pH/ORP Sensor


READ THIS PAGE BEFORE
Installations near flammable liquids or in haz- ardous area locations must be carefully eval- uated by qualified on site safety personnel. This sensor is not Intrinsically Safe or Explosion Proof.
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international stan- dards. Because these instruments are sophisticated tech- nical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
To secure and maintain an intrinsically safe installation, the certified safety barrier, transmitter, and sensor combi nation must be used. The installation system must com- ply with the governing approval agency (FM, CSA or BASEEFA/CENELEC) hazardous area classification requirements. Consult your analyzer/transmitter instruc tion manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal - lation is entirely the responsibility of the user.
Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2001
Emerson Process Management



MODEL 381 pH/ORP TABLE OF CONTENTS
Figure No. Title Page 1-1 Dimensional Drawing 3 1-2 Mounting Drawing 3 2-1 Model 381 Sensor Component Locator Diagram 6 2-2 Submersion Installation Diagram 8 2-3 Insertion Installation Diagram 8 2-4 Flow Through Installation Diagram (Code 03 or 04) 9 2-5 Installation Details Low Flow Conditions Model 381 (Code 03-04) 9 2-6 Retro-Fit Kit and Code-04 Flow Powered Cleaner, Model 381 10 2-7 Flow Through Diagram and Wiring (Code 03 or 04) 11 2-8 Remote Junction Box with Internal Preamp 12 2-9 Wiring with Weatherproof or Div. 2 (Code 2) Junction Box 13 4-1 Model 381 Ultrasonic Cleaner and Wiring 23 4-2 Dimensional and Component Locator Ultrasonic Generator 24
This manual contains instructions for installation and operation of the Model 381 pH/ORP Sensor.
The following list provides notes concerning all revisions of this document.
A 6/94-1/00 This is the initial release of the product manual. The manual has been reformatted to reflect the Emerson documentation style and updated to reflect any changes in the product offering.
B 5/02 Updated multiple drawings throughout manual.
D 10/07 Updated page 3 code 00 insertion is no longer available.
C 3/05 Updates

MODEL 381 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS
1.1 FEATURES AND APPLICATIONS The Model 381 Sensor measures the pH or the Oxidation Reduction Potential (ORP) of aqueous solu- tions in pipelines, open tanks or ponds. The sensor is used in most industrial applications including SO2 scrubbers, rinse tanks, plating baths, reverse osmosis units, and water and waste treatment plants. The sensor is housed in a molded PES body and has two O-ring seals with breach lock threads which secure the PES cover. This provides a waterproof union for long operating life and easy removal for rou- tine maintenance. The modular body design eliminates the need for internal mounting brackets, terminal brackets and screws. All components are screw-type or plug-in- place, allowing for fast simple service. The integral preamplifier amplifies the high imped- ance glass electrode signal at the sensor, providing a transmission capability of up to three miles. The Rosemount Analytical Inc. method of preamplification has become the industry standard for reliable pH measurement. In submersion applications at tempera- tures greater than 80°C, the preamplifier must be located in a remote junction box. A standard feature of the Model 381 is the double junction, gel-filled reference cell which reduces the need for special applications engineering where unknown reference cell contamination, such as sul- fides, may exist. The gel solution maintains its viscosi-
ty at high temperatures, and resists the effect of pumping and dilution, resulting in an extended service life. A choice of wood (Code 20) or ceramic (Code 21) liquid junction is available to provide resistance to fouling and subsequent loss of electrical contact between the reference electrode and process fluid. In flow applications where crystalline or viscous coat- ings, such as calcium carbonate or petroleum oils, may coat the electrode surfaces and impede the sen- sor's performance, the flow-powered cleaning option (Code 04) may be used. A specially designed flow chamber directs the process fluid in a circular path carrying three Teflon1 balls which clean the electrode surface, physically preventing accumulation of coating materials. Flow-powered cleaning is suitable for all hazardous area applications. In submersion (Code 02) applications, the ultrasonic cleaning option may be specified. This assembly includes the ultrasonic transducer and generator, mounting hardware, and 20 ft. of interconnection cable. Ultrasonic cleaning is not suitable for haz- ardous area applications. The Model 381 is compatible with all Rosemount Analytical pH/ORP analyzers. For use with the 1054/2054 Series, 2081 and other pH analyzers which utilize a PT-100 automatic temperature com- pensator order the Model 381 pHE.
1 Teflon is a registered trademark of E.I. du Pont de Nemours & Co.

MODEL 381 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS
1.2 SENSOR SPECIFICATIONS Materials of Construction:

Kynar/ceramic (Code 21). Process Connections: Submersion: 3/4 inch MNPT.
Insertion: 2 inch MNPT. Flow Through:3/4 inch MNPT. Dimensions: (see Figure 1-1). Measuring Range: pH 0-14
ORP –1500mV to +1500mV Temperature Compensation: (Model 381 pH only,
not required for Model 381 ORP). Automatic, 0 to 100°C (32 to 212°F). Pressure Rating: 790 kPa abs at 100°C
with common shield. Weight/Shipping Weight: 1.0 kg/1.7 kg (2.2 lb./ 3.7 lb.).
2 Viton is a registered trademark of E.I. du Pont de Nemours & Co. 3 Kynar is a trademark of Elf Atochem North America, Inc.
FIGURE 1-1. Dimensional Drawing


FIGURE 1-2. Mounting Drawing

MODEL 381 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS
1.3 ORDERING INFORMATION Model pH/ORP Sensor is housed in a PES body suitable for insertion, submer- sion or flow through installation. The sensor includes an integral preamplifier, measuring electrode, double junc- tion gel-filled reference cell, automatic temperature compensation for pH, and a choice of two cable lengths of either 15 ft. or 50 ft. (4.5 or 15.2 m).
CODE FUNCTION
pH Requires Code 30
ORP Requires Code 32 (not for use with 1054/1054A ORP Series)
pHE Requires Code 31 (for use with 1054/1054A/2054 Series and 2081)
04 Flow Through with Flow Powered Cleaner (not available with ORP)
10 General purpose pH
11 High pH Electrode, requires Code 21. For sodium concentrations greater than 1% or continuous measurement above 11 pH.
12 Ruggedized pH Electrode. For abrasive solutions.
13 Platinum ORP Electrode.
14 Gold ORP Electrode.
15 HF pH Electrode. For HF concentrations less than 100 ppm.
20 Wood/KYNAR Liquid Junction
21 Ceramic/KYNAR Liquid Junction
30 Automatic T.C., 3.0K ohm. For Models 1003, 1050 & 1181 pH
31 Automatic T.C., PT-100. For Models 1054/2054 Series, and 2081
32 Automatic T.C., NOT REQUIRED (For Models 1023, 1060 and 1181 ORP)
40 15 ft. cable for integral sensor preamp
41 15 ft. coax cable w/BNC connector for remote sensor preamp and integral analyzer preamp (1054A-54/2054-54 only)
42 15 ft. coax cable w/o BNC connector for integral analyzer preamp (1054, 1181-43/44 & 2081-05 only)
43 50 ft. cable for integral sensor preamp
50 Preamp for Models 1181 pH/ORP & 1050/1060
51 Preamp for Models 1003/1023
52 Preamp NOT REQUIRED
54 Preamp for Models 1054/2054 Series, and 2081
381 pH 00 10 20 30 40 51 EXAMPLE
02 Submersion · 03 Flow Through

4 This page left blank intentionally.

2.1 INSTALLATION. Prepare the sensor for instal lation as follows (see Figure 2-1):
2.2 SUBMERSION INSTALLATION (Code 02). To install the sensor in process proceed as follows (see Figure 2-2):


FIGURE 2-1. Model 381 Sensor Component Locator Diagram

2.3 INSERTION INSTALLATION (Code 00). To install the sensor in the side of a tank, in a pipeline, or in a pipe “tee”, proceed as follows (see Figure 2-3):
* In "guarded" flow position, solid particles in the flow stream are diverted from electrode.
2.4.1 Installation Procedures For Low Flow Conditions. It is extremely important that the measur- ing electrode and liquid junction of any pH/ORP sensor remain in contact with the process fluid at all times.
Under low flow conditions or where flow fluctuations are common, sensor contact with the process fluid may be interrupted unless piping procedures are followed to prevent it.
A customer supplied valve should be installed to elimi- nate head pressure when the sensor is being removed from the process. This valve may also serve as a grab sample take-off point.
2.4 FLOW THROUGH INSTALLATION (Code 03). To install the sensor in the flow cell proceed as follows (see Figure 2-4):
The installation drawing (Figure 2-5) shows the recom- mended piping procedures for the Model 381 pH/ORP sensor in low flow conditions.
2.5 FLOW POWERED CLEANER INSTALLATION (Code 04). To install the sensor in a flow cell with flow powered cleaner, proceed as follows (see Figure 2-6):
2.6 WIRING. Connect the wire lugs to the transmitter terminals as shown in Figures 2-7 thru 2-10.


FIGURE 2-2. Submersion Installation Diagram

FIGURE 2-3. Insertion Installation Diagram


FIGURE 2-4. Flow Through Installation Diagram (Code 03 or 04)

FIGURE 2-5. Installation Details Low Flow Conditions Model 381 (Code 03 and 04)



PRESSURE DROP AND RECOMMENDED FLOW RATES- FLOW POWERED CLEANER (CODE 04)
POSSIBLE ELECTRODE DAMAGE ABOVE THE THIS POINT
PRESSURE DROP
RECOMMENDED OPERATING RANGER
INEFFECTIVE CLEANING BELOW THIS POINT
1 2 3 4 5 6 7 8
5 10 15 20 25 30
FIGURE 2-6. Retro-Fit Kit Code 04 Flow Powered Cleaner, Model 381
kPa · gpm @ 60° · liters/min

ARE GIVEN

INCH
DWG. NO. REV. 40038105 L
FIGURE 2-7. Flow Through Diagram and Wiring (Code 03 or 04)


ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV. 40038103 J
FIGURE 2-8. Remote Junction Box with Internal Preamp


ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV. 40038116 I
DWG. NO. REV. 40038115 K
FIGURE 2-9. Wiring with Weatherproof or Div. 2 (Code 52) Junction Box

2.7 START-UP AND CALIBRATION 2.8 MODEL 381 pH and pHE 2.8.1 SENSOR PREPARATION. Shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped and read- ings with an accuracy of ± 0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be calibrated as a loop with its com- patible analyzer or transmitter.
2.9.2 Calibration
2.8.2 CALIBRATION USING BUFFER SOLUTIONS OR GRAB SAMPLES. The pH sensor-analyzer/ trans- mitter loop may be calibrated by submersing the sensor in a buffer solution (standard solutions of known pH val- ues) or in a process grab sample whose pH value maybe checked by a calibrated laboratory or portable pH meter. Please refer to the analyzer's or the trans- mitter’s respective instruction manuals for proper calibration procedures.
4. Adjust the standardize control of the instrument to the solution value shown in Table 2-1. The resulting potentials, measured with a clean platinum electrode and saturated KCl/AgCl reference electrode, should be within ±20 millivolts of the value shown in Table 2- 1. Solution temperature must be noted to insure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used.
2.9 MODEL 381 ORP. 2.9.1 MODEL 396P ORP. Most industrial applications have a number of ORP reactions occurring in sequence or simultaneously. There can be several components that are oxidized or reduced by the reagents that are used. Theoretically, the ORP poten- tial is absolute because it is the result of the oxidation- reduction equilibrium. However, the actual measured potential is dependent on many factors, including the condition of the surface of the ORP platinum elec- trode. Therefore, the sensor should be allowed 1-2 hours to become “conditioned” to the stream when first set-up or after being cleaned.
TABLE 2-1. ORP of Saturated Quinhydrone Solution
pH 4 Solution pH 7 Solution Temp °C 20 25 30 20 25 30 Millivolt Potential 268 264 260 94 87 80
(In Millivolts)

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE
3.1 GENERAL. This section provides instructions for performing periodic maintenance and for determining the cause of a malfunction (troubleshooting).
3.2 PERIODIC MAINTENANCE. The following period- ic maintenance is based on operation in a water sam- ple. Determine the schedule for the operation being maintained from past experience, or by analyzing data gathered since initial installation of the sensor.
3.2.1 Monthly. Perform the following maintenance monthly:
3.2.2 Annually. Perform the following maintenance annually (not required for combination electrode):

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE
Slowly and carefully add 56.2 milliliters of concentrated sulfuric acid. Add sufficient water to bring the total solu- tion volume up to 1000 ml. This standard ORP solution, although not as simple to prepare as the quinhydrone recipe, is much more stable, and will maintain its milli- volt value for approximately one year when stored in glass containers. This solution (ferric/ferrous ammoni- um sulfate) will produce a nominal ORP of 476 +20 mV at 25°C when used with a saturated KCl/AgCl refer- ence electrode and platinum measuring electrode. Some tolerance in mV values is to be expected due to the rather large liquid reference junction potentials which can arise when measuring this strongly acidic and concentrated solution. However, if the measuring electrodes are kept clean and in good operating condi- tion, consistently repeatable calibrations can be carried out using this standard solution.
3.3.1 Platinum Electrode Check. The platinum elec- trode may be checked as follows: There are two types of standard solutions which may be used to check the ORP electrode/transmitter system.
Type 1: One type of commonly used ORP standard solution is the saturated quinhydrone solution. A com- monly used ORP standard solution is a saturated quin- hydrone solution. This can be made by simply adding a few quinhydrone crystals to either a 4 pH or a 7 pH buffer. Quinhydrone is only sightly soluble so only a few crystals will be required. The solution will have a yellow color. The resulting potentials should be within ±20 millivolts of the value shown in Table 3-1. The ORP value of saturated quinhydrone solution is not stable for long periods of time and therefore new solutions should be made each time.
3.3.2 Cleaning Platinum Electrode. The electrode is never exposed to these undesirable compounds. In the event poisoning is suspected, the electrode can be restored to normal operation by simply cleaning the plat- inum electrode with baking soda. Polish it by rubbing it with a damp paper towel and baking soda until a bright, shiny appearance is attained.
TABLE 3-1. ORP of Saturated Quinhydrone Solutions
pH 4 pH 7 Temperature °C 20 25 30 20 25 30 Millivolt Potential 268 264 260 94 87 80

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE
TABLE 3-1. Troubleshooting
Trouble Probable Cause Remedy
Meter reads off scale. Reference electrolyte depleted. Replace reference electrolyte as instructed in Section 3.2.2.
Defective preamplifier. Check preamplifier as instructed in Section 3.4 and replace preamplifier if defective.
T.C. element shorted. Check T.C. element as instructed in Section 3.5, and replace body if defective.
Electrode not in solution or sample Make sure sensor is in solution or that stream is not full. sample stream is full (refer to Section 2.0 for installation details).
Open glass electrode. Replace glass electrode.
Plugged liquid junction. Replace liquid junction. Recharge reference cell.
Reference element open- Replace reference element. no contact.
Meter reads between 3 and 6 pH Electrode cracked. Replace electrode. regardless of actual pH of solution or sample.
Meter indication swings widely T.C. element shorted. Check T.C. element as instructed in in AUTO T.C. Mode. Section 3.5 and replace body if defective.
Span between buffers extremely T.C. element open. Check T.C. element as instructed in short in AUTO T.C. mode. Section 3.5 and replace body if defective.
Sluggish or slow meter indication Electrode coated. Clean electrode as instructed in for real changes in pH level. Section 3.2.1.
Electrode defective. Replace electrode.
Transmitter cannot be Electrode coated or cracked. Clean electrode as instructed in standardized. Section 3.2.1 and, if cracked, replace electrode.
Reference chamber is contaminated Replace gel solution as instructed in Section 3.2.2.
Defective preamplifier. Check preamplifier as instructed in Section 3.4 and replace if defective.
If trouble still persists, micro-junction Replace micro-junction electrode. electrode is defective.
Meter short spans. Old glass electrode (greater than 1 Replace electrode. year use).
Coated glass. Clean with soft cloth and clean water.

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE
3.4 PREAMPLIFIER TROUBLESHOOTING. To deter- mine if the preamplifier is operable, proceed as follows:
TABLE 3-2 TEMPERATURE vs RESISTANCE OF AUTO
1. Remove the sensor from process. 2. Remove the cover from the body by grasping the body and rotating the cover 1/4 turn in the count- er-clockwise direction. 3. When the cover breaks loose from the body, pull straight out on the cover to expose the com- ponents attached to the body and disconnect the BNC connector from the preamplifier, leaving the cable connected. Inspect the internal pre amplifier area for signs of moisture or corrosion; moisture can short out the preamplifier. Dry all parts thor- oughly before testing.
Temperature °C Resistance (Ohms) ± 1%
0 2670 100.0 10 2802 103.8 20 2934 107.7 25 3000 109.6 30 3066 111.5 40 3198 115.4 50 3330 119.2 60 3462 123.1 70 3594 126.9 80 3726 130.8 90 3858 134.6 100 3990 138.5
3.5 T Automatic Temperature Compensator. The temper- ature compensator element is temp erature sensitive and can be checked with an ohmmeter. Resistance increases with temperature.
The 3K element will read 3000 ohms ± 1% at 25°C (77°F) and a Pt-100 will read 110 ohms. Resistance varies with tem- perature for a 3K and Pt-100 element and can be determined according to Table 3-2 or the following formula:
Where RT = Resistance T = Temperature in °C
Refer to Table 3-3 for Ro and R1 values
Ro and R1 VALUES FOR TEMPERATURE
Temperature Compensation Element Ro R1
RT=Ro [l+R1 (T-20)]

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE
TABLE 3-4. Model 381 Spare Parts
RECOMMENDED PART NUMBER DESCRIPTION CODE QUANTITY 2000734 Liquid junction wood/Kynar 20 2 2000735 Liquid junction ceramic/Kynar 21 2 2001928 O-ring lubricant ALL 2 22692-00 Body, Uniloc standard T.C. 30 22692-03 Body, Pt100 T.C. 31 22692-06 Body w/o T.C. 32 22693-00 Cover, for integral preamp, w/15 ft. cable 40 22693-01 Cover, for remote preamp, w/15 ft. coax 41 22693-02 Cover for remote preamp, w/15 ft. coax, no BNC 42 22693-03 Cover, for integral preamp, w/15 ft. cable, Model 1054/2054 Series pHE-40 22693-04 Cover, for integral preamp w/50 ft. cable 43 22693-05 Cover, for integral preamp, w/50 ft. cable, Model 1054/2054 Series pHE-43 22693-06 Cover, for remote preamp, w/15 ft. coax, w/BNC Model 1054/2054 Series pHE-41 22694-00 General purpose pH electrode 10 2 22694-01 High pH electrode 11 2 22694-02 Ruggedized pH electrode 12 2 22694-03 HF resistive pH electrode 15 2 22698-00 Preamplifier, Models 1003 pH and 1023 ORP 51 1 22698-02 Preamplifier, Models 1181 pH/ORP, 1050 pH/1060 ORP 50 1 22698-03 Preamplifier, Models 1054/2054 Series 54 1 22723-00 Kit, cleaning balls 04 1 22731-00 Platinum ORP electrode 13 2 22751-00 (Qty. 5) O-ring, Viton 2-037, upper body kit (9550141) ALL 1 set 22751-01 (Qty. 5) O-ring, Viton 2-040, lower body kit (9550128) ALL 1 set 22811-01 Retro-fit kit, flow powered cleaner 00, 02 22892-00 Retro-fit kit, flow powered cleaner 03 23018-00 Micro junction, reference element ALL 2 32602-00 Flow cell 03,04 32605-00 Electrode shroud 02, 03 32606-00 Coupling nut 03,04 32792-00 Shroud, modified 04 32793-00 Ring, chamber 04 32794-00 Chamber 04 7900058 Gold ORP electrode 14 2 9210342 KCl Gel reference solution, 10 oz./250 ml ALL 2 9550146 (Qty. 1) O-ring, shroud assembly 02, 03, 04 2 9550147 (Qty. 1) O-ring, flow cell assembly 03, 04 2
TABLE 3-5. Accessories
PART NUMBER DESCRIPTION 1000856 Handrail mounting bracket for submersion service. For use with Models 1003 & 1023. 1000857 Handrail mounting bracket for submersion service. For use with remote NEMA 4X junction box. 23296-00 Remote Division 2 junction box for internal preamp for use with Models 1181pH/ORP, 1050/1060, 1003/1023 23296-01 Remote Division 2 junction box for internal preamp for use with Models 1054/1054A/2054pH ORP/2081pH. 22996-00 Remote weatherproof junction box for internal preamp. 22719-02 Remote weatherproof junction box without preamp, for cable extension. 22727-00 Ultrasonic cleaner with generator, 115 VAC/230 VAC 50/60 Hz, switch selectable. Includes mounting hardware and 20 feet of interconnecting cable. (For submersion service). 9200160 Extension cable, sensor or remote preamp to analyzer, 4 conductor, 22 AWG, shielded (Specify length). 22811-00 Flow cell with electrode shroud. 22892-00 Flow powered cleaner retrofit kit (for use with Code 03 sensors).

20 This page left blank intentionally.

MODEL 381pH/ORP SECTION 4.0 ULTRASONIC CLEANER
4.1 ULTRASONIC CLEANER (P/N 22727-00). This accessory provides an ultrasonic cleaner and mount- ing hardware for either continuous or inter-mittent cleaning of the measuring electrode and liquid junction on the Model 381. The information in the following paragraph is supplemental to the information in Sections 1.0 thru 3.0.
4.2 SPECIFICATIONS. The ultrasonic cleaner oper- ates on either 115 Vac or 230 Vac, 50/60 Hz. The mode is jumper selectable within the generator enclosure. The generator is shipped in the 115 Vac configuration.
The power output is approximately 1400 volts at 0.002 amp, at a frequency of 40 to 45 kHz. The transducer is supplied with 20 feet of interconnecting cable in a polypropylene protective sheath. Output connections are made at TB1-4 for R.F. (+) and TB1-2 for RF(-). Make sure the AC power line ground is properly con- nected to the generator’s input terminals.
4.3 INSTALLATION. The ultrasonic generator is shipped with special cushioning material. Before open- ing the container or removing any packing material, all items should be inspected for any possible shipping damage. If any damage is found, the carrier should be notified. Check the generator enclosure for damage to the finish; loosen the door clamps and inspect the cir- cuit board and components. If there is apparent dam- age, contact the carrier for instructions. The packing container and cushioning material should be retained should it ever become necessary to return the genera- tor to the factory. Install the ultrasonic cleaner as fol- lows:
4.4 OPERATION. Submerge the sensor into sample and start the sensor as instructed in Section 2.0. After measuring system is in operation, apply AC power to the ultrasonic cleaner. Over a period of time (usually several days), gradually increase the gain until the power setting is sufficient to keep the electrodes free of process buildup.

MODEL 381pH/ORP SECTION 4.0 ULTRASONIC CLEANER
4.4 Operation (continued)
4.5 MAINTENANCE. The ultrasonic cleaner should operate continuously (with the exception of carbonate and phosphate slurries) and requires no maintenance, calibration, or adjustment, except that the power adjustment should be as low as possible consistent with good cleaning action. The ultrasonic transducer is completely sealed and should not be operating while troubleshooting the sensor or transmitter.
NOTE When the ultrasonic cleaner is used on some slurry applications (such as car- bon- ates and phosphates), continuous ultra- sonic cleaning is not recommended as this may tend to compact the solids of the slur- ry on the electrode into a hard, cement like coating. Instead, it is recommended that a shock-type treatment be applied (i.e., for every three hours of operation, employ 30 minutes of ultrasonic cleaning). The inter- val and duration of the cleaning, as well as the power setting utilized, are determined by the individual application. Factors such as the concentration of carbonates or phosphates, flow velocities, temperature, etc., must all be considered.
1. Troubleshooting. In case problems arise in the functioning of either the generator or the trans- ducer, the entire assembly should be returned to the factory for servicing. No electronic repair work of any kind should be attempted in the field, or the warranty will be voided. Whether the unit is still under warranty or not, field troubleshooting is not recommended, and Rosemount Analytical cannot be liable for operation or safety of equipment that has been repaired or modified by persons other than its factory or their trained and authorized Field Engineers. If your unit’s warranty has expired, and field troubleshooting is still desired, waveforms, frequency and voltages can be obtained directly from Branson Ultrasonic, Eagle Road, Danbury, CT 06810, (203) 796-0400. The circuit is fundamentally identical for both the old and new version.
1. Output Power Adjustment. The output power adjustment on the ultrasonic generator is located on the powerstat. The powerstat is adjusted with the knob marked as PS: (Refer to Figure 4-2), and determines the cleaning efficiency of the unit by adjusting the power input of the generator. Proper adjustment is made by turning the powerstat knob counter-clockwise (minimum power position) and gradually increasing the power, by turning the knob clockwise, until the power is sufficient to keep the electrode clean. This is usually accom- plished over a period of several days.
4.6 REPLACEMENT PARTS. The ultrasonic generator may be ordered as a unit (P/N 22990-00). The ultra- sonic transducer is a sealed unit and cannot be conve- niently repaired in the field. While it is not likely that the transducer will fail electrically or mechanically, certain process streams may contain corrosive compounds which, in time could attack and ultimately destroy the 316 Stainless Steel faceplate. The number of spare transducers (P/N 2001463) needed on hand as replacement stock, should be determined on an indi- vidual basis, with regard to specific application condi- tions.

MODEL 381pH/ORP SECTION 4.0 ULTRASONIC CLEANER

MILLIMETER
INCH
DWG. NO. REV. 40038156 F
FIGURE 4-1. Model 381 Ultrasonic Cleaner and Wiring

MODEL 381pH/ORP SECTION 4.0 ULTRASONIC CLEANER
ARE GIVEN

INCH
DWG. NO. REV. 40037505 D
FIGURE 4-2. Dimensional and Component Locator Ultrasonic Generator

SECTION 5.0 RETURN OF MATERIAL
5.1 GENERAL. To expedite the repair and return of instruments, proper communication between the cus- tomer and the factory is important. The “Return of Materials Request” form is provided for you to copy and use in case the situation arises. The accuracy and completeness of this form will affect the processing time of your materials. Call 1 (949) 757-8500 for a Return Materials Authorization (RMA) number.
Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606
RMA No
5.2 WARRANTY REPAIR. The following is the proce- dure for returning instruments still under warranty.
Model No
5.3 NON WARRANTY REPAIR.

RETURN OF MATERIALS REQUEST •IMPORTANT! This form must be completed to ensure expedient factory service.
FROM: RETURN BILL TO: _____________________________ _____________________________ _____________________________
E R
N O
CUSTOMER/USER MUST SUBMIT MATERIAL SAFETY SHEET (MSDS) OR COMPLETE STREAM COMPOSITION, AND/OR LETTER CERTIFYING THE MATERIALS HAVE BEEN DISINFECTED AND/OR DETOXIFIED WHEN RETURNING ANY PROD- UCT, SAMPLE OR MATERIAL THAT HAVE BEEN EXPOSED TO OR USED IN AN ENVIRONMENT OR PROCESS THAT CON- TAINS A HAZARDOUS MATERIAL ANY OF THE ABOVE THAT IS SUBMITTED TO ROSEMOUNT ANALYTICAL WITHOUT THE MSDS WILL BE RETURNED TO SENDER C.O.D. FOR THE SAFETY AND HEALTH OF OUR EMPLOYEES. WE THANK YOU IN ADVANCE FOR COMPLIANCE TO THIS SUBJECT.
S E N D E R
I C E
T O
2. PART NO.__________________________2. MODEL_________________________________2. SER. NO.________________
3. PART NO.__________________________3. MODEL_________________________________3. SER. NO.________________
4. PART NO.__________________________4. MODEL_________________________________4. SER. NO.________________
R E A S O N
PLEASE CHECK ONE:
n REPAIR AND CALIBRATE n DEMO EQUIPMENT NO
n EVALUATION n OTHER (EXPLAIN)
n REPLACEMENT REQUIRED? n YES n NO
R E T U R N
R E P A
WARRANTY REPAIR REQUESTED:
n YES-REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO
CUSTOMER PURCHASE ORDER NO. _________________________________________________
S T A T U S
n NO-PROCEED WITH REPAIRS-INVOICE AGAINST P.O. NO
n NO-CONTACT WITH ESTIMATE OF REPAIR CHARGES: LETTER n
ADDRESS ___________________________________________________________________________________________________
______________________________________________________________ ZIP _________________________________________
RETURN AUTHORITY FOR CREDIT ADJUSTMENT [Please check appropriate box(s)]
n WRONG PART RECEIVED n REPLACEMENT RECEIVED
n DUPLICATE SHIPMENT REFERENCE ROSEMOUNT ANALYTICAL SALES ORDER NO
n RETURN FOR CREDIT RETURN AUTHORIZED BY:
Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2001

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workman- ship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, pH electrodes, membranes, liquid junctions, electrolyte, O-rings, etc. are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and / or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, parts(s), or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and consumables for the remaining portion of the period of the twelve (12) month warranty in the case of goods and part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such com- mercial unit when such goods, parts(s) or consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage, directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Rosemount Analytical Inc.
Uniloc Division 2400 Barranca Parkway Irvine, CA 92606
The shipping container should be marked:
Return for Repair Model
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following there- on):
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.


The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2010
Emerson Process Management
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Reference electrolyte depleted. | Replace reference electrolyte as instructed in Section 3.2.2. | p.21 | |
| Defective preamplifier. | Check preamplifier as instructed in Section 3.4 and replace preamplifier if defective. | p.21 | |
| T.C. element shorted. | Check T.C. element as instructed in Section 3.5, and replace body if defective. | p.21 | |
| Electrode not in solution or sample stream is not full. | Make sure sensor is in solution or that sample stream is full (refer to Section 2.0 for installation details). | p.21 | |
| Open glass electrode. | Replace glass electrode. | p.21 | |
| Plugged liquid junction. | Replace liquid junction. Recharge reference cell. | p.21 | |
| Reference element open- no contact. | Replace reference element. | p.21 | |
| Electrode cracked. | Replace electrode. | p.21 | |
| T.C. element open. | Check T.C. element as instructed in Section 3.5 and replace body if defective. | p.21 | |
| Electrode coated. | Clean electrode as instructed in Section 3.2.1. | p.21 | |
| Electrode defective. | Replace electrode. | p.21 | |
| Electrode coated or cracked. | Clean electrode as instructed in Section 3.2.1 and, if cracked, replace electrode. | p.21 | |
| Reference chamber is contaminated | Replace gel solution as instructed in Section 3.2.2. | p.21 | |
| If trouble still persists, micro-junction electrode is defective. | Replace micro-junction electrode. | p.21 | |
| Old glass electrode (greater than 1 year use). | Replace electrode. | p.21 | |
| Coated glass. | Clean with soft cloth and clean water. | p.21 |