1 Safety notes Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter. The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture. Do not expose the device to shocks or strong vibrations. The case should only be opened by qualified PCE Instruments personnel. Never use the instrument when your hands are wet. You must not make any technical changes to the device. The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents. The device must only be used with accessories from PCE Instruments or equivalent. Before each use, inspect the case for visible damage. If any damage is visible, do not use the device. Do not use the instrument in explosive atmospheres. The measurement range as stated in the specifications must not be exceeded under any circumstances. Do not touch the heating element or put your hand into the water when the heating element is turned on. During operation, do not touch the tank. This can cause burns. When opening the door, hot and humid air leaks from the device. Do not place any electrical equipment in or on top of the cabinet. Make sure that the power plug is disconnected before cleaning the water tank or changing the water. Wait for the water to cool down before changing it. Mind possible consequences of the leaking water vapour. Condensed water might affect the safety or the functionality of other devices. It might also cause damage to walls, ceilings or floors over time. Wet floors also become slippery. Wear proper safety equipment when using the device, especially gloves. Make sure that the samples are fastened properly. Only open and close the door by using the handle. Place the cabinet on an even surface. Use the adjustable feet and a water level to level it properly. Place the cabinet in an area with as little draft as possible. Do not place it in high- traffic areas in order to minimize draft. Non-observance of the safety notes can cause damage to the device and injuries to the user. We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of business. If you have any questions please contact PCE Instruments. The contact details can be found at the end of this manual.

Safety-related instructions the non-observance of which can cause damage to the device or personal injury carry a safety symbol.

2.1 Technical specifications

Standard Based on AMK.MB-005 standard (04/2015) Temperature control 50 … 55 °C (automatic) Heating power 3000 W Temperature probe PT 100 class A 4 Display 3.5“ TFT touch screen colour display (16 bit) 320 x 240 pixels Memory 1.5 GB (>1 M Readings) Sampling rate Max. 10 Hz (adjustable)
Interfaces USB (for USB sticks) Ethernet (optional) Environmental conditions 0 … +60 °C Power supply 230 V AC / 50 Hz Dimensions 1130 x 720 x 690 mm Weight Approx. 36 kg
2.2 Delivery contents
1 x vapour resistance test chamber PCE-DLT 10 incl. display unit, heating element and temperature probe 4 x fastening rails 4 x holding clips 1 x water tank 1 x instruction manual
2.3 Optional accessories
Safety symbols · cooling frame

3.1 Device


3.2 Display


4.1 Levelling the device
Place the cabinet on an even surface and use the adjustable feet and a water level to level it properly. ATTENTION: Make sure to place the cabinet in an area with as little draft as possible. Do not place it in high-traffic areas.
4.2 Preparation of the sample
Samples should have dimensions of 200 x 100 mm. Further information and requirements can be found in the relevant standards (such as AMK-MB-005 or DIN 68930).
4.3 Test preparation


During operation, do not touch the tank. This can cause burns. 3. Use a fastening rail and two holding clips to fasten the sample. Use the screws on the gripping jaws to fasten the sample properly and to prevent it from getting out of place. 4. Wait for the temperature display to reach the test temperature (50 … 55 °C). The temperature display now turns green.


the yellow vertical adjustment screw to adjust the distance.
5.1 Measurement
By using the vapour resistance test chamber, you can test the vapour resistance of samples based on the AMK-MB-005 standard. To do so, you have to perform three 30-minute tests for each probe and let it rest for 30 minutes between the individual tests. To perform a test, follow these steps:
1. Wait for the temperature display to reach the test temperature (50 … 55 °C). The temperature display now turns green. 2. Press the “Start” button. The 30-minute timer starts and can be seen in the timer display. Press the “Stop” button to stop the timer. Press the “Reset” button to cancel it. 3. Wait for the timer to run out. Once the timer hits “0”, the device emits an audible signal and the timer display starts to flash. Open the door and take out the fastening rail with the sample. Let it rest for 30 minutes. When performing a test, mind the following notes:

5.2 After the measurement
When you have finished all tests, turn off the heating element by using the key switch. After that, turn off the device with the main switch and disconnect the power plug. Open the door to speed up the vapour release. Once the heating element and the water have cooled down, you can remove the water tank and empty it. To do so, loosen the black vertical adjustment screw of the heating element and move the heating element upwards until it is outside the tank. Now you can pull out the water tank to the right side of the vapour resistance test chamber.
5.3 Data transfer
The display unit stores the temperature and the timer count on a regular basis. The data can be transferred via the USB interface. To do so, use a USB pen drive with the following properties:
The power consumption of the USB pen drive should not be higher than 100 mA. Pen drives with a high memory capacity might not be supported. If this is the case, you can use an USB hub with external power supply. However, we recommend using USB pen drives with a memory capacity of 2 GB. The USB pen drive must be formatted in FAT file system (not FAT32!!). To transfer the data to the USB pen drive, follow these steps:
1. Connect the USB pen drive to the USB interface of the display unit. 2. Touch the display and select “Menu”. 3. Select “File management”. 4. Select “Logging files”. 5. Below “File selection”, you can see the different measuring groups. If there are any readings stored in the measuring group, a “click to select” button is shown right next to it. Press the button to select the measuring group you wish to transfer. 6. A new screen appears where you can see all readings which are stored in the selected measuring group. You can select the readings you want to be transferred by selecting them manually or by using the buttons on the bottom left side which select or de-select all readings of the group. If a reading is selected, its checkbox is activated. To confirm the selection, press the check mark button on the bottom right. 7. Now you are back in the “Logging files” screen. Once you have selected the desired measuring groups and readings, press the “Export selected files” button to transfer the data to the USB stick. 8. If you want to delete the selected files, press the “Delete selected files” button.

6 Warranty You can read our warranty terms in our General Business Terms which you can find here: https://www.pce-instruments.com/english/terms
7 Disposal For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose. In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law. For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations. If you have any questions, please contact PCE Instruments.

PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 4 23, rue de Strasbourg Calle Mayor, 53 D-59872 Meschede 67250 SOULTZ-SOUS-FORETS 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548 Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542 info@pce-instruments.com [email protected] [email protected] www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol United States of America United Kingdom Italy
PCE Americas Inc. PCE Instruments UK Ltd PCE Italia s.r.l. 711 Commerce Way suite 8 Units 12/13 Southpoint Business Park Via Pesciatina 878 / B-Interno 6 Jupiter / Palm Beach Ensign Way, Southampton 55010 LOC. GRAGNANO 33458 FL Hampshire CAPANNORI (LUCCA) USA United Kingdom, SO31 4RF Italia Tel: +1 (561) 320-9162 Tel: +44 (0) 2380 98703 0 Telefono: +39 0583 975 114 Fax: +1 (561) 320-9176 Fax: +44 (0) 2380 98703 9 Fax: +39 0583 974 824 info@pce-americas.com info@industrial-needs.com [email protected] www.pce-instruments.com/us www.pce-instruments.com/english www.pce-instruments.com/italiano The Netherlands Chile Hong Kong
PCE Brookhuis B.V. PCE Instruments Chile S.A. PCE Instruments HK Ltd. Institutenweg 15 RUT: 76.154.057-2 Unit J, 21/F., COS Centre 7521 PH Enschede Santos Dumont 738, local 4 56 Tsun Yip Street Nederland Comuna de Recoleta, Santiago, Chile Kwun Tong Telefoon: +31 (0) 900 1200 003 Tel. : +56 2 24053238 Kowloon, Hong Kong Fax: +31 53 430 36 46 Fax: +56 2 2873 3777 Tel: +852-301-84912 [email protected] [email protected] jyi@pce-instruments.com www.pce-instruments.com/dutch www.pce-instruments.com/chile www.pce-instruments.cn China Turkey
PCE (Beijing) Technology Limited PCE Teknik Cihazları Ltd.Şti. 1519 Room, 4 Building Halkalı Merkez Mah. Men Tou Gou Xin Cheng Pehlivan Sok. No.6/C Men Tou Gou District 34303 Küçükçekmece - İstanbul 102300 Beijing Türkiye China Tel: 0212 471 11 47 Tel: +86 0755-32978297 Faks: 0212 705 53 93 [email protected] info@pce-cihazlari.com.tr www.pce-instruments.cn www.pce-instruments.com/turkish
Germany France Spain

русский, 中文 )
can be downloaded here: www.pce-instruments.com
Specifications are subject to change without notice.
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.