Table of Contents 67 sections Instruction Manual PN 51-381/rev.E March 2010 Model 381 · pH/ORP Sensor · ESSENTIAL INSTRUCTIONS · DANGER HAZARDOUS AREA INSTALLATION · … ShowHide
Instruction Manual PN 51-381/rev.E March 2010 Model 381p.1pH/ORP Sensorp.1ESSENTIAL INSTRUCTIONSp.2DANGER HAZARDOUS AREA INSTALLATIONp.2READ THIS PAGE BEFOREp.2PROCEEDING!p.2MODEL 381 pH/ORP SENSORp.4LIST OF FIGURESp.4About This Documentp.4Rev. Level Date Notesp.4SECTION 1.0 DESCRIPTION AND SPECIFICATIONSp.5HANDRAIL MOUNTING ACCESSORY HANDRAIL MOUNTING ACCESSORYp.6PN 1000856 PN 11275-00p.6MODEL 381 INSERTION, SUBMERSION, FLOW THROUGH SENSORp.7CODE FUNCTIONp.7CODE INSTALLATIONp.7CODE MEASURING ELECTRODEp.7CODE LIQUID JUNCTIONp.7CODE TEMPERATURE COMPENSATIONp.7CODE CABLEp.7CODE INTEGRAL PREAMPLIFIERp.7SECTION 2.0 INSTALLATIONp.9IMPORTANT!p.9PSIGp.14POSSIBLE ELECTRODE DAMAGE ABOVE THE THIS POINTp.14PRESSURE DROPp.14RECOMMENDED OPERATING RANGERp.14INEFFECTIVE CLEANING BELOW THIS POINTp.14FLOW RATEp.14WHEN INCH AND METRIC DIMSp.15ARE GIVENp.15MILLIMETERp.15INCHp.15DWG. NO. REV. 40038105 Lp.15DWG. NO. REV. 40038103 Jp.16DWG. NO. REV. 40038116 Ip.17DWG. NO. REV. 40038115 Kp.17SECTION 3.0 MAINTENANCEp.193.3 MODEL 381p.20T.C. ELEMENTp.223K PT-100p.22TABLE 3-3p.22COMPENSATION ELEMENTSp.223K 2934 .0045 PT-100 107.7 .00385p.22SECTION 4.0 ULTRASONIC CLEANERp.25DWG. NO. REV. 40038156 Fp.27DWG. NO. REV. 40037505 Dp.28MODEL 381 SECTION 5.0 RETURN OF MATERIALp.29SECTION 5.0 RETURN OF MATERIALp.29C U S T O Mp.30E Rp.30N Op.30S E N D E Rp.30I C Ep.30T Op.30R E A S O Np.30PLEASE CHECK ONE:p.30F O Rp.30DESCRIPTION OF MALFUNCTION:p.30R E T U R Np.30R E P Ap.30WARRANTY REPAIR REQUESTED:p.30I Rp.30CUSTOMER PURCHASE ORDER NO. _________________________________________________p.30S T A T U Sp.30WARRANTYp.31RETURN OF MATERIALp.31
Page 1 of the Emerson Process Management pH Sensor (381) user manual — Instruction Manual PN 51-381/rev.E March 2010 Model 381
Page 1 · click to enlarge
Page 1

Instruction Manual PN 51-381/rev.E March 2010 Model 381

pH/ORP Sensor

Emerson Process Management pH Sensor (381) — Instruction Manual PN 51-381/rev.E March 2010 Model 381 diagram (page 1)
Fig. 1 · Instruction Manual PN 51-381/rev.E March 2010 Model…
Page 2 of the Emerson Process Management pH Sensor (381) user manual — PROCEEDING!
Page 2 · click to enlarge
Page 2

ESSENTIAL INSTRUCTIONS

DANGER HAZARDOUS AREA INSTALLATION

READ THIS PAGE BEFORE

Installations near flammable liquids or in haz- ardous area locations must be carefully eval- uated by qualified on site safety personnel. This sensor is not Intrinsically Safe or Explosion Proof.

PROCEEDING!

Rosemount Analytical designs, manufactures, and tests its products to meet many national and international stan- dards. Because these instruments are sophisticated tech- nical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.

To secure and maintain an intrinsically safe installation, the certified safety barrier, transmitter, and sensor combi nation must be used. The installation system must com- ply with the governing approval agency (FM, CSA or BASEEFA/CENELEC) hazardous area classification requirements. Consult your analyzer/transmitter instruc tion manual for details.

  • Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-800-654-7768 and the requested manual will be provided. Save this Instruction Manual for future reference.

Proper installation, operation and servicing of this sensor in a Hazardous Area Instal - lation is entirely the responsibility of the user.

  • If you do not understand any of the instructions, contact your Rosemount representative for clarification.
  • Follow all warnings, cautions, and instructions marked on and supplied with the product.
  • Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
  • Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
  • To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
  • When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.
  • Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.

Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250

http://www.RAuniloc.com

© Rosemount Analytical Inc. 2001

Emerson Process Management

Page 3 of the Emerson Process Management pH Sensor (381) user manual — ESSENTIAL INSTRUCTIONS
Page 3 · click to enlarge
Page 3
Page 3 of the Emerson Process Management pH Sensor (381) user manual — ESSENTIAL INSTRUCTIONS
Page 4 of the Emerson Process Management pH Sensor (381) user manual — Rev. Level Date Notes
Page 4 · click to enlarge
Page 4

MODEL 381 pH/ORP TABLE OF CONTENTS

MODEL 381 pH/ORP SENSOR

LIST OF FIGURES

Figure No. Title Page 1-1 Dimensional Drawing 3 1-2 Mounting Drawing 3 2-1 Model 381 Sensor Component Locator Diagram 6 2-2 Submersion Installation Diagram 8 2-3 Insertion Installation Diagram 8 2-4 Flow Through Installation Diagram (Code 03 or 04) 9 2-5 Installation Details Low Flow Conditions Model 381 (Code 03-04) 9 2-6 Retro-Fit Kit and Code-04 Flow Powered Cleaner, Model 381 10 2-7 Flow Through Diagram and Wiring (Code 03 or 04) 11 2-8 Remote Junction Box with Internal Preamp 12 2-9 Wiring with Weatherproof or Div. 2 (Code 2) Junction Box 13 4-1 Model 381 Ultrasonic Cleaner and Wiring 23 4-2 Dimensional and Component Locator Ultrasonic Generator 24

About This Document

This manual contains instructions for installation and operation of the Model 381 pH/ORP Sensor.

The following list provides notes concerning all revisions of this document.

Rev. Level Date Notes

A 6/94-1/00 This is the initial release of the product manual. The manual has been reformatted to reflect the Emerson documentation style and updated to reflect any changes in the product offering.

B 5/02 Updated multiple drawings throughout manual.

D 10/07 Updated page 3 code 00 insertion is no longer available.

C 3/05 Updates

Page 5 of the Emerson Process Management pH Sensor (381) user manual — SECTION 1.0 DESCRIPTION AND SPECIFICATIONS
Page 5 · click to enlarge
Page 5

MODEL 381 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS

SECTION 1.0 DESCRIPTION AND SPECIFICATIONS

  • MOLDED POLYETHERSULFONE (PES) BODY provides compatibility with a variety of processes.
  • MODULAR DESIGN facilitates maintenance by eliminating terminal and mounting brackets and extraneous hardware.
  • INTEGRAL PREAMPLIFIER provides reliable signal transmission.
  • DOUBLE JUNCTION, GEL-FILLED, REFERENCE CELL provides high temperature performance, extended service life, and resists the effects of pumping, dilution and contamination.
  • OPTIONAL FLOW-POWERED OR ULTRASONIC CLEANING provides trouble-free service in difficult applications.

1.1 FEATURES AND APPLICATIONS The Model 381 Sensor measures the pH or the Oxidation Reduction Potential (ORP) of aqueous solu- tions in pipelines, open tanks or ponds. The sensor is used in most industrial applications including SO2 scrubbers, rinse tanks, plating baths, reverse osmosis units, and water and waste treatment plants. The sensor is housed in a molded PES body and has two O-ring seals with breach lock threads which secure the PES cover. This provides a waterproof union for long operating life and easy removal for rou- tine maintenance. The modular body design eliminates the need for internal mounting brackets, terminal brackets and screws. All components are screw-type or plug-in- place, allowing for fast simple service. The integral preamplifier amplifies the high imped- ance glass electrode signal at the sensor, providing a transmission capability of up to three miles. The Rosemount Analytical Inc. method of preamplification has become the industry standard for reliable pH measurement. In submersion applications at tempera- tures greater than 80°C, the preamplifier must be located in a remote junction box. A standard feature of the Model 381 is the double junction, gel-filled reference cell which reduces the need for special applications engineering where unknown reference cell contamination, such as sul- fides, may exist. The gel solution maintains its viscosi-

ty at high temperatures, and resists the effect of pumping and dilution, resulting in an extended service life. A choice of wood (Code 20) or ceramic (Code 21) liquid junction is available to provide resistance to fouling and subsequent loss of electrical contact between the reference electrode and process fluid. In flow applications where crystalline or viscous coat- ings, such as calcium carbonate or petroleum oils, may coat the electrode surfaces and impede the sen- sor's performance, the flow-powered cleaning option (Code 04) may be used. A specially designed flow chamber directs the process fluid in a circular path carrying three Teflon1 balls which clean the electrode surface, physically preventing accumulation of coating materials. Flow-powered cleaning is suitable for all hazardous area applications. In submersion (Code 02) applications, the ultrasonic cleaning option may be specified. This assembly includes the ultrasonic transducer and generator, mounting hardware, and 20 ft. of interconnection cable. Ultrasonic cleaning is not suitable for haz- ardous area applications. The Model 381 is compatible with all Rosemount Analytical pH/ORP analyzers. For use with the 1054/2054 Series, 2081 and other pH analyzers which utilize a PT-100 automatic temperature com- pensator order the Model 381 pHE.

1 Teflon is a registered trademark of E.I. du Pont de Nemours & Co.

Page 6 of the Emerson Process Management pH Sensor (381) user manual — PN 1000856 PN 11275-00
Page 6 · click to enlarge
Page 6

MODEL 381 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS

1.2 SENSOR SPECIFICATIONS Materials of Construction:

Emerson Process Management pH Sensor (381) — HANDRAIL MOUNTING ACCESSORY HANDRAIL MOUNTING ACCESSORY diagram (page 6)
Fig. 2
Body, Cover and Flow CellPolyethersulfone (PES). O-Rings: Viton 2. Measuring Electrode: Glass (and platinum for
Model 381 ORP). Liquid JunctionKynar 3 /wood (Code 20) or

Kynar/ceramic (Code 21). Process Connections: Submersion: 3/4 inch MNPT.

Insertion: 2 inch MNPT. Flow Through:3/4 inch MNPT. Dimensions: (see Figure 1-1). Measuring Range: pH 0-14

ORP –1500mV to +1500mV Temperature Compensation: (Model 381 pH only,

not required for Model 381 ORP). Automatic, 0 to 100°C (32 to 212°F). Pressure Rating: 790 kPa abs at 100°C

(100 psig at 212°F) CableFour-conductor, 22 GA, each (0.64 mm dia.),

with common shield. Weight/Shipping Weight: 1.0 kg/1.7 kg (2.2 lb./ 3.7 lb.).

2 Viton is a registered trademark of E.I. du Pont de Nemours & Co. 3 Kynar is a trademark of Elf Atochem North America, Inc.

FIGURE 1-1. Dimensional Drawing

Emerson Process Management pH Sensor (381) — HANDRAIL MOUNTING ACCESSORY HANDRAIL MOUNTING ACCESSORY diagram (page 6)
Fig. 3

HANDRAIL MOUNTING ACCESSORY HANDRAIL MOUNTING ACCESSORY

Emerson Process Management pH Sensor (381) — HANDRAIL MOUNTING ACCESSORY HANDRAIL MOUNTING ACCESSORY diagram (page 6)
Fig. 4

PN 1000856 PN 11275-00

FIGURE 1-2. Mounting Drawing

Page 7 of the Emerson Process Management pH Sensor (381) user manual — CODE INTEGRAL PREAMPLIFIER
Page 7 · click to enlarge
Page 7

MODEL 381 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS

1.3 ORDERING INFORMATION Model pH/ORP Sensor is housed in a PES body suitable for insertion, submer- sion or flow through installation. The sensor includes an integral preamplifier, measuring electrode, double junc- tion gel-filled reference cell, automatic temperature compensation for pH, and a choice of two cable lengths of either 15 ft. or 50 ft. (4.5 or 15.2 m).

MODEL 381 INSERTION, SUBMERSION, FLOW THROUGH SENSOR

CODE FUNCTION

pH Requires Code 30

ORP Requires Code 32 (not for use with 1054/1054A ORP Series)

pHE Requires Code 31 (for use with 1054/1054A/2054 Series and 2081)

CODE INSTALLATION

04 Flow Through with Flow Powered Cleaner (not available with ORP)

CODE MEASURING ELECTRODE

10 General purpose pH

11 High pH Electrode, requires Code 21. For sodium concentrations greater than 1% or continuous measurement above 11 pH.

12 Ruggedized pH Electrode. For abrasive solutions.

13 Platinum ORP Electrode.

14 Gold ORP Electrode.

15 HF pH Electrode. For HF concentrations less than 100 ppm.

CODE LIQUID JUNCTION

20 Wood/KYNAR Liquid Junction

21 Ceramic/KYNAR Liquid Junction

CODE TEMPERATURE COMPENSATION

30 Automatic T.C., 3.0K ohm. For Models 1003, 1050 & 1181 pH

31 Automatic T.C., PT-100. For Models 1054/2054 Series, and 2081

32 Automatic T.C., NOT REQUIRED (For Models 1023, 1060 and 1181 ORP)

CODE CABLE

40 15 ft. cable for integral sensor preamp

41 15 ft. coax cable w/BNC connector for remote sensor preamp and integral analyzer preamp (1054A-54/2054-54 only)

42 15 ft. coax cable w/o BNC connector for integral analyzer preamp (1054, 1181-43/44 & 2081-05 only)

43 50 ft. cable for integral sensor preamp

CODE INTEGRAL PREAMPLIFIER

50 Preamp for Models 1181 pH/ORP & 1050/1060

51 Preamp for Models 1003/1023

52 Preamp NOT REQUIRED

54 Preamp for Models 1054/2054 Series, and 2081

381 pH 00 10 20 30 40 51 EXAMPLE

02 Submersion · 03 Flow Through

Page 8 of the Emerson Process Management pH Sensor (381) user manual — MODEL 381 INSERTION, SUBMERSION, FLOW THROUGH SENSOR
Page 8 · click to enlarge
Page 8

4 This page left blank intentionally.

Page 9 of the Emerson Process Management pH Sensor (381) user manual — IMPORTANT!
Page 9 · click to enlarge
Page 9

SECTION 2.0 INSTALLATION

2.1 INSTALLATION. Prepare the sensor for instal lation as follows (see Figure 2-1):

IMPORTANT!

12. While holding the sensor in an up side down position (see Figure 2-1), remove the 1/4 inch plug from the electrode tip end of the sensor.
1. Remove the cover from the body by grasping the body and rotating the cover 1/4 turn counterclockwise.
13. Install the liquid junction in place of the 1/4 inch plug, using Teflon tape on the liquid junction threads.
2. When the cover breaks loose from the body, pull the cover straight out.
3. Lubricate the seals of the glass electrode with Oring lubricant, Rosemount Analytical P/N 2001928, and install the glass electrode in the body. Tighten the electrode nut by hand. DO NOT use tools.

2.2 SUBMERSION INSTALLATION (Code 02). To install the sensor in process proceed as follows (see Figure 2-2):

4. Make sure the double junction electrode is threaded tightly.
1. Wrap Teflon tape on 3/4 inch MNPT threads of cover and on standpipe threads. Seal must be water tight. Cable gland is not waterproof. Do not allow water to accumulate in the stand pipe.
5. Plug the preamplifier (or remote connector) onto the T.C. and reference electrode pins.
6. Connect the BNC connector from the glass electrode to the preamplifier.
7. Install the body O-rings. Make sure they are clean, and not twisted. Make sure the covers mating surfaces faces are clean.
2. Attach a 3/14 inch coupling to the sensor.
3. Attach electrode shroud to 2 inch MNPT.
4. Feed cable through the rigid standpipe.
5. Attach the rigid standpipe to 1 inch coupling.
8. Lubricate the body O-rings with O-ring lubricant, P/N 2001928.
9. Plug the cable connector to the preamplifier, making sure the cable is toward the center of the body.
6. Tighten all fittings and secure the standpipe to minimize sensor movement. Use flexible conduit at the top of the rigid standpipe to permit removal of the sensor for periodic maintenance.
10. Install the cover on the body so that the threads will engage.
7. Refer to Section 2.6 and wire the sensor to the transmitter as described.
11. Rotate the cover until the triangle on the body aligns with (or falls within) the adjacent mark on the cover.
Page 10 of the Emerson Process Management pH Sensor (381) user manual — SECTION 2.0 INSTALLATION
Page 10 · click to enlarge
Page 10
Emerson Process Management pH Sensor (381) — SECTION 2.0 INSTALLATION diagram (page 10)
Fig. 5

FIGURE 2-1. Model 381 Sensor Component Locator Diagram

Page 11 of the Emerson Process Management pH Sensor (381) user manual — SECTION 2.0 INSTALLATION
Page 11 · click to enlarge
Page 11

2.3 INSERTION INSTALLATION (Code 00). To install the sensor in the side of a tank, in a pipeline, or in a pipe “tee”, proceed as follows (see Figure 2-3):

1. Use Teflon tape on pipe joint, and install sensor as shown in Figure 2-3.

* In "guarded" flow position, solid particles in the flow stream are diverted from electrode.

2.4.1 Installation Procedures For Low Flow Conditions. It is extremely important that the measur- ing electrode and liquid junction of any pH/ORP sensor remain in contact with the process fluid at all times.

2. Tighten all fittings and sensor.
3. Refer to Section 2.6 for wiring instructions.

Under low flow conditions or where flow fluctuations are common, sensor contact with the process fluid may be interrupted unless piping procedures are followed to prevent it.

4. If desired, the cable may be installed in a con-duit. Flexible conduit MUST be used at the sensor for a short distance to permit removal for periodic maintenance. Use the 3/4 inch MNPT threads on the cover to attach the conduit to the sensor.

A customer supplied valve should be installed to elimi- nate head pressure when the sensor is being removed from the process. This valve may also serve as a grab sample take-off point.

2.4 FLOW THROUGH INSTALLATION (Code 03). To install the sensor in the flow cell proceed as follows (see Figure 2-4):

The installation drawing (Figure 2-5) shows the recom- mended piping procedures for the Model 381 pH/ORP sensor in low flow conditions.

1. Attach the flow cell’s 3/4 inch MNPT to the sample or process line. (Note flow direction on side of flow cell.)

2.5 FLOW POWERED CLEANER INSTALLATION (Code 04). To install the sensor in a flow cell with flow powered cleaner, proceed as follows (see Figure 2-6):

2. Holding the sensor with the electrode pointing up, seat the upper O-ring (P/N 9550146) flush with the sensor body above the 2-inch MNPT. Lubricate with O-ring lubricant (P/N 2001928).
1. Attach the flow cell’s 3/4 inch MNPT to sample or process line. Note flow direction shown on the side of the flow cell.
3. With the coupling nut thread in place, thread the electrode shroud onto the sensor body (see Figure 2-4), insuring that the O-ring is properly seated and does not become pinched or twisted.
2. Insert the ring as shown in Figure 2-6. It should snap in place.
3. Place the chamber into the flow cell. Notch must be aligned for the chamber to go in place. Place the Teflon balls (4 each) in the chamber.
4. Follow Steps 2 through 7 in Section 2.4 for sensor installation into flow assembly (see Figure 2-4).
4. Next, place the lower O-ring (P/N 9550147) in the flow cell and lubricate with O-ring lubricant (P/N 2001928).

2.6 WIRING. Connect the wire lugs to the transmitter terminals as shown in Figures 2-7 thru 2-10.

5. Place sensor, coupling nut and electrode shroud in the flow cell. Insure that the O-ring is seated properly.
6. Rotate sensor until the key on electrode shroud drops into either the “open-flow” or “guarded” flow* position.
7. Tighten the coupling nut and the flow cell fittings.
Page 12 of the Emerson Process Management pH Sensor (381) user manual — SECTION 2.0 INSTALLATION
Page 12 · click to enlarge
Page 12
Emerson Process Management pH Sensor (381) — SECTION 2.0 INSTALLATION diagram (page 12)
Fig. 6

FIGURE 2-2. Submersion Installation Diagram

Emerson Process Management pH Sensor (381) — SECTION 2.0 INSTALLATION diagram (page 12)
Fig. 7

FIGURE 2-3. Insertion Installation Diagram

Page 13 of the Emerson Process Management pH Sensor (381) user manual — SECTION 2.0 INSTALLATION
Page 13 · click to enlarge
Page 13
Emerson Process Management pH Sensor (381) — SECTION 2.0 INSTALLATION diagram (page 13)
Fig. 8

FIGURE 2-4. Flow Through Installation Diagram (Code 03 or 04)

Emerson Process Management pH Sensor (381) — SECTION 2.0 INSTALLATION diagram (page 13)
Fig. 9

FIGURE 2-5. Installation Details Low Flow Conditions Model 381 (Code 03 and 04)

Page 14 of the Emerson Process Management pH Sensor (381) user manual — FLOW RATE
Page 14 · click to enlarge
Page 14
Emerson Process Management pH Sensor (381) — PSIG diagram (page 14)
Fig. 10
Emerson Process Management pH Sensor (381) — PSIG diagram (page 14)
Fig. 11

PRESSURE DROP AND RECOMMENDED FLOW RATES- FLOW POWERED CLEANER (CODE 04)

PSIG

POSSIBLE ELECTRODE DAMAGE ABOVE THE THIS POINT

PRESSURE DROP

RECOMMENDED OPERATING RANGER

INEFFECTIVE CLEANING BELOW THIS POINT

1 2 3 4 5 6 7 8

5 10 15 20 25 30

FLOW RATE

FIGURE 2-6. Retro-Fit Kit Code 04 Flow Powered Cleaner, Model 381

kPa · gpm @ 60° · liters/min

Page 15 of the Emerson Process Management pH Sensor (381) user manual — MILLIMETER
Page 15 · click to enlarge
Page 15

WHEN INCH AND METRIC DIMS

ARE GIVEN

Emerson Process Management pH Sensor (381) — WHEN INCH AND METRIC DIMS diagram (page 15)
Fig. 12

MILLIMETER

INCH

DWG. NO. REV. 40038105 L

FIGURE 2-7. Flow Through Diagram and Wiring (Code 03 or 04)

Page 16 of the Emerson Process Management pH Sensor (381) user manual — WHEN INCH AND METRIC DIMS
Page 16 · click to enlarge
Page 16
Emerson Process Management pH Sensor (381) — DWG. NO. REV. 40038103 J diagram (page 16)
Fig. 13

WHEN INCH AND METRIC DIMS

ARE GIVEN

MILLIMETER

INCH

DWG. NO. REV. 40038103 J

FIGURE 2-8. Remote Junction Box with Internal Preamp

Page 17 of the Emerson Process Management pH Sensor (381) user manual — WHEN INCH AND METRIC DIMS
Page 17 · click to enlarge
Page 17
Emerson Process Management pH Sensor (381) — DWG. NO. REV. 40038116 I diagram (page 17)
Fig. 14

WHEN INCH AND METRIC DIMS

ARE GIVEN

MILLIMETER

INCH

DWG. NO. REV. 40038116 I

DWG. NO. REV. 40038115 K

FIGURE 2-9. Wiring with Weatherproof or Div. 2 (Code 52) Junction Box

Page 18 of the Emerson Process Management pH Sensor (381) user manual — DWG. NO. REV. 40038116 I
Page 18 · click to enlarge
Page 18

2.7 START-UP AND CALIBRATION 2.8 MODEL 381 pH and pHE 2.8.1 SENSOR PREPARATION. Shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped and read- ings with an accuracy of ± 0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be calibrated as a loop with its com- patible analyzer or transmitter.

2.9.2 Calibration

1. Make a temporary electrical connection between the sensor and the instrument.
2. Obtain a standard solution of saturated quinhydrone (PN R508-160Z). This can also be made quite simply by adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble therefore a few crystals will be required. (Refer to Section 4.3. for an alternate ORP standard solution).

2.8.2 CALIBRATION USING BUFFER SOLUTIONS OR GRAB SAMPLES. The pH sensor-analyzer/ trans- mitter loop may be calibrated by submersing the sensor in a buffer solution (standard solutions of known pH val- ues) or in a process grab sample whose pH value maybe checked by a calibrated laboratory or portable pH meter. Please refer to the analyzer's or the trans- mitter’s respective instruction manuals for proper calibration procedures.

3. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.

4. Adjust the standardize control of the instrument to the solution value shown in Table 2-1. The resulting potentials, measured with a clean platinum electrode and saturated KCl/AgCl reference electrode, should be within ±20 millivolts of the value shown in Table 2- 1. Solution temperature must be noted to insure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used.

2.9 MODEL 381 ORP. 2.9.1 MODEL 396P ORP. Most industrial applications have a number of ORP reactions occurring in sequence or simultaneously. There can be several components that are oxidized or reduced by the reagents that are used. Theoretically, the ORP poten- tial is absolute because it is the result of the oxidation- reduction equilibrium. However, the actual measured potential is dependent on many factors, including the condition of the surface of the ORP platinum elec- trode. Therefore, the sensor should be allowed 1-2 hours to become “conditioned” to the stream when first set-up or after being cleaned.

TABLE 2-1. ORP of Saturated Quinhydrone Solution

pH 4 Solution pH 7 Solution Temp °C 20 25 30 20 25 30 Millivolt Potential 268 264 260 94 87 80

5. Remove the sensor from the buffer, rinse and install in the process.

(In Millivolts)

Page 19 of the Emerson Process Management pH Sensor (381) user manual — SECTION 3.0 MAINTENANCE
Page 19 · click to enlarge
Page 19

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE

SECTION 3.0 MAINTENANCE

3.1 GENERAL. This section provides instructions for performing periodic maintenance and for determining the cause of a malfunction (troubleshooting).

7. Remove and discard the double junction electrode.
8. Remove and discard the O-rings on the body.
9. Install a new double junction electrode in the body (refer to Table 3-4 Spare Parts).

3.2 PERIODIC MAINTENANCE. The following period- ic maintenance is based on operation in a water sam- ple. Determine the schedule for the operation being maintained from past experience, or by analyzing data gathered since initial installation of the sensor.

10. Make sure the upper and lower body O-ring sealing surfaces are clean, and lubricate the new Orings with O-ring lubricant (P/N 2001928). Install the O-rings making sure they are not twisted.

3.2.1 Monthly. Perform the following maintenance monthly:

11. Install the preamplifier on the body and make sure all connections are secure and that both O-rings are in place.
1. Remove the sensor from the solution or sample and observe the condition of the electrode tip.
12. Install the cover on the body so the threads will engage.
2. If the electrode tip is coated, clean it with a soft, clean cloth or tissue moistened with a mild, nonabrasive detergent, and rinse the electrode tip with clean water. Detergents may be used to remove oil and grease coatings.
13. Rotate the cover until the triangle on the body aligns with (or falls within) the adjacent mark on the cover.
3. If Step 2 does not clean the electrode tip, remove the electrode and clean it with a dilute, 0.1 hydrochloric acid solution and rinse with clean water. HCI may be used to remove thin film and scale coatings not removed with a damp cloth or detergents.
14. While holding the sensor in an upright position, fill the reference chamber in the body (through the liquid junction opening) with gel solution (P/N 9210342). Care must be taken to remove as much air as possible from the reference cell.
4. Check calibration and install the sensor in the process as instructed in Section 2.0.

3.2.2 Annually. Perform the following maintenance annually (not required for combination electrode):

15. Wrap Teflon tape on a new liquid junction (see Table 3-4 Spare Parts) and install the liquid junction in the body.
1. Remove the sensor from the process.
16. Test the sensor per Section 2.7. Install the sensor as described in Section 2.0.
2. While holding the sensor in an upright position, remove and discard the liquid junction.
3. Drain the solution from the sensor body and rinse the inside of the body with clean, warm water.
4. Remove the cover from the body by grasping the body and rotating the cover ¼-turn in the counterclockwise direction.
5. When the cover breaks loose from the body, pull straight out on the cover and disconnect the cable connector from the preamplifier.
6. Disconnect the electrode BNC connector from the preamplifier and then unplug the preamplifier from the body.
Page 20 of the Emerson Process Management pH Sensor (381) user manual — 3.3 MODEL 381
Page 20 · click to enlarge
Page 20

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE

3.3 MODEL 381

Slowly and carefully add 56.2 milliliters of concentrated sulfuric acid. Add sufficient water to bring the total solu- tion volume up to 1000 ml. This standard ORP solution, although not as simple to prepare as the quinhydrone recipe, is much more stable, and will maintain its milli- volt value for approximately one year when stored in glass containers. This solution (ferric/ferrous ammoni- um sulfate) will produce a nominal ORP of 476 +20 mV at 25°C when used with a saturated KCl/AgCl refer- ence electrode and platinum measuring electrode. Some tolerance in mV values is to be expected due to the rather large liquid reference junction potentials which can arise when measuring this strongly acidic and concentrated solution. However, if the measuring electrodes are kept clean and in good operating condi- tion, consistently repeatable calibrations can be carried out using this standard solution.

3.3.1 Platinum Electrode Check. The platinum elec- trode may be checked as follows: There are two types of standard solutions which may be used to check the ORP electrode/transmitter system.

Type 1: One type of commonly used ORP standard solution is the saturated quinhydrone solution. A com- monly used ORP standard solution is a saturated quin- hydrone solution. This can be made by simply adding a few quinhydrone crystals to either a 4 pH or a 7 pH buffer. Quinhydrone is only sightly soluble so only a few crystals will be required. The solution will have a yellow color. The resulting potentials should be within ±20 millivolts of the value shown in Table 3-1. The ORP value of saturated quinhydrone solution is not stable for long periods of time and therefore new solutions should be made each time.

3.3.2 Cleaning Platinum Electrode. The electrode is never exposed to these undesirable compounds. In the event poisoning is suspected, the electrode can be restored to normal operation by simply cleaning the plat- inum electrode with baking soda. Polish it by rubbing it with a damp paper towel and baking soda until a bright, shiny appearance is attained.

  • Type 2: A second ORP standard solution is the FerricFerrous Ammonium Sulfate Solution (PN R508-16OZ), and it can be ordered as a spare part; otherwise, it can be prepared from the following recipe: Dissolve 39.2 grams of reagent grade ferrous ammonium sulfate, Fe(NH4)2 (SO4)2
  • 6H2O and 48.2 grams of reagent grade ferric ammonium sulfate, FeNH4(SO4)2
  • 12H2O, in approximately 700 milliliters of water (distilled water is preferred, but tap water is acceptable).

TABLE 3-1. ORP of Saturated Quinhydrone Solutions

pH 4 pH 7 Temperature °C 20 25 30 20 25 30 Millivolt Potential 268 264 260 94 87 80

Page 21 of the Emerson Process Management pH Sensor (381) user manual — 3.3 MODEL 381
Page 21 · click to enlarge
Page 21

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE

TABLE 3-1. Troubleshooting

Trouble Probable Cause Remedy

Meter reads off scale. Reference electrolyte depleted. Replace reference electrolyte as instructed in Section 3.2.2.

Defective preamplifier. Check preamplifier as instructed in Section 3.4 and replace preamplifier if defective.

T.C. element shorted. Check T.C. element as instructed in Section 3.5, and replace body if defective.

Electrode not in solution or sample Make sure sensor is in solution or that stream is not full. sample stream is full (refer to Section 2.0 for installation details).

Open glass electrode. Replace glass electrode.

Plugged liquid junction. Replace liquid junction. Recharge reference cell.

Reference element open- Replace reference element. no contact.

Meter reads between 3 and 6 pH Electrode cracked. Replace electrode. regardless of actual pH of solution or sample.

Meter indication swings widely T.C. element shorted. Check T.C. element as instructed in in AUTO T.C. Mode. Section 3.5 and replace body if defective.

Span between buffers extremely T.C. element open. Check T.C. element as instructed in short in AUTO T.C. mode. Section 3.5 and replace body if defective.

Sluggish or slow meter indication Electrode coated. Clean electrode as instructed in for real changes in pH level. Section 3.2.1.

Electrode defective. Replace electrode.

Transmitter cannot be Electrode coated or cracked. Clean electrode as instructed in standardized. Section 3.2.1 and, if cracked, replace electrode.

Reference chamber is contaminated Replace gel solution as instructed in Section 3.2.2.

Defective preamplifier. Check preamplifier as instructed in Section 3.4 and replace if defective.

If trouble still persists, micro-junction Replace micro-junction electrode. electrode is defective.

Meter short spans. Old glass electrode (greater than 1 Replace electrode. year use).

Coated glass. Clean with soft cloth and clean water.

Page 22 of the Emerson Process Management pH Sensor (381) user manual — 3K 2934 .0045 PT-100 107.7 .00385
Page 22 · click to enlarge
Page 22

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE

3.4 PREAMPLIFIER TROUBLESHOOTING. To deter- mine if the preamplifier is operable, proceed as follows:

TABLE 3-2 TEMPERATURE vs RESISTANCE OF AUTO

1. Remove the sensor from process. 2. Remove the cover from the body by grasping the body and rotating the cover 1/4 turn in the count- er-clockwise direction. 3. When the cover breaks loose from the body, pull straight out on the cover to expose the com- ponents attached to the body and disconnect the BNC connector from the preamplifier, leaving the cable connected. Inspect the internal pre amplifier area for signs of moisture or corrosion; moisture can short out the preamplifier. Dry all parts thor- oughly before testing.

T.C. ELEMENT

Temperature °C Resistance (Ohms) ± 1%

3K PT-100

0 2670 100.0 10 2802 103.8 20 2934 107.7 25 3000 109.6 30 3066 111.5 40 3198 115.4 50 3330 119.2 60 3462 123.1 70 3594 126.9 80 3726 130.8 90 3858 134.6 100 3990 138.5

4. Short the center pin of the BNC on the preamp-lifier to the outside of the BNC connector.
5. The meter indication should be approximately center scale (7 pH).
6. If the meter indication is not as specified in Step 5, proceed to the transmitter instruction manual for further troubleshooting. If the transmitter is operable with the preamplifier bypassed, then the preamplifier is defective.

3.5 T Automatic Temperature Compensator. The temper- ature compensator element is temp erature sensitive and can be checked with an ohmmeter. Resistance increases with temperature.

7. Unplug the preamplifier from the body.
8. Plug a new preamplifier into the body.

The 3K element will read 3000 ohms ± 1% at 25°C (77°F) and a Pt-100 will read 110 ohms. Resistance varies with tem- perature for a 3K and Pt-100 element and can be determined according to Table 3-2 or the following formula:

9. Plug the BNC connector of the glass electrode into the preamplifier.
10. Make sure all O-rings are clean, lubricated and in place.
11. Plug the cable connector into the preamplifier, making sure the cable is towards the center of the body.

Where RT = Resistance T = Temperature in °C

12. Install the cover on the body so that the threads engage.

Refer to Table 3-3 for Ro and R1 values

13. Rotate the cover until the triangle on the body aligns with (or falls within) the adjacent mark on the cover.

TABLE 3-3

Ro and R1 VALUES FOR TEMPERATURE

COMPENSATION ELEMENTS

Temperature Compensation Element Ro R1

14. Install the sensor as described in Section 2.0.

3K 2934 .0045 PT-100 107.7 .00385

RT=Ro [l+R1 (T-20)]

Page 23 of the Emerson Process Management pH Sensor (381) user manual — T.C. ELEMENT
Page 23 · click to enlarge
Page 23

MODEL 381pH/ORP SECTION 3.0 MAINTENANCE

TABLE 3-4. Model 381 Spare Parts

RECOMMENDED PART NUMBER DESCRIPTION CODE QUANTITY 2000734 Liquid junction wood/Kynar 20 2 2000735 Liquid junction ceramic/Kynar 21 2 2001928 O-ring lubricant ALL 2 22692-00 Body, Uniloc standard T.C. 30 22692-03 Body, Pt100 T.C. 31 22692-06 Body w/o T.C. 32 22693-00 Cover, for integral preamp, w/15 ft. cable 40 22693-01 Cover, for remote preamp, w/15 ft. coax 41 22693-02 Cover for remote preamp, w/15 ft. coax, no BNC 42 22693-03 Cover, for integral preamp, w/15 ft. cable, Model 1054/2054 Series pHE-40 22693-04 Cover, for integral preamp w/50 ft. cable 43 22693-05 Cover, for integral preamp, w/50 ft. cable, Model 1054/2054 Series pHE-43 22693-06 Cover, for remote preamp, w/15 ft. coax, w/BNC Model 1054/2054 Series pHE-41 22694-00 General purpose pH electrode 10 2 22694-01 High pH electrode 11 2 22694-02 Ruggedized pH electrode 12 2 22694-03 HF resistive pH electrode 15 2 22698-00 Preamplifier, Models 1003 pH and 1023 ORP 51 1 22698-02 Preamplifier, Models 1181 pH/ORP, 1050 pH/1060 ORP 50 1 22698-03 Preamplifier, Models 1054/2054 Series 54 1 22723-00 Kit, cleaning balls 04 1 22731-00 Platinum ORP electrode 13 2 22751-00 (Qty. 5) O-ring, Viton 2-037, upper body kit (9550141) ALL 1 set 22751-01 (Qty. 5) O-ring, Viton 2-040, lower body kit (9550128) ALL 1 set 22811-01 Retro-fit kit, flow powered cleaner 00, 02 22892-00 Retro-fit kit, flow powered cleaner 03 23018-00 Micro junction, reference element ALL 2 32602-00 Flow cell 03,04 32605-00 Electrode shroud 02, 03 32606-00 Coupling nut 03,04 32792-00 Shroud, modified 04 32793-00 Ring, chamber 04 32794-00 Chamber 04 7900058 Gold ORP electrode 14 2 9210342 KCl Gel reference solution, 10 oz./250 ml ALL 2 9550146 (Qty. 1) O-ring, shroud assembly 02, 03, 04 2 9550147 (Qty. 1) O-ring, flow cell assembly 03, 04 2

TABLE 3-5. Accessories

PART NUMBER DESCRIPTION 1000856 Handrail mounting bracket for submersion service. For use with Models 1003 & 1023. 1000857 Handrail mounting bracket for submersion service. For use with remote NEMA 4X junction box. 23296-00 Remote Division 2 junction box for internal preamp for use with Models 1181pH/ORP, 1050/1060, 1003/1023 23296-01 Remote Division 2 junction box for internal preamp for use with Models 1054/1054A/2054pH ORP/2081pH. 22996-00 Remote weatherproof junction box for internal preamp. 22719-02 Remote weatherproof junction box without preamp, for cable extension. 22727-00 Ultrasonic cleaner with generator, 115 VAC/230 VAC 50/60 Hz, switch selectable. Includes mounting hardware and 20 feet of interconnecting cable. (For submersion service). 9200160 Extension cable, sensor or remote preamp to analyzer, 4 conductor, 22 AWG, shielded (Specify length). 22811-00 Flow cell with electrode shroud. 22892-00 Flow powered cleaner retrofit kit (for use with Code 03 sensors).

Page 24 of the Emerson Process Management pH Sensor (381) user manual — T.C. ELEMENT
Page 24 · click to enlarge
Page 24

20 This page left blank intentionally.

Page 25 of the Emerson Process Management pH Sensor (381) user manual — SECTION 4.0 ULTRASONIC CLEANER
Page 25 · click to enlarge
Page 25

MODEL 381pH/ORP SECTION 4.0 ULTRASONIC CLEANER

SECTION 4.0 ULTRASONIC CLEANER

4.1 ULTRASONIC CLEANER (P/N 22727-00). This accessory provides an ultrasonic cleaner and mount- ing hardware for either continuous or inter-mittent cleaning of the measuring electrode and liquid junction on the Model 381. The information in the following paragraph is supplemental to the information in Sections 1.0 thru 3.0.

4.2 SPECIFICATIONS. The ultrasonic cleaner oper- ates on either 115 Vac or 230 Vac, 50/60 Hz. The mode is jumper selectable within the generator enclosure. The generator is shipped in the 115 Vac configuration.

The power output is approximately 1400 volts at 0.002 amp, at a frequency of 40 to 45 kHz. The transducer is supplied with 20 feet of interconnecting cable in a polypropylene protective sheath. Output connections are made at TB1-4 for R.F. (+) and TB1-2 for RF(-). Make sure the AC power line ground is properly con- nected to the generator’s input terminals.

4.3 INSTALLATION. The ultrasonic generator is shipped with special cushioning material. Before open- ing the container or removing any packing material, all items should be inspected for any possible shipping damage. If any damage is found, the carrier should be notified. Check the generator enclosure for damage to the finish; loosen the door clamps and inspect the cir- cuit board and components. If there is apparent dam- age, contact the carrier for instructions. The packing container and cushioning material should be retained should it ever become necessary to return the genera- tor to the factory. Install the ultrasonic cleaner as fol- lows:

4.4 OPERATION. Submerge the sensor into sample and start the sensor as instructed in Section 2.0. After measuring system is in operation, apply AC power to the ultrasonic cleaner. Over a period of time (usually several days), gradually increase the gain until the power setting is sufficient to keep the electrodes free of process buildup.

1. Mechanical Installation. Although the generator is designed to be located in a process environ-ment, corrosive atmospheres and excessive vibration should be avoided if possible. The ultrasonic transducer is shipped loose along with mounting hardware and mounting instructions. Check to make certain all connections are secure. Refer to Figure 4-1 and 4-2 for mounting dimensions and select a convenient location for the generator.
2. Electrical Installation. An external line switch should be installed to disconnect power from the generator for maintenance or periods of non-use. Apply input power (refer to Figure 4-2) as follows: Ground to TB1-8, hot to TB1-6, and neutral to TB1-7.
Page 26 of the Emerson Process Management pH Sensor (381) user manual — SECTION 4.0 ULTRASONIC CLEANER
Page 26 · click to enlarge
Page 26

MODEL 381pH/ORP SECTION 4.0 ULTRASONIC CLEANER

4.4 Operation (continued)

4.5 MAINTENANCE. The ultrasonic cleaner should operate continuously (with the exception of carbonate and phosphate slurries) and requires no maintenance, calibration, or adjustment, except that the power adjustment should be as low as possible consistent with good cleaning action. The ultrasonic transducer is completely sealed and should not be operating while troubleshooting the sensor or transmitter.

NOTE When the ultrasonic cleaner is used on some slurry applications (such as car- bon- ates and phosphates), continuous ultra- sonic cleaning is not recommended as this may tend to compact the solids of the slur- ry on the electrode into a hard, cement like coating. Instead, it is recommended that a shock-type treatment be applied (i.e., for every three hours of operation, employ 30 minutes of ultrasonic cleaning). The inter- val and duration of the cleaning, as well as the power setting utilized, are determined by the individual application. Factors such as the concentration of carbonates or phosphates, flow velocities, temperature, etc., must all be considered.

1. Troubleshooting. In case problems arise in the functioning of either the generator or the trans- ducer, the entire assembly should be returned to the factory for servicing. No electronic repair work of any kind should be attempted in the field, or the warranty will be voided. Whether the unit is still under warranty or not, field troubleshooting is not recommended, and Rosemount Analytical cannot be liable for operation or safety of equipment that has been repaired or modified by persons other than its factory or their trained and authorized Field Engineers. If your unit’s warranty has expired, and field troubleshooting is still desired, waveforms, frequency and voltages can be obtained directly from Branson Ultrasonic, Eagle Road, Danbury, CT 06810, (203) 796-0400. The circuit is fundamentally identical for both the old and new version.

1. Output Power Adjustment. The output power adjustment on the ultrasonic generator is located on the powerstat. The powerstat is adjusted with the knob marked as PS: (Refer to Figure 4-2), and determines the cleaning efficiency of the unit by adjusting the power input of the generator. Proper adjustment is made by turning the powerstat knob counter-clockwise (minimum power position) and gradually increasing the power, by turning the knob clockwise, until the power is sufficient to keep the electrode clean. This is usually accom- plished over a period of several days.

4.6 REPLACEMENT PARTS. The ultrasonic generator may be ordered as a unit (P/N 22990-00). The ultra- sonic transducer is a sealed unit and cannot be conve- niently repaired in the field. While it is not likely that the transducer will fail electrically or mechanically, certain process streams may contain corrosive compounds which, in time could attack and ultimately destroy the 316 Stainless Steel faceplate. The number of spare transducers (P/N 2001463) needed on hand as replacement stock, should be determined on an indi- vidual basis, with regard to specific application condi- tions.

Page 27 of the Emerson Process Management pH Sensor (381) user manual — ARE GIVEN
Page 27 · click to enlarge
Page 27

MODEL 381pH/ORP SECTION 4.0 ULTRASONIC CLEANER

WHEN INCH AND METRIC DIMS

Emerson Process Management pH Sensor (381) — WHEN INCH AND METRIC DIMS diagram (page 27)
Fig. 15

ARE GIVEN

MILLIMETER

INCH

DWG. NO. REV. 40038156 F

FIGURE 4-1. Model 381 Ultrasonic Cleaner and Wiring

Page 28 of the Emerson Process Management pH Sensor (381) user manual — MILLIMETER
Page 28 · click to enlarge
Page 28

MODEL 381pH/ORP SECTION 4.0 ULTRASONIC CLEANER

WHEN INCH AND METRIC DIMS

ARE GIVEN

Emerson Process Management pH Sensor (381) — WHEN INCH AND METRIC DIMS diagram (page 28)
Fig. 16

MILLIMETER

INCH

DWG. NO. REV. 40037505 D

FIGURE 4-2. Dimensional and Component Locator Ultrasonic Generator

Page 29 of the Emerson Process Management pH Sensor (381) user manual — MODEL 381 SECTION 5.0 RETURN OF MATERIAL
Page 29 · click to enlarge
Page 29

MODEL 381 SECTION 5.0 RETURN OF MATERIAL

SECTION 5.0 RETURN OF MATERIAL

5.1 GENERAL. To expedite the repair and return of instruments, proper communication between the cus- tomer and the factory is important. The “Return of Materials Request” form is provided for you to copy and use in case the situation arises. The accuracy and completeness of this form will affect the processing time of your materials. Call 1 (949) 757-8500 for a Return Materials Authorization (RMA) number.

5. Send the package prepaid to:

Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606

AttnFactory Repair

RMA No

Mark the packageReturned for Repair

5.2 WARRANTY REPAIR. The following is the proce- dure for returning instruments still under warranty.

Model No

5.3 NON WARRANTY REPAIR.

1. Contact the factory for authorization.
2. Complete a copy of the “Return of Materials Request” form as completely and accurately as possible.
1. Contact the factory for authorization.
2. Fill out a copy of the “Return of Materials Request” form as completely and accurately as possible.
3. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub-assemblies, the serial number on the unit must be supplied.
3. Include a purchase order number and make sure to include the name and telephone number of the right individual to be contacted should additional information be needed.
4. Carefully package the materials and enclose your “Letter of Transmittal” and the completed copy of the “Return of Materials Request” form. If possible, pack the materials in the same manner as it was received.
4. Do Steps 4 and 5 of Section 5.2.
Page 30 of the Emerson Process Management pH Sensor (381) user manual — WARRANTY DEFECT____________________________________________________________________________________
Page 30 · click to enlarge
Page 30

RETURN OF MATERIALS REQUEST •IMPORTANT! This form must be completed to ensure expedient factory service.

C U S T O M

FROM: RETURN BILL TO: _____________________________ _____________________________ _____________________________

E R

N O

CUSTOMER/USER MUST SUBMIT MATERIAL SAFETY SHEET (MSDS) OR COMPLETE STREAM COMPOSITION, AND/OR LETTER CERTIFYING THE MATERIALS HAVE BEEN DISINFECTED AND/OR DETOXIFIED WHEN RETURNING ANY PROD- UCT, SAMPLE OR MATERIAL THAT HAVE BEEN EXPOSED TO OR USED IN AN ENVIRONMENT OR PROCESS THAT CON- TAINS A HAZARDOUS MATERIAL ANY OF THE ABOVE THAT IS SUBMITTED TO ROSEMOUNT ANALYTICAL WITHOUT THE MSDS WILL BE RETURNED TO SENDER C.O.D. FOR THE SAFETY AND HEALTH OF OUR EMPLOYEES. WE THANK YOU IN ADVANCE FOR COMPLIANCE TO THIS SUBJECT.

T

S E N D E R

I C E

T O

SENSOR OR CIRCUIT BOARD ONLY(Please reference where from in MODEL / SER. NO. Column)

1. PART NO 1. MODEL 1. SER. NO

2. PART NO.__________________________2. MODEL_________________________________2. SER. NO.________________

3. PART NO.__________________________3. MODEL_________________________________3. SER. NO.________________

4. PART NO.__________________________4. MODEL_________________________________4. SER. NO.________________

R E A S O N

PLEASE CHECK ONE:

n REPAIR AND CALIBRATE n DEMO EQUIPMENT NO

n EVALUATION n OTHER (EXPLAIN)

F O R

n REPLACEMENT REQUIRED? n YES n NO

DESCRIPTION OF MALFUNCTION:

R E T U R N

R E P A

WARRANTY REPAIR REQUESTED:

n YES-REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO

I R

CUSTOMER PURCHASE ORDER NO. _________________________________________________

S T A T U S

n NO-PROCEED WITH REPAIRS-INVOICE AGAINST P.O. NO

n NO-CONTACT WITH ESTIMATE OF REPAIR CHARGES: LETTER n

NAME PHONE

ADDRESS ___________________________________________________________________________________________________

______________________________________________________________ ZIP _________________________________________

RETURN AUTHORITY FOR CREDIT ADJUSTMENT [Please check appropriate box(s)]

n WRONG PART RECEIVED n REPLACEMENT RECEIVED

n DUPLICATE SHIPMENT REFERENCE ROSEMOUNT ANALYTICAL SALES ORDER NO

n RETURN FOR CREDIT RETURN AUTHORIZED BY:

WARRANTY DEFECT

Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250

http://www.RAuniloc.com

© Rosemount Analytical Inc. 2001

Show 9 more labels / short text on this page
_____________________________ _____________________________ _____________________________ · _____________________________ _____________________________ _____________________________ · ______________________________________________________________________________________________________ · ______________________________________________________________________________________________________ · ______________________________________________________________________________________________________ · PHONE n ___________________________________________ · _____________________________________________________________________________________________________ · 24-6047 · Emerson Process Management
Page 31 of the Emerson Process Management pH Sensor (381) user manual — RETURN OF MATERIAL
Page 31 · click to enlarge
Page 31

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workman- ship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, pH electrodes, membranes, liquid junctions, electrolyte, O-rings, etc. are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and / or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, parts(s), or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and consumables for the remaining portion of the period of the twelve (12) month warranty in the case of goods and part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such com- mercial unit when such goods, parts(s) or consumables are capable of being renewed, repaired or replaced.

The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage, directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty or from any other cause. All other warranties, expressed or implied are hereby excluded.

IN CONSIDERATION OF THE STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

RETURN OF MATERIAL

Material returned for repair, whether in or out of warranty, should be shipped prepaid to:

Rosemount Analytical Inc.

Uniloc Division 2400 Barranca Parkway Irvine, CA 92606

The shipping container should be marked:

Return for Repair Model

The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following there- on):

1. Location type of service, and length of time of service of the device. 2. Description of the faulty operation of the device and the circumstances of the failure. 3. Name and telephone number of the person to contact if there are questions about the returned material. 4. Statement as to whether warranty or non-warranty service is requested. 5. Complete shipping instructions for return of the material.

Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.

If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.

Page 32 of the Emerson Process Management pH Sensor (381) user manual — WARRANTY
Page 32 · click to enlarge
Page 32
Emerson Process Management pH Sensor (381) — WARRANTY diagram (page 32)
Fig. 17

The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE

http://www.raihome.com

Specifications subject to change without notice.

Credit Cards for U.S. Purchases Only.

2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250

http://www.raihome.com

© Rosemount Analytical Inc. 2010

Emerson Process Management

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Reference electrolyte depleted.Replace reference electrolyte as instructed in Section 3.2.2. p.21
Defective preamplifier.Check preamplifier as instructed in Section 3.4 and replace preamplifier if defective. p.21
T.C. element shorted.Check T.C. element as instructed in Section 3.5, and replace body if defective. p.21
Electrode not in solution or sample stream is not full.Make sure sensor is in solution or that sample stream is full (refer to Section 2.0 for installation details). p.21
Open glass electrode.Replace glass electrode. p.21
Plugged liquid junction.Replace liquid junction. Recharge reference cell. p.21
Reference element open- no contact.Replace reference element. p.21
Electrode cracked.Replace electrode. p.21
T.C. element open.Check T.C. element as instructed in Section 3.5 and replace body if defective. p.21
Electrode coated.Clean electrode as instructed in Section 3.2.1. p.21
Electrode defective.Replace electrode. p.21
Electrode coated or cracked.Clean electrode as instructed in Section 3.2.1 and, if cracked, replace electrode. p.21
Reference chamber is contaminatedReplace gel solution as instructed in Section 3.2.2. p.21
If trouble still persists, micro-junction electrode is defective.Replace micro-junction electrode. p.21
Old glass electrode (greater than 1 year use).Replace electrode. p.21
Coated glass.Clean with soft cloth and clean water. p.21

Ask about the Emerson Process Management pH Sensor (381)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Process Management pH Sensor (381)

Be the first to ask — other owners or our team will answer.

Original document

Emerson Process Management pH Sensor (381) manual PDF

32 pages·492 KB·Stud Sensor·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…