fl Current Loop Wind Stations
WMS-22A Wind Speed indicator
WMS-22 Wind Speed and Direction Indicator
Manual


Remove the Packing List and verify that you have received all equipment, including the following (quantities in parentheses):
0 WMS-22A or WMS-22”. Wind Sensor
_-_ 0 Transmitter Housing Assembly, with I/O connection terminal strips
If you have any questions about the shipment, please call the OMEGA Customer Service Department. When you receive the shipment, inspect the container and equipment for signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
The carrier will not honor damage claims unless all shipping material is saved for inspection. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
Unpacking Instructions · l Operator’s Manual


Toll Free USA Canada One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047
/
Servicing Europe: United Kingdom Sales and Distribution Center
25 Swannington Road, Broughton Astley, Leicestershire
The OMEGA Complete Measurement and
Control Handbooks
& Encyclopedias
& Force V Pressure, Strain V Temperature
Acquistion Systems V Data
Call for Your FREE Handbook Request Form Today: (203)
359-RUSH


WindStations
Instruction Manual
WindStation measures wind speed and direction and converts each paramenter into 4-20
The WMS-22 Current Loop
mA output signals for use by process control or monitoring systems. No external power is required since the encoding electronics for wind speed and for wind direction are isolated and powered from their respective 2-wire current loops.
The WMS-22 wind sensor includes a three-cup anemometer and wind vane. The sensor is ruggedly constructed of UV-resistant ABS plastic and anodized aluminum parts. The cable that connects the pole-mounted wind sensor to the encoder electronics package may be extended up to a total of 250 feet. The encoded current loops signals may be transmitted over distances of up to several miles. The electronics package, which is typically mounted in an indoor locations, is supplied in a gasketed, wall-mounted
NEMA-4X enclosure.
The measurement time base for the wind speed-to-current conversion in the encoder is crystal controlled for stable long-term accuracy. Likewise, ratiometric signal conditioning is used to the direction measurement circuitry to minimize drift and non-linerity.
Two versions are offered; the WMS-22A measures wind speed only, while the WMS-22 measures both wind speed and direction. The WMS-22 systems consist of two subassemblies: a wind sensor and a signal conditioning electronics assembly.
2.1 The Wind Sensor The rotating assembly containing the three-cup anemometer and wind vane for sensing wind speed and directional information is called the wind sensor. The wind vane is mounted on a common axis with the anemometer and includes a tail fin with a nose weight that provides balance. The wind vane is coupled to an angular encoder which is housed within the cylindrical weather skirt.
2.2 The Encoder The encoder electronics package is housed in a sealed NEMA enclosure and contains two independent 4-20
mA data channels, one for wind speed and the second for wind direction. Each- is powered from its respective 2-wire measurement loop.


2.3 Specifications Wind Speed
I Measurement Range O-100 mph
~ Averaging Interval 2.1 Seconds Resolution 1 mph Threshold
f3% F.S.
Range O-360” Mechanical 350” Electrical Resolution 2 Degrees Accuracy
InA Supply Voltage Range 10 to 48 Vdc 2-Wire Loop Interface Screw Terminal Block
t o 20 ou t pu t span 4
3.1 Wind Speed Measurement The WMS-22 wind sensor is designed to be mounted on the end of a
1” outside diameter pole. It is supplied with 40 feet of cable for connection to the electronics housing. The anemometer’s precision ball bearing is protected from the weather and has lifetime lubrication. The counterweight at the end of the wind vane balances the weight of the moving mass over its supporting bearing. It is important that the wind sensor be installed in a location free from any obstructions that would distort the natural flow of air across the sensor.
mA) output of the wind speed channel represents a measured 100 mph wind speed. The full-scale span of the direction channel represents a full 360 degree swing of the wind vane. The potentiometer used as a direction sensor for wind direction has a small arc of resistance discontinuity (typically
5-10”) straddling the North direction (indicated by the set screw in the mounting base). When the wind vane is pointing in this region, the loop current will be encoded as an underscale value of approximately 3.6
mA.
4.1 Unpacking Empty the loose packing material from the corrugated shipping container and carefully lift out the wind sensor assembly.
List of Enclosed Materials


4.2 Wind Sensor Installation Choose the mounting location for the wind sensor and temperature sensor unit which is free from obstructions. Use extreme care to prevent contact with electrical power lines while erecting the unit. A typical installation will position the wind sensor approximately 10 meters above the highest obstacle within a 300 meter radius of the mounting locations. Position the WMS-22 sensor on the end of a
1” diameter TV mast tubing or a
314” Schedule 40 pipe. If the pole is metal it should be well grounded electrically to minimize the probability of any lightning damage. Turn the anemometer on the pole so that the setscrew on the anemometer base is facing True North. Tighten the setscrew to fix the directional orientation.
4.3 Transmitter Installation Mount the transmitter assembly in a covered location where it can be conveniently connected to its exterior wind sensor. The six-wire cable to the wind sensor may be extended to up to 250 feet in length without degrading the accuracy of its measurements. Orient the sensor so that the setscrew in its base is pointing to North. Use care not to puncture the jacket of the interconnecting cable if staples are used to secure it. The standard wind sensor is supplied with 40 feet of cable. In the event that the interconnecting cable needs to be extended, use good splicing techniques and waterproof the splice if it will be exposed to the weather.
I/O connection terminal strips are located at the top edge of the printed circuit board. Terminal functions with the corresponding colors of the wind sensor interconnecting cable are as follows:
Jl 1 1 0 0 1
- (Yellow) 5 lo l I Direction Return
06 0 Unused (Blue)
52 Wind Speed plc Loop + 01 0
- LOOD ~lc Sneed a2 0 Wind
53 Wind Direction plc Loop +
a2 0 Wind Direction plc Loop
5.0 OPERATIONAL CHECKOUT Connect the speed and wind direction process control loops to the appropriate locations of the terminal strip of the printed circuit board as shown on the earlier chart. For test purposes, you may wish to use an unregulated but filtered source of
lo- to 24-volt DC power having a precision 100 ohm resistor (preferably
f 1% accuracy) connected in a series with the supply. The loop powered encoding circuits requires the proper polarity of loop current; however, diodes are included to protect the circuitry against the case of a reversed polarity connection.
- (Black)


lo.9 current. The measured wind speed and direction at several cardinal calibration points should correspond to the following:
Connect the test leads of a digital voltmeter across the resistor to measure the plc
Wind Direction Wind Speed (mph) Loop Current
NE = 45 ” 12.5 6.0 0.6
E = 90 ” 25 8.0 0.8
SE = 135 ” 37.5 10.0 1.0
S = 180 ” 50 12.0 1.2
I 14.0 I 62.5 II SW = 225”
W = 270” 75 16.0 1.6
NW = 315” 87.5 18.0 1.8
N - I 100 I 20 . 0 I 2.0
5.1 Circuit Board Adjustments Multiple turn potentiometers on the printed circuit board are provided for independent gain and offset adjustment of both the speed and direction channels as follows:
The WMS-22 normally should not require adjustment of these controls. An exception occurs if for any reason the wind sensor is changed or if the cabling to the wind sensor exceeds 200 feet in length. In these cases the span control of the direction channel may require a slight adjustment. A movable jumper is included on the PC board direction channel to facilitate checking its offset and span
“JW” position and move it to the “JO” position. This should force the wind direction loop current to its zero scale or 4
adjustments. To use this feature, remove the jumper normally in place at the
mA level. If necessary, adjust R-25 to obtain this value, noting that counterclockwise rotation of its adjustment screw will cause the offset value to increase. Moving the jumper position to the “JF” location will cause the loop current to its full scale value of 20
mA. If necessary, adjust R-24 to obtain this value noting that counterclockwise rotation of its adjustment screw will cause an increase in the span. If either the span or offset trim are changed, it is suggested that both values be checked, since there is some slight interaction between the two adjustments. Be sure to return the jumper to the JW position at the conclusion of this adjustment procedure.
The Wind Speed channel should not normally require adjustment in the field.
(mA) V · 1000 (Volts)


- SOURC E
TYPICAL LO
VDC
PRODUCES OU
MOD E L Z OO - W S - 2 2 CURREN T L OO P W I NDS TAT I O N
TYP I CA L W I R I NG W I T H A S I NG L E PO W ER SOURC E
+ · + · 10 lo 48




WARRANTY ree of defects in mat satisfactory service for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. However, this WARRANTY is VOID, if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA ’s control. Components which wear or which are damaged by misuse are not warranted. These include contact points, fuses, and
triacs.
OMEGA is glad to offer suggestions on the use of its various products. Nevertheless, OMEGA only warrants that the parts manufactured by it will be as specified and free of defects.
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
Every precaution for accuracy has been taken in the preparation of this manual; however, OMEGA ENGINEERING, INC. neither assumes responsibility for any omissions or errors that may appear nor assumes liability for any damages that result from the use of the products in accordance with the information contained in the manual.
SPECIAL CONDITION: Should this equipment be used in or with any nuclear installation or activity, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the equipment in such a manner.
/ INQUIRIES Direct all warranty and repair requests/inquiries to the OMEGA ENGINEERING Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA ’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
CALIBRA- following information available BEFORE
FOR WARRANTY RETURNS, please have the FOR NON-WARRANTY REPAIRS OR
repair/calibra- contacting OMEGA:
TION, consult OMEGA for current
Charges. Have the following information 1.
1 available-BEFORE contaction OMEGA: PURCHASED, 1. Model and serial number of the product under warranty, and 2. Repair instructions and/or specific problems 3. relative to the product.
P.O. number under which the product was
P-0. number to cover the COST of the repair/calibration, Model and serial number of the product, and Repair instructions and/or specific problems relative to the product.
OMEGA ’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
not be cop ied , pho t ocop reproduced, translated, or reduced to any electronic medium or machine-readable form, in w ho le o r in pa rt, w it hou t written consent of OMEGA ENGINEERING, INC.
0 Copyright 1995 OMEGA ENGINEERING, INC. All rights reserved. This documentation m ay


I Need for Process Measurement and Control?
OMEGA...Of Course!
& Thermistor Probes, Connectors, Panels & Assemblies w Thermocouple, RTD
& Ice Point References m Calibrators w Wire: Thermocouple, RTD & Thermistor
& Process Monitors m Recorders, Controllers
& Strain Gages m Transducers
& Pressure Gauges B Load Cells
w Instrumentation B Displacement Transducers
& Flow Computers m Rotameters, Gas Mass Flowmeters
Turbine/Paddlewheel Systems @ w Air Velocity Indicators
pH Electrodes, Testers & Accessories g
& Pumps B Controllers, Calibrators, Simulators l8 Benchtop/Laboratory Meters
pH & Conductivity Equipment
& Compatibles w Plug-in Cards for Apple, IBM @t’ Communications-Based Acquisition Systems B Data Acquisition and Engineering Software
B Recorders, Printers B Datalogging Systems
& Strip Heaters @’ Cartridge m Heating Cable
B Flexible Heaters w Immersion & Band Heaters
& Control Instrumentation B Metering
& Tubing @’ Pumps w Refractometers
& Water Monitors B Air, Soil
& Wastewater Treatment m Industrial Water
pH, Conductivity & Dissolved Oxygen Instruments B


What the fault codes shown on the display mean and how to clear them.
| Code | Meaning | Fix | |
|---|---|---|---|
| P-0 | number to cover the COST of the | p.10 |
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.