Operating Instructions ProMinent ® Sigma/ 2 S2Ba (Basic Type) S2Ca (Control Type)
S2Ba
S2Ca
Please enter the identity code of the device here.
Two sets of operating instructions are necessary to ensure the ProMinent® Sigma/ 2 S2Ba or S2Ca metering pumps
are operated safely and reliably for their intended purpose:
This product specific Sigma/ 2 operating instructions manual and the ”General operating instructions
ProMinent® motor-driven metering pumps and hydraulic accessories” are only valid if read together!
Please completely read through these operating instructions first! · Do not discard!
The warranty shall be invalidated by damage caused by operating errors!
Part No. 987282 ProMinent Dosiertechnik GmbH · 69123 Heidelberg · Germany BA SI 018 10/03 GB
® · S2Ba S2Ca

Control elements: overview
1 Operating light (green) 2 Warning light (yellow) 3 Error warning light (red) 4 Stroke length adjuster 5 LCD display 6 UP key 7 DOWN key 8 P key 9 STOP/START key 10 i key 11 Relay output (optional) 12 Metering monitor socket 13 Float switch socket 14 External control socket 15 Diaphragm rupture indicator socket 16 Mains switch
15 14 13 12 11
In continuous display mode In settings mode (operating) (settings)
START
Press briefly Stop pump, Stop pump, start pump start pump
Press briefly Start batch (in “batch” operating mode only), Confirm entry- jump to next menu Cancel error option or continuous display
Press for 2 s Change to settings mode --- Press for 3 s --- Jump to continuous display Press for 10 s Software version displayed Press for 15 s Load factory settings (calibration) ---
Press x1 Toggle between continuous displays Toggle between “change individual digits” and change a figure"
Press x2 --- For “change individual digits”: jumps to first digit
Arrow keys UP and DOWN
Press separately Change directly alterable values Select other settings, (until “Set” appears) change individual digit or figure
Press simultaneously Prime (in “stroke rate” permanent display) ---
STOP/START key · P key · i key


Select the operating mode here first, then carry out the settings in the SET menu.
next page


p. 51
previous page


memory and tansfer factor 5 Operating mode "Batch" with
memory and transfer factor 5 Operating mode ”Batch” with
= UP and/or DOWN arrow keys, directly alterable values

Operating Instructions for ProMinent® Sigma/ 2 S2Ba/S2Ca © ProMinent Dosiertechnik GmbH, 2002
ProMinent Dosiertechnik GmbH Im Schuhmachergewann 5-11 69123 Heidelberg · Germany [email protected] www.prominent.de
Subject to technical modifications.
Imprint · ®





Identcode ordering system Please enter the identcode on the device label into the grey box below.
HM Main drive, diaphragm
Pump type: (figures 1+2 = back pressure [bar], figures 3-5 = feed rate [l/h]) 16050 16 bar; 50 l/h 16090 16 bar; 90 l/h 16130 16 bar; 130 l/h 12050* 10 bar; 48 l/h 12090* 10 bar; 86 l/h 12030* 10 bar; 125 l/h 07120 7 bar; 120 l/h 07220 7 bar; 220 l/h 04350 4 bar; 350 l/h * for SST versions, max. 12 bar
Diaphragm: 0 Standard diaphragm, PTFE version 1 Double diaphragm with diaphragm rupture indicator (retro fit possible)
Liquid end version: 0 No spring 1 With 2 valve springs, Hastelloy C, 0.1 bar 4 With pressure relief valve, Viton® seal, no valve spring 5 With pressure relief valve, Viton® seal, and valve spring
Hydraulic connection: 0 Standard threaded connector (according to technical data) 1 Union nut and PVC insert 2 Union nut and PP insert 3 Union nut and PVDF insert 4 Union nut and stainless steel insert 7 Union nut and PVDF hose nozzle 8 Union nut and stainless steel hose nozzle
Electrical power supply: S 3 ph, 230 V/400 V 50/60 Hz, 0.18 kW M 1 ph, AC, 230 V/50/60 Hz, 0.18 kW N 1 ph, AC 115 V 60 Hz, 0.18 kW L 3 ph, 230 V/400 V, 50 Hz, (EExe, EExde) P 3 ph, 230 V/400 V, 60 Hz, (EExe, EExde) R 3 ph, variable speed motor, 230/400 V, 0.37 kW V Variable speed motor with integrated speed control 1 pH, 230 V, 50/60 Hz Z 1 ph, variable speed control set 1 ph, 230 V, 50/60 Hz 1 No motor, with B14 flange (Gr. 71 (DIN)) 2 No motor, C 56 flange (NEMA) 3 No motor, B5 Gr. 63 (DIN) 4 No motor, B5, Gr. 70 (DIN)
Motor version: 0 IP 55 (standard) 1 Exe version (II 2G EExe II T3) 2 Exde version (II 2G EExde IIC T4) A ATEX type drive
Stroke sensor: 0 No stroke sensor (standard) 1 with stroke sensor (Hall sensor, no connection cable) 2 Pacing relay (reed relay) 3 Stroke sensor (Namur) intrinsically safe
Stroke length adjustment: 0 Manual (standard) 1 With stroke positioning motor, 230 V/50/60 Hz 2 With stroke positioning motor, 115 V/50/60 Hz 3 With stroke control motor, 0...20 mA 230 V/50/60 Hz 4 With stroke control motor 4...20 mA 230 V/50/60 Hz 5 With stroke control motor 0...20 mA 115 V/50/60 Hz 6 With stroke control motor 4...20 mA 115 V/50/60 Hz
S2Ba
Viton® is a registered trademark of DuPont Dow Elastomers.
ProMinent · ® · ®

Identcode ordering system Please enter the identcode on the device label into the grey box below.
HM Power end, diaphragm
Pump typ: (figures 1+2 =back pressure [bar],figures 3-5 =feed rate [l/h]) 16050 16 bar; 60 l/h 16090 16 bar; 95 l/h 16130 16 bar; 136 l/h 12050* 10 bar; 60 l/h 12090* 10 bar; 95 l/h 12130* 10 bar; 136 l/h 07120 7 bar; 148 l/h 07220 7 bar; 232 l/h 04350 4 bar; 350 l/h * for SST versions, max. 12 bar
Displacement element: 0 Standard diaphragm 1 Double diaphragm with rupture indicator incorporating “Pump stopping” function 2 Double diaphragm with rupture indicator incorporating “Pump alarm” function
Liquid end version: 0 No springs 1 With 2 valve springs,Hastelloy C,0.1 bar 4 With relief valve,Viton® seal,no valve spring 5 With relief valve,Viton® seal,with valve spring
Hydraulic connection: 0 Standard threaded connector (according to technical data) 1 Union nut and PVC insert 2 Union nut and PP insert 3 Union nut and PVDF insert 4 Union nut and stainless steel insert 7 Union nut and PVDF hose nozzle 8 Union nut and stainless steel hose nozzle
Electrical power supply: U 1 ph 100-230 V ±10 %, 50/60 Hz Cable and plug: A 2 m European B 2 m Swiss C 2 m Australian D 2 m USA Relays: 0 No relay 1 With fault indicating relay (N/C) 3 With fault indicating relay (N/O) 4 As 1 with pacing relay 5 As 3 with pacing relay F Power relay N/C G Power relay N/O
Control variant: 0 Manual + external with pulse control 1 Manual + external + pulse control + analogue 4 As 0 + process-timer 5 As 1 + process-timer P PROFIBUS®
S2Ca
Viton® is a registered trademark of DuPont Dow Elastomers.
®

1 Notes on safety of ProMinent ® metering pumps
General user instructions
Please read through the following user instructions carefully! They will help you get the best use out of the operating instruction manual.
The following are highlighted in the text:
Operating guidelines:
and safety guidelines:
Describes a potentially dangerous situation. If not avoided. could cause fatal or serous injury.
Describes a potentially dangerous situation. If not avoided. could cause slight or minor injury or damage to property.
Describes a potentially damaging situation. If not avoided. could cause damage to property.
Please also note the guidelines in “General Operating Instruction Manual for ProMinent® Motor- Driven Metering Pumps and Hydraulic Accessories”!
EX-pump only: The nameplates affixed to the title page are identical to those on the pump supplied to enable clear identification on the correct operating instruction manual for the pump.
Please give the order number and the serial number, which you will find on the nameplates of the pump itself, in the event of any query or spare part order. This facilitates identification of the pump.
1.1 Summary of additional safety instructions for the EX version of the pump (in accordance with ATEX)
Instructions · ®

- Diaphragm pumps with mechanically actuated diaphragm: no additional measures
necessary, however, version with diaphragm rupture indicator, version Ex“i”*, must be used on principle.
- For all metering pumps for metering inflammable media applies: starting up and
emptying only supervised by a competent person.
- Inflammable media may only be lifted with stainless steel liquid ends. - For exceptional
cases, where this is not possible, also PTFE with carbon may be used, our versions TT_ are made of conductive plastics. In this case, special attention must be paid by the user due to the lower mechanical stability.
®

Safety devices The following safety notice must be affixed to the pump (with PP, PC and PV liquid ends, acrylic glass cover, ...)
Because of danger of electrostatical
ignition use damp cloth only
for wiping plastic parts!
EC conformity declaration / certificates The EC conformity declaration for pumps for explosion-threatened workplaces is supplied with the pump (the copy of the EC conformity declaration for the pump itself does not apply to pumps designed for explosion-threatened workplaces).
The EC conformity declarations, the EC series prototype test reports and the operating manuals for the individual components are also supplied with the pump.
WARNING
WARNING
If handling media which must not come in contact with water, all traces of water must be removed from the liquid end before start-up! For this purpose, turn the pump through 180˚ and drain off the liquid end then flush with a suitable medium from above via the intake connection!
®

IMPORTANT
The lines must be secured such that no vibrations can occur!
Avoid reducing hose sizes!
Observe the permissible pressure of the hoses!
In addition, a shut-off valve should be provided on the delivery or intake side!
WARNING
The metering pump must be in a cleaned condition with the liquid end flushed if returned for repair!
®

2 Product description S2Ba/S2Ca
WARNING
Use for intended purpose
The identity code and serial number are given in addition to the standard technical specifications. These two numbers must always be quoted when making any enquiries as they enable distinct identification of the type of metering pump.
2.2.1 Functional description, drive
The ProMinent® Sigma/ 2 diaphragm-type metering pump is an oscillatory displacement pump with the stroke length adjustable in steps of 0.5 %. It is driven by electric motor (1). The rotary drive of the electric motor is stepped down by worm gear and transmitted via the eccentric roller (3) to the push rod (4) connected to fork (8) and converted into oscillatory movement. Return spring (5) presses the fork with push rod positively against the eccentric roller thus producing the return stroke. The stroke is adjusted by means of stroke adjustment knob (6) and spindle (7) by limiting the return stroke. Stroke movement is transmitted directly to the displacement diaphragm. Interacting with the valves, this diaphragm produces the overpressure and vacuum in the liquid end necessary for delivery. Flow is pulsating.
In the basic type, the electric motor is normally a 3-ph extended-range AC motor (refer to Section 3 for other options).
®


Fig. 2 61_01-101_00_25-01
b) At reduced stroke length
at max. stroke rate and stroke length

Angle reduced stroke length
Set stroke length dependent on the required delivery capacity.

A large stroke length and low metering frequency should be selected for very viscous media! A shorter stroke length and high frequency should be selected to achieve good mixing properties!

2.2.3 Metering capacity diagram
0 2 4 6 8 10 12 14
0 20 40 60 80 100 120
Metering capacity diagram S2Ba HM (60 Hz)
Metering capacity diagram S2Ba HM (60 Hz)
Stroke length in (%)
0 20 40 60 80 100 120
0 2 4 6 8 10 12 14
Metering capacity diagram S2Ba HM (50 Hz)
Metering capacity diagram S2Ba HM (50 Hz)
Stroke length in (%)

0 20 40 60 80 100 120
0 2 4 6 8 10 12 14
Metering capacity diagram S2Ca HM
Metering capacity diagram S2Ca HM
Stroke length in (%)

2.2.4 Functional description, delivery unit
The heart of the delivery unit is the DEVELOPAN® metering diaphragm (2). It hermetically seals the delivery chamber of the liquid end (4) and produces a displacement in the liquid end. The end disc (5) made of chemically resistant plastic together with safety diaphragm (13) separates the drive housing from the delivery unit and protects the drive from corrosion in the event of the diaphragm failing. Delivery is based on the interaction between intake valve (1) and head valve (3) of the same design together with the diaphragm movement. The valve balls can be supported with springs for metering viscous media.
The connection dimensions of valves and liquid ends of the same size but with different materials are identical. These parts can be interchanged as required.
Materials and dimensions are specified in Section 3, Technical data.

®

2.2.5 Integrated overflow valve with bleeder function
Task:
The task of the overflow valve is to protect the motor and gear unit against impermissible overpressure caused by the metering pump.
This function is produced by a spring-loaded ball.
A pressure relief mechanism for the bleeder function is provided.
Design and functional description (refer to Fig. No. 6 and 7)
Initially, the overflow valve illustrated under item 102 operates as a simple directly controlled safety valve. As soon as the pressure set with spring item 132 is exceeded, the effective pressure raises ball item 130. The liquid then flows off into the tank via hose connection item 128.

Overflow valve Sigma/ 2 liquid end, 10 bar PVT Identcode Type: 12050, 12090, 12130
Overflow valve Sigma/ 2 liquid end, 7 bar PVT Identcode Type: 07120, 07220
Overflow valve Sigma/ 2 liquid end, 4 bar PVT Identcode Type: 04350


Overflow valve Sigma/ 2 liquid end, 16 bar SST Identcode Type: 16050, 16090, 16130
Overflow valve Sigma/ 2 liquid end, 12 bar SST Identcode Type: 12050, 12090, 12130
Overflow valve Sigma/ 2 liquid end, 7 bar SST Identcode Type: 07120, 07220
Overflow valve Sigma/ 2 liquid end, 4 bar SST Identcode Type: 04350
The bleeder function is achieved by turning knob item 139 in counterclockwise direction as far as it will go towards ”open“: Priming aid for starting up pump against pressure. The force of spring item 132 relieves ball item 130 which is controlled by the lower spring force of bleeder spring item 133.
Once the pump has primed, turn knob item 139 in clockwise direction as far as it will go towards ”close“! The pump is now operational.

Corresponding to the type of pump, overflow valves are available for pressure stages pnom 4, 7, 10, 12 and 16 bar with (1.05 ... 1.15) xpnom opening pressure.
Material in contact with metered medium
Material Liquid end Suction/ Seals Balls Springs Integrated version discharge overload connector valve
PVT PVDF PVDF PTFE Ceramic / glass* Hastelloy C PDFE/Viton®
SST Stainless steel Stainless steel PTFE Stainless steel Hastelloy C Stainless steel/ 1.4571/1.4404 1.4581 1.4404 Viton®
*for 07120, 07220, 04350
Viton® is a registered trademark of DuPont Dow Elastomers.
Use for intended purpose/Use unintended purpose
Use for intended purpose
Protect the dosing pump against unacceptable high overpressure produced by the dosing pump.
Use the overflow value only in connection with liquid with a viscosity of up to max. 200 mPa s.
Use for unintended purpose
To protect the system from impermissible overpressure.
The pump must not be operated without the bypass line connected.
The bypass line must not be connected in the intake line (the bleeder function will no longer be guaranteed). The bypass line must be routed back into the supply tank.
When carrying out maintenance work on the overcurrent valve, pay attention to the tensioned state of the pressure spring item132! Wear safety goggles!
Function:
Monitors the seals in the working diaphragm. This liquid end can continue to function for a short period in emergency mode, i.e. full operating pressure, no leakage, even after diaphragm rupture.
Design and function description (see Fig. 8)
Liquid ends with diaphragm rupture sensors comprise a standard liquid end (item 100) a working diaphragm (item 200) and an auxiliary diaphragm (item 148) The auxiliary diaphragm is positioned between the back plate (item 201) and the interim plate (item 147) and forms a sealed compartment together with the working membrane (item 200).
The leak tightness of the working diaphragm, Item 200, is monitored with a diaphragm failure detector, Item 104, that triggers a contact signal in the event of diaphragm failure so that the pump is stopped in the S2Ca and the diaphragm failure is indicated on a LCD.
Technical data · ®

The liquid end can continue to operate in emergency mode, i.e. full operating pressure, no leakage, even after diaphragm rupture, until the diaphragm has been replaced. We offer two versions of the S3Ca with diaphragm rupture sensor:
A function plug is supplied which allows the pump to continue operating after a fault has occurred (diaphragm rupture, failure of the diaphragm rupture sensor).

®

Material in contact with metered medium
Lens, seals Intermediate disc Item 147, Item 148, 156, 159 intermediate bush Item 150
PVDF PTFE PVDF Stainless steel 1.4571 PTFE PVDF
Electrical data for the diaphragm breakage sensor
30 V DC/1 A or 125 V AC/0.6 A or 250 V AC/0.3 A
The diaphragm sensor is a N/C relay.
Before commencing operation, install the provided diaphragm breakage sensor together with the gasket (Item159) and make the electrical connections.
b) Namur sensor, intrinsically
5-25 V DC, Namur type and/or DIN 19234, zero volt design. Rated voltage: 8 V DC (Ri ~ 1 kΩ) Power consumption:
Active surface, uncovered > 3 mA Active surface, covered < 2 mA Rated switching distance: 1.5 mm The monitor/feeder must be capable of evaluating current changes in order to indicate diaphragm rupture!

3 Technical data 3.1 Technical data Sigma/ 2 3.1.1 Capacity data
Technical data S2Ba at 50 Hz operation
Feed rate at Max. Suction Admissible Connection Shipping maximum stroke lift priming suction / weight back pressure rate pressure discharge suction side side
Pump type Sigma/ bar l/h ml/stroke strokes/min. m Ws bar G - DN kg
12050 PVT 10 50 11.4 73 7 3 1"-15 15
12050 SST 12 48 11.4 73 7 3 1"-15 20
16050 SST 16 48 11.4 73 7 3 1"-15 20
12090 PVT 10 90 11.4 132 7 3 1"-15 15
12090 SST 12 86 11.4 132 7 3 1"-15 20
16090 SST 16 86 11.4 132 7 3 1"-15 20
12130 PVT 10 130 10.9 198 7 3 1"-15 15
12130 SST 12 125 10.9 198 7 3 1"-15 20
16130 SST 16 125 10.9 198 7 3 1"-15 20
07120 PVT 7 120 27.4 73 5 1 20* 16
07120 SST 7 120 27.4 73 5 1 20* 24
07220 PVT 7 220 27.7 132 5 1 20* 16
07220 SST 7 220 27.7 132 5 1 20* 24
04350 PVT 4 350 29.4 198 5 1 20* 16
04350 SST 4 350 29.4 198 5 1 20* 24
All performance data applies to water at 20 °C. The suction lift applies when the suction line and liquid end are full and correctly installed. * Sigma types 07120, 07220 and 04350, the valves used in the liquid end are DN 25 (G11/2). Since DN 20 is usually adequate for this tubing (see Technical Data, connection-suction/discharge side) the connector parts to be ordered by Identcode (e.g. washers) have already been reduced to DN 20, i.e. tubing and accessories can be DN 20.
Technical data S2Ba at 60 Hz operation
Feed rate at Max. Suction Admissible Connection Shipping maximum stroke lift priming suction / weight back pressure rate pressure discharge suction side side
Pump type Sigma/ bar psi I/h/gph strokes/min. m Ws bar G - DN kg
12050 PVT 10 145 60/ 15.9 87 7 3 1"-15 15
12050 SST 12 174 57/ 15.2 87 7 3 1"-15 20
16050 SST 16 232 57/ 15.2 87 7 3 1"-15 20
12090 PVT 10 145 108/ 28.5 156 7 3 1"-15 15
12090 SST 12 174 103/ 27 156 7 3 1"-15 20
16090 SST 16 232 103/ 27 156 7 3 1"-15 20
12130 PVT 10 145 156/ 41 232 7 3 1"-15 15
12130 SST 12 174 150/ 39.6 232 7 3 1"-15 20
16130 SST 16 232 150/ 39.6 232 7 3 1"-15 20
07120 PVT 7 100 144/ 38 87 5 1 20* 16
07120 SST 7 100 144/ 38 87 5 1 20* 24
07220 PVT 7 100 264/ 69.7 156 5 1 20* 16
07220 SST 7 100 264/ 69.7 156 5 1 20* 24
04350 PVT 4 58 420/ 111 232 5 1 20* 16
04350 SST 4 58 420/ 111 232 5 1 20* 24
All performance data applies to water at 20 °C. The suction lift applies when the suction line and liquid end are full and correctly installed. * Sigma types 07120, 07220 and 04350, the valves used in the liquid end are DN 25 (G11/2). Since DN 20 is usually adequate for this tubing (see Technical Data, connection-suction/discharge side) the connector parts to be ordered by Identcode (e.g. washers) have already been reduced to DN 20, i.e. tubing and accessories can be DN 20.
®

Feed rate at Max. Suction Admissible Connection Shipping maximum stroke lift priming suction / weight back pressure rate pressure discharge suction side side
Pump type Sigma/ bar l/h ml/stroke strokes/min. m Ws bar R - DN kg
12050 PVT 10 145 60/ 16 90 7 2 1"-15 17
12050 SST 12 174 60/ 16 90 7 2 1"-15 22
16050 SST 16 232 60/ 16 90 7 2 1"-15 22
12090 PVT 10 145 95/ 25 160 7 2 1"-15 17
12090 SST 12 174 95/ 25 160 7 2 1"-15 22
16090 SST 16 232 95/ 25 160 7 2 1"-15 22
12130 PVT 10 145 136/ 36 200 7 2 1"-15 17
12130 SST 12 174 136/ 36 200 7 2 1"-15 22
16130 SST 16 232 136/ 36 200 7 2 1"-15 22
07120 PVT 7 100 148/ 39 90 5 1 11/2 - 20* 18
07120 SST 7 100 148/ 39 90 5 1 11/2 - 20* 26
07220 PVT 7 100 232/ 61 160 5 1 11/2 - 20* 18
07220 SST 7 100 232/ 61 160 5 1 11/2 - 20* 26
04350 PVT 4 58 350/ 93 200 5 1 11/2 - 20* 18
04350 SST 4 58 350/ 93 200 5 1 11/2 - 20* 26
All performance data applies to water at 20 °C. The suction lift applies when the suction line and liquid end are full and correctly installed. * Sigma types 07120, 07220 and 04350, the valves used in the liquid end are DN 25 (G11/2). Since DN 20 is usually adequate for this tubing (see Technical Data, connection-suction/discharge side) the connector parts to be ordered by Identcode (e.g. washers) have already been reduced to DN 20, i.e. tubing and accessories can be DN 20.
Materials in contact with chemicals
Material Liquid end Suction/ Seals Balls Springs Integrated version discharge overload valve connector
PVT PVDF PVDF PTFE Ceramic / glass* Hastelloy C PDFE/Viton®
SST Stainless steel Stainless steel PTFE Stainless steel Hastelloy C Stainless steel/ 1.4571/1.4404 1.4581 1.4404 Viton®
* for 07120, 07220 and 04350
Temperature compatibility (medium temperature) of materials
SST 90 ° C 120 ° C
The specified temperatures (see above) can be exceeded temporarily, e.g. for sterilisation or flushing with hot water.
Under constant conditions and in minimum stroke length of 30 % corresponding to following notes, the reproducibility of the metered quantity is better than ±2 %.
All specifications refer to metered quantities with water at 20 °C and correct installation of the metering pump.


Model Connector A B C D D1** E E1** F Q G
FM 130 PVT DN 15 265/250* 160 G 1 A 104 124 326/329* 346/349* 82 101
FM 130 SST DN 15 265/250* 160 G 1 A 104 124 326/329* 346/349* 82 101
FM 350 PVT DN 25 304/289* 237 G 1 1/2 A 115 135 337/340* 357/360* 84 148
FM 350 SST DN 25 304/289* 237 G 1 1/2 A 115 135 337/340* 357/360* 84 148
FM 130 = Identcode type 16050,16090,16130 FM 350 = Identcode type 04350,07120,07220
* Dimensions for S2Ba ** Dimensions with diaphragm rupture sensor
®

3.1.4 Motor data
3 ph IP 55 230 V/400 V 50 Hz 0.25 kW 1.62/0.94 A S
3 ph IP 55 230 V/400 V 60 Hz 0.25 kW 1.42/0.82 A S
1 ph AC 230 V 50/60 Hz 0.18 kW 1.7/1.5 A M
1 ph AC 115 V 60 Hz 0.18 kW 3.3 A N
3 ph EXe or EXde 230 V/400 V 50 Hz 0.18 kW 1.1/0.7 A L
3 ph EXe or EXde 230 V/400 V 60 Hz 0.18 kW 1.1/0.6 A P
3 ph IP 55 230 V/400 V 50/60 Hz 0.37 kW R Version with external fan 1 PH 230 v; 50/60 Hz and PTC
1 ph IP 55 230 V 50/60 Hz 0.37 kW V Three phase motor with integrated speed changer, (see chapter 11)
For more details you can request the motor specification sheets. Custom motors and/or custom motor flanges are available on request.
Protection against accidental contact and moisture (IP)

Motor: IP 55 DIN EN 60034-5 (in accordance with DIN VDE 0470 Part 1, corresponds to
EN 60529 and IEC 529).
230 V ± 10 % 50/60 Hz 11.7 W
115 V ± 10 % 60 Hz 11.7 W
Cf. “Appendix” for terminal connection diagram.
230 V ± 10 % 50/60 Hz 6.5 W
115 V ± 10 % 60 Hz 6.5 W
Cf. “Appendix” for terminal connection diagram.
a) Reed contact (Identcode characteristic “Stroke sensor”: 1) Pin 1 (white) = 4.5 V to 24 V, max. 10 mA Pin 2 (brown) = OUT, open collector, 24 V, 20 mA Pin 3 (green) = GND Pulse width (low) ≥ 4 ms (depending on gearbox and power frequency)

b) Namur sensor, intrinsically (Identcode characteristic “Stroke sensor”: 3)
5-25 V DC, Namur type and/or DIN 19234, zero volt design. Rated voltage: 8 V DC (Ri ~ 1 kΩ) Power consumption:
Active surface, uncovered > 3 mA Active surface, covered < 1 mA Rated switching distance: 1.5 mm The monitor/feeder must be capable of evaluating current changes in order to indicate diaphragm rupture!
(power supply for the relay board)
Input voltage Mains frequency Power consumption
200/230 V AC (180-254 V) 50/60 Hz 10 mA (230 V/50 Hz)
100/115 V AC (90-134 V) 50/60 Hz 15 mA (115 V/60 Hz)
24 V DC (20-28 V) - 10 mA at 24 V DC
Voltage, max. 24 V DC
Current, max. 100 mA
Hook up duration 100 ms Standard adjustable
The contacts are zero volt.
The sound intensity level is < 70 dB (A) at maximum stroke, maximum stroking rate, maximum back pressure (water) in accordance with DIN EN 12639 (noise measurement in fluid pumps)

4 Start-up/Maintenance
WARNING
Observe the safety notes provided in Section 1.
All general guidelines in the accompanying ”General operating instructions ProMinent®
motordriven metering pumps and hydraulic accessories” apply.
Diaphragm pumps with mechanically actuated diaphragm: no additional measures necessary, however, version with diaphragm rupture indicator, version Ex“i”*, must be used on principle.
For all metering pumps for metering inflammable media applies: starting up and emptying only supervised by a competent person.
Inflammable media may only be lifted with stainless steel liquid ends. - For exceptional cases, where this is not possible, also PTFE with carbon may be used, our versions TT_ are made of conductive plastics. In this case, special attention must be paid by the user due to the lower mechanical stability.
WARNING
After loosening the liquid end screws (e.g. to change the valves or diaphragm), the screws must be retightened clockwise to the specified tightening torque.
What requires maintenance? · ®

General recommendation for maintenance intervals - every 3 months.
Shorter intervals are recommended if operated under load conditions (e.g. continuous operation).
The gear oil should be changed after approx. 5000 duty hours.
Gear oil ISO viscosity class VG 460, e.g. Mobil Gear 634,
ProMinent Order No. 555325 (Amount of oil approx. 0.5 l).
The metering diaphragm is a wearing part whose service life is dependent on following parameters:
The service life of the diaphragm is restricted in the case of abrasive media. In such cases, it is recommended to check the diaphragm more frequently and to install a diaphragm failure monitor.
Replacing diaphragm (see exploded diagrams in appendix!)
Flush liquid end first in the case of hazardous media. For this purpose, force water or a suitable flushing agent through the intake connection of the liquid end with a wash bottle.
Set stroke length to zero with the pump running. Switch off pump.
Release the six screws holding the liquid end, detach liquid end together with screws.
Release diaphragm from the push rod by jolting in counterclockwise direction and unscrew.
Screw on new diaphragm until it is firmly seated on the push rod. Mount the dosing head with
screws such that the suction connection lies at the bottom (observe the flow through direction / arrow marks on the valves). Switch on pump. Set stroke length to 100% and turn in screws then tighten crosswise to 7.5 ± 0.5 Nm. Check pump for leaks at max. pressure.
Maintenance intervals · ®

NOTE
The tightening torque of the liquid end screws should be rechecked after 24 hours of operation.
The tightening torques of the liquid end screws should be checked every 3 months for the PVT material version.

In the case of suction problems during installation, place the valves on a firm surface and tap the PTFE ball seat disk lightly with a brass rod and a hammer weighing about 300 g. Let the valves operate in the wet state.

approx. 300 g
Brass rod 9 mm diam. x approx. 200 mm
®

5 Features of the S2Ca metering pumps
5.1 Function description, motors
The pump has an integrated electronic overload cut-out in all versions. This responds as soon as the maximum admissible power consumption is reached and stops the motor.
If the motor stops due to a system overload the electronic controller detects the fault and
transmits a message which is displayed on the display panel and at the pump.
The fault signal can be cancelled by pressing the ”P“ key, altering the level for a brief period at
the pause input (switch function) or by reconnecting the pump to the mains.
If the motor has been switched off via the electronic overload cut-out, check that the pump is not continuously overloaded.
at 100 V at 230 V
nominal output 250 W 250 W
nominal current 3.2 A 2.0 A
peak current (in operation)* 10 A 5.5 A
making peak current 12 A 24 A
fuse, internal** 4.0 AT (1.5kA) 4.0 AT (1.5kA)
* internal switching ** Only genuine ProMinent fuses, item no. 732414, may be used!
Operating modes Operating modes are selected using the MODE menu (depending upon identity code, some operating modes may be absent).
“Analogue” operating mode: (Identity code, control variant: analogue) The stroke rate is controlled via an analogue electrical signal via the “external control” terminal. Signal processing is pre-selected at the controller.
“Manual” operating mode: The stroke rate is controlled manually via the controller.
“Contact” operating mode: This operating mode offers the opportunity to make fine adjustments with small increase/ decrease factors. Dosing can be activated by a pulse via the “external control” terminal or by a semiconductor element. With the “pulse control” option it is possible to pre-set a feed quantity (batch) or number of strokes (factor 0.01 to 99.99) via the control unit.
“Batch” operating function: This operating mode offers the option of working with larger transfer factors (up to 65535). Metering can be triggered by pressing the P key or a pulse from the “external control” terminal via a contact or semiconductor element. A batching quantity or number of strokes can be pre- selected via the control unit.
®

“PROFIBUS®” mode: (Identity code, control variant: PROFIBUS®) This operating mode provides the option of controlling the pump via the PROFIBUS® (see “Supplementary instructions for ProMinent® gamma/ L and ProMinent® Sigma versions with “PROFIBUS®”).
Functions The following functions can be selected using the SET menu:
“Calibrate” function (Identcode, stroke length adjustment: manual + calibration): The S3Ca can be operated in all operating modes including in calibrating mode. The corresponding continuous displays can show the actual feed quantity or the feed rate. Calibration is maintained within the stroke frequency range 0-180 strokes/ min. Calibration is also maintained when a stroke frequency is altered up to ±10 % scale divisions.
“Auxiliary frequency” function: It is possible to set a stroke rate in the SET menu, which may be activated via the “external control” terminal. This auxiliary frequency overrides all other pre-set stroke rate frequencies.
“Flow” function: Stops the S2Ca when the flow is insufficient. In the SET menu, the number of failed strokes is entered after which the pump will be turned off.
The following functions are available as standard:
“Float switch” function: Information on the liquid level in the feed chemical container is transmitted to the S2Ca. This option requires the installation of a 2-stage float switch. This is connected to the “float switch” terminal.
“Pause” function: The S2Ca can be stopped by remote control via the “external control” terminal. The “pause” function operates only via the “external control” terminal.
The following functions are activated by keystrokes:
“Stop” function: The S2Ca can be stopped by pressing the STOP/START key without disconnecting from the mains power supply.
“Prime” function: Priming (short term feed at maximum frequency) is activated by pressing both arrow keys at the same time (in “Stroke rate” permanent display).
Optional relay The S2Ca has two connection options (not in PROFIBUS® version).
Optional “fault-indicating relay” or “power relay”: In the event of fault signals, warning signals or float switch activation signals, connects an electrical circuit to trigger alarm sirens etc. The relay is retrofitted via an aperture in the power end.
“Fault indicating and pacing relay” option: In addition to the fault indicating relay the pacing relay can make a contact with every stroke. The relay is retrofitted via an aperture in the power end.
indicators The operating and error status is shown via the three LEDs and the “error” indicator on the LCD (see also section 9):
LCD indicator If a fault occurs “error” will appear along with an additional fault warning.
LED indicator Operating indicator (green) This indicator is lit as long as the S2Ca is operating correctly. It goes out briefly with every stroke.
Error indicator (red) This warning light appears if a fault occurs, e.g. “liquid levels low 2nd stage”.
Function and error · ®

Hierarchy of operating modes, functions and fault statuses
The different operating modes, functions and fault statuses each have a differing effect on whether and how the S2Ca functions. These effects are given below:
i.e.
(1) Socket for two-stage float switch with advance warning and de-activate function (with
function plug*, not illustrated)
(2) External socket for contact or analogue controller with zero volts deactivation via pause
function (with function plug* - not illustrated)
(3) Metering monitor socket for connection of ProMinent® metering monitor
(4) Mains switch (1-pin)
(5) Relay cable (fault indicating or pacing relay)
(6) Socket for diaphragm rupture sensor
(7) Mains lead with plug
*Must be plugged in unless cable attached.
®

Technical data, relay (control version)
Relay Fault-indicating relay Fault-indicating relay + pulse generator
Relay type Fault-indicating relay pulse generator Fault-indicating relay
Current, max. 2 A (ohmic) 100 mA 100 mA
Hook up duration 100 ms
Behaviour see Identcode see Identcode see Identcode
Service life > 200 000 cycles* > 50 x 106 (10 V 10 mA) > 200 000 cycles

fault indicating relay yellow/NO
Relay type Power relay
Voltage, max. 250 V (50/60 Hz)
Current, max. 16 A (ohmic)
Service life > 30 000 cycles*
* in the case of nominal load
The contacts are zero volt.
N/C fault indicating relay - the relay closes immediately after the power is switched on and opens in the event of a fault.
N/O fault indicating relay, the relay closes in the event of a fault.
Use suitable interference suppression (e.g. RC glands) when connecting inductive loads.

Wiring diagram View of cable connectors from front
Float switch cable
for Universal signal cable (5-core)


1 brown / pause
3 blue / analogue
Blue and black open -> alarm signal Brown and black open -> alarm signal + pump stops Pause function:
brown and black closed -> pump metering
brown and black open -> alarm + pump stopped

External/contact:
white and black closed -> start contact for pump (note pause function: brown and black closed?)
Analogue:
blue, black -> analogue output 0/4-20 mA (note pause function: brown and black closed?)
1 brown / 5 V
Auxiliary frequency:
grey and black closed -> pump metering at pre-set stroking rate
Circuit examples see p. 40
External/contact cable (2 core)


Contact open -> alarm -> pump stops with controller type 0 Contact closed -> metering stroke

Technical data, external contact
Semi-conductor switch elements (e.g. transistors in open-collector circuits) or contactors (relays) with a residual voltage of -0.7 V can be used as input switch elements.

Controller type 0 (see identity code)
Pin 1 = Pause input (activating function)
Voltage at open contacts: approx. 5 V Input resistance: 10 kΩ Controller: – zero volts contact (approx. 0.5 mA) – semi-conductor (residual voltage: < 0.7 V)
Pin 2 = Contact input
Voltage at open contacts: approx. 5 V Input resistance: 10 kΩ Controller: – zero volts contact (approx. 0.5 mA) – semi-conductor (residual voltage: < 0.7 V) Min. contact duration: 20 ms Max. pulse frequency: 25 pulses/s
Pin 3 = unused
Pin 4 = GND
Pin 5 = Auxiliary input
Voltage at open contacts: approx. 5 V Input resistance: 10 kΩ Controller: – zero volts contact (approx. 0.5 mA) – semi-conductor (residual voltage: < 0.7 V)
Controller type 1 (see identity code)
Pin 1 = Pause input (activating function)
Voltage at open contacts: approx. 5 V Input resistance: 10 kΩ Controller: – zero volts contact (approx. 0.5 mA) – semi-conductor (residual voltage: < 0.7 V)
Pin 2 = Contact input (not active during analogue operation)
Voltage at open contacts: approx. 5 V Input resistance: 10 kΩ Controller: – zero volts contact (approx. 0.5 mA) – semi-conductor (residual voltage: < 0.7 V) Min. contact duration: 20 ms Max. pulse frequency: 25 pulses/s
Pin 3 = Identical to input*
Pin 4 = GND
Pin 5 = Auxiliary input
Voltage at open contacts: approx. 5 V Input resistance: 10 kΩ Controller: – zero volts contact (approx. 0.5 mA) – semi-conductor (residual voltage: < 0.7 V)
* The metering pump makes its first metering stroke at approx. 0.4 mA (4.4 mA) and starts continuous operation at approx. 19.2 mA.
®

Block circuit diagram S2Ca
VDE cable:
VDE cable:

Circuit examples: universal signal cable by customer Cable Pump, inside

“External contact“ function (ProMinent® external/contact cable) two-core
Pulse rate, e.g. contact water meter
“External contact“ function (ProMinent® universal signal cable) five-core

Pulse rate, e.g. contact water meter

Permanent contact, e.g. from control room

Permanent contact, e.g. from control room

Analogue signal, e.g. from magnetic inductive flow meter

IMPORTANT
Delivery range:
1 relay circuit set with 2 screw fasteners 1 relay cable set with socket 1 seal
Disconnect the S3Ca from the mains power supply and rinse liquid end before commencing work (see section 10)!
When preparing the opening, ensure that the punch is not forced through the entire pump base! Pump circuits may become damaged.
Place the S2Ca on a firm surface with the relay opening press-out section at the top (see fig. 15) Place a punch (dia. 8-15 mm) in the centre of the relay opening press-out section , and strike briefly and sharply with a hammer (approx. 250 g) If necessary clean up the edges of the opening Remove the pressed out section from the S2Ca
Inserting the relay component Hold the relay component with your right hand gripping the left and right hand edges of the relay cover, and tilt the front end slightly to the left (see fig. 16) Push the relay component through the relay opening, holding the upper corner of the lower edge against the guide rail on the pump base, until the contact of the relay component has reached the controller contact. (test: can you still move the end of the circuit back and forth?) Gently push the relay component right into the opening Screw the relay cover firmly onto the housing using the screws provided Insert the relay cable plug seal into the relay cover and screw on the plug (see fig. 18) The pump is generally programmed ex-factory to ”drop-out action alarm relay“ and – if fitted – ”pull-up action clock pulse generator relay“. If a switching function is required, the pump can be reprogrammed at the Heidelberg plant (not necessary in PROFIBUS®-version).
Fig. 15 Fig. 16
Press-out relay opening · ®

Fig. 17 Fig. 18
Please acquaint yourself with the S2Ca control elements with the help of the “control elements and key functions” overview!
Indicators The LCD display supports the operation and setting of the S2Ca with a range of indicators:
Error Pause Aux Stop
Set Flow Calib Mem
The indicators are interpreted as follows:
Symbol for P key: The S2Ca is in settings mode.
Close symbol: In a continuous display: lock (if code has been set). In settings mode: indicates access to code menu.
Stop: The S2Ca has been stopped using the STOP/START key.
Aux: The S2Ca is pumping at the auxiliary frequency. In AUX menu: the S2Ca is in the AUX menu.
Pause: The S2Ca has been stopped using the “pause” function (external).
Error: A fault has occurred and the pump has been stopped.
Mem: An additional “memory” function has been set in the “contact” and “batch” operating modes. In CNTCT or BATCH menus (“mem” flashes): the memory function can be set.
Calib: The S2Ca is in the CALIB menu. In a continuous display (“calib” flashes”): Deviation from the stroking rate from the value at the time of the calibration by more than 10 percent (e.g. in the case of a stroke length of 40% when it has been set to less than 30% or more than 50%).
Flow: The S2Ca is in the FLOW menu.
Set: The S2Ca is in the SET menu.
Exclamation mark: The number of strokes reached is above the maximum value (99999) that can be shown in the LCD display.
When calibrated, the S2Ca displays the feed rate and the feed capacity in l and/or in l/h or in gal and/or g/h.
®

7 Settings
NOTE
Basic information for setting up the S2Ca
P
Set Set
3 s P 3 s continuous
Confirm entries Press the P key briefly; you will automatically move to the next menu option or to a continuous display.
Exit menu option without confirming Press the P key for 3 s: The entry is cancelled and you will return to a continuous display.
of a value Press the i key 1x; you can toggle between altering the digits of a value (“change individual digits” = standard) or incremental alteration of a value (“change a figure”).
Change adjustable values Press UP or DOWN arrow key; The flashing digit or figure will start to increase or decrease incrementally.
Confirm adjustable values For “change individual digits”: confirm each digit using the P key. When the last digit has been confirmed you will automatically move to the next menu option or to a continuous display.
For “change a figure”; press the P key 1x; you will simultaneously move to the next menu option or to a continuous display.
Correct wrongly set digits Press the i key 2x; you will go back to the first digit.

7.1 Check adjustable values
Before setting up the S2Ca you can check the current settings of adjustable values:
Press the i key (“i” as in “info”) when the S2Ca is in continuous display mode (There is no P key symbol in the LCD display).
Each time you press the i key you will see a different continuous display. The number of continuous displays depends upon the identity code, the selected operating mode and the connected accessories (see overview “continuous displays”).
If you hold down the P key for 2 seconds in any continuous display, the S2Ca will change to the settings mode. If CODE 1 is set, the code must be entered after pressing the P key.
The following menu options appear first in the settings mode (see also overview “operating/ settings diagram”):
In order to adapt the S2Ca to your process requirements you must:
CODE
P
7.3 Select operating mode (MODE menu)
The following operating modes are selected via the MODE menu (depending upon identity code, some operating modes may be absent):

- MANUAL
Batch Contact Manual Analog
P
- CONTACT
- BATCH
7.4 Settings for operating mode (SET menu)
NOTE
First select the operating mode in the MODE menu! Exceptions: Timer and PROFIBUS®.
In the SET menu you can adjust various settings depending upon the selected operating mode.
The following programmable function settings menus appear in all operating modes:
See also section 7.5!
Further settings menus depend upon the selected operating mode.
There are no other settings menus in the overall SET menu for the “manual” operating mode apart from those described in 7.5.
In addition to those settings menus described in 7.5, there is an additional ANALG menu in the overall SET menu for the “analogue” operating mode.
The stroke rate is controlled by an analogue electrical signal via the “external control” terminal.
You can select three signal-processing methods:
Set


Straight line:
The following symbol appears in the LCD display: . You can enter any stroke frequency ratio for the S2Ca in proportion to the electrical signal. You must enter two points P1 (I1, F1) and P2 (I2, F2). F1 is the stroke rate at which the pump should operate at current I1: the straight line and the ratio are fixed accordingly:

F1
I [mA] I 1 I 2
Draw a diagram like the one above - with values for (I1, F1) and (I2, F2) - in order to set the S2Ca to your required stroke rate!
Fmax · ®

Lower/upper band:
This processing mode allows you to control a pump via an electrical signal as shown in the diagrams below.
Lower band Upper band
e.g. Alkali pump e.g. acid pump
P2 F2
F1
P2
F2
P1
F1
I [mA] I 1 I 2
I [mA] I 1 I 2
Fig. 20 Fig. 21
Lower band:
The symbol appears in the LCD display. The S2Ca will operate below I1 at F1. Above I2, the S2Ca ceases to operate. Between I1 and I2 the stroke rate is between F1 and F2, proportional to the signal current.
Upper band:
The signal appears in the LCD display. The S2Ca will cease to operate below I1. Above I2, the S2Ca will operate at F2. Between I1 and I2 the stroke rate between is F1 and F2, proportional to the signal current.
The smallest processable difference between I1 and I2 is 4 mA.
Error processing In the “ER” (error) menu option you can activate an error processing function for the “curve” mode. An error message appears for signals below 3.8 mA and the S2Ca stops.
In addition to those settings menus described in 7.5, there is an additional CONTCT in the overall SET menu for the “contact” operating mode.
The operating mode “contact” allows you to activate a single stroke or a series of strokes. The strokes can be activated by a pulse or via the “external control” terminal. This operating mode is intended to transfer input pulses into a reduction (fraction) or small increase in strokes.
The factor resets to “1” if you change to a different operating mode.tity cod
Fmax · Fmax · ®

In the “contact - identity code: external with external pulse control” you can enter the number of pulses after which a stroke should be carried out. “Contact - identity code: external with external pulse control” is intended for small dosing quantities.
Set Set
Set
The number of strokes per pulse depends upon the factor, which you can enter. This allows you to vary to a certain extent the input pulses by a factor of 1.01 to 99.99 and/or reduce by a factor of 0.01 to 0.99:
“Number of strokes activated = factor x number of input pulses”
Examples Factor Pulse (sequential) Stroke number (sequential)
Increase 1 1 1
99.99 1 99.99
1.50 1 1.5 (1 / 2)
1.25 1 1.25 (1 / 1 / 1 / 2)
Reduction 1 1 1
0.50 2 1
0.10 10 1
0.01 100 1
0.25 4 1
0.40 2.5 (3 / 2) (1 / 1)
0.75 1.33 (2 / 1 / 1) (1 / 1 / 1)
Explanation of increase At a factor of 1 For every 1 pulse, 1 stroke is activated At a factor of 2 For every 1 pulse, 2 strokes are activated At a factor of 25 For every 1 pulse, 25 strokes are activated
Explanation of decrease At a factor of 1 After 1 pulse, 1 stroke is activated At a factor of 0.5 After 2 pulses, 1 stroke is activated At a factor of 0.1 After 10 pulses, 1 stroke is activated At a factor of 0.75 After 2 pulses, 1 stroke is activated, then after 1 pulse, 1 stroke is activated, then after 2 pulses, 1 stroke is activated etc.

NOTE
If a remainder occurs when the factor is processed, the S2Ca counts up the remainder values. When the sum reaches or exceeds “1” the S2Ca will activate a stroke. This ensures that the stroke number corresponds exactly to the factor throughout the dosing operation.
The number of input pulses which have not been processed are stored by the S2Ca in the stroke memory. The stroke buffer is limited to the batch size when memory is not activated (- with “Memory” at 65535 stokes). You can delete it by switching to a different operating mode.
You can optimally adapt the S2Ca to the respective process, e.g. in connection with contact water meters.
The “memory” extension function can be optionally activated (“mem” appears in the LCD display). When “Memory” is activated the S2Ca adds any strokes which could not be carried out up to the maximum capacity of the stroke buffer – 65535 strokes. If this maximum capacity is exceeded the pump will register a fault.
In addition to those settings menus described in 7.5, there is an additional BATCH menu in the overall SET menu for the “batch” operating mode.
Set Set
Set
The “batch” operating mode is a variant of the “contact” operating mode (see 7.4.3). You can pre-select a stroke number (no breaks, whole numbers only from 1 to 65535) as well as a feed quantity (batch). To switch between entries for “stroke number” and “feed quantity” press the i key 1x in the corresponding menu option (see also overview “operating/settings diagram”, fold-out page). The “batch” operating mode is intended for large dosing quantities.
Metering can be actuated by pressing the P-key or via a pulse from the “external control socket”.
The number of input pulses which have not been processed are stored by the S2Ca in the stroke memory. The stroke buffer is limited to the batch size when memory is not activated (- with “Memory” at 65535 stokes). You can delete it by switching to a different operating mode.
“Memory” extension function The “memory” extension function can be optionally activated (“mem” appears in the LCD display). When “Memory” is activated the S2Ca adds any strokes which could not be carried out up to the maximum capacity of the stroke buffer – 65535 strokes. If this maximum capacity is exceeded the pump will register a fault.

7.5 Settings for programmable functions (SET menu)
The following programmable function settings menus appear in all operating modes:
P P P P
Set Set
Calib Set
N
P P
L Set L Set
Calib
P
The S2Ca can also run in calibration mode. The corresponding continuous displays show the current dosing quantities or the feed rate.
The calibration is retained in the event of a change of the set stroke length by up to ±10 scale rates percent (if the stroke length is set to 40 % the range is 30 % to 50 %). If the stroke length is changed by more than ±10 scale rates percent the yellow warning light lights up, the perma- nent display flashes and the flashing “Calib” indicator appears.
If using a hazardous feed chemical, the following setting instructions ensure adequate safety precautions are taken!
Calibration Insert the suction tube into a measuring cylinder containing the feed chemical - the discharge tubing must also be correctly installed (operating pressure,…!)
Suck up the feed chemical (press both arrow keys at the same time) when the suction tube is empty
Note the liquid level in the measuring cylinder and the stroke length
Select the CALIB menu and go the first menu option using the P key
Select “ON” using an arrow key and change to the next menu option using the P key
To commence calibration, press the P key. The S2Ca starts to pump and displays the number of strokes (“STOP” appears at regular intervals)
After a sufficient number of strokes, stop the S2Ca with the P key
Calculate the dosed quantity (difference between the original quantity and the quantity remaining)
Enter this quantity in the next menu option and then change to the following menu option by pressing the P key.
Select the unit (“L” or “gal”) in the “UNIT” menu with an arrow key and push the P-button.
The S2Ca is calibrated. The corresponding continuous displays show the calibrated values. The total stroke number and total litres are set during calibration to “0”. The S2Ca is in the STOP status.

7.5.2 Settings for “auxiliary frequency” function (AUX menu)
P P P
Set
Freq.
The programmable function “auxiliary frequency” allows switching to a different stroke frequency, which can be set in the AUX menu. It can be activated via the “external control” terminal. When the auxiliary frequency is activated, “aux” appears in the LCD display.
This auxiliary frequency overrides the current stroke frequency set for the selected operating mode.
Set Set Flow Set Flow
The flow menu only appears when a dosing monitor is connected to the “dosing monitor” terminal. This dosing monitor registers each discharge stroke of the S2Ca at the discharge connector and transmits it back to the S2Ca. If this response transmission is serially omitted for a period set in the FLOW menu (due to failure or below-minimum dosing) the S2Ca stops.
The code menu is used to select whether you want to prevent access to parts of the settings options.
In the first menu option you can choose CODE 1 or CODE 2 (both use the same number).
In the CLEAR window you can delete the stored total stroke number and simultaneously the total litres (= set to “0”). You may then press the P key briefly to exit this window.
The values displayed are counted incrementally from the point of commissioning the pump, or from the last delete action.
N

8 Operating
This section describes all operating options available to you when the S2Ca is in continuous display mode (no P key symbol in the LCD display).
Set stroke length Stroke length is continually adjustable within a range of 0-100 %. The recommended stroke length range, which will practically guarantee technical reproducibility, is 30-100 % (SEK type: 50-100 %).
At low stroking rates (less than 1/3 maximum stroking rate) the controller switches to digital stroking mode. This ensures adequate cooling of the motor at low stroking rates.
The following operating options are available via the different keys (see also figure on the next page):
Stop/Start S2Ca To stop S2Ca: press STOP/START key. To start S2Ca: press STOP/START key.
Start batch Press the P key briefly in “batch” operating mode.
Load factory settings Press the P key for 15 s to load factory calibration settings! Current settings will be deleted.
Change to settings mode When you press the P key for 2 s in any continuous display the S2Ca will change to settings mode (see section 7). If CODE 1 is set, the code must be entered after pressing the P key.
Check adjustable values Each time you press the i key you will see a different continuous display. The number of continuous displays depends upon the identity code, the selected operating mode and the connected accessories.
Change directly alterable values To change a value (see below) directly in the corresponding continuous display, press one of the arrow keys until “set” appears in the LCD display. The delay has been programmed in to prevent inadvertent changing of values.
If CODE 2 has been set, this code must be entered after pressing the arrow key.
Directly alterable values are as follows:
Stroke rate In “manual”, “contact” and “batch” operating modes: The stroke rate can be altered in the “stroke rate” display.
Feed rate In “manual” operating mode The feed rate can be altered in the “feed rate” display.
Factor The factor is the number of strokes activated by an external pulse or a press of the P key (in “batch” mode only).
In “Batch” operating mode: You can alter the factor from the “remaining strokes” display. The S2Ca returns to the original continuous display a few seconds after the factor has been reset.
Display program version Press the P key for 10 seconds to display the program version. Example: “V1052” + “X1010”. In the case of “LOAD3” release the key immediately!
®

Batch size In “batch” operating mode: The batch size can by changed from the “batch size/remaining litres” display. The S3Ca returns to the original continuous display a few seconds after the batch size has been reset.
Priming The “priming” function is activated by pressing both arrow keys at the same time (in “Stroke rate” permanent display).
Cancel error Error messages are cancelled by pressing the P key briefly.
START
Change directly alterable values
Start batch (in "batch" operating mode only)
P
P
= Security lock (CODE 1)
= Security lock (CODE 2)
Remote control of the S2Ca is possible via a control cable or PROFIBUS® (see Section 5.3 and Chapter 7, ”Supplementary instructions for ProMinent® gamma/ L and ProMinent® Sigma versions with PROFIBUS®“ as well as your documentation in the annex).
Pump primes despite completed stroke action and no venting
Cause Crystalline deposits on the ball seat because valves have dried out.
Remedy Remove the suction sleeve from the chemical supply container and rinse out the liquid end thoroughly.
If still unsuccessful, dismantle valves and clean.
Fluid is seeping from the top plate
Cause The liquid end is not sealed against the pump diaphragm.
Remedy Tighten screws in the liquid end.
If unsuccessful, replace the diaphragm (see section 4.3).

Green LED indicator (operating display) is not lit
Cause Incorrect or no mains voltage.
Remedy Use the recommended mains voltage as given in the voltage specification on the nameplate.
Red LED display is lit, “Error” and “MINIM” flash in the display
Cause Fluid level in the chemical storage tank has reached “liquid level low, stage 2”.
Remedy Fill the chemical supply container.
Red LED display is lit, “Error” and “ANALG” flash in the display
Cause S2Ca is in “analogue” operating mode, a fault routine has been programmed in the ANALG menu and the operating current has fallen below 3.8 mA.
Remedy Remedy low operating current.
Switch fault routine “OFF” (see section 7.4.2.).
Red LED display is lit, “Error” and “CNTCT” flash in the display
Cause S2Ca is in “contact” or “batch” operating mode and the extended function “memory” has been set. In addition a very large factor has been entered, too many contacts have been input or the P- key has been pressed too often, resulting in an overflow of the stroke memory.
Remedy Press the P-key, saved data will be deleted.
Change S2Ca set up.
Red LED display is lit, “Error” and “FLOW” flash in the display
Cause Dosing monitor not properly connected.
Remedy Connect dosing monitor properly.
Press P-key.
Cause Dosing monitor has reported more defective strokes than have been set in the FLOW menu.
Remedy Press P-key.
Investigate the cause and remedy.
Red LED is lit. “Error” appears in the display and “MOTOR” is flashing
Cause The motor is not operating accurately because the backpressure is too high.
Remedy Reduce backpressure.
Press the P key (reset key).
Cause The motor is overheating.
Remedy Check ambient temperature (max. 40 ºC).
Allow motor to cool down.
Press the P key (reset key).
Cause Other motor fault.
Remedy Call ProMinent.
Press the P key (reset key).
Red LED is lit. “Error” appears in the display and “TEMPERATURE” is flashing
Cause The temperature in the pump housing is too high because the ambient temperature is too high.
Remedy Reduce ambient temperature.
Allow motor to cool down.
Press the P key (reset key).
Cause The temperature in the pump housing is too high because the pump is consuming too much power.
Remedy Check installation, rectify if necessary.
Allow motor to cool down.
Press the P key (reset key).
®

Red LED display is lit, “Error” appears in the display and “FAN” is flashing
Cause Fault related to fan in pump housing.
Remedy Check fan and replace if necessary.
Press the P key (reset function).
Red LED display is lit, “Error” appears in the display and “SYSTEM” is flashing
Cause Controller fault.
Remedy Disconnect the pump from the mains and reconnect. If the error message continues, send the pump back to ProMinent.
Red LED lights up, “Error” appears in the display and “Menu” flashes.
Cause Stroke buffer full
Remedy Rectify fault
Press P key (Make a note of what effect this has on your process).
Please contact your ProMinent branch or representative!
Yellow LED display is lit
Cause Liquid level in chemical storage tank has reached “liquid level low, stage 1”.
Remedy Fill chemical storage tank.
Yellow LED Display is lit and “calib” flashes
Cause The pump is calibrated and the stroke length deviates by more than ±10 % from the value at the time of calibration.
Remedy Reset the stroke length or calibrate the pump again to the desired stroke length.
Disconnect the pump from the mains.
Empty the liquid end by placing the pump on its head and allowing the chemical to drain out.
Rinse the liquid end with a suitable material, thoroughly rinse the liquid end after use with hazardous materials!
If decommissioning is only temporary, maintain the correct storage conditions:
Take particular care when disassembling the pump as the return spring (Item 5, Chapter 2.2) is under considerable mechanical tension!
Observe your locally applicable regulations (particularly with regard to electronic scrap and gear oil)!

11 Spare parts and accessories
CAUTION
The assembly and installation of ProMinent® metering pumps with other-make parts that are not tested and recommended by ProMinent are not permitted and can lead to personal injury and damage to property for which no liability shall be accepted!
Spare parts You will find the order numbers for the wearing parts in the exploded view drawings in the annex. If you require further parts, please order the “Spare parts list ProMinent® Sigma/ 2”.
Actuator for automatic stroke adjustment, setting time approx. 1 sec for 1% stroke length, feedback potentiometer 1k Ohm enclosure rating IP 54.
Adjuster comprising actuator and incorporating servo controller for stroke length adjustment via standard signal. Standard signal input 0/4-20 mA, corresponds to stroke length 0 – 100 %. Manual to automatic mode switch, switch for stroke length adjustment in manual mode, mechanical position display of 0/4-20 mA actual stroke length value output for remote display.
Variable speed motors with integrated speed changer (Identcode characteristic V) Power supply 1 ph 230 V, 50/60 Hz, 0.37 kW Optional external control via 0/4-20 mA
Speed controllers in metal casing (Identcode characteristic Z) The speed controller kit comprises a speed changer and a 0.37 kW variable speed motor.
Float switches 2-stage, with 2 m connection cable
Alarm relays For signalling faults
Alarm and pacing relays For cyclic timing of other devices and for signalling faults.
Control cables Universal control cable 5-core/2.5 and 10 m External contact cable 2-core/2.5 and 10 m
Foot valves With intake filter and non-return ball for connection at end of intake line.
Metering valves With spring-loaded non-return ball for metering in open or closed systems and for mounting the metering line.
Backpressure valves For exact metering at low operating pressure or as overflow safety valve.
Accumulators For pulsation damping in long metering lines.
Metering monitors For monitoring metering operations. After an adjustable number of non-acknowledged metering strokes, a fault is indicated and the metering pump shut down.
Intake assemblies With foot valve and float switch for expendable packing drums or metering tanks.
Metering tanks From 35 to 1000 l capacity with lockable screw cover and necessary accessories.
Manual/electric stirrers For mixing and preparing metering solutions.
®

Motor Datenblatt / Motor data sheet / Fiche technique pour moteur
Bestell Nr. 1011036 Hersteller ATB order no. / no. de commande producer / producteur
Motor-Typ AF 63/4C-7 Leistungsfaktor 0,66 motor type power factor type du moteur facteur de puissance
Maschinenart 3-Ph. Motor Wirkungsgrad 88 % type of machine efficiency désignation rendement
Schutzart IP55 Bemessungsfrequenz 50 Hz degree of protection rated frequency degré de protection fréquence nominale
Bauform IMB5 Bemessungsdrehzahl 1370 U/min mounting rated speed rpm construction vitesse nominale t/mn
Bemessungsleistung 0,25 kW Wärmeklasse F rated output temperature class puissance nominale class d’isolement
Bemessungsspannung / ∆ Anzugsstrom 3,2 fach rated voltage 400/230 V starting current fold tension nominale courant de démarrage fois
Bemessungsstrom 0,94/1,62 A Anzugsmoment 2,1 fach rated current starting torque fold courant nominale couple de démarrge fois
Geprüft nach DIN EN 60034 Kippmoment 2,3 fach tested in acc. with pull-out torque fold contrôlé selon couple de décrochage fois
PTB Nr. ohne EX-Schutz Umgebungstemperatur max. 40 °C without EX-protection ambient temperature température ambiante
Schaltung / ∆ connection branchement
Anmerkung 400/230 V comments 380-420/220-240 V (50 Hz) observation
Pumpentyp S2Ca HM _ _ T U
ProMinent Dosiertechnik GmbH · 69123 Heidelberg · Germany Nr./No. MD - 1011036 Datum/Date 28.02.2002
Motor datasheets are available for further information. Special motors and special motor flanges are possible on request.
®

For pumps without Ex protection:

the EC conformity declaration for the pump for explosion-threatened workplaces is supplied with the pump.
The EC conformity declarations, the EC series prototype test reports and the operating manuals for the individual components are also supplied with the pump.
®


Delivery unit Sigma/ 2 350-DN 25 PVT for identity code types: S2Ba/S2Ca 04350, 07120, 07220 Delivery unit Sigma/ 2 130-DN 15 PVT for identity code types: S2Ba/S2Ca 12050, 12090, 12130
Spare parts kit FM 350 PVT Order No. 740325 Spare parts kit FM 130 PVT Order No. 740324
* The listed items are constituent parts of the spare part kit.
** Special accessories (not included in spare part kit). Subject to technical modifications.
®

Delivery unit Sigma/ 2 350-DN 25 PVT for identity code types: S2Ba/S2Ca 04350, 07120, 07220 Delivery unit Sigma/ 2 130-DN 15 PVT for identity code types: S2Ba/S2Ca 12050, 12090, 12130
Overflow valve assy 12 bar PVA Order No. 1018572

Overflow valve assy 10 bar PVA Order No. 1018947
Overflow valve assy 7 bar PVA Order No. 740811
Overflow valve assy 4 bar PVA Order No. 740812
For other spare parts see delivery unit without overflow valve (see P. 56).
Fig. 23 61_05-104_01_00-03
Subject to technical modifications.
Delivery unit Sigma/ 2 350-DN 25 SST for identity code types: S2Ba/S2Ca 04350, 07120, 07220 Delivery unit Sigma/ 2 130-DN 15 SST for identity code types: S2Ba/S2Ca 12050, 12090, 12130, 16050, 16090, 16130

Overflow valve assy 16 bar SSA Order No. 1019246
Overflow valve assy 12 bar SSA Order No. 740813
Overflow valve assy 7 bar SSA Order No. 740815
Overflow valve assy 4 bar SSA Order No. 740814
For other spare parts see delivery unit without overflow valve (see P. 58).
Fig. 24 61_05-104_01_01-03_2
Subject to technical modifications.
®

Delivery unit Sigma/ 2 350-DN 25 SST for identity code types: S2Ba/S2Ca 04350, 07120, 07220 Delivery unit Sigma/ 2 130-DN 15 SST for identity code types: S2Ba/S2Ca 12050, 12090, 12130
Valve assy DN 10/SST* Order No. 809459
Diaphragm FM 050 Order No 1010279 Diaphragm FM 065 Order No. 1010282
Valve assy DN 10/SST* Order No. 809459
Spare parts kit FM 350 SST without valve Order No. 740327 Spare parts kit FM 350 SST with 2 valves Order No. 740329 Spare parts kit FM 130 SST without valve Order No. 740326 Spare parts kit FM 130 SST with 2 valves Order No. 740328
* The listed items are constituent parts of the spare parts kit.
** Special accessories (not included in spare parts kit). Subject to technical modifications.
Fig. 25 61_05-104_01_01-03_2_norm

Retrofit kit double diaphragm S2Ba/S2Ca
S1Ba Retrofit kit double diaphragm FM 050 Order No. 1009846
S1Ba Retrofit kit double diaphragm FM 065 Order No. 1009848
S1Ba Retrofit kit double diaphragm FM 120 Order No. 1009850
S1Ca Retrofit kit double diaphragm FM 050 Order No. 1009847
S1Ca Retrofit kit double diaphragm FM 065 Order No. 1009849
S1Ca Retrofit kit double diaphragm FM 120 Order No. 1009851


Lense Order No. 792807
Diaphragm failure detector *
Fig. 26 61_05-104_00_98-03_2 * These items belong to the retrofit kit.
Subject to technical modifications.
®

Stroke actuator wiring diagram
4 3 2 1
Wiring diagram for stroke length adjuster
N
E
V+
- Sig.
Pages referenced in this manual — manufacturer, support and downloads.