OS531, OS532, OS53x-CF, OS533, OS534, OS530L, OS530HR
OS523, OS524 OMEGASCOPE ®
Handheld Infrared Thermometer

Servicing North America: U.S.A.: One Omega Drive, Box 4047 ISO 9001 Certified Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: [email protected]
Servicing Europe: Czech Republic: Frystatska 184, 733 01 Karvina´, Czech Republic Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: [email protected] Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: [email protected] United Kingdom: One Omega Drive, River Bend Technology Centre ISO 9002 Certified Northbank, Irlam, Manchester M44 5BD United Kingdom Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.

Unpacking Instructions · n4

Remove the Packing List and verify that you have received all equipment, including the following (quantities in parentheses):
If you have any questions about the shipment, please call the Customer Service Department at: 1-800-622-2378 or 203-359-1660. We can also be reached on the Internet at www.omega.com e-mail: info@omega.com When you receive the shipment, inspect the container and equipment for signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
The carrier will not honor damage claims unless all shipping material is saved for inspection. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
Unpacking Instructions · Accessories






1.1 Introduction The OS530/OS520 series Handheld Infrared (IR) Thermometers provide non-contact temperature measurements up to 4500°F. They offer effective solutions for many non-contact temperature applications, including the following:

Table 1-1. OS530 Series Handheld Infrared Thermometer Features
General Description 1

* The temperature accuracy is 1% or 2% of Rdg or 3ºF (2ºC) whichever is greater.
General Description 1

** OS523 provides three field of views:
Distance to Spot Size Ratio
General Description 1


Figure 1-1. OS530/OS520 Series Handheld Infrared Thermometer Front View
The display is shown in more detail in Figure 1-2 and described in Table 1-2. There are no user-serviceable parts in the thermometer.
Refer to Chapter 3 for Laser Sight information.
General Description 1


LCK HAL LOBAT ATC LAL PRN °F °C
Table 1-2. Display Details
TC (Thermocouple Input) E (Emissivity) HAL (High Alarm Setpoint) MAX (Maximum Temperature) LAL (Low Alarm Setpoint) MIN (Minimum Temperature) AMB (Ambient Target Temp) dIF (Differential Temperature) PRN (Print Data) AVG (Average Temperature) MEM (Store Temperature Data) ➁ Data associated with one of the Display Modes ➂ Backlighting Icon - allows the display to be viewed under low ambient light ➃ Displays the units of measure in either °F or °C ➄ Main display - displays the current temperature ➅ Locks the trigger / Enables or Disables alarms ➆ ▲for incrementing data; is for turning on/off the backlighting ➇ ▼ for decrementing data; is for changing the units of measure from °F to °C or vice versa ➈ Function key for scrolling through the display modes ➉ Display Icons

TC: OS532, OS533, OS534
MEM: OS534, OS523, OS524

1.2.2 Rear of the Thermometer Figure 1-3 shows the various jacks that are used to connect a recorder or the ac adapter to the thermometer. The figure also shows the location of the tripod thread mount used for fixed point monitoring. More details are provided in Section 2.2.1.

Figure 1-3. OS530 Series Handheld Infrared Thermometer Rear View
General Description 1

General Description 1

2.1.1 Battery Operation Invert the thermometer and install 4 fresh AA size batteries as shown in Figure 2-1. Make sure the batteries’ polarities are correct, the batteries are not put in backwards, and are of the same type.
If the icon flashes, the batteries must be
replaced with fresh batteries immediately.
Figure 2-1. Installing the Batteries
2.1.2 ac Power Operation The thermometer may be operated on ac power using the optional ac adapter. 120Vac/60 Hz and 220Vac/50 Hz adapters are available. When operating on ac power the batteries supply backup power in case of ac power failure. The ac adapter input jack is shown in Figure 1-3.

2.2 Operating the Thermometer 1a. (Without the Laser Sighting) -Aim the thermometer at the target to be measured. Use the “V” groove (shown in Figure 1-1) on top of the thermometer to align the target to the thermometer’s field of view. Look down the “V” groove with one eye only, in order to guarantee proper sighting. Pull and hold the trigger.
1b. (With the Laser Sighting) - Set the laser power switch to the ON position. Aim at the target and pull the trigger. The laser beam and the red power indicator LED will turn on while the trigger is pulled. Refer to Chapter 3 for more details on the Laser Sighting.
Figure 2-2. OS530/OS520 Series with Built-in Laser Sighting
Figure 2-3. Field of View Positions
Field of View · Target


Figure 2-4. Field of View OS531, OS532, OS530L
SPOT DIA.* (IN) SPOT DIA.* (CM)
3' 5' 7' 20"
4.8"
1.0" 1.8" 2.4" 3.0" 3.6" 4.2"
1.0" @ 0" to 20"
1.2" 1.0"
6.0 4.0
2.5cm @ 51cm
8.0 10.0 12.2
160 120 80 40
Figure 2-5 Field of View OS533, OS530HR
** Measurement distance is from the outside surface of the rubber boot.
1' 2' · 4' · 8' 6'


Figure 2-6 Field of View OS534, OS523-1
6" 9" 12" 15"
1.17"
0.9"
.15" .39" .78"
.45"
3.9 9.9 19.9 29.9
*SPOT DIAMETER MEASURED AT 90% ENERGY
22.9 30.5 38.1
Figure 2-7 Field of View OS53x-CF
3"

SPOT DIA.* (MM) SPOT DIA.* (IN)
0' 3' 16' 10'
2.9"
1.9"
0.9"@ 0
1.2" 1.0"
0.9"
*SPOT DIAMETER MEASURED AT 90% ENERGY
5.0 0 1.0 3.0 1.5
Figure 2-8 Field of View OS523-2
SPOT DIA.* (IN)
0’ 3’ 16’ 10’ 2’ 5’
7.0" .35"@ 24"
4.0"
1.6"
.8"
.9"
*SPOT DIAMETER MEASURED AT 90% ENERGY
5.0 0 .61 1.5 1.0 3.0
Figure 2-9 Field of View OS523-3

SPOT DIA.* (MM) SPOT DIA.* (IN)
0' 16' 82' 50'
8.7" 0.5"@ 0
5.1"
0.9"
1.5"
*SPOT DIAMETER MEASURED AT 90% ENERGY
25 0 5 15
Figure 2-10 Field of View OS524
Determine the emissivity of the target (refer to Appendix B). Press the
key to increment the target emissivity. Press the key to
decrement the target emissivity.
on the display. This allows the thermometer to operate continuously
whether or not the trigger is pulled. To unlock the trigger, press the
key again. The icon is no longer displayed. When the
trigger is pulled, the Laser Sighting as well as the display backlight
will stay on .
38 130 221 · 13mm @ 0

2.2.1 Measurement Techniques You can use the IR Thermometer to collect temperature data in any one of five different ways: • Spot Measurement — Measures the temperature of discrete objects such as motor bearings, engine exhaust manifolds, etc.:
the key to lock the trigger.

Figure 2-11 Recorder Hookup
1. Mount the thermometer on a camera tripod and aim at a fixed point on the moving surface. 2. Pull the trigger and press the key to lock the trigger. 3. If necessary, adjust the emissivity. The thermometer is now set up for measuring the temperature of a moving surface. 4. To record the temperature profile of the moving surface, connect the IR thermometer to a strip chart recorder. Refer to Figure 2-11 for details. 5. After all data is taken, press the key to unlock the trigger.
It is recommended that you use the ac adapter for long term measurement of temperature.

Definition: Real Time Mode is the active operational mode of the thermometer. In this mode, the thermometer constantly measures and displays temperature.

Figure 2-12. General Operational Block Diagram
If the trigger is pulled two times in rapid sequence, it may reset the emissivity, high alarm, low alarm and target ambient temperature to the default values.

Table 2-1. Functional Flow Chart when the Trigger is Pulled (Real Time Mode)

° ° NOTE: The unit of measure ( F or C) stays on (does not flash) during Recall Mode.


(Model OS531, OS532 OS530L, OS530HR)
Figure 2-13. Visual Function Flow Chart

Use key to change temperature from °F to °C or vice versa.

2.3.1 Adjusting Emissivity Refer to Appendices B and C for information on emissivity.
emissivity or press the key to decrement the target
emissivity.
The Emissivity Display Mode (E) appears every time the trigger is pulled regardless of how the Display Mode was previously set.
The emissivity setting does not change when the thermometer is turned off. However, when the batteries are replaced, the emissivity is reset to 0.95, the default value.
2.3.2 Using the LOCK Function This function electronically locks the trigger mechanism:
icon will appear on the display. 2. Release the trigger. This allows the thermometer to operate continuously whether or not the trigger is pulled.
To unlock the trigger function, press the key again,
and the icon is no longer displayed.

2.3.3 Calculating Temperature Values The thermometer calculates the MAX, MIN, dIF, and AVG temperatures based on the current temperature.
is the maximum temperature since the temperature measurement session starts (pulling the trigger).
is the minimum temperature since the temperature measurement session starts.
is the difference between the MAX and MIN temperatures.
is the true average temperature since the temperature measurement session starts. The average temperature under continuous operation is accurate for a limited period of time (refer to the specifications). However, the AVG temperature function can be used indefinitely when the thermometer is operating intermittently.
“AVG ---” is displayed when either of the following conditions occur: 1. When the average temperature measurement reaches its time period as stated in the specifications. 2. When the thermometer is trying to measure a target temperature which is outside of its measuring temperature range.
To clear the “AVG ---” display, turn off the thermometer.
Every time the thermometer goes from the sleep mode to the Real Time mode (by pulling the trigger) the MAX, MIN, dIF, AVG and TC temperatures are updated.
During the time that the thermometer displays either MAX, MIN, dIF, AVG or thermocouple temperatures, press the key to
change all the temperatures from °F to °C or vice versa.
During the time that the thermometer displays either MAX, MIN, dIF, AVG or thermocouple temperatures, press the key to turn the display backlighting ON/OFF while the unit is in LOCK mode and trigger is released. In addition, the display backlight turns on automatically when the trigger is pulled.

2.3.6 Thermocouple Input (OS532, OS533, OS534) The thermometer accepts thermocouple input. It displays thermocouple temperature and the target temperature (via infrared) simultaneously. This function provides an accurate method of determining an unknown emissivity.
° F
thermometer as shown in Figure 1-3.
probe.
mode (E) appears.
the temperature reading matches the thermocouple
temperature measurement.
emissivity.
"TC--- is” is displayed when the thermocouple
input is open or out of range (0 to 1600°F).
° F

2.3.7 Using the Alarm Functions The thermometer provides audible and visible alarm indications. • To set the high alarm value:
the High Alarm Display Mode (HAL) appears.
the key to decrement the high alarm value.
icon appears.
If the temperature exceeds the high alarm setpoint,
you will hear a beep and the icon on the
display flashes.
and the icon disappears.
If you are not in High Alarm Display Mode (HAL) when
the high alarm goes off, you must press the key to
get into the High Alarm Display Mode. Then press the
key to disable the high alarm.
The high alarm setpoint does not change when the thermometer is turned off. However, when the batteries are replaced, it is reset to the default value as follows:

the Low Alarm Display Mode (LAL) appears.
the key to decrement the low alarm value.
icon appears.
If the temperature drops below the low alarm
setpoint, you will hear a beep and the icon on
the display flashes.
the icon disappears.
If you are not in Low Alarm Display Mode (LAL)
when the low alarm goes off, you must press the
key to get into the Low Alarm Display Mode.Then
press the key to disable the low alarm.
The low alarm setpoint does not change when the thermometer is turned off. However, when the batteries are replaced, it is reset to the default value of 0°F (1000°F for OS524).

2.3.8 Using Ambient Target Temperature Compensation (OS533, OS534, OS523, OS524)
Use the Ambient Target Temperature Compensation
(AMB) Display Mode when high accuracy readings under both of these conditions are required:
To set and activate the Ambient Target Temperature Compensation Mode:
trigger. Set the emissivity to 1.0 (refer to Section 2.3.1).
Mode (AVG) appears.
Temperature Display Mode (AMB) appears.
key or the key.
temperature compensation. The icon appears
on the display.
To disable this mode, press the key again.
The icon disappears.

Mode (E) appears.
To disable the Ambient Target Temperature
Compensation at a later time, you must press the
key to get into the Ambient Target
Temperature Display Mode. Then press the
key to disable it.
The target ambient temperature does not change when the thermometer is turned off. However, when the batteries are replaced, it is reset to the default value of 75°F.
2.3.9 Sending Temperature Data to a Serial Printer (OS533, OS534, OS523, OS524)
The thermometer can transmit temperature data to a Serial Printer via the RS-232 phone jack and the RS-232 cable.
this mode


printing interval (from 1 to 1999 seconds) is the time
between data points. The default value is 2 seconds.
To stop sending data, press the key again, and


(E) appears.
2.3.10 Sending Temperature Data to a Personal Computer (OS533, OS534, OS523, OS524)
The thermometer can transmit temperature data to a Personal Computer via the RS-232 phone jack and the RS-232 cable.


Figure 2-15. Personal Computer Hookup
2.3.10.1 Software Installation System Requirements To run IR_TEMPSOFT you must have an IBM-compatible personal computer with these minimum specifications:

Operating System: Win95, Win98, Win NT 4.0, Windows 2000 Processor : Pentium Class (Any speed) Ram: 32 Mega-Bytes Monitor: VGA, Or Higher Resolution Video Card: Supports VGA, 256 Color, 800 x 600 Resolution (Minimum) CommPort: 4800 Baud, 8 Databits, 1 Stop, No Parity. CD-ROM: Any Speed Hard Drive: 10 Mega-Bytes Available Space (For Installation Only)
How to Install IR_TEMPSOFT
Please follow this installation procedure to install IR_TEMPSOFT:
If you have any trouble with the installation of this

program please contact technical support at Omega Engineering Inc.
Connecting the IR-System to the PC
Make certain that your Infrared Thermometer is powered down before you attempt to make any connections to the personal computer. Making connections between powered devices and a powered PC can cause damage to communications ports, or other electronic circuitry in your device or computer.
thermometer. Point the thermometer to your target, and make certain that you are reading a temperature. Do not begin to print (send) data at this time.
Starting The IR_TEMPSOFT Software
Receiving Real-Time Temperature Data

waiting for the thermometer to transmit data through your RS-232 port.
pressing and holding the key on the thermometer
until the Print Data Display Mode (PRN) appears.
transmitting data. The "PRN" icon appears on the
thermometer display. You should now see the current
temperature, the downloaded parameters, and the
temperature data being charted.
If instead the "Scanning CommPort for IR Transmission" popup is still visible, then there is a communications error. Check the communications cable, and its connection at both the IR device and computers comm-port. Make certain that you are connected to your configured comm-port.
From the main menu, click on Help, and select On-Line Manual. The on-line Manual has more detailed information regarding all the features of the software.
To stop the data transmission from the thermometer, press
the key and the PRN icon will disappear.
The transmitter temperature data is the average temperature for every printing interval period.

2.3.11 Storing the Temperature Data on Command (OS534, OS523, OS524)
The thermometer can store up to 100 temperature data points on command. Each set of temperature data is broken down into the temperature value, emissivity, and high alarm setpoint for that temperature. This data is stored in non-volatile memory, so removing the batteries will not affect this data. To store temperature data:
key to lock the trigger. The icon will appear
on the display.
emissivity or press the key to decrement the target
emissivity.
Mode (MEM) appears.
press the key to decrement the memory location.
The memory location can be from 001 to 100.
memory location indicated. You will hear a beep to
verify that the data is stored.
You can write over any previously stored memory locations.
until the Emissivity Display Mode (E) appears.

2.3.12 Erasing the Temperature Data from Memory The user can erase all 100 temperature data points in memory at any time by using the following procedure:
icon will appear.
If the icon disappeared, repeat steps
The display freezes momentarily, and a beep sounds for about 1 second. Now the memory is cleared. The thermometer reverts to real time mode.
Erasing the temperature data does not erase or reset Emissivity, High and Low Alarm setpoints, printing interval and Ambient Target Temperature compensation
key to unlock the trigger.
1 and 2

Definition: Recall Mode is the passive operational mode of the thermometer. In this mode, you may review the most recently stored temperature data and parameters.
Display Turns Off in Approx. 10 Seconds
Real Time Mode (Active)
Figure 2-16. General Operational Block Diagram
In order to get into the Recall Mode of operation,
press the key only. Do not pull the
trigger; otherwise, you will get into the Real Time
(Active) Mode of operation.

Table 2-2. Functional Flow Chart (Recall Mode)

° ° NOTE: The unit of measure ( F or C) stays on (does not flash) during Recall Mode.

The thermometer stores the last temperature measured in
the real time mode (refer to Table 2-1). This temperature
can be recalled by pressing the key.
- Press the key to review the most recently stored
temperature data and parameters. You may review:
3.6.2 Downloading Previously Stored Temperature Data (OS534, OS523, OS524)
Up to 100 sets of stored temperature information can be downloaded to a serial printer or a personal computer. Each set of temperature information is broken down to the temperature value, the Emissivity, and the high alarm setpoint for that temperature.
1a. Turn on the Serial Printer and set the following parameters:
1b. Turn on the Personal Computer.
Calculated values · Set values

3. To download stored temperature data points from the thermometer, first make certain that it is not in printing mode. Make sure that the IR_TEMPSOFT is installed properly as explained in section 2.3.10. 4. On the main menu bar, click on “Command-> Download Stored Data”. The “Stored Data” window will appear. 5. The “Stored Data Points” window will appear. 6. Click on the “START RECEIVING” button to bring the program into receive mode. Release the trigger to put the thermometer into sleep mode.
Mode (PRN) appear.
the Personal Computer. The “PRN” icon appears on the thermometer display. When the stored data is completely
downloaded, you will hear a peep and the “PRN” icon disappears.

You can review all 100 stored temperature values on the thermometer display using the following procedure:
Display Mode (MEM) appear.
press the key to decrement the memory location.
The memory location can be from 001 to 100.
the lower portion of the display. If there is no data stored
in a memory location, the display shows “ ”.
and 3.
If no keys are pressed, the thermometer goes into sleep mode in approximately 10 seconds.

CAUTION
You may receive harmful laser radiation exposure if you do not adhere to the warnings listed below:
DO NOT ATTEMPT TO OPEN THE LASER SIGHTING.
(There are no user-serviceable parts in the module.)
Refer to the inside back cover for product warning label.

The Laser Sighting is built into the thermometer. It provides a visual indication of the field of view of the thermometer. Aiming at distant targets (up to 40 feet) becomes much easier by using the Laser Sighting. It is offered in two different models, laser dot, and laser dot/circle switchable.
OS53x-CF and OS523-3 — Thermometer with built-in Laser Dot
All other models — Thermometer with built-in Laser Dot/Circle Switchable
Figures 3-1 and 3-2 show the rear and front view of the thermometer with the built-in laser sight module.


Figure 3-1. Rear View
Figure 3-2. Front View
Laser Sighting 3 · of the Thermometer · of the Thermometer

The laser beam will stay on as long as the trigger is pulled. If the trigger is locked (the key is previously pressed) or released, the laser beam will turn off. In order to turn on the Laser Sighting again, pull the trigger again.
Figure 3-3 shows the two different laser configurations. The laser Dot indicates the center of the field of view of the thermometer. The laser Circle indicates the perimeter of the thermometer’s field of view.
The visibility of the laser beam depends on the ambient light levels.
Laser Dot Laser Circle
Figure 3-3. Two Laser Configurations
Laser Sighting 3

The Laser Sighting turns on only when used with the thermometer. The module does not turn on by itself.
The line of sight of the thermometer does not coincide with that of the Laser Sighting, as shown in Figure 3-4. The two lines of sight become less critical when measuring distant targets. For example, at 30 feet from the target and a 3 foot diameter target size, there is a 2.7% offset error with respect to the target size. For close-up targets, first make sure the target fills the laser circle, then point it with the center of the beam approximately 1" below the center of the target. A simple method to make infrared measurements is to scan the laser beam across the target area vertically and horizontally and recall measurements of maximum for hot and minimum for cold target (compared to the background) to obtain the correct temperature.
Figure 3-4 Lines of Sight of the Laser Sighting and Thermometer
Laser Sighting 3

The Sighting scope is an accessory for the thermometer. It provides a visual indication of the target being measured. Aiming at distant targets (up to 200 feet) becomes much easier by using the Sighting scope.
4.2 Installing and Operating the Sighting Scope 1. If the sighting scope is already installed on the thermometer, go onto step 5.
Since the sighting scope mounts on top of the thermometer, the line of sight of the thermometer does not coincide with that of the sighting scope, as shown in Fig. 4-1. The distance between the two lines of sight (111⁄16") becomes less critical compared to the target size when measuring distant targets (50 feet and longer).

Pair of Mounting Clamps Line of sight of the sighting scope

Figure 4-1. Installing the Sighting Scope
Sighting Scope 4

NOTE
When you change the batteries, all of the set parameters (i.e. emissivity, high alarm, low alarm, Target Ambient Temperature) will be reset to the default values. For your convenience, you may want to write down all of the set parameters BEFORE replacing the batteries.
The thermometer is powered by 4 standard AA size lithium batteries. To replace the batteries:
When the battery power is so low that accurate measurements are no longer possible, the thermometer shuts down and you must replace the batteries immediately. You will see and hear the following:
Do not open batteries, dispose of in fire, heat above 100°C (212°F), expose contents to water, recharge, put in backwards, mix with used or other battery types – may explode or leak and cause personal injury.

Although all lenses are quite durable, take care to prevent scratching when cleaning them. To clean the lens:
Do not use any ammonia or cleaners with ammonia on the lens, as damage may result. Do not wipe the surface dry, as this may cause scratching.
The thermometer can not be calibrated by the user. For precise calibration of the thermometer, call our Customer Service Department. It is recommended that the Infrared Thermometer to be sent to the factory once a year for recalibration.
Servicing and maintenance is not required to keep the laser sighting in proper operating condition. In the event of a malfunction, the unit should be returned to the manufacturer for repair.
5 Maintenance

Problem Solution
The thermometer does 1a. Properly install fresh batteries. not turn on (No Display)
1b. If operating under ac power, check that the ac adapter is plugged in properly to the ac wall outlet and to the thermometer.
1c. Make sure the batteries make good contact - remove and reinstall the batteries.
a. Press and hold the key.
b. Pull the trigger.
c. Release the trigger.
d. Release the key.
e. The version of the software is
displayed for about 1 second.
The emissivity display mode
immediately follows with the
emissivity of 0.95.
- The icon 1. Properly install fresh batteries. flashes. - The thermometer
beeps intermittently. - The thermometer
flashes “ “ in the Main Display.

The thermometer is Remove and reinstall the batteries or “locked up” (the disconnect and reconnect the ac display is “frozen”). adapter.
The display is either 1. Clean the thermometer lens. erratic or stays at one Refer to Section 4.2. reading.
a. Pull the trigger and press the
key to lock the trigger.
b. Hold down the key and
press the key until
“VER X.X” is displayed.
You can expect to see and hear the following:
Troubleshooting Guide 6

If you see an error code, either “ERR1”, “ERR2”, or “ERR3”, record the code and call our Customer Service Department. Provide Customer Service with the error code that is displayed in the upper left corner of the display. The Customer Service Department representative may ask you to return the thermometer to the factory.
c. After running the diagnostic
program, press the key to
unlock the trigger.
The temperature reading is erratic. The thermometer has just been moved from one extreme temperature to room temperature [0°C or 50°C (32°F or 122°F)] or vice versa.
The temperature reading is erratic. The thermometer has just been moved from room temperature (ambient temperature) to a temperature 10°C colder or warmer.
Troubleshooting Guide 6

The thermometer resets itself unexpectedly. The emissivity has been reset to .95. All other parameters are reset to the default values.
The trigger is pulled two times in rapid sequence. Wait at least 2 seconds between two successive trigger pulls. You may need to set the emissivity, low alarm, high alarm, target ambient temperature values again.
No Laser Beam 1. Make sure the trigger is pulled and the laser power switch is turned on. (The red power LED should be lit).
The Laser "line of sight"
does not coincide with the center of the target.
Troubleshooting Guide 6 · Laser Sighting

(Specifications are for all models except where noted)
THERMOMETER Measuring: OS530HR, -22°F to 250°F (-30°C to 121°C) Temperature OS531: 0°F to 750°F (–18°C to 400°C) Range: OS530L, OS533,OS532: 0°F to 1000°F (–18°C to 538°C) OS534 0°F to 1600°F (–18°C to 871°C) OS523 0°F to 2500°F (–18°C to 1371°C) OS524 1000°F to 4500°F (538°C to 2482°C) OS531: ±2% of reading or 3°F whichever is greater All others ±1% of reading or 3°F whichever is greater (2% Rdg for temp > 2000°F for OS524)
Accuracy (24°C or 75°F Ambient Temperature and at emissivity of 0.95 or greater):
Field of Vision: OS531, OS532, OS530L 10:1 OS533, OS530HR 20:1 OS534 30:1 OS53x-CF .15"@6" OS523-1 30:1 OS523-2 60:1 OS523-3 68:1 OS524 110:1
Spectral Response: 8 to 14 microns (2 to 2.5 microns, OS524) (5 to14 microns, OS530L, OS531, OS532)
Thermocouple Input Type K, 0 to 1600°F (-18 to 871°C) (OS532, OS533, OS534 only)
Input Connection SMP Connector
Thermocouple Display ±5°F (±3°C) Accuracy @ 75°F (24°C) Ambient Temperature
Thermocouple Display Response Time 2 seconds Operating Ambient 32°F to 122°F Temperature: (0°C to 50°C)

Average Temperature Accuracy Time Period (under continuous operation): 111⁄2 days
Calculated Temperature Maximum (MAX), Minimum (MIN), Values: Average (AVG), Differential (dIF), Thermocouple (TC)
Ambient Target Set and enabled via keypad Temperature Compensation: (OS533, OS534, OS523, OS524)
RS232 Output (for Standard on OS533, OS534, OS523, OS524 personal computers 4800 bits per second, 8 bits of data, and serial printers): 1 stop bit, no parity
RS-232 Cable: RJ12 to 25 pin D connector, Female RJ12 Pin # 25 pin D connector Pin #
3 3 5 7 4 & 5 jumpered 6 & 20 jumpered
OS533, OS534 Low alarm standard, with OS523, OS524: audible and visual indication
Data Storage: OS534,OS523 Up to 100 sets of data OS524: points on command. Each set of data consists of the temperature, the Emissivity and the high alarm value.
Specifications 7

Battery Storage –40°C to 50°C (-40°F to 122°F) Temperature
ac adapter: Optional - available in 120 Vac or 220Vac Class 2 Transformer, UL & CSA Listed
Alkaline Battery Life at 24°C (75°F) ambient temperature
(With Laser Sighting Dot or Circle): 6 hours (continuous operation)
(With LCD backlighting and no Laser Sighting): 17 hours (continuous operation)
Lithium Battery Life at 24°C (75°F) ambient temperature
(With Laser Sighting Dot or Circle): 30 hours (continuous operation)
(With LCD backlighting and no Laser Sighting): 3 days (continuous operation)
Specifications 7 · Battery Types · – +

Laser Dot 2 to 40 ft. Laser Circle 2 to 15 ft. Max. Output Optical Power:
Class II Laser Product European Classification: Class 2, EN60825-1 Maximum Operating Current: 25mA at 5.5 V FDA Classification: Complies with 21 CFR Chapter 1, Subchapter J Beam Diameter: 5 mm Beam Divergence:
(for the label layout, refer to the inside back cover)
Specifications 7

Press and hold down the
key & then press the • Allows you to go to the Diagnostic Routine.
and keys • Allows you to erase all 100 stored temperature data from the memory.
Key(s) Key(s) Functions · key

Glossary of Key Strokes 8

Thermal Radiation
Heat is transferred from all objects via radiation in the form of electromagnetic waves or by conduction or convection. All objects having a temperature greater than absolute zero (-459°F, -273°C, 0 K) radiate energy. The thermal energy radiated by an object increases as the object gets hotter. Measurement of this thermal energy allows an infrared thermometer to calculate the object’s temperature if the emissivity (blackness) is known. Generally, it is convenient to measure the amount of radiated energy in the infrared part of an object’s radiation spectrum.
Figure A-1 shows a block diagram of an infrared radiation thermometer. Energy from the object is focused by the lens onto the detector. As the detector heats up, it sends out an electrical signal, which in turn is amplified and sent to the circuitry of the thermometer. The thermometer software then calculates the temperature of the object.
OBJECT
AMPLIFIER
DETECTOR
OR
THERMOMETER CIRCUITRY AND
DISPLAY
SIGNAL OUTPUT TO SERIAL PRINTER
OR PERSONAL COMPUTER
Figure A-1. Infrared Thermometer Block Diagram

When thermal radiation falls on an object, part of the energy is transmitted through the object, part is reflected and part is absorbed. A blackbody is defined as an ideal object that absorbs all the radiation incident upon it. The best example of a real object that acts like a blackbody is a small hole drilled deep into a large opaque cavity. Thermal radiation entering the cavity is internally reflected and has little chance of escaping the cavity before it is fully absorbed.
Emissivity is defined as the ratio of energy radiated by an object to that of the energy radiated by a blackbody. By definition, the emissivity of a blackbody is 1. Most objects are considered gray objects with an emissivity between 0 and 1. Various emissivities for common materials are shown in Appendix B.
Objects radiate energy at different wavelengths, but not with constant intensity at each wavelength. Figure A-2 shows the energy radiated by a blackbody at various temperatures as a function of wavelength. As a body is heated, the intensity of the radiated energy increases and the peak of the curve shifts towards the shorter wavelength end of the spectrum. The total area under a spectral distribution curve is proportional to the total energy radiated by the blackbody at a given temperature.

Relative emission from a blackbody versus wavelength. The area under the curve corresponds to the total ener- gy, and is proportional to the absolute temperature to the 4th power. The peak of the spectral distribution curve shifts to shorter wavelengths as the temperature increases.
Figure A-2. Blackbody Spectral Distribution

Wien’s Displacement Law describes the exact mathematical relationship between the temperature of a blackbody and the wavelength of the maximum intensity radiation.
T
where λm = wavelength measured in microns
T = temperature in Kelvin
The net thermal power radiated by an object has been shown to depend on its emissivity, its temperature and that of the ambient temperature around the object. A very useful equation known today as the Stefan-Boltzmann Law has been shown both theoretically and empirically to describe the relationship.
I = thermal power in watts/meter2
ε = Emissivity σ = 5.6703 x 10-8 watts/meter2 x K4 (Stefan’s constant) T = temperature of object in Kelvin Ta = temperature of ambient surroundings in Kelvin
The infrared thermometer uses this equation directly in calculating the temperature of an object. The incident power is measured by the infrared detector. The emissivity of the object is determined by the user. The ambient temperature is measured by a sensor inside the thermometer. With all quantities known, the thermometer uses the Stefan-Bolzmann Law to calculate and output the temperature of the object.

Accurate measurement of temperature via infrared means depends strongly on the size of the object and the distance between the thermometer and the object. All optical devices (e.g. cameras, microscopes, infrared thermometers) have an angle of vision, known as a field of view or FOV, within which they see all objects. In particular, the thermometer will measure a fixed proportion of the energy radiated by all objects within its FOV. The user must guarantee that the distance between the thermometer and the object is defined so that only that object fills the FOV of the instrument.
Referring to Figure A-3, Objects “X” and “Y” are within the FOV of the thermometer. The measured temperature would fall somewhere between the actual temperatures of the two objects. In order to measure the temperature of Object “X” accurately, Object “Y” would need to be removed. In order to measure the temperature of Object “Y” accurately, the user would need to move closer to Object “Y” until it completely filled the FOV of the thermometer. Alternatively, the user could measure the temperature of Object “Y” with a thermometer with a smaller FOV.
OBJECT "X"
FOV ANGLE
OBJECT "Y"
DISTANCE
Figure A-3. Field of View of a Thermometer
The distance-to-spot size ratio (D⁄S) defines the field of view (FOV). Thus, a D⁄S = 10 gives you approximately a 1 foot spot size at a distance of 10 feet. For accurate spot size values, refer to the Field of View diagrams shown in Figures 2-4 through 2-6.

Table B-1 provides guidelines for estimating the emissivity of various common materials. Actual emissivity, especially of metals, can vary greatly depending upon surface finish, oxidation, or the presence of contaminants. Also, emissivity or infrared radiation for some materials varies with wavelength and temperature. To determine the exact emissivities for most applications, follow the procedures in Appendix C.
Table B-1. Emissivity Table
Material Emissivity (ε) Aluminum – pure highly polished plate 0.04 – 0.06 Aluminum – heavily oxidized 0.20 – 0.31 Aluminum – commercial sheet 0.09 Brass – dull plate 0.22 Brass – highly polished, 73.2% Cu, 26.7% Zn 0.03 Chromium – polished 0.08 – 0.36 Copper – polished 0.05 Copper – heated at 1110°F (600°C) 0.57 Gold – pure, highly polished or liquid 0.02 – 0.04 Iron and steel (excluding stainless) – polished iron 0.14 – 0.38 Iron and steel (excluding stainless) – polished cast iron 0.21 Iron and steel (excluding stainless) – polished wrought iron 0.28 Iron and steel (excluding stainless) – oxidized dull wrought iron 0.94 Iron and steel (excluding stainless) – rusted iron plate 0.69 Iron and steel (excluding stainless) – polished steel 0.07
Iron and steel (excluding stainless) – polished steel oxidized at
1110°F (600°C) 0.79 Iron and steel (excluding stainless) – rolled sheet steel 0.66 Iron and steel (excluding stainless) – rough steel plate 0.94 – 0.97 Lead – gray and oxidized 0.28 Mercury 0.09 – 0.12 Molybdenum filament 0.10 – 0.20 Nickel – polished 0.07 Nickel – oxidized at 1200°F – 2290°F 0.59 – 0.86 Platinum – pure polished plate 0.05 – 0.10 Platinum – wire 0.07 – 0.18 Silver – pure and polished 0.02 – 0.03 Stainless steel – polished 0.07 Stainless steel – Type 301 at 450°F – 1725°F 0.54 – 0.63 Tin – bright 0.06 Tungsten – filament 0.39 Zinc – polished commercial pure 0.05 Zinc – galvanized sheet 0.23

Asbestos Board 0.96
Asphalt, tar, pitch 0.95 – 1.00
Brick – red and rough 0.93
Brick – fireclay 0.75
Carbon – filament 0.53
Carbon – lampblack - rough deposit 0.78 – 0.84
Glass - Pyrex, lead, soda 0.85 – 0.95
Marble – polished light gray 0.93
Paints, lacquers, and varnishes – Black matte shellac 0.91
Paints, lacquers, and varnishes – aluminum paints 0.27 – 0.67
Paints, lacquers, and varnishes – flat black lacquer 0.96 – 0.98
Paints, lacquers, and varnishes – white enamel varnish 0.91
Porcelain – glazed 0.92
Quartz – opaque 0.68 – 0.92
Roofing Paper 0.91
Tape – Masking 0.95
Water 0.95 – 0.96
Wood – planed oak 0.90

In Appendix A, we showed how emissivity is an important parameter in calculating the temperature of an object via infrared means. In this section we discuss how to determine a specific emissivity value. If you know the material of the object, use Table B- 1 in Appendix B to look up its approximate emissivity. Most organic materials such as plastics, cloth, or wood have an emissivity of about 0.95. For this reason, we use 0.95 as the default emissivity setting in the OS530 Series Thermometer.
For objects of unknown material or for very precise measurements, use one of the following methods to determine a specific emissivity value.

Figure C-1. Determining Emissivity

Appendix - Determining an Unknown Emissivity C
Drilled out Area 'A' Target
Figure C-2. Determining Emissivity with a Drilled Hole
Area 'B' Target


A
D
ac Adapter Input Jack 1-6 Active Operation 2-8 Aiming Sight “V Groove” 1-4 Alarms 2-14, 2-15 Alkaline Batteries 2-1, 4-1, 6-3 Ambient Target Temperature
Diagnostic Program 5-2 Differential Measurement 2-6 Display
Icons:
ATC 1-5 Backlighting 1-5 HAL 1-5 LAL 1-5 LCK 1-5 LOBAT 1-5 PRN 1-5 Modes:
Compensation 2-16 Analog Output Jack 1-6
Backlighting Icon 1-5 Battery(s):
AMB 1-5 AVG 1-5 dIF 1-5 E 1-5 HAL 1-5 LAL 1-5 MAX 1-5 MEM 1-5 MIN 1-5 PRN 1-5 Problems 5-1
Compartment 1-4, 2-1 Installing AA size 2-1 Life of 6-3 Blackbody A-2, C-3 Beeping Sound 2-14, 2-23, 4-1, 5-2, 6-3 C
Carrying Case 1-2 Chart Recorder Hookup 2-7 Computer:
Emissivities:
Parameters:
Adjusting 2-11 Definition A-2 Figuring out Unknown C-1 Values B-1 Erasing Temperature Data ... 2-24 Error Codes 5-2, 5-3
Speed, Data, Parity, Stop Bit 2-17, 2-27 Personal, Hookup 2-19 Program (IR_TEMPSOFT)
2-19, 2-26

K
Field of View:
Keypad, 4-position 1-4 Keys:
Diagrams 2-3, 2-4 Positions 2-2 Fixed Point Monitoring over
▼& °F-°C 1-4, 1-5 FUNC (Function) 1-4, 1-5 LOCK (Lock) 1-4, 1-5 ▲& ❍-● 1-4, 1-5 Key Strokes 7-1
Time Measurement 2-7
Gray Bodies (Objects) A-2
Label Layout:
High Alarm Value, setting ... 2-14
Inside Back Cover Laser Sight Module:
Icons:
Laser Circle 3-3 Laser Dot 3-3 Installing onto
ATC 1-5 Backlighting 1-5 HAL 1-5 LAL 1-5 LCK 1-5 LOBAT 1-5 PRN 1-5 Installing
Thermometer 3-4 Line of Sight 3-4 Power Button 2-2, 3-2 Problems 5-1 - 5-4 Power Indicator LED 2-2, 3-2 Removing from
AA Batteries 2-1 Laser Sight Module 3-3, 3-4
Thermometer 3-4, 3-5 Warnings and Cautions 3-1 LCD, Backlit 1-4 Lens Cleaning 4-2 Lines of Sight of the Module
ac Adapter Input 1-6 Analog Output 1-6 RS-232 Phone 1-6
and Thermometer 3-4 Lithium Batteries 2-1, 4-1, 6-3 Lock Function 2-11 Low Alarm Value, setting 2-15
Index I · Jacks

S
Main Display 1-4 Modes:
Serial Printer Hookup 2-18 Sleep Mode 2-5, 2-8,
2-12, 2-28, 5-1 Spectral Distribution A-2 Spot Measurement 2-6 Static Surface Scan 2-6 Stefan-Boltzmann Law A-3 Storing Temperature Data 2-23
Real Time 2-8 Recall 2-23, 2-25 Moving Surface Scan 2-7
Optics A-4
T
Parameters, reviewing 2-27 PAS Code 5-2 Passive Operation 2-25 Personal Computer
Temperature Data:
Erasing 2-24 Storing 2-23 Thermal Radiation A-1 Thermometer:
Hookup 2-19 Power Contacts:
Front View 1-4 Rear View 1-6 Tripod Thread Mount 1-4, 1-6
Cleaning 4-2 On Laser Sight Module .. 1-4, 4-2 Power Indicator LED 2-2, 3-2 Printer, Serial, Hookup 2-17
“V” Groove Aiming Sight 1-4
Real Time Mode Block
Diagram 2-8 Recall Mode Block Diagram ... 2-26 Replacing AA Batteries 4-1 RS-232 Phone Jack 1-6 Rubber Boot:
Wein’s Displacement Law A-3 Wrist Strap 1-4
Display 1-4 IR Lens 1-4
Index I

WARRANTY/DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 25 months from date of purchase on the base unit and 13 months from date of purchase on Laser Sight Module. OMEGA WARRANTY adds an additional one (1) month grace period to the normal product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unautho- rized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear are not warranted, include but are not limited to contact points, fuses, and triacs. OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages. CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the fol- lowing information available BEFORE contacting OMEGA: 1. P.O. number under which the product was PURCHASED, 2. Model and serial number of the product under warranty, and 3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following informa- tion available BEFORE contacting OMEGA: 1. P.O. number to cover the COST of the repair, 2. Model and serial number of product, and 3. Repair instructions and/or specific problems relative to the product.
PATENT NOTICE: U.S. PAT. D357,194, B1 5,368,392, 5,524,984, 5,727,880, 5,465,838 5,823,678, 5,823,679, 6,267,500B1, 6,123,453, 6,341,891B1/Canada 75811 D OMEGA ENGINEERING, INC., 2,116,055, 2,114,806/Czech Republic 25372/France 0378411 to 0378446, 2,767,921, 2 773 213 B1/Germany M 94 06 478.4, G 94 22 197.9, G 94 22 203.7/Italy RM940000913/Japan 988,378/Netherlands 1007752, 25009-00/Spain mod. ut. 133292/Slovak Republic 24565/U.K. Registered 2041153, 9726133.3, EPO 0 644408, EP 1 085 307 A1 Other U.S. and International Patents Pending. OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2006 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
LASER RADIATION
IS EMITTED FROM
THIS APERTURE
® OMEGASCOPE CAUTION
LASER RADIATION - DO NOT STARE INTO BEAM
OUTPUT
Label Location - refer to Section 3.2 Warnings and Cautions - refer to Section 3.1

Process Measurement and Control?
OMEGA…Of Course! Shop online at omega.com
TEMPERATURE Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies Wire: Thermocouple, RTD & Thermistor Calibrators & Ice Point References Recorders, Controllers & Process Monitors Infrared Pyrometers
PRESSURE, STRAIN AND FORCE Transducers & Strain Gages Load Cells & Pressure Gages Displacement Transducers Instrumentation & Accessories
FLOW/LEVEL Rotameters, Gas Mass Flowmeters & Flow Computers Air Velocity Indicators Turbine/Paddlewheel Systems Totalizers & Batch Controllers
pH/CONDUCTIVITY pH Electrodes, Testers & Accessories Benchtop/Laboratory Meters Controllers, Calibrators, Simulators & Pumps Industrial pH & Conductivity Equipment
DATA ACQUISITION Data Acquisition & Engineering Software Communications-Based Acquisition Systems Plug-in Cards for Apple, IBM & Compatibles Datalogging Systems Recorders, Printers & Plotters
HEATERS Heating Cable Cartridge & Strip Heaters Immersion & Band Heaters Flexible Heaters Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL Metering & Control Instrumentation Refractometers Pumps & Tubing Air, Soil & Water Monitors Industrial Water & Wastewater Treatment pH, Conductivity & Dissolved Oxygen Instruments
Pages referenced in this manual — manufacturer, support and downloads.
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.