Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

3 SpotOn ™ Temperature Monitoring System Model 370 Operator’s Manual

  • 1 3M ™ SpotOn ™ Temperature Monitoring System
  • Model 370
  • Operator’s Manual

2 Système de surveillance de la température 3M ™ SpotOn ™

  • Modèle 370
  • Manuel d ’ utilisation
  • 3 3M ™ SpotOn ™ Temperaturüberwachungssystem
  • Modell 370
  • Bedienerhandbuch
  • 4 3M ™ SpotOn ™ Sistema di monitoraggio della temperatura
  • Modello 370
  • Manuale dell’operatore
  • 5 3M ™ SpotOn ™ Sistema de control de temperatura
  • Modelo 370
  • Manual del operador
  • 6 3M ™ SpotOn ™ -temperatuurcontrolesysteem
  • Model 370
  • Gebruikershandleiding
  • 7 3M ™ SpotOn ™ Temperaturövervakningssystem
  • Modell 370
  • Bruksanvisning 8 3M ™ SpotOn ™ temperaturovervågningssystem
  • Model 370
  • Brugervejledning
  • 9 3M ™ SpotOn ™ temperaturovervåkingssystem
  • modell 370
  • Brukerhåndbok
  • 0 3M ™ SpotOn ™ -lämpötilanseurantajärjestelmä
  • Malli 370
  • Käyttöopas
  • - Sistema de Monitorização da Temperatura 3M ™ SpotOn ™
  • Modelo 370
  • Manual do utilizador

3M

™ SpotOn

™

Temperature Monitoring System

Page 2 of the 3M 370 user manual — 370096 A New Revision - First Release All May 2012
Page 2 · click to enlarge
Page 2

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Revision History

3M 370 — Revision History diagram (page 2)
Fig. 1 · Revision History

370096 A New Revision - First Release All May 2012

Page 3 of the 3M 370 user manual — Operator’s Manual Model 370 3M
Page 3 · click to enlarge
Page 3

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Operator’s Manual Model 370 3M

™ SpotOn

Temperature Monitoring System

English 5

Français 23

Deutsch 47

Italiano 71

Español 95

Nederland 119

Svenska 143

Dansk 167

Norsk 191

Suomenkielinen 215

Português 239

™

Page 4 of the 3M 370 user manual
Page 4 · click to enlarge
Page 4

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Page 5 of the 3M 370 user manual — ™
Page 5 · click to enlarge
Page 5

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

English

  • 1 3M ™ SpotOn ™ Temperature Monitoring System
  • Model 370
  • Operator’s Manual

3M

™ SpotOn

™

Temperature Monitoring System

Page 6 of the 3M 370 user manual
Page 6 · click to enlarge
Page 6

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Page 7 of the 3M 370 user manual —  1
Page 7 · click to enlarge
Page 7

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

1

Table of Contents

Section 1: Introduction 3

About this manual 3 Intended use 3 Safety information and symbols 3 Proper use and maintenance 6 Read before servicing equipment 6

Section 2: Overview and Operation 7

Description of the SpotOn temperature monitoring system 7

How does it work? 8

SpotOn control unit 8 SpotOn temperature sensor 8 Back of Control Unit 9 Front of Control Unit 9 Setting up the SpotOn control unit 9 Instructions for use 10 Transferring the patient and sensor 11 Removing the sensor 11 To change temperature unit display (Celsius or Fahrenheit) 12 Shutting down the control unit 12

Section 3: Display Panel 13 Display screens 13 Start-up 13 Standby 13 Ready 14 Equilibration 14 Running 14 Control unit error 14 Sensor error 14

Section 4: Troubleshooting 15 On Mode 15 Displays 15 Errors 16

Section 5: General Maintenance, Storage and Service 17 Cleaning and Disinfecting Procedure 17 Cleaning the control unit and cable 17 Disinfecting the sensor cable 17 Storage 17 Service 17

Page 9 of the 3M 370 user manual — Defibrillation-proof type CF applied part
Page 9 · click to enlarge
Page 9

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Introduction 3

Section 1: Introduction

About this manual 3

Intended use 3

Safety information and symbols 3

Explanation of signal word consequences 4

Proper use and maintenance 6

Read before servicing equipment 6

About this manual This Operator’s Manual describes the setup, use, and maintenance of the 3M™ SpotOn™ temperature monitoring system. The SpotOn temperature monitoring system and the manual are to be used by healthcare professionals in clinical environments only. Read and follow all instructions, labeling and accompanying documents supplied with this SpotOn temperature monitoring system. Failure to follow instructions could lead to misuse of the device, device malfunction, or patient injury.

Intended use Measure, monitor, and trend body temperature of adult and pediatric patients.

Safety information and symbols

Date of manufacture

Manufacturer

Single use only

Container quantity

CAUTION

This system is subject to the European WEEE Directive 2002/96/EC. This product contains electrical and electronic components and must not be disposed of using standard refuse collection. Please consult local directives for disposal of electrical and electronic equipment.

Defibrillation-proof type CF applied part

Page 10 of the 3M 370 user manual — NOTICE: Indicates a situation which, if not avoided, could result in property
Page 10 · click to enlarge
Page 10

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

4 Introduction

Class II equipment

Authorized Representative in the European Community

CAUTION: Recycle to avoid environmental contamination This product contains recyclable parts. For information on recycling - please contact your nearest 3M Service Center for advice.

Direct current

See accompanying documents

Follow instructions for use

Keep dry

Hardware temperature limits

Sensor temperature limits

Explanation of signal word consequences

WARNING: Indicates a hazardous situation which, if not avoided, could result in death or serious injury.

CAUTION: Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.

NOTICE: Indicates a situation which, if not avoided, could result in property

damage only.

Page 11 of the 3M 370 user manual — 8. To reduce the risks associated with entanglement:
Page 11 · click to enlarge
Page 11

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Introduction 5

WARNINGS

1. To reduce the risks associated with access to critical patient information or equipment:
  • The SpotOn control unit is to be attached to other equipment by authorized service personnel only.
  • Do not re-install or re-locate the SpotOn control unit.
2. To reduce the risks associated with hazardous voltage, fire, and thermal energy hazards:
  • Use ONLY with 3M SpotOn system components (sensors, cables, and power supply). Do not substitute other devices for the SpotOn control unit, sensor, cables, or power supply.
  • Do not use the SpotOn temperature monitoring system (sensor, control unit, cables, or cords) in an MRI environment.

3. To reduce the risks associated with hazardous voltage and fire:

  • Keep power supply visible and accessible at all times. The plug on the power supply serves as the disconnect device. The power outlet shall be as close as practical and shall be easily accessible.
  • Use only the power supply specified for this product and certified for the country of use.
  • Use only a properly grounded power outlet; do not use extension cords or multiple portable socket outlets.
  • Do not allow the power supply to get wet.
  • Do not use the SpotOn control unit when it appears the unit or system components are damaged. Contact 3M Patient Warming technical support at 1-800-733-7775 or 952-9471200.
  • Do not service or modify the power supply, control unit, cables, sensor, or any part of the 3M SpotOn system. There are no user serviceable parts.

4. To reduce the risks associated with fire:

  • This system is not suitable for use in the presence of a flammable anesthetic mixture with air nitrous oxide.
5. To reduce the risks associate with pressure, and crushed or entangled cords and connectors:
  • Do not allow the patient to lie on any SpotOn sensor cable or connector.
  • Always position cables and cords away from the patient’s body.
  • Do not use a headband or other device to secure the SpotOn sensor to the patient.

6. To reduce the risks associated with incorrect system or sensor use:

  • Limit use of the SpotOn temperature sensor to 24 hours. Extended use may compromise skin, cause degradation of material or performance.
  • Avoid exposing the SpotOn temperature sensor, sensor cable, power supply, and control unit to surgical skin surface preparation solutions or other fluids.
  • Use an additional independent thermometer to measure body temperature during intentional hyperthermia or hypothermia therapy.
  • Confirm unanticipated temperature reading with an independent thermometer, if needed.
  • Do not reposition the SpotOn temperature sensor; repositioning may weaken the sensor adhesive, damage the sensor, or compromise the device performance.
  • Avoid placing the sensor in the center of the forehead as this may affect sensor accuracy.

7. To reduce the risks associated with exposure to biohazards:

  • Follow facilities policies and procedures for disposal of contaminated materials.
  • Always perform the decontamination procedure prior to returning the SpotOn temperature monitoring system for service and prior to disposal.

8. To reduce the risks associated with entanglement:

  • Do not leave pediatric patients unattended while using the SpotOn temperature monitoring system.
Page 12 of the 3M 370 user manual — 3. To avoid SpotOn temperature monitoring system damage that may impact performance:
Page 12 · click to enlarge
Page 12

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

6 Introduction

CAUTIONS 1. To reduce the risks associated with skin maceration:

  • Do not use the sensor on damaged or compromised skin.

2. To reduce the risks associated with cross-contamination:

  • Clean the sensor cable before connecting to a new sensor.
3. To reduce the risks associated with impact and facility medical device damage:
  • Do not use the SpotOn control unit as a handle to transport or to move the device to which it is attached.
  • Do not initiate temperature monitoring unless the SpotOn control unit is safely placed on a hard, flat surface or is securely mounted.

4. To reduce the risks associated with environmental contamination:

  • Follow applicable regulations when disposing of this device or any of its electronic components. NOTICES
1. The SpotOn temperature monitoring system meets medical electronic interference requirements. If radio frequency interference with other equipment should occur, connect the unit to a different power source.
2. Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.

3. To avoid SpotOn temperature monitoring system damage that may impact performance:

  • Do not store the SpotOn control unit and system components in a wet or damp place.
  • Do not spray cleaning solutions onto the control unit or into the sensor cable connector.
  • Do not immerse the SpotOn temperature monitoring control unit or system components in any liquid or subject them to any sterilization process.
  • Do not use solvents such as acetone or thinner to clean the control unit; avoid abrasive cleaners.
  • Clean control unit exterior with a damp soft cloth and 70% mixture of isopropyl alcohol and water. For full cleaning procedures see “Section 5: General Maintenance, Storage and Service” on page 17.
  • Do not immerse any of the devices or use a dripping wet cloth for cleaning.
4. The SpotOn temperature monitoring sensor is not made with natural latex material.
5. To the full extent permitted by law, the manufacturer and/or importer declines all responsibility for injury resulting from the unit being used in conjunction with unapproved system components. Proper use and maintenance Arizant Healthcare Inc., a 3M company assumes no responsibility for the reliability, performance, or safety of the temperature monitoring system if the following events occur:
  • Modifications or repairs are performed by unqualified personnel.
  • The unit is used in a manner other than that described in the Operator’s Manual.
  • The unit is installed in an environment that does not meet the appropriate electrical requirements.

Read before servicing equipment • All repair, calibration, and servicing of this equipment must be performed by 3M Patient Warming.

  • There are no user serviceable parts inside the equipment or power supply.
Page 13 of the 3M 370 user manual — Figure 2: SpotOn Temperature Monitoring System
Page 13 · click to enlarge
Page 13

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Overview and Operation 7

Section 2: Overview and Operation

Description of the SpotOn temperature monitoring system 7

How does it work? 8

Setting up the SpotOn control unit 9

Instructions for use 10

Transferring the patient and sensor 11

Removing the sensor 11

To change temperature unit display (Celsius to Fahrenheit) 12

Description of the SpotOn temperature monitoring system The SpotOn temperature monitoring system measures the core temperature of the patient. The system is comprised of the SpotOn temperature sensor, which connects to the SpotOn control unit through the sensor cable, a power supply, and an optional monitor cable. For additional sensors, sensor cable, power supply, control unit stand, monitor cable, hook-and-loop, or foam adhesive, contact 3M Patient Warming.

3M 370 — Overview and Operation 7 diagram (page 13)
Fig. 2 · Overview and Operation 7
3M 370 — Overview and Operation 7 diagram (page 13)
Fig. 3 · Overview and Operation 7

Power Supply

3M ™ SpotOn ™ Control Unit

3M 370 — Power Supply diagram (page 13)
Fig. 4 · Power Supply
3M 370 — Power Supply diagram (page 13)
Fig. 5 · Power Supply
3M 370 — Power Supply diagram (page 13)
Fig. 6 · Power Supply
3M 370 — Power Supply diagram (page 13)
Fig. 7 · Power Supply
3M 370 — Power Supply diagram (page 13)
Fig. 8 · Power Supply

Sensor Cable

Temperature Sensor

Sensor Connector

Figure 2: SpotOn Temperature Monitoring System

Page 14 of the 3M 370 user manual — 3) Equilibrated - isothermal zone established
Page 14 · click to enlarge
Page 14

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

8 Overview and Operation

How does it work? The SpotOn temperature monitoring system uses zero-heat-flux thermometry to accurately measure a person’s core temperature as characterized in Figure 3 below. 1) The SpotOn temperature monitoring system gently warms the sensor creating an isothermal zone under the sensor. 2) Once equilibrated to the core temperature, skin-surface heat loss to the environment is prevented and a zero-heat-flux condition is established. 3) When the temperature sensor reaches equilibrium with the patient’s core temperature, the SpotOn control unit displays an accurate, noninvasive measurement of the patient’s core temperature.

3M 370 — 8 Overview and Operation diagram (page 14)
Fig. 9 · 8 Overview and Operation

Skin Surface

Temperature

Sensor

Deep Tissue

1) Placement and connection

3M 370 — Skin Surface diagram (page 14)
Fig. 10 · Skin Surface

2) Equilibration - isothermal zone development

3M 370 — 2) Equilibration - isothermal zone development diagram (page 14)
Fig. 11 · 2) Equilibration - isothermal zone development

3) Equilibrated - isothermal zone established

Figure 3: Isothermal tunnel formation with zero-heat-flux technology

SpotOn control unit The SpotOn control unit displays noninvasive core temperature measurements obtained from the SpotOn temperature sensor. The current temperature is displayed numerically, and previous temperature data is displayed graphically as a temperature trend graph. When the sensor is plugged into the sensor cable, the control unit reads up to the last two hours of stored temperature data and displays the temperature trend graph. Disconnections of the sensor from the sensor cable are indicated by gaps in the temperature trend graph. The current temperature measurement can be continuously transmitted to a patient vital signs monitor through the optional SpotOn monitor cable using a standard YSI-400 type input.

SpotOn temperature sensor The SpotOn sensor is constructed of two layers of medical grade foam and a flexible circuit which contains a resistive warming circuit, two calibrated thermistors, and nonvolatile memory. The information that generates the temperature trend graph is stored on the SpotOn sensor. If the patient is transferred from one location to another, the SpotOn sensor is disconnected from the SpotOn sensor cable so the sensor is left in place on the patient. The SpotOn sensor may be reconnected to a different SpotOn sensor cable and control unit when the patient arrives at the next point of care. Once the SpotOn sensor is reconnected to the SpotOn sensor cable, the system will begin to reequilibrate and will regenerate the stored temperature trend graph and display the patient’s current temperature. The act of disconnecting and reconnecting the SpotOn sensor from the sensor cable is indicated by a gap in the temperature trend graph.

Page 15 of the 3M 370 user manual — Figure 6: Front of Control Unit
Page 15 · click to enlarge
Page 15

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Overview and Operation 9

Back of Control Unit To view information and connections on the back of the control unit, assure that controller is viewed from a distance no further than 15” (38 cm) from the operator.

3M 370 — Overview and Operation 9 diagram (page 15)
Fig. 12 · Overview and Operation 9
3M 370 — Overview and Operation 9 diagram (page 15)
Fig. 13 · Overview and Operation 9

Stand Knob

Digital Port (used by 3M personnel only)

°C/°F Temperature

Display Button

Monitor Cable Port

Power Input Port

Figure 5: Back of Control Unit

The digital port is used only by 3M Patient Warming personnel for testing purposes and to access variables within the control unit. The monitor cable port provides a resistance that corresponds to a YSI-400 thermistor at the displayed temperature. It is electrically isolated from the SpotOn control unit.

Front of Control Unit

Figure 6: Front of Control Unit

Setting up the SpotOn control unit

1. The SpotOn control unit must be properly installed and mounted by qualified personnel. NOTE: The SpotOn sensor and control unit are suitable for use in the patient environment.
2. Connect the power supply to the back of the SpotOn control unit (see Figure 5: Back of Control Unit). Plug the power supply into an appropriate outlet. The standby screen will illuminate.
3. Connect the SpotOn sensor cable to the front of the control unit. See Figure 6: Front of Control Unit for proper orientation and connection.
4. If desired, use the provided monitor cable to connect the SpotOn control unit to the patient monitor (see Figure 5: Back of Control Unit). NOTE: Temperature input on patient monitor typically labelled T1 or T2.
Page 16 of the 3M 370 user manual — Figure 9: Apply the SpotOn sensor to the
Page 16 · click to enlarge
Page 16

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

10 Overview and Operation

Instructions for use

1. Ensure the SpotOn system power supply is connected to the back of the control unit and the power supply is plugged into an appropriate outlet. If desired, ensure the provided monitor cable is connected to the back of the SpotOn control unit and to the YSI-400 type input on the patient monitor.
3M 370 — 10 Overview and Operation diagram (page 16)
Fig. 14 · 10 Overview and Operation
2. Use an alcohol wipe to clean and disinfect the patient’s forehead above the orbital ridge. Allow the patient’s forehead to completely dry.
3. Connect the SpotOn temperature sensor to the sensor cable connector. Ensure the sensor tab is properly connected and is fully inserted into the sensor cable (Figure 7). The display on the control unit will change to the ready screen (See “Ready” on page 14) once it confirms the sensor connection.

Figure 7: Connect the sensor to the

sensor cable

3M 370 — Figure 7: Connect the sensor to the diagram (page 16)
Fig. 15 · Figure 7: Connect the sensor to the
4. Remove backing liners on the SpotOn temperature sensor to expose adhesive (Figure 8).

Figure 8: Remove the backing liners

3M 370 — Figure 8: Remove the backing liners diagram (page 16)
Fig. 16 · Figure 8: Remove the backing liners
5. Position the sensor cable away from the patient’s face and apply the SpotOn temperature sensor to the patient’s forehead above the orbital ridge (Figure 9).

Figure 9: Apply the SpotOn sensor to the

patient’s forehead

Page 17 of the 3M 370 user manual — Figure 10: Remove the sensor
Page 17 · click to enlarge
Page 17

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Overview and Operation 11

6. Gently press the edge of the sensor to ensure good adhesion with the skin. OPTIONAL: Support the SpotOn temperature sensor cable with the provided sensor cable clip.
7. The control unit will sense patient attachment after several seconds and the control unit display will change to the equilibration screen (see “Equilibration” on page 14).
8. The patient’s temperature will display on the patient monitor once equilibrium is reached (approximately three minutes).

Transferring the patient and sensor

1. Prior to moving the patient, disconnect the sensor cable from the SpotOn sensor. The sensor may be left on the patient and reconnected to another SpotOn system once patient transfer is complete. NOTE: The SpotOn temperature sensor contains integrated memory that captures up to two hours of temperature data in five minute intervals. The SpotOn sensor will retain these temperatures and history.
2. Carefully reconnect a SpotOn sensor cable and control unit to the temperature sensor once transfer is complete.
3. The control unit will sense patient attachment after several seconds and the control unit display will change to the equilibration screen (see “Equilibration” on page 14).
4. The patient’s temperature will display on the patient monitor once equilibrium is reached (approximately three minutes).

Removing the sensor

1. Disconnect the sensor cable from the SpotOn temperature sensor. Disconnect the sensor cable clip from the bed or sheet, if used.
2. Starting at the edge of the SpotOn sensor, gently remove the SpotOn temperature sensor from the patient (Figure 10). NOTE: Do not use the sensor tab to remove the SpotOn temperature sensor from the patient. NOTE: If needed, use an alcohol swab along the sensor edge to help remove the sensor from the patient.
3M 370 — Removing the sensor diagram (page 17)
Fig. 17 · Removing the sensor
3M 370 — Removing the sensor diagram (page 17)
Fig. 18 · Removing the sensor
3M 370 — Removing the sensor diagram (page 17)
Fig. 19 · Removing the sensor
3M 370 — Removing the sensor diagram (page 17)
Fig. 20 · Removing the sensor

Sensor

Tab

3. Discard the SpotOn sensor and clean the sensor cable. (See Section 5: General Maintenance for details).

Figure 10: Remove the sensor

Page 18 of the 3M 370 user manual — °C/°F Temperature
Page 18 · click to enlarge
Page 18

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

12 Overview and Operation

To change temperature unit display (Celsius or Fahrenheit)

The SpotOn control unit can display temperature measurements in either Celsius or Fahrenheit. The unit will be factory preset to Celsius.

1. Plug the SpotOn control unit into an appropriate outlet.
2. Push the temperature display button on the back of the control unit to change to the desired measurement scale. NOTE: The trend display is always shown in Celsius.
3M 370 — 12 Overview and Operation diagram (page 18)
Fig. 21 · 12 Overview and Operation
3M 370 — 12 Overview and Operation diagram (page 18)
Fig. 22 · 12 Overview and Operation

°C/°F Temperature

Display Button

Figure 11: Temperature Display Button

Shutting down the control unit

The plug on the power supply serves as the disconnect device. Turn the control unit on and off by plugging or unplugging the power supply from the wall outlet. The power outlet shall be as close as practical and shall be easily accessible.

Page 19 of the 3M 370 user manual — ™
Page 19 · click to enlarge
Page 19

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Display Panel 13

Section 3: Display Panel

Display screens 13

Start-up 13

Standby 13

Ready 14

Equilibration 14

Running 14

Control unit error 14

Sensor error 14

Display screens

Start-up The start-up screen displays for approximately five seconds when the control unit is initially powered ON. The revision number of the control unit software is also displayed.

FW-x.x.xxx

Standby The SpotOn system graphic is displayed when the unit is not in use.

3M ™ SpotOn

™

Temperature Monitoring System

Page 20 of the 3M 370 user manual — E-xx
Page 20 · click to enlarge
Page 20

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

14 Display Panel

Ready The ready screen displays the serial number of the SpotOn sensor and indicates the SpotOn sensor cable and sensor are properly attached to the SpotOn control unit. The SpotOn sensor is now ready to be attached to the patient.

SN - 000000000

Equilibration A flashing yellow temperature display indicates the sensor has been connected to the patient and to the control unit and is in the process of equilibration. The graph below the temperature output indicates the progress of equilibration. This process will take approximately three minutes.

36 9 36 0

39 37 35

If you see a temperature in the upper left hand corner of the screen, that is the last recorded temperature that has been captured by the sensor on the patient’s head. The sensor captures and records a temperature every five minutes.

After the equilibration is complete, the patient’s body temperature is displayed on the SpotOn control unit in bold white numbers and can be automatically transmitted to the patient monitor via the monitor cable.

Running The running screen displays the deep tissue temperature of the patient in white. The trend graph at the bottom of the screen displays the last two hours of the patient’s temperature in five minute increments. The bars on the graph will begin filling from the right with the 36°C always visible. The blue bars indicate that the temperature dropped below 36°C (white indicates the temperature is above 36°C).

3M 370 — 39 37 35 diagram (page 20)
Fig. 23

Control unit error The SpotOn control unit error screen displays when there is a system error. Discontinue use of unit. Contact a biomedical technician.

E-xx

Sensor error The sensor error screen displays when there is a temperature sensor or cable error. See “Section 4: Troubleshooting” on page 15 for details.

E-xx

Page 21 of the 3M 370 user manual — 36 9 36 0
Page 21 · click to enlarge
Page 21

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Troubleshooting 15

Section 4: Troubleshooting

On Mode 15

Displays 15

Errors 16

The following conditions are listed in the order of which troubleshooting actions should be performed.

On Mode

Condition Cause Action

Unit does not power up. Unit is not plugged in, or power supply is not plugged into an appropriate outlet.

Make sure the power supply is plugged into the control unit. Make sure the control unit is plugged into an appropriate power outlet.

Unit failure. Contact 3M Patient Warming technical service.

Displays

Condition Cause Action

“Ready” screen continues to display while sensor is adhered to the patient.

Sensor not properly adhered to the patient.

Make sure the sensor is connected to the patient.

Low patient temperature (below 29°C).

Press and hold the °C/°F button on the back of the control unit for five seconds to force the control unit into equilibration mode. The patient’s temperature will then display.

Control unit failure. Discontinue use of unit. Contact 3M Patient Warming technical service.

SN - 000000000

“Running” screen does not appear on the control unit.

Control unit may still be equilibrating. Wait until the SpotOn control unit has finished equilibration. Continue use.

Control unit failure. Contact 3M Patient Warming technical service.

Patient temperature does not appear on the patient monitor.

Control unit may still be equilibrating. Wait until the SpotOn control unit has finished equilibration. Continue use.

36 9 36 0

Monitor cable is not plugged in. Make sure the monitor cable is plugged into the control unit and patient monitor.

  • Cable failure.
  • Replace monitor cable.
  • Contact 3M Patient Warming technical service.

Control unit failure. Contact 3M Patient Warming technical service after use.

The patient’s temperature output on the control unit and patient monitor does not read the same value.

  • Cable failure.
  • Replace monitor cable.
  • Contact 3M Patient Warming technical service.

39 37 35

Page 22 of the 3M 370 user manual — 16 Troubleshooting
Page 22 · click to enlarge
Page 22

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

16 Troubleshooting

Errors

Error Screen Condition Cause Action

E-1 through E-16, E-18 through E-99

Sensor not properly connected to the sensor cable.

Make sure the sensor is connected to the sensor cable.

Sensor failure. Replace sensor.

  • Cable failure.
  • Replace sensor cable.
  • Contact 3M Patient Warming technical service.

E-17 Sensor failure. Replace sensor.

  • Cable failure.
  • Replace sensor cable.
  • Contact 3M Patient Warming technical service.

Control unit failure. Discontinue use of the unit. Contact a biomedical technician.

E-100 and up Control unit failure. Discontinue use of the unit. Contact a biomedical technician.

E-xx · E-xx · E-xx

Page 23 of the 3M 370 user manual — Disinfecting the sensor cable
Page 23 · click to enlarge
Page 23

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

General Maintenance 17

Section 5: General Maintenance, Storage and Service

Cleaning and Disinfecting Procedure 17

Storage 17

Service 17

Cleaning and Disinfecting Procedure CAUTION:

  • Do not immerse any of the devices or use a dripping wet cloth for cleaning. Moisture may seep inside the device and damage the electrical components and lead to incorrect temperature reporting.
  • Do not spray cleaning solutions onto the control unit or into the sensor cable connector. Damage to the control unit or sensor cable connector may occur.

Cleaning the control unit and cable

Clean the SpotOn control unit on an as-needed basis or per facility policies and procedures for cleaning electronic equipment. Clean the sensor cable between each use.

1. Disconnect power supply from power outlet.
2. Use a slightly damp soft cloth moistened with a mild, nonabrasive cleaning solution to clean the device surfaces, cords, and cables. Avoid getting liquid into electronic ports.
3. Dry with a separate soft cloth.

Disinfecting the sensor cable

1. Clean the devices as described above.
2. Wipe down the control unit and cable using a damp soft cloth and 70% isopropyl alcohol solution. Avoid getting liquid into electronic ports.
3. Dry with a separate soft cloth.

Storage Notice: Do not store the SpotOn control unit and system components in a wet or damp place. Damage to the electrical components may occur.

Store all SpotOn components at room temperature and in a dry place when not in use.

Service All service must be performed by 3M Patient Warming or an authorized service technician. Call 3M Patient Warming Customer Service at 1-800-733-7775 for service information. Testing of the SpotOn system can be performed by 3M Patient Warming or call Customer Service for information about specific test procedures.

Page 24 of the 3M 370 user manual — 18 General Maintenance
Page 24 · click to enlarge
Page 24

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

18 General Maintenance

Page 25 of the 3M 370 user manual — FAX: +1-952-947-1400 +1-800-775-0002
Page 25 · click to enlarge
Page 25

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Technical Service and Order Placement 19

Section 6: Technical Service and Order Placement

When you call for technical support 19

Technical service and order placement 19

When you call for technical support Remember, we will need to know the serial number, software number and error code of your SpotOn control unit when you call us. The serial number label is located on the back of the control unit. (The firmware number displays on the control unit when the control unit is connected to a power outlet.)

Technical service and order placement

USA TEL: +1-952-947-1200 +1-800-733-7775

FAX: +1-952-947-1400 +1-800-775-0002

Outside of the USA Contact your local 3M Patient Warming representative.

Page 26 of the 3M 370 user manual — 20 Technical Service and Order Placement
Page 26 · click to enlarge
Page 26

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

20 Technical Service and Order Placement

Page 27 of the 3M 370 user manual — Accuracy 25°C to 43°C ± 0.2°C
Page 27 · click to enlarge
Page 27

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Specifications 21

Section 7: Specifications

Physical characteristics 21

Electrical characteristics 22

Temperature characteristics 22

Performance characteristics 22

Environmental conditions 22

Physical characteristics Dimensions of Control Unit 9.3 cm (3.7 in) high, extendable to 11.4 cm (4.5 in) high

7.1 cm (2.8 in) wide, 4.3 cm (1.7 in) deep

Weight of Control Unit 128 g (4.5 oz)

Dimensions of Sensor 4.1 cm (1.6 in) diameter, 0.5 cm (0.2 in) thick

Length of the Sensor Cable 400 cm (158 in)

Classification

MEDICAL—GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2, No. 60601-1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No.601-1:08; EN60601-1; Control No.4HZ8

Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class II, Type CF, Defibrillation-Proof, Ordinary Equipment. Conforms to EN12470. Not suitable for use in the presence of flammable anesthetic mixtures with air or nitrous oxide. Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only, in accordance with UL 60601-1, IEC 60601-1, and Canadian/CSA C22.2 No. 601.1. ANSI/AAMI ES 60601-1; 2005. EN60601-1. Classified under the Medical Device Directive as a Class IIb device.

Accuracy 25°C to 43°C ± 0.2°C

Sensor Material Medical grade foam and adhesive PET flexible circuit

Recommended Calibration Check Every 12 months

Sensor Shelf Life 36 months

Digital Port Serial output 0 [-12 to +12]

Monitor Cable Port Provides a resistance that corresponds to a YSI-400 thermistor at the displayed temperature. It is electrically isolated from the SpotOn control unit.

Page 28 of the 3M 370 user manual — Performance characteristics Equilibration time 3 minutes
Page 28 · click to enlarge
Page 28

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

22 Specifications

Electrical characteristics External power supply 100-240 VAC, 50-60 Hz Output; 5 VDC Class II, double insulated medical grade

Environmental conditions Ambient temperature range 10°C to 40°C (50°F to 104°F)

Storage and transport temperature range -20°C to 60°C (-4°F to 140°F)

Maximum heating power 2 W

Store all components at room temperature and in a dry place when not in use.

Temperature characteristics Over-temperature cutoff 43°C - Skin thermistor reading

Operating humidity 10 to 75% RH, noncondensing

Atmospheric pressure range 80 kPa to 106 kPa (12 PSI to 15 PSI)

44.5°C - Heater thermistor reading

Oxygen environment The sensor only is suitable for use in the presence of an oxygen enriched environment.

Performance characteristics Equilibration time 3 minutes

Time response Approximately three minutes

Measurement locations Patient’s lateral forehead above the orbital ridge

Measurement range 25°C - 43°C

Measurement readout Celsius or Fahrenheit

Leakage current CF classification. Meets leakage current requirements in accordance with UL 60601-1 and EN60601-1.

Page 29 of the 3M 370 user manual — Electrical characteristics
Page 29 · click to enlarge
Page 29

Made in the USA by 3M Health Care.

3M is a trademark of 3M Company, used under license in Canada. SPOTON is a trademark of Arizant Healthcare Inc., used under license in Canada. ©2013 Arizant Healthcare Inc. All rights reserved.

3M Deutschland GmbH, Health Care Business Carl-Schurz-Str. 1, 41453 Neuss, Germany

3M Health Care, 2510 Conway Ave., St. Paul, MN 55144 USA TEL 800-228-3957 | www.spotontemperature.com 202052A 04/13

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Unit does not power up.Unit is not plugged in, or power supply is not plugged into an appropriate outlet.Make sure the power supply is plugged into the control unit. Make sure the control unit is plugged into an appropriate power outlet. p.21
Unit does not power up.Unit failure.Contact 3M Patient Warming technical service. p.21
“Ready” screen continues to display while sensor is adhered to the patient. SN - 000000000Sensor not properly adhered to the patient.Make sure the sensor is connected to the patient. p.21
“Ready” screen continues to display while sensor is adhered to the patient. SN - 000000000Low patient temperature (below 29°C).Press and hold the °C/°F button on the back of the control unit for five seconds to force the control unit into equilibration mode. The patient’s temperature will then display. p.21
“Ready” screen continues to display while sensor is adhered to the patient. SN - 000000000Control unit failure.Discontinue use of unit. Contact 3M Patient Warming technical service. p.21
“Running” screen does not appear on the control unit.Control unit may still be equilibrating.Wait until the SpotOn control unit has finished equilibration. Continue use. p.21
“Running” screen does not appear on the control unit.Control unit failure.Contact 3M Patient Warming technical service. p.21
Patient temperature does not appear on the patient monitor. 36 9C 36 0 C 39 37 35Control unit may still be equilibrating.Wait until the SpotOn control unit has finished equilibration. Continue use. p.21
Patient temperature does not appear on the patient monitor. 36 9C 36 0 C 39 37 35Monitor cable is not plugged in.Make sure the monitor cable is plugged into the control unit and patient monitor. p.21
Patient temperature does not appear on the patient monitor. 36 9C 36 0 C 39 37 35Cable failure.• Replace monitor cable. • Contact 3M Patient Warming technical service. p.21
Patient temperature does not appear on the patient monitor. 36 9C 36 0 C 39 37 35Control unit failure.Contact 3M Patient Warming technical service after use. p.21
The patient’s temperature output on Cable failure. • Replace monitor cable. the control unit and patient monitor • Contact 3M Patient Warming technical does not read the same value. service. p.21

Ask about the 3M 370

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the 3M 370

Be the first to ask — other owners or our team will answer.

Original document

3M 370 manual PDF

29 pages·2.6 MB·Marine Instruments·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…