© Agilent Technologies, Inc. 2008
Safety Notices
No part of this manual may be reproduced in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without prior agree- ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
The material contained in this docu- ment is provided “as is,” and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec- tion with the furnishing, use, or per- formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.
A CAUTION notice denotes a haz- ard. It calls attention to an operat- ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
G3170-90001
Edition
Second edition, January 2008
Agilent Technologies, Inc. 5301 Stevens Creek Boulevard Santa Clara, CA 95052
Acknowledgements
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi- tions are fully understood and met.
Microsoft® and Windows® are U.S. regis- tered trademarks of Microsoft Corporation
Printed in USA



5975 Series Mass Selective Detectors Site Preparation Guide
Instrument Identification 6 Important Safety Warnings 7 Hydrogen Safety 9 Chemical Safety 10 Safety and Regulatory Certifications 11 Cleaning 14 Recycling the Product 14
This section provides information on how to identify specifically what type of MSD you are working with as well as how to properly clean and dispose of the instrument. Also included in this section are the “Important Safety Warnings” which list critical safety precautions for all users.

Each 5975 Series Mass Selective Detector (MSD) is identified by a unique 10-character serial number. This serial number is located on a label on the lower left side near the front of the instrument. On CI upgrades, a second serial number is located on the flow module.
When corresponding with Agilent Technologies about your instrument, be sure to include the model number and both full 10-character serial numbers.
Write the serial number of your 5975 MSD here for reference:
Table 1 shows the kinds of 5975 MSDs available. The product number on the serial number label identifies your MSD.
Table 1 5975 MSD versions
Model number Product number Pump type Ionization mode
5975 VL MSD G3170A Diffusion Electron Impact (EI) only
5975 inert MSD G3171A Standard turbo Electron Impact (EI) only
G3172A Performance turbo Electron Impact (EI) only
5975 inert XL MSD G3174A Performance turbo Electron Impact (EI) Negative Chemical Ionization (NCI) Positive Chemical Ionization (PCI
This site preparation guide will refer to 5975 Series MSDs unless noted otherwise.
General Information

Before moving on, there are several important safety notices that you should always keep in mind when using the 5975 Series Mass Selective Detector.
If the MSD is connected to a power source, even if the power switch is off, potentially dangerous voltages exist on:
With the power switch on, potentially dangerous voltages also exist on:
Connecting the MSD to power sources that are not equipped with protective earth contacts creates a shock hazard for the operator and can damage the instrument.
Interrupting the protective conductor inside or outside the MSD or disconnecting the protective earth terminal creates a shock hazard for the operator and can damage the instrument.
Make sure the power cords supplied with the MSD are appropriate for your country and site before using them. Maintain easy access to the power cords so they can be disconnected during maintenance.
General Information

The use of incorrect or makeshift fuses or the short-circuiting of fuse holders creates a shock hazard for the operator and can damage the instrument. Replace fuses only with fuses of identical current rating and type.
Excessive fluctuations in the line voltage can create a shock hazard and can damage the instrument. Make sure the supply voltage does not fluctuate more than +5% or -10% from the rated voltage. This equipment must be installed in a Category II environment as defined in IEC 664.
The printed circuit (PC) boards in the MSD can be damaged by electrostatic discharge. Do not touch any of the boards unless it is absolutely necessary. If you must handle them, wear a grounded wrist strap and take other antistatic precautions. Wear a grounded wrist strap any time you must remove the MSD covers.
Many parts of the MSD operate at temperatures high enough to cause serious burns.
You should always cool heated areas of the MSD to room temperature before working on them. They will cool faster if you first set the temperature of the heated zone to room temperature. Turn the zone off after it has reached the setpoint. If you must perform maintenance on hot parts, use a wrench and wear gloves. Whenever possible, cool the part of the instrument that you will be maintaining before you begin working on it.
WARNING The insulation around the GC inlets, detectors, valve box, and the insulation cups is made of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend the following safety procedures: ventilate your work area; wear long sleeves, gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation in a sealed plastic bag; wash your hands with mild soap and cold water after handling the insulation.
General Information

Hydrogen gas may be used as carrier gas, and/or as fuel for certain GC detectors. When mixed with air, hydrogen can form explosive mixtures.
A detailed discussion of hydrogen safety appears in Chapter 1 of the Agilent 5975 MSD Troubleshooting and Maintenance Manual. Read this material before using hydrogen with the MSD.
WARNING When using hydrogen (H2) as the carrier gas or fuel gas, be aware that hydrogen gas can flow into the oven and create an explosion hazard. Therefore, be sure that the supply is off until all connections are made, and ensure that the GC inlet and detector column fittings are either connected to a column or capped at all times when hydrogen gas is supplied to the instrument.
Hydrogen is flammable. Leaks, when confined in an enclosed space, may create a fire or explosion hazard. In any application using hydrogen, leak test all connections, lines, and valves before operating the instrument. Always turn off the hydrogen supply at its source before working on the instrument.
When using hydrogen gas, check the system for leaks as described by your local Environmental Health and Safety (EHS) requirements to prevent possible fire and explosion hazards. Always check for leaks after changing a tank or servicing the gas lines. Always make sure the vent line is vented into a fume hood.
General Information

General Information

The 5975 Series MSD conforms to the following safety standards:
The 5975 Series MSD conforms to the following regulations on Electromagnetic Compatibility (EMC) and Radio Frequency Interference (RFI):
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB—001 du Canada.
The 5975 Series MSD is designed and manufactured under a quality system registered to ISO 9001.
The Agilent Technologies 5975 Series MSD meets the following IEC (International Electro-technical Commission) classifications: Equipment Class I, Laboratory Equipment, Installation Category II, Pollution Degree 2.
This unit has been designed and tested in accordance with recognized safety standards and is designed for use indoors. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. Whenever the safety protection of the MSD has been compromised, disconnect the unit from all power sources and secure the unit against unintended operation.
Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard.
General Information

See accompanying instructions for more information.
Indicates a hot surface.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
Indicates potential explosion hazard.
Indicates radioactivity hazard.
Indicates electrostatic discharge hazard.
Indicates that you must not discard this electrical/electronic product in domestic household waste.
General Information · or

This device complies with the requirements of CISPR 11. Operation is subject to the following two conditions:
If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try one or more of the following measures:
1 Relocate the radio or antenna.
2 Move the device away from the radio or television.
3 Plug the device into a different electrical outlet, so that the device and the radio or television are on separate electrical circuits.
4 Make sure that all peripheral devices are also certified.
5 Make sure that appropriate cables are used to connect the device to peripheral equipment.
6 Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance.
7 Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment.
Sound pressure Lp < 70 dB according to EN 27779:1991.
Schalldruckpegel LP < 70 dB am nach EN 27779:1991.
General Information

To clean the unit, disconnect the power and wipe down with a damp, lint-free cloth.
For recycling, contact your local Agilent sales office.
General Information


5975 Series Mass Selective Detectors Site Preparation Guide
Overview 16 Space and Weight Requirements 20 Communications Requirements 22 Electrical Requirements 23 Air Conditioning and Environmental Requirements 28 Carrier and Reagent Gas Requirements 31 Laboratory Supply Requirements 33 Receiving the System 37 Installation and Verification 39
This section describes how to properly prepare your site for a new MSD. Follow these instructions carefully as delays due to improper site preparation may result in loss of instrument use during the warranty period.

Before the 5975 Series MSD system can be installed, the site must be properly prepared. Site preparation includes, but is not limited to, ensuring that adequate facilities are available. Among the site requirements are:
Delays due to inadequate site preparation could cause loss of instrument use during the warranty period. In extreme cases, Agilent Technologies may ask to be reimbursed for the additional time required to complete the installation. Agilent Technologies provides service during the warranty period and under maintenance agreements only if the specified site requirements are met.
Unless previous arrangements have been made with Agilent Technologies, site preparation is the customer’s responsibility. Customer responsibilities include, but are not limited to:

An Agilent Technologies service representative will install the MSD and verify its performance. The service representative’s responsibilities are limited to:

Agilent Technologies is not responsible for:

Additional information is contained in the following documentation:

Table 2 lists dimension and weight information for the MSD and related components. Your site must have enough bench space for the MSD, GC, data system, and accessories (Figure 1). In addition, there must be sufficient space around the system for ventilation and maintenance access. At least 30 cm (12 inches) to the left of the MSD and at least 20 cm (8 1/4 inches) behind the MSD must be kept clear.
Benches must be sturdy enough to support the weight of the entire system.
The foreline pump can be located on the laboratory bench or on the floor. It must be close to the MSD because it is connected by a 200-cm (79-inch) hose. The hose is stiff and cannot be bent sharply.

Figure 1 Top view of typical installation

Table 2 Product dimensions
Approximate Dimensions, cm (in)
Product Height Width Depth Weight, kg (lb)
7890 Series GC 54 (21) 59 (23) 54 (21) 50 (112)
6890 Series GC 50 (20) 58 (23) 55 (22) 50 (110)
6850 Series GC 50 (20) 29 (11) 54 (21) 23 (51)
Data System Data system size and weight depend on the components included in the data system. Reserve at least 100 cm (39 in) of bench space for the data system. A typical data system weight is 34 kg (75 lb).
Be sure to allow space for your carrier and reagent gases.
5975 Series MSD

Telephone
A telephone with a cord long enough to be used at the computer will allow the MSD operator to communicate with Agilent Technologies support personnel.
If you intend to connect your MSD system to your site’s LAN network, you must have an additional shielded twisted pair network cable.

You are responsible for providing appropriate electrical power and power outlets for all of the components in your 5975 Series MSD system. Power considerations include:
The 5975 Series MSD includes a full-range power supply that can operate without reconfiguration on either of two wide ranges of single-phase alternating current (AC) electrical power:
The foreline pump also draws its power from the MSD. However, a different foreline pump is supplied depending on the voltage range on which the MSD will be operating. The pump is supplied according to the standard voltage in the country from which the order originates. For example, if an order originates from an Agilent Technologies sales office in Germany, the foreline pump supplied will be configured to operate on the standard voltage and frequency of electrical power in Germany.

Electrical power for the MSD must be single-phase. The neutral wire cannot be used for safety grounding. The ground wire should carry zero current except for ground-fault current or static electric discharge. The entire system should share an isolated, noise-free electrical ground. This system ground should be electrically separate from the ground for the rest of the building, back to the main ground for the facility. Power configurations are provided in Table 3.
Table 3 Power configurations
Configuration Measurement Nominal voltage
Single phase, 120-127 VAC Line to neutral 120 or 127 VAC *
Line to ground 120 or 127 VAC *
Ground to neutral < 0.5 V rms
Single phase, 200-240 VAC Line to neutral 200, 220, or 240 VAC *
Line to ground 200, 220, or 240 VAC *
Ground to neutral < 0.5 V rms
* Varies with country and/or region

Table 4 lists the power requirements for the 5975 Series MSD and related equipment. Extra power capacity for future additions is recommended.
Each product listed requires a dedicated circuit. The gauge controller is powered by the MSD. The MSD, GC, and data system must each have a separate circuit breaker. All of the equipment must share a common ground.
Power must meet the stability specifications listed in Table 4. Use a line monitor to check power stability. If your line power is unstable, you may need to install a line conditioner.
Table 4 Power requirements
Product Line voltage Maximum continuous power consumption
5975 Series MSD* 120-127 VAC 60 Hz 900 VA (400 VA for foreline pump only)
200-240 VAC, 50/60 Hz 900 VA 15 A 1
7890A and 6890 GCs (standard oven) and ALS
220/230/240V single/split phase (-10%/+5%), 48-66 Hz
7890A and 6890 GCs (fast oven) and ALS
220/230/240 VAC‡ single/split phase (-10%/+5%), 48-66 Hz
6850 GC (standard oven)**
230 VAC, 48-63 Hz 2000 VA 8-12 A 1
230 VAC, 48-63 Hz 2400 VA 11 A 1
Supply circuit rating · Outlets required

Table 4 Power requirements (continued)
Product Line voltage Maximum continuous power consumption
ChemStation computer 100-120 or 220-240 VAC+5/-10%, 50/60 Hz ± 5% 1000 VA 15 A 3 to 5
* The MSD operates on either voltage range. The foreline and turbo pumps are supplied for the voltage range suitable to the country where the order originates
† Americas 120 V requires 20 A dedicated line. Americas 240 V requires 15 A dedicated line.
‡ Option 003, fast oven, is a unit with operating range 198 to 231 V. Most laboratories have 4-wire service resulting in 208 V at the wall receptacle. It is important to measure the line voltage at the receptacle for the GC.
** Requires an isolated ground and dedicated outlet.
The MSD is supplied with a power cord and plug appropriate for the country from which the order originates. For example, if an order originates from a Agilent Technologies sales office in Germany, the power cord and plug supplied will be compatible with the standard voltage and outlet configuration in Germany. See “Power Cords" on page 41 for illustrations of the power cords available.
Data system components also include power cords with plugs appropriate for the country where the order was placed.
Power cord lengths for the MSD and the data system components and accessories are approximately 2.3 m (7.5 ft).
Supply circuit rating · Outlets required

Additional electrical considerations include:

Air conditioning considerations include temperature, humidity, airborne dust, and exhaust venting. Each of these is considered in more detail in the following material.
The 5975 Series MSD is specified for operation under the following conditions:
Table 5 Temperature and humidity limits
Operation 15 °C to 35 °C (59 °F to 95 °F) 40% to 80%
Storage –20 °C to 70 °C (–4 °F to 158 °F) 0% to 95%
Environmental control systems must maintain these temperature and humidity ranges.
The MSD is rated for 1200 Watts (4000 BTU/h). The GC and the data system also contribute significantly to the cooling load although the exact amounts depend on their configurations. Additional allowances should be made for other heat sources such as heat from other equipment, heat from adjacent rooms, and heat from laboratory personnel.
Temperature Humidity (relative)

Agilent Technologies recommends a maximum airborne particle density of 55 μg/m3. If you suspect your site exceeds this limit, contact your local Agilent Technologies Customer Service Organization. Customer Engineers with special training and equipment can test for airborne particle density. They can offer suggestions for reducing airborne dust.
There are two sources of exhaust on the GC/MSD system: the foreline pump and the GC split vent. The foreline pump outputs gas removed from the vacuum manifold by the high vacuum pumps. The foreline pump exhaust will also contain traces of solvent and sample.
WARNING The pump exhaust contains carrier gas and traces of solvents, analytes, and foreline pump oil. The supplied oil trap stops only pump oil. It does not trap or filter out toxic chemicals. If you are using toxic solvents or toxic or flammable carrier gas, or analyzing toxic chemicals, do not install the oil trap. Install a hose to take the pump exhaust to a fume hood. Do not use the oil trap on any CI MSD.
The foreline pump exhaust cannot be vented into the laboratory if any hazardous materials will be introduced into the GC/MS system, including samples, solvents, carrier gases, and reagent gases. It must be vented external to the building or vacuum exhausted to a fume hood.
If a negative pressure vent is not available, the length of tubing from the foreline pump to an ambient pressure vent should not exceed 460 cm (15 ft). The exhaust can not be connected to a positive pressure vent.
Exhaust gas venting must comply with all local environmental and safety codes.

See your GC manuals for information on venting the split vent.
An auxiliary work space and fume hood are needed for some maintenance procedures.

Table 6 lists the specifications for the carrier and reagent gases. Pressures and flows are as measured at the bulkhead fittings on the back of the instrument, not at the gas supply. We recommend that additional shut-off valves be installed near the instruments.
Table 6 Carrier and reagent gas requirements
Purity Typical pressure range, kPa (psi)
Helium (required) 99.9995% or better and hydrocarbon-free
345 to 550 (50 to 80) 20 to 50 (column and split flow)
Methane reagent gas (required for CI operation)
Research or SFC grade, 99.999% 103 to 172 (15 to 25) 1 to 2
Isobutane reagent gas (optional)
Instrument grade, 99.99% 103 to 172 (15 to 25) 1 to 2
Ammonia reagent gas (optional)
Research or SFC grade, 99.9995% 34 to 55 (5 to 8) 1 to 2
Carbon dioxide reagent gas (optional)
SFC Grade, 99.995% 103 to 138 (15 to 20) 1 to 2
Typical flow, mL/min

You must supply appropriate regulators for your carrier and reagent gas.
See the Agilent Technologies Chemical Analysis Columns and Supplies Catalog or visit the Agilent Technologies Chemical Analysis website at http://www.agilent.com/chem for dual-stage regulators available from Agilent Technologies.
You must supply Swagelok or Swagelok-compatible fittings, ferrules, and connectors for the 1/8-inch tubing.

Laboratory supply requirements vary with applications.
Cleaning tasks for the MSD require the following HPLC-grade (or better) solvents:
Proper storage, handling, and disposal of these chemicals is required for personal and environmental safety.
Carbon disulfide may be used as a sample solvent in standard size injections (typically 2 µL or less).
You will need paper for printing the results of the testing done during installation and later for printing reports of your analyses. You will also need appropriate supplies for making backup copies of your data files.

The supplies and parts listed in Table 7 are used in the operation and maintenance of a 5975 Series MSD system. Keeping these parts on hand can reduce system downtime related to instrument maintenance and repair.
Table 7 Maintenance supplies and consumables
One-year maintenance kit 5183-0296
He/H2 gas filter for GC 5182-3468
Abrasive sheets (5/pk) 5061-5896
Cloths, lint-free (15/pk) 05980-60051
Cotton swabs (100/pk) 5080-5400
Foreline pump oil, Inland 45, 1 liter 6040-0834
EI filament (2/pk) 05972-60053
EI high temp filament G2590-60053
Nitrile rubber gloves, medium 9300-1751
EI high temp filament G2590-60053

Table 7 Maintenance supplies and consumables (continued)
PFTBA (EI calibration fluid) 05971-60571
PFDTD (CI calibration fluid) 8500-8130
EVAL A (evaluation sample) 05990-60045
OFN 100 fg/µL 5188-5347
Foreline pump oil, P3, 0.5 liter 6040-0621
He/H2 gas purifier for GC RMSH-2
Methane/Isobutane gas purifier for CI MSD G1999-80410
1/2-inch x 9/16-inch, open end 8710-0877
0.3-mm id, 85% Vespel 15% graphite, for 0.10-mm id columns
0.4-mm id, 85% Vespel 15% graphite, for 0.20-mm id and 0.25-mm id columns 5062-3508

Table 7 Maintenance supplies and consumables (continued)
0.5-mm id, 85% Vespel 15% graphite, for 0.32-mm id columns
0.8-mm id, 85% Vespel 15% graphite, for 0.53-mm id columns
0.27-mm id, 90% Vespel 10% graphite, for 0.10-mm id columns
0.37-mm id, 90% Vespel 10% graphite, for 0.20-mm id columns
0.40-mm id, 90% Vespel 10% graphite, for 0.25-mm id columns
0.47-mm id, 90% Vespel 10% graphite, for 0.32-mm id columns
0.74-mm id, 90% Vespel 10% graphite, for 0.53-mm id columns

When your MSD system is delivered, it is your responsibility to provide for removal of the shipping containers from the truck and their storage until installation. Contact your Agilent Technologies service representative as soon as your shipment arrives to arrange an installation date.
The shipping containers are large and heavy. The largest container in the shipment is 121.9 cm (48 inches) × 121.9 cm × 121.9 cm. It contains several smaller cartons that can be removed from the large container after delivery to facilitate moving them to the location where the instrument is to be installed. After removing the smaller cartons and discarding the outer container, the largest remaining carton, may be as large as:
All doorways, hallways, floors, and elevators must be able to accommodate the largest, heaviest container.
Once the shipping containers are unloaded, examine them for any obvious external damage. If any of the containers appear damaged, note on the carrier's bill of lading that there is Apparent damage - subject to inspection and test. Arrange for both the carrier's claims representative and your Agilent Technologies service representative to be present when the containers are unpacked.
Do not open any of the shipping containers unless a representative of Agilent Technologies is present. Opening any of the containers without an Agilent Technologies representative present will void the receiving warranty on the instrument.

It is your responsibility to store the containers until installation. If your site does not have adequate storage space, the containers may be stored at your expense in a bonded warehouse. Allow space for data system and accessory containers too.
The environment in the storage area should be between 5 °C and 50 °C (41 °F and 122 °F), 20% to 80% relative humidity, non-condensing and non-corrosive.
Do not open any shipping containers until an Agilent Technologies representative is present. Warranty claims for missing items will not be honored unless an Agilent Technologies representative is present to verify the contents of each container as it is unpacked.
The actual shipping containers become your property and should not be returned to Agilent Technologies.

Installation
Once the installation has begun, it should progress in a timely manner to completion. Delays due to inadequate site preparation could cause loss of instrument use during the warranty period. In extreme cases, Agilent Technologies may ask to be reimbursed for the additional time required to complete the installation.
The primary user and, if possible, all other users of the MSD should be present during installation to receive familiarization instruction from the Agilent Technologies service representative.
The final step in the installation process is system verification. Your Agilent Technologies service representative will test the system against Agilent Technologies specifications as documented for the product(s) or system(s) you have purchased. (Not all tests are performed for every system.)
The Agilent Technologies service representative will not test your system against your standards or samples. Further, the Agilent Technologies service representative will not set up your laboratory procedures. Assistance with laboratory procedures can be obtained from your local Agilent Technologies Applications Engineer (AE) on a consulting basis at additional cost.

See the appropriate sensitivity specifications for your instrument type. See the Agilent Technologies Web site at http://www.agilent.com/chem


5975 Series Mass Selective Detectors Site Preparation Guide
This appendix shows the power cords available for the 5975 Series MSD. See page 26 for information about ensuring that the correct power cord is supplied with your system.



Australia/New Zealand, AS 3112-1981 (Part number 8120-1369)
India/South Africa, BS 546 (Part number 8120-4211)

Japan, NEMA L6-20P (Part number G2025-60189)

Power Cords



© Agilent Technologies, Inc.
Printed in USA, January 2008
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.