Instruction Sheet PN 51A-400/rev.R December 2010

Model 400 and 400 VP

ENDURANCE

™ Conductivity Sensors

For additional information, please refer visit our website at

www.emersonprocess.com/raihome/liquid/.

CAUTION

SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING

WARNING

Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!

The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsi- bility of the user.

SENSOR SPECIFICATIONS

SPECIFICATIONSMODEL 400MODEL 400 VPFLOW CELL PN 24091-02 Wetted Materialstitanium, PEEK (glass-filled), 316 SST, EPDMtitanium, PEEK (glass-filled) , 316 SST, EPDMpolycarbonate, polyester, 316 SST, silicone Temperature Rangestnd: 32-221°F (0-105°C) -60: 32-392°F (0-200°C) (requires sensor-mounted junction box)32-221°F (0-105°C)32-158°F (0-70°C) Maximum Pressure250 psig (1825 kPa abs)250 psig (1825 kPa abs)90 psig (722 kPa abs) Vacuum Service7.3 psia (51 kPa abs)7.3 psia (51 kPa abs)

INSTALLATION

Keep 1/4 in. (6 mm) clearance between electrodes and piping. The electrodes must be completely submerged in the process liquid, i.e., to the level of the threaded connection. See Figures 1 - 6 for recommended orientation and installa- tion. Models 400 and 400VP sensors with 0.1 and 1.0/cm cell constants can be installed in 3/4-inch pipe tees. All Model 400 and 400VP sensors can be installed in 1-inch tees with a 3/4-inch bushing.

If the sensor is installed in a pipe tee or flow cell with the sample draining to open atmosphere, bubbles may accumulate on the electrodes. Trapped bubbles will cause errors. As bubbles accumulate, the conductivity reading normally drifts down. In the plastic flow cell, bubbles are readily visible. To control bubble formation, apply a small amount of back pres- sure to the flow cell or pipe tee.

Page 2 of the Emerson 400 user manual — Instruction Sheet PN 51A-400/rev.R December 2010
Page 2 · click to enlarge
Page 2
Emerson 400 — Instruction Sheet PN 51A-400/rev.R December 2010 diagram (page 2)
Fig. 1 · Instruction Sheet PN 51A-400/rev.R December 2010

FIGURE 1. Sensor Orientation FIGURE 2. Direct Insertion in a Pipe

Emerson 400 — Instruction Sheet PN 51A-400/rev.R December 2010 diagram (page 2)
Fig. 2 · Instruction Sheet PN 51A-400/rev.R December 2010
Emerson 400 — Instruction Sheet PN 51A-400/rev.R December 2010 diagram (page 2)
Fig. 3 · Instruction Sheet PN 51A-400/rev.R December 2010

FIGURE 3. Insertion in a Pipe Tee

FIGURE 4. Insertion in a Pipe Tee

Emerson 400 — Instruction Sheet PN 51A-400/rev.R December 2010 diagram (page 2)
Fig. 4 · Instruction Sheet PN 51A-400/rev.R December 2010
Emerson 400 — Instruction Sheet PN 51A-400/rev.R December 2010 diagram (page 2)
Fig. 5 · Instruction Sheet PN 51A-400/rev.R December 2010

FIGURE 5. Insertion in a Pipe Tee

FIGURE 6. Insertion in a Flow Cell (PN 24091-02)

° °

Page 3 of the Emerson 400 user manual — Instruction Sheet PN 51A-400/rev.R December 2010
Page 3 · click to enlarge
Page 3
Emerson 400 — Instruction Sheet PN 51A-400/rev.R December 2010 diagram (page 3)
Fig. 6 · Instruction Sheet PN 51A-400/rev.R December 2010

FIGURE 7. Model 54eC Wiring FIGURE 8. Model 56 and 1056 Wiring

Page 4 of the Emerson 400 user manual — TB1 TB2
Page 4 · click to enlarge
Page 4

TB1 TB2

RTN

WHITE

RCV B

SENSE

WHITE/RED

RCV A

RTD IN

RED

CLEAR

RSHLD

SHLD CLEAR

GRAY

DRVB

ORANGE

DRVA

CLEAR

DSHLD

FIGURE 9. Model 1066 Wiring

Emerson 400 — TB1 TB2 diagram (page 4)
Fig. 7
Emerson 400 — TB1 TB2 diagram (page 4)
Fig. 8

FIGURE 10. Model Xmt-C-10 Wiring (Panel)

FIGURE 11. Model Xmt-C-11 Wiring (Pipe or Wall)

Emerson 400 — TB1 TB2 diagram (page 4)
Fig. 9
Emerson 400 — TB1 TB2 diagram (page 4)
Fig. 10

FIGURE 12. Model 5081C Wiring

FIGURE 13. Model 6081-C Wiring

Page 5 of the Emerson 400 user manual — CALIBRATION
Page 5 · click to enlarge
Page 5
WIRING THROUGH A JUNCTION BOX Model 400-60 sensors have a junction box mounted on the end of the sensor. See Figure 14 for wiring instructions. If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point. Use cable 23747-00 (factory-terminated) or 9200275 (raw cable). NOTES: 1. THE GRAY SENSOR WIRE IS CONNECTED TO THE JUNCTION BOX, WHICH MAKES ELECTRICAL CONTACT WITH THE OUTER ELECTRODE. 2. TERMINALS IN JUNCTION BOX ARE NOT NUMBERED. REFER TO ANALYZER WIRING DIAGRAM FOR CONNECTIONS AT ANALYZER. FIGURE 14. Sensor-Mounted Junction Box WiringPIN OUT DIAGRAM FOR 400VP FIGURE 15. VP pin-out (viewed from connector end of sensor, looking down)PIN OUT DIAGRAM FOR 400VP FIGURE 15. VP pin-out (viewed from connector end of sensor, looking down)
Emerson 400 — CLEANING THE SENSOR diagram (page 5)
Fig. 11
Emerson 400 — CLEANING THE SENSOR diagram (page 5)
Fig. 12

CLEANING THE SENSOR

Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol (rubbing alcohol) can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.

CALIBRATION

ENDURANCE conductivity sensors are calibrated at the factory and do not need calibration when first placed in service. Simply enter the cell constant printed on the label into the analyzer.

After a period of service, the sensor may require calibration. The sensor can be calibrated against a solution having known conductivity or against a referee meter and sensor. If using a standard solution, choose one having conduc- tivity in the recommended operating for the sensor cell constant. Refer to the analyzer manual or product data sheet for recommended ranges. Do not use standard solutions having conductivity less than about 100 uS/cm. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25°C). Because 0.01/cm sensors must be calibrated in low conductivity solutions, they are best calibrated against a referee meter and sensor in a closed system.

For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website.

Page 6 of the Emerson 400 user manual — TROUBLESHOOTING
Page 6 · click to enlarge
Page 6

TROUBLESHOOTING

PROBLEMPROBABLE CAUSESOLUTION Off-scale readingWiring is wrong. Temperature element is open or shorted. Sensor is not in process stream. Variopol cable is not properly seated. Sensor has failed.Verify wiring. Check temperature element for open or short circuits. See Figure 16. Be sure sensor is completely submerged in process stream. Loosen connector and reseat. Perform isolation checks. See Figure 17. Noisy readingSensor is improperly installed in process stream. Variopol cable is not properly seated.Be sure sensor is completely submerged in process stream. Loosen connector and reseat. Reading seems wrong (lower or higher than expected)Bubbles trapped in sensor. Wrong temperature correction algorithm. Wrong cell constant.Be sure sensor is properly oriented in pipe or flow cell. See Figure 1. Apply back pressure to flow cell. Check that temperature correction is appropriate for the sample. See analyzer manual for more information. Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. Sluggish responseElectrodes are fouled. Sensor is installed in dead area in piping.Clean electrodes. Move sensor to a location more representative of the process liquid.

FIGURE 16. Checking Temperature Element Disconnect leads and measure resistance shown. The meas- ured resistance should be close to the value in the table.

FIGURE 17. Checking Continuity and Leakage Disconnect electrode leads and measure resistance and continuity as shown. Sensor must be dry when checking resistance between electrode leads.

Page 7 of the Emerson 400 user manual — NOTES
Page 7 · click to enlarge
Page 7

NOTES

Page 8 of the Emerson 400 user manual — NOTES
Page 8 · click to enlarge
Page 8
Emerson 400 — NOTES diagram (page 8)
Fig. 13

The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE

http://www.raihome.com

Specifications subject to change without notice.

Credit Cards for U.S. Purchases Only.

2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250

http://www.raihome.com

© Rosemount Analytical Inc. 2010

Emerson Process Management

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Off-scale readingWiring is wrong. Temperature element is open or shorted. Sensor is not in process stream. Variopol cable is not properly seated. Sensor has failed.Verify wiring. p.6
Off-scale readingCheck temperature element for open or short circuits. See Figure 16. p.6
Off-scale readingBe sure sensor is completely submerged in process stream. p.6
Off-scale readingLoosen connector and reseat. p.6
Off-scale readingPerform isolation checks. See Figure 17. p.6
Noisy readingSensor is improperly installed in process stream. Variopol cable is not properly seated.Be sure sensor is completely submerged in process stream. p.6
Noisy readingLoosen connector and reseat. p.6
Reading seems wrong (lower or higher than expected)Bubbles trapped in sensor. Wrong temperature correction algorithm. Wrong cell constant.Be sure sensor is properly oriented in pipe or flow cell. See Figure 1. Apply back pressure to flow cell. p.6
Reading seems wrong (lower or higher than expected)Check that temperature correction is appropriate for the sample. See analyzer manual for more information. p.6
Reading seems wrong (lower or higher than expected)Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. p.6
Sluggish responseElectrodes are fouled. Sensor is installed in dead area in piping.Clean electrodes. p.6
Sluggish responseMove sensor to a location more representative of the process liquid. p.6

Ask about the Emerson 400

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson 400

Be the first to ask — other owners or our team will answer.

Original document

Emerson 400 manual PDF

8 pages·392 KB·Stud Sensor·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…