Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Specifications

Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process
temperature is lowered to a safe level!
Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.
The wetted sensor materials may not be compatible with process composition and operating conditions. Application
compatibility is entirely the responsibility of the user.
The buffer in the protective boot may cause skin or eye irritation.
Special Conditions for safe use
build up of an electrostatic charge.
to earth. This must be taken into consideration at installation.


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
For sensor dimensions, see Figure 1. For sensor orientation, see Figure 2. For VP sensor or -02 sensor (no preamp) wiring, see Figures 3 thru 22. For -01 sensor (with preamp) wiring, see Figures 23 thru 28. Two Point Buffer Calibration
Select two stable buffer solutions, preferably pH 4.0 and 10.0 (pH buffers other than pH 4.0 and pH 10.0 can be used as long as the pH values are at least two pH units apart).
1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/transmitter. 2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water. 3. Repeat steps 1 and 2 using the second buffer solution. 4. The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.17 mV/pH. Over time the sensor will age, both in the process and in storage, and will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mV/pH.
For maximum accuracy, the sensor can be standardized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor’s slope but will simply adjust the analyzer’s reading to match that of the known process pH.
Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement.

Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
1. After making an electrical connection between the sensor and the instrument, obtain a standard solution of saturated quinhydrone. This can also be made quite simply by adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble, therefore only a few crystals will be required. 2. Immerse the sensor in the standard solution. Allow 1–2 minutes for the ORP sensor to stabilize. 3. Adjust the standardize control of the instrument to the solution value shown in the table below. The resulting potentials, measured with a clean platinum electrode and saturated KCl/AgCl reference electrode, should be within ±20 millivolts of the value shown in the table below. Solution temperature must be noted to ensure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used. 4. Remove the sensor from the buffer, rinse, and install in the process.
Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement.
Figure 1. Dimensional Drawing



TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 2. Flow Through and Insertion Installation

Figure 3. Wiring for 81 Figure 4. Wiring for 1181



Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 5. Wiring 81 through Remote Junction Box Figure 6. Wiring 1181, 1050/1060, and 1003/1023 through Remote Junction Box



TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 7. Wiring Details of 396P-01-50 for use with and without Junction Box (PN 22719-02) for 1181, 1050, and 1060.

Figure 8. Wiring for 2081 Figure 9. Wiring for 3081, 4081



Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 10. Wiring 2081 through Remote Junction Box Figure 11. Wiring 3081, 4081 through Remote Junction Box


Figure 12. Wiring for 1054 A/B and 2054


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 13. Wiring 1054 A/B and 2054 Through a Remote Junction Box


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 14. Wiring 54 through Remote Junction Box

Figure 15. Wiring for 54


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 16. Wiring Details of 396P-01-54 for use with and without Junction Box (PN 22719-02) for 1054, 1054A, 1054B, 2054, 2081, and Extension Cable (9200254).


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 17. Wiring of 396P-01-55 for use with 54, 81, 3081, and 4081.

Figure 18. Junction Box (PN 23550-00) for 54, 3081, 4081, and 81.


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 19. Wiring Details of 396P-01 Gray Cable for use with 54e pH/ORP-01-09

If present, reove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT and
1056, terminate Inner Drain with a wire nut.
Figure 20. Wiring Details of 396P-01 Blue Cable for use with 54e pH/ORP-01-09

If present, reove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056, terminate Inner Drain with a wire nut.

Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 21. Wiring Details of 396P-02 Gray Cable for use with 54e pH/ORP-01-09

If present, reove shrink sleeve from gray reference wire before connecting wire to terminal.
For XMT and 1056, terminate Inner Drain with a wire nut.
Figure 22. Wiring Details of 396P-02 Blue Cable for use with 54e pH/ORP-01-09

If present, reove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT and
1056, terminate Inner Drain with a wire nut.

TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 23. Wiring Details of 396PVP Gray Cable for use with 54e pH/ORP-01-09

For XMT and 1056 terminate Inner Drain with a wire nut. VP8 Cable part number 24281-xx shown.
Figure 24. Wiring Details of 396PVP Blue Cable for use with 54e pH/ORP-01-09

For XMT and 1056 terminate Inner Drain with a wire nut. VP8 Cable part number 24281-xx shown.

Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 25. Wiring Details of 396P-01 for use with 1055-01-10-22

Figure 26. Wiring Details of 396P-01 for use with 1055-01-10-22-32


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 27. Wiring Details of 396P-02 for use with 1055-01-10-22-32

Figure 28. Wiring Details of 396PVP for use with 1055-01-10-22-32


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 29. Wiring Details of 396P-01 Gray Cable for use with XMT-P-HT-10

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal.
For XMT and 1056 terminate Inner Drain with a wire nut.
Figure 30. Wiring Details of 396P-01 Blue Cable for use with XMT-P-HT-10

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal.
For XMT and 1056 terminate Inner Drain with a wire nut.

TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 31. Wiring Details of 396P-02 Gray Cable for use with XMT-P-HT-10

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056 terminate Inner Drain with a wire nut.
Figure 32. Wiring Details of 396P-02 Blue Cable for use with XMT-P-HT-10

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056 terminate Inner Drain with a wire nut.

Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 33. Wiring Details of 396PVP Gray Cable for use with XMT-P-HT-10

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056 terminate Inner Drain with a wire nut.
Figure 34. Wiring Details of 396PVP Blue Cable for use with XMT-P-HT-10

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056 terminate Inner Drain with a wire nut.

TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 35. Wiring through j-box for 1056

Figure 36. Wiring Details of 396P-01 Gray Cable for 1056/56

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056 terminate Inner Drain with a wire nut.

Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 37. Wiring Details of 396P-01 Blue Cable for 1056/56

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056 terminate Inner Drain with a wire nut.
Figure 38. Wiring Details of 396P-02 Blue Cable for 1056/56

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056 terminate Inner Drain with a wire nut.

TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 39. Wiring Details of 396P-02 Gray Cable for 1056/56

If present, remove shrink sleeve from gray reference wire before connecting wire to terminal. For XMT
and 1056 terminate Inner Drain with a wire nut.
Figure 40. Wiring Details of 396PVP Gray Cable for 1056/56

For XMT and 1056 terminate inner Drain with a wire nut. VP8 Cable part number 24281-xx shown

Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 41. Wiring Details of 396PVP Blue Cable for 1056/56

For XMT and 1056 terminate inner Drain with a wire nut. VP8 Cable part number 24281-xx shown
Figure 42. pH/ORP Sensor Wiring for 1056/1057/56/396VP-70 (Blue Cable)


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 43. Wiring Details of 396P-01 Gray Cable for 1057

Figure 44. Wiring Details of 396P-01 Blue Cable for 1057


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 45. Wiring Details of 396P-02 Gray Cable for 1057

Figure 46. Wiring Details of 396P-02 Blue Cable for 1057


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 47. Wiring Details of 396PVP Gray Cable for 1057

Figure 48. Wiring Details of 396PVP Blue Cable for 1057


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 49. pH/ORP Sensor Wiring for 1066/396P-01 (Gray Cable)

Figure 50. pH/ORP Sensor Wiring for 1066/396P-01 (Blue Cable)


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 51. pH/ORP Sensor Wiring for 1066/396P-02 (Gray Cable)

Figure 52. pH/ORP Sensor Wiring for 1066/396P-02 (Blue Cable)


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 53. pH/ORP Sensor Wiring for 1066/396PVP (Gray Cable)

Figure 54. pH/ORP Sensor Wiring for 1066/396PVP (Blue Cable)


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 55. pH/ORP Sensor Wiring for 1066/396PVP-70 (Blue Cable)

Figure 56. Wiring Details of 396P-01 Gray Cable for 5081-P-HT


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 57. Wiring Details of 396P-01 Blue Cable for 5081-P-HT

Figure 58. Wiring Details of 396P-02 Gray Cable for 5081-P-HT


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 59. Wiring Details of 396P-02 Blue Cable for 5081-P-HT

Figure 60. Wiring Details of 396PVP Gray Cable for 5081-P-HT


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 61. Wiring Details of 396PVP Blue Cable for 5081-P-HT

Figure 62. pH/ORP Sensor Wiring for 5081/398RVP-70 (Blue Cable)


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 63. Wiring Details of 396P-01 Gray Cable for use with 6081

Figure 64. Wiring Details of 396P-01 Blue Cable for use with 6081


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 65. Wiring Details of 396P-02 Gray Cable for use with 6081

Figure 66. Wiring Details of 396P-02 Blue Cable for use with 6081


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP
Figure 67. Wiring Details of 396PVP Gray Cable for use with 6081

Figure 68. Wiring Details of 396PVP Blue Cable for use with 6081


Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
Figure 69. pH/ORP Sensor Wiring for 6081/3400HTVP-70 (Blue Cable)


TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP

Part Number 1700716

Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013
22698-00 Preamplifier plug-in for junction box, for 1003
22698-02 Preamplifier plug-in for junction box, for 1181, 1050
22698-03 Preamplifier plug-in for junction box for 1054 A/B, 2054, 2081
22743-01 Pt100 preamplifier for 1181
22744-01 3K Preamplifier for 1181
23557-00 Preamplifier for junction box for 54, 3081, 81, 4081
9210012 Buffer solution, 4.01 pH, 16 oz.
9210013 Buffer solution, 6.86 pH, 16 oz.
9210014 Buffer solution, 9.18 pH, 16 oz
R508-80X ORP solution, 460 mv ± 10 at 20 °C
12707-00 Jet Spray Cleaner
25645-07 15 ft. VP6 Cable
24281-00 15 ft. VP8 Cable
24281-01 25 ft. VP8 Cable
24281-02 2.5 ft. VP8 Cable
24281-03 50 ft. VP8 Cable
24281-04 100 ft. VP8 Cable
24281-05 4 ft. VP8 Cable
24281-06 10 ft. VP8 Cable
24281-07 20 ft. VP8 Cable
24281-08 30 ft. VP8 Cable
Part Description

TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP Rev. J
facebook.com/EmersonRosemountAnalytical
Credit Cards for U.S. Purchases Only.
youtube.com/user/RosemountAnalytical
Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 rosemountanalytical.com
©2013 Rosemount Analytical, Inc. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
© Rosemount Analytical Inc. 2013