H e a r t S t a r t D e f i b r i l l a t o r
O W N E R ’ S M A N U A L Guide to Set Up, Operation, Maintenance, and Accessories
M5066A Edition 7

Intentionally blank.


Pads Cartridge Handle. Pull the handle to turn on the HeartStart and remove the cartridge’s hard cover.
Information Button. This blue “i-button” flashes when it has information you can access by pressing it. It also flashes at the beginning of a patient care pause when CPR coaching is enabled.
Speaker. When the device is being used, its voice instructions come from this speaker.
Beeper. The HeartStart “chirps” through this beeper to alert you when it needs attention.
Ready Light. This green light tells you the readiness of the HeartStart.
SMART Pads Cartridge. This disposable cartridge contains self- adhesive pads with attached cable. Shown with adult pads cartridge.
Blinking: standby mode (ready for use) Solid: in use Off: needs attention (HeartStart “chirps” and i-button flashes)
SMART Pads Cartridge Latch. Slide the latch to the right to release the pads cartridge for replacement.
Shock Button. When instructed by the HeartStart to deliver a shock, press this flashing orange button .
On/Off Button. Press this green button to turn on the HeartStart. To turn off the HeartStart, press the green button again and hold it down for one (1) second.
Battery. The non- rechargeable battery is inserted in a recess on the back of the HeartStart.
Infrared (IR) Communications Port. This special lens, or “eye,” is used to transfer HeartStart data directly to or from a computer.

Intentionally blank.

QUICK REFERENCE


Intentionally blank.

O W N E R ’ S M A N U A L
It is important to understand that survival rates for sudden cardiac arrest are directly related to how soon victims receive treatment. For every minute of delay, the chance of survival declines by 7% to 10%.
Defibrillation cannot assure survival. In some victims, the underlying problem causing the cardiac arrest is simply not survivable despite any available care.
Edition 7

Intentionally blank.

The information in this guide applies to the model M5066A HeartStart Defibrillator. Its technical contents apply to all models in the HeartStart HS1 family of defibrillators, including the HeartStart, the HeartStart OnSite, and the HeartStart First Aid Defibrillator. This information is subject to change. Please contact Philips at www.medical.philips.com/heartstart or contact your local Philips representative for information on revisions.
Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard Strasse 2 71034 Boeblingen, Germany (+49) 7031 463-2254
The Philips HeartStart Defibrillator is designed to be used only with Philips-approved accessories. The HeartStart may perform improperly if non- approved accessories are used.
Edition 7 Publication date: July 2007 Publication #: M5066-91901 Assembly #: 011872-0007 Printed in the U.S.A.
In the U.S.A., this device is subject to tracking requirements by the manufacturer and distributors. If the defibrillator has been sold, donated, lost, stolen, exported, or destroyed, notify Philips Medical Systems or your distributor.
© 2007 Philips Electronics North America Corp.
No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system or translated into any human or computer language in any form by any means without the consent of the copyright holder.
The HeartStart Defibrillator is manufactured by Philips Medical Systems, Seattle, Washington, USA.
This product is manufactured and sold under one or more of the following United States patents: US6047212, US6317635, US5892046, US5891049, US6356785, US5650750, US6553257, US5902249, US6287328, US6662056, US5617853, US5951598, US6272385, US6234816, US6346014, US6230054, US6299574, US5607454, US5803927, US5735879, US5749905, US5601612, US6441582, US5889388, US5773961, US6016059, US6075369, US5904707, US5868792, US5899926, US5879374, US5632280, US5800460, US6185458, US5611815, US6556864, US5607454, and other patents pending.
Unauthorized copying of this publication may not only infringe copyright but also reduce the ability of Philips Medical Systems to provide accurate and up-to-date information to users and operators alike.

Intentionally blank.


A Accessories for the HeartStart
B Glossary of terms
C Glossary of symbols/controls
D Warnings and precautions
G Testing and troubleshooting
H Additional technical information required for European conformity
E Technical information · F Configuration

The HeartStart Defibrillator M5066A is part of the Philips HeartStart HS1 family of defibrillators. Small, lightweight, and battery powered, it is designed for simple and reliable operation.
The HeartStart is used to treat the most common causes of sudden cardiac arrest (SCA), including ventricular fibrillation (VF). SCA is a condition that occurs when the heart unexpectedly stops pumping. SCA can occur to anyone – infant, child, adult, male or female – anywhere, at any time. Many victims of SCA do not have warning signs or symptoms.
VF is a chaotic quivering of the heart muscle that prevents it from pumping blood. The only effective treatment for VF is defibrillation. The HeartStart treats VF by sending a shock across the heart, so it can start beating regularly again. Unless this is successful within the first few minutes after the heart stops beating, the victim is not likely to survive.
The HeartStart should be used to treat someone you think may be a victim of SCA. A person in SCA:
If in doubt, apply the pads. Follow the voice instructions for each step in using the defibrillator.
Description · Sudden Cardiac Arrest · Indications for Use

The HeartStart is one part of a well-designed emergency response plan. Any emergency response plan should be under the oversight of a physician and should include training in cardiopulmonary resuscitation (CPR). Philips recommends that you train on the device you will be using.
Several national and local organizations offer combined CPR/defibrillator training. Contact your Philips representative, or visit us on-line at www.medical.philips.com for information about training programs in your area.
National and local requirements
Check with your local health department to see if there are any national or local requirements about owning and using a defibrillator.
Contact your local Philips representative for additional information about the HeartStart. They will be happy to answer any questions you may have and to provide you with copies of the clinical summaries of several key studies using Philips automated external defibrillators.*
You can also find the clinical summaries online at www.medical.philips.com/ heartstart. Technical information about all Philips HeartStart automated external defibrillators is also available online, in the Technical Reference Guide for HeartStart Defibrillators.
* Clinical summaries also include Heartstream ForeRunner and FR2 Defibrillators.
Training and practice · For more information

Check the contents of the HeartStart Defibrillator M5066A box to be sure it contains:
Training materials and optional accessories for the HeartStart are also available from Philips. See Appendix A for a description of these items.
Setting up the HeartStart
Setting-up the HeartStart Defibrillator is simple and quick.
* To replace a used cartridge or insert a different cartridge, first locate the latch at the top edge of the defibrillator, and slide it to the side. The pads cartridge will be released. Lift out the cartridge and replace as described in steps 2 and 3.
Package contents


5. The HeartStart will automatically run a self-test when the battery is inserted. Press the Shock button when instructed. When the self-test is over, the defibrillator will report the result, and tell you to push the green On/Off button in case of an emergency. (Do not push the green button unless this is an actual emergency.) Then the defibrillator will turn off and go to standby mode. The green Ready light will be blinking to show the HeartStart is ready for use.*
* As long as a battery is installed, turning the HeartStart “off” puts it into standby mode, which means that it is ready for use.

7. Store the HeartStart in accordance with your site’s emergency response protocol. Typically, this will be in a high-traffic area that is easy to access, convenient for checking the Ready light periodically, and easy to hear the alarm chirp if the battery power gets low or the defibrillator needs attention. Ideally, the HeartStart should be stored near a telephone, so the Emergency Response Team or Emergency Medical Services can be alerted as fast as possible in the event of a possible SCA. If possible, keep the spare SMART Pads Cartridge and other accessories with the defibrillator – in the carry case if one is used – for quick access when needed. In general, treat the HeartStart as you would any piece of electronic equipment, such as a computer. Be sure to store the defibrillator according to its specifications. See Appendix E for details. As long as a battery and a pads cartridge are installed, the green Ready light should be blinking to show that the HeartStart has passed its most recent self-test and is therefore ready to use.
* The illustration on the cover of the Quick Reference Guide is a 3-step guide to using the HeartStart. Detailed illustrated directions are inside, for reference in an emergency, or if you are hearing impaired or using the HeartStart where it is hard to hear the voice instructions.

It is always a good idea to have a spare battery and a spare pads set. Other things that are useful to keep with the HeartStart include:
Philips has a Fast Response Kit with all these items. See Appendix A for details.
If you may need to defibrillate an infant or a child under 25 kg (55 pounds) or 8 years old, it is recommended that you order the Infant/Child SMART Pads Cartridge, available separately. When the Infant/Child Pads Cartridge is installed in the HeartStart, the HeartStart automatically reduces the defibrillation energy to an energy level more appropriate for infants and children. In addition, if optional CPR coaching is selected, the HeartStart provides coaching appropriate for infants and children. Directions for using the Infant/Child SMART Pads are provided in Chapter 3, “Using the HeartStart.”
Recommended accessories

If you think someone is in SCA, act quickly and calmly. If someone else is available, ask him or her to call for emergency medical assistance while you get the HeartStart. If you are alone, follow these steps:
There are three basic steps to using the defibrillator to treat someone who may be in sudden cardiac arrest:
The following pages provide details about each step.
Overview


Turn on the HeartStart by pulling the SMART Pads Cartridge’s green handle.* Remove the hard cover from the pads cartridge and set it aside. Remain calm and follow the HeartStart’s instructions.
The HeartStart starts by directing you to remove all clothes from the patient’s chest. If necessary, rip or cut off the clothing to bare the person’s chest
* You can also turn on the HeartStart Defibrillator by pressing the green On/Off button.

Pull the tab at the top of the pads cartridge to peel off the film seal. Inside are two adhesive pads on a plastic liner. Remove the pads from the cartridge.
Peel one pad off the liner. Place the pad on the patient’s bare skin, exactly as shown in the picture on the pad. Press the pad down firmly. Then repeat this with the other pad. Be sure the pads have been removed from the liner before placing them.
Where to place pads on adults and children over 25 kg/55 pounds or 8 years old (anterior-anterior).
Where to place pads on infants or children under 25 kg/55 pounds or 8 years old (anterior-posterior).

As soon as the HeartStart detects that the pads are attached to the patient, it begins analyzing the patient’s heart rhythm. It tells you that no one should be touching the patient, and the Caution light begins flashing as a reminder.
If a shock is needed: The Caution light goes from flashing to solid, the orange Shock button starts flashing, and the defibrillator tells you to press the flashing orange button. Before you press the button, make sure no one is touching the patient. When you press the Shock button, the defibrillator tells you that the shock has been delivered. Then the HeartStart tells you it is safe to touch the patient, instructs you to begin CPR, and invites you to press the flashing blue i-button for CPR Coaching if desired.
If a shock is not needed: The HeartStart tells you it is safe to touch the patient and instructs you to perform CPR if needed. (If CPR is not needed – for example, if the patient is moving or regaining consciousness – follow your local protocol until emergency medical personnel arrive.) Then the HeartStart invites you to press the flashing blue i-button for CPR Coaching, if desired.
For CPR Coaching: Press the flashing blue i-button during the first 30 seconds of the patient care pause to activate CPR Coaching.* (If the Infant/Child SMART Pads Cartridge is inserted, CPR Coaching will provide coaching for infant/child CPR.) When the pause is over, the defibrillator tells you to stop CPR, so it can analyze the patient’s heart rhythm. The motion caused by CPR can interfere with analysis, so be sure to stop all motion when instructed.
* The default configuration for the HeartStart provides CPR Coaching when you press the i-button in this situation; however, the default setting can be revised by your Medical Director using Philips software available separately. See Appendix F for more information.

Treating infants and children

If the patient is under 55 pounds or 8 years old, and you have an Infant/Child Pads Cartridge:

With the Infant/Child Pads Cartridge inserted, the HeartStart automatically reduces the defibrillation energy from the adult dose of 150 joules to 50 Joules† and provides optional infant/child CPR Coaching. Place the pads exactly as shown on the illustration on the pads.
* Philips recommends that the HeartStart be stored with an adult pads cartridge installed, as pediatric cardiac arrest is not common. † This lower energy level may not be effective for treating an adult.

If the patient is under 55 pounds or 8 years old, but you do NOT have an Infant/Child Pads Cartridge:
If the patient is over 55 pounds or 8 years old, or if you are not sure of the exact weight or age:
When emergency medical services arrive
When Emergency Medical Services (EMS) personnel arrive to care for the patient, they may decide to apply another defibrillator to allow monitoring of the patient. The SMART Pads should be removed from the patient prior to using another defibrillator. EMS personnel may want a summary of the last-use data* stored in the HeartStart. To hear the summary data, hold down the i-button until the HeartStart beeps.
* See Chapter 4, “After using the HeartStart” for details about data storage.

Reminders • Remove any medicine patches and residual adhesive from the patient’s chest before applying the pads.


The HeartStart automatically stores data about its last clinical use in its internal memory. The stored data can be conveniently transferred to a personal computer or a handheld computer running the appropriate application in the Philips HeartStart Event Review data management software suite. Event Review software is for use by trained personnel only. Information about HeartStart Event Review is available online at www.medical.philips.com/goto/eventreview
* If you leave the battery in the HeartStart after using the defibrillator, then transfer the last-use data to a computer running HeartStart Event Review software, the software will calculate the local date and time of the device use. However, if you remove the battery prior to transferring the data, the software will only show elapsed time.
After each use · HeartStart data storage

Follow your local protocol with regard to prompt data transfer for medical review after using the HeartStart.* Details about data transfer and timing are provided in Event Review documentation.
The information automatically stored by the HeartStart includes a summary of last-use data and detailed data about its last clinical use. You can get a voice summary of information about the last use of the defibrillator by holding the i- button down until it beeps once. The HeartStart will tell you how many shocks were delivered and how long it has been since it was turned on. Summary data are available anytime the defibrillator is ready for use (the battery and pads are installed, and the defibrillator is not turned on) or while it is actually in use. Removing the battery erases the summary data for the last use.
Last-use data stored in internal memory include:
* The HeartStart automatically stores information about its last clinical use in its internal memory for at least 30 days, so the data can be downloaded to a computer running appropriate Event Review software. (If the battery is removed during this period, the defibrillator retains the files. When the battery is reinstalled, the last-use ECG recording will be kept in defibrillator memory for an additional 30 days.) After this time, the last- use ECG recordings will automatically be erased to prepare for a future use. † If ECG recordings from a previous use have not been erased, the maximum time for new
ECG recordings may be less.

The HeartStart is very simple to maintain. The defibrillator performs a self-test every day. In addition, a battery insertion self-test is run whenever a battery is installed in the device. The defibrillator’s extensive automatic self-test features eliminate the need for any manual calibration. The HeartStart has no user- serviceable parts.
Reminders:
Other than the checks recommended after each use of the HeartStart, maintenance is limited to periodically checking the following:
Routine Maintenance · Periodic checks

The outside of the HeartStart and its carry case can be cleaned with a soft cloth dampened in soapy water, chlorine bleach (2 tablespoons per quart or liter of water), or ammonia-based cleaners.
Reminders:
Disposing of the HeartStart
The HeartStart and its accessories should be disposed of in accordance with local regulations.
The HeartStart’s green Ready light is your guide to knowing if the defibrillator is ready for use.
More detailed testing and troubleshooting information is available in Appendix G.
Cleaning the HeartStart · Troubleshooting tips

A Accessories for the HeartStart
Accessories* for the HeartStart Defibrillator available separately from your Philips representative or on-line at www.medical.philips.com/heartstart include:
* Certain accessories require a prescription in the United States.


B
The terms listed in this Glossary are defined in the context of the Philips HeartStart Defibrillator and its use.
AED Automated external defibrillator (a semi-automatic defibrillator).
AED mode The standard treatment mode for the HeartStart Defibrillator. It provides voice instructions guiding the rescuer through applying the adhesive pads, waiting for rhythm analysis, and delivering a shock if needed.
analysis See “SMART analysis.”
arrhythmia An unhealthy, often irregular, beating of the heart.
artifact Electrical “noise” caused by sources such as muscle movements, CPR, patient transport, or static electricity that may interfere with rhythm analysis.
battery The sealed lithium manganese dioxide battery used to power the HeartStart Defibrillator. It is provided in a pack that fits into a compartment on the back of the defibrillator.
configuration The settings for all operating options of the HeartStart Defibrillator, including treatment protocol. The factory default configuration can be modified by authorized personnel using HeartStart Event Review software.
CPR Cardiopulmonary resuscitation. A technique for providing artificial respiration and heart compressions.
CPR Coaching Basic verbal instructions for performing cardiopulmonary resuscitation, including hand placement, rescue breathing, compression depth and timing, provided by the HeartStart when the flashing blue i-button is pressed during the first 30 seconds of a patient care pause.
defibrillation Termination of cardiac fibrillation by applying electrical energy.
ECG Electrocardiogram, a record of the electrical rhythm of the heart as detected through defibrillation pads.

fibrillation A disturbance of the normal heart rhythm that results in chaotic, disorganized activity that cannot effectively pump blood. Ventricular fibrillation (fibrillation in the lower chambers of the heart) is associated with sudden cardiac arrest.
HeartStart Event Review A suite of data management software applications for use by trained personnel to review and analyze HeartStart Defibrillator patient use and by authorized personnel to alter HeartStart configuration. Information is available from Philips Medical Systems on the internet at http://www.medical.philips.com/goto/ eventreview.
i-button A blue “information” button on the front of the HeartStart Defibrillator. If the i-button is pressed during the 30 seconds it flashes during a patient care pause, the HeartStart provides CPR Coaching;* if the i-button is pressed when it is flashing and the HeartStart is chirping, the HeartStart provides troubleshooting guidance. At other times, if the i-button is pressed and held until it beeps once, the HeartStart provides summary information about its last clinical use and device status. When the i-button is on solid (not flashing), it indicates the user may safely touch the patient.
A method of sending information using a special part of the light spectrum. It is used to transmit information between the HeartStart Defibrillator and a computer running HeartStart Event Review software.
NSA “No Shock Advised,” a decision made by the HeartStart Defibrillator that a shock is not needed, based on analysis of the patient’s heart rhythm.
NSA pause A pause provided by the HeartStart Defibrillator following an NSA decision. The pause can be configured to a “standard” NSA pause or a “SMART” NSA pause. During a standard NSA pause the defibrillator performs no background monitoring of patient rhythm. During a SMART NSA pause, the defibrillator conducts background monitoring and, if it detects an artifact-free shockable rhythm, will exit the pause and begin rhythm analysis. If the HeartStart detects artifact such as that created by CPR, or if the user presses the i-button for CPR Coaching during a SMART NSA pause, the defibrillator will not exit the pause for rhythm analysis in order to allow CPR to be completed uninterrupted.
non-shockable rhythm A heart rhythm that the HeartStart Defibrillator determines is not appropriate for defibrillation.
On/Off button A green button located on the front of the HeartStart Defibrillator. Pressing the On/Off button when the defibrillator is in standby mode turns the defibrillator on; pressing and holding the On/Off button for one second when the defibrillator is on turns the defibrillator off and disarms the defibrillator. In addition, pressing the
* Pressing the i-button for CPR Coaching during a SMART NSA pause turns off background monitoring.
infrared communications

On/Off button stops the battery insertion self-test that automatically runs when a battery is inserted.
pads See “SMART pads.”
patient care pause A defined pause to allow patient assessment, treatment, and/or CPR. See “NSA pause” and “protocol pause.”
periodic self-tests Daily, weekly, and monthly tests automatically conducted by the HeartStart Defibrillator when it is in its standby mode. The tests monitor many key functions and parameters of the defibrillator, including battery capacity, pads cartridge readiness, and the state of its internal circuitry.
protocol A sequence of operations performed by the HeartStart Defibrillator to direct patient care in the AED mode.
protocol pause A pause provided by the HeartStart Defibrillator after a shock series, during which the responder can administer CPR. The defibrillator does not conduct background monitoring of the patient’s heart rhythm during this pause.
Ready light A green LED showing the readiness for use of the HeartStart Defibrillator. A blinking Ready light means the defibrillator is ready for use; a solid Ready light means the defibrillator is being used.
rhythm analysis See “SMART analysis.”
Shock button An orange button with a lightning bolt symbol on it, located on the front of the HeartStart Defibrillator. The Shock button flashes when a shock is advised. You must press the button for the shock to be delivered.
shockable rhythm A heart rhythm that the HeartStart Defibrillator determines is appropriate for defibrillation, such as ventricular fibrillation and some ventricular tachycardias associated with sudden cardiac arrest.
shock series interval A configurable interval between shocks, used by the HeartStart Defibrillator to decide if the shocks are part of the same shock series.
SMART analysis The proprietary algorithm used by the HeartStart Defibrillator to analyze the patient’s heart rhythm and determine whether the rhythm is shockable.
The patented, low-energy defibrillation shock waveform used by the HeartStart Defibrillator. It is an impedance-compensated biphasic waveform. Used with the Adult SMART Pads, it delivers 150 Joules, nominal, into a 50 ohm load; used with the Infant/Child SMART Pads, it delivers 50 Joules, nominal, into a 50 ohm load.
SMART NSA pause See “NSA pause.”
SMART Pads The adhesive pads, supplied in a cartridge, used with the HeartStart Defibrillator. Pulling the handle on the cartridge turns on the defibrillator and opens the
SMART biphasic · waveform

cartridge. The pads are applied to the patient’s bare skin and used to detect the patient’s heart rhythm and to transfer the defibrillation shock. Only HeartStart SMART Pads can be used with the HeartStart Defibrillator.
standby mode The operating mode of the HeartStart Defibrillator when a battery has been installed, and the unit is turned off and ready for use when needed. Shown by blinking green READY light.
standard NSA pause See “NSA pause.”
The sudden stopping of the heart’s pumping rhythm, accompanied by loss of consciousness, absence of respiration, and lack of a pulse.
waveform See “SMART biphasic waveform.”
sudden cardiac arrest

Pads cartridge handle. Green. Pulling the handle turns on the defibrillator and opens pads cartridge for use.
Refer to operating instructions.
On/Off button. Green. Pressing the On/Off button when the defibrillator is in standby mode turns the defibrillator on; pressing and holding the On/Off button for one second when the defibrillator is on turns the defibrillator off and disarms the defibrillator. In addition, pressing the On/Off button stops the battery insertion self-test that automatically runs when a battery is inserted.
Information button (i-button). Blue. Pressing the i-button while it is flashing during a patient care pause provides CPR Coaching; pressing it while it is flashing and the defibrillator is chirping provides troubleshooting guidance. Pressing it until it beeps at other times provides summary information about the defibrillator’s last clinical use and device status.
Shock button. Orange. Flashes when the defibrillator is charged. If a shock is needed, the defibrillator directs the user to press the Shock button to deliver a shock to the patient.
Defibrillation protection. Defibrillation protected, type BF patient connection.
Meets the requirements of the European medical device directives 93/ 42/EEC.
Indicates that this device is optimized for Guidelines 2005.
symbol description

Certified by the Canadian Standards Association.
Reference order number.
Expiration date.
Lithium manganese dioxide battery.
One battery in package.
Do not crush the battery.
Do not expose the battery to high heat or open flames. Do not incinerate the battery.
Do not mutilate the battery or open the battery case.
Class 9 miscellaneous dangerous goods. (Symbol required on outer packaging by freight carrier regulations to identify shipments containing lithium batteries.)
Install the battery in the defibrillator before the date (MM-YYYY) shown on the associated label.
Do not expose to moisture.
Handle with care.
symbol description


These pads are disposable and are for single patient use only.
Store the pads at temperatures between 0° and 43° C (32° and 110° F).
Meets the requirements of the European electromagnetic compatibility directive 89/336/EEC.
Pads intended for use on infant or child under 8 years or 55 pounds (25 Kg).

Serial number.
Lot number.
Federal law (USA) restricts this device to sale by or on the order of a physician.
Dispose of in accordance with your country's requirements.
Printed on recycled paper.
symbol description

It is important to understand how to use your HeartStart Defibrillator safely. Please read these warnings and precautions carefully.
A warning describes something that could cause serious personal injury or death. A precaution describes something that could cause minor personal injury, damage to the HeartStart, loss of data stored in the HeartStart, or less chance of successful defibrillation.
flammable gases If the HeartStart is used to give a shock in the presence of flammable gases such as in an oxygen tent, there is a risk of explosion. Move supplemental oxygen and oxygen delivery devices away from the defibrillation pads. (However, it is safe to use the HeartStart on someone wearing an oxygen mask.)
battery The HeartStart M5070A battery is not rechargeable. Do not try to recharge, open, crush, or burn the battery, or it may explode or catch fire.
fluids Do not let fluids get into the HeartStart. Avoid spilling any fluids on the HeartStart or its accessories. Spilling fluids into the HeartStart may damage it or cause a fire or shock hazard. Do not sterilize the HeartStart or its accessories.
accessories Using damaged or expired equipment or accessories may cause the HeartStart Defibrillator to perform improperly, and/or injure the patient or the user.
patient handling Performing CPR or otherwise handling or moving the patient while the HeartStart is analyzing heart rhythm can cause an incorrect or delayed analysis. If the HeartStart tells you a shock is advised while you are handling or moving the patient, stop the vehicle or CPR and keep the patient as still as possible for at least 15 seconds. This will give the HeartStart time to reconfirm the analysis before telling you to press the Shock button.
cell phones The HeartStart can work correctly when it is fairly close to equipment like emergency two-way radios and cell phones. Normally, using a cell phone near the patient should not cause a problem for the HeartStart. However, it is best to keep such equipment only as close as necessary to the patient and the HeartStart.

pads Do not allow the pads to contact other electrodes or metal parts that are in contact with the patient.
device handling The HeartStart was designed to be sturdy and reliable for many different use conditions. However, handling the HeartStart too roughly can damage it or its accessories and will invalidate the warranty. Check the HeartStart and accessories regularly for damage, according to directions.
maintenance Improper maintenance may damage the HeartStart or cause it to function improperly. Maintain the HeartStart according to directions.
skin burns Do not let the pads touch each other or other electrodes, lead wires, dressings, medicine patches, etc. Such contact can cause electrical arcing and skin burns during a shock and may also divert the electrical current away from the patient’s heart. During a shock, air pockets between the skin and pads can cause skin burns. To help prevent air pockets, make sure pads stick well to the skin. Do not use dried out pads because they will not provide good contact with the skin.
patient handling Before delivering a shock, it is important to disconnect the patient from other medical electrical equipment, such as blood-flow meters, that may not incorporate defibrillation protection. In addition, make sure the pads are not in contact with metal objects such as a bed frame or stretcher.
Precautions

The specifications provided in the following tables are nominal values. Additional information can be found in the Technical Reference Manual for HeartStart Automated External Defibrillators, located online at www.medical.philips.com
size 2.80” H x 7.40” D x 8.30” W (7.1cm H x 19cm D x 21cm W).
weight Approximately 3.3 lbs (1.5 kg) with battery and pads cartridge installed.
Operating (battery and pads cartridge installed): 32° to 122° F (0° to 50° C) 0% to 95% RH (non-condensing).
Standby (battery and pads cartridge installed): 50° to 109° F (10° to 43° C) 10% to 75% RH (non-condensing).
Storage/shipping (with battery and pads cartridge): -4° to 140° F (-20° to 60° C) for up to 2 days 0% to 85% RH (non-condensing)
altitude Operates at 0 to 15,000 feet; can be stored at up to 8,500 feet, in standby mode.
Withstands 1 meter drop to any edge, corner, or surface.
vibration Operating: meets EN1789 random, road ambulance.
Standby: meets EN1789 swept sine, road ambulance.

sealing Drip proof per EN60529 class IPx1. Solid Objects per EN60529 class IP2x.
See Electromagnetic Conformity tables.
controls Green SMART Pads cartridge handle Green On/Off button Blue i-button Orange Shock button
indicators Ready light: green; blinks when the defibrillator is in standby mode (ready for use); solid when the defibrillator is being used. i-button: blue, flashes when information is available, on solid during patient care pause. Caution light: flashes when the defibrillator is analyzing, comes on solid when the defibrillator is ready to deliver a shock. Shock button: orange, flashes when the defibrillator is charged and ready to deliver a shock.
audio speaker Provides voice prompts and warning tones during normal use.
beeper Provides chirps when troubleshooting is needed.

waveform parameters Biphasic truncated exponential. Waveform parameters are automatically adjusted as a function of patient defibrillation impedance. In the diagram at left, D is the duration of phase 1 and E is the duration of phase 2 of the waveform, F is the interphase delay (500 µs), and Ip is the peak current. The HeartStart delivers shocks to load impedances from 25 to 180 ohms. The duration of each phase of the waveform is dynamically adjusted based on delivered charge, in order to compensate for patient impedance variations, as shown below:
adult defibrillation load phase 1 phase 2 peak delivered resistance (Ω) duration (ms) duration (ms) current (A) energy (J) 25 2.8 2.8 65 128 50 4.5 4.5 40 150 75 6.25 5.0 30 155 100 8.0 5.3 24 157 125 9.65 6.4 21 159 150 11.5 7.7 18 160 175 12.0 8.0 16 158
pediatric defibrillation (using M5072A infant/child reduced-energy defibrillator pads) load phase 1 phase 2 peak delivered resistance (Ω) duration (ms) duration (ms) current (A) energy (J) 25 4.1 2.8 28 35 50 5.1 3.4 20 46 75 6.2 4.1 15 52 100 7.2 4.8 12 54 125 8.3 5.5 10 56 150 9.0 6.0 9 57 175 9.0 6.0 8 55
Defibrillation Waveform · category specifications

energy* (pediatric doses indicated
Using HeartStart Adult SMART Pads: 150 J nominal (±15%) into a 50 ohm load. Using HeartStart Infant/Child SMART Pads: 50 J nominal (±15%) into a 50 ohm load. Sample pediatric energy doses:
are based on CDC growth charts for the 50th percentile weights for boys.)
age energy dose newborn 14 J/kg 1 year 5 J/kg 2 − 3 years 4 J/kg 4 − 5 years 3 J/kg 6 − 8 years 2 J/kg * National Center for Health Statistics in collaboration with the National Center for Chronic Disease Prevention and Health Promotion. CDC growth charts: weight-for-age percentiles, revised and corrected November 28, 2000. Atlanta, GA: Centers for Disease Control and Prevention © 2000.
charge control Controlled by Patient Analysis System for automated operation.
Shock button flashes, audio tone sounds.
shock-to-shock cycle time
patient care pause-to-shock time
Quick Shock. 8 seconds, typical, from end of patient care pause to shock delivery.
adult shock delivery vector
Via adhesive pads placed in the anterior-anterior (Lead II) position.
Via adhesive pads typically placed in the anterior-posterior position.

function Evaluates impedance of adhesive pads for proper contact with the patient’s skin, and evaluates the ECG rhythm and signal quality to determine if a shock is appropriate.
shockable rhythms Ventricular fibrillation (VF) and some ventricular tachycardias associated with a lack of circulation, including ventricular flutter and polymorphic ventricular tachycardia (VT). The HeartStart uses multiple parameters to determine if a rhythm is shockable. NOTE: For patient safety reasons, some very low-amplitude or low-frequency rhythms may not be interpreted as shockable VF rhythms. Also, some VT rhythms usually associated with circulation will not be interpreted as shockable rhythms.
non-shockable rhythms SMART Analysis is designed to detect non-shockable rhythms as defined by AHA/AAMI DF-80. See following table. On detection of any non-shockable rhythm, the HeartStart prompts user to perform CPR if needed.
pacemaker detection Pacemaker artifact is removed from the signal for rhythm analysis.
artifact detection If electrical “noise” (artifact) is detected which interferes with accurate rhythm analysis, analysis will be delayed until the ECG signal is clean.
analysis protocol Depending on results of analysis, either prepares for shock delivery or provides a pause. For details of protocol, see Appendix F, “Configuration.”
ECG analysis system · category specifications

meets AHA recommendationsb for adult defibrillation
rhythm class ECG test samplea size
observed 90% one-sided performance lower confidence limit
shockable rhythm — ventricular fibrillation
300 sensitivity >90% (meets AAMI DF80 requirement)
shockable rhythm — ventricular tachycardia
100 sensitivity >75% (meets AAMI DF80 requirement)
300 specificity >99% (meets AAMI DF80 requirement)
100 specificity >95% (meets AAMI DF80 requirement)
450 specificity >95% (meets AAMI DF80 requirement)
a. From Philips Medical Systems Heartstream ECG rhythm databases. b. American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy. Automatic External Defibrillators for Public
Access Use: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporation of New Waveforms, and Enhancing Safety. Circulation 1997;95:1677-1682. c. Supraventricular tachycardia (SVT) is specifically included in the non-shockable rhythm class, in accordance with AHA recommendationsb
and the AAMI standard DF80.

battery type 9 VDC, 4.2 Ah, lithium manganese dioxide. Disposable, long-life primary cell.
capacity When new, a minimum of 200 shocks or 4 hours of operating time at 77° F (25° C). (IEC 60601-2-4 2002)
shelf life (prior to insertion)
A minimum of 5 years from date of manufacture when stored and maintained according to directions provided in this Owner’s Manual.
standby life (after insertion)
Typically, 4 years when stored and maintained according to directions provided in this Owner’s Manual.
training life Supports 10 hours of use in training mode.
HeartStart Adult SMART Pads M5071A and Infant/Child SMART Pads M5072A
adult pads Disposable, adhesive defibrillation pads with a nominal active surface area of 85 cm2 each, provided in a snap-in cartridge with an integrated 54” (137.1 cm), typical, cable.
infant/child pads Disposable, adhesive defibrillation pads with a nominal active surface area of 85 cm2 each, provided in a snap-in cartridge with an integrated 40 inch (101.6 cm), typical, cable. Cartridge incorporates teddy bear icon on cover of seal for ready identification.
Use only HeartStart Adult SMART Pads M5071A or Infant/Child SMART Pads M5072A with the HeartStart Defibrillator.

By complying with your national regulations regarding disposal of electric, electronic, and battery waste, you can make a positive contribution to our shared environment. Such waste can introduce harmful elements into the environment as a whole and may also endanger human health.
defibrillator The defibrillator contains electronic components. Do not dispose of it as unsorted municipal waste. Collect such electronic waste separately and dispose of it at an appropriate recycling facility according to your country's regulations.
battery The battery cells contain chemicals. The chemistry used in each battery is identified by a symbol on the label; symbols are defined in the defibrillator User's Guide/Instructions for Use/Owner's Manual. Recycle the battery at an appropriate recycling facility.
pads The used pads may be contaminated with body tissue, fluid, or blood. Cut them off and dispose of them as infectious waste. Recycle the remaining cartridge components at an appropriate recycling facility in accordance with local regulations.
Environmental considerations · product information

The Philips HeartStart Defibrillator comes with a factory default configuration designed to meet the needs of most users. This configuration can only be changed by an authorized person using HeartStart Configure PDA software or Event Review software. This software is for use by trained personnel. Information about HeartStart data management products is available online at http://www.medical.philips.com/goto/eventreview
The following table includes the features of HeartStart Defibrillator operation that are not related to patient treatment.
parameter settings default default description
speaker volume 1, 2, 3, 4, 5, 6, 7, 8 8 The volume of the HeartStart’s speaker is set to 8, highest.
auto send periodic self-test (PST) data
On, Off On Enables the periodic self-test data to be broadcast through the device's infrared data port.
ECG out data On, Off On Enables the ECG data to be broadcast through the device's infrared data port.
Overview · Device options

Patient treatment protocol options
parameter settings default default description
At the start of the first patient care pause
Provides a voice reminder to make sure emergency medical services have been called, at the start of the first patient care pause.
shock series 1, 2, 3, 4 1 The automatic protocol pause for CPR is activated each time a shock is delivered.*
During the protocol pause, the HeartStart does not perform rhythm analysis.
The length of the protocol pause after a shock series is completed is determined by the protocol pause timer setting.
1.0, 2.0, ∞ (infinity)
1.0 A delivered shock must occur within 1 minute of the previous shock to be counted as part of the current shock series. NOTE: This parameter is only applicable when the shock series is not configured to the default 1 shock.
* A shock series begins when a shock is delivered after the HeartStart is turned on. A new shock series begins after a protocol pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous shock exceeds the shock series interval setting.

parameter settings default default description
0.5, 1.0, 1.5, 2.0, 2.5, 3.0 2.0 A 2-minute protocol pause for CPR automatically starts after voice instruction is given when a shock series is completed. After the protocol pause, the defibrillator returns to rhythm analysis. If the user presses the i-button for optional CPR coaching, the HeartStart provides coaching for 5 cycles of CPR, starting and ending with compressions, when the CPR Coaching parameters are also set to their default values. The number of CPR cycles varies for other protocol pause timer and CPR Coaching parameter settings. Note: Because the protocol pause ends upon completion of a CPR cycle in order to maximize the benefits of CPR, the actual duration of the pause may differ slightly from the timer setting.
During a SMART NSA pause, the defibrillator conducts background monitoring. If a potentially shockable rhythm is detected in a motionless patient, the defibrillator terminates the SMART NSA pause and resumes rhythm analysis. NOTE: If the HeartStart detects CPR in progress or if the responder has pressed the i-button for CPR Coaching, the SMART NSA pause will be converted to a standard NSA pause. During the standard NSA pause, the defibrillator does not perform rhythm analysis.
protocol pause timer · (minutes) · pause

parameter settings default default description
0.5, 1.0, 1.5, 2.0, 2.5, 3.0 2.0 A 2-minute NSA pause for CPR automatically starts after voice instruction is given when no shock is advised (NSA).*
If the user presses the i-button for optional CPR coaching, the HeartStart provides coaching for 5 cycles of CPR, starting and ending with compressions, when the CPR Coaching parameters are also set to their default values. The number of CPR cycles varies for other NSA pause timer and CPR Coaching parameter settings. Note: Because the NSA pause ends upon completion of a CPR cycle in order to maximize the benefits of CPR, the actual duration of the pause may differ slightly from the timer setting.
CPR4: Instructs the user that it is safe to touch the patient, to begin CPR, and to press the i-button for CPR Coaching.
The CPR reminder voice instructions provided at the beginning of a pause interval assures the user that it is safe to touch the patient, instructs the user to begin CPR, and invites the user to press the i-button for guidance in the basic steps of CPR. Note: CPR Coaching is available only with the CPR3 and CPR4 settings.
* If the shock series is configured to 2 or more, and a shock has been delivered as part of a series, the length of the first NSA pause within that shock series is determined by the protocol pause timer setting. Otherwise, the length of an NSA pause is determined by the NSA pause timer setting.
NSA pause timer · (minutes)

parameter settings default default description
CPR Coaching adult ventilation
Yes, No Yes Optional CPR Coaching includes rescue breaths at the rate determined by the CPR Coaching compression:ventilation ratio for adults when an adult pads cartridge is installed. NOTE: if this parameter is configured to NO, CPR Coaching will always be compressions-only when an adult pads cartridge is installed.
CPR Coaching infant/child ventilation
Yes, No Yes Optional CPR Coaching includes rescue breaths at the rate determined by the CPR Coaching compression:ventilation ratio for infants and children when an infant/ child pads cartridge is installed. NOTE: if this parameter is configured to NO, CPR Coaching will always be compressions-only when an infant/child pads cartridge is installed.
CPR Coaching compression:ventilation
30:2 adult and 30:2 infant/ child
If the user presses the i-button for optional CPR Coaching during a protocol pause or NSA pause, the HeartStart provides coaching in basic CPR for cycles of 30 compressions and 2 ventilations for adults, children, and infants. Pauses begin and end with compressions.
instruction · instruction · ratio


As long as a battery is installed, the HeartStart Defibrillator automatically tests itself every day and alerts you if it finds a problem. The self-test includes pads readiness testing. In addition, it runs a pads self-test each time a pads cartridge is inserted. It alerts you if it finds a problem. See the Technical Reference Manual, available online at www.medical.philips.com for a detailed discussion of the self- tests.
You can also test the defibrillator at any time by removing the battery for five seconds then reinstalling it. This test takes about one minute. Because the battery insertion self-test is very detailed and uses battery power, running it more often than necessary will drain the battery prematurely. It is recommended that you run the battery insertion self-test only:
If you need to use the defibrillator in an emergency while you are running a battery self-test, pull the SMART Pads cartridge handle to stop the test and to turn on the HeartStart for use.
The HeartStart’s green Ready light is the signal that tells you if the defibrillator is ready for use. The defibrillator also uses chirps and the i-button flashes to alert you to a problem.
Recommended action during an emergency
If for any reason the HeartStart does not turn on when you pull the SMART Pads cartridge handle, press the On/Off button.
If that does not turn on the defibrillator, remove the battery and replace it with a new battery if available and press the On/Off button to turn on the defibrillator. If
Testing · Troubleshooting

no spare battery is available, remove the installed battery for five seconds, then reinsert it and run a battery insertion self-test.
If the problem continues, do not use the HeartStart. Attend to the patient, providing CPR if needed, until Emergency Medical Services Personnel arrive.
Troubleshooting while the HeartStart is in use
(green Ready light is solid)
... to replace the battery
The battery is nearly depleted. The defibrillator will turn off if a new battery is not inserted.
Replace the battery with a new battery immediately.
... there is no cartridge
Insert a new pads cartridge.
installed, and ... to insert a pads cartridge
... to press the pads firmly to
the skin ... to make sure the pads have been removed from
the liner ... the pads should not be
clothing.
... to insert new
The pads cartridge has been opened and the pads peeled off the liner, but the pads have not been successfully attached to the patient. There may be a problem with the pads cartridge.
Replace the damaged pads cartridge. Pull up the handle on the cartridge cover, and replace pads on patient with new pads to continue with the rescue.
immediately · touching the patient’s · pads cartridge

... to stop all motion • The patient is being moved or jostled.
... the shock was not
... the shock button was not
Shock has been advised but the shock button has not been pressed within 30 seconds.
When next prompted, press the Shock button to deliver shock.
delivered · pressed

Troubleshooting while the HeartStart is not in use
(green Ready light is not on)
behavior possible cause recommended action
chirps or i-button flashes
no chirping and/or i-button does not flash
Remove the battery for five seconds then reinstall it to start the battery insertion self-test. If it fails, insert a new battery and repeat the test. If it fails again, do not use the defibrillator. Contact Philips for service.

Guidance and manufacturer’s declaration: The HeartStart is intended for use in the electromagnetic environment specified in the tables below. The customer or user of the HeartStart should assure that it is used in such an environment.
emissions test compliance electromagnetic environment – guidance
The HeartStart uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. The HeartStart is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

immunity test IEC 60601
test level compliance level electromagnetic environment -
± 6 kV contact
± 6 kV contact
There are no special requirements with respect to electrostatic discharge.a
± 8 kV air
± 8 kV air
power frequency (50/60 Hz) magnetic field
3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial/hospital environment. There are no special requirements for non-commercial/non-hospital environments.
radiated RF IEC 61000-4-3
10 V/m 80 MHz to 2.5 GHz
20 V/m Portable and mobile RF communications equipment should be used no closer to any part of the HeartStart, including cables, than is absolutely necessary.b,c The recommended separation distances for various transmitters and the AED are shown in the following table. Interference may occur in the vicinity of equipment marked with the following symbol:
a. Generally, AEDs are sometimes susceptible to interference generated by patient and/or responder motion in environments in which a high
static electric field is present (e.g., low humidity, synthetic carpets, etc.). As a safety measure, Philips AEDs incorporate a patented method to sense possible corruption of the ECG signal by such interference and to respond by directing the user to stop all motion. In these cases, it is important to minimize movement in the vicinity of the patient during rhythm analysis in order to ensure that the signal being analyzed accurately reflects the patient’s underlying heart rhythm. b. The ISM (industrial, scientific and medial) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz;
26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. c. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the HeartStart is used exceeds the applicable RF compliance level above, the HeartStart should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the HeartStart.
Electromagnetic immunity · guidance · electrostatic discharge

Recommended separation distances between portable and mobile RF communications equipment and the HeartStart
The HeartStart is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HeartStart can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the HeartStart as recommended below, according to the maximum output power of the communications equipment.
separation distance according to frequency of transmitter (m)
rated maximum output power of transmitter (W)
80 MHz to 800 MHz d = 0.6√ P
800 MHz to 2.5 GHz d = 1.15√ P
0.01 0.06 0.115
0.1 0.19 0.36
1 0.6 1.15
10 1.9 3.64
100 6.0 11.5
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

The HeartStart Quick Shock feature allows it to deliver a shock within 8 seconds, typical, following the prompt ending a CPR Interval. From shock to shock, the HeartStart takes
Environmental considerations · Shock cycle timing

Intentionally blank.

Intentionally blank.

Intentionally blank.

Intentionally blank.

Intentionally blank.

P O W E R T O S A V E A L I F E
Philips Medical Systems is part of
United States Philips Medical Systems 2301 Fifth Avenue, Suite 200 Seattle, WA, USA 98121 (800) 263-3342
Canada Philips Medical Systems 281 Hillmount Road Markham, Ontario L6C 2S3 (800) 291-6743
Europe, Middle East, and Africa Philips Medizin Systeme Boeblingen GmbH Cardiac and Monitoring Systems Hewlett-Packard Strasse 2 71034 Boeblingen, Germany +49 7031 463 2254
Latin America Philips Medical Systems 1550 Sawgrass Corporate Parkway, Suite 300 Sunrise, FL 33323, USA (954) 835-2660
Asia Pacific Philips Electronics Hong Kong Ltd. 30th Floor, Hopewell Centre 17, Kennedy Road, Wanchai Hong Kong (852) 2821 5888
Royal Philips Electronics