ).
  • Establish communications between MON2000 and the GC unit ( see Section 2.10 and Section 5.17 ).

2.10 COMMUNICATIONS

MON2000 can communicate to the GC unit locally via a serial port cable, Ethernet connection, or remotely via a modem. If performing a remote connection, ensure that you have configured the PC modem first.

To configure the communication and serial port settings for the GC unit, see Section 5.17.

2.10.1 Connect to the GC Unit

Emerson Process Management Chromatograph Software (MON2000) — 2.10.1 Connect to the GC Unit diagram (page 48)
Fig. 27 · 2.10.1 Connect to the GC Unit
1. Use the File > Connect menu or click the toolbar icon to access the Select GC for

Connect menu.

Emerson Process Management Chromatograph Software (MON2000) — 2.10.1 Connect to the GC Unit diagram (page 48)
Fig. 28 · 2.10.1 Connect to the GC Unit
2. Double-click the desired GC unit.

Checklist for Setting Up MON2000 JULY 2010

Page 49 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.10.2 Disconnect from GC Unit
Page 49 · click to enlarge
Page 49

INSTALLATION AND SETUP 2-23 MON2000

3. MON2000 appears the connection status dialog while dialing the selected unit.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-23 MON2000 diagram (page 49)
Fig. 29

If another user is already connected to the GC, the unit will “lock out” the second user.

When a PC connection is active, the GC Controller front panel will indicate a “System Lockout”. This status times out after 10 minutes.

The messages “Reading dictionary”, “Reading GC Application data”’ and “Logon” appears in the status bar and an information screen appears.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-23 MON2000 diagram (page 49)
Fig. 30

Once connected, the name of the GC unit displays in the lower left hand corner of the MON2000 main window.

2.10.2 Disconnect from GC Unit

Use this process to terminate an active PC connection to a GC unit.

Emerson Process Management Chromatograph Software (MON2000) — 2.10.2 Disconnect from GC Unit diagram (page 49)
Fig. 31 · 2.10.2 Disconnect from GC Unit

The menu-driven procedure for disconnecting is completely optional. MON2000 intelligently and automatically disconnects when you exit MON2000 or connect to a second GC Controller.

1. Use the File > Disconnect menu or click the toolbar icon to access this function.
2. MON2000 automatically terminates all open connections.

JULY 2010 Disconnect from GC Unit

Page 50 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.11.1 Configure Users
Page 50 · click to enlarge
Page 50

2-24 INSTALLATION AND SETUP

A Terminating communications message appears in the status bar.

If MON2000 does not detect an active connection to terminate, an “Invalid selection or not downloaded” message appears.

2.11 CUSTOMIZING MON2000

Use the following functions to customize users, the directory of GC applications, and PC communications parameters.

2.11.1 Configure Users

Use this function to create new user accounts or edit the existing account data. An extra level of security is assigned to the File>Users submenu. The File>Users submenu can only be accessed by first entering a Password into a dialog box titled 'Enter Administrator Password'. After gaining access to the Users dialog, the operator then has the ability to create new user accounts or edit existing account data. This feature is not available to the 'Regular' and 'Read Only' user.

1. Use the File > Users menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — 2.11.1 Configure Users diagram (page 50)
Fig. 32 · 2.11.1 Configure Users
2. The Enter Administrator Password dialog box appears.
Emerson Process Management Chromatograph Software (MON2000) — 2.11.1 Configure Users diagram (page 50)
Fig. 33 · 2.11.1 Configure Users

Customizing MON2000 JULY 2010

Page 51 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-25 MON2000
Page 51 · click to enlarge
Page 51

INSTALLATION AND SETUP 2-25 MON2000

users access to the MON2000 Users submenu, which allows Super users to add, edit, or change security levels.

Click the button. Then the Users dialog appears.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-25 MON2000 diagram (page 51)
Fig. 34
3. Use the File>Change Administrator menu to change the Administrator password.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-25 MON2000 diagram (page 51)
Fig. 35

The Change Administrator Password dialog box appears.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-25 MON2000 diagram (page 51)
Fig. 36

Insert the old password, enter the new password and confirm the change.

Then, click the button to apply your changes.

JULY 2010 Configure Users

Page 52 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-26 INSTALLATION AND SETUP
Page 52 · click to enlarge
Page 52

2-26 INSTALLATION AND SETUP

4. To configure a new user or edit existing user parameters,

(a) Click the appropriate Name cell and type

Super users can write changes to the GC unit, configure MON2000, and access the Users function.

the desired user name.

Regular and Read only users can only view data.

(b) Click the appropriate Type cell and use

the provided combo box to select the desired security level. Note that there should be at least one Super user. (c) Click the appropriate PIN cell and type

the desired PIN for this user. Note that the PIN is limited to 12 numeric characters.

If no PIN is entered, the user can log on to the MON2000 software program or the GC unit (via the front panel) with the assigned user name – that is, no password will be required.

5. To write user data to the online GC unit,

(a) Click the or press the F2 key. (b) MON2000 writes the data to the GC

Controller.

6. Click the button to apply your changes and return to the main screen.

Click the button to exit and return to the main screen without applying your changes.

Configure Users JULY 2010

Page 53 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-27 MON2000
Page 53 · click to enlarge
Page 53

INSTALLATION AND SETUP 2-27 MON2000

2.11.2 Set Up GC Directory

This function allows you to set up a directory listing of all GC units MON2000 can control for this application. From the GC Directory, you can configure the PC serial port communication parameters for a specific GC unit.

Do not delete any currently used applications from the GC Directory. If an entry is inadvertently deleted, you may need to reinstall the application software for that GC.

Data entered in the GC Directory is stored in the “Gcdir.dat” file (...\GC\BIN\). The Station Name is the user assigned name to a GC location. As GC applications are installed, MON2000 adds the application name to the GC Directory.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-27 MON2000 diagram (page 53)
Fig. 37

The GC unit name shown in the GC Status Bar of the MON2000 main window is entered via the System dialog (see Section 5.1).

To edit the GC directory,

1. Use the File > GC Directory menu to access this function.
2. The GC Directory dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-27 MON2000 diagram (page 53)
Fig. 38

JULY 2010 Set Up GC Directory

Page 54 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-28 INSTALLATION AND SETUP
Page 54 · click to enlarge
Page 54

2-28 INSTALLATION AND SETUP

3. Press the F1 key to open the MON2000 Application help file for detailed information about these settings.
Emerson Process Management Chromatograph Software (MON2000) — 2-28 INSTALLATION AND SETUP diagram (page 54)
Fig. 39
4. MON2000 is configured with two default directories:
  • Default 1: parameters are set for 9600 baud rate
5. Default 2: parameters are set for 19200 baud rate.
6. Click the appropriate Com ID cell and type the Modbus communication identification number (set by DIP switch positions on the

Set Up GC Directory JULY 2010

Page 55 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-29 MON2000
Page 55 · click to enlarge
Page 55

INSTALLATION AND SETUP 2-29 MON2000

GC Controller system interface board) for the GC unit.

To successfully connect to the GC unit, the COM IDs specified via the File>GC Directory and the Application>Serial Ports dialogs must match. See Section 5.17.1 for more information.

7. Click the Comments cell to enter any helpful information regarding a particular GC unit, such as location or purpose.
8. Use the dynamic pull-down menus to select the desired PC Port , Baud Rate , Data Bits , Stop Bits , Parity , and Handshaking settings.

The following parameters support an auto-detect setting: Baud Rate, Data Bits, Stop Bits, Parity, Protocol.

Set any of these parameters to “?” (autodetect) and MON2000 will conform to the settings used by the GC unit.

9. Use the provided cells to input RTS On Delay , RTS Off Delay , and Extra Delay values.

The Extra Delay field accepts numerical values (0 to 9000 milliseconds) for additional time to be added to the current communication timeout delay.

10.Use the Protocol pull-down menu to select the desired Modbus communications protocol.

11.If you plan to connect remotely to the GC unit via a modem or Ethernet connection, use the Connection Type (Direct/Remote)

JULY 2010 Set Up GC Directory

Page 56 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.11.3 Dial-up Connection
Page 56 · click to enlarge
Page 56

2-30 INSTALLATION AND SETUP

pull-down menu for a “Direct Connect” option or a previously configured modem.

2.11.3 Dial-up Connection

Use this command to run the Dial-up dialog to configure the GC Controller modem. The default properties from Windows® Control Panel are used so you must ensure that the modem connects at the GC Controller’s configured baud rate. It may be necessary to enter a modem initialization string.

The Direct Connect option allows you to connect from the PC to the GC Controller via a serial port, per its default settings.

If you select a modem, the modem property dialog appears, allowing you to change its attributes.

To configure the modem and make a connection,

1. Use the File>GC Directory menu, and the GC Directory appears.
Emerson Process Management Chromatograph Software (MON2000) — 2.11.3 Dial-up Connection diagram (page 56)
Fig. 40 · 2.11.3 Dial-up Connection
2. Click the Station Name data field and click the Modem Properties button or press the F2 button.

Dial-up Connection JULY 2010

Page 57 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-31 MON2000
Page 57 · click to enlarge
Page 57

INSTALLATION AND SETUP 2-31 MON2000

3. The Modem Properties dialog appears and the General Data Connection Modem Preferences are activated.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-31 MON2000 diagram (page 57)
Fig. 41

The General Data default settings are:

Table 2-1 General Data Connection Preferences

JULY 2010 Dial-up Connection

Show 4 more labels / short text on this page
Port Speed 19200 · Data Protocol Disabled · Compression Disabled · Flow Control None
Page 58 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-32 INSTALLATION AND SETUP
Page 58 · click to enlarge
Page 58

2-32 INSTALLATION AND SETUP

4. Click the Advanced tab to configure the Hardware settings.
Emerson Process Management Chromatograph Software (MON2000) — 2-32 INSTALLATION AND SETUP diagram (page 58)
Fig. 42

The Advanced Data default settings are:

Table 2-2 Advanced Data Connection Preferences

5. Click the button to apply your

changes, or click the button to discard the changes and return to the GC Directory dialog.

Dial-up Connection JULY 2010

Data Bits 8 · Parity None · Stop Bits 1

Page 59 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-33 MON2000
Page 59 · click to enlarge
Page 59

INSTALLATION AND SETUP 2-33 MON2000

6. Next, from the GC Directory dialog, scroll over to configure the following parameters:

The Baud Rate, Data Bits, and Stop Bits parameters were configured above.

Table 2-3 GC Directory Dialog Default Settings

Heading Pull-down Menu Selection

PC Port COM1, COM2,COM3

RTS On Delay 0

RTS On Delay 0

Extra Delay 0 (See note)

Telephone Complete number (Area Code) (XXX-XXXX)

Server Type - IP Address

XXX.XX.XX.XXX (10 digits)

  • Retries
  • Modem
  • IP Address 5 3

JULY 2010 Dial-up Connection

Show 4 more labels / short text on this page
Handshaking None · Connection Type (Direct/Remote) · Direct Connection Modem · IC Multiplier 10
Page 60 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-34 INSTALLATION AND SETUP
Page 60 · click to enlarge
Page 60

2-34 INSTALLATION AND SETUP

The Extra Delay is enabled for Ethernet connections.

The default value is “0” and the default 30 seconds timeout is used. If the value is changed to greater than “0”, the value is used as the timeout. The entered value is multiplied by 10 inside MON2000, therefore the maximum timeout is 100,000 ms or 100 seconds (ex: If timeout value needs to be 15 seconds, the user should enter 1,500).

Dial-up Connection JULY 2010

Page 61 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-35 MON2000
Page 61 · click to enlarge
Page 61

INSTALLATION AND SETUP 2-35 MON2000

7. Click the button to accept the changes. While connecting, MON2000 dials the number, attempts to connect via the selected modem and displays progress messages. A message box appears when a connection has been made or if the attempt to connect fails.

Once a connection has been established and while MON2000 remains connected via the modem, the Connect button is disabled and the Disconnect button is enabled.

If the connection fails at any time, MON2000 displays a message in the GC Status bar that it has been disconnected. You can end the connection from the File>Disconnect menu, by clicking the disconnect icon on the MON2000 toolbar, or by exiting MON2000.

If you are using the Modbus RTU protocol, the values for baud rate, data bits, stop bits, and parity must match the settings you configured in the Modem Properties dialog.

These values are required to correctly calculate character timing.

When you start MON2000 after the parameters are established, you can connect to the GC via the File>Connect menu or by clicking the Connection icon on the MON2000 toolbar.

JULY 2010 Dial-up Connection

Page 62 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-36 INSTALLATION AND SETUP
Page 62 · click to enlarge
Page 62

2-36 INSTALLATION AND SETUP

2.11.4 Modem Initialization Strings/Setup

This section provides “examples” of modem initialization strings and setups that have been tested and proven operable. This string of text characters, known as 'AT commands', has special meaning to the PC modem and is used by the modem for every telephone connection made to a selected GC unit.

To edit or enter a modem initialization string,

use the button and use the Control Panel > Modems > General > Properties > Connection > Advanced to access the Advanced Modem Connection Settings dialog. Enter the modem initialization string in the Extra Settings field.

Emerson Process Management Chromatograph Software (MON2000) — 2-36 INSTALLATION AND SETUP diagram (page 62)
Fig. 43

Modem Initialization Strings/Setup JULY 2010

Page 63 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Black Box - Modem 325
Page 63 · click to enlarge
Page 63

INSTALLATION AND SETUP 2-37 MON2000

By default, MON2000 provides a modem initialization string that serves the purpose for most Hayes-compatible modems. Therefore, in most cases, you need not enter a different string.

However, for some modems, you may want to consult the following pages which document initialization strings and modem setups that have been field-tested.

The default initialization string that MON2000 provides is sent to the PC modem first. Then the custom initialization string you provide is sent. In this manner, MON2000 sets the modem parameters that it requires, while you can customize the modem's operation to respond to special conditions.

Once the optimum modem initialization string has been determined, it should need no further revision unless a different brand or model of modem is installed.

Black Box - Modem 325

During development of the GC Controller, a Black Box Modem 325 was used to test modem communications with the GC unit. For that test and that modem, the modem initialization was set as follows:

\N0\C0\Q0 (where 0 = zero)

JULY 2010 Modem Initialization Strings/Setup

Page 64 of the Emerson Process Management Chromatograph Software (MON2000) user manual — U.S. Robotics Sportster
Page 64 · click to enlarge
Page 64

2-38 INSTALLATION AND SETUP

Interpretation of the AT commands for that particular modem was as follows:

  • \N0 = normal mode; no error control; data is buffered (versus ‘direct mode,’ ‘reliable mode, ’or ‘auto reliable mode’)
  • \C0 = disable auto reliable buffer (versus ‘buffer data for 4 seconds or 200 characters’)
  • \Q0 = disable flow control (versus ‘enable XON/XOFF flow control,’ ‘enable unilateral CTS flow control,’ or ‘enable bilateral CTS/ RTS flow control’)

U.S. Robotics Sportster

The following setup for U.S. Robotics Sportster 28.8 Kbps external FAX-modems was verified at the GC assembly and checkout lab:

  • At the GC Controller, the modem DIP switches were set to 5 and 8 UP, the rest DOWN.
  • At the PC, the modem DIP switches were set to 2, 4, 6, and 7 UP, the rest DOWN.

Modem Initialization Strings/Setup JULY 2010

Page 65 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-39 MON2000
Page 65 · click to enlarge
Page 65

INSTALLATION AND SETUP 2-39 MON2000

With these switch settings, initialization string &F was required. See Table 2-4 for further explanation of these switch settings and the parameters they control.

Table 2-4 DIP Switch Settings for U.S. Robotics Sportster Modem

Switch Function At GC At PC

1 UP (off) – Normal DTR; computer must provide DTR signal for modem to accept commands.

DOWN (on) – Modem ignores DTR.

2 UP (off) – Verbal (word) results.

DOWN (on) – Numeric results.

3 UP (off) – Suppress result codes.

DOWN (on) – Enable result codes.

4 UP (off) – Display keyboard commands.

DOWN (on) – Suppress echo.

5 UP (off) – Modem answers on first ring or higher.

DOWN (on) – Disable auto answer.

6 UP (off) – Modem sends carrier detect (CD) signal on connect, and drops CD on disconnect.

DOWN (on) – CD always on.

7 UP (off) – For power-on and ATZ reset, the Y or Y1 configuration is used from user-defined nonvolatile memory (NVRAM).

DOWN (on) – For power-on and ATZ reset, the generic template, &F0, is used from read only memory (ROM).

8 UP (off) – Disable AT command set recognition.

DOWN (on) – Enable AT command set recognition.

JULY 2010 Modem Initialization Strings/Setup

Page 66 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-40 INSTALLATION AND SETUP
Page 66 · click to enlarge
Page 66

2-40 INSTALLATION AND SETUP

GVC/MaxTech 28.8 Kbps Modem

The following setup for GVC 28.8 Kbps external Voice/Data/FAX modems was verified at Daniel Canada. At the GC Controller:

To make these settings for the modem at the GC Controller, you will need to use a terminal program (such as ProComm o other commercial modem software) to input the “AT” commands to the modem before connecting it to the GC Controller.

AT&F5 Factory default with v.42bis flow control disabled (must be first).

AT&D0 Modem ignores DTR (not necessary if using a cable that incorporates the DTR signal from the Controller).

ATM0 Turn OFF speaker.

ATS0=n n is number of rings for auto answer (e.g., ATS0=1, ATS0=2, etc.).

AT&C0 Force carrier detect high

AT%C0 Turn OFF compression.

If compression is turned ON with AT%C1, or OFF with AT%C0, then the modem at other end should match.

AT&W0 Write to stored profile “0” (zero).

AT&Y0 Use stored profile “0” (zero) when turned on.

Modem Initialization Strings/Setup JULY 2010

ATB8 9600 baud · Or · ATB11 19200 baud

Page 67 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.12 2350 TO 2350A RETROFIT INSTRUCTIONS
Page 67 · click to enlarge
Page 67

INSTALLATION AND SETUP 2-41 MON2000

At the PC, use &F5%C1M0 which indicates a factory default with v.42bis flow control disabled / compression ON / speaker OFF.

2.12 2350 TO 2350A RETROFIT INSTRUCTIONS

The 2350A GC Controller CPU assembly has been designed to include all digital I/O and COM3 and COM4 serial ports. This eliminates the requirement (and additional cost) for an I/O assembly or DSPI/O assembly.

In addition, BOS is now resident in the DiskOnChip, instead of an EPROM set. The DiskOnChip provides additional expanded memory for archiving data, instead of having to purchase a memory expansion board assembly. The BOS file in the DiskOnChip may be upgraded in the field through MON2000.

Optional boards may be plugged directly into the PC/104 Bus (connectors J19 and J20) on the CPU board assembly. The COM4A board provides four additional serial ports (COM5 through COM8). A modem board and/or Ethernet board may be also be plugged directly into the PC/104 Bus for additional communications requirements.

JULY 2010 2350 to 2350A Retrofit Instructions

Page 68 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-42 INSTALLATION AND SETUP
Page 68 · click to enlarge
Page 68

2-42 INSTALLATION AND SETUP

The PC/104 Bus is designed to allow any combination of option boards to be installed in any order, with the exception of the CSA approved Radicom modem board which must be the top board in the assembly. (see Figure 2-1).

Emerson Process Management Chromatograph Software (MON2000) — 2-42 INSTALLATION AND SETUP diagram (page 68)
Fig. 44

Figure 2-1 2350A Controller Card Cage Assembly

If you are using the CSA approved Radicom modem; ensure that it is the top card in the card cage assembly. The connection configuration of the Radicom modem requires installation at the top of the assembly.

2350 to 2350A Retrofit Instructions JULY 2010

Page 69 of the Emerson Process Management Chromatograph Software (MON2000) user manual — SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Page 69 · click to enlarge
Page 69

INSTALLATION AND SETUP 2-43 MON2000

2.13 CONVERSION PROCESS

Use the following process to retrofit a 2350 GC Controller to a 2350A GC Controller.

1. At the GC Controller site, remove the Controller enclosure's front panel.

SERIOUS PERSONAL INJURY OR DEATH POSSIBLE

Before removing the unit cover from the GC Controller, make certain the power supply switch is OFF and the AC power cord is disconnected. Observe all safety precautions when you are working in a hazardous environment.

Failure to observe all safety precautions could result in serious injury or death.

(a) For the explosion-proof Controller, the

front panel is secured by 16 screws. Remove those screws first. (b) Then carefully lower the front panel on

its bottom hinges. The front panel is heavy, so make sure it does not drop and cause damage. (c) For the rack mount Controller, the rear

of the enclosure is open; it allows access for most field wiring procedures without removing the enclosure.

2. Locate the GC Controller's Terminal Board for Field Wiring (TB). The TB is attached to the GC Controller's card cage assembly, facing the enclosure's front panel. (In the rack mount Controller, the TB faces outward toward the rear of the enclosure.)
3. Loosen the six screws that secure the TB. Then unplug the TB from its connections at the back, top of the board.

JULY 2010 Conversion Process

Page 70 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-44 INSTALLATION AND SETUP
Page 70 · click to enlarge
Page 70

2-44 INSTALLATION AND SETUP

4. Lower the TB down and out of the way, held in place by its ground straps at the bottom of the board. This exposes the Card Cage Assembly.
5. Loosen the four screws that secure the Card Cage Assembly to the chassis. Then remove the Card Cage Assembly away from its chassis mount so that it is easy to work on.
6. Locate the System Interface and Driver board. It is mounted to the top of the Card Cage Assembly.
7. Remove all cables connected to the System Interface Board (P/N 3-2350-005, P/N 3-2350-022, or P/N 3-2350-023).
8. Remove the I/O48 Board assembly (drawing P/N BE-12973) and cables from the top slot of the Cage Card Assembly. This board is not used on the 2350A GC Controller.
9. Remove the CPU Board assembly (P/N CE19281) and cables from the second slot of the card cage assembly. This board is not used on the 2350A GC Controller.

Conversion Process JULY 2010

Page 71 of the Emerson Process Management Chromatograph Software (MON2000) user manual — SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Page 71 · click to enlarge
Page 71

INSTALLATION AND SETUP 2-45 MON2000

10.Remove the DSPI/O board assembly (drawing P/N CE-12976) and cables from the third slot of the card cage assembly. This board is not used on the 2350A GC Controller.

11.Remove any other optional boards, such as memory expansion boards and associated cables. These parts are not used on the 2350A GC Controller.

12.Leave the Analog Board assembly (drawing P/N BE-18044) with attached cable in the card cage. This board is used on the 2350A basic configuration.

13.Change the fuse in the in-line fuse holder from 1 amp to the 2.5 amp, Slo-Blo fuse provided. The fuse is located in the cable (P/ N 2-3-2350-069) between the power supply and the System Interface Board.

14.Ensure that DIP Switches 6, 7, and 8 of S1 are all in the "OFF" position. If the 2350A is powered up with S8 in the OFF position, it will delete the current application (commonly called a "Cold Start").

15.To inspect or change the GC Controller's COM ID setup at the GC Controller site, locate the DIP switch as described in the following steps.

SERIOUS PERSONAL INJURY OR DEATH POSSIBLE

Before removing the unit cover from the GC Controller, make certain the power supply switch is OFF and the AC power cord is disconnected. Observe all safety precautions when you are working in a hazardous environment.

Failure to observe all safety precautions could result in serious injury or death.

16.For the explosion-proof Controller, the front panel is secured by 16 screws. Remove those screws first.

JULY 2010 Conversion Process

Page 72 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-46 INSTALLATION AND SETUP
Page 72 · click to enlarge
Page 72

2-46 INSTALLATION AND SETUP

(a) Then carefully lower the front panel on

its bottom hinges. The front panel is heavy, so make sure it does not drop and cause damage. The DIP switch is located on the lower left side of the front panel (see Figure G-2).

Emerson Process Management Chromatograph Software (MON2000) — 2-46 INSTALLATION AND SETUP diagram (page 72)
Fig. 45

Figure 2-2 Explosion-Proof Controller Dip Switch

Conversion Process JULY 2010

Page 73 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-47 MON2000
Page 73 · click to enlarge
Page 73

INSTALLATION AND SETUP 2-47 MON2000

17.For rack mount and panel mount Controllers, use a flat head screw driver to remove the access panel on the right side of the card cage assembly (see Figure 2-3).

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-47 MON2000 diagram (page 73)
Fig. 46

Figure 2-3 Right Side View Rack and Panel Mount Units

18.Inspect or change the DIP switch settings as necessary.

(a) See Table 2-5 as a guide. (b) Make sure you record in the GC

  • Controller's maintenance records any changes you make to the switch settings.
  • Switches "1" through "5" form a 5-bit binary number for setting the Modbus slave address (also known as COM ID or Device ID.)
  • Switch number "1" is the least significant bit, and switch number "5" is the most significant bit.

Switch to ON = 1

Switch to OFF = 0

JULY 2010 Conversion Process

Page 74 of the Emerson Process Management Chromatograph Software (MON2000) user manual — COM ID 1 2 3 4 5
Page 74 · click to enlarge
Page 74

2-48 INSTALLATION AND SETUP

  • Switch "6" is a spare for future use. Switches "7" and "8" are set as needed for the presence of an optional LOI (Local Operator Interface) connected via COM8 When the COM4A Board is installed. If the COM4A Board is not installed, the LOI is connected via COM4.

Table 2-5 Dip Switch Settings Switch Positions

COM ID 1 2 3 4 5

1 ON OFF OFF OFF OFF

2 OFF ON OFF OFF OFF

3 ON ON OFF OFF OFF

4 OFF OFF ON OFF OFF

5 ON OFF ON OFF OFF

6 OFF ON ON OFF OFF

7 ON ON ON OFF OFF

8 OFF OFF OFF ON OFF

RAM CLEAR

Dip Switch Setting Positions

Clears RAM when unit is powered down ON

Keeps RAM data when unit is powered down OFF

Conversion Process JULY 2010

Page 75 of the Emerson Process Management Chromatograph Software (MON2000) user manual — JULY 2010 BASIC 2350A CONFIGURATION
Page 75 · click to enlarge
Page 75

INSTALLATION AND SETUP 2-49 MON2000

2.14 BASIC 2350A CONFIGURATION

At this point, install, if any, all mounting hardware and optional PC/104 boards onto the 2350A CPU Board.

1. Install the 2350A CPU (P/N 3-2350-090) into slot 3 of the card cage assembly.
2. Install cables on the CPU Board in the following sequence:

(a) Digital I/O cable (P/N 3-2350-081) from

CPU J7 to the System Interface Board J2. (b) Digital I/O cable (P/N 3-2350-080) from

CPU J4 to the System Interface Board J3. (c) COM1, COM2, and the printer cable (P/

N 3-2350-083) from CPU J1 to the System Interface Board J4, J9, and J11. (d) COM3 and COM4 can be configured two

different ways. To configure a system WITHOUT a keyboard and display, connect cable (P/N 3-2350-084) from CPU J6 to the System Interface Board J8 and J10. This provides access to COM3 on J10 and COM4 on J11 of the Field Termination board. The serial port setup in the MON2000 Software for COM4 must be selected as a PC port. (e) For a system WITH a keyboard and

display, connect cable (P/N 3-2350-087) from CPU J6 to the System Interface J8 and J12. This provides access to COM3 on J10 and COM4 is dedicated for use as a serial interface to the keyboard and display. COM4 will not be available at J11 of the Field Termination board.

JULY 2010 BASIC 2350A CONFIGURATION

Page 76 of the Emerson Process Management Chromatograph Software (MON2000) user manual — BASIC 2350A CONFIGURATION JULY 2010
Page 76 · click to enlarge
Page 76

2-50 INSTALLATION AND SETUP

The serial port setup in the MON2000 Software for COM4 must be selected as Front Panel (see Figure 2-4). If all four serial ports are required for communications on a system with keyboard and display, an optional COM4A Board must be installed and a COM7 and COM8 cable (P/N 3-2350-086) connected (see Figure 2-5). The serial port setup in the MON2000 Software for COM8 must be configured as Front Panel. See Figure G-4b. (f) Reinstall the Analog cable to J6 on the

System Interface Board. (g) Place the "Unit Updated to 2350A" label,

included with the upgrade kit, on the instruction decal located on the inside, right wall of the explosion-proof units or on the card cage of the rack or panel mount units.

Emerson Process Management Chromatograph Software (MON2000) — 2-50 INSTALLATION AND SETUP diagram (page 76)
Fig. 47

BASIC 2350A CONFIGURATION JULY 2010

Page 77 of the Emerson Process Management Chromatograph Software (MON2000) user manual — JULY 2010 BASIC 2350A CONFIGURATION
Page 77 · click to enlarge
Page 77

INSTALLATION AND SETUP 2-51 MON2000

Figure 2-4 Serial Port 4 with COM4A Board

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-51 MON2000 diagram (page 77)
Fig. 48

Figure 2-5 Serial Port 8 with COM4A Board

JULY 2010 BASIC 2350A CONFIGURATION

Page 78 of the Emerson Process Management Chromatograph Software (MON2000) user manual — SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Page 78 · click to enlarge
Page 78

2-52 INSTALLATION AND SETUP

2.15 2350A OPTIONS

The following board configurations are optional for the Model 500 GC with 2350A Controller.

2.15.1 The COM4A Board

1. To add four additional communications ports at the GC Controller site, remove the Controller enclosure's front panel.

SERIOUS PERSONAL INJURY OR DEATH POSSIBLE

Before removing the unit cover from the GC Controller, make certain the power supply switch is OFF and the AC power cord is disconnected. Observe all safety precautions when you are working in a hazardous environment.

Failure to observe all safety precautions could result in serious injury or death.

(a) For the explosion-proof Controller, the

front panel is secured by 16 screws. Remove those screws first. (b) Then carefully lower the front panel on

its bottom hinges. The front panel is heavy, so make sure it does not drop and cause damage. (c) For the rack mount Controller, the rear

of the enclosure is open; it allows access for most field wiring procedures without removing the enclosure.

2. Locate the GC Controller's Terminal Board for Field Wiring (TB). The TB is attached to the GC Controller's Card Cage Assembly, facing the enclosure's front panel. (In the rack mount Controller, the TB faces outward toward the rear of the enclosure.)
3. Loosen the six screws that secure the TB. Then unplug the TB from its connections at the back, top of the board.

2350A Options JULY 2010

Page 79 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-53 MON2000
Page 79 · click to enlarge
Page 79

INSTALLATION AND SETUP 2-53 MON2000

4. Lower the TB down and out of the way, held in place by its ground straps at the bottom of the board. This exposes the Card Cage Assembly.
5. Loosen the four screws that secure the Card Cage Assembly to the chassis. Then remove the Card Cage Assembly away from its chassis mount so that it is easy to work on.
6. Locate the System Interface and Driver board. It is mounted to the top of the Card Cage Assembly.

(a) Disconnect the Analog cable from J6 on

the System Interface Board, then disconnect all CPU cables from the System Interface Board. (b) Remove the CPU assembly and install

the COM4A Board, with associated mounting hardware to J19 and J20 PC/ 104 Bus connector on the CPU assembly. (c) Connect COM5 (P22) and COM6 (P23)

on the Field Termination board via cable (P/N 3-2350-085) to J6 on COM4A. (d) Connect COM7 (P24) on the Field

Termination board and COM8 via cable (P/N 3-2350-086) to J3 on COM4A. (e) Install the CPU assembly into the third

slot of card cage. Connect the CPU cables as detailed in Section 6.3 Basic 2350A Configuration. (f) Install the DB9 connector from COM5 to

P22 on the Field Termination board with jackpost assemblies. (g) Install the DB9 connector from COM6 to

P23 on the Field Termination board with jackpost assemblies. (h) Install the DB9 connector from COM7 to

P24 on the Field Termination board with jackpost assemblies.

JULY 2010 The COM4A Board

Page 80 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-54 INSTALLATION AND SETUP
Page 80 · click to enlarge
Page 80

2-54 INSTALLATION AND SETUP

(i) Install the ten position connector from

COM8 to J12 on the System Interface Board. (j) Connect the Analog cable to J6 on the

System Interface Board.

With COM4A Boards installed, Serial Port 8 is usually assigned via MON2000 Applica- tion>Serial Ports menu. Select Front Panel from the Usage pull-down menu (see Figure 2-6 and Figure 2-7).

Figure 2-6 Serial Port 4 with COM4A Board

Figure 2-7 Serial Port 8 with COM4A Board

The COM4A Board JULY 2010

Page 81 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-55 MON2000
Page 81 · click to enlarge
Page 81

INSTALLATION AND SETUP 2-55 MON2000

2.15.2 2350A Modem Installation

To install the Internal Modem for the 2350A GC Controller, follow these steps:

Refer to P/N 3-9000-537 2350A GC Hardware manual for COM settings and pinouts.

1. To add a modem, at the GC Controller site, remove the Controller enclosure's front panel.

(a) For the explosion-proof Controller, the

front panel is secured by 16 screws. Remove those screws first. (b) Then carefully lower the front panel on

its bottom hinges. The front panel is heavy, so make sure it does not drop and cause damage. (c) For the rack mount Controller, the rear

of the enclosure is open; it allows access for most field wiring procedures without removing the enclosure.

2. Locate the GC Controller's Terminal Board for Field Wiring (TB). The TB is attached to the GC Controller's Card Cage Assembly, facing the enclosure's front panel. (In the rack mount Controller, the TB faces outward toward the rear of the enclosure.)
3. Loosen the six screws that secure the TB. Then unplug the TB from its connections at the back, top of the board.
4. Lower the TB down and out of the way, held in place by its ground straps at the bottom of the board. This exposes the Card Cage Assembly.

JULY 2010 2350A Modem Installation

Page 82 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-56 INSTALLATION AND SETUP
Page 82 · click to enlarge
Page 82

2-56 INSTALLATION AND SETUP

5. Loosen the four screws that secure the Card Cage Assembly to the chassis. Then remove the Card Cage Assembly away from its chassis mount so that it is easy to work on.
6. Locate the System Interface and Driver board. It is mounted to the top of the Card Cage Assembly.

(a) Disconnect the Analog cable from J6 on

the System Interface Board, then disconnect all CPU cables from the System Interface Board. (b) Remove the CPU assembly and install

the COM4A Board, with associated mounting hardware to J19 and J20 PC/ 104 Bus connector on the CPU assembly.

If an option board is already plugged into the PC/104 bus on the CPU, the modem and mounting hardware will be installed into the PC/104 connector on the option board.

Figure 2-8 Modem Piggy-backed on CPU Board

2350A Modem Installation JULY 2010

Page 83 of the Emerson Process Management Chromatograph Software (MON2000) user manual — J6 3-5, 4-6
Page 83 · click to enlarge
Page 83

INSTALLATION AND SETUP 2-57 MON2000

(c) Plug one end of the modem extension

cable (P/N 3-2350-075) into J1 of the modem assembly. The in-line jack on the remaining end of the modem extension cable attaches to the lower left inside wall of the card cage shield (after the CPU assembly is reinstalled and all cables reconnected to the System Interface board). No software setup is required for this board.

7. To use an existing Modem (1414):

(a) For operation with the 2350A CPU, set

the Standard 1414 Modem jumpers per the following table:

Table 2-6 1414 (Standard) Modem Jumper Settings

J6 3-5, 4-6

J7 7-8

(b) On the 2350A WinSystems CPU, set J21,

pin 13-14.

JULY 2010 2350A Modem Installation

Jumper Pin · J4/J5 Open · J9 Open

Page 84 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-58 INSTALLATION AND SETUP
Page 84 · click to enlarge
Page 84

2-58 INSTALLATION AND SETUP

2.15.3 2350A Ethernet Installation

To add an ethernet assembly:

1. To add an ethernet card at the GC Controller site, remove the Controller enclosure's front panel.

(a) For the explosion-proof Controller, the

front panel is secured by 16 screws. Remove those screws first. (b) Then carefully lower the front panel on

its bottom hinges. The front panel is heavy, so make sure it does not drop and cause damage. (c) For the rack mount Controller, the rear

of the enclosure is open; it allows access for most field wiring procedures without removing the enclosure. (d) Locate the GC Controller's Terminal

Board for Field Wiring (TB). The TB is attached to the GC Controller's Card Cage Assembly, facing the enclosure's front panel. (In the rack mount Controller, the TB faces outward toward the rear of the enclosure.)

2. Loosen the six screws that secure the TB. Then unplug the TB from its connections at the back, top of the board.
3. Lower the TB down and out of the way, held in place by its ground straps at the bottom of the board. This exposes the Card Cage Assembly.
4. Loosen the four screws that secure the Card Cage Assembly to the chassis. Then remove the Card Cage Assembly away from its chassis mount so that it is easy to work on.

2350A Ethernet Installation JULY 2010

Page 85 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-59 MON2000
Page 85 · click to enlarge
Page 85

INSTALLATION AND SETUP 2-59 MON2000

5. Locate the System Interface and Driver board. It is mounted to the top of the Card Cage Assembly.

(a) Disconnect the Analog cable from J6 on

the System Interface Board, then disconnect all CPU cables from the System Interface Board. (b) Remove the CPU assembly and install

the COM4A Board, with associated mounting hardware to J19 and J20 PC/ 104 Bus connector on the CPU assembly.

If an option board is already plugged into the PC/104 bus on the CPU, the ethernet card and mounting hardware will be installed into the PC/104 connector on the option board.

If you are using the CSA approved Radicom modem; ensure that it is the top card in the card cage assembly. The connection configuration of the Radicom modem requires installation at the top of the assembly.

(c) Plug one end of the Ethernet extension

cable (P/N 3-2350-088) into J5 of the Ethernet assembly.

6. The in-line jack on the remaining end of the Ethernet extension cable attaches to the lower left inside wall of the card cage shield, after the CPU assembly is reinstalled and all cables reconnected to the System Interface board ( see Table 2-7 and Table 2-8 below).

JULY 2010 2350A Ethernet Installation

Page 86 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-60 INSTALLATION AND SETUP
Page 86 · click to enlarge
Page 86

2-60 INSTALLATION AND SETUP

No software setup is required for this board.

Table 2-7 System Interface Board Cable Connections

J5 and J6 Analog I/O

J2 and J3 Digital I/O

J12 Front Panel and Keyboard (if installed)

Table 2-8 CPU Board Cable Connections

J1 Com1, Com2, and Printer

J4 and J7 Digital I/O

J6 Com3 and Com4

(a) If the 2350A BOS version is v1.84 or

later, ensure that J6 Jumper on pins 13- 14 of J1 has been removed. (see Figure 2-9 and Figure 2-10).

7. If a modem is installed, remove the phone cord.
8. Install the Ethernet card and set the jumpers as shown below.
Emerson Process Management Chromatograph Software (MON2000) — 2-60 INSTALLATION AND SETUP diagram (page 86)
Fig. 49

2350A Ethernet Installation JULY 2010

Show 6 more labels / short text on this page
Connection Ribbon Cable · J8 Com3 · J10 Com4 · J4 Printer · J11 Com2 · Connection Ribbon Cable
Page 87 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-61 MON2000
Page 87 · click to enlarge
Page 87

INSTALLATION AND SETUP 2-61 MON2000

Figure 2-9 Ethernet Jumper Configuration BOS v1.82 and Earlier

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-61 MON2000 diagram (page 87)
Fig. 50

Figure 2-10 Ethernet Jumper Configuration BOS v1.84 and Later

JULY 2010 2350A Ethernet Installation

Page 88 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-62 INSTALLATION AND SETUP
Page 88 · click to enlarge
Page 88

2-62 INSTALLATION AND SETUP

2.15.4 Ethernet TCP/IP Settings

If you plan to connect via an Ethernet connection, use the following instructions.

To configure the Ethernet TCP/IP settings for the 2350A GC Controller:

1. Obtain the IP Address, Subnet Mask, and Gateway addresses (numbers) for the Target 2350A GC Controller from your supervisor or Information Technology Department.

If the controller has never been configured for Ethernet communications, logon to the Model 500 GC using either the Direct Connect option, or connect remotely via a Modem.

2. Use the Application > TCP/IP menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — 2-62 INSTALLATION AND SETUP diagram (page 88)
Fig. 51
3. The TCP/IP dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 2-62 INSTALLATION AND SETUP diagram (page 88)
Fig. 52
4. Go to the Use DHCP/Specify an IP Address field and select Specify an IP Address .
5. In the IP Address field, enter the IP Address (number) obtained from your Supervisor or the Information Technology Department.

Ethernet TCP/IP Settings JULY 2010

Page 89 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-63 MON2000
Page 89 · click to enlarge
Page 89

INSTALLATION AND SETUP 2-63 MON2000

6. In the Subnet Mask field, enter the subnet mask address (number) obtained from your Supervisor or the Information Technology Department.
7. In the Gateway Address field, enter the Gateway Address (number) obtained from your Supervisor or the Information Technology Department.
8. Click the button to apply all changes.
9. Disconnect from the controller, using the toolbar icon or by using the File>Disconnect menu.

JULY 2010 Ethernet TCP/IP Settings

Page 90 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-64 INSTALLATION AND SETUP
Page 90 · click to enlarge
Page 90

2-64 INSTALLATION AND SETUP

2.15.5 MON2000 TCP/IP Settings

Emerson Process Management Chromatograph Software (MON2000) — 2-64 INSTALLATION AND SETUP diagram (page 90)
Fig. 53

To configure the Ethernet TCP/IP settings for the MON2000,

1. From the File menu, select GC Directory.
2. Ensure that an Ethernet location has been created or an existing location modified and that an Ethernet card installed in the PC has been selected in the field labeled Connection Type (Direct/Remote).
Emerson Process Management Chromatograph Software (MON2000) — 2-64 INSTALLATION AND SETUP diagram (page 90)
Fig. 54

MON2000 TCP/IP Settings JULY 2010

Page 91 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-65 MON2000
Page 91 · click to enlarge
Page 91

INSTALLATION AND SETUP 2-65 MON2000

3. Enter the same 'IP Address' into the Server Name/IP Address field of the GC Directory that was previously entered in the GC Controller ( see Section 2.11.2 ).
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-65 MON2000 diagram (page 91)
Fig. 55
4. Set the number of retries to three.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-65 MON2000 diagram (page 91)
Fig. 56

Click the button to apply your changes.

JULY 2010 MON2000 TCP/IP Settings

Page 92 of the Emerson Process Management Chromatograph Software (MON2000) user manual — CONNECT VIA GC EXTERNAL MODEM JULY 2010
Page 92 · click to enlarge
Page 92

2-66 INSTALLATION AND SETUP

2.16 CONNECT VIA GC EXTERNAL MODEM

Ensure you have a pre-existing PC connection to the GC unit before installing the external modem. If the GC unit is already on location, do this via another remote operator connection, or, onsite, via use of a portable PC and serial cable. Ideally, configure the GC unit and install the external modem prior to placement at the remote location.

2.16.1 HARDWARE SETUP

1. Halt any ongoing analysis runs.

Use the Control > Halt menu to access this function.

2. Disconnect AC power from the GC Controller.

SERIOUS PERSONAL INJURY OR DEATH POSSIBLE

Before removing the unit cover from the GC Controller, make certain the power supply switch is OFF and the AC power cord is disconnected. Observe all safety precautions when you are working in a hazardous environment.

Failure to observe all safety precautions could result in serious injury or death.

3. For the explosion-proof Controller, remove the cover from the GC Controller housing to expose the GC Controller Terminal Board for field wiring.

For the rack mount Controller, the rear of the enclosure is open; it allows access for most field wiring procedures without removing the enclosure.

CONNECT VIA GC EXTERNAL MODEM JULY 2010

Page 93 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-67 MON2000
Page 93 · click to enlarge
Page 93

INSTALLATION AND SETUP 2-67 MON2000

4. Connect a custom-made serial line cable (for RS-232 serial transmission) from the serial port of the external modem to an available serial port on the GC Controller Terminal Board for field wiring.

Special hardware modifications may be needed when operating an external GC modem from the GC serial ports 3 or 4 (COM3 or COM4), as follows:

  • DE-20782 (see drawing addendum of Model 500 Gas Chromatograph System Hardware Reference Manual , P/N 3-9000-537).
  • Transient protection modules M7 and/or M4, located on the back side of the GC Controller Terminal Board for field wiring, may need replacement with the transient protection modules for RS-232 signals (P/N 3-2350-027). Note that M4 serves COM3 and M7 serves COM4.
  • Make note of the serial port number you choose for connection. You will need this information later to make appropriate software settings.
  • Use an RS-232 serial cable no longer than 50 feet. Longer distances may result in spurious loss or corruption of transmitted data.
  • See Section 5.17 , for details of GC Controller serial port connectivity (i.e., port locations, pinouts, connector types, gender, cabling, etc.).
5. Connect a telephone line (from wall phone jack, if available) to the phone jack for the external modem.
6. Set external modem for Auto Answer (see external modem user manual).

JULY 2010 Hardware Setup

Page 94 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-68 INSTALLATION AND SETUP
Page 94 · click to enlarge
Page 94

2-68 INSTALLATION AND SETUP

2.16.2 Select Program Settings

Emerson Process Management Chromatograph Software (MON2000) — 2-68 INSTALLATION AND SETUP diagram (page 94)
Fig. 57

This function allows you to configure where database, GC applications, and Modbus Test program files are stored.

1. Use the File > Program Settings menu to access this function.
2. The Program Settings dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 2-68 INSTALLATION AND SETUP diagram (page 94)
Fig. 58
3. To use a Chromatogram file extension other than the default (e.g. .CGM ) or to select the location of saved data, GC database, and GC application files,
  • click in the Chromatogram file extension data field and type the desired file extension
  • click in the appropriate data field and type the desired directory,
  • click the button and use the directory tree to select the desired location

Select Program Settings JULY 2010

Or

Page 95 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-69 MON2000
Page 95 · click to enlarge
Page 95

INSTALLATION AND SETUP 2-69 MON2000

4. Use the provided check boxes to turn the following settings on/off. Note that the default settings are ON.

Table 2-9 Optional Program Settings

Editing replaces existing text in tables

If enabled, replaces selected table cell entry with user input. If disabled, appends user input to end of the selected table cell entry.

Prompt to confirm discard changes

Display confirmation dialog before exiting a dialog that contains changed data.

Prompt to confirm save changes

Display confirmation dialog before saving changed data.

Synchronize GC Controller’s time to PC on download, etc.

Program the GC Controller clock to match the PC clock upon download.

Tab from spreadsheet to next control

Move from field to field and button to button (rather than from cell to cell within a spreadsheet or tabular display).

Enable display in PPM If enabled, percentages less than 0.99% are displayed in PPM (parts per million) rather than as a percentage.

5. Click the button to apply your changes and return to the main screen.

Click the button to exit and return to the main screen without applying your changes.

JULY 2010 Select Program Settings

Setting Function

Page 96 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-70 INSTALLATION AND SETUP
Page 96 · click to enlarge
Page 96

2-70 INSTALLATION AND SETUP

2.16.3 Generate PC Config Report

This function allows you to generate or print out a report of the currently active GC application control settings. See Appendix A for a sample print-out of a PC Config Report.

Output data for the PC Config Report depends on the GC Controller and its application.

To view a PC Config Report already saved to disk, see Section 7.1.2.

1. Use the File > PC Config Report menu to access this function. The PC Config Report dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 2-70 INSTALLATION AND SETUP diagram (page 96)
Fig. 59
2. Click the check boxes (see Select column) to choose the data you want included in the configuration report.
Emerson Process Management Chromatograph Software (MON2000) — 2-70 INSTALLATION AND SETUP diagram (page 96)
Fig. 60

Click the button or press the F2 key to select all options.

Click the button or press F3 to deselect (i.e., clear) all checked options.

Generate PC Config Report JULY 2010

Page 97 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-71 MON2000
Page 97 · click to enlarge
Page 97

INSTALLATION AND SETUP 2-71 MON2000

3. Use the Output radio buttons to either save the report to disk, print the report with or without form feeds, or display the report via the report viewer ( see Section 7.1 ).

If you want to use the standard Windows® Print dialog to select a printer, deselect the Use default printer option.

The File option will open the Save As dialog, with CONFIG.TXT as the default file name.

The Printer (with form feeds) option will print the report, with form feeds, to your configured printer (see Section 3.7).

The Printer (without form feeds) option will print the report, without form feeds, to your configured printer (see Section 3.7).

The Screen option will launch the report viewer (see Section 7.1).

4. Click the button or press the F4 key to generate your customized PC Config Report.

If you are online with the GC Controller, a PC Config Report that includes all options can require up to 20 minutes to generate and save. Printing a full report can take longer.

If you press the ESC key, MON2000 will stop after the current option is completed.

A progress bar appears, indicating the completion status.

JULY 2010 Generate PC Config Report

Page 98 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-72 INSTALLATION AND SETUP
Page 98 · click to enlarge
Page 98

2-72 INSTALLATION AND SETUP

2.17 DOWNLOADING AN APPLICATION

This function allows the MON2000 operator to download applications to GC Controllers. GC applications with the .APP extension can only be down loaded to 2350A GC Controllers and Model 700 units. The Download feature is NOT available if the keylock switch is closed.

At the end of the download process, MON2000 displays the Set GC Time dialog, which allows you to accept or change the GC Controller's date and time.

Only download an application when performing an initial startup, downloading a new GC application, or after the GC unit has been out of service. You must first connect to the GC to use the Download function.

Emerson Process Management Chromatograph Software (MON2000) — 2-72 INSTALLATION AND SETUP diagram (page 98)
Fig. 61
1. Use the File > Download menu to access the Select GC Application File To Download dialog.
Emerson Process Management Chromatograph Software (MON2000) — 2-72 INSTALLATION AND SETUP diagram (page 98)
Fig. 62

Downloading an Application JULY 2010

Page 99 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-73 MON2000
Page 99 · click to enlarge
Page 99

INSTALLATION AND SETUP 2-73 MON2000

2. Click the desired .app file. The file name appears in the File name data field.

GC application files contain all control settings and application program functions (see Section 5) specific to a particular GC Controller.

For 2350A GC Controllers and the Model 700 controller, application files (.app files) are normally stored in the \GC\APP directory.

3. Click the button to download the selected application file to the connected GC unit.

Click the button to exit and return to the main window.

4. MON2000 reads the time at the GC Controller and the Set Time dialog appears appears for you to edit or accept ( see Section 4.7.2 for more information).
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-73 MON2000 diagram (page 99)
Fig. 63
5. A progress bar appears, indicating the completion status. See Table 2-10 for a listing of the automatically enabled function keys.

JULY 2010 Downloading an Application

Page 100 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-74 INSTALLATION AND SETUP
Page 100 · click to enlarge
Page 100

2-74 INSTALLATION AND SETUP

Table 2-10 Main Menu Function Keys Enabled by Download

F1 Help Access context-sensitive online help.

F2 Start Auto Sequence Initialize the Auto-Sequencing function. This function is also available via the Control > Auto Sequence menu.

F3 Halt Stop a function at the end of the current cycle. This function is also available via the Control > Halt menu.

F5 Edit TEVs Display the Timed Event Tables. This function is also available via the Application > Timed Events menu.

F6 Edit CDTs Display the Component Data Tables. This function is also available via the Application > Component Data menu.

F7 View Current CGM Display chromatogram of the stream currently being analyzed. This function is also available via the Chromatogram menu.

F8 View Last CGM Display chromatogram stored in the GC unit. This function is also available via the Chromatogram menu.

Downloading an Application JULY 2010

Key Function Description

Page 101 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-75 MON2000
Page 101 · click to enlarge
Page 101

INSTALLATION AND SETUP 2-75 MON2000

2.18 UPGRADING FLASH

Use this function to download a BOS file to a connected GC Controller.

Only download a BOS file when upgrading the GC firmware or after the GC unit has been out of service. You must first connect to the GC to use the Upgrade Flash function.

1. Use the File > Upgrade Flash menu to access the Select BOS File To Download dialog.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-75 MON2000 diagram (page 101)
Fig. 64
2. Click the desired BOS file. The file name appears in the File name data field.
3. Click the button to download the selected BOS file to the connected GC unit.

Click the button to exit and return to the main window.

JULY 2010 Upgrading Flash

Page 102 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-76 INSTALLATION AND SETUP
Page 102 · click to enlarge
Page 102

2-76 INSTALLATION AND SETUP

2.19 OFFLINE EDIT OF GC APPLICATION

Use this function to edit a GC application file that is saved on disk or has been downloaded (and is currently on, or being used by, the GC unit).

When selecting Offline Edit, you can choose either:

  • File Selection – Allows you to edit a GC application file stored on disk.
  • Upload Application – Allows you to upload the current GC application data from the GC Controller and save to file for future editing.
  • Upload BOS - Allows you to upload the current BOS (i.e., firmware) file from the GC Controller to save for future download.

See the following sections for detailed instructions.

Offline Edit of GC Application JULY 2010

Page 103 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-77 MON2000
Page 103 · click to enlarge
Page 103

INSTALLATION AND SETUP 2-77 MON2000

2.19.1 File Selection

This function allows you to edit a GC application file stored on disk.

1. Use the File > Offline Edit > File Selection menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-77 MON2000 diagram (page 103)
Fig. 65

The Open GC Application File dialog appears.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-77 MON2000 diagram (page 103)
Fig. 66

Using the directory tree, select the desired application file.

The GC application files (.APP) are normally stored in the \GC\App directory.

2. Click the button.
3. Edit this file by using the MON2000 Application functions ( see Section 5 for more information).

JULY 2010 File Selection

Page 104 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2.19.2 Upload Application
Page 104 · click to enlarge
Page 104

2-78 INSTALLATION AND SETUP

Emerson Process Management Chromatograph Software (MON2000) — 2-78 INSTALLATION AND SETUP diagram (page 104)
Fig. 67

When you have finished editing this file, you can download it to the GC Controller (see Section 2.17) or save for future use.

2.19.2 Upload Application

This function allows you to upload the application from an online GC Controller to a disk file and edit that file.

When connected to the GC, this function allows you to upload the current GC application data from the GC Controller and save to file for future editing. When the application upload is complete the connection will be terminated (i.e., disconnected).

1. Establish a connection to the GC.
2. Use the File > Offline Edit > Upload Application menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — 2.19.2 Upload Application diagram (page 104)
Fig. 68 · 2.19.2 Upload Application

Upload Application JULY 2010

Page 105 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-79 MON2000
Page 105 · click to enlarge
Page 105

INSTALLATION AND SETUP 2-79 MON2000

3. The Save Uploaded GC Application File menu appears.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-79 MON2000 diagram (page 105)
Fig. 69

Use this dialog to choose the directory and file name or create a special file name for this particular application upload. Click the

button.

4. MON2000 displays the message “Uploading C:\GC\APP\2350A.app” in the progress status bar.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-79 MON2000 diagram (page 105)
Fig. 70

MON2000 continues with the upload process and displays the message, “Uploading Segment 1 (through segment 15)” in the status bar and a progress bar appears.

Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-79 MON2000 diagram (page 105)
Fig. 71

Once all of the segments are uploaded and the application file (.app) has been saved,

JULY 2010 Upload Application

Page 106 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-80 INSTALLATION AND SETUP
Page 106 · click to enlarge
Page 106

2-80 INSTALLATION AND SETUP

MON2000 prompts you to save the parameters file.

5. The Save Parameter File dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 2-80 INSTALLATION AND SETUP diagram (page 106)
Fig. 72

Use this dialog to choose the directory location and file name or create your own

parameter file name. Click the button to continue.

6. MON2000 saves the parameter data to the selected directory. When the upload is complete, your connection to the GC unit is automatically terminated.
7. You are now viewing the uploaded application file off-line.

You can edit this application by using the MON2000 Application functions (see Section 5 for more information).

When you have finished editing this file, you can download it to the GC Controller (see Section 2.17) or save for future use.

Upload Application JULY 2010

Page 107 of the Emerson Process Management Chromatograph Software (MON2000) user manual — INSTALLATION AND SETUP 2-81 MON2000
Page 107 · click to enlarge
Page 107

INSTALLATION AND SETUP 2-81 MON2000

2.19.3 Upload BOS

When connected to the GC, this function allows you to upload the current BOS (i.e., firmware) file from the GC Controller to save for future download (see Section 2.18).

1. Use the File > Offline Edit > Upload BOS menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-81 MON2000 diagram (page 107)
Fig. 73
2. The Save Uploaded GC BOS File dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — INSTALLATION AND SETUP 2-81 MON2000 diagram (page 107)
Fig. 74

By default, GC BOS files are stored in C:\GC\APP folder.

JULY 2010 Upload BOS

Page 108 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 2-82 INSTALLATION AND SETUP
Page 108 · click to enlarge
Page 108

2-82 INSTALLATION AND SETUP

3. Use this dialog to choose the directory location and file name or create a file name for this particular upload. Click the

MON2000 will not perform a BOS upload during an analysis run. You must halt the analysis first (see Section 4.3) or wait until the GC is idle.

button to continue.

The message:

“Uploading C:\GC\APP\2350A.bos” appears in the status bar and a progress bar appears.

Emerson Process Management Chromatograph Software (MON2000) — 2-82 INSTALLATION AND SETUP diagram (page 108)
Fig. 75
4. MON2000 begins to upload the BOS to the selected file. When the upload is complete, normal MON2000 operations resume.

A BOS upload usually takes 10 minutes to complete.

Upload BOS JULY 2010

Page 109 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3.1 LOGGING ON
Page 109 · click to enlarge
Page 109

GETTING STARTED 3-1 MON2000

3 GETTING STARTED

This section shows you how to get started with the MON2000 software, covering such functions as logging on, navigation, and configuring your PC printer for MON2000.

3.1 LOGGING ON

1. Start the MON2000 software program by clicking the desktop icon or using the Windows ® Start menu ( see Section 2.7 ).
2. Click the User Name data field and type your assigned user name (up to 12 characters). Note that the user name is not case-sensitive and can include characters such as “*”, “-”, “.”, etc.

JULY 2010 Logging On

To log on,

Page 110 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-2 GETTING STARTED
Page 110 · click to enlarge
Page 110

3-2 GETTING STARTED

3. If required, click the User PIN data field and type your assigned PIN.

If you enter an incorrect user name or user PIN, a error message appears:

Emerson Process Management Chromatograph Software (MON2000) — 3-2 GETTING STARTED diagram (page 110)
Fig. 76

Click the button to exit and return to the Logon dialog.

4. When you have successfully logged on, the MON2000 main window appears.

Logging On JULY 2010

Page 111 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3.3 MON2000 USER INTERFACE
Page 111 · click to enlarge
Page 111

GETTING STARTED 3-3 MON2000

3.2 LOGGING OFF

Ensure that you exit the MON2000 program and shut down Windows before powering off the PC.

1. Disconnect from the GC unit ( see Section 2.10.2 ).
2. Exit the MON2000 program by
  • using the File > Exit menu
  • pressing the ESC key
  • clicking the button

3.3 MON2000 USER INTERFACE

Use the following sections to familiarize yourself with the MON2000 user interface.

JULY 2010 Logging Off

To log off,

Page 112 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-4 GETTING STARTED
Page 112 · click to enlarge
Page 112

3-4 GETTING STARTED

3.3.1 Main Window

The MON2000 main window typically looks like this:

titlebar, with connection data

toolbar GC Status bar

Use the available menus, icons (see toolbar area and Section 3.3.4), and function keys (see Section 3.4) to access the desired functions.

Main Window JULY 2010

menu bar · view area · MON2000 Status bar

Page 113 of the Emerson Process Management Chromatograph Software (MON2000) user manual — GETTING STARTED 3-5 MON2000
Page 113 · click to enlarge
Page 113

GETTING STARTED 3-5 MON2000

3.3.2 GC Status Bar

The status bar appears across the bottom of the application window. To display or hide the status bar, click the Status Bar command using the View menu (ALT + V, G).

Access this menu command from the View menu (ALT+ V, S) to display and hide the GC Status Bar, which displays the following:

Emerson Process Management Chromatograph Software (MON2000) — GETTING STARTED 3-5 MON2000 diagram (page 113)
Fig. 77
  • GC - name of the GC currently online
  • Alarm status (Unacknowledged)
  • Detector number
  • Mode (e.g., analysis; Calibration mode or Analysis mode)
  • Streams - number current streams
  • Next - next stream number to run)
  • Anly - analysis cycle
  • Run - run time
  • Date - current date is displayed
  • Time - current time is displayed

The left area of the status bar describes actions of menu items as you use the mouse pointer over the menu selection. This area similarly shows messages that describe the actions of toolbar buttons as you press and hold them. After viewing the description of the toolbar button command and you do not wish execute the command, move the mouse pointer away from the toolbar command, then release the mouse button.

JULY 2010 GC Status Bar

Page 114 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-6 GETTING STARTED
Page 114 · click to enlarge
Page 114

3-6 GETTING STARTED

The right areas of the status bar indicate which of the following keys are activated:

CAP the Caps Lock key is ON.

NUM the Num Lock key is ON

SCRL the Scroll Lock key is ON

Your connection status is also displayed in the titlebar.

If “Online Edit” appears, then MON2000 is connected to the GC and is recognizing the existing application.

If “Connected” appears, then MON2000 is connected to the GC but does not recognize the existing application.

If no connection data is displayed, there is no connection.

GC Status Bar JULY 2010

Indicator Description

Page 115 of the Emerson Process Management Chromatograph Software (MON2000) user manual — GETTING STARTED 3-7 MON2000
Page 115 · click to enlarge
Page 115

GETTING STARTED 3-7 MON2000

The GC Status bar consists of:

GC GC unit to which MON2000 is currently connected

A status flag indicates an active alarm, an unacknowledged alarm, or the File Edit mode (offline editing).

Det # detector monitoring the alarm status of the currently connected GC

A GC unit can have up to 2 detectors.

Mode mode of the Detector

  • Typical modes are:
  • Idle
  • Auto Cal
  • Auto Base
  • Auto Anly
  • FCal

Stream current stream being analyzed

Next next sample stream to be analyzed

Cycle total cycle time (seconds), selected before starting another analysis

Run time (seconds) elapsed since current cycle began

Date/Time date and time, per PC clock

JULY 2010 GC Status Bar

Item Description · Anly analysis time

Page 116 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3.3.5 Data Entry and Function Features
Page 116 · click to enlarge
Page 116

3-8 GETTING STARTED

3.3.3 View Menu

Use the View menu to toggle (i.e., switch ON or OFF) which main window component you wish to view. See Section 3.3.1 for an identification and definition of each component.

3.3.4 ToolTips

To view a ToolTip, move the mouse cursor to the Toolbar icon you want identified. MON2000 displays the ToolTip for five seconds.

3.3.5 Data Entry and Function Features

Use the data boxes, buttons, check boxes, and toggles to input data or to choose options and activate features.

click check box to turn setting ON or OFF

menus to select option

Emerson Process Management Chromatograph Software (MON2000) — 3.3.5 Data Entry and Function Features diagram (page 116)
Fig. 78 · 3.3.5 Data Entry and Function Features
Emerson Process Management Chromatograph Software (MON2000) — 3.3.5 Data Entry and Function Features diagram (page 116)
Fig. 79 · 3.3.5 Data Entry and Function Features

click radio button to select option

scroll to access additional settings

click push-buttons to activate features

View Menu JULY 2010

use dynamic pull-down · type input · in data fields

Page 117 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3.4 KEYBOARD SHORTCUTS
Page 117 · click to enlarge
Page 117

GETTING STARTED 3-9 MON2000

3.3.6 Shortcut to Save or Print Data

From any active dialog, right-click the displayed data to access this menu.

Choose the Save Sheet option to save data to disk or the Print Sheet option to queue data to your PC printer. MON2000 prints the report to your configured printer (see Section 3.7).

Microsoft Internet Explorer 5.0 (or later) is required to view a spreadsheet file saved in HTML format.

3.4 KEYBOARD SHORTCUTS

MON2000 supports keystrokes as well as a standard PC mouse interface. Use a mouse to perform point-and-click operations.

Refer to the following table for common keyboard shortcuts. See Table 3-2 for function keys that can be used after an application has been downloaded to the GC Controller.

Table 3-1 Frequently Used Keystrokes

  • Move cursor:
  • left or right in a data field
  • up or down in a menu or combo box
  • up or down (column), left or right (row) through displayed data entries
  • DELETE
  • Delete the character after cursor.
  • Delete selected rows from a table or return row values to the default settings.
  • ENTER
  • Add a blank item or row to selected table.
  • Activate the default control element (e.g., the OK button) in current window.

ESC Exit application or active window without saving data.

F1 Access context-sensitive help topic.

  • INSERT
  • Toggle between insert and type-over mode in selected cell.
  • Insert a new row above the highlighted row.

SHIFT+TAB Move to previous control element (e.g., button) or data field in window; see TAB description.

JULY 2010 Shortcut to Save or Print Data

Keystroke Action · ARROW keys

Page 118 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-10 GETTING STARTED
Page 118 · click to enlarge
Page 118

3-10 GETTING STARTED

Table 3-1 Frequently Used Keystrokes (Continued)

SPACE Toggle settings (via radio buttons or check boxes).

TAB Move to next control element (e.g., button) in window; use File > Program Settings (see Section 2.16.2) to use TAB key to move to next data field.

Table 3-2 Application-Specific Function Keys

F2 Start Auto Sequence

These keystrokes are available from the main menu only. Function key definitions will change per the active dialog/function.

Starts the Auto-Sequencing function (see Section 4.1).

Halts the GC program (e.g., an analysis run) at the end of the current cycle (see Section 4.3).

Displays the Timed Event Table (TEV) per specified stream (see Section 5.3).

Displays the Component Data Table (CDT) per specified stream (see Section 5.2).

F7 View Current CGM

Displays the chromatogram for the sample stream being analyzed (see Section 6.1).

F8 View Last GC CGM

Displays any chromatogram stored in the GC Controller (see Section 6.1).

Keyboard Shortcuts JULY 2010

Show 5 more labels / short text on this page
Keystroke Action · Function Key Action · F3 Halt · F5 Edit TEVs · F6 Edit CDTs
Page 119 of the Emerson Process Management Chromatograph Software (MON2000) user manual — GETTING STARTED 3-11 MON2000
Page 119 · click to enlarge
Page 119

GETTING STARTED 3-11 MON2000

3.5 PROCEDURES GUIDE

Use the following table to look up the related manual section and menu (and keystroke) for a given procedure.

Table 3-3 Menu and Manual Section per MON2000 Procedure

Task or Data Item Section(s) Menu [Keystroke]

24-Hour Average, component(s) measured

5.5.2 Application > Calculations > Averages

alarms, related components 5.2

Application > Component Data [F6] Application > Limit Alarms Application > Discrete Alarms Application > Discrete Outputs

5.6 5.7 5.12

alarms, stream number(s) programmed 5.6 Application > Limit Alarms

Analysis Report (on/off) 7.3 7.4 Reports > GC Printer Control Reports > MON2000 Printer Control

analysis time 5.3 Application > Timed Events [F5]

auto-calibration (on/off) 5.8 Application > Streams

Auto-Calibration Interval 5.8 Application > Streams

Auto-Calibration Start Time 5.8 Application > Streams

autocal time 5.8 Application > Streams

Baseline 5.8 Application > Streams, Baseline Values [F2]

base pressure used for calculations 5.8 Application > Streams

BTU-dry (on/off) 5.5.1 Application > Calculations > Control

calibration concentration 5.2 Application > Component Data [F6]

calibration cycle time 5.3 Application > Timed Events [F5]

calibration runs, number averaged 5.8 Application > Streams

calibration runs, number of 5.8 Application > Streams

calibration stream number 5.8 Application > Streams

communications baud rate, at GC 5.17 Application > Serial Ports

JULY 2010 Procedures Guide

Page 120 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-12 GETTING STARTED
Page 120 · click to enlarge
Page 120

3-12 GETTING STARTED

Table 3-3 Menu and Manual Section per MON2000 Procedure (Continued)

Task or Data Item Section(s) Menu [Keystroke]

communications baud rate, at PC 2.11.3 File > (COM ID)

component code and name 5.2 Application > Component Data [F6]

component full scale (for output) 5.1

Application > System (CGM Analog Output Config and Baseline Offset) Application > Analog Outputs

component(s) programmed for input 5.9 5.11 Application > Analog Inputs Application > Discrete Inputs

component(s) programmed for output 5.6 5.7 5.10 5.12

Application > Limit Alarms Application > Discrete Alarms Application > Analog Outputs Application > Discrete Outputs

component, retention time 5.2 Application > Component Data [F6]

component zero (for output) 5.10 Application > Analog Outputs

Compressibility (on/off) 5.5.1 Application > Calculations > Control

current date 4.7 Control > GC Time

current time 4.7 Control > GC Time

cycle time 5.3 Application > Timed Events [F5]

delete alarms 5.6 8.3 Application > Limit Alarms Log > Alarm Log

delete component from component list 5.2 Application > Component Data [F6], [DELETE]

delete inhibit, integration, peak width 5.3 Application > Timed Events [F5]

delete output(s) 5.6 5.7 5.10 5.12

Application > Limit Alarms Application > Discrete Alarms Application > Analog Outputs Application > Discrete Outputs

edit numeric value 5.4.1 Application > User Defined > Numeric

edit select values 5.4.2 Application > User Defined > Selection

edit text strings 5.4.3 Application > User Defined > Text Strings

Procedures Guide JULY 2010

Page 121 of the Emerson Process Management Chromatograph Software (MON2000) user manual — GETTING STARTED 3-13 MON2000
Page 121 · click to enlarge
Page 121

GETTING STARTED 3-13 MON2000

Table 3-3 Menu and Manual Section per MON2000 Procedure (Continued)

Task or Data Item Section(s) Menu [Keystroke]

existing alarm(s) 8.3 Log > Alarm Log

full-scale value (for input) 5.9 Application > Analog Inputs

GPM liquid equivalent (on/off) 5.5.1 Application > Calculations > Control

height or area measurement method 5.2 Application > Component Data [F6]

High Alarm 5.6 Application > Limit Alarms

(Analyzer) I.D. 2.11.2 5.1 File > GC Directory Application > System

inhibit on-off times 5.3 Application > Timed Events [F5]

input(s) being used 5.9 5.11 Application > Analog Inputs Application > Discrete Inputs

integration on-off times 5.3 Application > Timed Events [F5]

Low Alarm 5.6 Application > Limit Alarms

Mole percent (on/off) 5.5.1 Application > Calculations > Control

Normalization (on/off) 5.5.1 Application > Calculations > Control

outputs being used 5.6 5.7 5.10 5.12

Application > Limit Alarms Application > Discrete Alarms Application > Analog Outputs Application > Discrete Outputs

Peak Width, on time 5.3 Application > Timed Events [F5]

Ratio (on/off) 5.5.3 Application > Calculations > User Defined

Ratio Denominator 5.5.3 Application > Calculations > User Defined

Ratio, stream number(s) 5.5.3 Application > Calculations > User Defined

Relative Density (on/off) 5.5.1 Application > Calculations > Control

Response Factor 5.2 Application > Component Data [F6]

Response Factor, percent deviation 5.2 Application > Component Data [F6]

Retention Time, percent deviation 5.2 Application > Component Data [F6]

Rolling Average (on/off) 5.5.2 Application > Calculations > Averages

JULY 2010 Procedures Guide

Page 122 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-14 GETTING STARTED
Page 122 · click to enlarge
Page 122

3-14 GETTING STARTED

Table 3-3 Menu and Manual Section per MON2000 Procedure (Continued)

Task or Data Item Section(s) Menu [Keystroke]

Rolling Average, component for 5.5.2 Application > Calculations > Averages

Rolling Average, number of analyses 5.5.2 Application > Calculations > Averages

Rolling Average, stream number(s) 5.5.2 Application > Calculations > Averages

Spectrum Gain 5.3 Application > Timed Events [F5]

stream number(s) (for output) 5.6 5.7 5.10 5.12

Application > Limit Alarms Application > Discrete Alarms Application > Analog Outputs Application > Discrete Outputs

stream sequences skipped, number 5.1 5.8 Application > System Application > Streams

streams analyzed, number 5.1 5.8 Application > System Application > Streams

streams analyzed, sequence 5.1 5.8 Application > System Application > Streams

Valve on/off times 5.3 Application > Timed Events [F5]

Weight Percent (on/off) 5.5.1 Application > Calculations > Control

Wobbe value (on/off) 5.5.1 Application > Calculations > Control

Zero value (for input) 5.9 Application > Analog Inputs

Procedures Guide JULY 2010

Page 123 of the Emerson Process Management Chromatograph Software (MON2000) user manual — GETTING STARTED 3-15 MON2000
Page 123 · click to enlarge
Page 123

GETTING STARTED 3-15 MON2000

3.6 KEYLOCK SWITCH CONTROL

If your GC unit has the keylock control enabled, you will not be able to write changes or new data to the GC Controller. The following error dialog will display.

Emerson Process Management Chromatograph Software (MON2000) — GETTING STARTED 3-15 MON2000 diagram (page 123)
Fig. 80

To see which Modbus registers are affected by the keylock control, view the Communications section of the PC Config Report via the Report Viewer (see Appendix A for instructions and an example).

When closed ('on'), the GC keylock switch protects many of the GC Controller's parameters from being changed using MON2000.

Most of the dialog boxes in the Application’s menu allows you to view data but does not allow you to save changes when the keylock switch is closed. Several items in the Reports and Control menus are unavailable when the keylock switch is closed. The Users dialog cannot be downloaded to the GC Controller when the keylock switch is closed. A GC application cannot be downloaded to the GC Controller when the keylock switch is closed.

The keylock switch affects only online operations, i.e. those that are performed while connected to a GC Controller.

The keylock switch feature is only available on controllers with BOS version 1.6 and later. The GC Controller parameters protected by the keylock switch may vary slightly from version to version.

Since most of the GC Controller's parameters are protected by the keylock switch, it is easiest to list the operations that are available when the keylock switch is closed. The following operations are available when the keylock switch is closed:

JULY 2010 Keylock Switch Control

Page 124 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3.7 CONFIGURING YOUR PRINTER
Page 124 · click to enlarge
Page 124

3-16 GETTING STARTED

The stream sequences can be changed in the System dialog box. (GC applications version 1.7 and above.)

  • Automatic calibration in the Analog Outputs dialog box
  • Report display
  • GC report request (GC applications version 1.7 and above)
  • MON2000 printer control (GC applications version 1.7 and above)
  • View archive data
  • Trend data
  • Maintenance Log editing (GC applications version 1.7 and above)
  • Alarms (except clear/ack all active alarms)
  • Event log
  • Chromatogram viewer (including forced calibration using archived chromatograms)
  • Start Auto-Sequencing
  • Start single stream
  • Halt analysis
  • Start calibration
  • View and set GC time (GC applications version 1.7 and above)

3.7 CONFIGURING YOUR PRINTER

Use this function to configure the settings for the printer connected to your PC. This configuration will apply to any print job queued from MON2000.

Emerson Process Management Chromatograph Software (MON2000) — 3.7 CONFIGURING YOUR PRINTER diagram (page 124)
Fig. 81

The settings available depend on the printer type. Refer to the manufacture’s user manual for more information.

1. Use the File > Print Setup menu to access this function.

Configuring Your Printer JULY 2010

Page 125 of the Emerson Process Management Chromatograph Software (MON2000) user manual — GETTING STARTED 3-17 MON2000
Page 125 · click to enlarge
Page 125

GETTING STARTED 3-17 MON2000

2. The Printer Setup dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — GETTING STARTED 3-17 MON2000 diagram (page 125)
Fig. 82
3. Use the provided data fields, radio buttons, check and combo boxes to select the desired settings.

Your configuration will be cleared (i.e., the settings will return to the default values) when you exit MON2000.

Click the button to configure more advanced settings.

Refer to the printer operator manual for more information.

4. Click the button to queue the printer.

Click the button to abort and return to the main window.

JULY 2010 Configuring Your Printer

Page 126 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-18 GETTING STARTED
Page 126 · click to enlarge
Page 126

3-18 GETTING STARTED

3.8 USING ONLINE HELP

Currently, the online help feature contains all user information and instructions for each MON2000 function as well as the MON2000 system.

To access the MON2000 Application Help file,

  • Press the F1 key at any time to view help topics that discuss the currently active dialog or function.
  • Use the Help > Help Topics menu to view the help contents dialog.

Use the scroll bars and arrows to display more of a topic. You can also resize and/or move the topic dialog for better viewing convenience.

When applicable, a topic will contain links to other related topics, notes, and helpful popups. Use these links as you need.

Using Online Help JULY 2010

Page 127 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3.10 ABOUT MON2000
Page 127 · click to enlarge
Page 127

GETTING STARTED 3-19 MON2000

3.9 OPERATING MODES FOR MODEL 700

The Model 700 GC supports four different operating modes. Each mode allows the GC to analyze data from a given number of detectors, streams, and methods. See Table 3-4 for more information.

Table 3-4 Operating Modes for Model 700

0 1 1 1

1 2 1 1

2

3 2 1 2

3.10 ABOUT MON2000

To verify information such as the software version number, copyright date, and the call/ query ratio.

1. Use the Help > About MON2000 menu to access this dialog.

JULY 2010 Operating Modes for Model 700

Show 4 more labels / short text on this page
Mode ID Number · Detectors Supported · Streams Supported · Methods Supported
Page 128 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-20 GETTING STARTED
Page 128 · click to enlarge
Page 128

3-20 GETTING STARTED

2. MON2000 appears the About MON2000 dialog.
Emerson Process Management Chromatograph Software (MON2000) — 3-20 GETTING STARTED diagram (page 128)
Fig. 83

The Call/Query Ratio estimates the quality of your serial communications. If this ratio is less than or equal to 0.5, then the connection is problematic. The maximum value of 1 indicates a good connection.

The TAPI version represents the Telephony API software MON2000 accesses for dial-up (i.e., remote) communications.

3. Click the button to exit and return to the main window.

About MON2000 JULY 2010

Page 129 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4.1 AUTO SEQUENCE
Page 129 · click to enlarge
Page 129

CONTROL FUNCTIONS 4-1 MON2000

4 CONTROL FUNCTIONS

The options in the Control pull-down menu allow you to initiate various operating modes of the gas chromatograph (GC) unit.

4.1 AUTO SEQUENCE

Use this function to start continuous GC analysis runs that follow a predefined stream sequence. At the initial startup, the current stream (in the stream sequence) is the first stream analyzed. See Section 5.8 for detailed instructions on configuring the predefined sequence.

Emerson Process Management Chromatograph Software (MON2000) — 4.1 AUTO SEQUENCE diagram (page 129)
Fig. 84
1. Use the Control > Auto Sequence menu or press the F2 key to access this function.
2. The Start Auto Sequence dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 4.1 AUTO SEQUENCE diagram (page 129)
Fig. 85
3. Click the Purge check box to set the purging option. Note that the default setting is ON.

Purging allows sample gas to flow through the sample loop for 60 seconds prior to beginning the first analysis.

4. Click the button to continue.

Click the button to abort and return to the main window.

5. The Auto Sequence GC Analysis begins (if you selected the Purge Stream option, MON2000 performs this 60-second task first).

Use the Report Display option in the Reports menu to view the results of the Auto Sequence function.

JULY 2010 Auto Sequence

Page 130 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4.2 SINGLE STREAM
Page 130 · click to enlarge
Page 130

4-2 CONTROL FUNCTIONS

Use the status bar to monitor the progress of this function.

To change the operating parameters, such as the Timed Event Tables (TEVs) and Component Data Tables (CDTs), See Section 4 for information about Application functions.

Emerson Process Management Chromatograph Software (MON2000) — 4-2 CONTROL FUNCTIONS diagram (page 130)
Fig. 86

4.2 SINGLE STREAM

Use this function to start continuous GC analysis run(s) on a single stream.

Emerson Process Management Chromatograph Software (MON2000) — 4.2 SINGLE STREAM diagram (page 130)
Fig. 87
1. Use the Control > Single Stream menu to access this function.
2. The Start Single Stream Analysis dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 4.2 SINGLE STREAM diagram (page 130)
Fig. 88
3. Select the desired stream from the Stream menu.
4. Click the Purge check box to set the purging option. Note that the default setting is ON.

Purging allows sample gas to flow through the sample loop for 60 seconds prior to beginning the first analysis.

5. Click the Continuous operation check box to perform repetitive analysis runs. Note that the default setting is ON.

Deselect this option if you want to perform only one analysis run.

Single Stream JULY 2010

Page 131 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4.3 HALT
Page 131 · click to enlarge
Page 131

CONTROL FUNCTIONS 4-3 MON2000

6. Click the button to accept your selections and continue.

Click the button to abort and return to the main window.

7. The Single Stream GC Analysis begins (if you selected the Purge Stream option, MON2000 performs this 60-second task first).

Use the Report Display option in the Reports menu to view the results of the Single Stream function.

Use the status bar to monitor the progress of this function.

Emerson Process Management Chromatograph Software (MON2000) — CONTROL FUNCTIONS 4-3 MON2000 diagram (page 131)
Fig. 89

4.3 HALT

Use this function to stop the current operation mode at the end of the run in progress.

Emerson Process Management Chromatograph Software (MON2000) — 4.3 HALT diagram (page 131)
Fig. 90
1. Use the Control > Halt menu or press the F3 key to access this function.
2. MON2000 prompts you to confirm your selection.

Click the button to proceed.

Click the button to abort and return to the main window.

3. After the current run finishes, the function stops and “Idle” appears in the status bar.
Emerson Process Management Chromatograph Software (MON2000) — 4.3 HALT diagram (page 131)
Fig. 91

JULY 2010 Halt

Page 132 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4-4 CONTROL FUNCTIONS
Page 132 · click to enlarge
Page 132

4-4 CONTROL FUNCTIONS

4.4 CALIBRATION

Use this function to calibrate the GC unit.

Calibration runs are determined by the CDT and Streams settings. See Section 5.2 and Section 5.8 for detailed instructions on how to edit these settings.

1. Use the Control > Calibration menu to access this function.

If the GC unit is in Auto Sequence mode, calibration will not start until 2 or more analysis runs have been completed. This delay is required to complete the current analysis and the analysis of the stream currently purging through the valve.

2. The Start Calibration dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 4-4 CONTROL FUNCTIONS diagram (page 132)
Fig. 92

MON2000 displays the stream to be used for the calibration.

To select which stream is used for calibration, See Section 5.8.

3. Click the Purge check box to set the purging option. Note that the default setting is ON.

See Section 7.1 to select report parameters.

Calibration JULY 2010

Page 133 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4.5 BASELINE RUN
Page 133 · click to enlarge
Page 133

CONTROL FUNCTIONS 4-5 MON2000

Purging allows sample gas to flow through the sample loop for 60 seconds prior to beginning the first analysis.

4. Select the desired calibration type.

(a) Click the Manual radio button to

perform a manual calibration.

The CDT for the selected stream(s) will not be updated with raw data that are outside the acceptable deviations (as listed on the CDT). For more information, See Section 5.2.

For either choice, only VAR (variable) components on the CDT are updated. See Section 5.2 for details.

(b) Click the Forced radio button to perform

a manual calibration.

The CDT for the selected stream(s) will be updated with raw data that are outside the acceptable deviations (as listed on the CDT). For more information, See Section 5.2.

5. Click the button to accept your selections and continue.

Use the Report Display option in the Reports menu to view the results of the Calibration function.

Click the button to abort and return to the main window.

6. The Calibration begins.

Use the status bar to monitor the progress of this function.

4.5 BASELINE RUN

Use this function to initiate a single GC baseline run, to assess chromatogram peaks caused by the GC valve action alone. A baseline run is a calibration-like run during which no calibration or sample gas is injected (i.e., the valve is not fired).

Emerson Process Management Chromatograph Software (MON2000) — 4.5 BASELINE RUN diagram (page 133)
Fig. 93

JULY 2010 Baseline Run

Page 134 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4-6 CONTROL FUNCTIONS
Page 134 · click to enlarge
Page 134

4-6 CONTROL FUNCTIONS

1. Use the Control > Baseline Run menu to access this function.

If the GC unit is in Auto Sequence mode, MON2000 prompts that the continuous analysis must be stopped. Halt the analysis (See Section 4.3) and wait for an Idle status. Then continue initiating the baseline run.

2. The Start Baseline Run dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 4-6 CONTROL FUNCTIONS diagram (page 134)
Fig. 94

MON2000 displays the stream to be used for the baseline run.

To select which stream is used for the baseline calibration, See Section 5.8.

3. Click the button to continue.

Click the button to abort and return to the main window.

Use the status bar to monitor the progress of this function. Use the Report Display option in the Reports menu to view the results of the Baseline Run function?

Emerson Process Management Chromatograph Software (MON2000) — 4-6 CONTROL FUNCTIONS diagram (page 134)
Fig. 95

Baseline Run JULY 2010

Page 135 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4.6.1 Initiating an Auto BTU Start Up Run
Page 135 · click to enlarge
Page 135

CONTROL FUNCTIONS 4-7 MON2000

4.6 AUTO BTU START UP AND VALVE TIMING

The Auto BTU Start Up process, which takes about an hour to complete, automatically performs the following sequence of tasks:

1. Sets the timing for each valve.
2. Matches all the component peaks.
3. Adjusts the timed events based on peak integration times.
4. Runs a calibration.
5. Checks the range and order of response factors.
6. Adjusts the retention time deviations to avoid peak overlapping.

You can also initiate a valve timing run, which is the first step of the Auto BTU Start Up process, independently of that process. See “Starting a Valve Timing Run” on page 12 for more information.

Emerson Process Management Chromatograph Software (MON2000) — CONTROL FUNCTIONS 4-7 MON2000 diagram (page 135)
Fig. 96

4.6.1 Initiating an Auto BTU Start Up Run

To launch the Auto BTU Start Up process, do the following:

1. If the GC is currently performing an analysis, select Halt from the Control menu to put the GC in Idle mode after the current analysis run.

JULY 2010 Auto BTU Start Up and Valve Timing

Page 136 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4-8 CONTROL FUNCTIONS
Page 136 · click to enlarge
Page 136

4-8 CONTROL FUNCTIONS

2. Select Valve Timing from the Control menu. The Valve Timing window displays.
Emerson Process Management Chromatograph Software (MON2000) — 4-8 CONTROL FUNCTIONS diagram (page 136)
Fig. 97

MON2000 displays in the Stream field the calibration stream(s) to be used for the Auto BTU Start Up run.

3. Select the Auto BTU Start Up checkbox.
4. To use the default Component Data and Timed Event tables, select the Default CDT/ TEV checkbox.
Emerson Process Management Chromatograph Software (MON2000) — 4-8 CONTROL FUNCTIONS diagram (page 136)
Fig. 98

Initiating an Auto BTU Start Up Run JULY 2010

Page 137 of the Emerson Process Management Chromatograph Software (MON2000) user manual — CONTROL FUNCTIONS 4-9 MON2000
Page 137 · click to enlarge
Page 137

CONTROL FUNCTIONS 4-9 MON2000

It is strongly recommended that you use the Default CDT/TEV.

When the Default CDT/TEV checkbox is selected, the Valve # and CDT Component fields are read-only.

5. To use the GC’s existing CDT/TEV tables, select the Configured CDT/TEV checkbox.
Emerson Process Management Chromatograph Software (MON2000) — CONTROL FUNCTIONS 4-9 MON2000 diagram (page 137)
Fig. 99

You can enter a valve number in the Valve # field and select the component number from the drop-down list of the CDT Component field in the table below the checkboxes. The list of components are from the CDT of the corresponding calibration stream.

neo-Pentane is not available in the CDT Component drop-down list.

JULY 2010 Initiating an Auto BTU Start Up Run

Page 138 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4-10 CONTROL FUNCTIONS
Page 138 · click to enlarge
Page 138

4-10 CONTROL FUNCTIONS

6. To allow sample gas to flow through the sample loop for 60 seconds prior to beginning the first analysis, click the Purge stream for 60 seconds check box.

It is strongly recommended that you check the Purge stream for 60 seconds check box.

7. Click OK to begin the Auto BTU Start Up process. When the process starts, the GC Status window indicates that the GC is in Vlv Timing mode.
Emerson Process Management Chromatograph Software (MON2000) — 4-10 CONTROL FUNCTIONS diagram (page 138)
Fig. 100

If the Auto BTU Start Up run fails, an alarm will be triggered and displayed in the Alarm Log.

Initiating an Auto BTU Start Up Run JULY 2010

Page 139 of the Emerson Process Management Chromatograph Software (MON2000) user manual — CONTROL FUNCTIONS 4-11 MON2000
Page 139 · click to enlarge
Page 139

CONTROL FUNCTIONS 4-11 MON2000

The following six potential system alarms can be triggered during the Auto BTU Start Up process:

  • Valve Timing 1 Failure : Triggered if the timing for valve 1 cannot be determined after a ten second change of VALVE OFF time.
  • Valve Timing 2 Failure : Triggered if the timing for valve 2 cannot be determined after a ten second change of VALVE OFF time.
  • Valve Timing 3 Failure : Triggered if the timing for valve 3 cannot be determined after a ten second change of VALVE OFF time.
  • Auto BTU RT Failure - Missing Peaks : Triggered if not all the peaks are found after running the calibration stream to match all component peaks.
  • Auto BTU Response Factor Failure : Triggered if all of the response factors are not within +/- 50% of the default values after the calibration runs.
  • Auto BTU Calib Failure - RF Order : Triggered if the response factors are not in the proper order after the calibration runs.

A triggered alarm will be posted in the Alarm Log. See Section 8.3 for detailed instructions on reading Alarm Logs.

JULY 2010 Initiating an Auto BTU Start Up Run

Page 140 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4-12 CONTROL FUNCTIONS
Page 140 · click to enlarge
Page 140

4-12 CONTROL FUNCTIONS

4.6.2 Starting a Valve Timing Run

If the GC is in Auto Sequence mode, halt the analysis (See Section 4.3) and then return to the Valve Timing window.

1. Select Valve Timing from the Control menu. The Start Valve Timing Run dialog appears.
2. If the Auto BTU Start Up checkbox is selected, uncheck it. The valve timing feature is now enabled.
3. MON2000 displays in the Stream field the calibration stream(s) to be used for the valve timing run.
4. To allow sample gas to flow through the sample loop for 60 seconds prior to beginning the first analysis, click the Purge stream for 60 seconds check box..

Starting a Valve Timing Run JULY 2010

Page 141 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4.7.1 View GC Time
Page 141 · click to enlarge
Page 141

CONTROL FUNCTIONS 4-13 MON2000

5. You can enter a valve number in the Valve # field and select the component number from the drop-down list of the CDT Component field in the table below the checkboxes. The list of components are from the CDT of the corresponding calibration stream.
6. Click the button to accept your selections and to continue.

Click the button to abort and return to the main window.

7. The Valve Timing run begins.

If no change was made to a valve’s timing, the following message will be posted in the Alarm Log: “Valve Timing N Failure”, where N is the ID number for the valve. See Section 8.3 for detailed instructions on reading Alarm Logs. If all valves are successfully optimized, no messages will be recorded.

4.7 GC TIME

Emerson Process Management Chromatograph Software (MON2000) — 4.7 GC TIME diagram (page 141)
Fig. 101

Use the GC Time function to view and, if necessary, set the system date and time used by the GC Controller.

4.7.1 View GC Time

To view the system date and time used by the GC, do the following:

1. Use the Control > GC Time > View menu to select the View option.

JULY 2010 GC Time

Page 142 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4.7.2 Set GC Time
Page 142 · click to enlarge
Page 142

4-14 CONTROL FUNCTIONS

2. The following MON2000 dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 4-14 CONTROL FUNCTIONS diagram (page 142)
Fig. 102
3. Click the button to exit this dialog and return to the main window.

4.7.2 Set GC Time

To set the system date and time used by the GC,

1. Use the Control > GC Time > Set menu to select the Set option.
2. The Set GC Time dialog appears MON2000 indicates whether the GC Controller is on daylight saving or standard time, which depends on the configured system parameters.
3. Use the provided pull-down arrow to access an interactive calendar.
Emerson Process Management Chromatograph Software (MON2000) — 4.7.2 Set GC Time diagram (page 142)
Fig. 103 · 4.7.2 Set GC Time

Click the left/right arrows to view the previous or future months.

Click the desired day for the date. Note that a blue oval appears on the selected day.

4. Select which part of the time (hour, minute, or seconds) you wish to change. Use the

Set GC Time JULY 2010

Page 143 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4.8 STOP NOW
Page 143 · click to enlarge
Page 143

CONTROL FUNCTIONS 4-15 MON2000

provided scroll arrows to increase or decrease the highlighted value.

5. Click the button or press the enter key to write the selected date/time to the GC Controller.

Click the button to abort and return to the main window.

4.8 STOP NOW

Use this function to immediately stop all analysis runs.

Emerson Process Management Chromatograph Software (MON2000) — 4.8 STOP NOW diagram (page 143)
Fig. 104
1. Use the Control > Stop Now menu to access this function.
2. MON2000 prompts you to confirm your selection.
Emerson Process Management Chromatograph Software (MON2000) — 4.8 STOP NOW diagram (page 143)
Fig. 105

Click the button to proceed.

Click the button to abort and return to the main window.

This function forces the system to the Idle mode. If Stop Now is performed while an analysis is in progress, the components will continue to elute from the columns per the Idle mode. All resultant data will be lost.

Do not perform a Stop Now unless absolutely necessary. Whenever possible, use the Halt function.

JULY 2010 Stop Now

Page 144 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 4-16 CONTROL FUNCTIONS
Page 144 · click to enlarge
Page 144

4-16 CONTROL FUNCTIONS

3. Any GC function in progress is stopped. The Mode field in the status bar displays ‘Idle’.

Stop Now JULY 2010

Page 145 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.1 SYSTEM
Page 145 · click to enlarge
Page 145

APPLICATION FUNCTIONS 5-1 MON2000

5 APPLICATION FUNCTIONS

The options in the Application pull-down menu allow you to manipulate or edit the control settings that exist for particular gas chromatograph (GC) application functions. You can edit an application while online with the GC Controller or when using the Offline Edit function (see Section 2.19 for information on Offline Edit; see Section 2.17 to download an application first).

5.1 SYSTEM

Emerson Process Management Chromatograph Software (MON2000) — 5.1 SYSTEM diagram (page 145)
Fig. 106

Use this function to define or edit the GC stream sequence, unit name, and system description. You can also use this function to activate Metric Base Condition for calculations based on metric units.

The System dialog also includes a series of non- editable fields for reference or trouble shooting. See Table 5-1 for a summary of the items provided by the System dialog, along with the related functions and editing status.

1. Use the Application > System menu to access this function.

JULY 2010 System

Page 146 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-2 APPLICATION FUNCTIONS
Page 146 · click to enlarge
Page 146

5-2 APPLICATION FUNCTIONS

2. The System dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5-2 APPLICATION FUNCTIONS diagram (page 146)
Fig. 107

Click and edit any value presented in the white cells.

3. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

System JULY 2010

Page 147 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-3 MON2000
Page 147 · click to enlarge
Page 147

APPLICATION FUNCTIONS 5-3 MON2000

Table 5-1 Description of Items in System Dialog

Stream Sequence Defines the order of stream analysis for Detector 1.

Stream Sequence 2 Defines the order of stream analysis for Detector 2; only available with selected hardware.

Aux. Stream Sequence Defines the auxiliary order of stream analysis for Detector 1. Can be used to run an alternative stream sequence when a Discrete Input is set (see Section 5.11 for detailed instructions).

Aux. Stream Sequence 2 Additional alternative stream sequence provided for second detector in dual detector applications.

Auxiliary order of stream analysis for Detector 2. Can be used to run an alternative stream sequence when a Discrete Input is set (see Section 5.11 for detailed instructions).

Analyzer Name Defines the GC unit name that appears in the status bar of the main window when connected to the GC Con- troller. Can contain up to 12 characters.

Unit Type Identifies GC unit type (i.e., model number).

System Description A field to record miscellaneous reference information to further identify the system. Can contain up to 28 characters.

Chromatogram Buffer Size Provides number of bytes used to store a chromatogram.

Max Peaks Provides maximum number of peaks (application-spe- cific).

Item Function Edit · JULY 2010 System

Page 148 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-4 APPLICATION FUNCTIONS
Page 148 · click to enlarge
Page 148

5-4 APPLICATION FUNCTIONS

Table 5-1 Description of Items in System Dialog (Continued)

Number of Valves Identifies number of chromatographic valves (application-specific).

Daylight Savings Time Allows the GC Controller to adjust automatically for DST. “Normal” enables automatic DST adjustment (United States). “Override” disables automatic DST adjustment.

CGM Analog Output Displays chromatogram trace signals for strip chart generation from Detectors 1 and/or 2, trace 1 only. Identifies the analog output number.

If this field is set to 0, then no analog output will produce the CGM signal.

When assigning a CGM Analog Output, also check the Analog Outputs dialog (see Section 5.10). A CGM Analog Output setting that uses a previously assigned analog output number will produce erratic CGM trace output, possibly with incorrect scaling.

Do not assign this item analog output number 1 if the connected GC unit supports bargraph output (see Section 5.10.3 for details).

CGM Baseline Offset A fixed offset (percent of full scale) added to the CGM analog output to allow the user to conveniently position the chromatogram on a strip chart recorder. Displays chromatogram trace signals for strip chart generation from Detectors 1 and/or 2. Identifies the baseline offset.

CFG Baseline Number Identifies the Standard GC Application software from which this GC application or configuration was developed (i.e., the number of the standard configura- tion file that was used to generate the current applica- tion).

Metric Base Condition Enables (metric) or disables (off) whether MON2000 displays the GC Control Calculation in metric units. See Section 5.5.1 for details.

Item Function Edit · System JULY 2010

Page 149 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-5 MON2000
Page 149 · click to enlarge
Page 149

APPLICATION FUNCTIONS 5-5 MON2000

Table 5-1 Description of Items in System Dialog (Continued)

Max Archive Averages maximum number of averages (1-254)

Max Archive Avg. Records maximum number of archive records per average without an extended memory card (1-128)

Application Revision (non-edit) - revision level of the current application

Application Base Name (non-edit) - base name of the application

Archive Days (non-edit) – maximum number of days that archive records are kept (1-400)

BOS Revision (non-edit) – revision level of the BOS

Item Function Edit · JULY 2010 System

Page 150 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-6 APPLICATION FUNCTIONS
Page 150 · click to enlarge
Page 150

5-6 APPLICATION FUNCTIONS

5.2 COMPONENT DATA

Emerson Process Management Chromatograph Software (MON2000) — 5-6 APPLICATION FUNCTIONS diagram (page 150)
Fig. 108

Use this function to view and/or edit the Component Data Tables (CDTs) for a given GC application.

1. Use the Application > Component Data menu or press the F6 key to access this function.
2. The Component Data Tables menu appears, listing the CDTs available per associated stream(s).
Emerson Process Management Chromatograph Software (MON2000) — 5-6 APPLICATION FUNCTIONS diagram (page 150)
Fig. 109

Choose the desired CDT.

The number of available CDTs depends on the GC unit configuration.

The standard GC application contains 4 CDTs. MON2000 can operate GC applications with up to 5 CDTs.

To assign a CDT to a stream, see Section 5.8.

3. Click the button to accept your selection. The Component Data Table dialog appears.

Click the button to abort and return to the main window.

Component Data JULY 2010

Page 151 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-7 MON2000
Page 151 · click to enlarge
Page 151

APPLICATION FUNCTIONS 5-7 MON2000

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-7 MON2000 diagram (page 151)
Fig. 110
4. View or edit data. Double-click a given table cell or use one of the edit function buttons to make your changes.

To add a new component, move the cursor to the last table cell (last row, last column) and press ENTER.

Some table cells will change into pull-down menus when selected. Choose your data from the pull-down menu provided.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-7 MON2000 diagram (page 151)
Fig. 111

For more information on the editing functions available, see Section 5.2.1 through Section 5.2.4.

For descriptions of the standard data presented in a CDT, see Table 5-2.

5. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

JULY 2010 Component Data

Page 152 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-8 APPLICATION FUNCTIONS
Page 152 · click to enlarge
Page 152

5-8 APPLICATION FUNCTIONS

Table 5-2 Description of Standard Parameters in a CDT

Number (#) component number

An index number that acts as an identification label when used in other menus/dialogs (e.g., Limit Alarms, Analog Outputs). Up to 20 components can be defined per data table.

Component names assigned to the various components displayed

Components can be selected from the standard list, by pressing the F2 key (see Section 5.2.1), or entered by the user. Up to 15 characters can be used for the component name.

USR/STD whether the component data is standard or user- defined

If the component was added by the Select Standard Component function (see Section 5.2.1) and no changes were made to the component data, STD displays. If the component was added manually by the user or modified, USR displays.

Detector ID component detector number

Retention Time time in seconds that apex of the component peak will appear

Range = 0 to 6000 seconds

Ensure that the component retention times do not exceed the analysis time, as defined by the TEV Table (see Section 5.3). MON2000 does not automatically prevent the user from defining excessive component retention times.

Response Factor equal to the raw data of component peak divided by component concentration

The maximum value is 1.0E+38.

Component Data JULY 2010

Item Function Edit

Page 153 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-9 MON2000
Page 153 · click to enlarge
Page 153

APPLICATION FUNCTIONS 5-9 MON2000

Table 5-2 Description of Standard Parameters in a CDT (Continued)

Fixed/Variable whether a component response factor is fixed or variable

A fixed response factor will not be updated during calibration.

Calibration Concentration component concentration in calibration gas

To enter a PPM, type “Xp”; e.g., “4.0p” will appear as “4.0 ppm”. Values above 90 PPM are converted to percentages.

Analysis Method used to determine component raw data value

Use the pull-down menu to select either area, height, fixed, or the appropriate analog input (used with auxiliary analyzer).

RT Percent Dev. maximum acceptable percentage of deviation of the new retention time from the current retention time

RT Secs. Dev. maximum acceptable deviation time, in seconds, of the new retention time from the current retention time

RT Update Method determines when component retention time will be updated

Select Analysis to update after each run.

Select Calib to update only during final calibration run.

Resp. Factor % Dev. maximum acceptable percent deviation between the new response factor and the current response factor

Total Concentration total concentration of all components

JULY 2010 Component Data

Item Function Edit

Page 154 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.2.1 Select Standard Component(s)
Page 154 · click to enlarge
Page 154

5-10 APPLICATION FUNCTIONS

See Table 5-2 an asterisk (*) in the column heading indicates a column that contains standard component values. If any of these values is modified, the corresponding component becomes non-standard (i.e., user-defined). It is not recommended that you edit a standard component value.

5.2.1 Select Standard Component(s)

To select standard component(s) to be used by the displayed CDT,

To add a new row to the CDT, click the last cell in the last row and press the ENTER key.

1. Select the Component cell where you want the new component.
2. Click the button or press the

F2 key to access the Standard Components menu.

3. The Standard Components menu appears.
Emerson Process Management Chromatograph Software (MON2000) — 5.2.1 Select Standard Component(s) diagram (page 154)
Fig. 112 · 5.2.1 Select Standard Component(s)
4. Double-click the selected component to add it to the CDT. The component is automatically inserted into the selected table cell.

Click the button or press the ESC key to abort and return to the Component Data Table dialog.

Select Standard Component(s) JULY 2010

Page 155 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.2.3 Update Standard Component(s)
Page 155 · click to enlarge
Page 155

APPLICATION FUNCTIONS 5-11 MON2000

5.2.2 Select Standard Values

To read the values for the standard

components, click the button or press the F3 key to access the Standard Components dialog.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-11 MON2000 diagram (page 155)
Fig. 113

Click the button to return to the CDT.

5.2.3 Update Standard Component(s)

Click the or press the F5 key to upgrade the Standard Components list of this CDT to match the MON2000 Standard Component menu.

Typically, this function should only be performed when the MON2000 software is upgraded.

User-defined components are not affected by this update.

JULY 2010 Select Standard Values

Page 156 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-12 APPLICATION FUNCTIONS
Page 156 · click to enlarge
Page 156

5-12 APPLICATION FUNCTIONS

5.2.4 View Raw Data

To view the raw data for the displayed CDT,

1. Select the desired row from the currently displayed CDT.
2. Click the button or press the

F4 key to access the Raw Data dialog.

3. The Stream Selection dialog appears. Double-click the desired stream or click the

button to exit. Only the streams associated with this CDT will display in the list menu.

4. The Raw Data dialog appears, listing the peak raw data from the last run of the stream represented by the CDT.

When first displayed during a session, the Select Stream dialog highlights the stream that has the most recent raw data. Thereafter, the previously selected stream is highlighted.

Emerson Process Management Chromatograph Software (MON2000) — 5-12 APPLICATION FUNCTIONS diagram (page 156)
Fig. 114
5. The MON2000 operator can now copy the Retention Time from the Raw Data screen to the CDT as follows:

(a) Before pulling up the Raw Data screen,

click Component #1 of the CDT. (b) Click the Raw Data button or press the

(F4) key to access the Raw Data screen. (c) Press the ENTER key to copy the

Retention Time (the next component's

View Raw Data JULY 2010

Page 157 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-13 MON2000
Page 157 · click to enlarge
Page 157

APPLICATION FUNCTIONS 5-13 MON2000

RT is automatically targeted by MON2000). Then, press the DOWN ARROW key to advance to the next 'Raw Data' value. (d) Repeat Step (c) until all Retention Times

have been copied to the CDT.

6. Click the button to clear the Raw Data screen.

Following is an example of a Raw Data Table.

Table 5-3 Raw Data Table

Last of Fused Group 3

Some methods are a combination of the above types, e.g. 103 is LAST OF FUSED GROUP (3) and INHIBIT ON (100).

Resolved Rider Peak 5

Last of Fused Rider 7

Baseline Forced at Valley 60

Forced Integrate On 200

Forced Integrate End 300

JULY 2010 View Raw Data

Show 11 more labels / short text on this page
Description Method · Baseline Resolved 1 · Fused 2 · Tangent Skimmed 4 · Fused Rider 6 · Tailing Peak 8 · Forward Horizontal 20 · Backward Horizontal 40 · Inhibit On 100 · Plateau Detect 400 · Summation 500
Page 158 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.3 TIMED EVENTS
Page 158 · click to enlarge
Page 158

5-14 APPLICATION FUNCTIONS

5.2.5 Sort Retention Time

To sort the components listed in the displayed CDT by their retention times,

MON2000 sorts components by the corresponding detector number first, then by the Retention Time.

1. Click the button or press the

F6 key to access this function.

2. MON2000 automatically sorts the components.

Sorting the CDT components by retention time may affect one or more of the following:

  • Analog Outputs
  • Averages
  • Bargraphs
  • Limit Alarms
  • User Defined Calculations

5.3 TIMED EVENTS

Emerson Process Management Chromatograph Software (MON2000) — 5.3 TIMED EVENTS diagram (page 158)
Fig. 115

Use this function to view and/or edit the GC Timed Events Tables (TEVs) assigned to and used by particular gas streams.

Timed Events Description•

See Section 6.6.2 for details about editing from the Chromatogram > Chromatogram Viewer menu.

  • Analysis Time - Input total analysis time. LIMITS: 0.0 to 6000.0 seconds
  • Cycle Time - Input total cycle time. LIMITS: 0.0 to 6000.0 seconds
  • Off During Baseline (per sample valve) - toggle check boxes to select which valve(s) remain off during a baseline run

The Cycle Time must be at least 10 seconds greater than the Analysis Time.

Sort Retention Time JULY 2010

Page 159 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Integration Events
Page 159 · click to enlarge
Page 159

APPLICATION FUNCTIONS 5-15 MON2000

Valve Events

  • Type - use the pull-down menu to select between valve number (Valve #), discrete output channel number (DO #), and stream switch (Strm Sw); additional selections SSO1, SSO2, Bleed1, Bleed2, Block1 and Block2 are available for the Model 700 GC applications
  • Valve/DO # (for valves and discrete outputs only) - input the valve number or discrete output channel number. Use the provided combo box to select valve number by name
  • State (for all types except for stream switch) - use the pull-down menu to toggle ON\OFF
  • Time - input the time that the event is to occur. LIMITS: 0.0 to 6000.0 seconds

Integration Events

  • Type - use the pull-down menu to select between Inhibit, Integrate, Auto Zero, Slope Sensitivity, Peak Width, Single Baseline, FID Gain, Summation, or Fused Override
  • Value - For peak width or slope sensitivity, input the number of points to be used. LIMITS: Peak Width and Slope Sensitivity, 1 to 99
  • For FID gain, use the pull-down menu to toggle Low\High.
  • For Single Baseline, use the pull-down menu to toggle OFF\BGN (Beginning)\END.
  • Except for AutoZero, for all others use the pull-down menu to toggle ON\OFF. Auto Zero does not have 'Value' field.
  • Det - enter the appropriate detector number that the event will effect. Valid range is '1' or '2'
  • Time - input the time that the event is to occur. LIMITS: 0.0 to 6000.0 seconds

JULY 2010 Timed Events

Page 160 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-16 APPLICATION FUNCTIONS
Page 160 · click to enlarge
Page 160

5-16 APPLICATION FUNCTIONS

Spectrum Gain Events

  • Detector # - use the pull-down menu to select between detectors #1 and #2
  • Gain - input the gain value for that detector. LIMITS: 0 to 24
  • Time - input the time that the event is to occur. LIMITS: 0.0 to 6000.0 seconds

See Section 6.6 for editing TEVs from CGM Viewer.

To access the Timed events feature,

1. Use the Application > Timed Events menu or press the F5 key to access this function.
2. The Timed Event Tables menu displays, listing the TEVs available per associated stream(s).
Emerson Process Management Chromatograph Software (MON2000) — 5-16 APPLICATION FUNCTIONS diagram (page 160)
Fig. 116

Choose the desired TEV.

The number of available TEVs depends on the GC unit configuration.

The standard GC application contains four TEVs. MON2000 can operate GC Applications with up to five TEVs.

To assign a TEV to a stream, use the Streams function (see Section 5.8).

Timed Events JULY 2010

Page 161 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-17 MON2000
Page 161 · click to enlarge
Page 161

APPLICATION FUNCTIONS 5-17 MON2000

3. Click the button to accept your selection. The Timed Events Table dialog appears.

Click the button to abort and MON2000 returns you to the main window.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-17 MON2000 diagram (page 161)
Fig. 117
4. View or edit data.

To add a new timed event, move the cursor to the last table cell (last row, last column) and press the ENTER key.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-17 MON2000 diagram (page 161)
Fig. 118

Use the check boxes to toggle the listed sample valves ON or OFF.

Double-click a given table cell or use one of the edit function buttons to make your changes. Some table cells will change into pull-down menus when selected. Choose your data from the pull-down menu provided.

JULY 2010 Timed Events

Page 162 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-18 APPLICATION FUNCTIONS
Page 162 · click to enlarge
Page 162

5-18 APPLICATION FUNCTIONS

5. To sort data based on time, click the button for the corresponding table.
6. Click the button to accept your changes and return to the main window.

If you configure duplicate TEVs (i.e., 2 or more TEVs contain the same parameter settings), an error dialog will display but your data will be saved.

Click the button to abort and return to the main window.

Timed Events JULY 2010

Page 163 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.4.1 Numeric
Page 163 · click to enlarge
Page 163

APPLICATION FUNCTIONS 5-19 MON2000

5.4 USER DEFINED

The user-defined functions allow you to edit a user-defined operator entry for an application- specific variable. These variables may be used in reports, calculations, and/or for controlling the GC operations; however, user-defined variables are not used by all applications.

5.4.1 Numeric

Use this function to edit values for defined initialized variables. The User Defined Numerics dialog defines the names, types, and values for all application-specific numeric variables.

Emerson Process Management Chromatograph Software (MON2000) — 5.4.1 Numeric diagram (page 163)
Fig. 119 · 5.4.1 Numeric
1. Use the Application > User Defined > Numerics menu to access this function.
2. The User-Defined Numerics dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5.4.1 Numeric diagram (page 163)
Fig. 120 · 5.4.1 Numeric

Double-click the desired Value cell and type the new number.

3. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

JULY 2010 User Defined

Page 164 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.4.3 Text Strings
Page 164 · click to enlarge
Page 164

5-20 APPLICATION FUNCTIONS

5.4.2 Selection

Use this function to edit values for defined selection variables. The User Defined Selections dialog defines the names and values for all application-specific selection variables.

1. Use the Application > User Defined > Selection menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — 5-20 APPLICATION FUNCTIONS diagram (page 164)
Fig. 121
2. The User-Defined Selections dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5-20 APPLICATION FUNCTIONS diagram (page 164)
Fig. 122

Double-click the desired Option cell and use the provided pull-down menu to select the new setting.

3. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

5.4.3 Text Strings

Use this function to edit text strings for defined string variables that may be used in reports. The User Defined Text Strings dialog defines the symbol name, size, and assigned text string for all application-specific text variables.

Emerson Process Management Chromatograph Software (MON2000) — 5.4.3 Text Strings diagram (page 164)
Fig. 123 · 5.4.3 Text Strings
1. Use the Application > User Defined > Text Strings menu to access this function.

Selection JULY 2010

Page 165 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.5.1 Control
Page 165 · click to enlarge
Page 165

APPLICATION FUNCTIONS 5-21 MON2000

2. The User-Defined Text Strings dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-21 MON2000 diagram (page 165)
Fig. 124

Double-click the desired String cell and type the new text.

3. Click the button to accept your changes and return to the main window.

String length (i.e., how many characters you can enter) is limited to the number cited in the Size cell.

Click the button to abort and return to the main window.

5.5 CALCULATIONS

These options allow you to activate and define how the output of standard or user-defined chromatograph analysis data is used in various calculations.

5.5.1 Control

Emerson Process Management Chromatograph Software (MON2000) — 5.5.1 Control diagram (page 165)
Fig. 125 · 5.5.1 Control

Use this function to designate, by streams, the standard calculations (e.g., mole percent, liquid volume, gas density, Wobbe index, etc.) that should be performed from the GC analysis data. Calculations are enabled or disabled for each sample stream.

1. Use the Application > Calculations > Control menu to access this function.

JULY 2010 Calculations

Page 166 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-22 APPLICATION FUNCTIONS
Page 166 · click to enlarge
Page 166

5-22 APPLICATION FUNCTIONS

2. The Control Calculations dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5-22 APPLICATION FUNCTIONS diagram (page 166)
Fig. 126

Click the check boxes to toggle a calculation ON or OFF for a given stream.

Scroll towards the end of the list to set metric calculations.

You can use the arrow keys to move from one Stream cell to another, and press the space bar to toggle the calculation.

Emerson Process Management Chromatograph Software (MON2000) — 5-22 APPLICATION FUNCTIONS diagram (page 166)
Fig. 127

Before enabling a metric calculation, ensure that you also enable the Metric Base Conditions setting in the Systems dialog (see Section 5.1).

3. Click the button to accept your selections and return to the main window.

Click the button to abort and return to the main window.

Control JULY 2010

Page 167 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-23 MON2000
Page 167 · click to enlarge
Page 167

APPLICATION FUNCTIONS 5-23 MON2000

5.5.2 Averages

Use this function to designate, by streams and components, averages of standard calculations MON2000 should perform.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-23 MON2000 diagram (page 167)
Fig. 128
1. Use the Application > Calculations > Averages menu to access this function.
2. The Averages Calculations dialog appears.

Double-click the desired cell and either type the new data or use the provided pull-down menu to select the new setting.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-23 MON2000 diagram (page 167)
Fig. 129

When changing the reset time, note the following:

24-Hour Day, Reset Time

Monthly Day, Reset Time

Show 4 more labels / short text on this page
Average Setting Used · Hourly Reset Time · Weekly Weekday · JULY 2010 Averages
Page 168 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-24 APPLICATION FUNCTIONS
Page 168 · click to enlarge
Page 168

5-24 APPLICATION FUNCTIONS

To set a custom time interval per calculation, set the Type to Variable and enter the desired interval length in the Hours cell. A setting of “0” means an average calculation will be performed for each analysis run of that stream.

Any GC analysis variable with defined alarm limits cannot be averaged or archived while in an active alarm state (see Section 5.6 for more information on Limit Alarms).

To disable this feature, use the Control Calculations function to set Avg Limit Alarm Test to OFF (see Section 5.5.1).

3. Use the provided pull-down menu to add a new variable to the list.
4. To view an archive of averages for a given variable, select the desired variable and

click the button or press the F5 key.

The Archive Data dialog for that variable appears.

Emerson Process Management Chromatograph Software (MON2000) — 5-24 APPLICATION FUNCTIONS diagram (page 168)
Fig. 130

Averages JULY 2010

Page 169 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.5.3 User Defined
Page 169 · click to enlarge
Page 169

APPLICATION FUNCTIONS 5-25 MON2000

To save this archive to disk, right-click the archive table and select the Save Sheet option from the pop-up menu.

An error message displays when the last available stream or component is reached.

5. To copy the stream settings from a highlighted row and apply them to the next

row, click the button or press the F7 key. This feature also increments the Stream value to the next available stream (e.g., incrementing from Stream 2 to Stream 8), per the GC application.

6. To copy the component settings from a highlighted row and apply them to the next

row, click the button or press the F8 key. This feature also increments the Component value to the next available component (e.g., incrementing from “Ammonia” to “Benzene”), per the GC application.

7. To copy the Restart setting from a highlighted row and apply it to the next

row, click the button or press the F9 key.

8. Click the button to accept your selections and return to the main window.

Click the button to abort and return to the main window.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-25 MON2000 diagram (page 169)
Fig. 131

5.5.3 User Defined

Use this function to create and edit customized calculations on GC analysis data. MON2000 will allow up to 20 user-defined calculations.

1. Use the Application > Calculations > User Defined menu to access this function.

JULY 2010 User Defined

Page 170 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-26 APPLICATION FUNCTIONS
Page 170 · click to enlarge
Page 170

5-26 APPLICATION FUNCTIONS

2. The User-Defined Calculations dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5-26 APPLICATION FUNCTIONS diagram (page 170)
Fig. 132
3. Click the calculation name you wish to edit, or click in the next available blank Label cell. The Label field accepts 15 characters.
4. Click the button to edit or create the calculation function. The Edit User-Defined Calculation dialog appears.

Your user-defined calculation becomes a variable that can be used when:

  • calculating averages
  • assigning alarm limits
  • calculating analog outputs
  • generating bargraphs and reports
  • creating other user-defined calculations
Emerson Process Management Chromatograph Software (MON2000) — 5-26 APPLICATION FUNCTIONS diagram (page 170)
Fig. 133

User Defined JULY 2010

Page 171 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.6 LIMIT ALARMS
Page 171 · click to enlarge
Page 171

APPLICATION FUNCTIONS 5-27 MON2000

(a) To change the calculation function, click

the appropriate cell and use the pull- down menu to choose the desired operator.

The calculation completes each step from top to bottom.

To change or add a variable name, use the Variables pull-down menu.

To select a stream or component, click the appropriate cell and use the pull- down menu to choose the desired stream/ component.

Enter any constant values in the Value 1 and Value 2 cells.

(b) Click the button to accept your changes and return to the User- Defined Calculation dialog.

Click the button to abort and return to the User-Defined Calculation dialog.

5. Click the button to accept your new or edited calculation and return to the main window.

Click the button to abort and return to the main window.

5.6 LIMIT ALARMS

Emerson Process Management Chromatograph Software (MON2000) — 5.6 LIMIT ALARMS diagram (page 171)
Fig. 134

Use this function to set threshold limits for GC analysis data. When a limit is exceeded, an alarm is activated and logged. See Section 8.3 for information on Alarm Logs.

JULY 2010 Limit Alarms

Page 172 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-28 APPLICATION FUNCTIONS
Page 172 · click to enlarge
Page 172

5-28 APPLICATION FUNCTIONS

1. Use the Application > Limit Alarms menu to access this function.
2. The Limit Alarms dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5-28 APPLICATION FUNCTIONS diagram (page 172)
Fig. 135
3. To change the assigned variable, stream or component, click the appropriate cell and use the provided pull-down menu.

You can add a new variable only if an available row is blank. The number of allowed limit alarms is fixed by the GC application.

4. To assign the discrete hardware output that will be set when the alarm is active, click in the appropriate cell and use the provided pull-down menu.

Output values range from “1” to “N”, per the discrete hardware output identification number and the GC application. A discrete

Limit Alarms JULY 2010

Page 173 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-29 MON2000
Page 173 · click to enlarge
Page 173

APPLICATION FUNCTIONS 5-29 MON2000

output of “0 - Not used” indicates that no output is set.

A discrete output can be used to monitor one or more inputs.

If you are using discrete outputs to reflect the status of discrete inputs, ensure that the output assignments set here coordinate with those set in the Discrete Alarms dialog (see Section 5.7).

To set discrete outputs, see Section 5.12.

5. To change the alarm type, click the appropriate cell and use the provided pulldown menu.

All use all 4 limits (both Low and both High)

High use only High and High-High limits

Low use only Low and Low-Low limits

6. Input the desired high/low limits in the appropriate cells.
7. To copy the stream settings from a highlighted row and apply them to the next

An error message displays when the last available stream or component is reached.

row, click the button or press the F5 key. This feature increments the Stream value to the next available stream (e.g., incrementing from “Stream 2” to “Stream 8”), per the GC application.

JULY 2010 Limit Alarms

Type Purpose

Page 174 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Limit Alarms Description
Page 174 · click to enlarge
Page 174

5-30 APPLICATION FUNCTIONS

8. To copy the component settings from a highlighted row and apply them to the next

row, click the button or press the F6 key. This feature increments the Component value to the next available component (e.g., incrementing from “Ammonia” to “Benzene”), per the GC application.

9. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

Limit Alarms Description

  • Variable - selected variable assigned to the limit alarm
  • Stream - stream number assigned to the variable or the channel number if an analog input or analog output
  • Component - component name and number assigned to the variable
  • Discrete Output - number of the discrete hardware output that will be set when the alarm is active; zero indicates no discrete output will be set

Limit Alarms JULY 2010

Page 175 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.7 DISCRETE ALARMS
Page 175 · click to enlarge
Page 175

APPLICATION FUNCTIONS 5-31 MON2000

  • Type - High uses only the high and the highhigh limits, Low uses only the low and the low-low limits, and All uses both low and both high limits

Low-Low If the variable value falls below this limit, the low-low limit alarm is acti- vated. This alarm is for extreme situa- tions

Low If the variable value falls below this limit, the low limit alarm is activated.

High If the variable value rises above this limit, the high limit alarm is activated.

High-High If the variable value rises above this limit, the high-high limit alarm is acti- vated. This alarm is for extreme situa- tions.

5.7 DISCRETE ALARMS

Emerson Process Management Chromatograph Software (MON2000) — 5.7 DISCRETE ALARMS diagram (page 175)
Fig. 136

Use this function to assign discrete hardware inputs, used to set alarms, to one of the available discrete hardware outputs.

For more information on discrete outputs, see Section 5.12.

1. Use the Application > Discrete Alarms menu to access this function.

JULY 2010 Discrete Alarms

Type Purpose

Page 176 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-32 APPLICATION FUNCTIONS
Page 176 · click to enlarge
Page 176

5-32 APPLICATION FUNCTIONS

2. The Discrete Alarms dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5-32 APPLICATION FUNCTIONS diagram (page 176)
Fig. 137
3. Use the provided Discrete Input pull-down menu to select the identification number of the discrete hardware input to be used for alarm purposes.

Input values can range from “1” to “N”, per the discrete hardware output identification number and the GC application. A discrete input of “0 - Not used” indicates that no input has been assigned.

All alarms from digital inputs with assignments other than “0” are recorded in the Alarm Log.

To configure discrete inputs, see Section 5.11.

4. Use the provided Discrete Output pull-down menu to select the identification number of

Discrete Alarms JULY 2010

Page 177 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.7.1 Delayed Discrete Alarms
Page 177 · click to enlarge
Page 177

APPLICATION FUNCTIONS 5-33 MON2000

the discrete hardware output to be used for alarm purposes.

A discrete output can be used to monitor one or more inputs.

If you are using discrete outputs to reflect the status of GC analysis data alarms, ensure that the output assignments set here coordinate with those set in the Limit Alarms dialog (see Section 5.6).

Output values can range from “1” to “N”, per the discrete hardware output identification number and the GC application. A discrete output of “0 - Not used” indicates that no output has been assigned. To configure discrete outputs, see Section 5.12.

5.7.1 Delayed Discrete Alarms

A delayed discrete alarm can be set to trigger a discrete alarm and activate the appropriate discrete output if a previously specified discrete input has been closed for a user-defined amount of time.

JULY 2010 Delayed Discrete Alarms

Page 178 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-34 APPLICATION FUNCTIONS
Page 178 · click to enlarge
Page 178

5-34 APPLICATION FUNCTIONS

Emerson Process Management Chromatograph Software (MON2000) — 5-34 APPLICATION FUNCTIONS diagram (page 178)
Fig. 138

For example, if discrete input #1 is closed for 60 seconds, discrete alarm #1 will be activated, which will in turn activate discrete output #3. MON2000 will display an active alarm in the bottom row of the GC column of the GC Status Bar. MON2000 will also copy the alarm description in the Alarm Text column to the appropriate Alarm screens.

The delayed discrete alarm feature is only available with the following apps installed:

  • 700_Mode0_v192 or later;
  • 700_DewUS_v184 or later;
  • 700_Mode1_v193 or later;
  • 2350A_Mode0_v183 or later;
  • 2350A_Mode1_v184 or later.

Delayed Discrete Alarms JULY 2010

Page 179 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-35 MON2000
Page 179 · click to enlarge
Page 179

APPLICATION FUNCTIONS 5-35 MON2000

To set up a discrete alarm, do the following:

1. Select Applications > User Defined > Selection The User-Defined Selections window displays.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-35 MON2000 diagram (page 179)
Fig. 139
2. Locate the Delyd_alm variable under the Name column and set its Option to ON and click OK.
3. Select Applications > Discrete Alarms The Discrete Alarms window displays.

JULY 2010 Delayed Discrete Alarms

Page 180 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-36 APPLICATION FUNCTIONS
Page 180 · click to enlarge
Page 180

5-36 APPLICATION FUNCTIONS

4. For the appropriate alarm, select the discrete output that should be activated if the associated discrete input is closed.

By default, discrete outputs #1, #2, and #3 can be selected.

To make discrete output #4 available for selection, do the following:

(a) Select Applications > User Defined >

Selection.... The User-Defined Selections window displays. (b) Set EnableComAlarm to Disable. (c) Select Applications > Discrete Outputs....

The Discrete Outputs window displays. (d) Set the Switch field for discrete output

#4 to Auto.

To make discrete output #5 available for selection, do the following:

(a) Select Applications > User Defined >

Selection.... The User-Defined Selections window displays. (b) Set EnableCalMan to Disable. (c) Select Applications > Discrete Outputs....

The Discrete Outputs window displays. (d) Set the Switch field for discrete output

#5 to Auto.

All discrete inputs should be set to “0 - Not used”.

5. Enter the appropriate alarm text in the Alarm Text column, if desired. This text will be copied to the appropriate alarm screens if the discrete alarm is triggered.
6. Click OK to accept your changes.

Delayed Discrete Alarms JULY 2010

Page 181 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-37 MON2000
Page 181 · click to enlarge
Page 181

APPLICATION FUNCTIONS 5-37 MON2000

7. Select Applications > User Defined > Numeric The User-Defined Numerics window displays.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-37 MON2000 diagram (page 181)
Fig. 140
8. The following variables apply to the delayed discrete alarm:
  • delay_alm_da : Defines which discrete alarm will be activated.
  • delay_alm_di : Defines which discrete input will be used to initiate the delayed discrete alarm.

Although there are five discrete inputs available, closing Discrete Input #4 will result in an active Analyzer Failure alarm that will cause the GC to go to idle if Discrete Input #4 remains closed up to the end of the current analysis.

Discrete Input #5 can be used if the run_Aux_Seq option in the 'User-Defined Selections window is set to OFF.

  • delay_time : Defines the time, in seconds, between the moment the discrete input is closed and the delayed discrete alarm is activated. Range is 0 to 500 seconds. Default value is 60 seconds.

JULY 2010 Delayed Discrete Alarms

Page 182 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-38 APPLICATION FUNCTIONS
Page 182 · click to enlarge
Page 182

5-38 APPLICATION FUNCTIONS

5.8 STREAMS

This function allows you to:

Emerson Process Management Chromatograph Software (MON2000) — 5-38 APPLICATION FUNCTIONS diagram (page 182)
Fig. 141
  • assign CDTs and TEVs to a particular stream
  • designate a stream for analysis or calibration (or none)
  • control automatic calibration parameters, such as the total number of runs, runs to be averaged, starting times, and time between automatic calibrations and baseline runs
  • define baseline pressure and temperature conditions that are applicable to calculated GC analysis data, such as compressibility

Streams JULY 2010

Page 183 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-39 MON2000
Page 183 · click to enlarge
Page 183

APPLICATION FUNCTIONS 5-39 MON2000

To use the Streams function,

1. Use the Application > Streams menu to access this function.
2. The Streams dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-39 MON2000 diagram (page 183)
Fig. 142
3. To edit the settings displayed, click the appropriate cell, use the provided pull-down menus, or to choose the check boxes to enable or disable the automatic calibration and baseline calibration.

See the following tables for details.

JULY 2010 Streams

Page 184 of the Emerson Process Management Chromatograph Software (MON2000) user manual — TE2
Page 184 · click to enlarge
Page 184

5-40 APPLICATION FUNCTIONS

The standard settings include:

Table 5-4 Description of Standard Stream Settings

number label assigned by MON2000

This number correlates with the stream switch. A maximum of 12 streams can be con- trolled by MON2000 and a standard GC appli- cation.

Name text string label to identify a stream (12 character maximum)

Use how stream is used

Settings includecalibration (CAL); analysis (ANL); not used (NOT)

Det detector used by this stream

Control of the detector is application-specific. For example,

  • for Model 700/2350A, Modes 0 and 1, the Det col is “DET1” and read-only
  • for Model 700/2350A, Mode 2, the Det col is editable
  • for Model 700/2350A, Mode 3, the Det col is “DUAL” and read-only

CD

CDT associated with this stream (see Section 5.2)

CD2

CDT 2 For the Model 700/2350A Modes 2 and 3, an additional CD column displays.

TE

TEV associated with this stream (see Section 5.3)

TE2

TEV 2 For the Model 700/2350A Modes 2 and 3, an additional TE column displays.

Std Setting Description · Streams JULY 2010

Page 185 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-41 MON2000
Page 185 · click to enlarge
Page 185

APPLICATION FUNCTIONS 5-41 MON2000

For calibration (CAL) streams, you can edit the following control parameters:

Table 5-5 Description of Calibration Stream Settings

Tot number of runs made for each calibration

Range1 to 10

Avg number of last calibration runs to average

E.g., if 5 calibration runs are performed and Avg is set to “3”, then the last 3 runs are used to average the calibration results.

Start time the first automatic calibration is to be performed

Int number of hours between automatic calibrations

Cal Setting Description · JULY 2010 Streams

Page 186 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-42 APPLICATION FUNCTIONS
Page 186 · click to enlarge
Page 186

5-42 APPLICATION FUNCTIONS

Table 5-5 Description of Calibration Stream Settings (Continued)

Auto Cal Auto Base

enable/disable the automatic calibration and baseline run (first check box sets the automatic calibration; second check box sets the baseline run)

For the automatic calibration to work, at least one stream switch (valve event Strm Sw) must be included in the corresponding TEV (see Section 5.3).

If the automatic baseline run is enabled, the GC performs an additional calibra- tion run (before the calibration runs to be averaged) without the calibration gas. This run evaluates peaks caused by the GC valve action alone; any peak areas found are subtracted from the subse- quent analyses. To view or print the baseline run results, see Step 4.

Disabling the Auto Base setting will delete existing CDT baseline data for the associated stream.

You can edit the following parameters for a defined stream used in GC calculations. These parameters are also used for post- analysis calculations that use stream- specific base pressure and temperature values. Note that these parameters are not applicable for every application.

Cal Setting Description · Streams JULY 2010

Page 187 of the Emerson Process Management Chromatograph Software (MON2000) user manual — PSIA BAR
Page 187 · click to enlarge
Page 187

APPLICATION FUNCTIONS 5-43 MON2000

Table 5-6 Description of GC Calculation Parameters

PSIA BAR

base temperature in PSIA or bar units

To change the displayed system units, see Section 5.1 for details.

deg F deg C

base temperature in degrees Fahrenheit or Celsius

Opt Pres 1 Opt Pres 2 Opt Pres 3

If compressibility is activated for a par- ticular stream (Calculation Control function, (see Section 5.5.1), up to 3 optional pressures can be defined to calculate additional post-analysis data.

Only the base pressure can be edited. The base temperature is a constant 6 oF (15 oC) for all calculations.

4. To view or print the baseline run results,

click the button or press the F2 key. The Baseline Values dialog appears.

If a stream is not used, you cannot access baseline data for that stream.

Emerson Process Management Chromatograph Software (MON2000) — PSIA BAR diagram (page 187)
Fig. 143
5. Click the button to return to the main window.

Par Setting Description · JULY 2010 Streams

Page 188 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-44 APPLICATION FUNCTIONS
Page 188 · click to enlarge
Page 188

5-44 APPLICATION FUNCTIONS

5.9 ANALOG INPUTS

Emerson Process Management Chromatograph Software (MON2000) — 5-44 APPLICATION FUNCTIONS diagram (page 188)
Fig. 144

This function allows you to:

  • assign identifying labels
  • assign scale ranges
  • calibrate analog inputs for zero and full scale values

Electrical current signals ranging from 4 to 20mA ( %) are accepted as analog inputs.

To use the Analog Inputs function,

1. Use the Application > Analog Inputs menu to access this function.
2. The Analog Inputs dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5-44 APPLICATION FUNCTIONS diagram (page 188)
Fig. 145

To edit the settings displayed, click the appropriate cell or click and use the provided pull-down menus.

The 2350A GC Controller and the Model 700 unit supports up to four AI channels.

Analog Inputs JULY 2010

10 ±

Page 189 of the Emerson Process Management Chromatograph Software (MON2000) user manual — GC GRI PAZ
Page 189 · click to enlarge
Page 189

APPLICATION FUNCTIONS 5-45 MON2000

See the following table for details or refer to Step 3 for a more automated process.

Table 5-7 Description of Analog Input Settings

Channel channel number assigned

Number of available analog inputs is dependent upon the GC Controller model.

Label name used to identify the analog input (12 characters maximum)

Zero Scale value used as the zero scale when con- verting the analog input value

Full Scale value used as the full scale when convert- ing the analog input value

Fixed/Variable analog input operation mode

FXD = fixed VAR = variable

If the operation mode is fixed, the analog input is set to the number entered for the Fixed Value.

Fixed Value analog input value used during fixed operation

MON2000 accepts a Zero Scale calibration value that is higher than the Full Scale value.

Current Value current readout values of the analog inputs (values reflect scale assignments)

A higher Zero Scale value can be useful in applications that require an inverse display, or record of analog signal levels.

GC GRI PAZ

Gain Ratio Index (GRI) and Preamp Zero (PAZ) readouts from the GC preampli- fier’s 4 stages; use as a troubleshooting aid Normal preamp stage (gain) operating ranges for Models 500, 700 and 1000 GC units are shown in Table 5-8 below.

Zero Scale Calib Adjustment mA

calibrated analog inputs for known zero scale electrical current levels

JULY 2010 Analog Inputs

Setting Description

Page 190 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-46 APPLICATION FUNCTIONS
Page 190 · click to enlarge
Page 190

5-46 APPLICATION FUNCTIONS

Table 5-7 Description of Analog Input Settings (Continued)

Full Scale Calib Adjustment mA

calibrated analog inputs for known full scale electrical current levels

Table 5-8 Gain Operating Ranges for Models 500, 700 and 1000 GCs

GC GRI PAZ (12-bit AD)

1 0.0 to 0.0 600 to 800 0 to 4095 4800 to 6400 -32767 to 32767

2 0.8 to 1.1 600 to 800 0 to 4095 4800 to 6400 -32767 to 32767

3 0.8 to 1.1 600 to 800 0 to 4095 4800 to 6400 -32767 to 32767

4 0.8 to 1.1 1150 to 1500 0 to 4095 9200 to 12000 -32767 to 32767

3. To perform an automated input calibration,

(a) Select the desired channel by clicking

anywhere in that row. (b) Set the analog input signal to either its

minimum level or the anticipated Zero Value.

This step may require adjusting a rheostat, or another controlling device, at the source of the analog input signal.

(c) Set the anticipated Full Value.

(d) Click the button or press the F3 key.

You can abort this process at any time by

clicking the button.

Analog Inputs JULY 2010

Show 4 more labels / short text on this page
Setting Description · Value (12-bit AD) · PAZ (16-bit AD) · Value (16-bit AD)
Page 191 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-47 MON2000
Page 191 · click to enlarge
Page 191

APPLICATION FUNCTIONS 5-47 MON2000

4. The Zero Scale Adjustment dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-47 MON2000 diagram (page 191)
Fig. 146

Click the button to continue.

MON2000 accepts a Zero Scale calibration value that is higher than the Full Scale value. A higher Zero Scale value can be useful in the applications that require an inverse display, or record, of analog signal levels. 2350A GC Controllers and Model 700 units support up to 4 Analog channels.

5. The Full Scale Adjustment dialog appears. Adjust the Analog Input to its Full Scale value.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-47 MON2000 diagram (page 191)
Fig. 147

Click the button to continue.

6. MON2000 verifies that the recorded values are acceptable. A confirmation dialog displays.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-47 MON2000 diagram (page 191)
Fig. 148

Click the button to accept your changes.

Click the button to cancel and return to the Analog Input dialog.

JULY 2010 Analog Inputs

Page 192 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.10.1 Analog Output Dialog Description
Page 192 · click to enlarge
Page 192

5-48 APPLICATION FUNCTIONS

7. To perform a GRI (Gain Ratio Index)

calculation, click the button or press the F4 key.

Ensure that you halt any ongoing analysis runs (see Section 4.3) before performing a GRI calculation.

A GRI calculation confirms the GC Controller preamp operation. It can be performed after a preamp calibration has been completed, or as a trouble shooting procedure.

8. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

5.10 ANALOG OUTPUTS

Emerson Process Management Chromatograph Software (MON2000) — 5.10 ANALOG OUTPUTS diagram (page 192)
Fig. 149

This function allows you to:

  • assign variables to Analog Outputs
  • assign scale ranges
  • calibrate analog outputs for zero and full scale values

5.10.1 Analog Output Dialog Description

To use the Analog Outputs function,

1. Use the Application > Analog Outputs menu to access this function.

Analog Outputs JULY 2010

Page 193 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-49 MON2000
Page 193 · click to enlarge
Page 193

APPLICATION FUNCTIONS 5-49 MON2000

2. The Analog Outputs dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-49 MON2000 diagram (page 193)
Fig. 150

To edit the settings displayed, double-click the appropriate cell or click and use the provided pull-down menus.

When assigning analog outputs, first check the CGM Analog Output Config field in the System dialog (see Section 5.1).

An analog output assignment that uses the same analog output number as the CGM setting will produce erratic CGM trace output, possibly with incorrect scaling.

JULY 2010 Analog Output Dialog Description

Page 194 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-50 APPLICATION FUNCTIONS
Page 194 · click to enlarge
Page 194

5-50 APPLICATION FUNCTIONS

See the following table for details.

Table 5-9 Description of Analog Output Settings

number assigned to Analog Output

Number of available analog outputs is dependent upon the GC Controller model.

If the GC unit includes a Daniel Analog Expansion Module (AEM), P/N 1-0500- 001, reserve analog output number 1 (first row) for the variable Bargraph.

Variable type of GC analysis data on which to base signal level of analog output

Stream stream number assigned to the variable or the channel number if referencing an analog output

Component name of component monitored by this output

Current Value current readout values of the analog output (values reflect scale assignments)

Zero Scale value used to represent the minimum value (4mA) when scaling the analog output value

Full Scale value used to represent the maximum value (20mA) when scaling the analog output value

Fixed/Var analog output operation mode

Fixed = fixed Var = variable

If the operation mode is fixed, the analog output is set to the number entered for the Fixed Value.

Analog Output Dialog Description JULY 2010

Setting Description

Page 195 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-51 MON2000
Page 195 · click to enlarge
Page 195

APPLICATION FUNCTIONS 5-51 MON2000

Table 5-9 Description of Analog Output Settings (Continued)

Fixed Value analog output value used during fixed operation

Zero Adjustment value used to correct Zero Scale

Full Adjustment value used to correct Full Scale

When defining a new analog output, perform a calibration first to obtain accurate Zero and Full Adjustment values (see Section 5.10.4 or Section 5.10.5).

3. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

JULY 2010 Analog Output Dialog Description

Setting Description

Page 196 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.10.3 Changing the Bargraph
Page 196 · click to enlarge
Page 196

5-52 APPLICATION FUNCTIONS

5.10.2 Changing a Variable

Emerson Process Management Chromatograph Software (MON2000) — 5-52 APPLICATION FUNCTIONS diagram (page 196)
Fig. 151

To change a variable assignment, click the appropriate Variable cell. Use the provided pull-down menu and click the desired variable to select it.

5.10.3 Changing the Bargraph

Use this function to designate which AEM output analog signals can be used to drive the bargaph device inputs. Each of the 16 available AEM output signals can be assigned to represent various GC analysis data variables.

To edit the variables and corresponding settings displayed by the bargraph,

1. Click the button or press the

F5 key.

2. The Bargraph dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5.10.3 Changing the Bargraph diagram (page 196)
Fig. 152 · 5.10.3 Changing the Bargraph

Multiple streams can be entered by separating the stream numbers with commas (e.g., 2,3). Bargraph results will be produced for each stream.

To edit the settings displayed, double-click the appropriate cell or click and use the provided pull-down menus. See Section 5.10.1 for more details.

Changing a Variable JULY 2010

Page 197 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.10.4 Performing a Manual Calibration
Page 197 · click to enlarge
Page 197

APPLICATION FUNCTIONS 5-53 MON2000

3. To change a variable assignment, click the appropriate Variable cell. Use the provided pull-down menu and click the desired variable to select it.
4. Click the button to accept your changes and return to the Analog Outputs dialog.

Click the button to abort and return to the Analog Outputs dialog.

5.10.4 Performing a Manual Calibration

To manually calibrate an analog output,

1. Select the desired analog output by clicking anywhere in the corresponding row.
2. Set the Zero Scale and Full Scale values as desired.
3. Set the Fixed/Var parameter to “Fixed”.
4. Set Fixed Value equal to the Zero Scale value.
5. Set Zero Adjustment and Full Adjustment to “0.0”.
6. Click the button to accept your changes and exit from the Analog Outputs dialog.

Setting both adjustment values to “0.0” disables the scale adjustment.

See Appendix D for an alternative calibration method for analog outputs. Appendix D includes an example of calibrating an output scale of 0 to 1200.

7. Return to the Analog Outputs dialog. Wait until the Current Value of the analog output is equal to the Zero Scale value ( see Step 2 ).
8. Record the value, in engineering units, read by the receiving device (e.g., a voltmeter).
9. Set Fixed Value equal to the Full Scale value ( see Step 2 ).

JULY 2010 Performing a Manual Calibration

Page 198 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.10.5 Performing an Automated Calibration
Page 198 · click to enlarge
Page 198

5-54 APPLICATION FUNCTIONS

10.Click the button to accept your changes and exit from the Analog Outputs dialog.

11.Return to the Analog Outputs dialog. Wait until the Current Value of the analog output is equal to the Full Scale value (see Step 2).

12.Record the value, in engineering units, read by the receiving device (e.g., a voltmeter).

13.Set Zero Adjustment to the value recorded in Step 8.

14.Set Full Adjustment to the value recorded in Step 12.

15.Set the Fixed/Var parameter to “Var”.

16.Click the button to complete the calibration and return to the main window.

5.10.5 Performing an Automated Calibration

To perform an automated analog output calibration,

To determine the uncalibrated Zero Scale and Full Scale analog output levels, see Section 5.10.1.

1. Select the desired analog output by clicking anywhere in the corresponding row.
2. Click the button or press the

F8 key.

You can abort this process at any time by

clicking the button.

To prevent scale adjustment, set both adjustment values to zero (0.0).

3. The Zero Scale Adjustment dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5.10.5 Performing an Automated Calibration diagram (page 198)
Fig. 153 · 5.10.5 Performing an Automated Calibration

See Appendix D for additional instructions on calibrating analog outputs. Appendix D also includes a calibration example for an output scale of 0 to 1200.

Input the appropriate value and click the

button.

Performing an Automated Calibration JULY 2010

Page 199 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-55 MON2000
Page 199 · click to enlarge
Page 199

APPLICATION FUNCTIONS 5-55 MON2000

4. The Full Scale Adjustment dialog appears.

Input the appropriate value and click the

button.

5. If the values entered are within tolerance, data in the zero and full adjustment columns is updated. Otherwise, MON2000 displays an error message.

JULY 2010 Performing an Automated Calibration

Page 200 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-56 APPLICATION FUNCTIONS
Page 200 · click to enlarge
Page 200

5-56 APPLICATION FUNCTIONS

5.11 DISCRETE INPUTS

Use this function to assign labels to the GC discrete inputs and control their operational modes. The number of discrete inputs available depends on the GC application.

1. Use the Application > Discrete Inputs menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — 5-56 APPLICATION FUNCTIONS diagram (page 200)
Fig. 154
2. The Discrete Inputs dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5-56 APPLICATION FUNCTIONS diagram (page 200)
Fig. 155

To edit the settings displayed, double-click the appropriate cell or click and use the provided pull-down menus.

See the following table for details.

Table 5-10 Description of Discrete Input Settings

Channel channel number assigned

Label name used to identify the discrete input (12 characters maximum)

Switch sets operational mode for this discrete input

Auto = value is determined by application Off = value is set to OFF On = value is set to ON

Current Value current state of the discrete input

Discrete Inputs JULY 2010

Setting Description

Page 201 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.12 DISCRETE OUTPUTS
Page 201 · click to enlarge
Page 201

APPLICATION FUNCTIONS 5-57 MON2000

3. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

5.12 DISCRETE OUTPUTS

Use this function to assign labels to the GC discrete outputs and control their operational modes. The number of discrete outputs available depends on the GC application.

1. Use the Application > Discrete Outputs menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — 5.12 DISCRETE OUTPUTS diagram (page 201)
Fig. 156
2. The Discrete Outputs dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5.12 DISCRETE OUTPUTS diagram (page 201)
Fig. 157

To edit the settings displayed, double-click the appropriate cell or click and use the provided pull-down menus.

Signals routed to discrete outputs are assigned via the Limit Alarm and Discrete Alarm functions (see Section 5.6 and Section 5.7).

JULY 2010 Discrete Outputs

Page 202 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.13 VALVES
Page 202 · click to enlarge
Page 202

5-58 APPLICATION FUNCTIONS

See the following table for details.

Table 5-11 Description of Discrete Output Settings

Channel channel number assigned

Label name used to identify the discrete output (12 characters maximum)

Switch sets operational mode for this discrete output

Auto = value is determined by application Off = value is set to OFF On = value is set to ON

Current Value current state of the discrete output

On Time Off Time

time the digital output will be turned ON/OFF (MON2000 clock based on user PC system clock)

If the ON/OFF times are the same, the timing feature has no effect.

Emerson Process Management Chromatograph Software (MON2000) — 5-58 APPLICATION FUNCTIONS diagram (page 202)
Fig. 158
3. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

5.13 VALVES

This function allows you to:

  • assign identifying labels
  • monitor valve operation
  • control the operation modes

2350A GC Controllers support up to 12 Stream valves and 5 Analyzer valves, for a total of 17 valves.

The number of discrete outputs available depends on the GC application.

Model 700 controllers support up to Analytical Valves (AV-1 to AV-5 and S-1 to S-8 Stream valves

To use the Valve function,

1. Use the Application > Valves menu to access this function.

Setting Description · Valves JULY 2010

Page 203 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-59 MON2000
Page 203 · click to enlarge
Page 203

APPLICATION FUNCTIONS 5-59 MON2000

2. The Valves dialog appears, per GC type.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-59 MON2000 diagram (page 203)
Fig. 159
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-59 MON2000 diagram (page 203)
Fig. 160

Valves dialog for Model 700 GC Valves dialog for 2350A GCs

To edit the settings displayed, double-click the appropriate cell or click and use the provided pull-down menus.

JULY 2010 Valves

Page 204 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-60 APPLICATION FUNCTIONS
Page 204 · click to enlarge
Page 204

5-60 APPLICATION FUNCTIONS

See the following table for details.

Table 5-12 Description of Valve Settings

Channel channel number assigned

Label name used to identify the valve (12 characters maximum)

By default, the 2350A GC stream valves are shown first and are labeled S1 - S12 followed by the analytical valves (AV1 - AV5). Model 700 GCs analytical valves are shown first and are labeled AV1 - AV5 followed by the stream valves (S1 - S8).

Switch sets operational mode for this valve

Auto = valve operation is determined by the GC application Off = valve is set to OFF On = valve is set to ON

Usage (Model 700 and Model 500 GCs with 2350A Controllers)

sets which digital output from the Timed Events function (see Section 5.3) is assigned to this valve

Each Usage option is created per the following example: 1-AV1 - where the digit “1” is taken from the valve number set by the Valve/DO# option (see Timed Events, see Section 5.3), and the name “AV1” is taken from Label set in this Valve dialog.

The same Usage assignment cannot be used for multiple valves, unless these valves are set to “Unused”.

Current Value current state of the valve

3. Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

Setting Description · Valves JULY 2010

Page 205 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-61 MON2000
Page 205 · click to enlarge
Page 205

APPLICATION FUNCTIONS 5-61 MON2000

5.14 TEMPERATURE CONTROL

Use the Temperature Control function for monitoring the Temperature of the Oven (Detector/s and Columns) and the Stream Switching block to determine when the Model 700 is thermally stable. The bottom row labeled Temperature (C) displays the current temperatures.

The settings and values are shown in the following figure and table. The settings and values are preset at the factory and are based on the specific customer application. These values should not be changed unless recommended by Daniel Customer Service Personnel, or it is a factory application requirement.

1. Use the Application>Temperature Control menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-61 MON2000 diagram (page 205)
Fig. 161
2. The Temperature Control dialog displays.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-61 MON2000 diagram (page 205)
Fig. 162

JULY 2010 Temperature Control

Page 206 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.15 FID CONFIGURATION
Page 206 · click to enlarge
Page 206

5-62 APPLICATION FUNCTIONS

3. Use the pull down menu to select the appropriate mode setting (e.g. AUTOMATIC, MANUAL, or OUT OF SERVICE). Ensure that the temperature is constant for the Oven (i.e. Multivalve System block and column module kit) and the SSS.

Column Function Factory Default Setting

Column 1 Detector/s or Block Temperature 80o C (176.0o F)

Column 2 Column Temperature 80o C (176.0o F)

Column 3 Sample Stream Block Temperature 40o C (104.0o F)

Column 4 (Spare) Future Use

4. Click the OK button to apply the changes or click the Cancel button to discard your changes and return to the MON2000 Main window.

5.15 FID CONFIGURATION

Use this function to view the status information about the Model 700 Gas Chro- matograph’s Preamplifier Assembly, as well as to modify certain FID parameters.

Emerson Process Management Chromatograph Software (MON2000) — 5.15 FID CONFIGURATION diagram (page 206)
Fig. 163

To access the FID Configuration window, MON2000 must be connected to a GC that has its serial port #4 configured to use ‘ADC700’.

Emerson Process Management Chromatograph Software (MON2000) — 5.15 FID CONFIGURATION diagram (page 206)
Fig. 164

FID Configuration JULY 2010

Page 207 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-63 MON2000
Page 207 · click to enlarge
Page 207

APPLICATION FUNCTIONS 5-63 MON2000

1. Use the Application > FID Configuration menu to access this function. The FID Configuration window appears.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-63 MON2000 diagram (page 207)
Fig. 165

Information concerning the current status of the FID Preamplifier Assembly displays in the upper left corner.

2. To modify FID parameters, do the following:

(a) Halt the analysis.

JULY 2010 FID Configuration

Page 208 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-64 APPLICATION FUNCTIONS
Page 208 · click to enlarge
Page 208

5-64 APPLICATION FUNCTIONS

(b) Click Modify. The blue background of

the parameter fields will turn white, indicating that they are now modifiable.

Emerson Process Management Chromatograph Software (MON2000) — 5-64 APPLICATION FUNCTIONS diagram (page 208)
Fig. 166

The parameters that you can configure are:

  • FID ignition - Select Manual if you wish to control the ignition of the FID (default); select Auto to let the GC control the ignition of the FID.
  • Number of ignition attempts - Indicates the number of times the GC will try to light the flame. If an 'Auto' FID ignition sequence fails to light the flame after the specified number of attempts, the GC will close the hydrogen valve, switch the FID ignition parameter to Manual , and set an active alarm.
  • Wait time between tries.
  • Igniter on duration.
  • Flame On sense temperature.
  • Flame Out sense temperature.
  • Electrometer gain - Your choices are High and Low (default). (c) Click Save to accept the changes or

Cancel to reject them.

FID Configuration JULY 2010

Page 209 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-65 MON2000
Page 209 · click to enlarge
Page 209

APPLICATION FUNCTIONS 5-65 MON2000

3. If Flame status is OFF, due to a power failure or maintenance, for example, you can restart the flame by doing the following:

(a) Click H2 Open. The H2 valve status

parameter displayed in the FID PreAmp Status table changes to OPEN. (b) Click Light flame. The Flame status

parameter displayed in the FID PreAmp Status table will change to ON when the FID internal temperature exceeds the value set in the Flame On sense temperature field.

If FID ignition is set to Auto, the GC will automatically restart the flame if it goes out.

4. To enable or disable digital filtering, do the following:

(a) Right-click on the FID Configuration

window. The FID Advanced Configuration window appears.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-65 MON2000 diagram (page 209)
Fig. 167

JULY 2010 FID Configuration

Page 210 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.16 LOI STATUS VARIABLES
Page 210 · click to enlarge
Page 210

5-66 APPLICATION FUNCTIONS

(b) Click Modify to make the appropriate

selection from the Digital filtering drop- down list.

Emerson Process Management Chromatograph Software (MON2000) — 5-66 APPLICATION FUNCTIONS diagram (page 210)
Fig. 168

(c) Click Save to accept the change, or

Cancel to reject the change. (d) Click Close to leave the FID Advanced

Configuration window.

5. To reset the FID counts parameter, click Autozero FID .

5.16 LOI STATUS VARIABLES

Use this function to select and configure up to 25 GC parameters that you would like to monitor using the LOI’s Display mode (refer to the Model 700 Gas Chromatograph System Reference Manual for more information about this mode).

Emerson Process Management Chromatograph Software (MON2000) — 5.16 LOI STATUS VARIABLES diagram (page 210)
Fig. 169

LOI Status Variables JULY 2010

Page 211 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-67 MON2000
Page 211 · click to enlarge
Page 211

APPLICATION FUNCTIONS 5-67 MON2000

1. Use the Application > LOI Status Variables menu to access this function. The LOI Status Variables window appears.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-67 MON2000 diagram (page 211)
Fig. 170
2. If Include Mole Percents for current stream is checked, which is the default, the maximum number of variable you can select is five; if Include Mole Percents for current stream is not checked, you can choose up to 25 variables.
3. To select a new variable, click on the appropriate drop-down list under the Variable column.
4. Select a stream from the drop-down list in the Stream column. If you want to use the current stream, select ‘0’ in this column.

To copy the stream settings from a highlighted row and apply them to the next

row, click the button or press the F7 key. This feature also increments the Stream value to the next available stream (e.g., incrementing from Stream 2 to Stream 8), per the GC application.

5. Select a component from the drop-down list in the Component column. If you do not want to use a component, select ‘0’ in this column.

To copy the component settings from a highlighted row and apply them to the next

row, click the button or press

JULY 2010 LOI Status Variables

Page 212 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.17 SERIAL PORTS
Page 212 · click to enlarge
Page 212

5-68 APPLICATION FUNCTIONS

the F8 key. This feature also increments the Component value to the next available component (e.g., incrementing from “Ammonia” to “Benzene”), per the GC application.

6. Enter a value in the Precision column that indicates the number of decimal places to display for this particular variable. For component concentrations, the range of possible Precision values is between 2 and 6. For all other variables, the range of possible values is between 0 and 6.

5.17 SERIAL PORTS

Emerson Process Management Chromatograph Software (MON2000) — 5.17 SERIAL PORTS diagram (page 212)
Fig. 171

Use this function to configure and manipulate the serial port communication parameters the GC Controller uses to connect with MON2000.

1. Use the Application > Serial Ports menu to access this function.
2. The Serial Ports dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 5.17 SERIAL PORTS diagram (page 212)
Fig. 172

The Mode column applies only to the following CPU boards:

  • 2350A CPU board (P/N 2-3-2350-190)
  • Model 700 CPU board (P/N 2-3-0700-036)

Serial Ports JULY 2010

Page 213 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-69 MON2000
Page 213 · click to enlarge
Page 213

APPLICATION FUNCTIONS 5-69 MON2000

To edit the settings displayed, double-click the appropriate cell or click and use the provided pull-down menus.

For proper communications, these parameters, which are continuously downloaded to the GC, must be identical to the parameters set in the MON2000 GC Directory.

See the following table and individual sections for details.

Table 5-13 Description of Serial Port Settings

Port Serial port number for GC Controller.

The Model 700 and Model 500 with 2350A Controller (with the PC/104 COM 4A board installed) the controller supports 8 ports (9 ports with the modem).

With the Multifunction board installed, COM3 is the assigned port.

Usage Defines port usage – such as a slave config-uration (i.e.,User_Modbus, SIM_2251), PC communication port, a Report, or Serial Port – as well as the general type of serial communications protocol.

To accommodate installations that use Modbus registers and the SIM_2251 protocol along with a data acquisition system, MON2000 enables you to arrange data values for a more effective polling.

With the Multifunction board installed, COM3 is the assigned port.

JULY 2010 Serial Ports

Setting Description

Page 214 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-70 APPLICATION FUNCTIONS
Page 214 · click to enlarge
Page 214

5-70 APPLICATION FUNCTIONS

Table 5-13 Description of Serial Port Settings (Continued)

Baud Rate Baud rate setting.

Range: 1200, 2400, 4800, 9600, 19200, 38400.

For high performing PCs, set the baud rate to 19200. If you experience a com- munications failure at this rate, set the baud rate to 9600. Baud rate settings less than 9600 may result in real-time delivery that is unacceptably slow.

Data Bits Number of data bits.

Range: 7, 8. Default: 8.

Stop Bits Number of stop bits.

Range: 1, 2. Default: 1.

Parity Parity check method.

Range: None, Even, Odd. Default: None.

Handshake Handshaking method.

Range: None, RTS-CTS, XON-XOFF. Default: None.

RTS On Delay in milliseconds between RTS activation and the start of data transmis- sion .

Range: 0 to 5000.

RTS Off Delay in milliseconds between RTS termination and the end of data transmis- sion.

Range: 0 to 5000.

Protocol Transmission protocol.

RangeASCII, RTU Default: ASCII

Serial Ports JULY 2010

Setting Description

Page 215 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.17.1 Serial Ports Configuration
Page 215 · click to enlarge
Page 215

APPLICATION FUNCTIONS 5-71 MON2000

Table 5-13 Description of Serial Port Settings (Continued)

Com ID Modbus slave address.

To use the address defined by the GC Controller DIP switches, set to 0.

Range: R (Read only), W (Write only), RW (Read and write).

This parameter only functions with slave ports.

Mode Only applies to 2350A CPU board (P/N 2-3-2350-190) or Model 700 CPU board (P/N 2-3-0700-036).

Sets the communication protocol directly, instead of through jumper settings.

Range: RS232, RS422, RS485.

Click the button to accept your changes and return to the main window.

Click the button to abort and return to the main window.

5.17.1 Serial Ports Configuration

Each GC Controller port number is listed in the Port column of the Serial Ports dialog. The number of available ports is hardware-specific.

JULY 2010 Serial Ports Configuration

Setting Description · Read/Write Read/write setting

Page 216 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-72 APPLICATION FUNCTIONS
Page 216 · click to enlarge
Page 216

5-72 APPLICATION FUNCTIONS

These ports can be configured with the Usage settings to accommodate any one of the following end-connection devices:

The optional internal modem installed on a Model 700 or Model 500 uses COM9 only, except when a keyboard and display are installed. The optional internal modem would then use COM8 only.

For proper communications, these parameters, which are continuously downloaded to the GC, must be identical to the parameters set in the MON2000 GC Directory.

  • PC – direct serial line connection to the PC, a serial line connection to an external or internal modem or an ethernet connection
  • Report – direct serial line connection to a printer
  • User_Modbus or SIM_2251 – connection to a Data Collection System (DCS) or a multidrop serial data highway network

Specific configurations of these ports may also include settings for serial interface protocols (i.e., RS-232, RS-485, or RS-422).

These protocols can either be set from the Serial Ports window, or they can be configured through jumper settings that are located on the

Serial Ports Configuration JULY 2010

Page 217 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-73 MON2000
Page 217 · click to enlarge
Page 217

APPLICATION FUNCTIONS 5-73 MON2000

CPU board. See the appropriate GC hardware manual for instructions and drawings.

Communications between the GC and a DCS or multi-drop serial data highway can be established with Modbus protocol. For this, the Usage setting should be either User_Modbus or SIM_2251.

In any Modbus host-slave configuration, there must be one Host to which any one of the GCs can respond as a Slave. Thus, the controller could be connected to a maximum of 4 different serial data networks.

The serial interface protocol RS-232 will service most end-connection devices (i.e., PC and Report). DCS connections use either RS-485 or RS-422 protocols although, in some cases, the RS-232 protocol can be used.

To access this function,

1. From the Application>Serial Ports menu, select the Usage column and use the pulldown menu to select the appropriate configuration.

JULY 2010 Serial Ports Configuration

Page 218 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 1 (COM1) P2 J5
Page 218 · click to enlarge
Page 218

5-74 APPLICATION FUNCTIONS

The Mode column applies only to the following CPU boards:

  • 2350A CPU board (P/N 2-3-2350-190)
  • Model 700 CPU board (P/N 2-3-0700-036)

5.18 GC SERIAL PORT AND CABLE CONFIGURATIONS

This section provides more detailed information about the serial port connections of the 2350A GC Controller. It identifies serial port pin assignments and diagrams for designing RS-232 serial cables necessary for your application.

GC serial ports are found on the GC Controller Terminal Board for field wiring, and the connection points for external devices are as follows:

Table 5-14 Serial Ports on GC Terminal Board

Port DB-9 Plug Phoenix Plug

1 (COM1) P2 J5

2 (COM2) P3 J6

3 (COM3) n/a J10

4 (COM4) n/a J11

External modem connections can be made to any of the four serial ports. However, a possible exception exists for Serial Port 4: the internal modem for the Model 500 with the 2350A Controller, if installed, uses Serial Port 4.

GC Serial Port and Cable Configurations JULY 2010

Page 219 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-75 MON2000
Page 219 · click to enlarge
Page 219

APPLICATION FUNCTIONS 5-75 MON2000

Phoenix plug (bare-wire) connections are available to all four serial ports. Pin-outs are identical for all four serial port Phoenix plugs and jacks. Each Phoenix plug / jack (male) combination allows bare-wire connection and uses 9 pins as illustrated:

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-75 MON2000 diagram (page 219)
Fig. 173

Figure 5-1 Phoenix Connector Pinout for J5, J6, J10, J11

DB-9 plug connections available for serial ports 1 and 2 only. Two of the serial ports, as noted in Table 5-14, permit connection to an external modem via a DB-9 plug.

JULY 2010 GC Serial Port and Cable Configurations

Page 220 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.18.1 GC DB-9 Serial to External Modem DB-25
Page 220 · click to enlarge
Page 220

5-76 APPLICATION FUNCTIONS

Both of the GC Controller DB-9 jacks are female and have identical pin assignments.

Emerson Process Management Chromatograph Software (MON2000) — 5-76 APPLICATION FUNCTIONS diagram (page 220)
Fig. 174

A DB-9 male pin numbering scheme is also illustrated, but for reference purposes only.

Figure 5-2 DB-9 Connector and Pinout, P2 and P3

5.18.1 GC DB-9 Serial to External Modem DB-25

To make an RS-232 serial connection between one of the DB-9 serial ports of the GC, and an external modem with DB-25 serial port, you will need to manufacture a cable. The cable will need DB-9, male, and DB-25, male, plug cable ends as illustrated in Figure 5-3.

GC serial ports were wired to appear as DCE so that a straight-through serial cable could be used, instead of a null-modem cable, for direct serial connection between the GC Controller and the PC, which is Data Terminal Equipment (DTE).

The DB-9 jack on the GC serial port is wired to appear like a Data Communications Equipment (DCE). Therefore, you use a custom

GC DB-9 Serial to External Modem DB-25 JULY 2010

Page 221 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.18.2 GC Phoenix Plug to External Modem DB-25
Page 221 · click to enlarge
Page 221

APPLICATION FUNCTIONS 5-77 MON2000

“null-modem” type cable, as shown below, to make the connection between the GC and an external modem.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-77 MON2000 diagram (page 221)
Fig. 175

Figure 5-3 GC DB-9 Port to External Modem DB-25 Port

5.18.2 GC Phoenix Plug to External Modem DB-25

To make an RS-232 serial connection between one of the Phoenix Plug serial ports of the GC, and an external modem with DB-25 serial port, you will need to manufacture the cable and its DB-25, male plug cable end as illustrated in Figure 5-4.

JULY 2010 GC Phoenix Plug to External Modem DB-25

Page 222 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.18.3 Com ID
Page 222 · click to enlarge
Page 222

5-78 APPLICATION FUNCTIONS

Emerson Process Management Chromatograph Software (MON2000) — 5-78 APPLICATION FUNCTIONS diagram (page 222)
Fig. 176

Figure 5-4 Phoenix Plug Port to External Modem DB-25 Port

5.18.3 Com ID

The GC Controller Com ID is usually preset at the factory and defined by the DIP switch settings on the controller system interface board. Five switches are reserved for defining the Com ID; therefore, Com ID values range from 1 to 31.

Normally it is not necessary to change the Com ID value to anything other than 0. Setting the Com ID to zero allows the MON2000 software to automatically recognize the Com ID defined by the controller DIP switches. When connected, MON2000 displays the controller DIP setting as the Com ID in the main window titlebar. See Table 5-15 for more information.

Com ID JULY 2010

Page 223 of the Emerson Process Management Chromatograph Software (MON2000) user manual — R RW
Page 223 · click to enlarge
Page 223

APPLICATION FUNCTIONS 5-79 MON2000

Table 5-15 Description of Com ID Settings

Connection Usage Read/Write Protocol

PC or modem (direct serial connection)

R RW

ASCII RTU

R RW

ASCII RTU

Printer (direct serial connection)

1. User_Modbus and SIM_2251 are slave port protocols (the GC Controller can operate only as a slave). 2. User_Modbus, a PLC emulation Modbus protocol, can use scaling to convert floating point numbers to integers. 3. SIM_2251 emulates Daniel 2500 communication protocol and is a simulation of the 2251 GC Controller. 4. See Appendix I for instructions on how to connect a modem to the GC Controller.

JULY 2010 Com ID

PC User_Modbus SIM_2251 · DCS User_Modbus SIM_2251 · Report W ASCII

Page 224 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-80 APPLICATION FUNCTIONS
Page 224 · click to enlarge
Page 224

5-80 APPLICATION FUNCTIONS

5.18.4 Registers

To accommodate installations that use Modbus registers and the User_Modbus protocol, MON2000 enables you to assign GC analysis variables to registers.

This function is not required for normal GC operation. Only use this function if you are designing a special installation to directly access the GC Controller Modbus registers.

See Appendix F for more information regarding GC Modbus registers.

For a list of variable assignments made to all registers, print the communication section of the PC Config Report.

To assign GC analysis variables and scale ranges,

Emerson Process Management Chromatograph Software (MON2000) — 5-80 APPLICATION FUNCTIONS diagram (page 224)
Fig. 177
1. Use the Application>Serial Ports menu to access this function.
2. The Serial Ports dialog displays.

Registers JULY 2010

Page 225 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-81 MON2000
Page 225 · click to enlarge
Page 225

APPLICATION FUNCTIONS 5-81 MON2000

3. From the Serial Ports dialog, click the button or press the F3 key.
4. The Registers dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-81 MON2000 diagram (page 225)
Fig. 178
5. View or edit data. Double-click a given table cell or use one of the edit function buttons, described below, to make your changes.

For each default 0 FloatPt scale assigned, reserve 2 Modbus registers for a given variable.

Scale assignments other than 0 FloatPT require only one register per variable.

6. To change a variable assignment, click the appropriate Variable cell. Use the provided pull-down menu and click the desired variable to select it.

From the Variables menu, you can assign the results of optional base pressure(s) to the Modbus register. See Section 5.18.5.

JULY 2010 Registers

Page 226 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-82 APPLICATION FUNCTIONS
Page 226 · click to enlarge
Page 226

5-82 APPLICATION FUNCTIONS

7. To change the stream assignment, click the appropriate cell. Use the provided pulldown menu and click the desired stream to select it.

To copy the stream settings from a highlighted row and apply them to the next

row, click the button or press the F7 key. This feature also increments the Stream value to the next available stream (e.g., incrementing from Stream 2 to Stream 8), per the GC application. An error message displays when the last available stream is reached.

8. To change the component assignment, click the appropriate cell. Use the provided pulldown menu and click the desired option to select it.

To copy the component settings from a highlighted row and apply them to the next

row, click the button or press the F8 key.

This feature also increments the Component value to the next available component (e.g., incrementing from Ammonia to Benzene), per the GC application. An error message displays when the last available component is reached.

9. To select a scale and range,

Assigning a scale and corresponding range is an optional step.

(a) Select the scale/range to be replaced.

By assigning a scale range, a floating point value can be converted to an integer value per the user-defined scale assignment.

The default scale assignment is 0 FloatPt, which means the value is not converted to integer and is stored in two (2) adjacent Modbus registers.

Registers JULY 2010

Page 227 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-83 MON2000
Page 227 · click to enlarge
Page 227

APPLICATION FUNCTIONS 5-83 MON2000

(b) Click the button or press the F5 key. A menu list of all available scales and ranges appears. (c) Double-click the desired new scale and

corresponding range.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-83 MON2000 diagram (page 227)
Fig. 179

10.To redefine a scale or create a custom scale,

(a) Click the button or press the F2 key. (b) The Edit Scales dialog appears.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-83 MON2000 diagram (page 227)
Fig. 180

All 33 scaling options will display.

JULY 2010 Registers

Page 228 of the Emerson Process Management Chromatograph Software (MON2000) user manual — R F R Z –
Page 228 · click to enlarge
Page 228

5-84 APPLICATION FUNCTIONS

(c) Input the new Zero Scale or Full Scale

values as appropriate.

To convert floating point data to a whole integer, each scale definition follows this algorithm:

R F R Z –

SF SZ – ⎝ ⎠ ⎛ ⎞ = Dfp 3977 =

RF = range, Full Scale =4095 Rz = range, Zero Scale =0 SF = scale, Full =100 Sz = scale, Zero =0 Dfp = data, floating pt =97.13 (% from Methane)

(d) Click the button to accept your changes and return to the Registers dialog.

Click the button to abort and return to the Registers dialog.

integer · where · Registers JULY 2010

Page 229 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-85 MON2000
Page 229 · click to enlarge
Page 229

APPLICATION FUNCTIONS 5-85 MON2000

11.To check for conflicting Modbus register assignments,

(a) Click the button or press the

F6 key. (b) MON2000 displays the first encountered

error.

Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-85 MON2000 diagram (page 229)
Fig. 181

For this example, you would change the register number for Row 2. You would then verify all subsequent register numbers as well.

(c) Click the button to accept your changes and return to the Serial Ports dialog. MON2000 automatically checks for conflicting Modbus register assignments.

Click the button to abort and return to the Serial Ports dialog.

JULY 2010 Registers

Page 230 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-86 APPLICATION FUNCTIONS
Page 230 · click to enlarge
Page 230

5-86 APPLICATION FUNCTIONS

5.18.5 Setting Optional Base Pressures

As discussed in Section 5.18.4, you can assign GC analysis variables to various Modbus registers. Follow the procedure below to add the “Optional Base Pressure” variable to the Registers List.

Up to three (3) optional base pressures can be used for GC analysis calculations. These optional base pressure values are entered via the Streams function (see Section 5.8).

  • “Results - Opt. Base Press” – A group of 12 GC analysis calculations can be used to reflect the use of different base pressures. See Section C.3 for more information.

To assign the “Optional Base Pressure” variable,

1. Verify the values for optional base pressures that have already been entered.
Emerson Process Management Chromatograph Software (MON2000) — 5-86 APPLICATION FUNCTIONS diagram (page 230)
Fig. 182

(a) Use the Application > Streams menu to

access this function. (b) The Streams dialog box displays.

Emerson Process Management Chromatograph Software (MON2000) — 5-86 APPLICATION FUNCTIONS diagram (page 230)
Fig. 183

Setting Optional Base Pressures JULY 2010

Page 231 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-87 MON2000
Page 231 · click to enlarge
Page 231

APPLICATION FUNCTIONS 5-87 MON2000

(c) Scroll to the Opt Press columns and note

these values along with the corresponding optional pressure and stream number(s). (d) If you wish to enter different pressure

values, do so now and then click the

button to accept your changes.

Otherwise, click the button to return to the main window.

2. Use the Application > Serial Ports menu to access the Register function.
Emerson Process Management Chromatograph Software (MON2000) — APPLICATION FUNCTIONS 5-87 MON2000 diagram (page 231)
Fig. 184

JULY 2010 Setting Optional Base Pressures

Page 232 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-88 APPLICATION FUNCTIONS
Page 232 · click to enlarge
Page 232

5-88 APPLICATION FUNCTIONS

3. From the Serial Ports dialog, click the

button or press the F3 key. The Registers dialog appears.

4. From the Registers dialog, use the Variable pull-down menu to select the Optional Base Pressures option.
5. Use the corresponding component pulldown menu to select the desired optional base pressure option.
Emerson Process Management Chromatograph Software (MON2000) — 5-88 APPLICATION FUNCTIONS diagram (page 232)
Fig. 185
Emerson Process Management Chromatograph Software (MON2000) — 5-88 APPLICATION FUNCTIONS diagram (page 232)
Fig. 186

Setting Optional Base Pressures JULY 2010

Page 233 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-89 MON2000
Page 233 · click to enlarge
Page 233

APPLICATION FUNCTIONS 5-89 MON2000

6. To assign a scale, ensure that the correct

row is selected and click the button or press the F5 key. Double-click the desired scale option to apply.

For instructions on creating or editing a scale, see Section 5.18.4.

7. The “Optional Base Pressure” variable should be listed with the settings specified.
8. Click the button to accept your changes.

JULY 2010 Setting Optional Base Pressures

Page 234 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-90 APPLICATION FUNCTIONS
Page 234 · click to enlarge
Page 234

5-90 APPLICATION FUNCTIONS

5.18.6 Setting Optional Base Results

As discussed in Section 5.18.4, you can assign GC analysis variables to various Modbus registers. Follow the procedure below to add the “Results - Opt. Base Press” variable to the Registers List.

A group of 12 GC analysis calculations can be used to reflect the use of different base pressures. See Section C.3 for more information.

To assign the “Results - Opt. Base Press” variable,

1. Verify the values for optional base pressures that have already been entered.

(a) Use the Application > Streams menu to

access this function. (b) The Streams dialog displays.

Emerson Process Management Chromatograph Software (MON2000) — 5-90 APPLICATION FUNCTIONS diagram (page 234)
Fig. 187

(c) Scroll to the Opt Press columns and note

these values along with the corresponding optional pressure and stream number(s).

Setting Optional Base Results JULY 2010

Page 235 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-91 MON2000
Page 235 · click to enlarge
Page 235

APPLICATION FUNCTIONS 5-91 MON2000

(d) If you wish to enter different pressure

values, do so now and then click the

button to accept your changes.

Otherwise, click the button to return to the main window.

2. Use the Application > Serial Ports menu to access this function.
3. From the Serial Ports dialog, click the

button or press the F3 key. The Registers dialog appears.

JULY 2010 Setting Optional Base Results

Page 236 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-92 APPLICATION FUNCTIONS
Page 236 · click to enlarge
Page 236

5-92 APPLICATION FUNCTIONS

4. From the Registers dialog, use the Variable pull-down menu to select the Results - Opt. Base Press option.
5. Use the corresponding component pulldown menu to select the desired optional base pressure option.

Metric GC analysis variables are available only if the Metric Base Conditions option has been enabled (see Section 5.1).

Emerson Process Management Chromatograph Software (MON2000) — 5-92 APPLICATION FUNCTIONS diagram (page 236)
Fig. 188

See Table 5-16 to determine which number you should enter.

6. To assign a scale, ensure that the correct

row is selected and click the button or press the F5 key. Double-click the desired scale option to apply.

For instructions on creating or editing a scale, see Section 5.18.4.

7. The “Results - Opt. Base Press” variable should be listed with the settings specified.
8. Click the button to accept your changes.

Setting Optional Base Results JULY 2010

Page 237 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-93 MON2000
Page 237 · click to enlarge
Page 237

APPLICATION FUNCTIONS 5-93 MON2000

. Table 5-16 GC Analysis Variable Number per Optional Pressure

GC Analysis Variable Optional Pressure

U.S. Metric 1 2 3

Gross Heating Value or Btu (Dry)

Heating Value, Superior, MJ/m3 (Dry)

Gross Heating Value or Btu (Sat.)

Heating Value, Superior, MJ/m3 (Sat.)

Gross Heating Value or Btu (Act.)

Heating Value, Superior, MJ/m3 (Act.)

Net Btu (Dry) Heating Value, Inferior, MJ/m3 (Dry)

Net Btu (Sat.) Heating Value, Inferior, MJ/m3 (Sat.)

Net Btu (Act.) Heating Value, Inferior, MJ/m3 (Act.)

Real Relative Density Gas

Heating Value, Superior, kcal/m3 (Dry)

Gallons/1000 SCF C2 + Heating Value, Superior, kcal/m3 (Sat.)

Gallons/1000 SCF C3 + Heating Value, Superior, kcal/m3 (Act.)

Gallons/1000 SCF C4 + Heating Value, Inferior, kcal/m3 (Dry)

Gallons/1000 SCF C5 + Heating Value, Inferior, kcal/m3 (Sat.)

Gallons/1000 SCF C6 + Heating Value, Inferior, kcal/m3 (Act.)

JULY 2010 Setting Optional Base Results

Show 12 more labels / short text on this page
1 13 25 · 2 14 26 · 3 15 27 · 4 16 28 · 5 17 29 · 6 18 30 · 7 19 31 · 8 20 32 · 9 21 33 · 10 22 34 · 11 23 35 · 12 24 36
Page 238 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5.19 TCP/IP, SUBNET, AND GATEWAY MENU
Page 238 · click to enlarge
Page 238

5-94 APPLICATION FUNCTIONS

Table 5-16 GC Analysis Variable Number per Optional Pressure (Continued)

GC Analysis Variable Optional Pressure

U.S. Metric 1 2 3

Use the above index numbers to assign “Results - Opt. Base Press” to User_Modbus registers.

5.19 TCP/IP, SUBNET, AND GATEWAY MENU

Use this function to configure the TCP/IP, Subnet, and Gateway communication parameters for the 2350A GC Controller.

1. Use the Application > TCP/IP menu to access this function.
2. The TCP/IP dialog displays. See Appendix F for detailed instructions for the Ethernet card installation, TCP/IP, Subnet, and Gateway settings.

TCP/IP, Subnet, and Gateway Menu JULY 2010

Page 239 of the Emerson Process Management Chromatograph Software (MON2000) user manual — APPLICATION FUNCTIONS 5-95 MON2000
Page 239 · click to enlarge
Page 239

APPLICATION FUNCTIONS 5-95 MON2000

This page is intentionally left blank.

JULY 2010 TCP/IP, Subnet, and Gateway Menu

Page 240 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 5-96 APPLICATION FUNCTIONS
Page 240 · click to enlarge
Page 240

5-96 APPLICATION FUNCTIONS

TCP/IP, Subnet, and Gateway Menu JULY 2010

Page 241 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.1 CHROMATOGRAM VIEWER FUNCTION
Page 241 · click to enlarge
Page 241

CHROMATOGRAM VIEWER 6-1 MON2000

6 CHROMATOGRAM VIEWER

Use the Chromatogram Viewer to display and print live, archived, or saved chromatograms (from a .cgm file on disk). The 2350A GC Controller saves the most recent chromatograms for each stream running in the Stream Sequence, as well as the chromatogram from the most recent calibration run for each CDT. See the following sections for further instructions.

6.1 CHROMATOGRAM VIEWER FUNCTION

To access the Chromatogram Viewer,

1. Connect to the GC.
2. Use the Chromatogram > Chromatogram Viewer menu to access this function.
3. The Chromatogram Viewer dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 6.1 CHROMATOGRAM VIEWER FUNCTION diagram (page 241)
Fig. 189

JULY 2010 Chromatogram Viewer Function

Page 242 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.2 VIEWING A LIVE CHROMATOGRAM
Page 242 · click to enlarge
Page 242

6-2 CHROMATOGRAM VIEWER

4. Click the button to exit this dialog. Any display changes or open data will be lost. See Section 6.4 to save a chromatogram to disk.

6.2 VIEWING A LIVE CHROMATOGRAM

6.2.1 From Online GC

To view a live chromatogram trace from the online GC,

Emerson Process Management Chromatograph Software (MON2000) — 6.2 VIEWING A LIVE CHROMATOGRAM diagram (page 242)
Fig. 190
1. Check the View current CGM option.

If a chromatogram contains data for more than one detector, a Select Detector menu dialog displays. Double-click either “Detector 1”, “Detector 2”, or “Both”.

2. MON2000 plots the current, live chromatogram.

The corresponding data is dynamically displayed in the provided legends.

Emerson Process Management Chromatograph Software (MON2000) — 6.2 VIEWING A LIVE CHROMATOGRAM diagram (page 242)
Fig. 191

Use the Graph and Chromatogram functions to manipulate the display. To zoom in on a particular point, click and drag your mouse over that area; repeat as necessary. Use the Restore State or Zoom Out options from the right-click menu (see Section 6.9) to return to the original display.

You can display up to 8 chromatograms at one time. Each chromatogram is color-

Viewing a Live Chromatogram JULY 2010

Page 243 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.2.2 From GC Archive
Page 243 · click to enlarge
Page 243

CHROMATOGRAM VIEWER 6-3 MON2000

coded; use the Chromatogram pull-down menu to select a specific chromatogram.

Emerson Process Management Chromatograph Software (MON2000) — CHROMATOGRAM VIEWER 6-3 MON2000 diagram (page 243)
Fig. 192

6.2.2 From GC Archive

To view the most recent chromatogram trace from the online GC archive,

1. Click the button. The Available Chromatograms dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 6.2.2 From GC Archive diagram (page 243)
Fig. 193 · 6.2.2 From GC Archive
2. Double-click the desired entry.

If a chromatogram contains data for more than one detector, a Select Detector menu dialog displays. Double-click either “Detector 1”, “Detector 2”, or “Both”.

3. MON2000 plots the archived chromatogram.

The corresponding data is dynamically displayed in the provided legends.

Use the Graph and Chromatogram functions to manipulate the display. To

JULY 2010 From GC Archive

Page 244 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.2.3 From File on Disk
Page 244 · click to enlarge
Page 244

6-4 CHROMATOGRAM VIEWER

zoom in on a particular point, click and drag your mouse over that area; repeat as necessary. Use the Restore State or Zoom Out options from the right-click menu (see Section 6.9) to return to the original display.

You can display up to 8 chromatograms at one time. Each chromatogram is color- coded; use the Chromatogram pull-down menu to select a specific chromatogram.

Emerson Process Management Chromatograph Software (MON2000) — 6-4 CHROMATOGRAM VIEWER diagram (page 244)
Fig. 194

6.2.3 From File on Disk

To view a chromatogram trace from a file previously saved to disk,

1. Click the button. The Open Chromatogram File dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 6.2.3 From File on Disk diagram (page 244)
Fig. 195 · 6.2.3 From File on Disk
2. Use the provided directory tree to locate the desired CGM file or CGM Comparison file and click the button to open.

From File on Disk JULY 2010

Page 245 of the Emerson Process Management Chromatograph Software (MON2000) user manual — CHROMATOGRAM VIEWER 6-5 MON2000
Page 245 · click to enlarge
Page 245

CHROMATOGRAM VIEWER 6-5 MON2000

3. The Chromatogram window displays.
Emerson Process Management Chromatograph Software (MON2000) — CHROMATOGRAM VIEWER 6-5 MON2000 diagram (page 245)
Fig. 196

If a chromatogram contains data for more than one detector, a Select Detector menu dialog displays. Double-click either “Detector 1”, “Detector 2”, or “Both”.

JULY 2010 From File on Disk

Page 246 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6-6 CHROMATOGRAM VIEWER
Page 246 · click to enlarge
Page 246

6-6 CHROMATOGRAM VIEWER

4. MON2000 displays the chromatogram file.

The corresponding data is dynamically displayed in the provided legends.

Use the Graph and Chromatogram functions to manipulate the display. To zoom in on a particular point, click and drag your mouse over that area; repeat as necessary. Use the Save State, Restore State, or Zoom In or Out options from the right-click menu (see Section 6.9) to return to the original display for both Live and Archived CGM views. Press the keyboard HOME key to force the CMG view to the default state. Also, pressing the keyboard HOME key sets the CGM view to the default state if the current application analysis time is different from the previous application analysis time.

The Save State function is available when viewing a live or archived CGM.

Emerson Process Management Chromatograph Software (MON2000) — 6-6 CHROMATOGRAM VIEWER diagram (page 246)
Fig. 197

The MON2000 software looks for and finds the application’s analysis time. Pressing the ”Home” key returns the user to the default CGM state.

Emerson Process Management Chromatograph Software (MON2000) — 6-6 CHROMATOGRAM VIEWER diagram (page 246)
Fig. 198

From File on Disk JULY 2010

Right Click Menu

Page 247 of the Emerson Process Management Chromatograph Software (MON2000) user manual — CHROMATOGRAM VIEWER 6-7 MON2000
Page 247 · click to enlarge
Page 247

CHROMATOGRAM VIEWER 6-7 MON2000

You can display up to 8 chromatograms at one time. Each chromatogram is color- coded; use the Chromatogram pull-down menu to select a specific chromatogram

5. Click the button to save the

selected comparison file to disk.

The Save Chromatogram Comparison File dialog appears.

Emerson Process Management Chromatograph Software (MON2000) — CHROMATOGRAM VIEWER 6-7 MON2000 diagram (page 247)
Fig. 199
6. Enter the file name to be saved and click the

button to save the file to disk.

Click the button to exit this dialog.

JULY 2010 From File on Disk

Or,

Page 248 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.4 SAVING A CHROMATOGRAM TO DISK
Page 248 · click to enlarge
Page 248

6-8 CHROMATOGRAM VIEWER

6.3 REMOVING A CHROMATOGRAM FROM VIEW

To remove a chromatogram trace from the Viewer display (and close the cgm file, if reading from disk),

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
Emerson Process Management Chromatograph Software (MON2000) — 6-8 CHROMATOGRAM VIEWER diagram (page 248)
Fig. 200
2. Click the button. MON2000 removes the chromatogram from view and, if applicable, closes the .cgm file.

6.4 SAVING A CHROMATOGRAM TO DISK

To save a currently displayed chromatogram to disk,

This function will save a live chromatogram at the current point, thus overwriting the “last.cgm” file.

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button.

Removing a Chromatogram from View JULY 2010

Page 249 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.5.1 Editing the Chromatogram Graph
Page 249 · click to enlarge
Page 249

CHROMATOGRAM VIEWER 6-9 MON2000

MON2000 displays the Save Chromatogram File window.

Emerson Process Management Chromatograph Software (MON2000) — CHROMATOGRAM VIEWER 6-9 MON2000 diagram (page 249)
Fig. 201

Use the provided directory tree to select the desired file location and name, then click

the button.

3. MON2000 saves the specified chromatogram along with its corresponding graph settings.

6.5 GRAPH FUNCTIONS

Use the Graph buttons to access any of the following graph-related functions.

6.5.1 Editing the Chromatogram Graph

Use the Edit Graph function to change the span and interval for the x/y axis, as well as other display parameters.

1. Use the Chromatogram pull-down menu to select a specific chromatogram.

JULY 2010 Graph Functions

Page 250 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.5.2 Changing Cursor Size
Page 250 · click to enlarge
Page 250

6-10 CHROMATOGRAM VIEWER

2. Click the button. The Edit Graph dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 6-10 CHROMATOGRAM VIEWER diagram (page 250)
Fig. 202
3. Use the provided attributes and options to change the chromatogram display and printer speed as desired.
4. Click the button to accept your changes and return to the Chromatogram Viewer dialog.

Click the button to abort and return to the Chromatogram Viewer dialog.

5. The chromatogram display changes as specified.

Additional display options are available by right-clicking the display area. See Section 6.9 for more information.

6.5.2 Changing Cursor Size

To toggle the cursor size from coarse movement (less accurate) to fine movement (more accurate), click the button.

Changing Cursor Size JULY 2010

Page 251 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.6 CHROMATOGRAM FUNCTIONS
Page 251 · click to enlarge
Page 251

CHROMATOGRAM VIEWER 6-11 MON2000

6.5.3 Printing the Chromatogram

To print a currently displayed area of the chromatogram,

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button.

6.6 CHROMATOGRAM FUNCTIONS

Use the Chromatogram buttons to access any of the following chromatogram-related functions.

Emerson Process Management Chromatograph Software (MON2000) — 6.6 CHROMATOGRAM FUNCTIONS diagram (page 251)
Fig. 203
  • Edit - edit the Chromatogram
  • Results - view the CGM calculation results
  • Desc - save a description of the CGM
  • Save - save the displayed chromatogram to disk
  • Remove - removes a chromatogram from view
  • Forced Cal - perform a forced calibration
  • Cur/All - toggle the TEV and CGM Component spreadsheet display
  • Save CMP - save multiple CGMs

To save the displayed chromatogram to disk, see Section 6.4.

To remove a chromatogram from view, see Section 6.3.

To toggle the TEV and CGM Component spreadsheet display, see Section 6.6.8.

JULY 2010 Printing the Chromatogram

Page 252 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6-12 CHROMATOGRAM VIEWER
Page 252 · click to enlarge
Page 252

6-12 CHROMATOGRAM VIEWER

6.6.1 Editing the Chromatogram

Use the Edit Chromatogram function to change the x/y offset values and other display parameters such as color.

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button. The Edit Chromatogram dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 6-12 CHROMATOGRAM VIEWER diagram (page 252)
Fig. 204
3. Use the X Offset and Y Offset values to change the chromatogram display as desired.
4. Use the Color pull-down menu to assign a different color to this chromatogram.
5. Click the button to accept your changes and return to the Chromatogram Viewer dialog.

Click the button to abort and return to the Chromatogram Viewer dialog.

Editing the Chromatogram JULY 2010

Page 253 of the Emerson Process Management Chromatograph Software (MON2000) user manual — CHROMATOGRAM VIEWER 6-13 MON2000
Page 253 · click to enlarge
Page 253

CHROMATOGRAM VIEWER 6-13 MON2000

6.6.2 Editing TEVs from CGM Viewer

To edit the Timed Events Table (TEV) from the Chromatogram Viewer window,

In the edit mode, initial focus is applied to the currently selected event in the table, or to the event with the closest time.

1. Double-click the left mouse button, while hovering over the TEV legend, to edit the timed events from the CGM viewer (see the following figures).

If more than one CGM is displayed, only the selected CGM timed events will be available to edit.

Right-click and use the drop-down menu to modify, add, insert, and delete TEV’s via the normal editing procedures.

Emerson Process Management Chromatograph Software (MON2000) — CHROMATOGRAM VIEWER 6-13 MON2000 diagram (page 253)
Fig. 205
Emerson Process Management Chromatograph Software (MON2000) — CHROMATOGRAM VIEWER 6-13 MON2000 diagram (page 253)
Fig. 206

JULY 2010 Editing TEVs from CGM Viewer

Or,

Page 254 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6-14 CHROMATOGRAM VIEWER
Page 254 · click to enlarge
Page 254

6-14 CHROMATOGRAM VIEWER

Emerson Process Management Chromatograph Software (MON2000) — 6-14 CHROMATOGRAM VIEWER diagram (page 254)
Fig. 207

MON2000 displays the Timed Events Table dialog.

Emerson Process Management Chromatograph Software (MON2000) — 6-14 CHROMATOGRAM VIEWER diagram (page 254)
Fig. 208

Editing TEVs from CGM Viewer JULY 2010

Page 255 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.6.4 Editing TEVs from Cursor
Page 255 · click to enlarge
Page 255

CHROMATOGRAM VIEWER 6-15 MON2000

2. Click the button to save your changes, or press the F8 key.
3. Press the ESC key to discard changes.

Save changes and discard changes options are available from the right- click menu.

6.6.3 Editing Retention Times from CGM Viewer

To edit component retention from the Chromatogram Viewer window,

1. Double-click the left mouse button to edit the retention times from the CGM viewer

You must access the Component Data Table to add, insert, and delete retention times. Use the Application> Component Data menu or press the F6 key.

Right-click and use the drop-down menu to modify retention times via the normal editing procedures.

6.6.4 Editing TEVs from Cursor

To update Timed Events from the Chromatogram Viewer using the cursor,

1. Right-click the mouse button, with the mouse pointer hovered over the Event Time legend.
Emerson Process Management Chromatograph Software (MON2000) — 6.6.4 Editing TEVs from Cursor diagram (page 255)
Fig. 209 · 6.6.4 Editing TEVs from Cursor

From the pull-down menu, select Update Time from Cursor.

JULY 2010 Editing Retention Times from CGM Viewer

Or,

Page 256 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.6.5 Viewing the Chromatogram Results
Page 256 · click to enlarge
Page 256

6-16 CHROMATOGRAM VIEWER

Press the F9 key. MON2000 displays the message, “Reading GC Data” in the status bar, then actives the TEV Table.

Emerson Process Management Chromatograph Software (MON2000) — 6-16 CHROMATOGRAM VIEWER diagram (page 256)
Fig. 210
2. Click the button to save your changes, or press the F8 key.

Save changes and discard changes options are available from the right- click menu.

Press the ESC key to discard changes.

6.6.5 Viewing the Chromatogram Results

To read a table of calculation results for a selected chromatogram,

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button. A dialog appears asking if the results should be displayed using the calculation control flags.
3. Click the button to display the report using the calculations selected in the Application>Control menu.

Viewing the Chromatogram Results JULY 2010

Or,

Page 257 of the Emerson Process Management Chromatograph Software (MON2000) user manual — CHROMATOGRAM VIEWER 6-17 MON2000
Page 257 · click to enlarge
Page 257

CHROMATOGRAM VIEWER 6-17 MON2000

Click the button to display the report using ALL calculations.

Emerson Process Management Chromatograph Software (MON2000) — CHROMATOGRAM VIEWER 6-17 MON2000 diagram (page 257)
Fig. 211

Click the button to return to the Chromatogram Viewer dialog.

JULY 2010 Viewing the Chromatogram Results

Page 258 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6-18 CHROMATOGRAM VIEWER
Page 258 · click to enlarge
Page 258

6-18 CHROMATOGRAM VIEWER

6.6.6 Entering a Description

To enter a description for the selected chromatogram,

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button. The Edit Description dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 6-18 CHROMATOGRAM VIEWER diagram (page 258)
Fig. 212
3. Type the desired statement(s).
4. Press the ENTER key to accept this description.

Use the button or press the ESC key to exit this dialog without accepting your entry.

5. Click the button to save this description with the chromatogram file.

Entering a Description JULY 2010

Page 259 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.6.8 Toggling TEVs and CGM Components
Page 259 · click to enlarge
Page 259

CHROMATOGRAM VIEWER 6-19 MON2000

6.6.7 Forcing a Calibration

Use the Forced Cal function to calibrate the GC, using the raw data and stream number for the selected GC Archive chromatogram. The results are stored in the CDT under the corresponding stream number.

The selected GC Archive chromatogram must be from a good calibration gas run.

To perform a forced calibration,

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button.

6.6.8 Toggling TEVs and CGM Components

To toggle the display of the TEV and CGM Component speadsheets,

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button. If multiple CGMs are displayed in the graph area, the 'CUR/ALL' button toggles between showing all TEVs and all components for all CGMS to showing only the TEVs and components for the currently selected stream.

JULY 2010 Forcing a Calibration

Page 260 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.7 VIEWING BASELINE DATA
Page 260 · click to enlarge
Page 260

6-20 CHROMATOGRAM VIEWER

3. MON2000 toggles the display of the TEVs and Components spreadsheets and the CGM view.
Emerson Process Management Chromatograph Software (MON2000) — 6-20 CHROMATOGRAM VIEWER diagram (page 260)
Fig. 213
Emerson Process Management Chromatograph Software (MON2000) — 6-20 CHROMATOGRAM VIEWER diagram (page 260)
Fig. 214

Cur/All View 1 Cur/All View 2

6.7 VIEWING BASELINE DATA

Use the Baseline function to display the retention times and baselines for the selected chromatogram.

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button. This removes the integration lines from a displayed Archive CGM.
3. MON2000 displays a plot of the baseline values, and lists the corresponding retention times in the legend table.

Viewing Baseline Data JULY 2010

Page 261 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6.9 DISPLAY OPTIONS
Page 261 · click to enlarge
Page 261

CHROMATOGRAM VIEWER 6-21 MON2000

6.8 VIEWING RAW DATA

Use the Raw Data function to display the Raw Data Table for the selected chromatogram.

1. Use the Chromatogram pull-down menu to select a specific chromatogram.
2. Click the button. The Raw Data dialog displays and shows the raw data for all CGMs in the graph area.
Emerson Process Management Chromatograph Software (MON2000) — CHROMATOGRAM VIEWER 6-21 MON2000 diagram (page 261)
Fig. 215
3. Click the button to return to the Chromatogram Viewer dialog.

6.9 DISPLAY OPTIONS

Right-click the graph area to access these options, or use the corresponding keystroke(s). Options are listed in alphabetical order.

See Table 6-1 for a complete listing of option names, descriptions, and keystroke sequences.

JULY 2010 Viewing RAW Data

Page 262 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 6-22 CHROMATOGRAM VIEWER
Page 262 · click to enlarge
Page 262

6-22 CHROMATOGRAM VIEWER

Table 6-1 Display Options for Chromatogram Viewer

Keystroke Right-Click Option Description

CTRL C Copy to clipboard allows you to copy this graph to another application such as MSWord or Excel

F8 Cursor to nearest point snap cursor to nearest point in both the X and Y directions

CTRL V Paste from clipboard allows you to plot a range of points from another application such as MSWord or Excel

CTRL P Print Graph Area print currently displayed chromatograph area

CTRL S Print Series All print all displayed chromatograms

HOME Restore State restore last saved display settings for Live or Archived for the selected chromatogram

CTRL HOME Save State save current or archived display settings for the selected chromatogram

F4 Toggle Coarse/Fine toggle cursor from coarse (less accurate) to fine (more accurate)

F9 Toggle Lines/Dots toggle graph from line(s) to dots

CTRL F4 Toggle Mouse Position Tip graph cursor follows movement of mouse while tooltip displays exact coordinates of the current point

CTRL F9 Toggle Nearest Point Tip graph cursor follows movement of mouse cursor

NUM + Zoom In zoom in on entire graph

NUM – Zoom Out zoom out from entire graph

NUM → Zoom X In zoom in on X axis

NUM ← Zoom X Out zoom out from X axis

NUM ↑ Zoom Y In zoom in on Y axis

NUM ↓ Zoom Y Out zoom out from Y axis

Display Options JULY 2010

Page 263 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.1.1 View Report from Live Data
Page 263 · click to enlarge
Page 263

REPORTS 7-1 MON2000

7 REPORTS

The options listed in the Reports menu enable you to display, print, or store various pre- configured GC reports.

You must be currently online with the desired GC unit to generate a report from the controller.

See the following sections for more information on specific reports and related functions.

7.1 REPORT DISPLAY

This function allows you to immediately display, print, or store preconfigured reports on GC analysis data. Data is reported either real- time from the GC unit or from previously stored files.

7.1.1 View Report from Live Data

To display a report with live data from the GC unit,

Emerson Process Management Chromatograph Software (MON2000) — 7.1.1 View Report from Live Data diagram (page 263)
Fig. 216 · 7.1.1 View Report from Live Data
1. Use the Reports > Report Display menu to access this function.
2. The Report Display dialog appears, per GC type.
Emerson Process Management Chromatograph Software (MON2000) — 7.1.1 View Report from Live Data diagram (page 263)
Fig. 217 · 7.1.1 View Report from Live Data
Emerson Process Management Chromatograph Software (MON2000) — 7.1.1 View Report from Live Data diagram (page 263)
Fig. 218 · 7.1.1 View Report from Live Data

Report Display dialog for 2350A GC Controllers

Report Display dialog for Model 700 GC

JULY 2010 Report Display

Page 264 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7-2 REPORTS
Page 264 · click to enlarge
Page 264

7-2 REPORTS

3. Select the report type, stream, and update setting from the lists provided.

Table 7-1 Description of Report Display Settings

The contents of the GC analysis reports are defined by:

Reports list of pre-defined reports; Shows report type (Analysis, Raw Data, Calibration, etc...)

  • the GC application
  • functions such as Calculations, User Defined, Limit Alarms, and Streams

Select Stream list of available streams

For a single detector GC0–Current = last stream analyzed

For a dual detector GC: 0–Current Det1 = last stream analyzed, Detector 1 0–Current Det2 = last stream analyzed, Detector 2

Update Method list of update options that determine when the report display is refreshed

Anly = after each analysis Cal = after each calibration Fcal = after each final calibration None = not updated

4. Click the button or press the F2 key to generate and display the selected report.

To close this dialog without generating a

report, click the button to return to the main window.

View Report from Live Data JULY 2010

Setting Description

Page 265 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-3 MON2000
Page 265 · click to enlarge
Page 265

REPORTS 7-3 MON2000

5. MON2000 generates the report, per your specifications, and displays the results in the File Viewer dialog.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-3 MON2000 diagram (page 265)
Fig. 219

MON2000 creates a temporary HTML file to display the generated report. If you want to save this report to disk, see Section 7.1.4.

The display refreshes per the update setting configured in the Report Display dialog (see Section 7.4).

For example, in the Report Display dialog, if Stream is set to “0– Current” and Update is set to “Anly”, the report display will refres after the GC analyzes the stream selected via the Report Display dialog.

The refresh function displays the newly generated report and deletes the previous report (unless already saved to disk).

Check the GC status bar for the run time and current stream.

Use the scroll bars to see other areas of the report. Use the button to increase or decrease the font size displayed.

JULY 2010 View Report from Live Data

Page 266 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.1.2 View Report from File
Page 266 · click to enlarge
Page 266

7-4 REPORTS

See the following sections for instructions printing the displayed file or opening other report files from disk.

6. Click the button or press the ESC key to exit this function and return to the Report Display dialog.

7.1.2 View Report from File

MicroSoft Internet Explorer 5.0 (or later) is required to view a report file in HTML format.

To open a report file for viewing,

1. Use the Reports > Report Display menu to access this function.
2. The Report Display dialog appears.

If you are working in off-line mode, the Report Display dialog will appear blank.

Emerson Process Management Chromatograph Software (MON2000) — 7.1.2 View Report from File diagram (page 266)
Fig. 220 · 7.1.2 View Report from File

View Report from File JULY 2010

Page 267 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-5 MON2000
Page 267 · click to enlarge
Page 267

REPORTS 7-5 MON2000

3. Click the button or press the

F3 key to access the File Viewer dialog.

Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-5 MON2000 diagram (page 267)
Fig. 221
4. Click the button. The Open dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-5 MON2000 diagram (page 267)
Fig. 222
5. Use the provided directory tree to select the

desired report file. Click the button.

By default, the Open dialog displays the c:\GC\Save directory.

6. MON2000 opens the file and displays it in the File Viewer dialog.

Use the scroll bars to see other areas of the

report. Use the button to increase or decrease the font size displayed.

JULY 2010 View Report from File

Page 268 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.1.4 Save Report to Disk
Page 268 · click to enlarge
Page 268

7-6 REPORTS

Use the and buttons to browse through multiple report files.

7. Click the button or press the ESC key to return to the Report Display dialog.

7.1.3 Print Report

To print a report,

The refresh function displays a newly generated report and deletes the currently displayed report (see Section 7.1.1 for details).

You should print or save the current report immediately after it is displayed.

1. Display a live report or open a report file from disk. See Section 7.1.1 or Section 7.1.2 for details.
2. Click the button.
3. MON2000 prints the report to your configured printer ( see Section 3.7 ).

7.1.4 Save Report to Disk

To save a report to disk,

1. Display a live report or open a report file from disk. See Section 7.1.1 or Section 7.1.2 for details.

Print Report JULY 2010

Page 269 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-7 MON2000
Page 269 · click to enlarge
Page 269

REPORTS 7-7 MON2000

2. Click the button. The Save HTML Document dialog appears.

MicroSoft Internet Explorer 5.0 (or later) is required to view a report file in HTML format.

Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-7 MON2000 diagram (page 269)
Fig. 223
3. Select the directory in which you want to save this report and type the desired name.

You can save the report in any of these formats:

  • .txt, ASCII Tab-delimited
  • .cvs, Comma-deliminted
  • .xls, Excel
  • .htm, HTML
  • .xml, EXtensible Markup Language
4. Click the button.

JULY 2010 Save Report to Disk

Page 270 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7-8 REPORTS
Page 270 · click to enlarge
Page 270

7-8 REPORTS

7.1.5 Report Contents

This section includes examples of the following reports: Analysis, Raw Data, Calibration, 24-Hour Average, Weekly Average, Hourly Average, and Variable Average. After each sample report, a table describes the contents and identifies which functions affect the data.

Each report contains the header information, including:

Table 7-2 Contents Guide for Report Headers

Content Related Function(s) Menu Section

Date-Time standard GC application n/a n/a

Analysis Time Timed Events Table (TEV) Application > Timed Events [F5]

Cycle Time TEV Application > Timed Events [F5]

Stream Report Display, Select Stream

Reports > Report Display

Mode standard GC application n/a n/a

Cycle Start Time standard GC application n/a n/a

Analyzer System, Analyzer Name Application > System 5.1

Strm Seq System, Stream Sequence Application > System 5.1

User Defined, Text Strings Application > User Defined > Text Strings

Report Contents JULY 2010

Stream Name · Application > Streams · Report Header (optional)

Page 271 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-9 MON2000
Page 271 · click to enlarge
Page 271

REPORTS 7-9 MON2000

Analysis Report

Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-9 MON2000 diagram (page 271)
Fig. 224

JULY 2010 Report Contents

Page 272 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7-10 REPORTS
Page 272 · click to enlarge
Page 272

7-10 REPORTS

Table 7-3 Contents Guide for Analysis Report

Content Related Function(s) Menu Section

Component Name Component Data Table (CDT) Application > Component Data [F6]

GC Analysis Data variables (std)

Calculations, Control Application > Calculations > Control

GC Analysis Data variables (user)

Calculations, User Defined Application > Calculations > User Defined

Alarms Limit Alarms Application > Limit Alarms 5.6

Application > Calculations > Control

Report Contents JULY 2010

Show 4 more labels / short text on this page
Additional Base Pressures · Compress(ibility) enabled · Streams, Optional Pressure · Application > Streams
Page 273 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-11 MON2000
Page 273 · click to enlarge
Page 273

REPORTS 7-11 MON2000

Raw Data Report

Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-11 MON2000 diagram (page 273)
Fig. 225

Table 7-4 Contents Guide for Raw Data Report

Content Related Function(s) Menu Section

Retention Time Peak Area Peak Height Det(ector) Integration Start/End Peak Width 1/2 Height

standard GC application n/a n/a

Mthd (method) standard GC application, peak derivation method (non-edit)

JULY 2010 Report Contents

Chromatogram Viewer

Page 274 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7-12 REPORTS
Page 274 · click to enlarge
Page 274

7-12 REPORTS

Calibration Report

Emerson Process Management Chromatograph Software (MON2000) — 7-12 REPORTS diagram (page 274)
Fig. 226

Table 7-5 Contents Guide for Calibration Report

Content Related Function(s) Menu Section

Component Name Cal. Conc.

CDT Application > Component Data [F6]

Raw Data New RF RF % Dev. New RT RT % Dev.

CDT (additional columns) Application > Component Data [F6]

Alarms Limit Alarms Application > Limit Alarms 5.6

Streams used for analysis display “0”s in the calibration raw data columns (see Section 5.8)

Report Contents JULY 2010

Page 275 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-13 MON2000
Page 275 · click to enlarge
Page 275

REPORTS 7-13 MON2000

Final Calibration Report

Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-13 MON2000 diagram (page 275)
Fig. 227

Table 7-6 Contents Guide for Final Calibration Report

Content Related Function(s) Menu Section

Component Name Cal. Conc.

CDT Application > Component Data [F6]

Old RF New RF RF % Dev. Old RT New RT RT % Dev.

Application > Component Data [F6]

results of final calibration, adjustments to response factors (RFs) and retention times (RTs)

Alarms Limit Alarms Application > Limit Alarms 5.6

Report items marked with an asterisk (*) were updated during calibration.

JULY 2010 Report Contents

standard GC application · CDT (additional columns)

Page 276 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Sample 24-Hour Average Report
Page 276 · click to enlarge
Page 276

7-14 REPORTS

Average Reports

There are five types of Average Reports:

  • Hourly Average Report
  • 24-Hour Average Report
  • Weekly Average Report
  • Monthly Average Report
  • Variable Average Report

Sample 24-Hour Average Report

Emerson Process Management Chromatograph Software (MON2000) — Sample 24-Hour Average Report diagram (page 276)
Fig. 228 · Sample 24-Hour Average Report

Report Contents JULY 2010

Page 277 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.2 GC REPORT REQUEST
Page 277 · click to enlarge
Page 277

REPORTS 7-15 MON2000

Table 7-7 Contents Guide for Average Reports

Content Related Function(s) Menu Section

Item Number Analysis data/calcula- tion Stream Component

Calculations, Averages Application > Calculations > Averages

Old RF New RF RF % Dev. Old RT New RT RT % Dev.

results of ongoing or completed averages

7.2 GC REPORT REQUEST

This function allows you to send a report to a printer connected to the GC Controller.

Emerson Process Management Chromatograph Software (MON2000) — 7.2 GC REPORT REQUEST diagram (page 277)
Fig. 229
1. Use the Reports > GC Report Request menu to access this function.
2. The GC Report Request dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 7.2 GC REPORT REQUEST diagram (page 277)
Fig. 230

JULY 2010 GC Report Request

standard GC application · n/a n/a

Page 278 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.3 GC PRINTER CONTROL
Page 278 · click to enlarge
Page 278

7-16 REPORTS

3. Select the report you wish to print.

The GC Config Report is a snapshot of the PC Config Report. Contents include:

  • current alarms
  • streams
  • control calculations
  • CDTs
  • TEVs
  • GC serial port settings
  • analog outputs

See Appendix A for a sample PC Config Report.

For reports other than the GC Config, a Streams list displays.

Double-click the appropriate stream and the controller sends the report to the printer.

Click the button. Double-click the appropriate stream and the print process begins.

Click the button to abort and return to the main window.

7.3 GC PRINTER CONTROL

This function allows you to define schedules for automatic printing of standard GC reports to a printer connected to the GC Controller.

1. Use the Reports > GC Printer Control menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — 7.3 GC PRINTER CONTROL diagram (page 278)
Fig. 231

GC Printer Control JULY 2010

Page 279 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-17 MON2000
Page 279 · click to enlarge
Page 279

REPORTS 7-17 MON2000

2. The GC Printer Control dialog displays.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-17 MON2000 diagram (page 279)
Fig. 232
3. Use the provided data fields, check and combo boxes to select the desired settings. See Table 7-8 for detailed descriptions.

You can, for example, set a report to print after each analysis and every 12 hours.

4. Click the button to accept your selections and return to the main window.

Click the button to abort and return to the main window.

Table 7-8 Description of Settings for GC Printer Control

Anly Enables/disables report printing after each analysis run regardless of report type selected.

Cal Enables/disables report printing after each calibration run regardless of report type selected.

FCal Enable/disable report printing after every final calibration run regardless of report type selected.

Time Time at which report should print at 24-hour intervals.

To disable this feature, set Time to 00:00 or 12:00 AM.

The Time setting overrides the Hours setting and only applies to streams that are enabled (i.e., checked ON) for report printing.

JULY 2010 GC Printer Control

Setting Description

Page 280 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.4 MON2000 PRINTER CONTROL
Page 280 · click to enlarge
Page 280

7-18 REPORTS

Table 7-8 Description of Settings for GC Printer Control (Continued)

Hours hour interval at report should print

Use this setting for intervals less than 24 hours. For example, use this setting if you wish to print a report every 8 hours. Ensure that the Time value is “00:00” to disable the 24-hour setting or “12:00 AM” to disable the 12-hour setting.

This setting only applies to streams enabled (i.e., checked ON) for report printing.

Avg average type to be used in the report calculations; calculations are dependent on settings in the Averages Calculations dialog (see Section 5.5.2)

No = disabled Hr = hourly avg 24 = 24-hour avg Wk = weekly avg Mn = monthly avg Var = variable avg

streams to be used for data collection when the report is generated

enable/disable manual paper feed feature of printer

enable/disable real-time printing of alarms as they occur

enable/disable real-time printing of events as they occur

7.4 MON2000 PRINTER CONTROL

This function allows you to define schedules for automatic printing of standard GC reports to a printer connected to your PC. See Section 3.7 to configure the printer settings. The user must be on-line (connected to the controller) using either a serial port direct connection, modem or Ethernet card in order to print from the MON2000 printer.

MON2000 Printer Control JULY 2010

Show 4 more labels / short text on this page
Setting Description · Form Feed · Alarm Logging · Event Logging
Page 281 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-19 MON2000
Page 281 · click to enlarge
Page 281

REPORTS 7-19 MON2000

1. Use the Reports > MON2000 Printer Control menu to access this function.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-19 MON2000 diagram (page 281)
Fig. 233
2. The Printer Control dialog displays.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-19 MON2000 diagram (page 281)
Fig. 234

Use the provided data fields, check and combo boxes to select the desired settings. See Table 7-9 for detailed descriptions.

You can, for example, set a report to print after each analysis and every 12 hours.

3. Click the button to accept your selections and return to the main window.

Each report generation interrupts all current functions.

Click the button to abort and return to the main window.

Table 7-9 Description of Settings for MON2000 Printer Control

Anly enable/disable report printing after each analysis run regardless of report type selected

Cal enable/disable report printing after each calibration run regardless of report type selected

FCal enable/disable report printing after every final calibration run regardless of report type selected

Time time at which report should print (every 24 hours)

00:00 = disables 24-hour setting 12:00 AM = disables 12-hour setting

JULY 2010 MON2000 Printer Control

Setting Description

Page 282 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.5 ARCHIVE DATA
Page 282 · click to enlarge
Page 282

7-20 REPORTS

Table 7-9 Description of Settings for MON2000 Printer Control (Continued)

Hours hour interval at report should print

Use this setting for intervals less than 24 hours. For example, use this setting if you wish to print a report every 8 hours. Ensure that the Time value is “00:00” to disable the 24-hour setting or “12:00 AM” to disable the 12-hour setting.

Avg average type to be used in the report calculations; calculations are dependent on settings in the Averages Calculations dialog (see Section 5.5.2)

No = disabled Hr = hourly avg 24 = 24-hour avg Wk = weekly avg Mn = monthly avg Var = variable avg

streams to be used for data collection when the report is generated

enable/disable manual paper feed feature of printer

7.5 ARCHIVE DATA

This function enables you to view, print, or save various reports and records stored in the GC memory for the Model 500 and Model 700 analyzers. You can also reset (i.e., delete) the stored records from the GC memory.

Archive Data JULY 2010

Setting Description · Form Feed

Page 283 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.5.1 View Data Model 500/Model 700
Page 283 · click to enlarge
Page 283

REPORTS 7-21 MON2000

The standard GC application can store up to ninety-nine days of analysis runs and calibration records.

Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-21 MON2000 diagram (page 283)
Fig. 235

7.5.1 View Data Model 500/Model 700

Emerson Process Management Chromatograph Software (MON2000) — 7.5.1 View Data Model 500/Model 700 diagram (page 283)
Fig. 236 · 7.5.1 View Data Model 500/Model 700

To view, print, copy or save archived data,

1. Use the Reports > Archive Data > View Data menu to access this function.
2. MON2000 displays the Date Range Selection pop-up dialog.
Emerson Process Management Chromatograph Software (MON2000) — 7.5.1 View Data Model 500/Model 700 diagram (page 283)
Fig. 237 · 7.5.1 View Data Model 500/Model 700

Select All Dates radio button or choose Selected Dates. Select the Start date and End Date (date range) for the report to be archived.

JULY 2010 View Data Model 500/Model 700

Page 284 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7-22 REPORTS
Page 284 · click to enlarge
Page 284

7-22 REPORTS

3. Click the button to apply your selection.

Click the button to return to the View Data window.

MON2000 displays the message “Reading GC Data” in the status bar.

4. The Select Archive Record dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 7-22 REPORTS diagram (page 284)
Fig. 238

Double-click the desired archived report or record. (If no records are found, an information dialog displays.) Select a range (multiple) of records to archive by left- clicking the mouse at the beginning record and dragging the pointer to the last record to be archived. Then, press the ENTER key.

Left-click the mouse pointer on the beginning record, then move the mouse pointer to another location (either further up or down the list) and while holding down the Shift key, click the ending record. All highlighted records will be automatically selected. Press the Enter key to display all of the selected records.

MON2000 limits the number of viewable Archived Records to 100.

View Data Model 500/Model 700 JULY 2010

Or, · Or,

Page 285 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-23 MON2000
Page 285 · click to enlarge
Page 285

REPORTS 7-23 MON2000

5. The Calculation Results from Ext. Modbus window appears.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-23 MON2000 diagram (page 285)
Fig. 239

Use the button or press the ESC key to exit this function and return to the main window.

As applicable, use the scroll bars to see other areas of the record.

JULY 2010 View Data Model 500/Model 700

Page 286 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.5.2 Archive Export Data
Page 286 · click to enlarge
Page 286

7-24 REPORTS

6. To print, copy to the clipboard, or save the report to disk, right-click the Archive Data sheet and select the desired option from the popup menu.

Or, use the , or

buttons for these functions.

Use the Save File dialog to choose the directory location and file name.

You can save the report in any of these formats:

  • .txt, ASCII Tab-delimited
  • .cvs, Comma-deliminted
  • .xls, Excel
  • .htm, HTML
  • .xml, EXtensible Markup Language
7. Click the button to return to the main window.

7.5.2 Archive Export Data

Use this function to export data files for review at a later time.

Emerson Process Management Chromatograph Software (MON2000) — 7.5.2 Archive Export Data diagram (page 286)
Fig. 240 · 7.5.2 Archive Export Data

From the Reports>Archive Data>Export Data menu,

Archive Export Data JULY 2010

Page 287 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-25 MON2000
Page 287 · click to enlarge
Page 287

REPORTS 7-25 MON2000

1. Click Export Data and the Export Archive window displays.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-25 MON2000 diagram (page 287)
Fig. 241
2. The exported data file is saved in the default (GC>Directory). Use the pull-down menu to save the data file to a different directory.
3. Click an existing file (to write over an existing file) or enter a 'new' file name in the 'Data Field'.
4. Save the file as either:
  • .csv (spreadsheet application format)
  • .txt (Word Pad or Note Pad application format)
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-25 MON2000 diagram (page 287)
Fig. 242

JULY 2010 Archive Export Data

Page 288 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7-26 REPORTS
Page 288 · click to enlarge
Page 288

7-26 REPORTS

If the GC you are using is a Model 2350A, the Data Selection Range' dialog appears. Choose the date range as:

Emerson Process Management Chromatograph Software (MON2000) — 7-26 REPORTS diagram (page 288)
Fig. 243
  • All dates (radio button) or,
  • Selected Dates (radio button)
  • Start date
  • End date
Emerson Process Management Chromatograph Software (MON2000) — 7-26 REPORTS diagram (page 288)
Fig. 244

Also note that the Date Range dialog only displays when files are archived for more than one day.

5. Click the button to apply your

selection. Click the button to abort and return to the Trend Data window.

6. The Select Archive Records dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 7-26 REPORTS diagram (page 288)
Fig. 245
Emerson Process Management Chromatograph Software (MON2000) — 7-26 REPORTS diagram (page 288)
Fig. 246

Select Archive Record Default

Select Archive Record Selection

To Select all records, the operator may drag the slide bar down to the end of the list, then while holding down the 'Shift' key click the last report and MON2000 will automatically select all records in between. This also works for any number of records within the list.

Archive Export Data JULY 2010

Default Settings · Date Range Selection

Page 289 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.5.3 Reset
Page 289 · click to enlarge
Page 289

REPORTS 7-27 MON2000

The operator may select several records individually by holding down the 'Ctrl' key and clicking the mouse pointer on random selected reports. Reports that are several hours apart may be viewed in this manner.

Or, as the example .BMP above shows, the operator may click a report and drag the mouse pointer down, highlighting several reports in sequential order.

7.5.3 Reset

Emerson Process Management Chromatograph Software (MON2000) — 7.5.3 Reset diagram (page 289)
Fig. 247 · 7.5.3 Reset

To delete archived data and reset the GC Controller memory,

1. Use the Reports > Archive Data > Reset menu to access this function.
2. MON2000 displays a confirmation dialog.
Emerson Process Management Chromatograph Software (MON2000) — 7.5.3 Reset diagram (page 289)
Fig. 248 · 7.5.3 Reset

Click the button to continue.

3. MON2000 clears the GC Controller memory. New archived records will begin accumulating again as analysis and calibration runs occur.

JULY 2010 Reset

Page 290 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.6.1 Trend Data Model 700/2350A GCs
Page 290 · click to enlarge
Page 290

7-28 REPORTS

7.6 TREND DATA

This function allows you to view, print, or save graphical representations, or trend lines, of accumulated analysis data for the 2350A and the Model 700 GC Controllers.

7.6.1 Trend Data Model 700/2350A GCs

1. Use the Reports > Trend Data menu, then
Emerson Process Management Chromatograph Software (MON2000) — 7.6.1 Trend Data Model 700/2350A GCs diagram (page 290)
Fig. 249 · 7.6.1 Trend Data Model 700/2350A GCs

click the button to access this function.

2. The Trend Data dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 7.6.1 Trend Data Model 700/2350A GCs diagram (page 290)
Fig. 250 · 7.6.1 Trend Data Model 700/2350A GCs

Use the enabled Graph and Trend push- buttons to access the Trend Data features. See the following sections for detailed information.

3. Click the button to return to the main window.

Trend Data JULY 2010

Page 291 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-29 MON2000
Page 291 · click to enlarge
Page 291

REPORTS 7-29 MON2000

7.6.2 View Live Trend Online Model 700/2350A

To view a live trend from the online Model 700/ 2350A GC,

You cannot view a live trend if the corresponding analysis record does not exist in the GC memory.

An error message will display:

Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-29 MON2000 diagram (page 291)
Fig. 251
1. Click the button.
2. The Select menu for variable displays.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-29 MON2000 diagram (page 291)
Fig. 252

Double-click the desired variable. If the selected variable applies to a single component (e.g., Gallons/1000 SCF, Liquid Volume Percent, or Mole Percent), then the Standard Components menu appears.

JULY 2010 View Live Trend Online Model 700/2350A

Page 292 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7-30 REPORTS
Page 292 · click to enlarge
Page 292

7-30 REPORTS

3. The Select menu for streams appears. Double-click the left mouse button on the desired stream to make your selection.
Emerson Process Management Chromatograph Software (MON2000) — 7-30 REPORTS diagram (page 292)
Fig. 253
4. MON2000 displays the Date Range Selection pop-up dialog.

Select All Dates radio button or choose Selected Dates. From the pull-down menu, select the date range for the Trend report.

Click the to apply your selection.

Click the button to return to the trend window.

5. MON2000 reads data from the online GC unit and plots the current, live trend for the selected component and stream.

Plotting a trend line interrupts the normal storage of ongoing Auto Sequence analysis results.

A trend line graph generated from the maximum 1200 archived analysis records can require up to 8 minutes to plot. Any Auto Sequence analysis that occurs during the plotting will not be retained for future retrieval nor will be included in ongoing calculations.

View Live Trend Online Model 700/2350A JULY 2010

Or,

Page 293 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-31 MON2000
Page 293 · click to enlarge
Page 293

REPORTS 7-31 MON2000

The Trend Data window appears.

JULY 2010 View Live Trend Online Model 700/2350A

Page 294 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.6.3 View Trend from File on Disk
Page 294 · click to enlarge
Page 294

7-32 REPORTS

The corresponding data is dynamically displayed in the provided legend.

Emerson Process Management Chromatograph Software (MON2000) — 7-32 REPORTS diagram (page 294)
Fig. 254

Use the Graph and Trend Options functions to manipulate the display. To zoom in on a particular point, click and drag your mouse over that area; repeat as necessary. Use the Restore State or Zoom Out options from the right-click menu (see Section 7.6.10) to return to the original display.

7.6.3 View Trend from File on Disk

To view a trend from a file previously saved to disk, use the Reports > Trend Data menu, then click the button to access this function.

1. Click the button.
2. Use the provided directory tree to locate the

desired TRD file and click the button to open.

Emerson Process Management Chromatograph Software (MON2000) — 7.6.3 View Trend from File on Disk diagram (page 294)
Fig. 255 · 7.6.3 View Trend from File on Disk

View Trend from File on Disk JULY 2010

Page 295 of the Emerson Process Management Chromatograph Software (MON2000) user manual — REPORTS 7-33 MON2000
Page 295 · click to enlarge
Page 295

REPORTS 7-33 MON2000

3. MON2000 displays the trend file.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-33 MON2000 diagram (page 295)
Fig. 256

The corresponding data is dynamically displayed in the provided legend.

Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-33 MON2000 diagram (page 295)
Fig. 257

Use the Graph and Trend Options functions to manipulate the display. To zoom in on a particular point, click and drag your mouse over that area; repeat as necessary. Use the Restore State or Zoom Out options from the right-click menu (see Section 7.6.10) to return to the original display.

JULY 2010 View Trend from File on Disk

Page 296 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.6.5 Changing Cursor Size
Page 296 · click to enlarge
Page 296

7-34 REPORTS

7.6.4 Edit Graph Display

1. Click the button. The Edit Graph dialog appears.
2. Use the provided attributes and options to change the trend display as desired.
3. Click the button to accept your changes and return to the Trend Data dialog.

Click the button to abort and return to the Trend Data dialog.

4. The trend display changes as specified.

Additional display options are available by right-clicking the display area. See Section 7.6.10 for more information.

7.6.5 Changing Cursor Size

To toggle the cursor size from course movement (less accurate) to fine movement (more accurate), click the button.

Edit Graph Display JULY 2010

Page 297 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.6.7 Print Trend
Page 297 · click to enlarge
Page 297

REPORTS 7-35 MON2000

7.6.6 Describe Trend

1. Click the button. The Edit Description dialog appears.
2. Type the desired statement(s).
3. Press the ENTER key to save this description.

Use the button or press the ESC key to exit this dialog without saving your entry.

7.6.7 Print Trend

To print the displayed trend graph, click the

button.

MON2000 prints the report to your configured printer (see Section 3.7).

JULY 2010 Describe Trend

Page 298 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.6.9 Read Trend Archive
Page 298 · click to enlarge
Page 298

7-36 REPORTS

7.6.8 Save Trend

To save a currently displayed trend to disk,

The table is saved as a binary file with a .trd extension.

1. Click the button.
2. Use the provided directory tree to select the desired file location and name, then click

the button.

3. MON2000 saves the trend along with its graph settings.

7.6.9 Read Trend Archive

This feature is not available if the corresponding analysis record does not exist in the GC memory or if you are viewing this trend data offline from a file.

To read the Trend Archive,

1. Click the button.

If you are operating a Model 700 or 2350A GC, then an Archive Range Selection dialog displays.

Emerson Process Management Chromatograph Software (MON2000) — 7.6.9 Read Trend Archive diagram (page 298)
Fig. 258 · 7.6.9 Read Trend Archive

(a) Toggle the appropriate radio button to

select either All Dates or Selected Dates.

To specify a date range, use the Start Date and End Date pull-down menus.

(b) Click the button to continue.

Save Trend JULY 2010

Page 299 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7.6.10 Display Options
Page 299 · click to enlarge
Page 299

REPORTS 7-37 MON2000

2. The Archive dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — REPORTS 7-37 MON2000 diagram (page 299)
Fig. 259
3. Click the button to return to the Trend Data dialog.

7.6.10 Display Options

Right-click the graph area to access these options, or use the corresponding keystroke(s). Options are listed in alphabetical order.

JULY 2010 Display Options

Page 300 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 7-38 REPORTS
Page 300 · click to enlarge
Page 300

7-38 REPORTS

Table 7-10 Display Options for Trend Data

Keystroke Right-Click Option Description

CTRL C Copy to clipboard allows you to copy this graph to another application such as MS Word or Excel

F8 Cursor to nearest point snap cursor to nearest point in both the X and Y directions

CTRL V Paste from clipboard allows you to plot a range of points from another application such as Meshwork or Excel

CTRL P Print Plot Area print currently displayed chromatograph area

CTRL S Print Series All print all displayed chromatograms

HOME Restore State restore last saved display settings for the selected chromatogram

CTRL HOME Save State save current display settings for the selected chromatogram

F4 Toggle Coarse/Fine toggle cursor from coarse (less accurate) to fine (more accurate)

F9 Toggle Lines/Dots toggle graph from line(s) to dots

CTRL F4 Toggle Mouse Position Tip graph cursor follows movement of mouse while tooltip displays exact coordinates of the current point

CTRL F9 Toggle Nearest Point Tip graph cursor follows movement of mouse cursor

NUM + Zoom In zoom in on entire graph

NUM – Zoom Out zoom out from entire graph

NUM → Zoom X In zoom in on X axis

NUM ← Zoom X Out zoom out from X axis

NUM ↑ Zoom Y In zoom in on Y axis

NUM ↓ Zoom Y Out zoom out from Y axis

Display Options JULY 2010

Page 301 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8.1 MAINTENANCE LOG
Page 301 · click to enlarge
Page 301

LOGS 8-1 MON2000

8Y LOGS

The options in the Logs pull-down menu allows you to keep a maintenance record, keep a parameter record, and view the Alarm and Event Logs.

8.1 MAINTENANCE LOG

Use this function to track maintenance activities performed on a given GC unit.

Emerson Process Management Chromatograph Software (MON2000) — 8.1 MAINTENANCE LOG diagram (page 301)
Fig. 260

To read and edit the Maintenance Log,

1. Use the Logs > Maintenance Log menu to access this function. MON2000 retrieves the data from the GC unit.
2. The Maintenance Log dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 8.1 MAINTENANCE LOG diagram (page 301)
Fig. 261

JULY 2010 Maintenance Log

Page 302 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8-2 LOGS
Page 302 · click to enlarge
Page 302

8-2 LOGS

3. To add the log entry text,

(a) Click the “Add Message F4” button or

click the last ‘Log Message’ cell and press the RETURN key. The Maintenance Log message dialog displays.

Emerson Process Management Chromatograph Software (MON2000) — 8-2 LOGS diagram (page 302)
Fig. 262

(b) Add text in the appropriate Log Message

cell, then click the button to

apply the changes. Click the button to discard your changes and return to the Main window.

A Log Message can contain up to 1000 characters.

4. To change the log entry text,

However, the entire Maintenance Log can contain no more than 1100 characters.

(a) Click the appropriate Log Message cell.

(b) Click the button. The Edit Log Message dialog appears. (c) Type your edits.

Maintenance Log JULY 2010

Page 303 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8.2 PARAMETER LIST
Page 303 · click to enlarge
Page 303

LOGS 8-3 MON2000

(d) To accept your edits, press the RETURN

key. To cancel your edits, click the button or press the ESC key.

Emerson Process Management Chromatograph Software (MON2000) — LOGS 8-3 MON2000 diagram (page 303)
Fig. 263
5. To delete a log entry, click the desired log entry and press the DELETE key.
6. Click the button to write this data to the GC unit.

Click the button to abort and return to the main window.

8.2 PARAMETER LIST

Use this function to keep a record of the hardware components and associated parameters for a given GC unit.

Emerson Process Management Chromatograph Software (MON2000) — 8.2 PARAMETER LIST diagram (page 303)
Fig. 264

To read and edit the Parameter List,

1. Use the Logs > Parameter List menu to access this function. MON2000 retrieves the data from the GC unit.

JULY 2010 Parameter List

Page 304 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8-4 LOGS
Page 304 · click to enlarge
Page 304

8-4 LOGS

2. The Parameter List dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 8-4 LOGS diagram (page 304)
Fig. 265

Table 8-1 Items in Parameter List

Operational SO # PN # Purchaser Location Order # Date Model # Control Serial # Analyzer Serial # Approved by Date Customer Approved by Date

Drawings Analyzer Flow Sample Conditioning System Interconnect Wiring Outline and Dimensional System Wiring Other

Parameter List JULY 2010

Type Parameter

Page 305 of the Emerson Process Management Chromatograph Software (MON2000) user manual — LOGS 8-5 MON2000
Page 305 · click to enlarge
Page 305

LOGS 8-5 MON2000

Table 8-1 Items in Parameter List (Continued)

Analysis Settings Flows in cc/min Sample Loop 1 Sample Loop 2 Detector Valve Part Numbers

Sample System Sample Valve 1 Flow Rate (cc/min) Sample 1 Pressure (PSIG) Sample Valve 2 Flow Rate (cc/min) Sample 2 Pressure (PSIG) SCS Oven Temperature (°C) Solenoid Purge Flow (cc/min/stream)

Column OD (in.) ID (in.) Length Support Mesh Phase Connected Function

3. To save this data to disk, click the

button or press the F3 key. The Save Parameter File dialog appears.

JULY 2010 Parameter List

Type Parameter

Page 306 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8.3 ALARM LOG
Page 306 · click to enlarge
Page 306

8-6 LOGS

This feature only saves the files in .par format. Use the Save Parameter File dialog to choose the directory location and file name.

4. To open an existing parameter file from

disk, click the button or press the F4 key.

Emerson Process Management Chromatograph Software (MON2000) — 8-6 LOGS diagram (page 306)
Fig. 266

Use the Open Parameter File dialog to choose the desired file.

5. Click the button to write your changes to the GC unit.

Click the button to abort and return to the main window.

8.3 ALARM LOG

Use this function to read and/or clear the various entries in the Alarm Log. The Alarm Log is a circular buffer that contains 50 entries.

Emerson Process Management Chromatograph Software (MON2000) — 8.3 ALARM LOG diagram (page 306)
Fig. 267

To clear or acknowledge all active alarms, see Section 8.4.

Alarm Log JULY 2010

Page 307 of the Emerson Process Management Chromatograph Software (MON2000) user manual — LOGS 8-7 MON2000
Page 307 · click to enlarge
Page 307

LOGS 8-7 MON2000

To view the Alarm Log,

1. Use the Logs > Alarm Log menu to access this function. MON2000 retrieves the last 50 alarms from the GC unit.
2. The Alarm Log dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — LOGS 8-7 MON2000 diagram (page 307)
Fig. 268

Table 8-2 Alarm Log Attributes

indicates whether the alarm has been acknowledge

U = unacknowledged ? = status unknown

State indicates whether the alarm is currently active

SET = currently active CLEAR = inactive

Date/Time date and time the alarm condition began

JULY 2010 Alarm Log

Attribute Description

Page 308 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8-8 LOGS
Page 308 · click to enlarge
Page 308

8-8 LOGS

Table 8-2 Alarm Log Attributes (Continued)

Alarm Message 1 describes the alarm condition

If an alarm message is changed (see Section 5.6 and Section 5.7) all affected alarm entries, including those previously recorded, will include that change.

Alarm Message 2 displays the alarm limit and current condition values, as applicable

3. By default, MON2000 displays all recorded alarms.

To view only the unacknowledged alarms, click the Unacknowledged alarms only radio button.

To view only the active alarms, click the Active alarms radio button.

4. To acknowledge a single alarm,

An alarm triggered by a user-defined value will continue to display as an active alarm until that value is no longer in the alarm state.

(a) Ensure you have selected the correct

display option (see Step 3). (b) Select the alarm you want to

acknowledge.

(c) Click the button or press the F2 key.

Alarm Log JULY 2010

Attribute Description

Page 309 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8.5 EVENT LOG
Page 309 · click to enlarge
Page 309

LOGS 8-9 MON2000

5. To acknowledge all unacknowledged alarms,

(a) Ensure you have selected the correct

display option (see Step 3).

(b) Click the button or press the F3 key.

In the Active Alarm view, click the

button or press the F4 key.

If an alarm is cleared before the condition has been resolved, MON2000 redisplays the alarm entry as an active alarm.

6. Click the button to exit and return to the main window.

8.4 CLEAR OR ACKNOWLEDGE ACTIVE ALARMS

To clear or acknowledge all alarms, click the Logs pull-down menu and select the Clear/Ack All Active Alarms option.

Emerson Process Management Chromatograph Software (MON2000) — 8.4 CLEAR OR ACKNOWLEDGE ACTIVE ALARMS diagram (page 309)
Fig. 269

8.5 EVENT LOG

Use this function to track the system and operator events that have occurred with a given GC unit.

Emerson Process Management Chromatograph Software (MON2000) — 8.5 EVENT LOG diagram (page 309)
Fig. 270

To view the Event Log,

1. Use the Logs > Event Log menu to access this function. MON2000 retrieves the last 50 events from the GC unit.

JULY 2010 Clear or Acknowledge Active Alarms

Page 310 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8-10 LOGS
Page 310 · click to enlarge
Page 310

8-10 LOGS

2. The Event Log dialog appears, displaying each system or operator event and the corresponding data.
Emerson Process Management Chromatograph Software (MON2000) — 8-10 LOGS diagram (page 310)
Fig. 271

Table 8-3 Event Log Attributes

User Id user name

Date date event occurred

Time time event occurred

Event Message indicates event type (i.e., describes event)

Old Value if applicable, the value before the event

New Value if applicable, the value after the event

3. Click the button to exit and return to the main window.

Event Log JULY 2010

Attribute Description

Page 311 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.1 OVERVIEW
Page 311 · click to enlarge
Page 311

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-1 MON2000

9 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

The Data Collection feature is only available with the MON2000 PLUS program.

9.1 OVERVIEW

The MON2000 PLUS program provides configurable automatic collection and storage of analysis and calibration data from the gas chromatograph controller.

Configuration of the specific chromatographs to be polled, timing of polling, and specific data to be collected from each is defined by the user through the setup of polling control files.

An Auto-Sequencing module interprets the commands in a polling control file and performs the collection and storage of data.

Data collected is stored in text files (*.txt) or comma delimited files (*.csv). The emphasis is on the data being in a form suitable for import by standard spreadsheet and/or database programs.

The MON2000 PLUS program is not designed to provide database capability for the data collected.

Reports collected are stored in text files (.txt) formatted as ready-to-print or in comma- delimited files (.csv) that reduce white space and the output file size.

JULY 2010 Overview

Page 312 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-2 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 312 · click to enlarge
Page 312

9-2 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

All Model 500 and 1000 GC’s with 2350A and older 2350/2360 GC Controllers and Model 700 GC Controllers are supported in the acquisition of the following types of data:

  • Alarms - Selection of Alarm Log, Active Alarms, or Unacknowledged Alarms.
  • Averages - Direct Access to single averages and/or blocks of averages to include support for acquiring results from most recent averaging period as well as those for previous averaging periods (number limited by number supported by specific model and application). Access to average value, maximum value, minimum value, and number of samples for the period to be supported.
  • CGM Archive - Selection of Last Analysis or Last Calibration and Stream number.
  • CGM on Alarm - Poll the CGM when an alarm condition is set or cleared.
  • Event Log - Selection of Start/End time, Most Recent n Days, Since Last Collected, or Entire Log.
  • Maintenance Log - Selection of Start/End Time, Most Recent n Days, Since Last Collected, or Entire Log
  • Registers - Both User Modbus and SIM_2251 single and multiple register queries are supported.
  • Reports - Archive data accessible via Reports>Archive Data>View Data menu are available (selection based on most recent number of days, date range, single stream or all streams). Selections are based on the various GC model and are consistent with the capabilities of those models and specific applications.

Overview JULY 2010

Page 313 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2 DATA COLLECTION CONFIGURATION
Page 313 · click to enlarge
Page 313

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-3 MON2000

  • Condition Start - Start of conditional polls. The poll commands following the Condition Start are run only when the condition is met. The condition is defined through the status change of a Modbus register (SIM_2251 or User_Modbus).
  • Condition End - End of conditional polls. The poll commands following the Condition End are run normally.

Use these Data Collection functions to configure a polling control file and start the Auto-Sequencing module:

  • Configuration Open File - Open or edit an existing Polling Control file.
  • Configuration New File - Create a new Polling Control file and use the Commands to configure the polling settings.
  • Start Auto - Initiate the Auto-Sequencing module. During Auto-Sequencing a Status display is maintained on the screen.

9.2 DATA COLLECTION CONFIGURATION

Use the Data Collection>Configuration>New File or Data Collection>Configuration>Open File menu to access the Configuration for Data Collection window and the commands to configure the Polling Control file. This function is performed in the Online or Offline mode.

A Polling Control file contains commands to be executed in order to acquire desired sets of data from one or more gas chromatographs.

JULY 2010 Data Collection Configuration

Page 314 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-4 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 314 · click to enlarge
Page 314

9-4 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

Select a desired command from the list of available commands and input the necessary data to complete the command (see Table 9-1 for a list of commands). The completed command is added to the Polling Control file. Continue with the next desired command. This process is continued until all commands are selected and configured for the file. Edit the polling control file by adding, deleting, and inserting commands.

Data Collection Configuration JULY 2010

Page 315 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-5 MON2000
Page 315 · click to enlarge
Page 315

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-5 MON2000

Table 9-1 List of Commands

Await This command instructs the Auto-Sequencing module to await a specific time of day prior to con- tinuing processing the commands in this file. The user enters the desired time of day in hours and minutes.

The Connect/Logon command instructs the Auto-Sequencing module to establish communications with the indicated GC controller using the communication method and parameters as indicated in the GC directory entry and logon to the GC Controller. After selection of the Connect command the user selects the desired GC controller from a list of the GC directory entries.

Another Poll Command is Connect/No Logon which allows the Auto-Sequencing module to execute without logging on to a GC Controller.

The Delay (Seconds) command instructs the Auto-Sequencing module to Delay for a specified interval of time prior to proceeding. The user enters the time in seconds.

The Delay (Hours) command instructs the Auto-Sequencing module to Delay for a specified interval of time prior to proceeding. The user enters the time in hours.

Disconnect This command instructs the Auto-Sequencing module to terminate communications with the con- troller to which it is currently connected. No additional user provided information is necessary for this command. If this command is executed and no GC Controller is connected, then the command is ignored.

End This command marks the end of the polling control file and signals the Auto-Sequencing module to stop processing.

Poll The poll command instructs the Auto-Sequencing module to acquire a set of data of a single type from the GC Controller to which it is currently connected. The user first selects the data type to be acquired.

  • Poll Data Types
  • Alarms
  • Averages
  • CGM Archive
  • CGM on Alarm
  • Event Log
  • Maintenance Log
  • Registers
  • Reports
  • Condition Start
  • Condition End

Repeat The repeat command instructs the Auto-Sequencing module to restart processing of the commands in the polling control file at the beginning of the file. The repeat command can be specified with a count that specifies the number of times to execute the sequence of commands in the file. Without the count (or with a count of zero) the sequence of commands will be re-executed indefinitely until terminated manually by the user.

Run The run command instructs the Auto-Sequencing module to start execution of a user program. The user specifies whether the Auto-Sequencing module waits for completion of the user program before processing the poll control file commands, or continues immediately thus executing in parallel with the user program. This feature allows you to further process the collected data. As an example, the user program might perform some validity checking on the data, reformat it, and then and store it in an Access database on the MON2000 PLUS PC.

JULY 2010 Data Collection Configuration

Show 5 more labels / short text on this page
Command Description · Connect/ Logon · Connect/No Logon · Delay (Seconds) · Delay (Hours)
Page 316 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-6 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 316 · click to enlarge
Page 316

9-6 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

9.2.1 Create a New Polling Control File

To create and configure a new polling control file,

1. Use the Data Collection > Configuration > New File menu to access this function.
2. MON2000 PLUS displays the Configuration for Data Collection window.
3. From the Polling Output File(s) table select Report and/or Data type by clicking the appropriate check box. Enter the Default Filename field for the chosen selections.
Emerson Process Management Chromatograph Software (MON2000) — 9-6 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 316)
Fig. 272
4. From the Commands pull-down list, select a command.
Emerson Process Management Chromatograph Software (MON2000) — 9-6 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 316)
Fig. 273
5. Enter the necessary data and a description for the command.
6. Use the add, Insert, Delete, and Poll Command buttons to add/edit/modify the configuration.
7. When you have finished adding/editing/ modifying the commands, click the Save button.

Create a New Polling Control File JULY 2010

Page 317 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.2 Open an Existing Polling Control File
Page 317 · click to enlarge
Page 317

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-7 MON2000

8. The Save Polling Control File dialog appears. Enter a filename and click the Save button to save the *.apc file.
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-7 MON2000 diagram (page 317)
Fig. 274

9.2.2 Open an Existing Polling Control File

To edit or modify the configuration of an existing file,

1. From the MON2000 PLUS main window, use the Data Collection>Configuration> Open File menu.
2. Click the button and the Open Polling Control File window displays.
Emerson Process Management Chromatograph Software (MON2000) — 9.2.2 Open an Existing Polling Control File diagram (page 317)
Fig. 275 · 9.2.2 Open an Existing Polling Control File
3. Select the desired polling control file, then

click the button.

JULY 2010 Open an Existing Polling Control File

Page 318 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-8 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 318 · click to enlarge
Page 318

9-8 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

4. The Configuration for Data Collection window displays.
Emerson Process Management Chromatograph Software (MON2000) — 9-8 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 318)
Fig. 276
5. Use the buttons to modify the Data Collection configuration or Polling Control file.
6. When you have finished modifying the file, click the button to apply the changes.
7. The Save Polling Control File dialog appears.

Open an Existing Polling Control File JULY 2010

Page 319 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.3 Await Command
Page 319 · click to enlarge
Page 319

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-9 MON2000

8. A Save Polling Control message prompts you to replace the exiting file or rename the Polling Control file (*.apc) file. Click the

button to replace the existing file.

9. Click the button to discard your changes.

9.2.3 Await Command

This command instructs the auto-sequencing module to wait for a specific time of day prior to processing the commands in the file. The user enters the desired time of day in hours and minutes.

To configure the Await Command,

1. From the General Configuration Commands pull-down list select the Await command.
Emerson Process Management Chromatograph Software (MON2000) — 9.2.3 Await Command diagram (page 319)
Fig. 277 · 9.2.3 Await Command
2. Press the RIGHT ARROW key on your keyboard or LEFT-CLICK the mouse in the Time field and enter the time.
3. Press the RIGHT ARROW key on your keyboard or LEFT-CLICK the mouse in the Command Description field. Enter a description for the Await command in the Command Description field.

JULY 2010 Await Command

Or,

Page 320 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-10 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 320 · click to enlarge
Page 320

9-10 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

9.2.4 Connect/Logon Command

The Connect/Logon command instructs the Auto-Sequencing module to establish communications with the indicated GC controller, using the communication method and parameters as indicated in the GC directory entry and logon to the GC Controller. After selection of the Connect command the user selects the desired GC controller from a list of the GC directory entries.

Use the communication method and parameters as configured in the File>GC Directory entry.

To configure the Connect/Logon Command,

1. Select Connect/Logon from the General Configuration Commands pull-down list.
Emerson Process Management Chromatograph Software (MON2000) — 9-10 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 320)
Fig. 278
2. From the GC Entry list that is configured in the File>GC Directory menu, select the desired GC Controller.
3. Enter a description for the Connect/Logon command in the Command Description field.

Connect/Logon Command JULY 2010

Page 321 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.6 Delay (Seconds) Command
Page 321 · click to enlarge
Page 321

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-11 MON2000

9.2.5 Connect/No Logon Command

The Poll Command Connect/No Logon allows the Auto-Sequencing function to execute without logging on to a GC Controller.

To configure the Connect/No Logon Command,

1. Select Connect/No Logon from the General Configuration Command pull-down list.
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-11 MON2000 diagram (page 321)
Fig. 279
2. From the GC Entry list that is configured in the File>GC Directory menu, select the desired GC Controller.
3. Enter a description for the Connect/No Logon command in the Command Description field.

9.2.6 Delay (Seconds) Command

Prior to executing the auto-polling function, the Delay (Seconds)Command instructs the Auto-Sequencing module to delay for a specified interval of time. The user enters the time delay in one second increments.

JULY 2010 Connect/No Logon Command

Page 322 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.7 Delay (Hours) Command
Page 322 · click to enlarge
Page 322

9-12 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

To configure the Delay (Seconds) Command, 1. From the General Configuration List of Commands pull-down list, select Delay (Seconds).

Emerson Process Management Chromatograph Software (MON2000) — 9-12 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 322)
Fig. 280
2. Enter the time delay, in one second increments, in the Delay field.
3. Enter a description for the Delay (Seconds) command in the Command Description field.

9.2.7 Delay (Hours) Command

Prior to executing the auto-polling function, the Delay(Hours)Command instructs the Auto- Sequencing module to delay for a specified interval of time. The user enters the time delay in one hour increments.

To configure the Delay (Hours) Command,

1. From the General Configuration List of Commands pull-down list, select Delay (Hours) .
Emerson Process Management Chromatograph Software (MON2000) — 9.2.7 Delay (Hours) Command diagram (page 322)
Fig. 281 · 9.2.7 Delay (Hours) Command
2. Enter the time delay, in one hour increments, in the Delay field.
3. Enter a description for the Delay (Hours) command in the Command Description field.

Delay (Hours) Command JULY 2010

Page 323 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.9 End Command
Page 323 · click to enlarge
Page 323

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-13 MON2000

9.2.8 Disconnect Command

Use the Disconnect Command to terminate communications with a currently connected GC Controller.

To configure the Disconnect command,

1. From the General Configuration: List of Commands, use the Commands pull-down list and select Disconnect .
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-13 MON2000 diagram (page 323)
Fig. 282
2. Enter a description for the Disconnect Command in the Command Description field.

No additional input parameters are required for this command. Also note, if you are not connected to a GC Controller, this command is ignored.

9.2.9 End Command

The End Command marks the end of the polling control file and signals the Auto-Sequencing module to stop processing.

1. From the Configuration for Data Collection window, click the button to select another command.
2. Use the Command pull-down list, and select End .
Emerson Process Management Chromatograph Software (MON2000) — 9.2.9 End Command diagram (page 323)
Fig. 283 · 9.2.9 End Command
3. Enter a description for the End Command in the Command Description field.

JULY 2010 Disconnect Command

Page 324 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-14 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 324 · click to enlarge
Page 324

9-14 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

9.2.10 Poll Command: Alarms

To configure the Alarms Command, from the Configuration for Data Collection window,

Each Poll Command includes the criteria to define acquisition and storage of a single type of data.

1. Select POLL from the General Configuration Commands pull-down list.
Emerson Process Management Chromatograph Software (MON2000) — 9-14 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 324)
Fig. 284
2. Press the RIGHT ARROW key on your keyboard or LEFT-CLICK the mouse in the Type field and select Alarms from the pulldown list.
3. In the Command Description field, enter a description for the Alarms command.
4. Click the button. The Poll Command: Alarms dialog box displays.
Emerson Process Management Chromatograph Software (MON2000) — 9-14 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 324)
Fig. 285
5. To select Alarm Log, click the Alarm Log check box. RIGHT ARROW key on your keyboard or LEFT-CLICK the mouse in the Time Selection field.
Poll CommandAlarms JULY 2010
Page 325 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-15 MON2000
Page 325 · click to enlarge
Page 325

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-15 MON2000

6. Use the Time Selection pull-down list and choose one of the following:
  • Start/End Time
  • Most Recent
  • Since Last Collected
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-15 MON2000 diagram (page 325)
Fig. 286

Use the Up and Down arrows and enter the starting date, starting time, ending date and ending time in the appropriate data fields.

7. If Active Alarms or Unacknowledged Alarms is selected, no additional information is necessary and all of the entries of the selected type are returned.
8. From the Output File table, enter/edit a filename and file format for the Alarms Output file.
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-15 MON2000 diagram (page 325)
Fig. 287
9. Click the button to apply the selections.

MON2000 PLUS displays the Save Changes dialog.

Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-15 MON2000 diagram (page 325)
Fig. 288
JULY 2010 Poll CommandAlarms
Page 326 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.11 Poll Command: Averages
Page 326 · click to enlarge
Page 326

9-16 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

10.Click the button to save the changes and return to the Configuration for Data Collection window.

Click the button to discard your changes.

9.2.11 Poll Command: Averages

To configure the Averages Command, to acquire sets of one or more averages for the most recent averaging period and/or the previous averaging period (any available in the GC Controller),

MON2000 PLUS reads all the Averages configuration (set up from the Applications>Calculatioins>Averages menu) if connected to GC controller.

If you are using a Direct Connection (rather than an Ethernet Connection), the reading process could take a long time.

1. From the Configuration for Data Collection window, select the Poll command.
Emerson Process Management Chromatograph Software (MON2000) — 9.2.11 Poll Command: Averages diagram (page 326)
Fig. 289 · 9.2.11 Poll Command: Averages
2. Press the RIGHT ARROW key or click in the Type field, then select Averages from the pull-down list.
3. Press the RIGHT ARROW key on your keyboard, or left-click the mouse in the Command Description field. Enter a description for the Averages command.
Poll CommandAverages JULY 2010

Or,

Page 327 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-17 MON2000
Page 327 · click to enlarge
Page 327

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-17 MON2000

4. Click the button and the Poll Command: Averages dialog box displays.
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-17 MON2000 diagram (page 327)
Fig. 290
5. From the Averages Configuration dialog, for a single record, select Average - Single Record . If the Average - Single Record is selected, only the average values are included in the output file.
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-17 MON2000 diagram (page 327)
Fig. 291
JULY 2010 Poll CommandAverages

Or,

Page 328 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-18 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 328 · click to enlarge
Page 328

9-18 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

If Complete Output Complete Output is selected, click the check boxes to select Average, Min, Max, or Samples values.

All averages referenced by a single Poll command must be for the same averaging interval (e.g. hourly, daily, weekly, monthly, variable).

Emerson Process Management Chromatograph Software (MON2000) — 9-18 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 328)
Fig. 292

Use the Result Grouping pull-down list and select By Period or By Average.

If the number of averages is greater than one, then select whether the results are to be grouped by average, or grouped by period.

(a) When grouped by Average, the data from

all of the periods of a given average are stored in a single record in the output file with the appropriate time-stamp included with the data for each period. (b) When grouped by Period, a record for

each period contains the values from each of the averages for that period and a single time stamp value is included in the record.

Poll CommandAverages JULY 2010
Page 329 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.12 Poll Command: CGM Archive
Page 329 · click to enlarge
Page 329

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-19 MON2000

6. From the Output File table, enter/edit the filename and file format (*.txt or *.csv) for the Averages Output file.
7. Click the button to apply the configuration changes.
8. MON2000 PLUS displays the Save Changes dialog.
9. Click the button to write the changes and return to the Configuration for Data Collection window.

Click the button to discard your changes.

9.2.12 Poll Command: CGM Archive

Use the CGM Archive command to poll either the Last Analysis or Last Calibration CGM Archive for a specific stream. The output file is in *.cgm format and the file name is based on the Analyzer Name, Stream, and Date/Time. The CGM file is stored in the "\GC\SAVE\CGM Archive" folder.

To configure the CGM Archive Command,

1. From the Configuration for Data Collection window, click the ADD button to select another command.
2. Next, select the Poll command.
Emerson Process Management Chromatograph Software (MON2000) — 9.2.12 Poll Command: CGM Archive diagram (page 329)
Fig. 293 · 9.2.12 Poll Command: CGM Archive
3. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse in the Type field, then select CGM Archive.
JULY 2010 Poll CommandCGM Archive

Or,

Page 330 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-20 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 330 · click to enlarge
Page 330

9-20 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

4. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse in the Command Description field. Enter a description for the CGM Archive command.
5. Click the button and the Poll Command: CGM Archive dialog box appears.
Emerson Process Management Chromatograph Software (MON2000) — 9-20 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 330)
Fig. 294
6. Enter the Stream number for data collection.
7. Use the Analysis/Calibration pull-down list and select the CGM Archive for the Last Analysis or Last Calibration.
8. Click the button to apply the configuration changes. MON2000 PLUS returns you to the Configuration for Data Collection window.
9. Click the button to discard your changes.
Poll CommandCGM Archive JULY 2010

Or,

Page 331 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.14 Poll Command: Condition Start
Page 331 · click to enlarge
Page 331

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-21 MON2000

9.2.13 Poll Command: CGM on Alarm

Use the CGM on Alarm command to poll the CGM when an alarm condition is set or cleared. Setup continuous polls so that MON2000 PLUS can determine and retrieve the CGM with the alarm condition. The data is saved to an output file in CGM format and is stored in the "\GC\SAVE\ERRORS" folder. The file name is based on the Date/Time and ends with an 'a'.

To configure the CGM on Alarm Command,

1. From the Configuration for Data Collection

window, click the button to select another command.

Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-21 MON2000 diagram (page 331)
Fig. 295
2. Next, select the Poll command.
3. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse in the Type field, then select CGM on Alarm.
4. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse in the Command Description field. Enter a description for the CGM on Alarm command. No further details are needed.

9.2.14 Poll Command: Condition Start

Use the Condition Start command to start conditional polling.

To configure the Condition Start command,

1. From the General Configuration (List of Commands) table, select Poll from the Command column.
JULY 2010 Poll CommandCGM on Alarm
Page 332 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-22 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 332 · click to enlarge
Page 332

9-22 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

Emerson Process Management Chromatograph Software (MON2000) — 9-22 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 332)
Fig. 296
2. Select Condition Start from the pull-down list in the Type column.
3. Enter a description in the Command Description column.
4. Select a default output file format by clicking either the Report or Data checkbox on the Polling Output File(s) table.
5. Enter a default output file name.
6. Click the Poll Command Details button located at the bottom of the Configuration for Data Collection window. The Poll Command: Condition Start window appears.
Emerson Process Management Chromatograph Software (MON2000) — 9-22 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 332)
Fig. 297
7. Select Register/SIM_2251 or Register/ User_Modbus from the Condition Data Type column.
8. Enter the register number in the Condition Data Value column.
9. Select Change from the Condition Status column.
Poll CommandCondition Start JULY 2010
Page 333 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.15 Poll Command: Condition End
Page 333 · click to enlarge
Page 333

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-23 MON2000

10.To apply the configuration changes, click OK. To discard your changes and return to the Configuration for Data Collection window, click Cancel.

9.2.15 Poll Command: Condition End

Use the Condition End command to stop conditional polling.

To configure the Condition End command,

1. From the General Configuration (List of Commands) table, select Poll from the Command column.
Emerson Process Management Chromatograph Software (MON2000) — 9.2.15 Poll Command: Condition End diagram (page 333)
Fig. 298 · 9.2.15 Poll Command: Condition End
2. Select Condition End from the pull-down list in the Type column.
3. Enter a description in the Command Description column.
JULY 2010 Poll CommandCondition End
Page 334 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-24 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 334 · click to enlarge
Page 334

9-24 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

9.2.16 Poll Command: Event Log

1. From the General Configuration: List of Commands, use the Commands pull-down list and select the POLL command.
Emerson Process Management Chromatograph Software (MON2000) — 9-24 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 334)
Fig. 299
2. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse Type field and select Event LOG from the pulldown list.
3. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse in the Command Description field. Enter a description for the Event Log command.
4. Click the button and the Poll Command: Event Log dialog box displays.
Emerson Process Management Chromatograph Software (MON2000) — 9-24 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 334)
Fig. 300
5. From the Event Log Selection pull-down list, choose one of the following:

Event Log Selection Additional Selections

Start/End Time Start Date

Poll CommandEvent Log JULY 2010

Start Time

Page 335 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-25 MON2000
Page 335 · click to enlarge
Page 335

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-25 MON2000

Event Log Selection Additional Selections

Most Recent n Days Enter number of n Days

Since Last Collected No further information is needed

Entire Log No further information is needed

6. For each entry in the Event Log which satisfies the selection criteria, a record containing the following information is stored:

- User ID of user initiating the change (or SYSTEM) - Date/time of occurrence - Description of the event - Parameter value prior to the event (old value) - Parameter value after the event (new value)

7. From the Output File table, enter/edit the filename and file format (*.txt or *.csv) for the Event Log Output file.
8. Click the button to apply the configuration changes.
9. MON2000 PLUS displays the Save Changes dialog.

10.Click the button to write the changes and return to the Configuration for Data Collection window.

JULY 2010 Poll CommandEvent Log

Ending Date · Ending Time · Or,

Page 336 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.17 Poll Command: Maintenance Log
Page 336 · click to enlarge
Page 336

9-26 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

Click the button to discard your changes and return to the Configuration for Data Collection window.

9.2.17 Poll Command: Maintenance Log

To configure the Maintenance Log command,

1. From the General Configuration: List of Commands, use the Commands pull-down list and select the POLL command.
Emerson Process Management Chromatograph Software (MON2000) — 9.2.17 Poll Command: Maintenance Log diagram (page 336)
Fig. 301 · 9.2.17 Poll Command: Maintenance Log
2. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse Type field and select Maintenance Log from the pull-down list.
3. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse in the Command Description field. Enter a description for the Maintenance Log command.
Poll CommandMaintenance Log JULY 2010
Page 337 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-27 MON2000
Page 337 · click to enlarge
Page 337

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-27 MON2000

4. Click the button and the Poll Command: Maintenance Log dialog box displays.
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-27 MON2000 diagram (page 337)
Fig. 302
5. From the Maintenance Log Configuration Selection table, choose one of the following:

Maintenance Log Selection Additional Selections

  • Start/End Time
  • Start Date
  • Start Time
  • End Date
  • End Time

Most Recent n Days • Enter number of n Days

Since Last Collected • No further information is needed

Entire Log • No further information is needed

6. From the Output File table, enter/edit the filename and file format (*.txt or *.csv) for the Maintenance Log Output file.
7. Click the button to apply the configuration changes.
JULY 2010 Poll CommandMaintenance Log

Or,

Page 338 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.18 Poll Command: Registers
Page 338 · click to enlarge
Page 338

9-28 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

8. Click the button to discard your changes and return to the Configuration for Data Collection window.

9.2.18 Poll Command: Registers

Blocks of Modbus Register data can be specified for collection.

To configure the Registers Command,

1. From the General Configuration: List of Commands, use the Commands pull-down list and select the POLL command.
Emerson Process Management Chromatograph Software (MON2000) — 9.2.18 Poll Command: Registers diagram (page 338)
Fig. 303 · 9.2.18 Poll Command: Registers
2. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse Type field and select Registers from the pulldown list.

MON2000 PLUS reads all the configured Registers associated with the Modbus Type (Sim_2251 or User_Modbus) if you are connected to a GC Controller.

3. Press the RIGHT ARROW key on your keyboard, or LEFT-CLICK the mouse in the Command Description field. Enter a description for the Registers command.

If you are using a direct connection (rather than an Ethernet connection), the reading process could take a long time.

When initially building a “polling control” file, you may want to use the MON2000 PLUS “Offline Edit” function with an “as configured” version of the application open to avoid these delays.

Poll CommandRegisters JULY 2010
Page 339 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-29 MON2000
Page 339 · click to enlarge
Page 339

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-29 MON2000

4. Click the button and the

Poll Command: Registers dialog appears.

Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-29 MON2000 diagram (page 339)
Fig. 304

The basic configuration includes the following:

  • Modbus Type - SIM_2251 or USER MODBUS
  • Data Type Register or Coil
  • Starting Register/Coil
  • Number of Registers/ Coils
JULY 2010 Poll CommandRegisters
Page 340 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-30 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 340 · click to enlarge
Page 340

9-30 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

(a) If SIM_2251 is selected and the Starting

Register is in the range 701 - 999, then Record Selection includes the following:

Record Selection Additional Selections

Starting Record + n Starting Record number

Number of Records (n)

Total Number of Records

Most Recent n Records Number of Records (n)

Total Number of Records

Location of Last Written Record

Since Last Collection Total Number of Records

Location of Last Written Record

(b) If USER MODBUS is selected, and Data

Type is Register, then the user must enter a Data Format string to define for each parameter being requested (an integer 16-bit or floating-point 32-bit value). Floating-point values use two consecutive registers for storage.

Emerson Process Management Chromatograph Software (MON2000) — 9-30 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 340)
Fig. 305
Poll CommandRegisters JULY 2010

Date/Time Format

Page 341 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-31 MON2000
Page 341 · click to enlarge
Page 341

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-31 MON2000

(c) For the standard registers, the values

returned are stored in a single record in the output file, following the Date and Time when the registers are polled, the Starting Register, and Number of Registers. (d) For SIM_2251 registers 701-999, one

record is stored in the output file for each record retrieved with fields within the record, starting the date and time when the record is polled, starting Register (701-999), Record Number, and the data for the record.

5. From the Output File table, enter/edit the filename and file format (*.txt, *.dat or *.csv) for the Registers Output file. The DAT file format is for customized chroma data.
6. Click the button to apply the configuration changes.
7. MON2000 PLUS displays the Save Changes dialog.
8. Click the button to write the changes and return to the Configuration for Data Collection window.
9. Click the button to discard your changes and return to the Configuration for Data Collection window.
JULY 2010 Poll CommandRegisters

Or,

Page 342 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-32 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 342 · click to enlarge
Page 342

9-32 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

9.2.19 Poll Command: Reports

The archived analysis reports (as viewable under MON2000 when connected to a controller, from the Reports>Archived Data>View Data menu) can be retrieved and stored.

To configure the Reports Command,

1. From the Configuration for Data Collection window, click the ADD button to select another command.
Emerson Process Management Chromatograph Software (MON2000) — 9-32 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 342)
Fig. 306
2. Next, select the Poll command.
3. Press the RIGHT ARROW on your keyboard, or LEFT-CLICK the mouse in the Type field, then select Reports from the pull-down list.
4. Press the right arrow on your keyboard, or left-click the mouse in the Command Description field. Enter a description for the Reports command.
Poll CommandReports JULY 2010
Page 343 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-33 MON2000
Page 343 · click to enlarge
Page 343

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-33 MON2000

5. Click the button and the Poll Command: Reports dialog box appears.
Emerson Process Management Chromatograph Software (MON2000) — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-33 MON2000 diagram (page 343)
Fig. 307
6. Use the following selection criteria and then enter appropriate data:

Reports Selection Additional Selections

  • Start/End Time
  • Stream Number (0 - indicate all streams)
  • Start Date
  • Start Time
  • End Date
  • End Time
  • Most Recent n Days
  • Stream Number (0 - indicate all streams)
  • Enter number of n Days

Runs Since Last Collection • Stream Number (0 - indicate all streams)

7. Regardless of the selection criteria method used above, choose the order of returned reports in the Output Order (with Most Recent First or with Oldest First).
8. Next, select the Output Type, as a Complete Report or an Analysis - Single Record.
JULY 2010 Poll CommandReports
Page 344 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.2.20 Repeat Command
Page 344 · click to enlarge
Page 344

9-34 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

9. From the Output File table, enter/edit the filename and file format (*.txt or *.csv) for the Reports Output file.

10.Click the button to apply the configuration changes.

11.Click the button to discard your changes and return to the Configuration for Data Collection window.

9.2.20 Repeat Command

Use the Repeat command to instruct the Auto- Sequencing module to restart processing of the commands in the polling control file (from the beginning of the file). Specify a count for the number of times to execute the sequence of commands in the file. Without a specified count (or with a count of zero) the sequence of commands will be re-executed indefinitely until terminated manually by the user.

To configure the Repeat command,

1. From the General Configuration List of Commands pull-down list, select Repeat .
Emerson Process Management Chromatograph Software (MON2000) — 9.2.20 Repeat Command diagram (page 344)
Fig. 308 · 9.2.20 Repeat Command
2. Enter the number of times to repeat the Auto-Sequencing process, in the Repeat Count field.
3. Enter a description for the Repeat Command in the Command Description field.

Repeat Command JULY 2010

Or,

Page 345 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.3 DATA COLLECTION
Page 345 · click to enlarge
Page 345

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-35 MON2000

9.2.21 Run Command

The run command (future release) instructs the Auto-Sequencing module to start execution of a user program. The user specifies whether the Auto-Sequencing module waits for completion of the user program before processing the poll control file commands, or continues immediately thus executing in parallel with the user program. This feature allows you to further process the collected data. As an example, the user program might perform some validity checking on the data, reformat it, and then and store it in an Access database on the MON2000 PLUS PC.

9.3 DATA COLLECTION

The Data Collection Start Auto function allows you to open an existing polling control file, select to append or overwrite the existing output files, select whether to enable Transaction Logging, and start the automatic data collection.

JULY 2010 Run Command

Page 346 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-36 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 346 · click to enlarge
Page 346

9-36 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

9.3.1 Start Auto-Sequencing

To initiate Auto-Sequencing, use the Data Col- lection>Start Auto menu,

1. Click START AUTO and the Start AutoSequencing dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 9-36 MON2000 PLUS DATA COLLECTION/AUTO-POLLING diagram (page 346)
Fig. 309
2. Click the button and the Open Polling Control File dialog appears.
3. Select the desired polling control file, and

then click the button.

4. From the Output Files selection, click the desired radio button:
  • Append to the existing file
  • Overwrite the existing file
5. In the check box, select Enable Transaction Logging (if desired).

Start Auto-Sequencing JULY 2010

Page 347 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-37 MON2000
Page 347 · click to enlarge
Page 347

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-37 MON2000

6. Click the button to confirm the start of automatic operation. MON2000 PLUS begins the Auto-Sequencing process and displays the Auto-Sequencing Status window.
7. The Auto-Sequencing module begins with the first command in the selected polling control file and performs the requested operation. Processing continues sequentially through the file performing the operations requested by each command.

(a) If a Repeat command is encountered,

then the processing is restarted at the beginning of the file. (b) If an End command (or the end-of-file) is

encountered, the module ends the command processing.

JULY 2010 Start Auto-Sequencing

Page 348 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.3.3 Auto-Sequencing Status
Page 348 · click to enlarge
Page 348

9-38 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

9.3.2 Transaction Log

To enable the Transaction Logging functions, select the check box on the Start Auto- Sequencing dialog. If this option is chosen, one or more entries are made to the AutoPollCommsLog.txt file each time a command is executed.

Transaction Log entries include a time stamp and a description of the command being executed. If the command results in Modbus messages being exchanged between the PC and the controller, the poll message from the PC and the response from the controller are recorded.

9.3.3 Auto-Sequencing Status

During Auto-Sequencing a status display is maintained on the screen.

Emerson Process Management Chromatograph Software (MON2000) — 9.3.3 Auto-Sequencing Status diagram (page 348)
Fig. 310 · 9.3.3 Auto-Sequencing Status

Transaction Log JULY 2010

Page 349 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9.3.4 Status Log
Page 349 · click to enlarge
Page 349

MON2000 PLUS DATA COLLECTION/AUTO-POLLING 9-39 MON2000

As each command is processed the following general information is displayed (depending on the command being processed, the details of the exact data included will vary:

  • Polling Control file in use
  • Start time of operation
  • Command being processed
  • Description of command being processed
  • Status of completion of the command (level detail appropriate to the individual command)
  • Next Command to be processed
  • Status Log

Auto-Sequencing must be stopped before other MON2000 PLUS menu selections can be chosen.

Terminate Auto-Sequencing at any time by clicking the Abort button on the Auto- Sequencing status display screen. If this occurs, the Auto-Sequencing module will close all open output files and return to normal operation of the MON2000 PLUS program.

9.3.4 Status Log

When MON2000 PLUS detects connection or polling errors, the error condition is logged and a Status Log file is saved as: AutoPollStatusLog.txt.

The Status Log field on the Auto-Sequencing Status dialog indicates the error condition.

JULY 2010 Status Log

Page 350 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 9-40 MON2000 PLUS DATA COLLECTION/AUTO-POLLING
Page 350 · click to enlarge
Page 350

9-40 MON2000 PLUS DATA COLLECTION/AUTO-POLLING

This page is intentionally left blank.

Status Log JULY 2010

Page 351 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.1 STARTING WINMB
Page 351 · click to enlarge
Page 351

MODBUS TEST 10-1 MON2000

10 MODBUS TEST

The Modbus Test Program (WinMB) enables you to test the operation of the serial port com- munications, and determine Modbus register and log contents. Use this program as an aid to software debugging or for special installations. Via WinMB, you can troubleshoot with any device including the GC Controller, an ultrasonic meter, or a flow computer.

FUNCTION NOT REQIURED FOR NORMAL GC OPERATION

The Modbus test is reserved for advanced functions

The Modbus Test function is not required for normal GC operation. Skip this section unless you are developing software, engaging in a software debugging process, or designing a custom installation that directly accesses the GC Controller Modbus registers.

Traditionally, Modbus registers are polled by using a data collection system (DCS). To facilitate installation and debugging, the WinMB program emulates a DCS.

Emerson Process Management Chromatograph Software (MON2000) — FUNCTION NOT REQIURED FOR NORMAL GC OPERATION diagram (page 351)
Fig. 311

This section provides detailed instructions for using the WinMB program. Use this program only if you are familiar with Modbus communication protocol and the operation of the MON2000 software.

10.1 STARTING WINMB

To start WinMB from MON2000, pull down the File menu and click the Modbus Test option.

JULY 2010 Starting WinMB

Page 352 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.2 ESTABLISHING COMMUNICATIONS
Page 352 · click to enlarge
Page 352

10-2 MODBUS TEST

If MON2000 displays an error, verify the installation directory via the Program Settings function (see Section 2.16.2).

The Modbus Test Program main window appears.

Emerson Process Management Chromatograph Software (MON2000) — 10-2 MODBUS TEST diagram (page 352)
Fig. 312

10.2 ESTABLISHING COMMUNICATIONS

Use the following instructions to configure the desired communications settings.

If you are accessing WinMB via MON2000, WinMB defaults to the same communications and serial port settings as those set for MON2000. To set up a WinMB-specific configuration, see Section 10.2.3.

Establishing Communications JULY 2010

Page 353 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS TEST 10-3 MON2000
Page 353 · click to enlarge
Page 353

MODBUS TEST 10-3 MON2000

10.2.1 Comparison of Modbus Protocols

MON2000 software and the WinMB program accommodate two different Modbus protocols: SIM_2251 and User_Modbus. For each protocol, separate Modbus registers are reserved.

Thus some settings for MON2000 and WinMB depend on which Modbus protocol is used.

The protocol you need depends, ultimately, on the hardware used for data acquisition from the GC Controller Modbus register contents.

The following comparison should help clarify the differences between these protocols as well as the utility of each.

Table 10-1 Comparison of SIM_2251 and User_Modbus Protocols

serial slave port serial slave port

modified protocol that allows floating point numbers to be transmitted over Modbus via 2251 emulation slave type

standard Gould protocol that accommodates PLC Emulation LO-HI (PLC-LH)

most register contents are predefined; some registers can be user-defined

predefined Boolean (coils) user-defined Numeric (registers)

data types are predefined for registers 1000 to 9000 data types are user-defined

variables assigned to registers can be listed in the PC Config Report (for instructions, see Section 2.16.3; for an example report, see Appendix A). See Appendix F for more detail about individual registers.

variables assigned to registers can be listed in the PC Config Report (for instructions, see Section 2.16.3; for example report, see Appendix A)

When using WinMB, set Register Mode to “DANIEL” to view register contents.

When using WinMB, set Register Mode to “PLC- LH” or “PLC-HL” to view register contents.

It is not necessary to assign scales to regis- ters.

It may be necessary to assign scales to regis- ters, to convert floating point values to whole integer representations.

JULY 2010 Comparison of Modbus Protocols

SIM_2251 User_Modbus

Page 354 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10-4 MODBUS TEST
Page 354 · click to enlarge
Page 354

10-4 MODBUS TEST

10.2.2 Set GC Com Parameters via MON2000

Set the communication parameters for the GC Controller serial port to which your PC, or other device, is connected.

To determine or reset the communications parameters at the GC Controller,

1. Use the Application > Serial Ports menu to access the Serial Ports function. The Serial Ports dialog appears.
2. Verify the MON2000 serial port Usage setting.

(a) Determine the GC Controller serial port

to which you are directly connected. (b) In the Serial Ports dialog, ensure that

the Usage setting for the port is either “SIM-2251” or “User_Modbus”.

Consult accurate documentation of the GC installation or visually inspect the serial line connection at the GC Controller Terminal Board.

If the setting is incorrect, change it by using the provided combo box.

3. Verify the Baud Rate setting.

WinMB accepts the following baud rates: 300, 1200, 2400, 4800, 9600, 14400, 19200, 38400, 57600, 115200.

4. Ensure the remaining serial port parameters are correct. Revise as required and record for future reference.

Set GC Com Parameters via MON2000 JULY 2010

Page 355 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.2.3 Set Up Port via WinMB
Page 355 · click to enlarge
Page 355

MODBUS TEST 10-5 MON2000

Typical settings are:

Table 10-2 Typical GC Communications Parameters

Parameter RTU Setting ASCII Setting

Data Bits 8 7

Stop Bits 1 1

RTS On 0 0

RTS Off 0 0

Ptrcl RTU ASCII RTS On/Off settings are required only when line settling is needed before data transmission begins.

5. Check the Com ID setting. The Com ID is the slave address defined by MON2000.
6. Verify that the RW setting is either “R” or “RW”.

Record the true Device ID number for future use.

7. Click the button to accept your input and return to the main screen.

10.2.3 Set Up Port via WinMB

To determine or reset the communications parameters used by WinMB,

JULY 2010 Set Up Port via WinMB

Parity None Even · Handshake None None

Page 356 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10-6 MODBUS TEST
Page 356 · click to enlarge
Page 356

10-6 MODBUS TEST

1. Click the button to access the Port Setup function. The Port Setup dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 10-6 MODBUS TEST diagram (page 356)
Fig. 313
2. Define the WinMB serial port setup options.

Table 10-3 Typical WinMB Port Parameters

The serial port settings displayed are for the serial communication parameters at the PC. These settings must match the GC Controller settings (to verify, see Section 10.2.2).

Port COM1 or COM2

Data Bits 7 or 8

Parity Even or None

For direct connection to the GC Controller, ensure that the Port setting is the same as the Com ID number of the serial port used.

Control Options Read Timeout, 500 ms Try, 2

Register Mode Daniel (for SIM_2251)PLC-LH (for User_Modbus) PLC-LH

Protocol ASCII Modbus RTU Modbus

3. Click the button to accept your input and return to the main screen.

Set Up Port via WinMB JULY 2010

Show 4 more labels / short text on this page
Parameter Typical Setting · Baud Rate 9600 · Stop Bits 1 · Flow Control None
Page 357 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS TEST 10-7 MON2000
Page 357 · click to enlarge
Page 357

MODBUS TEST 10-7 MON2000

10.3 GETTING MODBUS DATA

Use the following steps to read or write register contents to the GC Controller (or any other device). See the following subsections for additional details.

To access previous settings, use the spin arrows to view and select.

Before retrieving data, print a PC Config Report (see Section 2.16.3) and check the Communication data for variable names assigned to the Modbus registers.

Emerson Process Management Chromatograph Software (MON2000) — MODBUS TEST 10-7 MON2000 diagram (page 357)
Fig. 314
1. In the Slave Addr data field, type the Comm ID of the GC unit (or other device). WinMB will accept a slave address of 1 to 247.

Enter “0” to use the broadcast mode. In the broadcast mode, WinMB polls all known devices. Each device interprets this message as an instruction to read and take action; however, a response message may not be received by WinMB.

Any change is applied to the corresponding register value at each device. Table 10-4 shows which function codes support the broadcast mode.

JULY 2010 Getting Modbus Data

Page 358 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10-8 MODBUS TEST
Page 358 · click to enlarge
Page 358

10-8 MODBUS TEST

2. Use the Function pull-down menu to select the desired read or write option.

Table 10-4 Available Modbus Function Codes

Function Code Description Broadcast

1 (Read Coil) read one or more coil values

3 (Read Reg) read one or more register values

5 (Set Single Coil) set (write) one coil value

5 (Set Single Reg) set (write) one register value

15 (Set Multiple Coils) set (write) multiple coil values

16 (Set Multiple Regs) set (write) multiple register values

3. In the Data Addr data field, type the starting register.

To ensure best data type assignments, review a PC Config Report (see Section 2.16.3 to view a live report; see Section 7.1 to view a report file; see Appendix A for a sample report).

Table 10-5 Default Data Type per Register Range for SIM_2251

Register Range Default Type

1000 – 2999 Boolean

3000 – 4999 Integer

5000 – 6900 Long

7000 – 8999 Float

To select data types for SIM_2251 registers 1000 to 9000, see Section 10.3.2.

4. In the Quantity data field, type the number of registers to be retrieved (range from 1 to 2016).

Getting Modbus Data JULY 2010

Page 359 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.3.1 Use Single Data Type
Page 359 · click to enlarge
Page 359

MODBUS TEST 10-9 MON2000

Boolean registers are not user-defined (for either SIM_2251 or User_Modbus) and primarily contain alarm flags useful for debugging. To view the contents of Boolean registers, select the “1 (Read Coil)” function code.

Also note that in Standard Modbus mode each register is 16 bits. Therefore, integers (SHORT) consist of 1 register while floats (FLOAT) and long integers (LONG) consist of 2 registers.

Numeric registers for User_Modbus can be user- defined (see Section 5.18.4). To view the contents of Numeric registers, select the “3 (Read Regs)” function code.

5. Type the desired repeat count (how many times WinMB will read or set the specified registers before ceasing transmission) in the Repeat data box.

You can repeat the poll from 1 to 9999 times. A Repeat value of “–1” produces an infinite polling loop that can be terminated

by clicking the button.

10.3.1 Use Single Data Type

Use this function to assign a data type to a group of registers you will read or edit.

To poll a group of registers with the same user- defined data type,

1. Follow Steps 1 through 5 in Section 10.3 .
2. Click the Use to decode registers radio button to choose this option.
Emerson Process Management Chromatograph Software (MON2000) — 10.3.1 Use Single Data Type diagram (page 359)
Fig. 315 · 10.3.1 Use Single Data Type
3. Use the Data Type pull-down menu to assign a data type to the selected registers ( see Section 10.3 , Steps 3 and 4).

JULY 2010 Use Single Data Type

Page 360 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.3.2 Use Template (Mixed Data Types)
Page 360 · click to enlarge
Page 360

10-10 MODBUS TEST

4. Click the button to retrieve the selected registers (i.e., the specified data addresses) from the GC Controller (or other device).

Click the button to cease transmission and return to the Modbus Function Selection options.

10.3.2 Use Template (Mixed Data Types)

Use this function to create a new template or use an existing template file to directly decode Modbus register data retrieved from the GC Controller (or other device). Via the configuration of a given template, WinMB imposes the specified data types, etc., onto the incoming registers. The Template function is best used when decoding mixed data types.

To create a new template or use an existing template,

1. Follow Steps 1 through 5 in Section 10.3 .
2. Use the Data Type radio buttons to select the desired template option, to decode registers or decode logs .
Emerson Process Management Chromatograph Software (MON2000) — 10.3.2 Use Template (Mixed Data Types) diagram (page 360)
Fig. 316 · 10.3.2 Use Template (Mixed Data Types)
3. Note that the Record No . setting is now available. Enter the desired record by typing in the provided data field or using the scroll arrows.

Use Template (Mixed Data Types) JULY 2010

Page 361 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS TEST 10-11 MON2000
Page 361 · click to enlarge
Page 361

MODBUS TEST 10-11 MON2000

Table 10-6 Using Record Numbers with Templates

Data Type Setting Other Setting(s) Result

  • register template
  • Enter Data Addr value.
  • Enter Record No. value.
  • Enter Quantity value.

Read Quantity fields (i.e., the number of fields specified by the Quantity setting) from the specified Record No. of the register (Data Addr).

log template • Enter Record No. value. Read all fields associated with the Record No.

  • Enter Data Addr value.
  • Enter “0” for the Record No. value.

Read all fields in all records for the specified log register (Data Addr).

To verify which record number should be entered, consult the Modbus specifications for your device. For more information on GC Modbus registers, see Appendix F.

4. Click the button. The Template File dialog appears with a new template displayed.
Emerson Process Management Chromatograph Software (MON2000) — MODBUS TEST 10-11 MON2000 diagram (page 361)
Fig. 317

JULY 2010 Use Template (Mixed Data Types)

Page 362 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10-12 MODBUS TEST
Page 362 · click to enlarge
Page 362

10-12 MODBUS TEST

5. To open an existing template file, click the

button. The Select Template Configuration File dialog appears.

Emerson Process Management Chromatograph Software (MON2000) — 10-12 MODBUS TEST diagram (page 362)
Fig. 318

Use the directory tree to locate the desired file (e.g., temp1.cfg), highlight it, and then

click the button.

6. To edit the displayed file,
  • Use the provided combo boxes to select a different data type. The Offset and Size values change to accommodate the data type .
  • Click the button to change all template data types to the type last selected.
7. To save the displayed file to disk, click the

button. The Select Template Configuration File dialog appears.

Emerson Process Management Chromatograph Software (MON2000) — 10-12 MODBUS TEST diagram (page 362)
Fig. 319

Use Template (Mixed Data Types) JULY 2010

Page 363 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.4 USING MODBUS DATA
Page 363 · click to enlarge
Page 363

MODBUS TEST 10-13 MON2000

Use the directory tree to specify the desired location. Either select an existing file or type the new file name (e.g., temp2.cfg) in the File name data field. Then click the

button.

8. Click the button to apply your selections and return to the main window.

Click the button to return to the main window without applying your selections.

10.4 USING MODBUS DATA

Use WinMB to poll the GC Controller Modbus registers (or, registers from another device), confirming that data is being accurately relayed from the GC Controller to the PC. Then, as necessary, assign data types to the returned data (see Section 10.4.3). You can save all settings to file for future reference.

To launch the WinMB program from MON2000, use the File > Modbus Test menu (see Section 10.1 for additional options).

JULY 2010 Using Modbus Data

Page 364 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.4.1 Set Log Parameters
Page 364 · click to enlarge
Page 364

10-14 MODBUS TEST

Emerson Process Management Chromatograph Software (MON2000) — 10-14 MODBUS TEST diagram (page 364)
Fig. 320

10.4.1 Set Log Parameters

The Log Data function allows you to log the polled data to a specified file.

The Log Data function is not required for transmission of Modbus data.

To set the log parameters for WinMB,

1. Click the button to access the Log Data function. The Log Data List dialog appears.

To disable this function, deselect the Enable Logging check box.

2. Click the Enable Logging check box to turn this function ON and to access the Log Data Parameters.

Set Log Parameters JULY 2010

Page 365 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS TEST 10-15 MON2000
Page 365 · click to enlarge
Page 365

MODBUS TEST 10-15 MON2000

3. Use the Logging Mode pull-down menu to select either a Sampling or Continuous mode.
Emerson Process Management Chromatograph Software (MON2000) — MODBUS TEST 10-15 MON2000 diagram (page 365)
Fig. 321

Continuous mode records the polled data continuously until the connection is terminated or the Log Data function is disabled (see Step 2).

Sampling mode records the polled data per the user-defined Time Interval (e.g., every 2 minutes).

4. Select the desired logging type.

Append adds this log to the file specified, preserving previously logged data.

Reset deletes the previously logged data and saves only this new log.

5. Click the button to apply your selections and return to the main window. Go to Step 6.

Click the button to return to the main window without applying your selections.

JULY 2010 Set Log Parameters

Page 366 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.4.2 Save Modbus Data
Page 366 · click to enlarge
Page 366

10-16 MODBUS TEST

6. The Save As dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 10-16 MODBUS TEST diagram (page 366)
Fig. 322

Use the directory tree to specify the desired location and choose a format type. Either select an existing file or type the new file name in the File name data field.

You can save log data in any of these formats:

7. Click the button to create this log file, apply your selections, and return to the main window.
  • .txt, ASCII tabdeliminated
  • .xls, Excel
  • .htm, HTML

Click the button to abort and return to the main window.

10.4.2 Save Modbus Data

To save the data table in a separate file,

1. Click the button to access the Save Data function. The Save Data Displayed dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 10.4.2 Save Modbus Data diagram (page 366)
Fig. 323 · 10.4.2 Save Modbus Data

Save Modbus Data JULY 2010

Page 367 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.4.4 Print Modbus Data
Page 367 · click to enlarge
Page 367

MODBUS TEST 10-17 MON2000

2. Use the directory tree to specify the desired location and either select an existing file or type the new file name in the File name data field.

You can save log data in any of these formats:

  • .txt, ASCII tabdeliminated
  • .xls, Excel
  • .htm, HTML
3. Click the button to save this data and return to the main window.

Click the button to return to the main window without saving this data.

10.4.3 Assign Scale Ranges for User_Modbus

This is an optional task that applies to applications using the User_Modbus protocol. By assigning scale ranges, floating point data can be converted to integer values.

Use the Serial Ports, Edit Register List function to assign scale ranges. See Section 5.18.4 for detailed instructions. See Appendix F for more information regarding GC Modbus registers.

10.4.4 Print Modbus Data

To print the transmitted data, click the

button. The standard Windows® print dialog appears.

MON2000 prints the report to your configured printer (see Section 3.7).

JULY 2010 Assign Scale Ranges for User_Modbus

Page 368 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10-18 MODBUS TEST
Page 368 · click to enlarge
Page 368

10-18 MODBUS TEST

10.5 TROUBLESHOOTING COMMUNICATION ERRORS

The WinMB Communication Error Log is maintained in a circular buffer that holds up to 512 entries.

WinMB tracks the errors for a given session but does not store them. When you exit WinMB, all errors are cleared.

To view any communication errors that occurred during the data transfer, from the File>Modbus Test menu,

1. Click the button to access the Communication Error Log function. The Error dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — 10-18 MODBUS TEST diagram (page 368)
Fig. 324

Double-click a Description cell to “scroll through” the displayed text.

2. To view all errors that have occurred in this

session, click the button.

To delete all entries to date, click the

button.

3. Click the button to return to the main window.

Troubleshooting Communication Errors JULY 2010

Page 369 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10.6.1 How to Access
Page 369 · click to enlarge
Page 369

MODBUS TEST 10-19 MON2000

10.6 USING MODBUS TEST ONLINE HELP

Use the WinMB online help file to quickly access reference terminology, function descriptions, and other related information.

See the following sections for more information.

10.6.1 How to Access

To access online help, use the File>Modbus Test menu to display the Modbus Help test dialog, then,

  • Press F1 on your PC keyboard
  • Click the button located on the toolbar.

The Modbus Test Program Help Index displays.

Emerson Process Management Chromatograph Software (MON2000) — 10.6.1 How to Access diagram (page 369)
Fig. 325 · 10.6.1 How to Access

JULY 2010 Using Modbus Test Online Help

Or,

Page 370 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 10-20 MODBUS TEST
Page 370 · click to enlarge
Page 370

10-20 MODBUS TEST

10.6.2 How to Navigate

To navigate within this help file,

  • Click the button to display an index, a topic-tree contents menu, and a search function.
  • Click the button to display an index where you can select a specific term.
  • Click the button to return to the last topic viewed.
  • Click the button to print the current topic.

Use the scroll bars and arrows to display more of a topic. You can also resize and/or move the topic window for better viewing convenience.

When applicable, a topic will contain links to other related topics. Use these links as you need.

How to Navigate JULY 2010

Page 371 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.1 HOW TO PRINT
Page 371 · click to enlarge
Page 371

PC CONFIG REPORT A-1 MON2000

AA APPENDIX A, PC CONFIG REPORT

This appendix explains how to print a PC Config Report and provides an example for reference.

A.1 HOW TO PRINT

Use the following instructions to print the current application settings for the online GC. To view a PC Config Report already saved to disk, see Section 7.1.2.

Output data for the PC Config Report depends on the GC Controller and its application.

1. Use the File > PC Config Report menu to access this function. The PC Config Report window appears.
2. Click the check boxes (see Select column) to choose the option you want included in the configuration report.
Emerson Process Management Chromatograph Software (MON2000) — A.1 HOW TO PRINT diagram (page 371)
Fig. 326

Click the button or press the F2 key to select all options.

Click the button or press F3 to deselect (i.e., clear) all checked options.

JULY 2010 How to Print

Page 372 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 EXAMPLE REPORT
Page 372 · click to enlarge
Page 372

A-2 PC CONFIG REPORT

3. Use the Output radio buttons to either save the report to disk, print the report with or without form feeds, or display on screen.

If you are online with the GC Controller, a PC Config Report that includes all options can require up to 20 minutes to generate and save. Printing a full report can take longer.

If you want to use the standard Windows® Print dialog to select a printer, uncheck the Use default printer option.

4. Click the button or press the F4 key to generate your customized PC Config Report and print or save this file.

If you press the ESC key, MON2000 will stop after the current option is completed.

By default, if a print option is selected, MON2000 will print the report to your configured printer (see Section 3.7).

A.2 EXAMPLE REPORT

This example PC Config Report, on the following pages, represents a typical report that includes all sections. It is provided for reference only.

Example Report JULY 2010

Page 373 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 373 · click to enlarge
Page 373

System Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

Aux. Stream Sequence

System Description 2350 Standard 1 - 64 avgs

Chromatogram Buffers Size 21000

Number of Valves 5

Daylight Saving Time Normal

CGM Analog Output Config 0

CFG Baseline Number 1

Metric Base Conditions Off

Max. Archive Averages 64

Max. Archive Avg. Records 64

CFG Revision Level 150

CFG Base Name USASTD1

Show 8 more labels / short text on this page
Description Value · -------------------------------------------------------------------------------- · Stream Sequence 1 · Analyzer Name 2350_001 · Unit Type 2350 · Max Peaks 64 · Baseline Offset 12 · ********************************************************************************
Page 374 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 374 · click to enlarge
Page 374

Component Data Table Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

Component Data Table #1

No. Component U/S DetID RetTime Resp Factor F/V CalConc AnalMthd

1 C6+ 47/35/17 STD 1 30.2 2013490 VAR 0.0289 % AREA

2 PROPANE STD 1 50.2 1316330 VAR 1.0 % AREA

3 i-BUTANE STD 1 65.7 1520820 VAR 0.3 % AREA

4 n-BUTANE STD 1 73.6 1545350 VAR 0.3 % AREA

5 NEOPENTANE STD 1 86.3 1651260 VAR 0.1 % AREA

6 i-PENTANE STD 1 109.4 1715050 VAR 0.0993 % AREA

7 n-PENTANE STD 1 121.6 1793230 VAR 0.0999 % AREA

8 NITROGEN STD 1 149.5 782820 VAR 2.5 % AREA

9 METHANE STD 1 153.1 637330 VAR 89.5719 % AREA

10 CARBON DIOXIDE STD 1 179.2 953088 VAR 1.0 % AREA

11 ETHANE STD 1 200.6 1060440 VAR 5.0 % AREA

No. Component RTDev SecDev Updt RFDev GRSBTU NETBTU MolWgt AGA8

1 C6+ 47/35/17 3 2 CAL 10 5288.7 4900.5 95.96 C6mix1

2 PROPANE 3 2 CAL 10 2521.9 2320.3 44.1 PROPANE

3 i-BUTANE 3 2 CAL 10 3259.4 3007.3 58.12 i-BUTANE

4 n-BUTANE 3 2 CAL 10 3269.8 3017.4 58.12 n-BUTANE

5 NEOPENTANE 3 3 CAL 10 3993.9 3691.4 72.15 i-PENTANE

6 i-PENTANE 3 4 CAL 10 4010.2 3707.6 72.15 i-PENTANE

7 n-PENTANE 3 4 CAL 10 4018.2 3715.5 72.15 n-PENTANE

8 NITROGEN 3 2 CAL 10 0 0 28.01 NITROGEN

9 METHANE 3 3 CAL 10 1012.3 911.5 16.04 METHANE

10 CARBON DIOXIDE 3 4 CAL 10 0 0 44.01 CO2

11 ETHANE 3 5 CAL 10 1773.7 1622.4 30.07 ETHANE

·

Page 375 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 8 NITROGEN 0 0 0 0 0.8093
Page 375 · click to enlarge
Page 375

No. Component Lb/Gal GPMFactor Reid Vapor RelDenGas RelDenLqd

1 C6+ 47/35/17 5.6681 0.4462 3.0194 3.3132 0.68

2 PROPANE 4.227 0.2756 188.64 1.5226 0.507

3 i-BUTANE 4.693 0.3271 72.591 2.0068 0.5629

4 n-BUTANE 4.869 0.3153 51.709 2.0068 0.584

5 NEOPENTANE 4.975 0.383 35.9 2.4911 0.5967

6 i-PENTANE 5.206 0.3661 20.445 2.4912 0.6244

7 n-PENTANE 5.261 0.3622 15.574 2.4912 0.6311

8 NITROGEN 6.748 0.1096 0 0.9672 0.8094

9 METHANE 2.5 0.1695 5000 0.5539 0.3

10 CARBON DIOXIDE 6.82 0.1704 0 1.5196 0.818

11 ETHANE 2.97 0.2675 800 1.0382 0.3562

No. Component HVSupMJ/m3 HVInfMJ/m3 HVSupMJ/kg HVInfMJ/kg M_RelDens

1 C6+ 47/35/17 196.98 182.52 48.557 44.99 0.6803

2 PROPANE 93.936 86.42 50.368 46.34 0.5078

3 i-BUTANE 121.41 112.01 49.388 45.57 0.5635

4 n-BUTANE 121.79 112.4 49.546 45.72 0.5846

5 NEOPENTANE 148.76 137.49 48.75 45.06 0

6 i-PENTANE 149.36 138.09 48.949 45.25 0.6249

7 n-PENTANE 149.66 138.38 49.045 45.35 0.6316

8 NITROGEN 0.8093

9 METHANE 37.708 33.948 55.575 50.034 0.3

10 CARBON DIOXIDE 0 0 0 0 0.8227

11 ETHANE 66.065 60.43 51.95 47.52 0.3581

· · ********************************************************************************

Page 376 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 376 · click to enlarge
Page 376

Timed Event Table Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

Timed Event Table #1

Analysis Time 225 Cycle Time 240 Sample Valve Y N

Valve # 2 ON @ 0.0

Inhibit ON 1 @ 0.0

Valve # 3 ON @ 2.0

Slope Sens 8 1 @ 3.0

Detector 1 6 @ 4.0

Valve # 1 ON @ 5.0

Strm Switch @ 7.0

Peak Width 4 1 @ 8.0

Valve # 1 OFF @ 10.0

Valve # 2 OFF @ 23.0

Inhibit OFF 1 @ 26.0

Inhibit ON 1 @ 36.0

Valve # 3 OFF @ 42.0

Inhibit OFF 1 @ 46.0

Inhibit ON 1 @ 97.0

Peak Width 8 1 @ 100.0

Inhibit OFF 1 @ 101.0

Inhibit ON 1 @ 139.0

Detector 1 1 @ 140.0

Valve # 3 ON @ 142.0

Peak Width 3 1 @ 143.0

Inhibit OFF 1 @ 146.0

Detector 1 5 @ 169.0

Inhibit ON 1 @ 170.0

Peak Width 8 1 @ 173.0

Inhibit OFF 1 @ 174.0

Inhibit ON 1 @ 223.0

· ********************************************************************************

Page 377 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 1 DISABLE
Page 377 · click to enlarge
Page 377

User-Defined Numeric Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Name Type Value

No entries found.

User-Defined Selection Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Name Options

No. Name

1 DISABLE

2 ENABLE

No. Name

1 DISABLE

2 ENABLE_NC

3 ENABLE_NO

Show 13 more labels / short text on this page
-------------------------------------------------------------------------------- · ******************************************************************************** · -------------------------------------------------------------------------------- · 1 EnableCalMan 2 · 2 EnableComAlm 2 · 4 Maint_mode 1 · 5 Run_Aux_Seq 1 · Options for EnableCalMan · --------------------------- · ---------------------- · Options for EnableComAlm · --------------------------- · ----------------------
Page 378 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 1 OFF
Page 378 · click to enlarge
Page 378

No. Name

1 OFF

2 ON

No. Name

1 OFF

2 ON

User-Defined Text Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Name Size Text

1 rpt_header1 50 CompanyDaniel Industries
Show 10 more labels / short text on this page
Options for Maint_mode · --------------------------- · ---------------------- · Options for Run_Aux_Seq · --------------------------- · ---------------------- · ******************************************************************************** · -------------------------------------------------------------------------------- · 2 rpt_header2 50 · ********************************************************************************
Page 379 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 379 · click to enlarge
Page 379

Calculation Control Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

Description 1 2 3 4 5 6 7 8

Avg Limit Alarm Test Y

Mole Percent Y

Liquid Volume N

Weight Percent N

Normalize Results N

Gas Density N

Liquid Density N

Relative Density Gas N

Relative Density Lqd N

Compressibility N

Dry Gross Heating BTU N

Sat Gross Heating BTU N

Actual Gross Heating BTU N

Dry Net Heating BTU N

Sat Net Heating BTU N

Actual Net Heating BTU N

Wobbe Index N

Reid Vapor Pressure N

Gallons/1000 SCF C2+ N

Gallons/1000 SCF C3+ N

Gallons/1000 SCF C4+ N

Gallons/1000 SCF C5+ N

Gallons/1000 SCF C6+ N

User Flag 1 N

Average Molecular Wgt N

HeatVal Sup Dry MJ/m3 N

HeatVal Sup Sat MJ/m3 N

HeatVal Sup Act MJ/m3 N

HeatVal Inf Dry MJ/m3 N

Page 380 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 380 · click to enlarge
Page 380

HeatVal Inf Sat MJ/m3 N

HeatVal Inf Act MJ/m3 N

HeatVal Sup Dry MJ/kg N

HeatVal Inf Dry MJ/kg N

HeatVal Sup Dry Kc/m3 N

HeatVal Sup Sat Kc/m3 N

HeatVal Sup Act Kc/m3 N

HeatVal Inf Dry Kc/m3 N

HeatVal Inf Sat Kc/m3 N

HeatVal Inf Act Kc/m3 N

HeatVal Sup Dry Kc/kg N

HeatVal Inf Dry Kc/kg N

Metric Rel Dens Lqd N

Metric Gas Density N

Metric Lqd Density N

Calculation Average Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

Time00:00
WeekdaySun
Day of Month1

No. Variable Name S C

No entries found.

No entries found.

No entries found.

No entries found.

No entries found.

Show 9 more labels / short text on this page
******************************************************************************** · Resets · -------------------------------------------------------------------------------- · Hourly · 24 Hour · Weekly · Monthly · Variable · ********************************************************************************
Page 381 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 381 · click to enlarge
Page 381

Calculation User-Defined Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Variable S C

Avg. Molecular Weight 1 1

+ Gallons/1000 SCF 1 1

* Gas Density lbm/1000 ft3 1 1

/ Heating Value Gross BTU Dry 1 1

- Mole Percent 1 1

Gallons/1000 SCF 1 2

+ Gallons/1000 SCF 1 3

+ User Calc Result 1 1

+ Value 116.2 1 1

Limit Alarms Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Variable S C DO Type

No entries found.

Discrete Alarms Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

Alarm Discrete Discrete Alarm Message

No. Input Output

Show 8 more labels / short text on this page
-------------------------------------------------------------------------------- · 1 label 1 · 2 label 2 · ******************************************************************************** · -------------------------------------------------------------------------------- · ******************************************************************************** · -------------------------------------------------------------------------------- · ********************************************************************************
Page 382 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 382 · click to enlarge
Page 382

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 383 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 383 · click to enlarge
Page 383

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 384 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 1 DI 1 AUTO
Page 384 · click to enlarge
Page 384

No. Zero Scale Full Scale

Discrete Input Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Label Switch

1 DI 1 AUTO

2 DI 2 AUTO

3 DI 3 AUTO

4 DI 4 AUTO

5 AUX_STRM_SEQ AUTO

Show 15 more labels / short text on this page
Scaling Values · -------------------------------------------------------------------------------- · 1 0 100 · 2 0 100 · 3 0 100 · 4 0 100 · 5 0 100 · 6 0 100 · 7 0 100 · 8 0 100 · 9 0 100 · 10 0 100 · ******************************************************************************** · -------------------------------------------------------------------------------- · ********************************************************************************
Page 385 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 1 DO 1 AUTO 00:00 00:00
Page 385 · click to enlarge
Page 385

Discrete Output Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Label Switch On-Time Off-Time

1 DO 1 AUTO 00:00 00:00

2 DO 2 AUTO 00:00 00:00

3 DO 3 AUTO 00:00 00:00

4 Common_Alarm AUTO 00:00 00:00

5 Calib-Maint AUTO 00:00 00:00

Valve Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Label Switch

1 Valve 1 AUTO

2 Valve 2 AUTO

3 Valve 3 AUTO

4 Valve 4 AUTO

5 Valve 5 AUTO

Serial Port Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

Baud Data Stop Hand- RTS RTS Comm Read

Port Usage Rate Bits Bits Parity shaking ON OFF Protocol ID Write

1 USER_MODBUS 9600 8 1 None None 0 0 RTU 0 RW

2 SIM_2251 9600 8 1 None None 0 0 RTU 3 RW

3 PC 9600 8 1 None None 0 0 RTU 0 RW

4 PC 9600 8 1 None None 0 0 RTU 0 RW

Show 6 more labels / short text on this page
-------------------------------------------------------------------------------- · ******************************************************************************** · -------------------------------------------------------------------------------- · ******************************************************************************** · -------------------------------------------------------------------------------- · ********************************************************************************
Page 386 of the Emerson Process Management Chromatograph Software (MON2000) user manual — S C
Page 386 · click to enlarge
Page 386

Global Data Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Variable name Indices Type No. of Chars.

S C

1 cal_run_num 1 1 Integer

2 sim_abtu 1 1 Float

3 sim_alarm 16 1 Integer

4 sim_anlytime 1 1 Integer

5 sim_cal 1 1 Integer

6 sim_cal_abtu 1 1 Float

7 sim_cal_comp 1 1 Float

8 sim_cal_dbtu 1 1 Float

9 sim_cal_gpm 1 1 Float

10 sim_cal_rden 1 1 Float

11 sim_cal_sbtu 1 1 Float

12 sim_cal_unno 1 1 Float

13 sim_cal_wobb 1 1 Float

14 sim_calcycle 1 1 Long

15 sim_char 1 1 Char 1

16 sim_compno 16 4 Integer

17 sim_compress 1 1 Float

18 sim_cycday 1 1 Integer

19 sim_cychour 1 1 Integer

20 sim_cycmin 1 1 Integer

21 sim_cycmonth 1 1 Integer

22 sim_cyctime 1 1 Long

23 sim_cycyear 1 1 Integer

24 sim_day 1 1 Integer

Page 387 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 387 · click to enlarge
Page 387

25 sim_dbtu 1 1 Float

26 sim_dummy 4 1 Float

27 sim_gpmwt 16 4 Float

28 sim_hour 1 1 Integer

29 sim_min 1 1 Integer

30 sim_molpct 16 4 Float

31 sim_month 1 1 Integer

32 sim_newdata 1 1 Integer

33 sim_rden_gas 1 1 Float

34 sim_rf 16 4 Float

35 sim_sbtu 1 1 Float

36 sim_smask 1 1 Integer

37 sim_stream 1 1 Integer

38 sim_totgpm 1 1 Float

39 sim_unnormal 1 1 Float

40 sim_wobbe 1 1 Float

41 sim_year 1 1 Integer

Streams Data Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Structure Name No. of Streams

No entries found.

******************************************************************************** · · ********************************************************************************

Page 388 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 388 · click to enlarge
Page 388

Component Data Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. Structure Name No. of Streams No. of Components

No entries found.

Data Files Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

No. File Name No. of Records

No. Field Name Type No. of Chars.

3 operator_id Char 4

5 ASCII_text Char 60

Show 14 more labels / short text on this page
per Stream · -------------------------------------------------------------------------------- · ******************************************************************************** · -------------------------------------------------------------------------------- · 1 Event_Log 240 · File Name Event_Log · --------------------------- · -------------------------------------------------- · 1 date Float · 2 time Float · 4 event_id Float · 6 prev_value Float · 7 curr_value Float · ********************************************************************************
Page 389 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 389 · click to enlarge
Page 389

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 390 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 390 · click to enlarge
Page 390

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 391 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 391 · click to enlarge
Page 391

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 392 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 392 · click to enlarge
Page 392

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 393 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 393 · click to enlarge
Page 393

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 394 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 394 · click to enlarge
Page 394

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 395 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 395 · click to enlarge
Page 395

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 396 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 396 · click to enlarge
Page 396

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 397 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 397 · click to enlarge
Page 397

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 398 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 398 · click to enlarge
Page 398

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 399 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 399 · click to enlarge
Page 399

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 400 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 400 · click to enlarge
Page 400

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 401 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 401 · click to enlarge
Page 401

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 402 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 402 · click to enlarge
Page 402

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 403 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 403 · click to enlarge
Page 403

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 404 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 404 · click to enlarge
Page 404

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 405 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 405 · click to enlarge
Page 405

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 406 of the Emerson Process Management Chromatograph Software (MON2000) user manual — Table of contents
Page 406 · click to enlarge
Page 406

Table of contents

Table of Contentsp.7Introductionp.191.1 Description of Manualp.19Section 1 - Introductionp.19Section 2 - Installation and Startupp.20Section 3 - Getting Startedp.20Section 4 - Control Functionsp.20Section 5 - Application Functionsp.21Section 6 - Chromatogram Viewerp.21Section 7 - Reportsp.21Section 8 - Logsp.21Section 9 - Data Collectionp.22Section 10 - Modbus Testp.22Appendix A, PC Config Reportp.23Appendix B, Component Data Tablep.23Appendix C, Data Computationsp.23Appendix D, Analog Output Cal. 2350A GCp.23Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.23Appendix F, Modbus Reg. List for 2350A GCp.231.2 Description of Online Helpp.241.3 Description of MON2000p.24Installation and Setupp.272.1 System Requirementsp.272.2 Installing MON2000p.292.3 Installing the CrypKey License Service for MON2000 PLUSp.302.4 Copying MON2000 to Disksp.302.5 Transferring a MON2000 PLUS Licensep.322.6 Uninstalling MON2000p.412.7 Starting MON2000p.432.8 Performing Your First Logonp.442.8.1 The Initial Logonp.442.8.2 Registering MON2000p.452.8.3 Update MON2000p.472.9 Checklist for Setting Up MON2000p.482.10 Communicationsp.482.10.1 Connect to the GC Unitp.482.10.2 Disconnect from GC Unitp.492.11 Customizing MON2000p.502.11.1 Configure Usersp.502.11.2 Set Up GC Directoryp.532.11.3 Dial-up Connectionp.562.11.4 Modem Initialization Strings/Setupp.62Black Box - Modem 325p.63U.S. Robotics Sportsterp.64GVC/MaxTech 28.8 Kbps Modemp.662.12 2350 to 2350A Retrofit Instructionsp.672.13 Conversion Processp.692.14 BASIC 2350A CONFIGURATIONp.752.15 2350A Optionsp.782.15.1 The COM4A Boardp.782.15.2 2350A Modem Installationp.812.15.3 2350A Ethernet Installationp.842.15.4 Ethernet TCP/IP Settingsp.882.15.5 MON2000 TCP/IP Settingsp.902.16 CONNECT VIA GC EXTERNAL MODEMp.922.16.1 Hardware Setupp.922.16.2 Select Program Settingsp.942.16.3 Generate PC Config Reportp.962.17 Downloading an Applicationp.982.18 Upgrading Flashp.1012.19 Offline Edit of GC Applicationp.1022.19.1 File Selectionp.1032.19.2 Upload Applicationp.1042.19.3 Upload BOSp.107Getting Startedp.1093.1 Logging Onp.1093.2 Logging Offp.1113.3 MON2000 User Interfacep.1113.3.1 Main Windowp.1123.3.2 GC Status Barp.1133.3.3 View Menup.1163.3.4 ToolTipsp.1163.3.5 Data Entry and Function Featuresp.1163.3.6 Shortcut to Save or Print Datap.1173.4 Keyboard Shortcutsp.1173.5 Procedures Guidep.1193.6 Keylock Switch Controlp.1233.7 Configuring Your Printerp.1243.8 Using Online Helpp.1263.9 Operating Modes for Model 700p.1273.10 About MON2000p.127Control Functionsp.1294.1 Auto Sequencep.1294.2 Single Streamp.1304.3 Haltp.1314.4 Calibrationp.1324.5 Baseline Runp.1334.6 Auto BTU Start Up and Valve Timingp.1354.6.1 Initiating an Auto BTU Start Up Runp.1354.6.2 Starting a Valve Timing Runp.1404.7 GC Timep.1414.7.1 View GC Timep.1414.7.2 Set GC Timep.1424.8 Stop Nowp.143Application Functionsp.1455.1 Systemp.1455.2 Component Datap.1505.2.1 Select Standard Component(s)p.1545.2.2 Select Standard Valuesp.1555.2.3 Update Standard Component(s)p.1555.2.4 View Raw Datap.1565.2.5 Sort Retention Timep.1585.3 Timed Eventsp.158Timed Events Descriptionp.158Valve Eventsp.159Integration Eventsp.159Spectrum Gain Eventsp.1605.4 User Definedp.1635.4.1 Numericp.1635.4.2 Selectionp.1645.4.3 Text Stringsp.1645.5 Calculationsp.1655.5.1 Controlp.1655.5.2 Averagesp.1675.5.3 User Definedp.1695.6 Limit Alarmsp.171Limit Alarms Descriptionp.1745.7 Discrete Alarmsp.1755.7.1 Delayed Discrete Alarmsp.1775.8 Streamsp.1825.9 Analog Inputsp.1885.10 Analog Outputsp.1925.10.1 Analog Output Dialog Descriptionp.1925.10.2 Changing a Variablep.1965.10.3 Changing the Bargraphp.1965.10.4 Performing a Manual Calibrationp.1975.10.5 Performing an Automated Calibrationp.1985.11 Discrete Inputsp.2005.12 Discrete Outputsp.2015.13 Valvesp.2025.14 Temperature Controlp.2055.15 FID Configurationp.2065.16 LOI Status Variablesp.2105.17 Serial Portsp.2125.17.1 Serial Ports Configurationp.2155.18 GC Serial Port and Cable Configurationsp.2185.18.1 GC DB-9 Serial to External Modem DB-25p.2205.18.2 GC Phoenix Plug to External Modem DB-25p.2215.18.3 Com IDp.2225.18.4 Registersp.2245.18.5 Setting Optional Base Pressuresp.2305.18.6 Setting Optional Base Resultsp.2345.19 TCP/IP, Subnet, and Gateway Menup.238Chromatogram Viewerp.2416.1 Chromatogram Viewer Functionp.2416.2 Viewing a Live Chromatogramp.2426.2.1 From Online GCp.2426.2.2 From GC Archivep.2436.2.3 From File on Diskp.2446.3 Removing a Chromatogram from Viewp.2486.4 Saving a Chromatogram to Diskp.2486.5 Graph Functionsp.2496.5.1 Editing the Chromatogram Graphp.2496.5.2 Changing Cursor Sizep.2506.5.3 Printing the Chromatogramp.2516.6 Chromatogram Functionsp.2516.6.1 Editing the Chromatogramp.2526.6.2 Editing TEVs from CGM Viewerp.2536.6.3 Editing Retention Times from CGM Viewerp.2556.6.4 Editing TEVs from Cursorp.2556.6.5 Viewing the Chromatogram Resultsp.2566.6.6 Entering a Descriptionp.2586.6.7 Forcing a Calibrationp.2596.6.8 Toggling TEVs and CGM Componentsp.2596.7 Viewing Baseline Datap.2606.8 Viewing RAW Datap.2616.9 Display Optionsp.261Reportsp.2637.1 Report Displayp.2637.1.1 View Report from Live Datap.2637.1.2 View Report from Filep.2667.1.3 Print Reportp.2687.1.4 Save Report to Diskp.2687.1.5 Report Contentsp.270Analysis Reportp.271Raw Data Reportp.273Calibration Reportp.274Final Calibration Reportp.275Average Reportsp.276Sample 24-Hour Average Reportp.2767.2 GC Report Requestp.2777.3 GC Printer Controlp.2787.4 MON2000 Printer Controlp.2807.5 Archive Datap.2827.5.1 View Data Model 500/Model 700p.2837.5.2 Archive Export Datap.2867.5.3 Resetp.2897.6 Trend Datap.2907.6.1 Trend Data Model 700/2350A GCsp.2907.6.2 View Live Trend Online Model 700/2350Ap.2917.6.3 View Trend from File on Diskp.2947.6.4 Edit Graph Displayp.2967.6.5 Changing Cursor Sizep.2967.6.6 Describe Trendp.2977.6.7 Print Trendp.2977.6.8 Save Trendp.2987.6.9 Read Trend Archivep.2987.6.10 Display Optionsp.299Logsp.3018.1 Maintenance Logp.3018.2 Parameter Listp.3038.3 Alarm Logp.3068.4 Clear or Acknowledge Active Alarmsp.3098.5 Event Logp.309MON2000 Plus Data Collection/Auto-Pollingp.3119.1 Overviewp.3119.2 Data Collection Configurationp.3139.2.1 Create a New Polling Control Filep.3169.2.2 Open an Existing Polling Control Filep.3179.2.3 Await Commandp.3199.2.4 Connect/Logon Commandp.3209.2.5 Connect/No Logon Commandp.3219.2.6 Delay (Seconds) Commandp.3219.2.7 Delay (Hours) Commandp.3229.2.8 Disconnect Commandp.3239.2.9 End Commandp.3239.2.10 Poll Command: Alarmsp.3249.2.11 Poll Command: Averagesp.3269.2.12 Poll Command: CGM Archivep.3299.2.13 Poll Command: CGM on Alarmp.3319.2.14 Poll Command: Condition Startp.3319.2.15 Poll Command: Condition Endp.3339.2.16 Poll Command: Event Logp.3349.2.17 Poll Command: Maintenance Logp.3369.2.18 Poll Command: Registersp.3389.2.19 Poll Command: Reportsp.3429.2.20 Repeat Commandp.3449.2.21 Run Commandp.3459.3 Data Collectionp.3459.3.1 Start Auto-Sequencingp.3469.3.2 Transaction Logp.3489.3.3 Auto-Sequencing Statusp.3489.3.4 Status Logp.349Modbus Testp.35110.1 Starting WinMBp.35110.2 Establishing Communicationsp.35210.2.1 Comparison of Modbus Protocolsp.35310.2.2 Set GC Com Parameters via MON2000p.35410.2.3 Set Up Port via WinMBp.35510.3 Getting Modbus Datap.35710.3.1 Use Single Data Typep.35910.3.2 Use Template (Mixed Data Types)p.36010.4 Using Modbus Datap.36310.4.1 Set Log Parametersp.36410.4.2 Save Modbus Datap.36610.4.3 Assign Scale Ranges for User_Modbusp.36710.4.4 Print Modbus Datap.36710.5 Troubleshooting Communication Errorsp.36810.6 Using Modbus Test Online Helpp.36910.6.1 How to Accessp.36910.6.2 How to Navigatep.370Appendix A, PC Config Reportp.371A.1 How to Printp.371A.2 Example Reportp.372Appendix B: Component Data Tablep.433Appendix C, Data Computationsp.445C.1 Data Acquisitionp.445C.2 Peak Detectionp.446C.3 Analysis Computationsp.449C.3.1 Conc. Analysis with Response Factorp.449Calibrationp.449Calc. in Mole Percent w/o Normalizationp.451C.4 Post Analysis Computationsp.453C.4.1 Liquid Equivalent Computationsp.453C.4.2 Heating Value Calculationsp.453C.4.3 Multi-Level Calibrationp.461C.4.4 Indirect Calibrationp.462Appendix D, Analog Output Cal. For 2350Ap.465D.1 Calibrating by Voltsp.465D.2 Calibrating by Percentagesp.469Appendix E, Upgrade 2350A GC S/W and 2350 EPROMSp.473E.1 Connect to GC and Halt Analysisp.473E.2 Offline Edit to Upload App. & Renamep.473E.3 Upgrade User-Defined Applicationsp.475E.4 Upgrade from Diskp.475E.4.1 Upgrade from GC Controllerp.478E.4.2 Configure GC Connectionp.481E.5 Disconnect Power and Disassemblep.482E.6 Replace EPROMS/Reset CPUp.484E.7 Reassembly Proceduresp.485E.8 Set-Up and Programmingp.486E.8.1 Procedurep.486E.8.2 Reprogramming the DiskOnChipp.488E.8.3 GC Reassemblyp.489E.9 Connect to GC for Upgraded App.p.490E.10 Guide to Standard Application Filesp.491E.10.1 Importance of Standard Application Filesp.491E.10.2 Standard Applications v1.50 and Laterp.492E.10.3 Standard Applications Prior to v1.50p.494Appendix F, Modbus Reg. List for 2350A GCp.495F.1 Introduction - SIM_2251 & User_Modbusp.495F.1.1 Notes on User_Modbusp.495F.1.2 Notes on SIM_2251 Modbusp.496F.2 User_Modbus Register Listp.497F.2.1 SIM_2251 Modbus Register Listp.500
Page 407 of the Emerson Process Management Chromatograph Software (MON2000) user manual — USER_MODBUS
Page 407 · click to enlarge
Page 407

Communication Report from 2350_001 2350_001.BIN

6/2/2000 15:02:49

Slave Name Slave ID Slave Type

USER_MODBUS 0 PLC Emulation - Low-Hi

SIM_2251 3 Daniel 2500 Emulation

USER_MODBUS

Modbus Variable Field Indices Data String Override

Reg. Name Name S C Type Length Keylock?

0 sysalarm_set 1 1 Application Checksum Failure No

1 sysalarm_set 2 1 ROM Checksum Failure No

2 sysalarm_set 3 1 RAM Diagnostics Failure No

3 sysalarm_set 4 1 A/D Converter Failure No

4 sysalarm_set 5 1 Detector Oven Failure No

5 sysalarm_set 6 1 Liquid Sample Valve Heater Failure No

6 sysalarm_set 7 1 Sample System Oven Failure No

7 sysalarm_set 8 1 Catalytic Converter Failure No

8 sysalarm_set 9 1 Heater 5 Failure No

9 sysalarm_set 10 1 Heater 6 Failure No

10 sysalarm_set 11 1 Heater 1 Controller Failure No

11 sysalarm_set 12 1 Heater 2 Controller Failure No

12 sysalarm_set 13 1 Heater 3 Controller Failure No

13 sysalarm_set 14 1 Heater 4 Controller Failure No

14 sysalarm_set 15 1 Heater 5 Controller Failure No

15 sysalarm_set 16 1 Heater 6 Controller Failure No

16 sysalarm_set 17 1 FID Flame out No

17 sysalarm_set 18 1 Warmstart Calibration Failure No

Show 4 more labels / short text on this page
----------------------------------------------------------------------------------- · ------------- · ----------------------------------------------------------------------------------- · Boolean(Coils)
Page 408 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 408 · click to enlarge
Page 408

18 sysalarm_set 19 1 Valve Timing Failure No

19 sysalarm_set 20 1 Excess Response Factor Deviation No

20 sysalarm_set 21 1 M200 Invalid Non-Volatile Data No

21 sysalarm_set 22 1 M200 Invalid A Module Data No

22 sysalarm_set 23 1 M200 Invalid B Module Data No

23 sysalarm_set 24 1 M200 Bad Options No

24 sysalarm_set 25 1 M200 Stack Overflow No

25 sysalarm_set 26 1 M200 Hardware Shutdown No

26 sysalarm_set 27 1 M200 Synchronization Failure No

27 sysalarm_set 28 1 Preamp Input 1 Out of Range - DET1 No

28 sysalarm_set 29 1 Preamp Input 2 Out of Range - DET1 No

29 sysalarm_set 30 1 Preamp Input 3 Out of Range - DET1 No

30 sysalarm_set 31 1 Preamp Input 4 Out of Range - DET1 No

31 sysalarm_set 32 1 Preamp Failure - DET1 No

32 sysalarm_set 33 1 Analog Output 1 HIGH No

33 sysalarm_set 34 1 Analog Output 2 HIGH No

34 sysalarm_set 35 1 Analog Output 3 HIGH No

35 sysalarm_set 36 1 Analog Output 4 HIGH No

36 sysalarm_set 37 1 Analog Output 5 HIGH No

37 sysalarm_set 38 1 Analog Output 6 HIGH No

38 sysalarm_set 39 1 Analog Output 7 HIGH No

39 sysalarm_set 40 1 Analog Output 8 HIGH No

40 sysalarm_set 41 1 Analog Output 9 HIGH No

41 sysalarm_set 42 1 Analog Output 10 HIGH No

42 sysalarm_set 43 1 Analog Output 11 HIGH No

43 sysalarm_set 44 1 Analog Output 12 HIGH No

44 sysalarm_set 45 1 Analog Output 13 HIGH No

45 sysalarm_set 46 1 Analog Output 14 HIGH No

46 sysalarm_set 47 1 Analog Output 15 HIGH No

47 sysalarm_set 48 1 Analog Output 16 HIGH No

48 sysalarm_set 49 1 Analog Output 1 LOW No

49 sysalarm_set 50 1 Analog Output 2 LOW No

50 sysalarm_set 51 1 Analog Output 3 LOW No

Page 410 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 410 · click to enlarge
Page 410

84 sysalarm_set 85 1 NA No

85 lmtalarm_set 1 1 No

86 lmtalarm_set 2 1 No

87 lmtalarm_set 3 1 No

88 lmtalarm_set 4 1 No

89 lmtalarm_set 5 1 No

90 lmtalarm_set 6 1 No

91 lmtalarm_set 7 1 No

92 lmtalarm_set 8 1 No

93 lmtalarm_set 9 1 No

94 lmtalarm_set 10 1 No

95 lmtalarm_set 11 1 No

96 lmtalarm_set 12 1 No

97 lmtalarm_set 13 1 No

98 lmtalarm_set 14 1 No

99 lmtalarm_set 15 1 No

100 lmtalarm_set 16 1 No

101 lmtalarm_set 17 1 No

102 lmtalarm_set 18 1 No

103 lmtalarm_set 19 1 No

104 lmtalarm_set 20 1 No

105 stream_data stream_togg 1 1 Stream 1 No

106 stream_data stream_togg 2 1 Stream 2 No

107 stream_data stream_togg 3 1 Stream 3 No

108 stream_data stream_togg 4 1 Stream 4 No

109 stream_data stream_togg 5 1 Stream 5 No

110 stream_data stream_togg 6 1 Stream 6 No

111 stream_data stream_togg 7 1 Stream 7 No

112 stream_data stream_togg 8 1 Stream 8 No

113 doutcur 1 1 No

114 doutcur 2 1 No

115 doutcur 3 1 No

116 doutcur 4 1 No

117 doutcur 5 1 No

Page 411 of the Emerson Process Management Chromatograph Software (MON2000) user manual — SIM_2251
Page 411 · click to enlarge
Page 411

0 stream_data avg_molwgt 1 1 Char 4 Stream 1 No

2 analysis molpct 1 1 Char 4 C6+ 47/35/17 0 - 1 No

3 analysis molpct 1 2 Integer PROPANE 0 - 1 No

4 analysis molpct 1 3 Integer i-BUTANE 0 - 1 No

5 analysis molpct 1 4 Integer n-BUTANE 0 - 1 No

6 analysis molpct 1 5 Float NEOPENTANE No

8 analysis molpct 1 6 Float i-PENTANE No

10 analysis molpct 1 7 Float n-PENTANE No

12 analysis molpct 1 8 Float NITROGEN No

14 analysis molpct 1 9 Float METHANE No

16 analysis molpct 2 9 Float METHANE No

18 analysis molpct 3 9 Float METHANE No

20 analysis molpct 4 9 Float METHANE No

22 analysis molpct 5 9 Float METHANE No

24 analysis molpct 5 10 Float CARBON DIOXIDE No

26 analysis molpct 5 11 Float ETHANE No

28 analysis molpct 6 11 Float ETHANE No

30 analysis molpct 7 11 Float ETHANE No

32 analysis molpct 8 11 Float ETHANE No

SIM_2251

Modbus Variable Field Indices Override

Reg. Name Name S C Keylock?

Numeric(Registers) · ·

Page 413 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 413 · click to enlarge
Page 413

3028 sim_compno 12 1 No

3029 sim_compno 13 1 No

3030 sim_compno 14 1 No

3031 sim_compno 15 1 No

3032 sim_compno 16 1 No

3033 sim_anlytime 1 1 No

3034 sim_stream 1 1 No

3035 sim_smask 1 1 No

3036 sim_month 1 1 No

3037 sim_day 1 1 No

3038 sim_year 1 1 No

3039 sim_hour 1 1 No

3040 sim_min 1 1 No

3041 sim_cycmonth 1 1 No

3042 sim_cycday 1 1 No

3043 sim_cycyear 1 1 No

3044 sim_cychour 1 1 No

3045 sim_cycmin 1 1 No

3046 sim_alarm 1 1 No

3047 sim_alarm 2 1 No

3048 sim_alarm 3 1 No

3049 sim_alarm 4 1 No

3050 sim_alarm 5 1 No

3051 sim_alarm 6 1 No

3052 sim_alarm 7 1 No

3053 sim_alarm 8 1 No

3054 sim_alarm 9 1 No

3055 sim_alarm 10 1 No

3056 sim_alarm 11 1 No

3057 sim_alarm 12 1 No

3058 sim_newdata 1 1 No

3059 sim_cal 1 1 No

Page 414 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 414 · click to enlarge
Page 414

5001 sim_cyctime 1 1 No

5002 sim_calcycle 1 1 No

7001 sim_molpct 1 1 No

7002 sim_molpct 2 1 No

7003 sim_molpct 3 1 No

7004 sim_molpct 4 1 No

7005 sim_molpct 5 1 No

7006 sim_molpct 6 1 No

7007 sim_molpct 7 1 No

7008 sim_molpct 8 1 No

7009 sim_molpct 9 1 No

7010 sim_molpct 10 1 No

7011 sim_molpct 11 1 No

7012 sim_molpct 12 1 No

7013 sim_molpct 13 1 No

7014 sim_molpct 14 1 No

7015 sim_molpct 15 1 No

7016 sim_molpct 16 1 No

7017 sim_gpmwt 1 1 No

7018 sim_gpmwt 2 1 No

7019 sim_gpmwt 3 1 No

7020 sim_gpmwt 4 1 No

7021 sim_gpmwt 5 1 No

7022 sim_gpmwt 6 1 No

7023 sim_gpmwt 7 1 No

7024 sim_gpmwt 8 1 No

7025 sim_gpmwt 9 1 No

7026 sim_gpmwt 10 1 No

7027 sim_gpmwt 11 1 No

7028 sim_gpmwt 12 1 No

7029 sim_gpmwt 13 1 No

32-Bit Integer · Floating-Point

Page 415 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 415 · click to enlarge
Page 415

7030 sim_gpmwt 14 1 No

7031 sim_gpmwt 15 1 No

7032 sim_gpmwt 16 1 No

7033 sim_dbtu 1 1 No

7034 sim_sbtu 1 1 No

7035 sim_rden_gas 1 1 No

7036 sim_compress 1 1 No

7037 sim_wobbe 1 1 No

7038 sim_unnormal 1 1 No

7039 sim_totgpm 1 1 No

7040 calcval 1 1 label 1 No

7041 calcval 2 1 label 2 No

7042 calcval 3 1 No

7043 calcval 4 1 No

7044 calcval 5 1 No

7045 sim_dummy 1 1 No

7046 sim_dummy 1 1 No

7047 sim_dummy 1 1 No

7048 sim_dummy 1 1 No

7049 sim_dummy 1 1 No

7050 sim_dummy 1 1 No

7051 sim_dummy 1 1 No

7052 sim_dummy 1 1 No

7053 sim_dummy 1 1 No

7054 sim_abtu 1 1 No

7055 avgs avg 1 1 Not Used No

7056 avgs avg 2 1 Not Used No

7057 avgs avg 3 1 Not Used No

7058 avgs avg 4 1 Not Used No

7059 avgs avg 5 1 Not Used No

7060 avgs avg 6 1 Not Used No

7061 avgs avg 7 1 Not Used No

7062 avgs avg 8 1 Not Used No

Page 416 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 416 · click to enlarge
Page 416

7063 avgs avg 9 1 Not Used No

7064 avgs avg 10 1 Not Used No

7065 avgs avg 11 1 Not Used No

7066 avgs avg 12 1 Not Used No

7067 avgs avg 13 1 Not Used No

7068 avgs avg 14 1 Not Used No

7069 avgs avg 15 1 Not Used No

7070 avg_archive1 avg 1 1 Not Used No

7071 avg_archive1 avg 2 1 Not Used No

7072 avg_archive1 avg 3 1 Not Used No

7073 avg_archive1 avg 4 1 Not Used No

7074 avg_archive1 avg 5 1 Not Used No

7075 avg_archive1 avg 6 1 Not Used No

7076 avg_archive1 avg 7 1 Not Used No

7077 avg_archive1 avg 8 1 Not Used No

7078 avg_archive1 avg 9 1 Not Used No

7079 avg_archive1 avg 10 1 Not Used No

7080 avg_archive1 avg 11 1 Not Used No

7081 avg_archive1 avg 12 1 Not Used No

7082 avg_archive1 avg 13 1 Not Used No

7083 avg_archive1 avg 14 1 Not Used No

7084 avg_archive1 avg 15 1 Not Used No

7085 aincur 1 1 AI 1 No

7086 aincur 2 1 No

7087 sim_cal_abtu 1 1 No

7088 sim_cal_dbtu 1 1 No

7089 sim_cal_sbtu 1 1 No

7090 sim_cal_wobb 1 1 No

7091 sim_cal_rden 1 1 No

7092 sim_cal_comp 1 1 No

7093 sim_cal_gpm 1 1 No

7094 sim_cal_unno 1 1 No

7095 sim_rf 1 1 No

Page 417 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 417 · click to enlarge
Page 417

7096 sim_rf 2 1 No

7097 sim_rf 3 1 No

7098 sim_rf 4 1 No

7099 sim_rf 5 1 No

7100 sim_rf 6 1 No

7101 sim_rf 7 1 No

7102 sim_rf 8 1 No

7103 sim_rf 9 1 No

7104 sim_rf 10 1 No

7105 sim_rf 11 1 No

7106 sim_rf 12 1 No

7107 sim_rf 13 1 No

7108 sim_rf 14 1 No

7109 sim_rf 15 1 No

7110 sim_rf 16 1 No

7111 sim_rf 1 1 No

7112 sim_rf 2 1 No

7113 sim_rf 3 1 No

7114 sim_rf 4 1 No

7115 sim_rf 5 1 No

7116 sim_rf 6 1 No

7117 sim_rf 7 1 No

7118 sim_rf 8 1 No

7119 sim_rf 9 1 No

7120 sim_rf 10 1 No

7121 sim_rf 11 1 No

7122 sim_rf 12 1 No

7123 sim_rf 13 1 No

7124 sim_rf 14 1 No

7125 sim_rf 15 1 No

7126 sim_rf 16 1 No

7127 avgs avg 1 1 Not Used No

7128 avgs avg 2 1 Not Used No

Page 419 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 419 · click to enlarge
Page 419

7162 avgs avg 36 1 Not Used No

7163 avgs maximum 1 1 Not Used No

7164 avgs maximum 2 1 Not Used No

7165 avgs maximum 3 1 Not Used No

7166 avgs maximum 4 1 Not Used No

7167 avgs maximum 5 1 Not Used No

7168 avgs maximum 6 1 Not Used No

7169 avgs maximum 7 1 Not Used No

7170 avgs maximum 8 1 Not Used No

7171 avgs maximum 9 1 Not Used No

7172 avgs maximum 10 1 Not Used No

7173 avgs maximum 11 1 Not Used No

7174 avgs maximum 12 1 Not Used No

7175 avgs maximum 13 1 Not Used No

7176 avgs maximum 14 1 Not Used No

7177 avgs maximum 15 1 Not Used No

7178 avgs maximum 16 1 Not Used No

7179 avgs maximum 17 1 Not Used No

7180 avgs maximum 18 1 Not Used No

7181 avgs maximum 19 1 Not Used No

7182 avgs maximum 20 1 Not Used No

7183 avgs maximum 21 1 Not Used No

7184 avgs maximum 22 1 Not Used No

7185 avgs maximum 23 1 Not Used No

7186 avgs maximum 24 1 Not Used No

7187 avgs maximum 25 1 Not Used No

7188 avgs maximum 26 1 Not Used No

7189 avgs maximum 27 1 Not Used No

7190 avgs maximum 28 1 Not Used No

7191 avgs maximum 29 1 Not Used No

7192 avgs maximum 30 1 Not Used No

7193 avgs maximum 31 1 Not Used No

7194 avgs maximum 32 1 Not Used No

Page 420 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 420 · click to enlarge
Page 420

7195 avgs maximum 33 1 Not Used No

7196 avgs maximum 34 1 Not Used No

7197 avgs maximum 35 1 Not Used No

7198 avgs maximum 36 1 Not Used No

7199 avgs minimum 1 1 Not Used No

7200 avgs minimum 2 1 Not Used No

7201 avgs minimum 3 1 Not Used No

7202 avgs minimum 4 1 Not Used No

7203 avgs minimum 5 1 Not Used No

7204 avgs minimum 6 1 Not Used No

7205 avgs minimum 7 1 Not Used No

7206 avgs minimum 8 1 Not Used No

7207 avgs minimum 9 1 Not Used No

7208 avgs minimum 10 1 Not Used No

7209 avgs minimum 11 1 Not Used No

7210 avgs minimum 12 1 Not Used No

7211 avgs minimum 13 1 Not Used No

7212 avgs minimum 14 1 Not Used No

7213 avgs minimum 15 1 Not Used No

7214 avgs minimum 16 1 Not Used No

7215 avgs minimum 17 1 Not Used No

7216 avgs minimum 18 1 Not Used No

7217 avgs minimum 19 1 Not Used No

7218 avgs minimum 20 1 Not Used No

7219 avgs minimum 21 1 Not Used No

7220 avgs minimum 22 1 Not Used No

7221 avgs minimum 23 1 Not Used No

7222 avgs minimum 24 1 Not Used No

7223 avgs minimum 25 1 Not Used No

7224 avgs minimum 26 1 Not Used No

7225 avgs minimum 27 1 Not Used No

7226 avgs minimum 28 1 Not Used No

7227 avgs minimum 29 1 Not Used No

Page 421 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 421 · click to enlarge
Page 421

7228 avgs minimum 30 1 Not Used No

7229 avgs minimum 31 1 Not Used No

7230 avgs minimum 32 1 Not Used No

7231 avgs minimum 33 1 Not Used No

7232 avgs minimum 34 1 Not Used No

7233 avgs minimum 35 1 Not Used No

7234 avgs minimum 36 1 Not Used No

7235 avg_archive1 avg 1 1 Not Used No

7236 avg_archive1 avg 2 1 Not Used No

7237 avg_archive1 avg 3 1 Not Used No

7238 avg_archive1 avg 4 1 Not Used No

7239 avg_archive1 avg 5 1 Not Used No

7240 avg_archive1 avg 6 1 Not Used No

7241 avg_archive1 avg 7 1 Not Used No

7242 avg_archive1 avg 8 1 Not Used No

7243 avg_archive1 avg 9 1 Not Used No

7244 avg_archive1 avg 10 1 Not Used No

7245 avg_archive1 avg 11 1 Not Used No

7246 avg_archive1 avg 12 1 Not Used No

7247 avg_archive1 avg 13 1 Not Used No

7248 avg_archive1 avg 14 1 Not Used No

7249 avg_archive1 avg 15 1 Not Used No

7250 avg_archive1 avg 16 1 Not Used No

7251 avg_archive1 avg 17 1 Not Used No

7252 avg_archive1 avg 18 1 Not Used No

7253 avg_archive1 avg 19 1 Not Used No

7254 avg_archive1 avg 20 1 Not Used No

7255 avg_archive1 avg 21 1 Not Used No

7256 avg_archive1 avg 22 1 Not Used No

7257 avg_archive1 avg 23 1 Not Used No

7258 avg_archive1 avg 24 1 Not Used No

7259 avg_archive1 avg 25 1 Not Used No

7260 avg_archive1 avg 26 1 Not Used No

Page 422 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 422 · click to enlarge
Page 422

7261 avg_archive1 avg 27 1 Not Used No

7262 avg_archive1 avg 28 1 Not Used No

7263 avg_archive1 avg 29 1 Not Used No

7264 avg_archive1 avg 30 1 Not Used No

7265 avg_archive1 avg 31 1 Not Used No

7266 avg_archive1 avg 32 1 Not Used No

7267 avg_archive1 avg 33 1 Not Used No

7268 avg_archive1 avg 34 1 Not Used No

7269 avg_archive1 avg 35 1 Not Used No

7270 avg_archive1 avg 36 1 Not Used No

7271 avg_archive1 maximum 1 1 Not Used No

7272 avg_archive1 maximum 2 1 Not Used No

7273 avg_archive1 maximum 3 1 Not Used No

7274 avg_archive1 maximum 4 1 Not Used No

7275 avg_archive1 maximum 5 1 Not Used No

7276 avg_archive1 maximum 6 1 Not Used No

7277 avg_archive1 maximum 7 1 Not Used No

7278 avg_archive1 maximum 8 1 Not Used No

7279 avg_archive1 maximum 9 1 Not Used No

7280 avg_archive1 maximum 10 1 Not Used No

7281 avg_archive1 maximum 11 1 Not Used No

7282 avg_archive1 maximum 12 1 Not Used No

7283 avg_archive1 maximum 13 1 Not Used No

7284 avg_archive1 maximum 14 1 Not Used No

7285 avg_archive1 maximum 15 1 Not Used No

7286 avg_archive1 maximum 16 1 Not Used No

7287 avg_archive1 maximum 17 1 Not Used No

7288 avg_archive1 maximum 18 1 Not Used No

7289 avg_archive1 maximum 19 1 Not Used No

7290 avg_archive1 maximum 20 1 Not Used No

7291 avg_archive1 maximum 21 1 Not Used No

7292 avg_archive1 maximum 22 1 Not Used No

7293 avg_archive1 maximum 23 1 Not Used No

Page 423 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 423 · click to enlarge
Page 423

7294 avg_archive1 maximum 24 1 Not Used No

7295 avg_archive1 maximum 25 1 Not Used No

7296 avg_archive1 maximum 26 1 Not Used No

7297 avg_archive1 maximum 27 1 Not Used No

7298 avg_archive1 maximum 28 1 Not Used No

7299 avg_archive1 maximum 29 1 Not Used No

7300 avg_archive1 maximum 30 1 Not Used No

7301 avg_archive1 maximum 31 1 Not Used No

7302 avg_archive1 maximum 32 1 Not Used No

7303 avg_archive1 maximum 33 1 Not Used No

7304 avg_archive1 maximum 34 1 Not Used No

7305 avg_archive1 maximum 35 1 Not Used No

7306 avg_archive1 maximum 36 1 Not Used No

7307 avg_archive1 minimum 1 1 Not Used No

7308 avg_archive1 minimum 2 1 Not Used No

7309 avg_archive1 minimum 3 1 Not Used No

7310 avg_archive1 minimum 4 1 Not Used No

7311 avg_archive1 minimum 5 1 Not Used No

7312 avg_archive1 minimum 6 1 Not Used No

7313 avg_archive1 minimum 7 1 Not Used No

7314 avg_archive1 minimum 8 1 Not Used No

7315 avg_archive1 minimum 9 1 Not Used No

7316 avg_archive1 minimum 10 1 Not Used No

7317 avg_archive1 minimum 11 1 Not Used No

7318 avg_archive1 minimum 12 1 Not Used No

7319 avg_archive1 minimum 13 1 Not Used No

7320 avg_archive1 minimum 14 1 Not Used No

7321 avg_archive1 minimum 15 1 Not Used No

7322 avg_archive1 minimum 16 1 Not Used No

7323 avg_archive1 minimum 17 1 Not Used No

7324 avg_archive1 minimum 18 1 Not Used No

7325 avg_archive1 minimum 19 1 Not Used No

7326 avg_archive1 minimum 20 1 Not Used No

Page 424 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 424 · click to enlarge
Page 424

7327 avg_archive1 minimum 21 1 Not Used No

7328 avg_archive1 minimum 22 1 Not Used No

7329 avg_archive1 minimum 23 1 Not Used No

7330 avg_archive1 minimum 24 1 Not Used No

7331 avg_archive1 minimum 25 1 Not Used No

7332 avg_archive1 minimum 26 1 Not Used No

7333 avg_archive1 minimum 27 1 Not Used No

7334 avg_archive1 minimum 28 1 Not Used No

7335 avg_archive1 minimum 29 1 Not Used No

7336 avg_archive1 minimum 30 1 Not Used No

7337 avg_archive1 minimum 31 1 Not Used No

7338 avg_archive1 minimum 32 1 Not Used No

7339 avg_archive1 minimum 33 1 Not Used No

7340 avg_archive1 minimum 34 1 Not Used No

7341 avg_archive1 minimum 35 1 Not Used No

7342 avg_archive1 minimum 36 1 Not Used No

7343 avg_archive1 avg 1 2 Not Used No

7344 avg_archive1 avg 2 2 Not Used No

7345 avg_archive1 avg 3 2 Not Used No

7346 avg_archive1 avg 4 2 Not Used No

7347 avg_archive1 avg 5 2 Not Used No

7348 avg_archive1 avg 6 2 Not Used No

7349 avg_archive1 avg 7 2 Not Used No

7350 avg_archive1 avg 8 2 Not Used No

7351 avg_archive1 avg 9 2 Not Used No

7352 avg_archive1 avg 10 2 Not Used No

7353 avg_archive1 avg 11 2 Not Used No

7354 avg_archive1 avg 12 2 Not Used No

7355 avg_archive1 avg 13 2 Not Used No

7356 avg_archive1 avg 14 2 Not Used No

7357 avg_archive1 avg 15 2 Not Used No

7358 avg_archive1 avg 16 2 Not Used No

7359 avg_archive1 avg 17 2 Not Used No

Page 425 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 425 · click to enlarge
Page 425

7360 avg_archive1 avg 18 2 Not Used No

7361 avg_archive1 avg 19 2 Not Used No

7362 avg_archive1 avg 20 2 Not Used No

7363 avg_archive1 avg 21 2 Not Used No

7364 avg_archive1 avg 22 2 Not Used No

7365 avg_archive1 avg 23 2 Not Used No

7366 avg_archive1 avg 24 2 Not Used No

7367 avg_archive1 avg 25 2 Not Used No

7368 avg_archive1 avg 26 2 Not Used No

7369 avg_archive1 avg 27 2 Not Used No

7370 avg_archive1 avg 28 2 Not Used No

7371 avg_archive1 avg 29 2 Not Used No

7372 avg_archive1 avg 30 2 Not Used No

7373 avg_archive1 avg 31 2 Not Used No

7374 avg_archive1 avg 32 2 Not Used No

7375 avg_archive1 avg 33 2 Not Used No

7376 avg_archive1 avg 34 2 Not Used No

7377 avg_archive1 avg 35 2 Not Used No

7378 avg_archive1 avg 36 2 Not Used No

7379 avg_archive1 maximum 1 2 Not Used No

7380 avg_archive1 maximum 2 2 Not Used No

7381 avg_archive1 maximum 3 2 Not Used No

7382 avg_archive1 maximum 4 2 Not Used No

7383 avg_archive1 maximum 5 2 Not Used No

7384 avg_archive1 maximum 6 2 Not Used No

7385 avg_archive1 maximum 7 2 Not Used No

7386 avg_archive1 maximum 8 2 Not Used No

7387 avg_archive1 maximum 9 2 Not Used No

7388 avg_archive1 maximum 10 2 Not Used No

7389 avg_archive1 maximum 11 2 Not Used No

7390 avg_archive1 maximum 12 2 Not Used No

7391 avg_archive1 maximum 13 2 Not Used No

7392 avg_archive1 maximum 14 2 Not Used No

Page 426 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 426 · click to enlarge
Page 426

7393 avg_archive1 maximum 15 2 Not Used No

7394 avg_archive1 maximum 16 2 Not Used No

7395 avg_archive1 maximum 17 2 Not Used No

7396 avg_archive1 maximum 18 2 Not Used No

7397 avg_archive1 maximum 19 2 Not Used No

7398 avg_archive1 maximum 20 2 Not Used No

7399 avg_archive1 maximum 21 2 Not Used No

7400 avg_archive1 maximum 22 2 Not Used No

7401 avg_archive1 maximum 23 2 Not Used No

7402 avg_archive1 maximum 24 2 Not Used No

7403 avg_archive1 maximum 25 2 Not Used No

7404 avg_archive1 maximum 26 2 Not Used No

7405 avg_archive1 maximum 27 2 Not Used No

7406 avg_archive1 maximum 28 2 Not Used No

7407 avg_archive1 maximum 29 2 Not Used No

7408 avg_archive1 maximum 30 2 Not Used No

7409 avg_archive1 maximum 31 2 Not Used No

7410 avg_archive1 maximum 32 2 Not Used No

7411 avg_archive1 maximum 33 2 Not Used No

7412 avg_archive1 maximum 34 2 Not Used No

7413 avg_archive1 maximum 35 2 Not Used No

7414 avg_archive1 maximum 36 2 Not Used No

7415 avg_archive1 minimum 1 2 Not Used No

7416 avg_archive1 minimum 2 2 Not Used No

7417 avg_archive1 minimum 3 2 Not Used No

7418 avg_archive1 minimum 4 2 Not Used No

7419 avg_archive1 minimum 5 2 Not Used No

7420 avg_archive1 minimum 6 2 Not Used No

7421 avg_archive1 minimum 7 2 Not Used No

7422 avg_archive1 minimum 8 2 Not Used No

7423 avg_archive1 minimum 9 2 Not Used No

7424 avg_archive1 minimum 10 2 Not Used No

7425 avg_archive1 minimum 11 2 Not Used No

Page 427 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 427 · click to enlarge
Page 427

7426 avg_archive1 minimum 12 2 Not Used No

7427 avg_archive1 minimum 13 2 Not Used No

7428 avg_archive1 minimum 14 2 Not Used No

7429 avg_archive1 minimum 15 2 Not Used No

7430 avg_archive1 minimum 16 2 Not Used No

7431 avg_archive1 minimum 17 2 Not Used No

7432 avg_archive1 minimum 18 2 Not Used No

7433 avg_archive1 minimum 19 2 Not Used No

7434 avg_archive1 minimum 20 2 Not Used No

7435 avg_archive1 minimum 21 2 Not Used No

7436 avg_archive1 minimum 22 2 Not Used No

7437 avg_archive1 minimum 23 2 Not Used No

7438 avg_archive1 minimum 24 2 Not Used No

7439 avg_archive1 minimum 25 2 Not Used No

7440 avg_archive1 minimum 26 2 Not Used No

7441 avg_archive1 minimum 27 2 Not Used No

7442 avg_archive1 minimum 28 2 Not Used No

7443 avg_archive1 minimum 29 2 Not Used No

7444 avg_archive1 minimum 30 2 Not Used No

7445 avg_archive1 minimum 31 2 Not Used No

7446 avg_archive1 minimum 32 2 Not Used No

7447 avg_archive1 minimum 33 2 Not Used No

7448 avg_archive1 minimum 34 2 Not Used No

7449 avg_archive1 minimum 35 2 Not Used No

7450 avg_archive1 minimum 36 2 Not Used No

7451 avg_archive1 avg 1 3 Not Used No

7452 avg_archive1 avg 2 3 Not Used No

7453 avg_archive1 avg 3 3 Not Used No

7454 avg_archive1 avg 4 3 Not Used No

7455 avg_archive1 avg 5 3 Not Used No

7456 avg_archive1 avg 6 3 Not Used No

7457 avg_archive1 avg 7 3 Not Used No

7458 avg_archive1 avg 8 3 Not Used No

Page 428 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 428 · click to enlarge
Page 428

7459 avg_archive1 avg 9 3 Not Used No

7460 avg_archive1 avg 10 3 Not Used No

7461 avg_archive1 avg 11 3 Not Used No

7462 avg_archive1 avg 12 3 Not Used No

7463 avg_archive1 avg 13 3 Not Used No

7464 avg_archive1 avg 14 3 Not Used No

7465 avg_archive1 avg 15 3 Not Used No

7466 avg_archive1 avg 16 3 Not Used No

7467 avg_archive1 avg 17 3 Not Used No

7468 avg_archive1 avg 18 3 Not Used No

7469 avg_archive1 avg 19 3 Not Used No

7470 avg_archive1 avg 20 3 Not Used No

7471 avg_archive1 avg 21 3 Not Used No

7472 avg_archive1 avg 22 3 Not Used No

7473 avg_archive1 avg 23 3 Not Used No

7474 avg_archive1 avg 24 3 Not Used No

7475 avg_archive1 avg 25 3 Not Used No

7476 avg_archive1 avg 26 3 Not Used No

7477 avg_archive1 avg 27 3 Not Used No

7478 avg_archive1 avg 28 3 Not Used No

7479 avg_archive1 avg 29 3 Not Used No

7480 avg_archive1 avg 30 3 Not Used No

7481 avg_archive1 avg 31 3 Not Used No

7482 avg_archive1 avg 32 3 Not Used No

7483 avg_archive1 avg 33 3 Not Used No

7484 avg_archive1 avg 34 3 Not Used No

7485 avg_archive1 avg 35 3 Not Used No

7486 avg_archive1 avg 36 3 Not Used No

7487 avg_archive1 maximum 1 3 Not Used No

7488 avg_archive1 maximum 2 3 Not Used No

7489 avg_archive1 maximum 3 3 Not Used No

7490 avg_archive1 maximum 4 3 Not Used No

7491 avg_archive1 maximum 5 3 Not Used No

Page 429 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 429 · click to enlarge
Page 429

7492 avg_archive1 maximum 6 3 Not Used No

7493 avg_archive1 maximum 7 3 Not Used No

7494 avg_archive1 maximum 8 3 Not Used No

7495 avg_archive1 maximum 9 3 Not Used No

7496 avg_archive1 maximum 10 3 Not Used No

7497 avg_archive1 maximum 11 3 Not Used No

7498 avg_archive1 maximum 12 3 Not Used No

7499 avg_archive1 maximum 13 3 Not Used No

7500 avg_archive1 maximum 14 3 Not Used No

7501 avg_archive1 maximum 15 3 Not Used No

7502 avg_archive1 maximum 16 3 Not Used No

7503 avg_archive1 maximum 17 3 Not Used No

7504 avg_archive1 maximum 18 3 Not Used No

7505 avg_archive1 maximum 19 3 Not Used No

7506 avg_archive1 maximum 20 3 Not Used No

7507 avg_archive1 maximum 21 3 Not Used No

7508 avg_archive1 maximum 22 3 Not Used No

7509 avg_archive1 maximum 23 3 Not Used No

7510 avg_archive1 maximum 24 3 Not Used No

7511 avg_archive1 maximum 25 3 Not Used No

7512 avg_archive1 maximum 26 3 Not Used No

7513 avg_archive1 maximum 27 3 Not Used No

7514 avg_archive1 maximum 28 3 Not Used No

7515 avg_archive1 maximum 29 3 Not Used No

7516 avg_archive1 maximum 30 3 Not Used No

7517 avg_archive1 maximum 31 3 Not Used No

7518 avg_archive1 maximum 32 3 Not Used No

7519 avg_archive1 maximum 33 3 Not Used No

7520 avg_archive1 maximum 34 3 Not Used No

7521 avg_archive1 maximum 35 3 Not Used No

7522 avg_archive1 maximum 36 3 Not Used No

7523 avg_archive1 minimum 1 3 Not Used No

7524 avg_archive1 minimum 2 3 Not Used No

Page 430 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A.2 Example Report
Page 430 · click to enlarge
Page 430

7525 avg_archive1 minimum 3 3 Not Used No

7526 avg_archive1 minimum 4 3 Not Used No

7527 avg_archive1 minimum 5 3 Not Used No

7528 avg_archive1 minimum 6 3 Not Used No

7529 avg_archive1 minimum 7 3 Not Used No

7530 avg_archive1 minimum 8 3 Not Used No

7531 avg_archive1 minimum 9 3 Not Used No

7532 avg_archive1 minimum 10 3 Not Used No

7533 avg_archive1 minimum 11 3 Not Used No

7534 avg_archive1 minimum 12 3 Not Used No

7535 avg_archive1 minimum 13 3 Not Used No

7536 avg_archive1 minimum 14 3 Not Used No

7537 avg_archive1 minimum 15 3 Not Used No

7538 avg_archive1 minimum 16 3 Not Used No

7539 avg_archive1 minimum 17 3 Not Used No

7540 avg_archive1 minimum 18 3 Not Used No

7541 avg_archive1 minimum 19 3 Not Used No

7542 avg_archive1 minimum 20 3 Not Used No

7543 avg_archive1 minimum 21 3 Not Used No

7544 avg_archive1 minimum 22 3 Not Used No

7545 avg_archive1 minimum 23 3 Not Used No

7546 avg_archive1 minimum 24 3 Not Used No

7547 avg_archive1 minimum 25 3 Not Used No

7548 avg_archive1 minimum 26 3 Not Used No

7549 avg_archive1 minimum 27 3 Not Used No

7550 avg_archive1 minimum 28 3 Not Used No

7551 avg_archive1 minimum 29 3 Not Used No

7552 avg_archive1 minimum 30 3 Not Used No

7553 avg_archive1 minimum 31 3 Not Used No

7554 avg_archive1 minimum 32 3 Not Used No

7555 avg_archive1 minimum 33 3 Not Used No

7556 avg_archive1 minimum 34 3 Not Used No

7557 avg_archive1 minimum 35 3 Not Used No

Page 431 of the Emerson Process Management Chromatograph Software (MON2000) user manual — PC CONFIG REPORT A-61 MON2000
Page 431 · click to enlarge
Page 431

PC CONFIG REPORT A-61 MON2000

7558 avg_archive1 minimum 36 3 Not Used No

TCP/IP Report from Model 2350A 4/28/2005 2:00:51 PM

Usage PC Comm ID 0 Host Name DANIELGC Use DHCP/Specify an IP address Specify an IP Address IP Address 172.16.23.128 Subnet Mask 255.255.240.0 Gateway 172.16.16.2 Read/Write RW *************************************************************************** *****

Temperature Control Report from Model 2350A 4/28/2005 2:00:52 PM

Loop 1 2 3 4 Mode Automatic Prop Gain 60 Reset 600 Rate 1 D Filter 5 Setpoint 82 84 40 350 *************************************************************************** ***** End of Report

JULY 2010 Example Report

*************************************************************************** ******* · *************************************************************************** *****

Page 432 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A-62 PC CONFIG REPORT
Page 432 · click to enlarge
Page 432

A-62 PC CONFIG REPORT

This page is intentionally left blank.

Example Report JULY 2010

Page 433 of the Emerson Process Management Chromatograph Software (MON2000) user manual — JULY 2010 B-1
Page 433 · click to enlarge
Page 433

A APPENDIX B: COMPONENT DATA TABLE

This appendix provides a sample standard component data table as well as a table of the ISO-related components.

  • Table B-1, Example Standard Component Data Table
  • Table B-2, ISO Component Data Table

All values depend on a base pressure of 14.73 PSIA and a base temperature of 60 oF (15.56 oC).

BTU components that are listed in Table B-1 reference GPA Standard 2145-09. The ISO component data table references ISO Standard 6976: 1995(E).

JULY 2010 B-1

Page 434 of the Emerson Process Management Chromatograph Software (MON2000) user manual — BTU AGA 8
Page 434 · click to enlarge
Page 434

B-2

MON 2000 Software for Gas Chromatographs

I.D. No.

Acetylene 26.04 0 0.899 0.615 0 0 1476.9 1426.5 Ethane 22

Air 28.9625 0 1 0.87586 7.3022 0.104759 0 0 Airmix1 26

Argon 39.95 0 1.3792 0 Argon 46

Ammonia 17.03 212 0.588 0.6173 5.15 0.0874 435.4 359.8 None *

Benzene 78.11 3.224 2.6969 0.8844 7.373 0.2798 3750.5 3599.2 n-Hexane *

Butanes 58.1222 62.1055 2.0068 0.573515 4.78155 0.32117 3264.64 3012.45 n-Butane 33

Butene-1 56.11 63.05 1.9372 0.6013 5.013 0.2956 3087 2885.4 n-Butane 28

Butenes 56.11 55.448 1.9372 0.6097 5.0833 0.2916 3077.4 2875.73 n-Butane 32

1,2-Butadiene 54.09 20 1.8676 0.658 5.486 0.2604 2946.7 2795.5 n-Butane 35

1,3-Butadiene 54.09 60 1.8676 0.6272 5.229 0.2732 2886.6 2735.3 n-Butane 34

C3+ 44.0956 188.62 1.5225 0.50719 4.2285 0.275429 2521.92 2320.36 Propane 47

C4+ 58.1222 51.567 2.0068 0.5842 4.8706 0.315183 3269.85 3017.97 n-Butane 48

C4=1 56.11 63.05 1.9372 0.6013 5.013 0.2956 3087 2885.4 n-Butane 29

C5+ 72.1488 15.576 2.4911 0.63071 5.2584 0.362396 4017.97 3715.58 n-Pentane 49

C6+ 47/35/17 95.9558 3.01891 3.31309 0.679907 5.66853 0.446214 5288.71 4900.62 C6mix1 08

C6+ 50/50/00 93.1887 3.29 3.21755 0.676145 5.63715 0.43619 5141.12 4762.99 C6mix2 09

C6+ Gpa 2261-99 93.1887 3.51579 3.21755 0.67556 5.63228 0.436267 5141.09 4762.99 C6mix3 10

BTU AGA 8

Table B-1 Example Standard Component Data Table

Component Data Table JULY 2010

Show 11 more labels / short text on this page
Sim 2251 · Daniel · Component · BTU Net Dry · Factor Gross Dry · Liquid Lb/Gal GPM · Gas Rel Dens · Vapor Rel Dens · Wt Reid · Name Mol · Component
Page 435 of the Emerson Process Management Chromatograph Software (MON2000) user manual — BTU AGA 8
Page 435 · click to enlarge
Page 435

B-3 MON 2000 Software for Gas Chromatographs

I.D. No.

C6+ 57/28/14 94.1904 3.37386 3.25214 0.677036 5.64458 0.439881 5194.53 4812.82 C6mix4 11

Carbon Monoxide 28.01 0 0.9671 0.801 6.68 0 321.2 321.2 CO 15

Carbon Dioxide 44.0095 0 1.5195 0.81716 6.8129 0.170618 0 0 CO2 17

Cis-2-Butene 56.11 45.54 1.9372 0.6271 5.228 0.2835 3079.3 2877.6 n-Butane 31

COS 60.08 0 None 42

CS2 76.14 0 2.6298 0 0 0 1267 1267 None 41

Ethane 30.069 800 1.0382 0.35628 2.9704 0.267369 1773.79 1622.75 Ethane 01

Ethylene 28.0532 0 0.9686 0 0 0 1603.4 1502.47 Ethane 21

Ethylene Oxide 44.05 0 1.49 0 0 0 1459.4 1410.2 None 36

Cyclohexane 84.16 3.264 2.9057 0.7834 6.531 0.3403 4492.1 4189.4 n-Hexane *

Cyclopentane 70.14 9.914 2.4215 0.7504 6.256 0.2961 3772.4 3520.2 n-Pentane *

Diisobutyl 114.23 1.101 3.9439 0.6979 5.819 0.5185 6247.9 5793.9 n-Octane *

2,3-Dimethbutan 86.18 7.404 2.9753 0.6664 5.556 0.4096 4756 4403.1 n-Hexane *

2,2-Dimethpenta 100.21 3.492 3.4596 0.6782 5.654 0.4682 5494.6 5091.4 n-Heptane *

2,4-Dimethpenta 100.21 3.292 3.4596 0.6773 5.647 0.4686 5499.4 5096 n-Heptane *

3,3-Dimethpenta 100.2 2.773 3.4596 0.6976 5.816 0.455 5501.5 5098.2 n-Heptane *

Ethyl Alcohol 46.07 2.3 1.5906 0.794 6.62 0.1839 1602.8 1451.5 None *

Ethylbenzene 106.17 0.371 3.6655 0.8718 7.268 0.3858 5234.3 4982 n-Octane *

3-Ethylpentane 100.21 2.012 3.4596 0.7028 5.859 0.4517 5513.4 5110.1 n-Heptane *

BTU AGA 8

Table B-1 Example Standard Component Data Table (Continued)

JULY 2010 Component Data Table

Show 11 more labels / short text on this page
Sim 2251 · Daniel · Component · BTU Net Dry · Factor Gross Dry · Liquid Lb/Gal GPM · Gas Rel Dens · Vapor Rel Dens · Wt Reid · Name Mol · Component
Page 436 of the Emerson Process Management Chromatograph Software (MON2000) user manual — BTU AGA 8
Page 436 · click to enlarge
Page 436

B-4

MON 2000 Software for Gas Chromatographs

I.D. No.

H2S 34.0809 395 1.1767 0.79886 6.6602 0.135156 638.57 588.15 H2S 40

HCL 36.46 925 1.2588 0.8558 7.135 0.1349 0 0 None *

Helium 4.0026 0 0.1382 0.12486 1.041 0.101559 0 0 Helium 13

Hydrogen 2.02 0 0.0696 0.07 0 0 325 274.4 Hydrogen 12

i-Butane 58.1222 72.644 2.0068 0.56283 4.6925 0.327158 3259.42 3006.94 i-Butane 03

i-Butene 56.11 63.4 1.9372 0.6004 5.006 0.296 3068.2 2866.5 n-Butane 27

i-Pentane 72.1488 20.474 2.4911 0.62514 5.212 0.365621 4010.16 3707.56 i-Pentane 05

i-Propylbenzene 120.19 0.188 4.1498 0.8663 7.223 0.4396 5976.6 5674 n-Nonane *

i-Octane 114.23 1.708 3.9439 0.6962 5.804 0.5199 6246.1 5792.2 n-Octane *

Methane 16.0425 5000 0.5539 0.3 2.5 0.169487 1012.34 911.5 Methane 00

Methyl Alcohol 32.04 4.63 1.1063 0.796 6.64 0.1275 868.7 767.9 None *

Methylcyclo C5 84.16 4.503 2.9057 0.7536 6.283 0.3538 4511.6 4209.1 n-Hexane *

Methylcyclo C6 98.19 1.609 3.39 0.774 6.453 0.4019 5228 4874.9 n-Heptane *

2-Methylhexane 100.21 2.271 3.4596 0.683 5.694 0.4647 5507.3 5104 n-Heptane *

3-Methylhexane 100.21 2.13 3.4596 0.6917 5.767 0.4589 5511.3 5107.8 n-Heptane *

m-Xylene 106.17 0.326 3.6655 0.8687 7.243 0.3871 5219.9 4967.8 n-Octane *

n-Butane 58.1222 51.567 2.0068 0.5842 4.8706 0.315183 3269.85 3017.97 n-Butane 04

n-Decane 142.2817 0.06148 4.9126 0.73458 6.1244 0.613636 7760.81 7206.63 n-Decane *

n-Heptane 100.2019 1.619 3.4597 0.68823 5.7379 0.461258 5515.33 5111.8 n-Heptane 45

BTU AGA 8

Table B-1 Example Standard Component Data Table (Continued)

Component Data Table JULY 2010

Show 11 more labels / short text on this page
Sim 2251 · Daniel · Component · BTU Net Dry · Factor Gross Dry · Liquid Lb/Gal GPM · Gas Rel Dens · Vapor Rel Dens · Wt Reid · Name Mol · Component
Page 437 of the Emerson Process Management Chromatograph Software (MON2000) user manual — BTU AGA 8
Page 437 · click to enlarge
Page 437

B-5 MON 2000 Software for Gas Chromatographs

I.D. No.

n-Hexane 86.1754 4.961 2.9754 0.66406 5.5364 0.411121 4766.9 4414.19 n-Hexane 39

n-Nonane 128.2551 0.1809 4.4283 0.72224 6.0215 0.562592 7012.49 6508.02 n-Nonane 38

n-Octane 114.2285 0.5349 3.944 0.70655 5.8907 0.512168 6263.46 5809.41 n-Octane 20

n-Pentane 72.1488 15.576 2.4911 0.63071 5.2584 0.362396 4017.97 3715.58 n-Pentane 06

Neopentane 72.15 35.9 2.4911 0.5967 4.975 0.383 3993.9 3691.4 i-Pentane 07

Nitrogen 28.0134 0 0.9672 0.80687 6.7271 0.10999 0 0 Nitrogen 14

NO2 46 0 None 19

N2O 44.02 0 None 18

Oxygen 31.9988 0 1.1048 1.1423 9.5238 0.088739 0 0 Oxygen 16

1-Pentene 70.14 19.115 2.4215 0.6457 5.383 0.3441 3835.4 3583.3 n-Pentane 37

Propane 44.0956 188.62 1.5225 0.50719 4.2285 0.275429 2521.92 2320.36 Propane 02

Propadiene 40.07 0 1.411 0 0 0 2254.2 2254.2 Propane 24

Propylene 42.0797 227.3 1.4529 0.5226 4.3571 0.255087 2338.4 2187.05 Propane 23

Propyne 40.07 0 1.411 0 0 0 2246.2 2246.2 Propane 25

Sulfur Dioxide 64.06 88 2.2117 1.397 11.65 0.1453 0 0 CO2 43

Neohexane 86.18 9.856 2.9753 0.654 5.453 0.4175 4747.2 4394.1 n-Hexane *

NO 30.01 0 None *

o-Xylene 106.2 0.264 3.6655 0.8848 7.377 0.3801 5222 4969.7 n-Octane *

p-Xylene 106.17 0.342 3.6655 0.8657 7.218 0.3885 5220.8 4968.6 n-Octane *

BTU AGA 8

Table B-1 Example Standard Component Data Table (Continued)

JULY 2010 Component Data Table

Show 11 more labels / short text on this page
Sim 2251 · Daniel · Component · BTU Net Dry · Factor Gross Dry · Liquid Lb/Gal GPM · Gas Rel Dens · Vapor Rel Dens · Wt Reid · Name Mol · Component
Page 438 of the Emerson Process Management Chromatograph Software (MON2000) user manual — BTU AGA 8
Page 438 · click to enlarge
Page 438

B-6

MON 2000 Software for Gas Chromatographs

I.D. No.

Styrene 104.15 0.24 3.5959 0.911 7.595 0.3622 5042.7 4841 n-Octane *

Toluene 92.14 1.032 3.1812 0.8718 7.268 0.3348 4485.4 4283.5 n-Heptane *

Trans-2-Butene 56.11 49.8 1.9372 0.61 5.086 0.2914 3075.1 2873.4 n-Butane 30

Triptane 100.21 3.374 3.4596 0.6946 5.791 0.4571 5496.2 5093 n-Heptane *

Water 18.0153 0.9505 0.62202 1 8.3372 0.057072 50.43 0 Water 44

BTU AGA 8

Table B-1 Example Standard Component Data Table (Continued)

Component Data Table JULY 2010

Show 11 more labels / short text on this page
Sim 2251 · Daniel · Component · BTU Net Dry · Factor Gross Dry · Liquid Lb/Gal GPM · Gas Rel Dens · Vapor Rel Dens · Wt Reid · Name Mol · Component
Page 439 of the Emerson Process Management Chromatograph Software (MON2000) user manual — (20°C)
Page 439 · click to enlarge
Page 439

B-7 MON 2000 Software for Gas Chromatographs

(25°C)

Acetylene 26.038 0.0949 0.0837 0.0837 1301.86 1301.37 1301.21 1301.05 1256.79 1256.94 1256.98 1257.03

Benzene 78.114 0.3017 0.272 0.253 3305.03 3302.86 3302.15 3301.43 3169.81 3169.56 3169.48 3169.38

Butanes 58.1222 0.2059 0.183 0.1743 2879.01 2875.17 2873.98 2872.8 2653.64 2653.01 2652.86 2652.72

Butene-1 56.108 0.1871 0.1732 0.1673 2721.55 2718.7 2717.75 2716.82 2541.25 2540.97 2540.86 2540.76

Butenes 56.108 0.1923 0.176 0.1717 2713.09 2710.23 2709.31 2708.36 2532.79 2532.49 2532.42 2532.27

1,2-Butadiene 54.092 0.2121 0.1924 0.1871 2597.13 2595.12 2594.45 2593.79 2461.91 2461.82 2461.78 2461.74

1,3-Butadiene 54.092 0.1844 0.1703 0.1643 2544.13 2542.1 2541.43 2540.77 2408.91 2408.8 2408.76 2408.72

C3+ 44.0956 0.1682 0.1534 0.147 2461.51 2458.25 2457.23 2456.16 2264.71 2264.52 2264.38 2264.25

C4+ 58.1222 0.2281 0.2049 0.1947 3081.63 3077.47 3076.32 3074.97 2841.63 2841.98 2841.83 2841.68

C4=1 56.108 0.1871 0.1732 0.1673 2721.55 2718.7 2717.75 2716.82 2541.25 2540.97 2540.86 2540.76

C5+ 72.1488 0.2999 0.2651 0.2505 3754.2 3749.68 3748.71 3746.71 3464.63 3468.87 3468.75 3468.52

C6+ 47/35/17 95.9558 0.389 0.3459 0.3331 4663.16 4657.69 4655.86 4654.08 4316.22 4315.67 4315.46 4315.27

C6+ 50/50/00 93.1887 0.3704 0.3305 0.3183 4533.05 4527.71 4525.93 4524.19 4194.99 4194.46 4194.25 4194.07

C6+ GPA 2261-99 93.1887 0.3943 0.3503 0.3373 4697.93 4692.42 4690.58 4688.78 4348.61 4348.06 4347.84 4347.66

Ammonia 17.0306 0.1225 0.1095 0.1049 384.57 383.51 383.16 382.81 316.96 316.86 316.82 316.79

Air 28.9625 0

Argon 39.948 0.0316 0.0283 0.0265 0

(20°C)

(15°C)

(0°C)

(25°C)

Table B-2 ISO Component Data Table

(20°C)

(15°C)

(0°C)

(20°C)

(15°C)

(0°C)

JULY 2010 Component Data Table

Show 27 more labels / short text on this page
CV Inf · Mol · kJ/ · CV Inf · Mol · kJ/ · CV Inf · kJ/Mol · kJ/Mol · CV Inf · kJ/Mol · Sup · CV Sup · kJ/Mol · CV Sup · kJ/Mol · CV Sup · kJ/Mol · Factor · Sum · Factor · Sum · Factor · Sum · Name Molar · Mass · Component
Page 440 of the Emerson Process Management Chromatograph Software (MON2000) user manual — (20°C)
Page 440 · click to enlarge
Page 440

B-8

MON 2000 Software for Gas Chromatographs

(25°C)

C6+ 57/28/14 94.1904 0.3781 0.3367 0.3243 4580.15 4574.76 4572.96 4571.2 4238.87 4238.34 4238.12 4237.94

CS2 76.143 0.2145 0.1949 0.1871 1104.06 1104.32 1104.41 1104.49 1104.06 1104.32 1104.41 1104.49

Cyclohexane 84.161 0.3209 0.2864 0.2757 3960.67 3956.02 3954.47 3952.96 3690.23 3689.42 3689.13 3688.86

Cyclopentane 70.14 0.255 0.2302 0.2236 3326.14 3322.19 3320.88 3319.59 3100.77 3100.03 3099.76 3099.51

2,3-Dimethbutan 86.177 0.3 0.2739 0.2569 4193.63 4188.6 4186.93 4185.28 3878.11 3877.57 3877.36 3877.17

Ethane 30.069 0.1 0.0922 0.0894 1564.34 1562.14 1561.41 1560.69 1429.12 1428.84 1428.74 1428.64

Ethylbenzene 106.167 0.4858 0.4207 0.4037 4613.14 4609.53 4608.32 4607.15 4387.77 4387.37 4387.2 4387.07

Ethylene 28.0532 0.0866 0.08 0.0775 1413.51 1412.11 1411.65 1411.18 1323.36 1323.24 1323.2 1323.15

Carbon Monoxide 28.01 0.0265 0.0224 0.02 282.8 282.91 282.95 282.98 282.8 282.91 282.95 282.98

Cis-2-Butene 56.108 0.1975 0.1817 0.1761 2714.9 2711.9 2711 2710 2534.6 2534.2 2534.1 2533.9

COS 60.076 0.1225 0.114 0.1095 548.01 548.15 548.19 548.23 548.01 548.15 548.19 548.23

Carbon Dioxide 44.0095 0.0819 0.0748 0.0728 0

Diisobutyl 114.23 0

2,2-Dimethpenta 100.21 0

2,4-Dimethpenta 100.21 0

3,3-Dimethpenta 100.2 0

Ethyl Alcohol 46.07 0

Ethylene Oxide 44.05 0

(20°C)

(15°C)

(0°C)

(25°C)

Table B-2 ISO Component Data Table

(20°C)

(15°C)

(0°C)

(20°C)

(15°C)

(0°C)

Component Data Table JULY 2010

Show 27 more labels / short text on this page
CV Inf · Mol · kJ/ · CV Inf · Mol · kJ/ · kJ/Mol · CV Inf · CV Inf · kJ/Mol · kJ/Mol · Sup · CV Sup · kJ/Mol · CV Sup · kJ/Mol · CV Sup · kJ/Mol · Factor · Sum · Factor · Sum · Factor · Sum · Name Molar · Mass · Component
Page 441 of the Emerson Process Management Chromatograph Software (MON2000) user manual — (20°C)
Page 441 · click to enlarge
Page 441

B-9 MON 2000 Software for Gas Chromatographs

(25°C)

i-Butane 58.1222 0.2049 0.1789 0.1703 2874.2 2870.58 2869.38 2868.2 2648.83 2648.42 2648.26 2648.12

i-Pentane 72.1488 0.251 0.228 0.2168 3535.98 3531.68 3530.24 3528.83 3265.54 3265.08 3264.89 3264.73

Methylcyclo C6 98.188 0.3808 0.3376 0.3256 4600.64 4602.35 4604.09 4609.34 4292.53 4292.78 4293.06 4293.82

n-Butane 58.1222 0.2069 0.1871 0.1783 2883.82 2879.76 2878.57 2877.4 2658.45 2657.6 2657.45 2657.32

H2S 34.0809 0.1 0.1 0.1 562.94 562.38 562.19 562.01 517.87 517.95 517.97 517.99

Hydrogen 2.0159 -0.004 -0.0048 -0.0051 286.63 286.15 285.99 285.83 241.56 241.72 241.76 241.81

i-Butene 56.108 0.1871 0.1703 0.1673 2704.8 2702 2701.1 2700.2 2524.5 2524.3 2524.2 2524.1

Methyl Alcohol 32.042 0.4764 0.3578 0.3286 766.59 765.09 764.59 764.09 676.44 676.22 676.14 676.06

Methylcyclo C5 84.161 0.313 0.2811 0.2702 3977.04 3972.46 3970.93 3969.44 3705.34 3705.59 3705.86 3706.6

Methane 16.0425 0.049 0.0447 0.0436 892.97 891.56 891.09 890.63 802.82 802.69 802.65 802.6

3-Ethylpentane 100.21 0

HCL 36.46 925 1.2588 0.8558 7.135 0.1349 0

Helium 4.0026 0.0006 0.0002 0

i-Propylbenzene 120.19 0

i-Octane 114.23 0

2-Methylhexane 100.21 0

3-Methylhexane 100.21 0

m-Xylene 106.167 0

(20°C)

(15°C)

(0°C)

(25°C)

Table B-2 ISO Component Data Table

(20°C)

(15°C)

(0°C)

(20°C)

(15°C)

(0°C)

JULY 2010 Component Data Table

Show 27 more labels / short text on this page
CV Inf · Mol · kJ/ · CV Inf · Mol · kJ/ · CV Inf · kJ/Mol · kJ/Mol · CV Inf · kJ/Mol · Sup · CV Sup · kJ/Mol · CV Sup · kJ/Mol · CV Sup · kJ/Mol · Factor · Sum · Factor · Sum · Factor · Sum · Name Molar · Mass · Component
Page 442 of the Emerson Process Management Chromatograph Software (MON2000) user manual — (20°C)
Page 442 · click to enlarge
Page 442

B-10

MON 2000 Software for Gas Chromatographs

(25°C)

n-Decane 142.2817 0.7523 0.645 0.614 6842.69 6834.9 6832.31 6829.77 6346.88 6346.14 6345.85 6345.59

n-Hexane 86.1754 0.3286 0.295 0.2846 4203.23 4198.24 4196.58 4194.95 3887.71 3887.21 3887.01 3886.84

n-Nonane 128.2551 0.6221 0.5385 0.5148 6182.91 6175.82 6173.46 6171.15 5732.17 5731.49 5731.22 5730.99

n-Octane 114.2285 0.5079 0.445 0.4278 5522.4 5516.01 5513.88 5511.8 5116.73 5116.11 5115.87 5115.66

n-Pentane 72.1488 0.2864 0.251 0.2345 3542.89 3538.6 3537.17 3535.77 3272.45 3272 3271.83 3271.67

Neohexane 86.177 0.2898 0.2627 0.255 4185.84 4180.83 4179.15 4177.52 3870.32 3869.8 3869.59 3869.41

Neopentane 72.15 0.2387 0.2121 0.2025 3521.72 3517.43 3516.01 3514.61 3251.28 3250.83 3250.67 3250.51

o-Xylene 106.167 0.5128 0.4427 0.4231 4602.17 4598.64 4597.48 4596.31 4376.8 4376.48 4376.34 4376.23

1-Pentene 70.14 0.249 0.2258 0.2191 3381.29 3377.75 3376.57 3375.42 3155.92 3155.59 3155.45 3155.34

Propane 44.0956 0.1453 0.1338 0.1288 2224.01 2221.1 2220.13 2219.17 2043.71 2043.37 2043.23 2043.11

Propadiene 40.065 0.1414 0.1304 0.1265 1945.25 1943.96 1943.53 1943.11 1855.1 1855.09 1855.08 1855.08

Propylene 42.0797 0.1378 0.1265 0.1225 2061.57 2059.43 2058.72 2058.02 1926.35 1926.13 1926.05 1925.97

n-Heptane 100.2019 0.4123 0.3661 0.3521 4862.87 4857.18 4855.29 4853.43 4502.28 4501.72 4501.49 4501.3

Nitrogen 28.0134 0.0224 0.0173 0.0173 0

NO2 46.0006 0

NO 30.006 0

N2O 44.02 0

Oxygen 31.9988 0.0316 0.0283 0.0265 0

(20°C)

(15°C)

(0°C)

(25°C)

Table B-2 ISO Component Data Table

(20°C)

(15°C)

(0°C)

(20°C)

(15°C)

(0°C)

Component Data Table JULY 2010

Show 27 more labels / short text on this page
CV Inf · Mol · kJ/ · CV Inf · Mol · kJ/ · kJ/Mol · CV Inf · CV Inf · kJ/Mol · kJ/Mol · Sup · CV Sup · kJ/Mol · CV Sup · kJ/Mol · CV Sup · kJ/Mol · Factor · Sum · Factor · Sum · Factor · Sum · Name Molar · Mass · Component
Page 443 of the Emerson Process Management Chromatograph Software (MON2000) user manual — (20°C)
Page 443 · click to enlarge
Page 443

B-11 MON 2000 Software for Gas Chromatographs

(25°C)

Toluene 92.141 0.3886 0.3421 0.3286 3952.72 3949.81 3948.84 3947.89 3772.42 3772.08 3771.95 3771.83

Trans-2-Butene 56.108 0.1975 0.1789 0.1761 2711.1 2708.3 2707.4 2706.4 2530.8 2530.5 2530.5 2530.3

Propyne 40.065 0

p-Xylene 106.167 0

Sulfur Dioxide 64.065 0.1549 0.1449 0.1414 0

Styrene 104.15 0

Triptane 100.21 0

Water 18.0153 0.2646 0.2345 0.2191 45.074 44.433 44.224 44.016 0

(20°C)

(15°C)

(0°C)

(25°C)

Table B-2 ISO Component Data Table

(20°C)

(15°C)

(0°C)

(20°C)

(15°C)

(0°C)

JULY 2010 Component Data Table

Show 27 more labels / short text on this page
CV Inf · Mol · kJ/ · CV Inf · Mol · kJ/ · CV Inf · kJ/Mol · kJ/Mol · CV Inf · kJ/Mol · Sup · CV Sup · kJ/Mol · CV Sup · kJ/Mol · CV Sup · kJ/Mol · Factor · Sum · Factor · Sum · Factor · Sum · Name Molar · Mass · Component
Page 444 of the Emerson Process Management Chromatograph Software (MON2000) user manual — B-12
Page 444 · click to enlarge
Page 444

B-12

MON 2000 Software for Gas Chromatographs

Component Data Table JULY 2010

Page 445 of the Emerson Process Management Chromatograph Software (MON2000) user manual — N PW =
Page 445 · click to enlarge
Page 445

DATA COMPUTATIONS C-1 MON2000

CA APPENDIX C, DATA COMPUTATIONS

C.1 DATA ACQUISITION

The 2350A GC Controller performs a 12-bit analog to digital (A/D) conversion every 25 milliseconds on channel 0 during an analysis period. This is the period when the left side of the display is counting the elapsed time of the analysis. The fixed number on the right side of the display is the total cycle time.

Exactly 40 equi-spaced data samples are taken every second for analysis by the controller. A sampling frequency of 40 Hertz (Hz) was chosen to reduce 60 Hz normal mode noise. After each point on the chromatograph signal is sampled, the resulting number is stored in a buffer area in the controller memory for processing. During the analysis, only the last 256 data points are available for processing. Because the data analysis is done as the signal is sampled (in real-time), only a limited number of past data samples is required to analyze any signal.

As a part of the data acquisition process, groups of incoming data samples are averaged together before storing to the controller memory for processing. Non-overlapping groups of N samples are averaged and stored and thus reduce the effective incoming data rate to 40/N samples/second. For example, if N = 5, then a total of 40/5 or 6 (averaged) data samples are stored every second. The value for the variable N is determined by the selection of a Peak Width parameter (PW). The relationship is:

N PW =

JULY 2010 Data Acquisition

Page 446 of the Emerson Process Management Chromatograph Software (MON2000) user manual — C.2 PEAK DETECTION
Page 446 · click to enlarge
Page 446

C-2 DATA COMPUTATIONS

where PW is given in seconds. All the various details in the analysis process are independent of the value of N. Allowable values of N are one to 63 which corresponds to values of PW from 2 to 63 seconds.

The variable N is known as the integration factor. This term is used because N determines how many points are averaged or integrated to form a single value. The integration of data upon input before storing serves two purposes. First, the statistical noise on the input signal is reduced by the square root of N. In the case of N = 4, a noise reduction of two would be realized. Secondly, the integration factor controls the bandwidth of the chromatograph signal. It is necessary to match the bandwidth of the input signal to that of the analysis algorithms in the 2350A. This prevents small, short duration perturbations from being recognized as true peaks by the program. It is therefore important to choose a Peak Width corresponding to the narrowest peak in a group under consideration.

C.2 PEAK DETECTION

For normal area or peak height concentration evaluation, the determination of a peak's start, peak point, and end is automatic. The manual determination of start and end points is used only for area calculations in the Forced Integration mode. Automatic determination of peak onset or start is initiated whenever Integrate Inhibit is turned off. Analysis is started in a region of signal quiescence and stability such that the signal level and activity can be considered as baseline values. It is important that this be the case because the assumption is made by the 2350A software.

Peak Detection JULY 2010

Page 447 of the Emerson Process Management Chromatograph Software (MON2000) user manual — DATA COMPUTATIONS C-3 MON2000
Page 447 · click to enlarge
Page 447

DATA COMPUTATIONS C-3 MON2000

Having initiated a peak search by turning Inhibit off, the 2350A Controller performs a point by point examination of the signal slope. This is achieved by using a digital slope detection filter which is a combination low pass filter and differentiator. The output of this detector is constantly compared to a system constant entered by the operator called Slope Sensitivity. A default value of eight is assumed if no entry is made. Lower values make peak onset detection more sensitive and higher values make detection less sensitive. Higher values (20 to 100) would be appropriate for noisy signals, e.g. high amplifier gain. Peak termination is determined by the same application of this detector to the signal, but in the reverse sense. Onset is defined where the detector output exceeds the baseline constant, but termination is defined subsequently where the detector output is less than the same constant.

Sequences of fused peaks are also automatically handled. This is done by testing each termination point to see if the region immediately following it satisfies the criteria of a baseline. A baseline region must have a slope detector value less than the magnitude of the baseline constant for a number of sequential points. When a baseline region is found, this terminates a sequence of peaks. A zero reference line for peak height and area determination is established by extending a line from the point of the onset of the peak sequence to the point of the termination. The values of these two points are found by averaging the four integrated points just prior to the onset point and just after the termination points respectively.

JULY 2010 Peak Detection

Page 448 of the Emerson Process Management Chromatograph Software (MON2000) user manual — C-4 DATA COMPUTATIONS
Page 448 · click to enlarge
Page 448

C-4 DATA COMPUTATIONS

The zero reference line will, in general, be non- horizontal and thus compensates for any linear drift in the system from the time the peak sequence starts until it ends. In a single peak situation, peak area is the area of the component peak between the curve and the zero reference line. The peak height is the distance from the zero reference line to the maximum point on the component curve. The value and location of the maximum point is determined from quadratic interpolation through the three highest points at the peak of the discrete valued curve stored in 2350A. This interpolation technique is used both for peaks as well as valleys (minimum points) in fused peak sequences. In the latter case, lines are dropped from the interpolated valley points to the zero reference line to partition the fused peak areas into individual peaks. The use of quadratic interpolation improves both area and height calculation accuracy and eliminates the effects of variations in the integration factor on these calculations.

While calibrating, the controller may run several analyses of the calibration stream. Using entries from an example program, five analyses would be run. Since only the last three are to be averaged, data from runs one and two are not saved. Results are stored on the third run. Data from runs four and five are added to that stored for run three.

Peak Detection JULY 2010

Page 449 of the Emerson Process Management Chromatograph Software (MON2000) user manual — C.3.1 CONC. ANALYSIS WITH RESPONSE FACTOR
Page 449 · click to enlarge
Page 449

DATA COMPUTATIONS C-5 MON2000

C.3 ANALYSIS COMPUTATIONS

There are two basic analysis algorithms included in the controller:

  • Area Analysis – calculates area under component peak
  • Peak Height Analysis – measures height of component peak

C.3.1 CONC. ANALYSIS WITH RESPONSE FACTOR

Calibration

The concentration calculations discussed as follows require a unique response factor for each component in an analysis. These factors may be manually entered by an operator or automatically calculated by calibrating the system.

Equation C-1 Response Factor Calculation

Htn Caln =

Caln = HRFn

Equation C-1 uses an external standard.

ARFn area response factor for component “n” in area per mole percent (%)

HRFn height response factor for component “n”

Arean area associated with component “n” in cal- ibration gas

Htn height associated with component “n” in mole percent in calibration gas

Caln amount of component “n” in mole percent of calibration gas

JULY 2010 Analysis Computations

Show 4 more labels / short text on this page
Arean · ARFn · or · where
Page 450 of the Emerson Process Management Chromatograph Software (MON2000) user manual — k ∑
Page 450 · click to enlarge
Page 450

C-6 DATA COMPUTATIONS

Calculated response factors are stored by the controller for use in the concentration calculations, and are printed out in the configuration and calibration reports.

Equation C-2 Average Response Factor

k ∑

RFAVGn area or height average response factor for component “n”

RFi area or height response factor for component “n” from the calibration run

k number of calibration runs actually used to calculate the response factors

The percent deviation of new RF average from old RF average is calculated in the following manner:

Equation C-3 Percent Deviation

deviation RFnew RFold – Rfold 100 × =

where the absolute value of percent deviation for alarm has been previously entered by the operator.

Conc. Analysis with Response Factor JULY 2010

Show 5 more labels / short text on this page
RFi · k ------------------ = · i 1 = · RFAVGn · where
Page 451 of the Emerson Process Management Chromatograph Software (MON2000) user manual — DATA COMPUTATIONS C-7 MON2000
Page 451 · click to enlarge
Page 451

DATA COMPUTATIONS C-7 MON2000

Calc. in Mole Percent w/o Normalization

Once response factors have been determined by the controller or entered by the operator, component concentrations are determined for each analysis using the following equations:

Equation C-4 Response Factor Calculation

Htn HRFn =

ARFn = CONCn

CONCn concentration of component “n” in mole percent

AREAn area of component “n” in unknown sample

ARFn response factor of component “n” calculated from area of calibration sample. Units are area per mole percent

Htn peak height of component “n” in unknown sample

HRFn response factor of component “n” calculated from peak height of calibration sample. Units are height per mole percent

JULY 2010 Conc. Analysis with Response Factor

Show 4 more labels / short text on this page
Arean · CONCn · or · where
Page 452 of the Emerson Process Management Chromatograph Software (MON2000) user manual — k ∑
Page 452 · click to enlarge
Page 452

C-8 DATA COMPUTATIONS

Component concentrations may be input through analog inputs 1 to 4 or may be fixed. If a fixed value is used, the calibration for that component is the mole percent that will be used for all analyses:.

Equation C-5 Concentration Calculations with Normalization

100 × =

k ∑

CONCNn normalized concentration of component “n” in percent of total gas concentration

CONCn non-normalized concentration of component “n” in mole percent

CONCi non-normalized concentration (in mole percent) from each of the “k” components to be grouped into this normalization

k number of components to be included in the normalization

Conc. Analysis with Response Factor JULY 2010

Show 5 more labels / short text on this page
CONCn · CONCNn · CONCi · i 1 = · where
Page 453 of the Emerson Process Management Chromatograph Software (MON2000) user manual — DRYBTU CF
Page 453 · click to enlarge
Page 453

DATA COMPUTATIONS C-9 MON2000

C.4 POST ANALYSIS COMPUTATIONS

C.4.1 Liquid Equivalent Computations

The equivalent liquid volume, in gallons per 1000 standard cubic feet (GPM) is given by:

Equation C-6 Equivalent Liquid Volume (GPM)

GPMn CONCNn LCFn × BASEPRS

14.73 × BASETEMP 459.67 + 60 459.67 + × =

GPMn gallons/1000 standard cubic feet of component “n”

CONCNn normalized (if selected) concentration of component “n”

LCFn liquid equivalent conversion factor for component “n” at 14.73 PSIA and 60 degrees F

BASE PRS base (contact) Pressure specified; defaults to 14.73

C.4.2 Heating Value Calculations

(a) Dry Gross BTU of Total Gas

Equation C-7 Dry Gross BTU Value of Total Gas

P ∑

CONCN ( )n BTU CF ( )n [ ]

DRYBTU CF

DRYBTU/ CF

uncorrected dry BTU content per cubic foot of total gas sample

CONCNn normalized (if selected) concentration of component “n”, calculated from peak analysis

BTU/CFn energy content per cubic foot of component “n”, stored in permanent memory

JULY 2010 Post Analysis Computations

Show 4 more labels / short text on this page
where · 100 -------------------------------------------------------------------------- = · n 1 = · where
Page 454 of the Emerson Process Management Chromatograph Software (MON2000) user manual — D R
Page 454 · click to enlarge
Page 454

C-10 DATA COMPUTATIONS

P total number of components to be used in calcula- tion of total BTU/CF

“100” removed the “100” factored into the calculation of the concentration earlier in the analysis

(b) Ideal Relative Density

Equation C-8 Ideal Relative Density of Total Gas Sample

P ∑

CONCn RDn ( )

TOTALRD

RDn relative Density of component “n”

TOTAL RD relative Density of total gas sample

CONCNn, P, 100 are defined in “1.” preceding

(c) Real (corrected) Relative Density

The ideal gas relative density, DI is corrected to the real gas relative Density, Dr, by dividing by the com- pressibility factor, Z, for gas mixture at 60 oF and one atmosphere pressure and multiplying by the compressibility factor of air at the same conditions:

Equation C-9 Real Gas Relative Density

DIZb air ( ) Zb gas ( ) =

D R

Heating Value Calculations JULY 2010

100 = · n 1 = · where

Page 455 of the Emerson Process Management Chromatograph Software (MON2000) user manual — CORRDRYBTU DRYBTU
Page 455 · click to enlarge
Page 455

DATA COMPUTATIONS C-11 MON2000

DI ideal gas relative density ()

Zb(air) compressibility factor of air, or 0.99959

Zb(gas) compressibility factor of gas mixture

(d) Compressibility Factor Dry BTU

Compressibility uses calculations from AGA Report No. 8 (1992) errata (1993) “Compressibility Factors of Natural Gas and other related Hydro-Carbon Gases”.

Equation C-10 Corrected Dry BTU Value

CORRDRYBTU DRYBTU

Z ----------------------- =

DRYBTU value from Equation C-7

Z compressibility factor (see Equation C-9)

BASE PRS base (contract) pressure specified; defaults to 14.73 PSIA

(e) Corrected Saturated BTU

Equation C-11 Corrected Saturated BTU Value of Total Gas

CORRSATBTU DRYBTU ( ) 0.9826 ( ) Z =

JULY 2010 Heating Value Calculations

where · where

Page 456 of the Emerson Process Management Chromatograph Software (MON2000) user manual — C-12 DATA COMPUTATIONS
Page 456 · click to enlarge
Page 456

C-12 DATA COMPUTATIONS

DRYBTU value from Equation C-7

CORRSATBTU corrected saturated BTU content per cubic foot of total gas sample at base conditions of BASE PRS and 60 °F

Z compressibility of total gas as calculated in Equation C-10

BASEPRS base (contract) pressure specified; defaults to 14.73 PSIA

(f) Compressibility and Base Pressure

Compressibility and base pressure corrections for Dry BTU are:

Equation C-12 Dry BTU at Base Pressure

Z ⎝ ⎠ ⎛ ⎞ = ContractPressure

BasePressure ⎝ ⎠ ⎛ ⎞

CORRDRYBTU value from Equation C-7

Z compressibility factor (see Equation C-10)

BASE PRESSURE base (contract) pressure specified; defaults to 14.73 PSIA

Equation C-13 BTU Calculations

GrossActualBTU corr ( ) GrossDryBTU corr ( ) 100 WVC ( ) 100 – ⎝ ⎠ ⎛ ⎞ × =

Heating Value Calculations JULY 2010

Show 4 more labels / short text on this page
where · CorrDryBTU DryBTUatBasePressure · where · (g) BTU Calculations
Page 457 of the Emerson Process Management Chromatograph Software (MON2000) user manual — k ∑
Page 457 · click to enlarge
Page 457

DATA COMPUTATIONS C-13 MON2000

WVC Water volume content (provided by a “live analog input”

Equation C-14 Wobbe Index Calculation

WI CORR GROSS ( )BTU RD =

W.I. Wobbe index value

CORRGROSSBTU Corrected Dry BTU for total gas Sample as calculated in Equation C-10

RD real Relative Density as calculated in Equation C-9

Equation C-15 Weight Percent Calculation

CONCn ( ) MWn ( )

All components in the sample must be measured in order to calculate weight percent.

100 × =

k ∑ MW i ( )

CONCi ( )

WTpercentn weight percent of component “n”

CONCn concentration in mole percent of component “n”

Mwn molecular weight of component “n”

k ∑

sum of weights of all components in sample

JULY 2010 Heating Value Calculations

Show 6 more labels / short text on this page
where · where · WTpercentn · i 1 = · where · i 1 =
Page 458 of the Emerson Process Management Chromatograph Software (MON2000) user manual — k ∑
Page 458 · click to enlarge
Page 458

C-14 DATA COMPUTATIONS

Equation C-16 Average Molecular Weight

k ∑ =

AVGMW CONCi ( ) MWi ( )

AVGMW average molecular weight

k ∑

sum of weights of all components in sample

Equation C-17 Liquid Volume Percent

WTpercentn ( ) Dn ( ) ÷

100 × =

k ∑

WTpercenti ( ) Di ( ) ÷

All components in the sample must be measured in order to calculate liquid volume percent from mole percent.

LV percent liquid volume

WT percent weight percent

k ∑

sum of all components in sample

Heating Value Calculations JULY 2010

Show 8 more labels / short text on this page
i 1 = · where · i 1 = · LVpercent · i 1 = · where · D density · i 1 =
Page 459 of the Emerson Process Management Chromatograph Software (MON2000) user manual — LRD T
Page 459 · click to enlarge
Page 459

DATA COMPUTATIONS C-15 MON2000

Equation C-18 Reid Vapor Pressure

k ∑

CONCi ( ) VPi ( )

RVP

RVP reid vapor pressure

CONCi normalized concentration of component “i” in mole percent

VPi vapor pressure at 100 F of component “i” (GPA2145 = 94)

Equation C-19 Liquid Relative Density

All components in the sample must be measured to calculate LRDT.

k ∑

LVpercenti ( ) LRDi ( )

LRD T

LRDT liquid relative density of sample, relative to water at 60 °F

LRDi liquid relative density of component “i” (GPA2145-94)

LVpercent liquid volume percent from Equation C-18

JULY 2010 Heating Value Calculations

Show 6 more labels / short text on this page
i 1 = · 100 ----------------------------------------------- = · where · 100 ------------------------------------------------------------- = · i 1 = · where
Page 460 of the Emerson Process Management Chromatograph Software (MON2000) user manual — GD RD ( ) 76.4976 ( ) =
Page 460 · click to enlarge
Page 460

C-16 DATA COMPUTATIONS

Equation C-20 Liquid Density

All components on sample must be measured to calculate Liquid Density.

k ∑

LVi ( ) LDi ( )

LD T

LDT liquid density of total sample in pounds per gallon

LDi liquid density of component “i”. (GPA 2145-94)

LVi liquid volume percent

Equation C-21 Gas Density

GD RD ( ) 76.4976 ( ) =

GD gas density in lb/1000 ft3

RD relative density (relative to air)

76.4976 density of air at 14.73 PSIA and 60 °F, in lb/1000 ft3

Heating Value Calculations JULY 2010

Show 4 more labels / short text on this page
100 ------------------------------------- = · i 1 = · where · where
Page 461 of the Emerson Process Management Chromatograph Software (MON2000) user manual — DATA COMPUTATIONS C-17 MON2000
Page 461 · click to enlarge
Page 461

DATA COMPUTATIONS C-17 MON2000

C.4.3 Multi-Level Calibration

The properties of each gas component can be viewed using the Component Data menu. Included with the component properties in the Component Data Table are four coefficients labeled Multi-Level Calib 'a', 'b', 'c', and 'd' for each component. If these parameters are all zero, then linear calibration is used. See Section C.3.1 for the response factor calculations.

If any of these parameters have a value other than zero, then multi-level, or polynomial calibration is used for that component.

The response factors are then calculated as:

Equation C-22 Multi-Level Calibration

ResponseFactor aP3 bP2 cP d + + + CalibrationConcentration = (mol %)

P peak size (Area or Height as selected by the Operator); from average calibration runs

CoefficientsA,B,C, and D

Calculated Offline. Entered after Multi-Level Calibration using several calibration gases (typically seven gases)

1.Note: If the coefficient values are correct values, the response factor will be close to one.

JULY 2010 Multi-Level Calibration

where

Page 462 of the Emerson Process Management Chromatograph Software (MON2000) user manual — C.4.4 Indirect Calibration
Page 462 · click to enlarge
Page 462

C-18 DATA COMPUTATIONS

The mole% value in the sample gas is then calculated as

Equation C-23 Mole% Value

Mole % = aP3 bP2 cP d + + + responsefactor

P peak size measured in Sample Gas

C.4.4 Indirect Calibration

Component gases which are NOT found in the calibration gas, but may be found in the sample gas, can be assigned a response factor which is a fixed multiple (the Relative Response Factor to a Reference Component) that IS found in the calibration gas.

The Relative Response Factors and Reference Component Values are included in the Component Data Table (following the Multi-Level Calibration coefficients) see Appendix B.

If the Reference Component is None, then normal (direct) calibration is used.

If the Reference Component is defined, (e.g. Propane) then the mole% value for the indirect component (e.g. neoC5), is calculated as:

Equation C-24 Mole% Value Indirect Component

mole% (neoC5) = mole% (Propane) P neoC5 ( ) P Propane ( ) ⎝ ⎠ ⎛ ⎞RRFneoC5 ( )

Indirect Calibration JULY 2010

where

Page 463 of the Emerson Process Management Chromatograph Software (MON2000) user manual — DATA COMPUTATIONS C-19 MON2000
Page 463 · click to enlarge
Page 463

DATA COMPUTATIONS C-19 MON2000

RRF Relative Response Factor

JULY 2010 Indirect Calibration

where · P peak size

Page 464 of the Emerson Process Management Chromatograph Software (MON2000) user manual — C-20 DATA COMPUTATIONS
Page 464 · click to enlarge
Page 464

C-20 DATA COMPUTATIONS

This page is intentionally left blank.

Indirect Calibration JULY 2010

Page 465 of the Emerson Process Management Chromatograph Software (MON2000) user manual — D.1 CALIBRATING BY VOLTS
Page 465 · click to enlarge
Page 465

ANALOG OUTPUT CAL. FOR 2350A D-1 MON2000

DA ALOG OUTPUT CAL. FOR 2350A APPENDIX D, ANALOG OUTPUT CAL. FOR 2350A

The initial analog output adjustment will be set at the factory before shipment at standard values (4-20 mA or 1-5 V). It may be necessary to check and/or adjust these values, depending on output cabling or impedance. The adjustment may require two persons if the units are some distance apart. It will require a good digital meter to check the zero and full scale values at the receiving end. The scale or span value can be adjusted by a PC with MON2000 software when the values are known at the receiving end.

It is possible to calibrate the analog outputs using different engineering units. Two examples are described below using both volts and percentages.

The 4-20 mA outputs are calibrated similarly except the current is measured instead of the voltage.

D.1 CALIBRATING BY VOLTS

This example assumes that an analog output channel needs to calibrated to a span of 1-5 V.

To calibrate an analog output by adjusting the voltage,

1. Use the Application > Analog Outputs menu to access this function.
2. The Analog Outputs dialog appears.

JULY 2010 Calibrating by Volts

Page 466 of the Emerson Process Management Chromatograph Software (MON2000) user manual — D-2 ANALOG OUTPUT CAL. FOR 2350A
Page 466 · click to enlarge
Page 466

D-2 ANALOG OUTPUT CAL. FOR 2350A

3. Select the channel to be calibrated.

Output channels 1 and 2 can be scaled from either 4-20 mA or 1-5 V. Channels 3 to 10 are scaled from 4-20 mA.

Although output channels for direct voltage measurement exist for analog output channels 1 and 2, it is recommended that they not be used for calibration purposes. Instead, use the current-out terminals and the 250-ohm resistor as described.

Calibrating by Volts JULY 2010

Page 467 of the Emerson Process Management Chromatograph Software (MON2000) user manual — ANALOG OUTPUT CAL. FOR 2350A D-3 MON2000
Page 467 · click to enlarge
Page 467

ANALOG OUTPUT CAL. FOR 2350A D-3 MON2000

4. To change a variable assignment, click the appropriate Variable cell. Use the provided pull-down menu and click the desired variable to select it.
5. For voltage values, set Zero Scale to “1.0” and Full Scale to “5.0”.

If the channel is calibrated in milliamperes, set Zero Scale to “4” and Full Scale to “20”.

6. Set the Fixed/Var parameter to “Fixed”.
7. Set the Fixed Value to “0.0”.
8. Set Zero Scale Adjustment and the Full Scale Adjustment to “0.0”.
9. Click the button to accept your changes and exit from the Analog Outputs dialog.

10.Connect a 250-ohm resistor across the signal at the input of the device where this analog output is used. Connect a Voltmeter across the resistor and measure the result.

Your Voltmeter reading should be close to 1.0 V but does not have to be exactly 1.0 V. If the reading is negative, reverse the leads.

11.Return to the Analog Outputs dialog and reselect the channel being calibrated.

12.Set the Fixed Value to the Full Scale value entered in Step 5.

JULY 2010 Calibrating by Volts

Page 468 of the Emerson Process Management Chromatograph Software (MON2000) user manual — D-4 ANALOG OUTPUT CAL. FOR 2350A
Page 468 · click to enlarge
Page 468

D-4 ANALOG OUTPUT CAL. FOR 2350A

13.Click the button to accept your changes and exit from the Analog Outputs dialog.

14.Look at the voltmeter again and record the full scale reading.

15.Return to the Analog Outputs dialog and reselect the channel being calibrated.

Your voltmeter reading should be very close to 5.0 Volts. MON2000 will adjust the scale so the span will be 4 Volts and scale accordingly.

16.Change the Zero Scale Adjustment to the Zero Scale voltage reading and the Full Scale Adjustment to the Full Scale voltage reading.

17.Click the button to accept your changes and exit from the Analog Outputs dialog.

18.Return to the Analog Outputs dialog and reselect the channel being calibrated.

19.To quickly verify your changes, enter the midpoint of your range (e.g., 3) for the Fixed

Value. Click the button to accept your change. The resulting voltage output should be close to span midpoint (e.g., 3).

20. Return to the Analog Outputs dialog and reselect the channel being calibrated.

21.Set the Fixed/Var parameter to “Var” and the Fixed Value to “0.0”.

22.Click the button to accept your changes.

The voltage output could change slightly from time to time and the MON2000 software will now maintain the correct span.

Calibrating by Volts JULY 2010

Page 469 of the Emerson Process Management Chromatograph Software (MON2000) user manual — ANALOG OUTPUT CAL. FOR 2350A D-5 MON2000
Page 469 · click to enlarge
Page 469

ANALOG OUTPUT CAL. FOR 2350A D-5 MON2000

D.2 CALIBRATING BY PERCENTAGES

This second example, shown below, demonstrates that it is also possible to calibrate your outputs as a percentage. For instance, you may wish to read directly values that may be scaled from 0 to 1200. Your voltage span is scaled from 1 to 5 volts so the range is approximately 4 Volts. Thus, 1200 divided by 4 equals 300 units per volt. If you equate the difference as a percentage, the midpoint in units equals 600, or approximately 3 volts.

With this information in mind, you want to set zero units equal to 1.00 V and 1200 units to 5.00 V and let the software adjust the span as you calibrate.

This example assumes that an analog output channel needs to calibrated to a span of 0 to 1200.

JULY 2010 Calibrating by Percentages

Page 470 of the Emerson Process Management Chromatograph Software (MON2000) user manual — D-6 ANALOG OUTPUT CAL. FOR 2350A
Page 470 · click to enlarge
Page 470

D-6 ANALOG OUTPUT CAL. FOR 2350A

To calibrate by percentages,

1. Use the Application > Analog Outputs menu to access this function.
2. The Analog Outputs dialog appears.
3. Select the channel to be calibrated.

Output channels 1 and 2 can be scaled from either 4-20 mA or 1-5 V. Channels 3 to 10 are scaled from 4-20 mA.

Although output channels for direct voltage measurement exist for analog output channels 1 and 2, it is recommended that they not be used for calibration purposes. Instead, use the current-out terminals and the 250-ohm resistor as described.

All can be scaled with a percentage.

Calibrating by Percentages JULY 2010

Page 471 of the Emerson Process Management Chromatograph Software (MON2000) user manual — ANALOG OUTPUT CAL. FOR 2350A D-7 MON2000
Page 471 · click to enlarge
Page 471

ANALOG OUTPUT CAL. FOR 2350A D-7 MON2000

4. To change a variable assignment, click the appropriate Variable cell. Use the provided pull-down menu and click the desired variable to select it.
5. Set Zero Scale to “0” and Full Scale to “1200”.
6. Set the Fixed/Var parameter to “Fixed”.
7. Set Fixed Value to “0.0”.
8. Set Zero Scale Adjustment and Full Scale Adjustment to “0.0”.
9. Click the button to accept your changes and exit from the Analog Outputs dialog.

10.Connect a 250-ohm resistor across the signal at the input of the device where this analog output is used. Connect a Voltmeter across the resistor and measure the result.

Your Voltmeter reading should be close to 1.0 V but does not have to be exactly 1.0 V. If the reading is negative, reverse the leads.

11.Return to the Analog Outputs dialog and reselect the channel being calibrated.

12.Set the Fixed Value to the Full Scale value entered in Step 5.

JULY 2010 Calibrating by Percentages

Page 472 of the Emerson Process Management Chromatograph Software (MON2000) user manual — D-8 ANALOG OUTPUT CAL. FOR 2350A
Page 472 · click to enlarge
Page 472

D-8 ANALOG OUTPUT CAL. FOR 2350A

13.Click the button to accept your changes and exit from the Analog Outputs dialog.

14.Look at the voltmeter again and record the full scale reading.

15.Return to the Analog Outputs dialog and reselect the channel being calibrated.

Your voltmeter reading should be very close to 5.0 Volts. MON2000 will adjust the scale so the span will be 4 Volts and scale accordingly.

16.Change the Zero Scale Adjustment to the zero scale voltage reading and the Full Scale Adjustment to the full scale voltage reading.

17.Click the button to accept your changes and exit from the Analog Outputs dialog.

18.Return to the Analog Outputs dialog and reselect the channel being calibrated.

19.To quickly verify your changes, enter the midpoint of your range (e.g., 600) for the

Fixed Value. Click the button to accept your change. The resulting voltage output should be close to span midpoint (e.g., 3).

20. Return to the Analog Outputs dialog and reselect the channel being calibrated.

21.Set the Fixed/Var parameter to “Var” and the Fixed Value to “0.0”.

22.Click the button to accept your changes.

The voltage output could change slightly from time to time and the MON2000 software will now maintain the correct span.

Calibrating by Percentages JULY 2010

Page 473 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.2 OFFLINE EDIT TO UPLOAD APP. & RENAME
Page 473 · click to enlarge
Page 473

UPGRADE 2350A GC S/W AND 2350 EPROMS E-1 MON2000

EA APPENDIX E, UPGRADE 2350A GC S/W AND

2350 EPROMS

To upgrade the Model 500 2350A (GC) software and the 2350 EPROMs, perform all steps in the order they appear in this appendix.

Before beginning the upgrade procedure, ensure that you have identified which standard (i.e., factory-released) application file was used to form the current user application for this GC. See Section E.10 for more information on standard applications.

E.1 CONNECT TO GC AND HALT ANALYSIS

1. Ensure that the MON2000 software is running and you are connected to the GC that is to be upgraded.

Use the File > Connect menu to access this function. See Section 2.10.1 for more information.

2. Use MON2000 to halt any ongoing analysis or calibration runs.

When halted GC runs have finished, the status for Detector 1 mode will indicate Idle (see GC Status bar; see Section 3.3).

Use the Control > Halt menu to access this function. See Section 4.3 for more information.

E.2 OFFLINE EDIT TO UPLOAD APP. & RENAME

1. Use MON2000 to upload (from the GC unit) to the PC the most current parameters that are in use at the GC.

Use the File > Offline Edit > Upload Application menu to access this function.

2. A menu of stored application files appears. Double-click the desired file.

JULY 2010 Connect to GC and Halt Analysis

Page 474 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E-2 UPGRADE 2350A GC S/W AND 2350 EPROMS
Page 474 · click to enlarge
Page 474

E-2 UPGRADE 2350A GC S/W AND 2350 EPROMS

3. The Save Uploaded GC Application File dialog appears.
Emerson Process Management Chromatograph Software (MON2000) — E-2 UPGRADE 2350A GC S/W AND 2350 EPROMS diagram (page 474)
Fig. 327

Use this dialog to assign a new file name to the GC Application file that is the target for uploading.

The existing file name of the GC application file for the connected GC displays in the File name field.

For clarity, you could name this file “OLD.bin” to distinguish it from the upgraded application file.

Press the ENTER key or click the button to continue.

4. The Upload Application process begins. This process usually takes about 15 minutes to complete.

MON2000 automatically disconnects from the GC after the Upload Application process has finished.

5. Use the File > Exit menu to close MON2000.

Or, you can exit MON2000 by pressing the

ALT-F X keys, or click the button on the main window.

Offline Edit to Upload App. & Rename JULY 2010

Page 475 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.4 UPGRADE FROM DISK
Page 475 · click to enlarge
Page 475

UPGRADE 2350A GC S/W AND 2350 EPROMS E-3 MON2000

E.3 UPGRADE USER-DEFINED APPLICATIONS

Follow the installation instructions provided in Section 2.2 to upgrade the MON2000 software and/or a standard application file.

Do not delete any existing MON2000 program or GC application files from the PC hard drive. The upgrade process uses information from these files and automatically overwrites any outdated program file.

Use the Update BIN utility that is installed with the MON2000 software to upgrade a GC application that contains user-defined data (such as components). The Update BIN program allows you to upgrade an application that is either on disk or “live” at the GC Controller. See the following sections for details.

E.4 UPGRADE FROM DISK

To upgrade a user-defined application file (*.BIN) located on your PC hard drive or on a floppy disk,

It is recommended that you save a copy of your application file before performing the upgrade.

JULY 2010 Upgrade User-Defined Applications

Page 476 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E-4 UPGRADE 2350A GC S/W AND 2350 EPROMS
Page 476 · click to enlarge
Page 476

E-4 UPGRADE 2350A GC S/W AND 2350 EPROMS

1. Launch the Update BIN software by either:
  • clicking the button and using the Programs > MON2000 > Update BIN menu
  • double-clicking the Update BIN icon on your PC desktop
2. The Update BIN software program launches inside a DOS window.
Emerson Process Management Chromatograph Software (MON2000) — E-4 UPGRADE 2350A GC S/W AND 2350 EPROMS diagram (page 476)
Fig. 328
3. Press the TAB key to access the Communication menu.
Emerson Process Management Chromatograph Software (MON2000) — E-4 UPGRADE 2350A GC S/W AND 2350 EPROMS diagram (page 476)
Fig. 329
4. Use the ARROW keys to select the File Selection option, and then press ENTER .
5. The File Selection screen displays.
Emerson Process Management Chromatograph Software (MON2000) — E-4 UPGRADE 2350A GC S/W AND 2350 EPROMS diagram (page 476)
Fig. 330

Upgrade from Disk JULY 2010

Page 477 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.3 Upgrade User-Defined Applications
Page 477 · click to enlarge
Page 477

UPGRADE 2350A GC S/W AND 2350 EPROMS E-5 MON2000

Use the TAB and ARROW keys to select the desired application file from disk. This file is the file you will be upgrading.

6. Press the TAB key to access the Upgrade Config menu.
7. Use the ARROW keys to select the Merge option, and then press ENTER .
Emerson Process Management Chromatograph Software (MON2000) — E.3 Upgrade User-Defined Applications diagram (page 477)
Fig. 331 · E.3 Upgrade User-Defined Applications
8. The File Selection screen displays.
Emerson Process Management Chromatograph Software (MON2000) — E.3 Upgrade User-Defined Applications diagram (page 477)
Fig. 332 · E.3 Upgrade User-Defined Applications

Use the TAB and ARROW keys to select the desired application file from disk. This file is the new application file you wish to merge into the existing file (see Step 5).

9. The Merge screen displays.
Emerson Process Management Chromatograph Software (MON2000) — E.3 Upgrade User-Defined Applications diagram (page 477)
Fig. 333 · E.3 Upgrade User-Defined Applications

JULY 2010 Upgrade from Disk

Page 478 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.4.1 Upgrade from GC Controller
Page 478 · click to enlarge
Page 478

E-6 UPGRADE 2350A GC S/W AND 2350 EPROMS

10.Use the ARROW+ENTER keys to select the desired application settings to upgrade.

Press the SHIFT + ENTER keys to select multiple settings.

Press the F2 key to select all the available settings.

Press the F3 key to clear your selected settings.

11.Press the F4 key to begin the upgrade.

Emerson Process Management Chromatograph Software (MON2000) — E-6 UPGRADE 2350A GC S/W AND 2350 EPROMS diagram (page 478)
Fig. 334

12.When the upgrade is done, press the ESC key or use the Communication > Exit menu to close the Update BIN program.

Ensure that you inspect the upgraded application file for any errors.

E.4.1 Upgrade from GC Controller

To upgrade a user-defined application file (*.BIN) currently being used by the GC Controller,

It is recommended that you upload a copy of your application file before performing the upgrade.

1. Launch the Update BIN software by either:
  • clicking the button and using the Programs > MON2000 > Update BIN menu
  • double-click the Update BIN icon on your PC desktop

Upgrade from GC Controller JULY 2010

Page 479 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.4.1 Upgrade from GC Controller
Page 479 · click to enlarge
Page 479

UPGRADE 2350A GC S/W AND 2350 EPROMS E-7 MON2000

2. The Update BIN software program launches inside a DOS window.
3. Use the ARROW keys to select the Connect option, and then press ENTER .
Emerson Process Management Chromatograph Software (MON2000) — E.4.1 Upgrade from GC Controller diagram (page 479)
Fig. 335 · E.4.1 Upgrade from GC Controller
4. The Connect screen displays, listing the available GC units.
Emerson Process Management Chromatograph Software (MON2000) — E.4.1 Upgrade from GC Controller diagram (page 479)
Fig. 336 · E.4.1 Upgrade from GC Controller

Use the ARROW keys to select the appropriate GC, and then press ENTER.

5. Update BIN attempts to connect to the GC Controller.

If the attempt fails, see Section E.4.2 for information on configuring the communications parameters for Update BIN.

JULY 2010 Upgrade from GC Controller

Page 480 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E-8 UPGRADE 2350A GC S/W AND 2350 EPROMS
Page 480 · click to enlarge
Page 480

E-8 UPGRADE 2350A GC S/W AND 2350 EPROMS

6. The File Selection screen displays.

Use the TAB and ARROW keys to select the desired application file from disk. This file is the new application file you wish to merge into the existing file (see Step 5).

7. The Merge screen displays.
8. Use the ARROW+ENTER keys to select the desired application settings to upgrade.

Press the SHIFT + ENTER keys to select multiple settings.

Press the F2 key to select all the available settings.

Press the F3 key to clear your selected settings.

9. Press the F4 key to begin the upgrade.

Upgrade from GC Controller JULY 2010

Page 481 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.4.2 Configure GC Connection
Page 481 · click to enlarge
Page 481

UPGRADE 2350A GC S/W AND 2350 EPROMS E-9 MON2000

10.When the upgrade is done, use the Communications > Disconnect menu to cease communications with the GC unit.

Emerson Process Management Chromatograph Software (MON2000) — E.4.2 Configure GC Connection diagram (page 481)
Fig. 337 · E.4.2 Configure GC Connection

11.Press the ESC key or use the Communication > Exit menu to close the Update BIN program.

Ensure that you inspect the upgraded application file for any errors.

E.4.2 Configure GC Connection

To configure the GC communications parameters for the Update BIN software,

1. Use the Communication > Directory menu to access a list of communication settings for each GC unit.
Emerson Process Management Chromatograph Software (MON2000) — E.4.2 Configure GC Connection diagram (page 481)
Fig. 338 · E.4.2 Configure GC Connection
2. The Directory screen displays.
Emerson Process Management Chromatograph Software (MON2000) — E.4.2 Configure GC Connection diagram (page 481)
Fig. 339 · E.4.2 Configure GC Connection
3. Use the ARROW keys to select the desired GC unit and then press the F2 key to edit its communications parameters.

You can also press F3 to edit the modem initialization string.

JULY 2010 Configure GC Connection

Page 482 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.5 DISCONNECT POWER AND DISASSEMBLE
Page 482 · click to enlarge
Page 482

E-10 UPGRADE 2350A GC S/W AND 2350 EPROMS

4. The Parameters screen displays.
Emerson Process Management Chromatograph Software (MON2000) — E-10 UPGRADE 2350A GC S/W AND 2350 EPROMS diagram (page 482)
Fig. 340
5. Use the ARROW keys and space bar to select the desired parameter and toggle between its options.
6. Press the ENTER key to accept your changes or press the ESC key to cancel.

E.5 DISCONNECT POWER AND DISASSEMBLE

PERSONELL AND EQUIPMENT

Failure to follow the instructions below, may cause injury to personnel or damage equipment.

Before you remove the cover from the GC Controller or access its internal electronics, switch OFF the AC power source to the GC Controller, either at the breaker box or the controlling power source switch.

Also, disconnect the AC power cord from J21 before continuing working on the unit.

Disconnect Power and Disassemble JULY 2010

Page 483 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.5 Disconnect Power and Disassemble
Page 483 · click to enlarge
Page 483

UPGRADE 2350A GC S/W AND 2350 EPROMS E-11 MON2000

1. Disconnect AC power from the GC Controller.

When you have access to the GC Controller Terminal Board (TB) for field wiring (accessed from the rear on rack-mount versions and behind the front cover of explosion-proof models) disconnect the AC power cord from J21 before continuing working on the unit.

2. For newer model 19-inch rack mount and retrofit GC Controllers, proceed as follows:

(a) Remove the (right) side access panel that

is secured to the chassis with four thumbscrews. (b) Unplug all cables connected to the

boards in slots 2 through 6 of the Card Cage Assembly in order to access the SBC53 CPU Board. (c) Proceed to Step 3 below.

For the explosion-proof GC Controller or older rack mount GC Controllers, proceed as follows:

(a) Access the GC Controller Terminal

Board for field wiring.

On the explosion-proof GC Controller, remove outer housing bolts and lower the cover on its bottom hinge. The TB faces the front.

On rack mount GC Controllers, the TB is exposed and faces the rear.

(b) Loosen the six thumbscrews that secure

the TB. Then gently unplug the TB from its mating DIN connectors at the back, top of the board. (c) Lower the TB down (held in place by its

ground straps at the bottom of the board) in order to expose the Card Cage Assembly.

JULY 2010 Disconnect Power and Disassemble

Page 484 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.6 REPLACE EPROMS/RESET CPU
Page 484 · click to enlarge
Page 484

E-12 UPGRADE 2350A GC S/W AND 2350 EPROMS

(d) Unplug the TB power supply cord from

its connection at the Card Cage Assembly power supply. (e) Loosen the four thumbscrews that

secure the Card Cage Assembly to the chassis. Then remove the Card Cage Assembly away from its chassis mount so that it is easy to work on. (f) Unplug all cables connected to the

boards in slots 2 through 6 of the Card Cage Assembly in order to access the SBC53 CPU Board.

3. Remove the SBC53 CPU Board from slot number 2 (second from the top) of the Card Cage Assembly.

E.6 REPLACE EPROMS/RESET CPU

1. For 2350 GC Controllers, locate the EPROMs that will be replaced: U18 and U19. They are near the card-connector edge of the SBC53 CPU Board.

Carefully note which EPROMs are labeled ODD and EVEN and which sockets hold them:

socket U18 EVEN EPROM P/N 8-2350-001

socket U19 ODD EPROM P/N 8-2350-002

2. Remove, one at a time, each of the old EPROMs, and replace each with its upgrade EPROM.
3. Locate Jumper Pin Set J14. It is near the two EPROMs you have just replaced.

Replace EPROMS/Reset CPU JULY 2010

Page 485 of the Emerson Process Management Chromatograph Software (MON2000) user manual — 3-2350-063 J12 SBC53
Page 485 · click to enlarge
Page 485

UPGRADE 2350A GC S/W AND 2350 EPROMS E-13 MON2000

4. To reset the CPU registers, remove the jumper from J14 pins 1 and 2, and place it to short J14 pins 2 and 3.
5. After briefly shorting J14 pins 2 and 3, remove the jumper again. Place it back in its original position, shorting J14 pins 1 and 2.

E.7 REASSEMBLY PROCEDURES

1. Place and securely plug the SBC53 CPU Board into slot number 2 of the Card Cage Assembly.
2. Reconnect all ribbon cables that were unplugged ( see Section E.5 ).

Because the ribbon cables are arranged in layers, it important to reconnect them in the order listed below:

Table E-1 2350 GC Ribbon Cables to Connect

System Interface Board connection

Cable Originating from Board

3-2350-063 J12 SBC53

3-2350-052 J9 SBC53

3-2350-062 J11 SBC53

3-2350-051 J4 SBC53

3-2350-064 J8 DSPI/O

3-2350-053 J10 DSPI/O

3-2350-054 J6 RTI1281

3-2350-055 J5 RTI1281

JULY 2010 Reassembly Procedures

Part Number

Page 486 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.8.1 Procedure
Page 486 · click to enlarge
Page 486

E-14 UPGRADE 2350A GC S/W AND 2350 EPROMS

3. Return and secure the Card Cage Assembly its original position in the chassis mounting. Tighten the four screws.
4. Return and secure the field wiring TB in its original position. Tighten the six screws.
5. Reconnect the AC power cord to J21 on the TB, and restore AC power to the 2350A GC Controller.

E.8 SET-UP AND PROGRAMMING

The purpose of this procedure is to define the CMOS table set-up and programming requirements for the CPU assembly and DiskOnChip.

E.8.1 Procedure

1. Install the blank DiskOnChip into socket U1 on the CPU assembly, noting the orientation of pin 1.
2. Plug the floppy drive cable into J10 on the CPU assembly and plug the board into the card cage.
3. Plug the keyboard cable into J1 on the CPU assembly.
4. Turn on the power. When the video displays on the monitor, press the DELETE button to access the CMOS setup window.

(a) The Standard CMOS Setup window

displays. Press the ENTER button. (b) Set time and date by either PAGE UP,

PAGE DOWN, or by entering the appropriate number value. After making the desired change, press ENTER.

Set-Up and Programming JULY 2010

Page 487 of the Emerson Process Management Chromatograph Software (MON2000) user manual — A:\IMAGE>CD\
Page 487 · click to enlarge
Page 487

UPGRADE 2350A GC S/W AND 2350 EPROMS E-15 MON2000

(c) Change Drive C:\ to NONE by pressing

PAGE DOWN. Verify that Drive A:\ is the only selected drive. Press ESCAPE, F10 (to save changes), Y (to verify that you want the changes to be made), then ENTER.

5. The CPU assembly will reboot and displays the Systems Configuration window. The speaker on the CPU will chirp and the monitor will prompt:

“ENTER NEW DATE (mm-dd-yy)". Press ENTER.

"ENTER NEW TIME:". Press ENTER.

The A:\> prompt is displayed. Verify that the programming disk is the current revision specified on the work order package and it insert into the floppy drive. Type INSTALL.

6. As the DiskOnChip is being programmed, the following information is displayed:

A:\> SYS C: SYSTEM TRANSFERRED

A:\>CD IMAGE A:\IMAGE>COPY *.*C:\ BOS.EXE PME16.EXE SCANDISK.EXE SCANDISK.INI AUTOEXEC.BAT BOSCHECK.EXE 6 FILE(S) COPIED

A:\IMAGE>CD\

A:\> A:\>

7. At the A:\> prompt, type C :, then ENTER .
8. Type PME16 , then ENTER .

JULY 2010 Procedure

Page 488 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.8.2 Reprogramming the DiskOnChip
Page 488 · click to enlarge
Page 488

E-16 UPGRADE 2350A GC S/W AND 2350 EPROMS

9. The Protected Mode Program screen displays. Verify that a "cold start - initial" message is displayed in the middle of the screen and LED D2 is blinking on the CPU assembly.

10.Press ESC to exit the program and power down.

11.Carefully remove the cable from J1 and J10 to prevent cracking the connector housing. Install the program label on the end of DiskOnChip opposite the serial number.

E.8.2 Reprogramming the DiskOnChip

Reprogram the DiskOnChip using the following instructions.

1. Install either the programmed DiskOnChip into socket U1 on the CPU assembly (noting the orientation of pin 1) or install the complete CPU assembly into card cage.
2. Plug the floppy drive cable into J10 on the CPU assembly and the keyboard cable into J1 on the CPU assembly.
3. Turn on the power. The CPU assembly reboots and displays the Systems Configuration window. The speaker on the CPU will chirp and the monitor will prompt

"ENTER NEW DATE (mm-dd-yy)". Press ENTER

"ENTER NEW TIME:". Press ENTER.

The A:\> prompt displays. Verify that the programming disk is the current revision specified on the work order package and insert into the floppy drive. Type C:, then ENTER to change to the ‘C’ drive.

Reprogramming the DiskOnChip JULY 2010

Page 489 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.8.3 GC Reassembly
Page 489 · click to enlarge
Page 489

UPGRADE 2350A GC S/W AND 2350 EPROMS E-17 MON2000

4. After the C:\> prompt is displayed, type ERASE *.* , then press ENTER . When prompted “ARE YOU SURE?” Type Y , then press ENTER .
5. At the C:\> prompt, type DIR , then press ENTER . Verify that only the Command.com file remains on the ‘C:’ drive, that the file size is 54,645 bytes, and that there are 16,105,472 bytes free.
6. Type A: , then ENTER to change back to the A:\ drive. Type INSTALL , then follow Steps 6 through 11 ( see Section E.8.1 ).

E.8.3 GC Reassembly

Use the following instructions to reassemble the GC Controller (CPU, PC/104 Bus and the Modem).

Emerson Process Management Chromatograph Software (MON2000) — E.8.3 GC Reassembly diagram (page 489)
Fig. 341 · E.8.3 GC Reassembly

Figure E-1 CPU, PC/104 and Modem

1. Place and securely plug the WinSystems CPU Board into slot number 2 of the Card Cage Assembly.
2. Reconnect all ribbon cables that were unplugged.

JULY 2010 GC Reassembly

Page 490 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E-18 UPGRADE 2350A GC S/W AND 2350 EPROMS
Page 490 · click to enlarge
Page 490

E-18 UPGRADE 2350A GC S/W AND 2350 EPROMS

3. Return and secure the Card Cage Assembly back into the chassis mounting. Tighten the four screws.
4. Return and secure the field wiring TB to the original position. Tighten the six screws.
5. Reconnect the AC power cord to J21 on the TB, and restore AC power to the GC Controller.

E.9 CONNECT TO GC FOR UPGRADED APP.

1. Use MON2000 software to connect with the GC unit.

Use the File > Connect menu to access this function. See Section 2.10.1 for more information.

2. Use MON2000 to download the newly upgraded GC application to the connected GC Controller.

(a) Use the File > Download menu to access

the Download CG Application File dialog.

Emerson Process Management Chromatograph Software (MON2000) — E-18 UPGRADE 2350A GC S/W AND 2350 EPROMS diagram (page 490)
Fig. 342

(b) Click the desired .bin file. The file name

appears in the File name data field.

(c) Click the button or press the

ENTER key to download the selected application file to the connected GC unit. A progress bar displays, indicating the completion status.

Connect to GC for Upgraded App. JULY 2010

Page 491 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.10.1 Importance of Standard Application Files
Page 491 · click to enlarge
Page 491

UPGRADE 2350A GC S/W AND 2350 EPROMS E-19 MON2000

E.10 GUIDE TO STANDARD APPLICATION FILES

This section of the appendix lists the various standard application, or BIN files that have been released by for the GC Controller.

E.10.1 Importance of Standard Application Files

After a new GC Controller has been loaded with a standard application, the standard application is customized by the entry of timed events, component table entries, and other operator entered information consistent with the performance characteristics of the chromatograph hardware to which the Controller is connected and the needs of the user. Thus the standard application becomes a user application and differs from the factory- released, standard application from which it originated.

At the time when software is upgraded, including application files, it becomes important to know which factory-released standard application formed the basis of the user application, so the user application can be upgraded, too, with newer version of the standard application file.

You can determine which standard application formed the basis of the current user application by inspecting various MON2000 screens while connected to the GC (in most cases, you need only inspect the System dialog).

JULY 2010 Guide to Standard Application Files

Page 492 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E-20 UPGRADE 2350A GC S/W AND 2350 EPROMS
Page 492 · click to enlarge
Page 492

E-20 UPGRADE 2350A GC S/W AND 2350 EPROMS

E.10.2 Standard Applications v1.50 and Later

If you are using a GC Application that is version 1.50 or later, you can quickly determine the original standard application name by viewing the variables listed in the System dialog.

Use the Application > System menu to access this dialog. In the System dialog, (see Table E- 2) find the value for the variable named CFG Base Name, and locate it in the column labeled “CFG Base Name”, then read the “Analyzer Name (BIN file)” to determine the application BIN file name.

Standard Applications v1.50 and Later JULY 2010

Page 493 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E.10.2 Standard Applications v1.50 and Later
Page 493 · click to enlarge
Page 493

UPGRADE 2350A GC S/W AND 2350 EPROMS E-21 MON2000

Table E-2 System Variables for Standard Application Files

Analyzer Name (BIN file)

Max. Archive Averages

2350_001 2350 USASTD1 2350 Standard 1, 64 avgs 64 8

2350_002 2350 USASTD2 2350 Standard 2, 32 avgs 32 8

2350_003 2350 USASTD3 2350 Standard 3, 128 avgs 128 5

2350_006 2350 USASTD6 2350 Std 6, 32 avgs, 10 strs

2350_007 2350 USASTD7 2350-7 12Str, 128 Avg, 260Mbus

2350ISO1 2350 ISOSTD1 2350 ISO Std 1, 64 avgs 64 8

2350ISO2 2350 ISOSTD2 2350 ISO Std 2, 32 avgs 32 8

2350ISO3 2350 ISOSTD3 2350 ISO Std 3, 128 avgs 128 8

2350FI02 2350 USSTDFILE02 12 stream with files

2360_001 2360 USADUAL1 2360 test applica- tion

2350EXMB 2350 USSTDFILE01 Extended Modbus 4 strm plus cal 150 5

2360ISO1 2360 ISODUALSTD1 2360 ISO Std 1, 64 avgs 64 8

  • Standard BTU applications use the BIN files 2350_001, 2350_002, and 2350_003 for 64, 32, and 128 averages respectively.
  • The Extended Modbus application uses BIN file 2350EXMB.
  • The standard dual application uses BIN file 2360_001.

JULY 2010 Standard Applications v1.50 and Later

Show 4 more labels / short text on this page
Unit Type · CFG Base Name · System Description · Stream Sequence
Page 494 of the Emerson Process Management Chromatograph Software (MON2000) user manual — E-22 UPGRADE 2350A GC S/W AND 2350 EPROMS
Page 494 · click to enlarge
Page 494

E-22 UPGRADE 2350A GC S/W AND 2350 EPROMS

E.10.3 Standard Applications Prior to v1.50

For versions of GC Applications prior to version 1.50, the CFG Base Name variable did not exist. Therefore, a combination of other variables must be inspected to determine the standard application that was used to formulate the current user application.

To determine the standard application used to create a user application prior to version 1.50, compare the values of three variables from the System dialog - Unit Type, Max Archive Averages, and System Description — with the values found in Table E-2. You will also need to compare the maximum number of streams, which can be determined from the Streams option under the Application menu.

There is one more possible complication. If the variable System Description was edited, it may not be the same value as described earlier in Table E-2. In such an instance, it would be impossible to distinguish between the 2350_00n (i.e., USASTD1, USASTD2, and USASTD3) and 2350ISOn (i.e., ISOSTD1, ISOSTD2, and ISOSTD3) applications.

To overcome this complication, you can inspect one more variable:

1. Use the Application> User Defined> Selection menu to access the Selection dialog.
2. In the Selection dialog, CVTABLE_pri will be the first variable listed if the application is one of the ISO applications.

Standard Applications Prior to v1.50 JULY 2010

Page 495 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F.1.1 NOTES ON USER_MODBUS
Page 495 · click to enlarge
Page 495

MODBUS REG. LIST FOR 2350A GC F-1 MON2000

FA APPENDIX F, MODBUS REG. LIST FOR 2350A GC

F.1 INTRODUCTION – SIM_2251 & USER_MODBUS

GC Controller Modbus registers that may be of interest to the developer are of two varieties, SIM_2251 and User_Modbus. Differences are summarized in the following table.

Table F-1 Comparison of SIM_2251 and User_Modbus

Serial slave port. Serial slave port.

Modified protocol that allows floating point numbers to be transmitted over Modbus via 2251 emulation slave type.

The standard Gould Modbus protocol that accommodates PLC Emulation LO-HI.

Nearly all register contents are predefined; a few 9000-series registers can be user- defined (i.e., read-write).

Boolean (coils) are predefined. Numeric (registers) are user-defined.

Variables assigned to registers can be listed in contents of a PC-Config Report (for instructions, see Section 2.16.3; for an example report, see Appendix A,). For more detail about register contents, see Section F.2.1.

Variables assigned to registers can be listed in contents of a PC-Config Report (for instructions, see Section 2.16.3; for an example report, see Appendix A,). Boolean (coils) contents also listed in Section F.2.

When using the Modbus Test software, set Register Mode to “DANIEL” to view register contents (see Section 10.2.3).

When using the Modbus Test software, set Register Mode to “PLC LH” to view register contents (see Section 10.2.3).

F.1.1 NOTES ON USER_MODBUS

Section F.2 lists only variables included in the User_Modbus Boolean Modbus registers. These registers are not user-defined and primarily contain alarm flags, possibly useful for debugging purposes. To use the Modbus Test software to view the contents of these registers, you will need to set the Function parameter to “1 (Read Coil)”. See Section 10.2 through Section 10.4 for details on using the Modbus Test software.

JULY 2010 Introduction – SIM_2251 & User_Modbus

SIM_2251 User_Modbus

Page 496 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F.1.2 NOTES ON SIM_2251 MODBUS
Page 496 · click to enlarge
Page 496

F-2 MODBUS REG. LIST FOR 2350A GC MON2000

All other User_Modbus registers can be defined by the user. To define User_Modbus register contents (through assignment of variables), see Section 5.18.4.

To obtain a complete list of register assignments, both SIM_2251 and User_Modbus, use Model 700 to produce a PC-Config Report.

To print a PC-Config Report, see Section 2.16.3; for an example report, see Appendix A, Section F.2 and Section F.2.1.

F.1.2 NOTES ON SIM_2251 MODBUS

To use the Modbus Test software and view the contents of SIM_2251 registers, you will need to set Register Mode to “Daniel”, as noted in Table F-1.

To obtain a complete list of register assignments, both SIM_2251 and User_Modbus, use MON2000 to produce a PC-Config Report.

Introduction – SIM_2251 & User_Modbus JULY 2010

Page 497 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS REG. LIST FOR 2350A GC F-3 MON2000________________________________________________________________________________
Page 497 · click to enlarge
Page 497

MODBUS REG. LIST FOR 2350A GC F-3 MON2000

F.2 USER_MODBUS REGISTER LIST

Emerson Process Management Chromatograph Software (MON2000) — MODBUS REG. LIST FOR 2350A GC F-3 MON2000________________________________________________________________________________ diagram (page 497)
Fig. 343

JULY 2010 User_Modbus Register List

Page 498 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F-4 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000
Page 498 · click to enlarge
Page 498

F-4 MODBUS REG. LIST FOR 2350A GC MON2000

Emerson Process Management Chromatograph Software (MON2000) — F-4 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000 diagram (page 498)
Fig. 344

User_Modbus Register List JULY 2010

Page 499 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS REG. LIST FOR 2350A GC F-5 MON2000________________________________________________________________________________
Page 499 · click to enlarge
Page 499

MODBUS REG. LIST FOR 2350A GC F-5 MON2000

Emerson Process Management Chromatograph Software (MON2000) — MODBUS REG. LIST FOR 2350A GC F-5 MON2000________________________________________________________________________________ diagram (page 499)
Fig. 345

JULY 2010 User_Modbus Register List

Page 500 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F-6 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000
Page 500 · click to enlarge
Page 500

F-6 MODBUS REG. LIST FOR 2350A GC MON2000

F.2.1 SIM_2251 MODBUS REGISTER LIST

The information in the following tables is derived from engineering specification number ES-17128-005, “Model 2251 Enhanced Specification Chromatograph Controller Modbus Communication Indices” and has been updated for the 2350A GC Controllers.

Emerson Process Management Chromatograph Software (MON2000) — F-6 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000 diagram (page 500)
Fig. 346

User_Modbus Register List JULY 2010

Page 501 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS REG. LIST FOR 2350A GC F-7 MON2000________________________________________________________________________________
Page 501 · click to enlarge
Page 501

MODBUS REG. LIST FOR 2350A GC F-7 MON2000

Emerson Process Management Chromatograph Software (MON2000) — MODBUS REG. LIST FOR 2350A GC F-7 MON2000________________________________________________________________________________ diagram (page 501)
Fig. 347

JULY 2010 User_Modbus Register List

Page 502 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F-8 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000
Page 502 · click to enlarge
Page 502

F-8 MODBUS REG. LIST FOR 2350A GC MON2000

Emerson Process Management Chromatograph Software (MON2000) — F-8 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000 diagram (page 502)
Fig. 348

User_Modbus Register List JULY 2010

Page 503 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS REG. LIST FOR 2350A GC F-9 MON2000________________________________________________________________________________
Page 503 · click to enlarge
Page 503

MODBUS REG. LIST FOR 2350A GC F-9 MON2000

Emerson Process Management Chromatograph Software (MON2000) — MODBUS REG. LIST FOR 2350A GC F-9 MON2000________________________________________________________________________________ diagram (page 503)
Fig. 349

JULY 2010 User_Modbus Register List

Page 504 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F-10 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000
Page 504 · click to enlarge
Page 504

F-10 MODBUS REG. LIST FOR 2350A GC MON2000

Emerson Process Management Chromatograph Software (MON2000) — F-10 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000 diagram (page 504)
Fig. 350

User_Modbus Register List JULY 2010

Page 505 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS REG. LIST FOR 2350A GC F-11 MON2000________________________________________________________________________________
Page 505 · click to enlarge
Page 505

MODBUS REG. LIST FOR 2350A GC F-11 MON2000

Emerson Process Management Chromatograph Software (MON2000) — MODBUS REG. LIST FOR 2350A GC F-11 MON2000________________________________________________________________________________ diagram (page 505)
Fig. 351

JULY 2010 User_Modbus Register List

Page 506 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F-12 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000
Page 506 · click to enlarge
Page 506

F-12 MODBUS REG. LIST FOR 2350A GC MON2000

Emerson Process Management Chromatograph Software (MON2000) — F-12 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000 diagram (page 506)
Fig. 352

User_Modbus Register List JULY 2010

Page 507 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS REG. LIST FOR 2350A GC F-13 MON2000________________________________________________________________________________
Page 507 · click to enlarge
Page 507

MODBUS REG. LIST FOR 2350A GC F-13 MON2000

Emerson Process Management Chromatograph Software (MON2000) — MODBUS REG. LIST FOR 2350A GC F-13 MON2000________________________________________________________________________________ diagram (page 507)
Fig. 353

JULY 2010 User_Modbus Register List

Page 508 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F-14 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000
Page 508 · click to enlarge
Page 508

F-14 MODBUS REG. LIST FOR 2350A GC MON2000

Emerson Process Management Chromatograph Software (MON2000) — F-14 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000 diagram (page 508)
Fig. 354

User_Modbus Register List JULY 2010

Page 509 of the Emerson Process Management Chromatograph Software (MON2000) user manual — MODBUS REG. LIST FOR 2350A GC F-15 MON2000________________________________________________________________________________
Page 509 · click to enlarge
Page 509

MODBUS REG. LIST FOR 2350A GC F-15 MON2000

Emerson Process Management Chromatograph Software (MON2000) — MODBUS REG. LIST FOR 2350A GC F-15 MON2000________________________________________________________________________________ diagram (page 509)
Fig. 355

JULY 2010 User_Modbus Register List

Page 510 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F-16 MODBUS REG. LIST FOR 2350A GC ________________________________________________________________________________MON2000
Page 510 · click to enlarge
Page 510

F-16 MODBUS REG. LIST FOR 2350A GC MON2000

This page is intentionally left blank.

User_Modbus Register List JULY 2010

Page 511 of the Emerson Process Management Chromatograph Software (MON2000) user manual — WARRANTY CLAIM PROCEDURES
Page 511 · click to enlarge
Page 511

WARRANTY CLAIM PROCEDURES

To make a warranty claim, you, the Purchaser, must:

1. Provide Daniel Measurement and Control, Inc. or Rosemount Analytical, Inc. with proof of the Date of Purchase and proof of the Date of Shipment of the product in question.
2. Return the product to Daniel Measurement Services (DMS) within 12 months of the date of original shipment of the product, or within 18 months of the date of original shipment of the product to destinations outside of the United States. The Purchaser must prepay any shipping charges. In addition, the Purchaser is responsible for insuring any product shipped for return, and assumes the risk of loss of the product during shipment.
3. To obtain warranty service or to locate the nearest DMS office, sales office, or service center, call (713) 827-6314, fax a request to (713) 8276312, or write to:

Daniel Measurement Services 11100 Brittmore Park Drive Houston, Texas 77041

You can also contact DMS via www.emersonprocess.com/daniel.

4. When contacting DMS for product service, the Purchaser is asked to provide information as indicated on the following page entitled "Customer Repair Report".
5. For product returns from locations outside the United States, it will be necessary for you to obtain the import consignment address so that DMS's customs broker can handle the importation with the U.S. Customs Service.
6. DMS offers both on call and contract maintenance service designed to afford single source responsibility for all its products.
7. DMS reserves the right to make changes at any time to any product to improve its design and to insure the best available product.
Page 512 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F.2.1 SIM_2251 Modbus Register List
Page 512 · click to enlarge
Page 512

This page is intentionally left blank.

Page 513 of the Emerson Process Management Chromatograph Software (MON2000) user manual — COMPANY NAME: ___________________________________________________________________________
Page 513 · click to enlarge
Page 513

CUSTOMER REPAIR REPORT

FOR SERVICE, COMPLETE THIS FORM, AND RETURN IT ALONG WITH THE AFFECTED EQUIPMENT TO CUSTOMER SERVICE AT THE ADDRESS INDICATED BELOW.

COMPANY NAME:

TECHNICAL CONTACT:_____________________________________ PHONE: _________________________

REPAIR P. O. #:________________________ IF WARRANTY, UNIT S/N:_____________________________

INVOICE ADDRESS: __________________________________________________________________________

SHIPPING ADDRESS: _________________________________________________________________________

RETURN SHIPPING METHOD: _________________________________________________________________

EQUIPMENT MODEL #:____________________ S/N:__________________FAILURE DATE: _____________

DESCRIPTION OF PROBLEM: __________________________________________________________________

WHAT WAS HAPPENING AT TIME OF FAILURE? _______________________________________________

ADDITIONAL COMMENTS:____________________________________________________________________

REPORT PREPARED BY:__________________________________ TITLE: _____________________________

IF YOU REQUIRE TECHNICAL ASSISTANCE, PLEASE FAX OR WRITE THE CUSTOMER SERVICE DEPARTMENT AT:

DANIEL MEASUREMENT SERVICES DIVISION OF EMERSON PROCESS MANAGEMENT PHONE: (713) 827-6314 ATTN: CUSTOMER SERVICE FAX: (713) 827-6312 11100 BRITTMOORE PARK DRIVE HOUSTON, TEXAS 77041

FOR FASTEST SERVICE CONTACT DANIEL MEASUREMENT SERVICES VIA OUR WEBSITE:

www.emersonprocess.com/daniel

Show 8 more labels / short text on this page
_____________________________________________________________________________________________ · _____________________________________________________________________________________________ · _____________________________________________________________________________________________ · _____________________________________________________________________________________________ · _____________________________________________________________________________________________ · _____________________________________________________________________________________________ · _____________________________________________________________________________________________ · _____________________________________________________________________________________________
Page 514 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F.2.1 SIM_2251 Modbus Register List
Page 514 · click to enlarge
Page 514
Page 514 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F.2.1 SIM_2251 Modbus Register List
Page 515 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F.2.1 SIM_2251 Modbus Register List
Page 515 · click to enlarge
Page 515

This page is intentionally left blank.

Page 516 of the Emerson Process Management Chromatograph Software (MON2000) user manual — F.2.1 SIM_2251 Modbus Register List
Page 516 · click to enlarge
Page 516

Daniel Measurement and Control, Inc., Daniel Measurement Services, Inc., and Rosemount Analytical Inc., Divisions of Emerson Process Management, reserves the right to make changes to any of its products or services at any time without prior notification in order to improve that product or service and to supply the best product or service possible.

www.emersonprocess.com

Emerson Process Management Chromatograph Software (MON2000) — F.2.1 SIM_2251 Modbus Register List diagram (page 516)
Fig. 356 · F.2.1 SIM_2251 Modbus Register List
Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson Process Management Chromatograph Software (MON2000)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Process Management Chromatograph Software (MON2000)

Be the first to ask — other owners or our team will answer.

Original document

Emerson Process Management Chromatograph Software (MON2000) manual PDF

516 pages·5.7 MB·Whiteboard Accessories·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…