Accessing Prescription Setting Screens
The information on this page is ONLY for health care professionals. Remove this page from the manual before giving the manual to the patient!
The ventilator has two levels of menu access, Full and Limited. Full Menu Access allows you to alter all available settings. Limited Menu access permits the user to alter only those prescription parameters that affect patient comfort, such as Rise Time, Flex, and Ramp Start Pressure, if those parameters are available as part of the prescription. The ventilator defaults to Full Menu Access mode.
Note: When the airflow is off and AC power is removed from the device for more than 5 minutes, the device will enter a low power mode to save battery life. When the device is in low power mode, the Setup key sequence will be ignored. Press the Start/Stop button, or connect AC power, or insert an SD card to exit the low power mode.
After accessing the Main Menu, if the device is in Limited Menu Access mode, you can use the following Setup key sequence to enter Full Menu Access mode and be able to change prescription settings:
When you perform this key sequence from the Monitor screen, the Main Menu screen appears and an audible indicator sounds indicating you are now in Full Menu Access mode.
When you perform this key sequence when the airflow is off, the Setup screen appears and an audible indicator sounds.
You can go into the Options menu and permanently change the Menu Access setting to Full Menu Access. Otherwise, the device will return to the Menu Access mode stored in the setting once you exit the menu screens or if one minute passes without pressing any device buttons. If you are in the Setup mode, and an SD card is in the device, “Write Event Log to SD Card” will appear in the menu.

R E F 1 0 5 4 7 2 8
1 0 7 2 9 1 9

© 2010 Koninklijke Philips Electronics N.V. All rights reserved.
















This chapter provides an overview of the Trilogy200 device.
The Trilogy200 system may include the following components. Some components are optional accessories that may not be packaged with the device.

O2 Inlet Quick Connects Active Exhalation Device (AED)
Proximal Pressure Line (pre-assembled to AED) Whisper Swivel II Flow Sensor
Chapter 1 Introduction

The Philips Respironics Trilogy200 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy200 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.).
The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.
The system is recommended to be used only with various combinations of Philips Respironics-approved patient circuit accessories, such as patient interface devices, humidifiers, water traps, and circuit tubing.

A warning indicates the possibility of injury to the user or operator.
Patient Monitoring Prior to placing a patient on the ventilator, a clinical assessment should be performed to determine:
For ventilator dependent patients, always have alternate ventilation equipment, such as a back-up ventilator, manual resuscitator, or similar device, available.
Ventilator dependent patients should be continuously monitored by qualified personnel. These personnel should be prepared to provide alternate therapy in the event of ventilator failure or inoperative equipment.
For ventilator dependent patients, do not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect and Low Peak Inspiratory Pressure alarms.
Test the operation of the circuit disconnect function daily and whenever a change is made to the patient circuit. An increase in circuit resistance can prevent proper operation of some alarms.
Speaking valves, Heat Moisture Exchangers (HMEs), and filters create additional circuit resistance and may affect the performance of alarms chosen for circuit disconnect protection.
Do not set the Low Peak Inspiratory Pressure alarm too low, or the system may not detect large circuit leaks or a patient disconnect.
Alternative Ventilation · Patient Disconnect Protection · Chapter 1 Introduction

Trilogy200 is a restricted medical device designed for use by Respiratory Therapists or other trained and qualified caregivers under the supervision of a physician.
The prescription and other device settings should only be changed on the order of the supervising physician.
The operator of the ventilator is responsible to read and understand this manual before use.
The device can provide therapies typically associated with both ventilator dependent and non-dependent patients. The mode of ventilation, circuit type, and alarm strategies should be chosen after a clinical evaluation of each patient’s needs.
C-Flex, Bi-Flex, and AVAPS are intended for use by adult patients.
SD Card Prescription Changes
When you change the device prescription, alarms, and other settings using the SD card, Trilogy200 requires that the caregiver review and verify the changes prior to the changes being used by the device. The caregiver or health care professional is responsible to ensure that the prescription settings are correct and compatible with the patient after using this feature. Installing the wrong prescription for a particular patient may result in improper therapy, lack of appropriate safety monitoring, and risk of death or injury to the patient.
This device is intended for use in the electromagnetic environment specified in Chapter 13 of this manual. The user of this device should make sure it is used in a compatible environment.
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated using the information provided in Chapter 13 of this manual.
The internal battery is NOT intended to serve as a primary power source. It should only be used when other sources are not available or briefly when necessary; for example, when changing power sources.
The ventilator has a two-stage low battery alarm. The medium priority alarm indicates that approximately 20 minutes of operation remain, and the high priority alarm indicates that less than 10 minutes of operation remain. Actual run time may be more or less than this and varies with battery age, environmental conditions, and therapy.
Immediately seek an alternate power source when the “Low Battery” alarm appears. Complete power failure and loss of therapy is imminent.

Operating and Storage Temperatures
Do not use this device if the ambient temperature is warmer than 40˚ C (104˚ F). If the device is used at room temperatures warmer than 40˚ C, the temperature of the airflow may exceed 43˚ C. This could cause system alarms, thermal irritation, or injury to the patient’s airway.
Bacteria Filter Philips Respironics recommends that a main line outlet bacteria filter (Part Number 342077) be used whenever the device is used for invasive therapy or if the ventilator may be used on multiple patients.
The ventilator should only be used with patient interfaces (e.g., masks, circuits and exhalation ports) recommended by Philips Respironics. Proper operation of the device, including alarms, with other circuits has not been verified by Philips Respironics and is the responsibility of the health care professional or respiratory therapist.
When adding any components to the breathing system, the flow resistance and dead space of the added components such as humidifiers, speaking valves, Heat Moisture Exchangers (HMEs) and filters should be carefully considered in relation to the potential for adverse effects on the patient’s ventilatory management and device alarms.
Passive Circuits An exhalation port is required when using a passive circuit.
For the passive circuit, at low expiratory pressures, the flow through the exhalation port may be inadequate to clear all exhaled gas from the tubing – some rebreathing may occur. Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suffocation.
Active Circuits Only use the active exhalation devices designed for Trilogy200. Philips Respironics has not verified proper operation of other active exhalation devices, and their use may result in improper or unsafe device operation.
With active exhalation circuits, the exhalation device must be operating properly for the ventilator to deliver therapy. The exhalation device should be inspected on a daily basis and replaced whenever necessary.
System Checkout Do not use the ventilator on a patient until a system checkout has been performed. See Chapter 10 of this manual.
To make sure the device is operating properly at start-up, always verify that the audible tone sounds and the alarm LEDs light red and then yellow momentarily. Contact Philips Respironics or an authorized service center for service if these indications do not occur at start-up.
Patient Circuits (General) · Chapter 1 Introduction

Remote Alarms When using a remote alarm, make sure you fully test the remote alarm connector and cable by verifying that:
–– Annunciated alarms on the ventilator are also annunciated on the remote alarm. –– Disconnecting the remote alarm cable from the ventilator or from the remote alarm results in an alarm notification at the remote alarm.
The remote alarm should be tested daily.
Oxygen When administering fixed-flow supplemental oxygen, the oxygen concentration may not be constant. The inspired oxygen concentration will vary, depending on the pressures, patient flows and circuit leak. Substantial leaks may reduce the inspired oxygen concentration to less than the expected value. Appropriate patient monitoring should be used, as medically indicated, such as an alarming pulse oximeter.
This device DOES NOT alarm for loss of the low flow oxygen supply.
This device is intended to be connected to a low flow (0-15 l/min) oxygen source such as an oxygen concentrator or other oxygen source equipped with a pressure regulator (set to 50 PSI or less) and a flow regulator/meter.
Do not connect the device to an unregulated or high pressure oxygen source.
The device may result in incorrect flow and tidal volume measurements and improper operation of related alarms if you add low flow oxygen directly into the patient circuit or mask instead of using the oxygen inlet on the back of the ventilator.
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open flame.
If oxygen is used with the device, the oxygen flow must be turned off when the device is not in use. Explanation of the Warning: When the device is not in operation and the oxygen flow is left on, oxygen delivered into the tubing may accumulate within the device’s enclosure.
Fire or Explosion The ventilator should not be operated in the presence of flammable gasses. This could cause a fire or explosion.

Alarms Respond immediately to any alarm. It may indicate a potentially life-threatening condition. Refer to the Alarms and Troubleshooting chapters for more information.
Visually monitor the patient and ventilator at all times during an Alarm Silence period. Allowing alarms to continue without intervention may result in harm to the patient.
If the high priority “Low Internal Battery” message appears, immediately connect the ventilator to an alternate power source. If no alternate power source is available, immediately place the patient on an alternate source of ventilation.
If the “Ventilator Inoperable” alarm occurs, immediately place the patient on an alternate source of ventilation.
You should not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect alarm.
Make sure the alarm volume is set loud enough to be heard by the caregiver. Consider the use of a remote alarm.
Trilogy200 offers the following circuit type selections:
The Passive circuit type provides an ESTIMATE of Vte.
Only the Active Flow circuit type directly measures exhaled tidal volume (Vte).
The Active PAP circuit type DOES NOT measure Vte and only provides for an indication of the delivered tidal volume (Vti).
If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if the device or detachable battery are dropped, if water is spilled into the enclosure, or if the enclosure is cracked or broken, discontinue use and contact Philips Respironics or an authorized service center for service.
Improperly Functioning Ventilator · Chapter 1 Introduction

Maintenance Follow the service recommendations provided in Chapter 7 of this manual.
Periodically inspect electrical cords, cables, and the detachable battery pack for damage or signs of wear. Discontinue use and replace if damaged.
Repairs and adjustments must be performed by Philips Respironics- authorized service personnel only. Unauthorized service could cause death or injury, invalidate the warranty, or result in costly device damage.
Cleaning (Refer to Chapter 7 for detailed cleaning instructions.)
To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the ventilator.
Do not immerse the device in any fluids or spray the device with water or cleaners. Clean the device with a cloth dampened with an approved cleaner.
If the device has been exposed to rain or dampness, dry the device including the area around the power cord connection with the power cord disconnected from the device before applying AC power.
A caution indicates the possibility of damage to the device.
Storage The internal and detachable batteries will self-discharge in storage. If it is desired to keep the batteries fully charged (for example, as a back- up ventilator), plug the device into AC power for about eight hours every 16 days. Alternatively, the ventilator may be left continuously connected to AC power without battery degradation.
Allowing the batteries to fully discharge will not harm the batteries or lose device settings, but may require a longer battery charge time prior to use.

Operating and Storage Temperatures
The device may only be operated at temperatures between 5˚ C and 40˚ C (41˚ F and 104˚ F).
Do not operate the device in direct sunlight or near a heating appliance because these conditions can increase the temperature of the airflow delivered to the patient.
Prolonged operation or storage at elevated temperatures may reduce the service life of the battery and other internal components of the ventilator.
The ventilator has an internal and detachable Lithium-Ion Battery. Do not expose the device or detachable battery to temperatures above 40˚ C (104˚ F) during use, or above 60˚ C (140˚ F) during storage. This will reduce battery life and may increase the risk of fire or damage the battery.
Condensation Condensation may affect operation or accuracy of the device. If the device has been exposed to either very hot or very cold temperatures during storage, allow it to adjust to ambient temperature before starting therapy.
Air Filter The reusable foam inlet filter is required to protect the ventilator from dirt and dust. Wash periodically and replace when damaged for proper operation.
Cooling Air Vents Do not block the cooling air vents located on the base and the rear of the device. This may cause the device to overheat in high ambient temperatures or at high therapy settings.
Battery Life The internal and detachable batteries wear out based on the amount of use (hours or full charge-discharge cycles). The battery capacity and life are also reduced by operation at higher temperatures.
Detachable Battery Only use the Philips Respironics Trilogy Detachable Battery with the ventilator.
Cleaning Do not steam autoclave the ventilator. Doing so will destroy the ventilator.
Do not immerse the device in liquid or allow any liquid to enter the enclosure or inlet filter.
Do not spray water or any other solutions directly onto the ventilator.
Do not use harsh detergents, abrasive cleaners, or brushes to clean the ventilator system. Use only cleaning agents and methods listed in this manual.
Chapter 1 Introduction

Patient Circuit Exhalation valves, patient circuits, and water traps are shipped clean, not sterile. Cleaning and disinfection of these parts should follow individual institution processes and conform to guidelines provided by Philips Respironics with each accessory.
External DC Power Do not use the same external battery to operate both the ventilator and any other equipment such as power chairs.
An external battery should only be connected to the ventilator using the Philips Respironics Trilogy External Battery Cable. This cable is fused, pre-wired, and properly terminated to ensure safe connection to a standard deep-cycle, lead acid battery. Use of any other adapter or cable may cause improper operation of the ventilator.
The ventilator should only be connected to an automotive electrical system using the Philips Respironics Trilogy Automotive Adapter (when available). This adapter is fused, filtered, and designed for safe connection to a standard automotive electrical system. Use of any other adapter or cable may cause improper operation of the ventilator.
Do not operate the ventilator from a car electrical system when starting the vehicle or jump-starting the vehicle. Electrical transients during starting may cause improper operation of the ventilator.
Do not use antistatic or conductive hoses or conductive patient tubing with the device.
Electrostatic Discharge (ESD)

If the patient has any of the following conditions, consult their health care professional before using the device in a non-invasive mode:
This ventilator provides both pressure control and volume modes of therapy. The device can provide non-invasive or invasive ventilation. It can be used to provide total therapy to patients as they progress from non-invasive to invasive ventilation.
When prescribed, the device provides numerous special features to help make patient therapy more comfortable. For example, the ramp function allows you to lower the pressure when trying to fall asleep. The air pressure will gradually increase until the prescription pressure is reached. Additionally, the Flex comfort feature provides increased pressure relief during the expiratory phase of breathing.
The ventilator can be operated using several different power sources, including an internal Lithium-Ion battery. This battery is automatically used when the detachable Lithium-Ion battery pack, external Lead Acid battery, or AC power are not available.
Chapter 1 Introduction

The following symbols appear on the device.
Alarm Indicator/Audio Pause ~ AC Power Indicator
Rear and Side Panels
Symbol Description ~ AC Power Connector
Secure Digital (SD) Card Slot
Consult accompanying instructions for use.
Type BF Applied Part
Class II (Double Insulated)

For Airline Use. Complies with RTCA-D0160F section 21, category M.
To have your device serviced, contact Philips Respironics Customer Service department at 1-724-387-4000 or 1-800-345-6443.
Drip Proof Equipment · Chapter 1 Introduction



This chapter describes the front and rear panel device controls and features.
The front panel contains the control buttons, visual indicators, and display screen.
Front Panel Controls and Display Screen
The following buttons are included on the front panel of the device.
This button turns the airflow on or off, starting or stopping therapy.
This button serves two purposes: it temporarily silences the audible portion of an alarm, and it also acts as an alarm indicator. When silencing an alarm, if the cause of the alarm is not corrected, the alarm sounds again after one minute. Each time the button is pressed, the alarm silence period resets to one minute. See Chapter 6 for more information.
To make sure the device is operating properly at start-up, always verify that the audible tone sounds and the alarm LEDs light red and then yellow momentarily. Contact Philips Respironics or an authorized service center for service if these indications do not occur at start-up.
This button allows you to navigate the display menu and edit device settings.
Chapter 2 System Description

These buttons allow you to select display options or perform certain actions specified on-screen.
Several power and alarm indicators appear on the front panel.
In the lower right corner of the front panel, a green LED (~) indicates that AC power is applied to the device. This light remains on as long as adequate AC power is available.
The Start/Stop, Up/Down, and Left/Right buttons all have a white LED that lights up if the keypad backlight is turned on in the device Options menu. See Chapter 5 for more information.
On the Alarm Indicator/Audio Pause button, a red light flashes to indicate a high priority alarm.
On the Alarm Indicator/Audio Pause button, a yellow light flashes to indicate a medium priority alarm. A solid yellow light indicates a low priority alarm.
The display screen allows you to view settings, system status information, real-time patient data, alarms, and logs. You can also modify certain settings on the display screen.
See Chapter 5 for more information on viewing and modifying device settings.

The ventilator’s side and rear panels contain the following connectors and features, shown at right.
You can plug the AC power cord into this connector, located on the right side of the ventilator.
The breathing circuit connector is located on the right side of the device. You can connect your circuit tubing system here. See Chapter 4 for details.
The porting block used here depends on the type of exhalation device you are using. The Universal Exhalation Porting Block is shown here. See Chapter 4 for more information.
On the left side of the device is a slot for the optional SD Card. You can have the patient record usage and therapy information from the device on the SD card.
Chapter 2 System Description
Right Side Panel · Left Side Panel

You can use this connector to connect the device to a computer running PC Direct or Sleepware software or to other Philips Respironics devices such as Alice 5 and AOM. Use the Trilogy RS232 Serial Cable to connect the ventilator to the external device or computer.
9 5 7 8 6
If you are using an optional remote alarm or nurse call system with the ventilator, you can connect the Philips Respironics remote alarm adapter cable or nurse call adapter cable to this connector.
You can connect a PC or router to this connector to upload therapy information to a secure web site so you can review therapy information remotely or remotely troubleshoot and service the device.
You can connect an external, stand-alone lead acid battery here, using the Trilogy External Battery cable.
If using low flow supplemental oxygen, connect the oxygen source to this connector using one of the O2 Inlet Quick Connects provided with the device.
Insert the filter supplied with the device into the air inlet.
If you are using the Philips Respironics Lithium-Ion detachable battery pack to power the device, attach it here.
Secure the power cord using the cord retainer to prevent someone from accidentally disconnecting the power cord. See Chapter 4 for more information.
Rear Panel

The device provides Pressure Control Ventilation (PCV) and Volume Control Ventilation (VCV) for non-invasive and invasive patients.
Pressure Control ventilation delivers a prescribed pressure to the patient according to set breath rate and set inspiration time parameters. This means that each breath is controlled so that a prescribed amount of pressure is delivered to the patient. The device offers six different Pressure Control modes of operation:
Volume Control ventilation delivers a prescribed inspired tidal volume to the patient according to set breath rate and set inspiratory time parameters. This means that each breath is controlled so that a prescribed tidal volume is delivered to the patient. The device offers three different Volume Control modes of operation:
Chapter 3 Modes, Features, and Alarms

There are four breath types that apply to the Volume Control and Pressure Control ventilation therapy modes:
A Spontaneous breath is triggered by the patient. Breaths are initiated by the patient’s inhalation effort, and air delivery is controlled based on the current pressure or volume setting. Breaths are terminated by either the ventilator settings or by the patient’s exhalation effort, depending on the mode selected.
A Mandatory breath (or machine breath) is completely controlled by the ventilator. The ventilator controls both the beginning (triggering) and end (cycling) of the inspiratory phase.
An Assisted breath is controlled by both the patient and the ventilator. Breaths are initiated by the patient’s effort and air delivery is controlled by the current pressure or volume settings. Volume Assisted breaths will deliver the prescribed Tidal Volume within the prescribed Inspiratory Time. Pressure Assisted breaths will deliver the prescribed Inspiratory Pressure for the prescribed Inspiratory Time. Breaths are terminated when the Inspiratory Time setting has been reached.
A Sigh breath is a breath where 150% of the prescribed volume is delivered. The device will deliver this breath once every 100 Mandatory or Assist breaths when the Sigh setting is enabled. Sigh breaths are only available in volume modes of ventilation.

The following table summarizes all of the therapy modes and the settings available in each mode. Some settings in the table are dependent upon other settings. For instance, if the circuit type is set to Active with Flow, then the Flow Trigger Sensitivity, Leak Compensation, and Flow Cycle settings will display.
CPAP S S/T T PC PC-SIMV CV AC SIMV
Dual Prescription √
Circuit Type √ CPAP √ IPAP √ AVAPS (On, Off)* √ IPAP Max Pressure √ IPAP Min Pressure √ EPAP √ Pressure √ Pressure Support (PS) √ √ PEEP √ Tidal Volume √ Breath Rate √ Inspiratory Time √ Trigger Type* √ Flow Trigger Sensitivity √ Leak Compensation √ Flow Cycle Sensitivity √ Ramp Length √ Ramp Start Pressure √ Flex * √ √ Rise Time √ Flow Pattern √ √ √ Sigh √ √ √ Circuit Disconnect √ Apnea √ Apnea Rate √ High Vte* √ Low Vte* √ High Vti* √ Low Vti* √ High Minute Ventilation √ Low Minute Ventilation √ High Respiratory Rate √ Low Respiratory Rate √ High Inspiratory Pressure √ √ √ Low Inspiratory Pressure √ √ √
* Trigger Type, AVAPS, and Flex settings are only available with the Passive circuit type. High Vte and Low Vte settings are only available for Passive and Active Flow circuit types. High Vti and Low Vti settings are only available with the Active PAP circuit type. Leak compensation is only available with the Active Flow circuit type.
Chapter 3 Modes, Features, and Alarms
Therapy Modes · Therapy Parameters

Pressure Control ventilation modes deliver a prescribed pressure to the patient.
Continuous Positive Airway Pressure (CPAP) Mode
In the Continuous Positive Airway Pressure (CPAP) mode, the device delivers a continuous pressure to the patient at all times. All breaths in this mode are Spontaneous breaths.
In Spontaneous (S) mode, the device delivers bi-level pressure support. This mode provides only spontaneous breaths. In this mode, an Inspiratory Positive Airway Pressure (IPAP) is delivered during inhalation and a lower Expiratory Positive Airway Pressure (EPAP) is delivered during exhalation. The following illustration describes these concepts.
IPAP = 26 cm H2O
EPAP = 6 cm H2O
PS = IPAP-EPAP = 20 cm H2O
5 seconds 4 seconds 8 seconds
5 10 15 20

In Spontaneous/Timed (S/T) mode, the device delivers bi-level pressure support. This mode provides Spontaneous and Mandatory breaths. A Mandatory breath is delivered if the patient does not spontaneously breathe within the prescribed Breath Rate (BPM) setting. This ensures that the patient receives a minimum number of breaths per minute. In this mode, an IPAP is delivered during inhalation and a lower EPAP is delivered during exhalation. The duration of a Spontaneous breath is determined by the patient effort. The duration of a Mandatory breath is determined by the inspiratory time setting. The following illustration describes these concepts.
IPAP = 26 cm H2O
EPAP = 6 cm H2O
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H2O
6 seconds 6 seconds
5 10 15 20
Chapter 3 Modes, Features, and Alarms

In Timed (T) mode, the device delivers bi-level pressure support. This mode delivers Mandatory breaths only. A Mandatory breath is delivered according to the prescribed BPM and inspiratory time settings. This also means that the ventilator will not respond to patient effort. In this mode, an IPAP is delivered during inhalation and a lower EPAP is delivered during exhalation. The following illustration describes these concepts.
IPAP = 26 cm H2O
= Mandatory Breath 2
EPAP = 6 cm H2O
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H2O
6 seconds 6 seconds 6 seconds 6 seconds
2 6 10
5 10 15 20 25
Timed (T) Mode · T Mode · Enter T mode

Pressure Control (PC) Mode
In Pressure Control (PC) mode, the device delivers bi-level pressure support. This mode delivers Assist and Mandatory breaths. This mode is identical to S/T mode, except that all breaths have a fixed inspiratory time.
IPAP = 26 cm H2O
EPAP = 6 cm H2O
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H2O
6 seconds 6 seconds 6 seconds
5 10 15 20
Pressure Control - Synchronized Intermittent Mandatory
The Pressure Control – Synchronized Intermittent Mandatory Ventilation (PC-SIMV) mode provides Spontaneous, Assist, and Mandatory breaths. This mode uses a time window to decide what type of breaths should be delivered. This time window is the maximum time between breaths according to the set Breath Rate.
When you enter PC-SIMV mode, the time window is started. If the patient does not trigger a breath, the ventilator will provide a Mandatory breath when the time expires and then begin the process again. Mandatory breaths deliver the Pressure setting during inhalation and a lower Peak End Expiratory Pressure (PEEP) during exhalation.
Chapter 3 Modes, Features, and Alarms

During the time window, if patient effort is detected, either a Spontaneous or an Assist breath is delivered for the first effort. If the last breath delivered was Mandatory, then a Spontaneous breath is delivered for the first patient effort of the window. If the last breath delivered was an Assist or Spontaneous breath, then an Assist breath is delivered for the first patient effort of the window. Assist breaths deliver the Pressure setting during inhalation and the PEEP setting during exhalation.
During the time window, if patient effort continues to be detected (after the first patient triggered breath), Spontaneous breaths are delivered for the remainder of the window. The figure below provides example breath patterns in PC-SIMV mode.
Pressure = 26 cm H2O
Pressure Support = 10 cm H2O
PEEP = 6 cm H2O
Rate = 5 BPM
12 seconds 12 seconds 12 seconds
S S
2 Sample Breath Patterns in PC-SIMV Mode
5 10 15 20 25 30 35 40 45

Volume Control ventilation modes deliver a prescribed inspired tidal volume according to a set Breath Rate and a set Inspiratory Time.
Control Ventilation (CV) Mode
In Control Ventilation (CV) mode, the device delivers volume control therapy. This mode provides only Mandatory breaths. A Mandatory breath is delivered according to the prescribed BPM setting for the prescribed Inspiratory Time. This also means that the ventilator will not respond to patient effort. In this mode, the Tidal Volume is delivered during inhalation and PEEP is delivered during exhalation. The following illustration describes these concepts.
Tidal Volume = 500 mL
= Mandatory Breath 2
Rate = 10 BPM
Inspiratory Time = 1.5 seconds
Flow Pattern = Square
6 seconds 6 seconds 6 seconds 6 seconds
Chapter 3 Modes, Features, and Alarms
Enter CV mode · Control Mode

In Assist Control (AC) mode, the device delivers volume control therapy. This mode provides Assist and Mandatory breaths. An Assist breath is started when there is patient effort, but it is ended when the Inspiratory Time setting has been met. A Mandatory breath is delivered if the patient does not spontaneously breathe within the prescribed BPM setting. This ensures that the patient receives a minimum number of breaths per minute. In this mode, the Tidal Volume is delivered during inhalation and PEEP is delivered during exhalation. The following illustration describes these concepts.
Tidal Volume = 500 mL
Rate = 10 BPM
= Assist Breath 3
Inspiratory Time = 2.0 seconds
Flow Pattern = Square
6 seconds 6 seconds 5 seconds
Synchronized Intermittent Mandatory Ventilation (SIMV) Mode
In Synchronized Intermittent Mandatory Ventilation (SIMV) mode, the device delivers both volume control and pressure control therapy. This mode provides Spontaneous, Assist, and Mandatory breaths. This mode uses a time window to decide what type of breaths should be delivered. This time window is the maximum time between breaths according to the set Breath Rate.
When you enter SIMV mode, the time window is started. If the patient does not provide any effort, the ventilator will provide a Mandatory Volume breath when the time expires and then begin the process again. Mandatory breaths deliver the Tidal Volume setting during inhalation and provide a lower Peak End Expiratory Pressure (PEEP) during exhalation.
= Mandatory Breath · Enter A/C mode · Assist Control Mode

During the time window, if patient effort is detected, either a Spontaneous or an Assist breath is delivered for the first effort. If the last breath delivered was Mandatory, then a Spontaneous breath is delivered for the first patient effort of the window. If the last breath delivered was an Assist or Spontaneous breath, then an Assist breath is delivered for the first patient effort of the window. Assist breaths deliver the prescribed Tidal Volume during inhalation and the PEEP setting during exhalation.
During the time window, if patient effort continues to be detected (after the first patient triggered breath), Spontaneous breaths are delivered for the remainder of the window. Spontaneous breaths deliver the prescribed Pressure Support setting above PEEP during inhalation and PEEP during exhalation. The figure below provides example breath patterns in SIMV mode.
Tidal Volume = 500 mL
Breath Rate = 5 BPM
Inspiratory Time = 3.0 seconds
Pressure Support = 10 cm H2O
PEEP= 6 cm H2O
12 seconds 12 seconds 12 seconds 12 seconds
S S
5 10 15 20 25 30 35 40 45
Chapter 3 Modes, Features, and Alarms

The device has several additional features that enhance patient comfort.
The device consists of a special comfort feature called Flex. The device provides the Flex feature in CPAP mode and S mode. This feature is only available when Auto-Trak is enabled.
When in CPAP mode, if C-Flex is enabled, it enhances patient comfort by providing pressure relief during the expiratory phase of breathing. In the following diagram, the dashed lines represent normal CPAP therapy in comparison to the bold line representing C-Flex. C-Flex levels of 1, 2, or 3 progressively reflect increased pressure relief.
CPAP
C-Flex in Comparison to Traditional CPAP Therapy
C-Flex pressure relief is determined by the C-Flex setting and the amount of patient flow. C-Flex returns to the set pressure by the end of exhalation, when the airway is most vulnerable to closure.
C-Flex · Inhalation Exhalation

In S mode, the Bi-Flex attribute adjusts therapy by inserting a small amount of pressure relief during the latter stages of inspiration and during the beginning part of exhalation. In the following diagram, the bold lines represent Bi-Flex in comparison to the dashed line representing normal BiPAP therapy. Bi-Flex levels of 1, 2, or 3 progressively reflect increased pressure relief that will take place at the end of inspiration and at the beginning of expiration.
IPAP
EPAP
Bi-Flex in Comparison to Traditional Bi-Level Therapy
Ramp
The device is equipped with a linear ramp function. In CPAP, S, S/T, T, and PC modes, the Ramp feature will reduce the pressure and then gradually increase (ramp) the pressure to the prescription pressure setting so patients can fall asleep more comfortably. The figure below illustrates how the ramp function works.
Chapter 3 Modes, Features, and Alarms

In S, S/T, PC, T, PC-SIMV, and SIMV modes, rise time is the amount of time it takes the device to change from the expiratory pressure setting to the inspiratory pressure setting. Rise time levels of 1, 2, 3, 4, 5, or 6 progressively reflect slowed response of the pressure increase that will take place at the beginning of inspiration. Adjust the rise time to find the most comfortable setting for the patient.
IPAP
EPAP
TIME
AVAPS Feature
Average Volume Assured Pressure Support (AVAPS) is a feature available in the S, S/T, PC, and T modes. It helps patients maintain a tidal volume (VT) equal to or greater than the target tidal volume (Volume setting in the ventilator) by automatically controlling the pressure support (PS) provided to the patient. The AVAPS feature adjusts PS by varying the IPAP level between the minimum (IPAP Min) and maximum (IPAP Max) settings. AVAPS averages VT and changes the PS value gradually. This occurs over several minutes. The rate of change is slow, so that the patient is not aware of breath-to-breath pressure changes.
As patient effort decreases, AVAPS automatically increases PS to maintain the target tidal volume. The IPAP level will not rise above IPAP Max, even if the target tidal volume is not reached. Conversely, as patient effort increases, AVAPS will reduce PS. IPAP will not fall below IPAP Min, even if the target tidal volume is exceeded.
If the Ramp function has been activated, it will take precedence over the AVAPS feature. Once ramp is complete, AVAPS will resume.
Rise Time · Rise Time

EPAP
TIME 15 30 45 60
VOLUME
TIME 15 30 45 60
Flow Pattern Types
Two flow patterns are available in Volume Control ventilation therapy modes:
With a square wave pattern, airflow is generally constant throughout inspiration of the breath.
TIME
Chapter 3 Modes, Features, and Alarms

With a ramp flow pattern, the airflow starts high and decreases throughout inspiration of the breath.
50% of Peak Flow
For the active circuit in volume modes, peak flow is required to be a minimum of 20 l/min. The wave form may be flattened when the combination of Inspiratory Time and Tidal Volume set points would result in a flow of less than 20 l/min. Therefore, for some settings, a Ramp flow pattern may provide a pattern that more closely resembles a Square flow pattern.
Ramp · Ramp Waveform Pattern

The sigh feature is available for Volume ventilation modes only. When the sigh feature is enabled, the ventilator delivers a sigh breath in place of every 100th mandatory or assisted breath delivered regardless of the mode of operation (i.e., AC, CV, and SIMV). The Sigh breath is delivered using a volume equal to 150% of the set volume that was in effect when the breath was initiated.
Example of Sigh Feature
Dual Prescription Feature
The device provides a dual prescription feature that allows you to enter a primary prescription and a secondary prescription for the patient if needed. For example, you can set a primary daytime prescription and secondary nighttime prescription. See Chapter 5 for more information on the dual prescription feature.
Chapter 3 Modes, Features, and Alarms

The device can be set to trigger breaths using the Auto-Trak or Flow Trigger sensitivity features.
An important characteristic of the device is its ability to recognize and compensate for unintentional leaks in the system and to automatically adjust its trigger and cycle algorithms to maintain optimum performance in the presence of leaks. This feature is known as Digital Auto-Trak Sensitivity. The following sections examine this function in detail by describing the leak tolerance function and sensitivity.
A microprocessor monitors the total flow of the patient circuit and calculates patient flow values.
The device uses two leak estimation algorithms. A conservation of mass algorithm is used to compute the average leak for a given pressure support relationship. This average leak is used when large leak variations are present in the system. Average leak is a high estimate during EPAP pressure and a low estimate during IPAP pressure. A better leak estimate, enabled by the digital system, is the parabolic leak algorithm. Parabolic leak is proportional to the square of the patient pressure; therefore, the leak estimate is correlated to the changing patient pressure. Both algorithms include unintentional circuit leak and are averaged over several breaths.
B. Patient Flow
The total circuit flow is comprised of the circuit leak and the patient flow. The calculated patient flow is the total flow minus the circuit leak. Patient flow is a primary input into the triggering and cycling mechanisms.
Digital Auto-Trak Sensitivity · Leak Tolerance

An essential feature of the device while operating in all modes is its ability to effectively sense spontaneous breathing efforts, which causes the ventilator to trigger to inspiration and cycle to expiration. Because no preset sensitivity threshold can assure patient and machine synchrony with changing breathing efforts and circuit leaks, the device continuously tracks patient breathing patterns and automatically adjusts sensitivity thresholds to ensure optimum sensitivity as breathing patterns change or as circuit leaks change. The algorithms used to ensure optimum sensitivity are the Volume Trigger, Shape Signal, Spontaneous Expiratory Threshold (SET), Flow Reversal, Maximum IPAP Time, and Volume Control Cycle.
Volume Trigger (Expiration to Inspiration):
The volume trigger is one method used to trigger inspiration during spontaneous breathing in all modes except T and CV. The volume trigger threshold is 6 ml of accumulated patient inspiratory volume. When patient effort generates inspiratory flow causing 6 ml of volume, inspiration is triggered.
Shape Trigger/Shape Cycle (Expiration to Inspiration) (Inspiration to Expiration):
The shape trigger/cycle is another method used to trigger inspiration and/or cycle from inspiration to expiration during spontaneous breathing in all modes except T and CV. This method continuously tracks patient inspiratory and expiratory flow and adjusts the spontaneous trigger and cycle thresholds for optimum sensitivity. The Shape Signal appears as a shadow image of the patient’s actual flow. The shape signal functions as a sensitivity threshold at either inspiration or expiration. When the patient’s flow rate crosses the shape signal the unit changes pressure levels. The following figure illustrates how the shape signal is super-imposed onto the actual waveform to trigger and cycle off IPAP. The shape signal is created by offsetting the signal from the actual patient flow by 15 l/min and delaying it for a 300 msec period. This intentional delay causes the shape signal to be slightly behind the patient’s flow rate.
Chapter 3 Modes, Features, and Alarms
Auto-Trak Sensitivity

A sudden change in patient flow will cross the shape signal, causing the pressure level to change.
EPAP
Cycle to EPAP Crossover Point
Trigger to IPAP Crossover Point
Tracking the patient’s flow pattern with the Shape Signal provides a sensitive mechanism to trigger to inspiration or cycle to expiration in response to changing breathing patterns and circuit leaks.

Spontaneous Expiratory Threshold (Inspiration to Expiration):
A second method used to cycle to expiration during spontaneous breathing in all modes except T, CV, AC, and SIMV, is called Spontaneous Expiratory Threshold (SET). The SET rises in proportion to the inspiratory flow rate on each breath. When the SET and actual patient flow value are equal, the unit cycles to expiration.
EPAP
Flow Reversal (Inspiration to Expiration):
As flow begins to decrease during inspiration, a flow reversal can occur due to a large leak around the mask or because the patient’s mouth is open. When the device senses this flow reversal, it automatically cycles to expiration.
Maximum IPAP Time (Inspiration to Expiration):
The maximum inspiratory time is determined by the adjustment of the Inspiratory time setting. A maximum IPAP time of 3.0 seconds acts as a safety mechanism to limit the time spent in inspiration during spontaneous breathing. For mandatory or assisted breaths, the maximum inspiratory time will equal the Inspiratory time setting up to 5.0 seconds.
Chapter 3 Modes, Features, and Alarms

Volume Control Cycle (Inspiration to Expiration) (Only available during Volume Control Therapy)
An Inspiratory Time setpoint limits the time spent in inspiration during breathing in all modes. Once the time limit is reached, the unit automatically cycles to expiration.
Flow trigger provides a manual setting that allows for breath initiation and termination based on a set flow trigger sensitivity and flow cycle sensitivity.
Flow Trigger Sensitivity (Expiration to Inspiration):
The flow trigger initiates when the patient’s inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. The method of the flow trigger is dependent upon the circuit type that is chosen.
Leak Compensation:
When using the Passive Circuit configuration, compensation for both the intentional and unintentional leak is included in the triggering method.
When using the Active PAP Circuit configuration, leak compensation is not available.
When using the Active Flow Circuit configuration, flow trigger with leak compensation may be enabled. The default setting when using the Active Flow Circuit is Leak Compensation On. The clinician has the option to turn off leak compensation; however, unintentional leak will not be compensated. Both options measure the flow at the proximal flow sensor.
Flow Cycle Sensitivity (Inspiration to Expiration):
This cycling method is only active if the Flow Trigger has been selected for the Trigger Type. As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle sensitivity setting, the device will cycle to expiration.
For example: if the flow cycle sensitivity setting is set to 75%, when the flow has decreased by 25% of the peak flow, the device will cycle to the EPAP/PEEP level.
Flow Trigger

All flows and volumes used in Trilogy are expressed in BTPS - Body Temperature atmospheric Pressure Saturated with H2O.
All pressures are expressed relative to atmospheric pressure.
This section describes all of the ventilator alarms and informational messages in order of priority, from high priority alarms to low priority alarms and finally informational messages. Refer to Chapter 6 for more information on alarms.
This is a high priority alarm. It may occur when a complete power failure has occurred and power was lost while the device was providing therapy. This may happen if the internal battery was the only power source in use and was completely depleted.
This is a high priority alarm. It occurs when the ventilator detects an internal error or a condition that may affect therapy. The device will shut down if the cause of the failure indicates that the device cannot deliver therapy safely. If the device can deliver therapy at a limited level, the device will continue to deliver limited therapy.
This is a high priority alarm. It occurs when the device cannot perform to specification, a backup safety feature is compromised, or the delivery of therapy is compromised. The device continues to operate (possibly in a reduced capacity mode). If the problem is not corrected, the device will generate a reminder message once per hour until the issue is corrected. Additionally, if therapy is stopped, a reminder message will immediately appear when therapy is turned on again.
Chapter 3 Modes, Features, and Alarms

This is a high priority alarm. It occurs when the device detects a problem with the patient circuit, such as pinched or detached tubing, water condensation in the proximal pressure lines, or problems with the active exhalation device.
This is a high priority alarm that only occurs with the passive circuit. It occurs when the system detects a problem with the leak device in the passive circuit.
This is a high priority alarm. It occurs when the delivered pressure exceeds the target patient pressure during the expiratory phase by 5 cmH2O. This may be due to pinched tubing or the patient having a fast breath rate. The device continues to operate. The alarm will automatically terminate when the delivered pressure comes within 5 cmH2O of the target patient pressure during the expiratory phase.
This is a high priority alarm. It occurs when the delivered pressure is 5 cmH2O or more below the target patient pressure during the expiratory phase. The device continues to operate. The alarm will automatically terminate when the delivered pressure comes within 5 cmH2O of the target patient pressure during the expiratory phase.
This is a high priority alarm. It occurs when there is a leak in the internal air delivery system that allows oxygen to build up inside the device. The alarm is generated when the internal oxygen concentration reaches 5% above ambient levels.

WARNING
This is a high priority alarm. It occurs when the breathing circuit is disconnected or has a large leak. The device continues to operate. The alarm will automatically terminate when the circuit is reconnected or the leak is fixed.
You should not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect alarm. The Apnea alarm is only intended for spontaneously breathing patients.
This is a high priority alarm. It occurs when the patient has not triggered a breath within the time specified in the apnea alarm setting. The device continues to operate. The alarm will automatically terminate when two consecutive patient breaths are detected that meet the apnea alarm time setting.
This is a high priority alarm. It occurs when the estimated exhaled tidal volume is greater than the High Vte alarm setting for three consecutive breaths. The device continues to operate. The alarm will automatically terminate when a breath occurs in which the exhaled tidal volume does not reach the High Vte alarm setting.
This is a high priority alarm. It occurs when the estimated exhaled tidal volume is lower than the Low Vte alarm setting for three consecutive breaths. The device continues to operate. The alarm will automatically terminate when a breath occurs in which the exhaled tidal volume exceeds the Low Vte alarm setting.
When AVAPS is on, this alarm will occur when the calculated tidal volume is less than 90% of the target tidal volume setting. The alarm will automatically terminate when a breath occurs in which the exhaled tidal volume is equal or greater to 90% of the target tidal volume setting.
Chapter 3 Modes, Features, and Alarms
High Vte Alarm · Low Vte Alarm

This is a high priority alarm. It occurs when the delivered tidal volume is greater than the High Vti alarm setting for three consecutive breaths. The device continues to operate. The alarm will automatically terminate when a breath occurs in which the delivered tidal volume does not reach the High Vti alarm setting.
This is a high priority alarm. It occurs when the delivered tidal volume is less than the Low Vti alarm setting for three consecutive breaths. The device continues to operate. The alarm will automatically terminate when a breath occurs in which the delivered tidal volume exceeds the Low Vti alarm setting.
This is a high priority alarm. It occurs when the respiratory rate is greater than the High Respiratory Rate alarm setting. The device continues to operate. The alarm will automatically terminate when the measured respiratory rate is less than the High Respiratory Rate alarm setting.
Low Respiratory Rate Alarm
This is a high priority alarm. It occurs when the respiratory rate is less than the Low Respiratory Rate alarm setting. The device continues to operate. The alarm will automatically terminate when the measured respiratory rate is greater than the Low Respiratory Rate alarm setting.
This alarm occurs in several stages and escalates from an audible beep for the first two occurrences to a high priority alarm if the problem continues. It is detected differently for volume and pressure therapy modes.
For volume modes, the alarm will sound if the measured patient pressure exceeds the High Inspiratory Pressure setting specified by the clinician. The alarm will automatically terminate when the peak inspiratory pressure is less than or equal to the High Inspiratory Pressure alarm setting.
High Vti Alarm · Low Vti Alarm

For pressure modes, the alarm occurs when the delivered pressure exceeds the target patient pressure by 5 cmH2O or more during the inspiratory phase. The device will automatically cycle to the expiratory phase and continue to operate. The alarm will automatically terminate when the delivered pressure falls within 5 cmH2O of the target patient pressure during the inspiratory phase.
This is a high priority alarm. It is detected differently for volume and pressure therapy modes.
For volume modes, the alarm will sound if the measured patient pressure is less than the Low Inspiratory Pressure setting specified by the clinician. The alarm will automatically terminate when the peak pressure at the end of the breath is greater than or equal to the Low Inspiratory Pressure alarm setting.
For pressure modes, the alarm occurs when the delivered pressure is 5 cmH2O or more below the target patient pressure during the inspiratory phase. The alarm will automatically terminate when the delivered pressure comes within 5 cmH2O of the target patient pressure during the expiratory phase.
High Minute Ventilation Alarm
This alarm is a high priority alarm. It occurs when the patient’s minute ventilation is greater than the High Minute Ventilation alarm setting. The device continues to operate. The alarm will automatically terminate when the calculated minute ventilation is less than the High Minute Ventilation alarm setting.
This alarm is a high priority alarm. It when the patient’s minute ventilation is less than the Low Minute Ventilation alarm setting. The device continues to operate. The alarm will automatically terminate when the calculated minute ventilation is greater than the Low Minute Ventilation alarm setting.
Chapter 3 Modes, Features, and Alarms

The Low Battery alarm occurs when the last battery available is low or nearly depleted. This alarm occurs in two stages. When approximately 20 minutes of battery run time remains, a medium priority alarm is generated, and the device continues to operate. If no action is taken and the battery continues to deplete, the alarm escalates to a high priority alarm when approximately 10 minutes of battery run time remains.
The ventilator has a two-stage low battery alarm. The medium priority alarm indicates that approximately 20 minutes of operation remain, and the high priority alarm indicates that less than 10 minutes of operation remain. Actual run time may be more or less than this and varies with battery age, environmental conditions, and therapy.
This alarm occurs when the estimated patient airstream temperature or the ventilator internal temperature is too high. The alarm occurs in several stages. The ventilator continues to operate. Internal fans are started when the medium priority alarm is generated. If the condition causing the high temperature is not corrected and the temperature continues to rise, the alarm will escalate to the high priority alarm.
Replace Detachable Battery Alarm
The Replace Detachable Battery alarm occurs when the detachable battery is nearing the end of its useful life or a failure in the detachable battery that prevents it from charging or discharging has been detected.
The alarm occurs in several stages, from low to high priority. The device continues to operate when the alarm is the low priority alarm. If the alarm is reset without removing the battery, the alarm will be regenerated once every hour until the detachable battery is removed. The device continues to operate, the detachable battery is not used, and the power source is switched to the next available power source if the alarm is the high priority alarm.
Ventilator Service Recommended Alarm
This is a medium priority alarm. It occurs when the device has detected an error, but the error will not affect device performance or safety. The device continues to operate. Therapy and safety are not compromised.
Low Battery Alarm

If the problem is not corrected, the device will generate a reminder message once per day, or whenever power is cycled, until the issue is corrected. Additionally, if the device is powered off, a reminder message will immediately appear when the device is turned on again.
AC Power Disconnected Alarm
This is a medium priority alarm. It occurs when the AC power source was lost, and the device has switched to an alternate power source (either a detachable or external battery, if connected, or the internal battery if no other source is available). The device continues to operate. If AC power returns, the ventilator will beep, but no message will appear on the display.
This is a low priority alarm. It occurs when a key becomes lodged inside the case of the device.
Info Message
This info message occurs when the detachable or internal battery becomes overheated while providing power for the device. The device continues to operate. The battery is not used and the power source is switched to the next available power source.
This info message occurs when the detachable or internal battery becomes too hot while charging or the device was in too cold an environment before charging started. The device continues to operate. Battery charging stops until the battery cools or warms sufficiently.
This info message occurs when the device has detected an error condition with the battery that prevents it from accepting a charge. The device continues to operate. Battery charging stops.
Chapter 3 Modes, Features, and Alarms

This info message occurs when a bad connection exists to the external battery or the external battery failed. The device continues to operate using power from the detachable battery, if available, or the internal battery.
This info message occurs when the affected battery is fully depleted. The device continues to operate using the next available power source.
External Battery Disconnected Info Message
This info message occurs when the external battery power source is lost and the device has switched to an alternate power source (either a detachable battery, if connected, or the internal battery if no other source is available). If external battery power returns, the ventilator will beep, but no message will appear on the display.
This info message occurs when the detachable battery power source is lost and the device has switched to an alternate power source (the internal battery if no other source is available). If detachable battery power returns, the ventilator will beep, but no message will appear on the display.
This info message indicates that the ventilator has started on battery power and no AC power is available. The device operator should verify that this is what is wanted.
Card Error Info Message
This info message occurs when an unusable SD card is inserted into the ventilator. The device continues to operate but data cannot be logged onto the SD card.

This chapter provides instructions on how to assemble the ventilator. It includes the following setup information:
Ventilator dependent patients should be continuously monitored by qualified personnel. These personnel should be prepared to provide alternate therapy in the event of ventilator failure or inoperative equipment.
Do not use the ventilator on a patient until a system checkout has been performed. See Chapter 10 of this manual.
For ventilator dependent patients, always have alternate ventilation equipment, such as a back-up ventilator, manual resuscitator, or similar device, available.
Do not operate the ventilator at temperatures below 5° C (41° F) or above 40° C (104° F).
Chapter 4 Ventilator Setup

Place the ventilator on a flat, level surface. Or, as an alternative, you can place the device in the optional in-use bag. The in-use bag is required for portable use in order to properly secure the ventilator.
Do not operate the ventilator while on its side, upside-down, or in any other orientation. Make sure that the air inlet on the back of the device is not blocked. If you block the air flow around the device, the ventilator may not work properly.
Install the Air Filter
The reusable foam inlet filter is required to protect the ventilator from dirt and dust. Wash periodically and replace when damaged for proper operation.
The device uses a gray foam filter that is washable and reusable. The reusable filter screens out normal household dust and pollens. The filter must be in place at all times when the device is operating. Two reusable gray foam filters are supplied with the device. If the filter is not already installed when you receive the device, you must install the filter before using the ventilator.

To install the filter, insert the gray foam filter into the filter area as shown.
The device can operate on AC or DC power. The ventilator accesses power from potential sources in the following order:
Installing the Filter

An AC power cord is provided with the device.
The device may only be operated at temperatures between 5˚ C and 40˚ C (41˚ F and 104˚ F).
Periodically inspect the power cord for damage or signs of wear. Discontinue use and replace if damaged.
Connecting the AC Power Cord
Chapter 4 Ventilator Setup

Securing the Power Cord Using the Cord Retainer
You can operate the ventilator using an external battery, detachable battery, or the internal battery.
Do not use the same external battery to operate both the ventilator and any other equipment such as power chairs.
The ventilator can operate from a 12 VDC deep cycle marine-type (lead acid) battery using the Philips Respironics Trilogy External Battery Cable. This cable is pre-wired and properly terminated to ensure safe connection of an external battery to the ventilator. Battery operating time depends on the characteristics of the battery and usage of the device.
An external battery should only be connected to the ventilator using the Philips Respironics Trilogy External Battery Cable. This cable is fused, pre-wired and properly terminated to ensure safe connection to a standard deep cycle lead acid battery. Use of any other adapter or cable may cause improper operation of the ventilator.
Due to a variety of factors, including battery chemistry, battery age, and use profile, the capacity of the external battery as shown on the device display is only an estimate of the actual remaining capacity.
Refer to the instructions supplied with the Philips Respironics Trilogy External Battery Cable for detailed information on how to operate the device using an external battery.
External Battery

Philips Respironics offers a detachable Lithium-Ion battery pack. To use the detachable battery pack, snap the battery into place on the back of the ventilator, as shown in the following illustration. When the device is not connected to an AC power source or an external battery, the detachable battery will power the device, if attached. The length of time the ventilator will operate on battery power depends on many factors such as device settings, battery charge level, and condition or age of the battery. When fully charged, a new battery can power the ventilator for approximately three hours under typical patient conditions.
The detachable and internal batteries wear out based on the amount of use (hours or full charge-discharge cycles). The battery capacity and life are also reduced by operation at higher temperatures.
Whenever the ventilator is connected to AC power, it will automatically recharge the detachable battery pack. A completely discharged detachable battery will reach 80% charge status within 8 hours, when charging at approximately 23˚ C ambient temperature.
Only use the Philips Respironics Trilogy Detachable Battery with the device.
Insert and securely latch the detachable battery into the device as shown below.


Attaching the Detachable Battery
Prolonged operation or storage at elevated temperatures may reduce the service life of the detachable or internal battery and other internal components of the ventilator.
Chapter 4 Ventilator Setup

One side of the detachable battery has a set of LEDs that indicate the amount of charge left on the battery. You can press the button below the LEDs to view how much charge remains:
All 5 LEDs are lit 80-100% capacity
4 LEDs are lit 60-79% capacity
3 LEDs are lit 40-59% capacity
2 LEDs are lit 20-39% capacity
1 LED is lit 10-19% capacity
1 LED flashes 1 to 9% capacity
0 LEDs are lit 0% capacity
The device contains an internal battery that can be used as a back- up power source. It is intended for use during short periods while switching between external power sources, emergency situations, or short durations when the user needs to be mobile. The length of time the ventilator will operate on internal power depends on many factors such as device settings, battery charge level, and condition or age of the battery. When fully charged, a new battery can power the ventilator for approximately three hours under typical patient conditions.
The internal battery is NOT intended to serve as a primary power source. It should only be used when other sources are not available, or briefly when necessary; for example when changing power sources.
Whenever the ventilator is connected to AC power, it will automatically recharge the internal battery. A completely discharged internal battery will reach 80% charge status within 8 hours when charging at approximately 23˚ C ambient temperature.
There are many power source indicators on the device and the display screen. These indicators are described in detail below.
When AC power is applied to the device, the green AC LED indicator (~) on the front of the device lights.

The internal, detachable, and external battery symbols that will display on the Monitoring screen are shown below. The detachable and external battery symbols will only appear on-screen if a detachable or external battery is attached to the device.
There are several DC power indicators that will display on the Monitoring screen to indicate which battery is in use (if applicable), if the batteries are low, charging, or discharged, etc. The following table explains all of the DC power indicators.
DC Power Indicator Description
Battery In Use Indicator A black box will appear around the battery that is in use. For instance, if the external battery is currently in use, the
symbol appears on the Monitoring screen.
Green Fully Charged Battery Indicator
When a battery is charged to greater than 90% of its capacity, all of the bars in the battery symbol will appear in green.
Partially Charged Battery Indicator
When a battery is partially charged, some of the bars in the battery symbol will appear in green, while others will be clear. For instance, if the external battery is 50% charged, the following symbol displays on- screen:
Yellow Low Battery Indicator (Medium Priority)
When the device detects that an in-use battery’s charge is low (has approximately 20 minutes of charge left), the inside of the box surrounding the battery symbol turns yellow. (In addition to the battery indicator on the Monitoring screen, a medium priority alarm message will display indicating “Low Battery.” See Chapter 6 for more information. The yellow indicator is for the last available battery source.
Chapter 4 Ventilator Setup

DC Power Indicator Description
Red Low Battery Indicator
When the device detects that an in-use battery’s charge is nearly depleted (has approximately 10 minutes of charge left), the inside of the box surrounding the battery symbol turns red. In addition to the battery indicator on the Monitoring screen, a high priority alarm message will display indicating “Low Battery.” See Chapter 6 for more information. The red indicator is for the last available battery source.
Yellow Battery Recharging Symbol ( )
Whenever AC power is applied to the device, the internal and detachable batteries will recharge as needed. If the internal battery is being recharged, the
symbol displays. If the detachable battery is being recharged, the
symbol displays.
Dispose of the batteries in accordance with local regulations.
When setting up the device for the first time or after annual service, apply AC power to the ventilator before turning on the blower. Attempting to use the ventilator without first applying AC power, such as installing a detachable battery pack and starting the blower, will cause the internal battery to be displayed in red as an empty battery. When in this state, the internal battery will not be usable until AC power is applied.

Complete the following steps to set up your patient circuit.
Philips Respironics recommends that a main line outlet bacteria filter (Part Number 342077) be used whenever the device is used for invasive therapy or if the ventilator may be used on multiple patients.

Connecting the Breathing Circuit to the Device
a. If your circuit type is Passive and you are using a Philips
Respironics Passive Exhalation Device:

Connecting the Passive Exhalation Device
Bacteria Filter Whisper Swivel II
Refer to the instructions included with the Passive Exhalation Device for more detailed setup information.
Chapter 4 Ventilator Setup

b. If your circuit type is Active PAP:
Active Exhalation Device with Proximal Pressure
A. Connect the active exhalation device with proximal
pressure to the flexible tubing that attaches to the ventilator and to the patient interface (e.g., the tracheostomy tube).
For ventilator dependent patients, always have alternate ventilation equipment, such as a back-up ventilator, manual resuscitator, or similar device, available.
A. Attach the proximal pressure line (.476 cm or 3/16”
line) to the proximal pressure port as shown in the next illustration.
B. If not attached, connect the other end of the proximal
pressure line to the Trilogy Universal Active PAP Tube Adapter as shown in the next illustration.
C. Connect the Trilogy Universal Active PAP Tube Adapter
to the top, striped port on the Universal Porting Block on the ventilator.
D. Connect the exhalation valve line (.317 cm or 1/8” line)
to the exhalation valve port on the top of the active exhalation device with proximal pressure.
E. Connect the other end of the exhalation valve line
to the exhalation valve port on the Universal Porting Block.


Connecting an Active Device with Proximal Pressure
Refer to the instructions included with the Active Exhalation Device for more information.
c. If your circuit type is Active Flow:

Connecting the Flow Sensor
A. Connect the active exhalation device to the flexible
tubing that attaches to the ventilator.
B. Connect the flow sensor to the patient interface (e.g.,
the tracheostomy tube).
Chapter 4 Ventilator Setup

C. Connect the flow lines and the exhalation valve line to
the Universal Porting Block on the ventilator as shown below.
Exhalation Valve port on the top of the active exhalation device, and connect the other end of the line to the Exhalation Valve port on the Universal Porting Block.
Universal Porting Block on Trilogy Device White Striped Flow Line Exhalation Valve Port

Connecting the Active Exhalation Device with Proximal Flow Sensor
Refer to the instructions included with the active exhalation device for more information.
If you are using an optional water trap, connect it to the patient circuit according to the manufacturer’s instructions.

To attach supplemental oxygen to the device:
Oxygen Inlet Quick Connect
Attaching the O2 Tubing
Press Tubing in and Latch Pops Up
This device is only compatible with a low flow oxygen source providing up to 15 l/min.
Do not connect an unregulated or high pressure oxygen source to this connector on the device.
Chapter 4 Ventilator Setup
Oxygen Inlet · Oxygen Source · Tubing

You can use a remote alarm as a nurse call system or in-house remote alarm system. It can generate an alarm at a distance of up to 91 meters (300 feet) from the ventilator. An audible tone sounds and a red light blinks to indicate that an alarm condition exists. The remote alarm will sound when any of the following conditions exist:
When using a remote alarm, make sure you fully test the remote alarm connector and cable by verifying that:
- Annunciated alarms on
the ventilator are also annunciated on the remote alarm.
- Disconnecting the remote
Refer to the Accessories chapter of this manual for information on which remote alarm systems are compatible with the device. Refer to the instructions included with the Remote Alarm Adapter Cable for information on how to connect a remote alarm to the ventilator.
alarm cable from the ventilator or from the remote alarm results in an alarm notification at the remote alarm.
The remote alarm should be tested daily.

This chapter explains how to scroll through the ventilation screens and change ventilation settings. It also describes how to connect the ventilator to the patient once the settings are complete.
There is a Keypad Lock feature that users can enable from the Options menu. It is intended to prevent accidental changes to device settings. This feature will lock the navigation keys (Up, Down, Stop, Left, and Right keys). If the Keypad Lock is enabled, a Keypad Unlock message will display on the bottom of the screen any time you press one of the navigation keys.
If the keypad is locked, you must unlock it before you can enter the Menu. To unlock the keypad and enter the menu, you must first hold the Right button for 5 seconds to unlock the keypad. An audible indicator sounds when the keypad is successfully unlocked. Once the display is unlocked, you can enter the Menu as you normally would by pressing the Up button.
Chapter 5 Viewing and Changing Settings


Monitor Screen – Detailed View Off
– The top section of the screen, called the Monitor panel, shows the therapy mode and, if you set up a dual prescription for the patient, the Prescription indicator appears, indicating Primary or Secondary prescription. The patient breath symbol also displays during a patient-triggered breath, and a bar graph displays the current pressure level.
– The center section of the screen displays the current date and time.
– The bottom section, called the Status panel, displays certain symbols that indicate features being used, such as Ramp, as well as battery status.

If Detailed View is turned on in the Options menu, the Monitor screen will look like the screen shown below.

Monitor Screen – Detailed View On
This screen contains more detailed information about the therapy.
– The top Monitor panel contains the Prescription indicator if a dual prescription exists, the therapy mode, a graph displaying the current pressure, and the current date and time. Additionally, this panel also displays patient pressure, respiratory rate (RR), exhaled tidal volume (Vte), and leak.
– The second panel in Detailed view is the Measured Settings panel. It provides patient-related data including Peak Inspiratory Pressure (PIP), Minute Ventilation, Peak Inspiratory Flow, Mean Airway Pressure (MAP), and I:E Ratio.
– The third panel is the Status panel and shows the same information displayed in the Detailed View Off screen, including features in use such as Ramp and battery status.
Chapter 5 Viewing and Changing Settings

This section describes the following indicators:
All of the indicators that may appear on the Monitor Panel are described in detail in the following table.
Prescription If you set up a dual prescription for the patient, the words “Primary” or “Secondary” appear in the top left corner of the panel to indicate which prescription is active.
Therapy Mode The current therapy mode displays at the top of the panel (for example, CPAP, S, S/T, etc.). If a special feature such as Flex, AVAPS, or Sigh is active, this feature will appear next to the therapy mode.
Date and Time If you are in Detailed view, the current date and time appear in the top right corner of the panel. (In Detailed View Off, they appear in the center panel.)
Patient Breath This symbol displays during a patient-triggered breath.
Airway Pressure Manometer and Peak Pressure Symbol
The manometer (bar graph) displays the airway pressure in the patient circuit at all times. The manometer bar moves to the right as pressure increases during inhalation, and moves to the left as pressure decreases during exhalation. The peak pressure is also indicated on this bar. It is positioned according to the maximum patient pressure reached during each breath. The Peak Pressure symbol appears as a blue bar on the manometer. If a High Inspiratory Pressure alarm occurs, the Peak Pressure symbol changes from blue to red.
Monitor Panel Indicators · Indicator Description

Low Pressure Indicator L
If you enable a volume therapy mode, this indicator appears below the manometer bar, indicating the low pressure alarm setting.
High Pressure Indicator H
If you enable a volume therapy mode, this indicator appears below the manometer bar, indicating the high pressure alarm setting.
Pressure This indicator displays the current patient pressure. This only appears in detailed view.
Respiratory Rate (RR) This indicator displays the measured respiratory rate in Breaths Per Minute (BPM). This only appears in detailed view.
Exhaled Tidal Volume (Vte)
This indicator displays the estimated exhaled tidal volume in milliliters and reflects compensation for BTPS. This only appears in detailed view when Passive Circuit is selected.
Inhaled Tidal Volume (Vti)
This indicator displays delivered tidal volume in milliliters and reflects compensation for BTPS. This only appears in detailed view when the Active with PAP Circuit is selected.
Leak This indicator displays the total leak (non-returned flow), between the unit outlet and the patient, averaged over the previous breath. This only appears in detailed view when the Passive Circuit is selected.
Chapter 5 Viewing and Changing Settings
Indicator Description

All of the indicators that may appear on the Measured Settings panel (available only in Detailed view), are described in the following table.
PIP Peak Inspiratory Pressure displays the maximum pressure delivered to the patient during the previous breath.
I:E Ratio Displays a comparison of the time spent in inspiration to the time spent in expiration during the previous breath.
Peak Flow Displays the maximum inspiratory flow delivered to the patient during the previous breath in l/min BTPS.
MAP Displays the Mean Airway Pressure, which is the weighted average of pressure in the patient’s airway over 6 breaths.
MinVent Minute ventilation displays the amount of air delivered to the patient over the last minute in l/min BTPS.
Measured Settings Panel · Indicator Description

All of the indicators that may appear on the Status Panel are described in the following table.
Indicates that the device is in Full Menu Access mode, which means you can adjust all prescription settings. Philips Respironics recommends that you change the device to Limited Menu Access mode before giving the device to the patient, so patients cannot adjust their prescription settings. Only trained health care professionals and clinicians should adjust prescription settings.
Displays when a Secure Digital (SD) memory card is inserted in the ventilator.
Displays when the ventilator detects an error with the SD card.
Displays at all times when an external battery is attached to the ventilator. The level of green shading shown in the symbol indicates the battery capacity and will go down as the battery charge level decreases. When the entire symbol is green, the battery is fully charged.
Displays at all times when a detachable battery is attached to the ventilator. The level of green shading shown in the symbol indicates the battery capacity and will go down as the battery charge level decreases. When the entire symbol is green, the battery is fully charged.
Displays at all times, indicating the status of the internal battery. The level of green shading shown in the symbol indicates the battery capacity and will go down as the battery charge level decreases. When the entire symbol is green, the battery is fully charged.
Chapter 5 Viewing and Changing Settings
Status Panel Indicators · Indicator Description

A black box displays around the battery that is currently supplying power to the ventilator when AC power is not available. (In the status panel shown above, the external battery is in use, so the
symbol displays.)
A yellow lightning bolt symbol displays with the Detachable or Internal battery symbol to indicate when the battery is charging. (In the status panel shown on the previous page, the detachable battery is being charged, so the
symbol displays.)
Displays when the Alarm Indicator/Audio Pause button has been pressed and Audio Pause is active. The alarm is silenced for one minute when the Alarm Indicator/Audio Pause button is pressed.
Displays when the Ramp feature is active.
Note: If a battery in use is very low (less than 20 minutes remaining), the inside of the box surrounding the battery symbol will change to yellow and all of the bar indicators in the battery will be empty. If a battery in use is near depletion (less than 10 minutes remaining), the inside of the box surrounding the battery symbol will change to red, and all of the bar indicators in the battery will be empty. These color changes only occur for the last available battery source.
Indicator Description

The illustration below shows the on-screen button panel on the Main Menu screen, in relation to the buttons on the front of the device.

Example On-Screen Button Panel
At the very bottom of the display screen is the on-screen button panel. This panel corresponds with the control buttons on the ventilator:
To navigate through all of the menu screens and settings:
Chapter 5 Viewing and Changing Settings
On-Screen · Button · Panel

Clinicians can view and change settings using the Menu screens when the device is in Full Menu Access mode. To enter the Menu screens from the Monitor screen, press the Up button on the ventilator. The Main Menu screen shown below appears.

Main Menu Screen Example
Choose from the following selections on the Main Menu screen:

From the Main Menu screen, use the Up/Down button to highlight the Settings and Alarms menu, and press the Right button to select the menu.
Device Settings Common to All Therapy Modes
Some of the settings on this menu will vary depending on the therapy mode you select. The section below describes all of the settings that are common to all therapy modes.
You can turn the dual prescription setting on or off. Enable the setting if you want to create two separate prescriptions for the patient. For instance, you may want to set up a daytime prescription and then a separate nighttime prescription. If you enable this setting, then the menu options on the Main Menu will change to include three new options:
The Main Menu screen will look like the screen below.

Main Menu Screen with Dual Prescription Enabled
Chapter 5 Viewing and Changing Settings
Dual Prescription Setting

You can change the Mode setting on the Settings and Alarms screen to one of the following therapy modes:
There are three circuit types you can select:
The Passive circuit type uses the Whisper Swivel II passive exhalation device. The Active PAP circuit type uses an active exhalation device with a proximal air pressure sensing connection. The Active Flow circuit type uses an active exhalation device with a proximal flow sensor.
When using the Passive circuit, the ventilator displays estimated patient pressures based on the resistance of the standard patient circuit (Whisper Swivel II with 1.8 meter tubing). Adding accessories to the patient circuit (humidifier, water trap, etc.) may cause an increase in circuit resistance and cause the device to display slightly higher pressures than what is actually delivered to the patient.
With the Active PAP or Active Flow circuit type selected, patient pressure is measured directly and is not affected by any change in circuit resistance.
Mode Setting · Circuit Type

The Passive circuit provides leak compensation. When using the Passive circuit in Volume Ventilation, the set Vti is delivered to the patient above the calculated circuit and cuff (or mask) leak. This is different from traditional active circuit ventilation where the cuff (or mask) leak reduces the tidal volume delivered to the patient. Volume ventilation with the Passive circuit delivers an inspiratory tidal volume close to the device setting regardless of leak; this should be considered when transitioning a patient from an active to a passive circuit. With a Passive circuit, Vte is estimated based on the calculated sum of circuit and cuff (or mask) leak.
The Active Flow circuit monitors proximal flow and proximal pressure. When using the Active Flow circuit configuration, flow trigger with leak compensation may be enabled. The default setting when using the Active Flow Circuit is Leak Compensation On. The clinician has the option to turn off leak compensation; however, unintentional leak will not be compensated. Both options measure the flow at the proximal flow sensor. Leak compensation is not available in the Active with PAP circuit configuration.
You should not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect alarm.
This setting enables or disables the circuit disconnect alarm. If enabled, an audible alarm will sound when a large, continuous air leak (such as mask removal) has been detected in the circuit.
You can choose Off to disable the alarm. Or, you can increase or decrease the setting from 5 to 60 seconds in 5 second increments. For example, a setting of 10 means that the alarm will sound after the circuit has been disconnected for 10 seconds.
This setting enables or disables the apnea alarm. If enabled, an audible alarm will sound when an apnea is detected.
You can choose Off to disable the alarm. Or, you can increase or decrease the setting from 10 to 60 seconds in 5 second increments. For example, a setting of 10 means that the alarm will sound if the time between spontaneous breaths exceeds 10 seconds.
Chapter 5 Viewing and Changing Settings
Circuit Disconnect · Apnea

This setting enables or disables the Low Vte alarm. The alarm activates when the estimated exhaled tidal volume is less than or equal to this setting. You can choose Off to disable this alarm, or you can increase or decrease the setting from 40 ml to 2000 ml in 5 ml increments. It cannot be set higher than the High Vte setting.
When AVAPS is On, the alarm activates when the calculated tidal volume is less than 90% of the target tidal volume setting. This alarm can be set to on or off.
This setting enables or disables the High Vte alarm. The alarm activates when the estimated exhaled tidal volume is greater than or equal to this setting. You can choose Off to disable this alarm, or you can increase or decrease the setting from 50 ml to 2000 ml in 5 ml increments. It cannot be set lower than the Low Tidal Volume setting, except to be turned off.
This setting enables or disables the Low Vti alarm. The alarm activates when the measured inhaled tidal volume is less than or equal to this setting. You can choose Off to disable this alarm, or you can increase or decrease the setting from 40 ml to 2000 ml in 5 ml increments. It cannot be set higher than the High Vti setting.
This setting enables or disables the High Vti alarm. The alarm activates when the measured inhaled tidal volume is greater than or equal to this setting. You can choose Off to disable this alarm, or you can increase or decrease the setting from 40 ml to 2000 ml in 5 ml increments. It cannot be set lower than the Low Vti setting, except to be turned off.

This setting enables or disables the Low Minute Ventilation alarm. The alarm activates when the calculated minute ventilation is less than or equal to this setting. You can choose Off to disable this alarm, or you can increase or decrease the setting from 0.1 l/min to 99 l/min.
This setting enables or disables the High Minute Ventilation alarm. The alarm activates when the calculated minute ventilation reaches or exceeds this setting. You can choose Off to disable this alarm, or you can increase or decrease the setting from 1 l/min to 99 l/min in 1 l/min increments. It cannot be set lower than the Low Minute Ventilation setting except to be turned off.
This setting enables or disables the Low Respiratory Rate alarm. The alarm activates when the measured respiratory rate is less than or equal to this setting. You can choose Off to disable this alarm, or you can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM increments. It cannot be set higher than the High Respiratory Rate.
This setting enables or disables the High Respiratory Rate alarm. The alarm activates when the measured respiratory rate reaches or exceeds this setting. You can choose Off to disable this alarm, or you can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM increments. It cannot be set lower than the Low Respiratory Rate except to be turned off.
Chapter 5 Viewing and Changing Settings

This setting configures the Low Inspiratory Pressure alarm. It is only user-settable in CV, AC, and SIMV modes. It cannot be set lower than PEEP + 2 cm H2O or higher than the High Inspiratory Pressure. For passive circuits, you can increase or decrease the Low Inspiratory Pressure from 6 to 40 cm H2O in increments of 1 cm H2O. For active circuits, you can increase or decrease the setting from 2 to 40 cm H2O in increments of 1 cm H2O. For pressure modes, this alarm is not user- settable.
This setting enables or disables the High Inspiratory Pressure alarm. It is only user-settable in CV, AC, and SIMV modes. The High Inspiratory Pressure cannot be set lower than the Low Inspiratory Pressure. You can increase or decrease the High Inspiratory Pressure from 10 to 80 cm H2O in increments of 1 cm H2O. For pressure modes, this alarm is not user-settable.
Additional Settings Specific to Therapy Modes
The Settings and Alarms menu also contains many additional settings specific to the various therapy modes. The specific settings for each therapy mode are described below.
Continuous Positive Airway Pressure (CPAP) Mode
In addition to the general settings described in the previous section of this manual, you can also set the following settings in CPAP mode.
You can increase or decrease the CPAP pressure setting from 4 to 20 cm H2O in increments of 1.
The device can be set to trigger breaths based on automatic flow thresholds or specific flow settings. You can select Auto-Trak or Flow Trigger as the Trigger Type. When Auto-Trak is selected, the Auto-Trak trigger initiates based on automatic flow thresholds.
Low Inspiratory Pressure · High Inspiratory Pressure

When Trigger Type is set to Flow Trigger, Flow Trigger Sensitivity Leak Compensation, and Flow Cycle Sensitivity become active, and the trigger initiates based on the Flow Trigger Sensitivity setting.
If you set the Trigger Type to Flow Trigger, the Flow Trigger Sensitivity setting displays. You can increase or decrease the setting from 1 to 9 l/min in 1 l/min increments.
The flow trigger initiates when the patient’s inspiratory effort creates a flow equal to or greater than the flow sensitivity setting.
If you are using an Active Flow circuit, you can turn Leak Compensation On or Off.
If you set the Trigger Type to Flow Trigger, the Flow Cycle Sensitivity setting displays. You can increase or decrease the setting from 10 to 90 percent (%) in 1% increments.
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the device will cycle to expiration. For example: if the flow cycle is set to 75%, when the flow has decreased by 25% of the peak flow, the device will cycle to the EPAP/PEEP level.

The Ramp Length allows you to set the ramp time. You can disable Ramp by selecting Off, or you can increase or decrease the Ramp Length setting from 5 to 45 minutes in 5-minute increments.
Chapter 5 Viewing and Changing Settings
Flow Cycle Sensitivity

You can increase or decrease the ramp start pressure in increments of 1 from 4 cm H2O to the CPAP pressure setting. The patient also has access to this setting, unless the ramp length is set to Off.
You can enable or disable the Flex setting. Off disables the setting and prevents the patient from using Flex. To enable the setting, set Flex to 1, 2, or 3. The patient also has access to this setting, if Flex is enabled. The Flex feature is not available when using an active circuit type.
The following settings, described in the CPAP mode section of this chapter, also are available in S mode:
In addition to those settings, the settings below are also available in S mode.
AVAPS is only available if Flex is not enabled.
You can disable AVAPS by selecting Off, or you can enable AVAPS by selecting On. If you select Off, the IPAP setting displays. If you select On, the IPAP Max Pressure and IPAP Min Pressure display.
Spontaneous (S) Mode

The IPAP setting displays if AVAPS is Off. You can increase or decrease the Inspiratory Positive Airway Pressure (IPAP) from 4 to 50 cm H2O in increments of 1. IPAP is limited to a maximum of 25 cm H2O when Flex is enabled. You cannot set the IPAP setting lower than the EPAP setting.
The IPAP Max Pressure setting displays if AVAPS is enabled. You can increase or decrease the setting from 4 to 50 cm H2O in increments of 1. The IPAP Max Pressure must be equal to or greater than the IPAP Min value.
The IPAP Min Pressure setting displays if AVAPS is enabled. You can increase or decrease the setting from 4 to 50 cm H2O in increments of 1. The IPAP Min Pressure must be equal to or greater than the EPAP value, and it must be less than or equal to the IPAP Max Pressure.
You can increase or decrease the Expiratory Positive Airway Pressure (EPAP) from 4 to 25 cm H2O in increments of 1. For active circuits, EPAP can be set to zero.
When AVAPS is disabled, the EPAP setting must be less than or equal to the IPAP setting. When AVAPS is enabled, the EPAP pressure must be less than or equal to the IPAP Min Pressure.
The Tidal Volume setting displays if AVAPS is enabled. You can increase or decrease the setting from 50 to 2000 ml in 5 ml increments. Use this setting to establish the target volume of gas which the ventilator will produce and deliver during each Spontaneous breath.
Chapter 5 Viewing and Changing Settings

You can adjust the rise time to find the most comfortable setting for the patient. Increase or decrease the setting from 1 to 6 until you find the right setting. The rise time levels from 1 to 6 progressively reflect slowed response of the pressure increase that will take place at the beginning of inspiration.
If the Apnea alarm is enabled, you can set the Apnea Rate from 4 to 60 BPM in 1 BPM increments. In S mode, the Apnea Rate is greater than or equal to 1:2 I:E ratio.
All of the settings described in the S Mode section are also available in S/T mode, except for the Flex setting. In addition to those settings, the settings below are also available in S/T mode.
In AC mode, you can increase or decrease the Breath Rate setting from 0 to 60 BPM, while in all other modes, you can increase or decrease the setting from 1 to 60 BPM in 1 BPM increments. Use the Breath Rate setting to establish the minimum rate of mandatory breaths that the ventilator will deliver per minute.
You can adjust the Inspiratory Time setting from 0.3 to 5.0 seconds in 0.1 second increments. Inspiratory Time is the duration for the inspiratory phase of a mandatory breath.
All of the settings available in S/T mode are available in T mode, except for the Trigger Type setting. Please refer to the descriptions in the S and S/T Mode sections of this chapter for detailed information on the T mode settings.
Spontaneous/Timed (S/T) Mode · Timed (T) Mode

Pressure Control (PC) Mode
All of the settings available in S/T mode are available in PC mode, except for the Flow Cycle Sensitivity setting (when Flow Trigger is enabled). Please refer to the descriptions in the S and S/T Mode sections of this chapter for detailed information on the PC mode settings.
Pressure Control Synchronized Intermittent Mandatory Ventilation
The following settings, described in the S and S/T mode sections of this chapter, also are available in PC-SIMV mode:
In addition to these, the following settings are also available in PC- SIMV mode.
You can increase or decrease the Pressure setting from 4 to 50 cm H2O in increments of 1. This is the pressure the ventilator will deliver during the inspiratory phase of a mandatory or assist breath.
You can increase or decrease the Pressure Support setting from 0 to 30 cm H2O in increments of 1. This is the pressure support the ventilator will deliver during the inspiratory phase of a Spontaneous breath.
Chapter 5 Viewing and Changing Settings
(PC-SIMV) Mode

The Positive End Expiratory Pressure (PEEP) setting can be increased from 0 to 25 cm H2O in active circuits and 4 to 25 cm H2O in passive circuits, in increments of 1. PEEP is the positive pressure maintained in the patient circuit during exhalation. The PEEP must be less than or equal to the pressure setting.
Control Ventilation (CV) Mode
The following settings, described in the previous sections of this chapter, are also available in CV mode:
In addition to these, the following settings are also available in CV mode.
You can choose either Ramp or Square for the Flow Pattern setting.
You can enable or disable the Sigh setting by selecting On or Off. A Sigh is a breath that is delivered every 100 breaths at 150% of the normal volume.

Assist Control (AC) Mode
The AC mode contains the following settings described in the S, S/T, PC-SIMV, and CV mode sections in this chapter. Please refer to the descriptions in those sections for detailed information.
Synchronized Intermittent Mandatory Ventilation (SIMV) Mode
The SIMV mode contains the following settings described in the S, S/T, PC-SIMV, and CV mode sections in this chapter. Please refer to the descriptions in those sections for detailed information.
Chapter 5 Viewing and Changing Settings

From the Main Menu screen, select the Options item.
The Options menu appears, shown in the screen below.

The following settings are available on the Options menu.
- cm H2O
Options Menu · - hPa · - mBar

All pressure units on the screens will display in the unit of measure selected here.
Make sure the alarm volume is set loud enough to be heard by the caregiver. Consider the use of a remote alarm.
- Off: No screen saver displays and the LCD backlight remains lit at your brightness setting.
- Breath: The display appears as a black screen, with only the patient breath indicator and manometer visible.
- Black: The display’s backlight is turned off and the display is black with no information visible.
Chapter 5 Viewing and Changing Settings

- Dim: The display’s backlight is decreased, so that the display is still visible but not as bright.
If enabled, the screen saver will display after 5 minutes of no keypad activity. Pressing any button on the device will exit the screen saver. And, any alarm or informational message will also exit the screen saver.

From the Main Menu screen, you can select Alarm Log to access the Alarm Log screen. An example is shown below.

The alarm log displays the alarms in chronological order with the most recent events displayed first. It lists the 20 most recent alarms or messages that appeared on the device display. When the device is in Limited Menu access mode, the alarm log cannot be cleared. It can be cleared when in Full Menu access mode. Depending on how many alarms have occurred, the alarm log may be several pages long. The entries in the alarm log use the same names that displayed when the alarm initially occurred and was displayed in the Alarm View.
Chapter 5 Viewing and Changing Settings
Alarm Log Screen

From the Main Menu screen, you can select Event Log to access the Event Log screen.
The event log displays a list of all events that have occurred, such as ventilator setting changes, ventilator inoperative conditions, alarms, etc., in chronological order with the most recent events displayed first. When the device is in Limited Menu access mode, the event log is not available. It can be viewed and cleared when in Full Menu access mode. You can page through the event log if it is multiple pages. The number of pages appears in the upper right corner of the panel.
In the event log descriptions, any description beginning with a 1: or 2: is a prescription change event. The 1 represents a change to a primary setting and the 2 represents a change to a secondary setting. This is followed by the setting that was changed.
The last two columns indicate setting and alarm changes. If the entry is a setting change, the first column shows the old setting value and the last column shows the new setting. If the entry is an alarm, the first column shows the value that triggered the alarm and the last column shows the number of seconds that the alarm was active.
From the Main Menu screen, you can select Information to access the Information screen. You can also view the Information screen by holding the Down key for 5 seconds. This causes the detailed view of the Monitor Screen and the Information Menu to be displayed temporarily. This key sequence is valid from the Monitor Screen while in Full or Limited Access.
The Information screen provides you with a summary of the current prescription settings, device settings, and system settings. You can use the Up/Down buttons to scroll through the information.

When you change the device prescription, alarms and other settings using the SD card, Trilogy200 requires that the caregiver review and verify the changes prior to the changes being used by the device. The caregiver or health care professional is responsible to ensure that the prescription settings are correct and compatible with the patient after using this feature. Installing the wrong prescription for a particular patient may result in improper therapy, lack of appropriate safety monitoring, and risk of death or injury to the patient.
With the Trilogy200, you can update the patient’s prescription using the SD Card. This feature lets you update a single prescription or both prescriptions if the dual prescription feature is enabled. The prescription update can occur either when the ventilator is off or on.

Ventilator Off Prescription Display Screen

Ventilator On Prescription Display Screen
Chapter 5 Viewing and Changing Settings

If the SD card is removed at any time during the prescription update, the process aborts and the screen returns to the initial state before the prescription update started.
A confirmation screen will appear if errors occur during this process. The following table summarizes the prescription errors, their possible causes, and the actions to take.

Prescription Change Failed Displayed when an error occurs during the prescription update. This is caused by an attempt to set a value for a prescription parameter that is not valid for the therapy mode or an attempt to set a prescription parameter to an invalid value.
Remove the card and have the prescription replaced with a valid prescription.
Prescription Failed – Circuit Type
Displayed when the circuit type in the new prescription doesn’t match the circuit type set for the ventilator.
If the circuit type in the prescription is correct, remove the card, change the circuit type on the ventilator, and re-insert the card. If the circuit type on the ventilator is correct, remove the card and replace the prescription on the card with a prescription containing the correct circuit type.
Prescription Failed – Card is Read Only
Displayed when the prescription is write-protected.
Remove the card and check that the small switch on the side of the SD Card is not in the Locked position. If you continue to receive this message, remove the card and have the prescription replaced with a valid prescription.
Prescription Failed – Serial Number
Displayed when the device serial number on the new prescription does not match the serial number for the device.
Remove the card and have the prescription replaced with the prescription with the correct serial number.
Prescription Failed – Version Displayed when the version of the prescription does not match the version accepted by the device.
Remove the card and have the prescription replaced with a prescription in the correct version.
Chapter 5 Viewing and Changing Settings
Message Cause Action

After you press the button and access the Monitor screen, you can view and change settings using the Menu screens.
To enter the Menu screens from the Monitor screen, press the Up button on the ventilator. The Main Menu screen appears, shown below.

You can choose from the following selections on the Main Menu screen:
Main Menu Screen

If your health care professional has set up a dual prescription for you, follow the steps below to change your prescription selection. Your health care professional might use this feature if, for instance, you need one prescription for daytime use and a second, different prescription for nighttime use.
A screen appears with a confirmation message, saying “Switch to Primary Settings” or “Switch to Secondary Settings” depending on which prescription you are changing to.
Chapter 5 Viewing and Changing Settings

To view or change the therapy settings available on the My Settings screen, use the Up/Down (Navigate) button to highlight the My Settings option on the Main Menu, and press the Right (Select) button. The My Settings Menu screen appears.
Follow the general instructions below to navigate and change any of the therapy settings. Detailed information about each setting follows.
2 To modify a setting once it is highlighted, press the Right (Modify) button.
You can change the following settings in the My Settings menu, if they are enabled by your health care professional.

Chapter 5 Viewing and Changing Settings

After you have finished adjusting your ventilator settings, perform the following steps to connect the ventilator to the patient.
Do not connect the ventilator to the patient until you perform the system checkout procedures defined in Chapter 10.
WARNING
Press the button to begin therapy. When you start therapy, the display backlight and the backlights on the buttons turn on, the red and yellow alarm LEDs turn on momentarily, and an audible indicator sounds to indicate that therapy has started. The Startup screen appears on the display.
To make sure the device is operating properly at start-up, always verify that the audible tone sounds and the alarm LEDs light red and then yellow momentarily. Contact Philips Respironics or an authorized service center for service if these indications do not occur at start-up.
After you have assembled the system, started therapy, and adjusted your ventilator settings as needed, you can connect the breathing circuit to the patient. The illustration below shows the breathing circuit connected to a mask. You can also connect the breathing circuit to a tracheostomy tube.

Example of a Fully Connected Patient Breathing Circuit with Passive Exhalation Device

This chapter describes the ventilator alarms and what you should do if an alarm occurs.
There are three types of alarms:
Additionally, the ventilator also displays informational messages and confirmation alerts that notify you of conditions that need attention but do not qualify as alarm conditions.
Chapter 6 Ventilator Alarms

When an alarm condition occurs:
Each of these is described in detail below.
The Alarm Indicator/Audio Pause button on the front of the ventilator lights up as follows whenever an alarm is detected:
Alarm Indicator/Audio Pause Button
The Alarm Indicator/Audio Pause button does not light up when informational messages or confirmation alerts display.

An audible indicator sounds whenever a power failure or a high, medium, or low priority alarm is detected. Additionally, an audible indicator sounds for informational messages and to confirm that certain actions have occurred (for example, when an SD card is inserted or removed from the device).
Chapter 6 Ventilator Alarms

When the ventilator detects an alarm, the Alarms and Messages Screen is displayed showing a description of the alarm condition. When an alarm message appears, it will be highlighted in red if it is a high priority alarm or in yellow if it is a medium or low priority alarm. (The highlight color matches the alarm LED color on the Alarm Indicator/Audio Pause button.) If an alarm is manually reset by the user, the Alarms and Messages screen is removed and the Monitoring Screen is re-displayed. If the alarm self-cancels, the Alarms and Messages screen remains displayed, but the highlight for the active alarm is removed, the LED is unlit, and the audible alarm stops. The screen below is an example of a possible alarm message.

Sample Alarms and Messages Screen
If a menu is displayed on the screen when an alarm occurs, the description of the newly generated alarm is displayed in the menu banner area. This is done so that the modification to the current parameter can be completed before addressing the alarm condition in case the modification affects the alarm condition. The screen below is an example of an alarm message displayed in the menu banner.


Alarm in Menu Banner Screen
The Alarms and Messages Screen will automatically display in place of the Monitor screen when exiting from the menu system using the Exit soft key when an alarm is displayed in the menu banner. If an alarm is manually reset by the user or self-cancels, the menu banner on-screen before the alarm occurred will reappear.
If a Ventilator Inoperative alarm occurs, the entire display screen turns red and the Ventilator Inoperative message appears on-screen, as shown below.

Ventilator Inoperative Alarm Screen
Chapter 6 Ventilator Alarms

To turn the ventilator off from a Ventilator Inoperative condition, use the normal power off sequence. When the Start/Stop button is selected, the following screen will display.

Ventilator Inoperative Power Off Screen
Select the Right button (Yes) to turn the ventilator off and stop the audible alarm. Selecting the Left button (No) will return the screen to the Ventilator Inoperative Alarm Screen without silencing the audible alarm.
When the ventilator detects an alarm condition, if you are using a remote alarm system, a signal is sent from the ventilator to activate the remote alarm. The Remote Alarm is discussed in Chapter 4.

This section describes the Audio Pause and Alarm Reset features.
When an alarm occurs, you can temporarily silence the audible indicator by pressing the Alarm Indicator/Audio Pause button. The alarm is silenced for 60 seconds and then will sound again if the cause of the alarm has not been corrected. Each time the Alarm Indicator/Audio Pause button is pressed, the alarm silence period resets to one minute.
When Audio Pause is active, the Alarm Indicator/Audio Pause symbol (
) appears if you are on the Monitor screen. Additionally, an “Audio Pause” message displays in the menu banner on the Alarm Display screen.
You can ‘pre-silence’ alarms that have not yet occurred by pressing the Alarm Indicator/Audio Pause button while no alarms are active. Then, if an alarm occurs, the audible indicator does not sound until the Audio Pause time limit has expired. This can be useful during patient setup, so caregivers can prevent alarms from sounding that they know will occur during the setup process. The LED and display will still show the alarm, but the audible alarm will not sound.
The Reset button is used to clear the currently active alarm(s) from the display and stop the LED and audible alarm indicator. This button should be selected after the situation causing the alarm(s) has been corrected. All active alarms are cancelled and alarm detection is restarted when this button is selected.
The ventilator self-cancels certain alarms if the cause of the alarm is corrected, shutting off the alarm LED, the audible alarm, and the alarm background color. However, the alarm text remains on the screen. You can manually reset an alarm by pressing the Left button (Reset). The Audio Pause function is cancelled if the alarm is manually reset.
Chapter 6 Ventilator Alarms

WARNING
As explained in Chapter 5, you can adjust the Alarm Volume from the Options menu. You can select Loud or Soft, depending on your preference.
Make sure the alarm volume is set loud enough to be heard by the caregiver. Consider the use of a remote alarm.
You should not rely on any single alarm to detect a circuit disconnect condition. The Low Tidal Volume, Low Minute Ventilation, Low Respiratory Rate, and Apnea alarms should be used in conjunction with the Circuit Disconnect alarm.
Complete the following steps when an alarm occurs:
), or an “Audio Pause” message appears in the menu banner on the Alarm Display screen.

The following table summarizes all of the high, medium, and low priority alarms and informational messages.
Alarm Priority Audible Visual (Alarm Indicator Button
Shuts down User: Press Start/Stop button and then press Right button to silence alarm. If using AC power, try plugging device into alternate AC power source. If loss of power continues, switch to DC power by connecting a fully charged detachable or external battery to the device. If there is still no power, connect patient to alternate source of ventilation and contact your health care professional.
Clinician: Restore AC power. If alarm continues, connect a fully charged detachable or external battery to device to restore power. If alarm still continues, have the device serviced.
High Red flashing button; “Ventilator Inoperative” message
Shuts down if can’t provide therapy safely. Or, continues to operate at a limited level.
User: Press Start/Stop button. If display is operational, Power Off confirmation screen appears. Select Right button to shut off device and silence alarm. Immediately remove patient from ventilator and connect them to alternate source of ventilation. Contact your health care professional for service.
Clinician: Restore power. If alarm continues, have the device serviced.
Chapter 6 Ventilator Alarms

Alarm Priority Audible Visual (Alarm Indicator Button
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. The alarm goes away until next reminder appears. Connect patient to alternate source of ventilation and contact your health care professional as soon as possible.
Clinician: If alarm continues, have the device serviced.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Check patient circuit to make sure it is set up as your health care professional specified. Make sure there is no kinked or pinched tubing. Make sure tubing is properly attached. Make sure all tubing and exhalation devices are clean and free of condensation (refer to the Troubleshooting section – tubing condensation). If using Active PAP or Active Flow Exhalation Device, make sure diaphragm is inserted correctly and is not wrinkled or kinked. If alarm continues, contact your health care professional.
Clinician: Verify patient status. Verify that breathing circuit setup is correct. Correct any errors. If alarm continues, have device serviced.

Alarm Priority Audible Visual (Alarm Indicator Button
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Check for occlusions in exhalation devices. Make sure exhalation device is clean and functioning properly. If alarm continues, contact your health care professional.
Clinician: Remove obstruction in leak device. If alarm continues, have the device serviced.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Make sure tubing is not kinked or pinched. Check patient’s breath rate. If alarm continues, contact your health care professional.
Clinician: Check patient’s tubing to make sure it is not kinked or pinched. If alarm continues, have device serviced.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Make sure tubing is not kinked or pinched. If alarm continues, contact your health care professional.
Clinician: Check patient’s tubing to make sure it is not kinked or pinched. If alarm continues, have device serviced.
Chapter 6 Ventilator Alarms

Alarm Priority Audible Visual (Alarm Indicator Button
Continues to operate when internal oxygen concentration reaches 5% above ambient levels.
User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Disconnect supplemental oxygen supply from the ventilator and contact your health care professional.
Clinician: Disconnect supplemental oxygen from device. Check external oxygen connection. If problem continues, have device serviced.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Check breathing circuit and reconnect it if it has become disconnected, or fix the leak. If alarm continues to occur, contact your health care professional. If device will not exit circuit disconnect, switch to an alternate source of ventilation.
Clinician: Reconnect tubing or fix leak. If alarm continues, have device serviced.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If using an Active PAP or Active Flow circuit, check proximal pressure line to ensure it is not pinched or has water condensation (refer to the Troubleshooting section – tubing condensation). If alarm continues, contact your health care professional.
Clinician: Verify patient status.

Alarm Priority Audible Visual (Alarm Indicator Button
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If alarm continues, contact your health care professional.
Clinician: Verify patient status.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If alarm continues, contact your health care professional.
Clinician: Verify patient status.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If alarm continues, contact your health care professional.
Clinician: Verify patient status.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If alarm continues, contact your health care professional.
Clinician: Verify patient status.
Chapter 6 Ventilator Alarms
Device Action · User/Clinician Action · and Display)

Alarm Priority Audible Visual (Alarm Indicator Button
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If alarm continues, contact your health care professional.
Clinician: Verify patient status.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If alarm continues, contact your health care professional.
Clinician: Verify patient status.
Escalates from Audible indicator to Medium priority and then High priority
Peak Pressure symbol turns red. When condition first occurs, a beep will sound. When condition occurs for third time, button flashes yellow and yellow “High Inspiratory Pressure” message appears. When condition occurs for 10th time, button flashes red and red “High Inspiratory Pressure” message appears.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Make sure tubing is not pinched or blocked. If alarm continues, contact your health care professional.
Clinician: Verify patient status. If problem continues, have device serviced.

Alarm Priority Audible Visual (Alarm Indicator Button
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Check patient circuit for leaks or disconnects. If using an Active PAP or Active Flow circuit, check proximal lines to ensure they are not pinched or have water condensation (refer to the Troubleshooting section – tubing condensation). If alarm continues, contact your health care professional.
Clinician: Verify patient status. If problem continues, have device serviced.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If alarm continues, contact your health care professional.
Clinician: Verify patient status.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Continue using device. If alarm continues, contact your health care professional.
Clinician: Verify patient status.
Chapter 6 Ventilator Alarms

Alarm Priority Audible Visual (Alarm Indicator Button
Low Battery Escalates from Medium to High
Medium Priority- Yellow flashing button. “Low Detachable Battery,“ “Low External Battery,” or “Low Internal Battery” message appears in yellow, On Status Panel, box around battery is yellow
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left button (Reset) to reset alarm. Switch to an alternate battery. Or, switch to AC power source and recharge low battery.
Clinician: Switch to alternate battery or AC power while you recharge low battery. If low battery is recharged and alarm continues, replace battery.
High Priority-Red flashing button. “Low Detachable Battery”, “Low External Battery,” or “Low Internal Battery” message appears in red. On Status panel, box around battery is red
Escalates from Medium to High
Yellow flashing button and yellow “High Temperature” message appears. If condition worsens, button flashes red and red “High Temperature” message appears.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Make sure device is not close to a heat source. Make sure tubing is not under any bedding. Check inlet filter and clean or replace it if needed. Make sure cooling vents are not blocked. If the ventilator is running off of the internal or detachable battery, move to a cooler location and/or power the device with AC power or a lead-acid battery.
If alarm continues, place patient on alternate source of ventilation and contact your health care professional.
Clinician: If alarm continues after the measures above have been taken, have device serviced.

Alarm Priority Audible Visual (Alarm Indicator Button
Low or High, depending on cause of alarm
“Replace Detachable Battery” message appears. If battery is nearing end of useful life, message appears with yellow background and button is solid yellow. If battery fails, message appears with red background and button flashes red.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left button (Reset) to reset alarm. Switch to an alternate battery or AC power source while replacing current detachable battery.
Clinician: Switch to an alternate battery or AC power source while replacing the current detachable battery.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Alarm goes away until next reminder appears. Contact your health care professional to report alarm and have device serviced.
Clinician: If alarm continues, have device serviced.
Switches to alternate power source
User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. If you were using AC power, check power cord and reconnect it if it has become disconnected. Make sure device is not connected to an overloaded AC circuit. If you were using an external, detachable, or internal battery, recharge the battery.
Clinician: Check batteries and recharge f necessary. Restore AC power if available.
Chapter 6 Ventilator Alarms

Alarm Priority Audible Visual (Alarm Indicator Button
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Check the keys to determine if they are lodged in the case. If alarm continues, place patient on alternate source of ventilation and contact your health care professional.
Clinician: Check keys to determine if they are lodged in the case. If alarm continues, place patient on alternate source of ventilation and have device serviced.
Battery Discharging Stopped Due to Temperature
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Move device to a cooler location. Make sure device is not close to a heat source. Make sure cooling vents are not blocked. If alarm is detected for internal battery and continues, place patient on alternate source of ventilation and contact your health care professional.
Clinician: If alarm is detected on the internal battery and continues after the measures above have been taken, place patient on alternate source of ventilation and have device serviced.
Device Action · User/Clinician Action · and Display)

Alarm Priority Audible Visual (Alarm Indicator Button
Battery Not Charging Due to Temperature
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Make sure device is not close to a heat source. Make sure cooling vents are not blocked. Move device to a cooler location. If device is too cold, allow it to warm up. If alarm continues, place patient on alternate source of ventilation and contact your health care professional.
Clinician: If alarm continues after the measures above have been taken, place patient on alternate source of ventilation and have device serviced
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. If condition continues for internal battery, place patient on alternate source of ventilation and contact your health care professional.
Clinician: If condition continues for internal battery, place patient on alternate source of ventilation and have device serviced.
Chapter 6 Ventilator Alarms

Alarm Priority Audible Visual (Alarm Indicator Button
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Check connection to external battery. Replace external battery with another external battery, if available. If alarm continues, place patient on alternate source of ventilation and contact your health care professional.
Clinician: If the alarm continues after the measures above have been taken, place patient on alternate source of ventilation and have device serviced.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left (Reset) button to reset alarm. Replace depleted battery with another or switch to AC power, if available.
Clinician: Replace depleted battery with another or switch to AC power.
Info • “External Batt Disconnected” message, and a box appears around battery in use.
Switches to alternate power source
User: Check connection of the external battery to ventilator. Check the charge available on external battery and recharge battery if necessary.
Clinician: Check external battery and recharge if necessary. Restore AC power if available.

Alarm Priority Audible Visual (Alarm Indicator Button
Info • “Detachable Batt Disconnected” message, and a box appears around battery in use.
Switches to alternate power source
User: Check connection of the detachable battery to ventilator. Check the charge available on detachable battery and recharge battery if necessary.
Clinician: Check detachable battery and recharge if necessary. Restore AC power if available.
Start On Battery Info • “Start On Battery” message appears.
Operates User: Check battery status and determine how long device can operate on battery power. Find AC power source as soon as possible.
Clinician: Restore AC power when available. If alarm continues, replace AC power cord to determine if there is a problem with the cord.
Card Error Info • A “Card Error” message appears.
Operates User: Press Alarm Indicator/ Audio Pause button to silence alarm. Or, press Left button (Reset) to reset alarm. Remove SD Card and use another card, if available. Check the write- protect switch on the card. If condition persists, contact your health care professional.
Clinician: Remove SD Card and use another card, if available. Ensure card meets specifications. Check write- protect switch on card. If condition persists, contact an authorized representative of Philips Respironics.
Chapter 6 Ventilator Alarms



The ventilator’s exterior surface and the exterior of the detachable battery pack (if using) should be cleaned before and after each patient use, and more often if needed.
To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the ventilator.
Do not immerse the device or allow any liquid to enter the enclosure or the inlet filter.
– Soapy water or a mild detergent
– Hydrogen Peroxide (3%)
– Isopropyl Alcohol (91%)
Do not use harsh detergents, abrasive cleaners, or brushes to clean the ventilator system. Use only the cleaning agents and methods described in this manual.
– 10% bleach solution (10% bleach, 90% water)
Chapter 7 Cleaning and Maintenance
– Water

Under normal usage, you should clean the gray foam filter at least once every two weeks and replace it with a new filter every six months or sooner if needed.
The reusable foam inlet filter is required to protect the ventilator from dirt and dust. Wash periodically and replace when damaged for proper operation.

Removing the Air Inlet Filter

Reinstalling the Filter

Follow your institution’s protocol for how often you replace the air inlet path foam. To replace the foam, complete the steps below.

Removing the Air Inlet Path Cover

Removing the Air Inlet Path Foam
Chapter 7 Cleaning and Maintenance

Cleaning the reusable circuit is important in the hospital and in the home. Circuits infected with bacteria may infect the user’s lungs. Clean the respiratory circuit on a regular basis. If you are using a disposable circuit, dispose of and replace it on a regular basis.
Follow your institution’s protocol for cleaning the circuit. Philips Respironics recommends that you perform the cleaning twice a week under normal conditions and more frequently as required. If the patient uses the device on a 24 hour a day basis, it may be convenient to have a second breathing circuit so you can switch circuits while one is being cleaned.
Exhalation devices, patient circuits, and water traps are shipped clean, not sterile. Cleaning and disinfection of these parts should follow individual institution processes and conform to guidelines provided by Philips Respironics with each accessory.
Clean the patient circuit twice a week, or follow your institution’s protocol.

Cleaning the Exhalation Devices
Disassemble the exhalation device from the patient circuit. Follow the detailed cleaning instructions included with your exhalation device.
Replacing the Bacteria Filter (Optional)
To remove a bacteria filter and replace it with a new one:
Philips Respironics recommends that a main line outlet bacteria filter (Part Number 342077) be used whenever the device is used for invasive therapy or if the ventilator may be used on multiple patients.
Chapter 7 Cleaning and Maintenance


Installing a New Bacteria Filter
Your new Philips Respironics Trilogy ventilator includes a blank PM Service Due label affixed to the bottom of your device. The label is blank, which allows you to document when service is due. Service is due every 10,000 hours or 24 months, whichever comes first depending on the usage of the device.

You will need to record both the date and hours on the blank label to maximize your in-service interval. Record the service date as 24 months from the date first placed on a patient. Record the hours as 10,000 hours for initial usage.
Use the device blower hours to determine when service is required. The device blower hours are listed in the device software in the information menu.
Service Date · Preventive Maintenance Sticker · Service Hours

The following table provides information on recommended periodic maintenance for your device.
Time / Hours of Service Recommended Maintenance
Prior to long term storage after initial use
While in storage after initial connection to AC Power, every 3 months
Every 10,000 hours or 2 Years, whichever comes first
Refer to the Trilogy Service Manual for additional information.
Chapter 7 Cleaning and Maintenance



The ventilator is designed to respond to most issues with an appropriate alarm or information message. These messages appear on the display screen. Chapter 6 describes many of the alarms and what action you should take when the alarms occur.
This chapter provides additional troubleshooting information for other common issues you may have.
Issue Why It Happened What To Do
Device does not turn on. Nothing happens when you press Start/Stop to begin therapy. The audible indicator does not sound and the backlights on the buttons do not light.
The AC power cord is not plugged in and the internal battery is not charged. Note: The device requires AC power to charge the internal battery when a new device is powered up for the first time.
Plug the AC power cord into the device and a wall outlet that is not controlled by a switch. This powers the system and charges the internal battery. If the device still does not turn on, contact an authorized service representative or Philips Respironics to have the device serviced. Please have the model number and serial number ready when you call.
Chapter 8 Troubleshooting

Issue Why It Happened What To Do
The batteries are not charging. The device is plugged in, but the detachable (if installed) and internal batteries are not showing full capacity and are not charging.
The batteries may not charge if the device is too hot or too cold or is operating at an ambient temperature outside of the specified valid range. The device cooling fan may not be working properly. If this continues for more than 15 minutes, an error will occur.
Make sure the device is not close to a heat source. Ensure the cooling air vents are not blocked. Bring the ventilator to ambient room temperature. If the problem continues, contact an authorized service representative or Philips Respironics to have the device serviced. Please have the model number and serial number ready when you call.
Unexplained changes in the performance of the device The device is not working properly or is making unusual sounds.
The device has been dropped or mishandled, or is in an area of high EMI emissions.
Press the Start/Stop button and then select the Right button to shut off the device for a few seconds. Restart the device by pressing the Start/Stop button. If the problem continues, relocate the device to an area with lower EMI emissions (e.g., away from electronic equipment such as cellular phones, cordless phones, computers, TVs, electronic games, hair dryers, etc.). Refer to Chapter 13. If the problem still occurs, contact an authorized service representative or Philips Respironics to have the device serviced. Please have the model number and serial number ready when you call.

Issue Why It Happened What To Do
Excessive Warm Air Out of Mask The air out of the mask is much warmer than usual.
The air filter may be dirty. Or, the device may be operating in direct sunlight or near a heater.
Clean or replace the air filter as described in Chapter 7. The temperature of the air may vary based on the room temperature. Make sure the device is properly ventilated. Keep it away from bedding or curtains that could block the flow of air around the device. Make sure it is away from direct sunlight and heating equipment. Ensure the cooling air vents are not blocked. If the problem continues, contact an authorized service representative or Philips Respironics to have the device serviced. Please have the model number and serial number ready when you call.
Mask Discomfort The mask feels uncomfortable to wear, there is significant air leakage around the mask, or the patient experiences other mask-related issues.
This could be due to improper headgear adjustment or improper mask fitting.
Make sure the patient is properly fitted with the correct size mask. If the problem continues, try fitting the patient with a different mask.
Runny Nose The patient has a runny nose.
This may be caused by a nasal reaction to the airflow. This only applies to non-invasive therapy and will not occur during invasive therapy.
Call the patient’s health care professional.
Chapter 8 Troubleshooting

Issue Why It Happened What To Do
Dryness The patient has throat or nose dryness.
The air is too dry. This only applies to non-invasive therapy and will not occur during invasive therapy.
Increase the room humidity. Consult with the patient’s health care professional about using a humidifier with the device. If using a humidifier, refer to the instructions included with the humidifier to make sure it is working properly.
Sinus or Ear Pain The patient experiences nasal, sinus, or ear pain.
The patient may have a sinus or middle ear infection. This only applies to non-invasive therapy and will not occur during invasive therapy.
Tell the patient to stop using the device and contact their health care professional.
Patient Tubing Condensation There is condensation in the circuit tubing or exhalation valve lines.
Moisture added to the circuit from a heated humidifier may condense. This condensation is typically referred to as “rain-out.”
Use an appropriate method of water management, such as a water trap when using a heated humidifier.
Exhalation Valve or Flow Tube Condensation (Active Flow Circuit only) There is condensation in the proximal flow or exhalation valve lines, and the condensation is causing the system to alarm or not properly trigger on patient effort.
Excessive humidity (from the heated humidifier, the patient, or the cleaning process) condensed in the tubing or exhalation valve diaphragm. Droplets of water affected the ability of the ventilator to measure proximal patient flow and pressure and trigger on a patient breath.
Remove tube(s) with condensation: the exhalation valve line must be removed from both ends of the circuit; the flow tube lines must be removed from the ventilator side. The tubes are permanently attached to the flow element, so the flow element must be removed from the patient circuit. If there is water built up in the exhalation diaphragm, disassemble the exhalation valve and clean/dry the assembly. Clear the flow tubes of water droplets using a low flow air source and re- assemble the circuit.

There are several accessories you can use with the ventilator.
Use of a humidifier may reduce nasal dryness and irritation by adding moisture to the airflow.
If you are using a humidifier with the ventilator, refer to the instructions provided with the humidifier for complete setup information.
The delivered oxygen concentration varies with changes in flow in the circuit. The following may have an impact on oxygen concentration:
To add oxygen to the circuit, the oxygen supply must comply with the local regulations for medical oxygen. The oxygen flow into the oxygen valve cannot exceed 15 l/min and the pressure cannot exceed 50 psi.
Chapter 9 Accessories

The remote alarm should be tested daily. When using a remote alarm, make sure you fully test the remote alarm connector and cable by verifying that:
- Annunciated alarms on
the ventilator are also annunciated on the remote alarm.
You can use a Philips Respironics Remote Alarm unit with your device. There is a dedicated adapter cable assembly for connecting the device to the Remote Alarm unit (REF 1045291). Refer to the instructions included with your Remote Alarm unit and adapter cable assembly if you have questions on how to use it with device.
- Disconnecting the remote
alarm cable from the ventilator or from the remote alarm results in an alarm notification at the remote alarm.

Ensure that the Nurse Call systems used do not exceed SELV (Safety Extra Low Voltage) levels as described in IEC 60601-1. SELV levels do not exceed a nominal value of 25VAC or 60VDC at rated supply voltage on the transformer or converter, between conductors in an earth- free circuit which is isolated from the supply mains by a safety extra low voltage transformer or by a device with equivalent separation.
You can use an institutional Nurse Call system with your device. There are several Philips Respironics cables available to connect a nurse call system to the device. Refer to the instructions included with your cable assembly if you have questions on how to connect it to the device.
The Trilogy’s Nurse Call feature is for use only in a medically supervised environment.
Using a Secure Digital (SD) Card
Do not rely on the audible indicator provided by the Nurse Call system as the primary indicator of the operating state of the device or of patient events.
An SD card is provided with the device to record device usage information. You can insert the SD card into the SD Data Card slot on the left side of the device. Make sure the label on the SD card faces the back of the ventilator.
To insert the SD card:
The Nurse Call feature should be considered a back-up to the Trilogy device’s primary alarm system. Do not rely solely on the Nurse Call feature.
Opening the SD Card Door
Chapter 9 Accessories

Inserting the SD Card
To remove the SD card, select the “Safely Remove SD Card” option from the main menu. After the “Remove SD Card” confirmation message appears, remove the card by pushing the card in until the card clicks to release. The card will be slightly ejected from the device.
You can use the Philips Respironics DirectView software to download the prescription data from the SD card to a computer. DirectView can be used by clinicians to receive and report stored data from the SD card. DirectView does not perform any automatic scoring or diagnosing of a patient’s therapy data.
An optional, in-use bag is included for use with the ventilator. The in-use bag is required for portable use in order to properly secure the ventilator. See the instructions included with the in-use bag for details on how to use the bag with your device.

When traveling, the carrying case is for carry-on luggage only. The carrying case will not protect the system if it is put through checked baggage.
For your convenience at security stations, there is a note on the bottom of the device stating that it is medical equipment and is suitable for airline use. It may be helpful to bring this manual along with you to help security personnel understand the device.
If you are traveling to a country with a line voltage different than the one you are currently using, a different power cord or an international plug adaptor may be required to make your power cord compatible with the power outlets of the country to which you are traveling.
The device is suitable for use on airlines when the device is operating from an internal, detachable, or external battery power source. It is not suitable for airline use when operating from AC power or from DC power sourced from the airplane.
Chapter 9 Accessories



This chapter details the test procedures that should be performed by the clinician prior to connecting the device to the patient. Test the Active PAP, Active Flow, and Passive circuit types if you want to do a complete checkout on the device. The tests should be performed as described in order to verify proper operation of the device. Some of the procedures in this chapter require you to change settings on the device. If you are not familiar with the procedures for accomplishing this, please refer to Chapter 4, Ventilator Setup and Chapter 5, Viewing and Changing Settings.
If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if the device or detachable battery are dropped, if water is spilled into the enclosure, or if the enclosure is cracked or broken, discontinue use and contact Philips Respironics or an authorized service center for service.
Chapter 10 System Checkout Procedures

Complete the following steps to set up the settings and alarms tests.
A. Settings And Alarms Menu
Modify the settings in the Settings and Alarms menu to match those shown below in Table 1. If necessary, refer to Chapter 5 for instructions on modifying ventilator settings.
Circuit Type Active PAP, Active Flow or Passive
AVAPS (passive circuit only) Off
IPAP 20 cm H2O
EPAP 4 cm H2O
Table 1 - Ventilator Settings in the Settings and Alarms Menu
Breath Rate 12 BPM
Inspiratory Time 1.6 seconds
Setting Value · Dual Prescription Off · Therapy Mode S/T

Trigger Type (passive circuit type)
Flow Trigger Sensitivity (Active PAP or Active Flow circuit type)
6.0 l/min
Leak Compensation (Active Flow circuit type)
Flow Cycle Sensitivity (Active PAP or Active Flow circuit type)
All other alarms Off
B. Options Menu
Modify the settings in the Options menu to match those shown below in Table 2.
Table 2 - Ventilator Settings in the Options Menu
All other settings Discretionary
C. Turn Device Power On
Press the Start/Stop button on the front of the ventilator. The system will begin operating using the defined ventilation settings.
Chapter 10 System Checkout Procedures

This procedure verifies that the High Tidal Volume alarm is working properly. For Passive and Active Flow circuits, this will verify the High Vte alarm. For Active PAP circuits, this will verify the High Vti alarm. It assumes that you have attached the test lung, verified the ventilator settings, and turned on ventilator power as described in the Initial Setup section.
A. Change Alarm Ventilator Setting
Modify the High Tidal Volume alarm setting to match the one shown below in Table 3.
High Vte/High Vti 50 ml
Table 3 - High Vte/Vti Alarm Setting
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
C. Modify Ventilator Alarm Settings
Modify the High Tidal Volume alarm setting to match the one shown below in Table 4.
Table 4 - Modify Vte/High Vti Alarm Setting
High Vte/High Vti 500 ml
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
Setting Value · Setting Value

E. Restore Ventilator Settings
Modify the ventilator settings and change the following value shown in Table 5.
High Vte/High Vti Off
Table 5 - Restore Ventilator Settings
This procedure verifies that the Low Tidal Volume alarm is working properly. For Passive and Active Flow circuits, this will verify the Low Vte alarm. For Active PAP circuits, this will verify the Low Vti alarm. It assumes that you have attached the test lung, verified the ventilator settings, and turned on ventilator power as described in the Initial Setup section.
A. Change Alarm Ventilator Setting
Modify the Low Tidal Volume alarm setting to match the one
shown below in Table 6.
Low Vte/Low Vti 500 ml
Table 6 - Low Vte/Low Vti Alarm Setting
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
C. Modify Ventilator Alarm Settings
Modify the Low Tidal Volume alarm setting to match the one shown below in Table 7.
Table 7 - Modify Low Vte/Low Vti Alarm Setting
Low Vte/Low Vti 50 ml
Chapter 10 System Checkout Procedures
Setting Value · Setting Value · Setting Value

D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
E. Restore Ventilator Settings
Modify the ventilator settings and change the following value shown in Table 8.
Table 8 - Restore Ventilator Settings
Low Vte/Low Vti Off
This procedure verifies that the Circuit Disconnect alarm is working properly. It assumes that you have attached the test lung, verified ventilator settings, and turned on ventilator power as described in the Initial Setup section.
A. Change Circuit Disconnect Ventilator Setting
Modify the Circuit Disconnect ventilator setting to match the
value shown below in Table 9.
Circuit Disconnect 10 seconds
Table 9 - Ventilator Settings
B. Disconnect Test Lung
Disconnect the test lung from the circuit.
C. Verify the Alarm
Wait approximately 10 seconds and verify the following alarm signals:
Setting Value · Setting Value

D. Reconnect Test Lung
Reconnect the test lung to the circuit.
E. Verify Reset
Wait at least 40 seconds and verify the following auto-reset conditions:
F. Restore Ventilator Settings
Modify the ventilator settings and change the following values shown below in Table 10.
Table 10 - Restore Ventilator Settings
This procedure verifies that the High Inspiratory Pressure alarm is working properly. It assumes that you have attached the test lung, verified ventilator settings, and turned on ventilator power as described in the Initial Setup section.
A. Change Ventilator Settings
Modify the ventilator settings and change the following values shown below in Table 11.
Table 11 - Ventilator Settings
Tidal Volume 500 ml
Breath Rate 12 BPM
Inspiratory Time 1.0 seconds
Chapter 10 System Checkout Procedures

PEEP 4 cm H2O
Low Inspiratory Pressure 6 cm H2O
High Inspiratory Pressure 10 cm H2O
All other alarms Off
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
appears on the screen, highlighted in yellow
C. Modify Ventilator Alarm Settings
Modify the High Inspiratory Pressure setting to match the one
shown below in Table 12.
High Inspiratory Pressure 60 cm H2O
Table 12 - Modify High Inspiratory Pressure Alarm Setting
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
This procedure verifies that the Low Inspiratory Pressure alarm is working properly. It assumes that you have attached the test lung, verified ventilator settings, and turned on ventilator power as described in the Initial Setup section.

A. Change Ventilator Settings
Modify the ventilator settings and change the following values shown below in Table 13.
Tidal Volume 500 ml
Breath Rate 12 BPM
Inspiratory Time 1.0 seconds
PEEP 4 cm H2O
Low Inspiratory Pressure 40 cm H2O
High Inspiratory Pressure 60 cm H2O
Table 13 - Ventilator Settings
All other alarms Off
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
C. Modify Ventilator Alarm Settings
Modify the Low Inspiratory Pressure setting to match the one shown below in Table 14.
Table 14 - Modify Low Inspiratory Pressure Alarm Setting
Low Inspiratory Pressure 6 cm H2O
Chapter 10 System Checkout Procedures

D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
Make sure the batteries are functioning properly and fully charged before patient use.
Batteries Function
A. Connect AC Power to the device and verify that the green AC
LED on the front panel is lit.
B. Verify that the detachable battery is properly installed.
C. Turn the device on and verify that both the detachable and
internal battery symbols appear on the display. Verify that if either battery is less than fully charged, the charge symbol will display on the respective battery.
D. Disconnect the AC Power source from the device.

E. Disconnect the detachable battery pack from the device.
F. Reconnect the Detachable Battery and AC Power source.
A. Connect AC Power to the device and verify that the green AC
LED is lit.
B. Connect the external battery cable to the external battery and to
the ventilator.
C. Verify that the external battery symbol is shown on the display
and some level of charge is present.
D. Disconnect the AC Power source from the device.
E. Reconnect the AC Power source.
Chapter 10 System Checkout Procedures

A. Press Clear to clear the log file.
B. Press Yes to confirm.
C. Press Finish to complete.
A. Press Clear to clear the log file.
B. Press Yes to confirm.
C. Press Finish to complete.
All portions of this checkout procedure should be completed prior to connection to the patient. If any of the tests fail to complete as indicated, if possible, correct the error, clear the alarm and resume testing. If correction of the failed portion is not possible, return the device to Philips Respironics or an authorized service center for service and repair.

Temperature 41° F to 104° F (5° C to 40° C) -4° F to 140° F (-20° C to 60° C)
Relative Humidity 15 to 95% (non-condensing) 15 to 95% (non-condensing)
Atmospheric Pressure 110 kPa to 60 kPa N/A
The operating range for the charging of the internal and detachable batteries is 10° C to 30° C. The internal and detachable batteries will power the ventilator for the full operating range of 5° C to 40° C.
Accuracies stated in this manual are based on specific environmental conditions. For stated accuracy, the environmental conditions are: Temperature: 20-30° C; Humidity: 50% relative; Altitude: nominally 380 meters.
Dimensions: 6.6” L x 11.2” W x 9.3” H (16.68 cm L x 28.45 cm W x 23.52 cm H)
This device is designed to conform to the following standards:
Chapter 11 Technical Specifications
Operating Storage

Type of Protection Against Electric Shock: Class II/Internally Powered Equipment
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Protection against Ingress of Water: Device: Drip Proof, IPX1
Fuses: There are no user-replaceable fuses.
Use only SD cards and SD card readers available from Philips Respironics or the following:
SanDisk® Standard SD Card - 1.0 GB - REF SDSDJ-1024
SanDisk ® Card Reader/Writer - SanDisk ImageMate - REF SDDR-99-A15

IPAP 4 to 501 cm H2O2 Greater of 2 cm H2O or 8% of setting
EPAP 0 to 25 cm H2O for Active Circuits 4 to 25 cm H2O for Passive Circuits Greater of 2 cm H2O or 8% of setting
CPAP 4 to 20 cm H2O Greater of 2 cm H2O or 8% of setting
PEEP 0 to 25 cm H2O for Active Circuits 4 to 25 cm H2O for Passive Circuits Greater of 2 cm H2O or 8% of setting4
Pressure 4 to 50 cm H2O Greater of 2 cm H2O or 8% of setting
Pressure Support 0 to 30 cm H2O3 Greater of 2 cm H2O or 8% of setting4
Tidal Volume 50 to 2000 ml5 Greater of 15ml or 10% of setting (Active Circuits) Greater of 15ml or 15% of setting (Passive Circuits)
Breath Rate 0 to 60 BPM for AC mode 1 to 60 BPM for all other modes Greater of ± 1 BPM or ±10% of the setting
Timed Inspiration 0.3 to 5.0 seconds ± .1 second
Rise Time 1 to 66 ±2 cm H2O (the device will increase to a pressure of .67* (IPAP - EPAP) ±2 cm H2O @ the set rise time multiplied by 100 ms for pressure supports less than or equal to 25.)
Ramp Start Pressure 0 to 25 cm H2O for Active Circuits 4 to 25 cm H2O for Passive Circuits 4 to 19 cm H2O in CPAP mode
8% of setting + 2% Full Scale
Ramp Length Off, 5 to 45 minutes ±2 minutes
Flex Off, 1 to 37 N/A
Flow Trigger Sensitivity 1 to 9 l/min N/A
Flow Cycle 10 to 90% N/A
Apnea Rate 4 to 60 BPM Greater of 1 BPM or 10% of setting
Specifications listed are based on using a standard circuit consisting of 1.8 meter tubing- REF 622038, one of the following exhalation devices: Whisper Swivel II - REF 332113; Active PAP Exhalation Device - REF 1053716; Active Flow Exhalation Device - REF 1049503 or Active Exhalation Device – REF 1065658; UPB Active PAP Exhalation Device - REF 1073862 and Trilogy Proximal Flow Sensor- REF 1050408 (if required).
1Limited to 25 cm H2O when using the Bi-Flex feature in S mode.
2Pressure units may be cm H2O, hPa, or mBar depending on device setup.
3The difference between the Inspiratory Pressure and the Expiratory Pressure must never be more than 30 cm H2O.
4Pressure Support and PEEP not to exceed 50 cm H2O.
5All flows and volumes are measured at BTPS conditions.
6The range of values correspond to tenths of seconds (e.g., a setting of 4 indicates a Rise Time of 0.4 seconds).
7Flex is not available when AVAPS is active. Flex is not available with Active Circuits.
Chapter 11 Technical Specifications
Parameter Range Accuracy


Vte/Vti 0 to 2000 ml Greater of 15ml or 15% of reading
Minute Ventilation 0 to 99 l/min Calculation based on measured Vte or Vti and Respiratory Rate
Respiratory Rate 0 to 80 BPM Greater of 1 BPM or 10% of reading
Peak Inspiratory Flow 0 to 200 l/min 3 l/min plus 15% of reading
Peak Inspiratory Pressure 0 to 99 cm H2O Greater than 2 cm H2O or 10% of reading
Mean Airway Pressure 0 to 99 cm H2O Greater than 2 cm H2O or 10% of reading
I:E Ratio 9.9- 1: 1-9.9 Calculation based on Inspiratory Time and Expiratory Time
All flows and volumes are measured at BTPS conditions.

Inspiratory Resistance (cm H2O) Expiratory Resistance (cm H2O)
30 < 3.0 < 3.5 < 2.0 < 2.0 < 2.0 < 1.5
60 < 9.0 < 10.5 < 5.0 < 3.0 < 4.5 < 4.0
Dispose of this device and batteries in accordance with local regulations.

The following terms and acronyms appear throughout this manual.
AC Power Alternating Current power
AC Assist Control therapy mode; AC mode delivers Assisted and Mandatory breaths with a user-defined inspired tidal volume. This is a Volume Control mode.
Apnea A condition marked by the temporary cessation of spontaneous breathing.
Assisted Breath Breath type in which the breath is patient-triggered and the ventilator controls how the air is delivered based on defined settings. An assisted breath is not cycled until the inspiratory time setting has been reached.
AVAPS Average Volume Assured Pressure Support feature (available in S, S/T, T, and PC therapy modes).
Blower Hours The total number of hours that the blower has been on over the life of the device. This value helps determine when the ventilator needs to be serviced. You cannot reset this value. It can only be reset by an authorized service center.
BPM Breaths Per Minute
Term/Acronym Definition · Chapter 12 Glossary

BTPS Body Temperature and Pressure Saturated; A standardization for lung volumes and flows to barometric pressure at sea level, body temperature, and saturated with water vapor reflecting the condition of air in the lung.
CPAP Continuous Positive Airway Pressure
CV Control Ventilation therapy mode; CV mode delivers Mandatory breaths with a user-defined inspired tidal volume. This is a Volume Control mode.
EPAP Expiratory Positive Airway Pressure
Flex The device provides the Flex comfort feature. The C-Flex™ feature provides increased pressure relief during the expiratory phase of breathing. The Bi-Flex® feature provides a level of pressure relief taking place at the end of inhalation and at the start of exhalation.
I:E Ratio The ratio of inspiratory time to expiratory time.
IPAP Inspiratory Positive Airway Pressure
LED Light Emitting Diode
l/min Liters Per Minute
Term/Acronym Definition · DC Direct Current · ESD Electrostatic Discharge

Mandatory Breath A Mandatory Breath is completely controlled by the ventilator. Breaths are initiated by the ventilator according to the breath Rate (BPM) setting. Breaths are cycled according to the Inspiratory time setting.
MAP Mean Airway Pressure. This is the average airway pressure over a full breath cycle. The MAP value displayed on the screen represents the average MAP value over six breaths and is updated at the end of each exhalation.
Minute Ventilation The volume of gas that moves in and out of the lungs in one minute. It is calculated by multiplying the tidal volume by the respiratory rate. The Minute Ventilation value displayed on the screen represents the average Minute Ventilation over six breaths and is updated at the end of each exhalation.
Operational Hours This is the total number of hours that the blower has been on since this value was last reset. You can reset this value each time you give the device to a new patient to help track their device usage.
OSA Obstructive Sleep Apnea
PC Pressure Control therapy mode; PC mode delivers Assisted and Mandatory breaths with a user-defined pressure. This is a Pressure Control mode.
PC SIMV Pressure Control Synchronous Intermittent Mandatory Ventilation therapy mode; PC-SIMV mode delivers Spontaneous, Assisted, and Mandatory breaths. This is a Pressure Control mode.
Peak Flow Maximum flow rate (in liters per minute) reached during a breath.
Peak Inspiratory Pressure (PIP)
Highest pressure reached during inspiration.
Term/Acronym Definition · Chapter 12 Glossary

PEEP Positive End Expiratory Pressure
Pressure Control Ventilation (PCV)
Ventilation in which breaths are controlled by operator-defined Pressure, Inspiratory Time, and Rise Time.
Ramp A feature that may increase patient comfort when therapy is started. The ramp feature reduces pressure and then gradually increases the pressure to the prescription setting so patients can fall asleep more comfortably.
Rise Time The Rise Time is the time it takes the ventilator to change from expiration to inspiration.
RR Respiratory Rate (the number of breaths per minute).
SD Card Secure Digital card; This card records usage and therapy data from the device.
Sigh A breath that is delivered every 100 mandatory or assisted breaths at 150% of the normal volume.
SIMV Synchronous Intermittent Mandatory Ventilation therapy mode; SIMV mode delivers Spontaneous, Assisted, and Mandatory breaths with a user-defined inspired tidal volume. This is a Volume Control mode.
Sleep State A low-power state entered when the ventilator is running on the internal or detachable battery and the blower has been off for five minutes. This state preserves battery power. The device is restored by plugging the unit into AC power, inserting an SD card, or by pressing the Start/Stop button.
Spontaneous Breath Breath type in which the breath is patient-triggered.
Term/Acronym Definition · PS Pressure Support

Therapy mode in which all breaths are spontaneous. The ventilator supports breathing with user-defined Pressure Support (PS) and Rise Time values.
Therapy mode that is similar to S mode, except that it can also deliver a mandatory breath if the patient does not spontaneously breathe within a set time.
Tidal Volume The amount of air passing in and out of the lungs for each breath.
Timed (T) Mode Timed pressure support therapy mode. In Timed mode, all breaths delivered are mandatory.
Volume Control Ventilation (VCV)
Ventilation in which breaths are controlled by an operator- defined volume, flow pattern, breath rate, and inspiratory time.
Vte Exhaled Tidal Volume
Vti Inhaled Tidal Volume



This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
Group 1 The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Class B The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies building used for domestic
purpose.
Chapter 13 EMC Information

This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
Immunity Test IEC 60601 Test
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
±2 kV for power supply
±2 kV for supply mains
Mains power quality should be
that of a typical home or hospital
environment.
±1 kV for input-output
±1 kV for input/output
±1 kV differential mode
±1 kV differential mode
Mains power quality should be
that of a typical home or hospital
±2 kV common mode
±2 kV for common mode
environment.
95% dip in UT) for
95% dip in UT) for
Mains power quality should be
that of a typical home or hospital
0.5 cycle
0.5 cycle
environment. If the user of the
40% UT (60% dip in UT) for
40% UT (60% dip in UT) for 5
device requires continued operation
during power mains interruptions,
it is recommended that the device
70% UT (30% dip in
70% UT (30% dip in UT) for
be powered from an uninterruptible
UT) for 25 cycles
power supply or a battery.
95% dip in UT)
95% dip in UT)
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical home or hospital
environment.

This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
Immunity Test IEC 60601 Test
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance: d = 1.2 P
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz outside ISM bandsa
10Vrms 150 kHz to 80 MHz in ISM bandsa
d = 1.2 P
Radiated RF IEC 61000-4-3
10 V/m 80 MHz to 2.5 GHz
10 V/m 26 MHz to 2.5 GHz
d = 1.2 P 80 MHz to 800 MHz d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device. b Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
Chapter 13 EMC Information

Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This Device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and this device as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Power Output of
Separation Distance According to Frequency of Transmitter
150 kHz to 80 MHz
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2 P
d = 2.3 P
d = 1.2 P
d = 1.2 P
0.01 0.12 0.12 0.12 0.23
0.1 0.38 0.38 0.38 0.73
1 1.2 1.2 1.2 2.3
10 3.8 3.8 3.8 7.3
100 12 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power of the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range of 80 MHz and 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.









Respironics, Inc. warrants that the Trilogy200 system shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of two (2) years from the date of sale by Respironics, Inc. to the dealer. If the product fails to perform in accordance with the product specifications, Respironics, Inc.will repair or replace – at its option – the defective material or part. Respironics, Inc. will pay customary freight charges from Respironics, Inc. to the dealer location only. This warranty does not cover damage caused by accident, misuse, abuse, alteration, and other defects not related to material or workmanship.
Respironics, Inc. disclaims all liability for economic loss, loss of profits,overhead, or consequential damages which may be claimed to arise from any sale or use of this product. Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
Accessories and replacement parts, including, but not limited to, circuits, tubing, leak devices, exhaust valves, filters and fuses, are not covered under this warranty. However, Respironics warrants that the Product’s internal and detachable battery (where supplied) will be free from defects in material and workmanship, under normal and proper use and when correctly maintained in accordance with applicable instructions, for a period of 90 days from the date of shipment by Respironics to the original purchaser. This warranty does not apply to batteries that are dropped, misused, altered or otherwise damaged after they are shipped.
This warranty is given in lieu of all other express warranties. In addition, any implied warranties – including any warranty of merchantability or fitness for the particular purpose – are limited to two years. Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
Limited Warranty

To exercise your rights under this warranty, contact your local authorized Respironics, Inc. dealer or contact Respironics, Inc. at:
1001 Murry Ridge Lane Murrysville, Pennsylvania 15668-8550
1-724-387-4000
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Device does not turn on. Nothing happens when you press Start/Stop to begin therapy. The audible indicator does not sound and the backlights on the buttons do not light. | Plug the AC power cord into the device and a wall outlet that is not controlled by a switch. This powers the system and charges the internal battery. If the device still does not turn on, contact an authorized service representative or Philips Respironics to have the device serviced. Please have the model number and serial number ready when you call. | p.140 | |
| The batteries are not charging. The device is plugged in, but the detachable (if installed) and internal batteries are not showing full capacity and are not charging. | Make sure the device is not close to a heat source. Ensure the cooling air vents are not blocked. Bring the ventilator to ambient room temperature. If the problem continues, contact an authorized service representative or Philips Respironics to have the device serviced. Please have the model number and serial number ready when you call. | p.141 | |
| Unexplained changes in the performance of the device The device is not working properly or is making unusual sounds. | Press the Start/Stop button and then select the Right button to shut off the device for a few seconds. Restart the device by pressing the Start/Stop button. If the problem continues, relocate the device to an area with lower EMI emissions (e.g., away from electronic equipment such as cellular phones, cordless phones, computers, TVs, electronic games, hair dryers, etc.). Refer to Chapter 13. If the problem still occurs, contact an authorized service representative or Philips Respironics to have the device serviced. Please have the model number and serial number ready when you call. | p.141 | |
| Excessive Warm Air Out of Mask The air out of the mask is much warmer than usual. | Clean or replace the air filter as described in Chapter 7. The temperature of the air may vary based on the room temperature. Make sure the device is properly ventilated. Keep it away from bedding or curtains that could block the flow of air around the device. Make sure it is away from direct sunlight and heating equipment. Ensure the cooling air vents are not blocked. If the problem continues, contact an authorized service representative or Philips Respironics to have the device serviced. Please have the model number and serial number ready when you call. | p.142 | |
| Mask Discomfort The mask feels uncomfortable to wear, there is significant air leakage around the mask, or the patient experiences other mask-related issues. | Make sure the patient is properly fitted with the correct size mask. If the problem continues, try fitting the patient with a difefrent mask. | p.142 | |
| Runny Nose The patient has a runny nose. | Call the patient’s health care professional. | p.142 |