Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
™ SpotOn
Temperature Monitoring System

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A


Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
™ SpotOn
Temperature Monitoring System
English 5
Français 23
Deutsch 47
Italiano 71
Español 95
Nederland 119
Svenska 143
Dansk 167
Norsk 191
Suomenkielinen 215
Português 239
™

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
™ SpotOn
Temperature Monitoring System

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Table of Contents
Section 1: Introduction 3
About this manual 3 Intended use 3 Safety information and symbols 3 Proper use and maintenance 6 Read before servicing equipment 6
Section 2: Overview and Operation 7
Description of the SpotOn temperature monitoring system 7
How does it work? 8
SpotOn control unit 8 SpotOn temperature sensor 8 Back of Control Unit 9 Front of Control Unit 9 Setting up the SpotOn control unit 9 Instructions for use 10 Transferring the patient and sensor 11 Removing the sensor 11 To change temperature unit display (Celsius or Fahrenheit) 12 Shutting down the control unit 12
Section 3: Display Panel 13 Display screens 13 Start-up 13 Standby 13 Ready 14 Equilibration 14 Running 14 Control unit error 14 Sensor error 14
Section 4: Troubleshooting 15 On Mode 15 Displays 15 Errors 16
Section 5: General Maintenance, Storage and Service 17 Cleaning and Disinfecting Procedure 17 Cleaning the control unit and cable 17 Disinfecting the sensor cable 17 Storage 17 Service 17


Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Section 1: Introduction
About this manual 3
Intended use 3
Safety information and symbols 3
Explanation of signal word consequences 4
Proper use and maintenance 6
Read before servicing equipment 6
About this manual This Operator’s Manual describes the setup, use, and maintenance of the 3M™ SpotOn™ temperature monitoring system. The SpotOn temperature monitoring system and the manual are to be used by healthcare professionals in clinical environments only. Read and follow all instructions, labeling and accompanying documents supplied with this SpotOn temperature monitoring system. Failure to follow instructions could lead to misuse of the device, device malfunction, or patient injury.
Intended use Measure, monitor, and trend body temperature of adult and pediatric patients.
Manufacturer
Single use only
Container quantity
CAUTION
This system is subject to the European WEEE Directive 2002/96/EC. This product contains electrical and electronic components and must not be disposed of using standard refuse collection. Please consult local directives for disposal of electrical and electronic equipment.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Class II equipment
Authorized Representative in the European Community
CAUTION: Recycle to avoid environmental contamination This product contains recyclable parts. For information on recycling - please contact your nearest 3M Service Center for advice.
See accompanying documents
Follow instructions for use
Keep dry
Hardware temperature limits
Sensor temperature limits
Explanation of signal word consequences
WARNING: Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION: Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
damage only.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
WARNINGS

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
CAUTIONS 1. To reduce the risks associated with skin maceration:
Read before servicing equipment • All repair, calibration, and servicing of this equipment must be performed by 3M Patient Warming.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Section 2: Overview and Operation
Description of the SpotOn temperature monitoring system 7
How does it work? 8
Setting up the SpotOn control unit 9
Instructions for use 10
Transferring the patient and sensor 11
Removing the sensor 11
To change temperature unit display (Celsius to Fahrenheit) 12
Description of the SpotOn temperature monitoring system The SpotOn temperature monitoring system measures the core temperature of the patient. The system is comprised of the SpotOn temperature sensor, which connects to the SpotOn control unit through the sensor cable, a power supply, and an optional monitor cable. For additional sensors, sensor cable, power supply, control unit stand, monitor cable, hook-and-loop, or foam adhesive, contact 3M Patient Warming.


3M ™ SpotOn ™ Control Unit





Temperature Sensor
Sensor Connector

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
How does it work? The SpotOn temperature monitoring system uses zero-heat-flux thermometry to accurately measure a person’s core temperature as characterized in Figure 3 below. 1) The SpotOn temperature monitoring system gently warms the sensor creating an isothermal zone under the sensor. 2) Once equilibrated to the core temperature, skin-surface heat loss to the environment is prevented and a zero-heat-flux condition is established. 3) When the temperature sensor reaches equilibrium with the patient’s core temperature, the SpotOn control unit displays an accurate, noninvasive measurement of the patient’s core temperature.

Temperature
Sensor
Deep Tissue
1) Placement and connection


Figure 3: Isothermal tunnel formation with zero-heat-flux technology
SpotOn control unit The SpotOn control unit displays noninvasive core temperature measurements obtained from the SpotOn temperature sensor. The current temperature is displayed numerically, and previous temperature data is displayed graphically as a temperature trend graph. When the sensor is plugged into the sensor cable, the control unit reads up to the last two hours of stored temperature data and displays the temperature trend graph. Disconnections of the sensor from the sensor cable are indicated by gaps in the temperature trend graph. The current temperature measurement can be continuously transmitted to a patient vital signs monitor through the optional SpotOn monitor cable using a standard YSI-400 type input.
SpotOn temperature sensor The SpotOn sensor is constructed of two layers of medical grade foam and a flexible circuit which contains a resistive warming circuit, two calibrated thermistors, and nonvolatile memory. The information that generates the temperature trend graph is stored on the SpotOn sensor. If the patient is transferred from one location to another, the SpotOn sensor is disconnected from the SpotOn sensor cable so the sensor is left in place on the patient. The SpotOn sensor may be reconnected to a different SpotOn sensor cable and control unit when the patient arrives at the next point of care. Once the SpotOn sensor is reconnected to the SpotOn sensor cable, the system will begin to reequilibrate and will regenerate the stored temperature trend graph and display the patient’s current temperature. The act of disconnecting and reconnecting the SpotOn sensor from the sensor cable is indicated by a gap in the temperature trend graph.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Back of Control Unit To view information and connections on the back of the control unit, assure that controller is viewed from a distance no further than 15” (38 cm) from the operator.


Digital Port (used by 3M personnel only)
°C/°F Temperature
Display Button
Monitor Cable Port
Power Input Port
Figure 5: Back of Control Unit
The digital port is used only by 3M Patient Warming personnel for testing purposes and to access variables within the control unit. The monitor cable port provides a resistance that corresponds to a YSI-400 thermistor at the displayed temperature. It is electrically isolated from the SpotOn control unit.
Setting up the SpotOn control unit

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Instructions for use

sensor cable


patient’s forehead

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A




Tab

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
To change temperature unit display (Celsius or Fahrenheit)
The SpotOn control unit can display temperature measurements in either Celsius or Fahrenheit. The unit will be factory preset to Celsius.


Display Button
Figure 11: Temperature Display Button
Shutting down the control unit
The plug on the power supply serves as the disconnect device. Turn the control unit on and off by plugging or unplugging the power supply from the wall outlet. The power outlet shall be as close as practical and shall be easily accessible.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Section 3: Display Panel
Display screens 13
Start-up 13
Standby 13
Ready 14
Equilibration 14
Running 14
Control unit error 14
Sensor error 14
Start-up The start-up screen displays for approximately five seconds when the control unit is initially powered ON. The revision number of the control unit software is also displayed.
Standby The SpotOn system graphic is displayed when the unit is not in use.
Temperature Monitoring System

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Ready The ready screen displays the serial number of the SpotOn sensor and indicates the SpotOn sensor cable and sensor are properly attached to the SpotOn control unit. The SpotOn sensor is now ready to be attached to the patient.
Equilibration A flashing yellow temperature display indicates the sensor has been connected to the patient and to the control unit and is in the process of equilibration. The graph below the temperature output indicates the progress of equilibration. This process will take approximately three minutes.
If you see a temperature in the upper left hand corner of the screen, that is the last recorded temperature that has been captured by the sensor on the patient’s head. The sensor captures and records a temperature every five minutes.
After the equilibration is complete, the patient’s body temperature is displayed on the SpotOn control unit in bold white numbers and can be automatically transmitted to the patient monitor via the monitor cable.
Running The running screen displays the deep tissue temperature of the patient in white. The trend graph at the bottom of the screen displays the last two hours of the patient’s temperature in five minute increments. The bars on the graph will begin filling from the right with the 36°C always visible. The blue bars indicate that the temperature dropped below 36°C (white indicates the temperature is above 36°C).

Control unit error The SpotOn control unit error screen displays when there is a system error. Discontinue use of unit. Contact a biomedical technician.
Sensor error The sensor error screen displays when there is a temperature sensor or cable error. See “Section 4: Troubleshooting” on page 15 for details.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Section 4: Troubleshooting
On Mode 15
Displays 15
Errors 16
The following conditions are listed in the order of which troubleshooting actions should be performed.
Unit does not power up. Unit is not plugged in, or power supply is not plugged into an appropriate outlet.
Make sure the power supply is plugged into the control unit. Make sure the control unit is plugged into an appropriate power outlet.
Unit failure. Contact 3M Patient Warming technical service.
“Ready” screen continues to display while sensor is adhered to the patient.
Sensor not properly adhered to the patient.
Make sure the sensor is connected to the patient.
Low patient temperature (below 29°C).
Press and hold the °C/°F button on the back of the control unit for five seconds to force the control unit into equilibration mode. The patient’s temperature will then display.
Control unit failure. Discontinue use of unit. Contact 3M Patient Warming technical service.
“Running” screen does not appear on the control unit.
Control unit may still be equilibrating. Wait until the SpotOn control unit has finished equilibration. Continue use.
Control unit failure. Contact 3M Patient Warming technical service.
Patient temperature does not appear on the patient monitor.
Control unit may still be equilibrating. Wait until the SpotOn control unit has finished equilibration. Continue use.
Monitor cable is not plugged in. Make sure the monitor cable is plugged into the control unit and patient monitor.
Control unit failure. Contact 3M Patient Warming technical service after use.
The patient’s temperature output on the control unit and patient monitor does not read the same value.
39 37 35

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Errors
Error Screen Condition Cause Action
E-1 through E-16, E-18 through E-99
Sensor not properly connected to the sensor cable.
Make sure the sensor is connected to the sensor cable.
Sensor failure. Replace sensor.
E-17 Sensor failure. Replace sensor.
Control unit failure. Discontinue use of the unit. Contact a biomedical technician.
E-100 and up Control unit failure. Discontinue use of the unit. Contact a biomedical technician.
E-xx · E-xx · E-xx

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Section 5: General Maintenance, Storage and Service
Cleaning and Disinfecting Procedure 17
Storage 17
Service 17
Clean the SpotOn control unit on an as-needed basis or per facility policies and procedures for cleaning electronic equipment. Clean the sensor cable between each use.
Storage Notice: Do not store the SpotOn control unit and system components in a wet or damp place. Damage to the electrical components may occur.
Store all SpotOn components at room temperature and in a dry place when not in use.
Service All service must be performed by 3M Patient Warming or an authorized service technician. Call 3M Patient Warming Customer Service at 1-800-733-7775 for service information. Testing of the SpotOn system can be performed by 3M Patient Warming or call Customer Service for information about specific test procedures.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Section 6: Technical Service and Order Placement
When you call for technical support 19
Technical service and order placement 19
When you call for technical support Remember, we will need to know the serial number, software number and error code of your SpotOn control unit when you call us. The serial number label is located on the back of the control unit. (The firmware number displays on the control unit when the control unit is connected to a power outlet.)
USA TEL: +1-952-947-1200 +1-800-733-7775
Outside of the USA Contact your local 3M Patient Warming representative.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Section 7: Specifications
Physical characteristics 21
Electrical characteristics 22
Temperature characteristics 22
Performance characteristics 22
Environmental conditions 22
Physical characteristics Dimensions of Control Unit 9.3 cm (3.7 in) high, extendable to 11.4 cm (4.5 in) high
Weight of Control Unit 128 g (4.5 oz)
Dimensions of Sensor 4.1 cm (1.6 in) diameter, 0.5 cm (0.2 in) thick
Classification
MEDICAL—GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2, No. 60601-1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No.601-1:08; EN60601-1; Control No.4HZ8
Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class II, Type CF, Defibrillation-Proof, Ordinary Equipment. Conforms to EN12470. Not suitable for use in the presence of flammable anesthetic mixtures with air or nitrous oxide. Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only, in accordance with UL 60601-1, IEC 60601-1, and Canadian/CSA C22.2 No. 601.1. ANSI/AAMI ES 60601-1; 2005. EN60601-1. Classified under the Medical Device Directive as a Class IIb device.
Sensor Material Medical grade foam and adhesive PET flexible circuit
Recommended Calibration Check Every 12 months
Sensor Shelf Life 36 months
Digital Port Serial output 0 [-12 to +12]
Monitor Cable Port Provides a resistance that corresponds to a YSI-400 thermistor at the displayed temperature. It is electrically isolated from the SpotOn control unit.

Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Electrical characteristics External power supply 100-240 VAC, 50-60 Hz Output; 5 VDC Class II, double insulated medical grade
Storage and transport temperature range -20°C to 60°C (-4°F to 140°F)
Maximum heating power 2 W
Store all components at room temperature and in a dry place when not in use.
Operating humidity 10 to 75% RH, noncondensing
Atmospheric pressure range 80 kPa to 106 kPa (12 PSI to 15 PSI)
Oxygen environment The sensor only is suitable for use in the presence of an oxygen enriched environment.
Time response Approximately three minutes
Measurement locations Patient’s lateral forehead above the orbital ridge
Measurement range 25°C - 43°C
Measurement readout Celsius or Fahrenheit
Leakage current CF classification. Meets leakage current requirements in accordance with UL 60601-1 and EN60601-1.

Made in the USA by 3M Health Care.
3M is a trademark of 3M Company, used under license in Canada. SPOTON is a trademark of Arizant Healthcare Inc., used under license in Canada. ©2013 Arizant Healthcare Inc. All rights reserved.
3M Deutschland GmbH, Health Care Business Carl-Schurz-Str. 1, 41453 Neuss, Germany
3M Health Care, 2510 Conway Ave., St. Paul, MN 55144 USA TEL 800-228-3957 | www.spotontemperature.com 202052A 04/13
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Unit does not power up. | Unit is not plugged in, or power supply is not plugged into an appropriate outlet. | Make sure the power supply is plugged into the control unit. Make sure the control unit is plugged into an appropriate power outlet. | p.21 |
| Unit does not power up. | Unit failure. | Contact 3M Patient Warming technical service. | p.21 |
| “Ready” screen continues to display while sensor is adhered to the patient. SN - 000000000 | Sensor not properly adhered to the patient. | Make sure the sensor is connected to the patient. | p.21 |
| “Ready” screen continues to display while sensor is adhered to the patient. SN - 000000000 | Low patient temperature (below 29°C). | Press and hold the °C/°F button on the back of the control unit for five seconds to force the control unit into equilibration mode. The patient’s temperature will then display. | p.21 |
| “Ready” screen continues to display while sensor is adhered to the patient. SN - 000000000 | Control unit failure. | Discontinue use of unit. Contact 3M Patient Warming technical service. | p.21 |
| “Running” screen does not appear on the control unit. | Control unit may still be equilibrating. | Wait until the SpotOn control unit has finished equilibration. Continue use. | p.21 |
| “Running” screen does not appear on the control unit. | Control unit failure. | Contact 3M Patient Warming technical service. | p.21 |
| Patient temperature does not appear on the patient monitor. 36 9C 36 0 C 39 37 35 | Control unit may still be equilibrating. | Wait until the SpotOn control unit has finished equilibration. Continue use. | p.21 |
| Patient temperature does not appear on the patient monitor. 36 9C 36 0 C 39 37 35 | Monitor cable is not plugged in. | Make sure the monitor cable is plugged into the control unit and patient monitor. | p.21 |
| Patient temperature does not appear on the patient monitor. 36 9C 36 0 C 39 37 35 | Cable failure. | • Replace monitor cable. • Contact 3M Patient Warming technical service. | p.21 |
| Patient temperature does not appear on the patient monitor. 36 9C 36 0 C 39 37 35 | Control unit failure. | Contact 3M Patient Warming technical service after use. | p.21 |
| The patient’s temperature output on Cable failure. • Replace monitor cable. the control unit and patient monitor • Contact 3M Patient Warming technical does not read the same value. service. | p.21 |