Instruction/ maintenance manual of the product P Siemens
Go to page of 24
F i n a l l y , h e a r i n g i n s t r u m e n t s t h a t t a l k t o e a c h o t h e r . A CURIS TM S A CURIS TM P with e2e wireless TM User Guide Behind-the-Ear System s 010604402_Siemens_UG_Acuris_Bte 19.
2 Congratulations ! As the owner of a new hearing syst em, you may once again enjoy the sounds of y our environment, and con versation with your family and friends. These instructions as well as guidance from y our hearing instrument specialist will help you understand how y our new hearing system works.
3 Contents Page Y our A CURIS TM BTE (behind-the-ear) hearing system 4 - ACURIS S - ACURIS P Battery inser tion and removal 5 Battery compar tment door lock 8 Inserting and removing your A CURIS BTE 9.
4 Y our A CURIS BTE (behind-the- ear) hearing syst em a j d g f s Hearing system f or severe hearing loss with manual or automatic v olume control h k A CURIS P A CURIS S a d g h k s Hearing system f or mild to moderate hearing loss T o achie ve optimum fitting, A CURIS P instruments are designed specifically for the left and right ears.
5 Batter y inser tion and remo val Handling the batter y The pro tective tab on your battery should only be removed when you ar e ready to use the battery . Inserting the batter y Open the battery compar tment with your fingernail as shown in the illustration.
6 Gently close the battery compar t- ment by pressing the bo ttom of the door upward with your finger tip. Do not f orce the door shut. If it does not close easily , check to see if the battery was inser ted properly . When the batter y door is complet ely closed, your hearing instrument is ready f or use.
7 Caution: K eep batteries and instruments awa y from the reach of small children. Batteries can be harmful if swallo wed. If swallo wed consult a physician or a hospital immediately .
8 Batter y compar tment door lock Y our instruments are equipped with battery compar tment door locks. They are int ended to pre vent the unintentional r emoval of batteries from their compartment. T o lock the batter y compartment, open the compartment door to the first stop.
9 Inser ting and removing y our A CURIS BTE Make sur e that your hearing instrument is turned off. Hold the earmold between y our thumb and index finger and insert it carefully into y our ear canal. Y ou may find it helpful to pull y our ear slightly upward or t o the back.
10 T urning y our A CURIS BTE on and of f fig 1 fig 2 Via the batter y compartment Y our ACURIS hearing system aut omatically turns on when a battery is inser ted and the compartment door is closed complet ely . T o turn of f Press the batt er y compar tment to the rear down t o the first stop to turn the instrument OFF .
11 Y our hearing system can also be programmed b y your hearing instrument specialist to be turned on and off using the program button. Via the push button T o turn your hearing instrument off, press and hold the pro- gram button f or appro ximately 2 seconds.
12 e2e wireless TM Unique to y our ACURIS hearing system is the e2e (ear -to-ear) wireless TM t echnology . The wireless functionality: - automatically adjusts the system t o change both hearing instruments simultaneously to adapt t o different list ening situations.
13 Wireless remote control - ePock et TM (accessor y) With the wireless function y ou also hav e the possibility of using the ePock et (bi-directional r emote control) fr om Siemens. With it, you can change the pr ogram and volume of your hearing syst em and read out the syst em settings.
1 2 3 4 Description of the hearing situation Hearing program 14 Using the program butt on Y our hearing instrument specialist has programmed your sy stem as follows: How to select the prog rams Y our ACURIS hearing system can hav e up to 4 hearing programs.
15 Using the directional microphone syst em According t o your hearing needs, your A CURIS hearing system can be adjust ed to focus on those sounds in fr ont of you (dir ectional mode) for when you are having a face-to-face conv ersation in a noisy environment.
16 Using the v olume control (Only A CURIS P) Y our ACURIS P hearing system is fully automatic. Ther e is no need to manually adjust the volume contr ol. However , if you pr efer manual volume control, ACURIS P hearing syst em can be prog rammed for manual adjustment.
17 A udio input Attaching an audio shoe Open the battery compar tment door to the first st op. Open the audio input flap and hook the audio shoe in from the fr ont. Gently press it int o place so that it attaches to the back and the curve of the audio shoe presses against the bott om of the hearing instru- ment.
18 T urning of f an instrument with audio shoe attached Press the audio shoe up in the back, until the battery compar t- ment opens and catches on the first stop. Removing the audio shoe Press the audio shoe t oward the front. Then unhook it and close the audio input flap.
19 The telecoil mode is int ended for t elephone use. In the telecoil mode, place the telephone r ecei- ver close t o your hearing instru- ment for best r eception. By tr ying various positions with the telepho- ne receiver , you will disco ver what works best for y ou.
20 Maintenance and care Y our new ACURIS hearing syst em is a highly sensitive system. Proper car e and handling will ensure the best possible performance. Ear wax ma y accumulate on your earmold. If not cleaned regularly y ou can damage the hearing systems or cause health problems.
21 Impor tant not es • Never e xpose your hearing system to e xtreme temperatur es or high humidity . Do not st ore it in direct sunlight. Do not wear it when taking a shower or when using hairspra y . • As with all electrical, non gas-encapsulated systems, the hearing system must no t be operated in explosion- endangered r ooms.
22 T roubleshoo ting tips Y our ACURIS hearing system is an e xtremely dependable device. How ever , should a problem occur , this can usually be correct ed quickly .
23 PROBLEM CAUSE POSSIBLE SOLUTION Sound is V olume control too Lower volume contr ol distorted or high intermittent Weak batt er y Replace battery “Buzzing” or Battery compar tment Gently close t.
Siemens Audiologische T echnik GmbH Gebbertstraße 125 9 1 058 Erlangen Germany www .hearing-siemens.com Order -No.: A91100-M5100-1080-04-7600 Printed in Germany 301105.CG 87 81 994 Subject to change without prior notice. With the CE marking Siemens confirms compliance with the European Dir ective 93/42/EEC concering medical devices.
An important point after buying a device Siemens P (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Siemens P yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Siemens P - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Siemens P you will learn all the available features of the product, as well as information on its operation. The information that you get Siemens P will certainly help you make a decision on the purchase.
If you already are a holder of Siemens P, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Siemens P.
However, one of the most important roles played by the user manual is to help in solving problems with Siemens P. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Siemens P along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center