Instructions for use



The following symbols are used on the packaging and on the type plate for the device and accessories.

General notes


Prior to initial use
1 On/Off switch 2 Battery status indicator 3 Micro USB port 4 Holder for atomiser 5 Filter cap with filter 6 Hose connector
Overview of atomiser and accessories

8 Compressed air hose 9 Atomiser 10 Medication container 11 Atomiser insert 12 Mouthpiece 13 Adult mask 14 Children’s mask 15 Nosepiece with comfort attachment 16 Replacement filter 17 USB power mains part 18 USB cable
Repairs

Before using the device for the first time
During the charging procedure, between one (battery al- most empty) and four (battery fully charged) LEDs on the battery status indicator [2] flash according to the battery charge status. As soon as the battery is fully charged, all four LEDs are permanently illuminated. After each use, fully charge the battery of the nebuliser again.
Switching on the nebuliser To turn on the nebuliser, proceed as follows: • Press the ON/OFF button on the nebuliser. The ON/ OFF button illuminates blue. The nebuliser is now op- erational. The battery must be sufficiently charged for the nebuliser to function correctly.
The blue LEDs on the battery status indicator show the current battery state.
1 blue LED flashes = Battery is almost flat, nebu-
liser will soon switch off.
1 blue LED lights up = Battery capacity 25%
2 blue LEDs light up = Battery capacity 50%
3 blue LEDs light up = Battery capacity 75%
4 blue LEDs light up = Battery capacity 100%

In addition to the accompanying adult [13] and children’s mask (from 2 years of age) [14], you also have the option of using baby mask (from ages 0-3 years) with your ne- buliser. The baby mask is not included in delivery with the IH 60. You can find an overview of all replacement items in chapter „12. Replacement parts and wearing parts“.
The most effective form of nebulisation is by using the mouthpiece. Nebulisation using a mask is only recom- mended if it is not possible to use a mouthpiece (e.g. for children who are not yet able to inhale medication using a mouthpiece). If you are using the adult mask [13], you can secure it to your head with the adult mask strap. There is also a strap for the children’s mask [14]. When using a mask to inhale, take care to ensure the mask fits well and the eyes are unobstructed.
The device is not suitable for continuous operation; after 30 minutes of operation it must be switched off for 30 minutes.
6.3 Closing the atomiser • Close the atomiser [9] by twisting the top clockwise against the medication container [10]. Ensure that the connection is correct.
During the treatment, hold the atomiser straight (vertical- ly), otherwise the atomisation will not work and faultless functionality is not guaranteed. Depending on the battery charge status and accessory used, the inhalation time stated may vary.
6.4 Connecting accessories to the atomiser • Connect the atomiser [9] to the desired accessory (mouthpiece [12], adult mask [13], children’s mask [14] or nosepiece [15]).

To replace the filter, proceed as follows:
Essential oils, cough syrups, gargling solutions and drops to be used as a rub or in a steam bath are wholly unsuit- able for inhalation using a nebuliser. These additions are often viscous and can impair the correct functioning of the device and therefore the effectiveness of the applica- tion in the long-term. For individuals with a hypersensitive bronchial system, medications containing essential oils may under certain conditions cause an acute bronchos- pasm (a sudden cramp-like restriction of the bronchi with shortness of breath). Consult your doctor or pharmacist in relation to this matter.
If the filter remains in the device after the cap has been removed, take the filter out of the device, e.g. with twee- zers or similar.
6.7 Cleaning See “Cleaning and disinfection” on page 8.
Adhere to the following hygiene instructions to avoid health risks. • The atomiser [9] and accessories are designed for mul- tiple use. Please note that different areas of application involve different requirements in terms of cleaning and hygienic preparation.

The device must be switched off, disconnected from the mains and allowed to cool down each time before cleaning.
The atomiser and the accessories used such as the mouthpiece, mask, etc. must be washed with hot but not boiling water after each use. Dry the parts carefully using a soft cloth. Put the parts together again when they are completely dry and place them in a dry, sealed container or disinfect them.
When cleaning, ensure that any residue is removed. Never use any substances for cleaning that could potentially be toxic if they came into contact with the skin or mucous membranes, or if they were swallowed or inhaled.
Use a soft, dry cloth and non-abrasive cleaning products to clean the device.
Do not use any abrasive cleaning products and never submerge the device in water.
Do not boil or autoclave the compressed air hose or the masks.
Cleaning

The empty, completely flat rechargeable batteries must be disposed of through specially designated collection boxes, recycling points or electronics retailers. You are legally required to dispose of the rechargeable batteries. The codes below are printed on rechargeable batteries containing harmful substances: Pb = Battery contains lead, Cd = Battery contains cadmium, Hg = Battery contains mercury
Drying • Place the individual parts on a clean, dry and absorbent surface and leave them to dry completely (at least 4 hours).

Type IH 60 Dimensions (W x H x D) 110 x 62 x 47 mm Weight 222 g Operating pressure Approx. 0.25 to 0.5 bar Atomiser filling volume
Min. 2 ml Max. 6 ml Medication flow rate Approx. 0.25 ml/min Sound pressure Max. 45 dBA (acc. to DIN EN 13544-1 section 26) Mains connection Input: 100 – 240 V~; 50 – 60 Hz; 0.5 A
Temperature: +10 °C to +40 °C Relative humidity: 10% to 95% Ambient pressure: 700 to 1060 hPa Storage and transportation conditions
Temperature: -20 °C to +60 °C Relative humidity: 10% to 95% Ambient pressure: 700 to 1060 hPa Aerosol properties 1) Aerosol delivery: 0.15 ml
2570 mAh 3.7 V Li-Ion
The serial number is located on the device or in the bat- tery compartment.
Subject to technical changes.

ments of European Directive 93/42/EEC for medical devices, as well as those of the Medizinproduktegesetz (German Medical Devices Act).
Measurements were performed using a sodium fluoride solution with a “Next Generation Impactor” (NGI). The diagram may therefore not be applicable to suspensions or highly viscous medications. You can obtain more detailed information from the manufacturer of your medication. 12. Replacement parts and wearing parts
If the device is not used according to the instructions specified, perfect functionality cannot be guaranteed! We reserve the right to make technical changes to improve and develop the product. This device and its accesso- ries comply with the European standards EN60601-1, EN60601-1-2 (CISPR 11, IEC61000-3-2, IEC61000-3-3, IEC61000-4-2, IEC61000-4-3, IEC61000-4-4, EC61000-4-5, IEC61000-4-6, IEC61000-4-7, IEC61000-4-8, IEC61000-4-11) and EN13544-1 and are subject to particular precautions with regard to electro- magnetic compatibility. This device meets the require-
Beurer guarantees the perfect functionality and complete- ness of this product.
The worldwide warranty period is 5 years, commencing from the purchase of the new, unused product from the seller.
The warranty only applies to products purchased by the buyer as a consumer and used exclusively for personal purposes in the context of domestic use. German law shall apply.
Particle size diagram

During the warranty period, should this product prove to be incomplete or defective in functionality in accordance with the following provisions, Beurer shall carry out a re- pair or a replacement delivery free of charge, in accord- ance with these warranty conditions.
If the buyer wishes to make a warranty claim, they should approach their local retailer in the first in- stance: see the attached “International Service” list of service addresses.
The buyer will then receive further information about the processing of the warranty claim, e.g. where they can send the product and what documentation is required.
A warranty claim shall only be considered if the buyer can provide Beurer, or an authorised Beurer partner, with - a copy of the invoice/purchase receipt, and - the original product.
The following are explicitly excluded from this warranty: - deterioration due to normal use or consumption of the
product; - accessories supplied with this product which are worn
out or used up through proper use (e.g. batteries, rechargeable batteries, cuffs, seals, electrodes, light sources, attachments and nebuliser accessories); - products that are used, cleaned, stored or maintained
improperly and/or contrary to the provisions of the in- structions for use, as well as products that have been opened, repaired or modified by the buyer or by a ser- vice centre not authorised by Beurer; - damage that arises during transport between manu-
facturer and customer, or between service centre and customer; - products purchased as seconds or as used goods; - consequential damage arising from a fault in this prod-
Subject to errors and changes
uct (however, in this case, claims may exist arising from product liability or other compulsory statutory liability provisions).
Repairs or an exchange in full do not extend the warranty period under any circumstances.





Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| The atomiser produces no or too little aerosol. | 1. Too much or too little medication in the atomiser. Minimum: 2 ml, Maximum: 6 ml. | p.10 | |
| The atomiser produces no or too little aerosol. | 2. Check nozzle for blockages. Clean nozzle if necessary (e.g. by rinsing out). Then start using the atomiser again. IMPORTANT: Carefully pierce the fine holes from the underside of the nozzle only. | p.10 | |
| The atomiser produces no or too little aerosol. | 3. Atomiser not held vertically. | p.10 | |
| The atomiser produces no or too little aerosol. | 4. U nsuitable medication fluid added for nebulisation (e.g. too viscous). The medication fluid should be pre- scribed by the doctor. | p.10 | |
| The output is too low. | Kinked hose, clogged filter, too much inhalation solution. | p.10 | |
| What med- ications are suitable for inhaling? | Please consult your doctor in rela- tion to this matter. As a rule, all medication that is suitable and approved for device inhalation can be inhaled. | p.10 | |
| There is inha- lation solution residue in the atomiser. | This is normal and is due to technical reasons. Stop inhalation once the atomiser starts to make a notably different sound. | p.10 | |
| What should be taken into account when using the device with infants and children? | 1. On infants and children, the mask should cover the mouth and nose to ensure effective inhalation. | p.11 | |
| What should be taken into account when using the device with infants and children? | 2. On children, the mask should also cover both the nose and mouth. It is not a good idea to carry out nebulisation on someone who is sleeping, as in this case not enough of the medication will reach the lungs. | p.11 | |
| What should be taken into account when using the device with infants and children? | Note: Inhalation should only be car- ried out under the supervision of an adult and with their assistance and the child should not be left alone. | p.11 | |
| Inhalation us- ing the mask takes longer. | This is due to technical reasons. You breathe less medication per breath through the holes of the mask than via the mouthpiece. The aerosol is mixed with ambient air via the holes. | p.11 | |
| Why should the atomiser be replaced regularly? | There are two reasons for this: 1. To ensure a therapeutically effec- tive particle spectrum, the nozzle hole must not exceed a specific diameter. Due to mechanical and thermal stresses, the plastic is subject to a certain degree of wear. The atomiser insert [11] is particularly sensitive. This can also alter the composition of the aero- sol droplets, which directly affects the effectiveness of the treatment. 2. Regularly changing the atomiser is also recommended for hygiene reasons. | p.11 | |
| Should each person have their own atomiser? | Yes, this is absolutely essential for hygiene reasons. | p.11 |