Coating Thickness Instrume nts ProTek II S SERIES User’s Guide ProTek II SF _____ ProTek II SN _____ ProTek II ST _____.
2 1.0 INTRODUCTION T he P ro T ek II ser ies Coat ing T hic kness I nstruments represent the latest techn ology for the non-destru ctive m easurem ent of surf ace coatings on all metals .
3 2.0. EQUIPMENT SUP PLIED 2.1 Standard Accessories The ProT ek II Series Gaug es are supplied w ith the f ollowin g standard acces sories: a. Bare test plate for setting Z ERO calibration. ProT ek II SF supplied w ith steel plate ProT ek II SN supplied w ith alum inum plate ProT ek II ST supplied w ith steel and alu mi num plates b.
4 3.0. KEYP AD FUNCTIONS 1. Turns the power on ON 2. Turns the power off 3. Selects display ed value or displayed su b-m enu option 1. Enters calibration mode 2. Increases display ed value in calibration m ode 3. Increases num ber of m easuremen ts in CA M mode 1.
5 4.0 OP ERATING PR OCEDURE 4.1 HANDLE P ROB E CAREFULLY 4.2 Taking Measurements 1. Press the ON k ey to turn on the pow er. T he LCD w ill m omen tarily flash SAB ERN and all of th e indicators w ill be illum inated. T he display will then show “- - -- --” alon g with the selected un its of meas ure indicator: a.
6 5.0 CALIBRATION 5.1 Introduction The Pr oT ek II series normally d o not require calibr ation before use. How ever, wh enever the product or part to be checked varies greatly f rom the supplied test.
7 Settin g The “Z ERO” Point 3. Af ter pressing th e key , the display w ill flash “ 2 PT CAL” fol low ed by “ 0 0 P LATE” . An au dible “beep” w ill also be heard. Place the probe on th e bare m etal part and press until th e shell contacts the surf ace.
8 6.0 CHANGING UNITS OF MEASURE To change the u nits of m easure from mils to m icrons, and vice versa, press the 0N + key. Each tim e this combination of key s are p ressed the units w ill chang e from on e to the other. The unit of m easure indicat or, located in the u pper right corn er of th e display , w ill also change.
9 Af ter the first m easurement is tak en, a nu mber w ill app ear on t he l eft side o f the d ispl ay alo ng with the thi ckness rea di ng o n the r ight. T he number o n t he l eft indi ca tes the num ber of m easuremen ts that m ust be taken to complete the current series for the CAM average.
10 4- P ress the ON key . T he gaug e enters to its previous setup w ithout any changes 8.0 CHANGING THE BATTERY. When a battery indicator appears on top of the display , replace the battery . Battery life is approxim ately 6 m o nths w hen the ProTek II series g auge s are us ed for about 80 readin gs per day.
An important point after buying a device Protec PROTEK II PROTEK II SF (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Protec PROTEK II PROTEK II SF yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Protec PROTEK II PROTEK II SF - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Protec PROTEK II PROTEK II SF you will learn all the available features of the product, as well as information on its operation. The information that you get Protec PROTEK II PROTEK II SF will certainly help you make a decision on the purchase.
If you already are a holder of Protec PROTEK II PROTEK II SF, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Protec PROTEK II PROTEK II SF.
However, one of the most important roles played by the user manual is to help in solving problems with Protec PROTEK II PROTEK II SF. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Protec PROTEK II PROTEK II SF along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center