Model 400 and 400 VP
ENDURANCE
™ Conductivity Sensors
For additional information, please refer visit our website at
www.emersonprocess.com/raihome/liquid/.
SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsi- bility of the user.
Keep 1/4 in. (6 mm) clearance between electrodes and piping. The electrodes must be completely submerged in the process liquid, i.e., to the level of the threaded connection. See Figures 1 - 6 for recommended orientation and installa- tion. Models 400 and 400VP sensors with 0.1 and 1.0/cm cell constants can be installed in 3/4-inch pipe tees. All Model 400 and 400VP sensors can be installed in 1-inch tees with a 3/4-inch bushing.
If the sensor is installed in a pipe tee or flow cell with the sample draining to open atmosphere, bubbles may accumulate on the electrodes. Trapped bubbles will cause errors. As bubbles accumulate, the conductivity reading normally drifts down. In the plastic flow cell, bubbles are readily visible. To control bubble formation, apply a small amount of back pres- sure to the flow cell or pipe tee.


FIGURE 1. Sensor Orientation FIGURE 2. Direct Insertion in a Pipe


FIGURE 3. Insertion in a Pipe Tee
FIGURE 4. Insertion in a Pipe Tee


FIGURE 5. Insertion in a Pipe Tee
FIGURE 6. Insertion in a Flow Cell (PN 24091-02)
° °


FIGURE 7. Model 54eC Wiring FIGURE 8. Model 56 and 1056 Wiring

RTN
WHITE
RCV B
SENSE
WHITE/RED
RCV A
RTD IN
RED
CLEAR
RSHLD
SHLD CLEAR
GRAY
DRVB
ORANGE
DRVA
CLEAR
DSHLD
FIGURE 9. Model 1066 Wiring


FIGURE 10. Model Xmt-C-10 Wiring (Panel)
FIGURE 11. Model Xmt-C-11 Wiring (Pipe or Wall)


FIGURE 12. Model 5081C Wiring
FIGURE 13. Model 6081-C Wiring



Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol (rubbing alcohol) can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.
ENDURANCE conductivity sensors are calibrated at the factory and do not need calibration when first placed in service. Simply enter the cell constant printed on the label into the analyzer.
After a period of service, the sensor may require calibration. The sensor can be calibrated against a solution having known conductivity or against a referee meter and sensor. If using a standard solution, choose one having conduc- tivity in the recommended operating for the sensor cell constant. Refer to the analyzer manual or product data sheet for recommended ranges. Do not use standard solutions having conductivity less than about 100 uS/cm. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25°C). Because 0.01/cm sensors must be calibrated in low conductivity solutions, they are best calibrated against a referee meter and sensor in a closed system.
For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website.

FIGURE 16. Checking Temperature Element Disconnect leads and measure resistance shown. The meas- ured resistance should be close to the value in the table.
FIGURE 17. Checking Continuity and Leakage Disconnect electrode leads and measure resistance and continuity as shown. Sensor must be dry when checking resistance between electrode leads.



The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2010
Emerson Process Management
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Off-scale reading | Wiring is wrong. Temperature element is open or shorted. Sensor is not in process stream. Variopol cable is not properly seated. Sensor has failed. | Verify wiring. | p.6 |
| Off-scale reading | Check temperature element for open or short circuits. See Figure 16. | p.6 | |
| Off-scale reading | Be sure sensor is completely submerged in process stream. | p.6 | |
| Off-scale reading | Loosen connector and reseat. | p.6 | |
| Off-scale reading | Perform isolation checks. See Figure 17. | p.6 | |
| Noisy reading | Sensor is improperly installed in process stream. Variopol cable is not properly seated. | Be sure sensor is completely submerged in process stream. | p.6 |
| Noisy reading | Loosen connector and reseat. | p.6 | |
| Reading seems wrong (lower or higher than expected) | Bubbles trapped in sensor. Wrong temperature correction algorithm. Wrong cell constant. | Be sure sensor is properly oriented in pipe or flow cell. See Figure 1. Apply back pressure to flow cell. | p.6 |
| Reading seems wrong (lower or higher than expected) | Check that temperature correction is appropriate for the sample. See analyzer manual for more information. | p.6 | |
| Reading seems wrong (lower or higher than expected) | Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. | p.6 | |
| Sluggish response | Electrodes are fouled. Sensor is installed in dead area in piping. | Clean electrodes. | p.6 |
| Sluggish response | Move sensor to a location more representative of the process liquid. | p.6 |