VNA Master ™
Handheld Vector Network Analyzer + Spectrum Analyzer
MS2026C MS2036C MS2028C MS2038C 5 kHz to 6 GHz 5 kHz to 6 GHz 5 kHz to 20 GHz 5 kHz to 20 GHz Vector Network Analyzer 9 kHz to 9 GHz 9 kHz to 20 GHz + Spectrum Analyzer
The Ultimate Handheld Vector Network + Spectrum Analyzer for Cable, Antenna and Signal Analysis Anytime, Anywhere
High Performance Handheld S-Parameters
Anritsu introduces the MS202x/3xC VNA Master + Spectrum Analyzer, the industry’s broadest frequency handheld solution to address cable, antenna, component and signal analysis needs in the field: with frequency coverage from 5 kHz to 6/20 GHz. Equally impressive, this broadband measurement tool offers the industry’s first 12-term error correction algorithm in a truly handheld, battery-operated, rugged multi-function instrument. And now the MS2036/38C models include a powerful spectrum analyzer which multiplies user convenience by combining spectrum analysis with the VNA into a single measurement powerhouse for the harsh RF and physical environments of field test. Whether it is for spectrum monitoring, broadcast proofing, interference analysis, RF and microwave measurements, regulatory compliance, or 3G/4G and wireless data network measurements, this VNA/Spectrum Analyzer marriage is the ideal instrument to making fast and reliable measurements in the field.



Frequency VNA Master Frequency Range: MS2026/36C: 5 kHz to 6 GHz MS2028/38C: 5 kHz to 20 GHz Frequency Accuracy: 1.5 ppm Frequency Resolution: 1 Hz to 375 MHz, 10 Hz to 6 GHz, and 100 Hz to 20 GHz
Typical Test Port Power VNA Master supports selection of either High (default) or Low test port power. Changing power after calibration can degrade the calibrated performance. Typical power by bands is shown in the following table.
Block Diagram As shown in the following block diagram, the VNA Master has a 2-port, 2-path architecture that automatically measures four S-parameters with a single connection.
S 21
DUT
S 22
S 12
Transmission Dynamic Range The transmission dynamic range (the difference between test port power and noise floor) using 10 Hz IF Bandwidth and High Port Power is shown in the following table.
Receiver
Port 1
Port 2
Reference
Receiver
The above illustration is a simplified block diagram of VNA Master’s 2-port, 2-path architecture.
Typical Sweep Speed The typical sweep speed for IF Bandwidth of 100 kHz, 1001 data points, and single display is shown in the following table. The three receiver architecture will simultaneously collect S21 and S11 (or S12 and S22 ) in a single sweep.
Bridge/ Coupler · Bridge/ Coupler · Switch

Round-trip cable loss measurements are convenient for field personnel testing installed cable or waveguide runs. This one-port technique provides one-way data after twice traversing the cable. The following two sets of uncertainty curves, less than 6 GHz on the left and greater than 6 GHz on the right, present worst-case uncertainty by DUT Match (i.e., Log Mag) when using VNA Master for one-port cable loss measurements. As a practical tip, consider using a two-port transmission measurement technique to improve upon these one-port cable loss uncertainties.
VNA Master (MS20xxC), 5 kHz - 6 GHz One Port Cable Loss (Log Mag / 2) Uncertainty vs. DUT Match
VNA Master (MS20xxC), 6 - 20 GHz One Port Cable Loss (Log Mag / 2) Uncertainty vs. DUT Match
0 1 2 3 4 5 6 7 8 9 10
0 1 2 3 4 5 6 7 8 9 10
Cable Loss, Log Mag / 2 (dB)
Cable Loss, Log Mag / 2 (dB)
These uncertainty curves show how frequency range, DUT Match, and cable loss impact worst-case uncertainty of round-trip cable loss measurements. The uncertainty curves, separated by frequency range, are shown for DUT Match conditions of 20, 25, and 30 dB. For DUT Match of 30 dB and cable loss of 4-5 dB (reflection measurement of 8-10 dB) the worst-case uncertainties are approximately ±1 dB.


OSLxx50 Calibration Components (N-Connectors) Corrected System Performance and Uncertainties: MS202x/3xC Model with 12-term SOLT calibration including isolation using either OSLN50 & OSLNF50 or OSLK50 & OSLKF50 Calibration Kits
Precision calibration standards come in a convenient configuration for field work.
Frequency Range (GHz) Directivity (dB)
≤ 5 > 42
≤ 15 > 36
≤ 20* > 32
* N Connector guaranteed to 18 GHz, typical > 18 GHz
Measurement Uncertainties The following graphs provide measurement uncertainty at 23 ºC ± 5 ºC for the above indicated connector type and calibration. Errors are worse-case contributions of residual directivity, source match, frequency response, network analyzer dynamic range, and connector repeatability. For two-port measurements, transmission tracking, crosstalk, and physical load match termination were added. Isolation calibration and an IF Bandwidth of 10 Hz is used.


-80 -70 -60 -50 -40 -30 -20 -10 0 0.01
-40 -35 -30 -25 -20 -15 -10 -5 0


-40 -35 -30 -25 -20 -15 -10 -5 0
-80 -70 -60 -50 -40 -30 -20 -10 0

OSLxx50 Calibration Components Corrected System Performance and Uncertainties: MS202x/3xC Model with 12-term SOLT calibration including isolation using either OSLN50 & OSLNF50 or OSLK50 & OSLKF50 Calibration Kits


* N Connector guaranteed to 18 GHz, typical > 18 GHz
Measurement Uncertainties The following graphs provide measurement uncertainty at 23 ºC ± 5 ºC for the above indicated connector type and calibration. Errors are worse-case contributions of residual directivity, source match, frequency response, network analyzer dynamic range, and connector repeatability. For two-port measurements, transmission tracking, crosstalk, and physical load match termination were added. Isolation calibration and an IF Bandwidth of 10 Hz is used.


-80 -70 -60 -50 -40 -30 -20 -10 0 0.01
-40 -35 -30 -25 -20 -15 -10 -5 0


-40 -35 -30 -25 -20 -15 -10 -5 0
-80 -70 -60 -50 -40 -30 -20 -10 0


3652A Calibration Kit (K-Connector) Corrected System Performance and Uncertainties: MS202x/3xC Model with 12-term SOLT calibration including isolation using 3652A Calibration Kit
* N Connector guaranteed to 18 GHz, typical > 18 GHz
Measurement Uncertainties The following graphs provide measurement uncertainty at 23 ºC ± 5 ºC for the above indicated connector type and calibration. Errors are worse-case contributions of residual directivity, source match, frequency response, network analyzer dynamic range, and connector repeatability. For two-port measurements, transmission tracking, crosstalk, and physical load match termination were added. Isolation calibration and an IF Bandwidth of 10 Hz is used.


-40 -35 -30 -25 -20 -15 -10 -5 0
-80 -70 -60 -50 -40 -30 -20 -10 0


-40 -35 -30 -25 -20 -15 -10 -5 0
-80 -70 -60 -50 -40 -30 -20 -10 0


3652A Calibration Kit (K-Connector) Corrected System Performance and Uncertainties: MS202x/3xC Model with 12-term SOLT calibration including isolation using 3652A Calibration Kit
* N Connector guaranteed to 18 GHz, typical > 18 GHz
Measurement Uncertainties The following graphs provide measurement uncertainty at 23 ºC ± 5 ºC for the above indicated connector type and calibration. Errors are worse-case contributions of residual directivity, source match, frequency response, network analyzer dynamic range, and connector repeatability. For two-port measurements, transmission tracking, crosstalk, and physical load match termination were added. Isolation calibration and an IF Bandwidth of 10 Hz is used.


-80 -70 -60 -50 -40 -30 -20 -10 0 0.01
-40 -35 -30 -25 -20 -15 -10 -5 0


-40 -35 -30 -25 -20 -15 -10 -5 0
-80 -70 -60 -50 -40 -30 -20 -10 0

–100 dBc/Hz @ 10 kHz offset from carrier (–104 dBc/Hz typical) –102 dBc/Hz @ 100 kHz offset from carrier (–107 dBc/Hz typical) –107 dBc/Hz @ 1 MHz offset from carrier (–114 dBc/Hz typical) –120 dBc/Hz @ 10 MHz offset from carrier (–129 dBc/Hz typical)
SSB Phase Noise at 1 GHz
Amplitude Accuracy (single sine wave input < Ref level, and > DANL, auto attenuation)
Displayed Average Noise Level (DANL) (RMS detection, VBW/Avg type = Log., Ref Level = –20 dBm for preamp Off and –50 dBm for preamp On)

Preamp Off (RF input terminated, 0 dB input attenuation) –90 dBm 9 kHz to 13 GHz –85 dBm 13 GHz to 20 GHz
Preamp On (RF input terminated, 0 dB input attenuation) –100 dBm 1 MHz to 20 GHz
Input-Related Spurious (0 dB attenuation, –30 dBm input, span < 1.7 GHz) –60 dBc, –70 dBc typical
Third-Order Intercept (TOI) (–20 dBm tones 100 kHz apart, –20 dBm Ref level, 0 dB input attenuation, preamp Off)
2.4 GHz +15 dBm
50 MHz to 20 GHz +20 dBm typical
< 4 GHz +5 dBm typical
4 GHz to 20 GHz +12 dBm typical
50 MHz –54 dBc
< 4 GHz –60 dBc typical
> 4 GHz –75 dBc typical
> 10 dB input attenuation < 20 GHz 1:5:1 typical


Field Strength (uses antenna calibration tables to measure dBm/m2 or dBmV/m) Occupied Bandwidth (measures 99% to 1% power channel of a signal) Channel Power (measures the total power in a specified bandwidth) ACPR (adjacent channel power ratio) C/I (carrier-to-interference ratio) Emission Mask (recall limit lines as emission mask)

Time Domain (Option 0002) (includes Distance Domain Option 0501)
The VNA Master can also display the S-parameter measurements in the time or distance domain using lowpass or bandpass processing analysis modes. The broadband frequency coverage coupled with 4001 data points means you can measure discontinuities both near and far with unprecedented clarity for a handheld tool. With this option, you can simultaneously view S-parameters in frequency, time, and distance domain to quickly identify faults in the field. Advanced features available with this option include step response, phasor impulse, gating, and frequency gated in time. The option includes computational routines that further enhance the Distance Domain results by compensating for cable loss, relative velocity of propagation, and dispersion compensation in waveguide.
Round-Trip (reflection) Fault Resolution (meters): (0.5 x c x Vp) / ∆F; (c is speed of light = 3E8 m/s, ∆F is F2 – F1 in Hz)
One-Way (transmission) Fault Resolution (meters): (c x Vp) / ∆F; (c is speed of light = 3E8 m/s, ∆F is F2 – F1 in Hz)
Horizontal Range (meters): 0 to (data points – 1) x Fault Resolution to a maximum of 3000 m (9843 ft.)
Windowing Rectangular, Nominal Side Lobe (NSL), Low Side Lobe (LSL), and Minimum Side Lobe (MSL)
Transmitter measurements in the field are possible when using this VNA Master software mode with a separately purchased Anritsu 560 series detector. A variety of detectors are available to 50 GHz, but the popular 560-7N50B covers 10 MHz to 20 GHz with a measurement range of –50 to +20 dBm with better than 0.5 dB flatness to 18 GHz. After zeroing the detector to ensure accuracy at low power levels, the software offers intuitive operation for absolute and relative readouts in dBm or Watts.
Part Numbers 560-7N50B 560-7S50B
Frequency Range 0.01 to 20 GHz 0.01 to 20 GHz
Impedance 50 Ω 50 Ω
Power Range –55 dBm to +16 dBm –55 dBm to +16 dBm
15 dB, < 0.04 GHz 22 dB, < 8 GHz 17 dB, < 18 GHz 14 dB, < 20 GHz
15 dB, < 0.04 GHz 22 dB, < 8 GHz 17 dB, < 18 GHz 14 dB, < 20 GHz
Display Range –80 to +80 dBm (10 pW to 100 kW)
Measurement Range –50 to +20 dBm (10 nW to 40 mW)
Input Connector N(m) WSMA(m)
Offset Range 0 to +60 dB
Frequency Response ±0.5 dB, < 18 GHz ±1.25 dB, < 20 GHz
±0.5 dB, < 18 GHz ±1.25 dB, < 20 GHz
Resolution 0.1 dB, 0.1 xW (x = n, µ, m based on detector power)
*See www.us.anritsu.com for additional detectors
Accuracy ±1 dB maximum for >–40 dBm using 560-7N50B detector
For highly secure data handling requirements, this software option prevents the storing of measurement setup or data information onto any internal file storage location. Instead, setup and measurement information is stored ONLY to the external USB memory location. A simple factory preset prepares the VNA Master for transportation while the USB memory remains behind in the secure environment. The VNA Master cannot be switched between secure and non-secure operation by the user once configured for secure data operation. As an additional security measure, with this option enabled the user can choose to blank the frequency values displayed on the screen.
Distance Domain · Return Loss

Vector Voltmeter (Option 0015)
For tower mounted amplifier tests, the MS20x/3xC series with optional internal bias tees can supply both DC and RF signals on the center conductor of the cable during measurements. For frequency sweeps in excess of 2 MHz, the VNA Master can supply internal voltage control from +12 to +32 V in 0.1 V steps up to 450 mA. To extend battery life, an external power supply can substitute for the internal supply by using the external bias inputs instead. Both test ports can be configured to supply voltage via this integrated bias tees option. Bias can be directed to VNA Port 1 or Port 2.
A phased array system relies on phase matched cables for nominal performance. For this class of application, the VNA Master offers this special software mode to simplify phase matching cables at a single frequency. The similarity between the popular vector voltmeter and this software mode ensures minimal training is required to phase match cables. Operation is as simple as configuring the display for absolute or relative measurements. The easy-to-read large fonts show either reflection or transmission measurements using impedance, magnitude, or VSWR readouts. For instrument landing system (ILS) or VHF Omni-directional Range (VOR) applications, a table view improves operator efficiency when phase matching up to twelve cables. The MS202x/3xC solution is superior because the signal source is included internally, precluding the need for an external signal generator.
Frequency Range 2 MHz to 6 GHz (MS20x6C) 2 MHz to 20 GHz (MS20x8C)
Internal Voltage/Current +12V to +32V at 450 ma. Steady rate
Internal Resolution 0.1V
External Voltage/Current ±50 V at 500 mA steady rate
CW Frequency Range 5 kHz to 20 GHz
Bias Tee Selections Internal, External, Off
Measurement Display CW, Table (Twelve Entries, Plus Reference)
Measurement Types Return Loss, Insertion
Port 1 Port 2
External Bias Input (±50 V, 500 mA Max)



Internal Bias +12 to +32 V 450 mA Max
The VNA Master offers optional integrated bias tee for supplying DC plus RF to the DUT as shown in this simplified block diagram. Connectivity is also provided for external supply (instead of internal) to preserve battery consumption.

High Accuracy Power Meter (Option 0019) Requires external USB power sensor.
Conduct precise measurements of CW and digitally modulated transmitters in the field using this VNA Master software mode with a separately purchased Anritsu USB power sensor. After specifying the center frequency and zeroing the sensor to ensure accuracy at low power levels, the software offers intuitive operation for absolute and relative readouts in dBm or Watts. Option 0019 supports the USB Power Sensors in the following table.

Frequency Range 50 MHz to 6 GHz 600 MHz to 4 GHz 50 MHz to 6 GHz 10 MHz to 8 GHz 10 MHz to 18 GHz
Connector Type N, male, 50 Ω Type N, female, Ω Type N, male, 50 Ω Type N, male, 50 Ω Type N, male, 50 Ω
Dynamic Range –30 dBm to +20 dBm (0.001 mW to 100 mW)
+3 dBm to +51.76 dBm (2 mW to 150 W)
–40 dBm to +23 dBm (0.1 µW to 200 mW)
–40 dBm to +20 dBm (0.1 µW to 100 mW)
–40 dBm to +20 dBm (0.1 µW to 100 mW)
VBW 100 Hz 100 Hz 100 Hz 50 kHz 50 kHz
Measurement True-RMS. Slot Power, Burst Average Power
Measurement Uncertainty ±0.16 dB1 ±0.17 dB2 ±0.16 dB1 ±0.18 dB3 ±0.18 dB3
Datasheet for Additional Specifications 11410-00414 11410-00483 11410-00424 11410-00504 11410-00504
Notes: 1) Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than –20 dBm with zero mismatch errors 2) Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load. Measurement results referenced to the input side of the sensor. 3) Expanded uncertainty with K=2 for power measurements of a CW signal greater than –20 dBm with zero mismatch errors
Spectrum Field Strength Occupied Bandwidth Channel Power Adjacent Channel Power (ACPR) AM/FM/SSB Demodulation (Wide/Narrow FM, Upper/Lower SSB), (audio out only) Carrier-to-Interference ratio (C/I) Spectrogram (Collect data up to one week) Signal Strength (Gives visual and aural indication of signal strength) Received Signal Strength Indicator (RSSI) (collect data up to one week) Gives visual and aural indication of signal strength Signal ID (up to 12 signals) Center Frequency Bandwidth Signal Type (FM, GSM, W-CDMA, CDMA, Wi-Fi) Closest Channel Number Number of Carriers Signal-to-Nose Ratio (SNR) > 10 dB
Application Options Bias-Tee (On/Off), Impedance (50 Ω, 75 Ω, Other)

Number of Channels 1 to 20 Channels (Power Levels)
Measurements Graph/Table, Max Hold (On/5 sec/Off), Frequency/Channel, Current/Maximum, Dual Color
Scanner Scan Channels, Scan Frequencies, Scan Customer List, Scan Script Master™
Custom Scan Signal Standard, Channel, # of Channels, Channel Step Size, Custom Scan
Frequency Accuracy ± 10 Hz + Time base error
Application Options Bias-Tee (On/Off), Impedance (50 Ω, 75 Ω, Other) )
Measurements

Built-in GPS provides location information (latitude, longitude, altitude) and Universal Time (UT) information for storage along with trace data so you can later verify that measurements were taken at the right location. The GPS option requires a separately ordered magnet mount GPS antenna (2000-1528-R), which is configured with a 15 foot (~5 m) cable to mount outside on a metallic surface. Frequency accuracy is enhanced for the Spectrum Analyzer when Options 0025 Interference Analyzer and 0027 Channel Scanner are engaged.
Setup On/Off, Antenna Voltage 3.3/5.0 V, GPS Info
GPS Time/Location Indicator Time, Latitude, Longitude and Altitude on display Time, Latitude, Longitude and Altitude with trace storage
High Frequency Accuracy Spectrum Analyzer, Interference Analyzer, CW Signal Generator when GPS Antenna is connected < ± 50 ppb with GPS On, 3 minutes after satellite lock in selected mode
GPS Lock – after antenna is disconnected < ± 50 ppb for 3 days, 0 °C to 50 °C ambient temperature
As an alternative to a sampling oscilloscope, verifying the performance and identifying discontinuities in high-data-rate differential cables is now possible with the VNA Master. After a full two-port calibration, connect your differential cable directly to the two test ports and reveal the Sd1d1 performance, which is essentially differential return loss, or any of the other differential S-Parameters, Sc1c1, Sd1c1, or Sc1d1. With optional time domain, you can convert frequency sweeps to distance. This capability is especially valuable for applications in high data rate cables where balanced data formats are used to isolate noise and interference.
Distance Domain Analysis is a powerful field test tool to analyze cables for faults, including minor discontinuities that may occur due to a loose connection, corrosion, or other aging effects. By using Frequency Domain Reflectometry (FDR), the VNA Master exploits a user-specified band of full power operational frequencies (instead of DC pulses from TDR approaches) to more precisely identify cable discontinuities. The VNA Master converts S-parameters from frequency domain into distance domain on the horizontal display axis, using a mathematical computation called Inverse Fourier Transform. Connect a reflection at the opposite end of the cable and the discontinuities appear versus distance to reveal any potential maintenance issues. When access to both ends of the cable is convenient, a similar distance domain analysis is available on transmission measurements.
Option 0501 Distance Domain will improve your productivity with displays of the cable in terms of discontinuities versus distance. This readout can then be compared against previous measurements (from stored data) to determine whether any degradations have occurred since installation (or the last maintenance activity). More importantly, you will know precisely where to go to fix the problem and so minimize or prevent downtime of the system.
Connector SMA, female

Size and Weight

Status (Temperature, Battery Info, S/N, Firmware Ver, IP Address, Options Installed) Self Test, Application Self Test GPS (see Option 0031)

Depth 78 mm (3.1 in) (MS202xC) 97 mm (3.8 in) (MS203xC)
Name, Date and Time, Ethernet Configuration, Brightness, Volume Language (English, French, German, Spanish, Chinese, Japanese, Korean, Italian, User defined) Reset (Factory Defaults, Master Reset, Update Firmware)
Weight, Including Battery 4.5 kg (9.9 lbs) (MS202xC) 4.8 kg (10.5 lbs) (MS203xC)
Safety Class EN 61010-1 Class 1
Delete Selected File, All Measurements, All Mode Files, All Content
Product Safety IEC 60950-1 when used with Anritsu supplied Power Supply

External Trace/Setup Memory Limited by size of USB Flash drive
Temperature, operating (ºC) (3.8.2.1 & 4.5.5.14) Passed, –10 ºC to 55 ºC, Humidity 85%
Mode Switching Auto-Stores/Recalls most recently used Setup Parameters in the Mode
Temperature, not operating (ºC) (3.8.2.2 & 4.5.5.1) Passed, –51 ºC to 71 ºC

Altitude, not operating (3.8.3 & 4.5.5.2) Passed*, 4600 m
Type N female (or K female with opt 0011, MS20x8C only) VNA port (x2)
Type BNC female Bias Tee port (enabled with opt 0010) (x2)
Spectrum Analyzer Connectors Type N, female (or K female with opt 0011) (MS203xC)
Salt exposure structural parts (3.8.8.2 & 4.5.6.2.2) Not Required***
External Power 5.5 mm barrel connector, 12 to 15 VDC, < 5.0 Amps
LAN Connection RJ48C, 10/100 Mbps, Connect to PC or LAN for Remote Access (Available with opt 0411 Ethernet)
* Qualified by similarity (tested on a similar product) ** Not defined in standard; must be invoked and defined by purchase description *** Not required for Class 2 equipment
USB Interface (2) Type A, Connect Flash Drive and Power Sensor
USB Interface 5-pin mini-B, Connect to PC for data transfer

European Union CE Mark, EMC Directive 89/336/EEC, 92/31/EEC, 93/68/EEC and Low Voltage Directive 73/23/EEC, 93/68/EEC
External Trigger BNC, female, 50 Ω, Maximum Input ± 5 VDC
Size 8.4 in, daylight viewable color LCD
Resolution 800 x 600

DC power from Universal 110/220V AC/DC Adapter 55 Watts running off AC/DC adaptor while charging battery

The ODTF-1 module is primarily intended for field use by technicians and engineers responsible for the deployment and maintenance of remote radio heads (RRH), and nicely complements the field diagnostic power of the VNA Master. The ODTF-1 module is fully compatible with the MS202x/3xC VNA Masters which are equipped with the Time Domain Option 0002 or Distance Domain Option 0501. Field operation of the ODTF-1 module with the VNA Master requires the normal DTF (RF/ microwave) mode along with simple modification of some of the setup parameters such as Vp, cable loss, and frequency.

VNA Master users need only to connect a short cable between the RF output of the VNA and the ODTF-1 module and perform a 1-port calibration at the end of the cable. Essentially the ODTF-1 module is simply a wavelength translator, RF test signals in, RF signals returning. The same physics that apply to the traditional DTF measurements apply to ODTF-1 meaning highly accurate measurements can be made with event resolution as good as 10 cm. The same trade-offs carry over as well so better event resolution translates to shorter maximum distance, and vice-versa. Max distance is specified at 1020 meters (3345 ft).
The battery life of the ODTF-1 module matches the battery life of the VNA Master. It can be charged with the same 40-168-R power supply so there is no need to maintain different power supplies.
Using a VNA Master equipped with Option 0002 or 0501, this ODTF-1 optical module translates the optical signals to the RF domain of the VNA, to display fault locations in standard optical fibers.
General Specifications
Wavelength 1550 nm typical
External DC Input +12.5 to +15 VDC, 3A maximum
Frequency Range 1 GHz to GHz
Electromagnetic Compatibility Meets European Community requirements for CE marking
Fiber Type Single Mode Fiber (SMF)
Temperature Operating 0 to 50 ºC
Event Resolution 10.2 cm (0.335 ft) maximum, or 150/(n*∆F), ∆F in MHz, n is IOR
Horizontal Range 1020 meter (3345 ft) maximum, or (#dp-1)*Event Resolution
Non-operating 0 to 70 ºC recommended
Optical Dynamic Range 30 dB
Size 15.7*5.37*18.6 cm (6.18*2.11*7.3 in.)
Optical Output Power 3 dBm typical
Weight < 1 kg (2.2 lbs)
Max RF Input Power + 5 dBm
RF Connector N(m) · Optical Connector FC/APC

Each VNA Master ships with a versatile test assistant: a copy of Anritsu’s Master Software Tools for Windows® 2000/XP/Vista/7. This allows an operator to add the processing capabilities of a PC and this software utility to the VNA Master to form a powerful and flexible measurement solution for network analysis. For automation, the VNA Master also supports remote programming via the Ethernet or USB interface.

Connect VNA Master to a PC for archiving and additional post-processing. A standard tilt-bail provides convenient use on a bench.
Powerful data management tool for storing and sifting through measurement results MST simplifies transfers, printing, and archival of displays and setups
Connect to a PC using USB2.0 (full-speed), Ethernet LAN, or Direct Ethernet Unleash powerful MST capabilities by using variety of popular remote interfaces
Store an unlimited number of setups, traces, and JPEGs (limited only by PC memory) Develop libraries of frequently used setups and typical results
Manipulate traces and further optimize displays Versatility to further analyze results without re-taking measurements
Update with the latest firmware Easily download and upgrade to newest features from www.us.anritsu.com
Remote programming via Ethernet or USB Increase throughput by automating repetitive or operator intensive tasks
Feature Benefit

MS2028C 1
MS2036C 1
MS2038C 1
VNA Master, 2-port, VNA 5 kHz to 6 GHz
VNA Master, 2-port, VNA 5 kHz to 20 GHz
VNA Master + Spectrum Analyzer, S/A 9 kHz to 9 GHz
VNA Master + Spectrum Analyzer, S/A 9 kHz to 20 GHz
Options Description
MS2026C-0002 MS2028C-0002 MS2036C-0002 MS2038C-0002 Time Domain (includes DTF capability)
MS2026C-0005 MS2028C-0005 – – Power Monitor (requires external detector)
MS2026C-0007 MS2028C-0007 MS2036C-0007 MS2038C-0007 Secure Data Operation
MS2026C-0010 MS2028C-0010 MS2036C-0010 MS2038C-0010 Built-in Bias-Tee
– MS2028C-0011 – MS2028C-0011 K(f) Test Port Connectors
MS2026C-0015 MS2028C-0015 MS2046C-0015 MS2038C-0015 Vector Voltmeter
MS2026C-0019 MS2028C-0019 MS2036C-0019 MS2038C-0019 High Accuracy Power Meter (requires external USB sensor)
– – MS2036C-0025 MS2038C-0025 Interference Analysis, 9 kHz to 9/20 GHz 2
– – MS2036C-0027 MS2038C-0027 Channel Scanner, 9 kHz to 9/20 GHz 2
MS2026C-0031 MS2028C-0031 MS2036C-0031 MS2038C-0031 GPS Receiver (requires GPS antenna, 2000-1528-R) MS2026C-0077 MS2028C-0077 MS2036C-0077 MS2038C-0077 Balanced/Differential S-Parameters, 1-port
MS2026C-0099 MS2028C-0099 MS2036C-0099 MS2038C-0099 Premium Calibration
MS2026C-0501 MS2028C-0501 MS2036C-0501 MS2038C-0501 Distance Domain (included in Option 0002)
MS202x/3xC Standard Accessories 10580-00220 VNA Master User’s Guide 65729 Soft Carrying Case 2300-498 Master Software Tools CD ROM 633-44 Rechargeable Battery, Li-Ion, 6.6 Ah 40-168-R AC-DC Adapter 806-141-R Automotive Cigarette Lighter 12 V DC adapter 3-2000-1498 USB A-type to Mini USB B-type cable, 3.05 m (10 ft.) 2000-1371-R Ethernet cable, 2.13 m (7 ft.) 3-806-152 Ethernet Crossover Cable, 2.13 m (7 ft.) 2000-1520-R USB Flash Drive
Ancillary Equipment ODTF-1 Optical Time Domain Module 15NNF50-1.5C Armored Test Port Cable, 1.5 meter, N(m) to N(f) OSLN50-1 Precision Open/Short/Load, DC to 6 GHz, 42 dB Return Loss

Optional Accessories
Directional Antennas 2000-1411-R 824 MHz to 896 MHz, N(f), 10 dBd, Yagi 2000-1412-R 885 MHz to 975 MHz, N(f), 10 dBd, Yagi 2000-1413-R 1710 MHz to 1880 MHz, N(f), 10 dBd. Yagi 2000-1414-R 1850 MHz to 1990 MHz, N(f), 9.3 dBd, Yagi 2000-1415-R 2400 MHz to 2500 MHz, N(f), 10 dBd, Yagi 2000-1416-R 1920 MHz to 2170 MHz, N(f), 10 dBd, Yagi 2000-1519-R 500 MHz to 3000 MHz, log periodic 2000-1617 600 MHz to 21000 MHz, N(f), 5-8 dBi to 12 GHz, 0-6 dBi to 21 GHz, log periodic
High Accuracy Power Sensor PSN50 High Accuracy Power Sensor, 50 MHz to 6 GHz MA24104A Inline High Power Sensor, 600 MHz to 4 GHz, True RMS MA24106A High Accuracy Power Sensor, 50 MHz to 6 GHz, True RMS MA24108A High Accuracy Power Sensor, 10 MHz to 8 GHz, True RMS MA24118A High Accuracy Power Sensor, 10 MHz to 18 GHz, True RMS MA24126A High Accuracy Power Sensor, 10 MHz to 26 GHz, True RMS
Portable Antennas 2000-1200 806 MHz to 866 MHz, SMA(m), 50 Ω 2000-1473 870 MHz to 960 MHz, SMA(m), 50 Ω 2000-1035 896 MHz to 941 MHz, SMA (m), 50 Ω (1/4 wave) 2000-1030 1710 MHz to 1880 MHz, SMA(m), 50 Ω (1/2 wave) 2000-1474 1710 MHz to 1880 MHz with knuckle elbow (1/2 wave) 2000-1031 1850 MHz to 1990 MHz, SMA(m), 50 Ω (1/2 wave) 2000-1475 1920 MHz to 1980 MHz and 2110-2170 MHz, SMA(m), 50 Ω 2000-1032 2400 MHz to 2500 MHz, SMA(m), 50 Ω (1/2 wave) 2000-1361 2400 MHz to 2500, 5000 to 6000 MHz, SMA(m), 50 Ω 2000-1616 20 MHz to 21000 MHz, N(f), 50 Ω 61532 Antenna Kit (Consists of: 2000-1030, 2000-1031, 2000-1032-R, 2000-1200, 2000-1035, 2000-1361, and carrying pouch)
Power Monitor Detectors 560-7N50B RF Detector, 0.01 to 20 GHz, Type-N(m) 560-7S50B RF Detector, 0.01 to 20 GHz, W-SMA(m)
Detector Extender Cables 800-109 Detector Extender Cable, 7.6m (25 ft) 800-111 Detector Extender Cable, 30.5m (100 ft.)
K Connector Components OSLK50 Precision integrated Open/Short/Load K(m), DC to 20 GHz, 50 Ω OSLKF50 Precision integrated Open/Short/Load K(f), DC to 20 GHz, 50 Ω 22K50 Precision K(m) Short/Open, 40 GHz 22KF50 Precision K(f) Short/Open, 40 GHz 28K50 Precision Termination, DC to 40 GHz, 50 Ω, K(m) 28KF50 Precision Termination, DC to 40 GHz, 50 Ω, K(f) 3652A K Calibration Kit, DC to 40 GHz
Bandpass Filters 1030-114-R 806 MHz to 869 MHz, N(m) to SMA(f), 50 Ω 1030-109-R 824 MHz to 849 MHz, N(m) to SMA (f), 50 Ω 1030-110-R 880 MHz to 915 MHz, N(m) to SMA (f), 50 Ω 1030-105-R 890 MHz to 915 MHz Band, 0.41 dB loss, N(m) to SMA(f), 50 Ω 1030-111-R 1850 MHz to 1910 MHz, N(m) to SMA (f), 50 Ω 1030-106-R 1710 MHz to 1790 MHz Band, 0.34 dB loss, N(m) to SMA(f), 50 Ω 1030-107-R 1910 MHz to 1990 MHz Band, 0.41 dB loss, N(m) to SMA(f), 50 Ω 1030-112-R 2400 MHz to 2484 MHz, N(m) to SMA (f), 50 Ω 1030-155-R 2500 MHz to 2700 MHz, N(m) to N(f), 50 Ω
N-Type Connectors OSLN50 Precision Integrated Open/Short/Load N(m), DC to 18 GHz, 50 Ω OSLNF50 Precision Integrated Open/Short/Load N(f), DCto 18 GHz, 50 Ω 22N50 Precision N(m) Short/Open, 18 GHz 22NF50 Precision N(f) Short/Open, 18 GHz 28N50-2 Precision Termination, DC to 18 GHz, 50 Ω, N(m) 28NF50-2 Precision Termination, DC to 18 GHz, 50 Ω, N(f) OSLN50-1 Precision N(m) Open/Short/Load, 42 dB, 6 GHz OSLNF50-1 Precision N(f) Open/Short/Load, 42 dB, 6 GHz SM/PL-1 Precision N(m) Load, 42 dB, 6 GHz SM/PLNF-1 Precision N(f) Load, 42 dB, 6 GHz
Attenuators 3-1010-122 20 dB, 5 W, DC to 12.4 GHz, N(m) to N(f) 42N50-20 20 dB, 5 W, DC to 18 GHz, N(m) to N(f) 42N50A-30 30 dB, 5 W, DC to 18 GHz, N(m) to N(f) 3-1010-123 30 dB, 50 W, DC to 8.5 GHz, N(m) to N(f) 1010-127-R 30 dB, 150 W, DC to 3 GHz, N(m) to N(f) 3-1010-124 40 dB, 100 W, DC to 8.5 GHz, N(m) to N(f), Uni-directional 1010-121 40 dB, 100 W, DC to 18 GHz, N(m) to N(f), Uni-directional 1010-128-R 40 dB, 150 W, DC to 3 GHz, N(m) to N(f)
TNC Connector Components 1091-53-R Precision TNC(m) Open, 18 GHz, 50 Ω 1091-54-R Precision TNC(m) Short, 18 GHz, 50 Ω 1015-55-R Precision TNC(m) Load, 18 GHz, 50 Ω 1091-55-R Precision TNC(f) Open, 18 GHz, 50 Ω 1091-56-R Precision TNC(f) Short, 18 GHz, 50 Ω 1015-54-R Precision TNC(f) Load, 18 GHz, 50 Ω
7/16 Connector Components 2000-1618-R Precision Open/Short/Load, 7/16(m), 6.0 GHz 2000-1619-R Precision Open/Short/Load, 7/16(f), 6.0 GHz

Manuals 10580-00305 VNA Master User’s Guide 10580-00306 VNA Master Programming Manual 10580-00289 VNA Measurement Guide 10580-00244 Spectrum Analyzer Measurement Guide 10580-00240 Power Meter Measurement Guide 10580-00215 ODTF-1 Optical Distance-to-Fault Module – Quick Start Guide
Recommended waveguide calibration procedure requires two offset shorts and a precise load. The waveguide/coax adapter, shown attached to test port #2, adapts the VNA Master test ports to the waveguide under test.

Related Literature, Application Notes 11410-00206 Time Domain for Vector Network Analyzers 11410-00214 Reflectometer Measurements – Revisited 11410-00270 What is Your Measurement Accuracy? 11410-00373 Distance-to-Fault 11410-00387 Primer on Vector Network Analysis 11410-00414 High Accuracy Power Meter, PSN50 11410-00424 USB Power Sensor MA24106A 11410-00483 Inline High Power Sensor MA24104A 11410-00504 Microwave USB Power Sensor MA241x8A 11410-00531 Practical Tips on Making “Vector Voltmeter (VVM)” Phase Measurements using VNA Master (Opt. 15) 11410-00472 Measuring Interference
Type Compatible Flanges 1/8 Offset Short
23UM70 24UM70 26UM70 35UM70N 5.85 to 8.20 GHz WR137, WG14 CAR70, PAR70, UAR 70, PDR70
23UM84 24UM84 26UM84 35UM84N 7.05 to 10.00 GHz WR112, WG15 CBR84, UBR84, PBR84, PDR84
23UM100 24UM100 26UM100 35UM100N 8.20 to 12.40 GHz WR90, WG16 CBR100, UBR100, PBR100, PDR100
23UM120 24UM120 26UM120 35UM120N 10.00 to 15.00 GHz WR75, WG17 CBR120, UBR120, PBR120, PDR120
23UA187 24UA187 26UA187 35UA187N 3.95 to 5.85 GHz WR187, WG12
CPR137F, CPR137G, UG-1356/U UG-1357/U, UG-1732/U, UG-1733/U, UG-343B/U, UG-344/U, UG-440B/U, UG-441/U
23UA137 24UA137 26UA137 35UA137N 5.85 to 8.20 GHz WR137, WG14
CPR112F, CPR112G, UG-1358/U, UG-1359/U, UG-1734/U, UG-1735/U, UG-52B/U, UG-51/U, UG-137B/U, UG-138/U
23UA112 24UA112 26UA112 35UA112N 7.05 to 10.00 GHz WR112, WG15
CPR90F, CPR90G, UG-1360/U, UG-1361/U, UG-1736/U, UG-1737/U, UG-40B/U, UG-39/U, UG-135/U, UG-136B/U
23UA90 24UA90 26UA90 35UA90N 8.20 to 12.40 GHz WR90, WG16
23UA62 24UA62 26UA62 35UA62N 12.40 to 18.00 GHz WR62, WG18 UG-541A/U, UG-419/U, UG-1665/U, UG1666/U
23UA42 24UA42 26UA42 35UA42K 17.00 to 26.50 GHz WR42, WG20 UG-596A/U, UG-595/U, UG-597/U UG-598A/U
[1] For Coaxial/Waveguide Adapter part numbers, N designates Type N and K designates K-Connector



12 Avenue du Québec, Bâtiment Iris 1-Silic 638, 91140 VILLEBON SUR YVETTE, France Phone: +33-1-60-92-15-50 Fax: +33-1-64-46-10-65 • Germany Anritsu GmbH Nemetschek Haus, Konrad-Zuse-Platz 1 81829 München, Germany Phone: +49 (0) 89 442308-0 Fax: +49 (0) 89 442308-55
11410-00548, Rev. B Printed in United States 2010-09 ©2010 Anritsu Company. All Rights Reserved ®Anritsu All trademarks are registered trademarks of
their respective companies. Data subject to change without notice. For the most recent specifications visit: www.anritsu.com
Pages referenced in this manual — manufacturer, support and downloads.
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.