Micro Motion ®
Model 2400S Transmitters for DeviceNet ™

©2008, Micro Motion, Inc. All rights reserved. ELITE and ProLink are registered trademarks, and MVD and MVD Direct Connect are trademarks of Micro Motion, Inc., Boulder, Colorado. Micro Motion is a registered trade name of Micro Motion, Inc., Boulder, Colorado. The Micro Motion and Emerson logos are trademarks and service marks of Emerson Electric Co. All other trademarks are property of their respective owners.


Startup Using ProLink II Transmitter User Interface Before You Begin
This chapter provides an orientation to the use of this manual, and includes a configuration overview flowchart and a pre-configuration worksheet. This manual describes the procedures required to start, configure, use, maintain, and troubleshoot the Micro Motion® Model 2400S transmitter for DeviceNet™ (the Model 2400S DN transmitter).
If you do not know what transmitter you have, see Section 1.3 for instructions on identifying the transmitter type from the model number on the transmitter’s tag.
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.
Transmitter type, user interface option, and output options are encoded in the model number located on the transmitter tag. The model number is a string of the following form:
In this string:
- C = DeviceNet
- 1 = Display with glass lens
- 3 = No display
- 4 = Display with non-glass lens

1.4 DeviceNet functionality
The Model 2400S DN transmitter implements the following DeviceNet functionality:
1.5 Determining version information
Table 1-1 lists the version information that you may need and describes how to obtain the information.
Component With ProLink II With DeviceNet tool (1)
With display
Transmitter software revision(2)
ProLink II title bar or View/Installed Options/ Software Revision
Identity Object (0x01) Instance 1 Attribute 198
Software revision corresponding to revision specified on ODVA certificate
Not available Identity Object (0x01) Instance 1 Attribute 4
Hardware revision Not available Identity Object (0x01) Instance 1 Attribute 105
(1) See Chapter 5 for more information.
(2) Also represents the core processor version.
Most of the procedures described in this manual require the use of a communication tool. The following communication tools can be used:

In this manual:
Startup Using ProLink II Transmitter User Interface Before You Begin
Refer to the configuration overview flowchart in Figure 1-1 to plan transmitter configuration. In general, perform configuration steps in the order shown here.

Figure 1-1 Configuration overview
Chapter 1 Before You Begin
Chapter 9 Pressure Compensation and Temperature Compensation Fill out pre-configuration
Chapter 8 Optional Configuration
Configure pressure compensation (optional)
Chapter 2 Flowmeter Startup
Configure temperature compensation (optional)
Chapter 10 Measurement Performance
Configure DeviceNet communications parameters
Zero the flowmeter (optional)
Chapter 6 Required Configuration
Configure status alarm severity
Configure petroleum measurement application or enhanced density application
The pre-configuration worksheet provides a place to record basic information about your flowmeter (transmitter and sensor) and your application. This information will affect your configuration options as you work through this manual. You may need to consult with transmitter installation or application process personnel to obtain the required information.
If you are configuring multiple transmitters, make copies of this worksheet and fill one out for each individual transmitter.

Pre-configuration worksheet Transmitter
Startup Using ProLink II Transmitter User Interface Before You Begin
Core processor (transmitter) software revision
Measurement units Mass flow
Installed applications Meter verification software Petroleum measurement application Enhanced density application
Table 1-2 lists documentation sources for additional information.
Topic Document
DeviceNet device profile Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile shipped with the product or available on the Micro Motion web site (www.micromotion.com)
Sensor installation Sensor documentation
Transmitter installation Micro Motion® Model 2400S Transmitters: Installation Manual
Hazardous area installation See the approval documentation shipped with the transmitter, or download the appropriate documentation from the Micro Motion web site (www.micromotion.com)

1.10 Micro Motion customer service
For customer service, phone the support center nearest you:
- In Japan, phone 3 5769-6803
- In other locations, phone +65 6777-8211 (Singapore)
- In the U.K., phone 0870 240 1978 (toll-free)
- In other locations, phone +31 (0) 318 495 555 (The Netherlands)
Customers outside the U.S.A. can also email Micro Motion customer service at International.MMISupport@EmersonProcess.com.

Startup Using ProLink II Transmitter User Interface Before You Begin
This chapter describes the following procedures:
The default node address for the Model 2400S DN transmitter is 63. The default baud rate is 125 kBaud.
If desired, you can use the hardware switches on the face of the device to change these two settings before bringing the transmitter online. See Sections 8.10.1 and 8.10.2 for more information.
The DeviceNet cable used to connect the Model 2400S DN transmitter to the network provides both power and communications. The transmitter is prewired with a male sealed Micro Connector (Eurofast).
To bring the transmitter online:
Operating the flowmeter without covers in place creates electrical hazards that can cause death, injury, or property damage.
To avoid electrical hazards, ensure that the transmitter housing cover and all other covers are in place before connecting the transmitter to the network.

When the transmitter receives power, it will automatically perform diagnostic routines, and the module LED flashes red and green. When the flowmeter has completed its power-up sequence, the status LED will show a solid green. See Section 7.4 for information on LED behavior. If the status LED exhibits different behavior, an alarm condition is present. See Section 7.5.
Note: If this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, the flowmeter is ready to receive process fluid approximately one minute after power-up. However, it may take up to ten minutes for the electronics in the flowmeter to reach thermal equilibrium. During this warm-up period, you may observe minor measurement instability or inaccuracy.

Startup Using ProLink II Transmitter User Interface Before You Begin
This chapter describes the user interface of the Model 2400S DN transmitter. The following topics are discussed:
The user interface of the Model 2400S DN transmitter depends on whether it was ordered with or without a display:
- Digital communications hardware switches, used to set the DeviceNet node address and baud rate
- Service port clips
For all other functions, either ProLink II or a customer-supplied DeviceNet tool is required.
- An LCD panel, which displays process variable data and also provides access to the off-line menu for basic configuration and management. Optical switches are provided for LCD control.
- An IrDA port which provides wireless access to the service port
Figures 3-1 and 3-2 show the user interface of the Model 2400S DN transmitter without and with a display. In both illustrations, the transmitter housing cover has been removed.
- Zero button

Figure 3-1 User interface – Transmitters without display


If the transmitter does not have a display, the transmitter housing cover must be removed to access all user interface features and functions.
If the transmitter has a display, the transmitter housing cover has a lens. All of the features shown in Figure 3-2 are visible through the lens, and the following functions may be performed through the lens (i.e., with the transmitter housing cover in place):
All other functions require removal of the transmitter housing cover.

For information on:
Startup Using ProLink II Transmitter User Interface Before You Begin
For some procedures, you must remove the transmitter housing cover. To remove the transmitter housing cover:
Removing the transmitter housing cover in a Division 2 or Zone 2 area while the transmitter is powered up can cause an explosion.
To avoid the risk of an explosion, disconnect the DeviceNet cable to remove power from the transmitter before removing the transmitter housing cover.
When replacing the transmitter housing cover, first grease the gasket, then replace the cover. Tighten the screws so that no moisture can enter the transmitter housing.
The Scroll and Select optical switches are used to navigate the display menus. To activate an optical switch, touch the lens in front of the optical switch or move your finger over the optical switch close to the lens. There are two optical switch indicators: one for each switch. When an optical switch is activated, the associated optical switch indicator is a solid red.
Attempting to activate an optical switch by inserting an object into the opening can damage the equipment.
To avoid damage to the optical switches, do not insert an object into the openings. Use your fingers to activate the optical switches.

3.5 Using the display
The display can be used to view process variable data or to access the transmitter menus for configuration or maintenance.
The display can be configured for the following languages:
Due to software and hardware restrictions, some English words and terms may appear in the non-English display menus. For a list of the codes and abbreviations used on the display, see Appendix D.
For information on configuring the display language, see Section 8.9.
In this manual, English is used as the display language.
In ordinary use, the Process variable line on the LCD panel shows the configured display variables, and the Units of measure line shows the measurement unit for that process variable.
If more than one line is required to describe the display variable, the Units of measure line alternates between the measurement unit and the additional description. For example, if the LCD panel is displaying a mass inventory value, the Units of measure line alternates between the measurement unit (for example, G) and the name of the inventory (for example, MASSI).
Auto Scroll may or may not be enabled:
For more information on using the display to view process variables or manage totalizers and inventories, see Chapter 7.

3.5.3 Using display menus
Startup Using ProLink II Transmitter User Interface Before You Begin
To enter the display menu system, see the flowchart shown in Figure 3-3.

The unlock sequence prevents unintentional entry to the offline menu. A prompt is shown for each step, and the user has 10 seconds to perform the action.
If no optical switch activity occurs for two minutes, the transmitter will exit the off-line menu system and return to the process variable display.
To move through a list of options, activate Scroll.
To select from a list or to enter a lower-level menu, Scroll to the desired option, then activate Select. If a confirmation screen is displayed:
To exit a menu without making any changes

3.5.4 Display password
Some of the display menu functions, such as accessing the off-line menu, can be protected by a display password. For information about enabling and setting the display password, refer to Section 8.9.
If a password is required, the word CODE? appears at the top of the password screen. Enter the digits of the password one at a time by using Scroll to choose a number and Select to move to the next digit.
If you encounter the display password screen but do not know the password, wait 60 seconds without activating any of the display optical switches. The password screen will time out automatically and you will be returned to the previous screen.
Certain configuration values, such as meter factors or output ranges, are entered as floating-point values. When you first enter the configuration screen, the value is displayed in decimal notation (as shown in Figure 3-4) and the active digit is flashing.
SX.XXXX
Sign For positive numbers, leave this space blank. For negative numbers, enter a minus sign (–).
Digits Enter a number (maximum length: eight digits, or seven digits and a minus sign). Maximum precision is four.
To change the value:
To change the sign of a value:
In decimal notation, you can change the position of the decimal point up to a maximum precision of four (four digits to the right of the decimal point). To do this:

To change from decimal to exponential notation (see Figure 3-5):
Startup Using ProLink II Transmitter User Interface Before You Begin
a. Select until the desired digit is flashing.
b. Scroll to the desired value. You can enter a minus sign (first position only), values between 0 and 3 (for the first position in the exponent), or values between 0 and 9 (for the second position in the exponent).
c. Select.
SX.XXXEYY
Digits Enter a four-digit number; three digits must fall to the right of the decimal point. E
Sign or Digit (0–3)
To change from exponential to decimal notation:
To exit the menu:
- Select to apply the change and exit.
- Scroll to exit without applying the change.
Sign · Digit (0–9) · Exponent indicator



Startup Using ProLink II Transmitter User Interface Before You Begin
ProLink II is a Windows-based configuration and management tool for Micro Motion transmitters. It provides access to most transmitter functions and data. Pocket ProLink is a version of ProLink II that runs on a Pocket PC.
This chapter provides basic information for connecting ProLink II or Pocket ProLink to your transmitter. The following topics and procedures are discussed:
The instructions in this manual assume that users are already familiar with ProLink II or Pocket ProLink software. For more information on using ProLink II, see the ProLink II manual. For more information on using Pocket ProLink, see the Pocket ProLink manual. Instructions in this manual will refer only to ProLink II.
To use ProLink II with the Model 2400S DN transmitter, ProLink II v2.5 or later is required. In addition, you must have either the ProLink II installation kit appropriate to your PC and connection type, or the equivalent equipment. See the ProLink II manual or quick reference guide for details.
To use Pocket ProLink, v1.3 or later is required. In addition:
ProLink II and Pocket ProLink provide a configuration upload/download function which allows you to save configuration sets to a file on the PC or Pocket PC. This allows:
Micro Motion recommends that all transmitter configurations be saved to a file as soon as the configuration is complete. See the ProLink II or Pocket ProLink manual for details.

4.4 Connecting to a Model 2400S DN transmitter
To connect to the Model 2400S DN transmitter using ProLink II or Pocket ProLink, you must use a service port connection.
The service port can be accessed via the service port clips or the IrDA port.
The service port clips have priority over the IrDA port:
Additionally, access via the IrDA port may be disabled altogether. In this case, it is not available for connections at any time. By default, access via the IrDA port is disabled. See Section 8.10.6 for more information.
The service port uses default connection parameters. Both ProLink II and Pocket ProLink automatically use these default parameters when Protocol is set to Service Port.
Additionally, to minimize configuration requirements, the service port employs an auto-detection scheme when responding to connection requests. The service port will accept all connection requests within the limits described in Table 4-1. If you are connecting to the service port from another tool, ensure that configuration parameters are set within these limits.
Parameter Option
Protocol Modbus ASCII or Modbus RTU(1)
Address Responds to both:
Standard rates between 1200 and 38,400
Stop bits 1, 2
Parity Even, odd, none
(1) Service port support for Modbus ASCII may be disabled. See Section 8.10.5.
(2) See Section 8.10.4 for information on configuring the Modbus address.
(3) This is the baud rate between the service port and the connecting program. It is not the DeviceNet baud rate.
To connect to the service port via the service port clips:
Baud rate(3)


Startup Using ProLink II Transmitter User Interface Before You Begin Startup Using ProLink II Transmitter User Interface Before You Begin Startup Using ProLink II Transmitter User Interface Before You Begin Startup Using ProLink II Transmitter User Interface Before You Begin
Removing the transmitter housing cover in a hazardous area can cause an explosion.
Because the transmitter housing cover must be removed to connect to the service port clips, the service port clips should be used only for temporary connections, e.g., for configuration or troubleshooting purposes.
When the transmitter is in an explosive atmosphere, use a different method to connect to your transmitter.

No other parameters are required.
a. Swap the leads between the two service port clips and try again.
b. Ensure that you are using the correct COM port.
c. Check all the wiring between the PC and the transmitter.
d. Verify the RS-485 to RS-232 signal converter.

4.4.4 Connecting via the IrDA port
To connect to the service port via the IrDA port:
No other parameters are required.
a. Ensure that you are using the correct port.
b. Ensure that the IrDA port is enabled.
ProLink II can be configured for the following languages:
To configure the ProLink II language, use the Tools menu. See Figure B-1.
In this manual, English is used as the ProLink II language.

Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
A customer-supplied DeviceNet tool can be used to communicate with the Model 2400S DN transmitter. This chapter provides basic information on using a customer-supplied DeviceNet tool.
However, because there are a variety of DeviceNet tools available, this chapter does not provide detailed information for using any one tool. For detailed information on your DeviceNet tool, see the documentation supplied with the tool.
To connect to the Model 2400S DN transmitter:
If required, use the digital communications hardware switches on the device to set the DeviceNet node address and baud rate for this transmitter. To do this, see Sections 8.10.1 and 8.10.2.
All DeviceNet devices employ a device profile with an object-instance-attribute structure.
In general, process and configuration data is stored in attributes, and operational functions are performed by using services or setting attributes to specific values.
Two standard services are used to read or write single attributes:
In this manual, these two services are referenced as the Get and Set services.
Other services are used to reset values to 0, start or stop calibrations, to acknowledge alarms, etc. These services are identified by name and by service code (a hexadecimal label).
Input assemblies are used to publish multiple values to the DeviceNet bus. A summary of the input assemblies is provided in Table 7-2. Output assemblies can be used to read data from the DeviceNet bus or to perform totalizer and inventory control. Summaries of the output assemblies are provided in Tables 7-9 and 9-1.

For complete documentation of the Model 2400S DN transmitter’s device profile, including input and output assemblies, see the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile.
Micro Motion supplies an Electronic Data Sheet (EDS) for the Model 2400S transmitter. The EDS file is named MMI2400S-MassFlow.eds. The EDS presents the device profile in a format designed to be read and interpreted by other devices.
DeviceNet tools fall into two basic categories:
If you are using a Type A tool:
If you are using a Type B tool, or if you want to access features that are not available through your tool’s user interface, you must reference the feature by class, instance, and attribute, use the appropriate service, and supply an attribute value if required. Depending on the attribute, the value may be a numeric or character value or a code. Values must be entered in the data type appropriate to the attribute.
For example:
a. Specify the Analog Input Point class.
b. Specify the Mass Flow instance.
c. Specify the cutoff attribute.
d. Use the Set service to set the attribute value to the desired cutoff.
- Use the Get service to read the value of the corresponding attribute.
- Use one of the input assemblies that contains the mass flow process variable.
This manual provides class, instance, attribute, data type, and service information for most configuration parameters and for all procedures. Complete documentation of the Model 2400S DN transmitter’s device profile is provided in the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile.

5.5 Default assemblies
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
The default assemblies used by the Model 2400S DN transmitter are listed and described in Table 5-1. To change the default assemblies, see the flowchart in Figure 5-1.
Connection type Assembly type Instance ID Description Size (bytes) Data type
Polled Input 6 Status Mass flow Mass total Mass inventory Temperature Density
Output 54 Reset all totalizer values
Cyclic Input 6 Status Mass flow Mass total Mass inventory Temperature Density
Figure 5-1 Changing the default DeviceNet assemblies
Class: Connection Object (0x95) Instance: 1 Attribute ID: 100 Data type: UINT Value: See Table 7-2 Service: Set
Polled connection:
Class: Connection Object (0x95) Instance: 1 Attribute ID: 101 Data type: UINT Value: See Tables 7-8 and 9-1 Service: Set
Polled connection:
Class: Connection Object (0x95) Instance: 1 Attribute ID: 102 Data type: UINT Value: See Table 7-2 Service: Set
Cyclic connection:
Input assembly · Output assembly · Input assembly



Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
This chapter describes the configuration procedures that are usually required when a transmitter is installed for the first time.
The following procedures are discussed:
This chapter provides basic flowcharts for each procedure. For more detailed flowcharts, see the flowcharts for your communication tool, provided in the appendices to this manual.
For optional transmitter configuration parameters and procedures, see Chapter 8.
Characterizing the flowmeter adjusts the transmitter to compensate for the unique traits of the sensor it is paired with. The characterization parameters, or calibration factors, describe the sensor’s sensitivity to flow, density, and temperature.
If the transmitter and sensor were ordered together, then the flowmeter has already been characterized. You need to characterize the flowmeter only if the transmitter and sensor are being paired together for the first time.
The characterization parameters that must be configured depend on your flowmeter’s sensor type: “T-Series” or “Other” (also referred to as “Straight Tube” and “Curved Tube,” respectively), as listed in Table 6-1. The “Other” category includes all Micro Motion sensors except T-Series.
The characterization parameters are provided on the sensor tag. See Figure 6-1 for illustrations of sensor tags.

Table 6-1 Sensor calibration parameters
Sensor type
Parameter
T-Series Other
K1 ✓ ✓
K2 ✓ ✓
FD ✓ ✓
D1 ✓ ✓
D2 ✓ ✓
Temp coeff (DT)(1)
FTG ✓
FFQ ✓
DTG ✓
DFQ1 ✓
DFQ2 ✓
(1) On some sensor tags, shown as TC.
(2) See the section entitled “Flow calibration values.”


Two factors are used to define flow calibration:
These values are concatenated on the sensor tag, but different labels are used for different sensors. As shown in Figure 6-1:

When configuring the flow calibration factor:
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
- Enter 19.000 for the flow calibration factor.
- Enter 5.13 for the temperature coefficient for flow.
To characterize the flowmeter:
DeviceNet tool ProLink II
Class: Sensor Information Object (0x67) Instance: 1 Attribute ID: 3 Data type: USINT Value: · 0: Curved tube · 1: Straight tube Service: Set
Device · Sensor type
Class: Calibration Object (0x65) Instance: 1 Attribute ID 1: Flow calibration factor Attribute ID 2: Temperature coefficient for flow Data type: REAL Service: Set
Class: Calibration Object (0x65) Instance: 1 Attribute ID 7: K1 Attribute ID 8: K2 Attribute ID 9: FD Attribute ID 12: D1 Attribute ID 13: D2 Attribute ID 17: DT Attribute ID 18: FTG Attribute ID 19: FFQ Attribute ID 20: DTG Attribute ID 21: DFQ1 Attribute ID 22: DFQ2 Data type: REAL Service: Set

6.3 Configuring the measurement units
For each process variable, the transmitter must be configured to use the measurement unit appropriate to your application.
To configure measurement units for process variables, see the menu flowcharts in Figure 6-3. For details on measurement units for each process variable, see Sections 6.3.1 through 6.3.4.
The measurement units used for totalizers and inventories are assigned automatically, based on the measurement unit configured for the corresponding process variable. For example, if kg/hr (kilograms per hour) is configured for mass flow, the unit used for the mass flow totalizer and mass flow inventory is kg (kilograms). DeviceNet codes used for the measurement units are listed in Tables C-12 through C-14.

Figure 6-3 Configuring measurement units
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Off-line maint > Off-line config
Class: Analog Input Point Object (0x0A) Instance: 1 Attribute ID: 102 Value: See Table 6-2 Service: Set
Class: Analog Input Point Object (0x0A) Instance: 2 Attribute ID: 102 Value: See Table 6-3 Service: Set
Class: Analog Input Point Object (0x0A) Instance: 3 Attribute ID: 102 Value: See Table 6-5 Service: Set
Class: Analog Input Point Object (0x0A) Instance: 4 Attribute ID: 102 Value: See Table 6-6 Service: Set

6.3.1 Mass flow units
The default mass flow measurement unit is g/s. See Table 6-2 for a complete list of mass flow measurement units.
Mass flow unit
Unit description Display ProLink II DeviceNet tool DeviceNet code
G/S g/s g/s 0x0800 Grams per second
G/MIN g/min g/min 0x140F Grams per minute
G/H g/hr g/hr 0x0801 Grams per hour
KG/S kg/s kg/s 0x0802 Kilograms per second
KG/MIN kg/min kg/min 0x0803 Kilograms per minute
KG/H kg/hr kg/hr 0x1410 Kilograms per hour
KG/D kg/day kg/day 0x0804 Kilograms per day
T/MIN mTon/min MetTon/min 0x0805 Metric tons per minute
T/H mTon/hr MetTon/hr 0x0806 Metric tons per hour
T/D mTon/day MetTon/day 0x0807 Metric tons per day
LB/S lbs/s lb/s 0x140B Pounds per second
LB/MIN lbs/min lb/min 0x140C Pounds per minute
LB/H lbs/hr lb/hr 0x140D Pounds per hour
LB/D lbs/day lb/day 0x0808 Pounds per day
ST/MIN sTon/min ShTon/min 0x0809 Short tons (2000 pounds) per minute
ST/H sTon/hr ShTon/hr 0x080A Short tons (2000 pounds) per hour
ST/D sTon/day ShTon/dayr 0x080B Short tons (2000 pounds) per day
LT/H lTon/hr LTon/h 0x080C Long tons (2240 pounds) per hour
LT/D lTon/day LTon/day 0x080D Long tons (2240 pounds) per day
The default volume flow measurement unit is l/s (liters per second).
Two different sets of volume flow measurement units are provided:
By default, only liquid volume flow units are listed. To access the gas standard volume flow units, you must first configure Volume Flow Type, and additional configuration is required. See Section 8.2 for more information.
Volume flow unit
Display ProLink II DeviceNet tool DeviceNet code Unit description
CUFT/S ft3/sec ft3/s 0x0814 Cubic feet per second
CUF/MN ft3/min ft3/min 0x1402 Cubic feet per minute
CUFT/H ft3/hr ft3/hr 0x0815 Cubic feet per hour

Table 6-3 Volume flow measurement units – Liquid continued
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Display ProLink II DeviceNet tool DeviceNet code Unit description
CUFT/D ft3/day ft3/day 0x0816 Cubic feet per day
M3/S m3/sec m3/s 0x1405 Cubic meters per second
M3/MIN m3/min m3/min 0x080F Cubic meters per minute
M3/H m3/hr m3/hr 0x0810 Cubic meters per hour
M3/D m3/day m3/day 0x0811 Cubic meters per day
USGPS US gal/sec gal/s 0x1408 U.S. gallons per second
USGPM US gal/min gal/min 0x1409 U.S. gallons per minute
USGPH US gal/hr gal/hr 0x140A U.S. gallons per hour
USGPD US gal/d gal/day 0x0817 U.S. gallons per day
MILG/D mil US gal/day MillionGal/dday 0x0820 Million U.S. gallons per day
L/S l/sec l/s 0x1406 Liters per second
L/MIN l/min l/min 0x0812 Liters per minute
L/H l/hr l/hr 0x0813 Liters per hour
MILL/D mil l/day MillionL/day 0x0821 Million liters per day
UKGPS Imp gal/sec ImpGal/s 0x0818 Imperial gallons per second
UKGPM Imp gal/min ImpGal/min 0x0819 Imperial gallons per minute
UKGPH Imp gal/hr ImpGal/hr 0x081A Imperial gallons per hour
UKGPD Imp gal/day ImpGal/day 0x081B Imperial gallons per day
BBL/S barrels/sec bbl/s 0x081C Barrels per second(1)
BBL/MN barrels/min bbl/min 0x081D Barrels per minute(1)
BBL/H barrels/hr bbl/hr 0x081E Barrels per hour(1)
BBL/D barrels/day bbl/day 0x081F Barrels per day(1)
BBBL/S Beer barrels/sec Beer bbl/s 0x0853 Beer barrels per second(2)
Beer bbl/min 0x0854 Beer barrels per minute(2)
BBBL/H Beer barrels/hr Beer bbl/hr 0x0855 Beer barrels per hour(2)
Beer bbl/day 0x0856 Beer barrels per day(2)
(1) Unit based on oil barrels (42 U.S. gallons). (2) Unit based on beer barrels (31 U.S. gallons).
Volume flow unit
Display ProLink II DeviceNet tool DeviceNet code Unit description
NM3/S Nm3/sec Nml m3/s 0x0835 Normal cubic meters per second
NM3/MN Nm3/min Nml m3/min 0x0836 Normal cubic meters per minute
NM3/H Nm3/hr Nml m3/hr 0x0837 Normal cubic meters per hour
NM3/D Nm3/day Nml m3/day 0x0838 Normal cubic meters per day
NLPS NLPS Nml l/s 0x083D Normal liter per second
BBBL/MN Beer barrels/min · BBBL/D Beer barrels/day

Table 6-4 Volume flow measurement units – Gas continued
Display ProLink II DeviceNet tool DeviceNet code Unit description
NLPM NLPM Nml l/min 0x1401 Normal liter per minute
NLPH NLPH Nml l/hr 0x083E Normal liter per hour
NLPD NLPD Nml l/day 0x083F Normal liter per day
SCFS SCFS Std ft3/s 0x0831 Standard cubic feet per second
SCFM SCFM Std ft3/min 0x0832 Standard cubic feet per minute
SCFH SCFH Std ft3/hr 0x0833 Standard cubic feet per hour
SCFD SCFD Std ft3/day 0x0834 Standard cubic feet per day
SM3/S Sm3/S Std m3/s 0x0839 Standard cubic meters per second
SM3/MN Sm3/min Std m3/min 0x083A Standard cubic meters per minute
SM3/H Sm3/hr Std m3/hr 0x083B Standard cubic meters per hour
SM3/D Sm3/day Std m3/day 0x083C Standard cubic meters per day
SLPS SLPS Std l/s 0x0840 Standard liter per second
SLPM SLPM Std l/min 0x0841 Standard liter per minute
SLPH SLPH Std l/hr 0x0842 Standard liter per hour
SLPD SLPD Std l/day 0x0843 Standard liter per day
The default density measurement unit is g/cm3. See Table 6-2 for a complete list of density measurement units.
Density unit
Unit description Display ProLink II DeviceNet tool DeviceNet code
SGU SGU SGU 0x0823 Specific gravity unit (not temperature corrected)
G/CM3 g/cm3 g/cm3 0x2F08 Grams per cubic centimeter
G/L g/l g/l 0x0828 Grams per liter
G/ML g/ml g/ml 0x0826 Grams per milliliter
KG/L kg/l kg/l 0x0827 Kilograms per liter
KG/M3 kg/m3 kg/m3 0x2F07 Kilograms per cubic meter
LB/GAL lbs/Usgal lb/gal 0x0824 Pounds per U.S. gallon
LB/CUF lbs/ft3 lb/ft3 0x0825 Pounds per cubic foot
LB/CUI lbs/in3 lb/in3 0x0829 Pounds per cubic inch
ST/CUY sT/yd3 ShTon/yd3 0x082A Short ton per cubic yard
D API degAPI degAPI 0x082B Degrees API

6.3.4 Temperature units
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
The default temperature measurement unit is °C. See Table 6-6 for a complete list of temperature measurement units.
Temperature unit
Unit description Display ProLink II DeviceNet tool DeviceNet code
°C °C degC 0x1200 Degrees Celsius °F °F degF 0x1201 Degrees Fahrenheit °R °R degR 0x1202 Degrees Rankine °K °K Kelvin 0x1203 Kelvin
The flowmeter does not measure pressure. You need to configure the pressure units if either of the following is true:
If you do not know whether or not you will use pressure compensation or the Gas Wizard, you do not need to configure a pressure unit at this time. You can always configure the pressure unit later.
The default pressure measurement unit is PSI. See Table 6-7 for a complete list of pressure measurement units.
Pressure unit
Unit description Display ProLink II DeviceNet tool DeviceNet code
FTH2O Ft Water @ 68°F FtH2O(68F) 0x082D Feet water @ 68 °F
INW4C In Water @ 4°C InH2O(4C) 0x0858 Inches water @ 4 °C
INW60 In Water @ 60°F InH2O(60F) 0x0859 Inches water @ 60 °F
INH2O In Water @ 68°F InH2O(68F) 0x082C Inches water @ 68 °F
mmW4C mm Water @ 4°C mmH2O(4C) 0x085A Millimeters water @ 4 °C
mmH2O mm Water @ 68°F mmH2O(68F) 0x082E Millimeters water @ 68 °F
mmHG mm Mercury @ 0°C mmHg(0C) 0x1303 Millimeters mercury @ 0 °C
INHG In Mercury @ 0°C InHg(0C) 0x1304 Inches mercury @ 0 °C
PSI PSI psi 0x1300 Pounds per square inch
BAR bar bar 0x1307 Bar
mBAR millibar mbar 0x1308 Millibar
G/SCM g/cm2 g/cm2 0x082F Grams per square centimeter
KG/SCM kg/cm2 kg/cm2 0x0830 Kilograms per square centimeter
PA pascals PA 0x1309 Pascals
KPA Kilopascals kPA 0x130A Kilopascals

Table 6-7 Pressure measurement units continued
Unit description Display ProLink II DeviceNet tool DeviceNet code
MPA megapascals MPA 0x085B Megapascals
TORR Torr @ 0C torr 0x1301 Torr @ 0 °C
ATM atms ATM 0x130B Atmospheres

Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
This chapter describes how to use the transmitter in everyday operation. The following topics and procedures are discussed:
Micro Motion suggests that you make a record of the process variables listed below, under normal operating conditions. This will help you recognize when the process variables are unusually high or low, and may help in fine-tuning transmitter configuration.
Record the following process variables:
To view these values, see Section 7.3. For information on using this information in troubleshooting, see Section 11.13.

7.3 Viewing process variables
Process variables include measurements such as mass flow rate, volume flow rate, mass total, volume total, temperature, and density.
You can view process variables with the display (if your transmitter has a display), ProLink II, or a DeviceNet tool.
By default, the display shows the mass flow rate, mass total, volume flow rate, volume total, temperature, density, and drive gain. If desired, you can configure the display to show other process variables. See Section 8.9.5.
The LCD panel reports the abbreviated name of the process variable (e.g., DENS for density), the current value of that process variable, and the associated unit of measure (e.g., G/CM3). See Appendix D for information on the codes and abbreviations used for display variables.
To view a process variable with the display:
- Appears on the process variable line, or
- Begins to alternate with the units of measure
See Figure 3-2.
The display precision can be configured separately for each process variable (see Section 8.9.5). This affects only the value shown on the display, and does not affect the actual value as reported by the transmitter via digital communications.
Process variable values are displayed using either standard decimal notation or exponential notation:
- If the value is less than the precision configured for that process variable, the value is displayed as 0 (i.e., there is no exponential notation for fractional numbers).
- If the value is too large to be displayed with the configured precision, the displayed precision is reduced (i.e., the decimal point is shifted to the right) as required so that the value can be displayed.
The Process Variables window opens automatically when you first connect to the transmitter. This window displays current values for the standard process variables (mass, volume, density, temperature, external pressure, and external temperature).
To view the standard process variables with ProLink II, if you have closed the Process Variables window, click ProLink > Process Variables.
To view API process variables (if the petroleum measurement application is enabled), click ProLink > API Process Variables.

To view enhanced density process variables (if the enhanced density application is enabled), click ProLink > ED Process Variables. Different enhanced density process variables are displayed, depending on the configuration of the enhanced density application.
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
There are two methods that can be used to view process variables with a DeviceNet tool:
Attribute ID
Data type Description
Class Instance
Analog Input Point Object (0x04)
1 (mass) 3 REAL Mass flow rate
100 REAL Mass total
101 REAL Mass inventory
102 UINT Mass flow measurement unit
103 UINT Mass total and mass inventory measurement unit
2 (liquid volume) 3 REAL Liquid volume flow rate
100 REAL Liquid volume total
101 REAL Liquid volume inventory
102 UINT Liquid volume flow measurement unit
103 UINT Liquid volume total and liquid volume inventory measurement unit
3 (density) 3 REAL Density
102 UINT Density measurement unit
4 (temperature) 3 REAL Temperature
102 UINT Temperature measurement unit
Gas Standard Volume Object (0x64)
1 (gas standard volume)
1 REAL Gas standard volume flow rate
2 REAL Gas standard volume total
3 REAL Gas standard volume inventory
5 REAL Gas standard volume flow measurement unit
6 REAL Gas standard volume total and gas standard volume inventory measurement unit

Table 7-1 Process data in DeviceNet objects continued
Data type Description
Class Instance
API Object (0x69)(1) 1 1 REAL Temperature-corrected density
2 REAL Temperature-corrected (standard) volume flow
3 REAL Temperature-corrected (standard) volume total
4 REAL Temperature-corrected (standard) volume inventory
5 REAL Batch weighted average density
6 REAL Batch weighted average temperature
Enhanced Density Object (0x6A)(2) 1 1 REAL Density at reference temperature
2 REAL Density (fixed SG units)
3 REAL Standard volume flow rate
4 REAL Standard volume flow total
5 REAL Standard volume flow inventory
6 REAL Net mass flow rate
7 REAL Net mass flow total
8 REAL Net mass flow inventory
9 REAL Net volume flow rate
10 REAL Net volume flow total
11 REAL Net volume flow inventory
13 REAL Density (fixed Baume units)
(1) Requires petroleum measurement application. See Section 8.13 (2) Requires enhanced density application. See Section 8.14.
Instance ID Data description Size (bytes) Data type Description
Mass flow and total
Volume flow and total
12 REAL Concentration · Basic process variables

Table 7-2 Summary of input assemblies continued
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Mass flow, mass totals, and other process variables
Volume flow, volume totals, and other process variables
Gas standard volume flow
Gas standard volume flow
Gas standard volume flow
Gas standard volume flow
Petroleum measurement application · Petroleum measurement application · Petroleum measurement application

Table 7-2 Summary of input assemblies continued

Table 7-2 Summary of input assemblies continued
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
(1) Available only if Gas Standard Volume is not enabled. (2) Available only if Gas Standard Volume is enabled. (3) Requires the petroleum measurement application. (4) Requires the enhanced density application. (5) Default variables are mass flow, temperature, density, volume flow, and drive gain, respectively. See Section 8.10.3 for information
on specifying the variables.
The user interface module provides three LEDs: a status LED, a module LED, and a network LED (see Figures 3-1 and 3-2).
For more information:
The module LED indicates whether or not the transmitter has power and is operating properly. Table 7-3 lists the different states of the module LED, defines each state, and provides recommendations for correcting problem states.
Enhanced density application · Enhanced density application · Configurable assembly

Table 7-3 Module LED states, definitions, and recommendations
Module LED state Definition Recommendations
Off No power Check the connection to the DeviceNet network.
Solid green No processor faults No action required.
Flashing green Needs DeviceNet configuration; may be in Standby state
Indicates an A006 alarm. Characterization parameters are missing. See Section 6.2.
Solid red Non-recoverable fault Power cycle the transmitter. If condition does not clear, call Micro Motion customer service.
Flashing red Recoverable fault Check for any status alarms.
Flashing red/green Device in self-test Wait until self-test is complete. Check the Identity Object (0x01) for device states.
The behavior of the network LED is standard, and is defined by the DeviceNet protocol. Table 7-4 lists the different states of the network LED and defines each state.
Network LED state Definition Recommendations
Off Device not online The device is not connected to the network.(1) Check the wiring if this LED is lit.
Solid green Device online and connected No action required.
Flashing green Device online but not connected The device is connected to the network, but has not been allocated by a host. No action required.
Solid red Critical link failure The most common cause is duplicate MAC IDs (node addresses) on the network. Check for duplicate MAC IDs. Other causes include incorrect baud rate setting or other network failure.
Flashing red Connection timeout Power cycle the device, or release and re-allocate the device from the DeviceNet master. If desired, increase the timeout value (Expected Packet Rate) in the DeviceNet Object (0x03).
Flashing red/green Communication faulted state Not implemented in the Model 2400S DN transmitter.
(1) If the transmitter is the only device on the network, and there is no host on the network, this is the expected LED state, and no action
is required.
You can view transmitter status using the status LED, ProLink II, or a DeviceNet tool. Depending on the method chosen, different information is displayed.
The status LED shows transmitter status as described in Table 7-5. Note that the status LED does not report event status or alarm status for alarms with severity level set to Ignore (see Section 8.8).

Table 7-5 Transmitter status LED
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Green No alarm Normal operating mode
Flashing yellow A104 alarm Zero or calibration in progress
ProLink II provides a Status window that displays:
Status information is located in the Diagnostics Object (0x66), Instance 1. This Object includes, among other data:
Use the Get service to read the required data. See Table C-7, or see the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile for detailed information.
Specific process or flowmeter conditions cause status alarms. Each status alarm has an alarm code.
Status alarms are classified into three severity levels: Fault, Information, and Ignore. Severity level controls how the transmitter responds to the alarm condition.
The transmitter maintains two status flags for each alarm:

In addition, the transmitter maintains alarm history for the 50 most recent alarm occurrences. Alarm history includes:
When the transmitter detects an alarm condition, it checks the severity level of the specific alarm and performs the actions described in Table 7-6.
Transmitter response
Alarm severity level (1)
Digital communications fault action
Status flags Alarm history
“Alarm active” record written to alarm history immediately
Activated after configured fault timeout has expired (if applicable)(2)
“Alarm active” record written to alarm history immediately
No action Not activated
(1) See Section 8.8 for information on setting the alarm severity level.
(2) See Sections 8.10.7 and 8.10.8 for more information on digital communications fault action and fault timeout.
When the transmitter detects that the alarm condition has cleared:
Operator action is required to return the second status flag to “acknowledged.” Alarm acknowledgment is not necessary. If the alarm is acknowledged, the “alarm acknowledged” record is written to alarm history.
The display shows information only about active Fault or Informational alarms, based on alarm status bits. Ignore alarms are filtered out, and you cannot access alarm history via the display.
To view or acknowledge alarms using the display menus, see the flowchart in Figure 7-1.
If the transmitter does not have a display, or if operator access to the alarm menu is disabled (see Section 8.9.3), alarms can be viewed and acknowledged using ProLink II or a DeviceNet tool. Alarm acknowledgment is not required.
Additionally, the display may be configured to enable or disable the Ack All function. If disabled, the Ack All screen is not displayed and alarms must be acknowledged individually.
Not activated

Figure 7-1 Viewing and acknowledging alarms with the display
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Scroll and Select simultaneously
ACK ALL (1)
(1) This screen is displayed only if the ACK ALL
function is enabled (see Section 8.9.3) and there are unacknowledged alarms.
NO ALARM
ACK
EXIT
7.6.2 Using ProLink II
ProLink II provides two ways to view alarm information:
The Status window displays the current status of the alarms considered to be most useful for information, service, or troubleshooting, including Ignore alarms. The Status window reads alarm status bits, and does not access alarm history. The Status window does not display acknowledgment information, and you cannot acknowledge alarms from the Status window.

In the Status window:
To use the Status window:
The Alarm Log window selects information from alarm history, and lists all alarms of the following types:
Ignore alarms are never listed.
You can acknowledge alarms from the Alarm Log window.
In the Alarm Log window:
To use the Alarm Log window:
- If the alarm was inactive, it will be removed from the list.
- If the alarm was active, it will be removed from the list as soon as the alarm condition clears.
Using the Diagnostics Object (0x66), you can view the status of a group of preselected alarms, view information about a specific alarm, acknowledge an alarm, and retrieve information from alarm history. For detailed information on the Diagnostics Object, see Table C-7, or see the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile.
To view the status of a group of preselected alarms, execute a Get for Attributes 12–17, 40, or 41.

To view information about a single alarm:
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
To acknowledge an alarm:
To retrieve information from alarm history:
- 1 = Alarm posted
- 2 = Alarm cleared
The totalizers keep track of the total amount of mass or volume measured by the transmitter over a period of time.
The inventories track the same values as the totalizers. Whenever totalizers are started or stopped, all inventories (including the API volume inventory and enhanced density inventories) are started or stopped automatically. However, when totalizers are reset, inventories are not reset automatically – you must reset inventories separately. This allows you to use the inventories to keep running totals across multiple totalizer resets.
You can view all totalizer and inventory values using any of the commmunication tools: the display, ProLink II, or a DeviceNet tool. Specific starting, stopping, and resetting functionality depends on the tool you are using.

7.7.1 Viewing current values for totalizers and inventories
You can view current values for the totalizers and inventories with the display (if your transmitter has a display), ProLink II, or a DeviceNet tool.
You cannot view current totalizer or inventory values with the display unless the display has been configured to show them. See Section 8.9.5.
To view a totalizer or inventory value, refer to Figure 7-2 and:
- MASSI (for Mass Inventory)
- LVOLI (for Liquid Volume Inventory)
- GSV I (for Gas Standard Volume Inventory)
- TCORI (for API Temperature Corrected Inventory)
- STDVI (for ED Standard Volume Inventory)
- NETVI (for ED Net Volume Inventory)
- STDMI (for ED Net Mass Inventory)


With ProLink II
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
To view current totals for the totalizers and inventories with ProLink II:
To view current totals for the totalizers and inventories with a DeviceNet tool, see Section 7.3.3.
Specific starting, stopping, and resetting functionality depends on the tool you are using.
If the required value is shown on the display, you can use the display to start and stop all totalizers and inventories simultaneously, or to reset individual totalizers. See the flowchart in Figure 7-3. You cannot reset any inventories with the display.

(1) Displayed only if configured as a display variable. (2) The petroleum measurement application or enhanced density application must be enabled. (3) The Event Setpoint screens can be used to define or change Setpoint A for Event 1 or Event 2 only. These screens are displayed
only for specific types of events. To change the setpoint for an event defined on mass total, you must enter the totalizer management menu from the mass total screen. To change the setpoint for an event defined on volume total, you must enter the totalizer management menu from the volume total screen. See Section 8.6.3 for more information. (4) The display must be configured to allow stopping and starting. See Section 8.9.3. (5) All totalizers and inventories will be stopped and started together, including API and enhanced density totalizers and inventories. (6) The display must be configured to allow totalizer resetting. See Section 8.9.3. (7) Only the totalizer currently shown on the display will be reset. No other totalizers will be reset, and no inventories will be reset.
Be sure that the totalizer you want to reset is displayed before performing this reset.

With ProLink II
The totalizer and inventory control functions available with ProLink II are listed in Table 7-7. Note the following:
a. Click View > Preferences.
b. Check the Enable Inventory Totals Reset checkbox.
c. Click Apply.
Inventory reset
Object Function Disabled Enabled
Starting and stopping as a group ✓ ✓
Totalizers Resetting all ✓ ✓
Resetting mass totalizer separately ✓ ✓
Resetting volume totalizer separately ✓ ✓
Resetting enhanced density totalizers separately ✓ ✓
Resetting API volume totalizer separately Not supported Not supported
Inventories Resetting all ✓
Resetting mass inventory separately ✓
Resetting volume inventory separately ✓
Resetting enhanced density inventories separately ✓
Resetting API volume inventory separately Not supported Not supported
To start or stop all totalizers and inventories:
To reset all totalizers:
To reset all inventories:
Totalizers and inventories

To reset an individual totalizer or inventory:
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Using a DeviceNet tool, three methods are available for totalizer and inventory control:
- Reset mass totalizer
- Reset mass inventory
- Reset liquid volume totalizer
- Reset liquid volume inventory
- Reset API reference volume total
- Reset API reference volume inventory
- Reset gas standard volume totalizer
- Reset gas standard volume inventory
- Reset ED standard volume total
- Reset ED net mass total
- Reset ED net volume total
- Reset ED standard volume inventory
- Reset ED net mass inventory
- Reset ED net volume inventory
Table 7-8 Totalizer and inventory control with a DeviceNet tool using explicit write
Stop all totalizers and inventories Analog Input Point Object (0x0A) Instance: 0 Attribute ID: 100 Service: Set Value: 0
Start all totalizers and inventories Analog Input Point Object (0x0A) Instance: 0 Attribute ID: 100 Service: Set Value: 1
Reset all totalizers Analog Input Point Object (0x0A) Instance: 0 Attribute ID: 101 Service: Set Value: 1

Table 7-8 Totalizer and inventory control with a DeviceNet tool using explicit write continued
Reset all inventories Analog Input Point Object (0x0A) Instance: 0 Attribute ID: 102 Service: Set Value: 1
Reset mass totalizer Analog Input Point Object (0x0A) Instance: 1 Service: Reset Total (0x32)
Reset mass inventory Analog Input Point Object (0x0A) Instance: 1 Service: Reset Inventory (0x33)
Reset liquid volume totalizer Analog Input Point Object (0x0A) Instance: 2 Service: Reset Total (0x32)
Reset liquid volume inventory Analog Input Point Object (0x0A) Instance: 2 Service: Reset Inventory (0x33)
Reset gas standard volume totalizer Gas Standard Volume Object (0x64) Instance: 1 Service: Reset Total (0x4B)
Reset gas standard volume inventory Gas Standard Volume Object (0x64) Instance: 1 Service: Reset Inventory (0x4C)
Reset API reference volume total API Object (0x69) Instance: 1 Service: Reset Total (0x4B)
Reset API reference volume inventory API Object (0x69) Instance: 1 Service: Reset Inventory (0x4C)
Reset ED standard volume total Enhanced Density Object (0x6A) Instance: 1 Service: Reset Total (0x4B)
Reset ED net mass total Enhanced Density Object (0x6A) Instance: 1 Service: Reset Total (0x4C)
Reset ED net volume total Enhanced Density Object (0x6A) Instance: 1 Service: Reset Total (0x4D)
Reset ED standard volume inventory Enhanced Density Object (0x6A) Instance: 1 Service: Reset Inventory (0x4F)
Reset ED net mass inventory Enhanced Density Object (0x6A) Instance: 1 Service: Reset Inventory (0x50)
Reset ED net volume inventory Enhanced Density Object (0x6A) Instance: 1 Service: Reset Inventory (0x51)

Table 7-9 Output assemblies used for totalizer and inventory control
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool



Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
This chapter describes transmitter configuration parameters that may or may not be used, depending on your application requirements. For required transmitter configuration, see Chapter 6.
Table 8-1 lists the parameters that are discussed in this chapter. Default values and ranges for the most commonly used parameters are provided in Appendix A.
Tool
Topic Subtopic ProLink II DeviceNet tool Display Section
Volume flow measurement for gas
✓ ✓ 8.2
Cutoffs ✓ ✓ 8.3
Damping ✓ ✓ 8.4
Flow direction ✓ ✓ 8.5
Events ✓ ✓ 8.6
Slug flow ✓ ✓ 8.7
Status alarm severity ✓ ✓ 8.8

Table 8-1 Configuration map continued
Topic Subtopic ProLink II DeviceNet tool Display Section
Display(1) Update period ✓ ✓ ✓ 8.9.1
Display language ✓ ✓ ✓ 8.9.2
Totalizer start/stop ✓ ✓ ✓ 8.9.3
Totalizer reset ✓ ✓ ✓
Auto scroll ✓ ✓ ✓
Scroll rate ✓ ✓ ✓
Offline menu ✓ ✓ ✓
Password ✓ ✓ ✓
Alarm menu ✓ ✓ ✓
Ack all ✓ ✓ ✓
Backlight on/off ✓ ✓ ✓ 8.9.4
Backlight intensity ✓ ✓
Display variables ✓ ✓ 8.9.5
Display precision ✓ ✓
DeviceNet node address ✓ (2) 8.10.1
DeviceNet baud rate ✓ (2) 8.10.2
DeviceNet configurable input assembly
✓ 8.10.3
Modbus address ✓ ✓ ✓ 8.10.4
Modbus ASCII support ✓ ✓ ✓ 8.10.5
IrDA port usage ✓ ✓ ✓ 8.10.6
Digital communications fault action
✓ ✓ 8.10.78.10. 7
Fault timeout ✓ ✓ 8.10.8
Device settings ✓ ✓ 8.11
Sensor parameters ✓ ✓ 8.12
✓ ✓ 8.13
✓ ✓ 8.14
(1) These parameters apply only to transmitters with a display. (2) Cannot be set with the display menus, but can be set with digital communications hardware switches on the face of the transmitter.
Two types of volume flow measurement are available:
Digital communication · Petroleum measurement application · Enhanced density application

Only one type of volume flow measurement can be performed at a time (i.e., if liquid volume flow measurement is enabled, gas standard volume flow measurement is disabled, and vice versa). Different sets of volume flow measurement units are available, depending on which type of volume flow measurement is enabled (see Tables 6-3 and 6-4). If you will use a gas standard volume flow unit, additional configuration is required.
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
The method used to configure volume flow measurement for gas depends on the tool you are using: ProLink II or a DeviceNet tool.
To configure volume flow measurement for gas using ProLink II:
The Gas Wizard is used to calculate the standard density of the gas that you are measuring.
To use the Gas Wizard:
a. Enable the Choose Gas radio button.
b. Select your gas.
a. Enable the Enter Other Gas Property radio button.
b. Enable the method that you will use to describe its properties: Molecular Weight, Specific Gravity Compared to Air, or Density.
c. Provide the required information. Note that if you selected Density, you must enter the value in the configured density units and you must provide the temperature and pressure at which the density value was determined.

The Gas Standard Volume Object is used to configure volume flow measurement for gas. See the flowchart in Figure 8-1.
Class: Gas Standard Volume Object (0x64) Instance: 1 Attribute ID: 5 Data type: UINT Value: See Table 6-4 Service: Set
Class: Gas Standard Volume Object (0x64) Instance: 1 Attribute ID: 8 Data type: REAL Service: Set
(1) See Section 8.3. (2) The Gas Wizard is provided only
with ProLink II. If you are not using ProLink II, you must supply the required reference density.
Class: Gas Standard Volume Object (0x64) Instance: 1 Attribute ID: 4 Data type: REAL Service: Set
Cutoffs are user-defined values below which the transmitter reports a value of zero for the specified process variable. Cutoffs can be set for mass flow, liquid volume flow, gas standard volume flow, and density.
See Table 8-2 for cutoff default values and related information. See Section 8.3.1 for information on how the cutoffs interact with other transmitter measurements.

Table 8-2 Cutoff default values
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Mass flow 0.0 g/s Recommended setting: 5% of the sensor’s rated maximum flowrate
Liquid volume flow 0.0 L/s Limit: the sensor’s flow calibration factor in liters per second, multiplied by 0.2
Gas standard volume flow
0.0 No limit
To configure cutoffs:
If you are using liquid volume flow measurement:
If you are using gas standard volume flow measurement, neither the mass flow cutoff nor the density cutoff is applied to the volume flow calculation.
A damping value is a period of time, in seconds, over which the process variable value will change to reflect 63% of the change in the actual process. Damping helps the transmitter smooth out small, rapid measurement fluctuations.
Damping can be configured for flow, density, and temperature.
When you change the damping value, the specified value is automatically rounded down to the nearest valid damping value. Valid damping values are listed in Table 8-3.
Before setting the damping values, review Section 8.4.1 for information on how the damping values affect other transmitter measurements.

Table 8-3 Valid damping values
Process variable Valid damping values
Flow (mass and volume) 0, 0.04, 0.08, 0.16, ... 40.96
Density 0, 0.04, 0.08, 0.16, ... 40.96
Temperature 0, 0.6, 1.2, 2.4, 4.8, ... 76.8
To configure damping values:
When configuring damping values, note the following:
Be sure to set damping values accordingly.
The flow direction parameter controls how the transmitter reports flow rate and how flow is added to or subtracted from the totalizers, under conditions of forward flow, reverse flow, or zero flow.
Options for flow direction include:
For the effect of flow direction on flow totals and flow values, see Table 8-4.

Table 8-4 Effect of flow direction on totalizers and flow values
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Flow direction value
Flow totals Flow values
Forward only Increase Positive
Reverse only No change Positive
Absolute value Increase Positive(2)
Negate/Forward only No change Negative
Flow direction value
Flow totals Flow values
Forward only No change Negative
Reverse only Increase Negative
Absolute value Increase Positive(2)
Negate/Forward only Increase Positive
(1) Process fluid flowing in same direction as flow direction arrow on sensor.
(2) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
(3) Process fluid flowing in opposite direction from flow direction arrow on sensor.
To configure flow direction:
An event occurs if the real-time value of a user-specified process variable varies above or below a user-specified value, or inside or outside a user-specified range. You can configure up to five events.
You may optionally specify one or more actions that will occur if the event occurs. For example, if Event 1 occurs, you may specify that the transmitter will stop all totalizers and inventories and reset the mass totalizer.
To define an event:
The following general steps are required:

DeviceNet code Description
Type
High (> A) 0 Default. Discrete event will occur if the assigned variable is greater than the setpoint (A).(1)
Low (< A) 1 Discrete event will occur if the assigned variable is less than the setpoint (A).(1)
In Range 2 Discrete event will occur if the assigned variable is greater than or equal to the low setpoint (A) and less than or equal to the high setpoint (B).(2)
Out of Range 3 Discrete event will occur if the assigned variable is less than or equal to the low setpoint (A) or greater than or equal to the high setpoint (B).(2)
(1) An event does not occur if the assigned variable is equal to the setpoint.
(2) An event occurs if the assigned variable is equal to the setpoint.
DeviceNet code Description
ProLink II label Display label
Start sensor zero START ZERO 1 Initiates a zero calibration procedure
Reset mass total RESET MASS 2 Resets the value of the mass totalizer to 0
Reset volume total RESET VOL 3 Resets the value of the liquid volume totalizer to 0(1)
Reset gas std volume total RESET GSV 21 Resets the value of the gas standard volume totalizer to 0(2)
Reset API ref vol total RESET TCORR 4 Resets the value of the API temperature-corrected volume totalizer to 0(3)
Reset ED ref vol total RESET STD V 5 Resets the value of the ED standard volume totalizer to 0(4)
Reset ED net mass total RESET NET M 6 Resets the value of the ED net mass totalizer to 0(4)
Reset ED net vol total RESET NET V 7 Resets the value of the ED net volume totalizer to 0(4)

Table 8-6 Event actions continued
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
ProLink II label Display label
Reset all totals RESET ALL 8 Resets the value of all totalizers to 0
Start/stop all totalization START STOP 9 If totalizers are running, stops all totalizers If totalizers are not running, starts all totalizers
Increment current ED curve
INCR CURVE 18 Changes the active enhanced density curve from 1 to 2, from 2 to 3, etc.(4)
(1) Displayed only if Volume Flow Type = Liquid. (2) Displayed only if Volume Flow Type = Gas. (3) Available only if the petroleum measurement application is installed. (4) Available only if the enhanced density application is installed.
Example Define Event 1 to be active when the mass flow rate in forward or backward direction is less than 2 lb/min. Additionally, if this occurs, all totalizers should be stopped.
Using ProLink II:
Using a DeviceNet tool:

8.6.2 Checking and reporting event status
There are several ways that event status can be determined:
For Event 1 or Event 2 only, the value of Setpoint A can be changed from the display, under the following circumstances:
Then, to reset Setpoint A from the display:
Slugs – gas in a liquid process or liquid in a gas process – occasionally appear in some applications. The presence of slugs can significantly affect the process density reading. The slug flow parameters can help the transmitter suppress extreme changes in process variables, and can also be used to identify process conditions that require correction.
Slug flow parameters are as follows:

If the transmitter detects slug flow:
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
To configure slug flow parameters:
The Model 2400S transmitter can report faults in the following ways:
Status alarm severity determines which methods the transmitter will use when a specific alarm condition occurs, as described in Table 8-7. (See Section 7.6 for a more detailed discussion.)
Transmitter action if condition occurs
“Alarm active” status bit set?
“Alarm active” record written to history?
Digital communications fault action activated? (1)
Severity level
Fault Yes Yes Yes
Informational Yes Yes No
Ignore Yes No No
(1) For some alarms, the digital communications fault action will not begin until the fault timeout has expired. To configure fault timeout,
see Section 8.10.8. Other fault reporting methods occur as soon as the fault condition is recognized. Table 8-8 includes information on which alarms are affected by the fault timeout

Some alarms can be reclassified. For example:
For a list of all status alarms and default severity levels, see Table 8-8. (For more information on status alarms, including possible causes and troubleshooting suggestions, see Table 11-2.)
To configure alarm severity:
a. Set the alarm index (Attribute 18).
b. Set the severity for that alarm (Attribute 19).
Default severity Configurable?
Affected by fault timeout?
Alarm code ProLink II message
A001 (E)EPROM Checksum Error (CP) Fault No No
A002 RAM Error (CP) Fault No No
A003 Sensor Failure Fault Yes Yes
A004 Temperature Sensor Failure Fault No Yes
A005 Input Overrange Fault Yes Yes
A006 Not Configured Fault Yes No
A008 Density Overrange Fault Yes Yes
A009 Transmitter Initializing/Warming Up Ignore Yes No
A010 Calibration Failure Fault No No
A011 Zero Too Low Fault Yes No
A012 Zero Too High Fault Yes No
A013 Zero Too Noisy Fault Yes No
A014 Transmitter Failed Fault No No
A016 Line RTD Temperature Out-of-Range Fault Yes Yes
A017 Meter RTD Temperature Out-of-Range Fault Yes Yes
A020 Calibration Factors Unentered (FlowCal) Fault Yes No
A021 Incorrect Sensor Type (K1) Fault No No
A029 PIC/Daughterboard Communication Failure Fault No No
A030 Incorrect Board Type Fault No No
A031 Low Power Fault No No
A032 Meter Verification/Outputs In Fault Fault No No
A033 Sensor OK, Tubes Stopped by Process Fault Yes Yes
A102 Drive Overrange/Partially Full Tube Info Yes No
A104 Calibration in Progress Info Yes(1) No
A105 Slug Flow Info Yes No
A107 Power Reset Occurred Info Yes No

Table 8-8 Status alarms and severity levels continued
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Affected by fault timeout?
Alarm code ProLink II message
A131 Meter Verification/Outputs at Last Value Info Yes No
A132 Simulation Mode Active Info Yes No
A133 PIC UI EEPROM Error Info Yes No
(1) Can be set to either Informational or Ignore, but cannot be set to Fault.
If your transmitter has a display, you can configure a variety of parameters that control the display functionality.
The Update Period (or Display Rate) parameter controls how often the display is refreshed with current data. The default is 200 milliseconds; the range is 100 milliseconds to 10,000 milliseconds (10 seconds).
To configure Update Period:
The display can be configured to use any of the following languages for data and menus:
To set the display language:
Table 8-9 lists the display functions and describes their behavior when enabled (shown) or disabled (hidden).

Table 8-9 Display functions
Parameter Enabled (shown) Disabled (hidden)
Totalizer start/stop Operators can start or stop totalizers using the display.
Operators cannot start or stop totalizers using the display.
Totalizer reset Operators can reset the mass and volume totalizers using the display.
Operators cannot reset the mass and volume totalizers using the display.
The display automatically scrolls through each process variable at a configurable rate.
Operators must Scroll to view process variables.
Off-line menu Operators can access the off-line menu (zero, simulation, and configuration).
Operators cannot access the off-line menu.
Operators must use a password to access the off-line menu.
Operators can access the off-line menu without a password.
Alarm menu Operators can access the alarm menu (viewing and acknowledging alarms).
Operators cannot access the alarm menu.
Operators are able to acknowledge all current alarms at once.
Operators must acknowledge alarms individually.
(1) If enabled, you may want to configure Scroll Rate.
(2) If enabled, the off-line password must also be configured.
To configure these parameters:
- You must enable Auto Scroll before you can configure Scroll Rate.
- You must enable the off-line password before you can configure the password.
The backlight of the LCD panel on the display can be turned on or off. To turn the backlight on or off,
Auto scroll(1) · Off-line password(2) · Acknowledge all alarms

In addition, if you are using ProLink II or a DeviceNet tool, you can control the intensity of the backlight. You can specify any value between 0 and 63; the higher the value, the brighter the backlight. To control the intensity of the backlight:
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
The display can scroll through up to 15 process variables in any order. You can configure the process variables to be displayed and the order in which they should appear. Additionally, you can configure display precision for each display variable. Display precision controls the number of digits to the right of the decimal place. Precision can be set to any value from 0 to 5.
a. Set the process variable index (Attribute 29) to the process variable to be configured.
b. Set the precision (Attribute 30) for that process variable.
Table 8-10 shows an example of a display variable configuration. Notice that you can repeat variables, and you can also specify None for any display variable except Display Var 1. For information on how the display variables will appear on the display, see Appendix D.
Display variable Process variable
Display variable 1(1)
Display variable 2 Mass totalizer
Display variable 3 Volume flow
Display variable 4 Volume totalizer
Display variable 5 Density
Display variable 6 Temperature
Display variable 7 External temperature
Display variable 8 External pressure
Display variable 9 Mass flow
Display variable 10 None
Display variable 11 None
Display variable 12 None
Display variable 13 None
Display variable 14 None
Display variable 15 None
(1) Display variable 1 cannot be set to None.
Mass flow

8.10 Configuring digital communications
The digital communications parameters control how the transmitter will communicate using digital communications. The following digital communications parameters can be configured:
The default node address for the Model 2400S DN transmitter is 63. The valid range of node addresses is 0–63.
The DeviceNet node address can be set using digital communications hardware switches or a DeviceNet tool.
To set the node address using digital communications hardware switches:
a. Rotate the arrow in the left switch to point to the digit 6.
b. Rotate the arrow in the center switch to point to the digit 0.
To set the node address using a DeviceNet tool:

8.10.2 DeviceNet baud rate
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
The default baud rate for the Model 2400S DN transmitter is 125 kBaud. Valid baud rates are listed in Table 8-11.
The baud rate can be set using a digital communications hardware switch or a DeviceNet tool. If the device cannot determine what its baud rate should be, it defaults to 500 kBaud.
To set the baud rate using the digital communications hardware switch:
Switch position Baud rate
3–9 (Program range) Controlled by DeviceNet system
To set the baud rate using a DeviceNet tool:
The Model 2400S transmitter provides 25 predefined input assemblies and one configurable input assembly. The configurable input assembly allows you to specify five process variables to be published to the network.
0 125 kBaud · 1 250 kBaud · 2 500 kBaud

The Assembly Object is used to configure the configurable input assembly. See the flowchart in Figure 8-2.
Note: The Modbus address is applicable only when you are connecting to the service port from a tool that uses Modbus protocol. After initial startup, service port connections are typically used only for troubleshooting or for specific procedures such as temperature calibration. ProLink II is typically used for service port connections, and by default ProLink II will use the standard service port address rather than the configured Modbus address. See Section 4.4 for more information.
The set of valid Modbus addresses depends on whether or not support for Modbus ASCII is enabled or disabled (see Section 8.10.5). Valid Modbus addresses are as follows:
To configure the Modbus address:
When support for Modbus ASCII is enabled, the service port can accept connection requests that use either Modbus ASCII or Modbus RTU. When support for Modbus ASCII is disabled, the service port cannot accept connection requests that use Modbus ASCII. Only Modbus RTU connections are accepted.

The primary reason to disable Modbus ASCII support is to allow a wider range of Modbus addresses for the service port.
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
To enable or disable Modbus ASCII support:
The IrDA port on the display can be enabled or disabled. If enabled, it can be set for read-only or read/write access.
To enable or disable the IrDA port:
To configure the IrDA port for read-only or read-write access:
Digital communications fault action controls how digital communications will be affected by fault conditions. Table 8-12 lists the options for digital communications fault action.
Option
Definition ProLink II label DeviceNet label DeviceNet code
Upscale Upscale 0 Process variables indicate the value is greater than the upper sensor limit. Totalizers stop counting.
Downscale Downscale 1 Process variables indicate the value is less than the lower sensor limit. Totalizers stop counting.
Zero Zero 2 Flow rates go to the value that represents zero flow. Density and temperature go to zero. Totalizers stop counting.
NAN 3 Process variables report IEEE NAN. Totalizers stop counting.
Flow to Zero Flow goes to zero 4 Flow rates go to the value that represents zero flow; other process variables are not affected. Totalizers stop counting.
None (default) None 5 Process variables reported as measured.
Not-A-Number (NAN)

To configure digital communications fault action:
By default, the transmitter activates the digital communications fault action as soon as the fault is detected. The fault timeout allows you to delay the digital communications fault action for a specified interval, for certain faults only. During the fault timeout period, digital communications behaves normally.
Note: The fault timeout applies only to the digital communications fault action. The “alarm active” status bit is set as soon as the fault is detected (all alarm severity levels), and the “alarm active” record is written to history immediately (Fault and Informational alarms only). For more information on alarm handling, see Section 7.6. For more information on alarm severity, see Section 8.8.
The fault timeout applies only to specific faults. Other faults are reported immediately, regardless of the fault timeout setting. For information on which faults are affected by the fault timeout, see Table 8-8.
To configure fault timeout:
The device settings are used to describe the flowmeter components. Table 8-13 lists and defines the device settings.
Parameter Description
Descriptor Any user-supplied description. Not used in transmitter processing, and not required. Maximum length: 16 characters.
Message Any user-supplied message. Not used in transmitter processing, and not required. Maximum length: 32 characters.
Date Any user-selected date. Not used in transmitter processing, and not required.
To configure device settings, you must use ProLink II. See Figure B-2. If you are entering a date, use the left and right arrows at the top of the calendar to select the year and month, then click on a date.

8.12 Configuring sensor parameters
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
The sensor parameters are used to describe the sensor component of your flowmeter. They are not used in transmitter processing, and are not required. The following sensor parameters can be changed:
To configure sensor parameters:
The API parameters determine the values that will be used in API-related calculations. The API parameters are available only if the petroleum measurement application is enabled on your transmitter.
Some applications that measure liquid volume flow or liquid density are particularly sensitive to temperature factors, and must comply with American Petroleum Institute (API) standards for measurement. The petroleum measurement enables Correction of Temperature on volume of Liquids, or CTL.
The following terms and definitions are relevant to the petroleum measurement application:
There are two derivation methods for CTL:

API reference tables
Reference tables are organized by reference temperature, CTL derivation method, liquid type, and density unit. The table selected here controls all the remaining options.
- If you specify a 5x, 6x, 23x, or 24x table, the default reference temperature is 60 °F, and cannot be changed.
- If you specify a 53x or 54x table, the default reference temperature is 15 °C. However, you can change the reference temperature, as recommended in some locations (for example, to 14.0 or 14.5 °C).
- If you specify an odd-numbered table (5, 23, or 53), CTL will be derived using method 1 described above.
- If you specify an even-numbered table (6, 24, or 54), CTL will be derived using method 2 described above.
- A tables are used with generalized crude and JP4 applications.
- B tables are used with generalized products.
- C tables are used with liquids with a constant base density or known thermal expansion coefficient.
- D tables are used with lubricating oils.
- Relative density (SG)
- Base density (kg/m3)
- Degrees API

Table 8-14 summarizes these options.
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
CTL derivation method Base temperature
Density unit and range
Table
Degrees API Base density Relative density
5A Method 1 60 °F, non-configurable 0 to +100
5B Method 1 60 °F, non-configurable 0 to +85
5D Method 1 60 °F, non-configurable –10 to +40
23A Method 1 60 °F, non-configurable 0.6110 to 1.0760
23B Method 1 60 °F, non-configurable 0.6535 to 1.0760
23D Method 1 60 °F, non-configurable 0.8520 to 1.1640
53A Method 1 15 °C, configurable 610 to 1075 kg/m3
53B Method 1 15 °C, configurable 653 to 1075 kg/m3
53D Method 1 15 °C, configurable 825 to 1164 kg/m3
6C Method 2 60 °F, non-configurable 60 °F Degrees API
24C Method 2 60 °F, non-configurable 60 °F Relative density
54C Method 2 15 °C, configurable 15 °C Base density in kg/m3
The API configuration parameters are listed and defined in Table 8-15.
Variable Description
Table type Specifies the table that will be used for reference temperature and reference density unit. Select the table that matches your requirements. See API reference tables.
User defined TEC(1)
Thermal expansion coefficient. Enter the value to be used in CTL calculation.
Read-only. Displays the unit used for reference temperature in the reference table.
Density units Read-only. Displays the unit used for reference density in the reference table.
Read-only unless Table Type is set to 53x or 54x. If configurable:
(1) Configurable if Table Type is set to 6C, 24C, or 54C.
(2) In most cases, the temperature unit used by the API reference table should also be the temperature unit configured for the transmitter
to use in general processing. To configure the temperature unit, see Section 6.3.
To configure the petroleum measurement application:
Temperature units(2) · Reference temperature

For the temperature value to be used in CTL calculation, you can use the temperature data from the sensor, or you can configure external temperature compensation to use either a static temperature value or temperature data from an external temperature device.
Micro Motion sensors provide direct measurements of density, but not of concentration. The enhanced density application calculates enhanced density process variables, such as concentration or density at reference temperature, from density process data, appropriately corrected for temperature.
The enhanced density calculation calculation requires an enhanced density curve, which specifies the relationship between temperature, concentration, and density for the process fluid being measured. Micro Motion supplies a set of six standard enhanced density curves (see Table 8-16). If none of these curves is appropriate for your process fluid, you can configure a custom curve or purchase a custom curve from Micro Motion.
The derived variable, specified during configuration, controls the type of concentration measurement that will be produced. Each derived variable allows the calculation of a subset of enhanced density process variables (see Table 8-17). The available enhanced density process variables can be used in process control, just as mass flow rate, volume flow rate, and other process variables are used. For example, an event can be defined on an enhanced density process variable.
The transmitter can hold up to six curves at any given time, but only one curve can be active (used for measurement) at a time. All curves that are in transmitter memory must use the same derived variable.
Name Description Density unit Temperature unit
Deg Balling Curve represents percent extract, by mass, in solution, based on °Balling. For example, if a wort is 10 °Balling and the extract in solution is 100% sucrose, the extract is 10% of the total mass.
Deg Brix Curve represents a hydrometer scale for sucrose solutions that indicates the percent by mass of sucrose in solution at a given temperature. For example, 40 kg of sucrose mixed with 60 kg of water results in a 40 °Brix solution.
Deg Plato Curve represents percent extract, by mass, in solution, based on °Plato. For example, if a wort is 10 °Plato and the extract in solution is 100% sucrose, the extract is 10% of the total mass.
g/cm3 °F · g/cm3 °C · g/cm3 °F

Table 8-16 Standard curves and associated measurement units continued
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
HFCS 42 Curve represents a hydrometer scale for HFCS 42 (high fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
HFCS 55 Curve represents a hydrometer scale for HFCS 55 (high fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
HFCS 90 Curve represents a hydrometer scale for HFCS 90 (high fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
Available process variables
Density at reference temperature
Standard volume flow rate
Concentration Net mass flow rate
Net volume flow rate
Density @ Ref Density at reference temperature Mass/unit volume, corrected to a given reference temperature
SG Specific gravity The ratio of the density of a process fluid at a given temperature to the density of water at a given temperature. The two given temperature conditions do not need to be the same.
Mass Conc (Dens) Mass concentration derived from reference density The percent mass of solute or of material in suspension in the total solution, derived from reference density
✓
Mass Conc (SG) Mass concentration derived from specific gravity The percent mass of solute or of material in suspension in the total solution, derived from specific gravity
✓
Volume Conc (Dens) Volume concentration derived from reference density The percent volume of solute or of material in suspension in the total solution, derived from reference density
✓

Table 8-17 Derived variables and available process variables continued
Available process variables
Density at reference temperature
Standard volume flow rate
Concentration Net mass flow rate
Net volume flow rate
✓
Volume Conc (SG) Volume concentration derived from specific gravity The percent volume of solute or of material in suspension in the total solution, derived from specific gravity
Conc (Dens) Concentration derived from reference density The mass, volume, weight, or number of moles of solute or of material in suspension in proportion to the total solution, derived from reference density
Conc (SG) Concentration derived from specific gravity The mass, volume, weight, or number of moles of solute or of material in suspension in proportion to the total solution, derived from specific gravity
✓
Complete configuration instructions for the enhanced density application are provided in the manual entitled Micro Motion Enhanced Density Application: Theory, Configuration, and Use. Because of the complexity of this procedure, Micro Motion recommends using ProLink II for detailed configuration.
If it is necessary to use a DeviceNet tool, refer to the enhanced density manual for application information, and to the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile for complete device profile information.
Basic information on setting up the enhanced density application using a DeviceNet tool is provided in Figure 8-3.
Specific gravity · ✓ ✓ ✓

Figure 8-3 Configuring the enhanced density application – DeviceNet tool
Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool Required Configuration Optional Configuration Using the Transmitter Using a DeviceNet Tool
Class: Enhanced Density Object (0x64) Instance: 1 Attribute ID: 39 Data type: BOOL Value: · 0: Disabled · 1: Enabled Service: Set
Enable the enhanced density application
Set transmitter density measurement unit to match curve unit · For standard curves, see Table 8-16 · For custom curves, see the information provided with the curve
Class: Analog Input Point Object (0x0A) Instance: 3 Attribute ID: 102 Value: See Table 6-5 Service: Set
Set transmitter temperature measurement unit to match curve unit · For standard curves, see Table 8-16 · For custom curves, see the information provided with the curve
Class: Analog Input Point Object (0x0A) Instance: 4 Attribute ID: 102 Value: See Table 6-6 Service: Set
Class: Enhanced Density Object (0x64) Instance: 1 Attribute ID: 15 Data type: USINT Value: See Table D-17 Service: Set
Set derived variable · For standard curves, use Mass Conc (Dens) · For custom curves, see the information provided with the curve
Class: Enhanced Density Object (0x64) Instance: 1 Attribute ID: 16 Data type: USINT Value: 0 – 5 Service: Set
Specify the active curve



Measurement Performance Defaults Troubleshooting Compensation
This chapter describes the following procedures:
The Model 2400S DN transmitter can compensate for the effect of pressure on the sensor flow tubes. Pressure effect is defined as the change in sensor flow and density sensitivity due to process pressure change away from calibration pressure.
There are two ways to configure pressure compensation:

9.2.2 Pressure correction factors
When configuring pressure compensation, you must provide the flow calibration pressure – the pressure at which the flowmeter was calibrated (which therefore defines the pressure at which there will be no effect on the calibration factor). Enter 20 PSI unless the calibration document for your sensor indicates a different calibration pressure.
Two additional pressure correction factors may be configured: one for flow and one for density. These are defined as follows:
Not all sensors or applications require pressure correction factors. For the pressure correction values to be used, obtain the pressure effect values from the product data sheet for your sensor, then reverse the signs (e.g., if the pressure effect is 0.000004, enter a pressure correction factor of –0.000004).
To enable and configure pressure compensation:
Set measurement unit(1)
ProLink > Configuration > Pressure
ProLink > Configuration > Pressure
Use static pressure value? No
assembly(2) Enter External
(1) Pressure measurement unit must be configured to match pressure
unit used by external device. See Section 6.3. (2) See Section 9.4.

Figure 9-2 Configuring pressure compensation with a DeviceNet tool
Measurement Performance Defaults Troubleshooting Compensation
(1) Pressure measurement unit must
be configured to match pressure unit used by external device. See Section 6.3. (2) See Section 9.4.
Class: Calibration Object (0x65) Instance: 1 Attribute ID: 27 Data type: BOOL Value: · 0 = disabled · 1 = enabled
Use static pressure value? No
External temperature compensation can be used with the petroleum measurement application or the enhanced density application.

There are two ways to configure external temperature compensation:
To enable and configure external temperature compensation:
View Menu > Preferences
Use static temp value?
ProLink > Configuration > Temperature
Apply (1) Temperature measurement unit must be
Enter Temperature units(1)
configured to match temperature unit used by external device. See Section 6.3. (2) See Section 9.4.

Figure 9-4 Configuring external temperature compensation with a DeviceNet tool
Class: Analog Input Point Object (0x0A) Instance: 4 Attribute ID: 102 Value: See Table 6-6 Service: Set
Measurement Performance Defaults Troubleshooting Compensation
(1) Temperature measurement unit
must be configured to match temperature unit used by external device. See Section 6.3. (2) See Section 9.4.
Class: Calibration Object (0x65) Instance: 1 Attribute ID: 25 Data type: BOOL Value: · 0 = disabled · 1 = enabled
Use static temp value? No
The DeviceNet output assemblies used to obtain external pressure and/or temperature data are listed in Table 9-1. Use standard DeviceNet methods to implement the required connection.
Instance ID Data description Size Bytes Data type
50 External pressure 4 bytes Bytes 0–3 REAL
51 External temperature 4 bytes Bytes 0–3 REAL
52 External pressure External temperature
8 bytes Bytes 0–3 Bytes 4–7



Measurement Performance Defaults Troubleshooting Compensation
This chapter describes the following procedures:
The Model 2400S transmitter supports the following procedures for the evaluation and adjustment of measurement performance:
Meter validation and calibration are available on all Model 2400S DN transmitters. Meter verification is available only if the meter verification option was ordered with the transmitter.
These three procedures are discussed and compared in Sections 10.2.1 through 10.2.4. Before performing any of these procedures, review these sections to ensure that you will be performing the appropriate procedure for your purposes.

10.2.1 Meter verification
Meter verification evaluates the structural integrity of the sensor tubes by comparing current tube stiffness to the stiffness measured at the factory. Stiffness is defined as the load per unit deflection, or force divided by displacement. Because a change in structural integrity changes the sensor’s response to mass and density, this value can be used as an indicator of measurement performance. Changes in tube stiffness are typically caused by erosion, corrosion, or tube damage.
Meter verification does not affect measurement in any way. Micro Motion recommends performing meter verification at regular intervals.
Meter validation compares a measurement value reported by the transmitter with an external measurement standard. Meter validation requires one data point.
If the transmitter’s mass flow, volume flow, or density measurement is significantly different from the external measurement standard, you may want to adjust the corresponding meter factor. A meter factor is the value by which the transmitter multiplies the process variable value. The default meter factors are 1.0, resulting in no difference between the data retrieved from the sensor and the data reported externally.
Meter factors are typically used for proving the flowmeter against a weights and measures standard. You may need to calculate and adjust meter factors periodically to comply with regulations.
The flowmeter measures process variables based on fixed points of reference. Calibration adjusts those points of reference. Three types of calibration can be performed:
Density and temperature calibration require two data points (low and high) and an external measurement for each. Zero calibration requires one data point. Calibration produces a change in the offset and/or the slope of the line that represents the relationship between the actual process value and the reported value.
Micro Motion flowmeters with the Model 2400S transmitter are calibrated at the factory, and normally do not need to be calibrated in the field. Calibrate the flowmeter only if you must do so to meet regulatory requirements. Contact Micro Motion before calibrating your flowmeter.

10.2.4 Comparison and recommendations
When choosing among meter verification, meter validation, and calibration, consider the following factors:
Measurement Performance Defaults Troubleshooting Compensation
- Meter verification requires approximately four minutes to perform. During these four minutes, flow can continue (provided sufficient stability is maintained); however, current process data will not be reported.
- Meter validation for density does not interrupt the process. However, meter validation for mass flow or volume flow requires process down-time for the length of the test.
- Calibration requires process down-time. In addition, density and temperature calibration require replacing the process fluid with low-density and high density fluids, or low-temperature and high-temperature fluids. Zero calibration requires stopping flow through the sensor.
- Meter verification does not require external measurements.
- Zero calibration does not require external measurements.
- Density calibration, temperature calibration, and meter validation require external measurements. For good results, the external measurement must be highly accurate.
- Meter verification is an indicator of sensor condition, but does not change flowmeter internal measurement in any way.
- Meter validation does not change flowmeter internal measurement in any way. If you decide to adjust a meter factor as a result of a meter validation procedure, only the reported measurement is changed – the base measurement is not changed. You can always reverse the change by returning the meter factor to its previous value.
- Calibration changes the transmitter’s interpretation of process data, and accordingly changes the base measurement. If you perform a zero calibration, you can return to the previous zero or the factory zero. However, if you perform a density calibration or a temperature calibration, you cannot return to the previous calibration factors unless you have manually recorded them.
Micro Motion recommends that you purchase the meter verification option and perform meter verification frequently.
The meter verification procedure can be performed on any process fluid. It is not necessary to match factory conditions. Meter verification is not affected by any parameters configured for flow, density, or temperature.
During the test, process conditions must be stable. To maximize stability:
If stability varies outside test limits, the meter verification procedure will be aborted. Verify the stability of the process and retry the test.

During meter verification, you can choose between setting digital communications process variable values to the configured fault indicator or the last measured value. The values will remain fixed for the duration of the test (approximately four minutes). Disable all control loops for the duration of the procedure, and ensure that any data reported during this period is handled appropriately.
To perform meter verification:
For a discussion of meter verification results, see Section 10.3.1.
Tools > Meter Verification > Structural Integrity Method
View previous test data
Enter optional test data
View report (option to print
Initialize and start meter
(1) If the graph was viewed at the beginning of the procedure,
clicking Back here will return to the beginning of the procedure (along the dotted line). (2) The results of the meter verification test are not saved until
Finish is clicked.

Figure 10-2 Meter verification procedure – Display menu

Measurement Performance Defaults Troubleshooting Compensation
(1) Either Unstable Flow or Unstable Drive Gain may
be displayed, indicating that the standard deviation of the flow or drive gain is outside limits. (2) Represents the percentage completion of the
x%(2) UNSTABLE FLOW(1)

Figure 10-3 Meter verification procedure – DeviceNet tool
Step 1 Set output state (optional)
See Table 10-1 for the device profile information for each step.
Step 2 Set uncertainty limit (optional)
Step 3 Start/abort procedure
Step 4 Check current algorithm state
Running? Step 5 Read percent complete
Step 6 Check algorithm abort state
Step 9 Check abort code
Step 7 Check inlet stiffness
Within limits? No (>0)
Step 8 Check outlet stiffness
Within limits? No (>0)
PASS

Table 10-1 DeviceNet interface for meter verification
Step number Step description Interface
Measurement Performance Defaults Troubleshooting Compensation
2 Set uncertainty limit Class: Diagnostics Object (0x66) Instance: 1 Attribute ID: 59 Data type: REAL Range: 0.0025 to 0.05 Default: 0.04 Service: Set
4 Check current algorithm state Class: Diagnostics Object (0x66) Instance: 1 Attribute ID: 54 Data type: USINT Service: Get
6 Check algorithm abort state Class: Diagnostics Object (0x66) Instance: 1 Attribute ID: 56 Data type: USINT Service: Get
9 Read abort code Class: Diagnostics Object (0x66) Instance: 1 Attribute ID: 55 Data type: USINT Codes: See Table C-7 Service: Get

10.3.1 Uncertainty limit and test results
The result of the meter verification test will be a percent uncertainty of normalized tube stiffness. The default limit for this uncertainty is ±4.0%. This limit is stored in the transmitter, and can be changed with ProLink II or a DeviceNet tool when optional test parameters are entered. For most installations, it is advisable to leave the uncertainty limit at the default value.
When the test is completed, the result will be reported as Pass, Fail/Caution (depending on the tool you are using), or Abort:
ProLink II provides more detailed test data. See Section 10.3.2.

10.3.2 Additional ProLink II tools for meter verification
In addition to the Pass, Fail/Caution, and Abort result provided by the procedure, ProLink II provides the following additional meter verification tools:
Measurement Performance Defaults Troubleshooting Compensation
More information about using ProLink II to perform meter verification can be found in the ProLink II manual and in the on-line ProLink II help system.
To perform meter validation:

Therefore, to adjust volume flow, you must set the meter factor for volume flow. Setting a meter factor for mass flow and a meter factor for density will not produce the desired result. The volume flow calculations are based on original mass flow and density values, before the corresponding meter factors have been applied.
a. Sample the process fluid and record the process variable value reported by the flowmeter.
b. Measure the sample using an external standard.
c. Calculate the new meter factor using the following formula:
NewMeterFactor ConfiguredMeterFactor ExternalStandard ActualFlowmeterMeasurement × =
If you are calculating the volume flow meter factor, note that proving volume in the field may be expensive, and the procedure may be hazardous for some process fluids. Therefore, because volume is inversely proportional to density, an alternative to direct sampling and measurement is to calculate the volume flow meter factor from the density meter factor. This method provides partial correction by adjusting for any portion of the total offset that is caused by density measurement offset. Use this method only when a volume flow reference is not available, but a density reference is available. To use this method:
a. Calculate the meter factor for density, using the preceding formula.
b. Calculate the volume flow meter factor from the density meter factor, as shown below:
1 MeterFactorDensity =
DensityFlowmeter DensityExternalStandard × =
MeterFactorVolume ConfiguredMeterFactorDensity
MeterFactorVolume

Example The flowmeter is installed and proved for the first time. The flowmeter mass measurement is 250.27 lb; the reference device measurement is 250 lb. A mass flow meter factor is determined as follows:
Measurement Performance Defaults Troubleshooting Compensation
MassFlowMeterFactor 1 250 250.27 × 0.9989 = =
The first mass flow meter factor is 0.9989.
One year later, the flowmeter is proved again. The flowmeter mass measurement is 250.07 lb; the reference device measurement is 250.25 lb. A new mass flow meter factor is determined as follows:
MassFlowMeterFactor 0.9989 250.25 250.07 × 0.9996 = =
The new mass flow meter factor is 0.9996.
Zeroing the flowmeter establishes the flowmeter’s point of reference when there is no flow. The meter was zeroed at the factory, and should not require a field zero. However, you may wish to perform a field zero to meet local requirements or to confirm the factory zero.
When you zero the flowmeter, you may need to adjust the zero time parameter. Zero time is the amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds.
For most applications, the default zero time is appropriate.
If the zero procedure fails, two recovery functions are provided:
- The display – see Figure B-7
- ProLink II – see Figure B-1
- A DeviceNet tool – use the Diagnostics Object (0x66), Instance 1, Service 0x52. For more information, see the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile.
If desired, you can use one of these functions to return the meter to operation while you are troubleshooting the cause of the zero failure (see Section 11.8).

10.5.1 Preparing for zero
To prepare for the zero procedure:
If fluid is flowing through the sensor during zero calibration, the calibration may be inaccurate, resulting in inaccurate process measurement.
To improve the sensor zero calibration and measurement accuracy, ensure that process flow through the sensor has completely stopped.
To zero the flowmeter:
- The zero button is not available.
- If the off-line menu has been disabled, you will not be able to zero the transmitter with the display. For information about enabling and disabling the off-line menu, see Section 8.9.3.
- You cannot change the zero time with the display. If you need to change the zero time, you must use ProLink II or a DeviceNet tool.
- You cannot change the zero time with the zero button. If you need to change the zero time, you must use ProLink II or a DeviceNet tool.
- The zero button is located on the user interface board, beneath the transmitter housing cover (see Figure 3-1). For instructions on removing the transmitter housing cover, see Section 3.3.
- To press the zero button, use a fine-pointed object that will fit into the opening (0.14 in or 3.5 mm). Hold the button down until the status LED on the user interface module begins to flash yellow.

Figure 10-4 Zero button – Flowmeter zero procedure
Measurement Performance Defaults Troubleshooting Compensation
Scroll and Select simultaneously
ZERO
ZERO/YES?
ZERO
EXIT

Figure 10-6 ProLink II – Flowmeter zero procedure
ProLink > Calibration > Zero Calibration
Wait until Calibration in Progress LED turns green

Figure 10-7 DeviceNet tool – Flowmeter zero procedure
Measurement Performance Defaults Troubleshooting Compensation
Class: Calibration Object (0x65) Instance: 1 Attribute ID: 3 Data type: UINT Units: seconds Service: Set
Modify zero time if
Class: Calibration Object (0x65) Instance: 1 Service: 0x4B (perform or abort zero) Value: · 0: Abort zero calibration · 1: Start zero calibration
Attribute ID 16, Bit 0x8000 Value: · 0: Zero complete · 1: Zero in progress Data type: USINT
Attribute ID 12, Bit 0x0020 Value: · 0: Zero succeeded · 1: Zero failed Data type: USINT
Class: Calibration Object (0x65) Instance: 1 Attribute ID 4: Zero standard deviation Attribute ID 5: Zero offset Data type: REAL Units: milliseconds
Density calibration includes the following calibration points:
- D1 calibration (low-density)
- D2 calibration (high-density)
- D3 calibration (optional)
- D4 calibration (optional)
For T-Series sensors, the optional D3 and D4 calibrations could improve the accuracy of the density measurement. If you choose to perform the D3 and D4 calibration:

The calibrations that you choose must be performed without interruption, in the order listed here.
You can calibrate for density with ProLink II or a DeviceNet tool.
Before beginning density calibration, review the requirements in this section.
During density calibration, the sensor must be completely filled with the calibration fluid, and flow through the sensor must be at the lowest rate allowed by your application. This is usually accomplished by closing the shutoff valve downstream from the sensor, then filling the sensor with the appropriate fluid.
D1 and D2 density calibration require a D1 (low-density) fluid and a D2 (high-density) fluid. You may use air and water. If you are calibrating a T-Series sensor, the D1 fluid must be air and the D2 fluid must be water.
For T-Series sensors, the D1 calibration must be performed on air and the D2 calibration must be performed on water.
For D3 density calibration, the D3 fluid must meet the following requirements:
For D4 density calibration, the D4 fluid must meet the following requirements:
To perform a D1 and D2 density calibration:
To perform a D3 density calibration or a D3 and D4 density calibration:

Figure 10-8 D1 and D2 density calibration – ProLink II
Measurement Performance Defaults Troubleshooting Compensation
Close shutoff valve downstream from sensor Fill sensor with D1 fluid Fill sensor with D2 fluid
ProLink Menu > Calibration > Density cal – Point 1
ProLink Menu > Calibration > Density cal – Point 2
Enter density of D1 fluid
Enter density of D2 fluid

Figure 10-9 D1 and D2 density calibration – DeviceNet tool
Close shutoff valve downstream from sensor
Bit 14 0ff? No

Figure 10-10 D3 or D3 and D4 density calibration – ProLink II
Measurement Performance Defaults Troubleshooting Compensation
Close shutoff valve downstream from sensor Fill sensor with D3 fluid
Fill sensor with D4 fluid
ProLink Menu > Calibration > Density cal – Point 3
ProLink Menu > Calibration > Density cal – Point 4
Enter density of D3 fluid
Enter density of D4 fluid

Figure 10-11 D3 or D3 and D4 density calibration – DeviceNet tool
Close shutoff valve downstream from sensor
Bit 7 0ff? No

10.7 Performing temperature calibration
Temperature calibration is a two-part procedure: temperature offset calibration and temperature slope calibration. The entire procedure must be completed without interruption.
Measurement Performance Defaults Troubleshooting Compensation
To perform temperature calibration, you must use ProLink II. See Figure 10-12.
Fill sensor with low-
Fill sensor with high-
Wait until sensor achieves
Wait until sensor achieves
ProLink Menu > Calibration > Temp offset cal
ProLink Menu > Calibration > Temp slope cal
Enter temperature of low-
Enter temperature of high-



Measurement Performance Defaults Troubleshooting Compensation
This chapter describes guidelines and procedures for troubleshooting the flowmeter. The information in this chapter will enable you to:
Using the service port clips to communicate with the transmitter in a hazardous area can cause an explosion.
Before using the service port clips to communicate with the transmitter in a hazardous area, make sure the atmosphere is free of explosive gases.
Refer to Table 11-1 for a list of troubleshooting topics discussed in this chapter.
Section Topic
Section 11.4 Transmitter does not operate
Section 11.5 Transmitter does not communicate
Section 11.6 Checking the communication device
Section 11.7 Diagnosing wiring problems
Section 11.7.1 Checking the DeviceNet cable and connector

Table 11-1 Troubleshooting topics and locations continued
Section Topic
Section 11.7.2 Checking grounding
Section 11.8 Zero or calibration failure
Section 11.9 Fault conditions
Section 11.10 Simulation mode for process variables
Section 11.11 Transmitter LEDs
Section 11.12 Status alarms
Section 11.13 Checking process variables
Section 11.14 Checking slug flow
Section 11.15 Checking the sensor tubes
Section 11.16 Checking the flow measurement configuration
Section 11.17 Checking the characterization
Section 11.18 Checking the calibration
Section 11.19 Checking the test points
Section 11.20 Checking sensor circuitry
To speak to a customer service representative, contact the Micro Motion customer service department. Contact information is provided in Section 1.10.
Before contacting Micro Motion customer service, review the troubleshooting information and procedures in this chapter, and have the results available for discussion with the technician.
If the transmitter is not receiving power, all three LEDs on the user interface will be off.
If at least one LED is lit, perform all of the procedures in Section 11.7.
If the procedures do not indicate a problem with the electrical connections, contact the Micro Motion customer service department.
If the transmitter does not appear to be communicating, the wiring may be faulty or the communications device may be incompatible. Check the wiring and the communication device. See Chapter 4 for ProLink II and Pocket ProLink, or Chapter 5 for a DeviceNet tool.
If you are trying to communicate via the IrDA port, ensure that the port is enabled, that read-write access is enabled, and that there is no active connection via the service port clips. See Section 8.10.6.

If the transmitter is communicating via the service port but not via DeviceNet, if you are experiencing intermittent DeviceNet communications, or if the transmitter appears to be operating normally but you cannot establish a DeviceNet connection:
Measurement Performance Defaults Troubleshooting Compensation
Ensure that your communication device is compatible with your transmitter.
ProLink II v2.5 or later is required. To check the version of ProLink II:
Pocket ProLink v1.3 or later is required. To check the version of Pocket ProLink:
The Model 2400S DN transmitter is compatible with all DeviceNet tools. Check that your DeviceNet tool is correctly configured and can make a connection to other devices on the network.
Use the procedures in this section to check the transmitter installation for wiring problems.
Removing the transmitter housing cover in explosive atmospheres while the device is powered can subject the transmitter to environmental conditions that can cause an explosion.
Before removing the transmitter housing cover in explosive atmospheres, be sure to remove power from the device and wait five minutes.
To check the DeviceNet cable and connector:


The sensor / transmitter assembly must be grounded. See your sensor installation manual for grounding requirements and instructions.
If a zero or calibration procedure fails, the transmitter will send a status alarm indicating the cause of failure. See Section 11.12 for specific remedies for status alarms indicating calibration failure.
If a fault is reported, determine the exact nature of the fault by checking the status alarms (see Section 7.6). Once you have identified the status alarm(s) associated with the fault condition, refer to Section 11.12.
Some fault conditions can be corrected by cycling power to the transmitter. A power cycle can clear the following:
Simulation allows you to define arbitrary values for mass flow, temperature, and density. Simulation mode has several uses:

If simulation mode is active, the simulated values are stored in the same memory locations used for process data from the sensor. Therefore, the simulated values will be used throughout transmitter functioning. For example, simulation will affect:
Measurement Performance Defaults Troubleshooting Compensation
Accordingly, do not enable simulation when your process cannot tolerate these effects, and be sure to disable simulation when you have finished testing.
Simulation does not change any diagnostic values.
Simulation mode is available only via ProLink II. To set up simulation, refer to Figure B-3 and follow the steps below.
a. Specify the type of simulation you want: fixed value, sawtooth (triangular wave), or sine wave.
b. Enter the required values.
To use simulation mode for problem location, enable simulation mode and check the signal at various points between the transmitter and the receiving device.
Be sure to disable simulation when testing is complete.
The user interface board includes three LEDs:
a. View the alarm code using the procedures described in Section 7.5.
b. Identify the alarm (see Section 11.12).
c. Correct the condition.
d. If desired, acknowledge the alarm using the procedures described in Section 7.6.

11.12 Status alarms
Status alarm codes are reported on the LCD panel (for transmitters that have a display), and status alarms can be viewed with ProLink II or a DeviceNet tool (see Section 7.6). All possible status alarms are listed in Table 11-2, along with the ProLink II message, possible causes, and suggested remedies.
You may find it useful to acknowledge all alarms before beginning the troubleshooting procedures. This will remove inactive alarms from the list and allow you to focus on active alarms.
Alarm code
ProLink II message Cause Suggested remedy
A001 (E)EPROM Checksum Error (CP)
An uncorrectable checksum mismatch has been detected
A002 RAM Error (CP) ROM checksum error or a RAM location cannot be written to
A003 Sensor Failure Continuity failure of drive circuit, LPO, or RPO, or LPO-RPO mismatch when driving
A004 Temperature Sensor Failure
Combination of A016 and A017
A005 Input Overrange The measured flow has exceeded the maximum flow rate of the sensor (ΔT > 200 μs)
A006 Not Configured Combination of A020 and A021

Table 11-2 Status alarms and remedies continued
Alarm code
ProLink II message Cause Suggested remedy
Measurement Performance Defaults Troubleshooting Compensation
A009 Transmitter Initializing/Warming Up
Transmitter in power-up mode
A010 Calibration Failure Mechanical zero: The resulting zero was greater than 3 μs. Temperature/Density Cals: many possible causes.
A016 Line RTD Temperature Out-Of-Range
The value computed for the resistance of the Line RTD is outside limits
A017 Meter RTD Temperature Out-of-Range
The value computed for the resistance of the Meter/Case RTD is outside limits
A020 Calibration Factors Unentered (FlowCal)
The flow calibration factor and/or K1 has not been entered since the last master reset
A021 Incorrect Sensor Type (K1)
The sensor is recognized as a straight tube but the K1 value indicates a curved tube, or vice versa
A029 PIC/Daughterboard Communication Failure
Transmitter electronics failure

Table 11-2 Status alarms and remedies continued
Alarm code
ProLink II message Cause Suggested remedy
A030 Incorrect Board Type
The loaded software is not compatible with the programmed board type
A031 Low Power The transmitter is not receiving enough power
A032 Meter Verification/Outputs In Fault
Meter verification in progress, with outputs set to fault
A033 Sensor OK, Tubes Stopped by Process
No signal from LPO or RPO, suggesting that sensor tubes are not vibrating
A102 Drive Overrange/ Partially Full Tube
The drive power (current/voltage) is at its maximum
A104 Calibration in Progress
A calibration procedure is in progress
A105 Slug Flow The density has exceeded the user-defined slug (density) limits
A107 Power Reset Occurred
The transmitter has been restarted
The process temperature is outside API-defined extrapolation limits
The process density is outside API-defined extrapolation limits
The configured values for density curves do not meet accuracy requirements
Enhanced density calculations are outside the configured data range
A131 Meter Verification/Outputs at Last Value
Meter verification in progress, with outputs set to last measured value
A132 Simulation Mode Active
Simulation mode is enabled
A133 PIC UI EEPROM Error
EEPROM data on the user interface module is corrupt

11.13 Checking process variables
Micro Motion suggests that you make a record of the process variables listed below, under normal operating conditions. This will help you recognize when the process variables are unusually high or low.
Measurement Performance Defaults Troubleshooting Compensation
To view these values:
For troubleshooting, check the process variables under both normal flow and tubes-full no-flow conditions. Except for flow rate, you should see little or no change between flow and no-flow conditions. If you see a significant difference, record the values and contact Micro Motion customer service for assistance. See Section 11.3.
Unusual values for process variables may indicate a variety of different problems. Table 11-3 lists several possible problems and suggested remedies.
Symptom Cause Suggested remedy
Steady non-zero flow rate under no-flow conditions
Misaligned piping (especially in new installations)
Open or leaking valve • Check or correct the valve mechanism.
Bad sensor zero • Rezero the flowmeter or restore the factory zero or prior zero. See Section 10.5.

Table 11-3 Process variables problems and remedies continued
Symptom Cause Suggested remedy
Erratic non-zero flow rate under no-flow conditions
Leaking valve or seal • Check pipeline.
Slug flow • See Section 11.14.
Plugged flow tube • Check drive gain and tube frequency. Purge the flow tubes.
Incorrect sensor orientation • Sensor orientation must be appropriate to process fluid. See the installation manual for your sensor.
Wiring problem • Check the sensor circuitry. See Section 11.20.
Vibration in pipeline at rate close to sensor tube frequency
Damping value too low • Check configuration. See Section 8.4.
Mounting stress on sensor • Check sensor mounting. Ensure: - Sensor is not being used to support
pipe. - Sensor is not being used to correct
pipe misalignment. - Sensor is not too heavy for pipe.
Sensor cross-talk • Check environment for sensor with similar (±0.5 Hz) tube frequency.
Erratic non-zero flow rate when flow is steady
Slug flow • See Section 11.14.
Damping value too low • Check configuration. See Section 8.4.
Plugged flow tube • Check drive gain and tube frequency. Purge the flow tubes.
Excessive or erratic drive gain • See Section 11.19.3.
Output wiring problem • Verify wiring between transmitter and receiving device. See the installation manual for your transmitter.
Problem with receiving device • Test with another receiving device.
Wiring problem • Check the sensor circuitry. See Section 11.20.
Inaccurate flow rate or batch total Bad flow calibration factor • Verify characterization. See Section 6.2.
Inappropriate measurement unit • Check configuration. See Section 11.16.
Bad sensor zero • Rezero the flowmeter or restore the factory zero or prior zero. See Section 10.5.
Bad density calibration factors • Verify characterization. See Section 6.2.
Bad flowmeter grounding • See Section 11.7.2.
Slug flow • See Section 11.14.
Wiring problem • Check the sensor circuitry. See Section 11.20.

Table 11-3 Process variables problems and remedies continued
Symptom Cause Suggested remedy
Measurement Performance Defaults Troubleshooting Compensation
Inaccurate density reading Problem with process fluid • Use standard procedures to check quality of process fluid.
Bad density calibration factors • Verify characterization. See Section 6.2.
Wiring problem • Check the sensor circuitry. See Section 11.20.
Bad flowmeter grounding • See Section 11.7.2.
Slug flow • See Section 11.14.
Sensor cross-talk • Check environment for sensor with similar (±0.5 Hz) tube frequency.
Plugged flow tube • Check drive gain and tube frequency. Purge the flow tubes.
Incorrect sensor orientation • Sensor orientation must be appropriate to process fluid. See the installation manual for your sensor.
RTD failure • Check for alarm conditions and follow troubleshooting procedure for indicated alarm.
Physical characteristics of sensor have changed
Temperature reading significantly different from process temperature
Temperature reading slightly different from process temperature
Sensor leaking heat • Insulate the sensor.
Unusually high density reading Plugged flow tube • Check drive gain and tube frequency. Purge the flow tubes.
Incorrect K2 value • Verify characterization. See Section 6.2.
Unusually low density reading Slug flow • See Section 11.14.
Incorrect K2 value • Verify characterization. See Section 6.2.
Unusually high tube frequency Sensor erosion • Contact Micro Motion. See Section 11.3.
Unusually low pickoff voltages Several possible causes • See Section 11.19.4.
Unusually high drive gain Several possible causes • See Section 11.19.3.
A slug flow alarm is posted whenever the measured process density is outside the configured slug flow limits (i.e., density is higher or lower than the configured normal range). Slug flow is typically caused by gas in a liquid process or liquid in a gas process. See Section 8.7 for a discussion of slug flow functionality.

If slug flow occurs:
- Raising the low slug flow limit or lowering the high slug flow limit will increase the possibility of slug flow conditions.
- Lowering the low slug flow limit or raising the high slug flow limit will decrease the possibility of slug flow conditions.
Corrosion, erosion, or damage to the sensor tubes can affect process measurement. To check for these conditions, perform the meter verification procedure. See Chapter 10.
Using an incorrect flow measurement unit can cause the transmitter to report unexpected process variable values, with unpredictable effects on the process. Make sure that the configured flow measurement unit is correct. Check the abbreviations; for example, g/min represents grams per minute, not gallons per minute. See Section 6.3.
A transmitter that is incorrectly characterized for its sensor might report inaccurate process variable values. Both the K1 and Flow Cal (FCF) values must be appropriate for the sensor. If these values are incorrect, the sensor may not drive correctly or may send inaccurate process data.
If you discover that any of the characterization data are wrong, perform a complete characterization. See Section 6.2.
Improper calibration can cause the transmitter to report unexpected process variable values. If the transmitter appears to be operating correctly but sends unexpected process variable values, an improper calibration may be the cause.
Micro Motion calibrates every transmitter at the factory. Therefore, you should suspect improper calibration only if the transmitter has been calibrated after it was shipped from the factory. Before performing a calibration, consider meter validation or meter verification and select the appropriate procedure (see Section 10.2). Contact Micro Motion customer service for assistance.
Some status alarms that indicate a sensor failure or overrange condition can be caused by problems other than a failed sensor. You can diagnose sensor failure or overrange status alarms by checking the flowmeter test points. The test points include left and right pickoff voltages, drive gain, and tube frequency. These values describe the current operation of the sensor.

11.19.1 Obtaining the test point values
To obtain the test point values:
Measurement Performance Defaults Troubleshooting Compensation
a. Click ProLink > Diagnostic Information.
b. Observe or record the values displayed for Tube Frequency, Left Pickoff, Right Pickoff, and Drive Gain.
Test point Class Instance Attribute
Drive gain Diagnostics Object (0x66) 1 20
Use the following guidelines to evaluate the test points:
Sensor (1)
Pickoff value
ELITE® CMF sensors 3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
F025, F050, F100 sensors 3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
F200 sensors 2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
H025, H050, H100 sensors 3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
H200 sensors 2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
R025, R050, or R100 sensors 3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
R200 sensors 2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
T-Series sensors 0.5 mV peak-to-peak per Hz based on sensor flow tube frequency
CMF400 I.S. sensors 2.7 mV peak-to-peak per Hz based on sensor flow tube frequency
(1) If your sensor is not listed, contact Micro Motion. See Section 11.3.
Tube period 21 · Left pickoff 23 · Right pickoff 24

11.19.3 Drive gain problems
Problems with drive gain can appear in several different forms:
See Table 11-6 for a list of possible problems and remedies.
Cause Possible remedy
Excessive slug flow • See Section 11.14.
Plugged flow tube • Purge the flow tubes.
Mechanical binding of sensor tubes • Ensure sensor tubes are free to vibrate. Possible problems include: - Pipe stress. Check for pipe stress and eliminate if present. - Lateral tube shift due to hammer effect. If this is a possibility,
contact Micro Motion. See Section 11.3. - Warped tubes caused by overpressurization. If this is a possibility,
contact Micro Motion.
Incorrect sensor type configured • Verify sensor type configuration, then verify sensor characterization. See Section 6.2.
Open drive or left pickoff sensor coil • Contact Micro Motion. See Section 11.3.
Drive board or module failure, cracked flow tube, or sensor imbalance
Low pickoff voltage can be caused by several problems. See Table 11-7.
Cause Possible remedy
Slug flow • See Section 11.14.
No tube vibration in sensor • Check for plugging.
Moisture in the sensor electronics • Eliminate the moisture in the sensor electronics.
Damaged sensor • Ensure sensor is free to vibrate (no mechanical binding). Possible problems include: - Pipe stress. Check for pipe stress and eliminate if present. - Lateral tube shift due to hammer effect. If this is a possibility,
contact Micro Motion. See Section 11.3. - Warped tubes caused by overpressurization. If this is a possibility,

11.20 Checking sensor circuitry
Problems with sensor circuitry can cause several alarms, including sensor failure and a variety of out-of-range conditions. Testing involves:
Measurement Performance Defaults Troubleshooting Compensation
a. Referring to Figure 11-2, loosen the four captive transmitter housing cover screws and remove the transmitter housing cover.
b. Loosen the two captive user interface screws.
c. Gently lift the user interface module, disengaging it from the connector on the transmitter.
d. Two captive screws (2.5 mm hex head) hold the transmitter in the housing. Loosen the screws and gently lift the transmitter away from the housing. Allow the transmitter to hang by the cable.
e. Check the cable for any signs of damage.
f. Ensure that the cable is fully plugged in and making a good connection. If it is not, reseat the cable, reassemble the transmitter and sensor, and check operation.

Figure 11-2 Exploded view of transmitter and connection to sensor

Feedthrough (mounted on sensor)

Figure 11-3 Accessing the feedthrough pins
Measurement Performance Defaults Troubleshooting Compensation


Feedthrough connector Snap clip (assembled) Pull tab to remove Feedthrough pins
In this test:
If a problem appears, or if any resistance is out of range, contact Micro Motion (see Section 11.3).
Circuit Pin pairs Nominal resistance range (1)
Drive Drive + and – 8–1500 Ω
Left pickoff Left pickoff + and – 16–1000 Ω
Right pickoff Right pickoff + and – 16–1000 Ω
Transmitter

Table 11-8 Nominal resistance ranges for flowmeter circuits continued
Circuit Pin pairs Nominal resistance range (1)
Flow tube temperature sensor RTD + and RTD – 100 Ω at 0 °C + 0.38675 Ω / °C
(1) Actual resistance values depend on the sensor model and date of manufacture. Contact Micro Motion for more detailed data.
Drive – Drive +
Return for RTD, LLC, composite RTD, or fixed resistor
LLC / Composite RTD / Fixed resistor(1)
(1) Lead length compensator (LLC) for all sensors except T-Series, CMF400 I.S., and F300. For T-Series sensors, functions
as composite RTD. For CMF400 I.S. and F300 sensors, functions as fixed resistor.
a. Check between the pin and the sensor case.
b. Check between the pin and other pins as described below:

With the DMM set to its highest range, there should be infinite resistance on each lead. If there is any resistance at all, there is a short to case or a short between pins. See Table 11-9 for possible causes and solutions. If the problem is not resolved, contact Micro Motion (see Section 11.3).
Measurement Performance Defaults Troubleshooting Compensation
Cause Possible remedy
Moisture inside the transmitter housing • Make sure that the transmitter housing is dry and no corrosion is present.
Liquid or moisture inside the sensor case • Contact Micro Motion. See Section 11.3.
Internally shorted feedthrough (sealed passage for wiring from sensor to transmitter)
Faulty cable connecting sensor to transmitter • Visually inspect the cable for damage. To replace cable, contact Micro Motion. See Section 11.3.
To return to normal operation:
a. Rotate the connector until it engages the pins.
b. Push down until the connector shoulder is flush with the feedthrough notch.
c. Replace the snap clip by sliding the clip tab over the connector shoulder (see instruction label).



Measurement Performance Defaults Troubleshooting Compensation
This appendix provides information on the default values for most transmitter parameters. Where appropriate, valid ranges are also defined.
These default values represent the transmitter configuration after a master reset. Depending on how the transmitter was ordered, certain values may have been configured at the factory.
The table below contains the default values and ranges for the most frequently used transmitter settings.
Type Setting Default Range Comments
Flow Flow direction Forward
Flow damping 0.64 sec 0.0–40.96 sec User-entered value is corrected to nearest lower value in list of preset values. For gas applications, Micro Motion recommends a minimum value of 2.56.
Flow calibration factor 1.00005.13 For T-Series sensors,this value represents the FCF and FT factors concatenated. See Section 6.2.2.
Mass flow units g/s
Mass flow cutoff 0.0 g/s Recommended setting is 5% of the sensor’s rated maximum flowrate.
Volume flow type Liquid volume
Volume flow units L/s
Volume flow cutoff 0/0 L/s 0.0–x L/s x is obtained by multiplying the flow calibration factor by 0.2, using units of liters per second.
Meter factors Mass factor 1.00000
Density factor 1.00000
Volume factor 1.00000

Table A-1 Transmitter default values and ranges continued
Density Density damping 1.28 sec 0.0–40.96 sec User-entered value is corrected to nearest value in list of preset values.
Density units g/cm3
Density cutoff 0.2 g/cm3 0.0–0.5 g/cm3
D2 1.00000
K1 1000.00
K2 50,000.00
FD 0.00000
Temp Coefficient 4.44
Slug flow Slug flow low limit 0.0 g/cm3 0.0–10.0 g/cm3
Slug flow high limit 5.0 g/cm3 0.0–10.0 g/cm3
Slug duration 0.0 sec 0.0–60.0 sec
Temperature Temperature damping 4.8 sec 0.0–38.4 sec User-entered value is corrected to nearest lower value in list of preset values.
Temperature units Deg C
Temperature calibration factor 1.00000T0.0000
Pressure Pressure units PSI
Flow factor 0.00000
Density factor 0.00000
Cal pressure 0.00000
T-Series sensor D3 0.00000
K3 0.00000
K4 0.00000
FTG 0.00000
FFQ 0.00000
DTG 0.00000
DFQ1 0.00000
DFQ2 0.00000
Events 1–5 Type Low
Setpoint 0.0
Setpoint units g/cm3
Variable Density

Table A-1 Transmitter default values and ranges continued
Measurement Performance Defaults Troubleshooting Compensation Measurement Performance Defaults Troubleshooting Compensation Measurement Performance Defaults Troubleshooting Compensation Measurement Performance Defaults Troubleshooting Compensation
Display Backlight on/off On
Backlight intensity 63 0–63
Update period 200 milliseconds 100–10,000 milliseconds
Variable 1 Mass flow rate
Variable 2 Mass total
Variable 3 Volume flow rate
Variable 4 Volume total
Variable 7 Drive gain
Display totalizer start/stop Disabled
Display totalizer reset Disabled
Display auto scroll Disabled
Display offline menu Enabled
Display offline password Disabled
Display alarm menu Enabled
Display acknowledge all alarms
Auto scroll rate 10 sec
Fault timeout 0 seconds 0.0–60.0 sec
Modbus ASCII support Enabled
IrDA port enabled/disabled Disabled
IrDA port write-protect Disabled
Floating-point byte order 3–4–1–2



Device Profile Index Display Codes Menus
This appendix provides the following menu flowcharts for the Model 2400S DN transmitter:
- Main menu – see Figure B-1
- Configuration menu – see Figures B-2 and B-3
For information on the codes and abbreviations used on the display, see Appendix D.
These menu flowcharts are based on:
Menus may vary slightly for different versions of these components.

Figure B-1 ProLink II main menu
Load from Xmtr to File Save to Xmtr from File
Connect to Device Disconnect
Options · ProLink II Language · Error Log On
Preferences · Use External Temperature · Enable Inventory Totals Reset · Enable External Pressure Compensation · Copper RTD
Configuration Process Variables Status Alarm Log Diagnostic Information Calibration Test ED Totalizer Control(2)
Totalizer Control Core Processor Diagnostics API Process Variables(3)
(1) For information about using the data logging
function, see the ProLink II manual. (2) Available only if the enhanced density
ED Process Variables(2)
application is installed. (3) Available only if the petroleum measurement
application is installed.

Figure B-2 ProLink II configuration menu
Device Profile Index Display Codes Menus
· Flow direction · Flow damp · Flow cal · Mass flow cutoff · Mass flow units · Vol flow cutoff(1)
· Density units · Density damping · Slug high limit · Slug low limit · Slug duration · Low density cutoff · K1 · K2 · FD · D1 · D2 · Temp coeff (DT)
· Temp units · Temp cal factor · Temp damping · External temperature
· Flow factor · Dens factor · Cal pressure · Pressure units · External pressure
· Vol flow units(1)
· Vol flow type · Std gas vol flow cutoff(2)
· Std gas flow units(2)
· Std gas density(2)
· Mass factor · Dens factor · Vol factor
· Sensor s/n · Sensor model num · Sensor matl · Liner matl · Flange
Mass flow · Lower sensor limit · Upper sensor limit · Min span
· FTG · FFQ · DTG · DFQ1 · DFQ2 · K3 · D3 · D4 · K4
· Tag · Date · Descriptor · Message · Floating pt ordering · Add comm resp delay · Transmitter serial #
Volume flow · Lower sensor limit · Upper sensor limit · Min span
Digital comm settings · Fault setting · Modbus address · Disable Modbus ASCII · Enable IrDA comm · Enable write protect IrDA port
Density · Lower sensor limit · Upper sensor limit · Min span
Temperature · Lower sensor limit · Upper sensor limit · Min span
Last measured value timeout
(1) Displayed only if Vol Flow Type is set to Liquid Volume. (2) Displayed only if Vol Flow Type is set to Standard Gas Volume. (3) All values on this panel are read-only, and are displayed only for informational purposes.

Figure B-3 ProLink II configuration menu continued
· Var1 · Var2 · … · Var 15
· Meter fingerprinting · Cryogenic modulus compensation · Meter verification
· Alarm · Severity
Mass flow · Wave form · Fixed value · Period · Minimum · Maximum
Display precision · Var · Number of decimals
Display options · Display start/stop totalizers · Display totalizer reset · Display auto scroll · Display offline menu · Display offline password · Display alarm menu · Display ack all alarms · Display back light on/off
Density · Wave form · Fixed value · Period · Minimum · Maximum
· Event name · Event type · Process variable · Low setpoint · High setpoint
· Start sensor zero · Reset mass total · Reset volume total · Reset all totals · Start/stop all totalization · Reset gas standard volume total · Reset API reference volume total · Reset ED reference volume total · Reset ED net mass total · Reset ED net volume total · Increment current ED curve
Temperatuare · Fixed value · Period · Minimum · Maximum
· Offline password · Auto scroll rate · Update period · Backlight intensity
· Table type · Units
Global config · Active curve · Derived variable · Lock/unlock ED curves
Process fluid density at specified temperature and concentration · Curve being configured · Curve fit max order · Temperature isotherms · Concentration
Curve specific config · Curve configured · Curve name · Reference temperature · Water reference temperature · Water reference density · Trim slope · Trim offset
Curve fit results · Accuracy
Process fluid density at reference temperature and specified concentration · Reference temperature · Concentration
Extrapolation · Alarm limit · Enable density low · Enable density high · Enable temperature low · Enable temperature high
(1) Used to assign events to actions, even though the
Model 2400S DN transmitter does not provide a discrete input. (2) Available only if the petroleum measurement
Concentration · Units · Special unit string
application is installed. (3) Available only if the enhanced density application is
installed.

Figure B-4 Display menu – Off-line menu, top level

Device Profile Index Display Codes Menus
(1) This option is displayed only if the meter verification software is installed on the transmitter.

(1) The option is displayed only if the
corresponding Engineering To Order (ETO) or application is installed on the transmitter.

Figure B-6 Display menu – Off-line maintenance – Configuration

(1) Either Vol or GSV is displayed, depending on Volume Flow Type. See Section 8.2. (2) Displayed only if the petroleum measurement application is installed. (3) Displayed only if the enhanced density application is installed. (4) Displayed only if Auto Scroll is enabled. (5) Displayed only Off-Line Password is enabled.

Figure B-7 Display menu – Off-line maintenance – Zero

Device Profile Index Display Codes Menus

Figure B-8 Display menu – Off-line maintenance – Meter verification

(1) Either Unstable Flow or Unstable Drive Gain may be
displayed, indicating that the standard deviation of the flow or drive gain is outside limits. Check the process and retry the procedure. (2) Represents the percentage completion of the
x%(2) UNSTABLE FLOW(1)

Device Profile Index Display Codes Menus
This appendix documents the most commonly used portions of the Model 2400S DN transmitter’s device profile, including class/instance/attribute information and required codes.
The following object classes and instances are documented:
The following codes are documented:
For measurement unit codes used for process variables, see Section 6.3.
For complete documentation of the device profile, see the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile.

C.2 Analog Input Point Object (0x0A)
Table C-1 Analog Input Point Object (0x0A) – Instance 1 (mass flow)
Attrib ID Name Data type Service Mem Description Comments
3 Value REAL Get V Current value of mass flow process variable
Based on Attribute 8
8 Value data type USINT Get V Data type used to report mass flow process variable
100 Process total REAL Get Reset(1)
V Current value of mass total
101 Inventory total REAL Get Reset(2)
V Current value of mass inventory
102 Value engineering units
UINT Set NV Mass flow measurement unit
See Table 6-2 for unit codes.
103 Total engineering units
UINT Get V Mass total and inventory units
Transmitter automatically determines this based on Attribute 102. See Table C-12 for unit codes.
105 Cutoff REAL Set NV Value below which mass flow will be reported as 0
106 Meter factor REAL Set NV A multiplier to the calculated mass flow
107 Flow direction USINT Set NV Determines how flow direction affects reported flow rate and flow totals
108 Reset mass total USINT Set V Resets the mass total • 1 = Reset
109 Reset mass inventory
USINT Set V Resets the mass inventory
(1) Service code 0x4B.
(2) Service code 0x4C.

Table C-2 Analog Input Point Object (0x0A) – Instance 2 (liquid volume flow)
Device Profile Index Display Codes Menus
3 Value REAL Get V Current value of liquid volume flow process variable
Based on Attribute 8
8 Value data type USINT Get V Data type used to report volume flow process variable
100 Process total REAL Get Reset(1)
V Current value of liquid volume total
101 Inventory total REAL Get Reset(2)
V Current value of liquid volume inventory
102 Value engineering units
UINT Set NV Liquid volume flow measurement unit
See Table 6-3 for unit codes.
103 Total engineering units
UINT Get V Liquid volume total and inventory units
Transmitter automatically determines this based on Attribute 102. See Table C-13 for unit codes.
105 Cutoff REAL Set NV Value below which liquid volume flow will be reported as 0
106 Meter factor REAL Set NV A multiplier to the calculated liquid volume flow
108 Reset volume total
USINT Set V Resets the volume total • 1 = Reset
109 Reset volume inventory
USINT Set V Resets the volume inventory
(1) Service code 0x4B.
(2) Service code 0x4C.
Attrib ID Name Data type Service Mem Description Comments
3 Value REAL Get V Current value of density process variable
Based on Attribute 8
8 Value data type USINT Get V Data type used to report density process variable
102 Value engineering units
UINT Set NV Density measurement unit See Table 6-5 for unit codes.

Table C-3 Analog Input Point Object (0x0A) – Instance 3 (density) continued
104 Damping REAL Set NV Density damping value Unit = seconds
105 Cutoff REAL Set NV Value below which density will be reported as 0
106 Meter factor REAL Set NV A multiplier to the calculated density
Attrib ID Name Data type Service Mem Description Comments
3 Value REAL Get V Current value of temperature process variable
Based on Attribute 8
8 Value data type USINT Get V Data type used to report mass flow process variable
102 Value engineering units
UINT Set NV Temperature measurement unit
See Table 6-6 for unit codes.
104 Damping REAL Set NV Temperature damping value
Table C-5 Gas Standard Volume Object (0x64) – Instance 1
Attrib ID Name Data type Service Mem Description Comments
1 Gas standard volume flow
REAL Get V Current value of gas standard volume flow process variable
2 Gas standard volume total
REAL Get Reset(1) V Current value of gas standard volume total
3 Gas standard volume inventory
REAL Get Reset(2) V Current value of gas standard volume inventory
REAL Set NV Reference density of gas being measured
5 Gas standard volume flow units
UINT Set NV Gas standard volume flow measurement unit
See Table 6-4 for unit codes.
6 Gas standard volume total and inventory units
UINT Get V Gas standard volume total and inventory units
Transmitter automatically determines this based on Attribute 102. See Table C-14 for unit codes.
Unit = seconds · 4 Reference density

Table C-5 Gas Standard Volume Object (0x64) – Instance 1 continued
Device Profile Index Display Codes Menus
7 Enable gas standard volume
BOOL Set NV Enable or disable gas standard volume measurement(3)
8 Gas standard volume low flow cutoff
REAL Set NV Value below which gas standard volume flow will be reported as 0
9 Reset gas standard volume total
USINT Set V Resets the gas standard volume total
10 Reset gas standard volume inventory
USINT Set V Resets the gas standard volume inventory
(1) Service code 0x4B. (2) Service code 0x4C. (3) If gas standard volume measurement is enabled, liquid volume measurement is disabled, and vice versa.
Table C-6 Calibration Object (0x65) – Instance 1
Attrib ID Name Data type Service Mem Description Comments
1 Flow calibration factor
REAL Set NV 6-character flow calibration factor
2 Temperature coefficient for flow
REAL Set NV 4-character temperature coefficient
3 Zero time UINT Set V Duration of the zero calibration procedure
4 Zero standard deviation
REAL Get NV The standard deviation result of the zero calibration service
5 Zero offset REAL Set NV The offset result of the zero calibration service
6 Calibration failed value
REAL Get V The value of the calibration parameter if one of the calibration services fails
7 K1 REAL Set NV Density calibration constant 1
8 K2 REAL Set NV Density calibration constant 2
9 FD REAL Set NV Flowing density calibration constant
10 K3 REAL Set NV Density calibration constant 3
11 K4 REAL Set NV Density calibration constant 4
12 D1 REAL Set NV The line-condition density of D1 calibration service
Unit = g/cm3

Table C-6 Calibration Object (0x65) – Instance 1 continued
13 D2 REAL Set NV The line-condition density of D2 calibration service
Unit = g/cm3
14 FD REAL Set NV The line-condition density of FD calibration service
Unit = g/cm3
15 D3 REAL Set NV The line-condition density of D3 calibration service
Unit = g/cm3
16 D4 REAL Set NV The line-condition density of D4 calibration service
Unit = g/cm3
17 Density temperature coefficient
REAL Set NV The DT or TC calibration factor
18 FTG REAL Set NV T-Series: flow TG coefficient
19 FFQ REAL Set NV T-Series: flow FQ coefficient
20 DTG REAL Set NV T-Series: density TG coefficient
21 DFQ1 REAL Set NV T-Series: density FQ coefficient #1
22 DFQ2 REAL Set NV T-Series: density FQ coefficient #2
REAL Set NV Temperature offset
REAL Set NV Temperature slope
25 Enable temperature comp
BOOL Set NV Enable or disable temperature compensation
REAL Set V The external temperature value from output assembly instance 51 or 52
27 Enable pressure compensation
BOOL Set NV Enable or disable pressure compensation
REAL Set V The external pressure value from output assembly instance 50 or 52
29 Pressure units UINT Set NV Units used by external pressure input
See Table 6-7 for unit codes.
30 Pressure factor flow
REAL Set NV The pressure correction factor for flow
31 Pressure factor density
REAL Set NV The pressure correction factor for density
32 Flow cal pressure
REAL Set NV The flow calibration pressure

C.5 Diagnostics Object (0x66)
Table C-7 Diagnostics Object (0x66) – Instance 1
Device Profile Index Display Codes Menus
1 Fault behavior USINT Set NV Specifies the behavior of the process variables when the device is in a fault state
2 Fault timeout USINT Set NV The amount of time after a fault ocurrs before the fault behavior (Attribute 1) is implemented
3 Slug time REAL Set NV The amount of time the density is outside the slug low limit and slug high limit before a slug flow condition is declared
4 Slug low limit REAL Set NV The lower limit of a slug flow condition
Unit = g/cm3
5 Slug high limit REAL Set NV The upper limit of a slug flow condition
Unit = g/cm3
6 Discrete event index
USINT Set V The index of the discrete event that is being configured. There are 5 discrete events with the index starting at 0.
0, 1, 2, 3, 4
7 Discrete event type
USINT Set NV The type of the selected discrete event
8 Discrete event Setpoint A
REAL Set NV Setpoint A of the selected discrete event
9 Discrete event Setpoint B
REAL Set NV Setpoint B of the selected discrete event
10 Discrete event process variable
USINT Set NV The process variable on which the selected discrete event is defined
See Table C-15 for process variable codes. All codes are valid except for 52 (Input voltage).
11 Discrete event status
USINT Get V Each bit contains the status of the corresponding discrete event:
Unit = seconds · Unit = seconds

Table C-7 Diagnostics Object (0x66) – Instance 1 continued

Table C-7 Diagnostics Object (0x66) – Instance 1 continued
Device Profile Index Display Codes Menus

Table C-7 Diagnostics Object (0x66) – Instance 1 continued

Table C-7 Diagnostics Object (0x66) – Instance 1 continued
Device Profile Index Display Codes Menus
18 Alarm index USINT Set V Used to configure or read alarm severity, or to acknowledge alarms
See Table C-16 for alarm index codes.
19 Alarm severity USINT Set NV The alarm severity of the alarm that corresponds with the alarm index
20 Drive gain REAL Get V The drive gain %
21 Raw tube period REAL Get V The tube frequency Unit = Hz
22 Live zero (mass flow)
REAL Get V The unfiltered value of mass flow
Configured mass flow units
23 LPO voltage REAL Get V The left pickoff voltage Unit = volts
24 RPO voltage REAL Get V The right pickoff voltage Unit = volts
REAL Get V The temperature on the board
26 Maximum electronics temperature
REAL Get V The maximum temperature of the electronics
27 Minimum electronics temperature
REAL Get V The minimum temperature of the electronics
28 Average electronics temperature
REAL Get V The average temperature of the electronics
29 Maximum sensor temperature
REAL Get V The maximum temperature of the sensor
30 Minimum sensor temperature
REAL Get V The minimum temperature of the sensor
31 Average sensor temperature
REAL Get V The average temperature of the sensor

Table C-7 Diagnostics Object (0x66) – Instance 1 continued
32 9-wire cable RTD resistance
REAL Get V The resistance of the 9-wire cable Unit = ohms
33 Meter RTD resistance
REAL Get V The resistance of the meter RTD
34 Number of power cycles
UINT Get V The number of transmitter power cycles
35 Power on time Unsigned 32 Get Reset(1) V The cumlative amount of time the tranmitter has been on since the last reset (Class 0x01, Attribute 0x05)
Seconds since last reset
36 Line RTD REAL Get V The resistance of the process line RTD
37 Actual target amplitude
REAL Get V The amplitude the transmitter is attempting to drive the sensor
38 Input voltage REAL Get V The number of volts on the power input terminals
39 Drive current REAL Get V The drive current Unit = milliamps

Table C-7 Diagnostics Object (0x66) – Instance 1 continued
Device Profile Index Display Codes Menus
42 Alarm status USINT Set V The status of the alarm selected in Attribute 18. Write 0x00 to acknowledge the alarm selected in Attribute 18.
43 Alarm count UINT Get V The number of inactive-to- active transitions of the alarm selected in Attribute 18
44 Alarm last posted
Unsigned 32 Get V The number of seconds since the last reset that the alarm selected in Attribute 18 was posted
Seconds since last reset
45 Alarm last cleared
Unsigned 32 Get V The number of seconds since the last reset that the alarm selected in Attribute 18 was cleared
Seconds since last reset
46 Alarm history index
USINT Set V The entry in the alarm history log
47 Alarm history alarm number
USINT Get V The alarm number that corresponds to the alarm history entry selected in Attribute 45
1 = A001, 2 = A002, etc.
48 Alarm history alarm status changed
USINT Get V The alarm status change that corresponds to the alarm history entry selected in Attribute 45
49 Alarm history alarm status changed timestamp
Unsigned 32 Get V The timestamp of the alarm status change that corresponds to the alarm history entry selected in Attribute 45
Seconds since last reset
54 Meter verification algorithm state
USINT Get V The current state of the meter verification routine

Table C-7 Diagnostics Object (0x66) – Instance 1 continued
55 Meter verification abort code
USINT Get V The reason the meter verification routine aborted
56 Meter verification algorithm state at abort
USINT Get V The state of the meter verification routine when it aborted
57 Meter verification percent complete
USINT Get V The progress of the meter verification routine
USINT Set NV The state of the outputs when the meter verification routine is running
58 Meter verification outputs state
59 Meter verification stiffness limit
REAL Set NV The setpoint of the stiffness limit. Represents percentage.
60 Meter verification validation counter
UINT Get NV Indicates the number of times the meter verification routine has successfully completed
61 Meter verification inlet stiffness out of limits
USINT Get V Is the inlet stiffness out of limits?
62 Meter verification outlet stiffness out of limits
USINT Get V Is the outlet stiffness out of limits?
% · Unitless

Table C-7 Diagnostics Object (0x66) – Instance 1 continued
Device Profile Index Display Codes Menus
63 Meter verification – current inlet stiffness, mean
REAL Get NV The current inlet stiffness calculated as a mean
64 Meter verification – current outlet stiffness, mean
REAL Get NV The current outlet stiffness calculated as a mean
65 Meter verification – current damping, mean
REAL Get NV The current damping calculated as a mean
66 Meter verification – current inlet mass, mean
REAL Get NV The current inlet mass calculated as a mean
67 Meter verification – current outlet mass, mean
REAL Get NV The current outlet mass calculated as a mean
68 Meter verification – current inlet stiffness, SD
REAL Get NV The current inlet stiffness calculated as a standard deviation
69 Meter verification – current outlet stiffness, SD
REAL Get NV The current outlet stiffness calculated as a standard deviation
70 Meter verification – current damping, SD
REAL Get NV The current damping calculated as a standard deviation
71 Meter verification – current inlet mass, SD
REAL Get NV The current inlet mass calculated as a standard deviation
72 Meter verification – current outlet mass, SD
REAL Get NV The current outlet mass calculated as a standard deviation
73 Meter verification – current inlet stiffness, factory cal of air, mean
REAL Get NV The inlet stiffness calculated as a mean during factory calibration of air
74 Meter verification – current outlet stiffness, factory cal of air, mean
REAL Get NV The outlet stiffness calculated as a mean during factory calibration of air
75 Meter verification – current damping, factory cal of air, mean
REAL Get NV The damping calculated as a mean during factory calibration of air

Table C-7 Diagnostics Object (0x66) – Instance 1 continued
76 Meter verification – current inlet mass, factory cal of air, mean
REAL Get NV The inlet mass calculated as a mean during factory calibration of air
77 Meter verification – current outlet mass, factory cal of air, mean
REAL Get NV The outlet mass calculated as a mean during factory calibration of air
78 Meter verification – current inlet stiffness, factory cal of water, mean
REAL Get NV The inlet stiffness calculated as a mean during factory calibration of water
79 Meter verification – current outlet stiffness, factory cal of water, mean
REAL Get NV The outlet stiffness calculated as a mean during factory calibration of water
80 Meter verification – current damping, factory cal of water, mean
REAL Get NV The damping calculated as a mean during factory calibration of water
81 Meter verification – current inlet mass, factory cal of water, mean
REAL Get NV The inlet mass calculated as a mean during factory calibration of water
82 Meter verification – current outlet mass, factory cal of water, mean
REAL Get NV The outlet mass calculated as a mean during factory calibration of water

Table C-7 Diagnostics Object (0x66) – Instance 1 continued
Device Profile Index Display Codes Menus
83 Factory flow signal offset at zero flow
REAL Get NV The flow signal offset at zero flow when calibrated at the factory
84 Discrete event action code
USINT Set V The action that will be performed by the event specified in Attribute 85
85 Discrete event assignment
USINT Set NV The discrete event that is assigned to the action referenced in Attribute 84
(1) Service code 0x4D.
Table C-8 Sensor Information Object (0x67) – Instance 1
Attrib ID Name Data type Service Mem Description Comments
1 Sensor serial number
UDINT Set NV The serial number of the sensor
2 Sensor type SHORT STRING
Get NV A string that represents the type of sensor
For example, F200, CMF025
3 Sensor type code
Unit = microseconds

Table C-8 Sensor Information Object (0x67) – Instance 1 continued
4 Sensor material USINT Set NV The material of the sensor’s case
5 Liner material USINT Set NV The material of the sensor’s liner
6 Flange type USINT Set NV The type of process connection on the sensor
Table C-9 Local Display Object (0x68) – Instance 1
Attrib ID Name Data type Service Mem Description Comments
1 Scroll rate USINT Set NV The rate at which each variable will be displayed
2 Backlight control BOOL Set NV Whether the backlight is on or off
USINT Set NV The brightness of the backlight
0 (off) to 63 (full on)
4 Display variable 1 USINT Set V Displays the variable associated with the code on the local display
See Table C-15 for codes. All codes are valid except for 251 (None).
Unit = seconds · 3 Backlight intensity

Table C-9 Local Display Object (0x68) – Instance 1 continued
Device Profile Index Display Codes Menus
5 Display variable 2
See Table C-15 for codes. All codes are valid. 6 Display variable 3
USINT Set NV Displays the variable associated with the code on the local display
7 Display variable 4
8 Display variable 5
9 Display variable 6
10 Display variable 7
11 Display variable 8
12 Display variable 9
13 Display variable 10
14 Display variable 11
15 Display variable 12
16 Display variable 13
17 Display variable 14
18 Display variable 15
19 Enable start/stop totals
BOOL Set NV Enable or disable the ability to start and stop totals from the local display
20 Enable reset totals
BOOL Set NV Enable or disable the ability to reset totals from the local display
21 Enable auto scroll
BOOL Set NV Enable or disable the auto scroll feature. The scroll rate is set using Attribute 1.
22 Enable offline menu
BOOL Set NV Enable or disable the offline menu
23 Enable alarm menu
BOOL Set NV Enable or disable the alarm menu
24 Enable ACK All alarms
BOOL Set NV Enable or disable the ability to acknowledge all the alarms at once
25 Enable IrDA write protect
BOOL Set NV Enable or disable the write-protect feature on the IrDA port

Table C-9 Local Display Object (0x68) – Instance 1 continued
26 Enable offline password
BOOL Set NV Enable or disable the password requirement to access the offline menu
27 Offline password UINT Set NV The offline password for entering the offline menu
28 Update period UINT Set NV The period in which the display is updated
29 Process variable index
USINT Set V The process variable in which the precision will be set in Attribute 30
See Table C-15 for codes.
30 Process variable precision
USINT Set NV The number of digits displayed to the right of the decimal point for the process variable selected with Attribute 29
31 Language USINT Set NV Display language selection
32 Enable IrDA port USINT Set NV Enable or disable the IrDA port
(1) Not available in this release.
Table C-10 API Object (0x69) – Instance 1
Attrib ID Name Data type Service Mem Description Comments
1 Temperature corrected density
REAL Get V Current value
2 Temperature corrected (standard) volume flow
REAL Get V Current value
3 Temperature corrected (standard) volume total
REAL Get Reset(1) V Current value
4 Temperature corrected (standard) volume inventory
REAL Get Reset(2) V Current value
5 Batch weighted average density
REAL Get V Current value
6 Batch weighted average temperature
REAL Get V Current value
Unit = milliseconds

Table C-10 API Object (0x69) – Instance 1 continued
Attrib ID Name Data type Service Mem Description Comments
Device Profile Index Display Codes Menus
7 CTL REAL Get V Current value
8 API reference temperature
REAL Set NV The reference temperature to use in the API calculations
9 API thermal expansion coefficient
REAL Set NV The thermal expansion coefficient to use in the API calculations
10 API 2540 CTL table type
USINT Set NV The table type to use in the API calculations
11 Reset API reference volume total
USINT Set V Resets the API reference volume total
12 Reset API reference volume inventory
USINT Set V Resets the API reference volume inventory
(1) Service code 0x4B. (2) Service code 0x4C.
Table C-11 Enhanced Density Object (0x6A) – Instance 1
Attrib ID Name Data type Service Mem Description Comments
1 Density at reference
REAL Get V Current value
2 Density (fixed SG units)
REAL Get V Current value
3 Standard volume flow rate
REAL Get V Current value
4 Standard volume total
REAL Get Reset(1) V Current value
5 Standard volume inventory
REAL Get Reset(2) V Current value
6 Net mass flow rate
REAL Get V Current value
7 Net mass flow total
REAL Get Reset(3) V Current value

Table C-11 Enhanced Density Object (0x6A) – Instance 1 continued
Attrib ID Name Data type Service Mem Description Comments
8 Net mass flow inventory
REAL Get Reset(4) V Current value
9 Net volume flow rate
REAL Get V Current value
10 Net volume flow total
REAL Get Reset(5) V Current value
11 Net volume flow inventory
REAL Get Reset(6) V Current value
12 Concentration REAL Get V Current value
13 Density (fixed Baume units)
REAL Get V Current value
16 Active calculation curve
USINT Set NV The number of the curve that is currently active
38 Curven ASCII string
Set NV The name of the active curve
39 Enable enhanced density application
47 Reset standard volume total
USINT Set V Resets the standard volume total
48 Reset standard volume inventory
USINT Set V Resets the standard volume inventory
49 Reset net mass total
USINT Set V Resets the net mass total • 1 = Reset
24 characters maximum

Table C-11 Enhanced Density Object (0x6A) – Instance 1 continued
Attrib ID Name Data type Service Mem Description Comments
Device Profile Index Display Codes Menus
50 Reset net mass inventory
USINT Set V Resets the net mass inventory
51 Reset net volume total
USINT Set V Resets the net volume total
52 Reset net volume inventory
USINT Set V Resets the net volume inventory
(1) Service code 0x4B. (2) Service code 0x4F. (3) Service code 0x4C. (4) Service code 0x50. (5) Service code 0x4D. (6) Service code 0x51.
Table C-12 Mass totalizer and mass inventory measurement unit codes
Code Description
0x2506 Short ton (2000 pounds)
0x080E Long ton (2240 pounds)
Code Description
0x0857 Beer barrel(2)
(1) Unit based on oil barrels (42 U.S gallons).
(2) Unit based on beer barrels (31 U.S gallons).

Table C-14 Gas standard volume totalizer and gas standard volume inventory
measurement unit codes
Code Description
0x0844 Standard cubic feet
0x0845 Normal cubic meters
0x0846 Standard cubic meters
Table C-15 Process variable codes
Code Description
0 Mass flow rate
5 Volume flow rate

Table C-15 Process variable codes continued
Code Description
Device Profile Index Display Codes Menus
49 Left pickoff amplitude
50 Right pickoff amplitude
62 Gas standard volume flow rate
63 Gas standard volume total
64 Gas standard volume inventory
Table C-16 Alarm index codes
Code Description
1 (E)EPROM checksum error (CP)
2 RAM error (CP)
4 Temperature sensor failure
9 Transmitter initializing/warming up
11 Zero too low
12 Zero too high
13 Zero too noisy
16 Line RTD Temperature out-of-range
17 Meter RTD temperature out-of-range
20 Incorrect sensor type (K1)
21 Invalid sensor type
22 NV error (CP)
23 NV error (CP)
24 NV error (CP)
25 Boot failure (core processor)

Table C-16 Alarm index codes continued
Code Description
26 Sensor/transmitter communications error
28 Core processor exception
29 Core processor communications error
30 Invalid board type
32 Meter verification fault alarm
33 Tubes not full
43 Data loss possible
44 Calibration in progress
71 Meter verification info alarm
72 Simulation mode active

Device Profile Index Display Codes Menus
This appendix provides information on the codes and abbreviations used on the transmitter display.
Table D-1 lists and defines the codes and abbreviations that are used for display variables (see Section 8.9.5 for information on configuring display variables).
Table D-2 lists and defines the codes and abbreviations that are used in the off-line menu.
Code or abbreviation Definition Comment or reference
BRD T Board temperature
EXT P External pressure
EXT T External temperature
GSV F Gas standard volume flow
GSV I Gas standard volume flow inventory
LPO_A Left pickoff amplitude
LZERO Live zero flow
MTR T Case temperature
NET M Net mass flow rate Enhanced density application only
NET V Net volume flow rate Enhanced density application only
NETMI Net mass inventory Enhanced density application only
NETVI Net volume inventory Enhanced density application only
PWRIN Input voltage Refers to power input to the core processor

Table D-1 Display codes used for display variables
Code or abbreviation Definition Comment or reference
RDENS Density at reference temperature
Enhanced density application only
RPO A Right pickoff amplitude
SGU Specific gravity units
STD V Standard volume flow rate Enhanced density application only
STD V Standard volume flow rate Enhanced density application only
STDVI Standard volume inventory Enhanced density application only
Petroleum measurement application only
Petroleum measurement application only
TCORR Temperature-corrected total Petroleum measurement application only
Petroleum measurement application only
TUBEF Raw tube frequency
Code or abbreviation Definition Comment or reference
ACK Display Ack All menu
ACK ALARM Acknowledge alarm
ACK ALL Acknowledge all
ACT Action Action assigned to the discrete input or to a discrete event
BKLT, B LIGHT Display backlight
CH A Channel A
CH B Channel B
CHANGE PASSW Change password Change the password required for access to display functions
CUR Z Current zero
CUSTODY XFER Custody transfer
DRIVE%, DGAIN Drive gain
DISBL Disable Select to disable

Table D-2 Display codes used in off-line menu
Code or abbreviation Definition Comment or reference
Ex Event x Refers to Event 1 or Event 2 when setting the setpoint.
Device Profile Index Display Codes Menus
ENABL Enable Select to enable
EVNTx Event x
FAC Z Factory zero
FCF Flow calibration factor
FLSWT, FL SW Flow switch
GSV Gas standard volume
GSV T Gas standard volume total
MTR F Meter factor
OFF-LINE MAINT Off-line maintenance menu
OFFLN Display off-line menu
r. Revision
SrC Source Variable assignment for outputs
VOL Volume or volume flow







cutoff 58 measurement unit
configuration 30 list 30
W Wiring problems 113
Z Zero calibration 99
failure 114 procedure 100
Volume flow, liquid



For the latest Micro Motion product specifications, view the PRODUCTS section of our web site at www.micromotion.com
Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T +1 303-527-5200 +1 800-522-6277 F +1 303-530-8459
Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556
Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T +65 6777-8211 F +65 6770-8003
Micro Motion United Kingdom Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U.K. T +44 0870 240 1978 F +44 0800 966 181
Micro Motion Japan Emerson Process Management 1-2-5, Higashi Shinagawa Shinagawa-ku Tokyo 140-0002 Japan T +81 3 5769-6803 F +81 3 5769-6844