Models 396/396VP/397/398/398VP
pH/ORP Sensors




READ THIS PAGE BEFORE PROCEEDING!
This sensor is not Intrinsically Safe. or Explosion Proof. Installations near flammable liquids or in hazardous area locations must be carefully evaluated by qualified on site safety personnel.
To secure and maintain an intrinsically safe installation, an appropriate transmitter/safety barrier/sensor combi nation must be used. The installation system must be in accordance with the governing approval agency (FM, CSA or BASEEFA/CENELEC) hazardous area classifi- cation requirements. Consult your analyzer/ transmitter instruc tion manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal lation is entirely the responsibility of the user.
The wetted sensor materials may not be com- patible with process com position and operat- ing conditions. Application compat ibility is entirely the responsibility of the user.
About This Document This manual contains instructions for installation and operation of the Models 396, 396VP, 397, 398, and 398VP TUpH pH/ORP Sensors. The following list provides notes concerning all revisions of this document.
0 3/99 This is the initial release of the product manual. The manual has been reformatted to reflect the Emerson documentation style and updated to reflect any changes in the product offering.
A 12/01 Revised wiring diagram on page 24.
B 2/02 Added info to page 1.
C 6/02 Updated multiple drawings.
D 8/02 Added drawing #40105549, rev. D.
E 10/02 Revised drawing #40039601, rev. J, on page 10.
F 4/03 Revised Model 397 specs on page 2, and revised drawing on page 27.
G 8/03 Added Silcore information.
H 3/04 Added Xmt wiring drawings.
I 10/04 Added 5081 wiring drawing and updated 1055 wiring.
J 2/06 Changed drawing 40039603 rev. C, on page 13. Added a note on page 15.
K 1/07 Miscellaneous revisions
L 11/10 Removed mention of patents and updated dnv logo



LIST OF FIGURES Number Title Page 1-1 Cross Section Diagram of the TUpH Reference Technology 2 2-1 Flow Through/Insertion Installation for Models 396, 396VP, 398, & 398VP 10 2-2 Dimensional Drawing For Models 396 & 398 11 2-3 Dimensional Drawing For Models 396VP & 398VP 12 2-4 Submersion Installations For Models 396, 396VP, 398, & 398VP 13 2-5 Dimensional Drawings For Model 397 in the Quik-Loc Unit 14 2-6 Recommended Flow Through and Insertion Installation For Model 397 15 2-7 Models 396/397/398 Standard Wiring option used with Models 1054A, 1054B, 2054, and Preamps PN 22698-02 & 22698-03 15 2-8 Models 396/397/398 Option -62 for wiring to Models 1054, 1181, 2081, 54, 81, 3081, 4081, SoluComp, and 2700 Preamp (PN 23054-03) and remote J-box PN 23555-00 15 2-9 Wiring Models 396-54 (50), 397-54 (50), & 398-54 (50) to J-Box (PN 23309-04 (03)) & Remote Preamplifier 16 2-10 Wiring Models 396-54, 397-54, & 398-54 (PT-100-RTD) to J-Box (PN 23555-00) & Remote Preamplifier 17 2-11 Wiring Models 396/397/398 to Model 1181pH 18 2-12 Wiring Models 396-54, 397-54, & 398-54 to Models 1054A pH, 1054B pH, & 2054pH-54 19 2-13 Wiring Models 396-54, 397-54, & 398-54 to Models 2081pH-05 & 1054pH-54 20 2-14 Wiring Models 396-54, 397-54, & 398-54 to Models 54, 81, 3081, & 4081 21 2-15 Wiring Models 396-54, 397-54, & 398-54 to SoluComp Model SCL-P-014 22 2-16 Wiring Models 396-54, 397-54, & 398-54 to Model 2700 Preamp 23 2-17 Wiring Models 396-( )-54-62-(71), 397-02-10-54-62, & 398-( )-54-62 to Model 1055-01-10-22-32 23 2-18 Wiring Models 396-54-62, 397-02-54-62, & 398-54-62 to Model 1055-22-32 23 2-19 Wire Functions and Pin Connections for Variopol Connector Cable 24 2-20 Wiring Models 396VP/398VP to Model 81 thru a remote J-box 24 2-21 Wiring Models 396VP/398VP to Model 81 24 2-22 Wiring Models 396VP/398VP to Model 1181 24 2-23 Wiring Models 396VP/398VP to Model 1181, 1050/1060, 1003/1023 thru a remote J-box 25 2-24 Wiring Models 396VP/398VP to Model 2081 25 2-25 Wiring Models 396VP/398VP to Model 3081 & 4081 25 2-26 Wiring Models 396VP/398VP to Model 2081 thru a remote J-box 25 2-27 Wiring Models 396VP/398VP to Model 3081 & 4081 thru a remote J-box 25 2-28 Wiring Models 396VP/398VP to Model 1054 26 2-29 Wiring Models 396VP/398VP to Model 1054A/B & 2054 26 2-30 Wiring Models 396VP/398VP to Model 1054 thru a remote J-box 26 2-31 Wiring Models 396VP/398VP to Model 1054A/B & 2054 thru a remote J-box 26 2-32 Wiring Models 396VP/398VP to Model 54/54e thru a remote J-box 27 2-33 Wiring Models 396VP/398VP to Model 54/54e 27 2-34 Wiring Models 396VP/398VP to Model SCL-(P/Q) 27 2-35 Wiring Models 396VP/398VP to Model 2700 27 2-36 Wiring Models 396VP/398VP to Model 1055-22-32 28 2-37 Wiring Models 396/397/398 to Model Xmt 29 2-38 Wiring Models 396VP/398VP to Model Xmt 29 2-39 Wiring Models 396/397/398 to Model 5081-P 30 2-40 Wiring Models 396-54-62/398-54-62 to Model 1055-01-10-22-32 30

SECTION 1.0 DESCRIPTION AND SPECIFICATIONS
1.1 TUpH™ FEATURES AND APPLICATIONS.
mounting configurations. The Model 396 features an optional front or rear 1 in. MNPT process connection for insertion, submersion, or flow through applications.
Rosemount Analytical has achieved a new industry standard for the life expectancy of pH sensors with the TUpH Sensors. The low maintenance, disposable Models 396, 396VP, 397, 398, and 398VP TUpH Sensors offer long life and high performance in the measurement of pH in aqueous solutions in pipelines, open tanks, or ponds. These TUpH sensors feature a large area reference junction for maximum resistance to process coatings, generally found in dirty, high solid applications. In addition, the secondary helical refer- ence pathway serves as added protection from poi- soning ions. The simplified construction, designed with user convenience in mind, does not require elec- trolyte replenishment or any component replacement. All models feature a large glass bulb for increased resistance to the effects of aging for longer life. All models are available only without an integral pream- plifier. The preamplifier must be in a remote location or integral to the analyzer/transmitter. Models with a “VP” after the model number indicate the variopol connector is used on the sensor in place of the cable. In this case, a separate cable with a mating VP con- nector must be used. All models are compatible with all Rosemount Analytical and most manufacturers instruments. The entire line of TUpH model sensors now incor- porate the new SILCORE technology contaminant barrier. This triple-seal barrier prevents moisture and material impurities from migrating to the pH sensor’s reference electrode’s metal lead wire. By preventing these contaminants from compromising the integrity of the pH measurement, sensor life is increased, espe- cially at higher temperatures where increased migra- tions occur. In addition, the SILCORE1 technology pro- vides added protection against sensor failure due to vibrations and shock by transferring damaging energy away from the glass-to-metal seal.
1.1.2 Model 397 TUpH and Quik-Loc Features and Applications. The Model 397 is housed in a highly chemical resistant polypropylene body and completely sealed with EP to eliminate process intru- sion. The Model 397 body is specifically designed for use with the Quik-Loc Kit which consists of an adapter and coupler. The PEEK (polyetheretherketone) adapter enables the Model 397 sensor to fit into a 1 in. MNPT Dixon coupler for quick and easy removal without sensor cable twisting. The 316 stainless steel DIxon coupler is sealed with EP and features locking arms. The Quik-Loc Kit is not recom- mended for use in processes with hazardous, corro- sive, or strong oxidizing chemicals due to a risk of spray and bodily hazards.


1.1.3 Model 398 and 398VP TUpH Features and Applications.
The chemical-resistant con- struction of Tefzel, titanium, and the TUpH reference junc- tion make the Models 398 and 398VP the ideal sensors for measuring pH in harsh process liquids. Use Model 398 or 398VP to measure pH in sour water strippers, in pulp bleaching towers that use chlorine dioxide, and in process streams containing a variety of organic sol- vents.

Models 398 and 398VP use the highest quality mate- rials to provide superior chemical resistance. The sen- sors are housed in a titanium tube and features an optional 1 inch MNPT process connector for insertion, submersion, or flow-through applications. The molded Tefzel TUpH construction is offered with a choice of seals (Viton, EPDM, or Kalrez). Combining high qual- ity materials with the TUpH reference technology and ACCUGLASS1 pH bulb allows for ultimate chemical resistance and makes Models 398 and 398VP the perfect choice for measuring pH in harsh, demanding processes.
1.1.1 Model 396 and 396VP TUpH Features and Applications.
The Models 396 and 396VP TUpH sensors are construct- ed of polypropylene and stain- less steel and is completely sealed by EP (ethylene propy- lene) to eliminate process intrusion. It is specifically designed for improved life in harsh, dirty applications such as lime slurry waste treatment and paper machine headbox and pigment/dye applications where large quantities of suspended solids are present. Installation is easily achieved through a wide variety of
1 ACCUGLASS and Silcore are trademarks of Rosemount Analytical


All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an AccuGlass pH
glass bulb. This sensor technology ensures superior performance while only requiring minimal maintenance.

1.2 PERFORMANCE & PHYSICAL SPECIFICATIONS




1.3 ORDERING INFORMATION
The Model 396 pH Sensor features your selection of glass electrode type, either the standard hemi bulb or the optional flat glass electrode combined with the coating resistant, polypropylene reference with gel filled electrolyte. The sensor is housed in a stainless steel body and is used with a 1 in. MNPT threaded process connector suitable for insertion, submersion, or flow through installations. The Model 396 is available without an integral preamplifier only and 15 ft of cable. Automatic temperature compensation is standard 3 K Balco or PT-100 RTD.


The Model 396VP insertion/submersion sensor is the same as Model 396 except, instead of a cable on the back end of the sensor, it features a Variopol connector. The watertight Variopol sensor-to-cable connector uses a mating connector cable (ordered separately).

The Model 398 Sensor, housed in a titanium tube and made with a Tefzel reference junction, can be used with a 1 inch MNPT process connector (purchased separately). The sensor is available with a hemi bulb glass pH electrode or a platinum and glass ORP electrode and with Pt100 or 3K temperature compensation. The 398 is provided with a standard 15 ft cable. A preamplifier must be used if the sensor is more than 15 ft from the analyzer/transmitter. Junction box kits with preamplifiers must be ordered sep- arately. Process connector must also be ordered separately.


The Model 398VP insertion/submersion sensor is the same as Model 398, except the cable is replaced with a Variopol connector. The watertight Variopol sensor-to- cable connector uses the mating connector cable (ordered separately).

The TUpH Model 397 pH Sensor is housed in a polypropylene body and is designed to be used with the Quik-Loc Kit. The sensor includes a large general purpose (GPHT) glass pH electrode and a large area polypropylene reference junction with gel filled reference electrolyte. The Model 397 is available without an integral preamplifier only and 15 ft (4.6 m) of cable.

PN 23757-01, Quik-Loc Kit: for use in 1-1/2 in. and 2 in. tees; insertion depth 2 in. (50 mm)
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054 series, 2054, and 2081 Analyzers PN 23054-03 includes preamplifier for Solu Cube Model 2700 3. Extension cables (used with remote junction boxes) Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped 4. Wiring Aids Choose one: PN 9120531 BNC splitter, used in place of option -62 or BNC coax prepping for connections to
junction box (PN 23555-00); Models 1181, 1054, 2081, 54, 81, 3081, 4081; Solu Comp (SCL-P-014); and Model 2700 Preamplifier (PN 23054-03) NONE: No wiring aids required for connections to Models 1054A, 1054B, 2054, or preamplifiers
PN 22698-02 or 22698-03

FOR FIRST TIME 396, 396VP, 398, AND 398VP INSTALLATIONS, ROSEMOUNT ANALYTICAL RECOMMENDS USING THE FOLLOWING GUIDE:
PN 23166-01, Titanium, 1 in. x 1 in. NPT process connector, with EPDM o-ring PN 9510066, Nylon, 1 in. x 1 in. NPT process connector (submersion only) Choose one (optional process connector o-rings)
PN 9550220, Kalrez o-ring, 2-214 PN 9550099, Viton o-ring, 2-214
Choose one: PN 23645-06, 15 ft cable with mating VP connector, prepped with BNC on analyzer end
PN 23645-07, 15 ft cable with mating VP connector, prepped without BNC on analyzer end* 3. Mounting Accessories (optional) Choose one: PN 915240-03 PVC flow through tee, 3/4 in. NPT process connection
PN 915240-04 PVC flow through tee, 1 in. NPT process connection PN 915240-05 PVC flow through tee, 1 1/2 in. NPT process connection PN 11275-01 Sensor handrail mounting assembly PN 2002011 1-1/2 in. CPVC Tee with 1 in. FNPT connection PN 23728-00 Low Flow Cell, acrylic 4. Remote Junction Boxes (optional, recommended for sensor to analyzer distances of more than 15 ft) Choose one: PN 23555-00 includes preamplifier for Models 54/e, 1055, 81, 3081, 4081
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054 series, 2054, 2081 Analyzers PN 23054-03 includes preamplifier for Solu Cube Model 2700 5. Extension cables (used with remote junction boxes) Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped 6. Wiring Aids Choose one: PN 9120531 BNC splitter, used in place of option -62 or BNC coax prepping for connections to junction box
(PN 23555-00); Models 54, 1055, 1054, 81,1181, 2081, 3081, 4081; SoluComp (SCL-P-014); and Model 2700 Preamplifier (PN 23054-03) NONE: No wiring aids required for connections to preamplifier PN 22698-02, preamplifier PN 22698-03,
or Models 1054A, 1054B, and 2054
*Used for connections to Models 54/e, 1055, 1054, 81, 1181, 2081, 3081, 4081, and remote junction box PN 23555-00.

ADDITIONAL ACCESSORIES FOR MODEL 396, 396VP, 398 AND 398VP TUpH SENSORS
1 Teflon is a registered trademark of E.I. du Pont de Nemours & Co.

2.1 UNPACKING AND INSPECTION. Inspect the out- side of the carton for any damage. If damage is detect- ed, contact the carrier immediately. Inspect the instru- ment and hardware. Make sure all the items in the pack- ing list are present and in good condition. Notify the fac- tory if any part is missing. If the sensor appears to be in satisfactory condition, proceed to Section 2.2, Mounting.
2.2.1 Flow Through and Insertion Mounting for Models 396, 396VP, 398, and 398VP. The Models 396, 396VP, 398, and 398VP Sensors can be used with a 1 inch MNPT process connector at the front of the sensor for mounting into a 1-1/2 inch tee or the process. See Figure 2-1 for instal lation configurations.
2.2.2 Submersion Mounting for Models 396, 396VP, 398, and 398VP. The Models 396, 396VP, 398, and 398VP Sensors also have a 1 inch MNPT process con- nector available for use on the back of the sensor. Tapered pipe threads in plastic fittings tend to loosen after installation. It is therefore recom mended that Teflon tape be used on the threads and that the tightness of the connection be checked frequently to assure that no loos- ening has occurred. To prevent rain water or condensa- tion from running into the sensor, a weatherproof junc- tion box is recommended (see Figure 2-4). The sensor cable must be run through a protective conduit for isola- tion from electrical interference or physical abuse from the process. The sensor should be installed within 80° of vertical, with the electrode facing down. The sensor’s cable should not be run with power or control wiring.
NOTE Save the original packing cartons and mate- rials as most carriers require proof of dam- age due to mishandling, etc. Also, if it is nec- essary to return the instrument to the facto- ry, you must pack the instrument in the same manner as it was received. Refer to Section 6 for return instructions. If the sensor is to be stored, the vinyl boot should be filled with pH buffer solution and replaced on sensor tip until ready to use.
Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.
2.2.3 Quik-Loc Mounting for Model 397. The Quik- Loc mounting is used with the Model 397 TUpH sen- sor only.
2.2 MOUNTING. Each sensor has been designed to be located in industrial process environments. Temperature and pressure limitations must not be exceeded at any time. A caution or warning label regarding this matter is attached to each sensor. For insertion, transfer the label as shown on label instruc- tions. See Figure 2-2. For submersion applications, first note limits then remove and discard label.
It is recommended that a thermometer, drain valve to relieve pressure and pres-sure gauge be inserted near the Quik-Loc assembly (see Figure 2-5).
NOTE Before mounting the sensor, shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can sim- ply be installed as shipped and readings with an accuracy of ±0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be calibrated as a loop with its compatible ana- lyzer or transmitter.
Once the Quik-Loc unit is installed the oper- ator should wait for the process to cool to a safe temperature, use the pressure drain valve to relieve all process pressure and observe the pressure on the pressure gauge for proper removal of the sensor without spray or bodily injury. The Quik-Loc kit used with the 397 TUpH sensor is not recom- mended for use with hazardous, corrosive, or strong oxidizing chemicals due to a risk of spray or bodily injury.

Wrap the pipe threads of the Twin-Kam Kamloc coupler with Teflon tape before placing it into the process pipe. The coupler can be connected to any 1 in. process connection and must be mounted within 80° of vertical, with the elec- trode facing down. Once the coupler is in place, the adapter should be positioned onto the back end of the sen- sor. (See Figure 2-5) Remove the parafilm wrapping from the two 0-rings on the sensor, grease the o-rings with the lube provided, and feed the sensor cable through the adapter. Once the adapter is slipped over the sensor’s back end, the retaining ring (which is included with every 397 sensor) should be installed on the black, grooved, back end of the sensor body. The retaining ring secures the sensor into the adapter. The adapter/sensor assembly is now ready to be inserted into the coupler. With both arms of the coupler in the released position, insert the adapter/sensor assembly into the coupler.
Once the adapter has been properly placed in the coupler, both arms should be positioned in the locked position. A Sur- Loc™ 1 spring arm has been provided on one arm so that the “arm cannot be opened until the spring is released.”2
Once the arms of the coupler have been locked in position, use tamper-evident safety wire on the metal rings of the Twin-Kam arms to prevent unauthorized and/or untrained personnel from using the Quik-Loc unit.
It is the responsibility of each com- pany using the Quik-Loc Kit/ 397 TUpH Sensor to train personnel of the injury risks associated with using a quick-release coupler that is placed in a hot or pressurized process.
The Quik-Loc unit should be used only within the pressure and temperature limits stated for the Model 397 sensor in Section 1.2.

FIGURE 2-1. Recommended Flow Through and Insertion Installation
For Models 396, 396VP, 398 and 398VP
1-1/2 inch Pipe Tee (PN 2002011) with 1 inch threaded connections.


INCH
DWG. NO. REV. 40039601 J
FIGURE 2-2. Dimensional Drawing For Models 396 and 398



Metal Process Connector PN 23166-xx (xx = 00 for 316 SST and xx = 01 for tita- nium) can be used for insertion or sub- mersion mounting of Models 396VP or 398VP sensors in 1-inch fittings.
The metal process connector gives the sensor var- ious insertions depths, depending on where the user locates the compression fitting. Also the threads can be switched to face the cable end of the sensor for connection to submersion pipes.

INCH


DWG. NO. REV. 40039603 C
DWG. NO. REV. 40039604 D

FIGURE 2-4. Submersion Installations for Models 396, 396VP, 398, and 398VP.


INCH




FIGURE 2-5. Dimensional Drawings For Model 397 in the Quik-Loc Unit


FIGURE 2-6. Recommended Flow-Through and Insertion Installation For Model 397
2.3 ELECTRICAL INSTALLATION.
The Models 396, 396VP, 397, 398, and 398VP are for use with a remote preamplifier. Each sensor comes with either a special 15 ft low noise coax cable or a Variopol (VP) con- nector, which is used with a mating Variopol cable. The cable should be handled carefully, and kept dry and free of corrosive chemicals at all times. Extreme care should be used to prevent the cable from being twisted, damaged or scraped by rough, sharp edges or surfaces. Please refer to Figures 2-7 thru 2-18 for wiring Models 396, 397, and 398.
Please refer to Figures 2-19 thru 2-36 for wiring Models 396VP and 398VP.
DO NOT CONNECT SENSOR CABLE TO POWER
LINES. SERIOUS INJURY MAY RESULT.
Remove electrical tape or shrink sleeve from gray refer- ence wire before connecting wire to terminal.


FIGURE 2-7. Models 396, 397, and 398 Standard Wiring Option used with Models 1054A, 1054B, 2054, and Preamplifiers PN 22698-02 and 22698-03. FIGURE 2-8. Models 396, 397, and 398 Option -62 for Wiring to Models 1054, 1181, 2081, 54, 81, 3081, 4081, SoluComp, and 2700 Preamplifier (PN 23054-03). Also for use with remote Junction Box PN 23555-00.



FIGURE 2-9. Wiring Details For Model 396-54 (50), 397-54 (50), and 398-54 (50).


INCH
DWG. NO. REV. 40038920 G
FIGURE 2-10. Wiring Details for Model 396-54, 397-54, and 398-54 (Pt-100 RTD). For Use With
Junction Box (PN 23555-00) and Remote Preamplifier.


FIGURE 2-11. Wiring Details For Models 396, 397, and 398. For Use With Model 1181 pH.


INCH
DWG. NO. REV. 40039606 D
FIGURE 2-12. Wiring Details For Models 396-54, 397-54, and 398-54.


INCH
DWG. NO. REV. 40039608 D
FIGURE 2-13. Wiring Details For Models 396-54, 397-54, and 398-54.


INCH
DWG. NO. REV. 40039605 F
FIGURE 2-14. Wiring Details For Models 396-54 , 397-54, and 398-54.
For Use With Models 54, 54e, 81, 3081, 4081, and 5081.

INCH

FIGURE 2-15. Wiring Details For Models 396-54 , 397-54, and 398-54.
For Use With SoluComp Model SCL-P-014.


FIGURE 2-16. Wiring Models 396-54, 397-54, and 398-54 to Model 2700 Preamplifier (PN 23054-03).

to Model 1055-01-10-22-32


See below for wire functions for Models 396VP and 398VP. The Variopol connector cable part numbers are PN 23645- 06 (with BNC) and PN 23645-07.



FIGURE 2-20. Wiring Model 396VP
for Variopol connector cable PN 23645-06 (with BNC) PN 23645-07
Remote Junction Box


FIGURE 2-22. Wiring Model 396VP and 398VP to
Model 81
Model 1181




FIGURE 2-24. Wiring Model 396VP and 398VP to Model 2081


and 398VP to Model 1181, 1050/1060, and 1003/1023 through a Remote Junction Box
FIGURE 2-26. Wiring Model 396VP and 398VP to Model 2081 through a Remote Junction Box
FIGURE 2-27. Wiring Model 396VP and 398VP to Model 3081, 4081, and 5081 through a Remote Junction Box



FIGURE 2-28. Wiring Model 396VP and 398VP to
Models 1054A, 1054B, and 2054
Model 1054


FIGURE 2-30. Wiring Model 396VP and 398VP to
to Models 1054A, 1054B, and 2054 through a
Model 1054 through a Remote Junction Box
Remote Junction Box


FIGURE 2-32. Wiring Model 396VP and 398VP to Models 54 and 54e through a Remote Junction Box


FIGURE 2-33. Wiring Model 396VP/398VP to Models 54/54e
398VP to Model SCL-(P/Q)






396VP 398VP
FIGURE 2-38. Wiring Models 396VP & 398VP to Model Xmt-P-XX-10



Model 396-54-62 1055 -01 -10 -22 -32
Model 398-54-62

SECTION 3.0 START UP AND CALIBRATION
3.1 SENSOR PREPARATION.
3.2 CALIBRATION USING BUFFER SOLUTIONS OR GRAB SAMPLES.
Shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped and readings with an accuracy of ±0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be calibrated as a loop with its compatible analyzer or transmitter.
The pH sensor-analyzer/ transmitter loop may be calibrated by submersing the sensor in a buffer solu- tion (standard solutions of known pH values) or in a process grab sample whose pH value maybe checked by a calibrated laboratory or portable pH meter.
Please refer to the analyzer's or the transmitters respective instruction manuals for proper cali- bration procedures.

SECTION 4.0 MAINTENANCE
4.0 MAINTENANCE. The Model 396, 396VP, 397, 398, and 398VP Sensors are disposable type sensors and therefore require minimum maintenance. Every sensor should be kept clean and free of debris and sediment at all times. The frequency of cleaning by wiping or brushing with a soft cloth or brush is deter- mined by the nature of the solution being measured. The sensor should be removed from the process peri- odically and checked in buffer solutions.
Replace the sensor if cleaning does not restore sensor operation.
SENSOR IS IN PRESSURIZED SYSTEM! May cause spray and bodily injury. Before removing sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level.
4.2 AUTOMATIC TEMPERATURE COMPENSATOR. The temperature compensator element is temp erature sensitive and can be checked with an ohmmeter. Resistance increases with temperature.
The 3K element will read 3000 ohms ±1% at 25°C (77°F) and a Pt-100 will read 110 ohms. Resistance varies with temperature for a 3K and Pt-100 element and can be determined according to Table 4-2 or the following formula:
If the sensor will not calibrate, refer to your analyzer/ transmitter instruction manual for proper test proce- dures. If it is determined that the sensor has failed, it should be discarded and replaced.
Where RT = Resistance T = Temperature in °C
4.1 ELECTRODE CLEANING. If the electrode is coat- ed or dirty, clean as follows:
Refer to Table 4-1 for Ro and R1 values
SAFETY WARNING
SENSOR IS IN PRESSURIZED SYSTEM may cause spray and bodily injury. Before removing sensor, be absolutely cer- tain that the process pressure is reduced to 0 psig and the process temperature is low- ered to a safe level.
3K 2934 .0045 PT-100 107.7 .00385
of Auto T.C. Elements
Temperature °C Resistance (Ohms) ±1%
3K PT-100 0 2670 100.0 10 2802 103.8 20 2934 107.7 25 3000 109.6 30 3066 111.5 40 3198 115.4 50 3330 119.2 60 3462 123.1 70 3594 126.9 80 3726 130.8 90 3858 134.6 100 3990 138.5
The solution used during the following check is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immedi- ately rinse with clean water.
RT=Ro [l+R1 (T-20)]

SECTION 5.0 TROUBLESHOOTING
TABLE 5-1. Troubleshooting
Trouble Probable Cause Remedy
Meter reads off scale. (Display T.C. element shorted. Check T.C. element as instructed reads overrange.) in Section 4.2 and replace sensor if defective.
Sensor not in process or sample Make sure sensor is in process with stream is low. sufficient sample stream (refer to Section 2.0 for installation details).
Open glass electrode. Replace sensor.
Reference element open - no contact. Replace sensor.
Display reads between 3 and 6 pH Electrode cracked. Replace sensor. regardless of actual pH of solution or sample.
Meter or display indication swings T.C. element shorted. Check T.C. element as instructed or jumps widely in AUTO T.C. Mode. in Section 4.2 and replace sensor if defective.
Span between buffers extremely T.C. element open. Check T.C. element as instructed short in AUTO T.C. Mode. in Section 4.2 and replace sensor if defective.
Sluggish or slow meter indication Electrode coated. Clean sensor as instructed in for real changes in pH level. Sections 4.1 Replace sensor if cracked.
Electrode at end of life. Replace sensor.
Transmitter cannot be standardized. Electrode coated or cracked. Clean sensor as instructed in Sections 4.1. Replace sensor if cracked.
Transmitter short spans between Electrode at end of life, due to Replace sensor. two different buffer values. old glass or high temperature exposure.
Coated glass. Clean sensor as instructed in Section 4.1. Replace sensor if cracked.

SECTION 6.0 RETURN OF MATERIAL
6.1 GENERAL.
6.3 NON-WARRANTY REPAIR.
To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA) number.
The following is the procedure for returning for repair instruments that are no longer under warranty:
6.2 WARRANTY REPAIR.
The following is the procedure for returning instru- ments still under warranty:
Rosemount Analytical Inc., Uniloc Division Uniloc Division 2400 Barranca Parkway Irvine, CA 92606
RMA No
Model No

Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira- tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes, membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90 days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man- ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por- tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit- able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli- gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per- sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri- od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT- NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Liquid Division 2400 Barranca Parkway Irvine, CA 92606
The shipping container should be marked:
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Emerson Process Management · Return for Repair · Model


The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2010
Emerson Process Management
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Meter reads off scale. (Display reads overrange.) | T.C. element shorted. | Check T.C. element as instructed in Section 4.2 and replace sensor if defective. | p.37 |
| Meter reads off scale. (Display reads overrange.) | Sensor not in process or sample stream is low. | Make sure sensor is in process with sufficient sample stream (refer to Section 2.0 for installation details). | p.37 |
| Meter reads off scale. (Display reads overrange.) | Open glass electrode. | Replace sensor. | p.37 |
| Meter reads off scale. (Display reads overrange.) | Reference element open - no contact. | Replace sensor. | p.37 |
| Display reads between 3 and 6 pH regardless of actual pH of solution or sample. | Electrode cracked. | Replace sensor. | p.37 |
| Meter or display indication swings or jumps widely in AUTO T.C. Mode. | T.C. element shorted. | Check T.C. element as instructed in Section 4.2 and replace sensor if defective. | p.37 |
| Span between buffers extremely short in AUTO T.C. Mode. | T.C. element open. | Check T.C. element as instructed in Section 4.2 and replace sensor if defective. | p.37 |
| Sluggish or slow meter indication for real changes in pH level. | Electrode coated. | Clean sensor as instructed in Sections 4.1 Replace sensor if cracked. | p.37 |
| Sluggish or slow meter indication for real changes in pH level. | Electrode at end of life. | Replace sensor. | p.37 |
| Transmitter cannot be standardized. | Electrode coated or cracked. | Clean sensor as instructed in Sections 4.1. Replace sensor if cracked. | p.37 |
| Transmitter short spans between two different buffer values. | Electrode at end of life, due to old glass or high temperature exposure. | Replace sensor. | p.37 |
| Transmitter short spans between two different buffer values. | Coated glass. | Clean sensor as instructed in Section 4.1. Replace sensor if cracked. | p.37 |