User's manual Bedienungsanleitung
Libretto d’istruzioni Manual de instrucciones
ENGLISH - page 16 Chapter II
X03051A00 - Ed. 04 - 02/13
Copyright ©



1 2 3 4
Multimètre digital portable

LEGENDE / CAPTION / BESCHREIBUNG / LEGENDA / LEYENDA
1 Borne d'entrée calibres 11, 12, 13, 14, 15 11 Mesure de tensions alternatives 2 Entrée de référence du multimètre 12 Mesure de tensions 500 mV 3 Borne d'entrée calibres mA 13 Mesure de tensions continues 4 Borne d'entrée calibre 10 A 14 Mesure de résistance 5 Mise sous tension (sélection fonctions secondaires) 15 Mesure de capacité 6 Changement de gamme 16 Mesure de courant jusqu’à 5 mA 7 Mesure en mode relatif 17 Mesure de courant jusqu’à 50 mA 8 Mesure de crêtes 18 Mesure de courant jusqu’à 500 mA 9 Gel de l'affichage 19 Mesure de courant jusqu’à 10 A 10 Mise hors tension 20 Envoi de données vers imprimante
1 Input terminal, ranges 11, 12, 13, 14, 15 11 AC voltage measurement 2 Multimeter reference input 12 500 mV voltage measurement 3 Input terminal, ranges mA 13 DC voltage measurement 4 Input terminal range 10 A 14 Resistance measurement 5 Power on (selects secondary functions) 15 Capacitance measurement 6 Range change 16 Current measurement up to 5 mA 7 Relative mode measurement 17 Current measurement up to 50 mA 8 Peak measurement 18 Current measurement up to 500 mA 9 Display hold 19 Current measurement up to 10 A 10 Power off 20 Data sending to a printer
1 Eingangsbuchse für Meßbereiche 11,12,13,14,15 11 Messung von AC-Spannungen 2 COM-Eingangsbuchse 12 Spannungsmessung bis 500 mV 3 Eingangsbuchsen mA 13 Messung von DC-Spannungen 4 Eingangsbuchse 10 A 14 Widerstandsmessung 5 Multimeter Einschalten (+ Zweitfunktionen) 15 Kapazitätsmessung 6 Bereichsumschaltung 16 Strommessung bis 5 mA 7 Relativ-Messung 17 Strommessung bis 50 mA 8 Spitzenwertmessung 18 Strommessung bis 500 mA 9 Anzeige speichern 19 Strommessung bis10 A 10 Multimeter Ausschalten 20 Sendung von Angaben zum Drucker
1 Borne de entrada calibres 11, 12, 13, 14, 15 11 Medida de tensiones alternativas 2 Entrada de referencia del multÍmetro 12 Medida de tensiones 500 mV 3 Borne de entrada calibres mA 13 Medida de tensiones continuas 4 Borne de entrada calibre 10 A 14 Medida de resistencias 5 Puesta en servicio (+ funciones segundarias) 15 Medida de capacidades 6 Cambio de calibre 16 Medida de corrientes hasta 5 mA 7 Medidas relativas 17 Medida de corrientes hasta 50 mA 8 Medidas de cresta 18 Medida de corrientes hasta 500 mA 9 Memorizacion de la representacion visual 19 Medida de corrientes hasta 10 A 10 Puesta fuera de servicio 20 Transmisión de los datos hacia una impresora
1 Boccola d’ingresso portate 11, 12, 13, 14, 15 11 Misura delle tensioni alternate 2 Ingresso di riferimento del multimetro 12 Misura delle tensioni 500 mV 3 Boccola d’ingresso portate mA 13 Misura delle tensioni continue 4 Boccola d’ingresso portata 10 A 14 Misura di resistenza 5 Accensione (scelta funzioni secondarie) 15 Misura della capacità 6 Cambiamento di portata 16 Misura di corrente fino a 5 mA 7 Misura in modalità relativa 17 Misura di corrente fino a 50 mA 8 Misura delle creste 18 Misura di corrente fino a 500 mA 9 Immobilizzazione della visualizzazione 19 Misura di corrente fino a 10 A 10 Spegnimento 20 Invio dati verso la stampante
Multimètre digital portable



You have just acquired a 5000 count-portable digital multimeter and we thank you for your confidence. This instrument complies with the specification of EN publication 61010-1, concerning safety requirements for electronic measuring apparatus. To get the best service from this instrument, read carefully this user’s manual and respect the detailed safety precautions. 1.1. Precautions and safety measures
1.1.1. Before use
- in an environment with pollution level 2, - at an altitude of less than 2000 m, - at a temperature between 0°C and 50°C, - with relative humidity of less than 80 % up to 40°C.
- Measurement category III for voltages no higher than 600 V (AC or DC)
in relation to the earth and between input terminals. - Measurement category IV for voltages no higher than 600 V (AC or DC)
in relation to the earth and between input terminals.
connected to mains. E.g.: protected electronic circuits CAT II : Measurement category II is for measurements performed on circuits directly connected
to the low voltage installation. E.g.: power supply to domestic appliances and portable tools. CAT III : Measurement category III is for measurements performed in the building installation E.g.: machine or industrial apparatus power supply. CAT IV : Measurement category IV is for measurements performed at the source of the low-
voltage installation. E.g.: energy inputs ∗ For your own safety, only use the measuring probes which have been delivered with the
instrument : they conform to the EN 61010-031 safety standard. Before use, check that they are in good condition. ∗ The safety of any system incorporating this instrument is the responsibility of the system
assembler. 1.1.2. During use
∗ Never exceed the protection limit values indicated in the specifications for each type of
measurement. ∗ When the multimeter is linked to measurement circuits, do not touch unused terminals. ∗ When the scale of the value to be measured is unknown, check that the scale initially
set on the multimeter is the highest possible or, wherever possible, choose the autoranging mode. ∗ Before changing functions, disconnect the test leads from the circuit under test.

∗ When performing current measurements, never change of range, do not connect or
disconnect leads without first isolating the current. If you do, there is a risk of generating surge currents which can blow the fuses or damage the instrument.
∗ In TV repair work, or when carrying out measurements on power switching circuits,
remember that high amplitude voltage pulses at the test count can damage the multimeter. Use of a TV filter will attenuate any such pulses.
∗ Never perform resistance, condenser and diode measurements on live circuits.
1.1.3. Symbols

Selective sorting of waste for recycling electric and electronic materials. In accordance with the WEEE 2002/96/EC directive : must not be treated as household waste.
Earth Dual insulation DANGER: Risk of electrical shock Consult the manual.
1.1.4. Opening the instrument
1.2. Protection devices ASYC II series instruments are fitted with various protection devices :
∗ Varistor protection for limiting transients of over 1500 Vpk at the VΩ terminal,
particularly 8 kV pulse streams as defined in French standard EN 61010-1 relative to safety.
∗ A PTC (Positive Temperature Coefficient) resistor protects against overvoltages up to
600 V during resistance, capacitance and diode measurements. This protection is reset automatically once overload is over.
∗ Two fuses provide protection during measurements of intensity type.
∗ IP protection rating of 67

1.3. Safety devices
∗ The battery unit and fuses cannot be accessed without first disconnecting the measuring
leads.
∗ When measuring voltages above 24 V, the symbol blinks on the display. ∗ If the maximum range is repeatedly exceeded, an intermittent audible signal indicates the
risk of electric shock. 1.4. Warranty
This equipment is warranted against any defects of manufacture or materials according to the general conditions of sale. During the warranty period (3 years), defective parts will be replaced, the manufacturer reserving the right to repair or replace the product. In the event of the equipment being returned to the after sale department or to a local agency, carriage to the centre shall be payable by the customer. The warranty does not cover the following :
The contents of this manual must not be reproduced in any form whatsoever without the consent of the manufacturer.
1.5. Maintenance and metrological verification Return your instrument to your distributor for any work to be done within or outside the guarantee. 1.6. Unpacking - Repacking
This equipment has been fully checked out mechanically and electrically before shipping. All precautions have been taken to ensure that the instrument arrives at its destination undamaged.
However, it is advisable to carry out a rapid check for damage sustained in shipping. If there is any evidence of damage, make this known immediately to the shipper. * Caution Should you need to return the multimeter, preferably use the original packaging and indicate the reasons as clearly as possible on an accompanying note.

This multimeter is one of the ASYC II (Advanced SafetY Concept, third generation) range, designed for a high degree of user safety, maximum protection and unrivalled performance.
2.1. Selector switch
It is a standalone, handheld professional measuring instrument, capable of measuring the following quantities (accessed by the eleven-position rotary selector switch) :
2.2. Keypad
A six-key keypad lets you :
2.3. Display
The display provides :
2.4. Power supply
This multimeter is powered by a standard 9 V battery which provides approximately 500 hours of operation.
2.5. Input terminals
Measurements are performed using two measuring leads supplied with the instrument connected to input terminals 1, 2, 3 and 4, as indicated in section 3.1.

3.1. Connecting the test leads
Connect the black lead to the COM socket (for all measurements). Depending on the position of the selector switch, connect the red lead as follows:
3.2. Switching on the instrument
Turn the selector switch to the required function. All segments of the display come on for a few seconds. The instrument is then ready for measurements.
3.3. Switching off the instrument
The instrument can be switched off manually by returning the selector switch to the OFF position, or automatically after approximately half an hour if no key is pressed or the switch is not operated.
) Note Automatic shutdown of the instrument is disabled in order to avoid interrupting the peak value measurements (Pk +/-), or the data print out.
For user safety, automatic shutdown is also disabled when a measured magnitude (voltage/current) present at the input exceeds dangerous level
( indicator displayed).
3.4. Special configurations
To adapt the configuration of the instrument to the measurement environment, the user can choose 50 Hz or 60 Hz rejection :
Switch on with the rotary switch while holding down the HOLD key.
The selection :

3.5. Multimeter maintenance
3.5.1. Fuse self-test
When fuse F1 (0.63 A) or F2 (11 A) is blown, the display shows “FUSE.1” or “FUSE.2”, accordingly.
If both fuses are blown, the display shows “FUSES”.
Replace the fuse or fuses concerned. ) Note Fuse F1 cannot be tested unless the switch is set to a mA position. Fuse F2 is located in the common circuit. Therefore, all measurements become impossible when it is out of service.
3.5.2. Battery self-test
When the BAT indication appears on the display, the instrument still has approximately 50 hours of operation, but specifications can no longer be guaranteed.
Replace the battery.
3.5.3. Replacing the battery or fuses
Disconnect the cords from the measured circuit, then multimeter.
Open the multimeter casing as follows (see last page of the manual) :
Before use, make sure that the gasket, then the front cover are carefully set back on the instrument.
3.5.4. Cleaning
Clean the multimeter using a damp cloth. Do not use abrasives or solvents.

This can be used to switch on the multimeter again after an automatic shutdown. It can also be used to access secondary functions associated with the selector switch positions. The flowcharts below define these various functions.
4.1.1. VAC position
4.1.2. mV position
SEL/ON

4.1.3. VDC position
SEL/ON
4.1.4. Ω position
SEL/ON
4.1.5. position
SEL/ON

4.1.6. 5 mA / 50 mA / 500 mA / 10 A positions
SEL/ON
SEL/ON
4.2. RANGE key
4.3. REL key
from subsequent measurements.
Long press : When in REL mode, a long press displays the reference being used. This value may be adjusted using the SEL/ON key (selection of digits and a sign) and the RANGE key (increment the selected digit).
4.4. Pk +/- key
The fast positive or negative peak measurement functions (≥ 1 ms) can be accessed by repeatedly pressing this key in the VDC, mVDC, 5 mADC, 50 mADC, 500 mADC, and 10 ADC functions.
DC current measurement · AC current measurement · AC+DC current measurement

4.5. HOLD key
Short press : Sets the display on the current value.
Long press : Accesses or quits the “autostore” mode. Can be accessed in the VDC, mV, VAC positions.
Set the probes on the point to be measured. An audio signal indicates if the measurement is stable. When you remove the probes, a second audible signal indicates that this stable value displayed has been stored.
4.6. PRINT key
This key operates with the optional serial interface for printer or PC :
Short press : Activates / deactivates « send measures to printer » mode at the rate defined by the user.
Long press : Adjusts the rate varying from 00000 s (a single transmission) up to 9h 59min 59s, using the SEL/ON key (selection of digits) and the RANGE key (increment the selected digit).
Autostore

Only those values assigned tolerances or limits are guaranteed values. Values without tolerances are given for information only (French standard NFC 42670). The technical specifications are guaranteed only after 30 min warm-up period. Except special indication, they are valid from 5 to 100 % of the range of measurement. {Accuracy : ‘’n%R + nD’’ means ‘’n% of the reading + n digits’’ as per IEC 485}
5.1. DC voltages

mV 500 mV 10 MΩ ± 850 VPK∗ 100 µV 5 V 11 MΩ 1 mV 50 V 0.1% R + 2D 10 mV 500 V 0.2% R + 2D 100 mV VDC
600 V 0.3% R + 2D 10 MΩ ± 850 VPK
Number of count : 5000 Range selection : automatic or manual for the 5V, 50V, 500V, 600V ranges Common mode rejection : at 50 and 60 Hz, better than 120 dB Serial mode rejection : at 50 and 60 Hz, better than 60 dB Additional error in Pk +/- mode for a pulse of ≥ 1 ms: 1% R ± 50 D For measurements performed on alternative signals, the selected range must tally with the max. value of the signal peak.

200 to 4 kHz 4 to 10 kHz 10 to 30 kHz 30 to 50 kHz
10 % typ.∗∗∗ //////////// 10 MΩ/1GΩ
// 100 pF∗∗ ±850VPK 100 µV
5 V 11 MΩ // 100 pF 1 mV
2 % R + 3 D 3 % R + 3 D
1,5 % R + 3 D∗∗∗∗ //////////////// /////////////
∗AC + DC only ∗∗20 Hz to 40 Hz = 1,5 % ∗∗∗το 20 kHz ∗∗∗∗up to 1 kHz ∗∗∗∗∗ applicable max. voltage : 600 VRMS CAT IV Max. Frequency = 15 000 [V ∗ kHz] / Input [V]
Number of count : 5000 Range selection : automatic or manual for the 5V, 50V, 500V, 600V∗ ranges Common mode rejection : at 50 and 60 Hz, better than 80 dB
Additional error according to crest factor : 0 % for a crest factor < 1.5 1 % for a crest factor of 1.5 to 2 4 % for a crest factor of 2 to 3

5.3. DC current

5 mA 1 µA 50 mA 700 mV 10 µA 500 mA 0.2 % R + 2 D 1.5 V F1 + F2
100 µA 10 A∗∗ 0.5 % R + 5 D 500 mV
F2 10 mA ∗ refer to fuse specifications section 6.1.1. ∗∗ acceptance overload: 20 A during 30 s. max. observing a break > 5 min between 2 tests
Number of count : 5000 Additional error in Pk +/- mode for a pulse width of ≥ 1 ms : 1 % R ± 50 D For measurements performed on alternative signals, the selected range must tally with the max. value of the signal peak.
5.4. AC currents (AC and AC + DC)

40 Hz to 5 kHz 5 kHz to 20 kHz 20 kHz to 30kHz Ranges
5 mA 1%R+3D 1.5% typ. 4% typ. 1 µA 10 mA
50 mA 1.5% typ. 4% typ. 10 µA 100 mA 700 mV
500 mA 1%R +3 D 1.5% typ. 4% typ.
10 A∗∗ 1.5%R+3 D Æ 2 kHz 1.5% typ. ///////////
∗ refer to fuse specifications section 6.1.1. ∗∗ acceptance overload: 20 A during 30 s. max. observing a break > 5 min between 2 tests
Specification : from 5 % to 100 % of the range Number of count : 5000 Additional error according to crest factor : 0 % for a crest factor < 1.5 1 % for a crest factor of 1.5 to 2 4 % for a crest factor of 2 to 3 Additional error in IAC+DC with a direct current at input : 1 %
5.5. Resistance / Continuity

Ranges Accuracy Measur. current Protection∗ Resolution 500 Ω/ 0.1% R + 5 D 1 mA 100 mΩ 5 kΩ 100 µA 1 Ω 50 kΩ 10 µA 10 Ω 500 kΩ 0.1% R + 3 D 1 µA 100 Ω 5 MΩ∗∗ 0.3% R + 3 D 100 nA 1k Ω 50 MΩ∗∗ 1% R + 3 D 50 nA
10 k Ω ∗ Overload protection can be reset automatically. ∗∗ The use of very short and shielded test leads is highly recommended for measurements in this range (> 1 MΩ).
Number of count : 5000 Range selection : automatic or manual (set in continuity mode) Maximum open circuit voltage : 7 V Detection threshold in continuity mode : 10 Ω to 20 Ω Response time in continuity mode : 1 ms

5.6. Capacitance ) Note Discharge all capacitors before taking measurements.

50 nF∗∗ 100 nA 10 pF 500 nF 1 µA 100 pF 5 µF 10 µA 1 nF 50 µF 100 µA
1.5 s 100 nF 5000µF 3 s/mF 1 µF 50 mF 2% R + 2D 1 mA 3 s/mF
10 µF ∗ Overload protection can be reset automatically. ∗∗ It is highly recommended to use very short and shielded test leads for measurements in
this range.
5.7. Diode threshold voltage measurement
Measurable voltages : 0 to 2 V Measurement current : 1 mA typical Resolution : 1 mV Protection : 600 VRMS, can be reset automatically
5.8. Frequencies
Selector switch setting : VAC + SEL/ON Measurement range : 0.62 Hz to 500 kHz Accuracy : 0.03 % Protection : 600 VRMS Display : 50.000 count
Range Sensitivity 5 V to 500 V 600 V 0.62 Hz to 5 kHz ∗ 2 % of range 100 V 5 kHz to 50 kHz 5 % of range 250 V 50 kHz to 500 kHz 10 % of range //////////////// ∗ Rectangular signals
5.9. Duty cycle %+, %-
T x 100 % + = θ
% - = θ
T
Resolution : 0.01 % Min. duration for θ or T-θ : 2 µs Max. duration for T : 0.8 s Min. duration for T : 100 µs
4 10 .3 −
Sensitivity : see frequency measurement mode Absolute additional error on θ, due to the zero crossing slope :
0.1 x P C
with C : range in V (C = 5000 V for 600VDC or 600 VAC ranges) with P : slope in V/s
θ · θ · T x 100

Adjustment This multimeter incorporates a non-volatile memory containing the adjustment characteristics for all measurement ranges. This enables the instrument to be re-adjusted via a serial link without opening the instrument. It is supplied with a certificate of verification.
Safety according to EN 61010-1
Environment Indoor use Altitude < 2000 m Reference temperature 18°C to 28°C Rated range of use 0°C to 50°C Limit range of operation -10°C to 60°C Storage temperature range -40°C to 70°C Temperature coefficient max 0.1 x accuracy /°K Relative humidity 0 to 80 % from 0 to 40°C (70 % max for 5 MΩ / 50 MΩ) 0 to 70 % from 40°C to 50°C 60 % above 50°C
Casing and circuit self-extinguishing materials Operating quality IEC 359 Electromagnetic compatibility Emission & Immunity : EN 61326-1 Max. influence in electromagnetic fields at 3 V/m according to EN 61000-4-3 : 3 % end of scale in VDC and VAC, ADC and AAC ranges 5 % end of scale in Ω range 20 % end of scale in capacity range Power supply 9 V alkaline battery (6LF22) typical life of 500 hours in VDC mode
Mechanical Dimensions 189 x 82 x 40 mm Weight 400 g
Packaging Dimensions 230 x 155 x 65 mm Weight 500 g
Display Liquid crystal display comprising:
- a 5000-count display + sign (digits 14 mm high) - a 34-bar analogue bargraph display - appropriate units for each type of measurement - triggered mode indicators (relative, ranging) - battery discharged indicator
Measurement rate Digital display 2 measurements/s Bargraph 20 measurements/s

6.1. Accessories
6.1.1. Supplied with the multimeter
One set of test leads with safety probes One 6F22 9 V battery One spare 11 A fuse, 10 x 38 mm, rupture capacity 30 kA / 1000 V One spare 0.63 A fuse, 5 x 20 mm, rupture capacity 1.5 kA / 500 V One operating manual One rubber shock-absorber sheath
6.1.2. Optional
0,5 to 240 AAC, ∅ 20 mm MN09 0,1 to 1200 AAC, ∅ 52 mm CI03 0,5 to 1000 AAC, ∅ 39 mm PAC 20 0,5 to 1000 AAC, ∅ 39 mm PAC 21
30 A / 300 mV HA030-1 50 A / 100 mV HA050
Set of RS232 serial link SX-ASYC2CHD ASYC II acquisition software SX-DMM2 ASYC II adjusting software SX-ASYC2C/B
Current clamps · Shunts · Miscellaneous



Fig. 1 Soulever la béquille située à l'arrière.
Fig. 2 L'enlever en la tournant.
Fig. 1 Lift the stand on the back. Fig. 2 Remove it by rotating.
Abbg. 1 Die Klappstütze auf der Geräterückseite ausklappen.
Abbg. 2 Klappstütze durch Drehung abnehmen.
Fig. 1 Sollevare la linguetta situata sul retro. Fig. 2 Sganciare la linguetta ruotandola.
Fig. 1 Levantar el apyo sssituato on la parte trasera.
Fig. 2 Levantarlo ginándolo.


Fig. 3 Avec le coin de la béquille, enlever le volet supérieur, à gauche.
Fig. 4 Avec la béquille, enlever le volet supérieur, à droite.
Fig. 3 With the corner of the stand, remove the upper box.
Fig. 4 With the corner of the stand, remove the upper box.
Abbg. 3 Mit der Ecke der Klappstütze können Sie die obere Gehäuseabdeckung links abnehmen.
Abbg. 4 Durchanheben können Sie nun die obere Gehäuseabdeckung rechts abnehmen.
Fig. 3 Utilizzare la stessa parte per togliere la copertura del commutatore.
Fig. 4 Utilizzare la stessa parte per togliere la copertura del commutatore.
Fig. 3 Con el apoyo levantar la parte superior de la caja.
Fig. 4 Con el apoyo levantar la parte superior de la caja.