Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS

Set-Up Guide

Agilent Technologies

Page 2 of the Agilent Technologies G1978B user manual — Restricted Rights Legend
Page 2 · click to enlarge
Page 2

Notices

© Agilent Technologies, Inc. 2008

Warranty

Safety Notices

No part of this manual may be reproduced in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without prior agree- ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

The material contained in this docu- ment is provided “as is,” and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec- tion with the furnishing, use, or per- formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.

CAUTION

A CAUTION notice denotes a haz- ard. It calls attention to an operat- ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

Manual Part Number

G1978-90070

Edition

First Edition, December 2008

Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051 USA

Windows® and MS Windows® are U.S. registered trademarks of Microsoft Corpora- tion.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi- tions are fully understood and met.

Windows NT® is a U.S. registered trade- mark of Microsoft Corporation.

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor- dance with the terms of such license.

Restricted Rights Legend

U.S. Government Restricted Rights. Soft- ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus- tomers. Agilent provides this customary commercial license in Software and techni- cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com- puter Software Documentation).

2 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Printed in USA

Page 3 of the Agilent Technologies G1978B user manual — 3 Methods
Page 3 · click to enlarge
Page 3

In This Guide

This guide explains how to install, maintain and troubleshoot your nanoelectospray ion source.

1 Installation

This chapter tells you how to install the multimode ion source.

2 Verification

This chapter tells you how to install the multimode ion source.

3 Methods

This chapter describes basic operation and maintenance for the multimode ion source.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 3

Page 4 of the Agilent Technologies G1978B user manual
Page 4 · click to enlarge
Page 4

4 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 6 of the Agilent Technologies G1978B user manual — Content
Page 6 · click to enlarge
Page 6

6 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Contents

Page 7 of the Agilent Technologies G1978B user manual — Agilent Technologies
Page 7 · click to enlarge
Page 7
Agilent Technologies G1978B — Installation diagram (page 7)
Fig. 2 · Installation

Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

1 Installation

Installation 8 Step 1. Prepare to install 8 Step 2. Install the HV control PCA and cables 9 Changing Sources 13 To remove the multimode source 13 To convert from multimode to ESI, APCI or APPI 14 To convert from ESI, APCI or APPI to the multimode source 15

This chapter contains instructions to install the multimode source on a 6100 Series Single Quad LC/MS system, and also to remove and replace the source.

Agilent Technologies

Page 8 of the Agilent Technologies G1978B user manual — Step 1. Prepare to install
Page 8 · click to enlarge
Page 8

Installation

This section explains how to install the G1978B multimode source on a 6100 Series Single Quad LC/MS instrument. The G1978A multimode source is not supported on these instruments.

The 6100 Series Single Quad LC/MS with multimode source is supported on ChemStation B.03.01 or higher, or the ChemStation B.01.03 with a patch.

Step 1. Prepare to install

The Multimode Enablement Kit, G1978-60451, is shipped with the multimode source. This kit needs to be installed before the multimode source is used.

1 Check that the Multimode Enablement Kit contains the following parts:

  • Multimode Bd HV cable, p/n G1960-60858
  • Multimode HV PCA, p/n G1960-61015
  • Multimode Bd Power/Data Cable, p/n G1960-60873
Agilent Technologies G1978B — Step 1. Prepare to install diagram (page 8)
Fig. 3 · Step 1. Prepare to install
Agilent Technologies G1978B — Step 1. Prepare to install diagram (page 8)
Fig. 4 · Step 1. Prepare to install
Agilent Technologies G1978B — Step 1. Prepare to install diagram (page 8)
Fig. 5 · Step 1. Prepare to install

Figure 1 From left to right: G1960-60858, G1960-61015 and G1960-60873

2 Install the APCI Enablement Kit, G1947-60451, which is shipped with the multimode source.

The APCI Enablement kit contains the following parts:

  • Fast APCI HV Supply, p/n G1946-80058

8 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

1 Installation Installation

Page 9 of the Agilent Technologies G1978B user manual — Step 2. Install the HV control PCA and cables
Page 9 · click to enlarge
Page 9

Installation 1 Step 2. Install the HV control PCA and cables

  • Valve BD-APCI supply, p/nG1960-60802
  • Valve BD-APCI Needle Interlock, p/n G1960-60856
Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 9)
Fig. 6 · Step 2. Install the HV control PCA and cables
Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 9)
Fig. 7 · Step 2. Install the HV control PCA and cables
Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 9)
Fig. 8 · Step 2. Install the HV control PCA and cables

Figure 2 From left to right: G1946-80058, G1960-60802 and G1960-60856

Step 2. Install the HV control PCA and cables

1 Turn off the system power and remove the system power cord.

Keep the power cord intact if the vacuum control switch box is used. The switch box is intended to keep the vacuum on while a CE works on the electronics. The switch box is for CE use only.

2 Remove the CDS cover, top, side, front, and the Aux Module cover.

3 Disconnect the ribbon cable that connects the valve PCA to the Vcap/Vchamber power supply. Then disconnect the Vcap and Vchamber cable from the power supply.

Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 9)
Fig. 9 · Step 2. Install the HV control PCA and cables
Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 9)
Fig. 10 · Step 2. Install the HV control PCA and cables

Figure 3 Disconnecting the Vcap/Vchamber power supply from the valve PCA (left) and the Vcap/Vchamber.

4 Place the multimode HV power supply PCA in the slot between the valve PCA and the Vcap/Vchamber. Secure the board by pressing it down into its slot and then attach it with two screws.

5 Connect the short gray cable from valve PCA to the multimode HV power supply.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 9

Page 10 of the Agilent Technologies G1978B user manual — Step 2. Install the HV control PCA and cables
Page 10 · click to enlarge
Page 10

1 Installation Step 2. Install the HV control PCA and cables

Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 10)
Fig. 11 · Step 2. Install the HV control PCA and cables

Figure 4 Connecting the valve PCA to the multimode HV power supply.

6 Install the APCI HV power supply. The APCI HV power supply is located at the end of the AUX Module.

7 Connect ribbon cable between valve PCA and Vcap/Vchamber power supply.

Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 10)
Fig. 12 · Step 2. Install the HV control PCA and cables

Figure 5 Connecting the valve PCA to the Vcap/Vchamber power supply.

8 Connect the Vcap and Vchamber cables to the Vcap/Vchamber power supply.

10 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 11 of the Agilent Technologies G1978B user manual — Step 2. Install the HV control PCA and cables
Page 11 · click to enlarge
Page 11

Installation 1 Step 2. Install the HV control PCA and cables

Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 11)
Fig. 13 · Step 2. Install the HV control PCA and cables

Figure 6 Connecting the Vcap and Vchamber cables to the power supply.

9 Connect the long ribbon cable, p/n G1960-60802, from the APCI HV power supply to the valve PCA.

Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 11)
Fig. 14 · Step 2. Install the HV control PCA and cables

Figure 7 Connecting the APCI HV power supply to the valve PCA.

10 Insert one end of the APCI Needle Interlock cable, G1960-60856, through the slot at the front of the system and then plug it to the APCI HV connector. Attach the other end to the chassis with the o-ring and the nut.

Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 11)
Fig. 15 · Step 2. Install the HV control PCA and cables
Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 11)
Fig. 16 · Step 2. Install the HV control PCA and cables

Figure 8 Connecting the APCI HV to the chassis.

11 Insert the cable, G1960-60858, to the top slot and attach it to the chassis. Plug the other two ends into the multimode HV PCA.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 11

Page 12 of the Agilent Technologies G1978B user manual — Step 2. Install the HV control PCA and cables
Page 12 · click to enlarge
Page 12

1 Installation Step 2. Install the HV control PCA and cables

Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 12)
Fig. 17 · Step 2. Install the HV control PCA and cables
Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 12)
Fig. 18 · Step 2. Install the HV control PCA and cables

Figure 9 Connecting the HV PCA to the chassis.

12 Close the AUX Module cover and reconnect all cables.

13 Install the multimode source to the system and connect all connectors.

Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 12)
Fig. 19 · Step 2. Install the HV control PCA and cables
Agilent Technologies G1978B — Step 2. Install the HV control PCA and cables diagram (page 12)
Fig. 20 · Step 2. Install the HV control PCA and cables

Figure 10 Installing the multimode source (left) and connecting all connectors.

14 Put back the side, top, front and CDS cover.

15 Plug the system power cord back on and turn the front switch on.

The pump down process will start.

16 Start the ChemStation program.

17 Click Method and Run Control View > MSD > Spray Chamber and verify that the source is an MM Source.

18 Go to MSD Tune window and start the Auto tune with a multimode source.

19 Verify that the multimode source is working correctly.

12 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 13 of the Agilent Technologies G1978B user manual — To remove the multimode source
Page 13 · click to enlarge
Page 13

Installation 1 Changing Sources

Changing Sources

To remove the multimode source

Do the following steps to remove the multimode source.

1 Set the source temperatures for vaporizer heater and drying gas heater to minimum values to cool the source.

Use the Tune > Instrument > Edit Spray Chamber menu item to display the Edit Spray Chamber dialog box. Set the drying gas flow, nebulizer gas flow, drying gas temperature and the vaporizer temperature to minimum values.

2 Wait approximately 20 minutes or until the source is cool.

3 Open the CDS door at the front of the MS to access the cables.

4 Disconnect the ESI high voltage charging electrode cable.

5 Disconnect the APCI Needle Interlock and multimode HV cable.

6 Unscrew the nebulizer gas line from the nebulizer.

7 Unscrew the LC sample tubing from the nebulizer.

8 Open the latch on the source and open the source.

9 Remove the multimode source from the spray chamber mount.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 13

Page 14 of the Agilent Technologies G1978B user manual — To convert from multimode to ESI, APCI or APPI
Page 14 · click to enlarge
Page 14

1 Installation To convert from multimode to ESI, APCI or APPI

10 Place the source shipping cover on the source.

To convert from multimode to ESI, APCI or APPI

1 If the source to be installed is an APPI source, disconnect the multimode high voltage PCA RS-232 serial cable from the serial port B connector of Smart Card.

2 Unscrew and remove the multimode spray shield with the field shaping electrodes.

3 Install the new source and the standard spray shield, making sure that the hole in the spray shield is in the 12 o'clock position.

4 For APCI and APPI ion source, connect the vaporizer heater cable and the APCI high voltage cable. For the APPI source, connect the RS-232 cable to the serial port B connector of Smart Card.

5 For all sources, reconnect the nebulizer gas line tubing and the LC/MS sample tubing.

14 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 15 of the Agilent Technologies G1978B user manual — To convert from ESI, APCI or APPI to the multimode source
Page 15 · click to enlarge
Page 15

Installation 1 To convert from ESI, APCI or APPI to the multimode source

To convert from ESI, APCI or APPI to the multimode source

1 Switch to the MSD Tune view.

2 Select Instrument/Set Spray Chamber and set all gas flows and temperatures to 0.

  • Drying Gas (L/min)
  • Nebulizer Pressure (psig)
  • Drying Gas Temperature (°C)
  • Vaporizer Temperature (APCI source only)
  • Lamp Off (APPI source only)

3 Wait for the source to cool (until temperatures are at least below 100°C).

4 Disconnect the nebulizer gas tubing from the currently installed ion source.

5 Disconnect the LC/MS sample inlet tubing.

6 If the APCI or APPI source is installed, remove the APCI vaporizer heater cable and APCI high voltage cable.

7 If the APPI source is installed, remove the serial port B RS-232 cable.

8 Remove the currently installed ion source.

9 Unscrew and remove the spray shield. See Figure 11.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 15

Page 16 of the Agilent Technologies G1978B user manual — To convert from ESI, APCI or APPI to the multimode source
Page 16 · click to enlarge
Page 16

1 Installation To convert from ESI, APCI or APPI to the multimode source

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 16)
Fig. 21 · To convert from ESI, APCI or APPI to the multimode s…

Figure 11 Standard spray shield and capillary cap for ESI or APCI

10 Remove the capillary cap. If needed, moisten a clean cloth with isopropyl alcohol and wipe the capillary cap. See Figure 12.

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 16)
Fig. 22 · To convert from ESI, APCI or APPI to the multimode s…

Figure 12 Spray shield removed.

11 Place the capillary cap back on the capillary.

12 Install the new spray shield with field shaping electrodes. See Figure 13.

16 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Standard spray shield · Capillary cap · Capillary cap

Page 17 of the Agilent Technologies G1978B user manual — To convert from ESI, APCI or APPI to the multimode source
Page 17 · click to enlarge
Page 17

Installation 1 To convert from ESI, APCI or APPI to the multimode source

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 17)
Fig. 23 · To convert from ESI, APCI or APPI to the multimode s…

Figure 13 Multimode spray shield

13 Screw the multimode spray shield into the holder for the spray shield. See Figure 14.

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 17)
Fig. 24 · To convert from ESI, APCI or APPI to the multimode s…

Field shaping electrode 9 o'clock position

Field shaping electrode 6 o'clock position

Figure 14 Multimode spray shield installed

14 Remove the shipping cover from the multimode source spray chamber.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 17

Page 18 of the Agilent Technologies G1978B user manual — To convert from ESI, APCI or APPI to the multimode source
Page 18 · click to enlarge
Page 18

1 Installation To convert from ESI, APCI or APPI to the multimode source

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 18)
Fig. 25 · To convert from ESI, APCI or APPI to the multimode s…

Figure 15 Multimode Spray Chamber

15 Install the spray chamber on the spray chamber mount.

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 18)
Fig. 26 · To convert from ESI, APCI or APPI to the multimode s…
Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 18)
Fig. 27 · To convert from ESI, APCI or APPI to the multimode s…

Figure 16 Multimode source with I-Button

16 Install the nebulizer on the multimode source spray chamber.

18 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

I-Button

Page 19 of the Agilent Technologies G1978B user manual — To convert from ESI, APCI or APPI to the multimode source
Page 19 · click to enlarge
Page 19

Installation 1 To convert from ESI, APCI or APPI to the multimode source

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 19)
Fig. 28 · To convert from ESI, APCI or APPI to the multimode s…

Figure 17 No nebulizer on top of the multimode source

17 Connect the 1/8-inch nebulizer gas tubing from the LC/MS mainframe to the nebulizer gas fitting. See Figure 18.

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 19)
Fig. 29 · To convert from ESI, APCI or APPI to the multimode s…

Nebulizer zero dead volume

Figure 18 Nebulizer with gas tubing connected

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 19

Sample Tubing · Nebulizer gas fitting

Page 20 of the Agilent Technologies G1978B user manual — To convert from ESI, APCI or APPI to the multimode source
Page 20 · click to enlarge
Page 20

1 Installation To convert from ESI, APCI or APPI to the multimode source

18 Connect the LC/MS sample tubing to the LC/MS diverter valve inlet filter. See Figure 19 on page 20.

WARNING The LC/MS Liquid Chromatograph diverter valve is an integral part of the G1978B safety system. The LC mobile phase flow must always be connected to the diverter valve inlet filter. Never bypass the diverter valve and connect directly to the nebulizer. If the diverter valve is used in a manner not specified by Agilent Technologies, the protections provided by the diverter valve may be impaired.

Agilent Technologies G1978B — To convert from ESI, APCI or APPI to the multimode source diagram (page 20)
Fig. 30 · To convert from ESI, APCI or APPI to the multimode s…

Figure 19 LC/MS sample tubing connected to LC/MS inlet filter

19 If you are installing the multimode source for the first time, follow the steps in “Step 2. Install the HV control PCA and cables” on page 9.

20 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

LC/MS sample tubing

Page 21 of the Agilent Technologies G1978B user manual — Agilent Technologies
Page 21 · click to enlarge
Page 21
Agilent Technologies G1978B — Verification diagram (page 21)
Fig. 31 · Verification

Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

2 Verification

To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM

Modes 22 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive

SIM Modes 24 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive

Scan Modes 27 To verify sensitivity for the G1978B Multimode using Multiple FIA method

in Mixed Mode operation 30 To do an autotune 32 Example of multimode verification report 33

This chapter contains instructions to verify your G1978B multimode a 6100 Series Single Quad LC/MS instrument.

Agilent Technologies

Page 22 of the Agilent Technologies G1978B user manual — Preparation
Page 22 · click to enlarge
Page 22

2 Verification To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes

To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes

Preparation

1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric flask. Use a clean graduated pipette.

2 Dilute to the 50 mL mark with 75:25 methanol/water with 0.1% formic or acetic acid.

3 Transfer 1 mL of the first dilution to a second 50 mL volumetric flask. Use a clean graduated pipette.

4 Dilute to the 50 mL mark with 75:25 methanol/water with 0.1% formic or acetic acid.

This provides the final 2 pg/µL reserpine concentration required for performance verification.

5 Transfer approximately 1 mL of the second dilution to a vial for use in the LC autosampler.

Table 1 6110 and 6120 Quad Performance Verification Summary Table

G1978B ES Positive SIM Mode

G1978B APCI Positive SIM Mode

Sample Reserpine, 5 ng/µL Reserpine, 5 ng/µL

Concentration after dilution 2 pg/µL 2 pg/µL

Injection volume 5 µL 5 µL

Total sample amount injected 10 pg 10 pg

Sample order number G2423A G2423A

Solvent 75:25 methanol/water with 0.1% formic or acetic acid

75:25 methanol/water with 0.1% formic or acetic acid

22 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 23 of the Agilent Technologies G1978B user manual — Sensitivity Verification for Multimode in ESI Positive SIM Mode
Page 23 · click to enlarge
Page 23

Verification 2 To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes

Table 1 6110 and 6120 Quad Performance Verification Summary Table

G1978B ES Positive SIM Mode

G1978B APCI Positive SIM Mode

Method name 6110SIMES_MM.M (6110 Quad)

6110SIMCI_MM.M (6110 Quad)

6120SIMES_MM.M (6120 Quad)

6120SIMCI_MM.M (6120 Quad)

Performance Specification 201 pk-pk 100: 1 rms 10: 1 pk-pk 50: 1 rms

Sensitivity Verification for Multimode in ESI Positive SIM Mode

1 Start up the ChemStation software.

2 Change the view to MSD Tune, and start an Autotune.

After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.

3 Change the view to Method and Run Control, and load the method 6120SIMES_MM.M (6120 model).

4 Edit the method to ensure that 75:25 methanol/water with 0.1% formic or acetic acid is selected as the LC solvent.

5 Place the vials into the LC autosampler.

  • Position #1: empty, uncapped vial
  • Position #2: vial of the solvent used for dilution (solvent blank)
  • Position #3: vial with the reserpine sample (2 pg/µL)

6 Run the method.

The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.

You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity.

7 Review the results.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 23

Page 24 of the Agilent Technologies G1978B user manual — Preparation
Page 24 · click to enlarge
Page 24

2 Verification To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes

When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes

Preparation

1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric flask. Use a clean graduated pipette.

2 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium formate.

3 Transfer 1mL of the first dilution to a 100 mL volumetric flask. Use a clean graduated pipette.

4 Dilute to the 100 mL mark with 75:25 methanol/water with 5 mM ammonium formate.

This provides the final 1 pg/µL reserpine concentration required for perfor mance verification.

5 Transfer approximately 1 mL of the second dilution to an autosampler vial.

Table 2 6130 and 6140 Quad Performance Verification Summary, SIM Mode

G1978B ES Positive SIM Mode

G1978B APCI Positive SIM Mode

Sample Reserpine, 5 ng/µL Reserpine, 5 ng/µL

Concentration after dilution 1 pg/µL 1 pg/µL

Injection volume 1 µL 1 µL

Total sample amount injected 1 pg 1 pg

Sample order number G2423A G2423A

24 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 25 of the Agilent Technologies G1978B user manual — Sensitivity Verification for Multimode in ESI, Positive SIM Mode
Page 25 · click to enlarge
Page 25

Verification 2 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes

Table 2 6130 and 6140 Quad Performance Verification Summary, SIM Mode

G1978B ES Positive SIM Mode

G1978B APCI Positive SIM Mode

Solvent 75:25 methanol/water with 5mM ammonium formate

75:25 methanol/water with 5mM ammonium formate

Method name 6130SIMES_MM.M (6130 model)

6130SIMCI_MM.M (6140 model)

6140SIMES_MM.M (6140 model)

6140SIMCI_MM.M (6140 model)

Performance Specification 201 pk-pk 100: 1 rms 10: 1 pk-pk 50: 1 rms

Sensitivity Verification for Multimode in ESI, Positive SIM Mode

1 Start up the ChemStation software.

2 Change the view to MSD Tune, and start an Autotune.

After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.

3 Change the view to Method and Run Control, and load the method 6130SIMES_MM.M (6130 model) or 6140SIMES_MM.M (6140 model).

4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.

5 Place the vials into the LC autosampler.

  • Position #1: empty, uncapped vial
  • Position #2: vial of the solvent used for dilution (solvent blank)
  • Position #3: vial with the reserpine sample (1 pg/µL)

6 Run the method.

The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.

You may need to further optimize the nebulizer pressure by running FIA

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 25

Page 26 of the Agilent Technologies G1978B user manual — Sensitivity Verification for Multimode in APCI, Positive SIM Mode
Page 26 · click to enlarge
Page 26

2 Verification To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes

experiments to achieve maximum instrument sensitivity 7 Review the results.

When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).

Sensitivity Verification for Multimode in APCI, Positive SIM Mode

1 Start the ChemStation software.

2 Change the view to MSD Tune, and start an Autotune.

After the autotune completes, you may need to wait up to 30 minutes before you continue, to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.

3 Change the view to Method and Run Control, and load the method 6130SIMCI_MM.M (6130 model) or 6140SIMCI_MM.M (6140 model)

4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.

5 Place the vials into the LC autosampler.

  • Position #1: empty, uncapped vial
  • Position #2: vial of the solvent used for dilution (solvent blank)
  • Position #3: vial with the reserpine sample (1 pg/µL)

6 Run the method.

You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity

The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.

7 Review the results.

When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 10:1 peak-to-peak (50:1 RMS).

26 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 27 of the Agilent Technologies G1978B user manual — Preparation
Page 27 · click to enlarge
Page 27

Verification 2 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes

Preparation

1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric flask. Use a clean graduated pipette.

2 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium formate.

3 Transfer 5 mL of the first dilution to a 50 mL volumetric flask. Use a clean graduated pipette.

4 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium formate.

This provides the final 10 pg/µL reserpine concentration required for performance verification.

5 Transfer approximately 1 mL of the second dilution to an autosampler vial.

Table 3 6130 and 6140 Quad Performance Verification Summary, Scan Mode

G1978B ES Positive Scan Mode

G1978B APCI Positive Scan Mode

Sample Reserpine, 5 ng/µL Reserpine, 5 ng/µL

Concentration after dilution 10 pg/µL 10 pg/µL

Injection volume 5 µL 5 µL

Total sample amount injected 50 pg 50 pg

Sample order number G2423A G2423A

Solvent 75:25 methanol/water with 5mM ammonium formate

75:25 methanol/water with 5mM ammonium formate

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 27

Page 28 of the Agilent Technologies G1978B user manual — Sensitivity Verification for Multimode in ESI, Positive Scan Mode
Page 28 · click to enlarge
Page 28

2 Verification To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes

Table 3 6130 and 6140 Quad Performance Verification Summary, Scan Mode

G1978B ES Positive Scan Mode

G1978B APCI Positive Scan Mode

Method name 6130SCNES_MM.M (6130 Quad)

6130SCNCI_MM.M (6130 Quad)

6140SCNES_MM.M (6140 Quad)

6140SCNCI_MM.M (6140 Quad)

Performance Specification 201 pk-pk 100: 1 rms 10: 1 pk-pk 50: 1 rms

Sensitivity Verification for Multimode in ESI, Positive Scan Mode

1 Start the ChemStation software.

2 Change the view to MSD Tune, and start an Autotune.

After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.

3 Change the view to Method and Run Control, and load the method 6130SCNES_MM.M (for 6130) or 6140SCNES_MM.M (for 6140).

4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.

5 Place the vials into the LC autosampler.

  • Position #1: empty, uncapped vial
  • Position #2: vial of the solvent used for dilution (solvent blank)
  • Position #3: vial with the reserpine sample (10 pg/µL)

6 Run the method.

You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity.

The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.

7 Review the results.

28 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 29 of the Agilent Technologies G1978B user manual — Sensitivity Verification for Multimode in APCI, Positive Scan Mode
Page 29 · click to enlarge
Page 29

Verification 2 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes

When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).

Sensitivity Verification for Multimode in APCI, Positive Scan Mode

1 Start the ChemStation software.

2 Change the view to MSD Tune, and start an Autotune.

After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.

3 Change the view to Method and Run Control, and load the method 6130SCNCI_MM.M (6130 model) or 6140SCNCI_MM.M (6140 model).

4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.

5 Place the vials into the LC autosampler.

  • Position #1: empty, uncapped vial
  • Position #2: vial of the solvent used for dilution (solvent blank)
  • Position #3: vial with the reserpine sample (10 pg/µL)

6 Run the method.

The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.

You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity

7 Review the results.

When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 10:1 peak-to-peak (50:1 RMS).

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 29

Page 30 of the Agilent Technologies G1978B user manual — Create and run a Multiple FIA Method Sequence
Page 30 · click to enlarge
Page 30

2 Verification To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation

To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation

Preparation

1 Do an autotune if necessary.

After the autotune has completed, you may need to wait up to 30 minutes before continuing to allow the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal from the calibrant.

You may need to further optimize the nebulizer pressure to achieve maximum instrument sensitivity.

2 Load the method MMCheckSL_ES.M and:

a Edit the method to ensure that 65:35 methanol/water with .2% acetic acid

is selected as the LC solvent.

b Save the method.

3 Repeat for MMCheckSL_CI.M and MMCheckSL_MX.M.

4 Place the vial into the LC autosampler.

  • Position 21: vial with the ESI + APCI LC Demo Sample (p/n G1978-85000)

Create and run a Multiple FIA Method Sequence

1 Select RunControl > Run Multiple FIA Methods.

2 In the Run Multiple FIA Methods dialog box, click Group > Add Group.

3 Type in a unique directory name for the data files to be stored.

4 In the Run Multiple FIA Methods dialog box, click Methods > Add Method, select MMCheckSL_ES.M and click OK.

5 Repeat to add MMCheckSL_CI.M.

6 Repeat to add MMCheckSL_MX.M.

7 Click Data File for the MMCheckSL_ES.M data file, then:

a Change the Subdirectory value to a unique subdirectory name in which

to store the data files.

b Change Data File to Multi_ES.

c Change Operator to the user’s name or identification code.

30 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 31 of the Agilent Technologies G1978B user manual — To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation
Page 31 · click to enlarge
Page 31

Verification 2 To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation

d Click OK.

8 Repeat for the MMCheckSL_CI.M data file. Use the same subdirectory and change Data File to Multi_CI.

9 Repeat for the MMCheckSL_MX.M data file. Use the same subdirectory and change Data File to Multi_MX as the data file name.

10 Click Run to start the sequence.

11 Review the results.

When the last method has finished, the report “Multimode Verification Report” prints.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 31

Page 32 of the Agilent Technologies G1978B user manual — To do an autotune
Page 32 · click to enlarge
Page 32

2 Verification To do an autotune

To do an autotune

Tuning the multimode source is done in MM-APCI mode only. Autotune is done from the same menu as with all sources.

  • From the MSD Tune view , select Instrument > Autotune menu item.

The tune report will have a header with either the title MM-APCI Positive Mode - Standard Scan or MM-APCI Negative Mode - Standard Scan. You can run the check tune after an autotune to validate that the instrument passes check tune criteria. Do the autotune after the system has had at least 8 hours to equilibrate vacuum and temperatures.

Agilent Technologies G1978B — To do an autotune diagram (page 32)
Fig. 32 · To do an autotune

Figure 20 Autotune report

32 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 33 of the Agilent Technologies G1978B user manual — Example of multimode verification report
Page 33 · click to enlarge
Page 33

Verification 2 Example of multimode verification report

Example of multimode verification report

Agilent Technologies G1978B — Example of multimode verification report diagram (page 33)
Fig. 33 · Example of multimode verification report

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 33

Page 34 of the Agilent Technologies G1978B user manual — Example of multimode verification report
Page 34 · click to enlarge
Page 34

2 Verification Example of multimode verification report

34 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 35 of the Agilent Technologies G1978B user manual — Agilent Technologies
Page 35 · click to enlarge
Page 35
Agilent Technologies G1978B — Methods diagram (page 35)
Fig. 34 · Methods

Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

3 Methods

To setup a method to use the multimode source 36 To create a method for positive/negative mixed mode operation 37 To create a method for alternating ESI and APCI operation 39

This chapter describes the tasks that you need to set up methods for the multimode source. See “Guidelines” in the Agilent G1978A/B Multimode Source Maintenance Guide for suggested source conditions for the multimode source for the different ionization modes.

Agilent Technologies G1978B — 3 Methods diagram (page 35)
Fig. 35 · 3 Methods

Agilent Technologies

Page 36 of the Agilent Technologies G1978B user manual — To setup a method to use the multimode source
Page 36 · click to enlarge
Page 36

3 Methods To setup a method to use the multimode source

To setup a method to use the multimode source

To have your method use a multimode source, follow these steps:

1 Open the MSD Spray Chamber dialog box by clicking Instrument > MSD Spray Chamber in the Method and Run Control view.

2 Set Method Spray Chamber to MM-ES+APCI.

3 Verify that Installed Spray Chamber is set to MM-ES+APCI.

4 Make any other changes that are necessary for your method.

5 Click the OK button.

6 Open The Set up MSD Signals dialog box by clicking Instrument > More > Set up MSD Signals in the Method and Run Control view.

7 Choose the desired ionization mode from the Ionization list. This list is only visible if the Method Spray Chamber was set to MM-ES+APCI. You may set the ionization mode to one of the following:

  • MM-ES
  • MM-APCI
  • MM-ES+APCI

8 Make any other changes that are necessary for your method.

9 Click the OK button.

WARNING The 6100 Series Single Quad LC/MS Liquid Chromatograph diverter valve is an integral part of the G1978B safety system. The LC mobile phase flow must always be connected to the diverter valve inlet filter. Never bypass the diverter valve and connect directly to the nebulizer. If the diverter valve is used in a manner not specified by Agilent Technologies, the protections provided by the diverter valve may be impaired and the system may catch fire.

36 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 37 of the Agilent Technologies G1978B user manual — To create a method for positive/negative mixed mode operation
Page 37 · click to enlarge
Page 37

Methods 3 To create a method for positive/negative mixed mode operation

To create a method for positive/negative mixed mode operation

1 Open the MSD Spray Chamber dialog box by clicking the Instrument > Set Up MSD Signals in the Method and Run Control view.

2 Select MM-ES_APCI from the Method Spray Chamber drop-down list.

3 Check that Installed Spray Chamber is also set to MM-ES+APCI.

4 Make any other changes that are necessary for your method.

5 Click the OK button.

6 Open the Set up MSD Signals dialog box by clicking the Instrument > MSD Spray chamber in the Method and Run Control view.

7 Modify the settings so that Signal 1 has Positive polarity and Signal 2 has Negative polarity as shown in Figure 21.

8 Make any other changes that are necessary for your method.

9 Click the OK button.

Fast positive/negative polarity switching is a very useful technique but it requires time for the ion chemistry to be established and the optics path to refill with ions. The gas density plays a role in the speed of refilling the ion path. The gas density is affected by source temperature. For a method running positive/negative switching, use a lower vaporizer temperature (150 to 200°C) and a lower Vcap (approximately 1000 V). These will greatly affect the quality of the results in positive/negative switching experiments.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 37

Page 38 of the Agilent Technologies G1978B user manual — To create a method for positive/negative mixed mode operation
Page 38 · click to enlarge
Page 38

3 Methods To create a method for positive/negative mixed mode operation

Agilent Technologies G1978B — To create a method for positive/negative mixed mode operation diagram (page 38)
Fig. 36 · To create a method for positive/negative mixed mode…

Polarity options for Signal 1 and Signal 2

Figure 21 Positive/Negative polarity switching using Set Up MSD Signals dialog box

38 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 39 of the Agilent Technologies G1978B user manual — To create a method for alternating ESI and APCI operation
Page 39 · click to enlarge
Page 39

Methods 3 To create a method for alternating ESI and APCI operation

To create a method for alternating ESI and APCI operation

1 Open The MSD Spray Chamber dialog box by clicking Instrument > MSD Spray Chamber in the Method and Run Control view.

2 Set Method Spray Chamber to MM-ES+APCI.

Agilent Technologies G1978B — To create a method for alternating ESI and APCI operation diagram (page 39)
Fig. 37 · To create a method for alternating ESI and APCI oper…

Figure 22 Method Spray Chamber set to MM-ES+APCI.

3 Check that Installed Spray Chamber is also set to MM-ES+APCI.

4 Make any other changes that are necessary for your method.

5 Click the OK button.

6 Open the Set up MSD Signals dialog box by clicking Instrument > Set Up MSD Signals in the Method and Run Control view.

7 Modify the settings so that Signal 1 has Ionization value of MM-ES and Signal 2 has Ionization value of MM-APCI as shown in Figure 23.

8 Make any other changes that are necessary for your method.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 39

Page 40 of the Agilent Technologies G1978B user manual — To create a method for alternating ESI and APCI operation
Page 40 · click to enlarge
Page 40

3 Methods To create a method for alternating ESI and APCI operation

9 Click the OK button.

NOTE In general, use the mixed mode operation (MM-ES+APCI setting with Signal 1) instead of alternating MM-ES (Signal 1) and MM-APCI (Signal 2) mode switching. Two times as many scans are obtained during elution of a chromatographic peak, and no delay is needed between scans, resulting in better data. It is rarely necessary to know whether a compound responds purely in ESI or APCI modes on a chromatographic time scale.

Agilent Technologies G1978B — To create a method for alternating ESI and APCI operation diagram (page 40)
Fig. 38 · To create a method for alternating ESI and APCI oper…

Ionization mode for Signal 1 and Signal 2

Figure 23 MM-ES and MM-APCI switching using Set Up MSD Signals dialog box

40 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Page 42 of the Agilent Technologies G1978B user manual — Index
Page 42 · click to enlarge
Page 42

42 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Index

Page 43 of the Agilent Technologies G1978B user manual — Index
Page 43 · click to enlarge
Page 43
Page 43 of the Agilent Technologies G1978B user manual — Index
Page 44 of the Agilent Technologies G1978B user manual — G1978-90070
Page 44 · click to enlarge
Page 44

www.agilent.com

In This Book

This book contains installation, operation, maintenance and troubleshooting instruction for the Multimode Source for 6100 Series Single Quad LC/MS.

© Agilent Technologies, Inc. 2008

Printed in USA First Edition, December 2008

*G1978-90070*

G1978-90070

Agilent Technologies

Ask about the Agilent Technologies G1978B

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies G1978B

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies G1978B manual PDF

44 pages·1014 KB·Computer Accessories·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…