Micro Motion ®
F-Series Sensor
Installation Manual



This manual describes how to install a Micro Motion® F-Series sensor. The following information is provided in this manual:
Customer service page 1
Definitions page 2
European installations page 2
Installation options page 2
Determining a Location page 8
Orienting the Sensor page 12
Mounting the Sensor page 13
Wiring page 15
Grounding page 19
Pressure Ratings at High Temperatures page 19
Purge Fittings page 20
Return Policy page 22
For technical assistance, phone the Micro Motion Customer Service department:
- In Japan, phone 3 5769-6803
- In other locations, phone +65 6777-8211 (Singapore)
- In the U.K., phone 0870 240 1978 (toll-free)
- In other locations, phone +31 (0) 318 495 555 (The Netherlands)
Customers outside the U.S.A. can also email Micro Motion customer service at International.MMISupport@EmersonProcess.com.
Refer to the transmitter configuration and use manual for troubleshooting help.
©2007, Micro Motion, Inc. All rights reserved. ELITE and ProLink are registered trademarks, and MVD and MVD Direct Connect are trademarks of Micro Motion, Inc., Boulder, Colorado. Micro Motion is a registered trade name of Micro Motion, Inc., Boulder, Colorado. The Micro Motion and Emerson logos are trademarks and service marks of Emerson Electric Co. All other trademarks are property of their respective owners.
Installation steps · Additional information

Full product specifications can be found in the F-Series Product Data Sheet, which is available from the Micro Motion web site at www.micromotion.com.
The term MVD™ transmitter refers to the following transmitter models:
The term high-temperature sensor refers to the following F-Series models:
The term extreme high-temperature sensor refer to the following F-Series models:
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EC declaration of conformity for directives that apply to this product.
The EC declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions are available on the internet at www.micromotion.com/atex or through your local Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive can be found on the internet at www.micromotion.com/library.
The F-Series sensor makes up one part of a Coriolis flowmeter. The other part is a transmitter. F-Series sensors are available with the following electronics interfaces:


Sensor with purge fittings

Calibration tag · Process connection


Sensor with purge fittings

Model 1700 or 2700 transmitter
Process connection


Sensor with purge fittings

Process connection


Sensor with purge fittings



Transmitter, core processor, or junction box (core processor shown)
Flow direction arrow

Choose a location for the sensor based on the requirements described in this section. The following general guidelines can help you select an appropriate location for the sensor.
For optimal performance, the sensor tubes should remain full of process fluid.
Make sure the hazardous area specified on the sensor approval tag is suitable for the environment in which the sensor is installed (see Figures 1–5). For installation in an area that requires intrinsic safety, refer to the appropriate Micro Motion approval documentation, shipped with the sensor or available from the Micro Motion web site at www.micromotion.com.
Improper installation in a hazardous area can cause an explosion. When installing in a hazardous area, refer to Micro Motion approvals instructions, shipped with the product or available from the Micro Motion web site. For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
The ambient and process temperature limits of the sensor are shown in Figures 6 and 7:
Determining a Location

Figure 6 F-Series sensor ambient and process temperature limits (all models except high-temperature and extreme high-temperature models)
Mount electronics remotely; use junction box
Ambient temperature of core processor or
transmitter in °F (°C)
Mount electronics remotely; use junction box
Maximum process temperature in °F (°C)
Notes: 1. When ambient temperature is below –40 °F (–40 °C), a core processor must be heated to bring its local ambient temperature to between –40 °F (–40 °C) and +140 °F (+60 °C). Long-term storage of electronics at ambient temperatures below –40 °F (–40 °C) is not recommended. 2. Temperature limits may be further restricted by hazardous area approvals. 3. The temperature extender option allows the sensor case to be insulated without covering the transmitter, core processor, or junction box, but does not affect temperature ratings.

Figure 7 F-Series sensor ambient and process temperature limits (high-temperature and extreme high-temperature models)
Ambient temperature of core processor or
transmitter in °F (°C)
Maximum process temperature in °F (°C)
Hazardous area approvals may impose additional limits on ambient and process temperature. For the ATEX “T” rating, refer to the ATEX documentation shipped with the sensor or available on the Micro Motion web site at www.micromotion.com. IECEx and NEPSI approvals also use the ATEX “T” rating. UL and CSA ambient temperature limits are listed in Table 1.
Sensor models Electronics °F °C UL F025, F050, F100, F200 Junction box or IFT9701 transmitter
–4 to +104 –20 to +40
CSA All models except high-temperature or extreme high-temperature models
Junction box IFT9701 transmitter
+140 maximum +60 maximum
Core processor or Model 1700/2700 transmitter –40 to +140 –40 to +60
High-temperature and extreme high-temperature models
All options –40 to +140 –40 to +60
Determining a Location · 662 (350) · 800 (427)

Maximum wiring distances
If the transmitter is mounted remotely from the sensor, the maximum distance between the sensor and transmitter depends on cable type. See Table 2.
For high-temperature and extreme high-temperature F-Series sensors, note the following:
Cable type Wire size Maximum length
Micro Motion 9-wire to an MVD transmitter or core processor
Not applicable 60 feet (20 meters)
Micro Motion 9-wire to all other transmitters
Not applicable 1000 feet (300 meters)
Micro Motion 4-wire Not applicable 1000 feet (300 meters)
Power wires (VDC) 22 AWG (0,35 mm2) 300 feet (90 meters)
20 AWG (0,5 mm2) 500 feet (150 meters)
18 AWG (0,8 mm2) 1000 feet (300 meters)
Signal wires (RS-485) 22 AWG (0,35 mm2) or larger 1000 feet (300 meters)
Micro Motion sensors do not require a straight run of pipe upstream or downstream.
After the sensor and transmitter have been installed, you must perform the zeroing procedure. During the zeroing procedure, flow through the sensor must be halted and the sensor tubes must be completely full of process fluid. A shutoff valve, downstream from the sensor, is recommended to halt flow during the zeroing procedure. For more information about zeroing, refer to the instruction manual shipped with the transmitter.
User-supplied 4-wire

The sensor will function properly in any orientation if the sensor tubes remain filled with process fluid. Micro Motion recommends orienting F-Series sensors as shown in Figure 8.
Slurries and self-draining applications
Gases Tubes up Horizontal pipeline
Liquids Tubes down Horizontal pipeline
Flag mount Vertical pipeline

The F-Series sensor is self-draining when it is installed in a vertical pipeline (see Figure 8). For CIP applications it is recommended to use the generally accepted flow velocity of at least 1.5 m/sec for cleaning the sensor.
The sensor has a flow direction arrow (see Figures 1–5) to help you configure the transmitter for flow direction. If possible, install the sensor so that the flow direction arrow matches actual process flow.
If the sensor is installed in a vertical pipeline, liquids and slurries should flow upward through the sensor. Gases may flow upward or downward.
Orienting the Sensor

Step 3 Mounting the Sensor
Use your common practices to minimize torque and bending load on process connections. Figure 9 illustrates how to mount the sensor. To reduce the risk of condensation or excessive moisture, the conduit opening should not point upward (if possible). The conduit opening of the junction box or core processor can be rotated freely to facilitate wiring.


High-temperature and extreme high-temperature F-Series sensors come with a 32″ (812 mm) pre-installed flexible conduit. This conduit is required for agency approval.
A factory-supplied Model 1700/2700 transmitter, core processor, or junction box is connected to the end of the flexible conduit. Mount the electronics to a wall or instrument pole using the supplied bracket (Figure 10).


Mounting bracket (wall mount)
Mounting bracket (instrument pole)
Use 4 × 5/16″ (8 mm) bolts to secure the bracket to a wall
Use 2 × 5/16″ (8 mm) U-bolts to secure the bracket to an instrument pole

Secure the electronics to the bracket with the 4 #10 screws.
Mounting the Sensor

Step 4 Wiring
Improperly sealed housings can expose electronics to moisture, which can cause measurement error or flowmeter failure. Inspect and grease all gaskets and O-rings. Fully close and tighten all housing covers and conduit openings.
The sensor has one of the following electronics configurations:
Follow the steps below to connect the 4-wire cable between the core processor and the transmitter.
- Prepare the cable and apply shielded heat shrink as described below. The shielded heat shrink provides a shield termination suitable for use in the gland when using cable whose shield consists of foil and not a braid. Proceed to step 2 of the wiring procedure, below.
- With armored cable, where the shield consists of braid, prepare the cable as described below, but do not apply heat shrink. Proceed to step 2 of the wiring procedure, below.


4 1/2 in (114 mm)
3/4 in (19 mm)
Gland nut Gland clamping
7/8 in (22 mm) 7/8 in (22 mm)
a. Strip 4 1/2 inches (114 mm) of cable jacket.
b. Remove the clear wrap that is inside the cable jacket, and remove the filler material between the
wires.
c. Remove the foil shield that is around the insulated wires, leaving 3/4 inch (19 mm) of foil or braid
and drain wires exposed, and separate the wires.
d. Wrap the shield drain wire(s) around the exposed foil twice. Cut off the excess wire.

e. Place the shielded heat shrink over the exposed shield drain wire(s). The tubing should completely
cover the drain wires.
f. Without burning the cable, apply heat (250 °F or 120 °C) to shrink the tubing.

Figure 13 Applying the heat shrink

g. Position gland clamping insert so the interior end is flush with the heat shrink.
h. Fold the cloth shield or braid and drain wires over the clamping insert and approximately 1/8 inch
(3 mm) past the O-ring.

i. Install the gland body into the core processor housing conduit opening.



Terminal 1 Power supply + (Red wire)
Terminal 4 RS-485B (Green wire)
Terminal 3 RS-485A (White wire)
Terminal 2 Power supply – (Black wire)
Follow the steps below to connect the 9-wire cable between the sensor and the transmitter or core processor.
Wiring

Step 5 Grounding
The sensor can be grounded via the piping if the joints in the pipeline are ground-bonded. If the sensor is not grounded via the piping, connect a ground wire to the internal or external grounding screw, which is located on the core processor housing or junction box.
Improper grounding can cause measurement error. Ground the flowmeter to earth, or follow ground network requirements for the facility.
If national standards are not in effect, follow these guidelines to ground the sensor:
Table 3 shows the pressure ratings for F-Series sensors at high temperatures.
EN-1092 flange Pressure rating (bar)
≤50°C 100°C 150°C 200°C 250°C 300°C 350°C 400°C 427°C
40 11 weld neck
B1 & D Flange, EN 1092-1 PN40 Forms B1 & D
316 40.0 40.0 36.3 33.7 31.8 29.7 28.5 27.4 26.9 1
100 11 weld neck
B2 & D Flange, EN 1092-1 PN100 Form B2 & D
316 100.0 100.0 90.9 84.2 79.5 74.2 71.4 68.5 67.3 1
100/ 160 11 weld neck
316 100.0 100.0 90.9 84.2 79.5 74.2 71.4 68.5 67.3 2
1092-1 PN100/160 Form B2 (316)
160.0 160.0 145.5 134.8 127.2 118.8 114.2 109.7 107.8 3
40 32/02 lap joint
B1 Flange, EN 1092-1 PN40 Form B1(lap
304 flange 40.0 34.4 30.8 28.0 26.0 24.1 23.0 22.0 21.4 4

If the sensor has purge fittings, they should remain sealed at all times. After a purge plug has been removed, the sensor case should be purged with argon or nitrogen and resealed.
Purging the case protects internal components. The sensor is purged of all oxygen and sealed at the factory. If the purge plugs are never removed, it is not necessary to purge or re-seal the sensor. For more information, contact Micro Motion Customer Service.
If a purge plug is removed from the sensor case, it will be necessary to repurge the case.
Removing a purge plug compromises the secondary containment of the sensor and could expose the user to process fluid. Take all necessary precautions when removing purge plugs.
Improper pressurization of the sensor case could result in personal injury. Removing a purge plug will require the sensor case to be repurged with a dry inert gas. Follow all instructions provided in the case purging procedure.
Read all instructions before performing the case purging procedure. It is not necessary to perform this procedure unless a purge plug has been removed.
Performing the purge procedure while the flowmeter is operating could affect measurement accuracy, resulting in inaccurate flow signals. Before performing the case purging procedure, shut down the process, or set control devices for manual operation.
Purge Fittings

Table 4 Time required to purge F-Series sensor cases
Purge rate ft3/hr (l/hr)
F050 20 (566) 4 1/2
F100 20 (566) 6
F200 20 (566) 15
F300 20 (566) 25
(1) If purge lines are being used, increase purge time to fill the additional volume.
Time(1) minutes · Sensor model

Micro Motion procedures must be followed when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Failure to follow Micro Motion procedures will result in your equipment being refused delivery.
Information on return procedures and forms is available on our web support system at www.micromotion.com, or by phoning the Micro Motion Customer Service department.
Only equipment that has not been removed from the original shipping package will be considered new and unused. New and unused equipment requires a completed Return Materials Authorization form.
All equipment that is not classified as new and unused is considered used. This equipment must be completely decontaminated and cleaned before being returned.
Used equipment must be accompanied by a completed Return Materials Authorization form and a Decontamination Statement for all process fluids that have been in contact with the equipment. If a Decontamination Statement cannot be completed (e.g., for food-grade process fluids), you must include a statement certifying decontamination and documenting all foreign substances that have come in contact with the equipment.
Return Policy



© 2007 Micro Motion, Inc. All rights reserved. P/N 20002298, Rev. C *20002298*
For the latest Micro Motion product specifications, view the PRODUCTS section of our web site at www.micromotion.com
Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T +1 303-527-5200 +1 800-522-6277 F +1 303-530-8459
Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556
Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T +65 6777-8211 F +65 6770-8003
Micro Motion United Kingdom Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U.K. T +44 0870 240 1978 F +44 0800 966 181