Instruction/ maintenance manual of the product 16 Fluke
Go to page of 12
PN 644226 July, 1997 Rev.1, 3/99 © 1997,1999 Fluke Corporation, Al l rights reserved. Printed in U .S.A. All p r oduct names are trademarks of their respective companies. 1 ® 16 Multimeter with Temperature Calibration Information W Warning To avoid electrical shock, remove test leads and any input signals from the Meter before opening the case.
16 Multimeter with Temperature 2 Accuracy specifications are given as follows: ±([% of reading] + [number of least significant digits]). Table 1. Specifications (cont) Function Range Resolution Accuracy f (50 to 400 Hz) 4000 mV 1 4.000 V 40.00 V 300.
Specifications 3 Table 1. Specifications (cont) Function Overload Protection 1 Input Impedance (Nominal) Common Mode Rejection Ratio (1 k Ω Unbalanced) 2 Normal Mode Rejection 2 f 600 V dc >5 M .
16 Multimeter with Temperature 4 Disassembly and Reassembly Referring to Figure 1, disassemble the Meter as follows. A Phillips-head screwdriver and small flat-blade screwdriver are required. 1. Remove the test leads and set the rotary knob to OFF. 2.
Recommended Test Equipment 5 Case, Top Decal Switch Assembly Actuator, Switch Main PCA Shield, Bottom Foot LCD Conn. Elastomeric Conn. Elastomeric LCD to PCA Support, Switch Assembly Battery Shock Absorber Case, Bottom Screw, THD Form, PH.P.STL, 5-14 x .
16 Multimeter with Temperature 6 1 2 3 Snap Snap Snaps ig2f.eps Figure 2. Removing and Reinserting the Printed Circuit Assembly.
Performance Test 7 Performance Test W Warning To avoid electric shock, do not execute the performance tests procedures unless the Meter is fully assembled. To ensure that the Meter is working properly and performs to specifications, use the following procedures.
16 Multimeter with Temperature 8 Table 3. Performance Tests Test No. Meter Range Input To Meter Switch Position Display Minimum Display Maximum 1 2 3 4 5 6 7 8 9 1 10 4.000 V ac 4.000 V ac, MIN MAX 4.000 V ac 4.000 V ac 4.000 V ac 4.000 V ac 40.00 V ac 40.
Calibration Adjustments 9 + + 5700A Ice Bath AK-80 Banana Plug Temperature Yellow Red Copper 0 ig6f.eps Figure 3. Performance Test for Temperature + _ 9V BATTERY + Clip to Base Contacts Adjust R4 for 3.997-4.003V DC Input Receptacles + ig1f.eps Figure 4.
16 Multimeter with Temperature 10 Temperature Note Adjust DC voltage (R4) before adjusting temperature (R38). The following describes three temperature calibration procedures: • Calibrating at room temperature. Use this procedure if it is not necessary to optimize the Meter’s performance.
Replacement Parts 11 ! CAL ENABLE 10A 100 mA 1000V 750V MAX ALL INPUTS 1kV MAX 500mA F 250V POWER AUTO 2ND MN MX RATE dB REL LOCAL THRESH ADDR BAUD HOLD LO HI COMP A A V FREQ DUAL DISPLAY MULTIMETER 45 V REF# REF V COM FUSED 80T-150U + _ 9V BATTERY + + R38 Clip to Base Contacts V m ig4f.
16 Multimeter with Temperature 12.
An important point after buying a device Fluke Fluke 16 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Fluke Fluke 16 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Fluke Fluke 16 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Fluke Fluke 16 you will learn all the available features of the product, as well as information on its operation. The information that you get Fluke Fluke 16 will certainly help you make a decision on the purchase.
If you already are a holder of Fluke Fluke 16, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Fluke Fluke 16.
However, one of the most important roles played by the user manual is to help in solving problems with Fluke Fluke 16. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Fluke Fluke 16 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center