2 mm Allen Wrench 4.5 mm Allen Wrench Phillips Screwdriver
The Counting Feature is Not Legal for Trade
OK
LOW
CAL Current Weight Total Weight
Column x 1 Platform Base x 1 RS-232 Adapter x 1 Display Indicator x 1
AC Adapter x 1
The scale should always be used in an environment that is free from excessive air currents, corrosives, vibration and temperature or humidity extremes. These factors will affect displayed weight readings.

Avoid placing the scale next to or near:

REMOVING THE SHIPPING SCREW
ASSEMBLING THE COLUMN
Cable Socket Cap Bolt
center of the indicator. (See Figure 4)
wrench and with the socket cap bolt using a 4.5 mm Allen wrench, but do not tighten.
moved to a new location to verify scale is balanced.

ATTACHING INDICATOR TO THE COLUMN
CONNECTING POWER
The scale can operate continuously by connecting the AC adapter to an outlet or operate eight hours on a fully charged battery.
Plug the AC adapter into the AC adapter socket on the back of the indicator. Plug the AC adapter plug into a standard 110 volt outlet.

The scale is equipped with a rechargeable battery.
When the battery voltage is low, a battery symbol will appear in the lower left-hand corner of the display. After 15 minutes, the backlight will start to flicker and the display will show Bat lo. The scale will automatically turn off 30 minutes after the symbol appears.

When the battery indicator ( ) appears in the lower left-hand corner of the display, the battery needs to be recharged.
Plug the AC adapter into a power source to charge the battery. Scale does not need to be turned on. It will take 12 hours to fully charge the battery.
The charging indicator is located on the left side of the display, above the word CHARGE. The indicator light color will change to indicate the battery status:
Recharge battery every three months when not in use.
If the scale is not used for an extended period of time, remove the battery from the battery compartment to avoid leakage. Store the battery in a sealed bag or box in a dry, temperate environment.

DISPLAY AND KEYPAD

8 9 14 16 15 10 11 12 13
15 ON/OFF Key

PRIOR TO USE
Single point calibration uses two calibration points, zero and full weight capacity of the scale.
and then the TARE key . Display will show .
ON/OFF key is located in the lower right hand corner of the display.
TARE key until display shows sigr.
TARE key to select cal.
Press the UNIT key until the desired weighing unit appears – lb, oz or kg.
the scale. Press the G/N key and the
The scale is pre-calibrated during production. You must use a precision test weight to properly calibrate the scale. The test weight CANNOT exceed the scale's capacity.
ACCT/TTL/Print key to change the active digits
and press the TARE key to change the value.
key to confirm.

If there is a minor weight displayed without anything
on the platform, press the ZERO key to clear the display. Zero indicator will be shown on the display.
TARE
The minimum weight must be at least 20 times the scale accuracy for accurate calculations. Reference the table below for minimum weights.
When weighing an item that is in a container, taring stores the container weight to memory so only the item weight is displayed.
To store tare weight into memory and set display to zero:
If the unit weight is too light, another scale is needed. Continuing to use this scale will result in less accurate results.
Use this mode to set a number for a weighing check. When the weight of the objects on the platform is between the preset checking numbers, the alarm sound beeps repeatedly.
Use this mode to weigh items.
PCS/SET key simultaneously. SEt H will be shown
on the display.
Use this mode to count parts of uniform weight.
The scale cannot determine how many pieces are on a
ACC/TTL key to move the active digit left and right.
platform without a sample.
Use the TARE key to increase the numeric value.

Use this mode to measure the cumulative weight or counting values of a sequence of items. The maximum number of samples is 99 entries or the scale weight capacity.
ACC/TTL key to move the active digit left and right.
Use the TARE key to increase the numeric value.
press the ACC/TTL Print key . ACC 1 will be shown on the display and the total value will be displayed for three seconds. Weighing or counting value will be stored into memory. If a printer is connected, the data will print.
press the ACC/TTL Print key . New data will be added to memory. Continue until all data is added.
ACC/TTL Print key to display the total value of items stored to memory. The number of accumulated values will be displayed, ACC #, followed by the total accumulated value for three seconds.
G/N key and the ACC/TTL Print key at the same time.

RS-232
Use this application to automatically add weighing and counting values to the scale memory and to print values if a printer is connected.
To use the RS-232 port, connect the gray RS-232 adapter to the gray RS-232 cable on the display indicator. (See Figure 9)
When in the automatic accumulation application, the auto indicator will be shown in the upper right hand corner of the display.
RS-232 output of weighing data:
Use this mode to weigh moving items.
9 Pin D Connector
ZERO key at the same time to enter the dynamic weighing mode. The hold indicator will appear in the upper right hand corner of the display.
ZERO key at the same time to exit the dynamic weighing mode.
, k g CR LF
-- HEADER 1 -- -- HEADER 2 -- -- WEIGH DATA -- -- WEIGH UNIT -- TERMINATOR
c s p CR LF P C S :
HEADER1: ST=STABLE, US=UNSTABLE HEADER2: NT=NET, GS=GROSS

Con2:
When the indicator is used in trade or a legally controlled application, it must be set up, verified and sealed in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met.
ht6 Tare1 Tare2 Tare3 Tare4 Tare5 Tare6 Termin
Header0=02H Header1 follow decimal point Decimal point=0, header1=22H Decimal point=1, header1=23H Decimal point=2, header1=24H Decimal point=3, header1=25H Decimal point=4, header1=26H
Header2 follow weigh status, default value=20H
If in net mode (tare value not 0), header2=header2|01H If gross weight "-", header2=header2|02H
If overload or gross weight "-", header2=header2|04H

If weighing unit=kg, header2=header2|10H
Header3 follow weighing unit
If weighing unit=g, header3=21H
If weighing unit=oz, header3=23H
Con3:
7 Unit1 Unit2 Status Termin
Header1 follow weight "+" or "-"
The local weights and measures official or authorized service agent must perform the verification procedure.
When weight "+", header1="+", when weight "-", header="-"
The local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings.
BACKLIGHT
At startup, the backlight will automatically turn on when the scale is in use. Enter this menu to change the backlight setting.

AUTO POWER-OFF
ENABLE DYNAMIC WEIGHING
When the auto power-off time is enabled, the scale will automatically turn off when there has been no load on the platform and the scale is stable at zero position for a specified period of time (in minutes). At startup, the auto power-off is disabled. Enter this menu to turn the auto power-off feature on or off.
At startup, the dynamic weighing mode is enabled. Enter this menu to turn the dynamic weighing mode on or off.
then the TARE key . Po CHF will be shown on the display.
At startup, the beeping sounds are set to beep when weights are inside the limits. Enter this menu to change the beeping sound setting.
At startup, the accumulation mode is enabled. Enter this
menu to turn the accumulation and auto accumulation modes on or off and to set the RS-232 communication.

ACC/TTL Print key is pressed.
Enter this menu to select a printer.
time to printer when the ACC/TTL Print key is pressed.
the TARE key . Po CHF will be shown on the display.
Baud rate is the RS-232 transmission speed. If using a printer, both the printer and the scale must be set to the same baud rate. At startup, the baud rate is set at 9,600. Enter this menu to change the baud rate setting
then the TARE key . Po CHF will be shown on the display.

CLEANING
Use a stainless steel cleaning solution for the stainless steel indicator housing and platform.
Apply cleaner to a clean, damp cloth and wipe surface.
Dry thoroughly.
Use a mild detergent for the display and keypad.