Instruction/ maintenance manual of the product MW600 Milwaukee
Go to page of 12
Milwa uk ee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter.
PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the meter. Install the battery into the battery clip connector while observing polarity. D.O. probes are shipped dry. Assemble the probe and prepare it for use, connect it to the meter and proceed as follows.
PROBE POLARIZATION: • A dissolved oxygen probe must be polarized to function properly. • To polarize the probe, the fully assambled probe must be connected to the meter and the meter must be on.
Zero Calibration: • Dip the probe into MA9070 zero oxygen solution and stir gently for 2-3 minutes. • Wait for the probe to reach it’s lowest stable reading. • Adjust the zero D.O. calibration trimmer (located on the front panel) until the display reads "0.
TAKING MEASUREMENTS: Verify the probe is polarized and the probe and meter have been calibrated. Remove the protective cap from probe. Immerse the tip of the probe in the sample to be tested. For accurate dissolved oxygen measurements a minimum water movement of 0.
5 °C 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 0 m 14.6 13.8 13.1 12.4 11.8 11.3 10.8 10.3 9.9 9.5 9.1 8.7 8.4 8.1 7.8 7.5 7.3 7.1 6.8 6.6 6.4 300 m 14.1 13.3 12.7 12.0 11.4 10.9 10.4 9.9 9.7 9.2 8.8 8.4 8.1 7.8 7.5 7.2 7.1 6.9 6.6 6.
6 SALINITY COMPENSATION: The t able below illustrates the change in the solubility of oxygen in air saturated water as a result of chloride concentration or salinity .
7 PROBE & MEMBRANE MAINTENANCE: The D.O. probe body is made of reinforced plastic for maximum durability. A thermistor temperature sensor provides temperature measurements of the tested sample. It is always recommended to keep the protective cap on the probe when not in use, to protect the membrane against damage and dirt.
The Platinum cathode should always be bright and untarnished. If it is tarnished or stained, due to contact with certain gases or a damaged membrane cap, the cathode should be cleaned. You can use a clean lint-free cardboard or cloth. Rub the cathode very gently side to side 4-5 times.
BATTERY REPLACEMENT: When the battery becomes weak the meter will display the low battery indicator " ". When this appears, only a few hours of battery life remain. A low battery will result in unreliable measurements. Prompt battery replacement is required.
SPECIFICATIONS: MW600 RANGE 0.0 to 19.9 mg/L RESOLUTION 0.1 mg/L ACCURACY (@25°C) ±1.5% Full Scale TEMPERATURE Automatic from 0 to 30°C COMPENSATION CALIBRATION Manual on 2 points (zero and slope) LCD 3½ digits with symbols PROBE MA840 (included) ENVIRONMENT 0 to 50°C, 95% RH max.
ISTMW600 01/10 Milwa uk ee Authorized Dealer:.
An important point after buying a device Milwaukee MW600 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Milwaukee MW600 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Milwaukee MW600 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Milwaukee MW600 you will learn all the available features of the product, as well as information on its operation. The information that you get Milwaukee MW600 will certainly help you make a decision on the purchase.
If you already are a holder of Milwaukee MW600, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Milwaukee MW600.
However, one of the most important roles played by the user manual is to help in solving problems with Milwaukee MW600. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Milwaukee MW600 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center