Set-Up Guide

© Agilent Technologies, Inc. 2008
Safety Notices
No part of this manual may be reproduced in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without prior agree- ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
The material contained in this docu- ment is provided “as is,” and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec- tion with the furnishing, use, or per- formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.
A CAUTION notice denotes a haz- ard. It calls attention to an operat- ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
G1978-90070
Edition
First Edition, December 2008
Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051 USA
Windows® and MS Windows® are U.S. registered trademarks of Microsoft Corpora- tion.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi- tions are fully understood and met.
Windows NT® is a U.S. registered trade- mark of Microsoft Corporation.
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor- dance with the terms of such license.
U.S. Government Restricted Rights. Soft- ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus- tomers. Agilent provides this customary commercial license in Software and techni- cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com- puter Software Documentation).
2 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Printed in USA

This guide explains how to install, maintain and troubleshoot your nanoelectospray ion source.
This chapter tells you how to install the multimode ion source.
This chapter tells you how to install the multimode ion source.
This chapter describes basic operation and maintenance for the multimode ion source.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 3

4 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide


6 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Contents


Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Installation 8 Step 1. Prepare to install 8 Step 2. Install the HV control PCA and cables 9 Changing Sources 13 To remove the multimode source 13 To convert from multimode to ESI, APCI or APPI 14 To convert from ESI, APCI or APPI to the multimode source 15
This chapter contains instructions to install the multimode source on a 6100 Series Single Quad LC/MS system, and also to remove and replace the source.

This section explains how to install the G1978B multimode source on a 6100 Series Single Quad LC/MS instrument. The G1978A multimode source is not supported on these instruments.
The 6100 Series Single Quad LC/MS with multimode source is supported on ChemStation B.03.01 or higher, or the ChemStation B.01.03 with a patch.
The Multimode Enablement Kit, G1978-60451, is shipped with the multimode source. This kit needs to be installed before the multimode source is used.
1 Check that the Multimode Enablement Kit contains the following parts:



Figure 1 From left to right: G1960-60858, G1960-61015 and G1960-60873
2 Install the APCI Enablement Kit, G1947-60451, which is shipped with the multimode source.
The APCI Enablement kit contains the following parts:
8 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
1 Installation Installation

Installation 1 Step 2. Install the HV control PCA and cables



Figure 2 From left to right: G1946-80058, G1960-60802 and G1960-60856
1 Turn off the system power and remove the system power cord.
Keep the power cord intact if the vacuum control switch box is used. The switch box is intended to keep the vacuum on while a CE works on the electronics. The switch box is for CE use only.
2 Remove the CDS cover, top, side, front, and the Aux Module cover.
3 Disconnect the ribbon cable that connects the valve PCA to the Vcap/Vchamber power supply. Then disconnect the Vcap and Vchamber cable from the power supply.


Figure 3 Disconnecting the Vcap/Vchamber power supply from the valve PCA (left) and the Vcap/Vchamber.
4 Place the multimode HV power supply PCA in the slot between the valve PCA and the Vcap/Vchamber. Secure the board by pressing it down into its slot and then attach it with two screws.
5 Connect the short gray cable from valve PCA to the multimode HV power supply.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 9

1 Installation Step 2. Install the HV control PCA and cables

Figure 4 Connecting the valve PCA to the multimode HV power supply.
6 Install the APCI HV power supply. The APCI HV power supply is located at the end of the AUX Module.
7 Connect ribbon cable between valve PCA and Vcap/Vchamber power supply.

Figure 5 Connecting the valve PCA to the Vcap/Vchamber power supply.
8 Connect the Vcap and Vchamber cables to the Vcap/Vchamber power supply.
10 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Installation 1 Step 2. Install the HV control PCA and cables

Figure 6 Connecting the Vcap and Vchamber cables to the power supply.
9 Connect the long ribbon cable, p/n G1960-60802, from the APCI HV power supply to the valve PCA.

Figure 7 Connecting the APCI HV power supply to the valve PCA.
10 Insert one end of the APCI Needle Interlock cable, G1960-60856, through the slot at the front of the system and then plug it to the APCI HV connector. Attach the other end to the chassis with the o-ring and the nut.


Figure 8 Connecting the APCI HV to the chassis.
11 Insert the cable, G1960-60858, to the top slot and attach it to the chassis. Plug the other two ends into the multimode HV PCA.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 11

1 Installation Step 2. Install the HV control PCA and cables


Figure 9 Connecting the HV PCA to the chassis.
12 Close the AUX Module cover and reconnect all cables.
13 Install the multimode source to the system and connect all connectors.


Figure 10 Installing the multimode source (left) and connecting all connectors.
14 Put back the side, top, front and CDS cover.
15 Plug the system power cord back on and turn the front switch on.
The pump down process will start.
16 Start the ChemStation program.
17 Click Method and Run Control View > MSD > Spray Chamber and verify that the source is an MM Source.
18 Go to MSD Tune window and start the Auto tune with a multimode source.
19 Verify that the multimode source is working correctly.
12 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Installation 1 Changing Sources
Do the following steps to remove the multimode source.
1 Set the source temperatures for vaporizer heater and drying gas heater to minimum values to cool the source.
Use the Tune > Instrument > Edit Spray Chamber menu item to display the Edit Spray Chamber dialog box. Set the drying gas flow, nebulizer gas flow, drying gas temperature and the vaporizer temperature to minimum values.
2 Wait approximately 20 minutes or until the source is cool.
3 Open the CDS door at the front of the MS to access the cables.
4 Disconnect the ESI high voltage charging electrode cable.
5 Disconnect the APCI Needle Interlock and multimode HV cable.
6 Unscrew the nebulizer gas line from the nebulizer.
7 Unscrew the LC sample tubing from the nebulizer.
8 Open the latch on the source and open the source.
9 Remove the multimode source from the spray chamber mount.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 13

1 Installation To convert from multimode to ESI, APCI or APPI
10 Place the source shipping cover on the source.
1 If the source to be installed is an APPI source, disconnect the multimode high voltage PCA RS-232 serial cable from the serial port B connector of Smart Card.
2 Unscrew and remove the multimode spray shield with the field shaping electrodes.
3 Install the new source and the standard spray shield, making sure that the hole in the spray shield is in the 12 o'clock position.
4 For APCI and APPI ion source, connect the vaporizer heater cable and the APCI high voltage cable. For the APPI source, connect the RS-232 cable to the serial port B connector of Smart Card.
5 For all sources, reconnect the nebulizer gas line tubing and the LC/MS sample tubing.
14 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Installation 1 To convert from ESI, APCI or APPI to the multimode source
1 Switch to the MSD Tune view.
2 Select Instrument/Set Spray Chamber and set all gas flows and temperatures to 0.
3 Wait for the source to cool (until temperatures are at least below 100°C).
4 Disconnect the nebulizer gas tubing from the currently installed ion source.
5 Disconnect the LC/MS sample inlet tubing.
6 If the APCI or APPI source is installed, remove the APCI vaporizer heater cable and APCI high voltage cable.
7 If the APPI source is installed, remove the serial port B RS-232 cable.
8 Remove the currently installed ion source.
9 Unscrew and remove the spray shield. See Figure 11.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 15

1 Installation To convert from ESI, APCI or APPI to the multimode source

Figure 11 Standard spray shield and capillary cap for ESI or APCI
10 Remove the capillary cap. If needed, moisten a clean cloth with isopropyl alcohol and wipe the capillary cap. See Figure 12.

Figure 12 Spray shield removed.
11 Place the capillary cap back on the capillary.
12 Install the new spray shield with field shaping electrodes. See Figure 13.
16 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Standard spray shield · Capillary cap · Capillary cap

Installation 1 To convert from ESI, APCI or APPI to the multimode source

Figure 13 Multimode spray shield
13 Screw the multimode spray shield into the holder for the spray shield. See Figure 14.

Field shaping electrode 9 o'clock position
Field shaping electrode 6 o'clock position
Figure 14 Multimode spray shield installed
14 Remove the shipping cover from the multimode source spray chamber.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 17

1 Installation To convert from ESI, APCI or APPI to the multimode source

Figure 15 Multimode Spray Chamber
15 Install the spray chamber on the spray chamber mount.


Figure 16 Multimode source with I-Button
16 Install the nebulizer on the multimode source spray chamber.
18 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
I-Button

Installation 1 To convert from ESI, APCI or APPI to the multimode source

Figure 17 No nebulizer on top of the multimode source
17 Connect the 1/8-inch nebulizer gas tubing from the LC/MS mainframe to the nebulizer gas fitting. See Figure 18.

Nebulizer zero dead volume
Figure 18 Nebulizer with gas tubing connected
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 19
Sample Tubing · Nebulizer gas fitting

1 Installation To convert from ESI, APCI or APPI to the multimode source
18 Connect the LC/MS sample tubing to the LC/MS diverter valve inlet filter. See Figure 19 on page 20.
WARNING The LC/MS Liquid Chromatograph diverter valve is an integral part of the G1978B safety system. The LC mobile phase flow must always be connected to the diverter valve inlet filter. Never bypass the diverter valve and connect directly to the nebulizer. If the diverter valve is used in a manner not specified by Agilent Technologies, the protections provided by the diverter valve may be impaired.

Figure 19 LC/MS sample tubing connected to LC/MS inlet filter
19 If you are installing the multimode source for the first time, follow the steps in “Step 2. Install the HV control PCA and cables” on page 9.
20 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
LC/MS sample tubing


Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM
Modes 22 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive
SIM Modes 24 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive
Scan Modes 27 To verify sensitivity for the G1978B Multimode using Multiple FIA method
in Mixed Mode operation 30 To do an autotune 32 Example of multimode verification report 33
This chapter contains instructions to verify your G1978B multimode a 6100 Series Single Quad LC/MS instrument.

2 Verification To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes
1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric flask. Use a clean graduated pipette.
2 Dilute to the 50 mL mark with 75:25 methanol/water with 0.1% formic or acetic acid.
3 Transfer 1 mL of the first dilution to a second 50 mL volumetric flask. Use a clean graduated pipette.
4 Dilute to the 50 mL mark with 75:25 methanol/water with 0.1% formic or acetic acid.
This provides the final 2 pg/µL reserpine concentration required for performance verification.
5 Transfer approximately 1 mL of the second dilution to a vial for use in the LC autosampler.
Table 1 6110 and 6120 Quad Performance Verification Summary Table
G1978B ES Positive SIM Mode
G1978B APCI Positive SIM Mode
Sample Reserpine, 5 ng/µL Reserpine, 5 ng/µL
Concentration after dilution 2 pg/µL 2 pg/µL
Injection volume 5 µL 5 µL
Total sample amount injected 10 pg 10 pg
Sample order number G2423A G2423A
Solvent 75:25 methanol/water with 0.1% formic or acetic acid
75:25 methanol/water with 0.1% formic or acetic acid
22 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Verification 2 To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes
Table 1 6110 and 6120 Quad Performance Verification Summary Table
G1978B ES Positive SIM Mode
G1978B APCI Positive SIM Mode
Method name 6110SIMES_MM.M (6110 Quad)
6110SIMCI_MM.M (6110 Quad)
6120SIMES_MM.M (6120 Quad)
6120SIMCI_MM.M (6120 Quad)
1 Start up the ChemStation software.
2 Change the view to MSD Tune, and start an Autotune.
After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.
3 Change the view to Method and Run Control, and load the method 6120SIMES_MM.M (6120 model).
4 Edit the method to ensure that 75:25 methanol/water with 0.1% formic or acetic acid is selected as the LC solvent.
5 Place the vials into the LC autosampler.
6 Run the method.
The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.
You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity.
7 Review the results.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 23

2 Verification To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes
When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes
1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric flask. Use a clean graduated pipette.
2 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium formate.
3 Transfer 1mL of the first dilution to a 100 mL volumetric flask. Use a clean graduated pipette.
4 Dilute to the 100 mL mark with 75:25 methanol/water with 5 mM ammonium formate.
This provides the final 1 pg/µL reserpine concentration required for perfor mance verification.
5 Transfer approximately 1 mL of the second dilution to an autosampler vial.
Table 2 6130 and 6140 Quad Performance Verification Summary, SIM Mode
G1978B ES Positive SIM Mode
G1978B APCI Positive SIM Mode
Sample Reserpine, 5 ng/µL Reserpine, 5 ng/µL
Concentration after dilution 1 pg/µL 1 pg/µL
Injection volume 1 µL 1 µL
Total sample amount injected 1 pg 1 pg
Sample order number G2423A G2423A
24 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Verification 2 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes
Table 2 6130 and 6140 Quad Performance Verification Summary, SIM Mode
G1978B ES Positive SIM Mode
G1978B APCI Positive SIM Mode
Solvent 75:25 methanol/water with 5mM ammonium formate
75:25 methanol/water with 5mM ammonium formate
Method name 6130SIMES_MM.M (6130 model)
6130SIMCI_MM.M (6140 model)
6140SIMES_MM.M (6140 model)
6140SIMCI_MM.M (6140 model)
1 Start up the ChemStation software.
2 Change the view to MSD Tune, and start an Autotune.
After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.
3 Change the view to Method and Run Control, and load the method 6130SIMES_MM.M (6130 model) or 6140SIMES_MM.M (6140 model).
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.
5 Place the vials into the LC autosampler.
6 Run the method.
The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.
You may need to further optimize the nebulizer pressure by running FIA
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 25

2 Verification To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes
experiments to achieve maximum instrument sensitivity 7 Review the results.
When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).
1 Start the ChemStation software.
2 Change the view to MSD Tune, and start an Autotune.
After the autotune completes, you may need to wait up to 30 minutes before you continue, to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.
3 Change the view to Method and Run Control, and load the method 6130SIMCI_MM.M (6130 model) or 6140SIMCI_MM.M (6140 model)
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.
5 Place the vials into the LC autosampler.
6 Run the method.
You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity
The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.
7 Review the results.
When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 10:1 peak-to-peak (50:1 RMS).
26 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Verification 2 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes
1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric flask. Use a clean graduated pipette.
2 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium formate.
3 Transfer 5 mL of the first dilution to a 50 mL volumetric flask. Use a clean graduated pipette.
4 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium formate.
This provides the final 10 pg/µL reserpine concentration required for performance verification.
5 Transfer approximately 1 mL of the second dilution to an autosampler vial.
Table 3 6130 and 6140 Quad Performance Verification Summary, Scan Mode
G1978B ES Positive Scan Mode
G1978B APCI Positive Scan Mode
Sample Reserpine, 5 ng/µL Reserpine, 5 ng/µL
Concentration after dilution 10 pg/µL 10 pg/µL
Injection volume 5 µL 5 µL
Total sample amount injected 50 pg 50 pg
Sample order number G2423A G2423A
Solvent 75:25 methanol/water with 5mM ammonium formate
75:25 methanol/water with 5mM ammonium formate
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 27

2 Verification To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes
Table 3 6130 and 6140 Quad Performance Verification Summary, Scan Mode
G1978B ES Positive Scan Mode
G1978B APCI Positive Scan Mode
Method name 6130SCNES_MM.M (6130 Quad)
6130SCNCI_MM.M (6130 Quad)
6140SCNES_MM.M (6140 Quad)
6140SCNCI_MM.M (6140 Quad)
1 Start the ChemStation software.
2 Change the view to MSD Tune, and start an Autotune.
After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.
3 Change the view to Method and Run Control, and load the method 6130SCNES_MM.M (for 6130) or 6140SCNES_MM.M (for 6140).
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.
5 Place the vials into the LC autosampler.
6 Run the method.
You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity.
The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.
7 Review the results.
28 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Verification 2 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes
When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).
1 Start the ChemStation software.
2 Change the view to MSD Tune, and start an Autotune.
After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.
3 Change the view to Method and Run Control, and load the method 6130SCNCI_MM.M (6130 model) or 6140SCNCI_MM.M (6140 model).
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.
5 Place the vials into the LC autosampler.
6 Run the method.
The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.
You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity
7 Review the results.
When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 10:1 peak-to-peak (50:1 RMS).
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 29

2 Verification To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation
To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation
1 Do an autotune if necessary.
After the autotune has completed, you may need to wait up to 30 minutes before continuing to allow the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal from the calibrant.
You may need to further optimize the nebulizer pressure to achieve maximum instrument sensitivity.
2 Load the method MMCheckSL_ES.M and:
a Edit the method to ensure that 65:35 methanol/water with .2% acetic acid
is selected as the LC solvent.
b Save the method.
3 Repeat for MMCheckSL_CI.M and MMCheckSL_MX.M.
4 Place the vial into the LC autosampler.
1 Select RunControl > Run Multiple FIA Methods.
2 In the Run Multiple FIA Methods dialog box, click Group > Add Group.
3 Type in a unique directory name for the data files to be stored.
4 In the Run Multiple FIA Methods dialog box, click Methods > Add Method, select MMCheckSL_ES.M and click OK.
5 Repeat to add MMCheckSL_CI.M.
6 Repeat to add MMCheckSL_MX.M.
7 Click Data File for the MMCheckSL_ES.M data file, then:
a Change the Subdirectory value to a unique subdirectory name in which
to store the data files.
b Change Data File to Multi_ES.
c Change Operator to the user’s name or identification code.
30 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Verification 2 To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation
d Click OK.
8 Repeat for the MMCheckSL_CI.M data file. Use the same subdirectory and change Data File to Multi_CI.
9 Repeat for the MMCheckSL_MX.M data file. Use the same subdirectory and change Data File to Multi_MX as the data file name.
10 Click Run to start the sequence.
11 Review the results.
When the last method has finished, the report “Multimode Verification Report” prints.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 31

2 Verification To do an autotune
Tuning the multimode source is done in MM-APCI mode only. Autotune is done from the same menu as with all sources.
The tune report will have a header with either the title MM-APCI Positive Mode - Standard Scan or MM-APCI Negative Mode - Standard Scan. You can run the check tune after an autotune to validate that the instrument passes check tune criteria. Do the autotune after the system has had at least 8 hours to equilibrate vacuum and temperatures.

Figure 20 Autotune report
32 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Verification 2 Example of multimode verification report

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 33

2 Verification Example of multimode verification report
34 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide


Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
To setup a method to use the multimode source 36 To create a method for positive/negative mixed mode operation 37 To create a method for alternating ESI and APCI operation 39
This chapter describes the tasks that you need to set up methods for the multimode source. See “Guidelines” in the Agilent G1978A/B Multimode Source Maintenance Guide for suggested source conditions for the multimode source for the different ionization modes.


3 Methods To setup a method to use the multimode source
To have your method use a multimode source, follow these steps:
1 Open the MSD Spray Chamber dialog box by clicking Instrument > MSD Spray Chamber in the Method and Run Control view.
2 Set Method Spray Chamber to MM-ES+APCI.
3 Verify that Installed Spray Chamber is set to MM-ES+APCI.
4 Make any other changes that are necessary for your method.
5 Click the OK button.
6 Open The Set up MSD Signals dialog box by clicking Instrument > More > Set up MSD Signals in the Method and Run Control view.
7 Choose the desired ionization mode from the Ionization list. This list is only visible if the Method Spray Chamber was set to MM-ES+APCI. You may set the ionization mode to one of the following:
8 Make any other changes that are necessary for your method.
9 Click the OK button.
WARNING The 6100 Series Single Quad LC/MS Liquid Chromatograph diverter valve is an integral part of the G1978B safety system. The LC mobile phase flow must always be connected to the diverter valve inlet filter. Never bypass the diverter valve and connect directly to the nebulizer. If the diverter valve is used in a manner not specified by Agilent Technologies, the protections provided by the diverter valve may be impaired and the system may catch fire.
36 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Methods 3 To create a method for positive/negative mixed mode operation
1 Open the MSD Spray Chamber dialog box by clicking the Instrument > Set Up MSD Signals in the Method and Run Control view.
2 Select MM-ES_APCI from the Method Spray Chamber drop-down list.
3 Check that Installed Spray Chamber is also set to MM-ES+APCI.
4 Make any other changes that are necessary for your method.
5 Click the OK button.
6 Open the Set up MSD Signals dialog box by clicking the Instrument > MSD Spray chamber in the Method and Run Control view.
7 Modify the settings so that Signal 1 has Positive polarity and Signal 2 has Negative polarity as shown in Figure 21.
8 Make any other changes that are necessary for your method.
9 Click the OK button.
Fast positive/negative polarity switching is a very useful technique but it requires time for the ion chemistry to be established and the optics path to refill with ions. The gas density plays a role in the speed of refilling the ion path. The gas density is affected by source temperature. For a method running positive/negative switching, use a lower vaporizer temperature (150 to 200°C) and a lower Vcap (approximately 1000 V). These will greatly affect the quality of the results in positive/negative switching experiments.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 37

3 Methods To create a method for positive/negative mixed mode operation

Polarity options for Signal 1 and Signal 2
Figure 21 Positive/Negative polarity switching using Set Up MSD Signals dialog box
38 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Methods 3 To create a method for alternating ESI and APCI operation
1 Open The MSD Spray Chamber dialog box by clicking Instrument > MSD Spray Chamber in the Method and Run Control view.
2 Set Method Spray Chamber to MM-ES+APCI.

Figure 22 Method Spray Chamber set to MM-ES+APCI.
3 Check that Installed Spray Chamber is also set to MM-ES+APCI.
4 Make any other changes that are necessary for your method.
5 Click the OK button.
6 Open the Set up MSD Signals dialog box by clicking Instrument > Set Up MSD Signals in the Method and Run Control view.
7 Modify the settings so that Signal 1 has Ionization value of MM-ES and Signal 2 has Ionization value of MM-APCI as shown in Figure 23.
8 Make any other changes that are necessary for your method.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 39

3 Methods To create a method for alternating ESI and APCI operation
9 Click the OK button.
NOTE In general, use the mixed mode operation (MM-ES+APCI setting with Signal 1) instead of alternating MM-ES (Signal 1) and MM-APCI (Signal 2) mode switching. Two times as many scans are obtained during elution of a chromatographic peak, and no delay is needed between scans, resulting in better data. It is rarely necessary to know whether a compound responds purely in ESI or APCI modes on a chromatographic time scale.

Ionization mode for Signal 1 and Signal 2
Figure 23 MM-ES and MM-APCI switching using Set Up MSD Signals dialog box
40 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide


42 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Index



In This Book
This book contains installation, operation, maintenance and troubleshooting instruction for the Multimode Source for 6100 Series Single Quad LC/MS.
© Agilent Technologies, Inc. 2008
Printed in USA First Edition, December 2008
Agilent Technologies